Agilent E3632A Specifications Sheet

Service Guide
3DUW1XPEHU( $SULO
)RU:DUUDQW\LQIRUPDWLRQUHIHUWRWKHEDFNRIWKHPDQXDO
&RS\ULJKW$JLOHQW7HFKQRORJLHV,QF
$OO5LJKWV5HVHUYHG
$JLOHQW($ '&3RZHU6XSSO\
Convenient bench-top features
• Dual range
• Easy-to-use knob control settings
• Highly visible vacuum-fluorescent display meters
• Hghl accuracy and high resolution
• Remote voltage sensing
• Overvoltage and overcurrent protection
• Output on/off
• Excellent load and line regulation and low ripple and noise
• Operating states storage
• Portable, ruggedized case with non-skid feet
Flexible system features
• GPIB (IEEE-488) and RS-232 interfaces are standard
• SCPI (Standard Commands for Programmable Instruments) compatibility
• I/O setup easily done from front-panel
• Software calibration, no internal adjustments required
Agilent E3632A DC Power Supply

The Front Panel at a Glance

1 15V/7A range selection key 2 30V/4A range selection key 3 Overvoltage protection key 4 Overcurrent protection key 5 Display limit key 6 Recall operating state key 7 Store operating state/Local key
2
8 Error/Calibrate key 9 I/O Configuration/Secure key
10 Output On/Off key 11 Control knob 12 Resolution selection keys 13 Voltage/current adjust selection key
1 15V/7A range selection key Selects the 15V/7A range and allows the full rated output
to 15V7A.
2 30V/4A range selection key Selects the 30V/4A range and allows the full rated output
to 30V/4A.
3 Overvoltage protection key Enables or disables the overvoltage protection function,
sets trip voltage level, and clears the overvoltage condition.
4 Overcurrent protection key Enables or disables the overcurrent protection function,
sets trip current level, and clears the overcurrent condition.
5 Display limit key Shows voltage and current limit values on the display and allows
knob adjustment for setting limit values.
6 Recall operating state key Reca l ls a p revio usly stored operati ng st ate fr om lo cation
"1", "2", or "3".
1
7 Store operating state / Local key
Stores an operating state in location "1", "2", or
"3" / or returns the power supply to local mode from remote interface mode.
2
8 Error / Calibrate key
Displays error codes generated during operations, self-test and calibration / or enables calibration mode (the power supply must be unsecured before performing calibration).
3
9 I/O Configuration / Secure key
Configures the power supply for remote interfaces
/ or secure or unsecure the power supply for calibration.
10 Output On/Off key Enables or disables the power supply outputs. This key toggles
between on and off.
11 Control knob Increases or decreases the value of the blinking digit by turning
clockwise or counter clockwise.
12 Resolution selection keys Move the blinking digit to the right or left.
13 Voltage/current adjust selection key Selects the knob control function for voltage
or current adjustment.
1
The key can be used as the "Local" key when the power supply is in the remote
interface mode.
2
You can enable the "calibration mode" by holding down this key when you turn on
the power supply.
3
You can use it as the "Secure" or "Unsecure" key when the power supply is in the
calibration mode.
3

Front-Panel Voltage and Current Limit Settings

You can set the voltage and current limit values from the front panel using the following method.
Use the voltage/current adjust selection key, the resolution selection keys, and the control knob to change the voltage and current limit value.
1 Select the desired range using the range selection keys after turning on the power
supply.
2 Press the key to show the limit values on the display.
3 Move the
and change the blinking digit value to the desired vltage limit by turning the control knob. If the display limit times out, press the key again.
4 Set the knob to current control mode using the voltage/current adjust selection key.
5 Move the blinking digit to the appropriate position using the resolution selection keys
and change the blinking digit value to the desired current limit by turning the control knob.
6 Press the key to enable the output. After about 5 seconds, the
display will go to the output monitor mode automatically to display the voltage and current at the output or the display will go to the monitor mode immediately by pressing the
Display Limit
blinking digit to the appropriate position using the resolution selection keys
Display Limit
Output On/Off
Output On/Off
key again.
Note
All front panel keys and controls can be disabled with remote interface commands. The Agilent E3632A must be in "Local" mode for the front panel keys and controls to function.
4

Display Annunciators

Adrs Power supply is addressed to listen or talk over a remote interface. Rmt Power supply is in remote interface mode. 15V Shows the 15V/7A range is selected. 30V Shows the 30V/4A range is selected. OVP The overvoltage protection function is enabled when the annunciator
turns on or the overvoltage protection circuit has caused the power supply to shutdown when the annunciator blinks.
OCP The overcurrent protection function is enabled when the annunciator
turns on or the overcurrent protection circuit has caused the power supply to shutdown when the annunciator blinks.
CAL The power supply is in calibration mode. Limit The display shows the limit values of voltage and current. ERROR Hardware or remote interf ace command errors are detected a nd the error
OFF The output of the power supply is disabled (See
Unreg The output of the power supply is unregulated (output is neither CV
CV The power supply is in constant voltage mode. CC The power supply is in constant current mode.
bit has not been cleared.
information
nor CC).
).
page 52 for more
To review the display annunciators, hold down key as you
Display Limit
turn on the power supply.
5

The Rear Panel at a Glance

1 Power-line voltage setting 2 Power-line fuse-holder assembly 3 AC inlet
Use the front-panel key to:
• Select the GPIB or RS-232 interface (see chapter 3).
• Set the GPIB bus address (see chapter 3).
• Set the RS-232 baud rate and parity (see chapter 3).
I/O Config
4 Power-line module 5 GPIB (IEEE-488) interface connector 6 RS-232 interface connector
6

In This Book

Specifications Chapter 1 lists the power supply’s specifications and describes how to interpret these specifications.
Quick Start
familiar with the front-panel features.
Calibration Procedures Chapter 3 provides performance verification and calibration procedures.
Theory of Operation Chapter 4 describes block and circuit level theory related to the operation of the power supply.
Service Chapter 5 provides guidelines for returning your power supply to Agilent Technologies for servicing, or for servicing it yourself.
Replaceable Parts Chapter 6 contains a detailed parts list of the power supply.
Backdating Chapter 7 describes the difference between this manual and older issues of this manual.
Schematics Chapter 8 contains the power supply's schematics, disassembly drawings, and component locator drawings.
If you have questions relating to the operation of the power supply, call 1-800-452-4844 in the United States, or contact your nearest Agilent Technologies Sales Office.
If your Agilent E3632A fails within three years of purchase, Agilent will repair or replace it free of charge. Call 1-800-258-5165 ("Express Exchange") in the United States, or contact your nearest Agilent Technologies Sales Office.
Chapter 2 prepares the power supply for use and helps you get
7
8

Contents

Chapter 1 Specifications
Performance Specifications 15 Supplemental Characteristics 17
Chapter 2 Quick Start
To Prepare the Power Supply for Use 23 To Check the Rated Voltages of the Power Supply 25 To Check the Rated Currents of the Power Supply 26 To Use the Power Supply in Constant Voltage Mode 28 To Use the Power Supply in Constant Current Mode 30 To Store and Recall the Instrument State 32 To Program Overvoltage Protection 34
Setting the OVP Level and Enable the OVP Circuit 34 Checking OVP Operation 35 Clearing the Overvoltage Condition 35
To Program Overcurrent Protection 37
Setting the OCP Level and Enable the OCP Circuit 37 Checking OCP Operation 38 Clearing the Overcurrent Condition 38
To Rack Mount the Power Supply 39
Contents
Chapter 3 Calibration Procedures
Agilent Technologies Calibration Services 43 Calibration Interval 43 Automating Calibration Procedures 44 Recommended Test Equipment 44 Test Considerations 45 Performance Verification Tests 46
Self-Test 46 Performance Verification Tests 46
Measurement Techniques 47
Setup for Most Tests 47 Electronic Load 48 Current-Monitoring Resistor 48 Programming 48
9
Contents
Contents
Chapter 3 Calibration Procedures (continued)
Constant Voltage (CV) Verifications 49
Constant Voltage Test Setup 49 Voltage Programming and Readback Accuracy 49 CV Load Regulation 50 CV Line Regulation 50 Normal Mode Voltage Noise (CV Ripple and Noise) 51 Load Transient Response Time 52
Constant Current (CC) Verifications 53
Constant Current Test Setup 53 Current Programming and Readback Accuracy 53 CC Load Regulation 54 CC Line Regulation 55 Normal Mode Current Noise (CC Ripple and Noise) 55 Common Mode Current Noise 56
Performance Test Record for Agilent Technologies E3632A 57
CV Performance Test Record 57 CC Performance Test Record 57
Calibration Security Code 58
To Unsecure the Power Supply for Calibration 59
To Unsecure the Power Supply Without the Security Code 60 Calibration Count 61 Calibration Message 61 General Calibration/Adjustment Procedure 62
Voltage and OVP Calibration 63
Current and OCP Calibration 65 Aborting a Calibration in Progress 68 Calibration Record for Agilent Technologies E3632A 69 Error Messages 70 Calibration Program 72
10
Chapter 4 Theory of Operation
Block Diagram Overview 79 AC Input and Bias Supplies 81 Floating Logic 82 D-to-A Converter 84 A-to-D Converter 85 Power Mesh and Control 86 Earth-Referenced Logic 88 Front Panel 88
Contents
Chapter 5 Service
Operating Checklist 91 Types of Service Available 92 Repacking for Shipment 93 Electrostatic Discharge (ESD) Precautions 94 Surface Mount Repair 94 To Replace the Power-Line Fuse 94 To Disconnect the Output Using an External Relay 95 Troubleshooting Hints 96 Self-Test Procedures 98
Chapter 6 Replaceable Parts
Replaceable Parts 102
To Order Replaceable Parts 102
Backdating and Part Changes 102 E3632-60002 Main PC Assembly 103 E3632-60004 Front-Panel Display PC Assembly 111 E3632-60005 Front Frame Assembly 112 E3632A Power Supply Assembly 112 Manufacturer’s List 113
Contents
Chapter 7 Backdating
Chapter 8
Mechanical Disassembly 103 Component Locator 120 Power Circuit and Protection Circuit Schematic 121 Bias Supply Schematic 122 ADC and DAC System Schematic 123 Floating Logic Schematic 124 Earth-Referenced Logic Schematic 125 Display and Keyboard Schematic 126
11
Contents
Contents
12
1
1
Specifications

Specifications

The
performance
are warranted in the temperature range of 0 to 40°C with a resistive load.
Supplemental
performance determined either by design or testing. Chapter 3 "Calibration Procedures" contains procedures for verifying the performance specifications.
specifications are listed in the following pages. Specifications
characteristics, which are not warranted but are descriptions of
14
Chapter 1 Specifications

Performance Specifications

Performance Specifications
Output Ratings (@ 0°C - 40°C)
Low range 0 to +15 V/0 to 7 A High range 0 to +30 V/0 to 4 A
[1]
Programming Accuracy
Voltage 0.05% + 10 mV Current 0.2% + 10 mA
12 months (@ 25°C ± 5°C), ±(% of output + offset)
1
Readback Accuracy
to actual output @ 25°C ± 5°C), ±(% of output + offset)
Voltage 0.05% + 5 mV Current 0.15 % + 5 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded, 20 Hz to 20 MHz)
Normal mode voltage <0.35 mV rms and 2 mV p-p Current <2 mA rms Common mode current <1.5 mA rms
Load Regulation, ±(% of output + offset)
Change in output voltage or current for any load change within ratings with remote sensing connected.
Voltage <0.01% + 2 mV Current <0.01% + 250 mA
Line Regulation, ±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage <0.01% + 2 mV Current <0.01% + 250 mA
[1]
12 months (over GPIB and RS-232 or front panel with respect
[1]
Accuracy specifications are after an 1-hour warm-up with no load and calibration at
25°C
15
Chapter 1 Specifications
Performance Specifications
Programming Resolution
Voltage 1 mV Current 0.5 mA
Readback Resolution
Voltage 0.5 mV Current 0.1 mA
Front Panel Resolution
Voltage 1 mV Current 1 mA
Transient Response Time
Less than 50 msec for output recover to within 15 mV following a change in output current from full load to half load or vice versa
Command Processing Time
Average time for output voltage to begin to change after receipt of digital data when the power supply is connected directly to the GPIB or RS-232 is less than 100 msec.
OVP and OCP Accuracy, (% of output + offset)
OVP 0.5% + 0.5 V OCP 0.5% + 0.5 A
Activation time occurs.
OVP <1.5 msec when the trip voltage is equal or greater than 3 V
OCP <10 msec
: Average time for output to start to drop after OVP or OCP condition
<10 msec when the trip voltage is less than 3 V
16
Chapter 1 Specifications

Supplemental Characteristics

Supplemental Characteristics
Output Programming Range (maximum programmable values)
Low range 0 to 15.45 V/0 to 7.21 A High range 0 to 30.9 V/0 to 4.12 A OVP 1 V to 32 V OCP 0 A 7.5 A
Remote Sensing Capability
Voltage drop Up to 1 V per each lead Load regulation Add 5 mV to spec for each 1-volt change in the + output lead
due to load current changes.
Load voltage Subtract voltage drop in load leads from specified output
voltage atiing.
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage 0.01% + 3 mV Current 0.02% + 3 mA
1
Stability, ±(% of output + offset)
Following a 1 hour warm-up, change in output over 8 hours under constant load, line, and ambient temperature
Voltage 0.02% + 1 mV Current 0.1% + 1 mA
Voltage Programming Speed
Maximum time required for output voltage to settle within 1% of its total excursion (for resistive load). Excludes command processing time.
Full load
Up 50 msec 20 msec Down 45 msec 400 msec
No load
17
Chapter 1 Specifications
Supplemental Characteristics
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the (+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output to the (+) sense and the (-) output and the (-) sene terminals.
AC Input Ratings (selectable via rear panel selector)
std 115 Vac ± 10%, 47 to 63 Hz opt 0E3 230 Vac ± 10%, 47 to 63 Hz opt 0E9 100 Vac ± 10%, 47 to 63 Hz
Maximum Input Power
500 VA with full load
Cooling
Fan cooled
Operating Temperature
0 to 40°C for full rated output. At higher temperatures, the output current is derated linearly to 50% at 55°C maximum temperature.
Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the output control is set to less than 1 V. If the output control is set to 1 V or higher, there is no overshoot.
Programming Language
SCPI (Standard Commands for Programmable Instruments)
State Storage Memory
Three (3) user-configurable stored states
Recommended Calibration Interval
1 year
18
Chapter 1 Specifications
Supplemental Characteristics
Dimensions
213 mmW x 133 mmH x 348 mmD (8.4 x 5.2 x 13.7 in)
*See below for detailed information.
Weight
Net 9.5 kg (21 lb) Shipping 12 kg (26 lb)
*
1
Figure 1-1. Dimensions of Agilent E3632A Power Supply
19
Chapter 1 Specifications
Supplemental Characteristics
20
2
2
Quick Start

