This chapter contains a list of all the recommended equipment required to perform the
performance tests and adjustments for the ESG Family Signal Generators.
Calibration Guide1-1
Equipment RequiredESG Family Signal Generators
Required Equipment
• HP/Agilent 8563E Option E35
Spectrum Analyzer (No Substitution)
Required Equipment
• HP/Agilent 8902A Measuring
Receiver
1
Option 010
Capable of W-CDMA ACP
measurements from 1.2 to 3.0 GHz
Option E35 Comprises:
FM accuracy: ±2% of reading ±1 digit
AM accuracy: ±2% of reading ±1 digit
Range: 250 kHz to 1300 MHz
Filters: 300 Hz high pass and
15 kHz and 3 kHz low pass
Detectors: Peak+
• HP/Agilent 54750A Digitizing
Oscilloscope
Bandwidth: 20 GHz
• HP/Agilent 54751/2A Oscilloscope
Plug-in
Two 20 GHz bandwidth channels
• HP/Agilent 54610B Digital
Oscilloscope
1
Bandwidth: 500 MHz
• HP/Agilent 5071A
Primary Frequency Standard
• HP/Agilent 8563E Option H35
Spectrum Analyzer
• HP/Agilent 8563E Option K35
Adjacent Channel Power Ratio Test
Set for W-CDMA
NOTEHP/Agilent 8563E Option K37
combines the Option K35 test set
with firmware ROM chips to
upgrade existing HP/Agilent 8560
E-series analyzers.
• HP/Agilent 8563E Options 001, 007
Spectrum Analyzer
1
Frequency range: 100 kHz to 4000 MHz
Relative level accuracy (harmonic and
spurious measurements) ±2 dB
Displayed average noise level (1 Hz BW)
−110 dBm at 100 kHz,
−130 dBm at 1 to 10 MHz, and
−134 dBm at 0.01 to 4 GHz
Frequency: 10 MHz
Stability: > 1×10
-10
/year
• HP/Agilent 8903B Options 051, 010
Audio Analyzer
1
Distortion accuracy: ±1 dB
Residual distortion:
−80 dB at 80 kHz BW
30 kHz low pass filter
AC level accuracy: ±4%
• HP/Agilent 89441A Options AY7,
AYA, AY9, UFG or UTH, and AYH
Vector Signal Analyzer
1,2
1 Hz digital resolution BW filter
(required for power accuracy)
RF input VSWR: 1:1.5
External timebase input
Second IF output
• HP/Agilent 8663A Signal Generator
Frequency Range: 11.5 to 2500 MHz
Output level: +16 dBm
Meets HP/Agilent 8663A
single-sideband phase noise
specifications
(No Substitution)
1.Equipped with GPIB
2.Used only with ESG-D models
1-2Calibration Guide
1
ESG Family Signal GeneratorsEquipment Required
Required Equipment
• HP/Agilent 8904A Function
Generator
1
Frequency range (sinewave):
0 Hz to 600 kHz
AC Amplitude (sinewave only):
Range: 0 to 10 Vp-p into a 50Ω load
Flatness (>630 mVp-p into 50Ω):
±0.1% (±0.009 dB), 0.1 Hz to 100 kHz
• HP/Agilent 53132A Option 050
Universal Counter
1
Frequency range: 4 GHz
Resolution: 0.1 Hz
External timebase input
• HP/Agilent 33120A Option 001
Arbitrary Waveform Generator
1
(2)
(No Substitution)
• HP/Agilent 438A Power Meter
1
Instrumentation accuracy: ±0.5%
Power reference accuracy: 0.9%
• HP/Agilent 8482A Power Sensor
Frequency range: 100 kHz to 4000 MHz
Power range: −30 dBm to +13 dBm
Maximum SWR:
100 kHz to 300 kHz: 1:1.6,
300 kHz to 1 MHz: 1:1.12,
1 MHz to 2 GHz: 1:1.1,
2 GHz to 4 GHz: 1:1.3
Calibration factor accuracy (RSS):
≤ 1.6%
• HP/Agilent E3615A Power Supply
Minimum Power: 15V
Minimum Current: 325 mA
• MD/MDC-174 Mixer
Frequency range: 0.001 to 2.8 GHz
• HP/Agilent 3458A Digital
Multimeter
1
dc Volts: 0.1 to 100 V
Input resistance: 10Ω to 1 KΩ
• HP/Agilent 8491A/B Option 006
Attenuator
Attenuation: 6 dB
• HP/Agilent 8491A/B Option 010
Attenuator
Attenuation: 10 dB
• HP/Agilent 8491A/B Option 020
Attenuator
Attenuation: 20 dB
• Personal Computer (included in
HP/Agilent E5500A/B)
Running MS Windows® 3.1/95/98 or
Windows NT®
• HP LaserJet 5L Printer
MS Windows supported
• HP 82341C GPIB Interface Card
with SICL HP I/O Library Installed
(No Substitution)
• Agilent Technologies E4400-60073
Digital Test Card (No Substitution)
• HP/Agilent 8495G Programmable
Step Attenuator
Attenuation: 0 to 70 dB
Frequency range: dc to 4 GHz
• HP/Agilent 11713A Step Attenuator
Driver
• HP/Agilent E5500A/B Phase Noise
Measurement System
2, 3
Frequency range: 0 to 4 GHz
• MD/MDC-164 Mixer
Frequency range: 0.5 to 9 GHz
Calibration Guide1-3
1.Equipped with GPIB
2.Includes Personal Computer
3.Required for ESG-AP/DP only
Equipment RequiredESG Family Signal Generators
Required Equipment
Low Noise Amplifiers for
Power Level Accuracy Adjustment and Performance Test
The power level accuracy adjustment and performance test both use low noise amplifiers.
They are required to maintain separation between a low-level signal and the signal
analyzer noise floor. The following amplifiers meet the requirements of making this
measurement.
A +15 Vdc external supply is required for each of the amplifiers. An HP/Agilent E3615A,
0 to 20 V power supply can power all recommended amplifiers simultaneously.
100 kHz to 10 MHz Low Noise Amplifier
Critical Specifications
Frequency Range: 100 kHz to 10 MHz
Noise Figure: 1.5 dB
Gain: 50 dB
1 dB Compression Minimum: +5 dBm
External Power Supply Required: +15 Vdc @ 25 mA
Possible Part Number and Supplier
Part Number: W50ATC
Supplier: Advanced Control Components
20 Meridian Road
Eatontown, NJ 07724
30 MHz to 2 GHz Low Noise Amplifier
Configuration
Two amplifiers are cascaded to obtain the necessary requirements. The specifications
listed below are the combined specifications of the cascaded amplifiers.
Critical Specifications
Frequency Range: 10 MHz to 2 GHz
Noise Figure: 1.8 dB
Gain: > 56 dB
1 dB Compression Minimum: +10 dBm
External Power Supply Required: +15 Vdc @ 300 mA
Possible Part Number and Supplier
Part Number: AFS3-00010300-18-5P-4 (2 required)
Supplier: MITEQ Inc.
100 Davids Drive
Hauppauge, NY 11788
1-4Calibration Guide
ESG Family Signal Generators
2Operation Verification
This chapter provides procedures that will either ensure that the signal generator is
operating correctly or will help to point to problem areas if it is not. Operation verification
does not ensure performance to specifications, but should provide a level of confidence that
the signal generator is operating correctly within a minimum amount of time. Operation
verification is appropriate for incoming inspection, after repair when a full calibrated
performance is not required, or whenever the integrity of the signal generator is in
question.
Calibration Guide2-1
Operation VerificationESG Family Signal Generators
Verification Procedures
Verification Procedures
Performthe following procedures in the order they are presented. The tables referenced by
the tests are located in the back of the chapter where they can be copied easily.
This procedure verifies that the signal generator powers up and that the internal
instrument check identifies no errors. The internal check evaluates the correctness of
operation and returns an error message if a problem is detected.
1. Turn power on to the signal generator by pressing the power switch. The green LED
will light. Let the signal generator warm up for one hour.
NOTEFor ESG-AP or ESG-DP signal generators, or those with Option 1E5, ERROR
514, Reference Oven Cold occurs whenever the signal generator is first
connected to AC line power. The
annunciator both turn on. The OVEN COLD annunciator automatically clears
after approximately 5 minutes. The error queue cannot be cleared, however,
until the OVEN COLD annunciator has turned off.
OVEN COLD annunciator and the ERR
2. Cycle the power to the signal generator. The green LED should again be lit and the
signal generator will perform a check.
