The information contained in this document is subject to change
without notice.
Hewlett-Packard makes no warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Hewlett-Packard
shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance,
or use of this material.
The following safety symbols are used throughout this manual.
Familiarize yourself with the symbols and their meaning before
operating this instrument.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
NOTENote calls out special information for the user’s attention. It provides
operational information or additional instructions of which the user
should be aware.
ii
The instruction documentation symbol. The product is
marked with this symbol when it is necessary for the
user to refer to the instructions in the documentation.
This symbol is used to mark the on position of the
power line switch.
This symbol is used to mark the standby position of the
power line switch.
This symbol indicates that the input power required is
AC.
WARNINGThis is a Safety Class 1 Product (provided with a protective
earthing ground incorporated in the power cord). The mains
plug shall only be inserted in a socket outlet provided with a
protected earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
WARNINGIf this product is not used as specified, the protection provided
by the equipment could be impaired. This product must be used
in a normal condition (in which all means for protection are
intact) only.
iii
Warranty
This Hewlett-Packard instrument product is warranted against defects
in material and workmanship for a period of three years from date of
shipment. During the warranty period, Hewlett-PackardCompanywill,
at its option, either repair or replace products which prove to be
defective.
For warranty service or repair, this product must be returned to a
service facility designated by Hewlett-Packard. Buyer shall prepay
shipping charges to Hewlett-Packard and Hewlett-Packard shall pay
shipping charges to return the product to Buyer. However, Buyer shall
pay all shipping charges, duties, and taxes for products returned to
Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and firmware designated
by Hewlett-Packard for use with an instrument will execute its
programming instructions when properly installed on that instrument.
Hewlett-Packard does not warrant that the operation of the
instrument, or software, or firmware will be uninterrupted or
error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse, operation
outside of the environmental specifications for the product, or improper
site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. HEWLETT-PACKARD SHALL NOT BE
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
TORT, OR ANY OTHER LEGAL THEORY.
iv
Where to Find the Latest Information
Documentation is updated periodically. Forthelatestinformationabout
HP ESA Spectrum Analyzers, including firmware upgrades and
application information, please visit the following Internet URL:
This chapter provides information on how to inspect your analyzer
when it arrives and what to expect when it is first powered on. It also
includes the analyzer power requirements and printing information.
1-1
Preparing for Use and Printing
Initial Inspection
Initial Inspection
Inspect the shipping container and verify that the contents are
complete. Keep the container and cushioning material until you have
inspected the analyzer.
Table 1-1 contains the accessories shipped with the analyzer. If the
contents are incomplete, or the analyzer does not pass the performance
tests, notify the nearest Hewlett-Packard office. If the shipping
container is damaged or the cushioning material shows signs of stress,
also notify the carrier. Keep the shipping materials for the carrier’s
inspection. The HP office will arrange for repair or replacement without
waiting for a claim settlement.
If the shipping materials are in good condition, retain them for possible
future use. You may wish to ship the analyzer to another location or to
return it to Hewlett-Packard for service. See “How to Return Your
Analyzer for Service” on page 4-7 for more information about shipping
materials.
If cleaning is necessary, use a damp cloth only.
WARNINGTo prevent electrical shock, disconnect the spectrum analyzer
mains before cleaning. Use a dry cloth or one slightly
dampened with water to clean the external case parts. Do not
attempt to clean internally.
1-2Chapter1
Table 1-1Accessories Supplied with the Analyzer
DescriptionHP Part NumberComments
Preparing for Use and Printing
Initial Inspection
Adapter, Type N (m) to
BNC (f)
Adapter, BNC (m)toBNC
(f), 75 Ω
Adapter, Type N (m) to
SMA (f)
Adapter, BNC (f) to SMA
(m)
Cable, BNC (m) to BNC
(m), 203 mm
Cable,SMA (m) toType N
(m), 220 mm
Power cableSee the table after
1250-0780Not shipped with Option 1DP. Two adapters
1250-2477Shipped only with Option 1DP. Two adapters
1250-1250Shipped only with Option 1DN for HP
1250-1200Shipped only with Option BAB.
HP 10502AShipped only with HP E4402B, E4403B,
8120-5148Shipped only with Option 1DN for HP
“Power Cable”.
are shipped with Option 1DN.
shipped with Option 1DQ.
E4402B, E4403B, E4404B, E4405B, E4407B
and E4408B. Not shipped with Option BAB.
E4404B, E4405B, E4407B and E4408B.
E4402B, E4403B, E4404B, E4405B, E4407B
and E4408B.
Shipped with analyzer.
Chapter 11-3
Preparing for Use and Printing
Power Requirements
Power Requirements
The analyzer is a portable instrument and requires no physical
installation other than connection to a power source. There is no need
to select a line voltage.
Table 1-2AC Power Requirements
Voltage90-132 Vrms (47 to 440 Hz)
Voltage195 - 250 Vrms (47 to 66 Hz)
Power Consumption, Onless than 300 W
Power Consumption, Standbyless than 5 W
Table 1-3DC Power Requirements
Voltage12-20 Vdc
Power Consumptionless than 200 W
Checking the Fuse
Where IEC regulations apply, use a 5 by 20 mm, rated F5A, 250 V IEC
approved fuse. This fuse may be used with input line voltages of 115 V
or 230 V. Its HP part number is 2110-0709.
Where UL/CSA regulations apply, use a 5 by 20 mm rated fast blow,
5 A, 125 V UL/CSA approved fuse (HP part number 2110-0756). Thisfuse may only be used with an input line voltage of 115 V.
The line fuse is housed in a fuse holder in the upper left hand corner of
the rear panel.
To remove the fuse, first disconnect the power cord from theinstrument. Then insert the tip of a screwdriver into the slot at the
middle of the fuse holder, and turn counterclockwise to extend the fuse
holder.
1-4Chapter1
Preparing for Use and Printing
Power Requirements
WARNINGFor continued protection against fire hazard, replace the line
fuse only with the same type and rating. The use of other fuses
or material is prohibited.
Power Cable
The analyzer is equipped with a three-wire power cable, in accordance
with international safety standards. When connected to an appropriate
power line outlet, this cable grounds the instrument cabinet.
WARNINGFailure to ground the analyzer properly can result in personal
injury. Before turning on the analyzer, you must connect its
protective earth terminals to the protective conductor of the
main power cable. Insert the main power cable plug only into a
socket outlet that has a protective earth contact. DO NOT
defeat the earth-grounding protection by using an extension
cable, power cable, or autotransformer without a protective
ground conductor.
If you are using an autotransformer, make sure its common
terminal is connected to the protective earth contact of the
power source outlet socket.
Various power cables are available to connect the analyzer to the types
of ac power outlets unique to specific geographic areas. The cable
appropriate for the area to which the analyzer is originally shipped is
included with the unit. You can order additional ac power cables for use
in different areas. The following table lists the available ac power
cables, illustrates the plug configurations, and identifies the geographic
area in which each cable is appropriate.
Chapter 11-5
Preparing for Use and Printing
Power Requirements
Plug Type
250V
LN
250V
L
125V
125V
250V
230V
220V
EE
NNLLEENNLL
NNLLEENNLL
E
N
E
N
N
E
HP
a
Cable
Part
Description
Plug
b
Length
cm (in.)
Cable
Color
For Use
in Country
Number
8120-1351
E
8120-1703
8120-1369
E
N
8120-0696
8120-1378
8120-1521
8120-4753
EE
8120-4754
8120-1689
L
8120-1692
8120-2104
L
8120-2296
8120-2956
L
Straight
BS 1363A
90
Straight
AS 3112
90
Straight
NEMA 5-15P
90
Straight
NEMA 5-15P
90
Straight
CEE 7/VII
90
Straight
SEV Type 12
90
Straight
SR 107-2-D
229 (90)
229 (90)
210 (79)
200 (78)
203 (80)
203 (80)
229 (90)
229 (90)
200 (78)
200 (78)
200 (78)
200 (78)
200 (78)
Mint Gray
Mint Gray
Gray
Gray
Jade Gray
Jade Gray
Gray
Gray
Mint Gray
Mint Gray
Gray
Gray
Gray
Option 900
United Kingdom, Hong
Kong, Cyprus, Nigeria,
Singapore, Zimbabwe
Option 901
Argentina, Australia,
New Zealand, Mainland
China
Option 903
United States, Canada,
Brazil, Colombia,
Mexico,Philippines,
Saudi Arabia, Taiwan
Option 918
Japan
Option 902
Continental Europe,
Central African Republic,
United Arab Republic
Option 906
Switzerland
Option 912
Denmark
8120-2957
250V
E
L
8120-4211
N
8120-4600
250V
E
L
N
8120-5182
8120-5181
a. E =earth ground, L = line, and N = neutral.
b. Plug identifier numbers describe the plug only. The HP part number is for the complete cable assembly.
90
Straight
IEC 83-B1
90
Straight
SI 32
90
200 (78)
200 (78)
200 (78)
200 (78)
200 (78)
Gray
Mint Gray
Mint Gray
Jade Gray
Jade Gray
Option 917
South Africa, India
Option 919
Israel
1-6Chapter1
formt118
Preparing for Use and Printing
Preparation
Preparation
WARNINGIf this product is to be energized via an external
autotransformer for voltage reduction, make sure that its
common terminal is connected to a neutral (earthed pole) of the
power supply.
