The Agilent E3633A and Agilent E3634A are high performance 200 watt single-output dual range programmable DC power supplies with both GPIBandRS-232 interfaces. Thecombination of bench-top and system features in these power supplies provides versatile solutions for your design and testrequirements.
Convenient bench-top features
•Single-output dual range
•Easy-to-use knob controlsettings
•Highly visible vacuum-fluorescent display meters
•High accuracy andhigh resolution
•Remote voltage sensing
•Overvoltage and overcurrent protection
•Output on/off
•Excellent load and line regulation and low ripple and noise
•Operating states storage
•Portable, ruggedized case with non-skid feet
•Front and Rear output terminals
•Retrieving/Scrolling error messages on the display
Flexible system features
•GPIB (IEEE-488) and RS-232 interfaces are standard
test and calibration / or enables calibration mode (the power supply must be
unsecured before performing calibration). See Service Guide for more details
on calibration.
9 I/O Configuration / Secure key3 Configures the power supply for remote
interfaces / or secure or unsecure the power supply for calibration. See
Service Guide for more details on how to secure or unsecure the power supply.
10 Output On/Off key Enables or disables the power supply output. This key
toggles between on and off.
11 Control knob Increases or decreases the value of the blinking digit by turning
clockwise or counter clockwise.
12 Resolution selection keys Move the blinking digit to the right or left.
13 Voltage/current adjust selection key Selects the knob control function for
voltage or current adjustment.
1
The key can be used as the ‘‘
Local
’’ key when the power supply is in the remote
interface mode.
2
You can enable the ‘‘calibration mode’’ by holding down this key when you
turn on the power supply.
3
You can use it as the ‘‘Secure’’ or ‘‘Unsecure’’ key when the power supply is
in the calibration mode.
*For Agilent E3633A Model **For Agilent E3634A Model
iii
Front-Panel Voltage and Current Limit Settings
You can set the voltage and current limit values from the front panel using the
following method.
Use the voltage/current adjust selection key, the resolution selection keys,
and the control knob to change the voltage and current limit values.
1 Select the desired range using the range selection keys after turning on the
power supply.
2 Press the
3 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired voltage limit
by turning the control knob. If the display limit times out, press the
again.
4 Set the knob to current control mode by pressing the key.
5 Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired current limit
by turning the control knob.
6 Press the
will go to output monitoring mode automatically to display the voltage and
current at the output or the display will go to output monitoring mode
immediately by pressing the
Display
key to show the limit values on the display.
Limit
Vol tage
Current
Output
key to enable the output. After about 5 seconds, the display
On/Off
Output
key again.
On/Off
Display
Limit
key
Note All front panel keys and controls can be disabled with remote interface commands.
The Agilent E3633A and Agilent E3634A must be in "Local" mode for the front panel
keys and controls to function.
iv
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
8VShows the 8V/20A range is selected. (Agilent E3633A model)
20VShows the 20V/10A range is selected. (Agilent E3633A model)
25VShows the 25V/7A range is selected. (Agilent E3634A model)
50VShows the 50V/4A range is selected. (Agilent E3634A model)
OVPThe overvoltage protection function is enabled when the
annunciator turns on or the overvoltage protection circuit has
caused the power supply to shutdown when the annunciator blinks.
OCPThe overcurrent protection function is enabled when the
annunciator turns on or the overcurrent protection circuit has
caused the power supply to shutdown when the annunciator blinks.
CALThe power supply is in calibration mode.
LimitThe display shows the limit values of voltage and current.
ERROR Hardware or remote interface command errors are detected and
the error bit has not been cleared.
OFFThe output of the power supply is disabled (See page 52 for more
information).
UnregThe output of the power supply is unregulated (output is neither CV
nor CC).
CVThe power supply is in constant voltage mode.
CCThe power supply is in constant current mode.
To review the display annunciators, hold down key as you turn on
Display
Limit
the power supply.
v
The Rear Panel at a Glance
1 Power-line voltage setting
2 Power-line fuse-holder assembly
3 AC inlet
• Select the GPIB or RS-232 interface (see chapter 3).
• Set the GPIB bus address (see chapter 3).
• Set the RS-232 baud rate and parity (see chapter 3).
vi
In This Book
General Information
power supply. This chapter also provides instructions for checking your power supply, connecting to ac power, and selecting power-line voltage.
Initial Operation
outputs and properly responds to operation from the front panel.
Front-Panel Operation
keys andhow they are used to operate the power supply from the front panel.This chapter also shows how to configure the power supply for the remote interface and gives a brief introduction to the calibrationfeatures.
Remote Interface Reference
help you program the power supply over theremote interface. This chapter also explains how to program for status reporting.
Error Messages
are working with the power supply. Each listing contains information to help you diagnose and solve the problem.
Application Programs
applications to help you develop programs for your application.
Tutorial
gives specific details on the operation and use of the Agilent E3633Aand AgilentE3634A power supplies.
Chapter 7 describes basic operation of linear power supplies and
Chapter 1 contains a general description of your
Chapter 2ensures that the power supply develops its rated
Chapter 3 describes in detailtheuse of front-panel
Chapter 4 containsreference informationto
Chapter 5 lists the error messages that may appear as you
Chapter 6 containssome remote interface
Specifications
If you have questions relating to the operation of the power supply, call
1-800-829-4444 in the United States, or contact your nearest Agilent
Technologies Sales Office.
If your Agilent E3633A or Agilent E3634A fails within one year of purchase,
Agilent will repair or replace it free of charge. Call 1-800-258-5165 ("Express Exchange") in the United States, or contact your nearest Agilent Technologies
Sales Office.
This is the User’s guide for your Agilent E3633A and E3634A DC power supplies. Unless otherwise stated, the information in thismanual applies toboth two models.
This chapter provides a general description of your powersupply. This chapter also contains instructions for initialinspection, location and cooling for bench and rack operation, selectingthe power-line voltage, and connecting your power supply to ac power.
Safety Considerations
This power supplyis a Safety Class I instrument, which means that ithas a protective earth terminal. Thatterminal must be connected to earth groundthrough a powersource with a 3-wire ground receptacle.
Before installation or operation, check the powersupply and review this manual for safety markings and instructions. Safety information for specific procedures is located at the appropriate places in this manual. See also ‘‘
Safety
’’ at the beginningof this manual for general safetyinformation.
Safety and EMC Requirements
This power supplyis designed to comply with the following safety and EMC(Electromagnetic Compatibility) requirements:
•IEC 1010-1(1990)/EN 61010-1(1993) + A2 (1995):Safety Requirements for
Electrical Equipment for Measurement, Control, and Laboratory Use
•CSA C22.2 No.1010.1-92: Safety Requirements for Electrical Equipmentfor
Measurement, Control,and Laboratory Use
•EN50082-1(1992):
IEC1000-4-2(1995): Electrostatic Discharge RequirementsIEC 1000-4-3(1995):Radiated Electromagnetic Field Requirements
IEC 1000-4-4(1995):Electrical Fast Transient/Burst RequirementsEN61000-4-5(1995): Surge RequirementsEN61000-4-6(1995): Conducted Radio Frequency Immunity RequirementsEN61000-4-8(1993): Magnetic Field RequirementsEN61000-4-11(1994): Voltage dips,short, interruption andvar Requirement
•Low Voltage Directive 73/23/EEC
•EMC Directive 89/336/EEC
2
Chapter 1 General Information
Options and Accessories
•EN 55011(1991) Group 1, Class A/CISPR 11(1990): Limits and Methods of Radio Interference Characteristics of Industrial, Scientific, and Medical (ISM) Radio - Frequency Equipment.
•ICES/NMB-001This ISM device complies with Canadian ICES-001.Cet appareil ISM est conforme à la norme NMB-001 du Canada.
Options and Accessories
Options
Option 0EM, 0E3, and 0E9determine which power-linevoltage is selected at the factory.The standard unit is configured for 115 Vac ± 10%,47-63 Hz input voltage. Formore information about changing the power-line voltage setting, see ‘‘
Power-Line Voltage Selection
OptionDescription
0EM115
0E3
0E9
1CM
0L2
Vac ± 10%, 47-63Hz input voltage230 Vac ± 10%,47-63 Hz input voltage100 Vac ± 10%,47-63 Hz input voltageRack mount kit (Agilent partnumber 5063-9243)Extra Englishmanual set (local language manual files are includedon the CD-ROM, Agilent part number 5964-8251)
’’, starting on page22 in this chapter.
1
Accessories
The accessories listed below may be ordered from your local Agilent Technologies Sales Office either with the power supplyor separately.
Agilent No. Description
10833A
10833B
34398A
34399A
GPIB cable, 1 m (3.3 ft.)GPIB cable, 2 m (6.6 ft.)RS-232, 9 pin (f) to 9 pin (f),2.5 m(8.2 ft.) cable; plus 9 pin (m) to 25 pin (f) adapterRS-232 adapter kit (contains 4 adapters):9 pin (m) to 25 pin (m) for use with PCor printer9 pin (m) to 25 pin (f) for usewith PC or printer9 pin (m) to 25 pin (m) for use with modem9 pin (m) to 9 pin (m) for usewith modem
3
Chapter 1 General Information
Description
Description
The Agilent E3633A and Agilent E3634A DC power supplies featurea combination of programming capabilities and linear power supply performance that makes them ideal for power systems applications. The power supply is programmable locally from the front panel or remotely over the GPIBand RS-232 interfaces. This power supply has two ranges, allowing more voltage at a lower current. An output range needed is selected from the front panel or over the remote interfaces.
Operational features include:
•Dual range of single-output: 8V/20A and 20V/10A (Agilent E3633A), 25V/7A and 50V/4A (Agilent E3634A)
•Constant voltage(CV) or constant current (CC) operation
•Overvoltage protection (OVP) and overcurrent protection (OCP)
•Three storage locations (1 to 3) for user-defined operating states
•Automatic turn-on self-test
•Remote sensing for load voltage at the front or rear panel terminals
•User calibration from the front panel or over the remote interfaces
The front panel operation permits:
•Easy-to-useof knob control
•Output range selection
•Enabling or disablingOVP and OCP features
•Setting the OVP and OCP trip levels
•Clearing OVP and OCP conditions
•Setting and displaying thevoltage and current limit values
•Saving and recalling operating states
•Returning the power supplyto local mode from remote interface mode
•Retrieving/Scrolling error messages on the display
•Calibrating the powersupply,including changing the calibration secure code
•Configuring the power supply for remote interfaces
•Enabling or disablingthe output
4
Chapter 1 General Information
Description
When operated over the remote interface, the power supply can be both a listener and a talker.Using an external controller, you caninstruct the power supply toset the output and to send the statusdata back over the GPIB or RS-232. Capabilitiesinclude the following features:
•Voltage and current programming
•Voltage and current readback
•Present and stored status readback
•Programmingsyntax error detection
•Complete self-test
The front-panelVFD (Vacuum-Fluorescent Display) includes:
•Displaying actual values of output voltage and current (meter mode)
•Or displaying the limit values of voltageand current (limitmode)
•Checking the operating status fromthe annunciators
•Checking the type of error from the error codes (messages)
1
Connections to thepower supply’s output and to chassis ground are made tobindingposts on the
Warning Floating the power supply output more than ±60 Vdc from the chassis presents an
electric shock hazard to the operator. Do not float the outputs more than ±60 Vdc
when metal shorting bars without insulation are used to connect the (+) output to the
(+) sense and the (-) output to the (-) sense terminals.
front panel
andto the
rear output
terminals.
5
Chapter 1 General Information
Description
Warning Outputs can be floated to maximum of ±240 Vdc provided that the metal shorting bars
without insulation are either replaced with insulated conductors or they are removed
from the terminals so there is no operator access to the output conductors without
insulation. All field wiring insulation must be adequate for the voltage present.
The powersupply isshipped with a detachable, 3-wire grounding type power cord.The ac line fuse isan extractor type on the rear panel. The power supply can be calibrated from the frontpanel directly or with a controller over the GPIB or RS-232 interface using calibration commands. Correction factors are stored inCalibration from the front panel or a controller eliminate the need to remove the top cover or even the need to remove the power supply from your systemcabinet. You canguard against unauthorizedcalibration by using the “Secured” calibration protection function.
nonvolatile
memory and areused during output programming.
6
Chapter 1 General Information
Installation
Installation
Initial Inspection
When you receive your power supply, inspect it for any obvious damage that may have occurred during shipment. If any damage is found, notify thecarrier and the nearest Agilent Sales Officeimmediately.Warranty information is shown in the front of this manual.
Keep theoriginal packingmaterials in case the power supply has to be returnedto Agilent Technologies inthe future. If you return the power supply for service,attach atag identifying theowner and model number. Also include a brief description of the problem.
Mechanical Check
This checkconfirms that there are no broken keys or knob,that the cabinet and panel surfaces are free of dents and scratches,andthat the display is not scratched or cracked.
