The information contained in this document is subject to change
without notice.
Agilent Technologies makesno warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Warranty
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of three years from
date of shipment. During the warranty period, Agilent Technologies
Company will, at its option, either repair or replace products that prove
to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies
shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware
designated by Agilent Technologies for use with an instrument will
execute its programming instructions when properly installed on that
instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or
error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse, operation
outside of the environmental specifications for the product, or improper
site preparation or maintenance.
ii
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT
BE LIABLE FOR ANY DIRECT,INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON
CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
iii
Safety Information
The following safety notes are used throughout this manual.
Familiarize yourself with these notes before operating this instrument.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
WARNINGThis is a Safety Class 1 Product (provided with a protective
earth ground incorporated in the power cord). The mains plug
shall be inserted only in a socket outlet provided with a
protected earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
WARNINGNo operator serviceable parts inside. Refer servicing to
qualified personnel. To prevent electrical shock do not remove
covers.
CAUTIONAlways use the three-prong AC power cord supplied with this product.
Failure to ensure adequate grounding may cause product damage.
This chapter introduces you to basics of GSM technology and the
general functionality of the ESA with the GSM measurement
personality installed. In addition, sources for additional information on
digital communications are listed.
1-1
Understanding GSM
GSM Standards
GSM Standards
The Global System for Mobile communication (GSM) digital
communications standard defines a voice and data over-air interface
between a mobile radio and the system infrastructure. This standard
was designed as the basis for a radio communications system. A base
station control center (BSC) is linked to multiple base transceiver
station (BTS) sites which provide the required coverage.
GSM 900, DCS 1800, and PCS 1900 are GSM-defined frequency bands.
The term GSM 900 is used for any GSM system operating in the
900 MHz band, which includes P-GSM, E-GSM, and R-GSM. Primary
(or standard) GSM 900 band (P-GSM) is the original GSM band.
Extended GSM 900 band (E-GSM) includesall theP-GSM band plus an
additional 50 channels. Railway GSM 900 band (R-GSM) includes all
the E-GSM band plus additional channels. DCS 1800 is an adaptation
of GSM 900, created to allow for smaller cell sizes for higher system
capacity. PCS 1900 is intended to be identical to DCS 1800 except for
frequency allocation and power levels. The term GSM 1800 is
sometimes used for DCS 1800, and the term GSM 1900 is sometimes
used for PCS 1900. For specifics on the bands, refer to Table 1-1.
The GSM digital communications standard employs an 8:1 Time
Division Multiple Access (TDMA) allowing eight channels to use one
carrier frequency simultaneously.The 270.833 kbits/second raw bit rate
is modulated on the RF carrier using Gaussian Minimum Shift
Keying (GMSK).
The standard includes multiple traffic channels, a control channel, and
a cell broadcast channel. The GSM specification defines a channel
spacing of 200 kHz.
The GSM framing structure is based on a hierarchical system
consisting of timeslots, TDMA frames, multiframes, superframes, and
hyperframes. One timeslot consists of 156.25 (157) bit periods including
tail, training sequence, encryption, guard time, and data bits. Eight of
these timeslots make up one TDMA frame. Either 26 or 51 TDMA
frames make up one multiframe. Frames 13 and 26 in the 26 frame
multiframe are dedicated to control channel signaling.
These principles of the GSM systems lead to the need for the
fundamental transmitter measurements such as:
• Power vs Time - which verifies that the transmitter output power
has the correct amplitude, shape, and timing for the TDMA
sequence.
• Phase and Frequency Error - which verifies the accuracy of the
transmitters 0.3 GMSK modulation process.
• Output RF Spectrum - which verifies that the RF carrier is
contained within the designated 200 kHz channel.
Mobile Stations And Base Transceiver Stations
The cellular system includes the following:
• base transceiver stations, referred to as BTS
(frequency ranges dependent on the standard; refer to Table 1-1)
• mobile stations, referred to as MS
(frequency ranges dependent on the standard; refer to Table 1-1)
Uplink And Downlink
Uplink is defined as the path from the mobile station to the base
transceiver station. Downlink is the path from the base transceiver
station to the mobile station.
What Is An ARFCN?
An ARFCN is the Absolute Radio Frequency Channel Number used in
the GSM system. Each RF channel is shared by up to eight mobile
stations using Time Division Multiple Access (TDMA). The ARFCN is
an integer (in a range dependent on the chosen standard, refer to
Table 1-1) which designates the carrier frequency.
1-4Chapter1
Understanding GSM
GSM Standards
What Is A Timeslot?
GSM utilizes Time Division Multiple Access (TDMA) with eight time
slots per RF channel which allows eight users to use a single carrier
frequency simultaneously. Users avoid one another by transmitting in
series. The eight users can transmit once every 4.62 ms for 1 timeslot
which is 577 µs long.The eight user timeslots are numbered from 0 to 7.
Typically, each 577 µs timeslot has a length of 156.25 bit periods, which
consists of 148 data bits and 8.25 guard bits. The 4.62 ms required to
cycle through eight timeslots is called a frame. In a TDMA system, the
shape of each transmitted burst must be controlled carefully to avoid
over-lapping bursts in time.
Chapter 11-5
Understanding GSM
What Does The GSM Measurement Personality And Hardware Do?
What Does The GSM Measurement Personality
And Hardware Do?
When configured with a GSM mode personality module, the ESA series
spectrum analyzer can help determine if a GSM transmitter or receiver
is working correctly. The following documents detail GSM standards
used when testing:
Table 1-2Documents detailing the GSM standards used when testing
GSM-900DCS-1800PCS-1900
Mobile stationETS 300 607-1
(11.10)
Base stationETS 300 609-1
(11.21)
Radiotransmission
and reception
NOTEThese standards are continually being updated by the communications
ETS 300 910
(5.05)
ETS 300 607-1
(11.10)
ETS 300 609-1
(11.21)
ETS 300 910
(5.05)
J-STD-007
J-STD-007
J-STD-007
industry. The software performs to the standard/date shown under
Mode Setup, Radio.
When you select the standard format for use, the instrument
automatically makes measurements using the measurement methods
and limits defined in the standards. Detailed measurement results are
displayed allowing you to analyze GSM system performance.
Measurement parameters may be altered for specialized analysis. This
is done within narrow limits because the standards specify the test
methods and settings.
For infrastructure testing, the instrument tests base station
transmitters and receivers in a non-interfering manner by using a
coupler or power splitter.
1-6Chapter1
Understanding GSM
What Does The GSM Measurement Personality And Hardware Do?
The instrument makes the following GSM one-button measurements
(measurements activated by a single key press):
• Cable Fault Location (available in SA mode)
• Monitor Band/Channel
• Out of Band Spurious
• Output RF Spectrum
• Phase and Frequency Error
• Power Steps
• Power vs Time
• Receive (Rx) Band Spurious
• Transmit (Tx) Band Spurious
• Transmitter Power
In addition, the following manual measurements (measurements which
are not activated by a single key press) can be performed:
• Loss/Gain (available in SA mode)
• Return Loss (VSWR) (available in GSM mode)
Chapter 11-7
Understanding GSM
Other Sources Of Measurement Information.
Other Sources Of Measurement Information.
Application Notes
The following application notes describe digital communications
measurements in much greater detail than discussed in this User
Guide:
• Application Note 1298
Digital Modulation in Communications Systems - An Introduction
Agilent Technologies part number 5965-7160E
• Application Note 1311
Understanding GSM Transmitter Measurements for Base
Transceiver stations and Mobile Stations
part number 5966-2833E
• Application Note 1313
Testing and Troubleshooting Digital RF Communications
Transmitter Designs
part number 5968-3578E
• Application Note 1314
Testing and Troubleshooting Digital RF Communications Receiver
Designs
part number 5968-3579E
• Application Note 150
Spectrum Analyzer Basics
part number 5952-0292
Additional Information
• Additional measurement application information is available from
your local Agilent Technologies sales and service office.
• The latest information on ESA spectrum analyzers is available from
the following URL:
http://www.agilent.com/find/esa/
1-8Chapter1
2Getting Started
This chapter introduces you to basic features of the instrument,
including the front panel keys, rear panel connections, and display
annotation. Equipment required for GSM measurements, available
documentation, and processes for installing and uninstalling
applications are also described.
2-1
Getting Started
Instrument Overview
Instrument Overview
This section provides information on only GSM mode features. For
those features not described here, refer to the Agilent ESA Spectrum
Analyzers User’s Guide.
Front-Panel Features
For further information on the features mentioned in the following
section, refer to Chapter 5 of this document.
Figure 2-1Front-Panel Feature Overview
1Mode keysThese keys allow you to select the measurement mode and mode
parameters such as input and trigger settings.
• MODE accesses menu keys to select the instrument mode. Each mode is
independent of all other modes.
• Mode Setup accesses menu keys that allow you to configure the parameters
specific to the current mode and affect all measurements within that mode.
2-2Chapter2
Rear-Panel Features
This section provides information on GSM rear panel features only. For
those features not described here, refer to the ESA-E Series SpectrumAnalyzers User’s Guide.