Quick Start

One of the first things you will want to do with your power supply is to become acquainted with its front panel. Written procedures in this chapter prepare the power supply for use and familiarize you with most front-panel operations.
• The power supply is shipped from the factory configured in the front-panel operation mode. At power-on, the power supply is automatically set to operate in the front-panel operation mode. When in this mode, the front­panel keys can be used. When the power supply is in remote operation mode, you can return to front-panel operation mode at any time by pressing the
Local
key if you did not previously send the front-panel lockout command. A change between front-panel and remote operation modes will not result in a change in the output parameters.
• The power supply has two output ranges of 15V/7A or 30V/4A. This feature allows more voltage at a lower current or more current at a lower voltage. The desired output range is selected from the front panel or over the remote interfaces. The range.
• When you press the key (the display of the power supply goes to the limit mode and the present limit values will be displayed. In this mode, you can also observe the change of the limit values when adjusting the knob. If you press the key again or let the display time-out after several seconds, the power supply will return the display to the meter mode (the In this mode, the actual output voltage and current will be displayed.
• The outputs of the power supply can be enabled or disabled from the front panel using the key. When the output is off, the annunciator turns on and the output is disabled.
• The display provides the present operating status of the power supply with annunciators and also informs the user of error codes. For example, the power supply is operating in CV mode in the 15V/7A range and controlled from the front panel, then the however, the power supply is remotely controlled, the also turn on, and when the power supply is being addressed over GPIB interface, the page 6 for more information.
15V or 30V annunciator indicates the presently selected
Display Limit
Output On/Off
CV and +15V annunciators will turn on. If,
Adrs annunciator will turn on. see “Display Annunciators”, on
Limit annunciator blinks), the
Display Limit
Limit annunciator turns off).
OFF
Rmt annunciator will
22
Chapter 2 Quick Start

To Prepare the Power Supply for Use

To Prepare the Power Supply for Use
The following steps help you verify that the power supply is ready for use.
Power
Check the list of supplied items.
1
Verify that you have received the following items with your power supply. If anything is missing, contact your nearest Agilent Technologies Sales Office.
One appropriate power cord for your location.
One User's Guide.
This Service Guide.
Certificate of Calibration.
Verify that the correct power-line voltage setting is selected and that
2
the correct power-line fuse is installed.
The line voltage is set to 100, 115 or 230 Vac from the factory according to the input power option selected when you ordered the power supply. Change the voltage setting if it is not correct for your location (see the next page for detailed information). For 100 or 115 Vac operation, the correct fuse is 4 AT (Agilent part number 2110-0996) and for 230 Vac operation, the correct fuse is
2.5 AT (Agilent part number 2110-0999).
Connect the power cord and turn on the power supply.
3
A power-on self-test occurs automatically when you turn on the power supply. The front-panel display will light up while the power supply performs its power-on self- test. After performing the self-test, the power supply will go into the power-on / reset state; all outputs are disabled (the on); the 15V/7A rage is selected (the + selected for voltage control. Notice that the turn on.
15V annunciator turns on); the knob is
OVP and OCP annunciators also
OFF annunciator turns
2
Output On/Off
Enable the outputs.
4
Press the key to enable the outputs. The turns off and the can be adjusted by turning the knob. Notice that the display is in the meter mode. "Meter mode" means that the display shows the actual output voltage and current.
Output On/Off
15V, OVP, OCP a nd CV annunc iato rs are li t. The blinking digit
OFF annunciator
23
Chapter 2 Quick Start
To Prepare the Power Supply for Use
1 Remove the power cord. Rem ove
the fuse-holder assembly with a flat­blade screwdriver from the rear
3 Rotate the power-line voltage
selector until the correct voltage appears.
2 Install the corect line fuse. Remove
the power-line voltage selector from the power-line module.
100 or 115 Vac, 4 AT fuse 230 Vac, 2.5 AT fuse
4 Replace the power-line voltage
selector and the fuse-holder assembly in the rear panel.
100, 115 or 230 Vac
Install the correct fuse and verify that the correct line voltage appears in the window.
24
Chapter 2 Quick Start

To Check the Rated Voltages of the Power Supply

To Check the Rated Voltages of the Power Supply
The following procedures check to ensure that the power supply develops its rated voltage outputs with no load and properly responds to operation from the front panel.
For each step, use the keys shown on the left margins.
2
Power
Output On/Off
Turn on the power supply.
1
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 15V/7A range is selected (the +15V
(the annunciator turns on); and the knob is selected for voltage control.
Enable the outputs.
2
OFF annunciator turns off and the 15V, OVP, O CP and CV annunciators are
The lit. The blinking digit can be adjusted by turning the knob. Notice that the display is in the meter mode. "Meter mode" means that the display shows the actual output voltage and current.
Check that the front-panel voltmeter properly responds to knob
3
control.
Turn the knob clockwise or counter clockwise to check that the voltmeter responds to knob control and the ammeter indicates nearly zero.
Ensure that the voltage can be adjusted from zero to the maximum
4
rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob until the voltmeter indicates 15.0 volts.
1
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the voltage.
25
Chapter 2 Quick Start

To Check the Rated Currents of the Power Supply

To Check the Rated Currents of the Power Supply
The following procedures check to ensure that the power supply develops its rated current outputs with a short and properly responds to operation from the front panel.
For each step, use the keys shown on the left margin.
Power
Output On/Off
Display Limit
Volt/Curr
Turn on the power supply.
1
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 15V/7A range is selected (the 15V
(the annunciator turns on); and the knob is selected for voltage control.
Connect a short across (+) and (-) output terminals with an insulated
2
test lead.
Enable the outputs.
3
OFF annunciator turns off and the 15V, OVP and OCP annunciator turns
The on. The
CV or CC annunciator is lit depending on the resistance of the test lead.
The blinking di g it can be adju sted by tu r ning the kno b. N otice th at t he displ ay is in the meter mode. "Meter mode" means that the display shows the actual output voltage and current.
Adjust the voltage limit value to 1.0 volt.
4
Set the display to the limit mode (the the voltage limit to 1.0 volt to assure CC operation. The
Lmt annunciator will be blinking). Adjust
CC annunciator will
light.
Check that the front-panel ammeter properly responds to knob control.
5
Set the knob to the current control, and turn the knob clockwise or counter clockwise when the display is in the meter mode (the
Lmt annunciator is off).
Check that the ammeter responds to knob control and the voltmeter indicates nearly zero (actually, the voltmeter will show the voltage drop caused by the test lead).
26
Chapter 2 Quick Start
To Check the Rated Currents of the Power Supply
Ensure that the current can be adjusted from zero to the maximum
6
rated value.
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter indicates 7.0 amps.
Note If an error has been detected during the output checkout procedures, the ERROR
annunciator will turn on. See "Error Messages" for more information, starting on page 121 in chapter 5 of the User's Guide.
1
2
1
You can use the resolution selection keys to move the blinking digit to the
right or left when setting the current.
27
Chapter 2 Quick Start

To Use the Power Supply in Constant Voltage Mode

To Use the Power Supply in Constant Voltage Mode
To set up the power supply for constant voltage (CV) operation, proceed as follows.
For each step, use the keys shown on the left margin.
Connect a load to the desired output terminals.
1
With power-off, connect a load to the desired output terminals.
Power
Display Limit
Volt/Curr
Turn on the power supply.
2
The power supply will go into the power-on / reset state; output is disabled (the
OFF annunciator turns on); the 15V/7A range is selected (the 15V annunciator
turns on); and the knob is selected for voltage control.
To operate the power supply in the 30V/4A range, press the key before
30V,4A
proceeding to the next step. The 30V annunciator turns on.
Set the display for the limit mode.
3
Notice that the
Limit annunciator blinks, indicating that the display is in the
limit mode. When the disp lay is in the limit mode, you can see the voltage and current limit values of the power supply.
In constant voltage mode, the voltage values between meter mode and limit mode are the same, but the current values are not. Moreover if the display is in the meter mode, you cannot see the change of current limit value when adjusting the knob. We recommend that you should set the display to "limit" mode to see the change of current limit value in constant voltage mode whenever adjusting the knob.
Adjust the knob for the desired current limit.
4
1
Check that the Lmt annunciator still blinks. Set the knob for current control. The second digit of ammeter will be blinking. The blinking digit can be changed using the resolution selection keys and the blinking digit can be adjusted by turning the knob. Adjust the knob to the desired current limit.
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting current.
28
Chapter 2 Quick Start
To Use the Power Supply in Constant Voltage Mode
Volt/Curr
Adjust the knob for the desired output voltage.
5
1
Check that the Limit annunciator still blinks. Set the knob for voltage control. The second digit of the voltmeter will be blinking. Change the blinking digit using the resolution selection keys and adjust the knob to the desired output voltage.
Display Limit
Return to the meter mode.
6
Press the key or let the display time-out after several seconds to return to the meter mode. Notice that the
Display Limit
Limit annunciator turns off and
the display shows "OUTPUT OFF" message.
Output On/Off
Enable the outputs
7
OFF annunciator turns off and the 15V or 30V, OVP, OCP, and CV
The annunciators are lit. Notice that the display is in the meter mode. In the meter mode, the display shows the actual output voltage and current.
Verify that the power supply is in the constant voltage mode.
8
If you operate the power supply in the constant voltage (CV) mode, verify that
CV annunciator is lit. If the CC annunciator is lit, choose a higher current limit.
Note During actual CV operation, if a load change causes the current limit to be exceeded,
the power supply will automatically crossover to constant current mode at the preset current limit and the output voltage will drop proportionately.
2
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting voltage.
29
Chapter 2 Quick Start

To Use the Power Supply in Constant Current Mode

To Use the Power Supply in Constant Current Mode
To set up the power supply for constant current (CC) operation, proceed as follows.
For each step, use the keys shown on the left margin.
Connect a load to the output terminals.
1
With power-off, connect a load to the (+) and (-) output terminals.
Power
Display Limit
Turn on the power supply.
2
The power supply will go into the power-on / reset state; the output is disabled
OFF annunciator turns on); the 15V/7A range is selected (the 15V
(the annunciator turns on); and the knob is selected for voltage control.
To operate the power supply in the 30V4A range, press key before proceeding to the next step. The
Set the display for the limit mode.
3
Notice that the
Limit annunciator blinks, indicating that the display is in the
30V annunciator turns on.
30V,4A
limit mode. When the disp lay is in the limit mode, you can see the voltage and current limit values of the selected supply.
In constant current mode, the current values between meter mode and limit mode are the same, but the voltage values are not. Moreover if the display is in the meter mode, you cannot see the change of voltage limit value when adjusting the knob. We recommend that you should set the display to "limit" mode to see the change of voltage limit value in constant current mode whenever adjusting the knob.
Adjust the knob for the desired voltage limit.
4
1
Check that the Limit annunciator still blinks and the second digit of voltmeter blinks to indicate the knob is selected for voltage control. The blinking digit can be changed using the reolution keys and the blinking digit can be adjusted by turning the knob. Adjust the knob for the desired voltage limit.
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the voltage.
30
Chapter 2 Quick Start
To Use the Power Supply in Constant Current Mode
Volt/Curr
Adjust the knob for the desired output current.
5
1
Check that the Limit annunciator still blinks. Set the knob for current control. The second digit of the ammeter will be blinking. Change the blinking digit using the resolution selection keys and adjust adjust the knob to the desired output current.
Display Limit
Return to the meter mode.
6
Press the key or let the display time-out after several seconds to return the meter mode. Notice that the
Display Limit
Limit annunciator turns off and the
display shows "OUTPUT OFF" message.
Output On/Off
Enable the outputs.
7
OFF annunciator turns off and the 15V or 30V, OVP, OCP, and CC
The annunciators are lit. Notice that the display is in the meter mode. In the meter mode, the display shows the actual output voltage and current.
Verify that the power supply is in the constant current mode.
8
If you operate the power supply in the constant current (CC) mode, verify that
CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage limit.
Note During actual CC operation, if a load change causes the voltage limit to be exceeded,
the power supply will automatically crossover to constant voltage mode at the preset voltage limit and the output current will drop proportionately.
2
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the voltage and current.
31
Chapter 2 Quick Start

To Store and Recall the Instrument State

To Store and Recall the Instrument State
You can store up to three different operating states in non-volatile memory. This also enables you to recall the entire instrument with just a few key presses from the front panel.
The memory locations are supplied with the reset state from the factory for front panel operation. Refer to the description of page 94 in the User’s Guide for more information.
The following steps show you how to store and recall an operating state.
Set up the power supply for the desired operating state.
1
The storage feature "remembers" the output range selection, the limit value settings of voltage and current, output on/off state , OVP and OCP on/off state, and OVP and OCP trip levels.
command, starting on
*RST
Store
Turn on the storage mode.
2
Three memory locations (numbered 1, 2, and 3) are available to store the operating states. The operating states are stored in non-volatile memory and are remembered when being recalled.
Store 1
This message appears on the display for approximately 3 seconds.
Store the operating state in memory location "3".
3
Turn the knob to the right to specify the memory location 3.
Store 3
To cancel the store operation, let the display time-out after about 3 seconds or press any other function key except the key. The power supply returns to the normal operating mode and to the function pressed.
Store
32
Chapter 2 Quick Start
To Store and Recall the Instrument State
Store
Recall
Save the operating state.
4
The operating state is now stored. To recall the stored state, go to the following steps.
done
This message appears on the display for approximately 1 second.
Turn on the recall mode.
5
Memory location "1" will be displayed in the recall mode.
recall 1
This message appears on the display for approximately 3 seconds.
Recall the stored operating state.
6
Turn the knob to the right to change the displayed storage location to "3".
recall 3
If this setting is not followed within 3 seconds with a key stroke, the power supply returns to normal operating mode and will not recall the instrument state 3 from memory.
Recall
2
Recall
Restore the operating state.
7
The power supply should now be configured in the same state as when you stored the state on the previous steps.
done
This message appears on the display for approximately 1 second.
33
Chapter 2 Quick Start

To Program Overvoltage Protection

To Program Overvoltage Protection
Overvoltage protection guards the load against output voltages that reach a specified value greater than the programmed protection level. It is accomplished by shorting the output via an internal SCR when the trip level is set to equal or greater than 3 volts, or by progamming the output to 1 volt when the trip level is set to less than 3 volts.
The following steps show how to set the OVP trip level, how to check OVP operation and how to clear overvoltage condition.