2-2Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
2. Check for Error Messages
1. When the display is lit, check to see if the ERR annunciator is turned on.
2. If the
Utility > Error Info > View Next Error Message. The first error message in the queue will be
ERR annunciator is turned on, review the error messages in the queue by pressing
shown in the text area of the display. Refer to the error messages guide for information
about the error message.
If there is more than one error message (each message will be designated as 1 of n),
continue pressing the
View Next Error Message softkey until you have seen all of the
messages.
3. When you have resolved all of the error messages, press Clear Error Queue(s) to delete the
messages. Cycle the power on the signal generator and then restart this procedure
(Check for Error Messages) at step 1.
3. Frequency Range and Accuracy Check
Frequency range is tested by determining the frequency accuracy relative to the timebase
at the frequency limits.
Connect the Test Equipment
Figure 2-1 Frequency Range and Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. For frequencies < 10 MHz:
a. Press
b. Press RF On/Off. The RF ON annunciator is displayed.
Calibration Guide2-3
Mod On/Off. The MOD OFF annunciator is displayed.
Amplitude. Enter 10 using the numeric keypad and press the dBm terminator
softkey.
Operation VerificationESG Family Signal Generators
Verification Procedures
4. For frequencies ≥ 10 MHz:
a. Press
Amplitude. Enter 0 using the numeric keypad and press the dBm terminator
softkey.
b. Check that the RF ON annunciator is displayed. If not, press RF On/Off.
Configure the Frequency Counter
1. For frequencies < 10 MHz use Input 2. Press 1MΩ.
2. For frequencies≥ 10 MHz and ≤ 500 MHz use Input 2. Press
3. For frequencies > 500 MHz use Input 1. Press
AUTO.
50Ω.
NOTESet the gate time of > 5 seconds for maximum counter accuracy. Verify that
the counter is phase-locked to the 10 MHz external reference.
Measure the Frequency Accuracy
1. Set the signal generator to the frequencies shown in Table 2-1, “Frequency Accuracy,”
on page 2-10 (to the maximum frequency of your signal generator).
2. Record the measured frequency in Table 2-1 and compare it to the corresponding limits.
4. Power Level Accuracy Check
Performing this check will provide a high level of confidence that the signal generator’s
power level circuitry is functioning correctly. This check does not test the signal generator
to warranted specifications. Test points have been reduced and the limits are degraded in
order to minimize measurement time and take into account a broad range of measurement
uncertainties. To have the signal generator tested to warranted specifications, a complete
power level accuracy performance test is required. If the complete performance test is
needed and you are unable to perform it, contact your nearest Agilent Technologies service
center for information concerning calibration.
Connect the Test Equipment for Setup 1
Figure 2-2 Power Level Accuracy Equipment Setup 1
2-4Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
3. Press Amplitude. Enter 13 using the numeric keypad and press the dBm terminator
softkey.
4. Press Mod On/Off. The MOD OFF annunciator is displayed.
5. Press
RF On/Off. The RF ON annunciator is displayed.
Configure the Power Meter
1. Zero and calibrate the power meter to the sensor’s reference calibration factor.
2. Set the power meter to measure absolute log power (dBm).
Measure the Power Level Accuracy
1. Set the power levels and frequencies for the signal generator (to the maximum
frequency of your signal generator). Refer to the appropriate table in the following list:
• Table 2-3, “Power Level Accuracy Setup 1 (Signal Generators with Option 1E6),” on
page 2-13.
2. Set the power sensor’s calibration factor for each frequency.
3. Record the measured power level in the appropriate table (Table 2-2 or Table 2-3) and
compare it to the corresponding limits.
Connect the Test Equipment for Setup 2
Figure 2-3 Power Level Accuracy Equipment Setup 2
Calibration Guide2-5
Operation VerificationESG Family Signal Generators
Verification Procedures
Configure the Signal Generator
1. Press Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
2. Press Amplitude. Enter −15 using the numeric keypad and press the dBm terminator
softkey.
Configure the Spectrum Analyzer
1. Preset the spectrum analyzer.
2. Set the analyzer to external 10 MHz reference.
3. Set the center frequency to 277 kHz.
4. Set the frequency span to 100 Hz. (This will result ina1Hzresolution bandwidth with
a digital filter.)
Measure the Power Level Accuracy
1. Connect the signal generator’s RF OUTPUT through the 6 dB attenuator to the
spectrum analyzer’s RF input.
2. Transfer the power level results for each frequency at −15 dBm from the Measured
Power column in Table 2-2 or Table 2-3 as appropriate, to the Power Meter Reading for
−15 dBm column in Table 2-4, “Power Level Accuracy Setup 2 (All Signal Generators),”
on page 2-15.
3. On the spectrum analyzer, select
Marker Normal Mode and then select the Peak Search
function. This activates the marker and sets it to the signal peak.
4. On the spectrum analyzer, ensure that the marker is at the signal peak and use the
MKR-> menu to set the marker to the reference level. If necessary, select Peak Search to
ensure that the marker is at the signal peak.
5. With the marker at signal peak, select the Marker Delta function. This will set the marker
to measure relative amplitude from a reference of 0 dB. If the marker does not read
0 dB, press
Marker Normal > Peak Search > Marker Delta until the marker reads 0 dB.
6. Decrease the signal generator amplitude in 10 dB steps as shown in Table 2-4. At each
10 dB step, select Peak Search to ensure that the marker is at the signal peak.
7. Measure the power levels shown in the Power Level Setting (dBm) column of Table 2-4
for the current frequency, and record the values in the Spectrum Analyzer Marker (dB)
column.
8. Return the spectrum analyzer reference level and the signal generator amplitude to
−15 dBm. Set the signal generator frequency and the spectrum analyzer’s center
frequency to the next frequency listed in Table 2-4 and repeat the process from step 3.
Continue until all of the frequencies have been measured and recorded (to the
maximum frequency of your signal generator).
2-6Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Calculate the Actual Power Level
1. Calculate and record the Actual Power Level, in Table 2-4, as the sum of the Power
Meter Reading for −15 dBm and the Spectrum Analyzer Marker (dB) value.
For example:
• Power Meter Reading for −15 dBm at 2.516 MHz = −14.95 dBm
• Spectrum Analyzer Marker (dB) at 2.516 MHz and Power Level Setting at
−85 dBm = −70.17 dB
• Actual Power Level at 2.516 MHz and −85 dBm: (−14.95) + (−70.17) = −85.12 dBm
2. Compare the calculated Actual Power Level value to the corresponding limits.
5. FM Accuracy Check
Connect the Test Equipment
Figure 2-4 FM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. Press
4. Press
5. Press
softkey.
6. Check that the MOD ON annunciator is displayed. If it is not, press Mod On/Off.
7. Press
Calibration Guide2-7
FM/ΦM > FM Off On. The FM annunciator is displayed.
FM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
FM Dev. Enter 100 using the numeric keypad and press the kHz terminator softkey.
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator
RF On/Off. The RF ON annunciator is displayed.
Operation VerificationESG Family Signal Generators
Verification Procedures
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies shown in Table 2-5, “FM Accuracy,” on
page 2-19.
2. Record the deviations measured and compare them to the limits shown in Table 2-5.
6. AM Accuracy Check
Connect the Test Equipment
Figure 2-5 AM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. Press
4. Press
softkey.
5. Check that the MOD ON annunciator is displayed. If it is not, press Mod On/Off.
AM > AM Off On. The AM annunciator is displayed.
AM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator
6. Press
2-8Calibration Guide
RF On/Off. The RF ON annunciator is displayed.
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to AM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies and depths shown in Table 2-6, “AM
Accuracy,” on page 2-19.
2. Record the AM depths measured and compare them to the limits shown in Table 2-6.
7. I/Q Modulation Check (ESG-D Only)
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
I/Q > I/Q Calibration. Check that Full is highlighted on the Calibration Type User Full
softkey. If not, press Calibration Type User Full.
3. Press
range. The message
Execute Cal to begin an I/Q calibration for the signal generator’s full frequency
I/Q Calibration in Progress is displayed until the calibration is
complete.
Calibration Guide2-9
Operation VerificationESG Family Signal Generators
Verification Procedures
Operation VerificationESG Family Signal Generators
Verification Procedures
2-20Calibration Guide
ESG Family Signal Generators
3Service Software
The ESG Family Support Software contains the program and supporting files necessary to
run the automated performance tests and adjustments for your signal generator. This
chapter lists the equipment required to run the software, and gives instructions for
installing, administering, using and un-installing the software. For a description of the
individual performance tests and adjustments, refer to Chapter 4, “Performance Tests,”
and Chapter 5, “Adjustments.”
