Agilent Technologies E364xA
Dual Output DC Power Supplies
The Agilent Technologies E3646A/E3647A (30 watt) and E3648A/E3649A
(50 watt) are high performance dual-output dual range programmable DC
power supplies with GPIB and RS-232 interfaces. The combination of benchtop and system features in these power supplies provides versatile solutions
for your design and test requirements.
Convenient bench-top features
• Dual-output dual range
• Output on/off
• High accuracy and high resolution
• Excellent load and line regulation
• Low ripple and noise
• Overvoltage protection
• Five Operating states storage
• Tracking operation
• Easy-to-use controls
• Remote voltage sensing
• Front and Rear output terminals
• Portable, ruggedized case with non-skid feet
• Error messages available on the display
Flexible system features
• GPIB (IEEE-488) and RS-232 interfaces are standard
• SCPI (Standard Commands for Programmable Instruments) compatibility
• I/O setup easily done from front-panel
• Software calibration, no internal physical adjustments
Agilent Technologies E364xA
Dual Output DC Power Supplies
The Front Panel at a Glance
1 Output1 selection Key
2 Output2 selection Key
3 Low voltage range selection Key
4 High voltage range selection Key
5 Overvoltage protection Key
6 Display limit Key
7 Resolution selection Keys
8 Voltage / Current adjust selection Key
2
9 Knob
10 Output On / Off Key
11 I/O Configuration Menu / Secure Key
12 View Menu / Calibrate Key
13 State Storage Menu / Local Key
14 Stored state Recall/Reset Menu
15 Tracking enabling/disabling Key
Front-Panel Menu/Key Reference
This section gives an overview of the front-panel keys/menus. The menus are
designed to automatically guide you through all parameters required to
configure a particular function or operation.
1 Output1 selection key Select the output1 voltage and current to be controlled and
monitored on the display.
2 Output2 selection key Select the output2 voltage and current to be controlled and
monitored on the display.
3 Low voltage range selection key Selects the low voltage range and allows its full
rated output to the output terminals.
4 High voltage range selection key Selects the high voltage range and allows its full
rated output to the output terminals.
5 Overvoltage protection key Enables or disables the overvoltage protection function,
sets trip voltage level, and clears the overvoltage condition.
6 Display limit key Shows voltage and current limit values on the display and allows
the knob adjustment for setting limit values.
7 Resolution selection keys Move the flashing digit to the right or left. Adjust the
scrolling speed of the text being displayed in the View menu.
8 Voltage/Current adjust selection key Selects the knob control function for voltage
or current adjustment.
9 Knob Increases or decreases the value of the flashing digit by turning clockwise or
counter clockwise.
10 Output On/Off key Enables or disables the power supply output. This key toggles
between on and off.
11 I/O Configuration / Secure key
/ or secures or unsecures the power supply for calibration.
12 View menu / Calibrate key
calibration string, and system firmware revision / or enables calibration mode.
13 State storage menu / Local key
a name to each of the storage locations / or returns the power supply to local mode
from remote interface mode.
14 Stored state Recall menu Recalls a stored operating state from location ‘‘1’’ through
‘‘5’’ and resets the power supply to the power-on state (*RST command).
15 Tracking enabling/disabling key Enable / disable the track mode of the outputs.
3
Configures the power supply for remote interfaces
2
Views the error codes and the text of the error message,
1
Stores up to five power supply’s states and assigns
1
The key can be used as the ‘‘Local’’ key when the power supply is in the remote
interface mode.
2
You can enable the ‘‘calibration mode’’ by holding down this key when you turn on the
power supply.
3
You can use it as the ‘‘Secure’’ or ‘‘Unsecure’’ key when the power supply is in the
calibration mode.
3
Front-Panel Voltage and Current Limit Settings
You can set th e voltage and current limit valu es from the front panel using the
following method.
Use the voltage/current adjust selection key, the resolution selection keys,
and the control knob to change the voltage and current limit values.
