Agilent Technologies E364xA
Dual Output DC Power Supplies
The Agilent Technologies E3646A/E3647A (30 watt) and E3648A/E3649A
(50 watt) are high performance dual-output dual range programmable DC
power supplies with GPIB and RS-232 interfaces. The combination of benchtop and system features in these power supplies provides versatile solutions
for your design and test requirements.
Convenient bench-top features
• Dual-output dual range
• Output on/off
• High accuracy and high resolution
• Excellent load and line regulation
• Low ripple and noise
• Overvoltage protection
• Five Operating states storage
• Tracking operation
• Easy-to-use controls
• Remote voltage sensing
• Front and Rear output terminals
• Portable, ruggedized case with non-skid feet
• Error messages available on the display
Flexible system features
• GPIB (IEEE-488) and RS-232 interfaces are standard
• SCPI (Standard Commands for Programmable Instruments) compatibility
• I/O setup easily done from front-panel
• Software calibration, no internal physical adjustments
Agilent Technologies E364xA
Dual Output DC Power Supplies
The Front Panel at a Glance
1 Output1 selection Key
2 Output2 selection Key
3 Low voltage range selection Key
4 High voltage range selection Key
5 Overvoltage protection Key
6 Display limit Key
7 Resolution selection Keys
8 Voltage / Current adjust selection Key
2
9 Knob
10 Output On / Off Key
11 I/O Configuration Menu / Secure Key
12 View Menu / Calibrate Key
13 State Storage Menu / Local Key
14 Stored state Recall/Reset Menu
15 Tracking enabling/disabling Key
Front-Panel Menu/Key Reference
This section gives an overview of the front-panel keys/menus. The menus are
designed to automatically guide you through all parameters required to
configure a particular function or operation.
1 Output1 selection key Select the output1 voltage and current to be controlled and
monitored on the display.
2 Output2 selection key Select the output2 voltage and current to be controlled and
monitored on the display.
3 Low voltage range selection key Selects the low voltage range and allows its full
rated output to the output terminals.
4 High voltage range selection key Selects the high voltage range and allows its full
rated output to the output terminals.
5 Overvoltage protection key Enables or disables the overvoltage protection function,
sets trip voltage level, and clears the overvoltage condition.
6 Display limit key Shows voltage and current limit values on the display and allows
the knob adjustment for setting limit values.
7 Resolution selection keys Move the flashing digit to the right or left. Adjust the
scrolling speed of the text being displayed in the View menu.
8 Voltage/Current adjust selection key Selects the knob control function for voltage
or current adjustment.
9 Knob Increases or decreases the value of the flashing digit by turning clockwise or
counter clockwise.
10 Output On/Off key Enables or disables the power supply output. This key toggles
between on and off.
11 I/O Configuration / Secure key
/ or secures or unsecures the power supply for calibration.
12 View menu / Calibrate key
calibration string, and system firmware revision / or enables calibration mode.
13 State storage menu / Local key
a name to each of the storage locations / or returns the power supply to local mode
from remote interface mode.
14 Stored state Recall menu Recalls a stored operating state from location ‘‘1’’ through
‘‘5’’ and resets the power supply to the power-on state (*RST command).
15 Tracking enabling/disabling key Enable / disable the track mode of the outputs.
3
Configures the power supply for remote interfaces
2
Views the error codes and the text of the error message,
1
Stores up to five power supply’s states and assigns
1
The key can be used as the ‘‘Local’’ key when the power supply is in the remote
interface mode.
2
You can enable the ‘‘calibration mode’’ by holding down this key when you turn on the
power supply.
3
You can use it as the ‘‘Secure’’ or ‘‘Unsecure’’ key when the power supply is in the
calibration mode.
3
Front-Panel Voltage and Current Limit Settings
You can set th e voltage and current limit valu es from the front panel using the
following method.
Use the voltage/current adjust selection key, the resolution selection keys,
and the control knob to change the voltage and current limit values.
1
Or
2
+
Low
Or
High
+
1 Select the desired output and voltage range using the output selection keys
and voltage range selection keys after turning on the power supply.
2 Press key to show the limit values on the display.
Display
Limit
3 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired voltage limit
by turning the control knob. If the display limit times out, press key again.
4 Set the knob to current control mode by pressing key.
Voltage
Curre n t
Display
Limit
5 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired current limit
by turning the control knob.
6 Press key to enable the output. After about 5 seconds, the display will
Output
On/ Off
go to output monitoring mode automatically to display the voltage and current
at the output.
Note All front panel keys and controls can be disabled with remote interface commands.
The power supply must be in "Local" mode for the front panel keys and controls to
function.
4
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
8V*/35V**Shows the low voltage range is selected.
20V*/60**Shows the high voltage range is selected.
OUT1The output1 is selected for front-panel or remote operation.
OUT2The output2 is selected for front-panel or remote operation.
OVP1The output1 overvoltage protection function is enabled when the
OVP2The output2 overvoltage protection function is enabled when the
CALThe power supply is in calibration mode.
LimitThe display shows the limit values of voltage and current.
ERRORHardware or remote interface command errors are detected and
OFFThe output of the power supply is disabled (See page 56 for more
UnregThe output of the power supply is unregulated (output is neither CV
CVThe power supply is in constant voltage mode.
CCThe power supply is in constant current mode.
TRACKThe output1 and output2 are in track mode.
OVP1 annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
OVP2 annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
the error bit has not been cleared.
information).
nor CC).
To review the display annunciators, hold down key as you turn on
Display
Limit
the power supply.
*For E3646A/48A models. **For E3647A/49A models.
5
The Rear Panel at a Glance
1 AC inlet
2 Power-line fuse-holder assembly
3 Power-line module
• Select the GPIB or RS-232 interface (see chapter 3).
• Set the GPIB address (see chapter 3).
• Set the RS-232 baud rate and parity (see chapter 3).
6
In This Book
Quick Start Chapter 1 helps you get familiar with a few of the power supply’s
the front panel feature.
General Information Chapter 2 contains a general description of your
power supply. This chapter also provides instructions for installation of your
power supply and the output connections.
Front-Panel Operation Chapter 3 describes in detail the use of front-panel
keys and how they are used to operate the power supply from the front panel.
This chapter also shows how to configure the power supply for the remote
interface and gives a brief introduction to the calibration features.
Remote Interface Reference Chapter 4 contains reference information to
help you program the power supply over the remote interface. This chapter
also explains how to program for status reporting.
Error Messages Chapter 5 lists the error messages that may appear as you
are working with the power supply. Each listing contains information to help
you diagnose and solve the problem.
Application Programs Chapter 6 contains some remote interface
applications to help you develop programs for your application.
Tutorial Chapter 7 describes basic operation of linear power supplies and
gives specific details on the operation and use of your power supply.
Specifications Chapter 8 lists the power supply’s specifications.
Service Information Contains guidelines to return your power supply to
Agilent Technologies for servicing, procedures for verification & calibration,
and replaceable parts list and component locator diagram & schematics.
If you have questions relating to the operation of the power supply, call
1-800-452-4844 in the United States, or contact your nearest Agilent
Technologies Sales Office.
If your power supply fails within three years of purchase, Agilent will repair
or replace it free of charge. Call 1-800-258-5165 ("Express Exchange") in
the United States, or contact your nearest Agilent Technologies Sales Office.
One of the first things you will want to do with your power supply is to become
acquainted with the front panel. The exercises in this chapter prepare the
power supply for use and help you get familiar with some of its front-panel
operations.
This chapter is intended for both the experienced and the inexperienced user
because it calls attention to certain checks that should be made prior to
operation.
Throughout this chapter the key to be pressed is shown in the left margin.
16
Chapter 1 Quick Start
Preliminary Checkout
Preliminary Checkout
The following steps help you verify that the power supply is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your power supply. If
anything is missing, contact your nearest Agilent Technologies Sales Office.
One power cord for your location.
This User’s Guide.
Quick Reference Guide.
Certificate of Calibration.
2 Connect the power cord and turn on the power supply.
The front-panel display will light up briefly while the power supply performs
its power-on self-test. The GPIB address is also displayed. To review the power-
on display with all annunciators turned on, hold down as you turn on
the power supply. If the power supply does not turn on properly, see page 20.
Display
Limit
1
3 Perform a complete self-test.
The complete self-test performs a more extensive set of tests than those
performed at power-on. Hold down as you turn on the power supply and
Display
Limit
hold down the key until you hear a long beep. The self-test will begin when
you release the key following the beep.
If the self-test fails, see the Service Information for instructions on returning
the power supply to Agilent Technologies for service.
Note The power supply is shipped from the factory with a power-line cord that has a plug
appropriate for your location. Your power supply is equipped with a 3-wire grounding
type power cord; the third conductor being the ground. The power supply is grounded
only when the power-line cord is plugged into an appropriate receptacle. Do not
operate your power supply without adequate cabinet ground connection.
17
Powe r
Output
On/ Off
Chapter 1 Quick Start
Output Checkout
Output Checkout
The following procedures check to ensure that the power supply develops its
rated outputs and properly responds to operation from the front panel. For
complete performance and verification tests, refer to the Service Information.
Note: If an error has been detected during the output checkout procedures,
th e
ERROR annunciator will turn on. See "Error Messages" starting on page
121 in chapter 5 for more information.
Voltage Output Checkout
The following steps verify basic voltage functions with no load.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the outputs are
disabled (the
selected. The annunciator
indication annunciator turn on (for example,
E3646A model); and the knob is selected for voltage control.
2 Enable the outputs.
OFF annunciator turns on); its output1 and low voltage range are
OVP1, OVP2, CV, OUT1, and low voltage range
8V annunciator turns on for the
The
OFF annunciator turns off and the CV annunciator turns on. Notice that
the display is in the meter mode. ‘‘Meter mode’’ means that the display shows
the actual output voltage and current.
3 Check that the front-panel voltmeter properly responds to knob
control for both low and high voltage range.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero. The flashing
digit can be adjusted by turning the knob.
1
4 Ensure that the voltage can be adjusted from zero to the full rated
value by adjusting the knob.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the voltage.
18
Powe r
Chapter 1 Quick Start
Output Checkout
Current Output Checkout
The following steps check basic current functions with a short across the
power supply’s output.
1 Turn on the power supply.
Make sure that the output is disabled. The
2 Connect a short across (+) and (-) output terminals with an insulated
test lead.
OFF annunciator is on
1
Output
On/ Off
Display
Limit
Voltage
Curre n t
Use a wire size sufficient to handle the maximum current (See "
Table 2-1 Wire
Rating" on page 32 in chapter 2).
3 Enable the output.
The CV or CC annunciator turns on depending on the resistance of the test
lead. Notice that the display is in the meter mode.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
Adjust the voltage limit to 1.0 volt to assure CC operation. The
will turn on. To go back to normal mode, press the key again or let the
Limit annunciator will be flashing).
Display
Limit
CC annunciator
display time out after several seconds.
5 Set the knob to the current control to check that the front-panel
ammeter properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the meter
mode (the
Limit annunciator is off). Check that the ammeter responds to knob
control and the voltmeter indicates nearly zero (the voltmeter will show the
voltage drop caused by the test lead). The flashing digit can be adjusted by
turning the knob.
1
6 Ensure that the current can be adjusted from zero to the full rated
value .
7 Turn off the power supply and remove the short from the output
ter minals.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
19
Chapter 1 Quick Start
If the Power Supply Does Not Turn On
If the Power Supply Does Not Turn On
Use the following steps to help solve problems you might encounter when
turning on the instrument. If you need more help, refer to chapter 5 for
instructions on returning the instrument to Agilent Technologies for service.
1 Verify that there is ac power to the power supply.
First, verify that the power cord is firmly plugged into the power receptacle on
the rear panel of the power supply. You should also make sure that the power
source you plugged the power supply into is energized. Then, verify that the
power supply is turned on.
2 Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the power
supply is shipped from the factory. Change the voltage setting if it’s not correct.
The settings are: 100, 115, or 230 Vac.
3 Verify that the correct power-line fuse is installed.
The correct fuse is installed for your country when the power supply is shipped
from the factory. See the table below to replace the fuse for your power supply.
ModelAgilent Part Number Part Description
E3646A/47A2110-1070Fuse 2.5A T 125V for 100 and 115 Vac
E3646A/47A2110-0457Fuse 1A T 250V for 230Vac
E3648A/49A2110-0996Fuse 4A T 250V for 100 and 115 Vac
E3648A/49A2110-1068Fuse 2A T 250V for 230 Vac
See the next page if you need to change the line voltage setting and the powerline fuse.