Calibration Guide3-1
Service SoftwareESG Family Signal Generators
Required Test Equipment
Required Test Equipment
The following equipment is required to run the software:
• Personal computer with the following specifications:
• 386/33 MHz CPU, or better
• 8 Mbytes of RAM
• Hard drive with at least 350 Mbytes free
• 16-color VGA monitor
• 3.5-inch disk drive
• MS Windows 3.x, MS Windows 95, or MS Windows NT.
• 2 Button Mouse (optional)
• Printer, MS Windows-supported (for test results documentation)
• HP 82341C/D GPIB Interface Card
• HP SICL Libraries must be installed and configured.
NOTENo other MS Windows programs may be running during installation. Close
any MS Windows programs that are currently running prior to starting the
installation.
3-2Calibration Guide
ESG Family Signal GeneratorsService Software
Installing the ESG_B or ESG_APDP Service Software
Installing the ESG_B or ESG_APDP Service Software
NOTEIf you are installing the ESG_B or ESG_APDP Service Software onto a
computer which already has service software older than version A.03.00, you
must first remove the old version before installing the new software.
Follow this procedure to install the ESG_B or ESG_APDP Service Software on your
personal computer.
1. Insert “Disk 1” into the disk drive.
2. To display the Run dialog box:
• For MS Windows 95 or Windows NT: Select the Start button, then select Run ... from
the pop-up menu.
• For MS Windows version 3.x: Open the Program Manager, then select Run ... from
the File drop-down menu.
3. From the Run dialog box, type
The Setup window is displayed as it loads files for the installation. Once these files are
loaded, the ESG Service Software’s Welcome screen is displayed.
4. Continue with the setup by selecting the Next button.
5. The ESG Service Software’s Important Information window is displayed. It contains
information that is vital to installing and using the software. In addition, any new
information may also be shown on this screen. Stop and read this information before
continuing with the software installation!
6. Continue with the setup by selecting the Next button.
The Choose Destination Location window is displayed. The default location for
installing the software is C:\HP_SVC\ESG_B (or C:\HP_SVC\ESG_APDP). Use this
as the software’s destination folder.
7. Continue with the setup by selecting the Next button.
The ESG Service Software’s Select Program Folder window is displayed. This
installation procedure will install the service software icons into a program folder. You
can enter a new folder name in the Program Folders text field or select a folder from the
Existing Folders field. It is recommended to use “ESG_B Service Software” (or
“ESG_APDP Service Software”) for the folder name.
a:setup and select the OK button.
NOTEThis documentation refers to the folder name as “ESG_B Service Software.”If
you use another name for the folder, be aware of this difference.
Calibration Guide3-3
Service SoftwareESG Family Signal Generators
Installing the ESG_B or ESG_APDP Service Software
8. Continue with the setup by selecting the Next button.
The ESG Service Software’s, Start Copying Files window is displayed. This window
shows the destination directory and the folder name settings. To change these settings,
select the Back button until the appropriate window is displayed.
NOTEThis is the last point that you can cancel the installation. If you select the
Next button, the installation proceeds until the software is completely
installed.
9. Select the Next button to go ahead with the installation process.
The graphic on the left of the screen shows the status of each file as it is copied to the
destination directory, the status of each disk that is being copied, and the status of the
overall memory of the destination. The graphic in the center of the screen keeps you
updated on the destination folder, the files that are being copied, and the progress of the
installation.
When the contents of the disk have been copied to their destination, you are notified to
insert each of the next disks by the Setup Needs the Next Disk dialog box.
10. Insert the next disk, as instructed, and select the OK button.
When the installation is complete, the Setup Complete window is displayed. You will be
asked to select whether you want to view the README file or the INFO file. (You
already read the INFO file when you started the installation.)
11. Select the README file and then select the Finish button.
The text editor will display the file. To close this file select the File pull down menu and
then select Exit.
12. Select the Yes, I want to restart my computer now radio button. Ensure that the last
disk has been removed and then choose the Finish button to restart MS Windows.
NOTEThe MS Windows program must be restarted before you can use the software.
When you select restart, the computer reboots.
If you do not want to restart MS Windows at this time, select the No, I will restart my
computer later radio button. If you select restart, the computer reboots and you can
start using the software.
3-4Calibration Guide
ESG Family Signal GeneratorsService Software
Uninstalling the Software
Uninstalling the Software
• MS Windows 95 or Windows NT:
1. Display the Control Panel program group by pressing Start > Settings >
Control Panel.
2. Select the Add/Remove Programs icon.
3. From the Install/Uninstall Tab in the Add/Remove Programs properties dialog box:
a. Select ESG_B, then select the Add/Remove button.
b. Select ESG Service Software, then select the Add/Remove button.
4. If the Details button is displayed, the program is removed but the path has not been
removed. Click the Details button and note the path. Then click OK.
5. Display the MS Windows Explorer dialog box by selecting Start > Programs >
Windows Explorer.
6. Select the path that you previously noted, and if you are not using these files for
another purpose, select Delete from the Files drop-down menu to delete the path.
• MS Windows version 3.x:
1. From the Program Manager, click on program group (HP Service Support was the
program’s default title) to highlight the program group name.
2. From the File drop-down menu, select Delete.
3. From the File Manager, select the subdirectory where the program is stored.
(The program’s default subdirectory was hp_svc.)
4. From the File drop-down menu, select Delete.
Calibration Guide3-5
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
ESG Family Support Software Administration
The following section shows you how to administer and run the ESG Family Support
Software. The software’s administrative functions allow for the addition or removal of all
necessary test equipment, software drivers, and test procedures. This would include
additional software drivers (as they are released) to extend the range of tests and
equipment supported by the ESG family test set, or maintenance releases of the current
drivers.
Software Configuration
Follow the instructions below to configure the software to run in either User Mode or
Administration Mode.
Start the software using the steps appropriate for the version of MS Windows that is
installed on your PC.
• For MS Windows version 3.x:
1. Open the Program Manager window.
2. Open the HP Service Support program group.
3. Open the HP Service Software icon.
• For MS Windows 95 or Windows NT:
1. Select Start.
2. Select HP Service Software for PC’s.
3. Select HP Service Software.
The service support software has two configurations: User and Administration. The
following sections explain the difference between the two configurations.
Figure 3-1The User Information Window
3-6Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
The User Configuration
Logging on in User Configuration only allows access to the instrument’s performance tests
and adjustments. Refer to “Starting the Software” on page 3-20.
The Administration Configuration
Logging on in Administration Configuration supports all administrative functions,
including the addition of new equipment and the installation of test procedures and device
drivers.
Administering the Service Support Software
1. Refer to the illustration above and fill in the fields in the User Information window:
a. In the User Name field, type in the word Admin. (Case sensitive.)
b. In the Password field, type in the word Falcon. (Case sensitive.)
2. Click OK. (Cancel closes the software application.)
Identifying the DUT
When the DUT Selection dialog box is displayed:
Calibration Guide3-7
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
1. Make sure that ESG_B is selected in the Select an Instrument Family list.
2. In the Select Model list, select the instrument model of the DUT to be adjusted or
tested.
3. In the Serial Number box, enter the complete serial number of the DUT.
4. In the Address box, enter the two-digit GPIB address of the DUT. (To display the
address on the signal generator, press
Utility > GPIB/RS-232. The GPIB Address softkey
shows the current address.)
5. Select the OK button.
6. The Select Test Equipment and Tests window appears on the display. Close this
window.
Adding Test Equipment
Before any performance tests can be run, information regarding your specific test
equipment must be entered into the software’s database. This information includes the
serial numbers, calibration dates, GPIB address, and traceability data for each individual
component. Only equipment which has been entered into the system will be available for
the test equipment selection process during the test setup. To add test equipment, follow
the instructions below:
1. Refer to Figure 3-2. In the File drop-down menu, select Test Equipment.
Figure 3-2Test Equipment in the File Drop-Down Menu
2. Refer to Figure 3-3. In the Test Equipment dialog box, select the Device Type for the
new equipment you are adding.
3-8Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Figure 3-3Adding Test Equipment Using the Test Equipment Menu
3. Select the model of the device.
4. Click Add.
5. In the New Test Equipment window (refer to Figure 3-4), enter the following
parameters into the appropriate box:
Serial Numberof the new equipment.
GPIB Addressof the new equipment. This address must be in the range of 0
through 31 and it should not conflict with any other instrument
address already present in the test setup.
NOTEThe power sensor must be assigned GPIB address −1 (negative 1).
Cal Due Datethe projected calibration due date of the new equipment.
Trace Numberthe calibration tracking number. This is the last required item.
Calibration Type(optional) is used only when special calibration data is associated
with the device being added, and only as a function setup by the
factory for equipment requiring specific calibration data. The only
devices currently requiring this feature are power sensors.
