Agilent E4401B Data Sheet

The ESA family of spectrum analyzers have proven and guaranteed performance with the flexibility to select the right level of functionality for your test needs. Take advantage of the best overall perfor­mance on a mid-performance spectrum analyzer.
Industry best typical performance
Warm up time: 5 minutes
Sensitivity: -166 dBm
Amplitude accuracy: ± 0.4 dB
Overall phase noise (all carrier frequencies
a
):
• -101 dBc/Hz (10 kHz)
• -122 dBc/Hz (100 kHz)
• -136 dBc/Hz (1 MHz)
Agilent ESA Series Spectrum Analyzers
Data Sheet
a. Add 20LogN for frequencies > 6.7 GHz, where N is the harmonic mixing mode.
Express analyzer configurations
Basic Analyzer Express Option BAS
Standard Analyzer Express Option STD
Communications Test Analyzer Express Option COM
Definitions and Conditions
The distinction between specifications and characteristics is described as follows.
• Specifications describe the performance of parameters covered by the product warranty. (The temperature range is 0 °C to 55 °C, unless otherwise noted.)
• Characteristics describe product performance that is useful in the application of the product, but is not covered by the product warranty.
• Typical performance describes additional product performance information that is not covered by the product warranty. It is performance beyond specification that 80% of the units exhibit with a 95% confidence level over the temperature range 20 to 30 °C. Typical performance does not include measurement uncertainty.
• Nominal values indicate the expected performance, or describe product performance that is useful in the application of the product, but is not covered by the product warranty.
• N/A (not applicable) - Not specified for this configuration
The following conditions must be met for the analyzer to meet its specifications.
• The analyzer is within the one year calibration cycle.
• If Auto Align All is selected:
• After 2 hours of storage within the operating
temperature range.
• 5 minutes after the analyzer is turned on with
sweep times less than 4 seconds.
• If Auto Align Off is selected:
• When the analyzer is at a constant temperature,
within the operating temperature range, for a minimum of 90 minutes.
• After the analyzer is turned on for a minimum
of 90 minutes and Align Now All has been run.
• When Align Now All is run:
• Every hour
• If the ambient temperature changes more than 3 °C
• If the 10 MHz reference changes
• If Auto Align All but RF is selected:
• When the analyzer is at a constant temperature, within the operating temperature range, for a minimum of 90 minutes.
• After the analyzer is turned on for a minimum of 90 minutes and Align Now RF has been run.
• When Align Now RF is run:
• Every hour
• If the ambient temperature changes more
than 3 °C
Table of Contents
Definitions and Conditions 2 Frequency Specifications 3 Amplitude Specifications 7 Tracking Generator Specifications 12 Quasi-Peak Detector Specifications 13 General Specifications 14 Option Ordering 16
2
3
a. 30 Hz characteristic. b. N = LO harmonic mixing mode.
g
Frequency Specifications
Frequency range
BAS configuration
Custom configuration N/A N/A
(75 input Option 1DP)
E4411B
E4403B E4408B
9 kHz - 1.5 GHz 9 kHz - 3 GHz 9 kHz - 26.5 GHz
1 MHz - 1.5 GHz
Frequency range
STD or COM configuration
E4402B E4404B E4405B E4407B
9 kHz - 3 GHz 9 kHz – 6.7 GHz 9 kHz – 13.2 GHz 9 kHz - 26.5 GHz
Low frequency extension Option UKB
100Hza - 26.5 GHz
Custom configuration
100 Hza - 3 GHz
100Hza - 6.7 GHz
100Hza - 13.2 GHz
External mixing
Option AYZ
Add 18 GHz - 325 GHz
Frequency range 100 Hz - 3 GHz 2.85 - 6.7 GHz 6.2 - 13.2 GHz 12.8 – 19.2 GHz 18.7 – 26.5 GHz
Band
Harmonic (Nb)
mixing mode
01234
1- 1- 2- 4- 4-
a. N = LO harmonic mixing mode. b. 20 to 30 °C. c. +5% of span + . Sweep points fixed at 401 for basic analyzer.
d. Not available in RBW < 1 kHz (Option 1DR).
