This chapter contains a list of all the recommended equipment required to perform the
performance tests and adjustments for the ESG Family Signal Generators.
Calibration Guide1-1
Equipment RequiredESG Family Signal Generators
Required Equipment
• HP/Agilent 8563E Option E35
Spectrum Analyzer (No Substitution)
Required Equipment
• HP/Agilent 8902A Measuring
Receiver
1
Option 010
Capable of W-CDMA ACP
measurements from 1.2 to 3.0 GHz
Option E35 Comprises:
FM accuracy: ±2% of reading ±1 digit
AM accuracy: ±2% of reading ±1 digit
Range: 250 kHz to 1300 MHz
Filters: 300 Hz high pass and
15 kHz and 3 kHz low pass
Detectors: Peak+
• HP/Agilent 54750A Digitizing
Oscilloscope
Bandwidth: 20 GHz
• HP/Agilent 54751/2A Oscilloscope
Plug-in
Two 20 GHz bandwidth channels
• HP/Agilent 54610B Digital
Oscilloscope
1
Bandwidth: 500 MHz
• HP/Agilent 5071A
Primary Frequency Standard
• HP/Agilent 8563E Option H35
Spectrum Analyzer
• HP/Agilent 8563E Option K35
Adjacent Channel Power Ratio Test
Set for W-CDMA
NOTEHP/Agilent 8563E Option K37
combines the Option K35 test set
with firmware ROM chips to
upgrade existing HP/Agilent 8560
E-series analyzers.
• HP/Agilent 8563E Options 001, 007
Spectrum Analyzer
1
Frequency range: 100 kHz to 4000 MHz
Relative level accuracy (harmonic and
spurious measurements) ±2 dB
Displayed average noise level (1 Hz BW)
−110 dBm at 100 kHz,
−130 dBm at 1 to 10 MHz, and
−134 dBm at 0.01 to 4 GHz
Frequency: 10 MHz
Stability: > 1×10
-10
/year
• HP/Agilent 8903B Options 051, 010
Audio Analyzer
1
Distortion accuracy: ±1 dB
Residual distortion:
−80 dB at 80 kHz BW
30 kHz low pass filter
AC level accuracy: ±4%
• HP/Agilent 89441A Options AY7,
AYA, AY9, UFG or UTH, and AYH
Vector Signal Analyzer
1,2
1 Hz digital resolution BW filter
(required for power accuracy)
RF input VSWR: 1:1.5
External timebase input
Second IF output
• HP/Agilent 8663A Signal Generator
Frequency Range: 11.5 to 2500 MHz
Output level: +16 dBm
Meets HP/Agilent 8663A
single-sideband phase noise
specifications
(No Substitution)
1.Equipped with GPIB
2.Used only with ESG-D models
1-2Calibration Guide
1
ESG Family Signal GeneratorsEquipment Required
Required Equipment
• HP/Agilent 8904A Function
Generator
1
Frequency range (sinewave):
0 Hz to 600 kHz
AC Amplitude (sinewave only):
Range: 0 to 10 Vp-p into a 50Ω load
Flatness (>630 mVp-p into 50Ω):
±0.1% (±0.009 dB), 0.1 Hz to 100 kHz
• HP/Agilent 53132A Option 050
Universal Counter
1
Frequency range: 4 GHz
Resolution: 0.1 Hz
External timebase input
• HP/Agilent 33120A Option 001
Arbitrary Waveform Generator
1
(2)
(No Substitution)
• HP/Agilent 438A Power Meter
1
Instrumentation accuracy: ±0.5%
Power reference accuracy: 0.9%
• HP/Agilent 8482A Power Sensor
Frequency range: 100 kHz to 4000 MHz
Power range: −30 dBm to +13 dBm
Maximum SWR:
100 kHz to 300 kHz: 1:1.6,
300 kHz to 1 MHz: 1:1.12,
1 MHz to 2 GHz: 1:1.1,
2 GHz to 4 GHz: 1:1.3
Calibration factor accuracy (RSS):
≤ 1.6%
• HP/Agilent E3615A Power Supply
Minimum Power: 15V
Minimum Current: 325 mA
• MD/MDC-174 Mixer
Frequency range: 0.001 to 2.8 GHz
• HP/Agilent 3458A Digital
Multimeter
1
dc Volts: 0.1 to 100 V
Input resistance: 10Ω to 1 KΩ
• HP/Agilent 8491A/B Option 006
Attenuator
Attenuation: 6 dB
• HP/Agilent 8491A/B Option 010
Attenuator
Attenuation: 10 dB
• HP/Agilent 8491A/B Option 020
Attenuator
Attenuation: 20 dB
• Personal Computer (included in
HP/Agilent E5500A/B)
Running MS Windows® 3.1/95/98 or
Windows NT®
• HP LaserJet 5L Printer
MS Windows supported
• HP 82341C GPIB Interface Card
with SICL HP I/O Library Installed