Table 3-1 is an example of the calibration information required
to ensure accuracy for measurements using the power sensor.
This window is accessed for data entry by selecting the words
CAL DATA from the Calibration Type field in the Edit or New
Test Equipment window, Figure 3-4.
Calibration Guide3-9
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
Table 3-1An Example of Calibration Data for Power Sensors
Frequency (MHz)
0.100097.6
0.300098.9
1.099.1
3.099.4
Figure 3-4Adding the Equipment Information Using the New Test Equipment
Window
Calibration
Factor (%)
6. Click OK.
NOTEThe serial number of the test equipment added will be displayed in the
Equipment field of the Test Equipment dialog box (Figure 3-3).
Removing Test Equipment
Removal of test equipment is accomplished using the Test Equipment dialog box. Refer to
Figure 3-5.
1. Select the Device Type of the test equipment to be removed.
3-10Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Figure 3-5 Removing and Editing Test Equipment Using the Test Equipment
Window
2. Select the model of the test equipment to be removed from the Models field.
3. Select the serial number of the test equipment to be removed from the Equipment field.
4. Click Remove.
5. Click Close.
Editing Test Equipment
Editing test equipment parameters is accomplished using the Test Equipment dialog box.
Refer to Figure 3-5.
1. Select the Device Type of the test equipment to be edited.
2. Select the model of the test equipment to be edited from the Models field.
3. Select the serial number of the test equipment to be edited.
4. Click Edit. The Edit Test Equipment window will appear. Change the field parameters
as necessary.
5. Click OK to save the changes and close the window.
6. Click Close in the Test Equipment window.
Calibration Guide3-11
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
Adding Device Drivers
Follow these instructions to add test equipment device drivers to the program:
CAUTIONThis and the following procedures: Adding/Removing Device (Test
Equipment) Drivers, Adding/Removing Test Drivers, and Adding/Removing
Datapacks are included for informational purposes. These procedures should
only be used when instructions are provided describing which drivers or
datapacks need replacement or removal.
1. Log into the software normally until the Select Test Equipment and Tests window is
displayed.
2. Click the Cancel button to close the Select Test Equipment and Tests window.
3. Refer to Figure 3-6. In the File drop-down menu, select Test Equipment Drivers.
Figure 3-6 Test Equipment Drivers in the File Drop-Down Menu
This selection allows for the addition or removal of software drivers for the test
equipment being used to verify the performance of the DUT.
4. To add a device driver to the existing list of test equipment drivers, click Add
(Figure 3-7).
3-12Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Figure 3-7 Adding a Device Driver Using the Test Equipment Drivers Window
5. Refer to Figure 3-8. Using the standard file search procedure, select the driver that you
are adding and click OK.
Figure 3-8 Using the Open Dialog Box to Search for a Device Driver File to
Add
The selected driver should now be displayed in the Test Equipment Drivers dialog box,
as seen in Figure 3-9.
Calibration Guide3-13
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
Removing Device Drivers
Removing device drivers is accomplished using the Test Equipment Drivers dialog box.
Refer to Figure 3-9.
1. Select the driver (.dll) file to be removed.
Figure 3-9 Removing a Device Driver Using the Test Equipment Drivers
Window
2. Ensure that the information displayed in the Version, Device Type, and Models
Supported fields reflects the correct information for the selected driver being removed.
3. Click Remove.
4. Click Close.
Adding Test Drivers
Adding software drivers is accomplished using the Test Drivers window.
1. Refer to Figure 3-10. In the File drop-down menu, select Test Drivers.
3-14Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Figure 3-10 Test Drivers in the File Drop-Down Menu
2. Refer to Figure 3-11. To add a test driver to the existing list of test drivers, click Add.
Figure 3-11 Adding a Test Driver Using the Test Drivers Window
Calibration Guide3-15
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
3. Refer to Figure 3-12. Using the standard file search procedure, select the test driver
that you are adding and click OK.
Figure 3-12 Using the Open Dialog Box to Search for a Test Driver File to Add
The selected driver should now be displayed in the Test Drivers dialog box, as seen in
Figure 3-11.
4. Click Close (Figure 3-11).
3-16Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Removing Test Drivers
Removing test drivers is accomplished using the Test Drivers dialog box. Refer to
Figure 3-13.
1. Select the driver (
.dll) file to be removed.
Figure 3-13 Removing a Test Driver Using the Test Drivers Window
2. Ensure that the information displayed in the Version, Required Devices, and Tests
Supported fields reflects the correct information for the selected driver being removed.
3. Click Remove.
4. Click Close.
Adding Datapacks
Adding datapacks for test procedures is accomplished using the Datapacks dialog box.
1. Refer to Figure 3-14. In the File drop-down menu, select Datapacks.
Calibration Guide3-17
Service SoftwareESG Family Signal Generators
ESG Family Support Software Administration
Figure 3-14 Datapacks in the File Drop-Down Menu
2. Refer to Figure 3-15. To add a datapack to the existing list of datapacks, click Add.
Figure 3-15 Adding a Datapack Using the Datapacks Window
3. Refer to Figure 3-16. Using the standard file search procedure, select the datapack that
you are adding and click OK.
3-18Calibration Guide
ESG Family Signal GeneratorsService Software
ESG Family Support Software Administration
Figure 3-16 Using the Open Dialog Box to Search for a Datapack File to Add
The selected driver should now be displayed in the Datapacks dialog box, as seen in
Figure 3-15.
4. Click Close (Figure 3-15).
Removing Datapacks
Removing datapacks is accomplished using the Datapacks dialog box. Refer to Figure 3-17.
1. Select the datapack (
Figure 3-17 Removing a Datapack Using the Datapacks Window
.000) file to be removed.
2. Click Remove.
3. Click Close.
Calibration Guide3-19
Service SoftwareESG Family Signal Generators
Running the Service Support Software
Running the Service Support Software
Starting the Software
1. Start the software using the steps appropriate for the version of MS Windows that is
installed on your PC.
• For MS Windows 95:
a. Select Start.
b. Select HP Service Support for PC’s.
c. Select HP Service Software.
• For MS Windows version 3.x:
a. Open the Program Manager window.
b. Open the HP Service Support program group.
c. Select the HP Service Support icon.
2. Once the User Information dialog box is displayed, enter the user name and the
password.
a. In the User Name box, enter User. (Case sensitive.)
b. In the Password box, enter User. (Case sensitive.)
c. Click OK.
3-20Calibration Guide
ESG Family Signal GeneratorsService Software
Running the Service Support Software
Identifying the DUT
When the DUT Selection dialog box is displayed,
1. Make sure that ESG_B is selected in the Select An Instrument Family list.
2. In the Select Model list, select the signal generator model of the DUT to be adjusted or
tested.
3. In the Serial Number box, enter the complete serial number of the DUT.
4. In the Address box, enter the two-digit GPIB address of the DUT. (To display the
address on the signal generator, press
Utility > GPIB/RS-232. The GPIB Address softkey
shows the current address.)
5. Select the OK button.
Calibration Guide3-21
Service SoftwareESG Family Signal Generators
Running the Service Support Software
Selecting Performance Tests or Adjustments and the Test
Equipment
When the Select Test Equipment and Tests window is displayed,
1. Select either the Performance Tests radio button to display the list of automated
performance tests or the Adjustments radio button to display the list of automated
adjustments.
2. From the list of performance tests or adjustments, select the tests or adjustments that
will be performed on the DUT. Select a test name by clicking on it (the DCFM
FREQUENCY OFFSET test will be selected for the sections following). The selected
test will be highlighted.
Select all of the performance tests or adjustments by selecting the Select All button.
Unselect all of the selected tests or adjustments by choosing the Unselect button.
As each test or adjustment is selected, the test equipment required to perform the test
or adjustment is listed in the Required Test Equipment box. As each test or adjustment
is removed from the list, the test equipment required to perform the test or adjustment
is removed from the list. Only the test equipment for the highlighted tests is displayed
in the list.
3-22Calibration Guide
ESG Family Signal GeneratorsService Software
Running the Service Support Software
3. Once the test is highlighted, identify all test equipment listed in the Required Test
Equipment box. To identify test equipment:
a. Select the type of device from the Device Type list.
b. Select the model number from the Model list for the device.
c. Select the device’s serial number from the Available Test Equipment list.
d. Select the Add button to add the device to the list in the Selected Test Equipment
box.
The following buttons are available for the Selected Test Equipment box:
AddCopies the tests and adjustments highlighted in the Available Test
Equipment box to the Selected Test Equipment box.
RemoveRemoves the highlighted test equipment from the Selected Test
Equipment box.
ClearRemoves all of the test equipment from the Selected Test Equipment
box.
