The Agilent E3631A is a high performance 80 watt-triple output DC power
supply with GPIB and RS-232 interfaces. The combination of bench-top and
system features in this power supply provides versatile solutions for your
design and test requirements.
Convenient bench-top features
• Triple output
• Easy-to-use knob control for voltage and current settings
• Highly visible vacuum-fluorescent display for voltage and current meters
• Tracking operation for
• Excellent load and line regulation and low ripple and noise
• Operating states storage
• Portable, ruggedized case with non-skid feet
Flexible system features
• GPIB (IEEE-488) and RS-232 interfaces are standard
• SCPI (Standard Commands for Programmable Instruments) compatibility
• I/O setup easily done from front-panel
25V outputs
±
Agilent E3631A
Triple Output
DC Power Supply
Page 3
The Front Panel at a Glance
1 Meter and adjust selection keys
2 Tracking enable/disable key
3 Display limit key
4 Recall operating state key
5 Store operating state/Local key
6 Error/Calibrate key
Meter and adjust selection keys Select the output voltage and current of any one
1
supply (+6V, +25V, or -25V output) to be monitored on the display and allow knob
adjustment of that supply.
2 Tracking enable / disable key Enables / disables the track mode of ±25V supplies.
3 Display limit key Shows the voltage and current limit values on the display and
allows knob adjustment for setting limit values.
4 Recall operating state key Recalls a previously stored operating state from
location “1”, “2”, or “3”.
5 Store operating state / Local key
or “3” / or returns the power supply to local mode from remote interface mode.
6 Error / Calibrate key
and calibration / or enables calibration mode (the power supply must be unsecured
before performing calibration).
2
Displays error codes generated during operations, self-test
7 I/O Configuration / Secure key
interfaces / or secure and unsecure the power supply for calibration.
1
Stores an operating state in location “1”, “2”,
3
Configures the power supply for remote
8 Output On/Off key Enables or disables all three power supply outputs. This key
toggles between two states.
9 Control knob Increases or decreases the value of the blinking digit by turning
clockwise or counter clockwise.
10 Resolution selection keys Move the flashing digit to the right or left.
11 Voltage/current adjust selection key Selects the knob function to voltage control or
current control.
1
The key can be used as the “Local” key when the power supply is in the remote
interface mode.
2
You can enable the “calibration mode” by holding down this key when you
turn on the power supply.
3
You can use it as the “Secure” or “Unsecure” key when the power supply is
in the calibration mode.
3
Page 5
Front-Panel Voltage and Current Limit Settings
You can set the voltage and current limit values from the front panel
using the following method.
Use the voltage/current adjust selection key, the resolution selection keys, and
the control knob to change the monitoring or limiting value of voltage or current
.
Note
1
Press the key after turning on the power supply.
Display Limit
2 Set the knob to the voltage control mode or current control mode using the
voltage/current adjust selection key.
3 Move the blinking digit to the appropriate position using the resolution selection keys.
4 Change the blinking digit to the desired value using the control knob.
5 Press the key to enable the output. After about 5 seconds, the
display will go to the output monitor mode automatically to display the voltage
and current at the output.
Output On/Off
All front panel keys and controls can be disabled with remote interface
commands. The Agilent E3631A must be in “Local” mode for the front panel
keys and controls to function.
4
Page 6
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
+6VDisplays the output voltage and current for +6V supply. Knob is active for
+6V supply.
+25VDisplays the output voltage and current for +25V supply. Knob is active for
-25VDisplays the output voltage and current for -25V supply. Knob is active for
CALpower supply is in calibration mode.
TrackThe outputs of +25V and -25V supplies are in track mode.
LimitThe display shows the voltage and current limit value of a selected supply.
ERRORHardware or remote interface command errors are detected and also the
OFFThe three ouputs of the power supply are disabled.
UnregThe displayed output is unregulated (output is neither CV nor CC).
CVThe displayed output is in constant-voltage mode.
CCThe displayed output is in constant-current mode.
+25V supply.
-25V supply.
error bit has not been.
To review the display annunciators, hold down key as you
Display Limit
turn on the power supply.
5
Page 7
The Rear Panel at a Glance
1 Power-line voltage setting
2 Power-line fuse-holder assembly
3 AC inlet
Use the front-panel key to:
• Select the GPIB or RS-232 interface (see chapter 3).
• Set the GPIB bus address (see chapter 3).
• Set the RS-232 baud rate and parity (see chapter 3).
General Information Chapter 1 contains a general description of your
power supply. This chapter also provides instructions for checking your
power supply, connecting to ac power, and selecting power-line voltage.
Initial Operation Chapter 2 ensures that the power supply develops its
rated outputs and properly responds to operation from the front panel.
Front-Panel Operation Chapter 3 describes in detail the use of
front-panel keys and how they are used to operate the power supply from
the front panel. This chapter also shows how to configure the power supply
for the remote interface and gives a brief introduction to the calibration
features.
Remote Interface Reference Chapter 4 contains reference information
to help you program the power supply over the remote interface. This
chapter also explains how to program for status reporting.
Error Messages Chapter 5 lists the error messages that may appear as you
are working with the power supply. Each listing contains information to help
you diagnose and solve the problem.
Application Programs Chapter 6 contains some remote interface
applications to help you develop programs for your application.
Tutorial Chapter 7 describes basic operation of linear power supplies and
gives specific details on the operation and use of the Agilent E3631A power
supplies.
Specifications Chapter 8 lists the power supply’s specifications.
If you have questions relating to the operation of the power supply, call
1-800-452-4844 in the United States, or contact your nearest Agilent
Technologies Sales Office.
This chapter provides a general description of your power supply. This
chapter also contains instructions for initial inspection, location and cooling
for bench and rack operation, selecting the power-line voltage, and
connecting your power supply to ac power.
Options 0EM, 0E3, and 0E9 determine which power-line voltage is selected at
the factory. The standard unit is configured for 115 Vac ± 10%, 47-63 Hz input
voltage. For more information about changing the power-line voltage setting,
see ‘‘Power-Line Voltage Selection’’, starting on page 22 in this chapter.
OptionDescription
0EM115 Vac ± 10%, 47-63 Hz input voltage
0E3230 Vac ± 10%, 47-63 Hz input voltage
0E9 100 Vac ± 10%, 47-63 Hz input voltage
1CMRack mount kit (Agilent part number 5062-3957)
0L2Extra English manual set (local language manual files are included
on the CD-ROM, Agilent part number 5964-8251.)
Accessories
The accessories listed below may be ordered from your local
Agilent Technologies Sales Office either with the power supply or separately.
Agilent No. Description
10833AGPIB cable, 1 m (3.3 ft.)
10833BGPIB cable, 2 m (6.6 ft.)
34398ARS-232, 9 pin (f) to 9 pin (f), 2.5 m (8.2 ft.) cable; plus 9 pin (m)
to 25 pin (f) adapter
34399ARS-232 adapter kit (contains 4 adapters):
9 pin (m) to 25 pin (m) for use with PC or printer
9 pin (m) to 25 pin (f) for use with PC or printer
9 pin (m) to 25 pin (m) for use with modem
9 pin (m) to 9 pin (m) for use with modem
16
Page 18
Chapter 1 General Information
Description
Description
The Agilent E3631A power supply features a combination of programming
capabilities and linear power supply performance that makes it ideal for
power systems applications. The triple power supply delivers 0 to ± 25 V
outputs rated at 0 to 1 A and 0 to +6 V output rated at 0 to 5 A. The ± 25V
supplies also provide 0 to ± 25 V tracking output to power operational
amplifiers and circuits requiring symmetrically balanced voltages. The 0 to
± 25V outputs track each other within ±(0.2% output + 20 mV) in the track
mode. The ± 25V outputs can also be used in series as a single 0 to 50 V/1 A
supply.
The voltage and current of each supply can be adjusted independently from
the front panel or programmed over the GPIB or RS-232 interface. Using the
front panel keys and the control knob, you can adjust the voltage and
current of a selected output; enable or disable track mode; store and recall
operating states; enable or disable three outputs; calibrate the power supply
including changing the calibration security; return the power supply to local
operating mode; and configure the power supply for remote interface
operation.
From the front-panel VFD (vacuum-fluorescent display), you can monitor
actual values of output voltage and current (meter mode) or voltage and
current limit values (limit mode), check the operating status of the power
supply from the annunciators, and check the type of error from the
displayed error codes (messages).
When operated over the remote interface, the power supply can be both a
listener and a talker. Using an external controller, you can instruct the
power supply to set outputs and to send the status data back over the GPIB or
RS-232. Readback capabilities include reading back output voltage and
current; present and stored status; and error messages. The following
functions are implemented over the GPIB or RS-232:
1
• Voltage and current programming
• Voltage and current readback
• Enable or disable track mode
• Present and stored status readback
• Programming syntax error detection
• Voltage and current calibration
• Output on or off
• Self-test
17
Page 19
Chapter 1 General Information
Description
The front panel includes a VFD for displaying the output voltage and
current. Two 4-digit voltage and current meters accurately show the actual
or limit values of a selected supply simultaneously. Three meter selection
keys choose the voltage and current of any one output to be monitored on
the display.
Connections to the power supply's output and to chassis ground are made to
binding posts on the front panel. The +25V and -25V supply's outputs share a
common output terminal which is isolated from chassis ground. The positive
and negative terminals of each output can be grounded, or each output can
be kept within ±240 Vdc from the chassis ground. The power supply is
shipped with a detachable, 3-wire grounding type power cord. The ac line
fuse is an extractor type on the rear panel.
The power supply can be calibrated from the front panel directly or with a
controller over the GPIB or RS-232 interface using calibration commands.
Correction factors are stored in non-volatile memory and are used during
output programming. Calibration from the front panel or a controller
eliminates the need to remove the top cover or even the need to remove the
power supply from your system cabinet. You can guard against unauthorized
calibration by using the “Secured” calibration protection function.
18
Page 20
Chapter 1 General Information
Installation
Installation
Initial Inspection
When you receive your power supply, inspect it for any obvious damage that
may have occurred during shipment. If any damage is found, notify the
carrier and the nearest Agilent Sales Office immediately. Warranty information
is shown in the front of this manual.
Keep the original packing materials in case the power supply has to be
returned to Agilent Technologies in the future. If you return the power supply
for service, attach a tag identifying the owner and model number. Also include
a brief description of the problem.
Mechanical Check
This check confirms that there are no broken keys or knob, that the cabinet
and panel surfaces are free of dents and scratches, and that the display is not
scratched or cracked.
Electrical Check
Chapter 2 describes an initial operation procedure which, when successfully
completed, verifies to a high level of confidence that the power supply is
operating in accordance with its specifications. Detailed electrical
verification procedures are included in the Service Guide.
1
Cooling and Location
Cooling
The power supply can operate without loss of performance within the
temperature range of 0 °C to 40 °C, and with derated output current from
40 °C to 55 °C. A fan cools the power supply by drawing air through the rear
panel and exhausting it out the sides. Using an Agilent rack mount will not
impede the flow of air.
Bench Operation
Your power supply must be installed in a location that allows sufficient
space at the sides and rear of the power supply for adequate air circulation.
The rubber bumpers must be removed for rack mounting.
19
Page 21
Chapter 1
General Information
Installation
Rack Mounting
The power supply can be mounted in a standard 19-inch rack cabinet using
one of three optional kits available. A rack-mounting kit for a single
instrument is available as Option 1CM (P/N 5063-9243). Installation
instructions and hardware are included with each rack-mounting kit. Any
Agilent System II instrument of the same size can be rack-mounted beside the
Agilent E3631A power supply.
Remove the front and rear bumpers before rack-mounting the power
supply.