Electrical Check
Chapter 2 describes an initial operation procedure which, when successfully completed, verifies to a high level of confidence that the power supply isoperating in accordance with its specifications. Detailed electrical verification procedures are included in the
Service Guide
.
1
Cooling and Location
Cooling
The power supply can operate without loss of performance within the temperature range of 0 °Cto 40 °C,and with derated output current from40 °C to 55 °C. A fan cools the power supply by drawing air through the rear panel and exhausting it out the sides. Using an Agilent rack mount willnot impede the flow of air.
Bench Operation
Your power supply must be installed in a location that allows sufficient spaceatthe sides and rear of thepower supply for adequate air circulation. The rubber bumpers must be removed for rack mounting.
7
Chapter 1 General Information
Installation
Rack Mounting
The power supply can be mounted in a standard 19-inch rackcabinet using one of three optionalkits available. A rack-mounting kit for a single instrument is available as Option 1CM (P/N 5063-9243). Installation instructions and hardwareare included with each rack-mountingkit. Any Agilent System II instrument of thesame sizecan be rack-mounted beside the Agilent E3633Aor E3634A DCpower supply.
Remove the front and rear bumpers before rack-mounting the power supply.
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
8
Chapter 1 General Information
Installation
To rack mount two instruments of the same depth side-by-side, order
lock-link kit 5061-9694 and flange kit 5063-9214.
1
To install two instruments in a sliding support shelf, order support shelf
5063-9256, and slide kit 1494-0015.
9
Chapter 1 General Information
Input Power Requirements
Input Power Requirements
You canoperate your power supply from a nominal 100 V, 115 V, or 230 V single phase ac power sourceat 47 to 63 Hz. An indication on the rear panel showsthe nominal input voltage set for the power supply at the factory.If necessary, you can change the power-line voltage setting according to the instructions on the next page.
Power-Line Cord
The power supply is shipped from the factory with a power-line cord that has a plug appropriate for your location. Contact the nearest Agilent Sales and Service Officeif the wrongpower-line cord is included with your power supply.Your power supply is equipped with a 3-wiregrounding type power cord; the third conductor being the ground. The power supply is grounded only whenthe power-line cord is plugged into an appropriate receptacle. Do not operate your power supply without adequate cabinet ground connection.
Power-Line Voltage Selection
Power-line voltage selection is accomplished by adjusting two components:power-line voltage selector and power-line fuse on the power-line module of the rear panel.To change the power-line voltage, proceed as follows:
10
Chapter 1 General Information
Input Power Requirements
1
Remove the power cord. Remove the
fuse-holder assembly with a flat-blade
screwdriver from the rear panel.
2
Install the correct line fuse. Remove
the power-line voltage selector from the
power-line module.
100 or 115 Vac, 6.3 AT fuse
230 Vac, 3.15 AT fuse
1
3
Rotate the power-line voltage selector
until the correct voltage appears.
100, 115, or 230 Vac
4
Replace the power-line voltage selector
and the fuse-holder assembly in the rear
panel.
11
Chapter 1 General Information
Input Power Requirements
12
2
Initial Operation
Initial Operation
There arethree basic tests in this chapter. The automatic power-on test includes a self-test that checks the internal microprocessors and allows the user visually to check the display.The output check ensuresthat the power supply develops its rated outputsand properly respondsto operation fromthe front panel.For complete performance and/or verification tests, refer to the
Service Guide
This chapter is intended for both the experienced and the inexperienced userbecause it calls attention to certain checks that should be madeprior tooperation.
Throughout this chapter the key to be pressed is shown in the left margin.
.
14
Chapter 2 Initial Operation
Preliminary Checkout
Preliminary Checkout
The following steps help you verify that the power supply is ready for use.
1Check the list of supplied items.
Verify that you have received the following items with your power supply. If anything is missing, contact your nearest Agilent Technologies Sales Office.
Onepower cord for your location.
ThisUser’s
One Certificateof Calibration.
2Verify the power-line voltage setting on the rear panel.
The power-line voltage isset to the proper value for your country when the power supply is shipped from the factory. Change the voltage setting if it is not correct. The settings are: 100, 115,or 230 Vac.
3Verify that the correct power-line fuse is installed.
The correct fuse is installed for your country when the power supply is shipped from the factory. For 100 or 115 Vac operation, you must usea 6.3 AT fuse. For 230 Vac operation, you must use a 3.15 AT fuse.
4Connect the power-line cord and turn on your power supply.
The front-panel display will light up and apower-on self-test occurs automaticallywhen you turn on the power supply.
See ‘‘Power-Line Voltage Selection’’,starting on page 22 in chapter 1 if you need to changethepower-line voltage or the power-linefuse.
Service
Guide.
Guide.
2
To replace the 6.3ATfuse, order Agilent part number2110-1030.To replace the 3.15 AT fuse, order Agilent part number2110-1031.
15
Chapter 2 Initial Operation
Power-On Checkout
Power-On Checkout
The power-on test includesan automatic self-test that checks the internalmicroprocessors and allows the uservisually to check the display.You willobserve the following sequence on the display after pressing the front panel power switch to on.
1The front-panel display will light up briefly while the instrument
performs its power-on self-test.
To review thepower-ondisplay with all annunciators turned on, hold down
Display
key as you turn onthe powersupply
Limit
2The GPIB address or RS-232 is also displayed for about one second.
ADDR 05 (or RS-232)
The GPIB address is setto ‘‘5’’ when the power supply is shipped from the factory for remote interface configuration. If this is not the first time the power supply is turned on, a different interface(RS-232) or a different GPIB addressmay appear.
See "Remote Interface Configuration" in chapter 3 starting on page 58 if you needto change the remote interface configuration.
.
3The “8V”* or “
25V
”**, “
OVP
”, “
OCP
” and “
OFF
” annunciatorsare on.
All others are off.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF
annunciator turns on); the 8/20A* or 25V/7A** range is selected(the
8V
* or
25V
** annunciator turns on); and the knob is selected for voltage
Output
On/Off
control.Notice that the
4 Enable the outputs.
The
OFF
annunciator turns offand the 8V* or
annunciators are lit. The
OVP
and
blinking
OCP
annunciator also turn on.
25V
**,
OVP, OCP
, and CV
digit canbe adjusted by turning the knob. Notice that the display isin the meter mode.‘‘Meter mode’’means that the display shows the actual output voltage and current.
Note If the power supply detects an error during power-on self-test, the ERROR
annunciator will turn on.
See "Error Messages" for more information starting on page
123 in chapter 5
*For Agilent E3633A Model **For Agilent E3634A Model
16
Power
Output
On/Off
Chapter 2 Initial Operation
Output Checkout
Output Checkout
The following procedures check to ensure that the power supply develops its rated outputs and properly responds to operation from the front panel.Forcomplete performance and verification tests, refer to the
For eachstep, use the keys shownontheleft margins.
Voltage Output Checkout
The following steps verify basic voltage functions with no load.
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for voltage
control.
2Enable the outputs.
The
OFF
annunciator turns offand the 8V* or annunciators are lit. The blinking digit can be adjusted by turning the knob.Notice that the display isin the meter mode.‘‘Meter mode’’means that the display shows the actual output voltage and current.
25V
**,
Service Guide
OVP, OCP
, and CV
.
2
3Check that the front-panel voltmeterproperlyresponds to knob
control for the 8V/20A* or 25V/7A** range.
Turn the knob clockwiseor counter clockwise to check that the voltmeter responds to knob control and the ammeter indicates nearly zero.
1
4Ensure that the voltage can beadjusted from zero to the full rated
value.
Adjust the knob untilthe voltmeter indicates 0 voltsand thenadjust the knob
8.0
until the voltmeterindicates ‘‘
1
You can use the resolution selection keys to move the blinking digitto the
volts’’*or ‘‘
25.0
volts’’**
right or left whensetting thevoltage.
*For Agilent E3633A Model **For Agilent E3634A Model
17
Power
Output
On/Off
Display
Limit
Vol tage
Current
Chapter 2 Initial Operation
Output Checkout
Current Output Checkout
The following steps check basic current functionswith a short across thepower supply’s output.
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for voltage
control.
2Connect a short across (+) and (-) output terminals with an insulated
test lead.
Use a wire size sufficient to handle the maximumcurrent (See "Table 7-1 Wire Rating" on page 153 in chapter 7).
3Enable the output.
The
OFF
annunciator turns offand the 8V* or
annunciators are lit. The
CV
or CC annunciator turns on depending on the resistance of the test lead. The blinking digit can be adjusted byturning the knob. Notice that the display is in the meter mode. ‘‘Meter mode’’ means that the display shows the actual output voltage and current.
4Adjust the voltage limit value to 1.0 volt.
Set the display to the limitmode (the
Limit
the voltage limit to 1.0 volt to assure CCoperation. The turn on. To go back tonormalmode,press the time out for several seconds .
5Set the knob to the current control to check that the front-panel
ammeter properly responds to knob control.
Turn the knob clockwise or counter clockwisewhen thedisplay is in the meter
Limit
mode (the
annunciator is off). Check that the ammeter responds to knob control and the voltmeter indicates nearly zero (the voltmeter will show thevoltage drop caused by the test lead).
25V
**,
OVP
, and
OCP
annunciator will be blinking). Adjust
CC
Display
key again or let the display
Limit
annunciatorwill
*For Agilent E3633A Model **For Agilent E3634A Model
18
Chapter 2 Initial Operation
Output Checkout
1
6Ensure that the current canbe adjusted from zero to the full rated
value.
Adjust the knob until the ammeter indicates 0 amps and thenuntil the ammeter indicates 20.0 amps* or 7.0 amps**.
7Turn off the power supply and remove the short from the output.
1
You can use the resolution selection keys to move the blinking digitto the
right or left whensetting thecurrent.
Note If an error has been detected during the output checkout procedures, the ERROR
annunciator will turn on.
123 in chapter 5
See "Error Messages" for more information starting on page
2
*For Agilent E3633A Model **For Agilent E3634A Model
19
Chapter 2 Initial Operation
Output Checkout
20
3
Front-Panel Operation
Front-Panel Operation
So far you have learned how to install your power supply and perform initial operation. During the initial operation, youwere briefly introduced tooperatingfrom the front panel as you learned how to check basic voltage andcurrent functions. This chapter will describein detail the use of these front-panel keys and show how theyare used to accomplish power supplyoperation.
•Front-PanelOperation Overview,on page 23
•Constant Voltage Operation, on page24
•Constant Current Operation, on page 26
•Storing and Recalling Operating States, on page 28
•Programming Overvoltage Protection, on page 30
•ProgrammingOvercurrent Protection, on page 34
•Remote Voltage Sensingat the Front and Rear Terminals,on page 37
•Disabling the Output, on page 40
•Disabling the Output Usingan External Relay, on page 41
•Knob Locking, on page41
•System-Related Operations, on page 42
•Remote Interface Configuration, on page46
•HP-IB Interface Configuration, on page 51
•RS-232 Interface Configuration,on page 52
•Calibration Overview, on page 56
Throughout this chapter the key to be pressed is shown in the left margin.
Note See "Error Messages", starting on page 123 in chapter 5 if you encounter any errors
during front-panel operation.
22
Chapter 3 Front-Panel Operation
Front-Panel Operation Overview
Front-Panel Operation Overview
The following section describes an overview of the front-panel keys before operating your power supply.
•The powersupply isshipped fromthe factory configured in the front-panel operation mode. At power-on,the power supply is automatically set tooperate in the front-panel operation mode. When in this mode, the front-panel keys can be used. When the power supply is in remote operation mode, you can return to front-panel operation mode at any timeby pressing the
Store
(Local
Local
command. Achange between front-panel and remoteoperation modes willnot resultin a change in the output parameters.
•The power supply has two output ranges. This feature allows more voltage ata lower current or more current at a lower voltage. The desired outputrange is selected fromthe front panel or over the remote interfaces.The or
20V
the presently selected range.
•When you press key (the power supply goes to the displayed.In thismode,you canalso observe the change of the limit values when adjusting the knob.If you pressthe time-outafter several seconds, the power supplywill return the display to the meter mode (the output voltage and current will be displayed.
•The output ofthe power supply can be enabled or disabled from the front panel by pressing turns on and the output is disabled.
•The display providesthe present operating status of the power supply with annunciators and also informs the user oferror codes. For example, the powersupply isoperating inCVmode in the 8V/20A* or 25V/7A** range and controlled from the frontpanel, then thewill turn on. If, however, the power supply is remotely controlled, the annunciator will also turn on, and when the power supplyis beingaddressedover HP-IB interface, the
Annunciators
) key if you did not previouslysend the front-panel lockout
for the E3633A and
Display
Limit
Output
On/Off
25V
or
50V
for the E3634Aannunciator indicates
Limit
annunciator blinks), the display of the
limit
mode and the present limit values willbe
Display
key again or let the display
Limit
Limit
annunciator turns off). In this mode, theactual
key. When the output is off, the
CV
and8V* or
Adrs
annunciator will turn on. See “
OFF
25V
’ on page 5for more information.