NOTEIf necessary, the test equipment GPIB address can be changed after it is
added to the Selected Test Equipment box. Change the GPIB address by
pressing the right arrow on the keyboard until the GPIB address selection in
the Selected Test Equipment box is selected. (The GPIB address is selected
when it has a dark box around the selection.) Then, type the new GPIB
address and press Enter to change the address.
e. Repeat the previous steps for each test equipment instrument.
4. Once the test equipment has been identified, select the Add button located above the
Selected Testsbox to add each of the highlighted performance tests or adjustments. The
tests or adjustments that were highlighted are now displayed in the list in the Selected
Tests box.
The following buttons are available for the Selected Tests box:
AddCopies the tests and adjustments highlighted in the list of tests and
adjustments to the Selected Tests box.
RemoveRemoves the highlighted tests and adjustments from the Selected Tests
box.
ClearRemoves all of the tests and adjustments from the Selected Tests box.
Calibration Guide3-23
Service SoftwareESG Family Signal Generators
Running the Service Support Software
5. If the appropriate tests or adjustments are listed in the Selected Tests box and the
appropriate test equipment is listed in the Selected Test Equipment box, select the OK
button.
Defining the Location where the Test Results are Saved
1. When the Save As dialog box is displayed, select the File Name box and enter the file
name into which you would like to save the test results.
The results file name suffix is
You may select the directory into which you would like to save the file by changing the
drive and folder in this window. This file is a text file and can be viewed using many text
editors. Notepad, a standard accessory in MS Windows, is an example of a text editor
that can be used to view text files.
.log. The results are saved automatically to the.log file.
3-24Calibration Guide
ESG Family Signal GeneratorsService Software
Running the Service Support Software
2. Once you have selected the drive, folder, and assigned a file name, select the OK button
to save these settings for when the tests are complete.
Running the Tests and Adjustments
Once the HP Service Support Software window is displayed:
1. Select the Run button to start the automated tests or adjustments displayed in the
Selected Tests box.
The software steps through the tests or adjustments sequentially.
2. Follow the instructions displayed on the PC.
A description for each automated performance test or adjustment can be found in
Chapter 4, “Performance Tests,” and Chapter 5, “Adjustments.”
Calibration Guide3-25
Service SoftwareESG Family Signal Generators
Running the Service Support Software
Reviewing the Test and Adjustment Results
Once the tests have finished running, you will want to review the results of the tests. The
HP Service Support Software window displays the DUT model number and serial number,
the selected tests, the results of the selected test, and the current test information. Six
buttons, which are used to control the testing, are also displayed.
The Selected Tests box displays the name of the selected tests, the pass/fail status (P/F) of
each test that has been run, the total number of points that each test checks, and the
number of points that passed and failed for each test. The pass/fail status indicates a
failure if any point in that test fails.
The Selected Test Results box shows the results of the test that is highlighted in the
Selected Tests box. The Selected Test Results box shows the pass/fail status (P/F), the
lower limits (LL), the measured value (Result), the upper limits (UL), and the measured
units (for example, kHz, mV, or dBm) for each test point checked by the performance tests
and some adjustments. “**” indicates values not displayed by the adjustments.
The Current test box shows the results of the test that is currently running. The Current
test box shows the pass/fail status (P/F), the lower limits (LL), the measured value
(Result), the upper limits (UL), and the measured units (for example, kHz, mV, or dBm) for
each test point checked by the performance tests and some adjustments. “**” indicates
values not displayed by the adjustments.
The following buttons are also displayed on the HP Service Support Software window.
These buttons are used to control the testing. Only the buttons that are appropriate are
active. For example, if the test is already in progress, the Run button would not be active.
RunStart running the highlighted test when initially starting the testing. Also
used to start testing at the same point where the test was stopped.
StopStop the test that is currently running. The test stops after making the
next measurement.
3-26Calibration Guide
ESG Family Signal GeneratorsService Software
Running the Service Support Software
RestartRerun a test that was running when the testing was stopped. This restarts
the test from the beginning.
Next TestQuit running the current test and give it a Fail status. Then, continue
testing with the next test.
RerunRestart the testing at the beginning of the first test.
AbortQuit testing. Abort all tests.
Printing the Test and Adjustment Results
Once the tests are complete, a dialog box is displayed that asks if you want to print the log
file.
The log file is the file in which the test (or adjustment) results are stored. Select the Yes
button to print the test results using the printer connected to LPT1. Choosing the No
button allows you to exit the program without printing the test results.
There are two other methods of printing the test results.
The first method is selecting the Print Log File selection from the File drop-down menu.
The software asked you to define the computer path and file name when the tests were
performed. The default directory is the “log” subdirectory of the destination directory
where you installed the software. (C:\HP_SVC\EGS_B\ was the default destination
directory.)
The second alternative method of printing is opening the file in a text editor and printing
the file in the text editor.
Exiting the Software
Exit the software by selecting Exit from the file drop-down menu.
Pressing Alt-F4 on the keyboard is an alternate method of exiting the software.
Calibration Guide3-27
Service SoftwareESG Family Signal Generators
Running the Service Support Software
3-28Calibration Guide
ESG Family Signal Generators
4Performance Tests
Unless stated otherwise, the procedures in this chapter enable you to test the electrical
performance of the signal generator to its specifications.
Calibration Guide4-1
PerformanceTestsESG Family Signal Generators
Calibration Cycle
This instrument requires periodic verification of performance. Under normal use and
environmental conditions, the instrument should be calibrated every two years. Normal
use is defined as about 2,000 hours of use per year.
PerformanceTest Records
Unless stated differently, the ESG Family Support Software can print a test record that
contains the results of the automated performance tests. For each of the two manual
performance tests, “Timebase Aging Rate (ESG-AP, ESG-DP or Option 1E5 only)” and
“Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal
Generators),” a performance test record is located at the end of this chapter so it can be
photocopied for use.
Recommended Test Equipment
The recommended test equipment table, located in Chapter 1, “Equipment Required,” is
the complete list of equipment required to perform the procedures in this chapter. Test
equipment is also listed with each test and is shown in the test setup. However, to avoid
clutter in the figures, the controller required to run the support software is not shown in
the test setup for each automated test. The controller must be connected via GPIB to all of
the test equipment in the currently selected automated tests. For performance tests, other
equipment can be substituted for the recommended models if it meets or exceeds the
critical specifications listed in Chapter 1, “Equipment Required.” For the adjustment
procedures, no equipment substitutions are available.
4-2Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
Support Software
Support Software
The ESG Family Support Software runs performance tests and where applicable,
generates reports of the results for the tests. The following manual tests are not listed in
the software:
• Timebase Aging Rate (ESG-AP,
ESG-DP or Option 1E5 only)
• Measuring Phase Noise and Residual FM
(ESG-AP and ESG-DP Series Signal Generators)
This chapter shows how to run the software to test and verify the performance of a signal
generator. Chapter 3, “Service Software,” explains how to set up the software for a
particular set of test instruments and interfaced components in a test environment.
Running the Performance Tests
1. Load the support software into the personal computer. Refer to Chapter 3, “Service
Software,” for installation instructions.
2. Click on the installed support software icon.
3. Fill in the fields in the User Information window (Figure 4-1):
a. In the User Name field (item 1), enter User. (Case sensitive.)
b. In the Password field (item 2), enter User. (Case sensitive.)
4. Click OK.
Figure 4-1The User Information Window
5. In the DUT (Device Under Test) Selection Window (Figure 4-2):
a. In the Select Model list (item 3), choose the correct model number of your signal
generator.
b. In the Serial Number box (item 4), enter the complete serial number of the DUT.
c. In the Address box (item 5), enter the two-digit GPIB address of the signal generator.
(To display the address on the signal generator, press
GPIB Address. The GPIB Address softkey shows the current address.)
Utility > GPIB/RS-232 >
6. Click OK.
Calibration Guide4-3
PerformanceTestsESG Family Signal Generators
Support Software
Figure 4-2The DUT Selection Window
7. In the Select Test Equipment and Tests window (Figure 4-3), create a list of test
equipment that will verify the signal generator’s performance:
a. In the Device Type list (item 6), select a device type that you want to add to the
equipment list.
b. In the Model list (item 7), select the model of the device type that you want to add to
the equipment list.
c. In the Available Test Equipment box (item 8), select the serial number of an
instrument. There may be more than one instrument available for any model. Make
sure that the instrument calibration due date has not passed. If the due date has
passed, use another available instrument. Also, alert your system administrator that
the instrument is due for calibration.
d. Click Add (item 9) to add the instrument to the test equipment list.
e. Continue adding test equipment to the list by repeating these steps.