FrontRear (bottom view)
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
20
Page 22
Chapter 1
General Information
Installation
To rack mount two instruments of the same depth side-by-side, order
lock-link kit 5061-9694 and flange kit 5063-9214.
1
To install two instruments in a sliding support shelf, order support shelf
5063-9256, and slide kit 1494-0015.
21
Page 23
Chapter 1 General Information
Input Power Requirements
Input Power Requirements
You can operate your power supply from a nominal 100 V, 115 V, or 230 V
single phase ac power source at 47 to 63 Hz. An indication on the rear panel
shows the nominal input voltage set for the power supply at the factory. If
necessary, you can change the power-line voltage setting according to the
instructions on the next page.
Power-Line Cord
The power supply is shipped from the factory with a power-line cord that
has a plug appropriate for your location. Contact the nearest Agilent Sales and
Service Office if the wrong power-line cord is included with your power
supply. Your power supply is equipped with a 3-wire grounding type power
cord; the third conductor being the ground. The power supply is grounded
only when the power-line cord is plugged into an appropriate receptacle. Do
not operate your power supply without adequate cabinet ground connection.
Power-Line Voltage Selection
Power-line voltage selection is accomplished by adjusting two components:
power-line voltage selector and power-line fuse on the power-line module of
the rear panel. To change the power-line voltage, proceed as follows:
22
Page 24
Chapter 1 General Information
Input Power Requirements
1
1Remove the power cord. Remove the
fuse-holder assembly with a flat-blade
screwdriver from the rear panel.
3 Rotate the power-line voltage selector until
the correct voltage appears.
2Install the correct line fuse. Remove the
power-line voltage selector from the
power-line module.
4 Replace the power-line voltage selector and
the fuse-holder assembly in the rear panel.
23
Page 25
24
Page 26
2
2
Initial Operation
Page 27
Initial Operation
There are three basic tests in this chapter. The automatic power-on test
includes a self-test that checks the internal microprocessors and allows the
user visually to check the display. The output check ensures that the power
supply develops its rated outputs and properly responds to operation from
the front panel. For complete performance and/or verification tests, refer to
the Service Guide.
This chapter is intended for both the experienced and the inexperienced
user because it calls attention to certain checks that should be made prior to
operation.
Throughout this chapter the key to be pressed is shown in the left margin.
26
Page 28
Chapter 2 Initial Operation
Preliminary Checkout
Preliminary Checkout
The following steps help you verify that the power supply is ready for use.
1 Verify the power-line voltage setting on the rear panel.
The power-line voltage is set to the proper value for your country when the
power supply is shipped from the factory. Change the voltage setting if it is
not correct. The settings are: 100, 115, or 230 Vac.
2 Verify that the correct power-line fuse is installed.
The correct fuse is installed for your country when the power supply is
shipped from the factory. For 100 or 115 Vac operation, you must use a
2.5 AT fuse. For 230 Vac operation, you must use a 2.0 AT fuse.
3 Connect the power-line cord and turn on your power supply.
The front-panel display will light up and a power-on self-test occurs
automatically when you turn on the power supply.
See “Power-Line Voltage Selection”, starting on page 22 in chapter 1 if you
need to change the power-line voltage or the power-line fuse.
To replace the 2.5 AT fuse, order Agilent part number 2110-0913.
To replace the 2 AT fuse, order Agilent part number 2110-0982.
2
27
Page 29
Chapter 2 Initial Operation
Power-On Checkout
Power-On Checkout
The power-on test includes an automatic self-test that checks the internal
microprocessors and allows the user visually to check the display. You will
observe the following sequence on the display after pressing the front panel
power switch to on.
1 All segments of the display including all annunciators will turn on
for about one second.
Output On/Off
To review the annunciators, hold down the key as you
Display Limit
turn on the power supply.
2 The GPIB address or RS-232 message will then be displayed for
about one second.
ADDR 5 (or RS-232)
The GPIB address is set to “5” when the power supply is shipped from the
factory for remote interface configuration. If this is not the first time the
power supply is turned on, a different interface (RS-232) or a different GPIB
address may appear.
See ‘‘Remote Interface Configuration’’, for more information starting on
page 48 in chapter 3 if you need to change the remote interface configuration.
3The “OFF” and “+6V” annunciators are on. All others are off.
The power supply will go into the power-on / reset state; all outputs are
disabled (the OFF annunciator turns on); the display is selected for the +6V
supply (the +6V annunciator turns on); and the knob is selected for voltage
control.
4 Enable the outputs.
Press the
Output On/Off
key to enable the outputs. The OFF
annunciator turns off and the +6V and CV annunciators are lit. The
blinking digit can be adjusted by turning the knob. Notice that the
display is in the meter mode. “Meter mode” means that the display
shows the actual output voltage and current.
Note
If the power supply detects an error during power-on self-test, the ERROR
annunciator will turn on. See “Error Messages”, for more information
starting on page 113 in chapter 5
28
Page 30
Chapter 2 Initial Operation
Output Checkout
Output Checkout
The following procedures check to ensure that the power supply develops
its rated outputs and properly responds to operation from the front panel.
For complete performance and verification tests, refer to the Service Guide.
For each step, use the keys shown on the left margins.
Voltage Output Checkout
The following steps verify basic voltage functions with no load.
2
Power
Output On/Off
1 Turn on the power supply.
The power supply will go into the power-on / reset state; all outputs are
disabled (the OFF annunciator turns on); the display is selected for the +6V
supply (the +6V annunciator turns on); and the knob is selected for voltage
control.
2 Enable the outputs.
The OFF annunciator turns off and the +6V and CV annunciators are lit. The
blinking digit can be adjusted by turning the knob. Notice that the display is
in the meter mode. “Meter mode” means that the display shows the actual
output voltage and current.
3 Check that the front-panel voltmeter properly responds to knob
control for the +6V supply.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero.
4 Ensure that the voltage can be adjusted from zero to the maximum
rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the
knob until the voltmeter indicates 6.0 volts.
1
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the voltage.
29
Page 31
Chapter 2 Initial Operation
Output Checkout
+25V
-25V
5 Check the voltage function for the +25V supply.
Select the meter and adjust selection key for the +25V supply. The CV
annunciator is still lit and the +25V annunciator will turn on. Repeat steps
(3) and (4) to check the voltage function for the +25V supply.
6 Check the voltage function for the -25V supply.
Select the meter and adjust selection key for the -25V supply. The CV
annunciator is still lit and the -25V annunciator will turn on. Repeat steps
(3) and (4) to check the voltage function for the -25V supply.
30
Page 32
Chapter 2 Initial Operation
Output Checkout
Current Output Checkout
The following steps check basic current functions with a short across the
appropriate supply’s output.
Power
Output On/Off
Display Limit
Vol/Cur
1 Turn on the power supply.
2
The power supply will go into the power-on / reset state; all outputs are
disabled (the OFF annunciator turns on); the display is selected for the +6V
supply (the +6V annunciator turns on); and the knob is selected for voltage
control.
2 Connect a short across (+) and (-) output terminals of the +6V
supply with an insulated test lead.
3 Enable the outputs.
The OFF annunciator turns off and the +6V annunciator turns on. The CV or
CC annunciator is lit depending on the resistance of the test lead. The
blinking digit can be adjusted by turning the knob. Notice that the display is
in the meter mode. “Meter mode” means that the display shows the actual
output voltage and current.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the limit mode (the Lmt annunciator will be blinking).
Adjust the voltage limit to 1.0 volt to assure CC operation. The CC
annunciator will light.
5 Check that the front-panel ammeter properly responds to knob control
for the +6V supply.
Set the knob to the current control, and turn the knob clockwise or counter
clockwise when the display is in the meter mode (the Lmt annunciator is
off). Check that the ammeter responds to knob control and the voltmeter
indicates nearly zero (actually, the voltmeter will show the voltage
drop caused by the test lead).
31
Page 33
Chapter 2 Initial Operation
Output Checkout
6 Ensure that the current can be adjusted from zero to the maximum
rated value.
1
Adjust the knob until the ammeter indicates 0 amps and then until the ammeter
indicates 5.0 amps.
+25V
-25V
Note
7 Check the current function for the +25V supply.
Disable the outputs by pressing the key and connect a
Output On/Off
short across (+) and (COM) output terminals of the ±25V supply with an
insulated test lead. Repeat steps (3) through (6) after selecting the meter and
adjust selection key for the +25V supply.
8 Check the current function for the -25V supply.
Disable the outputs by pressing the key and connect a
Output On/Off
short across (-) and (COM) output terminals of +25V supply with an insulated
test lead. Repeat steps (3) through (6) after selecting the meter and adjust
selection key for the -25V supply.
If an error has been detected during the output checkout procedures, the
ERROR annunciator will turn on. See “Error Messages”, for more
information starting on page 113 in chapter 5
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the current.
32
Page 34
3
3
Front-Panel Operation
Page 35
Front-Panel Operation
So far you have learned how to install your power supply and perform
initial operation. During the initial operation, you were briefly introduced to
operating from the front panel as you learned how to check basic voltage
and current functions. This chapter will describe in detail the use of these
front-panel keys and show how they are used to accomplish power supply
operation.
• Front-Panel Operation Overview, page 35
• Constant Voltage Operation, page 36
• Constant Current Operation, page 38
• Tracking Operation, page 40
• Storing and Recalling Operating States, page 41
• Disabling the Outputs, page 43
• Knob Locking, page 43
• System-Related Operations, page 44
• Remote Interface Configuration, page 48
• GPIB Interface Configuration, page 53
• RS-232 Interface Configuration, page 54
• Calibration Overview, page 58
Note
Throughout this chapter the key to be pressed is shown in the left margin.
See “Error Messages”, starting on page 113 chapter 5 if you encounter any
errors during front-panel operation
34
Page 36
Chapter 3 Front-Panel Operation
Front-Panel Operation Overview
Front-Panel Operation Overview
The following section describes an overview of the front-panel keys before
operating your power supply.
• The power supply is shipped from the factory configured in the
front-panel operation mode. At power-on, the power supply is
automatically set to operate in the front-panel operation mode. When in
this mode, the front-panel keys can be used. When the power supply is in
remote operation mode, you can return to front-panel operation mode at
any time by pressing the key if you did not previously send the
front-panel lockout command. A change between front-panel and remote
operation modes will not result in a change in the output parameters.
Local
3
• When you press the key (the Lmt annunciator
blinks), the display of the power supply goes to the limit mode and
the present limit values of the selected supply will be displayed. In
this mode, you can also observe the change of the limit values when
adjusting the knob. If you press the key again or let the
display time-out after several seconds, the power supply will return the
display to the meter mode (the Lmt annunciator turns off). In this mode,
the actual output voltage and current will be displayed.
• All outputs of the power supply can be enabled or disabled from the front
panel using the key. When the output of the power supply
is off, the OFF annunciator turns on and the three outputs are disabled.
• The display provides the present operating status of the power supply
with annunciators and also informs the user of error codes. For example,
the +6V supply is operating in CV mode and controlled from the front
panel, then the CV and +6V annunciators will turn on. If, however, the
power supply is remotely controlled, the Rmt annunciator will also turn
on, and when the power supply is being addressed over GPIB interface,
the Adrs annunciator will turn on. See ‘‘Display Annunciators’’ on page 5
for more information.
Display Limit
Display Limit
Output On/Off
35
Page 37
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Constant Voltage Operation
To set up the power supply for constant voltage (CV) operation, proceed as
follows.
1 Connect a load to the desired output terminals.
With power-off, connect a load to the desired output terminals.
Power
Output
Display
On/Off
Limit
2 Turn on the power supply.
The power supply will go into the power-on / reset state; all outputs are
disabled (the OFF annunciator turns on); the display is selected for the +6V
supply (the +6V annunciator turns on); and the knob is selected for voltage
control.