Setting the OVP Level and Enable the OVP Circuit

Power
Turn on the power supply.
1
The power supply will go into the power-on / reset state; the output is disabled (the
OFF annunciator turns on); the 15V/7A range is selected (the 15V
annunciator turns on); and the knob is selected for voltage control.
Output On/Off
Over Voltage
Enable the output.
2
OFF annunciator turns off and the display will go to the meter mode.
The
Enter the OVP menu and set the trip level.
3
level 32.0 v
You will see the above message on the display when you enter the OVP menu. Adjust the knob for the desired OVP trip level.
Note You cannot set the trip levels to lower than 1.0 volt.
Over Voltage
Enable the OVP circuit.
4
ovp on
Over Voltage
You will see the above message after pressing the key.
Over Voltage
34
Chapter 2 Quick Start
To Program Overvoltage Protection
Exit the OVP menu.
5
changed
Over Voltage
Display Limit
The "CHANGED" message is displayed for a second to show that the new OVP trip level is now in effect. If the OVP settings are not changed, "NO CHANGE" will be displayed. The power supply will exit the OVP menu and the display will return to the meter mode. Check that the
OVP annunciator turns on.

Checking OVP Operation

To check OVP operation, raise the output voltage to near the trip point. Then very gradually increase the output by turning the knob until the OVP circuit trips. This will cause the power supply output to drop to near zero. The annunciator is blinking an d the
CC annunicator turns on. The "OVP TRIPPED"
message also appears on the display.

Clearing the Overvoltage Condition

When the OVP circuit trips (the "OVP TRIPPED" message is shown on the display), the source such as a battery, disconnect it first. The following steps show how to clear the overvoltage conditions and get back to normal mode operation. In the following steps, the display will go back to "OVP TRIPPED" if you let the display time out after about several seconds.
or
Readjust the OVP trip level or the output voltage level.
1
Lower the output voltage level below the OVP trip point after pressing the
Display Limit
Over Voltage
the key.
OVP annunicator blinks. When it was caused by external voltage
or raise the OVP trip level by using the knob after pressing
2
OVP
Over Voltage
Move to the clear mode.
2
ovp on
You will see the above message after pressing the key. If you changed the output voltage level, press the key twice. Turn the
Over Voltage
Over Voltage
knob to the right until the "OVP CLEAR" appears on the display.
35
Chapter 2 Quick Start
To Program Overvoltage Protection
Over Voltage
Clear the overvoltage condition and exist this menu.
3
Now, when you press the key again, the "DONE" message is displayed for a second and the
Over Voltage
OVP annunciator will not blink any more. The
output will return to meter mode.
36
Chapter 2 Quick Start

To Program Overcurrent Protection

To Program Overcurrent Protection
Overcurrent protection guards the load against output currents that reach a specified value greater than the programmed protection level. It is accomplished by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how to check OCP operation and how to clear overcurrent condition.

Setting the OCP Level and Enable the OCP Circuit

2
Power
Output On/Off
Over Current
Over Current
Over Current
Turn on the power supply.
1
The power supply will go into the power-on / reset state; the output is disabled (the
OFF annunciator turns on); the 15V/7A range is selected (the 15V
annunciator turns on); and the knob is selected for voltage control.
Enable the output.
2
OFF annunciator turns off and the display will go to the meter mode.
The
Enter the OCP menu and set the trip level.
3
level 7.5 a
You will see the above message on the display when you enter the OCP menu. Adjust the knob for the desired OCP trip level.
Enable the OCP circuit.
4
ocp on
You will see above message after pressing the key .
Exit the OCP menu.
5
Over Current
changed
37
Chapter 2 Quick Start
To Program Overcurrent Protection
The "CHANGED" message is displayed for a second to show that the new OCP trip level is now in effect. If the OCP settings are not changed, "NO CHANGE" will be displayed. The power supply will exit the OCP menu and the display will return to the meter mode. Check that the
OCP annunciator turns on.

Checking OCP Operation

To check OCP operation, raise the output current to near the trip point. Then very gradually increase the output by turning the knob until OCP circuit trips. This will cause the power supply's output current to drop to zero and the
OCP
annunciator is blinking. The "OCP TRIPPED" message also appears on the display.

Clearing the Overcurrent Condition

When the OCP circuit trips (the "OCP TRIPPED" message is shown on the display), the source such as a battery, disconnect it first. The following steps show how to clear the overcurrent conditions and get back to normal mode operation. In the following steps, the display will go back to "OCP TRIPPED" if you let the display time out after about several seconds.
OCP annunicator blinks. When it was caused by external voltage
Over Current
Display Limit
Over Current
or
Readjust the OCP trip level or the output current level.
1
Lower the output current level below the OCP trip point after pressing the
Display Limit
Over Current
the key.
Move to the clear mode.
2
or raise the OCP trip level by using the knob after pressing
ocp on
You will see the above message after pressing the key. If you changed the output current level, press the key twice. Turn the
Over Current
Over Current
knob to the right until the "OCP CLEAR" appears on the display.
Clear the overcurrent condition and exit this menu.
3
Now, when you press the key again, the "DONE" message is displayed for a second and the
Over Current
OCP annunciator will not blink any more. The
output will return to meter mode.
38
Chapter 2
Quick Start
To Rack Mount the Power Supply
To Rack Mount the Power Supply
The power supply can be mounted in a standard 19-inch rack cabinet using one of three optional kits available. A rack-mounting kit for a single instrument is available as Option 1CM (P/N 5063-9243). Installation instructions and hardware are included with each rack-mounting kit. Any Agilent System II instrument of the same size can be rack-mounted beside the Agilent E3632A power supply. To rack mount the power supply, follow these procedures.
Remove the front and rear bumpers before rack-mounting the power supply.
2
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
39
Chapter 2
Quick Start

To Rack Mount the Power Supply

To rack mount two instruments of the same depth side-by-side, order lock-link kit 5061-9694 and flange kit 5063-9214.
To install two instruments in a sliding support shelf, order support shelf 5063­9256, and slide kit 1494-0015.
40
3
Calibration Procedures

Calibration Procedures

This chapter contains procedures for verification of the power supply's performance and calibration (adjustment). The chapter is divided into the following sections:
• Agilent Technologies Calibration Services, page 43
• Calibration Interval, page 43
• Automating Calibration Procedures, page 44
• Recommended Test Equipment, page 44
• Test Considerations, page 45
• Performance Verification Tests, page 46
• Measurement Techniques, page 47
• Constant Voltage (CV) Verifications, page 49
• Constant Current (CC) Verifications, page 53
• Performance Test Record for Agilent E3632A, page 57
• Calibration Security Code, page 58
• Calibration Count, page 61
• Calibration Message, page 61
• General Calibration/Adjustment Procedure, page 62
• Aborting a Calibration in Progress, page 68
• Calibration Record for Agilent E3632A, page 69
• Error Messages, page 70
• Calibration Program, page 72
The performance verification tests for constant voltage (CV) and constant current (CC) operations use the power supply's specifications listed in chapter 1, "Specifications", starting on page 13.
42
Chapter 3 Calibration Procedures

Agilent Technologies Calibration Services

Closed-Case Electronic Calibration The power supply features closed­case electronic calibration since no internal mechanical adjustments are required for normal calibration. The power supply calculates correction factors based upon the input reference value you enter. The new correction factors are stored in non-volatile memory until the next calibration adjustment is performed. (Non-volatile memory does not change when power has been off or after a remote interface reset.)
Agilent Technologies Calibration Services
When your power supply is due for calibration, contact your local Agilent Technologies Service Center for a low-cost calibration. The Agilent E3632A power supply is supported on calibration processes which allow Agilent Technologies to provide this service at competitive prices.
3

Calibration Interval

The power supply should be calibrated on a regular interval determined by the accuracy requirements of your application. A 1-year interval is adequate for most applications. Agilent Technologies does not recommend extending calibration intervals beyond 1 year for any application. Agilent Technologies recommends that complete re-adjustment should always be performed at the calibration interval. This will increase your confidence that the Agilent E3632A will remain within specification for the next calibration interval. This criteria for re-adjustment provides the best long-term stability.
43
Chapter 3 Calibration Procedures

Automating Calibration Procedures

Automating Calibration Procedures
You can automate the complete verification procedures outlined in this chapter if you have access to programmable test equipment. You can program the instrument configurations specified for each test over the remote interface. You can then enter readback verification data into a test program and compare the results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote operation is similar to the local front-panel procedure. You can use a computer to perform the adjustment by first selecting the required setup. The calibration value is sent to the power supply and then the calibration is initiated over the remote interface. The power supply must be unsecured prior to initiating the calibration procedure. An Agilent BASIC program for calibration over the GPIB interface is listed at the end of this chapter.
For further details on programming the power supply, see chapters 3 and 4 in the Agilent E3632A User's Guide.

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, use the accuracy requirements shown to select substitute calibration standards.
Table 3-1 Recommended Test Equipment
Instrument Requirements
GPIB Controller Full GPIB capabilities
Oscilloscope
RMS Voltmeter 20 MHz Measure rms ripple & noise
Digital Voltmeter
Electronic Load
Resistive Loads (R
Current monitoring Resistor (Shunt)
100 MHz with 20MHz bandwidth
Resolution: 0.1 mV Accuracy: 0.01%
Voltage Range: 50 Vdc Current Range: 10 Adc Open and Short Switches Transient On/Off
2.19, 200 W
)
L
7.59, 200 W
0.019, 0.01%
Recommended
Model
Agilent 82341C Interface card
Agilent 54602B
Agilent 34401A Measure dc voltages
Agilent 6063B
Programming and readback accuracy
Display transient response and ripple & noise waveform
Measure load and line regulations and transient response time.
Measure ripple and noise
Constant current test setup
Use
44
Chapter 3 Calibration Procedures

Test Considerations

Test Considerations
To ensure proper instrument operation, verify that you have selected the correct power-line voltage prior to attempting any test procedure in this chapter. page 24 for more information.
For optimum performance verification, all test procedures should comply with the following recommendations.
• Assure that the calibration ambient temperature is stable and between 20°C and 30°C.
• Assure ambient relative humidity is less than 80%.
• Allow a 1-hour warm-up period before verification or calibration.
• Keep cables as short as possible, consistent with the impedance requirements.
3
Caution
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages may be present at the outputs of the power supply.
45
Chapter 3 Calibration Procedures

Performance Verification Tests

Performance Verification Tests

The performance verification tests use the power supply's specifications listed in chapter 1, "Specifications", starting on page 13.
You can perform two different levels of performance verification tests:
Self-Test A series of internal verification tests that provide high
confidence that the power supply is operational.
Performance Verification Tests These tests can be used to verify the
power supply's specifications following repairs to specific circuits.

Self-Test

A powe r-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the power supply is operational.
The complete self-test is enabled by pressing the key (actually any front panel keys except the key) and the power-line switch simultaneously and then continuing to press the key for 5 seconds. The complete self-test will be finished in 2 more seconds.
You can also perform a self-test from the remote interface (see chapter 3 in the Agilent E3632A User's Guide).
• If the self-test is successful, "PASS" is displayed on the front panel.
• If the self-test fails, "FAIL" is displayed and the ERROR annunciator turns
on. If repair is required, see chapter 5, "Service", starting on page 89, for further details.
• If self-test passes, you have a high confidence that the power supply is operational.
Error
Recall
Recall
Performance Verification Tests
These tests can be used to verify the power supply's specifications following repairs to specific circuits. The following sections explain all verification procedures in detail. All of the performance test specifications are shown in each test.
46
Chapter 3
Calibration Procedures

Measurement Techniques

Measurement Techniques

Setup for Most Tests

Most tests are performed at the front terminals as shown in the following figure. Measure the dc voltage directly at the (+) and (-) terminals on the front panel.
3
Figure 3-1. Performance Verification Test Setup
Electronic Load
Many of the test procedures require the use of a variable load resistor capable of dissipating the required power. Using a variable load resistor requires that switches be used to connect, disconnect, and short the load resistor. An electronic load, if available, can be used in place of a variable load resistor and switches. The electronic load is considerably easier to use than load resistors. It eliminates the need for connecting resistors or rheostats in parallel to handle power, it is much more stable than carbon-pile load, and it makes easy work of switching between load conditions as is required for the load regulation and load transient response tests. Substitution of the electronic load requires minor changes to the test procedures in this chapter.
47
Chapter 3
Calibration Procedures
Measurement Techniques
General Measurement Techniques
To achieve best results when measuring load regulation, peak to peak voltage, and transient response time of the power supply, measuring devices must be connected through the hole in the neck of the binding post at (A) while the load resistor is plugged into the front of the output terminals at (B). A measurement made across the load includes the impedance of the leads to the load. The impedance of the load leads can easily be several orders of the magnitude greater than the power supply impedance and thus invalidate the measurement. To avoid mutual coupling effects, each measuring device must be connected directly to the output terminals by separate pairs of leads.
Front Panel Terminal Connections (Side View)

Current-Monitoring Resistor

To eliminate output current measurement error caused by the voltage drops in the leads and connections, connect the current monitoring resistor between the (-) output terminal and the load as a four-terminal device. Connect the current-monitoring leads inside the load-lead connections directly at the monitoring points on the resistor element (see RM in Figure 3-1).