NOTEYou can correct mistakes while you are entering instruments in the test
equipment list:
• To delete a single entry in the equipment list, click on the unwanted
selection in the Selected Test Equipment box and then click Remove (item
11).
• To delete the entire list of equipment, click Clear (item 12).
4-4Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
Support Software
Figure 4-3The Select Test Equipment and Tests Window
8. Create a list of performance tests that you want the software to run (Figure 4-3):
a. In the Available Tests list (item 13), select a performance test.
b. Click Add (item 14) to add the test to the list of tests that the software will run.
You can view the Required Test Equipment list (item 15) for a performance test by
clicking on the test title in the Available Tests (item 13) box.
c. Continue adding performance tests to the list by repeating these steps.
NOTEYou can correct mistakes while you are entering tests in the performance test
list:
• To delete a single entry in the test list, select the unwanted test in the
Selected Tests box (item 16) and then click Remove (item 17).
• To delete the entire list of equipment, click Clear (item 18).
9. When you are done, click OK (item 19).
Calibration Guide4-5
PerformanceTestsESG Family Signal Generators
Support Software
10.To save the test results to a file, make the following selections (Figure 4-4):
a. Click on the Drives window (item 20) and select the drive where you want to save the
test results.
b. Click on the Save file as type window (item 21) and select the file type for saving the
test results.
c. Click on the File name window (item 22) and type in a name for the test results file.
11.Click OK (item 23).
Figure 4-4The Save As Window
12.In the Main Test and Results window (Figure 4-5), verify that the selected DUT (item
24), serial number (item 24), and selected tests (item 25) are correct. Then click Run.
The selected performance tests are listed in the order in which they will run. You
cannot change this order.
During the testing sequence the window shows the following information:
• the model and serial number of the signal generator under test (item 24)
• the list of selected performance tests with the current test highlighted (item 25)
• the performance test currently running (item 26)
• the data points taken (item 27)
• the total number of data points that will be measured during the test (item 28)
• the number of points that have currently passed the test (item 29)
• the number of points that have currently failed the test (item 30)
• the lower limits (LL) (item 31) and upper limits (UL) (item 32) for each test point
• the measured results (item 33) and the pass/fail (P/F) (item 34) indication
4-6Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
Support Software
You can scroll the results window, vertically and horizontally. You can also click on the
following buttons during the test sequence:
RunStart running the highlighted test when initially starting the testing.
Also used to start testing at the same point where the test was stopped.
StopStop the test that is currently running. The test stops after making the
next measurement.
RestartRerun a test that was running when the testing was stopped. This
restarts the test from the beginning.
Next TestQuit running the current test and give it a Fail status. Then, continue
testing with the next test.
RerunRestart the testing at the beginning of the first test.
AbortQuit testing. Abort all tests.
Tobegin a performance test sequence with a new DUT, click File and New Session. This
allows you to reselect the DUT, the equipment list, and the performance test list.
Figure 4-5 The Main Test and Results Window
13.When the software has run all of the selected performance tests, the status bar
indicates that the tests are completed (item 35).
Calibration Guide4-7
PerformanceTestsESG Family Signal Generators
Support Software
Printing the Performance Test Results
1. To print all the results of the entire test sequence, click File and Print.
Exiting the Support Software Program
1. To exit the performance tests program, click File and Exit.
4-8Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
1. Internal FM Accuracy and Distortion
1. Internal FM Accuracy and Distortion
These automated tests verify the FM accuracy and distortion specifications. Accuracy is
verified directly with a measuring receiver, and distortion is verified by measuring the
demodulated output of the measuring receiver, using the audio analyzer. The audio
analyzer provides a more accurate verification with better resolution, as opposed to the
measuring receiver alone.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8903B Audio Analyzer
Equipment Setup
Figure 4-6Internal FM Accuracy and Distortion Tests Setup
Calibration Guide4-9
PerformanceTestsESG Family Signal Generators
2. Internal AM Accuracy and Distortion
2. Internal AM Accuracy and Distortion
These automated tests verify the AM accuracy and distortion specifications.
For frequencies at or below 1300 MHz, accuracy is verified directly with a measuring
receiver, and distortion is verified by measuring the demodulated output from the
measuring receiver with an audio analyzer. The audio analyzer provides a more accurate
verification with better resolution than using the measuring receiver alone.
For frequencies above 1300 MHz, the RF output from the signal generator is mixed with
the output from another signal generator (local oscillator) to produce a suitable input
frequency for the measuring receiver.
For a frequency of 4 GHz, verification is accomplished by covering two different frequency
spans through the use of two mixers.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8903B Audio Analyzer
• HP/Agilent 8663A Signal Generator
• MD/MDC-174 Mixer (1 to 2800 MHz)
• MD/MDC-164 Mixer (0.5 to 9 GHz)
Equipment Setups
Figure 4-7Accuracy and Distortion for AM ≤ 1300 MHz Test Setup
4-10Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
2. Internal AM Accuracy and Distortion
Figure 4-8Accuracy and Distortion for AM > 1300 MHz and ≤ 2500 MHz Test Setup
Figure 4-9Accuracy and Distortion for AM > 2500 MHz Test Setup
Calibration Guide4-11
PerformanceTestsESG Family Signal Generators
3. Phase Modulation Accuracy and Distortion
3. Phase Modulation Accuracy and Distortion
These automated tests verify the phase modulation accuracy and distortion specifications.
Accuracy is verified directly with a measuring receiver, and distortion is verified by
measuring the demodulated output from the measuring receiver, using the audio analyzer.
The audio analyzer provides a more accurate verification with better resolution than using
the measuring receiver alone.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8903B Audio Analyzer
Equipment Setup
Figure 4-10Accuracy & Distortion for Phase Modulation Tests Setup
4-12Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
4. FM Frequency Response
4. FM Frequency Response
This automated test verifies the FM frequency response specifications. The equipment
measures the variations in frequency deviations due to changes in the applied FM rate; dc
to 100 kHz. The variations are expressed relative to a reference signal; 1 kHz rate set at
100 kHz deviation, in dB. Each frequency is tested with this sequence:
1. A reference signal is measured.
2. The set deviation is measured for each applied rate change.
3. Each deviation measurement is compared to the reference signal deviation.
The comparison is a calculation of the difference in the deviations relative to the reference
rate. The relative value is the FM frequency response.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8904A Function Generator
Equipment Setup
Figure 4-11 FM Frequency Response Tests Setup
Calibration Guide4-13
PerformanceTestsESG Family Signal Generators
5. AM Frequency Response
5. AM Frequency Response
This automated test verifies the AM frequency response specifications. The equipment
measures the variations in modulation depth due to changes in the applied AM rate; dc to
10 kHz. The response is relative to a 1 kHz rate set at the test depth, and expressed in dB.
Each frequency is tested with the following sequence:
1. A reference signal is measured.
2. The set depth is measured for each applied rate change.
3. Each depth measurement is compared to the reference signal depth.
The comparison is a calculation of the difference in the depths relative to the reference
rate. The relative value is the AM Frequency Response.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8904A Function Generator
• HP/Agilent 8663A Signal Generator
• MD/MDC-164 Mixer (0.5 to 9 GHz)
Equipment Setups
Figure 4-12 Frequency Response for AM ≤ 1300 MHz Test Setup
4-14Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
5. AM Frequency Response
Figure 4-13 Frequency Response for AM > 1300 MHz Test Setup
Calibration Guide4-15
PerformanceTestsESG Family Signal Generators
6. Phase Modulation Frequency Response
6. Phase Modulation Frequency Response
This automated test verifies the phase modulation frequency response specifications. The
equipment measures the variations in phase deviations due to changes in the applied
ΦM rate; dc to 100 kHz. The response is relative to a 1 kHz reference at the same phase
deviation, and expressed in dB. Each frequency is tested with the following sequence:
1. A reference signal is measured.
2. The set deviation is measured for each applied rate change.
3. Each deviation measurement is compared to the reference signal deviation.
The comparison is a calculation of the difference in the deviations relative to the reference
rate. The relative value is the ΦM frequency response.
Recommended Equipment
• HP/Agilent 8902A Measuring Receiver Option 010
• HP/Agilent 8904A Function Generator
Equipment Setup
Figure 4-14 Phase Modulation Frequency Response Test Setup
4-16Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
7. DCFM Frequency Offset Relative to CW
7. DCFM Frequency Offset Relative to CW
This automated test verifies the carrier frequency offset, relative to CW. The equipment
measures the RF output frequency. For each test point, a comparison is made between a
CW reference frequency without DCFM selected, and then with the DCFM selected; the
difference is the carrier frequency offset.