3 Enable the outputs.
The OFF annunciator turns off and the +6V and CV annunciators are lit. The
blinking digit can be adjusted by turning the knob. Notice that the display is
in the meter mode. “Meter mode” means that the display shows the actual
output voltage and current.
To set up the power supply for +25V supply or -25V supply operation,
you should press the or key to select the display and adjust
+25V-25V
for +25V supply or -25V supply before proceeding to the next step.
4 Set the display for the limit mode.
Notice that the Lmt annunciator blinks, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage
and current limit values of the selected supply.
In constant voltage mode, the voltage values between the meter mode and
limit mode are the same, but the current values are not. Further if the
display is in the meter mode, you cannot see the change of current limit
value when adjusting the knob. We recommend that you should set the
display to “limit” mode to see the change of current limit value in the
constant voltage mode whenever adjusting the knob.
36
Page 38
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Vol
/
Cur
Vol
/
Cur
Display
Limit
5 Adjust the knob for the desired current limit.
1
Check that the Lmt annunciator still blinks. Set the knob for current
control. The second digit of ammeter will be blinking. Adjust the knob to
the desired current limit.
6 Adjust the knob for the desired output voltage.
1
Set the knob for voltage control. The second digit of the voltmeter will be
blinking. Adjust the knob to the desired output voltage.
7 Return to the meter mode.
Press the key or let the display time-out after several
Display Limit
seconds to return to the meter mode. Notice that the Lmt annunciator
turns off and the display returns to the meter mode. In the meter mode,
the display shows the actual output voltage and current of the selected
supply.
8 Verify that the power supply is in the constant voltage mode.
If you operate the +6V supply in the constant voltage (CV) mode, verify that
CV and +6V annunciators are lit. If you operate the power supply for the
+25V supply or the -25V supply, the +25V or -25V annunciator will turn on.
If the CC annunciator is lit, choose a higher current limit.
3
Note
During actual CV operation, if a load change causes the current limit to be
exceeded, the power supply will automatically crossover to the constant
current mode at the preset current limit and the output voltage will drop
proportionately.
1
You can use the resolution selection keys to move the blinking digit to the right or
left when setting the voltage and current.
37
Page 39
Chapter 3 Front-Panel Operation
Constant Current Operation
Constant Current Operation
To set up the power supply for constant current (CC) operation, proceed as
follows.
1 Connect a load to the output terminals of the desired supply.
With power-off, connect a load to the desired output terminals.
Power
Output On/Off
Display Limit
2 Turn on the power supply.
The power supply will go into the power-on / reset state; all outputs are disabled
(the OFF annunciator turns on); the display is selected for the +6V supply
(the +6V annunciator turns on); and the knob is selected for voltage control.
3 Enable the outputs.
The OFF annunciator turns off and the +6V and CV annunciators are lit. The
blinking digit can be adjusted by turning the knob. Notice that the display is
in the meter mode. “Meter mode” means that the display shows the actual
output voltage and current.
To set up the power supply for +25V supply or -25V supply operation, you
should press the or key to select the display and adjust for +25V
+25V-25V
supply or -25V supply before proceeding to the next step.
4 Set the display for the limit mode.
Notice that the Lmt annunciator blinks, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the selected supply.
In constant current mode, the current values between the meter mode and
limit mode are the same, but the voltage values are not. Further if the
display is in the meter mode, you cannot see the change of voltage limit
value when adjusting the knob. We recommend that you should set the
display to “limit” mode to see the change of voltage limit value in the
constant current mode whenever adjusting the knob.
38
Page 40
Chapter 3 Front-Panel Operation
Constant Current Operation
Vol/Cur
Display Limit
Note
5 Adjust the knob for the desired voltage limit.
1
Check that the knob is still selected for voltage control and the Lmt
annunciator blinks. Adjust the knob for the desired voltage limit.
6 Adjust the knob for the desired output current.
1
Set the knob for current control. The second digit of the ammeter will be
blinking. Adjust the knob to the desired current output.
7 Return to the meter mode.
Press the key or let the display time-out after several seconds
Display Limit
to return the meter mode. Notice that the Lmt annunciator turns off and the
display retur ns to the meter mode. In the meter mode, the display shows the
actual output voltage and current of the selected supply.
8 Verify that the power supply is in the constant current mode.
If you operate the +6V supply in the constant current (CC) mode, verify that
CC an d +6V annunciators are lit. If you operate the power supply for the +25V
supply or the -25V supply, the +25V or -25V annunciator will turn on.
If the CV annunciator is lit, choose a higher voltage limit.
During actual CC operation, if a load change causes the voltage limit to be
exceeded, the power supply will automatically crossover to constant voltage
mode at the preset voltage limit and the output current will drop
proportionately.
3
1
You can use the resolution selection keys to move the blinking digit to the
right or left when setting the voltage and current.
39
Page 41
Chapter 3 Front-Panel Operation
Tracking Operation
Tracking Operation
The ±25V supplies provide 0 to ±25 V tracking outputs. In the track mode,
two voltages of the ±25V supplies track each other within ±(0.2% output +20
mV) for convenience in varying the symmetrical voltages needed by
operational amplifiers and other circuits using balanced positive and
negative inputs. The state of track mode is stored in volatile memory; the
track is always off state when power has been off or after a remote interface
reset.
To operate the power supply in the track mode, proceed as follows:
1 Set the +25V supply to the desired voltage as described in previous
section “Constant Voltage Operation”(see page 36 for detailed
information).
Track
2 Enable the track mode.
Track
The key must be depressed for at least 1 second to enable the
track mode. When the track mode is first enabled, the -25V supply will
be set to the same voltage level as the +25V supply. Once enabled, any
change of the voltage level in either the +25V supply or the -25V supply
will be reflected in other supply. The current limit is independently set
for each of the +25V or the -25V supply and is not affected by the track
mode.
3 Verify that the ±25V supplies track each other properly.
You can verify that the voltage of the -25V supply tracks that of the +25V
supply within ±(0.2% of output + 20 mV) from the front-panel display by
comparing the voltage values of the +25V supply and the -25V supply.
In the track mode, if the CC annunciator is lit when the display is selected
for the +25V supply, choose a higher current limit for the +25V supply.
If the CC annunciator is lit when the display is selected for the -25V
supply, choose a higher current limit for the -25V supply.
40
Page 42
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Storing and Recalling Operating States
You can store up to three different operating states in non-volatile memory.
This also enables you to recall the entire instrument configuration with just
a few key presses from the front panel.
The memory locations are supplied from the factory for front panel
operation with the following states: display and knob selection for +6V
output;
disabled; and track off state.
and 0 V and 1 A for the ±25V supplies.
The following steps show you how to store and recall an operating state.
1 Set up the power supply for the desired operating state.
The storage feature “remembers” the display and knob selection state, the
limit values of voltage and current for three outputs, output on/off state, and
track on/off state.
values of voltage and current limits for three outputs; output
*RST
values for +6V supply are 0 V and 5 A
*RST
3
Store
2 Turn on the storage mode.
Three memory locations (numbered 1, 2 and 3) are available to store the
operating states. The operating states are stored in non-volatile memory and
are remembered when being recalled.
STORE1
This message appears on the display for approximately 3 seconds.
3 Store the operating state in memory location “3”.
Turn the knob to the right to specify the memory location 3.
STORE3
To cancel the store operation, let the display time-out after about 3
seconds or press any other function key except the key. The
power supply returns to the normal operating mode and to the
function pressed.
Store
41
Page 43
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Store
Recall
4 Save the operating state.
The operating state is now stored. To recall the stored state, go to the
following steps.
DONE
This message appears on the display for approximately 1 second.
5 Turn on the recall mode.
Memory location “1” will be displayed in the recall mode.
RECALL1
This message appears on the display for approximately 3 seconds.
6 Recall the stored operating state.
Turn the knob to the right to change the displayed storage location to “3”.
RECALL3
If this setting is not followed within 3 seconds with a key
Recall
stroke, the power supply returns to normal operating mode and will
not recall the instrument state 3 from memory.
Recall
7 Restore the operating state.
The power supply should now be configured in the same state as when you
stored the state on the previous steps.
DONE
This message appears on the display for approximately 1 second.
This section gives information on topics such as self-test, error conditions,
and front-panel display control. This information is not directly related to
setting up the power supply but is an important part of operating the power
supply.
Self-Test
A power-on self-test occurs automatically when you turn on the power
supply. This test assures you that the power supply is operational. This test
does not perform the extensive set of tests that are included as part of the
complete self-test described below. If the power-on self-test fails, the
ERROR annunciator turns on.
•A complete self-test performs a series of tests and takes approximately 2
seconds to execute. If all tests pass, you can have a high confidence that the
power supply is operational.
•If the complete self-test is successful, “PASS” is displayed on the front panel.
If the self-test fails, “FAIL” is displayed and the ERROR annunciator turns
on. See the Service Guide for instructions on returning the power supply to
Agilent Technologies for service.
• Front-panel operation:
The complete self-test is enabled by pressing the key (actually
any front panel keys except the key) and the power-line switch
Error
simultaneously and then continuing to press the key for 5
Recall
Recall
seconds. The complete self-test will be finished in 2 more seconds.
• Remote interface operation:
*TST?
Returns “0” if the complete self-test passes or “1” if it fails.
44
Page 46
Chapter 3 Front-Panel Operation
System-Related Operations
Error Conditions
When the front-panel ERROR annunciator turns on, one or more command
syntax or hardware errors have been detected. A record of up to 20 errors
can be stored in the power supply's error queue. See chapter 5 “Error
Messages”, starting on page 113 for a complete listing of the errors.
• Errors are retrieved in first-in-first-out (FIFO) order. The first error returned
is the first error that was stored. When you have read all errors from the
queue, the ERROR annunciator turns off. The power supply beeps once
each time an error is generated.
• If more than 20 errors have occurred when you operate the power supply
over the remote interface, the last error stored in the queue (the most recent
error) is replaced with -350, “Too many errors”. No additional errors are
stored until you remove errors from the queue. If no errors have occurred
when you read the error queue, the power supply responds with +0, “No error” over the remote interface or “NO ERRORS” from the front panel.
• The error queue is cleared when power has been off or after a
status) command has been executed. The
clear the error queue.
• Front-panel operation:
If the ERROR annunciator is on, press the key repeatedly to read
the errors stored in the queue. All errors are cleared when you read all
errors.
ERROR-113
• Remote interface operation:
SYSTem:ERRor?
Errors have the following format (the error string may contain up to 80
characters).
Reads one error from the error queue
(reset) command does not
*RST
Error
*CLS
(clear
3
-113,"Undefined header"
45
Page 47
Chapter 3 Front-Panel Operation
System-Related Operations
Display Control
For security reasons, you may want to turn off the front-panel display. From
the remote interface, you can display a 12-character message on the front
panel.
The display can be enabled / disabled from the remote interface only.
• When the display is turned off, outputs are not sent to the display and all
annunciators are disabled except the ERROR annunciator. Front-panel
operation is otherwise unaffected by turning off the display.
• The display state is stored in volatile memory; the display is always enabled
when power has been off, after a remote interface reset, or after returning to
local from remote.
• You can display a message on the front panel by sending a command from
the remote interface. The power supply can display up to 12 characters of
the message on the front panel; any additional characters are truncated.
Commas, periods, and semicolons share a display space with the preceding
character, and are not considered individual characters. When a message is
displayed, outputs are not sent to the display.
• Sending a message to the display from the remote interface overrides the
display state; this means that you can display a message even if the display is
turned off.