Frequency Specifications
span
sweep pts. – 1
4
Communications test
Basic analyzer
Frequency reference
Frequency reference error = ± [(aging rate x time since last adjustment )+ settability + temperature stability]
Frequency readout accuracy (start, stop, center, marker) = ±(frequency indication x frequency reference error + SPc +15% of RBW + 10 Hz + 1 Hz x Na)
Aging rate ±2 x 10–6/year ±1 x 10–7/year
Temperature stability ±5 x 10–6/year
Standard analyzer
±2 x 10–6/year
±1 x 10–7/year
(Opt. 1D5)
±5 x 10–6/year
±1 x 10–8/year
(Opt. 1D5)
analyzer or ESA with
Option 1D5
b
±1 x 10–8/year
b
±5 x 10–7/year
Settability ±5 x 10–7/year ±1 x 10–8/year
±1 x 10–8/year
(Opt. 1D5)
Span coefficient (SP)
c
External reference
Marker frequency counter
d
Accuracy = ±(marker frequency x frequency reference error + counter resolution)
0.75 % x span
10 MHz
[0.5 % + 1/ (sweep
points –1) ] x span
10 MHz
[0.5 % + 1/ (sweep
points –1) ] x span
1 - 30 MHz
Counter resolution = selectable from 1 Hz to 100 kHz
Frequency span
Range = 0 Hz (zero span), 100 Hz to maximum frequency range of the analyzer
Linear scale 1% of span
±[0.5% x span + 2 x span/(sweep points – 1)]
Accuracy
Logarithmic scale
N/A
2% of span, nominal
5
a. RBW 1 kHz, 2 sweep points. b. TV trigger available with option B7B in custom configuration for ESA-E. c. Only available for spans < 5MHz. d. For resolution bandwidths < 1 kHz or > 3 MHz, not compatible with the rms detector. e. Firmware revision A.08.00 and later.
Frequency Specifications
Sweep time and trigger
Range
Accuracy (Span = 0 Hz)
Trigger type
b
Delayed trigger range
Sweep (trace) points
Range
Span = 0 Hz
Span 100 Hz
Span = 0 Hz
Span 100 Hz
Basic analyzer
4 ms – 4000 s
4 ms – 4000 s
Free Run, Single, Line, Video, Offset, Delayed, External
N/A
401
401
Standard analyzer
or ESA with Option AYX
50 nsa – 4000 s
± 1%
Gate (1D6)
N/A
1 us to 400 s
Communications test analyzer
or ESA with Option B7D/B7E
25 nsa - 4000 s
1 ms– 4000 s
RF burst (B7E)
2 - 8192
101 - 8192
Basic analyzer Standard analyzer
Resolution bandwidths (1-3-10 sequence)
Range
d
(-3 dB) (-6 dB EMI)
1 kHz – 5 MHz
9 KHz, 120 kHz 9 KHz, 120 kHz 200 Hz, 9 kHz,120 kHz
1 kHz – 5 MHzd 1 Hz to 5 MHzd
With 1DRc (-3dB) Add 100 Hz, 300 Hz Add 10 Hz - 300 Hz
(-6 dB EMI)
With 1DR and 1D5
e
Add 200 Hz Add 200 Hz
N/A
Add 1 Hz and 3 Hz
Accuracy
1 Hz to 300 Hz 1 kHz to 3 MHz 5 MHz
± 10% ± 15% ± 30%
Selectivity (60 dB/3 dB bandwidth ratio)
100 Hz to 300 Hz 1 kHz to 5 MHz
< 15:1 synchronously tuned four poles, approximately Gaussian
< 5:1 digital, approximately Gaussian
Video bandwidths (1-3-10 sequence)
Range with 1DR
Adds 1, 3, 10 Hz for RBWs less than 1 kHz
30 Hz to 3 MHz
Communications test analyzer
or ESA with Option 1DR and 1D5
Included
Included
Loading...
+ 11 hidden pages