8V
annunciator
** annunciators
Rmt
Display
3
*For HP E3633A Model **For HP E3634A Model
23
Power
Display
Limit
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Constant Voltage Operation
To set up the power supply for constant voltage (CV) operation, proceed asfollows.
•Front-panel operation:
1Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2Turn on the power supply.
The power supply will go into the (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for
20V,10A50V,4A
Press * or ** key to operatethe power supply in the 20V/10A* or 50V/4A** range before proceeding to thenext step. The annunciator turns on.
3Set the display to the limit mode.
Notice that the
Limit
annunciator blinks, indicating that the display is in the limit mode. Whenthedisplayis in thecurrent limit values of thepower supply.
power-on / reset
limit
mode, you can seethe voltage and
state; the output is disabled
20V
voltage
* or
50V
control.
**
Vol tage
Current
In
constant voltage
mode,thevoltage valuesbetween the meter and limit modesare the same,but the current values arenot. Moreover, if the display is inthe meter mode,youcannot see the change of current limit value when adjusting the knob. We recommend that you should setthe display to “limit” mode to seethechange of current limit value in the constant voltage mode whenever adjustingthe knob.
1
4Adjust the knob for the desired current limit.
Check that the
Limit
annunciatorstill blinks. Set the knob for
current
control.The second digit of the ammeter willbe blinking. Theblinking digitcan be changed using theresolution selection keys and the blinking digit can be adjusted by turning the knob. Adjust the knob to the desired current limit.
1
You can use the resolution selection keys to move the blinking digitto the
right or left whensetting current.
*For HP E3633A Model **For HP E3634A Model
24
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Vol tage
Current
Display
Limit
Output
On/Off
1
5Adjust the knob for the desired output voltage.
Check that the The second digit of the voltmeterwill be
Limit
annunciator still blinks. Set the knob for
blinking
. Change the blinkingdigit
voltage
control.
using the resolution selection keys and adjust the knob to the desired output voltage.
6Return to the meter mode.
Display
Press key or letthe display time-out after several seconds to return to
Limit
the meter mode. Notice that the
Limit
annunciator turns off and the display
shows “OUTPUTOFF” message.
3
7Enable the output.
The
OFF
annunciator turnsoff and the 8V* (or
OCP
and CV annunciators are lit. Notice that the display is in the
In the meter mode, the display shows the actual outputvoltage and current.
Refer to “Programming Overvoltage Protection” and “Programming Overcurrent Protection” sections, starting on information on
OVP
and
OCP
annunciators.
8Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify thatthe
CV
annunciator is lit. If the CCannunciator is lit, choose a higher current
limit.
25V
**) or
20V
* (or
50V
**),
OVP
meter
mode.
page30 and page34 for more
,
Note During actual CV operation, if a load change causes the current limit to be exceeded,
the power supply will automatically crossover to the constant current mode at the
preset current limit and the output voltage will drop proportionately.
•Remote interface operation:
CURRent {<current>|MIN|MAX}Set the current
VOLTage {<voltage>|MIN|MAX}Set the voltage
OUTPut ONEnable the output
1
You can use the resolution selection keys to move the blinking digitto the
right or left whensetting voltage.
*For HP E3633A Model **For HP E3634A Model
25
Power
Display
Limit
Chapter 3 Front-Panel Operation
Constant Current Operation
Constant Current Operation
To set up the power supply for constant current (CC) operation, proceed as follows.
•Front-panel operation:
1Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2Turn on the power supply.
The power supply will go into the (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for
To operate the power supply in the 20V/10A* or 50V/4A** range, press *
50V,4A
or ** key before proceedingto the next step. The annunciator turns on.
3Set the display to the limit mode.
Notice that the
Limit
annunciator blinks, indicating that the display is in the limit mode. Whenthedisplayis in thecurrent limit values oftheselectedsupply.
power-on / reset
limit
mode, you can seethe voltage and
state; the output is disabled
20V
voltage
* or
50V
control.
**
20V,10A
In
constant current
mode,thecurrent values between the meter mode and limit mode are the same, but the voltage values are not. Moreover, if the display is in the meter mode, you cannot see the change of voltagelimitvalue when adjusting the knob. We recommend thatyou should setthe display to “limit” mode to see the change of voltage limit value in the constant current mode whenever adjustingthe knob.
1
4Adjust the knob for the desired voltage limit.
Check thatthe
Limit
annunciator still blinks and the second digit of voltmeter blinks to indicate the knob is selected for voltage control. The blinking digit can be changed using the resolution keys and the blinking digit can be adjusted by turning the knob.Adjust the knobfor the desired voltage limit.
1
You can use the resolution selection keys to move the blinking digitto the
right or left whensetting thevoltage.
*For HP E3633A Model **For HP E3634A Model
26
Chapter 3 Front-Panel Operation
Constant Current Operation
Vol tage
Current
Display
Limit
Output
On/Off
1
5Adjust the knob for the desired output current.
Check that the
Limit
annunciator still blinks. Set the knob for current
control
. The second digit of the ammeter willbe blinking. Change the blinkingdigit using the resolution selection keys and adjust the knob to the desired output current.
6Return to the meter mode.
Display
Press key or let the display time-out after several seconds to return the
Limit
meter mode. Notice that the
Limit
annunciator turns off and the display shows
“OUTPUT OFF” message.
3
7Enable the output.
The
OFF
annunciator turnsoff and the 8V* (or
OCP
and CC annunciators are lit. Notice that the display is in the meter mode.
In the
meter
mode, the display shows the actual output voltage and current.
Refer to “Programming Overvoltage Protection” and “Programming Overcurrent Protection” sections, starting on information on
OVP
and
OCP
annunciators.
8Verify that the power supply is in the constant current mode.
If you operate the power supply in the constantcurrent (CC) mode, verify that the
CC
annunciator is lit. If the CVannunciator is lit, choose a highervoltage
limit.
25V
**) or
20V
* (or
50V
**),
OVP
page30 and page34 for more
,
Note During actual CC operation, if a load change causes the voltage limit to be exceeded,
the power supply will automatically crossover to constant voltage mode at the preset
voltage limit and the output current will drop proportionately.
•Remote interface operation:
VOLTage {<voltage>|MIN|MAX}Set the voltage
CURRent {<current>|MIN|MAX}Set the current
OUTPut ONEnable the output
1
You can use the resolution selection keys to move the blinking digitto the
right or left whensetting thecurrent.
*For HP E3633A Model **For HP E3634A Model
27
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Storing and Recalling Operating States
Store
You can storeup to three different operating states in
non-volatile
memory. This also enablesyou to recall the entire instrument configuration with just a few key presses fromthe front panel.
The memory locations are supplied with the reset statesfrom the factory for
*RST
front-panel operation. Refer to the description of
command, starting on page96 in chapter 4 for more information. The following steps show you how to store and recall an operating state.
•Front-panel operation:
1Set up the power supply for the desired operating state.
The storage feature “remembers” output rangeselection, the limit value settings of voltage and current, output on/off state,OVP and OCPon/off state, and OVP and OCP trip levels.
2Turn on the storage mode.
Three memory locations (numbered 1, 2 and 3) are availableto store the operating states. The operating states are stored in
non-volatile
memory and
are remembered when being recalled.
STORE 1
This message appears on the display for approximately 3 seconds.
3Store the operating state in memory location “3”.
Turn the knob to the right to specify thememory location 3.
STORE 3
To cancel the store operation
or press any other function key except the to the normaloperating mode and to the function pressed.Save the operating state.
28
, let the display time-out after about 3 seconds
Store
key. The power supply returns
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Store
Recall
Recall
4Save the operating state.
The operating state is now stored. To recall the stored state, go to the following steps.
DONE
5Turn on the recall mode.
Memory location “1” willbe displayed in the recallmode.
RECALL 1
This message appears on the display for approximately 3 seconds.
6Recall thestored operating state.
Turn the knob to the right to change the displayed storage location to 3.
RECALL 3
If this setting is not followed within 3 seconds with key stroke, the powersupply returns to normal operating mode and willnot recall the instrument state 3 frommemory.
7Restore the operating state.
The power supply should now be configured in the same state aswhen you stored the state on the previous steps.
Recall
3
DONE
This message appears on the display for approximately 1 second.
•Remote interface operation:
*SAV {1|2|3}Store an operating state to a specified location
*RCL {1|2|3}Recall a previously stored state from a specified location
29
Power
Output
On/Off
Over
Vol tage
Chapter 3 Front-Panel Operation
Programming Overvoltage Protection
Programming Overvoltage Protection
Overvoltage protection guards the load against output voltages that reach aspecified value greater than theprogrammed protection level. It isaccomplished by shorting the output via an internal SCR whenthe trip levelis set to equal or greater than 3volts,or byprogamming the output to 1 voltwhen the trip level isset tolessthan 3 volts.
The following steps show how to set the OVP trip level,how to check OVP operation, and how to clear overvoltage condition.
•Front-panel operation:
Setting the OVP Level and Enable the OVP Circuit
1Turn on the power supply.
The powersupply will go into the (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for
2Enable the output.
The
OFF
annunciator turns offand the display will go to the meter mode.
3Enter the OVP menu and set the trip level.
power-on / reset
state; the output is disabled
voltage
control.
Over
Vol tage
LEVEL 22.0V (E3633A)
LEVEL 55.0V (E3634A)
You willsee the abovemessage on the display when you enter the OVP menu.Adjust the control knob for thedesired OVP trip level.
Note that youcannot set the trip levels to lowerthan 1.0volt.
4Enable the OVP circuit.
OVP ON
You willsee the above message after pressing key.
*For HP E3633A Model **For HP E3634A Model
30
Over
Vol tage
Over
Vol tage
Chapter 3 Front-Panel Operation
Programming Overvoltage Protection
5Exit the OVP menu.
CHANGED
The “CHANGED”message is highlightedfor a second to show that the newOVP trip level is now in effect. If the OVP settings are not changed, “NO CHANGE”will be displayed.The power supply will exit the OVP menu andthe
OVP
display willreturn to the meter mode.Check that theon.
annunciator turns
Display
Limit
Over
Vol tage
Checking OVP Operation
3
To check OVP operation, raise the outputvoltageto near the trip point. Then verygradually increase the output byturning the knob until the OVP circuit
OVP
trips. This willcause the power supply output to drop to near zero, the annunciatorto blink, and the
CC
annunciator to turn on. The “OVP TRIPPED”
message also appears on the display.
Clearing the Overvoltage Condition
When the OVP condition occurs (the “OVP TRIPPED” messageis shown on the display), the OVP annunciator blinks. When it was caused by an externalvoltage source such as a battery, disconnectit first. Clear the overvoltage condition by adjusting output voltage level or by adjusting OVP trip level.
The following steps show how to clear the overvoltage condition and get backto normal mode operation. In thefollowing steps, the display will go back to “OVP TRIPPED” if you letthe display time out after about several seconds.
•Adjust output voltage level
1Lower the output voltage level.
Lower the output voltage level below theOVP trip point after pressing key. The
OVP
and
Limit
annunciators are blinking.
2Move to the clear mode.
Display
Limit
OVP CLEAR
Over
Press key twice to move to the OVP CLEAR mode. The“OVP ON”
Vol tage
message appears on the display. Turn the knob to the right until the above message appears on the display.
31
Over
Vol tage
Over
Vol tage
Over
Vol tage
Over
Vol tage
Chapter 3 Front-Panel Operation
Programming Overvoltage Protection
3Clear the overvoltage condition and exit this menu.
Now, when you press key again, the “DONE” message isdisplayed for a second and the
OVP
Over
Vol tage
annunciator will not blink any more. The output will return
to meter mode.
•Adjust OVP trip level
1Raise the OVP trip level.
Over
Press key and turn the knob to raisethe OVP trip level.
Vol tage
2Move to the OVP CLEAR mode.
OVP CLEAR
Over
Press key to moveto the OVP CLEAR mode. The “OVP ON” message
Vol tage
appears on the display.Turn the knob to the right untilthe above message appears on the display.
3Clear the overvoltage condition and exit this menu.
Now, when you presskey again, the “DONE’’ message is displayed for a second and thereturn to the meter mode.
Over
Vol tage
OVP
annunciator willnot blink any more. The output will
•Remote interface operation:
VOLT:PROT {<voltage>|MIN|MAX}Set the OVP level
VOLT:PROT:STAT {OFF|ON)Disable or enable the OVP circuit
VOLT:PROT:CLEClear the tripped OVP circuit
Note The power supply’s OVP circuit contains a crowbar SCR, which effectively shorts the
output of the power supply whenever the overvoltage condition occurs. If external
voltage source such as a battery is connected across the output, and the overvoltage
condition inadvertently occurs, the SCR will continuously sink a large current from
the source; possibly damaging the power supply. To avoid this a diode must be
connected in series with the output as shown in Figure 3-1 on next page.