The display state is automatically turned on when you return to the local
(front panel) operation. Press the key to return to the local state
Local
from the remote interface
• Remote interface operation:
DISPlay {OFF|ON}
DISPlay:TEXT
<quoted string>
DISPlay:TEXT:CLEar
Disable / enable the display
Display the string enclosed in quotes
Clear the displayed message
The following statement shows how to display a message on the front panel
from a Agilent Technologies controller.
"DISP:TEXT ’HELLO’"
46
Page 48
Chapter 3 Front-Panel Operation
System-Related Operations
Firmware Revision Query
The power supply has three microprocessors for control of various internal
systems. You can query the power supply to determine which revision of
firmware is installed for each microprocessor.
You can query the firmware revision from the remote interface only.
• The power supply returns four fields separated by commas and the fourth
field is a revision code which contains three numbers. The first number is
the firmware revision number for the main processor; the second is for the
input/output processor; and the third is for the front-panel processor.
• Remote interface operation
*IDN?
Be sure to dimension a string variable with at least 40 characters.
Returns “
HEWLETT-PACKARD,E 3631A,0, X.X-X.X-X .X
SCPI Language Version
The power supply complies with the rules and regulations of the present
version of SCPI (Standard Commands for Programmable Instruments). You
can determine the SCPI version with which the power supply is in
compliance by sending a command from the remote interface.
You can query the SCPI version from the remote interface only.
• Remote interface operation:
SYSTem:VERSion?
Returns a string in the form “YYYY.V” where the “Y’s” represent the year of
the version, and the “V” represents a version number for that year (for
example, 1995.0).
3
”
47
Page 49
Chapter 3 Front-Panel Operation
Remote Interface Configuration
Remote Interface Configuration
Before you can operate the power supply over the remote interface, you
must configure the power supply for the remote interface. This section gives
information on configuring the remote interface. For additional information
on programming the power supply over the remote interface, See chapter 4 “Remote Interface Reference”, starting on page 63.
Remote Interface Selection
The power supply is shipped with both an GPIB (IEEE-488) interface and an
RS-232 interface on the rear panel. Only one interface can be enabled at a
time. The GPIB interface is selected when the power supply is shipped
from the factory.
The remote interface can be selected from the front-panel only.
• The interface selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
• If you select the GPIB interface, you must select a unique address for the
power supply. The current address is displayed momentarily on the front
panel when you turn on the power supply.
1
• Your GPIB bus controller has its own address. Be sure to avoid using the
bus controller’s address for any instrument on the interface bus.
Agilent Technologies controllers generally use address “21”.
• If you enable the RS-232 interface, you must select the baud rate and
parity to be used. “RS-232” is displayed momentarily on the front panel
when you turn on the power supply if you have selected this interface.
1
Refer to "GPIB Interface Configuration" starting on page 53 for more
information on connecting the power supply to a computer over the GPIB
interface.
2
Refer to "RS-232 Interface Configuration" starting on page 54 for more
information on connecting the power supply to a computer over the RS-232
interface.
2
48
Page 50
Chapter 3 Front-Panel Operation
Remote Interface Configuration
GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique address.
You can set the power supply’s address to any value between 0 and 30. The
current address is displayed momentarily on the front panel when you turn
on the power supply. The address is set to “05” when the power supply is
shipped from the factory.
The GPIB address can be set from the front-panel only.
• The address is stored in non-volatile memo ry, and does not change when
power has been off or after a remote interface reset.
• Your GPIB bus controller has its own address. Be sure to avoid the bus
controller’s address for any instrument on the interface bus.
Agilent Technologies controllers generally use address “21”.
Baud Rate Selection (RS-232)
You can select one of six baud rates for RS-232 operation. The rate is set to
9600 baud when the power supply is shipped from the factory.
The baud rate can be set from the front-panel only.
• Select one of the following: 300, 600, 1200, 2400, 4800, 9600 baud. The
factory setting is 9600 baud.
• The baud rate selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
Parity Selection (RS-232)
You can select the parity for RS-232 operation. The power supply is
configured for no parity and 8 data bits when shipped from the factory.
The parity can be set from the front-panel only.
• Select one of the following: None (8 data bits), Even (7 data bits), or
Odd (7 data bits). When you set the parity, you are indirectly setting the
number of data bits.
3
• The parity selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
49
Page 51
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the GPIB Address
To configure the power supply for the GPIB interface, proceed as follows:
I/O Config
I/O Config
I/O Config
1 Turn on the remote configuration mode.
GPIB / 488
You will see the above message on the front-panel display if the power
supply has not been changed from the default setting. If “RS-232” appears,
choose “GPIB / 488” by turning the knob to the right.
2 Select the GPIB address.
ADDR05
The address is set to “05” when the power supply is shipped from the
factory. Notice that a different GPIB address may appear if the power
supply has been changed from the default setting.
3 Turn the knob to change the GPIB address.
The displayed address is changed when turning the knob to the right or left.
4 Save the change and turn off the I/O configuration mode.
CHANGE SAVED
Note
The address is stored in non-volatile memory, and does not change when
power has been off or after a remote interface reset. The power supply
displays a message to show that the change is now in effect. If the GPIB address
is not changed, “NO CHANGE” will be displayed for one second.
To cancel the I/O configuration mode without any changes during the
GPIB address selection, press the “I/O Config” key until the “NO
CHANGE” message is displayed.
50
Page 52
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the Baud Rate and Parity (RS-232)
To configure the power supply for the RS-232 interface, proceed as follows:
I/O Config
I/O Config
1 Turn on the remote configuration mode.
GPIB / 488
You will see the above message on the display if the power supply has not
been changed from the default setting.
Notice that if you changed the remote interface selection to RS-232 before,
“RS-232” message will be displayed.
2 Choose the RS-232 interface.
RS-232
You can choose the RS-232 interface by turning the knob to the left.
3 Select the RS-232 interface and choose the baud rate.
9600BAUD
The rate is set to 9600 baud when the power supply is shipped from the
factory. Choose from one of the following by turning the knob to the right or
left: 300, 600, 1200, 2400, 4800, or 9600 baud.
3
I/O Config
4 Save the change and choose the parity.
NONE8BITS
The power supply is configured for 8 data bits with no parity when shipped
from the factory. Choose from one of the following by turning the knob to
the right or left: None 8 Bits, Odd 7 Bits, or Even 7 Bits. When you set
parity, you are indirectly setting the number of the data bits.
51
Page 53
Chapter 3 Front-Panel Operation
Remote Interface Configuration
I/O Config
Note
5 Save the change and turn off the I/O configuration mode.
CHANGE SAVED
The RS-232 baud rate and parity selections are stored in non-volatile
memory, and does not change when power has been off or after a remote
interface reset. The power supply displays a message to show that the
change is now in effect. If the baud rate and the parity are not changed,
“NO CHANGE” will be displayed for one second.
To cancel the I/O configuration mode without any changes during the
baud rate and parity selection, press the “I/O Config” key until the “NO
CHANGE” message is displayed.
52
Page 54
Chapter 3 Front-Panel Operation
GPIB Interface Configuration
GPIB Interface Configuration
The GPIB connector on the rear panel connects your power supply to the
computer and other GPIB devices. Chapter 1 lists the cables that are
available from Agilent Technologies. An GPIB system can be connected
together in any configuration (star, linear, or both) as long as the following
rules are observed:
• The total number of devices including the computer is no more than 15.
• The total length of all the cables used is no more than 2 meter times the
number of devices connected together, up to a maximum of 20 meters.
3
Note
IEEE-488 states that you should exercise caution if your individual cable
lengths exceed 4 meters
Do not stack more than three connector blocks together on any GPIB
connector. Make sure that all connectors are fully seated and that the lock
screws are firmly finger tightened.
53
Page 55
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
RS-232 Interface Configuration
You connect the power supply to the RS-232 interface using the 9-pin (DB-9)
serial connector on the rear panel. The power supply is configured as a DTE
(Data Terminal Equipment) device. For all communications over the RS-232
interface, the power supply uses two handshake lines: DTR (Data Terminal
Ready, on pin 4) and DSR (Data Set Ready, on pin 6).
The following sections contain information to help you use the power supply
over the RS-232 interface. The programming commands for RS-232 are
explained on page 87.
RS-232 Configuration Overview
Configure the RS-232 interface using the parameters shown below. Use
the front-panel key to select the baud rate, parity, and
number of data bits (see page 51 for more information to configure from
the front panel).
• Parity and Data Bits:None / 8 data bits (factory setting)
Even / 7 data bits, or
Odd / 7 data bits
• Number of Start Bits:1 bit (fixed)
• Number of Stop Bits:2 bits (fixed)
RS-232 Data Frame Format
A character frame consists of all the transmitted bits that make up a single
character. The frame is defined as the characters from the start bit to the
last stop bit, inclusively. Within the frame, you can select the baud rate,
number of data bits, and parity type. The power supply uses the following
frame formats for seven and eight data bits.
54
Page 56
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
Connection to a Computer or Terminal
To connect the power supply to a computer or terminal, you must have the
proper interface cable. Most computers and terminals are DTE (Data Terminal Equipment) devices. Since the power supply is also a DTE device,
you must use a DTE-to-DTE interface cable. These cables are also called
null-modem, modem-eliminator, or crossover cables.
The interface cable must also have the proper connector on each end and
the internal wiring must be correct. Connectors typically have 9 pins (DB-9
connector) or 25 pins (DB-25 connector) with a “male” or “female” pin
configuration. A male connector has pins inside the connector shell and a
female connector has holes inside the connector shell.
If you cannot find the correct cable for your configuration, you may have to
use a wiring adapter. If you are using a DTE-to-DTE cable, make sure the
adapter is a “straight-through” type. Typical adapters include gender
changers, null-modem adapters, and DB-9 to DB-25 adapters.
The cable and adapter diagrams shown below can be used to connect the
power supply to most computers or terminals. If your configuration is
different than those described, order the Agilent 34399A, 34399A Adapter Kit.
This kit contains adapters for connection to other computers, terminals, and
modems. Instructions and pin diagrams are included with the adapter kit.
DB-9 Serial Connection If your computer or terminal has a 9-pin
serial port with a male connector, use the null-modem cable included with
the Agilent 34398A Cable Kit. This cable has a 9-pin female connector on each
end. The cable pin diagram is shown below.
3
Instrument
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
1
2
3
4
5
6
7
8
9
DB9
Female
5182-4794
Cable
1
2
3
4
5
6
7
8
9
DB9
Female
PC
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
55
Page 57
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
DB-25 Serial Connection If your computer or terminal has a 25-pin serial
port with a male connector, use the null-modem cable and 25-pin adapter
included with the Agilent 34398A Cable Kit. The cable and adapter pin
diagram are shown below.
5182-4794
InstrumentPC
DCD
1
RX
2
TX
3
DTR
4
5
GND
6
DSR
7
RTS
8
CTS
9
RI
DB9
DB9
Female
Male
Cable
1
2
3
4
5
6
7
8
9
DB9
Female
1
2
3
4
5
6
7
8
9
DB9
Male
5181-6641
Adapter
2
3
4
5
6
7
8
20
DB25
Female
TX
RX
RTS
CTS
DSR
GND
DCD
DTR
DB25
Male
DTR/DSR Handshake Protocol
The power supply is configured as a DTE (Data Terminal Equipment)
device and uses the DTR (Data Terminal Ready) and DSR (Data Set Ready)
lines of the RS-232 interface to handshake. The power supply uses the DTR
line to send a hold-off signal. The DTR line must be TRUE before the power
supply will accept data from the interface. When the power supply sets the
DTR line FALSE, the data must cease within 10 characters.
To disable the DTR/DSR handshake, do not connect the DTR line and tie the
DSR line to logic TRUE. If you disable the DTR/DSR handshake, also select a
slower baud rate to ensure that the data is transmitted correctly.
