The Agilent E3631A isa high performance 80 watt-tripleoutput DCpower supply withGPIB and RS-232 interfaces. The combination of bench-top and system features in thispower supply provides versatile solutions for your design and test requirements.
Convenient bench-top features
•
Triple output
•
Easy-to-use knob controlfor voltage and current settings
•
Highly visible vacuum-fluorescent display for voltage and current meters
•
Tracking operation for ±25V outputs
•
Excellent load and line regulation and low ripple and noise
•
Operating states storage
•
Portable, ruggedized case with non-skid feet
Flexible system features
•
GPIB (IEEE-488) and RS-232 interfaces are standard
•
SCPI (Standard Commands for Programmable Instruments)compatibility
•
I/O setup easilydone fromfront-panel
Agilent E3631A Triple OutputDC PowerSupply
The Front Panel ata Glance
1 Meter and adjust selection keys
2 Tracking enable/disable key
3 Display limit key
4 Recall operating state key
5 Store operating state/Local key
6 Error/Calibrate key
1 Meter and adjust selection keys Select the output voltage and current of any one
supply (+6V, +25V, or -25V output) to be monitored on the display and allow knob
adjustment of that supply.
2 Tracking enable / disable key Enables / disables the track mode of ±25V supplies.
3 Display limit key Shows the voltage and current limit values on the display and
allows knob adjustment for setting limit values.
4 Recall operating state key Recalls a previously stored operating state from
location “1”, “2”, or “3”.
5 Store operating state / Local key
or “3” / or returns the power supply to local mode from remote interface mode.
6 Error / Calibrate key
and calibration / or enables calibration mode (the power supply must be unsecured
before performing calibration).
2
Displays error codes generated during operations, self-test
7 I/O Configuration / Secure key
interfaces / or secure and unsecure the power supply for calibration.
1
Stores an operating state in location “1”, “2”,
3
Configures the power supply for remote
8 Output On/Off key Enables or disables all three power supply outputs. This key
toggles between two states.
9 Control knob Increases or decreases the value of the blinking digit by turning
clockwise or counter clockwise.
10 Resolution selection keys Move the flashing digit to the right or left.
11 Voltage/current adjust selection key Selects the knob function to voltage control or
current control.
1
Thekey canbeused as the “Local”key when the powersupply is in the remote
interface mode.
2
You can enablethe “calibration mode” by holding down this key whenyou
turn onthe power supply.
3
Youcan use it as the “Secure”or“Unsecure” key when the power supplyis
in the calibration mode.
3
Front-Panel Voltage and CurrentLimit Settings
You can set the voltage and current limit values from the front panel
using the following method.
Use the voltage/current adjust selection key, the resolution selection keys, and
the control knob to change the monitoring or limiting value of voltage or current
.
Note
1 Press the key after turning on the power supply.
Display Limit
2 Set the knob to the voltage control mode or current control mode using the
voltage/current adjust selection key.
3 Move the blinking digit to the appropriate position using the resolution selection keys.
4 Change the blinking digit to the desired value using the control knob.
5 Press the key to enable the output. After about 5 seconds, the
display will go to the output monitor mode automatically to display the voltage
and current at the output.
Output On/Off
All front panelkeys andcontrols canbedisabled withremote interface commands. TheAgilent E3631Amust be in “Local” mode for the front panel keys and controls to function.
4
Display Annunciators
AdrsPower supply is addressed to listen or talk over a remote interface.
RmtPower supply is in remote interface mode.
+6VDisplays the output voltage and current for +6V supply. Knob is active for
+6V supply.
+25VDisplays the output voltage and current for +25V supply. Knob is active for
+25V supply.
-25VDisplays the output voltage and current for -25V supply. Knob is active for
-25V supply.
CALpower supply is in calibration mode.
TrackThe outputs of +25V and -25V supplies are in track mode.
LimitThe display shows the voltage and current limit value of a selected supply.
ERRORHardware or remote interface command errors are detected and also the
error bit has not been.
OFFThe three ouputs of the power supply are disabled.
UnregThe displayed output is unregulated (output is neither CV nor CC).
CVThe displayed output is in constant-voltage mode.
CCThe displayed output is in constant-current mode.
To review the display annunciators, hold down key as you
Display Limit
turn on the power supply.
5
The Rear Panelata Glance
1 Power-line voltage setting
2 Power-line fuse-holder assembly
3 AC inlet
Use the front-panel key to:
•
Select the GPIB or RS-232 interface (see chapter 3).