32
Chapter 3 Front-Panel Operation
Programming Overvoltage Protection
Figure 3-1. Recommended Protection Circuit for Battery Charging
3
33
Power
Output
On/Off
Over
Current
Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
Programming Overcurrent Protection
Overcurrent protection guards the loadagainst output currents that reach a specified value greater than theprogrammed protection level. It isaccomplished by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level,how to check OCP operation and how to clear overcurrentcondition.
•Front-panel operation:
Setting the OCP Level and Enable the OCP Circuit
1Turn on the power supply.
The power supply will go into the (the
OFF
annunciator turnson); the 8V/20A* or 25V/7A** range is selected (the
8V
* or
25V
** annunciator turns on); and the knob is selected for
2Enable the output.
The
OFF
annunciator turns offand the display will go to the meter mode.
3Enter the OCP menu and set the trip level.
power-on / reset
state; the output is disabled
voltage
control.
Over
Current
LEVEL 22.0 A
LEVEL 7.5 A
(E3633A)
(E3634A)
You will see the above message on the display when you enter the OCP menu.Adjust the knob for the desired OCP trip level.
4Enable the OCP circuit.
OCP ON
You willsee the above message after pressing the key.
*For HP E3633A Model **For HP E3634A Model
34
Over
Current
Over
Current
Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
5Exit the OCP menu.
CHANGED
The “CHANGED” message isdisplayed for a second to showthat the new OCP trip level is now in effect. If the OCP settings are not changed, “NO CHANGE” will be displayed. The power supplywill exit the OCP menu and the display
OCP
will return to themeter mode. Check that the
annunciator turns on.
Display
Limit
Over
Current
Checking OCP Operation
To check OCP operation, raise the output current tonear the trippoint.Then
3
verygradually increase the output byturning the knob until the OCP circuit trips. This will cause the power supply’s output current to drop tozero and the
OCP
annunciator to blink. The “OCP TRIPPED” message also appears on the
display.
Clearing the Overcurrent Condition
When the OCP condition occurs (the“OCP TRIPPED” message is shown on the display), the OCP annunciator blinks. When it was caused by externalvoltage sources such as a battery, disconnect it first. Clear theovercurrent condition by adjusting output current level or by adjusting OCP trip level.
The following steps show how to clear the overcurrent condition and get back to normal mode operation. In thefollowing steps, the display will go back to “OCP TRIPPED” if you let the display timeout after about several seconds.
•Adjust output current level
1Lowerthe output current level.
Display
Press key andset the knobfor
Limit
current
control by pressing key,
then lower the outputcurrent level below the OCP trip point.
2Move to the clear mode.
Vol tage
Current
OCP CLEAR
Over
Press key twice to move to the OCP CLEAR mode. The “OCP ON”
Current
message appears on the display. Turn the knob to the right until the above message appears on the display.
35
Over
Current
Over
Current
Over
Current
Over
Current
Chapter 3 Front-Panel Operation
Programming Overcurrent Protection
3Clear the overcurrent condition and exit this menu.
Now, when you presskey again, the “DONE’’ message is displayed for a second and thereturn to meter mode. The knob is selected for
Over
Current
OCP
annunciator willnot blink any more. The output will
current
control.
Notice that the power supply is operated in the constant current (CC) mode.
•Adjust OCP trip level
1Raise the OCP trip level.
Over
Press key and turn the knob to raisethe OCP trip level.
Current
2Move to the OCP CLEAR mode.
OCP CLEAR
Press the key to move to the OCP CLEAR mode.The “OCP ON” message appears on the display.Turn the knob to the right untilthe above message appears on the display.
3Clear the overcurrent condition and exit this menu.
Now, when you presskey again, the “DONE’’ message is displayed for a second and theOCP annunciator will not blink any more. The output willreturn to the meter mode.
Over
Current
•Remote interface operation:
CURR:PROT {<current>|MIN|MAX}Set the OCP level
CURR:PROT:STAT {OFF|ON}Disable or enable the OCP circuit
CURR:PROT:CLEClear the tripped OCP circuit
36
Chapter 3 Front-Panel Operation
Remote Voltage Sensing at the Front and Rear Terminals
Remote Voltage Sensing at the Front and Rear Terminals
Remote voltage sensing is used to maintainregulation at the load and reduce the degradation of regulation that would occur due to the voltage drop in the leads between thepower supply and the load.
By connecting the power supply for remote voltage sensing, voltage is sensed at the load rather thanat the power supply’soutput terminals. This will allow the power supply to automatically compensate for the voltagedrop inapplicationswith long lead lengthsas well asto accurately read back thevoltage directly across the load.
When the power supply is connected forremote sensing, the OVP circuit senses the voltage at the
CV Regulation
The voltage load regulation specification in chapter 8 applies at theoutput terminals ofthe power supply.When remote sensing, add 5 mV to thisspecification for each 1 V drop between the positive sensing point and (+)output terminal due to the change in load current. Because the sense leads are part of thepower supply’s feedback path, keep the resistance of the sense leadsat or below0.5
sensing
Ω
per lead to maintain the above specified performance.
points (load) and not the output terminals.
3
Output Rating
The rated output voltage and current specifications in chapter 8 apply at the output terminalsof the powersupply. With remote sensing,any voltage dropped in the load leads must beadded to the load voltage to calculate maximum output voltage. The performance specifications are not guaranteed when the maximum output voltage is exceeded. If the excessivedemand on
Unreg
the power supply forces the power supplyto loseregulation, the annunciator willturn on to indicatethatthe output is unregulated.
Output Noise
Any noise picked up on the sense leads also appearsat the output of the power supply and may adversely affect the voltage loadregulation.Twist the senseleads to minimizeexternal noise pickup and run them parallel and close to the load leads.In noisy environments it may be necessary to shield the sense leads.Ground the shield at the power supply end only.
ofthe sense conductors
.
Do not use the shield as one
37
Chapter 3 Front-Panel Operation
Remote Voltage Sensing at the Front and Rear Terminals
Stability
Using remote sensing under certain combinations of load lead lengths andlarge load capacitances may cause your application to form a filter, which becomespart of the voltage feedback loop. The extra phase shift created by this filter can degrade the power supply’sstability,resulting inpoor transient response or loop instability.In severe cases, it may causeoscillations. Tominimizethis possibility, keepthe load leads as short as possible andtwistthem together. As the sense leads are partof thepower supply’s programmingfeedback loop, accidental open-connectionsof sense or load leads during remote sensing operation have various unwanted effects.Provide secure and permanent connections.
Remote Voltage Sensing Connections
Connections between the power supply sensingand output terminals should be removed, and usingshielded two-wire cable, the power supply sensing terminals should be connectedto the load as shownin Figure 3-2.
the shield as one of thesensing conductorsand the otherend shouldbeleft unconnected
⊥
) only. Opening a sensing lead causes the powersupply output voltage to
(decrease at the load leads. Observe polarity when connecting the sensing leads to the load.
For local voltage sensing connections, the (+) and (-) sense terminals must be connected to the (+) and (-) output terminals respectively.
. Connect oneend ofthe sensing lead shield to the chassis ground
Do not use
Note When you make the remote voltage sensing connections at the front or rear panel
terminals, make sure to disconnect all the connections to the load and sense leads
at the other end terminals. Do not make the sensing connections at both front and
rear terminals at the same time. It will cause to damage the power supply seriously.
Not e During remote sensing setup, it is strongly recommended to power off (by presssing
power ON/OFF button) the power supply to avoid undesirable damage to the load
or the power supply.
38
Chapter 3 Front-Panel Operation
Remote Voltage Sensing at the Front and Rear Terminals
Figure 3-2. Remote Voltage Sensing Connections
Remote Voltage Sensing at the Rear Panel
External sense terminalsare also available on the back of the power supply thatallow the rear output voltages to be sensed at the load, which compensates for impedance lossesin the load wiring. The front panel binding postsare paralleled with the rear output terminals.
The rear outputterminals accept wires sizes fromAWG 22 to AWG 10.To minimize the possibility of instability onthe output,keep load leads as short
as possible and bundle or twist the leads tightly together to minimizeinductance.
3
Figure 3-3. Rear local sensing connections
Note For rear local voltage sensing connections, the front shorting bars must be removed
first and connect the sense wires as shown in Figure 3-3.
39
Chapter 3 Front-Panel Operation
Disabling the Output
Disabling the Output
The output of thepower supply can be disabled or enabled from the front panel.When the power supply isin the “Off” state, the
the output is disabled. The
OFF
annunciator turnsoff when the power supply returns to the“On” state. When the output is disabled, the voltage value is 0 voltsand the current value is 0.02 amps.
The output state is stored in
volatile
memory; the output isalways disabled
when power has been off or after a remote interfacereset.
While theoutput is disabled, the range selectionkeys, the control knob, resolution selectionkeys,andadjust selection key are still working. Ifthedisplay isinthemetermode, you cannot see the changes of outputvoltage and current settings on thedisplay whenturning the knob.To seeor check the changes when the outputs are disabled, the display should bein the limitmode.
OFF
annunciator turns on and
•Front-panel operation:
You can disable the output by pressing key. This key toggles between
Output
On/Off
output “Off” and “On” states.
•Remote interface operation:
OUTP {OFF|ON}Disable or enable the output
40
Chapter 3 Front-Panel Operation
Disabling the Output Using an External Relay
Disabling the Output Using an External Relay
When the output of the power supply isturnedoff,it is implemented by setting the output to 0 volts and 0.02 amps. This gives a zerooutput voltage without actually disconnecting the output. Todisconnect the output an external relay must be connected between the output and the load. A TTL signal of either lowtrue or high true is provided to control an external relay. This signal can only be controlled with the remote command output is available on the RS-232connection pin 1 and pin 0.
When the (4.5 V) and pin 9 is low (0.5 V). The levels are reversed when the
OUTPut:RELay
Note TTL output of pin 1 or pin 9 of the RS-232 connector is available only after installing
two jumpers inside the power supply. See the Service Guide for more information.
OUTPut:RELay
state is“OFF”.
state is “ON”, theTTL output of pin 1 is high
OUTPut:RELay {OFF|ON}
. The TTL
3
Note Do not use the RS-232 interface if you have configured the power supply to output
relay control signals. Internal components on the RS-232 circuitry may be damaged.
Knob Locking
The control knob can be locked to prevent fromany unwanted changesoccurring during anexperiment,or when you leave the powersupplyunattended. To lock the knob, move the blinking digit to the right or left using
<>
the resolution selection key
Noticethat the knob and front panelkeysare disabled when in the remote interface mode.
or untilthe blinking digit disappears.
41
Chapter 3 Front-Panel Operation
System-Related Operations
System-Related Operations
This section gives information on topics such as self-test, error conditions, and front-panel display control. This information is not directly related to setting up thepower supply but is an important part of operatingthe power supply.
Self-Test
A
power-on
This test assures you that thepower supply is operational. This test does notperformthe extensive set of tests that are includedas part of the complete self-test described below. If the power-on self-test fails,theturns on.
•A
seconds toexecute. If all tests pass, you can have a high confidence that the power supply is operational.
•If the
If the self-test fails, “FAIL” is displayed and the on. SeetheHewlett-Packard for service.
self-test occurs automatically when you turn on the power supply.
ERROR
complete
self-test performs a series of tests and takes approximately2
complete
self-test is successful, “PASS” is displayed on the front panel.
Service Guide
ERROR
for instructions on returning the power supply to
annunciator turns
annunciator
•Front-panel operation:
To performthe complete front panelturn on the power supply andThe self-test willbegin when you release the key following the beep.
•Remote interface operation:
self-test
hold down
, hold down the key as you
the key
untilyouhear a long beep
Recall
*TST?
Returns “0” if the complete self-test passes or “1” if it fails.
42
.
Chapter 3 Front-Panel Operation
System-Related Operations
Error Conditions
When the front-panelsyntax or hardware errorshave been detected.A record of up to 20 errorscan be stored in the power supply’s error queue. See chapter 5 “Error Messages”,starting on
page123for a completelisting of the errors.
•Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the first error that was stored. Errors are cleared as you read themoverthe remote interface. When you have read all errors fromthequeue, the
ERROR
annunciator turnsoff. The power supply beepsonce each time an
error isgenerated.
•If more than 20 errors have occurred when you operate the power supplyover the remote interface, the last error stored in the queue (the most recent error) is replaced with -350, “stored until you remove errors from the queue. If no errorshave occurred when you read the errorqueue, the powersupply responds with +0, “
error
” over the remote interface or“NO ERRORS” fromthe front panel.
•The errorqueue is cleared when powerhas beenoff or aftera status) command has been executed. The clear the error queue.
ERROR
annunciator turns on, one or more command
Too many errors
”. No additional errors are
No
*CLS
(clear
*RST
(reset) command does not
3
•Front-panel operation:
Store
Press (supply is in remote operation mode.
Local
Local
) key to return to front-panel operation mode if the power
3: ERR -102
If the ERRORannunciatoris on, presskey to view theerrors. Use the knob to scroll through the error numbers. Press to view the text of theerror message. Press or key to increase or decrease the scrolling
<>
speed of the text on the display. Allerrors are cleared when you exitthe menu or let the display timeout for about 30 seconds.