The power supply sets the DTR line FALSE in the following cases:
1 When the power supply’s input buffer is full (when approximately 100
characters have been received), it sets the DTR line FALSE (pin 4 on the
RS-232 connector). When enough characters have been removed to make
space in the input buffer, the power supply sets the DTR line TRUE, unless
the second case (see next) prevents this.
56
Page 58
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
2 When the power supply wants to “talk” over the interface (which means that
it has processed a query) and has received a <new line> message
terminator, it will set the DTR line FALSE. This implies that once a query has
been sent to the power supply, the bus controller should read the response
before attempting to send more data. It also means that a <new line> must
terminate the command string. After the response has been output, the
power supply sets the DTR line TRUE again, unless the first case (see
above) prevents this.
The power supply monitors the DSR line to determine when the bus
controller is ready to accept data over the interface. The power supply
monitors the DSR line (pin 6 on the RS-232 connector) before each
character is sent. The output is suspended if the DSR line is FALSE. When
the DSR line goes TRUE, transmission will resume.
The power supply holds the DTR line FALSE while output is suspended. A
form of interface deadlock exists until the bus controller asserts the DSR
line TRUE to allow the power supply to complete the transmission. You can
break the interface deadlock by sending the <Ctrl-C> character, which
clears the operation in progress and discards pending output (this is
equivalent to the IEEE-488 device clear action).
3
For the <Ctrl-C> character to be recognized reliably by the power supply
while it holds DTR FALSE, the bus controller must first set DSR FALSE.
RS-232 Troubleshooting
Here are a few things to check if you are having problems communicating
over the RS-232 interface. If you need additional help, refer to the
documentation that came with your computer.
• Verify that the power supply and your computer are configured for the
same baud rate, parity, and number of data bits. Make sure that your
computer is set up for 1 start bit and 2 stop bits (these values are fixed
on the power supply).
• Make sure to execute the
power supply in the remote mode.
• Verify that you have connected the correct interface cable and adapters.
Even if the cable has the proper connectors for your system, the internal
wiring may be incorrect. The Agilent 34398A Cable Kit can be used to
connect the power supply to most computers or terminals.
• Verify that you have connected the interface cable to the correct serial
port on your computer (COM1, COM2, etc.).
SYSTem:REMote
command to place the
57
Page 59
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Overview
This section gives an overview of the calibration features of the power
supply. For more detailed discussion of the calibration procedures, see the
Service Guide.
Calibration Security
This feature allows you to enter a security code to prevent accidental or
unauthorized calibrations of the power supply. When you first receive your
power supply, it is secured. Before you can calibrate the power supply, you
must unsecure it by entering the correct security code.
• The security code is set to “HP003631” when the power supply is shipped
from the factory. The security code is stored in non-volatile memory, and
does not change when power has been off or after a remote interface reset.
• To secure the power supply from the remote interface, the security code
may contain up to 12 alphanumeric characters as shown below. The first
character must be a letter, but the remaining characters can be letters or
numbers. You do not have to use all 12 characters but the first character
must always be a letter.
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
• To secure the power supply from the remote interface so that it can be
unsecured from the front panel, use the eight-character format shown
below. The first two characters must be “H P” and the remaining
characters must be numbers. Only the last six characters are recognized
from the front panel, but all eight characters are required. To unsecure
the power supply from the front panel, omit the “H P” and enter the
remaining numbers as shown on the following
H P _ _ _ _ _ _ (8 characters)
If you forget your security code, you can disable the security feature by
adding a jumper inside the power supply, and then entering a new code.
See the Service Guide for more information.
58
Page 60
Chapter 3 Front-Panel Operation
Calibration Overview
To Unsecure for Calibration You can unsecure the power supply for
calibration either from the front panel or over the remote interface. The
power supply is secured when shipped from the factory, and the security
code is set to “HP003631”.
• Front-Panel Operation
SECURED
If the power supply is secured, you will see the above message for one
second by holding the
the power supply. To unsecure the power supply, press the key
Calibrate
key for 5 seconds when you turn on
Secure
after the “CAL MODE” message is displayed in the calibration mode,
enter the security code using the knob and resolution selection keys,
and then press the key.
Secure
000000 CODE
3
When you press the key to save the change, you will see the
Secure
message below for one second if the security code is correct. The
unsecured setting is stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset. To exit the
calibration mode, turn the power off and on.
Notice that if the security is incorrect, the power supply returns to the
code entering mode for you to enter the correct code.
UNSECURED
• Remote Interface Operation:
CALibrate:SECure:STATe, {OFF|ON},
<code>
To unsecure the power supply, send the above command with the same
code used to secure. For example,
"CAL:SEC:STAT OFF, HP003631"
59
Page 61
Chapter 3 Front-Panel Operation
Calibration Overview
To Secure Against Calibration You can secure the power supply against
calibration either from the front panel or over the remote interface. The
power supply is secured when shipped from the factory, and the security
code is set to “HP003631”.
Be sure to read the security code rules on page 58 before attempting to
secure the power supply.
• Front-Panel Operation:
UNSECURED
If the power supply is unsecured, you will see the above message for one
second by holding the key for 5 seconds when you turn on
the power supply. To secure the power supply, press the key
Calibrate
Secure
after the “CAL MODE” message is displayed in the calibration mode,
enter the security code using the knob and resolution selection keys,
and then press key.
Secure
Notice that you should omit the “H P” and enter the remaining numbers
as shown below.
000000 CODE
When you press the key to save the change, you will see the
Secure
message below. The secured setting is stored in non-volatile memory,
and does not change when power has been off or after a remote
interface reset. To exit the calibration mode, turn the power off and on.
SECURED
• Remote Interface Operation:
CALibrate:SECure:STATe {OFF|ON},
<code>
To secure the power supply, send the above command with the same code
as used to unsecure. For example,
"CAL:SEC:STAT ON, HP003631"
60
Page 62
Chapter 3 Front-Panel Operation
Calibration Overview
To Change the Security Code To change the security code, you must
first unsecure the power supply, and then enter a new code.
Be sure to read the security code rules on page 58 before attempting to
secure the power supply.
• Front-Panel Operation:
To change the security code, first make sure that the power supply is
unsecured. Press the key after the “CAL MODE” message is
displayed in the calibration mode, enter the new security code using the
knob and resolution selection keys, then press the key.
Changing the code from the front panel also changes the code required
from the remote interface.
• Remote Interface Operation:
Secure
Secure
3
CALibrate:SECure:CODE
To change the security code, first unsecure the power supply using the old
security code. Then, enter the new code. For example,
"CAL:SEC:STAT OFF, HP003631"
"CAL:SEC:CODE ZZ001443"
"CAL:SEC:STAT ON, ZZ00143"
<new code>
Unsecure with old code
Enter new code
Secure with new code
61
Page 63
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Count
You can determine the number of times that your power supply has been
calibrated. Your power supply was calibrated before it left the factory. When
you receive your power supply, read the count to determine its initial value.
The calibration count feature can be performed from the remote interface
only.
• The calibration count is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
• The calibration count increments up to a maximum of 32,767 after which
it wraps-around to 0. Since the value increments by one for each
calibration point, a complete calibration will increase the value by 6
counts.
• Remote Interface Operation:
CALibrate:COUNt?
Calibration Message
You can use the calibration message feature to record calibration
information about your power supply. For example, you can store such
information as the last calibration date, the next calibration due date, the
power supply's serial number, or even the name and phone number of the
person to contact for a new calibration.
You can record and read information in the calibration message from the
remote interface only.
• The power supply should be unsecured before sending a calibration
message.
• The calibration message may contain up to 40 characters.
• The calibration message is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
• Remote Interface Operation:
CALibrate:STRing
The following command string shows how to store a calibration message.
"CAL:STR 'CAL 05-1-95'"
<quoted string>
Store the cal message
62
Page 64
4
4
Remote Interface Reference
Page 65
Remote Interface Reference
• SCPI Command Summary, page 65
• Simplified Programming Overview, page 70
è
• Using the APPLy Command, page 73
• Output Setting and Operation Commands, page 74
• Triggering Commands, page 79
• System-Related Commands, page 82
• Calibration Commands, page 85
• RS-232 Interface Commands, page 87
• The SCPI Status Registers, page 88
• Status Reporting Commands, page 98
• An Introduction to the SCPI Language, page 102
è
• Halting an Output in Progress, page 107
• SCPI Conformance Information, page 108
• IEEE-488 Conformance Information, page 111
If you are a first-time user of the SCPI language, you may want to refer to
these sections to become familiar with the language before attempting to
program the power supply.
64
Page 66
Chapter 4 Remote Interface Reference
SCPI Command Summary
SCPI Command Summary
This section summarizes the SCPI (Standard Commands for Programmable
Instruments) commands available to program the power supply over the
remote interface. Refer to the later sections in this chapter for more
complete details on each command.
Throughout this manual, the following conventions are used for SCPI
command syntax.
• Square brackets ([ ]) indicate optional keywords or parameters.
• Braces ({ }) enclose parameters within a command string.
• Triangle brackets (< >) indicate that you must substitute a value or a code
for the enclosed parameter.
• A vertical bar ( | ) separates one of two or more alternative parameters.
This section gives an overview of the basic techniques used to program the
power supply over the remote interface. This section is only an overview
and does not give all of the details you will need to write your own
application programs. Refer to the remainder of this chapter and also
chapter 6, Application Programs, for more details and examples. Also refer
to the programming reference manual that came with your computer for
details on outputting command strings and entering data.
Using the APPLy Command
The
program the power supply over the remote interface. For example, the
following statement executed from your computer will set the +6V supply to
an output of 3 V rated at 1 A:
"APPL P6V, 3.0, 1.0"
command provides the most straightforward method to
APPLy
Using the Low-Level Commands
Although the
program the power supply, the low-level commands give you more flexibility
to change individual parameters. For example, the following statements
executed from your computer will set the +6V supply to an output of 3 V
rated at 1 A:
command provides the most straightforward method to
APPLy
"INST P6V"
"VOLT 3.0"
"CURR 1.0"
70
Select +6V output
Set output voltage to 3.0 V
Set output current to 1.0 A
Page 72
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Reading a Query Response
Only the query commands (commands that end with “?”) will instruct the
power supply to send a response message. Queries return either output
values or internal instrument settings. For example, the following statements
executed from your computer will read the power supply's error queue and
print the most recent error:
dimension statement
"SYST:ERR?"
bus enter statement
print statement
Dimension string array (80 elements)
Read error queu
Enter error string into computer
Print error string
Selecting a Trigger Source
The power supply will accept a “bus” (software) trigger or an immediate
internal trigger as a trigger source. By default, the “
selected. If you want the power supply to use an immediate internal trigger,
you must select “
executed from your computer will set the +6V supply to an output of 3 V/1 A
immediately:
"INST P6V"
"VOLT:TRIG 3.0"S
"CURR:TRIG 1.0"
"TRIG:SOUR IMM"
"INIT"
IMMediate
”. For example, the following statements
Select the +6V output
et the triggered voltage level to 3.0 V
Set the triggered current level to 1.0 A
Select the immediate trigger as a source
Cause the trigger system to initiate
” trigger source is
BUS
4
71
Page 73
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Programming Ranges and Output Identifiers
Output setting commands require a parameter for programming ranges and
an output name or an output number as the identifier of each output and
most queries will return a parameter. The programming range for a
parameter varies according to the selected output of the power supply. The
following table lists the programming ranges, output names, and output
numbers for each output.
Refer to this table to identify parameters when programming the power
supply.