20
Chapter 1 Quick Start
Line Voltage Conversion
Line Voltage Conversion
Warning Shock Hazard Operating personnel must not remove power supply covers.
Component replacement and internal adjustments must be made only by
qualified service personnel.
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
1 Remove AC line power.
2 Remove the cover (Refer to General Disassembly on page 174).
3 Set two sections of the line voltage selector switch on the PC board for the
desired line voltage (See Figure 1-1 below).
4 See the next page to check the rating of the power-line fuse and replace with
the correct one if necessary.
5 Replace the cover and mark the power supply clearly with a tag or label
indicating the correct line voltage and fuse that is in use.
1
(TOP VIEW)
Figure 1-1. Line Voltage Selector (set for 115Vac)
100V
11 5V
230V
21
Chapter 1 Quick Start
Line Voltage Conversion
1 Remove the power cord. Remove
the fuse-holder assembly with a flatblade screwdriver from the rear panel.
3 Replace with the correct fuse.
2 Remove the fuse-holder from the
assembly.
4 Replace the fuse-holder assembly in
rear panel.
Verify that the correct line voltage is selected and the power-line fuse is good.
22
Chapter 1 Quick Start
To Rack Mount the Instrument
To Rack Mount the Instrument
You can mount the power supply in a standard 19-inch rack cabinet using one
of three optional kits available. Instructions and mounting hardware are
included with each rack-mounting kit. Any Agilent Technologies System II
instrument of the same size can be rack-mounted beside the Agilent E3646A,
E3647A, E3648A, or E3649A.
Note: Remove the front and rear rubber bumpers before rack-mounting the
instrument.
1
FrontRear (bottom view)
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
23
Chapter 1 Quick Start
To Rack Mount the Instrument
To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and
flange kit 5063-9214.
To install one or two instruments in a sliding support shelf, order shelf 5063-9256,
and slide kit 1494-0015.
24
2
General Information
General Information
This manual describes the operation of the Agilent Technologies Model
E3646A, E3647A, E3648A, and E3649A DC power supplies.
This chapter contains a general description of your power supply. This chapter
also provides instructions for installation of your power supply and the output
connections. Unless otherwise stated, the information in this manual applies
to all the four models. This chapter is divided into the following sections:
• Safety Considerations‚ on page 27
• Options and Accessories‚ on page 28
• Description‚ starting on page 29
• Installation‚ on page 31
• Output Connections‚ on page 32
26
Chapter 2 General Information
Safety Considerations
Safety Considerations
This power supply is a Safety Class I instrument, which means that it has a
protective earth terminal. That terminal must be connected to earth ground
through a power source with a 3-wire ground receptacle.
Before installation or operation, check the power supply and review this
manual for safety markings and instructions. Safety information for specific
procedures is located at the appropriate places in this manual. See also
‘‘Safety’’ at the beginning of this manual for general safety information.
Safety and EMC Requirements
This power supply is designed to comply with the following safety and EMC
(Electromagnetic Compatibility) requirements:
• IEC 1010-1(1990)/EN 61010-1(1993) + A2 (1995): Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory Use
• CSA C22.2 No.1010.1-92: Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
• EN61326-1(1997):
EN 61000-4-2(1995): Electrostatic Discharge Requirements
EN 61000-4-3(1996): Radiated Electromagnetic Field Requirements
EN 61000-4-4(1995): Electrical Fast Transient/Burst Requirements
EN61000-4-5(1995): Surge Requirements
EN61000-4-6(1996): Conducted Radio Frequency Immunity Requirements
EN61000-4-8(1993): Magnetic Field Requirements
EN61000-4-11(1994): Voltage dips, short, interruption and var Requirement
EN 55011(1991) Group 1, Class A/CISPR 11(1990): Limits and Methods of
Radio Interference Characteristics of Industrial, Scientific, and Medical
(ISM) Radio - Frequency Equipment
• Low Voltage Directive 73/23/EEC
• EMC Directive 89/336/EEC
2
27
Chapter 2 General Information
Options and Accessories
Options and Accessories
Options
Option ‘‘0E3’’ and ‘‘0E9’’ determine which power-line voltage is selected at the
factory. The standard unit is configured for 115 Vac ± 10%, 47-63 Hz input
voltage. For more information about changing the power-line voltage setting,
see Line Voltage Conversion on page 21.
OptionDescription
0E3230 Vac ± 10%, 47-63 Hz input voltage
0E9100 Vac ± 10%, 47-63 Hz input voltage
1CMRack mount kit (Agilent part number 5063-9243)
910Extra manual (same language as the selected language manual
when you order the power supply)*
Accessories
The accessories listed below may be ordered from your local Agilent
Technologies Sales Office either with the power supply or separately.
Agilent No.Description
10833AGPIB cable, 1 m (3.3 ft.)
10833BGPIB cable, 2 m (6.6 ft.)
34398A RS-232, 9 pin (f) to 9 pin (f), 2.5 m (8.2 ft.) cable; plus 9 pin (m) to
25 pin (f) adapter
34399A RS-232 adapter kit (contains 4 adapters):
9 pin (m) to 25 pin (m) for use with PC or printer
9 pin (m) to 25 pin (f) for use with PC or printer
9 pin (m) to 25 pin (m) for use with modem
9 pin (m) to 9 pin (m) for use with modem
*To order an extra English User’s guide, order Agilent part number
E3646-90001.
28
Chapter 2 General Information
Description
Description
This power supply features a combination of programming capabilities and
linear power supply performance that makes it ideal for power systems
applications. The power supply may be programmed locally from the front
panel or remotely over the GPIB and RS-232 interfaces. This power supply has
two ranges, allowing more voltage at a lower current or more current at a lower
voltage. The output range is selected from the front panel or over the remote
interfaces.
Operational features include:
• Dual-output dual range
• Constant voltage (CV) or constant current (CC) operation
• Overvoltage protection (OVP)
• Five storage locations (1 to 5) for user-defined operating states
• Automatic turn-on self-test
• Remote sense at rear panel terminals
• User calibration from the front panel or over the remote interfaces
2
The front panel operation permits:
• Easy-to-use control features
• Output (1 or 2) and output voltage range selection
• Enabling or disabling OVP
• OVP trip level setting and condition clearing
• Setting and displaying the voltage and current limit values
• Operating state storage/recall
• Resetting the power supply to power-on state
• Returning the power supply to local mode from remote mode
• Retrieving/Scrolling error messages on the display
• Changing display mode (V-V, I-I, or V-I display)
• Viewing the errors, calibration string, or system firmware revision
• Enabling or disabling Tracking mode
• Calibrating the power supply, including changing calibration secure code
• Remote interface configuration
• Enabling or disabling the output
29
Chapter 2 General Information
Description
When operated over the remote interface, the power supply can be both a
listener and a talker. Using an external controller, you can instruct the power
supply to set the output and to send the status data back over the GPIB or
RS-232. Capabilities include the following features:
• Voltage and current programming
• Voltage and current readback
• Present and stored status hardback
• Programming syntax error detection
• Complete self-test
The front-panel VFD (Vacuum-Fluorescent Display) includes:
• Displaying actual values of output voltage and current (meter mode)
• Or displaying the limit values of voltage and current (limit mode)
• Checking the operating status from the annunciators
• Checking the type of error from the error codes (messages)
Front panel binding posts are available to connect load wires for bench
operation. Connections to the power supply’s output and to chassis ground are
made to the rear output terminals.
Warning Floating the power supply output more than ±60 Vdc from the chassis presents an
electric shock hazard to the operator. Do not float the outputs more than ±60 Vdc
when uninsulated sense wires are used to connect the (+) output to the (+) sense and
the (-) output to the (-) sense terminals on the back of the unit.
1. Float voltage +/-60 Vdc Max to (
(shorting conductors without insulation)
2. Float voltage +/-240 Vdc Max to (
(Insulated shorting conductors)
(Rear Output Terminals)
30
Chapter 2 General Information
Installation
Installation
Initial Inspection
When you receive your power supply, inspect it for any obvious damage that
may have occurred during shipment. If any damage is found, notify the carrier
and the nearest Agilent Technologies Sales Office immediately. Warranty
information is shown in the front of this manual.
Keep the original packing materials in case the power supply has to be returned
to Agilent Technologies in the future. If you return the power supply for service,
attach a tag identifying the owner and model number. Also include a brief
description of the problem.
Mechanical Check
This check confirms that there are no broken terminals or knob and that the
cabinet and panel surfaces are free of dents and scratches. Verify that the
display is not scratched or cracked.
Electrical Check
Chapter 1 describes quick operation procedure that verifies to a high level of
confidence that the power supply is operating in accordance with its
specifications. More complete verification procedures are included in the
Service Information.
2
Cooling and Location
Cooling
The power supply can operate at rated specifications within the temperature
range of 0 °C to 40 °C, Power supply loading is derated from 40 °C to 55 °C. A
fan cools the power supply by drawing air through the sides and exhausting it
out the back. Using an Agilent rack mount will not impede the flow of air.
Bench Operation
Your power supply must be installed in a location that allows sufficient space
at the sides and rear of the power supply for adequate air circulation. The
rubber bumpers must be removed for rack mounting.
Cleaning
No cleaning is required for this product. If you wish to remove dust from the
enclosure, use a dry cloth.
31
Chapter 2 General Information
Output Connections
Output Connections
Warning Before attempting to connect wires to the rear output terminals, make sure to turn
off the power supply first to avoid damage to the circuits being connected.
Front panel binding posts are available to connect load wires for bench
operation and are paralleled with the rear panel (+) and (-) connections. Both
front and rear panel terminals are optimized for noise, regulation, and transient
response as documented in chapter 8.
Available connections on the rear output terminals include the (+) and (-)
output, the (+) and (-) sense terminals, and an earth ground terminal. The rear
output terminals accept wire sizes from AWG 24 to AWG 14.
Note: When making load connections from the rear output terminals, four load
wire connection is recommended to keep good CV load regulation if carrying
full-rated current of the power supply.
Current Ratings
The following table lists the characteristics of AWG (American Wire Gage)
copper wire.
Table 2-1 Wire Rating
AWG10 121416182022242628
Suggested max imum
Curren t(amps)*
mW/ft1.001.592.534.026.3910.216.125.740.864.9
mW/m3.35.28.313.221.033.552.884.3133.9 212.9
*Single con ductor in free air at 30 °C with insulation
4025201310753.52.51.7
Warning To satisfy safety requirements, load wires must be heavy enough not to overheat when
carrying the maximum short-circuit output current of the power supply. If there is
more than one load, then any pair of load wires must be capable of safety carrying
the full-rated current of the power supply.
32
Chapter 2 General Information
Output Connections
Voltage Drops
The load wires must also be large enough to avoid excessive voltage drops due
to the impedance of the wires. In general, if the wires are heavy enough to carry
the maximum short circuit current without overheating, excessive voltage
drops will not be a problem. The voltage drops across the load wires should
be limited to less than two volts. Refer to Table 2-1 to calculate the voltage drop
for some commonly used AWG copper wire.
Load Consideration
Capacitive Loading
The power supply will be stable for almost any size load capacitance. However,
large load capacitors may cause transient response ringing. Certain
combinations of load capacitance, equivalent series resistance, and load lead
inductance may result in instability (oscillation). If this occurs, the problem
may often be solved by either increasing or decreasing the size of the capacitive
load.
A large load capacitor may cause the power supply to cross into CC or
unregulated mode momentarily when the output voltage is reprogrammed. The
slew rate of the output voltage will be limited to the current setting divided by
the total load capacitance (internal and external).
2
Inductive loading
Inductive loads present no loop stability problems in constant voltage mode.
In constant current mode, inductive loads form a parallel resonance with the
power supply’s output capacitor. Generally this will not affect the stability of
the power supply, but it may cause ringing of the current in the load.
Pulse Loading
In some applications the load current varies periodically from a minimum to
a maximum value. The constant current circuit limits the output current. Some
peak loading exceeding the current limit can be obtained due to the output
capacitor. To stay within the specifications for the output, the current limit
should be set greater than the peak current expected or the supply may go into
CC mode or unregulated mode for brief periods.
33
Chapter 2 General Information
Output Connections
Reverse Current Loading
An active load connected to the power supply may actually deliver a reverse
current to the supply during a portion of its operating cycle. An external source
can not be allowed to pump current into the supply without risking loss of
regulation and possible damage. These effects can be avoided by pre-loading
the output with a dummy load resistor. The dummy load resistor should draw
at least the same amount of current from the supply as the active load may
deliver to the supply. The value of the current for the dummy load plus the
value of the current the load draws from the supply must be less than the
maximum current of the supply.