Programming

Most performance tests can be performed only from the front panel. However, an GPIB or RS-232 controller is required to perform the voltage and current programming accuracy and readback accuracy tests.
The test procedures are written assuming that you know how to program the power supply either from the front panel or from an GPIB or RS-232 controller. Complete instructions on front panel and remote programming are given in the Agilent E3632A User's Guide.
48
Chapter 3 Calibration Procedures

Constant Voltage (CV) Verifications

Constant Voltage (CV) Verifications

Constant Voltage Test Setup

If more than one meter or a meter and an oscilloscope are used, connect each to the (+) and (- ) terminals b y a sepa rate pai r of leads to a void mutua l coupling effects. Use coaxial cable or shielded 2-wire cable to avoid noise pick-up on the test leads.

Voltage Programming and Readback Accuracy

This test verifies that the voltage programming and GPIB or RS-232 readback functions are within specifications. Note that the readback values over the remote interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test to avoid round off errors.
Turn off the power supply and connect a digital voltmeter between the (+) and
1
(-) terminals of the output to be tested as shown in Figure 3-1.
Turn on the power supply. Select the 30V/4A range and enable the outputs by
2
sending the commands:
VOLT:RANG P30V OUTP ON
Program the output voltage to zero volt and current to full rated value (4.0A)
3
by sending the commands:
VOLT 0 CURR 4
Record the ou tput voltage r eading o n the digital v oltmeter (DVM) . The reading
4
should be within the limits of (0V ± 10 mV). Also, note that the
Rmt annunciators are on.
and
Readback the output voltage over the remote interface by sending the
5
command:
MEAS:VOLT?
CV, Adrs, Limit,
3
Record the value displayed on the controller. This value should be within the
6
limits of (DVM ± 5 mV)
49
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications
Program the output voltage to full rated value (30.0V) by sending the command.
7
VOLT 30.0
Record the output voltage reading on the digital voltmeter (DVM). The readings
8
should be within the limits of (30V
Readback the output voltage over the remote interface by sending the
9
command:
MEAS:VOLT?
Record the value displayed on the controller. This value should be within the
10
limits of (DVM

0mV).

5mV).

CV Load Regulation

This test measures the change in the output voltage resulting from a change in the output current from full to no load.
Turn off the power supply and connect a digital voltmeter between the (+) and
1
(-) terminals of the output to be tested as shown in Figure 3-1.
Turn on the power supply. Select the 30V/4A range, enable the output, and set
2
the display to the limit mode. When the display is in the limit mode, program the output current to the full rated value (4.0A) and the voltage to the full rated value (30.0V)
Operate the electronic load in constant current mode and set its current to
3
4.0A. Check that the front panel
the load so that the output current drops slightly until the lights. Record the output voltage reading on the digital voltmeter.
Operate the electronic load in open mode (input off). Record the output voltage
4
reading on the digital voltmeter again. The difference between the digital voltmeter readings in steps (3) and (4) is the CV load regulation. The difference of the readings should be within the limit of 5mV.
CV annunciator remains lit. If not lit, adjust
CV annunciator

CV Line Regulation

This test measures the change in output voltage that results from a change in ac line voltage from the minimum value (10% below the nominal input voltage) to maximum value (10% above the nominal input voltage).
Turn off the power supply and connect a digital voltmeter between the (+) and
1
(-) terminals of the output to be tested as shown in Figure 3-1.
50
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications
Connect the ac power line through a variable voltage transformer.
2
Turn on the power supply. Select the 30V/4A range, enable the output, and set
3
the display to the limit mode. When the display is in the limit mode, program the current to the full rated value (4.0A) and the voltage to full rated value (30.0V).
Operate the electronic load in constant current mode and set its current to
4
4.0A. Check that the that the output current drops slightly until the
Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
5
90 Va c f or no mi nal 10 0 V ac , o r 20 7 V ac fo r n omin al 23 0 Vac ). Re co rd th e o ut pu t reading on the digital voltmeter.
Adjust the autotranformer to high line voltage (127 Vac for nominal 115 Vac,
6
110 Vac for nominal 100 Vac, or 253 Vac for nominal 230 Vac). Record the voltage reading on the digital voltmeter. The difference between the digital voltmeter readings in steps (5) and (6) is the CV line regulation. The difference of the readings should be within the limit of 5mV.
CV annunciator remains lit. If not lit, adjust the load so
CV annunciator lights.
3

Normal Mode Voltage Noise (CV Ripple and Noise)

The normal mode voltage noise is in the form of ripple related to the line frequency plus some random noise. The normal mode voltage noise is specified as the rms or peak-to-peak output voltage in a frequency range from 20 Hz to 20 MHz.
Turn off the power supply and connect the output to be tested as shown in Figure
1
3-1 to an oscilloscope (ac coupled) between (+) and (-) terminals. Set the oscilloscope to AC mode and bandwidth limit to 20 MHz. Connect a resistive load (7.5) as shown in Figure 3-1.
Turn on the power supply. Select the 30V/4A range, enable the output, and set
2
the display to the limit mode. When the display is in the limit mode, program the current to the full rated value (4.0A) and the voltage to the full rated value (30.0V).
Check that the front panel
3
down slightly.
Note that the waveform on the oscilloscope does not exceed the peak-to-peak
4
limit of 2 mV.
Disconnect the oscilloscope and connect an AC rms voltmeter in its place. The
5
rms voltage reading does not exceed the rms limit of 0.35 mV.
CV annunciator remains lit. If not lit, adjust the load
51
Chapter 3 Calibration Procedures
Constant Voltage (CV) Verifications

Load Transient Response Time

This test measures the time for the output voltage to recover to within 15 mV of nominal output voltage following a load change from full load to half load, or half load to full load.
Turn off the power supply and connect the output to be tested as shown in Figure
1
3-1 with an oscilloscope. Operate the electronic load in constant current mode.
Turn on the power supply. Select the 30V/4A enable the outputs and set the
2
display to the limit mode. When the display is in the limit mode, program the current to the full rated value (4.0A) and the voltage to the full rated value (3.0V).
Set the electronic load to transient operation mode between one half of the
3
output's full scale value and the output's full rated value at a 1 kHz rate with 50% duty cycle.
Set the the oscilloscope for ac coupling, internal sync, and lock on either the
4
positive or negative load transient.
Adjust the the oscilloscope to display transients as shown in Figure 3-2. Note
5
that the pulse width (t more than 50
52
msec for the output.
) of the transients at 15 mV from the base line is no
2-t1
Figure 3-2. Transient Response Time
Chapter 3 Calibration Procedures

Constant Current (CC) Verifications

Constant Current (CC) Verifications

Constant Current Test Setup

Follow the general setup instructions in the "Measurement Techniques" section, starting on page 47 and the specific instructions will be given in the following paragraphs.

Current Programming and Readback Accuracy

This test verifies that the current programming and GPIB or RS-232 readback functions are within specifications. Note that the readback values over the remote interface should be identical to those displayed on the front panel. The accuracy of the current monitoring resistor must be 0.01% or better.
You should program the power supply over the remote interface for this test to avoid round off errors.
Tu rn of f th e po wer su pply and con nect a 0 .01 current monitoring resistor (R
1
across the output to be tested and a digital voltmeter across the current monitoring resistor (R
Turn on the power supply. Select the 15V/7A range and enable the output by
2
sending the commands:
VOLT:RANG P15V OUTP ON
Program the output voltage to full rated voltage (15.0V) and output current to
3
zero amp by sending the commands:
VOLT 15 CURR 0
Divide the voltage drop (DVM reading) across the current monitoring resistor
4
) by its resistance to convert to amps and record this value (IO). This value
(R
M
should be within the limits of(0A
Rmt annunciators are on.
and
).
M
10mA). Also, note that the

CC, Adrs, Limit,
3
)
M
Readback the output current over the remote interface by sending the
5
command:
MEAS:CURR?
Record the value displayed on the controller. This value should be within the
6
5mA).
limit of (I

O
53
Chapter 3 Calibration Procedures
Constant Current (CC) Verifications
Program the output current to the full rated value (7.0A) by sending the
7
commands:
CURR 7.0
Divide the voltage drop (DVM reading) across the current monitoring resistor
8
(R
) by its resistance to convert to amps and record this value (IO). This value
M
should be within the limit of (7A
Readback the output current over the remote interface by sending the
9

4mA).
command:
MEAS:CURR?
Record the value displayed on the controller. This value should be within the
10
O
5.5mA).

limit (I

CC Load Regulation

Th is test m e asures th e change in ou tpu t current r e sultin g from a chang e in the load from full-rated output voltage to short circuit.
Turn off the power supply and connect the output to tested as shown in Figure
1
3-1 with the digital voltmeter connected across the 0.01 current monitoring resistor (R
Turn on the power supply. Select the 15V/7A range, enable the output, and set
2
).
M
the display to the limit mode. When the display is in the limit mode, program the output voltage to the full rated value (15.0V) and the output current to the full rated value (7.0A).
Operate the electronic load in constant voltage mode and set its voltage to
3
15.0V. Check that the CC annunciator is on. If it is not, adjust the load so that
the output voltage drops slightly. Record the current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.
Operate the electronic load in short (input short) mode. Record the current
4
reading again by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (3) and (4) is the load regulation current. The difference of the readings should be within the limit of 0.95mA.
54
Chapter 3 Calibration Procedures
Constant Current (CC) Verifications

CC Line Regulation

This test measures the change in output current that results from a change in ac line voltage from the minimum value (10% below the nominal input voltage) to the maximum value (10% above nominal voltage).
Turn off the power supply and connect the output to be tested as shown in Figure
1
3-1 with the digital voltmeter connected across the current monitoring resistor
).
(R
M
Connect the ac power line through a variable voltage transformer.
2
Turn on the power supply. Select the 15V/7A range, enable the output, and set
3
the display to the limit mode. When the display is in the limit mode, program the voltage to the full rated value (15.0V) and the current to the full rated value (7.0A).
Operate the electronic load in constant voltage mode and set its voltage to
4
15.0V. Check that the that the output voltage drops slightly until the
CC annunciator remains lit. If not lit, adjust the load so
CC annunciator lights.
3
Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
5
90 Va c f or no mi nal 10 0 V ac , o r 20 7 V ac fo r n omin al 23 0 Vac ). Re co rd th e o ut pu t current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.
Adjust the transformer to 10% above the nominal line voltage (127 Vac for a
6
115 Vac nominal input, 110 Vac for a 100 Vac nominal input or 253 Vac for a 230 Vac nominal input). Record the current reading again by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (5) and (6) is the load regulation current. The difference of the readings should be within the limit of 0.95mA.

Normal Mode Current Noise (CC Ripple and Noise)

The normal mode current noise is specified as the rms output current in a frequency range 20 Hz to 20 MHz with the power supply in constant current operation.
Turn off the power supply and connect the output to be tested as shown in Figure
1
3-1 with a load resistor (2.19) across output terminals to be tested. Connect a rms voltmeter across the load resistor. Use only a resistive load for this test.
55
Chapter 3 Calibration Procedures

Common Mode Current Noise

Turn on the power supply. Select the 15V/7A range, enable the output, and set
2
the display to the limit mode. When the display is in the limit mode, program the current to full rated value (7.0A) and the voltage to the full rated value (15.0V).
The output current should be at the full-rated rating with the CC annunciator
3
on. If not lit, adjust the load so that the output voltage drops slightly until the CC annunciator lights.
Divide the reading on the rms voltmeter by the load resistance to obtain rms
4
current. The readings should be within the limit of 2mA.
Common Mode Current Noise
The common mode current is that ac current component which exists between the output or output lines and chassis ground. Common mode noise can be a problem for very sensitive circuitry that is referenced to earth ground. When a circuit is referenced to earth ground, a low level line-related ac current will flow from the output terminals to earth ground. Any impendance to earth ground will create a voltage drop equal to the output current flow multiplied by the impendance.
Turn off the power supply and connect a 100K9 resistor (R
1
) and a 2200 pF
S
capacitor in parallel between the (-) terminal and chassis ground.
Connect a digital voltmeter across R
2
Turn on the power supply. Select the 15V/7A range, enable the output, and set
3
.
S
the display to the limit mode. When the display is in the limit mode, program the output to the full rated value (15.0V and 7.0A).
Record the voltage across RS and convert it to current by dividing by the
4
resistance (DVM reading/100K9). Note that the current is less than 1.5
56
mA.
Chapter 3 Calibration Procedures

Performance Test Record for Agilent E3632A

Performance Test Record for Agilent E3632A

CV Performance Test Record

Test Description
CV Programming A ccuracy @ 0 volts
(DVM reading)
CV Readback Accuracy @ 0 volts DVM + 0.0050 V DVM - 0.0050 V
CV Programming Accuracy @ Fu ll Sc al e
(DVM reading)
CV Readback Accuracy @ Full Scale DVM + 0.0200 V DVM - 0.0200 V
CV Load Regulation Maximum change: <5 mV
CV Line Regulation Maximum change: <5 mV
CV Ripple/Noise <2 mV p-p, 0.35 mV rms
Load Transient Response Time <50
Actual Result
Upper Limit Lower Limit
+0.0100 V -0.0100 V
+30.025V 29.9750 V
Specifications
msec

CC Performance Test Record

Test Description
CC Programming A ccuracy @ 0 amps (IO) +0.0100 A -0.0100 A
CC Readback Accuracy @ 0 amps I
CC Programming Accuracy @ Full Scale (I
CC Readback Accuracy @ Full Scale I
CC Load Regulation Maximum change: <0.95 mA
CC Line Regulation Maximum change: <0.95 mA
CC Ripple/Noise <2 mA
Common Mode Current Noise <1.5
Actual Result
O
) 7.0240 A 6.9760 A
O
O
Specifications
Upper Limit Lower Limit
+ 0.0050 A IO - 0.0050 A
+ 0.0155 A IO - 0.0155 A
mA rms
3
57
Chapter 3 Calibration Procedures

