Agilent E3646A Data Sheet

Agilent
E3640A – E3649A
Programmable DC Power Supplies
Data Sheet
Reliable Power, Repeatable Results
Single and dual outputs
Dual range output
Front and rear output terminals
Over-voltage protection
Remote Sensing
GPIB and RS-232 standard
Save and recall functions
Great Performance, Outstanding Price
With the output power of 30 to 100 W, the Agilent E364xA Series programmable DC power supplies provide great performance at a great price. All ten models deliver clean power, excellent regulation, fast transient response and built-in GPIB and RS-232 interfaces. They are designed to meet the needs of R&D design verifi cation, production testing, QA verifi cations, and other demanding applications with Agilent Technologies’s quality and reliability.
Steady Output
With 0.01 percent load and line regulation, Agilent E364xA power supplies are able to maintain a steady output when power line and load changes occur. They also specify normal mode voltage noise and low common mode current noise. The low normal mode noise specifi cation assures clean power for precision circuitry applications, and the low common mode current provides isolation from power line current injection. Agilent E364xA power supplies specify less than 90 msec of voltage settling time at any output load condition.
Remote Interface
Agilent E364xA power supplies support any PC with a GPIB (IEEE-488) card or RS-232 interface. Every model ships standard with both GPIB and RS-232. The easy-to-use SCPI (Standard Commands for Programmable Instruments) allow fast and simple programming procedures. Besides, the user manual provides suffi cient information on programming for all end users, from beginners to veterans.
Broad Support
VXI plug and play software drivers are available for Agilent VEE, National Instruments LabViewTM and LabWindowsTM. With these drivers, integration of the E364xA into your system can never be any easier. The drivers are supported under Microsoft® Windows 98® and NT®.
Front Panel Operation
An easy-to-use rotary knob and self­guiding keypads allow you to set the output at your desired resolution without any effort. Also, both voltage and current levels can be set to a maximum resolution of 10 mV/1 mA from the front panel. In addition, you can store and recall for up to fi ve complete power supply setups using the internal non-volatile memory.
The output on/off button sets the output to zero. If you own a dual output model, you can view two voltages or currents that are displayed simultaneously.
Versatile Power
Agilent E364xA power supplies give you the fl exibility to select from dual output ranges. Output load is protected against overvoltage, which can be easily monitored and adjusted from the front panel and remote interface. Remote sensing is available in the rear terminal to eliminate errors caused by voltage drops on the load leads. These power supplies offer new versatile binding posts on the front panel and screw-type terminals on the rear panel. New front panel binding posts allow you to use safety test leads as well as conventional banana clips and stripped wires. An optional rackmount kit is available. The Agilent E364xA Series employs a cooling fan with automatic speed control to reduce the acoustic noise.
LabView and LabWindows are registered trademarks of National Instruments. Microsoft Windows 98 and Windows NT are US registered trademerks of Microsoft Corporation.
2
Set the overvoltage protection and enable the overvoltage protection circuit
Highly visible vacuum fl ourescent display
Selectable dual range provides fl exibility and convenience
Store and recall up to fi ve different operating statuses in non­volatile storage locations
1 year warranty protects your investment
Electronic calibration from the front panel
Rotary knob for quick and analog-like control of voltage and current
Tough handle for easy carriage and unit prop-up
Versatile binding posts offer fl exibility to use safety test leads, banana plugs or stripped wires
Built-in GPIB and RS-232 interfaces
Cooling fan with automatic speed control for low acoustic noise
Rear output and sensing terminals
3
Agilent E3640A – E3649A Programmable DC Power Supply Specifi cations
Model Number E3640A E3641A E3642A E3643A E3644A E3645A
Maximum Power 30 W 50 W 80 W
# of Outputs 11111 1
DC Output Rating
(@ 0 ºC to 40 ºC)
Net Weight 5.3 kg 5.2 kg 6.3 kg 6.2 kg 6.6 kg 6.7 kg
Dimensions (without bumper) 212.6 mm W x 88.5 mm H x 348.3 mm D
Model Number E3646A E3647A E3648A E3649A
Maximum Power 60 W 100 W
# of Outputs 2222
DC Output Rating
(@ 0 ºC to 40 ºC)
Net Weight 8.2 kg 8.0 kg 9.2 kg 9.1 kg
Dimensions (without bumper) 212.8 mm W x 133.0 mm H x 348.3 mm D
Load and Line Regulation ± (% of output + offset)
Voltage < 0.01% + 3 mV
Current < 0.01% + 250 µA
Ripple and Noise (20 Hz to 20 MHz)
Normal Mode Voltage < 5 mVpp/0.5 mVrms for 8 V/20 V models
Normal Mode Current < 4 mArms
Common Mode Current < 1.5 µArms
Accuracy
1
12 Months (@ 25 ºC ± 5 ºC), ± (% output + offset)
Programming
Voltage < 0.05% + 10 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.2% + 10 mA
Readback
Voltage < 0.05% + 5 mV (< 0.1% + 25 mV for output 2 of E3646/47/48/49 A)
Current < 0.15% + 5 mA (< 0.15% + 10 mA for output 2 of E3646/47/48/49 A)
Resolution
Program < 5 mV/1 mA
Readback < 2 mV/1 mA
Meter 10 mV/1 mA
Transient Response Less than 50 µsec for output to recover to within 15 mV following a change in output current from full load to half
Settling Time
2
OVP
Accuracy, ± (% output + offset) Activation Time
3
Temperature Coeffi cient per ºC ± (% output + offset)
Voltage < 0.01% + 3 mV (< 0.02% + 5 mV for output 2 of E3646/47/48/49A)
Current < 0.02% + 3 mA
1
Accuracy specifi cations are valid after a 1-hour warm-up with no load and calibration at 25 ºC.