Figure 2-2Rear-Panel Feature Overview
Getting Started
Instrument Overview
1DSP and Fast
ADC
DSP and Fast ADC (Option B7D) provides digital signal processing
and fast ADC required for many of the digital demodulation
measurements in the GSM and cdmaOne measurement
personalities.It must beordered with Option B7E and Option 1D5.
2RF Comms
Hardware
RF Communications Hardware (Option B7E) provides the RF down
convertor hardware required for digital demodulation
measurements. It must be ordered with Option B7D and
Option 1D5
3Ext Ref InAccepts an external 1 MHz to 30 MHz reference frequency source.
410 MHz REF INAccepts an external frequency source to provide the 10 MHz, −15 to +10
dBm frequency reference used by the analyzer.
Chapter 22-3
.
Getting Started
Instrument Overview
510 MHz OutProvides a 10 MHz, 0 dBm minimum, timebase reference signal phase
locked to the Ext Ref In.
610 MHz REF
OUT
7Ext Frame SyncAccepts an external 0 to 5 V TTL trigger.
Provides a 10 MHz, 0 dBm minimum, timebase reference signal.
Display Annotation
This section describes GSM screen annotations which differ from SA
mode annotations. Refer to the ESA-ESeries Spectrum AnalyzersUser’sGuide for information on screen annotations not described here.
The following screen provides an example of a screen annotation:
Figure 2-3Screen Annotation: Example
NOTEAnnotations are not always identical for different measurementscreens
and the annotations displayed depend on what is required to display
critical measurement parameters. To view examples of the
measurement displays, refer to the ESA-E Series Spectrum AnalyzersGSM Measurement Personality Measurement Guide.
2-4Chapter2
Getting Started
Instrument Overview
1
2
3Current MeasurementDisplays the name of the current measurement.
4
5
6
7
8
9
10
Active Function Area
Measurement Bar
Base/Mobile
ARFCN
Base/Standard
TSC
Averaging Indicator
Trig Source
Burst Sync
Displays parameter entries. If you press a key which
activates a function, the parameter descriptionandvalue will
appear in the active function area.
Displays information about measurements including some
mode setup parameters.
Indicates which digital communications system platform you
are testing, base station systems or a mobile unit.
Active Radio Frequency Channel Number.
Displays the radio band currently in use.
Training sequence code. Displays the training sequence code
used on the assigned channel.
Displays the number of the current average (averaging
progress) when averaging is turned on.
Displays the triggering mode selected for the current
measurement. Possible values are External, Free, RF Burst
and Frame Timer.
Displays the type of burst sync selected for the current
measurement. Possible values are Training Seq, RF Amptd,
and None.
11
Pass/Fail Status
Indicates the status of the measurement limit pass/fail test
when one or more limit test function is turned on.
Chapter 22-5
Getting Started
Options Required
Options Required
Installing the GSM measurement personality firmware and making the
associated measurements require certain basic equipment. This section
lists GSM compatible Agilent ESA Spectrum Analyzers and required
hardware options.
Compatible spectrum analyzers
The GSM measurement personality is not compatible with all ESA
spectrum analyzer models. Table 2-1 lists the models that are
compatible and the upper frequency range of each.
Table 2-1GSM Compatible Agilent ESA Spectrum Analyzers
Additional hardware options must be installed in the spectrum
analyzer before GSM measurements can be made. Table 2-2 lists the
hardware options available and the measurements that they enable.
For optimum performance of cdmaOne measurements, Option B74
should be installed in your Agilent ESA-E Series Spectrum Analyzer.
Not all of the options can be installed by the user. Some of the options
require that the instrument be returned to the factory or an Agilent
Technologies service center. In addition, some of the options require
Performance Verification and Adjustments to be performed after
installation. Refer to Table 2-2 for option specific information.
NOTEWhen transporting the instrument, use the original packaging or
comparable packaging. If the shipping container is damaged, any part
is missing, or you do not have an appropriate shipping container,notify
Agilent Technologies at one of the addresses shown on Table 6-1 on
page 6-6.
2-6Chapter2
Table 2-2 GSM Hardware Options and measurements
Getting Started
Options Required
Required/recommended
option
Option
Number
Measurement
GSM Measurement PersonalityBAHRequired for all measurements.
Memory extensionB72
DSP and Fast ADC
B7D
a
Recommended options which enable
Phase and Frequency Error and full
RF Communications Hardware
B7E
a
triggering for all measurements.
Note: these options must be ordered
together.
50 Ohm Tracking Generator
Time Gated Spectrum Analysis
1DN
1D6
b
b
Required for cable fault measurement.
Required for Output RF Spectrum
Modulation Swept measurement. The
measurements will execute without the
option but not to specification and the
message “Gate option not installed.
Results may not be accurate” displayed.
Fast Time Domain SweepAYXRequired for Output RF Spectrum,
Transmitter Power, and Power vs Time
measurements if options B7D/B7E arenot
ordered.
High Stability Frequency
1D5
a
Improves frequency reference accuracy.
Reference
RF and Digital Communication
Hardware Option bundle
Option B74
Includes the
following
options:
a
Includes necessary hardwarefor the GSM
measurement personality. The Cable
FaultLocation measurement also requires
option 1DN.
1D6
B72
1D5
B7D
B7E
1DS
1DR
a. Service center or factory installation, calibration required.
b. Factory installation only.
NOTEIf the appropriate hardware is not present, the measurement softkey
under the
Measure menu will be grayed out and that measurement will
not be available.
Chapter 22-7
Getting Started
GSM Documentation For The ESA-E Series Spectrum Analyzer
GSM Documentation For The ESA-E Series
Spectrum Analyzer
Spectrum analyzers with GSM Installed
When you purchase your instrument already loaded with the GSM
measurement personality, you receive the following materials:
Table 2-3Personality Documentation
Part NumberPart DescriptionNotes
Refer to the ESA
WEB site for the
current part
number.
Spectrum analyzers without GSM Installed
If your instrument is ordered without measurement personalties
installed, you can order the GSM measurement personality as Option
B74 instrument upgrade package. With the purchase of a personality
option upgrade, you receive the documents listed in Table 2-3 and the
document shown in Table 2-4.
User’s Guide
ESA-E Series Spectrum Analyzers
GSM Measurement Personality
ESA-E Series Spectrum Analyzers
GSM Measurement Personality
Programming Commands
ESA-E Series Spectrum Analyzers
GSM Measurement Personality
Agilent ESA Series Spectrum
Analyzers Specifications Guide
Agilent ESA Series Spectrum
Analyzers Documentation and
Instrument Driver CD-ROM
BAH Option manuals
BAH Option manuals
BAH Option manuals
Includes specifications for
all optional measurement
personalities
Does not include service
documentation or software
2-8Chapter2
GSM Documentation For The ESA-E Series Spectrum Analyzer
Table 2-4Personality Documentation
Part NumberPart Description Notes
Getting Started
Refer to the ESA
WEB site for the
current part
number.
Understanding Digital Communications
Measurements
Additional measurement application information is available through
your local Agilent Technologies sales and service office. Refer to “Other
Sources Of Measurement Information.” on page 1-8 for applicable
application notes.
Agilent ESA Spectrum Analyzers Update
For the latest information about this instrument, including firmware
upgrades, application information, and product information, please
visit the URL listed below
Updating the Firmware
Updated versions ofthe ESA-E Series Spectrum Analyzerfirmware will
be available via several sources. Information on the latest firmware
revision can be accessed from the following URL:
RF and Digital Communications
Hardware Installation Note
B74 Option manual
http://www.agilent.com/find/esa/
Chapter 22-9
Getting Started
Installing Optional Measurement Personalities
Installing Optional Measurement
Personalities
Active License Key
The measurement personality software you have purchased with your
instrument has been installed and the license key has been enabled at
the factory. With any future purchase of a new personality software,
you will receive a certificate that displays the unique license key
number. The license key enables you to install, or reinstall, any
measurement personality you have purchased.
You will want to keep a copy of your license key number in a secure
location. Please enter your license key numbers in the box provided
below for future reference. If you should lose your license key number,
get in touch with your local Agilent Technologies service or sales office
for assistance. For the location of these offices, refer to Table 6-1 on
page 6-6.
Active License Key Numbers for Instrument with Serial # ________
For Option______________ the license number is ________________________
For Option______________ the license number is ________________________
For Option______________ the license number is ________________________
You will need to use a license key number only under the following
conditions:
• If you purchase an additional measurement software package.
• If the controller board is repaired or replaced.
2-10Chapter2
Getting Started
Installing Optional Measurement Personalities
Installing the Licensing Key
If you are installing a new option, follow these steps to install the
unique license key number for the measurement personality software
that you want to install in your instrument:
1. Press
When you press
System, More, Licensing, Option.
Option the alpha editor will be activated. For
instructions on using the alpha editor, refer to the Agilent ESA
Spectrum Analyzers User’s Guide.
2. Use the alpha editor to enter the three letter designation for the
software option that you wish to install in the instrument.