Calibration Security Code

Calibration Security Code
This feature allows you to enter a security code (electronic key) to prevent accidental or unauthorized calibrations of the power supply. When you first receive your power supply, it is secured. Before you can calibrate the power supply, you must unsecure it by entering the correct security code. A procedure to unsecure the power supply is given on the following page.
The security code is set to "HP003632" when the power supply is shipped
from the factory. The security code is stored in non-volatile memory, and does not change when power has been off or after a remote interface reset.
• To secure the power supply from the remote interface, the security code may contain up to 12 alphanumeric characters as shown below. The first character must be a letter, but the remaining characters can be letters or numbers. You do not have to use all 12 characters but the first character must always be a letter.
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
• To secure the power supply from the remote interface so that it can be unsecured from the front panel, use the eight-character format shown below. The first two characters must be "H P" and the remaining characters must be numbers. Only the last six characters are recognized from the front panel, but all eight characters are required. To unsecure the power supply
from the front panel, omit the "H P" and enter the remaining numbers as shown on the following pages.
H P _ _ _ _ _ _ (8 characters)
If you forget your security code, you can disable the security feature by adding a jumper inside the power supply, and then entering a new code. See
the procedure on page 60.
58
Chapter 3 Calibration Procedures
Calibration Security Code

To Unsecure the Power Supply for Calibration

The power supply can use a calibration security code to prevent unauthorized or accidental calibration. This procedure shows you how to unsecure the power supply for calibration from the front panel.
Calibrate
Power
Secure
Turn on the front-panel calibration mode.
1
secured
Turn on the calibration mode by pressing the "Calibrate" key while simultaneously turning on the power supply then continue to hold the "Calibrate" key for about 5 seconds until a beep is heard.
If the power supply is secured, you will see the above message from the front panel for approximately one second. The "CAL MODE'' message is then displayed on the front panel.
Move to the security code by pressing the "Secure" key.
2
000000 code
Enter the security code using the knob and resolution selection keys.
3
003632 code
The security code is set to "Agilent Technologies003632" when the power supply is shipped from the factory. The security code is stored in non-volatile memory and does not change when the power has been off or after a remote interface reset.
To enter the security code from the front panel, enter only the last six digits. To enter the security code from the remote interface, you may enter up to 12 characters.
Use the resolution selection keys to move left or right between digits. Use the knob to change the digits. Notice that the security code may be different if the security code has been changed from the default setting.
3
59
Chapter 3 Calibration Procedures
Calibration Security Code
Secure
Power
Unsecure the power supply.
4
unsecured
The power supply is unsecured when you press the key. You will see the above message from the front panel for one second. The "CAL MODE" message is displayed on the front panel after above message.
Turn off the calibration mode.
5
Turn off the power supply to exit the calibration mode.
To re-secure the power supply (following calibration), perform this procedure again.
Secure

To Unsecure the Power Supply Without the Security Code

To unsecure the power supply without the correct security code (when you forget the security code), follow the steps below. See "Electrostatic Discharge (ESD) Precautions" in chapter 5 before beginning this procedure.
Disconnect the power cord and all load connections from front terminals.
1
Remove the instrument cover. Refer to the disassembly drawing on page 119.
2
Connect the power cord and turn on the calibration mode by pressing the
3
"Calibrate" key while simultaneously turning on the power supply then continue to hold the "Calibrate" key for about 5 seconds until a beep is heard. Be careful not to touch the power line connections.
Apply a short between the two exposed metal pads on JP5 (located near U13).
4
The JP5 is outlined with a circle on the component locator drawing on page 120.
While maintainin g the short, move to the security code and e nter any un secure
5
code in the calibration mode. The power supply is now unsecured.
Remove the short at JP5. (An error occurs if not removed.)
6
Turn off and reassemble the power supply.
7
Now you can enter a new security code. Be sure you take note of the new security code.
60
Chapter 3 Calibration Procedures

Calibration Count

Calibration Count
The calibration count feature provides an independent "serialization" of your calibrations. You can determine the number of times that your power supply has been calibrated. By monitoring the calibration count, you can determine whether an unauthorized calibration has been performed. Since the value increments by one for each calibration parameter (see Table 3-2 on the next page), a complete calibration increases the value by 5 counts.
• The calibration count is stored in non-volatile memory and does not change when power has been off or after a remote interface reset. Your power supply was calibrated before it left the factory. When you receive the power supply, read the calibration count to determine its value.
• The calibration count increments up to a maximum of 32,767 after which it wraps around to 0. No way is provided to program or reset the calibration count.

Calibration Message

3
You can use the calibration message feature to record calibration information about your power supply. For example, you can store such information as the last calibration date, the next calibration due date, the power supply's serial number, or even the name and phone number of the person to contact for a new calibration.
You can record and read information in the calibration message from the remote interface only.
• The calibration message may contain up to 40 characters.
• The calibration message is stored in non-volatile memory and does not change when power has been off or after a remote interface reset.
61
Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

General Calibration/Adjustment Procedure
The calibration procedures from the front panel are described in this section. For voltage calibration, disconnect all loads from the power supply and connect a DVM across the output terminals. For current calibration, disconnect all loads from the power supply, connect an appropriate current monitoring resistor (0.01 across the terminals of the monitoring resistor.
Note that the power supply should be calibrated after 1-hour warm-up with no load connected.
The following table shows calibration parameters and points which should be used to calibrate the output voltage and current.
Table 3-2. Parameters for Calibration
Calibration Parameter
CAL SETUP 1 Voltage
CAL SETUP 2 OVP None
CAL SETUP 3 Current
CAL SETUP 4 OCP None
9) across the output terminals, and connect a DVM
Voltage/
Current
Calibration Point
mnemonic
V LO
V MI
V HI
V LO
I MI
V HI
Note
Note
You can terminate any CAL SETUP without changing its calibration constants by turning off power.
Perform the voltage calibration prior to the OVP calibration and the current calibration prior to the OCP calibration.
62
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
To calibrate the output voltages and currents of the power supply from the front panel, proceed as follows:
Unsecure the power supply.
1
To calibrate the voltage and current, you must unsecure the power supply according to the procedure given on page 59.
Disconnect all loads from the power supply and connect a DVM across
2
output terminals.
Calibrate
Power
Calibrate
Calibrate
Turn on the calibration mode.
3
cal mode
Turn on the calibration mode by pressing the "Calibrate" key while simultaneously turning on the power supply then continue to hold the "Calibrate" key for about 5 seconds until a beep is heard. Make sure that the power supply is in "CV" mode. If the power supply is not in "CV" mode, an error occurs.

Voltage and OVP Calibration

Move down a level to the voltage calibration mode.
4
cal setup 1
The display shows the above message to indicate that the power supply is ready for the voltage calibration.
Calibrate DAC and select the low voltage calibration point.
5
30 left
3
The "START BITCAL" message is displayed for about 3 seconds to indicate that the power supply is ready for DAC calibration. Then it counts down numbers from 30 to 0.
63
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
v lo +0.5000 v
Then, the display shows the low voltage calibration point.
Read the DVM and change the low voltage value on the display to match
6
the measured voltage.
For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V using the knob and resolution selection keys.
v lo +0.4500 v
Calibrate
Calibrate
Pressing the "Calibrate" key saves the change and selects the middle
7
voltage calibration point.
v Mi +15.000v
If the entered number is within an acceptable range, an "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" "MAX VALUE" message appears for one second and the display shows the low voltage calibration point again. The display now shows the middle voltage calibration point.
Read the DVM and change the middle voltage value on the display to
8
match the measured voltage.
For example, if the DVM reads 14.995 V, adjust the voltage to 14.995 V using the knob and arrow keys.
v Mi +14.995v
Pressing the "Calibrate" key saves the change and selects the high
9
voltage calibration point.
v hi 29.500 v
64
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
If the entered number is within an acceptable range, a "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" or "MAX VALUE" message appears for one second and the display shows the middle voltage calibration point again. The display now shows the high voltage calibration point.
Read the DVM and change the first voltage value on the display to match
10
the measured voltage.
For example, if the DVM reads 28.995V, adjust the current to 28.995 V using the knob and arrow keys.
Calibrate
v hi 28.995 v
Pressing the "Calibrate" key saves the new voltage calibration
11
constants, and goes to the OVP calibration mode.
cal setup 2
A "CALIBRATING" message appears for one second to indicate that the voltage calibration is progressing and new voltage calibration constants of "SETUP 1" are stored. Then, the display shows above message to indicate that the power supply is ready for the OVP calibration.
If t he c alibration fails, a "DAC CA L FAIL" or " A DC CAL FA IL" message appea rs for one second and the display shows the "CAL SETUP 1" for voltage calibration again.

Current and OCP Calibration

Connect an appropriate shunt (0.019) across the output terminals, and connect a digital voltmeter across the shunt resistor for the current calibration.
3
65
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
Pressing the "Calibrate" key saves the new calibration constants for
12
OVP circuit and goes to the current calibration mode.
cal setup 3
A "CALIBRATING" message appears for about several seconds to indicate that the OVP calibration is progressing and new calibration constants of "SETUP 2" are stored. Then, the display shows the above message to indicate that the power supply is ready for the current calibration.
If the cal ibration fa ils, a "OVP CAL FAIL" message appears for one se cond and the display shows the "CAL SETUP 2" for OVP calibration again.
Calibrate
Calibrate
Select the low current calibration point.
13
i lo 0.200 a
The display shows the low current calibration point.
Read the DVM and change the low current value on the display to match
14
the computed current (DVM reading
For example, if the computed value is 0.199 A, adjust the current to 0.199 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
¸ by shunt resistance).
i lo +0.199 a
Pressing the "Calibrate" key saves the change and selects the middle
15
current calibration point.
i mi 3.500 a
If the entered number is within an acceptable range, an "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" or "MAX VALUE" message appears for one second and the display shows the low current calibration point again. The display now shows the middle current calibration point.
66
Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
Read the DVM and change the middle current value on the display to
16
match the computer current (DVM reading
For example, if the computered value is 3.499 A, adjust the current to 3.499 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
¸ by shunt resistance).
i mi 3.499 a
Calibrate
Pressing the "Calibrate'' key saves the change and selects the high
17
current calibration point.
i hi 6.900 a
If the entered number is within an acceptable range, an "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" or "MAX VALUE" message appears for one second and the display shows the m id dl e c u rr e nt ca li b ra ti on p oi nt ag a in . T he di sp l ay no w s h ow s t he h ig h c ur r en t calibration point
Read the DVM and change the high current value on the display to
18
match the computed current (DVM reading
For example, if the computed value is 6.899 A, adjust the current to 6.899 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
¸ by shunt resistance).
i hi 6.899 a
3
67
Chapter 3 Calibration Procedures

Aborting a Calibration in Progress

Calibrate
Calibrate
Power
Pressing the "Calibrate" key saves the new calibration constants for
19
the output current and goes to the OCP calibration mode.
cal setup 4
A "CALIBRATING" message appears for one second to indicate that the current calibration is progressing and new calibration constants of "SETUP 3" are stored. Then, the display shows the above message to indicate that the power supply is ready for the OCP calibration.
If the calibration fails, an "DAC CAL FAIL" or "ADC CAL FAIL" message appears for one second and the display shows the "CAL SETUP 3" for current calibration again.
Pressing the "Calibrate" key saves the new OCP calibration constants
20
and return to the calibration mode.
cal mode
A "CALIBRATING" message appears for several seconds to indicate that the OCP calibration is progressing and new OCP calibration constants of "SETUP 4" are stored. Then the display will return to the calibration mode.
Turn off the power supply to exit the calibration mode.
21
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has already been initiated. You can abort a calibration at any time by turning the power supply off from the front panel. When performing a calibration from the remote interface, you can abort a calibration by issuing a remote interface device clear message or by pressing the front-panel "Local" key.
68
Chapter 3 Calibration Procedures

Calibration Record for Agilent E3632A

Calibration Record for Agilent E3632A
Step Calibration Description
1 Unsecure the power supply (see page 59).
2
3 Move down menu to "CAL SETUP 1" (press "Calibrate" key).
4
5
6
7 "CAL SETUP 2" now appears on the display (press "Calibrate" key). V OVP calibration
8
9
10
11
12 "CAL SETUP 4" now appears on the display (press "Calibrate" key). A OCP calibration
13 Press "Calibrate" key, then press "Power" switch. Exit CAL MODE
Turn on "CAL MODE"(simultaneously press "Calibrate" and "Power" keys) until it beeps.
Calibrate DAC and select the low point for voltage calibration; "START BITCAL" appears for 3 seconds and the display counts down numbers from 30 to 0. Then, " V LO 0.5000 V" appears on the display(press "Calibrate" key and wait about 30 seconds; then change the display to match the DVM reading).
"V MI 15.000 V" appears on the display (press "Calibrate" key; then change the display to match the DVM reading).
"V HI 29.500 V" appears on the display (press "Calibrate" key; then change the display to match the DVM reading).
"CAL SETUP 3" now appears on the display (press "Calibrate" key; then connect 0.019resistor across the output terminals).
"I LO 0.200 A" appears on the display (press "Calibrate" key; then change the display to match the computed current through 0.019resistor).
"I MI 3.500 A" appears on the display (press "Calibrate" key; then change the display to match the computed current through 0.019resistor).
"I HI 6.9000 A" appears on the display (press "Calibrate" key; then change the display to match the computed current through 0.019resistor).
Measurement
Mode (DVM)
Supply being
Adjusted
Voltage Calibration
DAC and low
V
voltage point calibration
Middle voltage
V
point calibration
High voltage point
V
calibration
Current calibration
Low current point
A
calibration
Middle current
A
point calibration
High current point
A
calibration
3
69
Chapter 3 Calibration Procedures