CAUTIONThis instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
CAUTIONVentilation Requirements:Wheninstallingtheproductinacabinet, the
convection into and out of the product must not be restricted. The
ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the product by 4 °C for every 100
watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
NOTEInstall the instrument so that the detachable power cord is readily
identifiable and is easily reached by the operator. The detachable power
cord is the instrument disconnecting device. It disconnects the mains
circuits from the mains supply before other parts of the instrument.
The front panel switch is only a standby switch and is not a LINE
switch. Alternatively, an externally installed switch or circuit breaker
(which is readily identifiable and is easily reached by the operator) may
be used as a disconnecting device.
Chapter 11-7
Preparing for Use and Printing
Turning on the Analyzer for the First Time
Turning on the Analyzer for the First Time
Before using your analyzer, plug the power cord into the analyzer.
Press ❙ (the On key).
When you turn on the analyzer, a box will appear on the lower-left
corner of the screen. The box will display your analyzer product
number, the firmware revision number, and a URL for accessing
product support information on the World Wide Web. Press
the menu keys, to remove the box from the display.
NOTERecord the firmware revision number and keep it for reference. If you
should ever need to call Hewlett-Packard for service or with any
questions regarding your analyzer, it will be helpful to have the
firmware date readily available.
To make sure your analyzer meets specifications, make sure it meets
operating temperature conditions, and allow a 5 minute warm-up
before making calibrated measurements.
ESC, above
NOTEIt is normal to hear clicking when the Auto Alignment function is on.
During the interval between sweeps portions of the analyzer’s circuitry
are realigned. Some of the circuitry is controlled by relays. It is the
rapid switching of these relays between sweeps which causes the
clicking sound. Under normal operation, these relays will last over
50 years.
To eliminate the clicking sound, turn off the automatic alignment. (See
the
Alignments key description in Chapter 5.) The Align Now All
function should, however, be performed periodically. For more
information on how often to perform Align Now All when the Auto
Alignment function is off, refer to “Specifications” in the HP ESA
Spectrum Analyzers Calibration Guide.
If you wish to use an external 10 MHz source as the reference
frequency, connect an external reference source to the
10 MHz REF IN
connector on the rear panel. The signal level should be greater than
–15 dBm.
1-8Chapter1
Preparing for Use and Printing
Printer Setup and Operation
Printer Setup and Operation
Equipment
• HP ESA Spectrum Analyzer equipped with Options A4H (HP-IB and
Parallel Interface) or 1AX (RS-232 and Parallel Interface).
• IEEE 1284 compliant printer cable (such as HP C2950A).
• Supported printer equipped with a parallel interface. (A supported
printer is one that accepts Printer Control Language Level 3 or 5).
— PCL3 printers include most HP DeskJet printers.
— PCL5 printers include most HP LaserJet printers and the
HP 1600C DeskJet printer.
Interconnection and Setup
1. Turn off the printer and the analyzer.
2. Connect the printer to the analyzer parallel I/O interface connector
using an IEEE 1284 compliant parallel printer cable.
3. If appropriate, configure your printer using configuration menus or
switches. Refer to your printer’s documentation for more specific
information on configuring your printer.
4. Turn on the analyzer and printer.
5. Press
Print Setup on the front panel and then press the Printer Type
menu key. Printer Type accesses the following keys:
NoneNone disables the analyzer from attempting to print
to a printer. This is the appropriate setting if no
printer is connected to the analyzer.
CustomCustom allows you to access the Define Custom menu
keys. The
Define Custom menu keys allow you to
specify printer characteristics such as PCL Level
and printer color capability.
AutoAuto enables the analyzer to automatically attempt
to identify the connected printer when the
is pressed or when
Printer Type is set to Auto.
Print key
Chapter 11-9
Preparing for Use and Printing
Printer Setup and Operation
6. Press Printer Type to access the Printer Type menu keys. Press Auto to
make the analyzer attempt to identify the connected printer. When
you press
Auto, the analyzer will respond in one of the three
following ways:
• The
Print Setup menu will be displayed with the Auto key selected
and no new message will be displayed in the display status line.
This indicates that the analyzer has successfully identified the
connected printer and no further setup is required. As long as
Auto remains selected in the Printer Type menu, the analyzer will
attempt to identify the printer when the front panel
Print key is
pressed. The selected printer will be displayed by pressing
System, More 1 of 3, Show System.
• The
Print Setup menu will be displayed with the Custom key
selected and one of the following diagnostic messages will be
displayed in the display status line:
Unknown printer, Define Custom to set up printer
No printer response, Define Custom to set up
printer
Invalid printer response, Define Custom to set up
printer
This indicates that the analyzer was unable to automatically
identify the connected printer, and
the
Printer Type menu. Press Print Setup, Define Custom to select
Custom has been selected in
specific printer characteristics such as the printer language
(PCL3 or PCL5) and color printing capability. Once you have set
these characteristics to match those of your connected printer, the
printer setup process is complete. As long as
selected in the
Printer Type menu,the analyzer will not attempt to
Custom remains
automatically identify the connected printer when the front panel
Print key is pressed.
• The
Print Setup menu will be displayed with the None key selected
and the following message will appear in the display status line:
Unsupported printer, Printer Type set to None
This indicates that the analyzer has successfully identified the
connected printer, but the printer is not supported by the
analyzer. As long as
None is selected in the Printer Type menu, the
analyzer will respond to any print command by displaying the
message Printer Type is None in the display status line.
1-10Chapter1
Preparing for Use and Printing
Printer Setup and Operation
Testing Printer Operation
When you have completed the analyzer’s printer setup, press Print
Setup, Print (Screen) and then press Print on the front panel. If the
printer is ready and the printer setup was successful, a printout of the
analyzer display will be printed. If the printer is not ready, the message
Printer Timeout will appear on the analyzer display. Printer
Timeout will remain on the display until the printer is ready or until
you press
ESC to cancel the printout request.
Chapter 11-11
Preparing for Use and Printing
Printer Setup and Operation
1-12Chapter1
2Getting Started
This chapter introduces you to basic features of the analyzer, including
front panel and rear panel descriptions, and an overview of the display
annotation. You will also learn how to make a basic measurement.
2-1
Getting Started
What You’ll Find in this Chapter
What You’ll Find in this Chapter
This chapter introduces the basic functions of the analyzer. In this
chapter you will:
• Get acquainted with the front-panel and rear-panel features.
• Get acquainted with the function keys.
• Learn about display annotation.
• Make a basic measurement (using the internal alignment signal).
• Learn about file menus.
• Learn about analyzer accuracy and the internal alignment process.
• Learn about the analyzer battery for retaining internal memory.
NOTEBefore using your analyzer, please read Chapter 1, Preparing for Use
and Printing, which describes how to set up your analyzer and how to
verify that it is operational. Chapter 1 describes many safety
considerations that should not be overlooked.
2-2Chapter2
Front-Panel Features
Figure 2-1Front Panel Feature Overview
Getting Started
Front-Panel Features
1Viewing Angle keys allow you to adjust the display so
that it can be optimally viewed from different angles.
2Esc. The Esc (escape) key cancels any entry in progress.
Esc will abort a print (if one is in progress) and clear
error messages from the status line at the bottom of the
display. It also clears input and tracking generator
overload conditions.
3Menu keys are the unlabeled keys next to the screen.
The menu key labels are the annotation on the screen
next to the unlabeled keys. Most of the labeled keys on
the analyzer front panel (also called front panel keys)
access menus of keys having related functions.
4FREQUENCY Channel, SPAN X Scale, and AMPLITUDE Y
Scale are the three large keys that activate the primary
analyzer functions and access menus of related
functions.Thesecondarylabels on these keys (Channel,
Chapter 22-3
Getting Started
Front-Panel Features
X Scale, and Y Scale) are used in some measurements
5CONTROL functions access menus that allow you to
adjust the resolutionbandwidth,adjustthesweeptime,
and control the instrument display. They also set other
analyzer parameters needed for making
measurements.
6MEASURE accesses a menu of keys that automate some
common analyzer measurements. Once a measurement
is running,
for defining your measurement.
Restart access additional measurement control
Meas Setup accesses additional menu keys
Meas Control and
functions.
7SYSTEM functions affect the state of the entire
spectrum analyzer. Various setup and alignment
routines are accessed with the
System key.
The green
Preset key resets the analyzer to a known
state.
The File key menu allows you to save and load traces,
states, limit-line tables, and amplitude correction
factors to or from analyzer memory or the floppy disk
drive. The
function defined under
Save key immediately executes the Save
File in the Front-Panel Key
Reference chapter.
The Print Setup menu keys allow you to configure
hardcopy outputs. The
Print key immediately sends
hardcopy data to the printer. See Chapter 6 for more
details.
8MARKER functions control the markers, read out
frequencies and amplitudes along the analyzer trace,
automatically locate the signals of highest amplitude,
and access functions like
9The Media Door on the right side of the front panel
Marker Noise and Band Power.
accesses the 3.5 inch disk drive and the Earphone
connector. The earphone connector provides a
connection for an earphone jack which bypasses the
internal speaker.
10The Data Control Keys, which include the step keys,
knob, and numeric keypad, allow you to change the
numeric value of an active function.