Remote Voltage Sensing Connections
Remote voltage sensing is used to maintain regulation at the load and reduce
the degradation of regulation that would occur due to the voltage drop in the
leads between the power supply and the load.
When the power supply is connected for remote sensing, the OVP circuit senses
the voltage at the sensing points (load) and not the output terminals.
Connections between the power supply sensing and output terminals should
be removed, and using shielded two-wire cable, the power supply sensing
terminals should be connected to the load as shown in Figure 2-1. Do not use
the shield as one of the sensing conductors and the other end should be left
unconnected. Connect one end of the sensing lead shield to the chassis ground
() only. Opening a sensing lead causes the power supply output voltage to
decrease at the load leads. Observe polarity when connecting the sensing leads
to the load.
Figure 2-1. Remote Voltage Sensing Connections
34
Figure 2-2. Local Sensing Connections
Chapter 2 General Information
Output Connections
Stability
Using remote sensing under certain combinations of load lead lengths and
large load capacitances may cause your application to form a filter, which
becomes part of the voltage feedback loop. The extra phase shift created by
this filter can degrade the power supply’s stability, resulting in poor transient
response or loop instability. In severe cases, it may cause oscillations. To
minimize this possibility, keep the load leads as short as possible and twist
them together. As the sense leads are part of the power supply’s programming
feedback loop, accidental open-connections of sense or load leads during
remote sensing operation have various unwanted effects. Provide secure and
permanent connections.
CV Regulation
The voltage load regulation specification in chapter 8 applies at the output
terminals of the power supply. When remote sensing, add 5 mV to this
specification for each 1 V drop between the positive sensing point and (+)
output terminal due to the change in load current. Because the sense leads are
part of the power supply’s feedback path, keep the resistance of the sense leads
at or below 0.5
per lead to maintain the above specified performance.
2
Output Rating
The rated output voltage and current specifications in chapter 8 apply at the
output terminals of the power supply. With remote sensing, any voltage
dropped in the load leads must be added to the load voltage to calculate
maximum output voltage. The performance specifications are not guaranteed
when the maximum output voltage is exceeded. If the excessive demand on
the power supply forces the power supply to lose regulation, the
annunciator will turn on to indicate that the output is unregulated.
Output Noise
Any noise picked up on the sense leads also appears at the output of the power
supply and may adversely affect the voltage load regulation. Twist the sense
leads to minimize external noise pickup and run them parallel and close to the
load leads. In noisy environments it may be necessary to shield the sense leads.
Ground the shield at the power supply end only. Do not use the shield as one of the sense conductors.
Unreg
35
Chapter 2 General Information
Output Connections
Multiple Loads
When connecting multiple loads to the power supply, each load should be
connected to the output terminals using separate connecting wires. This
minimizes mutual coupling effects between loads and takes full advantage of
the low output impedance of the power supply. Each pair of wires should be
as short as possible and twisted or bundled to reduce lead inductance and noise
pick-up. If a shield is used, connect one end to the power supply ground
terminal and leave the other end disconnected.
If cabling considerations require the use of distribution terminals that are
located remotely from the power supply, connect output terminals to the
distribution terminals by a pair of twisted or shielded wires. Connect each load
to the distribution terminals separately.
36
3
Front-Panel Operation and Features
Front-Panel Operation and Features
So far you have learned how to install your power supply and do quick start.
During the quick start, you were briefly introduced to operating from the front
panel as you learned how to check basic voltage and current functions. This
chapter describes in detail the use of the front-panel keys and shows how they
are used to accomplish power supply operation.
This chapter is divided into the following sections:
• Front-Panel Operation Overview‚ on page 39
• Constant Voltage Operation‚ starting on page 40
• Constant Current Operation‚ starting on page 42
• Tracking Operation‚ starting on page 44
• View Menu Operation‚ starting on page 45
• Configuring the Remote Interface‚ starting on page 48
• Storing and Recalling Operating States‚ starting on page 50
• Programming Overvoltage Protection‚ starting on page 52
• Disabling the Output‚ on page 56
• System-Related Operations‚ starting on page 57
• GPIB Interface Reference‚ on page 60
• RS-232 Interface Reference‚ starting on page 61
• Calibration Overview‚ starting on page 64
Throughout this chapter the key to be pressed is shown in the left margin.
Note See ‘‘Error Messages’’, starting on page 121 in chapter 5 if you encounter any errors
during front-panel operation.
38
Chapter 3 Front-Panel Operation and Features
Front-Panel Operation Overview
Front-Panel Operation Overview
The following section describes an overview of the front-panel keys before
operating your power supply.
• The power supply is shipped from the factory configured in the front-panel
operation mode. When in this mode, the front-panel keys can be used. At
power-on, the power supply is automatically set to operate in the front-panel
operation mode and the output1 is selected for front panel operation. Press
2
key for the output2 front panel operation.
• When the power supply is in remote operation mode, you can return to frontpanel operation mode at any time by pressing the
did not previously send the front-panel lockout command. A change
between front-panel and remote operation modes will not result in a change
in the output parameters.
• When you press key (the
Display
Limit
Limit annunciator flashes), the display of
the power supply goes to the limit mode and the present limit values will
be displayed. In this mode, you can also observe the change of the limit
values when adjusting the knob. If you press the key again or let the
display time-out after several seconds, the power supply will return the
display to the meter mode (the
Limit annunciator turns off). In this mode,
the actual output voltage and current will be displayed.
• The output of the power supply can be enabled or disabled from the front
panel by pressing key. When the output is off, the
Output
On/ Off
turns on and the output is disabled.
• The display provides the present operating status of the power supply with
annunciators and also informs the user of error codes. For example, the
power supply is operating in CV mode in the 8V/3A range and controlled
from the front panel, then the
CV and 8V annunciators will turn on. If,
however, the power supply is remotely controlled, the
also turn on, and when the power supply is being addressed over GPIB
interface, the
Adrs annunciator will turn on. See ‘‘Display Annunciators’’,
starting on page 5 for more information.
• The display provides the present operating status of the power supply with
annunciators and also informs the user of error codes.
Store
(Local) key if you
Local
Display
Limit
OFF annunciator
Rmt annunciator will
3
39
Powe r
Display
Limit
Chapter 3 Front-Panel Operation and Features
Constant Voltage Operation
Constant Voltage Operation
To set up the power supply for constant voltage (CV) operation, proceed as
follows.
• Front-panel operation:
1 Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); its low voltage range is selected (annunciator
(the
for the range presently selected turns on, for example, the
on for the E3646A model); and the knob is selected for voltage control. At
power-on, the output1 is selected and the
Press to operate the power supply in the high voltage range before
proceeding to the next step. The
High
20V or 60V annunciator turns on depending
OUT1 annunciator turns on.
on which power supply you are using.
3 Set the display to the limit mode.
Notice that the
Limit annunciator flashes, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the power supply.
8V annunciator turns
Voltage
Curre n t
In constant voltage mode, the voltage values between the meter and
limit modes are the same, but the current values are not. Moreover, if the
display is in the meter mode, you cannot see the change of current limit
value when adjusting the knob. We recommend that you should set the
display to “limit” mode to see the change of current limit value in the
constant voltage mode whenever adjusting the knob.
1
4 Adjust the knob for the desired current limit.
Check that the Limit annunciator still flashes. Set the knob for current control.
The flashing digit can be changed using the resolution selection keys and the
flashing digit can be adjusted by turning the knob. Adjust the knob to the
desired current limit.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting current.
40
Chapter 3 Front-Panel Operation and Features
Constant Voltage Operation
Voltage
Curre n t
1
5 Adjust the knob for the desired output voltage.
Check that the
Limit annunciator still flashes. Set the knob for voltage control.
Change the flashing digit using the resolution selection keys and adjust the
knob for the desired output voltage.
Display
Limit
Output
On/ Off
6 Return to the meter mode.
Press or let the display time-out after several seconds to return to the
meter mode. Notice that the
“
Display
Limit
OU TPU T OFF” message.
Limit annunciator turns off and the display shows
7 Enable the output.
OFF annunciator turns off and the CV annunciator turns on. Notice that
The
the display is in the meter mode.
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify that
the
CV annunciator is lit. If the CC annunciator is lit, choose a higher current
limit.
Note During actual CV operation, if a load change causes the current limit to be exceeded,
the power supply will automatically crossover to the constant current mode at the
preset current limit and the output voltage will drop proportionately.
3
• Remote interface operation:
CURRent {<current>|MIN|MAX}Set the current
VOLTage {<voltage>|MIN|MAX}Set the voltage
OUTPut ONEnable the output
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting voltage.
41
Powe r
Display
Limit
Chapter 3 Front-Panel Operation and Features
Constant Current Operation
Constant Current Operation
To set up the power supply for constant current (CC) operation, proceed as
follows.
• Front-panel operation:
1 Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); its low voltage range is selected (annunciator
(the
for the range presently selected turns on, for example, the
on for the E3646A model); and the knob is selected for voltage control. At
power-on, the output1 is selected and the
Press to operate the power supply in the high voltage range before
proceeding to the next step. The
High
20V or 60V annunciator turns on depending
OUT1 annunciator turns on.
on which power supply you are using.
3 Set the display to the limit mode.
Notice that the
Limit annunciator flashes, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the selected supply.
8V annunciator turns
In constant current mode, the current values between the meter mode
and limit mode are the same, but the voltage values are not. Moreover, if
the display is in the meter mode, you cannot see the change of voltage
limit value when adjusting the knob. We recommend that you should set
the display to “limit” mode to see the change of voltage limit value in the
constant current mode whenever adjusting the knob.
1
4 Adjust the knob for the desired voltage limit.
Check that the Limit annunciator still flashes and the knob is selected for
voltage control. The flashing digit can be changed using the resolution keys
and the flashing digit can be adjusted by turning the knob. Adjust the knob for
the desired voltage limit.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the voltage.
42
Chapter 3 Front-Panel Operation and Features
Constant Current Operation
Voltage
Curre n t
1
5 Adjust the knob for the desired output current.
Check that the
Limit annunciator still flashes. Set the knob for current control.
Change the flashing digit using the resolution selection keys and adjust the
knob to the desired output current.
Display
Limit
Output
On/ Off
6 Return to the meter mode.
Press or let the display time-out after several seconds to return to the
meter mode. Notice that the
“
Display
Limit
OU TPU T OFF” message.
Limit annunciator turns off and the display shows
7 Enable the output.
OFF annunciator turns off and the CC annunciator turns on. Notice that
The
the display is in the meter mode.
8 Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that
the
CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage
limit.
Note During actual CC operation, if a load change causes the voltage limit to be exceeded,
the power supply will automatically crossover to constant voltage mode at the preset
voltage limit and the output current will drop proportionately.
3
• Remote interface operation:
VOLTage {<voltage>|MIN|MAX}Set the voltage
CURRent {<current>|MIN|MAX}Set the current
OUTPut ONEnable the output
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
43
Track
Chapter 3 Front-Panel Operation and Features
Tracking Operation
Tracking Operation
This power supply provides tracking outputs. In the track mode, two voltages
of the output1 and the output2 supplies track each other within the voltage
programming accuracy as described on page 159 in chapter 8 for convenience
in varying the symmetrical voltages needed by operational amplifiers and other
circuits using balanced two outputs. The track mode is always off state when
power has been off or after a remote interface reset.
• Front-panel operation:
1 Set either the output1 or the output2 supply to the desired voltage.
2 Enable the track mode.
hold_key
Hold down key until the
Track
Track annunciator turns on. For example, when
the track mode is first enabled with the output1 selected, the output2 supply
will be set to the same voltage level as the output1 supply. Once enabled, any
change of the voltage level in either the output1 or the output2 supply will be
reflected in other supply. The current limit is independently set for each of the
output1 or the output2 supply and is not affected by the track mode.
Track
3 Exit the track mode.
The
Track annunciator turns off.
Note The tracking accuracy is dependent upon the voltage programming accuracy of each
output. See ‘‘Performance Specifications’’ for more information starting on page 159
in chapter 8.
• Remote interface operation:
OUTPut:TRACk[:STATe]{OFF|ON} Disable/enable the track mode
OUTPut:TRACk[:STATe]? Queries the track mode status
44
Chapter 3 Front-Panel Operation and Features
View Menu Operation
View Menu Operation
From the View menu, you can read the Errors, Firmware Revision, and
Calibration String. In addition, you can change the front-panel display mode
to
V-V display, I-I display, or V-I display as desired. With this V-V or I-I display
mode setting, you can see voltages or currents on the output1 and the output2,
simultaneously.