2
Maximum time required for the output voltage to change from 1% to 99% or vice versa following the receipt of VOLTage or APPLy command via direct
GPIB or RS-232 interface.
3
Average time for output to start and drop after an OVP condition occurred.
0 to 8 V/3 A or 0 to 20 V/1.5 A
Tw o 0 to 8 V/3 A or 0 to 20 V/1.5 A
0 to 35 V/0.8 A or 0 to 60 V/0.5 A
< 1.5 msec, OVP 3 V/< 10 msec, OVP < 3 V
0 to 8 V/5 A or 0 to 20 V/2.5 A
Tw o
0 to 35 V/0.8 A or
0 to 60 V/0.5 A
0 to 35 V/1.4 A or 0 to 60 V/0.8 A
Tw o 0 to 8 V/5 A or 0 to 20 V/2.5 A
< 8 mVpp/1 mVrms for 35 V/60 V models
load or vice versa.
< 90 msec
< 0.5% + 0.5 V
0 to 8 V/8 A or 0 to 20 V/4 A
0 to 35 V/2.2 A or 0 to 60 V/1.3 A
Tw o
0 to 35 V/1.4 A or
0 to 60 V/0.8 A
4
Stability, constant load & temperature ± (% output + offset), 8 hrs
Voltage < 0.02% + 2 mV
Current < 0.1% + 1 mA
Remote Sensing
Max. voltage drop in each load lead
AC Input
(47 Hz – 63 Hz)
Warranty 1 year
Product Regulation Designed to comply with UL3111-1; certifi ed to CSA 22.2 No. 1010.1;
conforms to IEC 1010-1; complies with EMC directive 89/336/EEC (Group 1, Class A)
100 Vac ± 10% (Opt 0E9)/115 Vac ± 10% (Std)/230 Vac ± 10% (0E3)
1 V
Agilent E364xA Series Power Supplies E3640A 30-Watt Single Power Supply E3641A 30-Watt Single Power Supply E3642A 50-Watt Single Power Supply E3643A 50-Watt Single Power Supply E3644A 80-Watt Single Power Supply E3645A 80-Watt Single Power Supply E3646A 60-Watt Dual Power Supply E3647A 60-Watt Dual Power Supply E3648A 100-Watt Dual Power Supply E3649A 100-Watt Dual Power Supply
Accessories Included
User’s Guide, Quick Reference Guide and AC power cord
Power Options
Opt. 0E3 230 Vac ± 10% Opt. 0EM 1150 Vac ± 10% Opt. 0E9 100 Vac ± 10%
Other Options
Opt. 1CM Rackmount kit* – Single Output (P/N 5063-9240) – Dual Output (P/N 5063-9243) Opt. 0L2 Extra Manual Opt. UK6 Commercial calibration with test result data E3600A-100 Test lead kit
Rackmount Kits*
Agilent E3640A/41A/42A/43A/44A/45A To rackmount two instruments side-by-side Lock-link Kit (P/N 5061-9694) Flange Kit (P/N 5063-9212) To rackmount one or two instruments in a sliding support shelf Support Shelf (P/N 5063-9255) Slide Kit (P/N 1494-0015) required for support shelf For a single instrument, also order Filter Panel (P/N 5002-3999)
Agilent E3646A/47A/48A/49A To rackmount two instruments side-by-side Lock-link Kit (P/N 5061-9694) Flange Kit (P/N 5063-9214) To rackmount two instruments in a sliding support shelf Support Shelf (P/N 5063-9256) Silde Kit (P/N 1494-0015)
* Rackmounting with 1CM or lock-link/ fl ange kit requires Agilent or customer support rails Agilent Support Rails-E3663AC
5
Agilent Email Updates
t
www.agilent.com/fi nd/emailupdates
Get the latest information on the products and applica­tions you select.
Agilent Direc
www.agilent.com/fi nd/agilentdirect
Quickly choose and use your test equipment solutions with confi dence.
Agilent
Open
www.agilent.com/fi nd/open
Agilent Open simplifi es the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system­ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development.
is the US registered trademark of the LXI Consortium.
Remove all doubt
Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confi dence in your measurements.
Agilent offers a wide range of additional expert test and measurement services for your equip­ment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management.
For more information on repair and calibration services, go to
www.agilent.com/find/removealldoubt
www.agilent.com
For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus
Phone or Fax
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800 (fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Taiwan:
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacifi c Countries:
(tel) (65) 6375 8100 (fax) (65) 6755 0042 Email: tm_ap@agilent.com
Revised: 1⅛ /06
Product specifi cations and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2002-2007 Printed in USA, October 26, 2007 5968-7355EN