Data control keys are used to change values for
functions such as center frequency, start frequency,
resolution bandwidth, and marker position.
The data controls will change the active function in a
2-4Chapter2
Getting Started
Front-Panel Features
manner prescribed by that function. For example, you
can change center frequency in fine steps with the
knob, in discrete steps with the step keys, or to an exact
value with the numeric keypad.
The Knob allows continuous change of functions such
as center frequency, reference level, and marker
position. It also changes the values of many functions
that change in increments only. Clockwise rotation of
the knob increases values. For continuous changes, the
extent of alteration is determined by the size of the
measurement range; the speed at which the knob is
turned affects the rate at which the values are changed.
Among other things, the knob enables you to change
the center frequency, start or stop frequency, or
reference level. For slow sweeps, the analyzer uses a
smooth panning feature which is designed to move the
trace display to the latest function value as the knob is
turned. When either center frequency or reference level
is adjusted, the signal will shift right or left or up or
down with the rotation of the knob before a new sweep
is actually taken. An asterisk is placed in the message
block (the upper right-hand corner of the analyzer
display) to indicate that the data on the screen does not
reflect data at the current setting.
The Numeric Keypad allows entry of exact values for
many of the analyzer functions. You may include a
decimal point in the number portion. If not, the decimal
point is placed at the end of the number.
Numeric entries must be terminated with a units key.
When a numeric entry is begun, the menu keys show
the units key labels. The units keys change depending
on what the active function is. For example, the units
keys for frequency span are
whereas the units for reference level are
mV, and µV.
NOTEIf an entry from the numeric keypad does not coincide with an allowed
GHz, MHz, kHz, and Hz,
+dBm, −dBm,
function value (for example, that of a 12 MHz bandwidth), the analyzer
defaults to the nearest allowable value.
The Step Keys (⇓⇑) allow discrete increases or
decreases of the active function value. The step size
depends upon the analyzer measurement range or on a
preset amount. Each press results in a single step
change. For those parameters with fixed values, the
next value in a sequence is selected each time a step
key is pressed. Changes are predictable and can be set
Chapter 22-5
Getting Started
Front-Panel Features
for some functions. Out-of-range values or
out-of-sequence values will not occur using these keys.
11VOLUME. The VOLUME knob adjusts the volume of the
internal speaker. The speaker is turned on and off with
the
Speaker On Off key in the Det/Demod menu.
12EXT KEYBOARD. The EXT KEYBOARD connector is a
6-pin mini-DIN connector for future use with PC
keyboards. It is not currently supported.
13PROBE POWER provides power for high-impedance ac
probes or other accessories.
14Return. The Return key accesses the previously selected
menu. Continuing to press
Return accesses earlier
menus.
15AMPTD REF OUT provides an amplitude reference signal
of 50 MHz at –20 dBm. HP E4402B, HP E4403B,
HP E4404B, HP E4405B, HP E4407B, and
HP E4408B only
16Tab Keys are used to move around in the Limit editor
and the Correction editor, and to move within the fields
of the dialog box accessed by the
17INPUT 50Ω (INPUT 75Ω for Option 1DP) is the signal
File menu keys.
input for the analyzer.
18The Next Window key can be used to select the active
window in functions which support split-screen display
modes, such as Zone markers. In such modes, pressing
Zoom allows you to switch between the split-screen and
full-sized display of the active window.
19Help. Press the Help key and then any front panel or
menu key to get a short description of the key function
and the associated SCPI command. The next key you
press will remove the help window from the display.
2-6Chapter2
Getting Started
Front-Panel Features
20RF OUT 50Ω (for Option 1DN) or RF OUT 75Ω
(for Option 1DQ) is the source output for the built-in
tracking generator. Option 1DN or 1DQ only
CAUTIONIf the tracking generator output power is too high, it may damage the
device under test. Do not exceed the maximum power that the device
under test can tolerate.
21The ❙ (On) key turns the analyzer on, while the O
(Standby) key turns most of the analyzer off. An
instrument alignment is performed (if
Auto Align is on)
every time the analyzer is turned on. After turning on
the analyzer, allow 5 minutes of warm-up time to
ensure the analyzer will meet all specifications.
NOTEThe instrument continues to draw power even if the line power switch
is in standby. The detachable power cord is the instrument
disconnecting device. It disconnects the mains circuits from the mains
supply before other parts of the instrument. The front-panel switch is
only a standby switch and is not a LINE switch (disconnecting device).
Chapter 22-7
Getting Started
Rear-Panel Features
Rear-Panel Features
Figure 2-2Rear Panel Feature Overview
1Power input is the input for the AC line power source.
Make sure that the line-power source outlet has a
protective ground contact.
2DC Power is the input for the DC power source. Refer to
“Power Requirements” in the Specifications Chapter of
the HP ESA Spectrum Analyzer CalibrationGuide.
CAUTIONAC line power and DC power should not be plugged in simultaneously.
3Line Fuse. The fuse is removed by twisting
counterclockwise 1/4 turn. Replace only with a fuse of
the same rating. See the label on the rear panel.
2-8Chapter2
Getting Started
Rear-Panel Features
4Service Connector. The service connector is for service use
only.
5Inputs/Outputs
5aVGA OUTPUT drives an external VGA
compatible monitor with a signal that
has 31.5 kHz horizontal, 60 Hz vertical
synchronizing rate, non-interlaced.
5bGATE/HI SWP OUT (TTL) indicates when
the analyzer is sweeping.
5cGATETRIG/EXT TRIG IN(TTL) accepts the
positive edge of an external voltage
input that triggers the analyzer
internal sweep source or the gate
function (Time Gate, Option 1D6).
Table2-1 and Table2-2 show the appropriate rear panelslotsto be used for the
optional cards available with the HP ESA spectrum analyzers. Refer to Table
2-1 if you have an HP ESA-L Series spectrum analyzer. Refer to Table 2-2 if
you have an HP ESA-E Series spectrum analyzer.
1. The Frequency Extension Assembly comes standard with the HP E4408B.
Serial
(Opt 1AX)
FADC
(Opt AYX)
IF and
Sweep
Ports
(Opt A4J)
FM Demod
(Opt BAA)
Frequency
Extension
1
Chapter 22-9
Getting Started
Rear-Panel Features
Table 2-2HP ESA-E Series (E4401B, E4402B, E4404B, E4405B, E4407B)
Slot #HP-IB
(Opt A4H)
2
1
2AA A AA–
3AA PAA–
4AA A AP–
5–– – P A–
6–– – A AP
1. The Frequency Extension Assembly comes standard with the HP E4404B, E4405B and E4407B.
2. The CPU heatsink invades the space allocated to Slot 1. Cards installed in this space must be“L”
shaped to avoid interference.
PP – A––
6HP-IB and parallel (Option A4H) are optional
Serial
(Opt 1AX)
FADC
(Opt AYX)
IF and
Sweep
Ports
(Opt A4J)
FM Demod
(Opt BAA)
Frequency
Extension
1
interfaces. HP-IB supports remote instrument
operation. The parallel port is for printing only.
7RS-232 and parallel (Option 1AX) are optional
interfaces. RS-232 supports remote instrument
operation. The parallel port is for printing only.
NOTEPrinting is only supported from the parallel port.
NOTEOnly one optional interface (Option A4H or Option 1AX) can be
installed at a time.
8IF and Sweep Ports (Option A4J):
SWP OUT provides a voltage ramp corresponding to the
sweep of the analyzer (0 V to 10 V).
HI SWP IN (TTL) can be grounded to stop sweeping.
HI SWP OUT (TTL) indicates when the analyzer is
sweeping.
AUX VIDEO OUT provides detected video output (before
the analog-to-digital conversion) proportional to
vertical deflection of the trace. Output is from 0 V to
1 V. Amplitude-correction factors are not applied to this
signal. The output signal will be blanked occasionally
during retrace by the automatic alignment routine.
Select a very long sweep time to minimize this, or turn
off the
Auto Align, All function (and use Align Now, All
manually to maintain calibration.) Refer to the
Alignments key description in Chapter 6 for more
2-10Chapter2
Getting Started
Rear-Panel Features
information on alignment key functions.
AUX IF OUT is a 50 Ω, 21.4 MHz IF output that is the
down-converted signal of the RF input of the analyzer.
Amplitude-correction factors are not applied to this
signal. This output is taken after the resolution
bandwidth filters and step gains and before the log
amplifier. The output signal will be blanked
occasionally during retrace by the automatic alignment
routine. Select a very long sweep time to minimize this,
or turn off the
Now, All manually to maintain calibration.) Refer to the
Alignments key description in Chapter 6 for more
Auto Align, All function (and use Align
information on alignment key functions.
9FM Demod (Option BAA) allows you to demodulate,
display, and measure deviation on FM signals. You can
listen to audio signals on a built-in speaker or with an
earphone.
10Frequency Extension Assembly controls the
microwave front-end components in the HP E4404B,
E4405B, E4407B and E4408B.
11Card Slot Identification Numbers. Refer to
Table 2-1 and Table 2-2 for card slot versus option card
compatibility information.
1210 MHz REF IN accepts an external frequency source to
provide the 10 MHz, −15 to +10 dBm frequency
reference used by the analyzer.