• After the display mode is set to V-V or I-I display mode, you can’t change the
limit values. If you attempt to change the limit values by turning the knob
or pressing one of the following keys: , , , , or ,
the front panel display will automatically return to the normal display to
allow you to change the limit values.
• After the display mode is set to V-V or I-I display mode, all the annunciators
remain unchanged.
• To exit the view menu without any changes, press key until
CHANGE is displayed, or select EXIT by turning the knob and press
key.
Voltage
<
Curre n t
>
View
LowHigh
NO
View
3
View
View
View
Changing Display Mode
1 Enter the View Menu.
display
If
ERRORS appears when you enter the view menu, one or more command
syntax or hardware errors have been detected. Then turn the knob until
DISPLAY is displayed.
2 Select the display mode you want to display.
v-I display
Select one of the following display modes by turning the knob:
I-I DISPLAY, or V-I DISPLAY.
3 Save the change and exit the menu.
V-V DISPLAY,
45
View
Chapter 3 Front-Panel Operation and Features
View Menu Operation
Viewing the Errors
1 Press key twice to view the errors.
Total number of the errors detected will be displayed shortly before the above
message is displayed. The above number ‘‘1’’ stands for the first error in queue
and the ‘‘-103’’ is error code.
>
2 Scroll through the error numbers and view the text of the error
message by using the knob and key.
Press key to increase the scrolling speed of the text on the display.
View
1: err -103
>
>
View
View
View
3 Clear the errors and exit the menu.
All errors are cleared when you exit the menu by pressing key or let the
View
display time out for about 30 seconds.
See ‘‘Error Messages’’ for more information starting on page 121 in chapter 5.
Viewing the Firmware Revision
1 Enter the View menu and select
FW REvision
2 View the firmware revision of your power supply.
REV X.X-Y.Y-Z.Z
The first number is the firmware revision number for the main processor; the
second is for the input/output processor; and the third is for the front-panel
processor. See “Firmware Revision Query” on page 59 for more information.
FW REV ISION.
46
View
Chapter 3 Front-Panel Operation and Features
View Menu Operation
Viewing the Calibration String
1 Enter the View menu and select
cal string
CAL STRING.
View
2 Scroll through the text of the message.
cal12-24-99
NO STRING is displayed if no calibration message is stored. Press to
increase the scrolling speed.The Cal String can’t be changed from the front
panel. See “Calibration Message” on page 68 for more information.
3
>
47
I/O
Config
Chapter 3 Front-Panel Operation and Features
Configuring the Remote Interface
Configuring the Remote Interface
This power supply is shipped with both a GPIB (IEEE-4888) interface and an
RS-232 interface. The GPIB interface is selected when the power supply is
shipped from the factory. Only one interface can be enabled at a time. To exit
the I/O configuration mode without any changes, press key until the ‘‘
CHANGE’’ message is displayed.
• You can set the GPIB address, parity, and baud rate from the front panel
only.
• The current selection is highlighted for emphasis. All other choices are
dimmed.
• The interface selection is stored in non-volatile memory, and does not
change when power has been off or after a power-on reset (
command).
GPIB Configuration
1 Turn on the remote configuration mode.
I/O
Config
*RST
NO
GPIB / 488
If “
RS-232” appears, select “GPIB / 488” by turning the knob.
I/O
Config
2 Select the GPIB address.
ADDR 05
You can set the power supply’s address to any value between 0 and 30. The
factory setting is address ‘‘5’’.
I/O
Config
3 Save the change and exit the menu.
SAVED
Note Your computer's GPIB interface card has its own address. Be sure to avoid
using the computer's address for any instrument on the interface bus.
Agilent Technologies GPIB interface cards generally use address ‘‘21’’.
48
I/O
Config
I/O
Config
Chapter 3 Front-Panel Operation and Features
Configuring the Remote Interface
RS-232 Configuration
1 Turn on the remote configuration mode.
GPIB / 488
Notice that if you changed the remote interface selection to RS-232 before,
“
RS-232” message is displayed.
2 Choose the RS-232 interface.
3
RS-232
You can choose the RS-232 interface by turning the knob.
3 Select the baud rate
Select one of the following: 300, 600, 1200, 2400, 4800, or 9600 (factory setting)
baud.
I/O
Config
I/O
Config
9600 BAUD
4 Select the parity and number of stop bits.
Select one of the following: None (8 data bits, factory setting), Odd (7 data
bits), or Even (7 data bits). When you set the parity, you are also indirectly
setting the number of the data bits.
NONE 8 BITS
5 Save the change and exit the menu.
SAVED
49
Sto re
Chapter 3 Front-Panel Operation and Features
Storing and Recalling Operating States
Storing and Recalling Operating States
You can store up to five different operating state in non-volatile storage
locations. When shipped from the factory, storage locations ‘‘1’’ through ‘‘5’’
are empty. You can name a location from the front panel or over the remote
interface but you can only recall a named state from the front panel.
The following steps show you how to store and recall an operating state.
To cancel the store/recall operation, select the ‘‘
knob then press the key pressed or let the display time-out.
• Front-panel operation:
Storing Operating State
1 Set up the power supply for the desired operating state.
The storage feature “remembers” output voltage range selection, the limit
value settings of voltage and current, output on/off state, OVP on/off state, and
OVP trip levels.
2 Turn on the storage mode.
EXIT’’ menu by turning the
Sto re
STORE STATE
From the front panel, you can assign names (up to 10 characters) to each of
the five stored states. Turn the knob until the ‘‘
press to select the locations, then press to name the locations.
Sto reSto re
NAME STATE’’ appears and
Name STATE
1:p15v_test
3 Select the storage location.
Turn the knob to the right to specify the memory location 2.
2: STatE2
50
Chapter 3 Front-Panel Operation and Features
Storing and Recalling Operating States
Sto re
Recall
Recall
4 Save the operating state
DONE
Recalling a Stored State
1 Turn on the recall mode.
Memory location “1” will be displayed in the recall mode.
1: p15v_test
2 Select the stored operating state.
2: state2
reset
You can select the above
on state without turning power off/on or without using ‘‘
the remote interface. See page 92 for more details on ‘‘
3 Recall the stored operating state.
RESET mode to reset the power supply to the power-
3
*RST’’ command over
*RST’’ command.
done
• Remote interface operation:
Use the following commands to store and recall power supply state.
*SAV {1|2|3|4|5}Store an operating state to a specified location
*RCL {1|2|3|4|5}Recall a previously stored state from a specified
loc atio n
‘‘MEM:STATE:NAME 1, ‘P15V_TEST’’’
Name the storage location 1 as ‘‘P15V_TEST’’.
51
Powe r
12
Over
Voltage
or
Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
Programming Overvoltage Protection
Overvoltage protection guards the load against output voltages reaching values
greater than the programmed protection level. It is accomplished by shorting
the output via an internal SCR when the trip level is set to equal or greater than
3 volts, or by programming the output to 1 volt when the trip level is set to less
than 3 volts.
The following steps show how to set the OVP trip level, how to check OVP
operation, and how to clear overvoltage condition.
• Front-panel operation:
Setting the OVP Level and Enable the OVP Circuit
1 Turn on the power supply.
2 Select the output to be programmed.
3 Enter the OVP menu and set the desired trip level.
Over
Voltage
Over
Voltage
LEVEL 22222.0V (E3646A)
Use the knob and the resolution selection key or to set the desired
trip level. Note that you cannot set the trip levels to lower than 1.0 volt.
4 Enable the OVP circuit.
<>
OVP ON
5 Exit the OVP menu.
CHANGED
If the OVP settings are not changed, “
power supply will exit the OVP menu and the display will return to the meter
mode. Check that the
OVP1 and OVP2 annunciators turn on if the OVP circuit
is enabled.
52
NO CHANGE” will be displayed. The
Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
Checking OVP Operation
To check OVP operation, raise the output voltage to near the trip point. Then
very gradually increase the output by turning the knob until the OVP circuit
trips. This will cause the power supply output to drop to near zero, the
OVP1
or OVP2 annunciator to flash depending on the output operated, and the CC
annunciator to turn on. The “
OVP1 (or OVP 2) T RI P” message also appears on
the display.
Clearing the Overvoltage Condition
When the OVP condition occurs, the OVP1 or OVP2 annunciator flashes. When
it was caused by an external voltage source such as a battery, disconnect it
first. Clear the overvoltage condition by adjusting output voltage level or by
adjusting OVP trip level.
The following steps show you how to clear the overvoltage condition and get
back to normal mode operation. In the following steps, the display will go back
to “
OVP1 (or OVP2) TRIP” if you let the display time out after about several
seconds.
By Adjusting output voltage level
3
Display
Limit
Over
Voltage
Over
Voltage
1 Lower the output voltage level below the OVP trip point.
OVP and Limit annunciators are flashing after key is pressed.
The
Display
Limit
2 Check that you lowered the voltage level below the OVP trip point.
The OVP trip point is displayed. Do not adjust the trip point at this step.
3 Select the OVP CLEAR mode by turning the knob.
OVP
OVP ON
ON
OVPOVP
ONON
OVP
OVP CLEAR
OVPOVP
53
Over
Voltage
Over
Voltage
Over
Voltage
Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
4 Clear the overvoltage condition and exit this menu.
done
done
donedone
The annunciator for OVP operation will not flash any more. The output will
return to meter mode.
By Adjusting OVP trip level
1 Raise the OVP trip level higher than the level tripped.
2 Select the OVP CLEAR mode by turning the knob.
OVP
OVP ON
OVPOVP
ON
ONON
Over
Voltage
OVP
OVP CLEAR
OVPOVP
3 Clear the overvoltage condition and exit this menu.
done
done
donedone
The annunciator for OVP operation will not flash any more. The output will
return to the meter mode.
• Remote interface operation:
VOLT:PROT {<voltage>|MIN|MAX} Set the OVP level
VOLT:PROT:STAT {OFF|ON)Disable or enable the OVP circuit
VOLT:PROT:CLEClear the tripped OVP circuit
54
Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
Note The power supply’s OVP circuit contains a crowbar SCR, which effectively shorts the
output of the power supply whenever the overvoltage condition occurs. If external
voltage source such as a battery is connected across the output, and the overvoltage
condition inadvertently occurs, the SCR will continuously sink a large current from
the source; possibly damaging the power supply. To avoid this a diode must be
connected in series with the output as shown in Figure 3-1.
3
Figure 3-1. Recommended Protection Circuit for Battery Charging
55
Chapter 3 Front-Panel Operation and Features
Disabling the Output
Disabling the Output
The output of the power supply can be disabled or enabled from the front panel.
• When the power supply is in the “Off” state, the OFF annunciator turns on
and the output is disabled. The
supply returns to the “On” state. When the output is disabled, the voltage
value is 0 volts and the current value is 0.02 amps. This gives a zero output
voltage without actually disconnecting the output.
• The output state is stored in volatile memory; the output is always disabled
when power has been off or after a remote interface reset.
• While the output is disabled, the control knob is locked to prevent from any
unwanted changes occurring. But the other front panel keys are working.
• To lock the control knob, move the flashing digit to the right or left using
<>
or resolution selection keys until the flashing digit disappears.
To see or check the changes while the output is disabled, press before
returning to meter mode.
• Front-panel operation:
OFF annunciator turns off when the power
Display
Limit
Output
output Off
On/ Off
• Remote interface operation:
OUTP {OFF|ON}
Disconnecting the Output Using an External Relay
To disconnect the output, an external relay must be connected between the
output and the load. A TTL signal of either low true or high true is provided to
control an external relay. This signal can only be controlled with the remote
command
232 connection pin 1 and pin 0.
When the
(4.5 V) and pin 9 is low (0.5 V). The levels are reversed when the
OUTPut:RELay state is “OFF”. The TTL output of pin 1 or pin 9 of the RS-232
connector is available only after installing two jumpers (JP1 and JP2) inside
the power supply. See the Service Information to locate them.
Note Do not use the RS-232 interface if you have configured the power supply to output
relay control signals. Internal components on the RS-232 circuitry may be damaged.
56
OUTPut:RELay {OFF|ON}. The TTL output is available on the RS-
OUTPut:RELay state is “ON”, the TTL output of pin 1 is high
Chapter 3 Front-Panel Operation and Features
System-Related Operations
System-Related Operations
This section gives information on system-related topics such as storing power
supply states, reading errors, running a self-test, displaying messages on the
front panel, and reading firmware revisions.
State Storage
The power supply has five storage locations in non-volatile memory to store
power supply states. The locations are numbered 1 through 5. You can assign
a name to each of the locations for use from the front panel.