1
Or
2
+
Low
Or
High
+
1 Select the desired output and voltage range using the output selection keys
and voltage range selection keys after turning on the power supply.
2 Press key to show the limit values on the display.
Display
Limit
3 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired voltage limit
by turning the control knob. If the display limit times out, press key again.
4 Set the knob to current control mode by pressing key.
Voltage
Curre n t
Display
Limit
5 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired current limit
by turning the control knob.
6 Press key to enable the output. After about 5 seconds, the display will
Output
On/ Off
go to output monitoring mode automatically to display the voltage and current
at the output.
Note All front panel keys and controls can be disabled with remote interface commands.
The power supply must be in "Local" mode for the front panel keys and controls to
function.
4
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
8V*/35V**Shows the low voltage range is selected.
20V*/60**Shows the high voltage range is selected.
OUT1The output1 is selected for front-panel or remote operation.
OUT2The output2 is selected for front-panel or remote operation.
OVP1The output1 overvoltage protection function is enabled when the
OVP2The output2 overvoltage protection function is enabled when the
CALThe power supply is in calibration mode.
LimitThe display shows the limit values of voltage and current.
ERRORHardware or remote interface command errors are detected and
OFFThe output of the power supply is disabled (See page 56 for more
UnregThe output of the power supply is unregulated (output is neither CV
CVThe power supply is in constant voltage mode.
CCThe power supply is in constant current mode.
TRACKThe output1 and output2 are in track mode.
OVP1 annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
OVP2 annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
the error bit has not been cleared.
information).
nor CC).
To review the display annunciators, hold down key as you turn on
Display
Limit
the power supply.
*For E3646A/48A models. **For E3647A/49A models.
5
The Rear Panel at a Glance
1 AC inlet
2 Power-line fuse-holder assembly
3 Power-line module
• Select the GPIB or RS-232 interface (see chapter 3).
• Set the GPIB address (see chapter 3).
• Set the RS-232 baud rate and parity (see chapter 3).
6
In This Book
Quick Start Chapter 1 helps you get familiar with a few of the power supply’s
the front panel feature.
General Information Chapter 2 contains a general description of your
power supply. This chapter also provides instructions for installation of your
power supply and the output connections.
Front-Panel Operation Chapter 3 describes in detail the use of front-panel
keys and how they are used to operate the power supply from the front panel.
This chapter also shows how to configure the power supply for the remote
interface and gives a brief introduction to the calibration features.
Remote Interface Reference Chapter 4 contains reference information to
help you program the power supply over the remote interface. This chapter
also explains how to program for status reporting.
Error Messages Chapter 5 lists the error messages that may appear as you
are working with the power supply. Each listing contains information to help
you diagnose and solve the problem.
Application Programs Chapter 6 contains some remote interface
applications to help you develop programs for your application.
Tutorial Chapter 7 describes basic operation of linear power supplies and
gives specific details on the operation and use of your power supply.
Specifications Chapter 8 lists the power supply’s specifications.
Service Information Contains guidelines to return your power supply to
Agilent Technologies for servicing, procedures for verification & calibration,
and replaceable parts list and component locator diagram & schematics.
If you have questions relating to the operation of the power supply, call
1-800-452-4844 in the United States, or contact your nearest Agilent
Technologies Sales Office.
If your power supply fails within three years of purchase, Agilent will repair
or replace it free of charge. Call 1-800-258-5165 ("Express Exchange") in
the United States, or contact your nearest Agilent Technologies Sales Office.
One of the first things you will want to do with your power supply is to become
acquainted with the front panel. The exercises in this chapter prepare the
power supply for use and help you get familiar with some of its front-panel
operations.
This chapter is intended for both the experienced and the inexperienced user
because it calls attention to certain checks that should be made prior to
operation.
Throughout this chapter the key to be pressed is shown in the left margin.
16
Chapter 1 Quick Start
Preliminary Checkout
Preliminary Checkout
The following steps help you verify that the power supply is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your power supply. If
anything is missing, contact your nearest Agilent Technologies Sales Office.