(No Substitution)
• Agilent Technologies E4400-60073
Digital Test Card (No Substitution)
• HP/Agilent 8495G Programmable
Step Attenuator
Attenuation: 0 to 70 dB
Frequency range: dc to 4 GHz
• HP/Agilent 11713A Step Attenuator
Driver
• HP/Agilent E5500A/B Phase Noise
Measurement System
2, 3
Frequency range: 0 to 4 GHz
• MD/MDC-164 Mixer
Frequency range: 0.5 to 9 GHz
Calibration Guide1-3
1.Equipped with GPIB
2.Includes Personal Computer
3.Required for ESG-AP/DP only
Equipment RequiredESG Family Signal Generators
Required Equipment
Low Noise Amplifiers for
Power Level Accuracy Adjustment and Performance Test
The power level accuracy adjustment and performance test both use low noise amplifiers.
They are required to maintain separation between a low-level signal and the signal
analyzer noise floor. The following amplifiers meet the requirements of making this
measurement.
A +15 Vdc external supply is required for each of the amplifiers. An HP/Agilent E3615A,
0 to 20 V power supply can power all recommended amplifiers simultaneously.
100 kHz to 10 MHz Low Noise Amplifier
Critical Specifications
Frequency Range: 100 kHz to 10 MHz
Noise Figure: 1.5 dB
Gain: 50 dB
1 dB Compression Minimum: +5 dBm
External Power Supply Required: +15 Vdc @ 25 mA
Possible Part Number and Supplier
Part Number: W50ATC
Supplier: Advanced Control Components
20 Meridian Road
Eatontown, NJ 07724
30 MHz to 2 GHz Low Noise Amplifier
Configuration
Two amplifiers are cascaded to obtain the necessary requirements. The specifications
listed below are the combined specifications of the cascaded amplifiers.
Critical Specifications
Frequency Range: 10 MHz to 2 GHz
Noise Figure: 1.8 dB
Gain: > 56 dB
1 dB Compression Minimum: +10 dBm
External Power Supply Required: +15 Vdc @ 300 mA
Possible Part Number and Supplier
Part Number: AFS3-00010300-18-5P-4 (2 required)
Supplier: MITEQ Inc.
100 Davids Drive
Hauppauge, NY 11788
1-4Calibration Guide
ESG Family Signal Generators
2Operation Verification
This chapter provides procedures that will either ensure that the signal generator is
operating correctly or will help to point to problem areas if it is not. Operation verification
does not ensure performance to specifications, but should provide a level of confidence that
the signal generator is operating correctly within a minimum amount of time. Operation
verification is appropriate for incoming inspection, after repair when a full calibrated
performance is not required, or whenever the integrity of the signal generator is in
question.
Calibration Guide2-1
Operation VerificationESG Family Signal Generators
Verification Procedures
Verification Procedures
Performthe following procedures in the order they are presented. The tables referenced by
the tests are located in the back of the chapter where they can be copied easily.
This procedure verifies that the signal generator powers up and that the internal
instrument check identifies no errors. The internal check evaluates the correctness of
operation and returns an error message if a problem is detected.
1. Turn power on to the signal generator by pressing the power switch. The green LED
will light. Let the signal generator warm up for one hour.
NOTEFor ESG-AP or ESG-DP signal generators, or those with Option 1E5, ERROR
514, Reference Oven Cold occurs whenever the signal generator is first
connected to AC line power. The
annunciator both turn on. The OVEN COLD annunciator automatically clears
after approximately 5 minutes. The error queue cannot be cleared, however,
until the OVEN COLD annunciator has turned off.