• You can store the power supply state in any of the five locations. However,
you can only recall a state from a location that contains a previously stored
state.
• The power supply stores the following: the state of output range selection,
flashing digit position on the display, the limit value settings of voltage and
current, output on/off state, OVP on/off state and trip levels.
• When shipped from the factory, storage locations ‘‘1’’ through ‘‘5’’ are empty.
• You can assign a name to the storage locations. You can name a location
from the front panel or over the remote interface but you can only recall a
named state from the front panel. From the remote interface, you can only
recall a stored state using a number (1 through 5).
• The name can contain up to 9 characters. A letter (A-Z) or numbers (0-9)
can be used for the first character to name a state. The underscore character
(‘‘_ ’’) can be used for the remaining 8 characters. Blank spaces are not
allowed. An error is generated if you specify a name with more than 10
characters.
•A power-on reset (
stored in memory. Once a state is stored, it remains until it is overwritten.
*RST command) does not affect the configurations
3
• Front-Panel Operation:
Sto re
STORE STATE, NAME STATE, EXIT
To reset the power supply to the power-on reset state without using the
*RST command or turning power off/on, select the ‘‘RESET’’ from the
following.
Recall
5 states, RESET, exit
57
Chapter 3 Front-Panel Operation and Features
System-Related Operations
• Remote Interface Operation:
Use the following commands to store and recall power supply state.
*SAV {1|2|3|4|5}
*RCL {1|2|3|4|5}
To assign a name to a stored state to be recalled from the front panel, send
the following command. From the remote interface, you can only recall a
stored state using a number (1 through 5).
‘‘
MEM:STATE:NAME 1, ‘P15V_TEST’’’
Self-Test
A power-on self-test occurs automatically when you turn on the power supply.
This test assures you that the power supply is operational. This test does not
perform the extensive set of tests that are included as part of the complete selftest described below. If the power-on self-test fails, the
turns on .
A complete self-test performs a series of tests and takes approximately 2
seconds to execute. If all tests pass, you can have a high confidence that the
power supply is operational.
If the complete self-test is successful, “
If the self-test fails, ‘‘
FAIL’’ is displayed and the ERROR annunciator turns on.
PA SS ” is displayed on the front panel.
See the Service Information for instructions on returning the power supply to
Agilent Technologies for service.
ERROR annunciator
• Front-panel operation:
To perform the complete front panel self-test, hold down the key as
you turn on the power supply and h old downthe
key until you hear a long
Display
Limit
beep. The self-test will begin when you release the key following the beep.
• Remote interface operation:
*TST?’’
‘‘
Returns “0” if the complete self-test passes or “1” if it fails.
Error Conditions
When the front-panel ERROR annunciator turns on, one or more command
syntax or hardware errors have been detected. A record of up to 20 errors can
be stored in the power supply’s error queue. See ‘‘Error Messages’’ for more
information starting on page 121 in chapter 5.
58
Chapter 3 Front-Panel Operation and Features
System-Related Operations
Firmware Revision Query
The power supply has three microprocessors for control of various internal
systems. You can query the power supply to determine which revision of
firmware is installed for each microprocessor.
• The power supply returns three revision numbers. The first number is the
firmware revision number for the main processor; the second is for the
input/output processor; and the third is for the front-panel processor.
• Front-Panel Operation:
ViewView
REV X.X-Y.Y-Z.Z
Press and turn the knob until FW REVISION appears and then press
• Remote interface operation:
*IDN?
The above command returns a string in the form:
‘‘
Be sure to dimension a string variable with at least 40 characters.
This power supply complies with the rules and regulations of the present
version of SCPI (Standard Commands for Programmable Instruments). You
can determine the SCPI version with which the power supply is in compliance
by sending a command from the remote interface.
You can query the SCPI version from the remote interface only.
• Remote interface operation:
‘‘SYST:VERS?’’Query the SCPI version
Returns a string in the form “YYYY.V” where the “Y’s” represent the year
of the version, and the “V” represents a version number for that year
(for example, 1997.0).
59
Chapter 3 Front-Panel Operation and Features
GPIB Interface Reference
GPIB Interface Reference
The GPIB connector on the rear panel connects your power supply to the
computer and other GPIB devices. Chapter 2 lists the cables that are available
from Agilent Technologies. A GPIB system can be connected together in any
configuration (star, linear, or both) as long as the following rules are observed:
Each device on the GPIB (IEEE-488) interface must have a unique address.
You can set the power supply’s address to any value between 0 and 30. The
address is set to ‘‘5’’ when the power supply is shipped from the factory. The
GPIB address is displayed at power-on.
You can set the GPIB address from the front panel only.
• The address is stored in non-volatile memory, and does not change when
power has been off or after a power-on reset (
• Your computer's GPIB interface card has its own address. Be sure to avoid
using the computer's address for any instrument on the interface bus.
Agilent Technologies GPIB interface cards generally use address ‘‘21’’.
• The total number of devices including the computer is no more than 15.
• The total length of all the cables used is no more than 2 meter times the
number of devices connected together, up to a maximum of 20 meters.
• Do not stack more than three connector blocks together on any GPIB
connector. Make sure that all connectors are fully seated and that the lock
screws are firmly finger tightened.
See page 48 for more information on configuring the power supply for remote
interface from the front panel.
*RST command).
Note IEEE-488 states that you should exercise caution if your individual cable
lengths exceed 4 meters.
60
Chapter 3 Front-Panel Operation and Features
RS-232 Interface Reference
RS-232 Interface Reference
The power supply can be connected to the RS-232 interface using the 9-pin
(DB-9) serial connector on the rear panel. The power supply is configured as
a DTE (Data Terminal Equipment) device. For all communications over the
RS-232 interface, the power supply uses two handshake lines: DTR (Data
Terminal Ready, on pin 4) and DSR (Data Set Ready, on pin 6).
The following sections contain information to help you use the power supply
over the RS-232 interface. The programming commands for RS-232 are
explained on page 97.
RS-232 Configuration Overview
Configure the RS-232 interface using the parameters shown below. Use the
front-panel key to select the baud rate, parity, and number of data bits
(See page 48 for more information).
• Parity and Data Bits: None / 8 data bits (factory setting)
• Number of Start Bits: 1 bit (fixed)
• Number of Stop Bits: 2 bits (fixed)
I/O
Config
Even / 7 data bits, or
Odd / 7 data bits
3
RS-232 Data Frame Format
A character frame consists of all the transmitted bits that make up a single
character. The frame is defined as the characters from the start bit to the last stop bit, inclusively. Within the frame, you can select the baud rate, number of
data bits, and parity type. The power supply uses the following frame formats
for seven and eight data bits.
61
Chapter 3 Front-Panel Operation and Features
RS-232 Interface Reference
Connection to a Computer or Terminal
To connect the power supply to a computer or terminal, you must have the
proper interface cable. Most computers and terminals are DTE (Data Terminal Equipment) devices. Since the power supply is also a DTE device, you must
use a DTE-to-DTE interface cable. These cables are also called null-modem,
modem-eliminator, or crossover cables.
The interface cable must also have the proper connector on each end and the
internal wiring must be correct. Connectors typically have 9 pins (DB-9
connector) or 25 pins (DB-25 connector) with a “male” or “female” pin
configuration. A male connector has pins inside the connector shell and a
female connector has holes inside the connector shell.
If you cannot find the correct cable for your configuration, you may have to
use a wiring adapter. If you are using a DTE-to-DTE cable, make sure the
adapter is a “straight-through” type. Typical adapters include gender changers,
null-modem adapters, and DB-9 to DB-25 adapters.
The cable and adapter diagrams shown below can be used to connect the
power supply to most computers or terminals. If your configuration is different
than those described, order the Agilent 34399A Adapter Kit. This kit contains
adapters for connection to other computers, terminals, and modems.
Instructions and pin diagrams are included with the adapter kit.
DB-9 Serial Connection If your computer or terminal has a 9-pin serial
port with a male connector, use the null-modem cable included with the
Agilent 34398A Cable Kit. This cable has a 9-pin female connector on each
end. The cable pin diagram is shown below.
5182-4794
Instrument
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
1
2
3
4
5
6
7
8
9
DB9
Female
Cable
1
2
3
4
5
6
7
8
9
DB9
Female
PC
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
62
Chapter 3 Front-Panel Operation and Features
RS-232 Interface Reference
DB-25 Serial Connection If your computer or terminal has a 25-pin serial
port with a male connector, use the null-modem cable and 25-pin adapter
included with the Agilent 34398A Cable Kit. The cable and adapter pin
diagram are shown below.
5182-4794
InstrumentPC
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
1
2
3
4
5
6
7
8
9
DB9
Fem ale
Cable
1
2
3
4
5
6
7
8
9
DB9
Fem ale
1
2
3
4
5
6
7
8
9
DB9
Male
5181-6641
Adapter
2
3
4
5
6
7
8
20
DB25
Fem ale
TX
RX
RTS
CTS
DSR
GND
DCD
DT R
DB25
Male
RS-232 Troubleshooting
Here are a few things to check if you are having problems communicating over
the RS-232 interface. If you need additional help, refer to the documentation
that came with your computer.
• Verify that the power supply and your computer are configured for the same
baud rate, parity, and number of data bits. Make sure that your computer is
set up for 1 start bit and 2 stop bits (these values are fixed on the power
supply).
• Make sure to execute the
supply in the remote mode.
• Verify that you have connected the correct interface cable and adapters.
Even if the cable has the proper connectors for your system, the internal
wiring may be incorrect. The Agilent 34398A Cable Kit can be used to
connect the power supply to most computers or terminals.
• Verify that you have connected the interface cable to the correct serial port
on your computer (COM1, COM2, etc.).
SYSTem:REMote command to place the power
3
63
Chapter 3 Front-Panel Operation and Features
Calibration Overview
Calibration Overview
This section gives an overview of the calibration features of the power supply.
For more detailed discussion of the calibration procedures, see the Service Information.
Calibration Security
This feature allows you to enter a security code to prevent accidental or
unauthorized calibrations of the power supply. When you first receive your
power supply, it is secured. Before you can calibrate the power supply, you
must unsecure it by entering the correct security code.
• The table 3-1 below shows the security code for each model when the power
supply is shipped from the factory. The security code is stored in non-volatile memory, and does not change when power has been off or after a
power-on reset (
• The security code may contain up to 11 alphanumeric characters or
underscore character (‘‘_’’). A letter (A-Z) or number (0-9) can be used for
the first character. You do not have to use all 11 characters.
*RST command).
_ _ _ _ _ _ _ _ _ _ _ (11 characters)
• When you secure the power supply from the remote interface, use maximum
8 alphanumeric characters to unsecure the power supply from the front
panel also. For example,
e3646A (less than 9 characters)
If you forget your security code, you can disable the security feature by
adding a jumper inside the power supply, and then entering a new code.
See the Service Information for more information.
Table 3-1. Factory setting security codes
ModelSecurity CodeModelSecurity Code
E3646A003646E3647A003647
E3648A003648E3649A003649
64
Powe r
View
Calibrate
Chapter 3 Front-Panel Operation and Features
Calibration Overview
To Unsecure for Calibration
You can unsecure the power supply either from the front panel or over the
remote interface. The power supply is secured when shipped from the factory.
See the table 3-1 for the factory setting secure code for your power supply.
1 Select the calibration mode.
SECURED
I/O
Config
Secure
I/O
Config
Secure
Powe r
If the power supply is secured, the above message is displayed as you turn on
the power supply by holding down (
key until you hear a long beep. And a message ‘‘
View
Calibrate) key and hold down the
Calibrate
CAL MODE’’ is displayed.
2 Enter the security code.
000000
Enter the security code using the control knob and resolution selection keys.
3 Save the change and exit the menu.
UNSECURED
You will see the above message if the security code is correct. And a message
‘‘
CAL MODE’’ is displayed. To exit the calibration mode, turn the power off
and on.
Notice that if you enter the wrong secure code, “INVALID” is displayed and
the code entering mode is displayed for you to enter the correct code.
• Remote Interface Operation:
3
CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply
To unsecure the power supply, send the above command with the same
code used to secure. For example,
‘‘
CAL:SEC:STAT OFF, ‘003646’’’(E3646A model)
65
Powe r
View
Calibrate
Chapter 3 Front-Panel Operation and Features
Calibration Overview
To Secure Against Calibration
You can secure the power supply against calibration either from the front panel
or over the remote interface. The power supply is secured when shipped from
the factory.
Be sure to read the security code rules on page 64 before attempting to secure
the power supply.