•Remote interface operation:
SYSTem:ERRor? Read and clear one error from the error queue
Errors have the following format (the error string may contain up to 80 characters).
-102, ‘‘Syntax error
’’
Error
>
43
Chapter 3 Front-Panel Operation
System-Related Operations
Display Control
For security reasons, you may want to turn off thefront-panel display.Fromthe remote interface, you can display a12-character message on the front panel.
•The display can be
•When the display isturned off, outputs are not sent to the display and allannunciators are disabledexcept the operation is otherwise unaffected by turning off the display.
•The display state is stored in volatile memory; the display is always enabled when power has been off, after aremote interface reset, or after returning tolocal from remote.
•You can display a message on the front panel by sending a command fromthe remote interface. The power supply can display up to12 characters ofthe message on the front panel; any additional characters are truncated.Commas, periods, and semicolons share a display space with the preceding character, and are not considered individual characters. Whena message isdisplayed, outputs are not sent to the display.
•Sending a message to the display fromthe remoteinterface overrides the displaystate; this means thatyou can display a message even if the display is turned off.
•The display state is automatically turned on when you return to thelocal(front panel) operation. Pressfrom the remote interface.
enabled /disabled
Store
Local
from the remote interface only.
ERROR
(
Local
annunciator. Front-panel
) key to return to the local state
•Remote interface operation:
DISP {OFF|ON}Disable / enable the display
DISP:TEXT <quoted string>Display the string enclosed in quotes
DISP:TEXT:CLEClear the displayed message
The following statement shows how to displaya message on the front panelfrom a Hewlett-Packard controller.
‘‘
DISP:TEXT ‘HELLO
’ ’’
44
Chapter 3 Front-Panel Operation
System-Related Operations
Firmware Revision Query
The power supply has three microprocessors for control of various internalsystems.You can querythe power supply to determine which revision of firmwareis installed for each microprocessor.
•You can query the firmware revision from the remote interface only.
•The power supply returns four fields separated by commas and the fourthfield is arevision code which contains threenumbers. The first number is the firmware revision number for the main processor; the second is for the input/output processor;and the third isfor the front-panel processor.
•Remote interface operation:
*IDN? Returns
‘‘HEWLETT-PACKARD,E3633A,0,X.X-X.X-X.X’’ (E3633A)
‘‘HEWLETT-PACKARD,E3634A,0,X.X-X.X-X.X’’ (E3634A)
Be sure to dimension a string variable with at least 40 characters.
SCPI Language Version
The power supply complieswith the rules and regulationsof the present version ofSCPI (Standard Commands for Programmable Instruments). You can determine the SCPI version with which the power supply is in compliance by sendinga command from the remote interface.
You can query the SCPI version from the remote interface only.
3
•Remote interface operation:
SYST:VERS?Query the SCPI version
Returns a string in theform “YYYY.V” where the “Y’s” represent the year of the version, and the “V” representsa version number for that year (for example,
1996.0).
45
Chapter 3 Front-Panel Operation
Remote Interface Configuration
Remote Interface Configuration
Before you can operate the power supply over the remote interface, you must configure the power supply for the remote interface. This sectiongivesinformation on configuringthe remote interface. For additional informationon programming the power supply over the remote interface, Interface Reference", starting on page 73 in chapter 4.
Remote Interface Selection
The power supply isshipped with both an HP-IB (IEEE-488)interface and an RS-232 interface on the rear panel. Only oneinterface can beenabled ata time. The HP-IB interface is selected when the power supply isshippedfromthe factory.
The remote interface can be selected from the front-panel only.
•The interface selection is stored in non-volatile memory, and does not changewhen power has been off or after a remote interfacereset.
•If you select the HP-IB interface, you must select a uniqueaddress for the power supply. The current address is displayed momentarily on the front panel when you turn on the power supply.
•Your HP-IBbus controller hasits own address. Be sure to avoid using the bus controller’s address for any instrument on the interfacebus. Hewlett-Packard controllers generally use address “21”.
•If you enable the RS-232 interface, you must select the baudrate and parity to be used. “RS-232” is displayed momentarily on the front panel when youturn on the power supply if you have selected thisinterface.
1
See "Remote
2
1
Refer to "HP-IB Interface Configuration" starting on page 51 for more information
on connecting the power supply to a computer over the HP-IB interface.
2
Refer to "RS-232 Interface Configuration" starting on page 52 for more information
on connecting the power supply to a computer over the RS-232 interface.
46
Chapter 3 Front-Panel Operation
Remote Interface Configuration
HP-IB Address
Eachdevice on the HP-IB (IEEE-488) interfacemust have a uniqueaddress.You can set the power supply’s addressto any value between0 and 30.The current address is displayed momentarily on thefront panel when you turn onthe power supply.The address is set to “05” when the power supply is shipped from the factory.
The HP-IB address can be set from the front-panel only.
•The addressis stored in non-volatile memory, and doesnot change when power has been off or after a remote interfacereset.
•Your HP-IB bus controller has its own address.Besureto avoid the bus controller’s address for anyinstrument on the interface bus. Hewlett-Packard controllers generally use address “21”.
Baud Rate Selection (RS-232)
You can select one of six baud rates for RS-232 operation. The rate is set to 9600 baud when the power supply is shipped from the factory.
The baud rate can be set from the front-panel only.
•Select one of the following: 300, 600, 1200, 2400, 4800, 9600 baud. The factory setting is
•The baud rate selection is stored in changewhen power has been off or after a remote interfacereset.
9600
baud.
non-volatile
memory, and does not
3
Parity Selection (RS-232)
You canselect the parity for RS-232 operation.The power supply isconfigured for no parity and 8databits when shipped fromthefactory.
The parity can be set from the front-panel only.
•Select one of the following: bits), or Odd (7 data bits).When youset the parity,you are indirectly settingthe number of data bits.
•The parity selection is stored in when power has been off or after a remote interfacereset.
None
(8data bits,
non-volatile
factory setting
memory, anddoes not change
), Even (7 data
47
I/O
Config
I/O
Config
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the HP-IB Address
To configurethe power supply for the HP-IB interface, proceed as follows:
1Turn on the remote configuration mode.
HP-IB / 488
You willsee the above message on thefront-panel display if the power supply has not been changed fromthe factory setting. If “RS-232” appears,choose “HP-IB / 488” by turning the knob to theright.
2Move to the HP-IB address setting mode.
ADDR 05
The address isset to “05” when the power supply isshipped fromthe factory.Notice that a different HP-IB address may appearif the power supply has been changed from the factory setting.
3Turn the knob to change the HP-IB address.
The displayed addressis changed when turning the knob to the right or left.
I/O
Config
4Save the change and turn off theI/O configuration mode.
CHANGE SAVED
The address is stored in non-volatile memory, anddoes not change when power has been off or after a remote interface reset. The power supply displays a message to show that the change is now in effect. If the HP-IB address is not changed, “NO CHANGE” willbe displayed for one second.
Note To exit the I/O configuration mode without any further changes, press the “I/O
Config” key until the “NO CHANGE” message is displayed.
48
I/O
Config
I/O
Config
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the Baud Rate and Parity (RS-232)
To configurethe power supply for the RS-232 interface, proceed as follows:
1Turn on the remote configuration mode.
HP-IB / 488
You willsee the above message on the display if the powersupply has not been changed from the factory setting.
Noticethat if you changed the remote interface selectionto RS-232 before,“RS-232” message will be displayed.
2Choose the RS-232 interface.
RS-232
You can choose the RS-232 interface by turning the knob to the left.
3Move to the RS-232 interface setting mode and select the baud rate.
3
I/O
Config
9600 BAUD
The rate is set to 9600baud when the power supply is shipped from the factory.Choosefrom oneof thefollowing by turning the knob to the right or left: 300, 600, 1200, 2400, 4800, or
4Save the change and choose the parity.
9600
baud.
NONE 8 BITS
The power supply isconfigured for 8 data bits withno parity when shipped from the factory.Choose from one of the following byturning the knob to the right or left: are indirectly setting the number of the data bits.
None 8Bits
, Odd 7 Bits, or Even 7 Bits. When you set parity, you
49
Chapter 3 Front-Panel Operation
Remote Interface Configuration
I/O
Config
5Save the change and turn off theI/O configuration mode.
CHANGE SAVED
The RS-232 baud rate and parity selections are stored innon-volatile memory, and does not change when power has been off or aftera remote interface reset. The power supply displays a message to show that the change is now in effect.If the baudrate and the parity are not changed, “NO CHANGE” will be displayed for one second.
Note To exit the I/O configuration mode without any further changes, press the “I/O
Config” key until the “NO CHANGE” message is displayed.
50
Chapter 3 Front-Panel Operation
HP-IB Interface Configuration
HP-IB Interface Configuration
The HP-IB connector on the rear panel connects your power supply to thecomputerand other HP-IB devices. Chapter 1 lists the cables that are available fromHewlett-Packard.An HP-IB system canbe connected together in any configuration (star, linear, or both) as long as the following rules are observed:
•The total number of devices including the computer is no more than 15.
•Thetotal length ofall the cables used isnomorethan 2 meter times thenumber ofdevices connected together, up to amaximum of 20 meters.
Note IEEE-488 states that you should exercise caution if your individual cable lengths
exceed 4 meters.
Do not stack morethan three connector blocks together on any HP-IBconnector. Make sure that all connectors are fully seated and thatthe lock screws are firmly finger tightened.
3
51
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
RS-232 Interface Configuration
You connectthe power supply to the RS-232 interface using the 9-pin (DB-9)serial connector on the rear panel. Thepower supplyis configured as a DTE(Data Terminal Equipment) device. For all communications over the RS-232interface, thepower supply uses two handshake lines:DTR (Data Terminal Ready, on pin 4) and DSR (Data SetReady, on pin 6).
The following sectionscontain information to help you use the power supply over the RS-232interface. The programming commands for RS-232 areexplained on
RS-232 Configuration Overview
Configure the RS-232 interfaceusing the parameters shown below. Usethe front-panel (see page47 for more information to configure fromthe front panel).
page101.
I/O
key to select the baud rate,parity,and number of data bits
Config
•Baud Rate: 300, 600, 1200, 2400, 4800, or
•Parity and Data Bits:
•Number of Start Bits:
•Number of Stop Bits:
None / 8 data bits (factory setting
Even / 7 data bits, orOdd / 7 data bits
1 bit (fixed2 bits (fixed
9600
baud (
factory setting
)
)
)
)
RS-232 Data Frame Format
A charactercharacter. The frame is defined as the charactersfromthe
stop bit
databits, and parity type. Thepower supply uses the following frame formatsfor seven and eight data bits.
frame
consists of all thetransmitted bits that make up a single
start bit
, inclusively. Within the frame, you can select the baud rate, number of
to thelast
52
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
Connection to a Computer or Terminal
To connect the power supply to a computer or terminal, you must have the proper interface cable. Most computers and terminals are DTE(
Equipment
use aDTE-to-DTE interface cable. These cables are also called null-modem,modem-eliminator, or crossover cables.
The interface cable must also havethe proper connector on each end and theinternal wiring must be correct. Connectors typically have 9 pins (DB-9 connector) or 25 pins (DB-25connector) with a “male” or “female” pin configuration. A male connector has pins insidethe connector shell and a femaleconnector has holes inside the connector shell.
If you cannotfind the correct cable for your configuration, you may have to use aadapter is a “null-modem adapters, and DB-9 to DB-25 adapters.
The cable and adapter diagrams shown below can beused to connect thepowersupply to most computers or terminals. If your configuration isdifferent than those described, order the HP 34399AAdapter Kit. This kitcontainsadapters for connection to other computers, terminals,and modems. Instructionsand pin diagrams are includedwith the adapter kit.
) devices. Sincethe power supply is also a DTE device, you must
wiring adapter
straight-through
. If you are using aDTE-to-DTE cable, make sure the
” type. Typical adapters include gender changers,
Data Terminal
3
DB-9 Serial Connection
port with a male connector, use thenull-modem cable included with the HP 34398A Cable Kit. This cable has a 9-pin femaleconnector on each end. Thecable pindiagram isshown below.
If your computerorterminalhas a 9-pin serial
53
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
DB-25 Serial Connection
port with a maleconnector, use thenull-modemcableand 25-pinadapterincluded withthe HP34398A Cable Kit. Thecableandadapter pin diagramare shown below.
If your computer or terminal has a 25-pin serial
DTR / DSR Handshake Protocol
The power supply is configuredas a DTE(and usesthe DTR (the RS-232 interface to handshake.The power supply usesthe DTR line to send a hold-off signal. The DTR line mustbe TRUE before the power supply willaccept data from the interface.When the powersupply sets the DTR line FALSE, the data must cease within 10 characters.
To disablethe DTR/DSR handshake, do not connect theDTR line and tie the DSR line to logic TRUE. If you disable the DTR/DSR handshake, also select a slower baud rate to ensure that the data istransmittedcorrectly.