3. Press
4. Press
When you press
Done on the alpha editor menu.
License Key.
Licensing Key the alpha editor will be activated. For
instructions on using the alpha editor, refer to the Agilent ESA
Spectrum Analyzers User’s Guide.
5. Use the alpha editor to enter the 12 character licensing key number
for the software option that you wish to install in the instrument.
6. Press
7. Press
Done on the alpha editor menu.
Activate to turn on the licensing key. You may now install the
measurement personality option software.
Chapter 22-11
Getting Started
Installing Optional Measurement Personalities
Using Install Key
You may want to install a software revision, install new measurement
software or reinstall measurement software that you have previously
uninstalled, or uninstall measurement software. Before you can install
software, you will need a set of installation diskettes.
If you have ordered a measurement personality, you will receive the
installation disk set in the option upgrade package. If you are updating
an existing, previously installed measurement option, you may order
the diskettes from Agilent Technologies or create a set from the Agilent
internet site shown in “Updating the Firmware” on page 2-9. When you
order the updated software disk set, you will need to order Option UE2.
(Option UE2 is a firmware update and is needed to ensure that the
firmware and the software are compatible.) A set of diskettes can be
ordered from your local Agilent Technologies service or sales office.
Refer to Table 6-1 on page 6-6 for the location of these offices. To create
a disk set refer to “Creating Software Installation Disks” below.
Creating Software Installation Disks
To create the installation disks on-line, visit the Agilent internet site
shown in “Updating the Firmware” on page 2-9. Follow the instructions
provided on the internet site for downloading the current measurement
personalty software and creating the installation disk set. The
instructions for creating the disk set will step you through the process
to create a firmware disk set when you create the measurement
personalty software disk set. (A firmware update may be needed to
ensure that the firmware and the software are compatible.) After you
have created the disk set, follow the on-line instructions to install the
firmware. After successfully installing the firmware update, proceed
with the following instructions for installing the measurement
personalty software in your instrument.
Installing Personality/Software Options
This procedure gives steps to install a new software option in an ESA-E
Series Spectrum Analyzer using the internal floppy drive of the
instrument. Screen messages display the update progress and give
directions. The instrument will not need to be re-calibrated after this
procedure since no changes are made to calibration or adjustment files.
If you have a problem with the installation process, refer to
“Troubleshooting the Installer” on page 2-14.
NOTEWhen the installer starts up, it examines the instrument to ensure that
all the required software and hardware options are present. If they are
not, the installer will generate and error and you will not be able to
install the personality.
1. If this is the installation of new personality option software, you
2-12Chapter2
Getting Started
Installing Optional Measurement Personalities
must enter the License Key for the new option. For instructions on
entering the License Key, refer to the “Installing the Licensing Key”
on page 2-11.
When you have completed entering the license key number, continue
with the next step.
2. Insert disk one of the installation disk set into the disk drive located
on the right side of the ESA front panel.
3. Press
System, More, Personalities, and Install. The instrument will
then load the installer off of the floppy drive. If there is no floppy in
the drive, the incorrect diskis inserted,or thereis no installer on the
disk, the error “No install disk present in disk drive” will be shown.
Once the instrument has loaded the installer, the screen will change
to the installer screen and the
Install Pers. menu will be shown. For
more information on the installer screen and menu, refer to
“Installer Screen and Menu” on page 2-15.
4. When the installer first starts up, it will show a popup message.
Select
NOTEOnce the installer has begun installing a personality, any error will
Verify Disks.
cause the whole personality (including a previously installed version) to
be removed from the instrument. Because of this, it is very important
that you verify the disks prior to installing them. If any of the disks or
files are bad, you will not be able to use the personality until you obtain
a new installation disk set and run the install using them.
5. When prompted, insert the next disk and press
When Verify Disks is running, the
Install Now and Exit Install keys
Verify Disks again.
will be grayed out.
6. When the verification is complete, press Install Now and the
installation of the personality will begin. Some of the disks may take
only a short time to load or be skipped entirely, while others can take
up to about 30 minutes to load.
When installer is running, the
Verify Disks and Exit Install keys will
be grayed out.
7. When prompted, insert the next disk and press Install Now again.
8. Once the installation is complete, press
Chapter 22-13
Exit Install.
Getting Started
Installing Optional Measurement Personalities
Troubleshooting the Installer
If the installation process stalls or fails in another way, follow these
steps to resolve your problem.
1. If the instrument stops the update process before all the disks are
loaded proceed as follows:
a. Press Exit Install to abort the process.
b. Return to step 2 under “Installing Personality/Software Options”
and start the installation process again.
2. If the instrument fails after repeating the installation procedure, get
in touch with your nearest Agilent sales and service office listed in
Table 6-1 on page 6-6 for assistance. Please provide the following
information:
Model Number:
Serial Number:
State that you are having trouble installing a software option
update.
2-14Chapter2
Getting Started
Installing Optional Measurement Personalities
Installer Screen and Menu
The top portion of the install screen is a table in which the files that are
about to be installed are listed. The bottom portion of the screen
contains information needed to track the progress of the install.
1File Table displays the files to be installed and various file
information. If there are more then six files, Next Item and Prev Item
allow you to scroll the table to view additional items.
2File Name displays the name of the files on the installation disk.
3Current Version displays the version of the file that is currently
installed in the instrument. (This field will be blank if this file is not
currently installed in the instrument or if the file is a data file that has
no version.)
4Upgrade Version shows the version of the file on the install disk.
This is the version of the file that will replace the currently installed
version.
5Status is updated to reflect what the installer is doing to the current
file as the install progress. The valid messages seen in this column are
listed in Table 2-5 on page 2-16.
6Data Field contains a status bar and various status information.
7Processing disk shows the disk that is currently being read.
8Processing item shows the file that is being processes by item
number.
9Bytes free on C is the number of bytes currently free on the
instrument C: drive.
Chapter 22-15
Getting Started
Installing Optional Measurement Personalities
10Status Bar contains a status bar that runs from 0 to 100% and tracks
the progress of the current step and a message line displays the step
that is currently being executed.
11Bytes in package lists the number of bytes in the install package/ fill.
12Bytes left in package lists the number of bytes left to be read.
13Message and error popup window that displays over the status bar.
Information in this box will prompt you for action required to proceed
to the next phase of the installation.It may also inform you of errors in
the installation process and may prompt you for action required to
correct the problem.
Table 2-5Installer Status Messages
FailedThis means that something has gone wrong while processing
this item. It is a fatal error and the installation can not be
completed. The installer will try to get the system back to a
good state which may entailcompletely removing the currently
installed personality.
LoadingThe file is currently being copied from the install media to the
instrument’s file system.
VerifyingThis may mean one of two things:
1. If “VerifyDisks” was pressed then Verifyingmeans that the
installer is currently reading the install media and
comparing the known checksums to ensure the data is good.
2. If “Install Now” waspressed, then Verifying means that the
installer is reading what was just loaded to ensure the
checksum is correct.
LoadedThis means that the data has been placed on the instrument
disk but has not yet been registered with the firmware.
InstalledThis means that the data has been loaded into the instrument
and registered. The install for this file is complete.
SkippingThis means that the installer has determined that this file
does not need to be loaded into the instrument.
2-16Chapter2
3Setting Up GSM Mode
This chapter introduces the basic measurement procedure including
mode setup and changing measurement frequency.
3-1
Setting Up GSM Mode
Preparing To Make Measurements
Preparing To Make Measurements
At initial power up, the analyzer is in spectrum analyzer (SA) mode,
with the Meas Off selected in the
FREQUENCY Channel menu displayed.
MEASURE menu and the
To access the GSM measurement personality, press the
panel key and select the
GSM menu key.
MODE front
Initial settings
Before making a measurement, make sure the mode setup,
measurement setup, and frequency channel parameters are set to the
appropriate settings. For further information refer to “Mode Setup” on
page 5- 32, “FREQUENCY Channel” on page 5- 6, and “Meas Setup” on
page 5- 11.
• Resetting all parameters:
To set all instrument parameters (including mode setup and
measurement setup parameters) to factory default values, press the
Preset front panel key. Note that the mode is changed from GSM to
SA when the
must use the
The
Preset front panel key can also be used to return the instrument
to a set of user preset values. Refer to the ESA Spectrum Analyzers
User’s Guide for further information.
• Resetting mode setup parameters:
Mode setup parameters apply to all measurements in GSM mode. To
reset them to factory default values, press
Restore Mode Setup Defaults.
Preset front panel key is used. After using Preset you
MODE key to return to GSM mode.
Mode Setup then
• Resetting measurement setup parameters:
Measurement setup parameters affect the current measurement
only. To reset them to factory default values (for the current
measurement only), press
3-2Chapter3
Meas Setup then Restore Meas Defaults.
Setting Up GSM Mode
Preparing To Make Measurements
Making A Measurement
GSM measurements are intended to be used as “one button”
measurements. This means that the appropriate measurement can be
selected and run by a single key press once the instrument has been
connected to the equipment to be tested. The measurement is made
automatically using default parameters defined by the selected
standard and the tuning plan.