• Front-Panel Operation:
1 Select the calibration mode.
UNSECURED
I/O
Config
Secure
I/O
Config
Secure
Powe r
If the power supply is unsecured, the above message is displayed as you turn
on the power supply by holding down (
key until you hear a long beep. And a message ‘‘
View
Calibrate) key and hold down the
Calibrate
CAL MODE’’ is displayed.
2 Enter the security code.
000000
Enter the desired security code using the control knob and resolution selection
keys.
3 Save the change and exit the menu.
secured
The secured setting is stored in non-volatile memory, and does not change
when power has been off or after a power-on reset (
• Remote Interface Operation:
CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply
To secure the power supply, send the above command with the same code
as used to unsecure. For example,
‘‘
CAL:SEC:STAT ON, ‘003646’’’(E3646A model)
*RST command).
66
Chapter 3 Front-Panel Operation and Features
Calibration Overview
To Change the Security Code To change the security code, you must first
unsecure the power supply, and then enter a new code. Be sure to read the
security code rules on page 64 before attempting to secure the power supply.
• Front-Panel Operation:
To change the security code, first make sure that the power supply is
unsecured. Go to the security code entry, press (
“
CAL MODE” message is displayed, enter the new security code using the
control knob and resolution selection keys, then press (
I/O
Config
Secure) key after the
Secure
I/O
Config
Secure) key.
Secure
Changing the code from the front panel also changes the code as seen from
the remote interface.
• Remote Interface Operation:
CAL:SEC:CODE <new code>Change the security code
To change the security code, first unsecure the power supply using the old
security code. Then, enter the new code as shown below.
‘‘CAL:SEC:STAT OFF, ‘003646’’’ Unsecure with old code
‘‘CAL:SEC:CODE ‘ZZ001443’’’Enter new code
‘‘CAL:SEC:STAT ON, ‘ZZ001443’’’Secure with new code
Calibration Count
You can determine the number of times that your power supply has been
calibrated. Your power supply was calibrated before it left the factory. When
you receive your power supply, read the count to determine its initial value.
The calibration count feature can be performed from the remote interface
only.
• The calibration count is stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset.
• The calibration count increments up to a maximum of 32,767 after which it
wraps-around to 0. Since the value increments by one for each calibration
point, a complete calibration will increase the value by 5 counts.
3
• Remote Interface Operation:
CAL:COUN?Query the number of times of calibration
67
Chapter 3 Front-Panel Operation and Features
Calibration Overview
Calibration Message
The power supply allows you to store one message in calibration memory in
the mainframe. For example, you can store such information as the date when
the last calibration was performed, the date when the next calibration is due,
the power supply’s serial number, or even the name and phone number of the
person to contact for a new calibration.
• You can record a calibration message only from the remote interface and
only when the power supply is unsecured. You can read the message from
either the front-panel or over the remote interface. You can read the
calibration message whether the powers supply is secured or unsecured.
• The calibration message may contain up to 40 characters. From the front
panel, you can view 11 characters of the message at a time.
• Storing a calibration message will overwrite any message previously stored
in memory.
• The calibration message is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
• Front-Panel Operation:
To read calibration message from the front panel, press and turn the
knob until ‘‘
text of the message. Press to increase the scrolling speed.
CAL STRING’’ is displayed. Press to scroll through the
>
View
View
ViewView
CAL STRING
• Remote Interface Operation:
To store the calibration m essage, send the following command.
‘‘
CAL:STR ‘CAL 12-05-99’’’
68
4
Remote Interface Reference
Remote Interface Reference
• SCPI Command Summary‚ starting on page 71
SCPI
• Simplified Programming Overview‚ starting on page 77
• Using the
• Output Setting and Operation Commands‚ starting on page 81
• Triggering‚ starting on page 87
• System-Related Commands‚ starting on page 90
• State Storage Commands‚ on page 93
• Calibration Commands‚ starting on page 94
• Interface Configuration Commands‚ on page 97
• The SCPI Status Registers‚ starting on page 98
• Status Reporting Commands‚ starting on page 108
SCPI
• An Introduction to the SCPI Language‚ starting on page 111
• Halting an Output in Progress‚ on page 116
• SCPI Conformance Information‚ starting on page 117
• IEEE-488 Conformance Information‚ on page 120
APPLy Command‚ on page 80
SCPI
If you are a first-time user of the SCPI language, you may want to refer to
these sections to become familiar with the language before attempting to
program the power supply.
70
Chapter 4 Remote Interface Reference
SCPI Command Summary
SCPI Command Summary
This section summarizes the SCPI (Standard Commands for Programmable
Instruments) commands available to program the power supply over the
remote interface. Refer to the later sections in this chapter for more complete
details on each command.
Throughout this manual, the following conventions are used for SCPI
command syntax.
• Square brackets ([ ]) indicate optional keywords or parameters.
• Braces ({ }) enclose parameters within a command string.
• Triangle brackets (< >) indicate that you must substitute a value or a code
for the enclosed parameter.
• A vertical bar ( | ) separates one of two or more alternative parameters.
This section gives an overview of the basic techniques used to program the
power supply over the remote interface. This section is only an overview and
does not give all of the details you will need to write your own application
programs. Refer to the remainder of this chapter and also chapter 6,
‘‘Application Programs’’, for more details and examples. Also refer to the
programming reference manual that came with your computer for details on
outputting command strings and entering data.
Using the APPLy Command
The APPLy command provides the most straightforward method to program
the power supply over the remote interface. For example, the following
statement executed from your computer will set the power supply to an output
of 3 V rated at 1 A:
‘‘APPL 3.0, 1.0’’
Using the Low-Level Commands
Although the APPLy command provides the most straightforward method to
program the power supply, the low-level commands give you more flexibility
to change individual parameters. For example, the following statements
executed from your computer will set the power supply to an output of 3 V
rated at 1 A:
‘‘VOLT 3.0’’Set output voltage to 3.0 V
‘‘CURR 1.0’’Set output current to 1.0 A
4
77
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Reading a Query Response
Only the query commands (commands that end with “ ? ”) will instruct the
power supply to send a response message. Queries return either output values
or internal instrument settings. For example, the following statements
executed from your computer will read the power supply’s error queue and
print the most recent error:
bus enter statementEnter error string into computer
print statementPrint error string
Selecting a Trigger Source
The power supply will accept a ‘‘bus’’ (software) trigger or an immediate
internal trigger as a trigger source. By default, the ‘‘BUS’’ trigger source is
selected. If you want the power supply to use an immediate internal trigger,
you must select ‘‘
executed from your computer will set to an output of 3 V/1 A immediately:
‘‘VOLT:TRIG 3.0’’Set the triggered voltage level to 3.0 V
‘‘CURR:TRIG 1.0’’Set the triggered current level to 1.0 A
‘‘TRIG:SOUR IMM’’Select the immediate trigger as a source
‘‘INIT’’Cause the trigger system to initiate
IMMediate’’. For example, the following statements
78
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Power Supply Programming Ranges
The SOURce subsystem requires parameters for programming values. The
available programming value for a parameter varies according to the desired
output range of the power supply. The following table lists the programming
values available and
power supply.
Refer to this table to identify programming values when programming the power
supply.
MINimum, MAXimum, DEFault and reset values of your
E3646AE3647AE3648AE3649A
0 - 8V/
Range
0 V to
8.24 V
0 A to
3.09 A
3A
0 - 20V/
1.5A
Range
0 V to
20.60 V
0 A to
1.545 A
0 - 35V/
0.8A
Range
0 V to
36.05 V
0 A to
0.824 A
0 - 60V/
0.5A
Range
0 V to
61.8 V
0 A to
0.515 A
0 - 8V/
5A
Range
0 V to
8.24 V
0 A to
5.15 A
0 - 20V/
2.5A
Range
0 V to
20.60 V
0 A to
2.575 A
0 - 35V/
1.4A
Range
0 V to
36.05 V
0 A to
1.442 A
0 - 60V/
0.8A
Range
0 V to
61.8 V
4
0 A to
0.824 A
79
Chapter 4 Remote Interface Reference
Using the A PPLy Command
Using the APPLy Command
The APPLy command provides the most straightforward method to program
the power supply over the remote interface. You can select the output voltage
and current in one command.
APPLy {<voltage>| DEF | MIN | MAX}[,{<current>| DEF | MIN | MAX}]
This command is combination of VOLTage and CURRent commands.
APPLy command changes the power supply’s output to the newly
The
programmed values only if the programmed values are valid within the
presently selected range. An execution error will occur if the programmed
values are not valid within the selected range.
You can substitute ‘‘
specific value for the voltage and current parameters. For more details of
parameters, see Table 4-1 through Table 4-4 for each model.
If you specify only one parameter of the
regards it as voltage setting value.
MINimum’’, ‘‘MAXimum’’, or ‘‘DEFault’’ in place of a
APPLy command, the power supply
APPLy?
Query the power supply’s present voltage and current setting values and
returns a quoted string. The voltage and current are returned in sequence as
shown in the sample string below (the quotation marks are returned as part of
the string).
‘‘8.00000,3.00000’’ (E3646A model)
80
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output Setting and Operation Commands
This section describes low-level commands used to program the power supply.
Although the
program the power supply, the low-level output setting commands give you
more flexibility to change the individual parameters.
CURRent{<current>| MINimum | MAXimum | UP | DOWN}
Program the immediate current level of the power supply. The immediate level
is the current value of the output terminals.
The
CURRent command changes the output of the power supply to the newly
programmed value regardless of the output range presently selected.
You can substitute ‘‘
the current parameter.
selects the highest current values allowed for the selected range.
This command also increases or decreases the immediate current level using
the ‘‘
UP’’ or ‘‘DOWN’’ parameter by a predetermined amount. The command
CURRent:STEP sets the amount of increase or decrease. Notice that a new
increment setting will cause an execution error -222 (Data out of range) when
the maximum or the minimum rated current is exceeded.
APPLy command provides the most straightforward method to
MINimum’’ or ‘‘MAXimum’’ in place of a specific value for
MIN selects the lowest current values of ‘‘0’’ volts. MAX
4
CURRent? [MINimum | MAXimum]
Return the presently programmed current level of the power supply.
CURR? MAX and CURR? MIN return the highest and lowest programmable
current levels for the selected range.
CURRent:STEP {<numeric value>| DEFault}
Set the step size for current programming with the CURRentUP and CURRent
DOWN commands. See the example in the next page.
To set the step size to the minimum resolution, set the step size to ‘‘
The minimum resolution of the step size is approximately 0.052 mA (E3646A),
0.014 mA (E3647A), 0.095 mA (E3648A), and 0.027 mA (E3649A), respectively.
The
CURR:STEP? DEF returns the minimum resolution of your instrument.
The immediate current level increases or decreases by the value of the step
size. For example, the output current will increase or decrease 10 mA if the
step size is 0.01. At
*RST, the step size is the value of the minimum resolution.
DEFault’’.
81
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
CURRent:STEP? [DEFault]
Return the value of the step size currently specified. The returned parameter
is a numeric value. ‘‘
in unit of amps.
DEFault’’ gives the minimum resolution of the step size
CURRent:TRIGgered {<current>| MINimum | MAXimum}
Program the pending triggered current level. The pending triggered current
level is a stored value that is transferred to the output terminals when a trigger
occurs. A pending triggered level is not affected by subsequent
commands.
CURRent
CURRent:TRIGgered? [MINimum | MAXimum]
Query the triggered current level presently programmed. If no triggered level
is programmed, the
CURRent level is returned.
Example The following program segments show how to use the CURR UP or CURR DOWN
command to increase or decrease the output current with the
command.
‘‘
CURR:STEP 0.01’’Set the step size to 0.01 A
‘‘CURRUP’’ Increase the output current
‘‘CURR:STEP0.02’’ Set the step size to 0.02 A
‘‘CURRDOWN’’ Decrease the output current
Select the output to be programmed one of the two outputs by the output
identifier. The outputs of the power supply are considered as two logical
instruments. The
and select an output. When one output is selected, the other output is
unavailable for programming until selected. The commands which are affected
by the
INSTrument command are output setting command (SOURce),
measurement command (
(
CALibration).
INSTrument command provides a mechanism to identify
MEASure), and calibration command
INSTrument[:SELect]?
Return the currently selected output by the INSTrument{:SELect] or
INSTrument:NSELect command. The returned value is ‘‘OUTP1’’ or
‘‘OUTP2’’.
82
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INSTrument:NSELect {1 | 2}
Select the output to be programmed one of the two outputs by a numeric value
instead of the output identifier used in the
INSTrument[:SELect] command. ‘‘1’’ selects output1, and ‘‘2’’ selects
output2.