Thepowersupply sets the DTR lineFALSE in the following two cases:
1
When the power supply’s input buffer is full (when approximately 100charactershave been received), it sets the DTR line FALSE (pin 4 on the RS-232 connector). When enough characters have been removed to make spaceinthe input buffer,the power supply sets the DTR line TRUE,unless thesecond case (see next) preventsthis.
2
When the power supply wants to “talk” over the interface (which means that it hasprocesseda query) and has received a it will set the DTR line FALSE. This implies that once a query has been sent to the power supply, the bus controller should read the response before attempting to send more data. It also meansthat a the command string.After the response has been output, the power supply sets the DTR line TRUE again, unless the first case (see above)prevents this.
Data Terminal Ready
Data Terminal Equipment
) and DSR (
<new line>
Data Set Ready
message terminator,
<new line>
must terminate
) device
) lines of
54
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
The powersupply monitorsthe DSRline to determine when the bus controller is ready to accept data over the interface. Thepower supply monitors the DSRline (pin 6 onthe RS-232 connector) before each character is sent.The output is suspended if the DSRline is FALSE. When the DSR line goes TRUE, transmission willresume.
The power supplyholds the DTR line FALSE while output is suspended. A formof interface deadlock exists until the buscontroller asserts the DSR line TRUE to allow the power supply to complete the transmission. You can break the interface deadlock by sending the operation in progress and discards pending output (this is equivalent to the IEEE-488 device clear action).
For the <Ctrl-C> character toberecognized reliably bythe powersupplywhileit holds DTRFALSE,the bus controller must first set DSRFALSE.
<Ctrl-C>
character, which clears the
RS-232 Troubleshooting
Here are a few things to check if you are having problems communicating over the RS-232 interface. If you need additional help, refer to the documentation that came with your computer.
•Verify that the power supply and yourcomputer are configuredforthe same baud rate,parity, and numberof data bits.Make sure that your computeris set up for 1 start bit and 2 stop bits (thesevalues are fixed on the power supply).
•Make sure to execute the supply in the remote mode.
•Verify that you have connected the correct interfacecable and adapters.Even if the cable hasthe proper connectors for your system,the internalwiring may be incorrect. The the power supply to most computers or terminals.
•Verify thatyou have connected the interface cable to the correct serial port on your computer (COM1, COM2,etc).
SYSTem:REMote
HP 34398A Cable Kit
command toplace the power
can be used to connect
3
55
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Overview
This section gives an overview of the calibration features of the power supply.For more detailed discussion of the calibration procedures, see the
Guide
.
Calibration Security
This feature allowsyou to enter a security code to prevent accidental or unauthorized calibrations of the power supply. When you first receiveyour power supply, itis secured. Before you can calibrate thepower supply, you must unsecure it by entering thecorrect security code.
•The
•To secure thepower supply from the remote interface, the security code
security code
supply is shipped from the
volatile
remote interface reset.
may contain up to 12 alphanumeric characters as shown below. Thefirst character must bea letter, but theremaining characters can be letters or numbers. You do not haveto use all 12 characters but thefirst character must always be a letter.
memory, and does not change when power has been off or after a
is set to “HP003633”* or “HP003634”** when the power
factory
. The security code is stored in
Service
non-
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
•To secure the powersupply fromthe remote interface so that it can be unsecured from the front panel,use the eight-character format shown below. The first twocharacters must be “H P” and the remaining characters must be numbers. Only the lastsix characters are recognized fromthe front panel, but all eight characters are required.
To
unsecure
the remaining numbers as shown on thefollowing pages.
If you forget your securitycode,you can disable the security featureby adding a jumper insidethepower supply,and thenentering a new code.See theService Guide for more information.
*For HP E3633A Model **For HP E3634A Model
56
the power supply from the front panel, omit the “H P” and enter
H P _ _ _ _ _ _ (8 characters)
Chapter 3 Front-Panel Operation
Calibration Overview
To Unsecure for Calibration
You can unsecure the power supply for
calibration either from the front panel or over theremoteinterface.
The power supplyis secured when shipped from the factory, and the securitycodeis set to “HP003633”* or“HP003634”**
•Front-Panel Operation:
SECURED
If the power supply is secured, you will seethe above message for one second by holding supply. To unsecure the powersupply,press (“CAL MODE” message is displayed in the calibration mode, enter the security code usingthe knob and resolution selection keys, and then press the (
Secure
) key.
Error
Calibrate
(
Calibrate
) key for 5 seconds when you turn on the power
I/O
Config
Secure
Secure
) key after the
000000 CODE
I/O
When you press (message below for one second if the security code iscorrect. The unsecured settingis stored in been off or after a remote interface reset.
the power off and on.
Config
Secure
Secure
non-volatile
) key to save the change, you will see the
memory, and does not change when power has
To exit the calibration mode, turn
I/O
Config
Secure
3
UNSECURED
Noticethat if the security isincorrect, thepowersupply displays an“INVALID” message for a second and returns tothe code entering mode foryoutoenter the correct code.
•Remote Interface Operation:
CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply
To unsecure the power supply, send the above command with the same code usedto secure. Forexample,
‘‘
CAL:SEC:STAT OFF, HP003633’’ (E3633A)or
‘‘
CAL:SEC:STAT OFF, HP003634’’ (E3634A)
*For HP E3633A Model **For HP E3634A Model
57
Chapter 3 Front-Panel Operation
Calibration Overview
To Secure Against Calibration
You can secure the powersupply against calibration either from the front panel or over the remote interface. Thepower supply is secured when shipped from the factory, and the security code is set to “HP003633”* or “HP003634”**.
Be sure to readthe security code rules on page56 before attempting to securethe power supply.
•Front-Panel Operation:
UNSECURED
If the power supply is unsecured, you will seethe above message for onesecond by holding power supply. To secure the power supply, press (“CAL MODE” message is displayed in the calibration mode, enter the security code using the (
Secure
) key.
Noticethat you shouldomit the “HP” and entertheremaining numbers asshown below.
Error
(
Calibrate
control knob
Calibrate
and
)key for 5 seconds when you turn on the
resolution selection keys
I/O
Config
Secure
Secure
, and then press
) key after the
000000 CODE
I/O
When you press (message below. The secured settingis stored in not change when power has been offor after a remote interface reset. To exitthe calibration mode, turn thepower off and on.
Config
Secure
Secure
) key to save the change, you willsee the
non-volatile
memory, and does
I/O
Config
Secure
SECURED
•Remote Interface Operation:
CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply
To secure the power supply, send the above command with the same code asused to unsecure. For example,
‘‘
CAL:SEC:STAT ON, HP003633’’ (E3633A) or
‘‘
CAL:SEC:STAT ON, HP003634’’ (E3634A)
*For HP E3633A Model **For HP E3634A Model
58
Chapter 3 Front-Panel Operation
Calibration Overview
To Change the Security Code
To change the security code, you must first
unsecure the power supply,and then enter a new code.
Be sure to readthe security code rules on page56 before attempting to securethe power supply.
•Front-Panel Operation:
To changethe security code, first make sure that thepower supplyisunsecured. Pressdisplayed in the calibration mode,enter the new security code using the
knob
and
resolution selectionkeys
I/O
Config
Secure
(
Secure
) key after the “CAL MODE” message is
control
, then press
I/O
Config
Secure
(
Secure
) key.
Changing the code from the front panel alsochanges the code required fromthe remote interface.
•Remote Interface Operation:
CAL:SEC:CODE <new code>Change the security code
To change the security code,first unsecure the powersupply using the old security code. Then, enter the new code. For example,
‘‘
CAL:SEC:STAT OFF, HP003633* or HP003634**
‘‘
CAL:SEC:CODE ZZ001443
‘‘
CAL:SEC:STAT ON, ZZ001443
’’
’’
’’
Unsecure with old code
Enter new code
Secure with new code
3
*For HP E3633A Model **For HP E3634A Model
59
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Count
You can determine the number of times that your power supply hasbeen calibrated. Your power supply was calibrated before it left thefactory.When you receive your power supply,read the countto determineits initial value.
The calibration countfeature canbeperformed from the remoteinterface only.
•The calibration countis stored in
when power has been off or after a remote interfacereset.
•The calibration count increments upto a maximum of 32,767 after which it
wraps-around to 0. Since the value increments by one for each calibration point, a complete calibration will increase the value by 5 counts.
•Remote Interface Operation:
CAL:COUN?Query the number of times of calibration
non-volatile
memory, and does not change
Calibration Message
You can use the calibration message feature torecord calibration information about your power supply. For example, you can store such information as the last calibration date, thenext calibration due date, the power supply’s serial number,or even the name and phonenumber of the person to contact for anew calibration.
Youcan record and read informationin the calibration messagefrom the remote interface only.
•The power supply should be unsecured before sending a calibration
message.
•The calibration message may contain up to40 characters.
•The calibration message is stored in non-volatile memory, and doesnot
changewhen power has been off or after a remote interfacereset.
•Remote Interface Operation:
CAL:STR <quoted string>Store the cal message
The following command stringshows how to store a calibration message.‘‘
CAL:STR ‘CAL 12-05-98
60
’ ’’
4
Remote Interface Reference
Remote Interface Reference
•SCPI Command Summary, on page 63
SCPI
•Simplified Programming Overview, on page 68
•Using the
•Output Setting and Operation Commands, on page 72
•Triggering Commands, on page 79
•System-Related Commands, on page 82
•Calibration Commands, on page 86
•RS-232 InterfaceCommands, on page 89
•The SCPI Status Registers,on page 90
•Status Reporting Commands, on page 98
SCPI
•An Introduction to the SCPI Language, on page 101
•Halting an Output in Progress, on page 106
•SCPI Conformance Information, on page 107
•IEEE-488 Conformance Information, on page 110
APPLy
Command, on page 71
SCPI
If youarea first-time userof the SCPI language, you may want to refer to these sections to become familiar withthelanguage beforeattempting toprogram the power supply.
62
Chapter 4 Remote Interface Reference
SCPI Command Summary
SCPI Command Summary
SCPI
This section summarizes the SCPI (
Instruments
remote interface. Refer to the latersectionsin thischapter for more complete details on each command.
Throughout this manual, the following conventions areused for SCPI command syntax.
•Square brackets ([ ]) indicate optional keywords or parameters.
•Braces ({ }) enclose parameters within a command string.
•Triangle brackets (< >)indicate that you must substitute a value or a code for the enclosed parameter.
•A vertical bar ( | ) separates oneof two or more alternativeparameters.
First-time SCPI users,see page101.
) commandsavailable to programthe power supply overthe
This section gives an overview of the basic techniques used to programthepowersupply over the remote interface. Thissection is only an overview and does not give allof the details you will need to write your own application programs. Refer to the remainder of this chapter and also chapter 6, “ApplicationPrograms”, for more details and examples. Also refer to theprogramming reference manual that camewith your computer for details on outputting command strings and entering data.
Using the APPLy Command
The
APPLy
command provides the most straightforward method to program the power supply over the remote interface. For example, the following statement executed from your computer will set the power supply to an output of 3 V rated at 1 A:
‘‘APPL 3.0, 1.0
’’
Using the Low-Level Commands
Although the program the power supply, the low-level commands give you more flexibility to change individual parameters. For example, the following statementsexecuted from your computer willset the power supply to an output of 3 V rated at 1 A:
‘‘VOLT 3.0
‘‘CURR 1.0
APPLy
’’’’
command provides the most straightforward method to
Set output voltage to 3.0 V
Set output current to 1.0 A
68
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Reading a Query Response
Onlythe query commands (commands that end with “ ? ” ) willinstruct the powersupply tosend a response message. Queries return either output values or internalinstrument settings.For example, the following statements executed from your computer willread the power supply’s error queue and print the most recent error:
bus enter statementEnter error string into computer
print statementPrint error string
’’
Read error queue
Selecting a Trigger Source
The power supply willaccept a ‘‘bus’’(software) trigger or an immediate internal trigger as a trigger source. By default, the ‘‘BUS’’ trigger sourceisselected. If you want the power supply to usean immediate internal trigger,you must select ‘‘executed fromyour computer will set to an output of 3 V/1 Aimmediately:
‘‘VOLT:TRIG 3.0
‘‘CURR:TRIG 1.0
‘‘TRIG:SOUR IMM
‘‘INIT
IMMediate
’’
’’.For example, the followingstatements
’’’’’’
Set the triggered voltage level to 3.0 V
Set the triggered current level to 1.0 A
Select the immediate trigger as a source
Cause the trigger system to initiate
4
69
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Power Supply Programming Ranges
The
SOURce
available programmingvalue for a parameter varies according to the desired output rangeof the power supply. The followingtable lists the programmingvalues available and Agilent E3633Aand E3634A powersupplies.
Refer to this table to identify programming values when programming the power
supply.