Recommended Equipment
• HP/Agilent 53132A Frequency Counter Option 050
Equipment Setup
Figure 4-15 DCFM Frequency Offset Test Setup
Calibration Guide4-17
PerformanceTestsESG Family Signal Generators
8. Residual FM (ESG-A and ESG-D only)
8. Residual FM (ESG-A and ESG-D only)
This automated test verifies the residual frequency modulation (FM) specification. The
signal generator’s RF output signal is mixed with a signal from a low noise external LO.
The resultant IF signal is demodulated as it passes through the measuring receiver,
bypassing the measuring receiver’s internal mixer which is inherently more noisy. Then
the audio analyzer measures the demodulated audio signal for an RMS voltage. The
measurement is scaled and displayed as RMS mV, but read as residual FM in Hz.
9. Harmonic, Subharmonic, and Nonharmonic Spurious Signals
9. Harmonic, Subharmonic, and
Nonharmonic Spurious Signals
These automated tests verify that the harmonic, subharmonic, and nonharmonic spurious
signals are within specifications. The signal generator’s output signal is set to values
where harmonic and spurious signal performance problems are most likely to occur. Then
the spectrum analyzer checks the level of the various harmonic and spurious signals and
compares each measurement to a signal of known frequency and amplitude.
Figure 4-18 Harmonic, Subharmonic, and Nonharmonic Spurious Signals Tests Setup
4-20Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
10. Power Level Accuracy
10. Power Level Accuracy
A power meter is used to verify performance over the +13 to –15 dBm range. The absolute
power level measured at –10 dBm is used as a reference for all lower level measurements.
For power levels below −15 dBm, a vector signal analyzer is used to make relative power
measurements. A low noise amplifier (LNA) and step attenuator are connected in series to
control the absolute power level input to the signal analyzer. For power level settings
≥ –60 dBm, the step attenuator is set to approximately cancel the gain of the LNA. Below
–60 dBm, the step attenuator is set to 0 dB, which keeps the signal level well above the
signal analyzer’s noise floor.
Relative measurements are accomplished by setting the DUT amplitude. For example:
Set the DUT to a specific frequency, and an amplitude of –10 dBm. For the purpose of this
example, assume the power meter measures –10.1 dBm at these settings. The signal
analyzer does a peak search, sets the reference level, and sets the marker to delta mode
(0 dB). The DUT amplitude then changes to –20 dBm, and the signal analyzer makes a
relative power measurement (assume it to be –10.15 dB). The absolute power measured is
–10.1 dBm + (–10.15 dB), or –20.25 dBm.
To determine power level accuracy below –60 dBm, the signal analyzer makes the 0 dB
marker delta measurement with the amplitude set to –60 dBm, and all lower level
measurements are made relative to this reference. The previously measured error for the
–60 dBm setting is then accounted for in the measurement.
Forfrequencies above 2 GHz, an HP/Agilent 8563E signal analyzer is used to downconvert
the frequency to 321.4 MHz.
Recommended Equipment
• HP/Agilent 438A Power Meter
• HP/Agilent 8482A Power Sensor
• HP/Agilent 89441A Vector Signal Analyzer
• HP/Agilent 8563E Option 001 Signal Analyzer
• HP/Agilent 8491A/B Option 006 Attenuator (6 dB)
• HP/Agilent 8495G Programmable Step Attenuator (0 to 70 dB)
• HP/Agilent 11713A Step Attenuator Driver
See page 1-4 for Details on the following:
• Low Frequency Noise Amplifier (LNA)
• High Frequency Low Noise Amplifier (LNA)
Calibration Guide4-21
PerformanceTestsESG Family Signal Generators
10. Power Level Accuracy
Equipment Setups
Figure 4-19 High-Power, Power Level Accuracy Setup
Figure 4-20 Low-Power, Power Level Accuracy (< 10 MHz) Setup
4-22Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
10. Power Level Accuracy
Figure 4-21 Low-Power, Power Level Accuracy (≥ 10 MHz and ≤ 2 GHz) Setup
Figure 4-22 Low-Power, Power Level Accuracy (> 2 GHz) Setup
11. Timebase Aging Rate
(ESG-AP, ESG-DP or Option 1E5 only)
This manual test verifies the accuracy of the signal generator’s internal timebase. The
time required for a 360° phase change is measured both before and after a specified
waiting period. The aging rate is inversely proportional to the absolute value of the
difference in the measured times. The results of this test can be recorded on the test record
provided on page 4-48.
The overall accuracy of the internal timebase is a function of:
TBC ±AR ±TE ±LE where:
TBC = timebase calibration
AR = aging rate
TE = temperature effects
LE = line effects
After the timebase is adjusted, the timebase frequency should stay within the aging rate
for the following situations:
• The timebase oven does not cool down.
• The instrument keeps the same orientation, with respect to the earth’s magnetic field.
• The instrument stays at the same altitude.
• The instrument does not receive any mechanical shock.
If the timebase oven cools (the instrument power switch is set to off), you may have to
readjust the timebase frequency after a new warm-up cycle. Typically, however, the
timebase frequency returns to within ±1 Hz of the original frequency.
NOTEThe internal timebase can be tested after the AC power is reconnected for 10
minutes. For best accuracy, retest after the instrument has been on for 24
hours.
Frequency changes due to either a change in orientation, with respect to the earth’s
magnetic field, or to a change in altitude usually go away when the instrument is returned
to its original position. A frequency change due to mechanical shock usually appears as a
fixed frequency error.
9. Write the test results on the performance test record located at the end of this chapter.
Compare the results to the limits in the test record.
NOTEIf the absolute frequency of the standard and the timebase oscillator are
extremely close, you can reduce the measurement time (T1 and T3) by
measuring the time required for a phase change of less than 360°. In step 4,
change 1 cycle to 0.5 cycle for 180°, or 0.25 cycle for 90°.
4-26Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
12. Digital Modulation Level Accuracy
12. Digital Modulation Level Accuracy
This automated test verifies the level accuracy of the signal generator’s I and Q
modulation inputs. Two arbitrary waveform generators provide the I and Q modulation
inputs to the signal generator. A power meter measures the RF output power with and
without QPSK modulation applied to the signal generator. The difference in power
measurements is the level accuracy. The QPSK modulation is loaded into the two arbitrary
waveform generators from a data file in the performance tests software. The vector signal
analyzer verifies the arbitrary waveform generator’s outputs are properly scaled and
synchronized before the relative power level accuracy is measured.
• HP/Agilent 89441A Options AY7, AY9, AYA, UFG or UTH Vector Signal Analyzer
Equipment Setup
Figure 4-24 Digital Modulation Level Accuracy Test Setup
Calibration Guide4-27
PerformanceTestsESG Family Signal Generators
13. Internal Digital Modulation Quality
13. Internal Digital Modulation Quality
NOTEThis test is only required for instruments with UN3, UN4, or UN8.
This automated test verifies the RF modulation quality of the signal generator’s internal
I/Q modulation. A vector signal analyzer is connected to the signal generator’s RF output.
The internal baseband generator modulates the RF carrier in each of the available digital
modulation formats. The vector signal analyzer measures the appropriate error parameter
for the modulation generated (EVM for PHS, PDC and NADC formats; global phase error
for GSM format).
Recommended Equipment
• HP/Agilent 89441A Options AYA, AY9, UFG or UTH Vector Signal Analyzer
Equipment Setup
Figure 4-25 Internal Digital Modulation Quality Test Setup
This automated test verifies the RF modulation quality of the Option UN8 signal
generator’s internal real time I/Q baseband modulation. A vector signal analyzer is
connected to the signal generator’s RF output. The Option UN8 internal baseband
generator modulates the RF carrier in each of the TDMA digital modulation formats. The
parameters of the individual TDMA formats common to Options UN3, UN4 and UN8 are
tested elsewhere. The vector signal analyzer measures the appropriate error parameter for
the modulation generated:
• EVM for QPSK, π/4DQPSK, and QAM16
• Phase error for MSK
• Frequency Shift Error (%) for FSK2
Recommended Equipment
• MD/MDC-174 Mixer (0.001 to 2.8 GHz)
• HP/Agilent 8663A Signal Generator
• HP/Agilent 89441A Options AY9, AYA, AYH, UFG or UTH Vector Signal Analyzer
This test measures the accuracy of the rear-panel I and Q output signals created by the
internal baseband generator in Option UN3, UN4, or UN8 signal generators. The vector
signal analyzer is configured to measure the I and Q baseband signals and demodulate
each modulation type separately.
3. Follow the instructions as they appear on the controller’s display.
Calibration Guide4-31
PerformanceTestsESG Family Signal Generators
16. Pulse Modulation On/Off Ratio
16. Pulse Modulation On/Off Ratio
This automated test verifies the pulse modulation on/off ratio. The signal generator is
configured for an external pulse input. The spectrum analyzer measures the RF output
power with and without the external pulse applied. The power difference is the on/off ratio.