Error Messages

Error Messages
The following tables are abbreviated lists of error messages for the E3632A. The errors listed are the most likely errors to be encountered during calibration and adjustment. A more complete list of error messages and descriptions is contained in chapter 5 of the E3632A User's Guide.
System Error Messages
Error Error Message
-330 Self-test failed
-350 Too many errors 501 Isolator UART framing error 502 Isolator UART overrun error 511 RS-232 framing error 512 RS-232 overrun error 513 RS-232 parity error 514 Command allowed only with RS-232 521 Input buffer overflow 522 Output buffer overflow 550 Command not allowed in local
Self-Test Error Messages
Error Error Message
601 Front panel does not respond 602 RAM read/write failed 603 A/D sync stuck 604 A/D slope convergence failed 605 Cannot calibrate rundown gain 606 Rundown gain out of range 607 Rundown too noisy 608 Serial configuration readback failed 624 Unable to sense line frequency 625 I/O processor does not respond 626 I/O processor failed self-test 630 Fan test failed 631 System DAC test failed 632 Hardware test failed
70
Chapter 3 Calibration Procedures
Error Messages
Calibration Error Messages
Error Error Message
701 Cal security disabled by jumper 702 Cal secured 703 Invalid secure code 704 Secure code too long 705 Cal aborted 708 Cal output disabled 712 Bad DAC cal data 713 Bad readback cal data 714 Bad OVP cal data 715 Bad OCP cal data 716 Bad OVP DNL error correction data 717 Cal OVP or OCP status enabled 740 Cal checksum failed, secure state 741 Cal checksum failed, string data 742 Cal checksum failed, store/recall data in location 0 743 Cal checksum failed, store/recall data in location 1 744 Cal checksum failed, store/recall data in location 2 745 Cal checksum failed, store/recall data in location 3 746 Cal checksum failed, DAC cal constants 747 Cal checksum failed, readback cal constants 748 Cal checksum failed, GPIB address 749 Cal checksum failed, internal data 750 Cal checksum failed, DAC DNL error correction data
3
71
Chapter 3 Calibration Procedures

Calibration Program

Calibration Program
This section contains an Agilent BASIC program for calibration over the GPIB interface. This program makes software adjustments to the E3632A power supply using a current shunt and a digital mutimeter which is connected to the controller. In this program a 0.01 ohm current shunt is used. Be sure to change the value of the variable "Current_shunt" to the value of the current shunt used and the GPIB address for the power supply and the digital voltmeter.
10 ! 20 ! This program was written on a PC with Agilent Basic for Windows. 30 ! It will make software adjustments to the E3632A Power Supply 40 ! on the GPIB bus using a Agilent 34401A Digital Multimeter and a 50 ! current shunt. In the program a 0.01 ohm current shunt is 60 ! used to measure current. Be sure to change the value of 70 ! the variable 'Current_shunt' to the value of the current 80 ! shunt used. 90 ! 100 CLEAR SCREEN 110 DIM Cal_msg$[40],Error$[40],Sec_code$[10] 120 REAL Dmm_rdg,Current_shunt 130 Current_shunt=.01 ! Current Shunt value in Ohms 140 Sec_code$="HP003632" ! Assign the security code 150 ASSIGN @Dmm TO 722 ! Assign address 22 to the Dmm 160 ASSIGN @Pwrsupply TO 705 ! Assign address 5 to the Power Supply 170 CLEAR 7 ! Clear GPIB, Dmm and Power Supply 180 OUTPUT @Pwrsupply;"*CLS" ! Clear Power Supply errors 190 OUTPUT @Dmm;"*RST" ! Reset Dmm 200 OUTPUT @Pwrsupply;"*RST" ! Reset Power Supply 210 OUTPUT @Pwrsupply;"CAL:STR?" ! Read the calibration message 220 ENTER @Pwrsupply;Cal_msg$ 230 PRINT TABXY(5,2),"Calibration message of Power Supply is: ";Cal_msg$ 240 ! 250 ! Set the Calibration security to off, and check to be sure 260 ! it is off. If not successful, print message to screen and end. 270 ! 280 OUTPUT @Pwrsupply;"VOLT:PROT:STAT OFF" 290 OUTPUT @Pwrsupply;"CURR:PROT:STAT OFF" 300 OUTPUT @Pwrsupply;"CAL:SEC:STAT OFF, ";Sec_code$ 310 OUTPUT @Pwrsupply;"CAL:SEC:STAT?" 320 ENTER @Pwrsupply;A 330 IF A=1 THEN 340 PRINT TABXY(5,5),"****** Unable to Unsecure the Power supply ******" 350 GOTO 2290 360 END IF
72
Chapter 3 Calibration Procedures
Calibration Program
continued
370 ! 380 ! Perform the DAC error correction, voltage calibration and OVP calibration. 390 ! Alert the operator to hook up the connection before calibrating. 400 ! 410 ! Alert operator to connect lead 420 PRINT TABXY(10,10),"*********************************************************" 430 PRINT TABXY(10,11)," Prepare for E3632A DAC DNL error correction and" 440 PRINT TABXY(10,12)," Voltage/OVP calibration. Connect the output to the DMM." 450 PRINT TABXY(10,13)," Observe Polarity!" 460 PRINT TABXY(10,14),"*********************************************************" 470 PRINT TABXY(10,16),"Press 'C' to Continue, 'I' to go to CURRENT calibration or" 480 PRINT TABXY(10,17),"'X' to eXit, then press 'Enter'" 490 Ch$="C" 500 INPUT Ch$ 510 IF Ch$="X" OR Ch$="x" THEN GOTO 2250 520 IF Ch$="I" OR Ch$="i" THEN 530 CLEAR SCREEN 540 GOTO 1460 550 END IF 560 CLEAR SCREEN 570 PRINT TABXY(10,7),"BEGIN DAC ERROR CORRECTION" 580 WAIT 4 590 CLEAR SCREEN 600 OUTPUT @Pwrsupply;"OUTP ON" ! Turn on Power Supply output 610 OUTPUT @Pwrsupply;"CAL:DAC:ERROR" ! Perform DAC DNL error correction 620 WAIT 29 ! Allow DAC error correction to finish 630 OUTPUT @Pwrsupply;"OUTPUT OFF" ! Turn off Power Supply output 640 OUTPUT @Pwrsupply;"SYST:ERR?" 650 ENTER @Pwrsupply;Error$ 660 ! 670 ! Check to see if there is an error. If there is an error, 680 ! display the error and exit the program. 690 ! 700 CLEAR SCREEN 710 IF Error$="+0,""No error""" THEN 720 PRINT "DAC DNL Error Correction completed for Power Supply " 730 ELSE 740 PRINT "ERROR:";Error$;"DAC DNL Error not corrected " 750 BEEP 760 GOTO 2250 770 END IF 780 PRINT TABXY(10,5),"DAC DNL ERROR CORRECTION COMPLETE" 790 PRINT TABXY(10,7),"BEGIN VOLTAGE CALIBRATION" 800 WAIT 4 810 OUTPUT @Pwrsupply;"OUTPUT ON" 820 CLEAR SCREEN 830 OUTPUT @Pwrsupply;"CAL:VOLT:LEV MIN" ! set output to minimum cal value 840 WAIT 2 ! allow output to settle 850 OUTPUT @Dmm;"MEAS:VOLT:DC?" ! measure output with Dmm and 860 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 870 PRINT Dmm_rdg
3
73
Chapter 3 Calibration Procedures
Calibration Program
continued
880 OUTPUT @Pwrsupply;"CAL:VOLT:DATA ";Dmm_rdg ! send stored value to Power Supply 890 OUTPUT @Pwrsupply;"CAL:VOLT:LEV MID" ! set output to middle cal value 900 WAIT 2 ! allow output to settle 910 OUTPUT @Dmm;"MEAS:VOLT:DC?" ! measure output with Dmm and 920 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 930 PRINT Dmm_rdg 940 OUTPUT @Pwrsupply;"CAL:VOLT:DATA ";Dmm_rdg ! send stored value to Power Supply 950 OUTPUT @Pwrsupply;"CAL:VOLT:LEV MAX" ! set output to maximum cal value 960 WAIT 2 ! allow output to settle 970 OUTPUT @Dmm;"MEAS:VOLT:DC?" ! measure output with Dmm and 980 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 990 PRINT Dmm_rdg 1000 OUTPUT @Pwrsupply;"CAL:VOLT:DATA ";Dmm_rdg ! send stored value to Power Supply 1010 OUTPUT @Pwrsupply;"OUTP OFF" 1020 OUTPUT @Pwrsupply;"SYST:ERR?" 1030 ENTER @Pwrsupply;Error$ 1040 ! 1050 ! Check to see if there is an error. If there is an error, 1060 ! display the error and exit the program. 1070 ! 1080 CLEAR SCREEN 1090 IF Error$="+0,""No error""" THEN 1100 PRINT "Voltage calibration completed for Power Supply " 1110 ELSE 1120 PRINT "ERROR:";Error$;"Voltage not Calibrated" 1130 BEEP 1140 GOTO 2250 1150 END IF 1160 PRINT TABXY(10,5),"VOLTAGE CALIBRATION COMPLETE" 1170 PRINT TABXY(10,7),"BEGIN OVP CALIBRATION" 1180 WAIT 4 1190 CLEAR SCREEN 1200 OUTPUT @Pwrsupply;"OUTP ON" ! Turn on Power Supply output 1210 OUTPUT @Pwrsupply;"CAL:VOLT:PROT" ! Perform OVP circuit calibration 1220 WAIT 9 ! Allow OVP calibration to finish 1230 OUTPUT @Pwrsupply;"OUTP OFF" ! Turn off Power Supply output 1240 OUTPUT @Pwrsupply;"SYST:ERR?" 1250 ENTER @Pwrsupply;Error$ 1260 ! 1270 ! Check to see if there is an error. If there is an error, 1280 ! display the error and exit the program. 1290 ! 1300 CLEAR SCREEN 1310 IF Error$="+0,""No error""" THEN 1320 PRINT "OVP calibration completed for Power Supply " 1330 ELSE 1340 PRINT "ERROR:";Error$;"OVP not Calibrated" 1350 BEEP 1360 GOTO 2250 1370 END IF
74
Chapter 3 Calibration Procedures
Calibration Program
continued
1380 CLEAR SCREEN 1390 PRINT TABXY(10,5),"DAC ERROR CORRECTION AND VOLTAGE/OVP CALIBRATION COMPLETE" 1400 WAIT 4 1410 ! 1420 ! Perform the Current calibration and OCP calibration. Alert the operator to 1430 ! hook up the connection before calibrating. 1440 ! 1450 ! Alert operator to connect lead 1460 PRINT TABXY(10,10),"*********************************************************" 1470 PRINT TABXY(10,11)," Connect a CURRENT SHUNT to the Dmm input for measuring" 1480 PRINT TABXY(10,12)," current. Connect the output to the shunt. Observe Polarity!" 1490 PRINT TABXY(10,13),"*********************************************************" 1500 PRINT TABXY(10,15),"Press 'C' to Continue, or 'X' to eXit, then 'Enter':" 1510 Ch$="C" 1520 INPUT Ch$ 1530 IF Ch$="X" OR Ch$="x" THEN GOTO 2250 1540 OUTPUT @Pwrsupply;"OUTP ON" ! Turn on Power Supply output 1550 CLEAR SCREEN 1560 PRINT TABXY(10,7),"BEGIN CURRENT/OCP CALIBRATION" 1570 WAIT 4 1580 CLEAR SCREEN 1590 OUTPUT @Pwrsupply;"CAL:CURR:LEVel MIN" ! set output to minimum cal value 1600 WAIT 2 ! allow output to settle 1610 OUTPUT @Dmm;"MEAS:VOLT:DC? " ! measure output with Dmm and 1620 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 1630 Dmm_rdg=Dmm_rdg/Current_shunt ! scale reading to amps 1640 PRINT Dmm_rdg 1650 OUTPUT @Pwrsupply;"CAL:CURR:DATA ";Dmm_rdg ! send stored value to Power Supply 1660 OUTPUT @Pwrsupply;"CAL:CURR:LEVel MID" ! set output to middle cal value 1670 WAIT 2 ! allow output to settle 1680 OUTPUT @Dmm;"MEAS:VOLT:DC? " ! measure output with Dmm and 1690 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 1700 Dmm_rdg=Dmm_rdg/Current_shunt ! scale reading to amps 1710 PRINT Dmm_rdg 1720 OUTPUT @Pwrsupply;"CAL:CURR:DATA ";Dmm_rdg ! send stored value to Power Supply 1730 OUTPUT @Pwrsupply;"CAL:CURR:LEVel MAX" ! set output to maximum cal value 1740 WAIT 2 ! allow output to settle 1750 OUTPUT @Dmm;"MEAS:VOLT:DC?" ! measure output with Dmm and 1760 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg 1770 Dmm_rdg=Dmm_rdg/Current_shunt ! scale reading to amps 1780 PRINT Dmm_rdg 1790 OUTPUT @Pwrsupply;"CAL:CURR:DATA ";Dmm_rdg ! send stored value to Power Supply 1800 OUTPUT @Pwrsupply;"OUTP OFF" ! Turn off Power Supply output 1810 OUTPUT @Pwrsupply;"SYST:ERR?" 1820 ENTER @Pwrsupply;Error$ 1830 ! 1840 ! Check to see if there is an error. If there is an error, 1850 ! display the error and exit the program.
3
75
Chapter 3 Calibration Procedures
Calibration Program
continued
1860 ! 1870 CLEAR SCREEN 1880 IF Error$="+0,""No error""" THEN 1890 PRINT "Current calibration completed for Power Supply " 1900 ELSE 1910 PRINT "ERROR:";Error$;"Current not Calibrated" 1920 BEEP 1930 GOTO 2250 1940 END IF 1950 CLEAR SCREEN 1960 PRINT TABXY(10,5),"CURRENT CALIBRATION COMPLETE" 1970 PRINT TABXY(10,7),"BEGIN OCP CALIBRATION" 1980 WAIT 4 1990 CLEAR SCREEN 2000 OUTPUT @Pwrsupply;"OUTP ON" ! Turn on Power Supply output 2010 OUTPUT @Pwrsupply;"CAL:CURR:PROT" ! Perform OCP calibration 2020 WAIT 9 ! Allow OCP calibration to finish 2030 OUTPUT @Pwrsupply;"OUTP OFF" ! Turn off Power Supply output 2040 OUTPUT @Pwrsupply;"SYST:ERR?" 2050 ENTER @Pwrsupply;Error$ 2060 ! 2070 ! Check to see if there is an error. If there is an error, 2080 ! display the error and exit the program. 2090 ! 2100 CLEAR SCREEN 2110 IF Error$="+0,""No error""" THEN 2120 PRINT "OCP calibration completed for Power Supply " 2130 ELSE 2140 PRINT "ERROR:";Error$;"OCP not Calibrated" 2150 BEEP 2160 GOTO 2250 2170 END IF 2180 CLEAR SCREEN 2190 PRINT TABXY(10,5),"CURRENT/OCP CALIBRATION COMPLETE" 2200 ! 2210 ! Create a time stamp and output to power supply 2220 ! 2230 Cal_msg$="Last Calibrated "&DATE$(TIMEDATE)&" "&TIME$(TIMEDATE) 2240 OUTPUT @Pwrsupply;"CAL:STR """;Cal_msg$;"""" 2250 OUTPUT @Pwrsupply;"CAL:SEC:STAT ON, ";Sec_code$ 2260 OUTPUT @Pwrsupply;"VOLT:PROT:STAT ON" 2270 OUTPUT @Pwrsupply;"CURR:PROT:STAT ON" 2280 DISP "Calibration terminated. " 2290 END
End of Program
76
4
4
Theory of Operation