Table 4-1. Agilent E3633A Programming Ranges
VoltageProgramming Range 0 V to 8.24V 0 V to 20.60 V
CurrentProgramming Range 0 A to 20.60 A 0 A to 10.30 A
subsystem requires parameters for programmingvalues. The
MINimum, MAXimum, DEFault
0 - 8V/20A Range 0 - 20V/10A Range
MAX Value 8.24 V 20.60 V
MIN Value 0 V 0 V
DEFault Value 0 V 0 V
*RST Value 0 V
MAX Value 20.60 A 10.30 A
MIN Value 0 A 0 A
DEFault Value 20 A 10 A
*RST Value 20.00 A
and reset values of the
Table 4-2. Agilent E3634A Programming Ranges
0 - 25V/7A Range 0 - 50V/4A Range
VoltageProgramming Range 0 V to 25.75V 0 V to 51.5 V
MAX Value 25.75 V 51.5 V
MIN Value 0 V 0 V
DEFault Value 0 V 0 V
*RST Value 0 V
CurrentProgramming Range 0 A to 7.21 A 0 A to 4.12 A
MAX Value 7.21 A 4.12 A
MIN Value 0 A 0 A
DEFault Value 7.0 A 4.0 A
*RST Value 7.00 A
70
Chapter 4 Remote Interface Reference
Using the APPLy Command
Using the APPLy Command
The
APPLy
command provides the most straightforward method to program the power supply overthe remote interface. You can select the output voltage and current in one command.
APPLy {<voltage>| DEF | MIN | MAX}[,{<current>| DEF | MIN | MAX}]
This command iscombinationof as the newlyprogrammed values are within the presently selected range, the output voltageand current are changed as soon as the command is executed.
The
APPLy
command changes the power supply’s output to the newly programmed values only if the programmed values are valid within the presently selected range. An execution error willoccurif the programmed values are not valid within the selected range.
You can substitute ‘‘specific value for thevalues of ‘‘0’’ volts and ‘‘0’’ amps. the selected range.
The
default
regardless of the presently selected range. For more details of parameters, see Table 4-1 for the Agilent E3633A model and Table 4-2 for the Agilent E3634Amodel.
If you specify only one parameter of the regards it asvoltage setting value.
values of voltage and current are ‘‘0’’ volts and ‘‘20’’*or‘‘7’’** amps
MINimum
voltage
’’, ‘‘
and
VOLTage
MAXimum
current
MAX
and
CURRent
’’, or ‘‘
DEFault
parameters.
selects the highestvalues allowed for
APPLy
command, the power supply
commands. As long
’’ in place of a
MIN
selects thelowest
4
APPLy?
This commandqueries the power supply’spresent voltage and current setting values and returns a quoted string. The voltage and current arereturned in sequence as shown in the sample string below (the quotation marks are returned as part of the string).
‘‘8.00000,20.00000’’* or ‘‘25.00000,7.00000’’**
In the above string, the first number 8.00000 isthe voltage setting value and the second number 20.00000 is the current setting value.
*For Agilent E3633A Model **For Agilent E3634A Model
71
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output Setting and Operation Commands
This section describes low-level commands used to program the power supply.
’’ or ‘‘
APPLy
DOWN
Although the programthe power supply, the low-level output setting commandsgive youmore flexibility to change theindividual parameters.
CURRent{<current>|MINimum | MAXimum|UP|DOWN}
This command programs the immediate current level of the power supply. The immediate level isthe currentvalue ofthe outputterminals.
The
CURRent
programmed value regardlessof the output range presently selected.You can substitute ‘‘
the current parameter.selects the highest current values allowed for the selected range.
This command also increases or decreasesthe immediate current level using the ‘‘
UP
CURRent:STEP
increment setting will cause anexecution error -222 (Data out of range) when the maximum or the minimum rated current is exceeded.
command provides the most straightforward method to
command changes the output of the power supply to the newly
MINimum
MIN
’’ parameter by a predetermined amount. The command
sets theamount of increase or decrease. Noticethat a new
’’ or ‘‘
MAXimum
selects the lowest current values of ‘‘0’’ amps.
’’ in place of aspecific value for
MAX
CURRent
Example
The following program segments show how to use the
CURRDOWN
CURR:STEP
‘‘
CURR:STEP0.01
‘‘
CURRUP
‘‘
CURR:STEP0.02
‘‘
CURRDOWN
72
command to increase or decrease the output current with the command.
’’
’’
’’
’’
Set the step size to 0.01 A
Increase the output current
Set the step size to 0.02 A
Decrease the output current
CURRUP
or
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
CURRent? [MINimum | MAXimum]
This query returns the presently programmed current level of the power supply. programmable current levels for the selected range.
CURRent:STEP {<numeric value>|DEFault}
This command sets the stepsize for current programming with the
UP
To set the step size to the minimum resolution, set the step size to ‘‘The minimum resolution of the step size is approximately 0.32 mA(E3633A) and 0.13 mA (E3634A) respectively. The minimum resolution of yourinstrument. The immediate current level increasesor decreases by thevalue of the step size. For example, the output current willincrease or decrease 10 mAif the step size is0.01.
This command isuseful when you program the powersupply to the allowed minimum resolution. At resolution.
CURR?MAX
and
CURRentDOWN
and
CURR? MIN
commands. See the example in the previous page.
*RST
return the highest and lowest
CURR:STEP? DEF
, the step size is the value of the minimum
returnsthe
CURRent
DEFault
’’.
CURRent:STEP? {DEFault}
This query returns the value of the step size currently specified. The returned parameter isa numeric value. ‘‘step size in unit of amps.
CURRent:TRIGgered {<current>| MINimum | MAXimum}
This command programs the pending triggeredcurrent level.The pending triggered current level is astored value that is transferred to theoutput terminals when a trigger occurs. A pending triggered level is not affected bysubsequent
CURRent:TRIGgered? [MINimum | MAXimum]
This query returns the triggered current level presently programmed. If no triggered level is programmed, the
:TRIG? MAX
programmable
CURRent
and
commands.
CURR:TRIG? MIN
triggered current levels.
DEFault
CURRent
return the highestand lowest
’’ gives the minimum resolution of the
level isreturned.
CURR
4
73
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
CURRent:PROTection {<current>|MINimum|MAXimum}
This command sets the current level at which the overcurrent protection (OCP) circuit will trip. If the peak output current exceeds theOCP level,thenthe outputcurrent is programmed to zero. TheQuestionableStatus register ‘‘OC’’ bit is set (see
CURR:PROT:CLE
removed.
page91). An overcurrent condition can be cleared with the
command after the condition that caused the OCP trip is
CURRent:PROTection? {MINimum|MAXimum}
This query returns the overcurrent protection triplevel presently programmed.
CURR:PROT? MAX
programmable overcurrent trip levels.
and
CURR:PROT? MIN
return the maximum and minimum
CURRent:PROTection:STATe {0|1|OFF|ON}
This command enablesordisables the overcurrent protection functionof thepower supply. An overcurrent condition can be cleared with the
CURR:PROT:CLE
removed. At
command after the condition that caused the OCP trip is
*RST
, thisvalue isset to‘‘ON’’.
CURRent:PROTection:STATe?
This query returns the state of the overcurrent protection function. Thereturned parameter is ‘‘0’’(OFF) or ‘‘1’’ (ON).
CURRent:PROTection:TRIPped?
This query returns a ‘‘1’’ if the overcurrent protection circuit is tripped and not cleared or a ‘‘0’’ if not tripped.
CURRent:PROTection:CLEar
This command causes the overcurrent protection circuit to be cleared.After thiscommand, the output current is restored to the state it was inbefore the current protection trippedand the OCP trip level remains unchanged to the value presently programmed. Before sending thiscommand, lower the output current below the trip OCP point, or raise the OCP trip level above the output setting.
must beremovedfirst beforeproceeding this command.
Note that the overcurrent condition caused by an externalsource
74
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage {<voltage>| MINimum | MAXimum|UP|DOWN}
This command programs the immediate voltage level of the power supply.Theimmediate level is the voltage valueof the output terminals.
The
VOLTage
programmed value regardlessof the output range presently selected.You can substitute ‘‘
the voltage parameter.selects the highest voltage values allowed for theselected range.
This command also increases or decreases the immediate voltage level using the ‘‘
UP
VOLTage:STEP
increment setting will cause anexecution error -222 (Data out of range) when the maximum or the minimum rated voltageis exceeded.
command changes the output of the power supply to the newly
’’ or ‘‘
MINimum
MIN
DOWN
’’ parameter by a predetermined amount. The command
sets the amount of increase or decrease. Notice that anew
’’ or ‘‘
MAXimum
selects the lowest voltage values of ‘‘0’’ volts.
’’ in place of aspecific value for
MAX
VOLTage
Example
The following program segments show how to use the
VOLT DOWN
VOLT:STEP
‘‘
VOLT:STEP 0.01
‘‘
VOLT UP
‘‘
VOLT:STEP 0.02
‘‘
VOLT DOWN
VOLTage? [MINimum | MAXimum]
This query returns the presently programmed voltage level ofthe power supply.
programmable
command to increase or decrease the output voltage with the command.
’’
’’
VOLT? MAX
voltage levels for the selected range.
’’
’’
and
VOLT? MIN
return the highest and lowest
Set the step size to 0.01 V
Increase the output voltage
Set the step size to 0.02 V
Decrease the output voltage
VOLT UP
or
4
75
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage:STEP {<numeric value>|DEFault}
This command sets the step size for voltageprogramming with the and
VOLT DOWN
To set the step size to the minimum resolution, set the stepsize to ‘‘The minimum resolution of the step size is approximately 0.36 mV(E3633A) and 0.95 mV (E3634A) respectively. The minimum resolution of your instrument. Theimmediate voltage level increasesor decreases by the valueof the step size. For example, the output voltagewillincrease or decrease10mV if the step sizeis 0.01.
This command isuseful when you program the powersupply to the allowed minimum resolution. At resolution.
commands. See the above example in the previous page.
VOLT:STEP? DEF
*RST
, the step size is the value of the minimum
returns the
VOLT UP
DEFault
’’.
VOLTage:STEP? {DEFault}
This query returns the value of the step size currently specified. The returned parameter is a numeric value.‘‘size in unit of volts.
DEFault
’’ givesthe minimum resolution step
VOLTage:TRIGgered {<voltage>| MINimum | MAXimum}
This command programs the pending triggeredvoltage level. The pendingtriggered voltage level isa stored value that is transferred to the output terminals when a trigger occurs. A pending triggered level is not affected bysubsequent
VOLTage
commands.
VOLTage:TRIGgered? [MINimum | MAXimum]
This query returnsthe triggered voltage level presently programmed. If no triggered levelis programmed, the and
VOLT:TRIG? MIN
voltage levels.
return the highest and lowest programmable triggered
VOLT
level is returned.
VOLT:TRIG? MAX
VOLTage:PROTection {<voltage>|MINimum|MAXimum}
This command sets the voltage level at which the overvoltage protection (OVP)circuit willtrip. If thepeak output voltage exceeds the OVP level,then the power supply output is shorted by an internal SCR. The Questionable Status register ‘‘OV’’bit is set (see with the trip is removed.
VOLT:PROT:CLE
page91). An overvoltage condition can be cleared
command after the condition that caused the OVP
76
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage:PROTection? {MINimum|MAXimum}
This query returns the overvoltageprotection trip level presently programmed.
VOLT:PROT? MAX
programmable overvoltage trip levels.
VOLTage:PROTection:STATe {0|1|OFF|ON}
This command enablesor disablesthe overvoltage protectionfunction. Anovervoltage condition can be cleared with the after the condition that causedthe OVP trip is removed. Atset to ‘‘ON’’.
VOLTage:PROTection:STATe?
This query returns the state of the overvoltage protection function. The returned parameter is ‘‘0’’(OFF) or ‘‘1’’ (ON).
VOLTage:PROTection:TRIPped?
This query returns a ‘‘1’’ if the overvoltage protection circuit is tripped and not cleared or a ‘‘0’’ if not tripped.
VOLTage:PROTection:CLEar
This command causes the overvoltage protection circuit to be cleared. After this command, theoutput voltage is restored to the state it was in before the protection feature occurred and the OVP trip level remains unchanged to the value presently programmed. Before sending thiscommand, lower the output voltage below the trip OVP point, or raise the OVPtrip level above the output setting.
be removed first before proceeding this command.
Note that theovervoltage condition caused by anexternal sourcemust
This command selects an output range to be programmed by the identifier. For example, when 8V/20A* range is selected, the voltage and current are limited to 8.24 voltsand 20.60 amps. When 20V/10A* range is selected, the limited to 20.60 volts and 10.30 amps. See programming rangesof the Agilent E3634Amodel.‘‘P20V’’* or ‘‘HIGH’’is theidentifier for the 20V/10A* range and‘‘P8V’’* or‘‘LOW’’ is for the 8V/20A* range.
*RST
At
*For Agilent E3633A Model **For Agilent E3634A Model
, the 8V/20A*or 25V/7A** range isselected.
maximum programmable
maximum programmable
voltage and current are
page70 for more details of the
77
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
VOLTage:RANGe?