•
Set the GPIB bus address (see chapter 3).
•
Set the RS-232 baud rate and parity (see chapter 3).
power supply.Thischapter also provides instructions for checking yourpower supply, connectingto ac power,and selecting power-line voltage.
Initial Operation
rated outputs and properly responds to operation from the front panel.
Front-Panel Operation
front-panel keys and howthey are used tooperate the power supply from the front panel. This chapter also shows how to configure thepower supply for the remote interface and gives a brief introductionto the calibration features.
Remote Interface Reference
to help you program the power supply over the remote interface. Thischapter also explains how to program for status reporting.
Error Messages
are working with the power supply. Each listing contains information to helpyou diagnose and solve the problem.
Application Programs
applications to help you develop programs for your application.
Tutorial
gives specific details on the operation and use of the Agilent E3631Apower supplies.
Chapter 7 describes basic operation of linear power supplies and
Chapter 1 contains a general description of your
Chapter 2 ensures thatthepower supplydevelops its
Chapter 3 describes in detail the use of
Chapter 4 contains reference information
Chapter 5 lists the error messages that may appear as you
Chapter 6 containssome remote interface
Specifications
If you havequestions relating to the operation of thepowersupply,call
1-800-829-4444
Technologies SalesOffice.
Chapter 8 liststhe power supply’s specifications.
in the United States, or contact your nearest Agilent
Thischapter provides a general description of your power supply. Thischapter also contains instructions for initial inspection, location and cooling for bench and rackoperation, selecting the power-line voltage, and connecting your power supply to acpower.
14
Chapter 1 General Information
Safety Considerations
Safety Considerations
Thispower supplyis a Safety Class I instrument, which means that ithas a protective earth terminal. Thatterminal must be connected to earth ground through a powersource with a 3-wire ground receptacle.
Before installation or operation, check the powersupply and review thismanual for safety markings and instructions. Safety information for specific procedures islocated at the appropriate places in thismanual. See also
“
Safety” at the beginning of this manual for general safety information.
Safety and EMC Requirements
Thispower supplyis designed to comply with the following safety and EMC(Electromagnetic Compatibility) requirements:
•IEC 1010-1(1990)/EN 61010-1(1993): Safety Requirementsfor ElectricalEquipment for Measurement, Control, and Laboratory Use
•CSA C22.2 No.1010.1-92: Safety Requirements for Electrical Equipment for Measurement, Control, and LaboratoryUse
•UL 1244: Electrical and Electronic Measuring and Testing Equipment
•EMC Directive 89/336/EEC:Council Directive entitled Approximation of the Laws of the Member States relating to Electromagnetic Compatibility
•EN 55011(1991) Group 1, Class A/CISPR 11(1991): Limits and Methods of Radio Interference Characteristics of Industrial, Scientific, and Medical (ISM) Radio - Frequency Equipment
•EN 50082-1(1992) /
IEC801-2(1991): Electrostatic Discharge RequirementsIEC 801-3(1984):Radiated Electromagnetic Field RequirementsIEC801-4(1988): Electrical Fast Transient/Burst Requirements
•ICES/NMB-001ThisISM device complies with Canadian ICES-001.Cet appareil ISM est conforme à la norme NMB-001 du Canada.
1
15
Chapter 1 General Information
Options and Accessories
Options and Accessories
Options
Options 0EM, 0E3, and 0E9 determine which power-line voltage is selected at the factory.The standard unit isconfigured for 115 Vac ± 10%,47-63 Hz input voltage. Formore information about changing the power-line voltage setting, see ‘‘
Rack mount kit (Agilent part number 5062-3957)Extra English manual set (local language manual files are included
on the CD-ROM, Agilent part number 5964-8251.)
’’, starting on page 22 inthis chapter.
Accessories
The accessories listed below may be ordered from your localAgilentTechnologies Sales Office either with the power supplyor separately.
Agilent No.Description10833A
10833B34398A
34399A
16
GPIB cable, 1 m (3.3 ft.)GPIB cable, 2 m (6.6 ft.)RS-232, 9 pin (f) to 9 pin (f), 2.5 m(8.2ft.) cable; plus 9 pin (m)
to 25 pin (f) adapter RS-232 adapter kit (contains 4 adapters):9 pin (m) to 25 pin (m) for use with PCor printer9 pin (m) to 25 pin (f) for usewith PCor printer9 pin (m) to 25 pin (m) for use with modem9 pin (m) to 9 pin (m) for usewith modem
Chapter 1 General Information
Description
Description
The Agilent E3631A power supplyfeatures a combination of programming capabilities and linear power supply performance thatmakes it ideal forpower systems applications. The triplepower supply delivers0to ± 25V outputs rated at 0 to 1 Aand 0 to +6 V output rated at 0 to 5A. The ± 25Vsupplies also provide 0 to ± 25 V tracking output to power operationalamplifiers and circuits requiring symmetrically balanced voltages. The 0 to ± 25V outputs track each other within ±(0.2% output + 20mV) inthetrack mode. The ± 25V outputs can also be used in series as a single 0 to 50 V/1 Asupply.