Even though the measurements are designed as one-button
measurements, you may change the default settings using various
setup keys. However, changing the default settings may produce
measurement results that are outside of the parameters of the selected
standard and tuning plan requirements.
Most measurements can be performed using the simple four-step
procedure outlined in the table below. Most measurements are
performed using only the primary keys listed in conjunction with a
minimum of setup keys. Measurement setup keys (
used for non-standards compliant testing. For more information see
“Initial settings” above.
Meas Setup) can be
Chapter 33-3
Setting Up GSM Mode
Preparing To Make Measurements
Table 3-1Four Step Procedure For Performing Measurements
StepPrimary KeySetup KeysRelated Keys
1.Select and setup modeMODEMode SetupSystem
2.Select and setup
measurement
MEASUREMeas Setup,
Restore Meas Defaults,
FREQUENCY Channel
3.Select and setup viewView/TraceSpan X Scale,
Amplitude Y Scale
4.Saving and printing
results
File
Print
Print SetupSave
The primary keys required for performing one button measurements
are shown in the following diagram.
, Display
Meas Control
Restart
, Search
Marker
,
3-4Chapter3
Setting Up GSM Mode
Preparing To Make Measurements
Saving Measurement Results
To save measurement results, follow the process shown below. For
additional information on file management in the spectrum analyzer,
refer to the ESA Spectrum Analyzers User’s Guide.
1. Press
2. If you want to change the file name, press
File, Save, Type, More, and Measurement Results.
Name, and use the Alpha
Editor the enter the new name. For more information on using the
Alpha Editor, refer to the ESA Spectrum Analyzers User’s Guide.
3. Press
Save Now to complete the file saving process.
4. If you have used the default file name and wish to save additional
measurement results, press Save. The current measurement result
will be saved with the next default file name.
5. If you have not used the default file name and wish to save
additional measurement results, repeat steps 1 through 3.
Chapter 33-5
Setting Up GSM Mode
Preparing To Make Measurements
3-6Chapter3
4Menu Maps
This chapter provides a visual representation of the front panel keys
and their associated menu keys. For function key descriptions, refer to
Chapter 5 , “Front-Panel Key Reference.”
4-1
Menu Maps
What You Will Find In This Chapter
What You Will Find In This Chapter
This chapter provides menu maps for the front panel keys having
associated menus. The key menus appear in alphabetical order as
follows:
Det/DemodPage 4-3
DisplayPage 4-4
FREQUENCY ChannelPage 4-5
Input/OutputPage 4-6
MEASUREPage 4-7
Measurement Setup—Cable Fault LocationPage 4-8
Measurement Setup—Monitor Band/ChannelPage 4-9
Measurement Setup—Out of Band SpuriousPage 4-10
Measurement Setup—Output RF SpectrumPage 4-11
Measurement Setup—Phase and Frequency ErrorPage 4-12
Measurement Setup—Power StepsPage 4-13
Measurement Setup—Power vs TimePage 4-14
Measurement Setup—Receive Band SpuriousPage 4-15
Measurement Setup—Transmit Band SpuriousPage 4-16
Measurement Setup—Transmitter PowerPage 4-17
MODEPage 4-18
Mode SetupPage 4-19
TrigPage 4-20
View/TracePage 4-21
4-2Chapter4
Menus
Detector and Demodulation Menu
Menu Maps
Menus
Chapter 44-3
Menu Maps
Menus
Display Menu
4-4Chapter4
Frequency/Channel Menu
Menu Maps
Menus
Chapter 44-5
Menu Maps
Menus
Input/Output Menu
4-6Chapter4
Measure Menu
Menu Maps
Menus
Chapter 44-7
Menu Maps
Menus
Measurement Setup Menus
Cable Fault Location Measurement Setup Menu
4-8Chapter4
Monitor Band/Channel Measurement Setup Menu
Menu Maps
Menus
Chapter 44-9
Menu Maps
Menus
Out Of Band Spurious Measurement Setup Menu
4-10Chapter4
Output RF Spectrum Measurement Setup Menu
Menu Maps
Menus
Chapter 44-11
Menu Maps
Menus
Phase And Frequency Error Measurement Setup Menu
4-12Chapter4
Power Steps Measurement Setup Menu
Menu Maps
Menus
Chapter 44-13
Menu Maps
Menus
Power vs Time Measurement Setup Menu
4-14Chapter4
Receive Band Spurious Measurement Setup Menu
Menu Maps
Menus
Chapter 44-15
Menu Maps
Menus
Transmit Band Spurious Measurement Setup Menu
4-16Chapter4
Transmitter Power Measurement Setup Menu
Menu Maps
Menus
Chapter 44-17
Menu Maps
Menus
Mode Menu
4-18Chapter4
Mode Setup Menu
Menu Maps
Menus
Chapter 44-19
Menu Maps
Menus
Trigger Menu
4-20Chapter4
View/Trace Menu
Menu Maps
Menus
Chapter 44-21
Menu Maps
Menus
4-22Chapter4
5Front-Panel Key Reference
This chapter details the front-panel keys and menu keys that appear on
the menu-maps presented in the previouschapter. The front-panel keys
are listed alphabetically and are described with their associated menu
keys. The menu keys are arranged as they appear on your analyzer
menus.
5-1
Front-Panel Key Reference
Key Descriptions and Locations
Key Descriptions and Locations
This chapter provides information on GSM front panel keys which do
not exist in SA mode or keys which have different functionality from SA
mode equivalents. For information on SA mode keys not described in
this chapter, refer to the ESA Spectrum Analyzers User’s Guide.
DisplayPage 5-3
FREQUENCY ChannelPage 5-6
MarkerPage 5-9
MEASUREPage 5-10
Meas Setup—Cable Fault LocationPage 5-12
Meas Setup—Monitor Band/ChannelPage 5-15
Meas Setup—Out of Band SpuriousPage 5-17
Meas Setup—Output RF SpectrumPage 5-19
Meas Setup—Phase and Frequency ErrorPage 5-21
Meas Setup—Power StepsPage 5-23
Meas Setup—Power vs TimePage 5-24
Meas Setup—Receive Band SpuriousPage 5-26
Meas Setup—Transmit Band SpuriousPage 5-28
Meas Setup—Transmitter PowerPage 5-29
MODEPage 5-31
Mode SetupPage 5-32
View/TracePage 5-37
5-2Chapter5
Front-Panel Key Reference
Display
The Display front panel key leads to one of the GSM measurement
display menus depending on which measurement is selected when the
key is pressed.
Transmitter Power
Current DataThe values for Mean Transmit Power, Max Pt, and Min Pt for the
current trace are calculated and displayed on the results screen by
default.
of these values. The measurement executes slightly more quickly when
the parameter is set to off.
OnSets calculation and display on.
OffSets calculation and display off.
The default value is On.
Current Data is used to toggle on/off the calculation and display
Display
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Power Steps
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Power vs Time
AnnotationShows/hides any annotations on the graph area of the display.
OnShows annotations.
OffHides annotations.
The default value is On.
Chapter 55-3
Front-Panel Key Reference
Display
Current DataThe values for Mean Transmit Power, Max Pt, and Min Pt for the
current trace are calculated and displayed on the results screen by
default.
Current Data is used to toggle on/off the calculation and display
of these values. The measurement executes faster when the parameter
is set to off.
OnSets calculation and display on.
OffSets calculation and display off.
The default value is On.
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Output RF Spectrum
Table DisplayWhen the measurement method is set to multiple, the results table can
be changed to show any of the following sets of values:
ResDisplays the actual measurement results.
LimDisplays the absolute and relative limits for all offsets.
MargDisplays the margin by which the result meets
specifications. Negative numbers indicate that the
result has failed.
The default value is Res.
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Cable Fault Location
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
5-4Chapter5
Front-Panel Key Reference
Display
Transmit (Tx) Band Spurious
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Receive (Rx) Band Spurious
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Out Of Band Spurious
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Monitor Band/Channel
Display LineTurns on/off an adjustable horizontal line that is used as a visual
reference line.
OnDisplays the line.
OffTurns off the line.
The default value is Off.
TitleLeads to a menu enabling the title string to be changed or cleared.
PreferencesLeads to a menu enabling the graph window graticule or annotation to
be turned on/off.
Chapter 55-5
Front-Panel Key Reference
FREQUENCY Channel
FREQUENCY Channel
The FREQUENCY Channel front panel key provides access to menu keys
which control the center frequency or channel number to be used for
measurements. These parameters apply to all measurements in the
current mode.
ARFCNAbsolute RF Channel Number. Used to enter the channel to be
measured for the selected band. It is coupled with
ARFCN sets Center Freq to the value corresponding to that ARFCN.
Range is limitedto valid channels for theselected band. Screen displays
show the
values do not correlate, the
changing
ARFCN if it correlates with the Center Freq value. If the two
Center Freq value is displayed. Note that
ARFCN does not affect the currently displayed Center Freq for
any of the band measurements.