One power cord for your location.
This User’s Guide.
Quick Reference Guide.
Certificate of Calibration.
2 Connect the power cord and turn on the power supply.
The front-panel display will light up briefly while the power supply performs
its power-on self-test. The GPIB address is also displayed. To review the power-
on display with all annunciators turned on, hold down as you turn on
the power supply. If the power supply does not turn on properly, see page 20.
Display
Limit
1
3 Perform a complete self-test.
The complete self-test performs a more extensive set of tests than those
performed at power-on. Hold down as you turn on the power supply and
Display
Limit
hold down the key until you hear a long beep. The self-test will begin when
you release the key following the beep.
If the self-test fails, see the Service Information for instructions on returning
the power supply to Agilent Technologies for service.
Note The power supply is shipped from the factory with a power-line cord that has a plug
appropriate for your location. Your power supply is equipped with a 3-wire grounding
type power cord; the third conductor being the ground. The power supply is grounded
only when the power-line cord is plugged into an appropriate receptacle. Do not
operate your power supply without adequate cabinet ground connection.
17
Powe r
Output
On/ Off
Chapter 1 Quick Start
Output Checkout
Output Checkout
The following procedures check to ensure that the power supply develops its
rated outputs and properly responds to operation from the front panel. For
complete performance and verification tests, refer to the Service Information.
Note: If an error has been detected during the output checkout procedures,
th e
ERROR annunciator will turn on. See "Error Messages" starting on page
121 in chapter 5 for more information.
Voltage Output Checkout
The following steps verify basic voltage functions with no load.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the outputs are
disabled (the
selected. The annunciator
indication annunciator turn on (for example,
E3646A model); and the knob is selected for voltage control.
2 Enable the outputs.
OFF annunciator turns on); its output1 and low voltage range are
OVP1, OVP2, CV, OUT1, and low voltage range
8V annunciator turns on for the
The
OFF annunciator turns off and the CV annunciator turns on. Notice that
the display is in the meter mode. ‘‘Meter mode’’ means that the display shows
the actual output voltage and current.
3 Check that the front-panel voltmeter properly responds to knob
control for both low and high voltage range.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero. The flashing
digit can be adjusted by turning the knob.
1
4 Ensure that the voltage can be adjusted from zero to the full rated
value by adjusting the knob.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the voltage.
18
Powe r
Chapter 1 Quick Start
Output Checkout
Current Output Checkout
The following steps check basic current functions with a short across the
power supply’s output.
1 Turn on the power supply.
Make sure that the output is disabled. The
2 Connect a short across (+) and (-) output terminals with an insulated
test lead.
OFF annunciator is on
1
Output
On/ Off
Display
Limit
Voltage
Curre n t
Use a wire size sufficient to handle the maximum current (See "
Table 2-1 Wire
Rating" on page 32 in chapter 2).
3 Enable the output.
The CV or CC annunciator turns on depending on the resistance of the test
lead. Notice that the display is in the meter mode.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the
Adjust the voltage limit to 1.0 volt to assure CC operation. The
will turn on. To go back to normal mode, press the key again or let the
Limit annunciator will be flashing).
Display
Limit
CC annunciator
display time out after several seconds.
5 Set the knob to the current control to check that the front-panel
ammeter properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the meter
mode (the
Limit annunciator is off). Check that the ammeter responds to knob
control and the voltmeter indicates nearly zero (the voltmeter will show the
voltage drop caused by the test lead). The flashing digit can be adjusted by
turning the knob.
1
6 Ensure that the current can be adjusted from zero to the full rated
value .
7 Turn off the power supply and remove the short from the output
ter minals.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
19
Chapter 1 Quick Start
If the Power Supply Does Not Turn On
If the Power Supply Does Not Turn On
Use the following steps to help solve problems you might encounter when
turning on the instrument. If you need more help, refer to chapter 5 for
instructions on returning the instrument to Agilent Technologies for service.