This query returns the currently selected range. The returned parameter for the Agilent E3633A is ‘‘P20V’’ (HIGH) or ‘‘P8V’’ (LOW) and the parameter for the Agilent E3634A is ‘‘P50V’’ (HIGH) or ‘‘P25V’’ (LOW).
MEASure:CURRent?
This command queries the current measured across the current sense resistor inside the power supply.
MEASure[:VOLTage]?
This command queriesthe voltage measured at the sense terminalsof the power supply.
78
Chapter 4 Remote Interface Reference
Triggering Commands
Triggering Commands
The power supply’striggering systemallows a change involtage and current when receiving a trigger,to select a trigger source, and to insert a trigger. Triggering the powersupply isa multi-step process.
•First, you mustspecify the source from which thepower supply will accept
the trigger. The power supply will accept a bus (software) trigger or animmediate trigger from the remote interface.
•Then, you can setthetime delay between the detection of the trigger on the
specified trigger source and the start of any corresponding output change.
Notice that the time delay is valid for only the bus trigger source.
•Finally, youmust providean
source isselected, the selected output is set to the triggered level immediately.But if the trigger source is the bus, the power supply is set to the triggeredlevelafter receiving the Group Execute Trigger (GET) orcommand.
Trigger Source Choices
Youmustspecify the source from which the power supply willaccept a trigger.The trigger is stored in volatile memory; the source is set to bus when the power supply has been off or after a remote interface reset.
INITiate
command. If the
IMMediate
*TRG
4
Bus (Software) Triggering
•To select the bustrigger source,send the following command.
TRIG:SOUR BUS
•To trigger thepower supply from the remote interface (GPIB or RS-232)
after selecting the bus source,send the
*TRG
is sent, the trigger action starts after the specified time delay if any
delay is given.
•Youcan also trigger the power supply from the GPIB interface by sending
the IEEE-488 Group Execute Trigger(GET)message. The following statement shows how to send a GET from a Agilent Technologies controller.
TRIGGER 705
(group execute trigger)
*TRG
(trigger) command. When the
79
Chapter 4 Remote Interface Reference
Triggering Commands
•To ensure synchronization when the bus source is selected, send the
*WAI
(wait) command. When the waits for all pending operations to complete before executing any additionalcommands. For example, the following command string guarantees that the first trigger is accepted and is executed before thesecond trigger isrecognized.
command is executed, the power supply
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
•You can usethe (operation complete) command to signal when the operation is complete.
*OPC?
The is complete. TheEvent register when the operation is complete.
*OPC?
command returns ‘‘1’’ to the outputbuffer when the operation
(operation completequery) command or the
*OPC
command sets the ‘‘OPC’’bit (bit 0) in the Standard
Immediate Triggering
•To select theimmediate trigger source, send the followingcommand.
TRIG:SOUR IMM
•When the command immediately transfers the
VOLT
IMMediate
or
CURR
is selected as a trigger source, an
VOLT:TRIG
value. Any delay is ignored.
or
CURR:TRIG
INITiate
value to
*WAI
*OPC
80
Chapter 4 Remote Interface Reference
Triggering Commands
Triggering Commands
INITiate
This command causes thetrigger system to initiate. This command completes one full trigger cycle when the trigger source is an immediate and initiates the trigger subsystem when the trigger source isbus.
TRIGger:DELay {<seconds>| MINimum | MAXimum}
This command sets the time delay between the detection of an event on the specified trigger source and the start of any corresponding trigger action on the power supply output. Select from 0 to 3600 seconds.
MAX
= 3600 seconds. At
TRIGger:DELay?
This command queries the trigger delay.
TRIGger:SOURce {BUS | IMMediate}
This command selects the source fromwhich the powersupply willaccept a trigger. Thepower supply will accept a bus (software) trigger or an internalimmediate trigger. At
*RST
, this value is set to 0seconds.
*RST
, the bus trigger source is selected.
MIN
= 0 seconds.
4
TRIGger:SOURce?
This command queries the present trigger source. Returns ‘‘BUS’’ or ‘‘IMM’’.
*TRG
This command generates a trigger to the trigger subsystem that has selected a bus (software) trigger as its source the same effect as theGroup Execute Trigger (GET) command. For RS-232 operation, make sure the power supply isin the remote interface mode by sending the
SYST:REM
command first.
(TRIG:SOUR BUS)
.The command has
81
Chapter 4 Remote Interface Reference
System-Related Commands
System-Related Commands
DISPlay {OFF | ON}
This command turns the front-panel display off or on. When the display is turned off, outputs are not sent to the display and all annunciators are disabledexcept the
The display state is automatically turned on when you return to the local mode. Press
DISPlay?
This command queries the front-panel display setting. Returns ‘‘0’’ (OFF) or ‘‘1’’ (ON).
DISPlay:TEXT <quoted string>
This command displaysa message on the front panel. The power supply willdisplay up to12 characters in a message; any additionalcharacters aretruncated. Commas, periods, and semicolons share a display spacewith the preceding character, and arenot considered individual characters.
Store
Local
ERROR
(
Local
annunciator.
) key to return to the local statefrom the remote interface.
DISPlay:TEXT?
This command queries the messagesent to the front panel and returns a quoted string.
DISPlay:TEXT:CLEar
This command clears the message displayed on the front panel.
OUTPut {OFF | ON}
This command enables ordisables the outputs of the power supply. When the output is disabled, thevoltage valueis 0 V and the currentvalue is 1 mA. At
*RST
, the output state isOFF.
OUTPut?
This command queries the output state of the power supply. The returned value is ‘‘0’’ (OFF) or ‘‘1’’ (ON).
82
Chapter 4 Remote Interface Reference
System-Related Commands
OUTPut:RELay {OFF | ON}
This command sets the state of two TTL signals on the RS-232 connector. These signals are intended for use with an external relay and relay driver. The TTL output is available on the RS-232connector pin 1and pin 9. When the
OUTPut:RELay
9 is low (0.5 V). The levels arereversed when the‘‘OFF’’. At
Note TTL output of pin 1 or pin 9 of the RS-232 connector is available only after installing
two jumpers inside the power supply. See the Service Guide for more information.
Note Do not use the RS-232 interface if you have configured the power supply to output
relay control signals. Internal components on the RS-232 circuitry may be damaged.
OUTPut:RELay?
This command returnsthe state of the TTL relay logic signals. See also
OUTP:REL
SYSTem:BEEPer
This command issues a single beep immediately.
state is ‘‘ON’’, the TTL outputof pin 1 is high (4.5V) and pin
OUTPut:RELay
*RST
, the
OUTPut:RELay
command.
state is OFF.
state is
4
SYSTem:ERRor?
This command queries the power supply’s error queue. When the front-panel
ERROR
errors have been detected. Up to 20 errors can be stored in the error queue.
See ‘‘Error Messages’’ for a complete listingof the errors in chapter 5.
•Errors are retrieved in first-in-first-out (FIFO) order. The first error returned
•If more than 20errors have occurred, the last error stored in the queue (the
•The errorqueue is cleared when powerhas beenoff or aftera
annunciator turns on, one or more command syntax or hardware
is the first error that was stored. When you haveread all errors from the queue, the time an error isgenerated.
most recenterror) is replaced with -350, ‘‘Too many errors’’. No additional errorsare stored untilyou remove errorsfrom the queue. If no errors haveoccurred when you read the error queue, the power supply responds with
+0, ‘‘Noerror’’.
status) command has been executed. The clear the error queue.
ERROR
annunciator turns off. The power supplybeeps once each
*CLS
(clear
*RST
(reset) command
does not
83
Chapter 4 Remote Interface Reference
System-Related Commands
SYSTem:VERSion?
This command queries the power supply to determine the present SCPI version. The returned value is of a stringin the form YYYY.V where the ‘‘Y’s’’represent the year of the version,andthe ‘‘V’’ represents a version number for that year(for example, 1996.0).
*IDN?
This query command reads the power supply’s identification string. The power supply returns four fields separated by commas.The first field isthe manufacturer’s name, the second field is the model number, the third field is not used (always ‘‘0’’), and the
fourth field
is a revision codewhich contains three numbers. The first number is the firmware revision number for the mainpower supplyprocessor; the second is for the
input/output
processor; and the
third is for the front-panel processor.The command returns a string with the following format (be sure to dimension
a string variablewith at least
40 characters
):
HEWLETT-PACKARD,E3633A or E3634A,0,X.X-X.X-X.X
*RST
This command resets the power supply to its power-on state as follows:
This query performs a the self-testpasses or ‘‘1’’ or any non-zero value if it fails. If the self-test fails,an error message is also generated with additional information on whythe test failed.
*SAV { 1 | 2 | 3 }
This command stores the present state of the power supply to the specified location in
3) are available to store operating states of the power supply. The state storage feature‘‘remembers’’ the states or values of the following commands:
To recall a stored state, you must use the same memory location usedpreviously to store the state.
non-volatile
complete
memory. Three memorylocations (numbered 1, 2 and
self-test ofthe power supply. Returns ‘‘0’’ if
and
*RCL { 1 | 2 | 3 }
This command recalls a previously storedstate. To recalla stored state,you must use the samememory location used previously to store the state.
Note DISP {OFF|ON} can be stored and recalled in remote interface mode only. Going
to local mode automatically sets the display state to ON.
4
85
Chapter 4 Remote Interface Reference
Calibration Commands
Calibration Commands
Seechapter 3 ‘‘CalibrationOverview’’, starting onpage56 for an overview of thecalibration features of the power supply. Formoredetaileddiscussion of the calibration procedures, see the Service Guide.
Note When you calibrate the power supply, you should not set the OVP and OCP to ON
state in order to prevent OVP or OCP from tripping.
CALibration:COUNt?
This command queries the power supply to determine the number oftimes ithas been calibrated. Your power supply was calibrated before it left the factory. When you receive your power supply, read the count to determine its initialvalue. Since the value increments by onefor each calibration point, a complete calibration will increase thevalue by5 counts.
CALibration:CURRent[:DATA] <numeric value>
This command can only be used after calibration is unsecured and the output state is ON.It enters a current value that you obtained by reading an external meter.You must first select the minimum calibration level(
) for the valuebeingentered. You must then select themiddle and
MIN
maximumcalibration levels () for the valuebeingentered. Three successive values mustbe selected and entered. The power supply then computes new calibration constants.Theseconstants are then stored in
This command can only be used after calibration is unsecured and the output state is ON. It sets thepower supply to a calibration point that is entered with
CAL:CURR
low-end point (MIN) must beselected and entered first.
CALibration:CURRent:PROTection
This command calibratesthe overcurrent protection circuit of the power supply. It takes about 10 seconds to execute the command. The calibration must be unsecured and the output shorted before calibrating the overcurrent protection. The power supply automatically performs the calibration andstores the new overcurrentconstant in
current calibration precedes beforesending thiscommand.
86
command. Duringcalibration, threepoints must be entered and the
nonvolatile
memory.
Notice that
Chapter 4 Remote Interface Reference
Calibration Commands
CALibration:DAC:ERRor
This command corrects the differential nonlinearity error of the internalDACwithout an external meter. You must send this command before calibrating the voltage. It takesabout 30 secondsto execute the command.
CALibration:SECure:CODE <new code>
This command enters a new security code. To change the security code, first unsecurethe power supply using theold security code. Then, enter the new code. The calibration code may contain up to 12 charactersover the remoteinterface but the first character must always be a letter.
CALibration:SECure:STATe {OFF | ON},<code>
This command unsecures or secures the power supply for calibration. The calibration code may contain up to 12 characters over the remote interface.
CALibration:SECure:STATe?
This command queries the secured state for calibration of the power supply.The returned parameter is ‘‘0’’ (OFF) or ‘‘1’’ (ON).
CALibration:STRing <quoted string>
This command records calibration information about your power supply. For example, you can store such information as the last calibration date, the nextcalibration due date, or the power supply’s serial number. The calibration message may contain up to40 characters.The power supply should be unsecured before sending a calibration message.
4
CALibration:STRing?
This command queries the calibration messageand returns a quoted string.
CALibration:VOLTage[:DATA] <numeric value>
This command can only beused after calibration is unsecured and the output state isON. It entersa voltage value that you obtained by reading an externalmeter.You must first select the minimum calibration level(
) for the valuebeingentered.You must then selectthe middle and
MIN
maximum calibration levels (for the value being entered. Threesuccessive values must be selected andentered. The power supply thencomputes new voltage calibrationconstants.These constants are then stored in
This command can only be used after calibration is unsecured and the output state is ON. It sets thepower supply to a calibration point that is entered with
CAL:VOLT
low-end point (MIN) must beselected and entered first.
command. Duringcalibration, threepoints must be entered and the
CALibration:VOLTage:PROTection
This command calibratesthe overvoltage protection circuit of the power supply. It takes about 10 seconds to execute the command. The calibration must be unsecured and the output be opened before calibrating the overvoltage protectioncircuit. The power supply automatically performs the calibrationand stores the new overvoltage constantin
voltage calibration precedes beforesending this command.
nonvolatile
memory.
Notice that
88
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.