Center FreqSets the channel frequency to be measured for the selected band. It is
coupled with
corresponding to that frequency. If the
not exactly correlate with an
display the closest
ARFCN—setting Center Freq sets ARFCN to the channel
Center Freq value entered does
ARFCN, the softkey label changes to
ARFCN to the selected frequency, along with a > or <
symbol indicating whether the frequency is above or below that
Center Freq—setting
ARFCN.
BMT ARFCNEnables quick selection from a subset of the available ARFCN numbers
Device is set to MS, ARFCN values are the same but Center Freq values
Top, Middle, or Bottom frequency in the selected band as follows.
BottomMiddleTop
Center
Freq
ARFCN
Center
Freq
ARFCN
Device is set to BTS. When
Center
Freq
ARFCN
are different.
Auto ARFCNInvokes the Auto ARFCN routine which locates the current GSM signal
and sets the channel RF Channel Frequency, ARFCN, and Band
accordingly.
5-6Chapter5
Front-Panel Key Reference
FREQUENCY Channel
TimeslotSet timeslot on/off. Set on to select a specific timeslot over which to
make the demodulation measurements. The timeslot feature is only
supported in External and Frame trigger source modes:
ExternalSet the value to the number of the timeslot to be
measured, from 0 to 7.
FrameSet the value to the number of the timeslot to be
measured, from 0 to 7.
The key is grayed out:
• If Options B7D: DSP with fast ADC and B7E: Digital Demod RF are
not installed.
• If trigger source is not set to external or frame.
Ref BurstUsed to select the type of burst to be used as the frame reference burst.
Possible options are
Normal (TCH & CCH), Sync (SCH) or Access (RACH).
The key is grayed out:
• If Options B7D: DSP with fast ADC and B7E: Digital Demod RF are
not installed.
• If trigger source is not set to frame.
Ref TSC (Std)Training sequence code.
The key is grayed out:
• If Options B7D: DSP with fast ADC and B7E: Digital Demod RF, are
not installed.
Ref Burst other than Normal (TCH & CCH) is selected.
•If
• If trigger source is not external or frame.
Chapter 55-7
Front-Panel Key Reference
FREQUENCY Channel
Default settings
The following table shows the default settings for the Frequency
Channel menu keys.
Frequency Channel Menu Keys: Default Settings
ARFCN38
Center Freq942.6 MHz (Mid for E-GSM)
BMT ARFCNN/A
Auto ARFCNN/A
TimeslotOff
Burst TypeNormal
TSCAuto
5-8Chapter5
Front-Panel Key Reference
Marker
Marker
The functions accessed by the Marker key are identical to those accessed
in SA mode. For further information refer to the Agilent ESA Spectrum
Analyzers User’s Guide.
Chapter 55-9
Front-Panel Key Reference
MEASURE
MEASURE
The MEASURE front panel key provides access to menu keys which
enable you to make the following ESA-E Series Spectrum Analyzer
GSM measurements:
•
Transmitter Power
• Power Steps
• Power vs Time
• Phase And Frequency Error
• Output RF Spectrum
• Cable Fault Location
• Transmit (Tx) Band Spurious
• Receive (Rx) Band Spurious
• Out Of Band Spurious
• Monitor Band/Channel
5-10Chapter5
Front-Panel Key Reference
Meas Setup
Meas Setup
The Meas Setup front panel key displays a menu that allows you to
enter custom setup parameters for a measurement. The setup menu
displayed depends on the measurement selected from the
menu.
MEASURE
Chapter 55-11
Front-Panel Key Reference
Cable Fault Location
Cable Fault Location
To access the keys for setting up a cable fault location measurement,
press
MEASURE, Cable Fault Location, and then the front panel
Meas Setup key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp.
CalibrateUsed to calibrate the spectrum analyzer, removing any errors
introduced by the cabling and components of the test setup before
making the measurement.
Cable TypeUsed to select the type of cable being tested. Provides access to the
Cable Editor form, displaying a table listing relevant cable types, plus
the following keys:
CableUsed to select an entry in the cable type table. Accepts
numeric values.
Vel FactorThe velocity factor for the cable type selected. The
field’s numeric value represents a proportion of the
speed of light. The value can be edited and saved using
the
Store Table key.
Cable Loss/mThe amount of signal lost per meter for the cable type
selected, expressed in dB. The value can be edited and
saved using the
Store TableUsed to save any changes made to the Cable,
Vel Factor or Cable Loss/m parameters.
5-12Chapter5
Store Table key.
Front-Panel Key Reference
Cable Fault Location
Max RangeUsed to define the range over which faults will be measured. The
smaller the value of this fieldthe more closely faults can be viewed. The
value should therefore be set as close as possible to the length of the
test apparatus.
The default value is 20m.
Trace PointsThe number of trace points used in the measurement. Changing the
number of trace points affects the minimum and maximum measurable
distances. As a general rule, the longer the cable, the more trace points
you should use. It is possible to select up to 8192 trace points but due to
the computationally intensive FFT, the higher the value, the slower the
measurement becomes. To ensure the fastest possible speed, use a
value that is a power of 2—for example 512, 1024 and so on.
WindowUsed to select the FFT windowing function to be used. The cable fault
measurement uses an FFT to convert the analyzer frequency trace into
a distance trace. To get the best possible results from the FFT, it is
important to apply the most suitable windowing function to the
frequency trace before performing the FFT. Possible values are
Rectangular, Flat Top, Gaussian, and Hanning. The default value is
Flat Top. The following table describes these values.
FunctionDescriptionMax Side-Lobe
Level
Rectangular Results in no
windowing.
Flat Top
(default)
GaussianA five term cosine
HanningHas a good frequency
Restore Meas
Defaults
The five term flat-top
window. A good window
to use when making
amplitude
measurements of
relatively pure tones.
window that resembles
a Gaussian window.
resolution and
reasonably good
side-lobe roll-off, but
poor main-lobe flatness
and relatively large
side-lobe peaks.
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
–13.261 dB 20 dB/decade–3.9224 dB
–95.1 dB+/–0.002077456085 dB
–125.4 dB–0.680056 dB
–31.46730784 dB 60 dB/decade–1.423622 dB
Side-Lobe
Roll Off
Max Main
Chapter 55-13
Front-Panel Key Reference
Cable Fault Location
AdvancedProvides access to the following advanced measurement parameter
settings:
Res BWUsed to set the resolution bandwidth used for the
measurement and to define whether it is automatic or
manual. Default values are 3 MHz and Auto.
Video BWUsed to set the video bandwidth used for the
measurement and to define whether it is automatic or
manual. Default values are 3 MHz and Auto.
Center FreqUsed to activate the centerfrequency function, allowing
you to select a frequency to be centered on the display.
Start FreqUsed to set the frequency at the left side of the
graticule. The left and right sides of the graticule
correspond to the start and stop frequencies. When
these frequencies are activated their values are
displayed below the graticule in place of center
frequency and span.
Stop FreqUsed to set the frequency at the right side of the
graticule. The left and right sides of the graticule
correspond to the start and stop frequencies. When
these frequencies are activated their values are
displayed below the graticule in place of center
frequency and span.
TG AmplitudeUsed to set the power level of the internal source.
5-14Chapter5
Front-Panel Key Reference
Monitor Band/Channel
Monitor Band/Channel
Toaccess the keys for setting up a monitor band/channel measurement,
press
Meas Setup key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
MEASURE, Monitor Band/Channel, and then the front panel
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp.
MethodUsed to select the method used for monitoring. Possible values are
Channel and Band and the default value is Band.
Channel SpanUsed to toggle between monitoring the current ARFCN by itself (One)
or the ARFCN and the three channels on either side of it (Three). The
default value is One.
Channel Span is grayed out when Method is set to
Band.
BandUsed to toggle between monitoring transmit (Tx) or receive (Rx). The
Chan Setup/ Band
Setup
default value is Tx.
Chan Setup is displayed when Method is set to Channel. Band Setup is
displayed when
Band is grayed out when Method is set to Channel.
Method is set to Band. Both provide access to the
following keys:
Res BWUsed to change the 3 dB resolution bandwidth in a 1, 3,
10 sequence. The default value is 100 kHzfor Bandand
10 kHz for Channel.
Video BWUsed to change the analyzer post detect filter in a 1, 3,
10 sequence. The default value is 100 kHzfor Bandand
10 kHz for Channel.
Chapter 55-15
Front-Panel Key Reference
Monitor Band/Channel
Max Hold Displays a trace of the maximum values for a
measurement. The default value is Off.
DetectorUsed to change the type of video detection used.
Possible values are Peak, Sample and Neg Peak. The
default value is Peak.
Method is set to Band.
Detector is grayed out when
Restore Meas
Defaults
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
5-16Chapter5
Front-Panel Key Reference
Out Of Band Spurious
Out Of Band Spurious
Toaccess the keys for setting up an out of band spurious measurement,
press
Setup key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeGrayed out for this measurement.
MS IdleUsed to specify that you wish to measure a mobile station that is in idle
mode—that is, where no call is in progress. The measurement uses the
filter bandwidths and sweep times appropriate to the specifications.