1 Verify that there is ac power to the power supply.
First, verify that the power cord is firmly plugged into the power receptacle on
the rear panel of the power supply. You should also make sure that the power
source you plugged the power supply into is energized. Then, verify that the
power supply is turned on.
2 Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the power
supply is shipped from the factory. Change the voltage setting if it’s not correct.
The settings are: 100, 115, or 230 Vac.
3 Verify that the correct power-line fuse is installed.
The correct fuse is installed for your country when the power supply is shipped
from the factory. See the table below to replace the fuse for your power supply.
ModelAgilent Part Number Part Description
E3646A/47A2110-1070Fuse 2.5A T 125V for 100 and 115 Vac
E3646A/47A2110-0457Fuse 1A T 250V for 230Vac
E3648A/49A2110-0996Fuse 4A T 250V for 100 and 115 Vac
E3648A/49A2110-1068Fuse 2A T 250V for 230 Vac
See the next page if you need to change the line voltage setting and the powerline fuse.
20
Chapter 1 Quick Start
Line Voltage Conversion
Line Voltage Conversion
Warning Shock Hazard Operating personnel must not remove power supply covers.
Component replacement and internal adjustments must be made only by
qualified service personnel.
Line voltage conversion is accomplished by adjusting two components: the line
voltage selection switch and the power-line fuse on the rear panel.
1 Remove AC line power.
2 Remove the cover (Refer to General Disassembly on page 174).
3 Set two sections of the line voltage selector switch on the PC board for the
desired line voltage (See Figure 1-1 below).
4 See the next page to check the rating of the power-line fuse and replace with
the correct one if necessary.
5 Replace the cover and mark the power supply clearly with a tag or label
indicating the correct line voltage and fuse that is in use.
1
(TOP VIEW)
Figure 1-1. Line Voltage Selector (set for 115Vac)
100V
11 5V
230V
21
Chapter 1 Quick Start
Line Voltage Conversion
1 Remove the power cord. Remove
the fuse-holder assembly with a flatblade screwdriver from the rear panel.
3 Replace with the correct fuse.
2 Remove the fuse-holder from the
assembly.
4 Replace the fuse-holder assembly in
rear panel.
Verify that the correct line voltage is selected and the power-line fuse is good.
22
Chapter 1 Quick Start
To Rack Mount the Instrument
To Rack Mount the Instrument
You can mount the power supply in a standard 19-inch rack cabinet using one
of three optional kits available. Instructions and mounting hardware are
included with each rack-mounting kit. Any Agilent Technologies System II
instrument of the same size can be rack-mounted beside the Agilent E3646A,
E3647A, E3648A, or E3649A.
Note: Remove the front and rear rubber bumpers before rack-mounting the
instrument.
1
FrontRear (bottom view)
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
23
Chapter 1 Quick Start
To Rack Mount the Instrument
To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and
flange kit 5063-9214.
To install one or two instruments in a sliding support shelf, order shelf 5063-9256,
and slide kit 1494-0015.
24
2
General Information
General Information
This manual describes the operation of the Agilent Technologies Model
E3646A, E3647A, E3648A, and E3649A DC power supplies.
This chapter contains a general description of your power supply. This chapter
also provides instructions for installation of your power supply and the output
connections. Unless otherwise stated, the information in this manual applies
to all the four models. This chapter is divided into the following sections:
• Safety Considerations‚ on page 27
• Options and Accessories‚ on page 28
• Description‚ starting on page 29
• Installation‚ on page 31
• Output Connections‚ on page 32
26
Chapter 2 General Information
Safety Considerations
Safety Considerations
This power supply is a Safety Class I instrument, which means that it has a
protective earth terminal. That terminal must be connected to earth ground
through a power source with a 3-wire ground receptacle.
Before installation or operation, check the power supply and review this
manual for safety markings and instructions. Safety information for specific
procedures is located at the appropriate places in this manual. See also
‘‘Safety’’ at the beginning of this manual for general safety information.