INSTrument:NSELect?
Return the currently selected output by the INSTrument[SELect]or
INSTrument[SELect] command. The returned parameter is ‘‘1’’ for output1,
‘‘2’’ for output2.
INSTrument:COUPle[:TRIGger] {ON | OFF}
Enable or disable a coupling between two logical outputs of the power supply.
The couple command consists of an optional subsystem node followed by a
signal parameter. The only valid parameter for the optional subsystem node is
TRIGger subsystem. If no node follows the couple command, TRIGger
subsystem is assumed to be coupled.
The parameter indicates to which logical outputs the specified coupling is to
apply. A list of outputs specifies a particular set of logical outputs to be coupled.
At
*RST, all outputs are uncoupled.
INSTrument:NSELect or
4
Example The following program segment shows how to use the INSTrument:COUPle
command to couple two outputs with voltage and current triggered levels. The
power supply is set to the newly programmed values as set by the
VOLTage:TRIGgered and CURRent:TRIGgered commands.
INST:SEL OUT1’’ Select the output1
‘‘
VOLT:TRIG 6’’ Set triggered level to 6 V
‘‘
CURR:TRIG 3’’ Set triggered level to 3 A
‘‘
INST:SEL OUT2’’ Select the output2
‘‘
VOLT:TRIG 18’’ Set triggered level to 18 V
‘‘
CURR:TRIG 0.7’’ Set triggered level to 0.7 A
‘‘
INST:COUP:TRIG ON’’ Couple the output1 and the output2
‘‘
TRIG:SOUR IMM’’ Set trigger to immediate
‘‘
INIT’’ Trigger the power supply to output the
‘‘
trigger values for the output1 and output2
supplies.
83
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INSTrument:COUPle[:TRIGger]?
Query the output coupling state of the power supply. The returned value is ‘‘0’’
(OFF) or ‘‘1’’ (ON).
MEASure:CURRent?
Query the current measured across the current sense resistor inside the power
supply.
MEASure[:VOLTage]?
Query the voltage measured at the sense terminals of the power supply.
OUTPut:TRACk[:STATe] {ON | OFF}
Enable or disable the power supply to operate in the track mode. See “Tracking
Operation” on page 44 for more information on tracking operation. At
the track mode is disabled.
*RST,
OUTPut:TRACk[:STATe]?
Query the tracking mode state of the power supply. The returned value is ‘‘0’’
(OFF) or ‘‘1’’ (ON).
VOLTage {<voltage>| MINi m um | MA Xim um | UP | DOWN }
Program the immediate voltage level of the power supply. The immediate level
is the voltage value of the output terminals.
The
VOLTage command changes the output of the power supply to the newly
programmed value regardless of the output range presently selected.
This command also increases or decreases the immediate voltage level using
the ‘‘
UP’’ or ‘‘DOWN’’ parameter by a predetermined amount. The command
VOLTage:STEP sets the amount of increase or decrease. Notice that a new
increment setting will cause an execution error -222 (Data out of range) when
the maximum or the minimum rated voltage is exceeded.
VOLTage? [MINimum | MAXimum]
Query the presently programmed voltage level of the power supply.
84
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage:STEP {<numeric value> | DEFault}
Set the step size for voltage programming with the VOLT UP and VOLT DOWN
commands. See the example below.
To set the step size to the minimum resolution, set the step size to ‘‘
The minimum resolution of the step size is approximately 0.35 mV (E3646A),
The immediate voltage level increases or decreases by the value of the step
size. For example, the output voltage will increase or decrease 10 mV if the
step size is 0.01. At
VOLTage:STEP? [DEFault]
Return the value of the step size currently specified. The returned parameter
is a numeric value. ‘‘
of volts.
Example The following program segments show how to use the VOLT UP or VOLT DOWN
command to increase or decrease the output voltage with the
command.
‘‘
VOLT:STEP 0.01’’ Set the step size to 0.01 V
‘‘VOLT UP’’ Increase the output voltage
‘‘VOLT:STEP 0.02’’ Set the step size to 0.02 V
‘‘VOLT DOWN’’ Decrease the output voltage
*RST, the step size is the value of the minimum resolution.
DEFault’’ gives the minimum resolution step size in unit
DEFault’’.
VOLT:STEP
4
VOLTage:TRIGgered {<voltage>| MINimum | MAXimum}
Program the pending triggered voltage level. The pending triggered voltage
level is a stored value that is transferred to the output terminals when a trigger
occurs. A pending triggered level is not affected by subsequent
commands.
VOLTage:TRIGgered? [MINimum | MAXimum]
Query the triggered voltage level presently programmed. If no triggered level
is programmed, the
* For E3646A/48A models **For E3647A/49A models
VOLT level is returned.
VOLTage
85
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage:PROTection {<voltage>|MINimum|MAXimum}
Set the voltage level at which the overvoltage protection (OVP) circuit will trip.
If the peak output voltage exceeds the OVP level, then the power supply output
is shorted by an internal SCR. An overvoltage condition can be cleared with
the
VOLT:PROT:CLE command after the condition that caused the OVP trip
is removed.
VOLTage:PROTection? [MINimum | MAXimum]
Query the overvoltage protection trip level presently programmed.
VOLTage:PROTection:STATe {0 | 1 | OFF | ON}
Enable or disable the overvoltage protection function. At *RST, this value is
set to ‘‘ON’’.
VOLTage:PROTection:STATe?
Query the state of the overvoltage protection function. The returned parameter
is ‘‘0’’ (OFF) or ‘‘1’’ (ON).
VOLTage:PROTection:TRIPped?
Return a ‘‘1’’ if the overvoltage protection circuit is tripped and not cleared or
a ‘‘0’’ if not tripped.
VOLTage:PROTection:CLEar
Cause the overvoltage protection circuit to be cleared. After this command,
the output voltage is restored to the state it was in before the protection feature
occurred and the OVP trip level remains unchanged to the value presently
programmed. Before sending this command, lower the output voltage below
the trip OVP point, or raise the OVP trip level above the output setting. Note
that the overvoltage condition caused by an external source must be removed
first before proceeding this command.
Select an output range to be programmed by the identifier. For example,
‘‘P20V’’ or ‘‘HIGH’’ is the identifier for the 20V/1.5A range and ‘‘P8V’’ or ‘‘LOW’’
is for the 8V/3A range (for E3646A model). At
selected.
*RST, low voltage range is
VOLTage:RANGe?
Query the currently selected range. The returned parameter is ‘‘P8V’’ or ‘‘P35V’’
for low voltage range, or ‘‘P20V’’ or ‘‘P60V’’ for high voltage range.
86
Chapter 4 Remote Interface Reference
Triggering
Triggering
The power supply’s triggering system allows a change in voltage and current
when receiving a trigger, to select a trigger source, and to insert a trigger.
Triggering the power supply is a multi-step process.
• First, you must specify the source from which the power supply will accept
the trigger. The power supply will accept a bus (software) trigger or an
immediate trigger from the remote interface.
• Then, you can set the time delay between the detection of the trigger on the
specified trigger source and the start of any corresponding output change.
Notice that the time delay is valid for only the bus trigger source.
• Finally, you must provide an INITiate command. If the IMMediate
source is selected, the selected output is set to the triggered level
immediately. But if the trigger source is the bus, the power supply is set to
the triggered level after receiving the Group Execute Trigger (GET) or
command.
Trigger Source Choices
You must specify the source from which the power supply will accept a trigger.
The trigger is stored in volatile memory; the source is set to bus when the power
supply has been off or after a remote interface reset.
*TRG
4
Bus (Software) Triggering
• To select the bus trigger source, send the following command.
‘‘TRIG:SOUR BUS’’
• To trigger the power supply from the remote interface (GPIB or RS-232)
after selecting the bus source, send the
*TRG is sent, the trigger action starts after the specified time delay if any
delay is given.
• You can also trigger the power supply from the GPIB interface by sending
the IEEE-488 Group Execute Trigger (GET) message. The following
statement shows how to send a GET from a Hewlett-Packard controller.
*TRG (trigger) command. When the
‘‘TRIGGER 705’’ (group execute trigger)
87
Chapter 4 Remote Interface Reference
Triggering
• To ensure synchronization when the bus source is selected, send the *WAI
(wait) command. When the
waits for all pending operations to complete before executing any additional
commands. For example, the following command string guarantees that the
first trigger is accepted and is executed before the second trigger is
recognized.
*WAI command is executed, the power supply
‘‘TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI’’
• You can use the
(operation complete) command to signal when the operation is complete.
The
*OPC? command returns ‘‘1’’ to the output buffer when the operation
is complete. The
Event register when the operation is complete.
Immediate Triggering
• To select the immediate trigger source, send the following command.
*OPC? (operation complete query) command or the *OPC
*OPC command sets the ‘‘OPC’’ bit (bit 0) in the Standard
‘‘TRIG:SOUR IMM’’
• When the IMMediate is selected as a trigger source, an INITiate
command immediately transfers the
VOLT or CURR value. Any delay is ignored.
VOLT:TRIG or CURR:TRIG value to
88
Chapter 4 Remote Interface Reference
Triggering
Triggering Commands
INITiate
Cause the trigger system to initiate. This command completes one full trigger
cycle when the trigger source is an immediate and initiates the trigger
subsystem when the trigger source is bus.
TRIGger:DELay {<seconds>| MINimum | MAXimum}
Set the time delay between the detection of an event on the specified trigger
source and the start of any corresponding trigger action on the power supply
output. Select from 0 to 3600 seconds.
At
*RST, this value is set to 0 seconds.
TRIGger:DELay?[MINimum | MAXimum]
Query the trigger delay.
TRIGger:SOURce {BUS | IMMediate}
Select the source from which the power supply will accept a trigger. The power
supply will accept a bus (software) trigger or an internal immediate trigger. At
*RST, the bus trigger source is selected.
MIN = 0 seconds. MAX = 3600 seconds.
4
TRIGger:SOURce?
Query the present trigger source. Returns ‘‘BUS’’ or ‘‘IMM’’.
*TRG
Generate a trigger to the trigger subsystem that has selected a bus (software)
trigger as its source
as the Group Execute Trigger (GET) command. For RS-232 operation, make
sure the power supply is in the remote interface mode by sending the
SYST:REM command first.
(TRIG:SOUR BUS). The command has the same effect
89
Chapter 4 Remote Interface Reference
System-Related Commands
System-Related Commands
DISPlay {OFF | ON}
Turn the front-panel display off or on. When the display is turned off, outputs
are not sent to the display and all annunciators are disabled except the
annunciator.
The display state is automatically turned on when you return to the local mode.
Press (
Store
Local) key to return to the local state from the remote interface.
Local
DISPlay:MODE {VI | VV | II}
Set the front-panel display mode of the power supply. V-V display mode shows
the voltages of the two outputs. I-I display mode shows the currents of the two
outputs. V-I display mode shows the voltage and current for each output. At
*RST, the power supply is set to the V-I display mode.
DISPlay:MODE?
Query the state of the display mode. Returns a string VV, II, or VI.
DISPlay?
Query the front-panel display setting. Returns ‘‘0’’ (OFF) or ‘‘1’’ (ON).
ERROR
DISPlay:TEXT <quoted string>
Display a message on the front panel. The power supply will display up to 11
characters in a message; any additional characters are truncated. Commas,
periods, and semicolons share a display space with the preceding character,
and are not considered individual characters.
DISPlay:TEXT?
Query the message sent to the front panel and returns a quoted string.
DISPlay:TEXT:CLEar
Clear the message displayed on the front panel.
OUTPut {OFF | ON}
Enable or disable the outputs of the power supply. When the output is disabled,
the voltage value is 0 V and the current value is 1 mA. At
*RST, the output state
is OFF.
OUTPut?
Query the output state of the power supply. The returned value is ‘‘0’’ (OFF)
or ‘‘1’’ (ON).
90
Chapter 4 Remote Interface Reference
System-Related Commands
OUTPut:RELay {OFF | ON}
Set the state of two TTL signals on the RS-232 connector pin 1 and pin 9. These
signals are intended for use with an external relay and relay driver. At
the
OUTPUT:RELay state is OFF. See ‘‘Disconnecting the Output Using an
External Relay’’, on page 56 for more information.
Note Do not use the RS-232 interface if you have configured the power supply to output
relay control signals. Internal components on the RS-232 circuitry may be damaged.
OUTPut:RELay?
Query the state of the TTL relay logic signals.
SYSTem:BEEPer
Issue a single beep immediately.
SYSTem:ERRor?