This involves some longer sweep times of greater than 60 seconds.
Possible values are On and Off. The default value is Off.
grayed out unless Device Type is set to MS in the Radio Mode setup
menu.
MEASURE, Out Of Band Spurious, and then the front panel Meas
MS Idle is
Inspect SpurUsed to select a spur number to be examined once the measurement
has finished running.
Inspect Spur is grayed out when a measurement is
in progress. When pressed, the key provides access to the following
keys:
Inspect SpurWhen set to On, the inspect spur functionality is made
available and a spur number can be supplied for
further examination. Possible values are On and Off.
The default value is Off.
Sweep TimeThe default value is 2 seconds and the range 4 ms to
500 seconds.
Res BWThe default value is 1 MHz and the range 1 kHz to
5 MHz
Video BWThe default value is 3 MHz and the range 1 kHz to
3 MHz.
Limits....Accesses the limits form for changing the limits for the currently
selected radio band and device. The measurement is restarted if any
limits value is changed.
Restore Meas
Defaults
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
Chapter 55-17
Front-Panel Key Reference
Out Of Band Spurious
AdvancedProvides access to the following advanced measurement parameter
setting:
Max Mxr LvlUsed to set the maximum power level at the RF mixer
input for the measurement. The instrument uses this
value to automatically set the required attenuation to
maintain the mixer input below the critical level. The
default value is 5 dBm and the range –100 dBm to
10 dBm.
5-18Chapter5
Front-Panel Key Reference
Output RF Spectrum
Output RF Spectrum
Toaccess the keys for setting up an output RF spectrum measurement,
press
Setup key.
Avg BurstsUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
MEASURE, Output RF Spectrum and then the front panel Meas
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
MeasureUsed to select the measurement method. Possible values are multi
offset, single offset and swept. The default value is multi offset.
Meas TypeUsed to switch measurement type between Output RF Spectrum due to
modulation (
(
Switch). The default value is Mod.
Ofs Freq ListUsed to select which set of frequency offsets to scan in multi offset
mode. Possible values are
Short. Grayed out unless
Offset FreqUsed to select a frequency offset from the carrier at which to perform a
Mod) and Output RF Spectrum due to switching transients
Standard and Short. The default value is
Measure is set to Multi Offset.
single offset Output RF Spectrum measurement. The default value is
250 kHz and the range –6 MHz to +6 MHz.
Trig SourceUsed to select the trigger source for the measurement. Possible values
are Free Run, RF Burst, External and Frame. The default value is RF
Burst. RF Burst and Frame are grayed out when hardware
Option B7D/B7E is not installed.
Chapter 55-19
Front-Panel Key Reference
Output RF Spectrum
Burst SyncUsed to select how the measurement will synchronize with the correct
part of the burst. Possible values are Training Seq, RF Amptd and
None. The default value is Training Seq. Training Seq is grayed out
when hardware Option B7D/B7E is not installed. The parameter is not
applicable to the swept measurement.
W/band NoiseWhen set to Off, the analyzer is tuned to the carrier and –1800 kHz to
+1800 kHz either side of the center frequency is swept. When set to On,
the whole of the relevant band +2 MHz either side is swept. The default
value is Off. The parameter is grayed out unless
and
Meas Method to Swept.
Meas Type is set to Mod
Restore Meas
Defaults
AdvancedProvides access to the following advanced measurement parameter
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
settings:
Ref Pwr AvgUsed to specify the number of averages to be taken
when determining the reference power (at 0 Hz offset).
This can be done by using a number of averages equal
to that specified in the
Avg Bursts parameter (Auto) or
by specifying a number manually (Man). The default
values is Auto (10). The parameter is grayed out unless
Meas Method is set to Single Offset.
Max Mxr LvlUsed to set the maximum power at the input mixer for
the out of band spurious measurement.
Mod BW/
Switch BWUsed to deviate from the GSM standard by specifying
custom RBW settings for the indicated frequency
offsets—that is, <1800 kHz and ≥1800 kHz.
5-20Chapter5
Front-Panel Key Reference
Phase And Frequency Error
Phase And Frequency Error
To access the keys for setting up a phase and frequency error
measurement, press
panel
Meas Setup key. The measurement is grayed out when the
hardware Options B7D and B7E are not installed.
Avg BurstsUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
MEASURE, Phase & Frequency, and then the front
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
Avg TypeUsed to select the type of averaging.
MaximumThe maximum values are restored (the maximum of all
values is retained).
MeanThe mean values are restored (the average of all values
is retained).
The default value is Maximum.
Trig SourceUsed to select the trigger source for the measurement. Possible values
are Free Run, RF Burst, External and Frame. The default value is Free
Run.
Chapter 55-21
Front-Panel Key Reference
Phase And Frequency Error
Burst SyncUsed to select how the measurement will synchronize with the correct
part of the burst. Possible values are Training Seq, RF Amptd and
External. The default value is Training Seq. The parameter is not
applicable to the swept measurement.
Limit TestUsed to turn limit checking on and off. This is not the same as limit
line—the numeric phase and frequency error results are checked
against the
RMS Phase Limit and Frequency Limit parameters to see if
they meet the limit requirements. The relevant PASS/FAIL annotation
is displayed in the measurement bar. The default value is On.
Limits....Provides access to the following:
RMS Phase Limit The RMS phase error must be less than or equal to
this limit. If the RMS phase error exceeds it, the
measurement display shows the RMS phase error in
red and appends a red “F” to the value. The default
value is 5 degree and the range from 0 degree to
100 degree.
Restore Meas
Defaults
Peak Phase Limit The peak phase error must be less than or equal to
this limit. If the peak phase error exceeds it the
measurement display shows the peak phase error in
red and appends a red “F” to the value. The default
value is 20 degree and the range from 0 degree to
100 degree.
Frequency Limit The frequency error must be less than or equal to this
limit. If the frequency error exceeds it the
measurement display shows the frequency error in red
and appends a red “F” to the value. The default value is
0.05 ppm (BTS) and 0.1 ppm (MS) and the range from
0 ppm to 100 ppm.
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
5-22Chapter5
Front-Panel Key Reference
Power Steps
Power Steps
To access the keys for setting up a power steps measurement, press
MEASURE, Power Steps, and then the front panel Meas Setup key.
Sweep TimeUsed to set the length of the measurement interval. The default value
is 2 seconds and the range 5ms to 400 seconds.
Marker DeltaUsed to adjust the marker separation distance. The distance is in time
as the power steps measurement is performed in zero span. The results
are updated as soon as the marker positions are changed.
Marker CenterUsed to adjust the delta marker’s position on the trace without
affecting the separation distance. The position is in time as the power
steps measurement is performed in zero span. The results are updated
as soon as the marker positions are changed.
Limit TestUsed to turn limit checking on and off. This is not the same as limit
line—the numeric power steps result is checked against the
Upper Limit and Lower Limit parameters to see if they meet the limit
requirements. The relevant PASS/FAIL annotation is displayed in the
measurement bar. The default value is On.
Limits....Provides access to the following:
Upper LimitThe absolute power step value must be less than or
equal to this limit. If it exceeds it, the power step result
is displayed in red and appended with an “F”. The
default value is 3.5 dB and the range 0 dB to 200 dB.
Lower LimitThe absolute power step value must be greater than or
equal to this limit. If it is less, the power step result is
displayed in red and appended with an “F”. The default
value is 0.5 dB and the range 0 dB to 200 dB.
Restore Meas
Defaults
AdvancedProvides access to the following advanced measurement parameter
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
settings:
Res BWUsed to change the 3 dB resolution bandwidth in a 1, 3,
10 sequence. The default value is 1 MHz and the range
1 kHz to 5 MHz.
Video BWUsed to change the analyzer post detect filter in a 1, 3,
10 sequence. The default value is 300 kHz and the
range 30Hz to 3 MHz.
Chapter 55-23
Front-Panel Key Reference
Power Vs Time
Power Vs Time
To access the keys for setting up a power vs time measurement, press
MEASURE, Power vs Time, and then the front panel Meas Setup key.
Avg BurstsUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and Off and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
Avg TypeUsed to specify the type of trace and result averaging to use.
VideoSelects video averaging.
PowerSelects power averaging.
The default value is Power.
Meas TimeUsed to set the number of slots to be displayed on the trace in each
measurement cycle.
Meas Time is closely tied to sweep time. The default
value is 1 slot and the range 1 to 50 slots.
Trig SourceUsed to select the trigger source for the measurement. Possible values
are Free Run,RF Burst,External, and Frame.The default value is Free
Run. RF Burst and Frame are grayed out when the hardware
Option B7D/B7E is not installed.
Burst SyncUsed to select how the measurement will synchronize with the correct
part of the burst. Possible values are Training Seq, RF Amptd, and
External. The default value is Training Seq. Training Seq is grayed out
when the hardware Option B7D/B7E is not installed. The parameter is
not applicable to the swept measurement.