The voltage and current of each supply can be adjusted independently fromthe front panel or programmed overthe GPIB or RS-232 interface. Using the front panel keys and the control knob, you can adjust the voltageand current of a selected output;enable or disable track mode; store andrecalloperating states;enable or disable three outputs;calibrate the powersupply including changing the calibration security; return the power supplyto local operatingmode; and configure the power supply for remoteinterface operation.
From the front-panel VFD (vacuum-fluorescent display),you canmonitor actual values ofoutput voltageand current (current limit values (supply from the annunciators, and check the type oferror from the displayed error codes (messages).
When operated over the remote interface, the power supply can be both a listener and a talker. Using an external controller, you can instructthe powersupply to set outputs and to send the status data back over the GPIB or RS-232. Readback capabilitiesinclude reading back output voltage and current; present and stored status; and error messages.The following functions are implemented over the GPIBor RS-232:
limit mode
),check the operating statusof the power
meter mode
) or voltage and
1
•Voltage and current programming
•Voltage and current readback
•Enable or disable track mode
•Present and stored status readback
•Programmingsyntax error detection
•Voltage and current calibration
•Output on or off
•Self-test
17
Chapter 1 General Information
Description
The front panel includes a VFD for displaying the output voltage andcurrent. Two 4-digit voltage and currentmeters accurately show the actual or limit values ofa selected supply simultaneously. Three meter selection keyschoosethe voltage and current of any oneoutput to be monitored on the display.
Connections to the power supply's output and to chassis ground are made to bindingposts on the front panel. The +25V and -25V supply's outputs share a common output terminal which is isolated from chassis ground. The positive and negative terminals of each output can begrounded, or each output can be kept within ±240 Vdc from the chassisground. The power supply isshipped with a detachable,3-wire grounding type power cord.The ac line fuse isan extractor type on the rear panel.
The power supply can be calibrated from the front panel directly or witha controller over the GPIB or RS-232 interface using calibration commands.Correctionfactors are storedinoutput programming. Calibration from the front panel or a controller eliminates the need to remove the top cover or even the need to remove the powersupply fromyour system cabinet. You can guard against unauthorized calibration by using the “Secured” calibration protection function.
non-volatile
memory and are used during
18
Chapter 1 General Information
Installation
Installation
Initial Inspection
When you receive your power supply, inspect it for any obvious damagethatmay have occurred during shipment. If any damage is found,notify the carrier and the nearest Agilent Sales Office immediately.Warranty information is shown in the front of this manual.
Keep the original packing materials in case the power supply has to be returned to Agilent Technologies in the future. If you return the power supply for service, attach a tag identifying the owner and modelnumber. Also include a brief description ofthe problem.
Mechanical Check
This checkconfirmsthat there are no broken keys or knob,that the cabinet and panel surfaces are free of dents and scratches,andthat the display is not scratched or cracked.
Electrical Check
Chapter 2 describes an initial operation procedure which, when successfully completed, verifies to a high level of confidence that the power supply isoperating in accordance with its specifications. Detailed electricalverification procedures are included in the
Service Guide
.
1
Cooling and Location
Cooling
The power supply can operate without loss of performance withinthe temperature range of 0 °Cto 40 °C,and with derated output current from40 °C to 55 °C. A fan cools the power supply by drawing air through the rear panel and exhausting it out the sides. Using an Agilent rack mount willnot impede the flow of air.
Bench Operation
Your power supply must be installed in a location that allowssufficient space at the sides and rear of the power supply for adequate air circulation.The rubber bumpers must be removed for rackmounting.
19
Chapter 1 General Information
Installation
Rack Mounting
The power supply can be mounted in a standard19-inch rack cabinet using one of three optional kits available. A rack-mounting kit for asingleinstrument is available as Option1CM (P/N 5063-9243). Installationinstructions and hardware are included with each rack-mounting kit. Any Agilent System II instrument of the same size can be rack-mounted beside the AgilentE3631A power supply.