Theory of Operation

This chapter provides block diagram level descriptions of the power supply. The descriptions provide a basic understanding of circuit operation and are intended as an aid in troubleshooting. It is assumed in the following discussions that you are familiar with the operating and programming instructions presented in the E3632A User's Guide. Subjects covered include the following:
• Block Diagram Overview, page 79
• AC Input and Bias Supplies, page 81
• Floating Logic, page 82
• D-to-A Converter, page 84
• A-to-D Converter, page 85
• Power Mesh and Control, page 86
• Earth-Referenced Logic, page 88
• Front Panel, page 88
78
Chapter 4 Theory of Operation

Block Diagram Overview

Block Diagram Overview
This discussion pertains to the block diagram on the next page. The power supply's circuitry is divided into two major blocks: the floating circuitry and the ground referenced circuitry. All power mesh and control circuits, display circuit, and digital circuits are contained in the floating circuitry. This circuitry also contains the power supply's main controller. The earth referenced circuitry provides the interface between the user and the power supply.
The floating circuitry can be viewed in four pieces; the DAC system, the digital logic section (floating logic), the power mesh and control section, and the front panel (display and keyboard) section.
The floating logic receives digital signals from the earth-referenced logic and the DAC converts them to analog signals which are sent to the power control circuits in order to program the power supply's output voltage and current. The power supply can also be commanded to send measurement and status data back to the remote interface controller and/or the VFD (vacuum fluorescent display) display on the front panel. The data is processed and sent back via the floating logic and earth-referenced logic.
The power mesh and control circuits contains voltage and current control circuits which allows the power supply to operate in either the constant voltage (CV) or constant current (CC) mode. The control circuits compare the power supply's output voltage or current with the programmed value and generates a control signal which varies the conduction of the series pass transistor to raise or lower the output as required.
The front panel circuits consist of VFD control, display high voltage drivers, and keyboard scanning. Communication between the front panel and floating logic circuits is accomplished through a 4-wire bi-directional serial interface.
The earth referenced circuitry uses a controller configured as a slave to the main controller. This controller establishes external I/O communication with the main controller through a bi-directional, optically isolated, serial communications link. The earth referenced controller controls low-level GPIB (IEEE-488) and RS-232 interface operation.
Separate reference and bias supplies are provided for the floating and ground reference circuitry. The front panel operates from the floating circuitry with its logic common different from the main controller logic common.
4
79
Chapter 4 Theory of Operation
Block Diagram Overview
Block Diagram
80
Chapter 4 Theory of Operation

AC Input and Bias Supplies

AC Input and Bias Supplies
Referring to the schematic shown on page 121 and 122, the ac mains are connected by a fused power module. This module incorporates the functions of mains connection, fusing, and line voltage selection (100/115/230 Vac). The line voltage selection function of the module selects which primary winding of power transformer is energized. The transformer secondary windings are connected to the main pc board through connectors.
The bias supplies are consists of four sections; +5 Vdc and -5.1 Vdc for the power circuits and floating logic; earth referenced logic. Power-on reset signals are provided by the +5 Vdc supply of the floating logic.
±17.4 Vdc for the display circuits are produced by rectifier CR4, filter
The capacitors C19 and C24, and voltage regulators U12 and U16. A separate winding of transformer provides a center tapped 6 Vrms filament supply for the display.
The floating +5 Vdc is produced from the separate winding of transformer. The +5_REF signal is derived from +15 Vdc supply and the TURN_ON RESET signal is derived from +5 Vdc supply. The FAN FAIL signal is asserted when the fan current through R15 is not detected. The TURN_ON RESET signal holds the main controller and other logic in a reset state until the +5 Vdc logic power is fully operational. This signal is generally active only following application of line power to the instrument.
The +5V dc earth referenced supply is produced by rectifier CR7, filter capacitor C21, and regulator U10. The GPIB (IEEE-488) and RS-232 computer interfaces are powered from this supply.
±17.4 Vdc for the display; and +5 Vdc for the
4
81
Chapter 4 Theory of Operation

Floating Logic

Floating Logic
Referring to the schematic shown on page 124, the floating common logic controls operation of the entire instrument. All output functions and bus command interpretation is performed in the main controller U17. The front panel and the earth referenced logic operate as slaves to U17. The floating common logic is comprised of the main controller U17, custom gate array U16, the program ROM U13, RAM U14, calibration EEPROM U15, and the 12 MHz clock oscillator. Non-volatile EEPROM U15 stores calibration constants, calibration secure code, and calibration count, and store/recall variables. Power-on reset is provided to the main controller by the voltage regulator U11.
The main controller U17 is a 16-bit micro controller. It controls such features as receive and transmit serial port, timer/counter ports, an 8-bit pulse width modulated DAC port, and selectable input 10-bit successive approximation a­to-d convert ports. A conventional address/data bus is used to transfer data between the main controller and external ROM and RAM. When the address latch enable (ALE) signal goes high, address data is present on the address/ data bus. ASIC U18 latches the address data and decodes the correct chip enable (low true) for external ROM and RAM accesses and for read/write accesses to the internal registers of U18. The system memory map is shown below.
0000H - 1FF7HU15 32k x 8 RAM 1FF8
- 1FFFHU16 Gate Array
H
- FFFFHU14 Program ROM
2000
H
Program ROM U13 contains four 64k x 8 data banks of data. Banks are selected by controlling A16 and A17 ROM address bits directly from the main controller port bits.
Custom gate array U18 performs address latching and memory map decoding functions as discussed above. In addition, U18 contains a variety of internal read/write registers. The read (XRD) and write (XWR) signals transfer data out of and into U18 when it is addressed. There are four internal registers in U18: an internal configuration register, an 8 bit counter register, a serial transmit/ receive register, and an internal status register.
Th e c ou nt er r eg is te r is u se d t o ca pt ur e t he A DC slo pe co un t at th e C OM P i np ut . The COMP input functions as both a clocked comparator and the slope counter input for the ADC. In both cases the counter register captures the lower 8 bits of a 24-bit counter. The upper 16 bits of the count are captured by the SYNC input to U17.
82
Chapter 4 Theory of Operation
Floating Logic
The serial register is used to send and receive serial data bytes from the main controller to the DAC system, or to communicate with the front panel controller. The serial register is multiplexed to these two circuits. The transmission rate is selected to 1.5 M bits/second for the DAC system and 93.75 k bits/second for communication with the front panel controller. The general serial interface is a 3-bit interface as shown below.
U18 Internal Signal Configuration Signals Front Panel Signals
Serial Clock Data OUT (send) Data IN (receive)
SERCK SERDAT SERRBK
XFPSK
FPDI
FPDO
Serial data is received simultaneously as serial data is clocked out. Front panel data is exchanged in both directions whenever a byte is sent from U18. The input data of DAC is strobed to outputs by U17 signal SERSTB. Interrupts from the front panel are detected by U18 and signaled to the CHINT. The main controller FPINT signals the front panel controller that U18 has data to send.
The power supply's calibration data are stored in a 256 x 16 bit non-volatile electrically erasable ROM U15. This non-volatile ROM read/write data is accessed by a 4-bit serial protocol controlled by U17.
The main controller has an on-chip 10-bit successive approximation ADC. The FLASH input is used to sample the residual charge on the main integrating ADC output of U26.
Port bits are also configured to measure the input power line frequency (LSENSE). Frequencies from 55 Hz to 66 Hz are measured as 60 Hz. All other line input frequencies are assumed to be 50 Hz.
The main controller communicates with the earth referenced controller U4 through an optically isolated (U6 and U7) asynchronous serial link. Data is sent in an 11-bit frame at a rate of 187.5 k bits/ second. When the RS-232 interface is selected, data is sent across the isolated link at 93.75 k bits/second. The 11­bit data frame is configured for one start bit, eight data bits, one control bit, and one stop bit.
4
83
Chapter 4 Theory of Operation

D-to-A Converter

D-to-A Converter
Referring to the schematic shown on page 123, all reference voltages of power circuits are derived from the internal voltage reference of system DAC U21. The system DAC track/hold amplifier outputs are used to provide controllable reference voltages to three power circuits. The system DAC is programmed and responds to the main controller via internal 3-wire serial data bus SERCLK, SERRBK, and SERSTB. The system DAC is mutiplexed to 6 track/hold amplifiers through U27. Each track/hold amplifier is refreshed approximately every 1 msec to maintain its output setting. Changes to track/hold amplifier outputs are accomplished by dwelling on that position for an extended period.
84
Chapter 4 Theory of Operation

A-to-D Converter

A-to-D Converter
Referring to the schematic shown on page 123, the analog-to-digital converter (ADC) is used to change dc voltages into digital information. The circuitry consists of an integrator amplifier (U26 and U29), current steering switch U33, resistors (R70, R71, and R96), voltage reference U32, ADC controller U18, and residue ADC U17.
The ADC method used by the Agilent E3632A is called multislope III. Multislope III is a charge balancing continuously integrating analog-to-digital converter. The input voltage continuously forces charge onto the integrator capacitors C49 and C51 through R71.
Switch U33 steers fixed positive or negative reference currents onto the integrator capacitors to cancel, or balance the accumulated input charge. The level shifted (R97 and R98) output of the integrator is checked every 2.66 by the U18 COMP input. Logic state machines in U18 control the U33 current steering to continuously seek an approximate 2.5 V level on the integrator amplifier output, FLASH. If the ADC input voltage is between integrator output (FLASH) will remain within the 0 V to 5 V range of the U17 on-chip ADC. The U17 ADC input (FLASH) is clamped to 0 V or 5 V by R53 and CR16 to protect U17.
The integrator amplifier is formed by U26 and U29. Resistors R61 and R62 affect the amplifier stability. Amplifier oscillation may occur if their values are incorrect. Amplifier U29 improves the offset voltage characteristics of integrator amplifier U26.
Each analog-to-digital conversion occurs continuously. The ADC starts by clearing the integrator slope count in U18. At the end of the integration period, the slope count is latched. The slope count provides the most significant bits of the input volta ge conversion. The lea st significant bits are converte d by the on-chip ADC of U17.
U39 provides a stable +5 V reference voltage for ADC. U32A amplifies the voltage reference to +10 V while amplifier U32B inverts the +10 V reference to
-10 V. The reference voltage forces precision slope currents for the integrating ADC through R70 and R96.
15 V, the
sec
4
85
Chapter 4 Theory of Operation

Power Mesh and Control

Power Mesh and Control
Refer to the schematics shown on page 121.
For the power mesh and control circuit, a preregulator is added ahead of the series pass transistor to minimize the power dissipated in the series pass transistor by controlling the dc level across the input filter capacitor, depending on the output voltage.
To achieve this, tap switching is accomplished by a SCR and one bridge diode and the SCR control circuit in each power circuit; CR14, CR17, CR18, CR19, and CR11. By turning on or off the SCR, these circuits allow the input capacitors (C36, C65, and C46) to charge to one of two discrete voltage levels, depending on the output voltage required. When the SCR is not fired, the bridge diode conducts and the low voltage of two discrete voltage levels is developed across the input filter capacitors.
The SCR control circuit determines whether SCR is to be fired by monitoring the output voltage and comparing this value against internally derived reference levels.
The series pass transistor is part of a feedback loop which consists of the driver and the Constant Voltage/Constant Current error amplifier. The feedback loop provides "fine and fast" regulation of the output while the preregulator feedback loop handles large, relatively slow, regulation demands.
The series pass transistor is made to alter its conduction to maintain a constant output voltage or current. The voltage developed across the current sampling resistors is the input to the constant current error amplifier. The constant voltage error amplifier obtains its input from differential amplifier which senses the output voltage. Any changes in output voltage or current are detected and amplified by the constant voltage or constant current error circuit and applied to the series pass transistor in the correct phase and amplitude to counteract the change in output voltage or current.
86
Chapter 4 Theory of Operation
Power Mesh and Control
Two error amplifiers are included in a CV/CC supply, one for controlling output voltage, the other for controlling output current. Since the constant voltage amplifier tends to achieve zero output impedance and alters the output current whenever the load resistance changes, while the constant current amplifier causes the output impedance to be infinite and changes the output voltage in response to any load resistance change, it is obvious that the two amplifiers can not operate simultaneously. For any given value of load resistance, the power supply must act either as a constant voltage source or as a constant current source - it can not be both; transfer between these two modes is accomplished at a value of load resistance equal to the ratio of the output voltage control setting to the output current control setting.
Full protection against any overload condition is inherent in the Constant Voltage/Constant Current design principle since there is not any load condition that can cause an output which lies outside the operating region. For either constant voltage or constant current operation, the proper choice of front panel voltage and current limit settings insures optimum protection for the load device as well as full protection for the power supply.
The diodes connected across the output terminals in reverse polarity protect the output electrolytic capacitor and the series pass transistors from the effects of a reverse voltage applied across the output terminals.
4
87
Chapter 4 Theory of Operation