1310 MHz REF OUT provides a 10 MHz, 0 dBm minimum,
timebase reference signal.
14Power On Selection selects an instrument power
preference. This preference applies after power has
been absent for > 20 seconds. The
PWR NORM position
causes the instrument to remain off when power is
applied. The
on. The
PWR ALWAYS ON position causes it to turn
PWR ALWAYS ON mode is useful if an external
power switch is used to control a rack of several
instruments.
15DC Fuse protects the analyzer from drawing too much
DC power. Replace only with a fuse of the same rating.
See the label on the rear panel.
Chapter 22-11
Getting Started
Display Annotation
Display Annotation
Here is an example of the annotation that may appear on an analyzer
display. The display annotation is referenced by numbers which are
listed in the following table. The Function Key column indicates which
key activates the function related to the annotation. Refer to
Chapter 6 for more information on a specific function key.
Figure 2-3Screen Annotation
27
26
25
24
23
22
21
20
19
3456987
2
1
10
1211
13
1817141516
Table 2-3Screen Annotation
ItemDescriptionFunction Key
1Detector modeDetector
2Reference levelRef Level
3Active function blockRefer to the description of the
4Screen titleChange Title
5Time and date displayTime/Date On Off
2-12Chapter2
pl727
activated function.
Table 2-3Screen Annotation
ItemDescriptionFunction Key
6RF attenuationAttenuation Auto Man
7Marker frequencyMarker Count On Off
8Marker amplitudeMarker
9HP-IB annunciatorsRLTS
10Data invalid indicatorSweep (Single)
Getting Started
Display Annotation
See below for more information
11Pop-up Informational
messages
12Key menu titleDependent on key selection.
13Key menuSee key label descriptions in
14Frequency span or stop
frequency
15Sweep timeSweep Time Auto Man
16Video bandwidthVideo BW Auto Man
17Frequency offsetFreq Offset
18Display status lineDisplays instrument status and
19Resolution bandwidthResolution BW Auto Man
20Center frequencyor start
frequency
21Auto alignment routine
is on
See Chapter 4, “Troubleshooting”
Chapter 6.
Span or Stop Freq
error messages.
See Chapter 4, “Troubleshooting”
Center Freq or Start Freq
Auto Align
See below for more information
22Trigger/SweepTrig, Sweep
See below for more information
23Trace modeTrace
24Video averageVideo Average On Off
25Display lineDisplay Line On Off
26Amplitude offsetRef Lvl Offst
27Amplitude scaleScale Type Log Lin
Item 21 refers to the auto alignment mode. AA indicates that auto
alignment of all analyzer parameters, except the tracking generator
and FM demodulation options, will occur. AB indicates that auto
Chapter 22-13
Getting Started
Display Annotation
alignment of all analyzer functions except the RF section (and tracking
generator and FM demodulation options) will occur. No indicator will
appear if auto alignment is off.
Item 22 refers to the trigger and sweep modes of the analyzer. The first
letter F indicates the spectrum analyzer is in free-run trigger mode.
The second letter C indicates the spectrum analyzer is in
continuous-sweep mode.
Item 23 refers to the trace modes of the analyzer. The first letter W
indicates that the analyzer is in clear-write mode. The second letter is
1, representing trace 1. The trace 2 trace mode is S2, indicating trace 2
(2) is in the store-blank mode (S). The trace mode annotation for trace 3
is displayed under the trace mode annotation of trace 1. The trace 3
trace mode is S3, indicating trace 3 (3) is in the store blank mode (S).
A # in front of display annotation indicates that the function is
uncoupled.
Refer to the following table for the screen annotation codes for trace,
trigger, and sweep modes.
The keys labeled FREQUENCY Channel, System, and Marker are all
examples of front panel keys. Pressing most front panel keys accesses
menus of functions that are displayed along the right side of the
display. These are called menu keys.
Menu keys list functions other than those accessed directly by the front
panel keys. To activate a menu key function, press the key immediately
to the right of the annotation on the screen. The menu keys that are
displayed depend on which front panel key is pressed and which menu
level is enabled.
If a menu key function’s value can be changed, it is called an active
function. The function label of the active function is highlighted after
that key has been selected. Forexample,press
calls up the menu of related amplitude functions. Note the function
labeled
highlighted.
Ref Level (the default selected key in the Amplitude menu) is
Ref Level also appears in the active function block,
indicating that it is the active amplitude function and can now be
changed using any of the data entry controls.
AMPLITUDE Y Scale. This
A menu key with On and Off in its label can be used to turn the menu
key’s function on or off. To turn the function on, press the menu key so
that On is underlined. To turn the function off, press the menu key so
that Off is underlined. In the manual, when On should be underlined, it
will be indicated as
Function (On).
A function with Auto and Man in the label can either be auto-coupled or
have its value manually changed. The value of the function can be
changed manually using the numeric keypad, knob, or step keys. To
auto-couple a function, press the menu key so that Auto is underlined.
In the manual, when
Function (Auto).
Auto should be underlined, it will be indicated as
In some menus of keys, one key label will alwaysbe highlighted to show
which key has been selected. For example, when you press Det/Demod,
Demod you will access a menu of keys in which some of the keys are
grouped together by a blue bar (on analyzers with a color display) on
the left side of the menu. The
Off key, which is the Det/Demod, Demod
menu default key, will be highlighted. When you press another key
within the blue bar region, such as
AM, the highlight will move to that
key to show it has been selected.
2-16Chapter2
Getting Started
Key Overview
In some menus of keys, one key label will alwaysbe highlighted to show
which key has been selected but the menu is immediately exited when a
selection is made. For example, when you press the
Print Setup menu key) it will bring up its own menu of keys. The Portrait
Orientation key (a
key, which is the Orientation menu default key, will be highlighted.
When you press another Orientation menu key, the highlight will move
to that key to show it has been selected.
A summary of all front panel keys and their related menu keys can be
found in Chapter 6, “Front-Panel Key Reference”.
Chapter 22-17
Getting Started
Making a Measurement
Making a Measurement
Let’s begin using the analyzer by measuring an input signal. Since the
internal 50 MHz amplitude reference signal is readily available, we will
use it as our test signal.
NOTEYou cannot hurt the analyzer by using the alignment signal and
pressing any of the keys described in this section. Don’t be afraid to
play with the knob, step keys, or numeric keypad. (If you have
experimented with other keys and wish to return to a known state,
press the green Preset key)
1. First, turn the instrument on by pressing
On. Wait for the power-up
process to complete.
2. Press System, Power On/Preset, Preset (Factory).
3. Press the green
pressing
from the
NOTEFor the HP E4401B and E4411B, the above key label is Amptd Ref
(f=50 MHz).
Input, Amptd Ref Out (f=50 MHz) (On) and connecting a cable
AMPTD REF OUT to the INPUT 50 Ω.
Preset key. Turn on the internal 50 MHz signal by
4. Set the frequency.
Press the
FREQUENCY Channel key. Center appears on the left side
of the screen, indicating that the center-frequency function is active.
The
Center Freq menu key label is highlighted to indicate that center
frequency is the active function. The active function block is the
space on the screen within the graticule where the active function
messages appear. Active function values can be changed using the
knob, step keys, or numeric keypad. Set the center frequency to
50 MHz with the DATA keys by pressing 50 MHz. The knob and step
keys can also be used to set the center frequency.
5. Set the span.
Press
block, and the
SPAN X Scale. Span is now displayed in the active function
Span menu key label is highlighted to indicate it is the
active function. Reduce the span to 20 MHz by using the knob,
pressing the down key (⇓), or pressing 20 MHz.
6. Set the amplitude.
When the peak of a signal does not appear on the screen, it may be
necessary to adjust the amplitude level on the screen. Press
AMPLITUDE Y Scale. Ref Level 0.0 dBm appears in the active
function block (48.75 dBmV with Option 1DP). The
Ref Level menu
key label is highlighted to indicate that reference level is the active
2-18Chapter2
Getting Started
Making a Measurement
function. The reference level is the top graticule line on the display
and is set to 0.0 dBm (48.75 dBmV with Option 1DP). Changing
the value of the reference level changes the amplitude level of the
top graticule line.
If desired, use the reference level function to place the signal peak at
the top of the screen using the knob, step keys, or numeric keypad.
Figure 2-4 demonstrates the relationship between center frequency
and reference level. The box in the figure represents the analyzer
display. Changing the center frequency changes the horizontal
placement of the signal on the display. Changing the reference level
changes the vertical placement of the signal on the display.
Increasing the span increases the frequency range that appears
horizontally on the display.
NOTEAnalyzers with Option 1DP default to display amplitude values in
dBmV when in a log scale mode.
Figure 2-4Relationship between Frequency and Amplitude
7. Set the marker.
Marker functions measure the frequency and amplitude of a signal.
You can place a diamond-shaped marker on the signal peak to find
the signal’s frequency and amplitude.
To activate a marker, press the
Marker key (located in the MARKER
section of the front panel). The Normal label is highlighted to show
that the marker is the active function. Turn the knob to place the
marker at the signal peak. You can also use the
Search key, which
automatically places a marker at the highest point on the trace.
Chapter 22-19
Getting Started
Making a Measurement
Readouts of marker frequency and amplitude appear in the active
function block and in the upper right corner of the display. Look at
the marker readout to determine the amplitude of the signal.