5-24Chapter5
Front-Panel Key Reference
Power Vs Time
Max HoldUsed to display a max hold trace to allow monitoring over time. Max
Hold is grayed out and set to Off for all views except Frame. The default
value is Off. The
grayed out when
Limit TestUsed to turn limit checking on and off. This is not the same as limit
Max Hold Trace softkey in the Marker Trace menu is
Max Hold is set to Off.
line—the numeric phase and frequency error results are checked
against the
RMS Phase Limit and Frequency Limit parameters to see if
they meet the limit requirements. The relevant PASS/FAIL annotation
is displayed in the measurement bar. The default value is On.
Restore Meas
Defaults
AdvancedProvides access to the following advanced measurement parameter
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
settings:
Res BWUsed to change the 3 dB resolution bandwidth in a 1, 3,
10 sequence. The default value is 100 kHzfor Bandand
10 kHz for Channel.
Video BWUsed to change the analyzer post detect filter in a 1, 3,
10 sequence. The default value is 100 kHzfor Bandand
10 kHz for Channel.
Trace PointsUsed to change the number of trace points used in the measurement.
Set to a lower number to decrease the resolution of the trace and the
mask, but increase measurement speed. The default value is 801 and
the range 101 to 801.
Chapter 55-25
Front-Panel Key Reference
Receive Band Spurious
Receive Band Spurious
Toaccess the keys for setting up a receive band spurious measurement,
press
MEASURE, Rx Band Spurs, and then the front panel Meas Setup
key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and Off and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
Preamp GainUsed to specify the gain of any external preamp that may be present at
the input of the analyzer during the measurement. The default value is
20 dB and the range –20 dB to 50 dB.
LimitThe upper limit value for testing spurs. By default the parameter is
based on the ETSI specification. It changes depending on the
band/device in use. The default value is –36 dBm and the range
–200 dBm to 100 dBm.
5-26Chapter5
Front-Panel Key Reference
Receive Band Spurious
Restore Meas
Defaults
AdvancedProvides access to the following advanced measurement parameter
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
settings:
Res BWUsed to change the 3 dB resolution bandwidth in a 1, 3,
10 sequence. The default value is 100 kHz and range
1 kHz to 5 MHz.
Video BWUsed to change the analyzer post detect filter in a 1, 3,
10 sequence. The default value is 100 kHz and range
1 kHz to 3 MHz.
Chapter 55-27
Front-Panel Key Reference
Transmit Band Spurious
Transmit Band Spurious
To access the keys for setting up a transmit band spurious
measurement, press
Meas Setup key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and Off and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
MEASURE, Tx Band Spurs, and then the front panel
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
Meas TypePossible values are Examine and Full. The default value is Full.
ExamineIn continuous measure, after doing one full search
across all segments, it stops on the worst segment and
continuously updates that segment.
FullIn continuous measure, it repeatedly does a full search
of all segments.
Idle ModeSet to on when measuring with no active carriers. The specification
document outlines different limits for cases where there is an active
carrier present and absent.
LimitThe upper limit value for testing spurs. By default the parameter is
Idle Mode is coupled to the Limit parameter.
based on the ETSI specification. It changes depending on the
band/device in use. The default value is –36 dBm and the range
–200 dBm to 100 dBm.
Restore Meas
Defaults
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
5-28Chapter5
Front-Panel Key Reference
Transmitter Power
Transmitter Power
To access the keys for setting up a transmitter power measurement,
press
Meas Setup key.
Avg NumberUsed to specify the number of data acquisitions that will be averaged.
After the specified number of average counts, the averaging mode
(terminal control) setting determines the averaging action.
OnSets measurement averaging on.
OffSets measurement averaging off.
Default values are 10 and On, and range values are 1 to 1000.
Avg ModeUsed to select the type of termination control used for the averaging
function. This determines the averaging action after the specified
number of data acquisitions (average count) is reached.
MEASURE, Transmitter Power, and then the front panel
ExponentialAfter the average count is reached, each successive
data acquisition is exponentially weighted and
combined with the existing average.
RepeatAfter reaching the averagecount, theaveraging is reset
and a new average is started.
The default value is Exp. Grayed out for Multiple measurement
method—only Repeat is supported.
Avg TypeUsed to specify the type of trace and result averaging to use.
VideoSelects video averaging.
PowerSelects power averaging.
The default value is Power.
Threshold LvlThe mean carrier power is calculated based on the trace above the
threshold level. The threshold level may be described in dB (relative to
the carrier) or dBm (absolute). A green line on the trace area will be
displayed at the y-position associated with the current threshold level
value, converted to the current amplitude display units. The default
value is –6 dB and relative, and the range –60 dB to 60 dB.
Trig SourceUsed to select the trigger source for the measurement. Possible values
are Free Run, RF Burst, External and Frame. The default value is Free
Run. RF Burst and Frame are grayed out when the hardware
Option B7D/B7E is not installed.
Chapter 55-29
Front-Panel Key Reference
Transmitter Power
Burst SyncUsed to select how the measurement will synchronize with the correct
part of the burst. Possible values are Training Seq, RF Amptd, and
None. The default value is Training Seq. Training Seq is grayed out
when hardware Option B7D/B7E is not installed.
Meas TimeUsed to set the number of timeslots to be displayed on the trace in each
measurement cycle.
Meas Time is closely tied to sweep time. The default
value is 1 timeslot and the range 1 to 8 timeslots.
Restore Meas
Defaults
AdvancedProvides access to the following advanced measurement parameter
Restores all Meas Setup parameter values for the current measurement
to their factory defaults.
settings:
Res BWUsed to change the 3 dB resolution bandwidth in a
1, 3, 10 sequence. The default value is 300 kHz and
range 1 kHz to 5 MHz.
Video BWUsed to change the analyzer post detect filter in a 1, 3,
10 sequence. The default value is 300 kHz and range
30Hz to 3 MHz.
5-30Chapter5
Front-Panel Key Reference
MODE
MODE
The MODE front panel key provides access to menu keys for selecting
the analyzer measurement mode.
The basic spectrum analyzer comes with only SA mode
installed—measurement personality firmware must be installed for
other
SAAccesses the base-instrument spectrum analyzer mode menu keys.
GSMAccesses the GSM measurement personality menu keys and associated
functions.
MODE menu keys to be labelled and functional.
Chapter 55-31
Front-Panel Key Reference
Mode Setup
Mode Setup
The Mode Setup front panel key provides access to front panel key for
setting up parameters that apply to all measurements for the selected
mode. This contrasts to the
to parameters which are measurement specific.
RadioProvides access to a screen enabling the following parameters to be
changed:
Band/Standard Select from P-GSM 900, E-GSM 900, R-GSM 900,
DCS 1800, or PCS 1900.
DeviceSelect the device to test: MS (mobile station),BTS (base
station) or uBTS M1-3 (micro base station class
M1-M3).
Freq Hopping Repetition Factor Set frequency hopping repetition factor
to On or Off. If set On, adefault value of 3 is provided in
the following field. The value has a range of 1 - 100. The
key is grayed out for the Power Steps, Phase And
Frequency and Transmit Band Spurious
measurements.
Meas Setup hardkey which provides access
DCS/PCS Overlap Priority Set to DCS or PCS. Determines which
band/standard is automatically selected when entering
an ARFCN in the range 512 - 810.
InputProvides access to a screen enabling the following parameters to be
changed:
RF CarriersPossible values are Single and Multiple. Select Single if
there is a single RF carrier present at the RF Input.
Select
Multiple if there is more than one carrier present
at the RF Input. This will reject any adjacent channels
for the modulation accuracy measurements. The key is
grayed out for Phase And Frequency Error and Monitor
Band/Channel measurements.
RF Input Range Possible values are Auto and Manual. If Auto is chosen,
the instrument automatically sets the attenuator and
reference level based on the total power level of the
input signal carriers. If there are multiple carriers
present, the total power might overdrive the front end.
In this case you need to set the
RF Input Range to Manual
and enter the expected Max Total Pwr. Manual is also
used if you want to hold the input attenuation constant
(for the best relative power accuracy). For single
carriers it is generally recommended that
Range is set to Auto. The key is grayed out for the
RF Input
5-32Chapter5
Front-Panel Key Reference
Mode Setup
Phase And Frequency Error measurement.
Max Total Power To set the maximum total power at the UUT (Unit
Under Test). This is the maximum expected value of
the mean carrier power referenced to the output of the
UUT (may include multiple carriers). The Max Total
Power setting is coupled to the
Total Power is unavailable when the
set to
Auto. The key is grayed out for the Phase And
Input Atten setting. Max
RF Input Range is
Frequency Error measurement.
Input Attenuation To set the input attenuator setting. The input
attenuator setting is coupled to the
Max Total Power
setting. The Input Attenuation key reads out the actual
internal input attenuatorvalue that will be usedfor the
current measurement. If more than one input
attenuator value is used in a single measurement, the
value used at the carrier frequency will be displayed.
Input Attenuation is unavailable when the
Range is set to Auto. The key is grayedout forthe Phase
RF Input
And Frequency Error measurement.