Remove the front and rearbumpers before rack-mounting thepower supply.
FrontRear (bottom view)
To remove the rubber bumper, stretch a corner and then slide it off.
To rack mount a single instrument, order adapter kit 5063-9243.
20
Chapter 1 General Information
Installation
To rack mount two instruments of the same depth side-by-side, order
lock-link kit 5061-9694 and flange kit 5063-9214.
1
To install two instruments in a sliding support shelf, order support shelf
5063-9256, and slide kit 1494-0015.
21
Chapter 1 General Information
Input Power Requirements
Input Power Requirements
You can operate your powersupply froma nominal 100 V, 115 V,or230 Vsingle phase ac power sourceat47 to 63 Hz. An indication on the rear panel shows the nominal input voltage set for the powersupply at the factory.If necessary, you can changethe power-linevoltage setting according to theinstructions on the next page.
Power-Line Cord
The power supply is shipped from the factory with a power-line cord thathas a plug appropriate for your location. Contact the nearest Agilent Sales and Service Officeif the wrong power-line cord is included with your power supply. Your power supply isequipped with a 3-wire grounding type power cord; the third conductor being the ground. The power supplyis grounded onlywhen the power-line cord is plugged into an appropriate receptacle. Do not operate your power supply withoutadequate cabinet ground connection.
Power-Line VoltageSelection
Power-line voltage selection is accomplished by adjusting two components:power-line voltage selector and power-line fuse on the power-line module of the rear panel.To change the power-line voltage, proceed as follows:
22
Chapter 1 General Information
Input Power Requirements
1
1Remove the power cord.
fuse-holder assembly with a flat-blade screwdriver from the rear panel.
3
Rotate the power-line voltageselectoruntil
the correct voltage appears.
Remove the
2Install the correct line fuse. Remove the
power-line voltage selector fromthe power-line module.
4
Replace the power-line voltage selector and
the fuse-holderassembly in the rear panel.
23
24
2
2
Initial Operation
Initial Operation
There arethree basic tests inthis chapter. The automatic power-on testincludes a self-test that checks the internal microprocessors and allows the user visuallyto check the display.The output check ensuresthat the power supply develops its rated outputs and properlyresponds to operation fromthe front panel. For complete performance and/or verification tests, refer tothe
Service Guide
Thischapter isintended for both the experienced and the inexperienced user because itcalls attention to certain checks that should be made prior tooperation.
Throughout this chapter the key to be pressed is shown in the left margin.
.
26
Chapter 2 Initial Operation
Preliminary Checkout
Preliminary Checkout
The following steps help you verify that the power supply is ready for use.
1Verify the power-line voltage setting on the rear panel.
The power-line voltage isset to the proper value for your country when the power supply is shipped from the factory.Change thevoltage setting if itisnot correct. The settings are: 100, 115, or 230 Vac.
2Verify that the correct power-line fuse is installed.
The correct fuse isinstalled for yourcountry when the powersupply isshipped from the factory. For 100 or 115Vac operation, you must use a
2.5 AT fuse. For 230 Vac operation, you must use a 2.0 ATfuse.
3Connect the power-line cord and turn on your power supply.
The front-panel display will light up and apower-on self-test occurs automaticallywhen you turn on the power supply.
See“Power-LineVoltage Selection”, starting on page 22 in chapter1 if you need to changethepower-line voltage or the power-linefuse.
Toreplace the 2.5ATfuse, order Agilent part number2110-0913.To replacethe 2 ATfuse, order Agilent part number 2110-0982.
2
27
Chapter 2 Initial Operation
Power-On Checkout
Power-On Checkout
The power-on testincludesan automatic self-test that checksthe internalmicroprocessors and allows the uservisuallyto check the display.You willobservethe following sequence on the display after pressing the front panel power switch to on.
Output On/Off
1All segments of the display includingall annunciators will turn on
for about one second.
To review theannunciators, hold downthe key as you
Display Limit
turn onthe power supply.
2The GPIB address or RS-232 message will then be displayed for
about one second.
ADDR 5 (or RS-232)
The GPIBaddress isset to “5” when the power supply isshipped from the factory for remote interface configuration. If this isnot the first time thepower supply is turned on, a different interface (RS-232) or a different GPIBaddress may appear.
See ‘‘Remote Interface Configuration’’, for more information starting on page 48 in chapter3 if you need to change theremote interface configuration.
3The“OFF” and“+6V” annunciators are on. All others are off.