Query the power supply’s error queue. A record of up to 20 errors is stored in
the power supply’s error queue. Errors are retrieved in first-in-first-out (FIFO)
order. The first error returned is the first error that was stored. When you have
read all errors from the queue, the
are cleared. See ‘‘Error Messages’’, starting on page 121 for more details.
ERROR annunciator turns off and the errors
*RST,
4
SYSTem:VERSion?
Query the power supply to determine the present SCPI version. The returned
value is of a string in the form YYYY.V where the ‘‘Y’s’’ represent the year of
the version, and the ‘‘V’’ represents a version number for that year (for example,
1997.0).
*IDN?
Read the power supply’s identification string. The power supply returns four
fields separated by commas. The first field is the manufacturer’s name, the
second field is the model number, the third field is not used (always ‘‘0’’), and
the fourth field is a revision code which contains three numbers. The first
number is the firmware revision number for the main power supply processor;
the second is for the input/output processor; and the third is for the front-panel
processor.
The command returns a string with the following format (be sure to dimension
a string variable with at least 40 characters):
Perform a complete self-test of the power supply. Returns ‘‘0’’ if the self-test
passes or ‘‘1’’ or any non-zero value if it fails. If the self-test fails, an error
message is also generated with additional information on why the test failed.
*RST
Reset the power supply to its power-on state. The table below shows the state
of the power supply after a
RESET from the Recall menu or *RST command
from the remote interface.
Comman dE3646A
CURR
CURR:STEP
CURR:TRIG
DISP
OUTP
OUTP:REL
TRIG:DEL
TRIG:SOUR
VOLT
VOLT:STEP
VOLT:TRIG
VOLT:PROT
VOLT:PROT:STAT
VOLT:RANG
state
3 A0.8 A5 A1.4 A
0.052 mA 0.014 mA 0.095 mA 0.027 mA
3 A0.8 A5 A1.4 A
0.35 mV1.14 mV0.38 mV1.14 mV
22.0 V66.0 V22.0 V66.0 V
P8V (Low)P35V (Low)P8V (Low)P35V (Low)
E3647A
state
ON
OFF
OFF
0
BUS
0 V
0 V
ON
E3648A
state
E3649A
state
Note: The voltage and current step sizes listed above are typical value.
92
Chapter 4 Remote Interface Reference
State Storage Commands
State Storage Commands
The power supply has five storage locations in non-volatile memory to store
power supply states. The locations are numbered 1 through 5. You can also
assign a name to each of the locations (1 through 5) for use from the front panel.
*SAV { 1 | 2 | 3 | 4 | 5 }
Store (Save) the present state of the power supply to the specified location.
Any state previously stored in the same location is overwritten (no error is
generated).
• A power-on reset (
stored in memory. Once a state is stored, it remains until it is overwritten
or specifically deleted.
• The state storage feature ‘‘remembers’’ the states or values of the following
commands:
CURR, CURR:STEP, CURR:TRIG, OUTP, OUTP:REL, TRIG:DEL,
TRIG:SOUR, VOLT, VOLT:STEP, VOLT:TRIG, VOLT:PROT,
VOLT:PROT:STAT, and VOLT:RANG
*RST command) does not affect the configurations
4
*RCL { 1 | 2 | 3 | 4 | 5 }
Recall the power supply state stored in the specified storage location. When
shipped from the factory, storage locations ‘‘1’’ through ‘‘5’’ are empty.
Note:
DISP {OFF|ON} can be stored and recalled in remote interface mode
only. Going to local mode automatically sets the display state to ON.
Assign a name to the specified storage location. From the remote interface,
you can only recall a stored state using a number (1 through 5). The
query returns a quoted string containing the name currently assigned to the
specified storage location. If the specified location has no name assigned, an
empty string (‘‘ ’’) is returned. The name can contain up to 11 characters. The
first character can be a alphanumeric. Blank spaces are not allowed. An error
is generated if you specify a name with more than 11 characters. See ‘‘State Storage’’, on page 57 for more information. An example is shown below.
‘‘
MEM:STATE:NAME 1,‘P15V_TEST’’’
If you do not specify a name (note that the name parameter is optional), no
name is assigned to that state. This provides a way to erase a name (however,
the stored state is not deleted).
:NAME?
93
Chapter 4 Remote Interface Reference
Calibration Commands
Calibration Commands
See ‘‘Calibration Overview’’, starting on page 64 for an overview of the
calibration features of the power supply. An example program for calibration
is listed on page 96. For more detailed discussion on the calibration
procedures, see the Service Information.
Note When you calibrate the power supply, you should NOT set the OVP to ON state in
order to prevent OVP from tripping.
CALibration:COUNt?
Query the power supply to determine the number of times it has been
calibrated. Your power supply was calibrated before it left the factory. When
you receive your power supply, read the count to determine its initial value.
Since the value increments by one for each calibration point, a complete
calibration will increase the value by 3 counts.
CALibration:CURRent[:DATA] <numeric value>
This command can only be used after calibration is unsecured and the output
state is ON. It enters a current value that you obtained by reading an external
meter. You must first select the minimum calibration level (
MIN) for the value being entered, then select the middle and maximum
calibration levels (
value being entered. Three successive values must be selected and entered.
The power supply then computes new calibration constants. These constants
are then stored in non-volatile memory.
This command can only be used after calibration is unsecured and the output
state is ON. It sets the power supply to a calibration point that is entered with
CAL:CURR command. During calibration, three points must be entered and the
low-end point (MIN) must be selected and entered first.
CALibration:SECure:CODE <quoted new code>
Enter a new security code. To change the security code, first unsecure the
power supply using the old security code. Then, enter the new code. The
calibration code may contain up to 11 characters over the remote interface.
See “Calibration Overview” on page 64 for more information.
94
Chapter 4 Remote Interface Reference
Calibration Commands
CALibration:SECure:STATe {OFF | ON},<quoted code>
Unsecure or secure the power supply with a security for calibration.
CALibration:SECure:STATe?
Query the secured state for calibration of the power supply. The returned
parameter is ‘‘0’’ (OFF) or ‘‘1’’ (ON).
CALibration:STRing <quoted string>
Record calibration information about your power supply. For example, you
can store such information as the last calibration date, the next calibration due
date, or the power supply’s serial number. The calibration message may
contain up to 40 characters. The power supply should be unsecured before
sending a calibration message.
CALibration:STRing?
Query the calibration message and returns a quoted string.
CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unsecured and the output
state is ON. It enters a voltage value that you obtained by reading an external
meter. You must first select the minimum calibration level (
MIN) for the value being entered. You must then select the middle and
maximum calibration levels (
for the value being entered. Three successive values must be selected and
entered. The power supply then computes new voltage calibration constants.
These constants are then stored in non-volatile memory.
This command can only be used after calibration is unsecured and the output
state is ON. It sets the power supply to a calibration point that is entered with
CAL:VOLT command. During calibration, three points must be entered and the
low-end point (MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
Calibrate the overvoltage protection circuit of the power supply. It takes about
10 seconds to execute the command. The calibration must be unsecured and
the output be opened before calibrating the overvoltage protection circuit. The
power supply automatically performs the calibration and stores the new
overvoltage constant in nonvolatile memory. Notice that voltage calibration
precedes before sending this command.
95
Chapter 4 Remote Interface Reference
Calibration Commands
Calibration
Example 1 Select the output to be calibrated and enable the output of the power
supply by sending the commands:
‘‘INST:SEL {OUT1|OUT2}’’
OUTP ON’’
‘‘
2 Disable the voltage protection function.
‘‘VOLT:PROT:STAT OFF’’
3 Unsecure the power supply with the secure code before calibration.
‘‘CAL:SEC:STAT OFF, ‘<code>’’’
4 For voltage calibration, connect a digital voltmeter (DVM) across the
power supply’s output terminals.
5 Set the power supply to low-end (MIN) calibration point.
‘‘CAL:VOLT:LEV MIN’’
6 Enter the reading you obtained from the DVM.
‘‘CAL:VOLT:DATA 0.549’’
7 Set the power supply to middle (MID) calibration point.
‘‘CAL:VOLT:LEV MID’’
8 Enter the reading you obtained from the DVM.
‘‘CAL:VOLT:DATA 11.058’’
9 Set the power supply to high (MAX) calibration point.
‘‘CAL:VOLT:LEV MAX’’
10 Enter the reading you obtained from the DVM.
‘‘CAL:VOLT:DATA 21.566’’
11 Set the power supply to overvoltage protection calibration point.
‘‘CAL:VOLT:PROT’’
12 For current calibration, connect an appropriate current monitoring
resistor (shunt) across the output terminals and connect the DVM
across the shunt resistor.
13 Repeat steps (5) through (9) by substituting ‘‘CURR’’ for ‘‘VOLT’’ for
current calibration. For example, ‘‘
14 Repeat steps (1) through (13) for the other output calibration.
15 Record calibration information such as next calibration due date or
contact person for future reference. The calibration string may contain
up to 40 characters.
CAL:CURR:LEV MIN’’
‘‘CALibration:STRing ‘<string>’’’
Notice: You should wait for the DVM reading to stabilize for accurate calibration.
96
Chapter 4 Remote Interface Reference
Interface Configuration Commands
Interface Configuration Commands
See also "Configuring the Remote Interface" in chapter 3 starting on page 48.
SYSTem:INTerface {GPIB | RS232}
Select the remote interface. Only one interface can be enabled at a time. The
GPIB interface is selected when the power supply is shipped from the factory.
SYSTem:LOCal
Place the power supply in the local mode during RS-232 operation. All keys on
the front panel are fully functional.
SYSTem:REMote
Place the power supply in the remote mode for RS-232 operation. All keys on
the front panel, except the ‘‘
Local’’ key, are disabled while in the remote mode.
It is very important that you send the
power supply in the remote mode. Sending or receiving data over the
RS-232 interface when not configured for remote operation can cause
unpredictable results.
SYSTem:RWLock
Place the power supply in the remote mode for RS-232 operation. This
command is the same as the
front panel are disabled, including the ‘‘
<Ctrl -C>
Clear the operation in progress over the RS-232 interface and discard any
pending output data. This is equivalent to the IEEE-488 device clear action
over the GPIB interface.
SYST:REM command except that all keys on the
SYST:REM command to place the
Local’’ key.
4
97
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The SCPI Status Registers
All SCPI instruments implement status registers in the same way. The status
system records various instrument conditions in three register groups: the
Status Byte register, the Standard Event register, and the Questionable Status
register groups. The status byte register records high-level summary
information reported in the other register groups. The diagram on the
subsequent pages illustrates the SCPI status system used by the power supply.
What is an Event Register?
An event register is a read-only register that reports defined conditions within
the power supply. Bits in an event register are latched. Once an event bit is set,
subsequent state changes are ignored. Bits in an event register are
automatically cleared by a query of that register (such as
STAT:QUES:EVEN?) or by sending the *CLS (clear status) command. A reset
(
*RST) or device clear will not clear bits in event registers. Querying an event
register returns a decimal value which corresponds to the binary-weighted sum
of all bits set in the register.
*ESR? or
What is an Enable Register?
An enable register defines which bits in the corresponding event register are
logically ORed together to form a single summary bit. Enable registers are both
readable and writable. Querying an enable register will not clear it. The
(clear status) command does not clear enable registers but it does clear the
bits in the event registers. To enable bits in an enable register, you must write
a decimal value which corresponds to the binary-weighted sum of the bits you
wish to enable in the register.
*CLS
What is a Multiple Logical Output?
The two-logical outputs of the power supply include an INSTrument summary
status register and an individual instrument
output. The
in turn reports to bit 13 of the Questionable status register. This is shown
pictorially on the next page.
98
ISUMmary registers report to the INSTrument register, which
ISUMmary register for each logical
Chapter 4 Remote Interface Reference
The SCPI Status Registers
Using such a status register configuration allows a status event to be crossreferenced by output and type of event. The
which output(s) have generated an event. The
INSTrument register indicates
ISUMmary register is a pseudo-
questionable status register for a particular logical output.
STA Tus:QUEStionable:INSTrument:ISUMmary1
Event Registers Enable Registers
Voltage
Current
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
O v erv o lt age
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
STA Tus:QUEStionable:INSTrument:ISUMmary2
Event Registers Enable Registers
Voltage
Current
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
O v erv o lt age
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
0
1
"OR"
+
9
0
1
"OR"
STAT us:QUEStionable:INSTrument
Event Registers Enable Registers
Not Used
INST1 Ev ent Summary
INST2 Event Summary
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
1
2
+
9
To STA Tus:QUEStionable, bit 13
4
"OR"
+
99
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.