Table 4-1. Agilent E3631A Programming Ranges and Output Identifiers
Output
+6V output+25V output-25V output
VoltageProgramming
CurrentProgramming
Output identifierP6VP25VN25V
Output number12 3
Range
MAX value6.18 V25.75 V-25.75 V
MIN value0 V0 V0 V
*RST value
(DEFault value)
Range
MAX value5.15 A1.03 A1.03 A
MIN value0 A0 A0 A
*RST value
(DEFault value)
0 to 6.18 V0 to +25.75 V0 to -25.75 V
0 V0 V0 V
0 to 5.15 A0 to 1.03 A0 to 1.03 A
5 A1 A1 A
72
Page 74
Chapter 4 Remote Interface Reference
Using the APPLy Command
Using the APPLy Command
The
APPLy
the power supply over the remote interface. You can select the specific
output, output voltage, and output current all in one command.
APPLy
{P6V | P25V | N25V}[,{<voltage>| DEF | MIN | MAX}[,{<current> | DEF | MIN | MA X}] ]
This command is combination of
VOLTage
the current of the specified output are changed as soon as the command is
executed.
You can identify each output by the output name (P6V, P25V or N25V) as
described in Table 4-1. For the voltage and current parameters of the
command, the ranges depend on the output currently selected. You can
substitute “MINimum”, “MAXimum”, or “DEFault” in place of a specific
value for the voltage an d current parameters. MIN selects the lowest voltage
and current values allowed for the selected output. MAX selects the highest
voltage and current values allowed. The default voltage values are 0 volts
for all outputs. The default current values are 5 A for +6V output and 1 A for
±25V outputs. The default voltage and current values are exactly the same as
the
*RST
If you specify only one value for the parameter, the power supply regards it
as voltage setting value. If you do not specify any value for the parameter,
the
APPLy
INSTrument
command provides the most straightforward method to program
INSTrument:SELect
, and
[SOURce:]CURRent
values. See Table 4-1 for details of parameters.
command only selects the output specified and acts as the
command.
commands. The values of voltage and
, [
SOURce:
]
APPLy
4
APPLy? [{P6V | P25V | N25V}]
This command queries the power supply's present voltage and current
values for each output and returns a quoted string. The voltage and current
are returned in sequence as shown in the sample string below (the quotation
marks are returned as part of the string). If any output identifier is not
specified, the voltage and the current of the currently selected output are
returned.
"5.000000,1.000000"
In the above string, the first number 5.000000 is the voltage limit value and
the second number 1.000000 is the current limit value for the specified
output.
73
Page 75
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output Setting and Operation Commands
This section describes the low-level commands used to program the power
supply. Although the
method to program the power supply, the low-level commands give you
more flexibility to change individual parameters.
See page 102 for programming ranges, output identifiers, and MIN / MAX
values in the following commands.
Output Selection Commands
INSTrument[:SELect] {P6V | P25V | N25V}
This command selects the output to be programmed among three outputs by
the output identifier. The outputs of the power supply are considered three
logical instruments. The
identify and select an output. When one output is selected, the other outputs
are unavailable for programming until selected. The commands which are
affected by the
(
SOURce
(
CALibration
output and “N25V” is for -25V output.
INSTrument
), measurement commands (
). “P6V” is the identifier for +6V output, “P25V” is for +25V
command provides the most straightforward
APPLy
INSTrument
command are output setting commands
command provides a mechanism to
MEASure
), and calibration commands
INSTrument[:SELect]?
This query returns the currently selected output by the
[
:SELect
parameter is “P6V”, “P25V”, or “N25V”.
INSTrument:NSELect {1 | 2 | 3}
This command selects the output to be programmed among three outputs by
a numeric value instead of the output identifier used in the
[
:SELect
“3” selects -25V output.
INSTrument:NSELect?
This query returns the currently selected output by the
:NSELect
is “1” for +6V output, “2” for +25V output or “3” for -25V output.
This command defines a coupling between various logical outputs of the
power supply. The couple command consists of an optional subsystem node
followed by a single parameter. The only valid parameter for the optional
subsystem node is
command,
The parameter indicates to which logical outputs the specified coupling is
to apply. “ALL” indicates that specified coupling is to apply to all outputs.
“NONE” indicates that specified coupling is to be removed. A list of outputs
specifies a particular set of logical outputs to be coupled. At
outputs are uncoupled. Noticethat
supplies can be coupled.
TRIGger
TRIGger
subsystem is assumed to be coupled.
subsystem. If no node follows the couple
, all
*RST
must be off before the ±25V
TRACk
INST:COUP
Example (1)
Note If you select the bus trigger source in the above program (see page 79 for
The following program segment shows how to use the
COUPle
outputs with voltage and current triggered levels. The power supply is set
to the newly programmed values as set by the
the detailed information), you must send the *TRG or Group Execute
Trigger (GET) command to start the trigger action after sending the
INITiate command.
command to couple two outputs between the +6V and the +25V
VOLTage:TRIGgered
commands.
Select the +6V output
Set triggered level to 5 V
Set triggered level to 3 A
Select the +25V output
Set triggered level to 20 V
Set triggered level to 0.5 A
Couple the +6V and +25V supply
Set trigger to immediate
Trigger the power supply to
output the trigger values for
the +6V and the +25V supplies
INSTrument:
4
and
75
Page 77
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INSTrument:COUPle[:TRIGger]?
This query returns the currently coupled output. Returns “ALL”, “NONE”, or
a list. If any output is not coupled, “NONE” is returned. If all of three
outputs are coupled, “ALL” is returned. If a list of outputs is coupled, the list
is returned.
Measurement Commands
MEASure:CURRent[:DC]? [{P6V | P25V | N25V}]
This command queries the current measured at the output terminals of the
power supply. The physical outputs of measurement are specified by the
output identifier. If any output identifier is not specified, the current of the
currently selected output is returned.
MEASure[:VOLTage][:DC]? [{P6V | P25V | N25V}]
This command queries the voltage measured at the output terminals of the
power supply. If any output identifier is not specified, the voltage of the
currently selected output is returned.
76
Page 78
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output On/Off and Tracking Operation Commands
OUTPut[:STATe] {OFF | ON}
This command enables or disables all three outputs of the power supply. The
state of the disabled outputs is a condition of less than 0.6 volts of opposite
polarity with no load and less than 60 mA of opposite direction with a short
circuit. At
OUTPut[:STATe]?
This command queries the output state of the power supply. The returned
value is “0” (OFF) or “1” (ON).
OUTPut:TRACk[:STATe] {OFF | ON}
This command enables or disables the power supply to operate in the track
mode. When the track mode is first enabled, the -25V supply will be set to
the same voltage level as the +25V supply. Once enabled, any change of the
programmed voltage level in either +25V supply or -25V supply will be
re fle c ted in th e oth er supply. The
supply to the non-track mode. The ±25V supplies must not be coupled to
enable “Track”. At
, the output state is off.
*RST
, the track mode is disabled.
*RST
TRACk OFF
command returns the power
4
OUTPut:TRACk[:STATe]?
This command queries the track mode state of the power supply. The
returned value is “0” (OFF) or “1” (ON).
This command directly programs the immediate current level of the power
supply. The immediate level is the current limit value of the output selected
with the
This query returns the presently programmed current limit level of the
selected output.
the maximum and minimum programmable current levels of the selected
output.
This command programs the pending triggered current level of the power
supply. The pending triggered current level is a stored value that is
transferred to the output terminals when a trigger occurs. A pending
triggered level is not affected by subsequent
This command directly programs the immediate voltage level of the power
supply. The immediate level is the voltage limit value of the selected output
with the
This command programs the pending triggered voltage level of the power
supply. The pending triggered voltage level is a stored value that is
transferred to the output terminals when a trigger occurs. A pending
triggered level is not affected by subsequent
This query returns the presently programmed triggered voltage level. If no
triggered level is programmed, the
:TRIGgered? MAXimum
and
VOLTage:TRIGgered? MINimum
VOLTage
level is returned.
VOLTage
return
the maximum and minimum programmable triggered voltage levels.
78
Page 80
Chapter 4 Remote Interface Reference
Triggering Commands
Triggering Commands
The power supply's triggering system allows a change in voltage and current
when receiving a trigger, to select a trigger source, and to insert a trigger.
Triggering the power supply is a multi-step process.
• First, you must select an output with the
command and then configure the power supply for the triggered output
level by using
commands.
• Then, you must specify the source from which the power supply will
accept the trigger. The power supply will accept a bus (software) trigger
or an immediate trigger from the remote interface.
• Then, you can set the time delay between the detection of the trigger on
the specified trigger source and the start of any corresponding output
change. Notice that the time delay is valid for only the bus trigger source.
• Finally, you must provide an
IMMediate
level immediately. But if the trigger source is the bus, the power supply is
set to the triggered level after receiving the Group Execute Trigger (GET)
or
*TRG
CURRent:TRIGgered
INITiate[:IMMediate]
source is selected, the selected output is set to the triggered
command.
INSTrument:SELect
and
VOLTage:TRIGgered
command. If the
Trigger Source Choices
You must specify the source from which the power supply will accept a
trigger. The trigger is stored in volatile memory; the source is set to bus
when the power supply has been off or after a remote interface reset.
4
Bus (Software) Triggering
• To select the bus trigger source, send the following command.
TRIGger:SOURce BUS
• To trigger the power supply from the remote interface (GPIB or RS-232)
after selecting the bus source, send the
the
any delay is given.
is sent, the trigger action starts after the specified time delay if
*TRG
(trigger) command. When
*TRG
79
Page 81
Chapter 4 Remote Interface Reference
Triggering Commands
• You can also trigger the power supply from the GPIB interface by
sending the IEEE-488 Group Execute Trigger (GET) message. The
following statement shows how to send a GET from a Agilent Technologies
controller.
TRIGGER 705 (group execute trigger)
• To ensure synchronization when the bus source is selected, send the
(wait) command. When the
*WAI
supply waits for all pending operations to complete before executing any
additional commands. For example, the following command string
guarantees that the first trigger is accepted and is executed before the
second trigger is recognized.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
command is executed, the power
*WAI
• You can use the
(operation complete) command to signal when the operation is
*OPC
complete. The
operation is complete. The
the Standard Event register when the operation is complete.
Immediate Triggering
• To select the immediate trigger source, send the following command.
This command causes the trigger system to initiate. This command
completes one full trigger cycle when the trigger source is an immediate and
initiates the trigger subsystem when the trigger source is bus.
This command sets the time delay between the detection of an event on the
specified trigger source and the start of any corresponding trigger action on
the power supply output. Select from 0 to 3600 seconds. MIN = 0 seconds.
MAX = 3600 seconds. At
TRIGger[:SEQuence]:DELay?
This command queries the trigger delay.
TRIGger[:SEQuence]:SOURce {BUS | IMMediate}
This command selects the source from which the power supply will accept
a trigger. The power supply will accept a bus (software) trigger or an internal
immediate trigger. At
*RST
, this value is set to 0 seconds.
*RST
, the bus trigger source is selected.
4
TRIGger[:SEQuence]:SOURce?
This command queries the present trigger source. Returns “BUS” or “IMM”.
*TRG
This command generates a trigger to the trigger subsystem that has selected a
bus (software) trigger as its source (
command has the same effect as the Group Execute Trigger (GET)
command. For RS-232 operation, make sure the power supply is in the
remote interface mode by sending the
TRIGger:SOURce BUS
SYSTem:REMote
). The
command first.
81
Page 83
Chapter 4 Remote Interface Reference
System-Related Commands
System-Related Commands
DISPlay[:WINDow][:STATe] {OFF | ON}
This command turns the front-panel display off or on. When the display is
turned off, outputs are not sent to the display and all annunciators are disabled
except the ERROR annunciator.
The display state is automatically turned on when you return to the local
mode. Press the key to return to the local state from the remote
interface.
DISPlay[:WINDow][:STATe]?