OVEN COLD annunciator and the ERR
2. Cycle the power to the signal generator. The green LED should again be lit and the
signal generator will perform a check.
2-2Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
2. Check for Error Messages
1. When the display is lit, check to see if the ERR annunciator is turned on.
2. If the
Utility > Error Info > View Next Error Message. The first error message in the queue will be
ERR annunciator is turned on, review the error messages in the queue by pressing
shown in the text area of the display. Refer to the error messages guide for information
about the error message.
If there is more than one error message (each message will be designated as 1 of n),
continue pressing the
View Next Error Message softkey until you have seen all of the
messages.
3. When you have resolved all of the error messages, press Clear Error Queue(s) to delete the
messages. Cycle the power on the signal generator and then restart this procedure
(Check for Error Messages) at step 1.
3. Frequency Range and Accuracy Check
Frequency range is tested by determining the frequency accuracy relative to the timebase
at the frequency limits.
Connect the Test Equipment
Figure 2-1 Frequency Range and Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. For frequencies < 10 MHz:
a. Press
b. Press RF On/Off. The RF ON annunciator is displayed.
Calibration Guide2-3
Mod On/Off. The MOD OFF annunciator is displayed.
Amplitude. Enter 10 using the numeric keypad and press the dBm terminator
softkey.
Operation VerificationESG Family Signal Generators
Verification Procedures
4. For frequencies ≥ 10 MHz:
a. Press
Amplitude. Enter 0 using the numeric keypad and press the dBm terminator
softkey.
b. Check that the RF ON annunciator is displayed. If not, press RF On/Off.
Configure the Frequency Counter
1. For frequencies < 10 MHz use Input 2. Press 1MΩ.
2. For frequencies≥ 10 MHz and ≤ 500 MHz use Input 2. Press
3. For frequencies > 500 MHz use Input 1. Press
AUTO.
50Ω.
NOTESet the gate time of > 5 seconds for maximum counter accuracy. Verify that
the counter is phase-locked to the 10 MHz external reference.
Measure the Frequency Accuracy
1. Set the signal generator to the frequencies shown in Table 2-1, “Frequency Accuracy,”
on page 2-10 (to the maximum frequency of your signal generator).
2. Record the measured frequency in Table 2-1 and compare it to the corresponding limits.
4. Power Level Accuracy Check
Performing this check will provide a high level of confidence that the signal generator’s
power level circuitry is functioning correctly. This check does not test the signal generator
to warranted specifications. Test points have been reduced and the limits are degraded in
order to minimize measurement time and take into account a broad range of measurement
uncertainties. To have the signal generator tested to warranted specifications, a complete
power level accuracy performance test is required. If the complete performance test is
needed and you are unable to perform it, contact your nearest Agilent Technologies service
center for information concerning calibration.
Connect the Test Equipment for Setup 1
Figure 2-2 Power Level Accuracy Equipment Setup 1
2-4Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
3. Press Amplitude. Enter 13 using the numeric keypad and press the dBm terminator
softkey.
4. Press Mod On/Off. The MOD OFF annunciator is displayed.
5. Press
RF On/Off. The RF ON annunciator is displayed.
Configure the Power Meter
1. Zero and calibrate the power meter to the sensor’s reference calibration factor.
2. Set the power meter to measure absolute log power (dBm).
Measure the Power Level Accuracy
1. Set the power levels and frequencies for the signal generator (to the maximum
frequency of your signal generator). Refer to the appropriate table in the following list:
• Table 2-3, “Power Level Accuracy Setup 1 (Signal Generators with Option 1E6),” on
page 2-13.
2. Set the power sensor’s calibration factor for each frequency.
3. Record the measured power level in the appropriate table (Table 2-2 or Table 2-3) and
compare it to the corresponding limits.
Connect the Test Equipment for Setup 2
Figure 2-3 Power Level Accuracy Equipment Setup 2
Calibration Guide2-5
Operation VerificationESG Family Signal Generators
Verification Procedures
Configure the Signal Generator
1. Press Frequency. Enter 277 using the numeric keypad and press the kHz terminator
softkey.
2. Press Amplitude. Enter −15 using the numeric keypad and press the dBm terminator
softkey.
Configure the Spectrum Analyzer
1. Preset the spectrum analyzer.
2. Set the analyzer to external 10 MHz reference.
3. Set the center frequency to 277 kHz.
4. Set the frequency span to 100 Hz. (This will result ina1Hzresolution bandwidth with
a digital filter.)