NOTEThis test does not test the high performance pulse circuitry used in
Option 1E6 instruments. See “20. Pulse Rise/Fall Time (Option 1E6 only)” on
page 4-36.
Figure 4-29 Pulse Modulation On/Off Ratio Test Setup
4-32Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
17. Burst Modulation On/Off Ratio (ESG-D only)
17. Burst Modulation On/Off Ratio (ESG-D only)
This automated test verifies the burst modulation on/off ratio. The signal generator is
configured for an external burst input. The spectrum analyzer measures the RF output
power with and without the external burst applied. The power difference is the on/off ratio.
Figure 4-30 Burst Modulation On/Off Ratio Test Setup
Calibration Guide4-33
PerformanceTestsESG Family Signal Generators
18. CDMA Adjacent Channel Power (Option UN5 only)
18. CDMA Adjacent Channel Power (Option UN5 only)
This automated test verifies the energy at an offset to the main channel relative to the
total energy in the main channel. Post-processing averaging techniques are used to
improve repeatability.
Figure 4-31 Adjacent Channel Power Test Setup (Option UN5)
4-34Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
19. Alternate Timeslot Power Settling Time (OptionUNA only)
19. Alternate Timeslot Power Settling Time
(Option UNA only)
This test measures the time required for a waveform envelope to settle within the
tolerance of its final amplitude after a sharp amplitude transition provided by the
Alternate Timeslot Power feature (Option UNA). Settling time is the result of heating and
cooling effects in the step attenuator and ALC loop. The rise and fall response times of the
ALC loop and each attenuator segment is characterized and combined to construct the
settling curve for any arbitrary transition.
Figure 4-32 Alternate Timeslot Power Settling Time Test Setup (Option UNA)
Procedure
1. Connect the equipment as shown in Figure 4-32.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.
Calibration Guide4-35
PerformanceTestsESG Family Signal Generators
20. Pulse Rise/Fall Time (Option 1E6 only)
20. Pulse Rise/Fall Time (Option 1E6 only)
This test checks the rise/fall time performance of the RF carrier when pulse modulated by
way of the PULSE IN path. The oscilloscope captures the rising or falling edge of the RF
pulse envelope, and calculations are made from the digitized data. Transition time
calculated is for the 10% to 90% portion of the envelope.
Recommended Equipment
• HP/Agilent 54750A Digitizing Oscilloscope
• HP/Agilent 54751A or 54752A Plug-in
• HP/Agilent 33120A Function Generator
• HP/Agilent 8491A/B Opt 010 Attenuator (10 dB)
• HP/Agilent 8491A/B Opt 020 Attenuator (20 dB)
Equipment Setup
Figure 4-33 Pulse Rise/Fall Time Setup (Option 1E6)
4-36Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)
21. Measuring Phase Noise and Residual FM
(ESG-AP and ESG-DP Series Signal Generators)
The signal generator’s automated performance tests do not include a test for either phase
noise or residual FM. These tests must be done separately and the results attached to the
test results obtained with the performance test software. Either output a plot for each
measured frequency, or copy the data onto the test record provided on page 4-48.
Measure phase noise and residual FM on an HP/Agilent E5500 system or its equivalent, at
0.5, 1, 2, 3, and 4 GHz as applicable to the model under test.
Measuring Phase Noise
Because of the excellent phase noise performance of ESG-P series signal generators, a
special procedure and extensive equipment is required. The recommended method is to use
an HP/Agilent E5500 series phase noise measurement system. This system is
recommended over the older HP/Agilent 3048 system due to speed issues and the fact that
the older system can not directly measure residual FM. Most of the rest of this discussion
refers to the E5500 system, although the principles apply to both.
The limitation of any measurement of phase noise is that the resulting measurement can
never be better than the phase noise of the source used to down-convert the signal. In an
E5500 system this source is typically either an HP/Agilent 8644B or an
HP/Agilent 8662A/63A. Both have advantages and disadvantages. The 8644B has better
phase noise above 10 kHz from the carrier, while the 8662/3 has better close-in phase
noise, but is worse above 10 kHz. Other sources may have different characteristics. To fully
characterize a signal generator to a high degree of accuracy, both sources may be needed.
The following sections illustrate three different measurements of the same device under
test (DUT):
• “Using an HP/Agilent 8644B” on page 4-38
• “Using an HP/Agilent 8663A Synthesizer” on page 4-39
• “Using Electronic Frequency Locking (EFC)” on page 4-40
Calibration Guide4-37
PerformanceTestsESG Family Signal Generators
21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)
Using an HP/Agilent 8644B
Figure 4-34 shows the phase noise results of the DUT using the 8644B as the
down-converting source. Note that while it has excellent phase noise above 10 kHz, the
results below 10 kHz rapidly degrade. The 60 Hz related noise spikes are typically due to
the DCFM method of phase locking the DUT to the E5500.
Figure 4-34
4-38Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)
Using an HP/Agilent 8663A Synthesizer
Figure 4-35 shows the results of the same DUT measured with a 8663A Synthesizer. In
this measurement, the DCFM method was also used to phase lock the DUT to the E5500
system. Again, this produces 60 Hz-related spikes as well as other spurs at 20 and
100 kHz. Notice that the phase noise above 10 kHz is also worse than with the
HP/Agilent 8644B. The DCFM method of testing phase noise with an HP/Agilent 8662/3
should not be used.
Figure 4-35
Calibration Guide4-39
PerformanceTestsESG Family Signal Generators
21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)
Using Electronic Frequency Locking (EFC)
Figure 4-36 shows the result of using the electronic frequency control (EFC) method of
phase-locking the DUT to the E5500 system (see the E5500 manual for set up). This
method produces the best result, but is not available on all sources (such as the
HP/Agilent 8644B). Note that the close-in phase noise is now much improved and the
60 Hz related spikes are eliminated. Although the two large spikes still exist, they are the
result of the HP/Agilent 8662/3 and are not characteristics of the DUT.
Figure 4-36
To get an accurate idea of the actual phase noise of the DUT, it may be necessary to
combine the two plots, always using the lowest result of both (see Figure 4-37). Keep in
mind that the actual phase noise results are probably better than what is displayed.
The results shown on the E5500 system may be worse than the actual DUT. If a tested
device has the same phase noise characteristics as the system’s down-converting source,
the displayed result will be 3 dB worse than either device. Toeliminate this error,measure
three different sources. This results in three equations and three unknowns, and the
software can then sort out the actual results for each device. This 3-source substitution
method generates a correction table that can be used for future measurements, effectively
eliminating the combined error (of up to 3 dB) for subsequent measurements. For best
results, use this method. Refer to the E5500 documentation for instructions on how to
implement a 3-source substitution. When the plots in this section were taken, this method
was not used. If it had been used, portions of the plots could have been up to 3 dB better.
Another way of stating the above: for any phase noise measurement where a 3-source
substitution is not performed, the actual phase noise of one of the two sources involved
(DUT or down-converting source) is at least 3 dB better than shown.
4-40Calibration Guide
ESG Family Signal GeneratorsPerformanceTests
21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)
Figure 4-37
Measuring Residual FM
Residual FM is closely related to phase noise. Good phase noise typically implies good
residual FM. Directly measuring residual FM at very low levels is difficult, but the
HP/Agilent E5500 system can integrate the phase noise results to determine an accurate
value. This is done under the E5500 software Trace Integration menu by selecting a data
type of
In the ESG-P series, residual FM is defined only within a 300-3000 Hz bandwidth and
with a specified frequency response within this band. That frequency response is specified
by a CCITT filter characteristic. In the E5500 system, software revision 5.0 or higher, use
of this filter response can be achieved by simply checking the
the
If you use the older HP/Agilent 3048 phase noise system, there is no direct way to measure
residual FM, but because residual FM and phase noise are so closely related and
interdependent, it is safe to assume that if the DUT’s phase noise meets specification,
looks typical (between 300 and 3000 Hz), and shows no unexpected large spikes in the
response, then it will most likely also meet the residual FM specification.
Snu(f) or Spectral density of frequency fluctuations.
This check verifies the functionality of the dual arbitrary waveform generator. A
triangular waveform is downloaded into the signal generator and then output to the
rear-panel I and Q outputs. An oscilloscope is used to monitor the rear-panel I and Q
outputs.
NOTEThis is not a performance test. This check is only provided to ensure that the
dual arbitrary waveform generator is operational. The results are verified
visually on the oscilloscope, and there is no data that is automatically logged
or that can be printed out.