Safety and EMC Requirements
This power supply is designed to comply with the following safety and EMC
(Electromagnetic Compatibility) requirements:
• IEC 1010-1(1990)/EN 61010-1(1993) + A2 (1995): Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory Use
• CSA C22.2 No.1010.1-92: Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
• EN61326-1(1997):
EN 61000-4-2(1995): Electrostatic Discharge Requirements
EN 61000-4-3(1996): Radiated Electromagnetic Field Requirements
EN 61000-4-4(1995): Electrical Fast Transient/Burst Requirements
EN61000-4-5(1995): Surge Requirements
EN61000-4-6(1996): Conducted Radio Frequency Immunity Requirements
EN61000-4-8(1993): Magnetic Field Requirements
EN61000-4-11(1994): Voltage dips, short, interruption and var Requirement
EN 55011(1991) Group 1, Class A/CISPR 11(1990): Limits and Methods of
Radio Interference Characteristics of Industrial, Scientific, and Medical
(ISM) Radio - Frequency Equipment
• Low Voltage Directive 73/23/EEC
• EMC Directive 89/336/EEC
2
27
Chapter 2 General Information
Options and Accessories
Options and Accessories
Options
Option ‘‘0E3’’ and ‘‘0E9’’ determine which power-line voltage is selected at the
factory. The standard unit is configured for 115 Vac ± 10%, 47-63 Hz input
voltage. For more information about changing the power-line voltage setting,
see Line Voltage Conversion on page 21.
OptionDescription
0E3230 Vac ± 10%, 47-63 Hz input voltage
0E9100 Vac ± 10%, 47-63 Hz input voltage
1CMRack mount kit (Agilent part number 5063-9243)
910Extra manual (same language as the selected language manual
when you order the power supply)*
Accessories
The accessories listed below may be ordered from your local Agilent
Technologies Sales Office either with the power supply or separately.
Agilent No.Description
10833AGPIB cable, 1 m (3.3 ft.)
10833BGPIB cable, 2 m (6.6 ft.)
34398A RS-232, 9 pin (f) to 9 pin (f), 2.5 m (8.2 ft.) cable; plus 9 pin (m) to
25 pin (f) adapter
34399A RS-232 adapter kit (contains 4 adapters):
9 pin (m) to 25 pin (m) for use with PC or printer
9 pin (m) to 25 pin (f) for use with PC or printer
9 pin (m) to 25 pin (m) for use with modem
9 pin (m) to 9 pin (m) for use with modem
*To order an extra English User’s guide, order Agilent part number
E3646-90001.
28
Chapter 2 General Information
Description
Description
This power supply features a combination of programming capabilities and
linear power supply performance that makes it ideal for power systems
applications. The power supply may be programmed locally from the front
panel or remotely over the GPIB and RS-232 interfaces. This power supply has
two ranges, allowing more voltage at a lower current or more current at a lower
voltage. The output range is selected from the front panel or over the remote
interfaces.
Operational features include:
• Dual-output dual range
• Constant voltage (CV) or constant current (CC) operation
• Overvoltage protection (OVP)
• Five storage locations (1 to 5) for user-defined operating states
• Automatic turn-on self-test
• Remote sense at rear panel terminals
• User calibration from the front panel or over the remote interfaces
2
The front panel operation permits:
• Easy-to-use control features
• Output (1 or 2) and output voltage range selection
• Enabling or disabling OVP
• OVP trip level setting and condition clearing
• Setting and displaying the voltage and current limit values
• Operating state storage/recall
• Resetting the power supply to power-on state
• Returning the power supply to local mode from remote mode
• Retrieving/Scrolling error messages on the display
• Changing display mode (V-V, I-I, or V-I display)
• Viewing the errors, calibration string, or system firmware revision
• Enabling or disabling Tracking mode
• Calibrating the power supply, including changing calibration secure code
• Remote interface configuration
• Enabling or disabling the output
29
Loading...
+ 178 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.