Measure the Power Level Accuracy
1. Connect the signal generator’s RF OUTPUT through the 6 dB attenuator to the
spectrum analyzer’s RF input.
2. Transfer the power level results for each frequency at −15 dBm from the Measured
Power column in Table 2-2 or Table 2-3 as appropriate, to the Power Meter Reading for
−15 dBm column in Table 2-4, “Power Level Accuracy Setup 2 (All Signal Generators),”
on page 2-15.
3. On the spectrum analyzer, select
Marker Normal Mode and then select the Peak Search
function. This activates the marker and sets it to the signal peak.
4. On the spectrum analyzer, ensure that the marker is at the signal peak and use the
MKR-> menu to set the marker to the reference level. If necessary, select Peak Search to
ensure that the marker is at the signal peak.
5. With the marker at signal peak, select the Marker Delta function. This will set the marker
to measure relative amplitude from a reference of 0 dB. If the marker does not read
0 dB, press
Marker Normal > Peak Search > Marker Delta until the marker reads 0 dB.
6. Decrease the signal generator amplitude in 10 dB steps as shown in Table 2-4. At each
10 dB step, select Peak Search to ensure that the marker is at the signal peak.
7. Measure the power levels shown in the Power Level Setting (dBm) column of Table 2-4
for the current frequency, and record the values in the Spectrum Analyzer Marker (dB)
column.
8. Return the spectrum analyzer reference level and the signal generator amplitude to
−15 dBm. Set the signal generator frequency and the spectrum analyzer’s center
frequency to the next frequency listed in Table 2-4 and repeat the process from step 3.
Continue until all of the frequencies have been measured and recorded (to the
maximum frequency of your signal generator).
2-6Calibration Guide
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Calculate the Actual Power Level
1. Calculate and record the Actual Power Level, in Table 2-4, as the sum of the Power
Meter Reading for −15 dBm and the Spectrum Analyzer Marker (dB) value.
For example:
• Power Meter Reading for −15 dBm at 2.516 MHz = −14.95 dBm
• Spectrum Analyzer Marker (dB) at 2.516 MHz and Power Level Setting at
−85 dBm = −70.17 dB
• Actual Power Level at 2.516 MHz and −85 dBm: (−14.95) + (−70.17) = −85.12 dBm
2. Compare the calculated Actual Power Level value to the corresponding limits.
5. FM Accuracy Check
Connect the Test Equipment
Figure 2-4 FM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. Press
4. Press
5. Press
softkey.
6. Check that the MOD ON annunciator is displayed. If it is not, press Mod On/Off.
7. Press
Calibration Guide2-7
FM/ΦM > FM Off On. The FM annunciator is displayed.
FM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
FM Dev. Enter 100 using the numeric keypad and press the kHz terminator softkey.
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator
RF On/Off. The RF ON annunciator is displayed.
Operation VerificationESG Family Signal Generators
Verification Procedures
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies shown in Table 2-5, “FM Accuracy,” on
page 2-19.
2. Record the deviations measured and compare them to the limits shown in Table 2-5.
6. AM Accuracy Check
Connect the Test Equipment
Figure 2-5 AM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
3. Press
4. Press
softkey.
5. Check that the MOD ON annunciator is displayed. If it is not, press Mod On/Off.
AM > AM Off On. The AM annunciator is displayed.
AM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator
6. Press
2-8Calibration Guide
RF On/Off. The RF ON annunciator is displayed.
ESG Family Signal GeneratorsOperation Verification
Verification Procedures
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to AM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies and depths shown in Table 2-6, “AM
Accuracy,” on page 2-19.
2. Record the AM depths measured and compare them to the limits shown in Table 2-6.
7. I/Q Modulation Check (ESG-D Only)
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
I/Q > I/Q Calibration. Check that Full is highlighted on the Calibration Type User Full
softkey. If not, press Calibration Type User Full.
3. Press
range. The message
Execute Cal to begin an I/Q calibration for the signal generator’s full frequency
I/Q Calibration in Progress is displayed until the calibration is
complete.
Calibration Guide2-9
Operation VerificationESG Family Signal Generators
Verification Procedures