This command queries the front-panel display setting. Returns “0” (OFF) or
“1” (ON).
DISPlay[:WINDow]:TEXT[:DATA] <quoted string>
This command displays a message on the front panel. The power supply will
display up to 12 characters in a message; any additional characters are
truncated. Commas, periods, and semicolons share a display space with the
preceding character, and are not considered individual characters.
Local
DISPlay[:WINDow]:TEXT[:DATA]?
This command queries the message sent to the front panel and returns a
quoted string.
DISPlay[:WINDow]:TEXT:CLEar
This command clears the message displayed on the front panel.
SYSTem:BEEPer[:IMMediate]
This command issues a single beep immediately.
82
Page 84
Chapter 4 Remote Interface Reference
System-Related Commands
SYSTem:ERRor?
This command queries the power supply's error queue. When the front-panel
ERROR annunciator turns on, one or more command syntax or hardware
errors have been detected. Up to 20 errors can be stored in the error queue.
See “Error Messages” in chapter 5.
• Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. When you have read all errors
from the queue, the ERROR annunciator turns off. The power supply
beeps once each time an error is generated.
• If more than 20 errors have occurred, the last error stored in the queue
(the most recent error) is replaced with -350, “Too many errors”. No
additional errors are stored until you remove errors from the queue. If no
errors have occurred when you read the error queue, the power supply
responds with +0, “No error”.
• The error queue is cleared when power has been off or after a
(clear status) command has been executed. The
does not clear the error queue.
SYSTem:VERSion?
This command queries the power supply to determine the present SCPI
version. The returned value is of a string in the form YYYY.V where the “Y’s”
represent the year of the version, and the “V” represents a version number for
that year (for example, 1995.0).
*IDN?
This query command reads the power supply's identification string. The
power supply returns four fields separated by commas. The first field is the
manufacturer's name, the second field is the model number, the third field is
not used (always “0”), and the fourth field is a revision code which contains
three numbers. The first number is the firmware revision number for the
main power supply processor; the second is for the input/output processor;
and the third is for the front-panel processor.
The command returns a string with the following format (be sure to
dimension a string variable with at least 40 characters):
HEWLETT-PACKARD,E3631A,0,X.X-X.X-X.X
(reset) command
*RST
*CLS
4
83
Page 85
Chapter 4 Remote Interface Reference
System-Related Commands
*RST
This command resets the power supply to its power-on state as follows:
Command
State
CURR[:LEV][:IMM]Output dependent value*
CURR[:LEV]:TRIGOutput dependent value*
DISP[:STAT]ON
INST[:SEL]P6V
INST:COUPNONE
OUTP[:STAT]OFF
OUTP:TRACOFF
TRIG:DEL0
TRIG:SOURBUS
VOLT[:LEV][:IMM]0
VOLT[:LEV]:TRIG0
The reset operation sets the current of +6V output to 5 A and the current of
*
+25V and -25V outputs to 1 A.
*TST?
This query performs a complete self-test of the power supply. Returns “0” if
the self-test passes or “1” or any non-zero value if it fails. If the self-test fails,
an error message is also generated with additional information on why the
test failed.
*SAV { 1 | 2 | 3 }
This command stores the present state of the power supply to the specified
location in non-volatile memory. Three memory locations (numbered 1, 2
and 3) are available to store operating states of the power supply. The state
storage feature “remembers” the states or values of
. To recall a stored state, you must use the same memory location
,
, and
used previously to store the state.
*RCL {1 | 2 | 3 }
This command recalls a previously stored state. To recall a stored state, you
must use the same memory location used previously to store the state. You
recall
states or values of the power supply from a memory location
*RST
that was not previously specified as a storage location.
84
Page 86
Chapter 4 Remote Interface Reference
Calibration Commands
Calibration Commands
See chapter 3 “Calibration Overview”, starting on page 58 for an overview
of the calibration features of the power supply. For more detailed discussion
of the calibration procedures, see the Service Guide.
CALibration:COUNt?
This command queries the power supply to determine the number of times it
has been calibrated. Your power supply was calibrated before it left the
factory. When you receive your power supply, read the count to determine its
initial value. Since the value increments by one for each calibration point,
a complete calibration for three outputs will increase the value by six counts.
CALibration:CURRent[:DATA] <numeric value>
This command can only be used after calibration is unsecured. It enters a
current value of a selected output that you obtained by reading an external
meter. You must first select a calibration level (
value being entered. Two successive values (one for each end of the
calibration range) must be selected and entered. The power supply then
computes new calibration constants. These constants are then stored in
non-volatile memory.
CAL:CURR:LEV
) for the
4
CALibration:CURRent:LEVel {MINimum | MAXimum}
Before using this command, you must select the output which is to be
calibrated by using
used after calibration is unsecured. It sets the power supply to a calibration
point that is entered with
During calibration, two points must be entered and the low-end point (MIN)
must be selected and entered first.
CALibration:SECure:CODE <new code>
This command enters a new security code. To change the security code, first
unsecure the power supply using the old security code. Then, enter the new
code. The calibration code may contain up to 12 characters over the remote
interface but the first character must always be a letter.
INSTrument
CALibration:CURRent[:DATA
command. This command can only be
] command.
85
Page 87
Chapter 4 Remote Interface Reference
Calibration Commands
CALibration:SECure:STATe {OFF | ON>}, <code>
This command unsecures or secures the power supply for calibration. The
calibration code may contain up to 12 characters over the remote interface.
CALibration:SECure:STATe?
This command queries the secured state for calibration of the power supply.
The returned parameter is “0” (OFF) or “1” (ON).
CALibration:STRing <quoted string>
This command records calibration information about your power supply.
For example, you can store such information as the last calibration date, the
next calibration due date, or the power supply’s serial number. The
calibration message may contain up to 40 characters. The power supply
should be unsecured before sending a calibration message.
CALibration:STRing?
This command queries the calibration message and returns a quoted string.
CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unsecured. It enters a
voltage value of a selected output that you obtained by reading an external
meter. You must first select a calibration level (
CAL:VOLT:LEV
) for the
value being entered. Two successive values (one for each end of the
calibration range) must be selected and entered. The power supply then
computes new voltage calibration constants. These constants are then
stored in non-volatile memory.
CALibration:VOLTage:LEVel {MINimum | MAXimum}
Before using this command, you must select the output which is to be
calibrated by using
INSTrument
command. This command can only be
used after calibration is unsecured. It sets the power supply to a calibration
point that is entered with
CALibration:VOLTage[:DATA]
command.
During calibration, two points must be entered and the low-end point (MIN)
must be selected
86
Page 88
Chapter 4 Remote Interface Reference
RS-232 Interface Commands
RS-232 Interface Commands
Use the front-panel “I/O configuration” key to select the baud rate, parity,
and the number of data bits (See chapter 3 “Remote Interface Configuration”,
starting on page 48).
SYSTem:LOCal
This command places the power supply in the local mode during RS-232
operation. All keys on the front panel are fully functional.
SYSTem:REMote
This command places the power supply in the remote mode for RS-232
operation. All keys on the front panel, except the “Local” key, are disabled.
It is very important that you send the
place the power supply in the remote mode. Sending or receiving data
over the RS-232 interface when not configured for remote operation can
cause unpredictable results.
SYSTem:RWLock
This command places the power supply in the remote mode for RS-232
operation. This command is the same as the
except that all keys on the front panel are disabled, including the “Local”
key.
Ctrl-C
This command clears the operation in progress over the RS-232 interface
and discard any pending output data. This is equivalent to the IEEE-488
device clear action over the GPIB interface.
SYSTem:REMote
SYSTem:REMote
command to
command
4
87
Page 89
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The SCPI Status Registers
All SCPI instruments implement status registers in the same way. The status
system records various instrument conditions in three register groups: the
Status Byte register, the Standard Event register, and the Questionable Status
register group. The status byte register records high-level summary
information reported in the other register groups. The diagrams on the
subsequent pages illustrate the SCPI status system used by the power supply.
An example program is included in chapter 6, “Application Programs,”
which shows the use of the status registers. You may find it useful to refer
to the program after reading the following section in this chapter.
What is an Event Register?
An event register is a read-only register that reports defined conditions
within the power supply. Bits in an event register are latched. Once an event
bit is set, subsequent state changes are ignored. Bits in an event register are
automatically cleared by a query of that register (such as
STAT:QUES:EVEN?
reset (
an event register returns a decimal value which corresponds to the
binary-weighted sum of all bits set in the register.
) or device clear will not clear bits in event registers. Querying
*RST
) or by sending the
(clear status) command. A
*CLS
*ESR?
or
What is an Enable Register?
An enable register defines which bits in the corresponding event register are
logically ORed together to form a single summary bit. Enable registers are both
readable and writable. Querying an enable register will not clear it. The
(clear status) command does not clear enable registers but it does clear the
bits in the event registers. To enable bits in an enable register, you must write
a decimal value which corresponds to the binary-weighted sum of the bits you
wish to enable in the register.
*CLS
What is a Multiple Logical Output?
The three-logical outputs of the power supply include an INSTrument
summary status register and an individual instrument ISUMmary register for
each logical output. The ISUMmary registers report to the INSTrument
register, which in turn reports to bit 13 of the Questionable status register.
This is shown pictorially on the next page.
88
Page 90
Chapter 4 Remote Interface Reference
The SCPI Status Registers
Using such a status register configuration allows a status event to be crossreferenced by output and type of event. The INSTrument register indicates
which output(s) have generated an event. The ISUMmary register is a pseudoquestionable status register for a particular logical output.
The Questionable Status register provides information about unexpected
operation of the power supply. Bit 4 reports a fault with the fan, and bit 13
summarizes questionable outputs for any of the three supplies. For example
if one of the three supplies is in constant voltage mode and due to an
overload looses regulation, bit 13 is set (latched). Send the command
STAT:QUES?
enable registers you wish to summarize with bit 13. Send
:INST:ENAB 14
STAT:QUES:INST:ISUM<n>:ENAB 3
Questionable Instrument Summary register, where n is 1, 2, or 3.
Table 4-2. Bit Definitions - Questionable Status Register
0-3 Not used0Always set to 0.
4 FAN16The fan has a fault condition.
5-12 Not Used0Always set to 0.
13 ISUM8192Summary of QUES:INST and QUES:INST:ISUM registers.
14-15 Not Used0Always set to 0.
to read the register. To make use of bit 13 you must first
STAT:QUES
to enable the Questionable Instrument register. Then send
for each supply to enable the
Bit
Decimal
Value
Definition
4
The Questionable Instrument Status Register
The Questionable Instrument register provides information about
unexpected operations for each of the three supplies. For example if the +6V
supply is in the constant voltage mode and looses regulation, then bit 1 set
indicating a possible overload in the +6V supply. The +25V supply is
reported as bit 2, and the -25V supply as bit 3. Send the command
QUES:INST?
to read the register. The
STAT:QUES:INST:ISUM<n>
STAT
registers must be enabled to make use of the Questionable Instrument
register. Send
STAT:QUES:INST:ISUM<n>:ENAB 3
to enable output n.
The Questionable Instrument Summary Register
There are three Questionable Instrument Summary registers, one for each
supply output. These registers provide information about voltage and
current regulation. Bit 0 is set when the voltage becomes unregulated, and
bit 1 is set if the current becomes unregulated. For example if a supply
which is operating as a voltage source (constant voltage mode) momentarily
goes to constant current mode, bit 0 is set to indicate that the voltage output
is not regulated. To read the register for each supply, send
:INST:ISUM<n>?
, where n is 1, 2, or 3.