If another function is activated, the marker frequency and
amplitude can still be identified by looking at the marker readout in
the upper right corner of the display.
Measurement Summary
1. Press System, Power On/Preset, Preset (Factory). Press the green
Preset key. Turn on the internal 50 MHz signal by pressing Input,
Amptd Ref Out (f=50 MHz) (On).
NOTEFor the HP E4401B and E4411B, the above key label is Amptd Ref
(f=50 MHz).
2. Set the center frequency by pressing the following keys:
5,0, MHz.
3. Set the span by pressing the following keys:
Span, 2,0, MHz.
Frequency,
4. The amplitude reference signal is about −27 dB (25 dBmV for
Option 1DP) for the HP E4411B and E4401B. The amplitude
reference signal is about – 20 dB for the HP E4402B, E4403B,E4404B, E4405B, E4407B and E4408B. If desired, adjust the
reference level: press
AMPLITUDE Y Scale to activate the reference
level, and use the knob or step keys to change the reference level and
bring the signal to the top of the screen.
5. Determine the amplitude and frequency of the signal. You can either
press
Search or press Marker and move the marker to the signal
peak. Read the amplitude and frequency. The display should look
like the one in Figure 2-5. Frequency is displayed horizontally, and
amplitude (power) is displayed vertically.
2-20Chapter2
Figure 2-5Reading the Amplitude and Frequency
Getting Started
Making a Measurement
Chapter 22-21
Getting Started
File Menu Functions
File Menu Functions
This section describes how to use the functions located under the
front-panel
You will learn how to do all of the following:
• format a floppy disk
• save a file
• load a file
• rename a file
• copy a file
• delete a file
NOTENote that the display examples in this section are typical to the
HP E4401B, you may see some variations of the display dependent on
your analyzer.
File key.
Format a Floppy Disk
You can format a floppy disk in the analyzer. The format is MS-DOS.
Note that it is not necessary to format your floppy disk with the
analyzer; preformatted disks can be used with the analyzer.
1. Place the disk you wish to format into the floppy drive (A:\) of the
analyzer.
2. Access the file manager menu by pressing
Figure 2-6.
File, then File Manager. See
2-22Chapter2
Figure 2-6File Manager Menu
Getting Started
File Menu Functions
3. Start the format process by pressing
CAUTIONTo abort disk format press any key, but Enter.
Format, then Enter.
When a disk is formatted all data on the disk is destroyed. A dialog box
appears on the analyzer display to warn you, allowing you the
opportunity to abort formatting.
4. Press
Enter, a second time to format the disk. The format process
takes approximately three minutes.
You are now ready to save files to the disk.
Chapter 22-23
Getting Started
File Menu Functions
Save a File
You can save files (traces, states, limits, corrections, or screens) to a
floppy disk (A:\), or the internal drive (C:\). In this example you will
save a trace to the internal C: drive.
Step 1. Set up the analyzer traces
1. Turn on the internal 50 MHz reference signal of the analyzer.
For the HP E4401B and HP E4411B, use the internal 50 MHz
reference signal of the analyzer as the signal being measured. Press
Preset, Input, Amptd Ref (On).
For all other models connect a cable between the front-panel
REF OUT to the analyzer INPUT, then press Preset, Input, Amptd Ref
Out (On).
2. Set the center frequency to 75 MHz and the span to 200 MHz: press
FREQUENCY, 75 MHz, SPAN, 200 MHz.
The reference signal appears on the display.
3. View trace 1 and put it into maximum hold by pressing
Trace 1 2 3 (1 is underlined), Max Hold. Your analyzer display should
look similar to Figure 2-7.
Figure 2-7Viewing Trace 1
AMPTD
View/Trace,
2-24Chapter2
4. Activate trace 2 and put it into minimum hold by pressing
View/Trace, Trace 1 2 3 (2 is underlined), Min Hold. Your analyzer
display should look similar to Figure 2-8.
Figure 2-8Viewing Traces 1 and 2
Getting Started
File Menu Functions
5. Activate trace 3 by pressing
Clear Write. Your analyzer display should look similar to Figure 2-9.
Figure 2-9Viewing Traces 1, 2, and 3
View/Trace, Trace 1 2 3 (3 is underlined),
Chapter 22-25
Getting Started
File Menu Functions
Step 2. Save the file
1. To access the Save menu press
File, Save.
2. Select the type of file you want to save. In this example you are
saving a trace, press Trace.
3. Select the trace you wish to save (1, 2, 3, or all traces). In this
example you will save trace 3, press 3.
4. Select the type file you wish to save the trace as, trace plus state
(Trace + State) or comma separated value (CSV). In this example you
will save as trace plus state, press
Trace + State. Note that the file
extension for this file type is TRC.
5. Enter a filename using the Alpha Editor softkeys. For this example
you will name the file TRACE.
6. Select the drive you wish to save to by pressing Tab →. This will
highlight [-C-] on the analyzer display. In this example you are
saving to the C: drive. Press
NOTEIf the correct drive is not listed in the Path: field, highlight “..” at the
Select.
top of the directory list. This enables you to move up a directory. Press
Select, to highlight the desired drive, ([-A-] or [-C-])the press Select
again.
7. Press
NOTEYou need to select a unique filename. The analyzer will not allow you to
Enter, to save the file to the C: drive.
overwrite an existing file. If you select a file name that already exists,
the analyzer will display the error message: File already exists.
2-26Chapter2
Load a file
1. Reset the analyzer by pressing Preset.
Getting Started
File Menu Functions
2. To access the Load menu press
File, Load.
3. Select the type of file you want to load (setup, state, trace, limits, or
corrections). In this example you are loading a trace file, press Trace.
NOTENot all file types can be loaded back into the analyzer. For example,
bitmap Screen files and CSV (comma separated value) Trace files are
designed for use with a PC.
4. Select the drive where your file is located by pressing
Tab →. In this
example drive C:, use the front-panel knob to highlight [-C-],
then press
Select.
5. Select the file you want to load into the analyzer. In this example the
file is TRACE.TRC. by moving the cursor with the front-panel knob to
highlight the file name.
6. Press Enter to load the specified file. Yourdisplay should look similar
to Figure 2-10.
Figure 2-10File Loaded
Chapter 22-27
Getting Started
File Menu Functions
NOTEYou should notice that the trace that you saved, in this example trace 3,
is loaded to trace 1. If you wish to verify this condition, remove the
signal input. For HP E4401B and E4411B,press
Input, Amptd Ref On Off
(Off), for all other models press Input, Amptd Ref Out On Off (Off).
To see that trace 1 is in view mode blank traces 2 and 3 by pressing
View/Trace, Trace 2 , Blank, Trace 3, Blank. Then press Trace 1, notice that
trace 1 is the only trace displayed in view mode (the view key is
selected, and the trace is not sweeping). Your display should look
similar to Figure 2-11.
Figure 2-11Verifying Trace 1
2-28Chapter2
Getting Started
File Menu Functions
Rename a File
1. To access the Rename menu press File, File Manager, Rename.
2. Select the type of file you want to rename (setup, state, trace, limits,
screens, or corrections). In this example you are renaming a trace
file, press
Trace.
3. Select the drive where you file is located, by pressing
Tab →. In this
example drive C:, use the front-panel key to highlight [-C-], then
press
Select.
4. Select the file you want to rename. In this example the file is
TRACE.TRC. by moving the cursor with the front-panel knob to
highlight the file name.
5. Press Tab → to enter the Alpha Editor menu. For this example,
rename the file to RENAME using the Alpha Editor softkeys.
6. Press Enter, your file is now renamed and visible within the directory
displayed on your analyzer.
Chapter 22-29
Getting Started
File Menu Functions
Copy a File
1. To access the Copy menu press File, File Manager, Copy.
2. Put a formatted floppy in the A: drive.
3. Select the type of file you want to rename (setup, state, trace, limits,
screens, or corrections). In this example you are copying a trace file
from the C: drive to the A: drive, press Trace.
4. Select the drive where your file is located, by pressing
Tab → to
highlight the From:Path: field. Select the drive, in this example
drive C:, use the front-panel knob or step keys to highlight [-C-],
then press
Select.
5. Select the file you wish to copy by highlighting the filename using
the front-panel knob or step keys. In this example the file RENAME.
6. Press
Tab → to move to the To:Path: field. select the drive where
you want to copy the file. Select the drive, in this example drive A:,
use the front-panel knob or step keys to highlight [−Α−], then press
Select.
NOTEIf the correct drive is not listed in the Path: field, highlight “..” at the
top of the directory list. This enables you to move up a directory. Press
Select, to highlight the desired drive, ([-A-] or [-C-]) then press Select
again.
7. Copy the file by pressing
Enter. You have now copied the file
RENAME from the C: drive to the A: drive.
2-30Chapter2
Getting Started
File Menu Functions
Delete a File
1. To access the Delete menu press File, File Manager, Delete.
2. Select the type of file you want to delete (setup, state, trace, limits,
screens, or corrections). In this example you are deleting a trace file,
press
Trace.
3. Select the drive where the file you wish to delete is located, by
pressing Tab →. In this example drive C:, use the front-panel knob or
step keys to highlight [-C-], then press
NOTEIf the correct drive is not listed in the Path: field, highlight “..” at the
top of the directory list. This enables you to move up a directory. Press
Select, to highlight the desired drive, ([-A-] or [-C-])the press Select
again.