Earth-Referenced Logic

Earth-Referenced Logic
Referring to the schematic shown on page 125, the earth referenced logic circuits schematic provides all rear panel input/output capability. Microprocessor U4 handles GPIB (IEEE-488) control through bus interface chip U3 and bus receiver/driver chips U1 and U2. The RS-232 interface is also controlled through microprocessor U4. RS-232 transceiver chip U17 provides the required level shifting to approximate charge-pump power supplies using C5 and C15. Communication between the earth referenced logic interface circuits and the floating logic is accomplished through an optically-isolated bi-directional serial interface. Isolator U7 couples data from U4 to processor U17. Isolator U6 couples data from U17 to microprocessor U4.
±9 volt logic levels through on-chip

Front Panel

Re fer rin g to the sche mati c shown on p a ge 127, th e front p a nel circu its consi st of vacuum fluorescent display control, display high voltage drivers, and keyboard scanning. Communication between the front panel and floating logic circuits is accomplished through a 4-wire bi-directional serial interface. The main controller U17 can cause a hardware reset to front-panel controller by signal IGFPRES. The front panel logic operates from -12.4 volts (logic 1) and -
17.4 volts (logic 0). The front panel logic high supply (-12.4 volts) is produced by the -17.4 volts bias supply and the voltage regulator U2 on the front panel board. The four serial communication signals are level shifted by the comparator U8 from the floating logic 0 V to 5 V levels to the -17.4 V to -12.4 V levels present on the front panel assembly. U6 acts as the serial shift register interface for the front-panel controller U7 on the front panel board.
Display anode and grid voltages are +17.4 volts for an "on" segment and -17.4 volts for an "off" segment. The -11.2 V cathode bias for the display is provided by filament winding center tap bias circuit VR1, R18, and C25 on the main board. Keyboard scanning is accomplished through a conventional scanned row-column key matrix. Keys are scanned by outputing data at front-panel controller U7 port pins P0.0 through P0.3 to poll each key column for a key press. Column read-back data are read by the microprocessor at port pins P1.0 through P1.3 for decoding and communication to the floating logic circuits.
88
5
Service
5

Service

This chapter discusses the procedures involved for returning a failed power supply to Agilent Technologies for service or repair. Subjects covered include the following:
• Operating Checklist, page 91
• Types of Service Available, page 92
• Repacking for Shipment, page 93
• Electrostatic Discharge (ESD) Precautions, page 94
• Surface Mount Repair, page 94
• To Replace the Power-Line Fuse, page 94
• To Disconnect the Output Using an External Relay, page 95
• Troubleshooting Hints, page 96
• Self-Test Procedures, page 98
90
Chapter 5 Service

Operating Checklist

Operating Checklist
Before returning your power supply to Agilent Technologies for service or repair check the following items:

Is the Power Supply Inoperative?

Verify that the ac power cord is connected to the power supply.
Verify that the front-panel power switch is depressed.
Verify that the power-line fuse is installed:
Use the 4 AT, 250 V fuse for 100 or 115 Vac operation. Use the 2.5 AT, 250 V fuse for 230 Vac operation.
Verify the power-line voltage setting.
See "To prepare the power supply for use" on page 23.

Does the Power Supply Fail Self-Test?

Verify that the correct power-line voltage is selected.
See "To prepare the power supply for use" on page 23.
Remove all load connections to the power supply.
Ensure that all terminal connections are removed while the self-test is performed.
5
91
Chapter 5 Service

Types of Service Available

Types of Service Available
If your power supply fails within three years of original purchase, Agilent Technologies will repair or replace it free of charge. If your unit fails after your three year warranty expires, Agilent will repair or replace it as a very competitive price. Agilent will make the decision locally whether to repair or replace your unit.

Standard Repair Service (worldwide)

Contact your nearest Agilent Technologies Service Center. They will arrange to have your power supply repaired or replaced.

Express Exchange (U.S.A. only)

You can receive a replacement Agilent E3632A via overnight shipment for low downtime.
Call 1-800-258-5165 and ask for "Express Exchange."
1
• You will be asked for your shipping address and a credit card number to guarantee return of your failed power supply.
• If you do not return your failed power supply within 45 days, your credit card will be billed for a new Agilent E3632A.
• If you choose not to supply a credit card number, you will be asked to send your failed unit to a designated Agilent Service Center. After the failed unit is received, Agilent will send your replacement unit.
Agilent will immediately send a replacement Agilent E3632A to you via
2
overnight shipment.
• The replacement unit will have a different serial number than your failed unit.
• If you can not accept a new serial number for the replacement unit, use the Standard Repair Service option described above.
92
Chapter 5 Service

Repacking for Shipment

• If your failed unit was "in-warranty,'' your replacement unit continues the original three year warranty period. You will not be billed for the replacement unit as long as the failed unit is received by Agilent.
• If your three year warranty has expired, Agilent Technologies will bill you for the Agilent E3632A exchange price - less than a new unit price. Agilent warrants exchange units against defects for 90 days.
Repacking for Shipment
For the Express Exchange Service described on the previous page, return your failed Agilent E3632A to the designated Agilent Service Center using the shipping carton of the exchange unit. A shipping label will be supplied. Agilent will notify you when your failed unit has been received.
If the instrument is to be shipped to Agilent for service or repair, be sure to:
• Attach a tag to the power supply identifying the owner and indicating the required service or repair. Include the instrument model number and full serial number.
• Place the power supply in its original container with appropriate packaging material.
• Secure the container with strong tape or metal bands.
If the original shipping container is not available, place your unit in a container which will ensure at least 4 inches of compressible packaging material around all sides for the power supply. Use static-free packaging materials to avoid additional damage to your unit.
Agilent Technologies recommends that you always insure shipments.
93
5
Chapter 5 Service

Electrostatic Discharge (ESD) Precautions

Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when serving the power supply or any electronic device.
• Disassemble instruments only in a static-free work area.
• Use a conductive work area to dissipate static charge.
• Use a conductive wrist strap to dissipate static charge accumulation.
• Minimize handling.
• Keep replacement parts in original static-free packaging.
• Remove all plastic, styrofoam, vinyl, paper, and other static-generating materials from the immediate work area.
•Use only anti-static solder suckers.

Surface Mount Repair

Surface mount components should only be removed using soldering irons or disordering stations expressly designed for surface mount components. Use
of conventional solder removal equipment will almost always result in permanent damage to the printed circuit board and will void your Agilent Technologies factory warranty.

To Replace the Power-Line Fuse

The power-line fuse is located within the power supply's fuse-holder assembly on the rear panel (see on page 24). For 100 or 115 Vac operation, you must use a 4 AT slow-blow fuse (Agilent part number 2110-0996). For 230 Vac operation, you must use a 2.5 AT slow-blow fuse (Agilent part number 2110-0999).
94
Chapter 5 Service

To Disconnect the Output Using an External Relay

To Disconnect the Output Using an External Relay
When the output of the E3632A is turned off, it is implemented by setting the output to 0 volts and 0.02 amps. This gives a zero output voltage without actually disconnecting the output. To disconnect the output, an external relay must be connected between the output and the load. A TTL signal of either low true or high true is provided to control an external relay. This signal can only be controlled with the remote command output is available on the RS-232 connector pin 1 and pin 9.
When the and pin 9 is low (0.5 V). The levels are reversed when the is "OFF".
OUTPut:RELay
state is "ON", the TTL ou tput of pin 1 is high (4.5 V)
OUTPut:RELay {OFF|ON}
OUTPut:RELay
. The TTL
state
Note
Note
TTL output of pin 1 or pin 9 of the RS-232 connector is available only after installing two jumpers inside the power supply. See below for more information.
Do not use the RS-232 interface if you have configured the power supply to output relay control signals. Internal components on the RS-232 circuitry may be damaged.

Installation Procedure

Th e a ssembly draw ings and sc h ematics a re lo cate d in chapte r 8, " Schematic s".
Remove the front and rear bumpers and take off the cover (See the mechanical
1
disassembly drawing on page 119).
Install JP3 and JP4 located adjacent to the connector P5 (The JP3 and JP4 are
2
outlined with a circle in the component locator diagram on page 120). A bare wire may be used.
Reassemble the power supply.
3
5
95
Chapter 5 Service

Troubleshooting Hints

Troubleshooting Hints
This section provides a brief check list of common failures. Before troubleshooting or repairing the power supply, make sure that the failure is in the instrument rather than any external connections. Also make sure that the instrument is accurately calibrated. The power supply's circuits allow troubleshooting and repair with basic equipment such as a digit multimeter and a 100 MHz oscilloscope.
Caution
This instrument contains CMOS integrated circuits which are susceptible to failure due to electrostatic discharge. Refer to the ``Electrostatic Discharge (ESD) Precautions'' section earlier in this chapter for further handling precautions.

Unit is Inoperative

Verify that the ac power cord is connected to the power supply.
Verify that the front-panel power switch is depressed.
Verify that the power-line fuse is installed:
Use the 4 AT, 250 V fuse for 100 or 115 Vac operation.. Use the 2.5 AT, 250 V fuse for 230 Vac operation..
Verify the power-line voltage setting.
See "To prepare the power supply for use" on page 23.

Unit Reports Errors 740 to 750

These errors may be produced if you accidentally turn off power the unit during a calibration or while changing a non-volatile state of the instrument. Recalibration or resetting the state should clear the error. If the error persists, a hardware failure may have occurred.

Unit Fails Self-Test

Verify that the correct power-line voltage setting is selected. Also, ensure that all terminal connections are removed while the self-test is performed. Failure of the DAC U21 on the top board will cause many self-test failures.
96
Chapter 5 Service
Troubleshooting Hints

Bias Supplies Problems

Check that the input to the voltage regulators of the bias supplies is at least 1 V greater than their output.
Circuit failures can cause heavy loads of the bias supplies which may pull down the regulator output voltages.
Check the voltages of bias supplies as tabulated below.
Table 5-1. Bias Supplies Voltages
Bias Supply Minimum Maximum Check At
+5V Floating
-5.1V Floating +15V Floating
-15V Floating
Some circuits produce their own local bias supplies from the main bias supplies. Be sure to check that these local bias supplies are active. In particular, the ADC (analog-to-digital converter), ac input, and front panel sections have local bias supplies. Always check that the power supplies are free of ac oscillations using an oscilloscope. Failure of bias supplies will cause many self­test failures.
+4.75 V
-4.75 V +14.25 V
-14.25 V
+5.25 V
-5.25 V +15.75 V
-15.75 V
U11 pin 2 Anode of CR5 Anode of CR9 Cathode of CR10
97
5
Chapter 5 Service

Self-Test Procedures

Self-Test Procedures

Power-On Self-Test

Each time the power supply is powered on, a set of self-tests are performed. These tests check that the minimum set of logic and measurement hardware are functioning properly. The power-on self-test performs checks 601 through 604 and 624 through 634.

Complete Self-Test

Ho ld any fro n t panel k e y except t he "Error ' ' key fo r mo re t han 5 se con ds w hile turning on the power to perform a complete self-test. The power supply beeps when the test starts. The tests are performed in the order shown below.
601
602
603
604
605
Front Panel Does not respond The main controller U17 on the top board attempts to establish serial communications with the front panel controller U7 on the front panel board. During this test, U7 turns on all display segments. Communication must function in both directions for this test to pass. If this error is detected during power-on self-test, the power supply will beep twice. This error is only readable from the remote interface.
RAM read/write failed This test writes and reads a 55 board pattern to each address of ram U14. Any incorrect readback will cause
a test failure. This error is only readable from the remote interface.
A/D sync stuck The main controller issues an A/D sync pulse to U17 and U18 to latch the value in the ADC slope counters. A failure is detected when a sync interrupt is not recognized and subsequent time-out occurs.
A/D slope convergence failed The input amplifier is configured to the measure zero (MZ) state in the 10 V range. This test checks whether the ADC integrator produces nominally the same number of positive and negative slope
decisions (
Cannot calibrate rundown gain This test checks the nominal gain between integrating ADC and the U17 on-chip ADC. This error is reported if the procedure can not run to completion due to a hardware failure.
±10%) during a 20 ms interval.
and AAh checker
h
98
Chapter 5 Service
Self-Test Procedures
606
607
608
624
625
626
Rundown gain out of range This test checks the nominal gain between the integrating ADC and the U17 on-chip ADC. The nominal gain is checked to
±10% tolerance.
Rundown too noisy This test checks the gain repeatability between the integrating ADC and the U17 on-chip ADC. The gain test (606) is performed
eight times. Gain noise must be less that
Serial configuration readback failed This test re-sends the last 3 byte serial configuration data to all the serial path (SERDAT, SERBCK, SERCLK). The data is then clocked back into U18 on the top board and compared against the original 3 bytes sent. A failure occurs if the data do not match. This tests checks the serial data path through U22.
Unable to sense line frequency This test checks that the LSENCE logic input U17 is togging. If no logic input detected, the power supply will assume a 50 Hz line operation for all future measurements.
I/O processor did not respond This test checks that communications can be established between U17 and U4 through the optically isolated (U6 and U7) serial data link. Failure to establish communication in either direction will generate an error. If this condition is detected at power-on self-test, the power supply will beep and the error annunciator will be on.
I/O processor failed self-test This test causes the earth referenced processor U4 to execute an internal, ram test. Failure will generate an error.
±64 lsb's of the U17 on-chip ADC.
5
630
631
632
Fan test failed This test checks if the fan current is flowing. If the current is not detected at power-on self-test, the power supply will beep and the error annunciator will be on.Fan test fail could likely induce overtemperature condition in the power supply.
System DAC t e s t f a i l e d This test checks if the DAC hardware is functional. The main controller U17 sends a reference voltage data to DAC and converts the DAC output to digital data to see if the digital data is within a valid range.
Hardware test failed This test checks the the status of voltage and current error amplifiers for the power circuit. If both amplifiers are not operational, the power supply will beep and the error annunciator will be on.
99
Loading...