STAT:QUES
91
Page 93
Chapter 4 Remote Interface Reference
The SCPI Status Registers
To determine the operating mode (CV or CC) for the power supply send
STAT:QUES:INST:ISUM<n>:COND?
the output. Bit 1 true indicates the output is in constant voltage mode, bit 0
true indicates constant current mode, both bits true indicates neither the
voltage nor the current is regulated, and both bits false indicates the outputs
of the power supply are off.
The Questionable Status Event register is cleared when:
• You execute the
(clear status) command.
*CLS
, where n is 1, 2, or 3 depending on
• You query the event register using
[:EVENt]?
For example, 16 is returned when you have queried the status of the
questionable event register, the FAN condition is questionable.
The Questionable Status Enable register is cleared when:
• You execute
For example, you must send the
(Status Questionable Event register) command.
STATus:QUEStionable:ENABle 0
STATus:QUEStionable
STAT:QUES:ENAB 16
command.
to enable th e FAN bit.
92
Page 94
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The Standard Event Register
The Standard Event register reports the following types of instrument
events: power-on detected, command syntax errors, command execution
errors, self-test or calibration errors, query errors, or when an
command is executed. Any or all of these conditions can be reported in the
Standard Event Summary bit (ESB, bit 5) of Status Byte register through the
enable register. To set the enable register mask, you write a decimal value to
the register using the
(Event Status Enable) command.
*ESE
An error condition (Standard Event register bits 2, 3, 4, or 5) will
always record one or more errors in the power supply's error queue. Read
the error queue using the
Table 4-3. Bit Definitions - Standard Event Register
Bit
0OPC1
1Not Used0Always set to 0.
2QYE4
3DDE8
4EXE16
5CME 32
6Not Used0Always set to 0.
7PON128
Decimal
Value
SYSTem:ERRor?
Definition
Operation Complete. All commands prior to and including an
*OPC command have been executed.
Query Error. The power supply tried to read the output buffer
but it was empty . Or, new command line was received before
a previous query had been read. Or, both the input and output
buffers are full.
Device Error. A self-test or calibration error occurred (see error
numbers 601 through 748 in chapter 5).
Execution Error. An execution error occurred (see error numbers
-211 through -224 in chapter 5).
Command Error. A command syntax error occurred (see error
number -101 through -178 in chapter 5).
Power On. Power has been turned off and on since the last
time the event register was read or cleared
command.
*OPC
4
93
Page 95
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The Standard Event register is cleared when:
• You execute the
• You query the event register using the
(clear status) command.
*CLS
*ESR?
(Event Status register)
command.
For example, 28 (4 + 8 + 16) is returned when you have queried the status of
the Standard Event register, QYE, DDE, and EXE conditions have occurred.
The Standard Event Enable register is cleared when:
• You execute the
*ESE 0
command.
• You turn on the power and have previously configured the power supply
using the
*PSC 1
command.
• The enable register will not be cleared at power-on if you have previously
configured the power supply using the
For example, you must send the
*ESE 24
*PSC 0
command.
(8 + 16) to enable DDE and EXE
bits.
The Status Byte Register
The Status Byte summary register reports conditions from the other status
registers. Query data that is waiting in the power supply's output buffer is
immediately reported through the “Message Available” bit (bit 4) of Status
Byte register. Bits in the summary register are not latched. Clearing an event
register will clear the corresponding bits in the Status Byte summary
register. Reading all messages in the output buffer, including any pending
queries, will clear the message available bit.
Table 4-4. Bit Definitions - Status Byte Summary Register
Bit
0-2Not Used0Always set to 0.
3QUES8
4MAV16Data is available in the power supply output buffer.
5ESB32
6RQS64The power supply is requesting service (serial poll).
7Not Used0Always set to 0.
Decimal
Value
Definition
One or more bits are set in the questionable status
register (bits must be “enabled” in the enable register).
One or more bits are set in the standard event register
(bits must be “enabled” in the enable register).
94
Page 96
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The Status Byte Summary register is cleared when:
• You execute the
• Querying the Standard Event register (
bit 5 in the Status Byte summary register.
For example, 24 (8 + 16) is returned when you have queried the status of the
Status Byte register, QUES and MAV conditions have occurred.
The Status Byte Enable register (Request Service) is cleared when:
• You execute the
• You turn on the power and have previously configured the power supply
using the
• The enable register will not be cleared at power-on if you have previously
configured the power supply using
For example, you must send the
bits.
*PSC 1
(clear status) command.
*CLS
*SRE 0
command.
command.
*SRE 96
command) will clear only
*ESR?
*PSC 0
.
(32 + 64) to enable ESB and RQS
Using Service Request (SRQ) and Serial POLL
You must configure your bus controller to respond to the IEEE-488 service
request (SRQ) interrupt to use this capability. Use the Status Byte enable
register (
low-level IEEE-488 service request signal. When bit 6 (request service) is set
in the Status Byte, an IEEE-488 service request interrupt message is
automatically sent to the bus controller. The bus controller may then poll
the instruments on the bus to identify which one requested service (the
instrument with bit 6 set in its Status Byte).
command) to select which summary bits will set the
*SRE
4
The request service bit is cleared only by reading the Status Byte using an
IEEE-488 serial poll or by reading the event register whose summary bit
is causing the service request.
To read the Status Byte summary register, send the IEEE-488 serial poll
message. Querying the summary register will return a decimal value which
corresponds to the binary-weighted sum of the bits set in the register. Serial
poll will automatically clear the “request service” bit in the Status Byte
summary register. No other bits are affected. Performing a serial poll will
not affect instrument throughput.
95
Page 97
Chapter 4 Remote Interface Reference
The SCPI Status Registers
Caution The IEEE-488 standard does not ensure synchronization between your
bus controller program and the instrument. Use the *OPC? command to
guarantee that commands previously sent to the instrument have
completed. Executing a serial poll before a *RST,*CLS, or other
commands have completed can cause previous conditions to be reported.
Using *STB? to Read the Status Byte
The
processed like any other instrument command. The
returns the same result as a serial poll but the “request service” bit (bit 6) is not cleared.
The
*STB?
interface hardware and will be executed only after previous commands have
completed. Polling is not possible using the
*STB?
(Status Byte query) command is similar to a serial poll but it is
*STB?
*STB?
command is not handled automatically by the IEEE-488 bus
*STB?
command does not clear the Status Byte summary register.
command. Executing the
command
Using the Message Available Bit (MAV)
You can use the Status Byte “message available” bit (bit 4) to determine
when data is available to read into your bus controller. The power supply
subsequently clears bit 4 only after all messages have been read from the
output buffer.
To Interrupt Your Bus Controller Using SRQ
1 Send a device clear message to clear the power supply's output buffer (e.g.,
CLEAR 705
2 Clear the event registers with the
3 Set up the enable register masks. Execute the
Standard Event register and the
4 Send the
to ensure synchronization.
5 Enable your bus controller's IEEE-488 SRQ interrupt.
96
).
*OPC?
(clear status) command.
*CLS
command to set up the
*ESE
command for the Status Byte.
*SRE
(operation complete query) command and enter the result
Page 98
Chapter 4 Remote Interface Reference
The SCPI Status Registers
To Determine When a Command Sequence is Completed
1 Send a device clear message to clear the power supply's output buffer (e.g.,
CLEAR 705
).
2 Clear the event registers with the
3 Enable the “operation complete” bit (bit 0) in the Standard Event register by
executing the
4 Send the
to ensure synchronization.
5 Execute your command string to program the desired configuration,
and then execute the
command. When the command sequence is completed, the “operation
complete” bit (bit 0) is set in the Standard Event register.
6 Use a serial poll to check to see when bit 5 (standard event) is set in the
Status Byte summary register. You could also configure the power supply
for an SR Q i nt er rup t b y s en di ng
*ESE 1
*OPC?
command.
(operation complete query) command and enter the result
(operation complete) command as the last
*OPC
*SRE 32
(clear status) command.
*CLS
(St at us By te en abl e r eg is te r, bi t 5 ).
Using *OPC to Signal When Data is in the Output Buffer
Generally, it is best to use the “operation complete” bit (bit 0) in the
Standard Event register to signal when a command sequence is completed.
This bit is set in the register after an
you send
supply's output buffer (query data), you can use the “operation complete” bit
to determine when the message is available. However, if too many messages
are generated before the
buffer will fill and the power supply will stop processing commands.
after a command which loads a message in the power
*OPC
command executes (sequentially), the output
*OPC
command has been executed. If
*OPC
4
97
Page 99
Chapter 4 Remote Interface Reference
Status Reporting Commands
Status Reporting Commands
See diagram “SCPI Status System”, on page 90 in this chapter for detailed
information of the status register structure of the power supply.
SYSTem:ERRor?
This query command reads one error from the error queue. When the
front-panel ERROR annunciator turns on, one or more command syntax or
hardware errors have been detected. A record of up to 20 errors can be
stored in the power supply’s error queue. See “Error Messages” in chapter 5
• Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. When you have read all errors
from the queue, the ERROR annunciator turns off. The power supply
beeps once each time an error is generated.
• If more than 20 errors have occurred, the last error stored in the queue
(the most recent error) is replaced with -350, “Too many errors”. No
additional errors are stored until you remove errors from the queue. If no
errors have occurred when you read the error queue, the power supply
responds with +0, “No error”.
• The error queue is cleared when power has been off or after a
(clear status) command has been executed. The
does not clear the error queue.
STATus:QUEStionable[:EVENt]?
This command queries the Questionable Status event register. The power
supply returns a decimal value which corresponds to the binary-weighted
sum of all bits in the register.
STATus:QUEStionable:ENABle <enable value>
This command enables bits in the Questionable Status enable register. The
selected bits are then reported to the Status Byte.
STATus:QUEStionable:ENABle?
This command queries the Questionable Status enable register. The power
supply returns a binary-weighted decimal representing the bits set in the
enable register.
98
(reset) command
*RST
*CLS
Page 100
Chapter 4 Remote Interface Reference
Status Reporting Commands
STATus:QUEStionable:INSTrument[:EVENt]?
This command queries the Questionable Instrument event register. The
power supply returns a decimal value which corresponds to the
binary-weighted sum of all bits in the register and clears the register.
This command sets the value of the Questionable Instrument enable register.
This register is a mask for enabling specific bits from the Questionable
Instrument event register to set the Instrument Summary bit (ISUM, bit 13)
of the Questionable Status register. The “ISUM” bit of the Questionable
Status register is the logical OR of all the Questionable Instrument event
register bits that are enabled by the Questionable Instrument enable register.
STATus:QUEStionable:INSTrument:ENABle?
This query returns the value of the Questionable Instrument enable register.
This query returns the value of the Questionable Instrument Isummary event
register for a specific output of the three-output power supply. The
particular output must be specified by a numeric value. n is 1, 2, or 3. See
Table 4-1 on page 72 for the output number. The event register is a read-only
register which holds (latches) all events. Reading the Questionable
Instrument Isummary event register clears it.
This query returns the CV or CC condition of the specified instrument. If “2” is
returned, the queried instrument is in the CV operating mode. If “1” is returned,
the queried instrument is in the CC operating mode. If “0” is returned, the
outputs of the instrument are off or unregulated. If ‘3” is returned, the
instrument is in the hardware failure. n is 1, 2, or 3.
This command sets the value of the Questionable Instrument Isummary
enable register for a specific output of the three-output power supply. The
particular output must be specified by a numeric value. n is 1, 2, or 3. See
Table 4-1 on page 72 for the output number. This register is a mask for
enabling specific bits from the Questionable Instrument Isummary event
register to set the Instrument Summary bit (bit 1, 2, and 3) of the
Questionable Instrument register. These bits 1, 2, and bit 3 are the logical OR
of all the Questionable Instrument Isummary event register bits that are
enabled by the Questionable Instrument Isummary enable register.
99
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.