4. Select the file you want to delete. In this example the file is
RENAME.TRC, by moving the cursor with the front-panel knob or step
keys to highlight the file name.
Select.
5. Press
Enter, your file is now deleted and is no longer visible in the
directory displayed on your analyzer.
Chapter 22-31
Getting Started
Analyzer Accuracy and the Internal Alignment Process
Analyzer Accuracy and the Internal Alignment
Process
Data from the internal alignment routine is necessary for spectrum
analyzer operation. The internal alignment routine runs continuously
to ensure that the analyzer is using current alignment data that
improves the analyzer frequency and amplitude accuracy. Press the
System, Alignments keys to view the alignment menus. For more
detailed information on the alignment keys, refer to Chapter 6.
Warm-Up Time
In order for the analyzer to meet its specifications, allow it to warm up
for 5 minutes after being turned on before attempting to make any
calibrated measurements. If
analyzer specifications for the conditions required to maintain
calibration.
Auto Align Off is selected, refer to the
2-32Chapter2
Getting Started
Analyzer Battery Information
Analyzer Battery Information
The analyzer uses a lithium battery to enable the internal memory to
retain data. The date when the battery was installed is on a label on the
rear panel of the analyzer. See Figure 2-12.
The minimum life expectancy of the battery is 7 years at 25 °C,
or 1 year at 55 °C. If you experience problems with the battery or the
recommended time period for battery replacement has elapsed, see
“How to Return Your Analyzer for Service” in Chapter 4.
If you wish to replace the battery yourself, you can purchase the service
documentation that provides all necessary test and maintenance
information.
You can order the service documentation for an HP ESA spectrum
analyzer through your HP Sales and Service office. The documentation
is described under “Service Documentation and Adjustment Software
(Option 0BW)” in Chapter 7 of this manual.
After replacing the analyzer battery, write the date of battery
replacement on the rear-panel label.
Figure 2-12Rear-Panel Battery Information Label
Chapter 22-33
Getting Started
Analyzer Battery Information
2-34Chapter2
3Functional Testing
3-1
Functional Testing
What You'll Find in This Chapter
What You'll Find in This Chapter
This chapter describes the functional tests and provides information on
how to perform them.
What Are the Functional Tests?
Functional tests are tests of various instrument parameters that give a
high degree of confidence that the instrument is operating correctly.
They are recommended as a check of instrument operation for incoming
inspection or after a repair. Measurement uncertainty analysis is not
available for functional tests, and the instrument is checked against
limits that are wider than the published specifications. The functional
tests are designed to test an instrument operating within the
temperature range defined by the instrument specifications using a
minimum set of test equipment. If a test does not pass, performance
tests must be run to confirm a problem exists.
Functional Test Versus Performance Verification
Performance verification tests check a wide range of instrument
parameters and provide the highest level of confidence that the
instrument is operating satisfactorily. They are used to verify that the
instrument conforms to published specifications. They are time
consuming and require extensive test equipment. The functional tests
check a much smaller range of parameters and a limited number of
data points for each parameter. They require only limited test
equipment.
3-2Chapter3
Functional Testing
Test Descriptions
Test Descriptions
Each of the following test descriptions include the test limits (pass/fail
criteria), a description of what the test does or what it measures, a list
of equipment required for the performance of the test, an illustration of
the test setup used, and a step by step test procedure. The tests are
designed to be run on an instrument operating within the operational
temperature range defined by the instrument specifications. Only
perform tests after the specified warm-up time.
The tests included in this chapter are as follows:
Displayed Average Noise Level
Frequency Readout and Marker Count Accuracy
Frequency Response
Reference Level Accuracy
Resolution Bandwidth Switching Uncertainty
Scale Fidelity
Second Harmonic Spurious Responses
Tracking Generator Level Flatness
(HP E4401B and HP E4411B)
Tracking Generator Level Flatness
1 MHz to 500 MHzHP E4401B, E4411B– 119 dBm1
501 MHz to 1.0 GHz– 117 dBm2
1.01 GHz to 1.5 GHz– 113 dBm3
10 MHz to 1.0 GHzHP E4402B, E4403B– 117 dBm4
HP E4404B, E4405B, E4407B,
E4408B
1.01 GHz to 2.0 GHzHP E4402B, E4403B– 116 dBm6
HP E4404B, E4405B, E4407B,
E4408B
2.01 GHz to 3.0 GHzHP E4402B, E4403B– 113 dBm8
HP E4404B, E4405B, E4407B,
E4408B
3.01 GHz to 6.0 GHzHP E4404B, E4405B, E4407B,
E4408B
6.01 GHz to 6.7 GHzHP E4404B– 110 dBm11
– 116 dBm5
– 115 dBm7
– 112 dBm9
– 112 dBm10
TR Entry
6.01 GHz to 12.0 GHzHP E4405B, E4407B, E4408B– 110 dBm12
12.01 GHz to 13.2 GHzHP E4405B– 107 dBm13
12.01 GHz to 22 GHzHP E4407B, E4408B– 107 dBm14
22.01 GHz to 26.5 GHzHP E4407B, E4408B– 101 dBm15
Frequency RangeModel (75 Ω Inputs)Maximum
(75 Ω Inputs)
1 MHz to 500 MHzHP E4401B, E4411B– 65 dBmV16
501 MHz to 1.0 GHz– 60 dBmV17
1.01 GHz to 1.5 GHz– 53 dBmV18
3-4Chapter3
TR Entry
Functional Testing
Displayed Average Noise Level
Test Description
The Displayed Average Noise Level is measured within the frequency
range specified. The analyzer input is terminated into either 50 Ω or
75 Ω, depending on analyzer options.
The test tunes the analyzer frequency across the band, uses the marker
to locate the frequency with the highest response and then reads the
average noise in zero span.
Required Equipment
AdaptersRecommended Model
Termination, 50 Ω Type-N(m)HP 908A
Additional Equipment for 75 Ω Input
Termination, 75 Ω Type-N(m)HP 909E
Figure 3-1
Adapter, Type-N(f), to BNC (m), 75-ohm1250-1534
e4411
SPECTRUM ANALYZER
RF Out
Termination
Input
Termination
wl719a
Procedure (1 MHz to 500 MHz) HP E4401B and E4411B
1. Connect the equipment as shown in Figure 3-1.
2. Press
Preset on the analyzer. Wait for the preset routine to finish. Set
Single
BW/Avg, Video Average (On), 5, Enter
Single
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take 5 sweeps, then stop).
4. Press Search and record the Measured Frequency in Table 3-1 (a) for
1 MHz to 500 MHz.
5. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
6. Press Center Freq, and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (a) for 1 MHz to
500 MHz.
7. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
8. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display so that it is centered on the average trace noise,
ignoring any residual responses.
9. Record the display line amplitude setting as TR Entry 1 (TR Entry
16 for a 75 Ω Input) in Table 3-1. The Average Noise Level should be
less than the Maximum for the appropriate input impedance.
3-6Chapter3
Functional Testing
Displayed Average Noise Level
Procedure (501 MHz to 1.0 GHz) HP E4401B and
E4411B
Single
BW/Avg, Video Average (On), 5, Enter
Single
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take five sweeps, then stop).
3. Press Search and record the Measured Frequency in Table 3-1 (b) for
501 MHz to 1.0 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (b) for 501 MHz to
1.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Chapter 33-7
Functional Testing
Displayed Average Noise Level
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 2 (TR Entry
17 for a 75 Ω Input) in Table 3-1. The Average Noise Level should be
less than the Maximum for the appropriate input impedance.
Procedure (1.01 GHz to 1.5 GHz) HP E4401B and
E4411B
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take 5 sweeps, then stop).
3. Press Peak Search and record the marker frequency as the Measured
Frequency in Table 3-1 (c) for 1.01 GHz to 1.5 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (c) for 1.01 GHz to
1.5 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
3-8Chapter3
Functional Testing
Displayed Average Noise Level
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. Record the display line amplitude setting as TR Entry 3 (TR Entry
18 for a 75 Ω Input) in Table 3-1. The average noise level should be
less than the Maximum for the appropriate input impedance.
Procedure (10 MHz to 1 GHz) HP E4402B, E4403B,
E4404B, E4405B, E4407B, E4408B
1. Press Preset on the analyzer. Wait for the preset routine to finish. Set
the analyzer by pressing the following keys:
Single
BW/Avg, Video Average (On), 5, Enter
Single
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take five sweeps, then stop).
3. Press Search and record the Measured Frequency in Table 3-1 (d) for
10 MHz to 1.0 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
Chapter 33-9
Functional Testing
Displayed Average Noise Level
5. Press Center Freq and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (d) for 10 MHz to
1.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. If the analyzer is an HP E4402B or an E4403B, record the display
line amplitude setting as TR Entry 4 in Table 3-1. Otherwise, record
the display line amplitude setting as TR Entry 5 in Table 3-1. The
average noise level should be less than the Maximum.
3-10Chapter3
Functional Testing
Displayed Average Noise Level
Procedure (1.01 GHz to 2 GHz) HP E4402B, E4403B,
E4404B, E4405B, E4407B, E4408B
1. Press Preset on the analyzer. Wait for the preset routine to finish. Set
the analyzer by pressing the following keys:
Single
BW/Avg, Video Average (On), 5, Enter
Single
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take five sweeps and then stop).