NOTEAs the Max Total Power and Input Attenuation settings are coupled
together, changing the input
Attenuation by x dB, and vice-versa for a given measurement. When you
switch to a different measurement, the
but the
Input Attenuation may change if the two measurements have
Max Total Power by x dB changes the Input
Max TotalPower is kept constant,
different mixer margins. Thus, you can directly set the analyzer input
attenuation, or you can set it indirectly by specifying the maximum
expected power at the UUT (Max Total Power setting).
External Gain/Attenuation Used to enter the external attenuator/gain
setting for any devices between the UUT and the
analyzer. This enables the instrument to display the
measurement results referenced to the output of the
UUT.
Internal Preamp Used to control the analyzer’s in-built preamplifier for
better sensitivity. Possible values are
Off and On.
The key is grayed out:
• If Option 1DS: RF Preamp is not installed.
• For the Phase And Frequency Error measurement.
Chapter 55-33
Front-Panel Key Reference
Mode Setup
Default settings
The following table shows the default settings for the Input screen.
TriggerProvides access to the mode setup menu for the following trigger
0
sources:
•
External
• RF Burst
• Frame Delay
Pressing the trigger key enables you to set the Delay, Level, and Slope
for each trigger source where applicable. Delay, Level, and Slope are
grayed out when not applicable. Note that the trigger source for each
measurement may be selected separately (under the
DelayFor trigger delay use positive values. For pre-trigger,
Meas Setup key.
use negative values.
LevelFor the RF Burst selection, the level is relative to the
peak level of the RF signal.
SlopeSelect from Positive or Negative to trigger on the
leading edge (
Pos) or the trailing edge (Neg) of the
signal.
The following keys are grayed out:
• For trigger delay: the
RF Burst key is grayed out when Option
B7E: Digital Demod RF is not installed or when Options
B7D: DSP with Fast ADC and B7E: Digital Demod RF are not
installed.
• For trigger level: the
RF Burst key is grayed out when Option
B7E: Digital Demod RF is not installed.
• For trigger slope: the RF Burst key is grayed out when Option
B7E: Digital Demod RF is not installed.
5-34Chapter5
Front-Panel Key Reference
Default settings
The following table shows the trigger default settings.
Trigger Default Settings
External Delay0.000 s
External LevelΝ/Α
External SlopePos
RF Burst Delay0.000 s
RF Burst Level−6.00 dBc
RF Burst SlopePos
Frame Delay0.000 s
Frame LevelN/A
Mode Setup
DemodThe
Burst Alignment Selects the sync alignment to the GSM standard or
RF Sync DelaySets the delay or offset applied when in RF Amptd
Burst Search Threshold Sets the relative power threshold from the peak
Opt Freq RefAccessed menu keys allowing you to set the reference
Frame SlopeN/A
Demod key accesses a screen displaying the following parameters:
offset by 1/2 bit.
alignment mode. When
Burst Sync =RF Amptd, RF Sync
Delay provides a common delay or offset of the burst
alignment.
power, which is used by the burst alignment algorithm
to determine the burst rising edge and falling edge
when
Burst Sync = RF Amptd.
oscillator in the Option B74 RF assembly or the
external oscillator as the reference oscillator (time
base) source for the DDRF assembly. Possible values
are:
IntSets the reference oscillator in the
Option B74 RF assembly as the
reference oscillator source.
Ext Sets the external oscillator as the
reference oscillator source.
Opt Freq RefAllows you to enter the frequency of the external
reference oscillator being supplied to the Option B74
RF assembly Ext Ref In jack. This provides the proper
Chapter 55-35
Front-Panel Key Reference
Mode Setup
frequency division in the Option B74 RF assembly to
achieve the required 10 MHz reference frequency.
Opt Freq RefAllows you to enter the frequency source for the
Option B74 RF assembly 10 MHz Ref Out jack to the
option reference oscillator.
OnEnables the 10 MHz Ref Out.
OffDisables the 10 MHz Ref Out.
The key is grayed out if Option B7E: Digital Demod RF
is not installed.
Default settings
The following table shows the default settings for the Demod screen.
PropertiesProvides access to a screen displaying the GSM Application Version,
–10.00 dB
default value is OFF.
When Opt Freq Ref = Int,
default value is N/A
DSP Bootrom Version and DSP GSM Code Version.
5-36Chapter5
Front-Panel Key Reference
View/Trace
View/Trace
The ViewTrace front panel key leads to one of the GSM measurement
view/display menus depending on which measurement is selected when
the key is pressed. Used toselect between different measurement views
where appropriate.
Power vs Time
MaskUsed to analyze a complete burst and compare it to a set of upper and
lower limit lines.
MonitorUsed to monitor up to an entire frame without the pass/fail limits.
Transmit (Tx) Band Spurious
Allows you to view the lowersegment after the measurement completes
(single measurement and Meas Type Full) or continually measure the
lower segment (Meas Type Examine).
Chapter 55-37
Front-Panel Key Reference
View/Trace
5-38Chapter5
6If You Have a Problem
This chapter includes information on how to check for a problem with
your ESA spectrum analyzer, and how to return it for service. It also
includes descriptions of all of the analyzer built-in messages.
6-1
If You Have a Problem
If you have a Problem
If you have a Problem
Your analyzer is built to provide dependable service. However Agilent
Technologies worldwide sales and service organization is ready to
provide the support if you:
• Experience a problem.
• Desire additional information.
• Wish to order parts, options, or accessories.
In general, a problem can be caused by a hardware failure, a software
error, or a user error. Follow these general steps to determine the cause
and to resolve the problem.
1. Performthe quick checks listed in “Check the Basics” in this chapter.
It is possible that a quick check may eliminate your problem
altogether.
2. If the problem is a hardware problem, you have two options:
a. Repair it yourself; see the “Service Options” section in this
chapter.
WARNINGThere are no operator serviceable parts inside the analyzer.
Refer servicing to qualified personnel. To prevent electrical
shock do not remove covers.
b. Return the analyzer to Agilent Technologies for repair; if the
analyzer is still under warranty or is covered by an Agilent
Technologies maintenance contract, it will be repaired under the
terms of the warranty or plan (the warranty is at the front of this
manual).
If the analyzer is no longer under warranty or is not covered by
an Agilent Technologies maintenance plan, Agilent Technologies
will notify you of the cost of the repair after examining the
instrument. See “How to Call Agilent Technologies” and “How to
Return Your Analyzer for Service” for more information.
6-2Chapter6
If You Have a Problem
If you have a Problem
3. If the problem is a software problem, you have two options:
a. Reinstall the firmware and or the measurement personality DLP.
b. Return the analyzer to Agilent Technologies for repair; if the
analyzer is still under warranty or is covered by an Agilent
Technologies maintenance contract, it will be repaired under the
terms of the warranty or plan (the warranty is at the front of this
manual).
If the analyzer is no longer under warranty or is not covered by
an Agilent Technologies maintenance plan, Agilent Technologies
will notify you of the cost of the repair after examining the
instrument. See “How to Call Agilent Technologies” and “How to
Return Your Analyzer for Service” for more information.
Chapter 66-3
If You Have a Problem
Before You Call Agilent Technologies
Before You Call Agilent Technologies
Check the Basics
A problem can often be resolved by repeating the procedure that you
were following when the problem occurred. Before calling Agilent
Technologies or returning the analyzer for service, please perform the
following checks:
• Check the line fuse.
• Check that there is power at the receptacle?
• Check that the analyzer turned on. Make sure the fan is running,
which indicates that the power supply is on.
• If the display is dark or dim, press the upper
the upper-left corner of the front panel. If the display is too bright,
adjust the lower
Viewing Angle key in the upper-left corner of the
front panel.
• If other equipment, cables, and connectors are being used with your
ESA spectrum analyzer, make sure they are connected properly and
operating correctly.
• Review the procedure for the measurement being performed when
the problem appeared. Are all the settings correct?
• If the analyzer is not functioning as expected, return the analyzer to
a known state by pressing the
Preset key.
Viewing Angle key in
Some analyzer settings are not affected by a Preset. If you wish to
reset the analyzer configuration to the state it was in when it was
originally sent from the factory, press
Preset (Factory).
System, Power On/Preset,
• Is the measurement being performed and are the results that are
expected within the specifications and capabilities of the analyzer?
Refer to the “Specifications” chapters in the ESA SpectrumAnalyzers Specification Guide for analyzer specifications.
• In order to meet specifications, the analyzer must be aligned. Either
Auto Align All must be selected (press
Align, All), or the analyzer must be manually aligned at least once
System, Alignments, Auto
per hour, or whenever the temperature changes more than 3˚
centigrade. When
Auto Align, All is selected, AA appears on the left
edge of the display (in SA mode only).
• Is the analyzer displaying an error message? If so, refer to the
Agilent ESA Spectrum Analyzers User’s Guide.
• If the necessary test equipment is available, perform the
performance verification tests in the ESA Spectrum AnalyzersCalibration Guide. Record all results on a Performance Verification
Test Record form which follows the tests.
6-4Chapter6
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.