3. Press Search and record the Measured Frequency in Table3-1 (e) for
1.01 GHz to 2 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (e) for 1.01 GHz to
2 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Chapter 33-11
Functional Testing
Displayed Average Noise Level
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. If the analyzer is an HP E4402B or an E4403B, record the display
line amplitude setting as TR Entry 6 in Table 3-1. Otherwise, record
the display line amplitude setting as TR Entry 7 in Table 3-1.The
average noise level should be less than the Maximum.
Procedure (2.01 GHz to 3.0 GHz) HP E4402B, E4403B,
E4404B, E4405B, E4407B, E4408B
1. Press Preset on the analyzer. Wait for the preset routine to finish. Set
the analyzer by pressing the following keys:
Single
BW/Avg, Video Average (On), 5, Enter
Single
Wait until AVG 5 is displayed to the left of the graticule (the
analyzer will take five sweeps and then stop).
3. Press Search and record the Measured Frequency in Table 3-1 (f) for
2.01 GHz to 3.0 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq and set the center frequency of the analyzer to the
Measured Frequency recorded in Table 3-1 (f) for 2.01 GHz to
3.0 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
3-12Chapter3
Functional Testing
Displayed Average Noise Level
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.
7. Press the following keys on the analyzer:
Display, Display Line (On)
Adjust the display line so that it is centered on the average trace
noise, ignoring any residual responses.
8. If the analyzer is an HP E4402B or an E4403B, record the display
line amplitude setting as TR Entry 8 in Table 3-1. Otherwise, record
the display line amplitude setting as TR Entry 9 in Table 3-1. The
Average Noise Level should be less than the maximum.
Chapter 33-13
Functional Testing
Displayed Average Noise Level
Procedure (3.01 GHz to 6.0 GHz) HP E4404B, E4405B,
E4407B, E4408B
1. Press Preset on the analyzer. Wait for the preset routine to finish. Set
the analyzer by pressing the following keys:
Marker Count
Accuracy with Counter
Resolution at 1 Hz
1.489999999 GHz1.490000001 GHz
Test Description
The frequency readout accuracy of the analyzer is tested with an input
signal of known frequency. Frequency reference error is eliminated by
using the same frequency standard for the analyzer and the
synthesized sweeper.
Chapter 33-23
Functional Testing
Frequency Readout Accuracy
Required Equipment
InstrumentCritical Specifications
Recommended Model
(for this test)
Signal Sources
Synthesized Sweeper10MHz-1.5GHz.
External Reference Input
HP 8340A/B or 836XX
Series
Adapters
Type-N(f), to APC
1250-1745
3.5(f)
Cables
Type-N,152-cm
HP 11500D
(60-in)
BNC,122-cm (48-in)HP 10503A
Additional
Equipment for
75-ohm Input
Pad, minimum lossHP 11852B
Type-N(f), to
1250-1534
BNC(m), 75 Ω
Figure 3-2
10 MHz
Ref
Input
SYNTHESIZED
SWEEPER
RF Output
ADAPTER
Type-N Cable
SPECTRUM
ANALYZER
RF Input
10 MHz
Ref
Output
wl71a
3-24Chapter3
Functional Testing
Frequency Readout Accuracy
Procedure
1. Connect the equipment as shown in Figure 3-2. Remember to
connect the 10MHz REF OUT of the analyzer to the 10MHz REF IN
of the synthesized sweeper.
2. Perform the following steps to set up the equipment:
a. Press
INSTRUMENT PRESET on the synthesized sweeper, then set
the controls as follows:
CW, 1.490 GHz
POWER LEVEL, – 10dBm
b. Press Preset on the analyzer, then wait for the preset routine to
finish. Set the analyzer by pressing the following keys:
Frequency, 1.490 GHz
SPAN, 10MHz
BW/Avg, Resolution BW Auto Man, 100kHz
Video BW Auto Man, 30kHz
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
inputs, or damage to connectors will occur.
3. Press
Search on the analyzer to measure the frequency readout
accuracy. If the instrument is functioning correctly, the result in the
active function should be between 1.48988 MHz and 1.49012 MHz.
Record the Marker Frequency Readout in Table 3-2.
4. Change the analyzer Span to 100kHz, BW/Avg, Resolution BW Auto
Man, 1 kHz, Video BW Auto Man, 1 kHz.
5. Repeat step 3. If the instrument is functioning correctly, the result
should be between 1.4899988 MHz and 1.4900012 MHz.
NOTEThe Frequency Readout Accuracy in now complete. Continue with the
Marker Count Accuracy functional check.
Chapter 33-25
Functional Testing
Marker Count Accuracy
Marker Count Accuracy
Procedure
1. Press Preset on the analyzer, then wait for the preset routine to
finish. Set the analyzer to measure the marker count accuracy by
pressing the following keys:
The output of the synthesized sweeper is fed through a power splitter to
a power sensor and the analyzer. The synthesized sweeper’s power
level is adjusted at 50 MHz to place the displayed signal at the analyzer
center horizontal graticule line. Measurements are made at various
points depending on the model being tested. The signal source
amplitude is measured with a power meter to eliminate errors due to
source flatness. The power meter is zeroed and calibrated before
starting the measurement
Chapter 33-27
.
Functional Testing
Frequency Response
Required Equipment
InstrumentCritical
Specifications
(for this test)
Signal Sources
Synthesized SweeperHP 8340A/B or 83630A
Adapters
Type-N (f) to BNC (f)1250-1474
APC 3.5 (f) to
APC 3.5 (f)
BNC(f) to SMA(m)1250-1237
Type-N(m), to APC
3.5(m)
Cables
(2) APC 3.5mm (36in)8120-4921
BNC(m) both ends,
(48in)
Recommended Model
Series
5061-5311
1250-1743
HP 10503A
Additional
Equipment for
75-ohm Input
Pad, minimum lossHP 11852B
Type-N (f), to
BNC (m)
Misc
Power MeterEPM-441A (HP E4418A)
Power Sensor, 75 ΩHP 8483A
Power Sensor, 50 ΩHP 8485A
Power Splitter11667B
1250-1534
3-28Chapter3
Figure 3-3
Functional Testing
Frequency Response
SYNTHESIZED
SWEEPER
POWER
SPECTRUM
ANALYZER
METER
RF
Output
POWER
SPLITTER
BURIED
POWER
SENSOR
Channel
B
Mechanical Adapter
(75 Inputs Only)W
Input
ADAPTER
pl728
Procedure
1. Zero and calibrate the power meter and power sensor as described in
the power meter operation manual.
2. Connect the equipment as shown in Figure 3-3 for a 50 Ω analyzer
input.
CAUTIONUse only 75 Ω cables, connectors, or adapters on instruments with 75 Ω
connectors or damage to the connectors will occur.
3. Set the synthesized sweeper controls as follows:
FREQUENCY, 50 MHz
AMPLITUDE, –8 dBm
4. Press Preset on the analyzer and wait for the preset routine to finish.
Set the analyzer by pressing the following keys.
FREQUENCY, 50 MHz
CF Step Auto Man, 50 MHz
SPAN, 20 kHz
AMPLITUDE, –10 dBm
AMPLITUDE, More, Amptd Units, dBm (75 Ω RF Input Only)
AMPLITUDE, –10 dBm, Attenuation Auto Man, 10 dB
Scale/Div, 2 dB
BW/Avg, 10 kHz
Video BW Auto Man, 3 kHz
Chapter 33-29
Functional Testing
Frequency Response
Search
FREQUENCY, Signal Track (On)
5. Adjust the synthesized sweeper amplitude for a marker amplitude
reading of –14 dBm +/– 0.10 dB.
NOTEThe power level of the synthesized sweeper remains unchanged for the
duration of the test. For each new test frequency, the power sensor cal
factor should be entered to minimize measurement errors.
6. Refer to Table 3-3, Frequency Response Report. Enter the marker
readout amplitude for 50 MHz as displayed on the analyzer in the
Analyzer Amplitude column.
7. Enter the power meter reading in the Power Meter Amplitude
column.
8. Compute the flatness error at 50 MHz using the following equation
and record the results in the Flatness Error column:
Flatness Error = Analyzer Amplitude – Power Meter Amplitude
9. Perform the following steps for each center frequency setting listed
in Table 3-3.
a. Tune the source to the next frequency listed in the Center
Frequency column.
b. Enter the power sensor cal factor for the new test frequency.
c. Tune the analyzer center frequency by pressing the ⇑ key or press
FREQUENCY, Center Freq, “n”, and MHz (where “n” is the next test
frequency in Table 3-3).
d. Press Search.
e. Enter the power meter reading in the Power Meter Amplitude
column.
f. Enter the analyzer reading in the Analyzer Amplitude column.
g. Compute the flatness error using the following equation and
record the results in the Flatness Error column:
Flatness Error = Analyzer Amplitude – Power Meter Amplitude
The flatness error should be less than the specified amount.
A 50 MHz CW signal is applied to the Input of the analyzer. The amplitude of
the source and the analyzer’sreference level are decreased in 10 dB steps. The
analyzer marker functions are used to measure the amplitude difference
between steps. Reference Level Accuracy is tested in both Log and Linear
Scale Modes. Most of the error is contributed from the output attenuator
inaccuracy in the synthesized sweeper and not the analyzer.
3-32Chapter3
Required Equipment
Functional Testing
Reference Level Accuracy
InstrumentCritical Specifications
Recommended Model
(for this test)
Signal Sources
Synthesized SweeperHP 8340A/B or 836XX
Series
Adapters
Type-N(m), to BNC(f)1250-1476
Cables
Type-N,152-cm
HP 11500D
(60-in)
BNC,122-cm (48-in)HP 10503A
Additional
Equipment for
75-ohm Input
Pad, minimum lossHP 11852B
Type-N(f), to BNC(m)1250-1534
Figure 3-4
10 MHz
Ref
Input
SYNTHESIZED
SWEEPER
RF Output
ADAPTER
Type-N Cable
SPECTRUM
ANALYZER
RF Input
10 MHz
Ref
Output
wl71a
Chapter 33-33
Functional Testing
Reference Level Accuracy
Log Mode Procedure
1. Connect the equipment as shown in Figure 3-4.
2. Press
PRESET on the synthesized sweeper. Set the synthesized
sweeper controls as follows:
CW, 50 MHz
Power Level, –30 dBm
3. Press Preset on the analyzer, then wait for the preset routine to
finish. Press
System, Alignments, Auto Align, Off. Set the analyzer by
pressing the following keys:
FREQUENCY, 50 MHz
SPAN, 50 kHz
AMPLITUDE, –30 dBm (50 Ω Input only)
AMPLITUDE, 21.76 dBmV (75 Ω Input only)
Attenuation (Man), 5 dB
BW/Avg, Resolution BW Auto Man, 3 kHz
Video BW Auto Man, 30 Hz
4. Press Search on the analyzer. Adjust the amplitude on the
synthesized sweeper until the marker amplitude on the analyzer
reads –30 dBm +/– 0.10 dB. Enter the synthesized sweeper power
level as the Synthesized Sweeper Amplitude reference in Table 3-4.
NOTEUnder these analyzer conditions, the sweep time is 1.7 seconds.
Therefore, the marker amplitude updates are fairly slow when
adjusting the synthesizer output power.
5. Now that the reference has been established in step 4, adjust the
synthesized sweeper power level and the analyzer reference level
according to Table 3-4. (The synthesized sweeper output power and
the analyzer’s reference level will be changed in 10 dB steps.)
6. On the analyzer, press
Search, Marker, Delta.
Single and wait for a sweep to finish. Press
7. For each new synthesized sweeper power level and analyzer
reference level change, press the following keys on the analyzer:
Single
Search
Record the Analyzer Marker Amplitude reading in Table 3-4.
1. Set the power level on the synthesized sweeper to –30 dBm.
2. Set the analyzer by pressing the following keys:
Sweep, Sweep (Cont)
AMPLITUDE, –30 dBm (50 Ω Input)
AMPLITUDE, +21.76 dBmV (75 Ω Input)
Scale Type (Lin)
AMPLITUDE, More 1 of 2, Amptd Units, dBm (50 Ω Input)
AMPLITUDE, More 1 of 2, Amptd Units, dBmV (75 Ω Input)
Marker, Off
3. Adjust the amplitude on the synthesized sweeper until the marker
amplitude on the analyzer reads –30 dBm +/– 0.10 dB. Enter the
synthesized sweeper power level as the Synthesized Sweeper
Amplitude reference in Table 3-5.
NOTEUnder these analyzer conditions, the sweep time is 1.7 seconds.
Therefore, the marker amplitude updates are fairly slow when
adjusting the synthesizer output power.
4. Now that the reference has been established in step 4, adjust the
synthesized sweeper power level and the analyzer reference level
Chapter 33-35
Table 3-5
Functional Testing
Reference Level Accuracy
according to Table 3-5. (The synthesized sweeper output power and
the analyzer’s reference level will be changed in 10 dB steps.)
5. On the analyzer, press Single and wait for a sweep to finish. Press
Search, Marker, Delta.
6. For each new synthesized sweeper power level and analyzer
reference level change, press the following keys on the analyzer:
Single
Search
Record the Analyzer Marker Amplitude reading in Table 3-5.
1 kHz0 (Ref)0 (Ref)
3 kHz–0.3 dB0.3 dB
10 kHz–0.3 dB0.3 dB
30 kHz–0.3 dB0.3 dB
100 kHz–0.3 dB0.3 dB
300 kHz–0.3 dB0.3 dB
1 MHz–0.3 dB0.3 dB
3 MHz–0.3 dB0.3 dB
5 MHz–0.6 dB0.6 dB
Test Description
To measure the resolution bandwidth switching uncertainty an
amplitude reference is taken with the resolution bandwidth set to
1 kHz. The resolution bandwidth is changed to settings between 5 MHz
and 3 kHz and the amplitude variation is measured at each setting
using the marker delta function and compared to the specification. The
span is changed as necessary to maintain approximately the same
aspect ratio.
Required Equipment
(No Equipment Required.)
Procedure
NOTEThe 50 MHz reference output will automatically be switched internally
on the HP E4401B and HP E4411B and will not require any external
connections. All other ESA analyzers require that the AMPTD REF
OUT be connected to the INPUT to perform this test.
Chapter 33-37
Functional Testing
Resolution Bandwidth Switching Uncertainty
1. Press Preset on the analyzer. Wait for the preset routine to finish. Set
the analyzer by pressing the following keys:
Input, Amptd Ref (f=50 MHz) (On) (HP E4401B and HP E4411B)
Input, Amptd Ref Out (f=50 MHz (On) (HP E4402B, E4403B, E4404B,
E4405B, E4407B and E4408B)
FREQUENCY, 50 MHz
SPAN, 50 kHz
AMPLITUDE, –20 dBm
Scale/Div, 1 dB
More 1 of 2, Amptd Units, dBm (75 Ω Input only)
BW/Avg, 1 kHz
Video BW Auto Man, 1 kHz
2. Press AMPLITUDE and use the Knob to adjust the reference level until
the signal appears five divisions (mid-screen) below the reference
level. Press the following keys on the analyzer:
Table 3-6
Search
Marker, Delta
FREQUENCY, Signal Track (On)
3. Set the analyzer Resolution Bandwidth and Span according to Table
3-6.
4. Press Search, then record the ∆ Mkr 1 amplitude reading in Table
3-6.
5. Repeat step 3 and 4 for each of the remaining resolution bandwidth
and span settings listed in Table 3-6. The ∆ Mkr 1 amplitude reading
should be within the range indicated in the Test Limits Table.
A 50 MHz CW signal is applied from a synthesized sweeper to the input
of the analyzer. The source is adjusted for a response at the reference
level. The synthesized sweeper amplitude is adjusted to achieve a
nominal amplitude below the reference level. The analyzer’s amplitude
marker is compared to the actual source change to determine the scale
fidelity error. Most of the error is the output source attenuator
inaccuracy from the synthesized sweeper.
3-40Chapter3
Required Equipment
Functional Testing
Scale Fidelity
InstrumentCritical Specifications
(for this test)
Recommended
Model
Signal Sources
Synthesized SweeperOutput Level Accuracy
0 to –15 dBm: ±1.0 dB
HP 8340A/B or 836XX
Series
–16 dBm to –63 dBm 1.4 dB
≤ –64 dBm: 2.0 dB
Adapters
(2) Type-N(m), to
1250-1476
BNC (f)
Cables
Type-N,152-cm
HP 11500D
(60-in)
BNC,122-cm (48-in)HP 10503A
Additional
Equipment for 75 Ω
Input
Pad, minimum lossHP 11852B
Type-N(f), to BNC
(m)
Figure 3-5
10 MHz
Ref
Input
SYNTHESIZED
SWEEPER
RF Output
SPECTRUM
ANALYZER
RF Input
ADAPTER
Type-N Cable
10 MHz
Ref
Output
Procedure
1. Connect the equipment as shown in Figure 3-5.
1250-1534
wl71a
Chapter 33-41
Functional Testing
Scale Fidelity
2. Preset the synthesized sweeper. Set the synthesized sweeper
controls as follows:
Power Level, 0 dBm (50 Ω Input)
Power Level, 4 dBm (75 Ω Input)
CW, 50 MHz
3. Press
finish. Press
Preset on the analyzer, then wait for the preset routine to
System, Alignments, Auto Align, Off. Set the analyzer by
pressing the following keys
FREQUENCY, 50 MHz
SPAN, 45 kHz
AMPLITUDE, Attenuation Auto Man, 10 dB
BW/Avg, Resolution BW Auto Man, 3 kHz
Video BW, 1 kHz
Search
4. Adjust the amplitude on the sythesized sweeper until the marker
amplitude on the analyzer reads 0 dBm +/– 0.10 dB. Record the
synthesized sweeper output level as the reference in Table 3-7.
NOTE75 Ω Input only. Adjust the synthesized sweeper amplitude until the
analyzer’s marker reads 48.8 dBmV +/– 0.10 dB.
5. On the analyzer, press the following keys:
Single
Search
Marker, Delta
6. Record the marker delta reading in Table 3-7. At each new
synthesized sweeper power level press
marker amplitude level.