The power supply willgo into thedisabled (the supply (the +
OFF
annunciator turns on); the display is selected for the +6V
6V
annunciator turns on);and the knob isselected for
power-on / reset
state; all outputsare
voltage
control.
4Enable the outputs.
Note
Press the
Output On/Off
annunciator turns off and the
blinking
digit can be adjusted by turning the knob. Notice that the
key to enable the outputs. The
+6V
and CV annunciators are lit. The
OFF
displayis in themeter mode. “Meter mode”means that the display shows the actual output voltage and current.
If the power supplydetects anerrorduring power-on self-test, theannunciator willturn on.
See “ErrorMessages”, for moreinformation
starting on page 113 in chapter5
28
ERROR
Chapter 2 Initial Operation
Output Checkout
Output Checkout
Power
Output On/Off
The following procedures check to ensure that the power supply developsits rated outputs and properly responds to operation from the front panel.For complete performance and verification tests, refer to the
For eachstep, use the keys shownontheleft margins.
Service Guide
Voltage OutputCheckout
The following steps verify basic voltage functions with no load.
1Turn on the power supply.
The power supply willgo into thedisabled (the supply (the control.
2Enable the outputs.
The
OFF
blinking
in the meter mode. “Meter mode” means thatthe display shows the actual output voltage and current.
3Check that the front-panel voltmeterproperlyresponds to knob
control for the +6V supply.
Turn the knob clockwiseor counter clockwise to check that the voltmeter responds to knob control and the ammeter indicates nearly zero.
OFF
annunciator turns on); the display is selected for the +6V
+6V
annunciator turns on); and the knob isselected for
annunciatorturns off and the
digit can be adjusted by turning the knob. Notice that the display is
power-on / reset
+6V
and
CV
state; all outputs are
voltage
annunciators are lit. The
2
.
4Ensure that the voltage can beadjusted from zero to the maximum
rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob until the voltmeterindicates 6.0 volts.
1
Youcan use the resolution selection keys tomove theblinkingdigit tothe rightor
left when settingthe voltage.
1
29
Chapter 2 Initial Operation
Output Checkout
+25V
-25V
5Check the voltage function for the +25V supply.
Select the meter and adjust selection key for the +25V supply. The CVannunciator isstill lit and the
+25V
annunciatorwill turn on. Repeat steps
(3) and (4) to check the voltagefunction for the+25V supply.
6Check the voltage function for the -25V supply.
Select the meter and adjust selection key for the -25V supply. TheCVannunciator isstill lit and the
-25V
annunciator will turn on. Repeat steps
(3) and (4) to check the voltagefunction for the-25V supply.
30
Chapter 2 Initial Operation
Output Checkout
Current OutputCheckout
The following steps check basic current functionswith a short across theappropriate supply’s output.
Power
Output On/Off
Display Limit
Vol/Cur
1Turn on the power supply.
The power supply willgo into thedisabled (the supply (the control.
2Connect a short across (+) and (-) output terminals of the +6V
supply with an insulated test lead.
3Enable the outputs.
The
OFF
CC
annunciator islit depending on the resistanceof the test lead. The
blinking
in the meter mode. “Meter mode” means thatthe display shows the actualoutput voltage and current.
4Adjust the voltage limit value to 1.0 volt.
Set the display to the Adjust the voltagelimit to 1.0 volt to assure CC operation. The CCannunciator willlight.
5Check that the front-panel ammeter properlyresponds toknob control
for the +6V supply.
Set the knob to the clockwise when the display isin the meter mode (the off). Check that the ammeter responds to knob control and the voltmeter indicates nearly zero (actually,the voltmeter willshow the voltagedrop caused by the test lead).
OFF
annunciator turns on); the display is selected for the
+6V
annunciator turns on);and the knob isselected for
annunciator turns off and the
digit can be adjusted by turning the knob. Notice that the display is
limit
mode (the
current
power-on / reset
+6V
Lmt
annunciator will be
control, and turn the knob clockwiseor counter
state; all outputsare
annunciator turns on. TheCV or
blinking
Lmt
annunciator is
+6V
voltage
).
2
31
Chapter 2 Initial Operation
Output Checkout
6Ensure that the current canbe adjusted from zero to the maximum
rated value.
1
Adjust the knob until the ammeter indicates 0 amps and thenuntil the ammeter indicates 5.0 amps.
+25V
-25V
Note
7Check the current function for the +25V supply.
Disable the outputsby pressing the key and connect a
Output On/Off
short across (+) and (COM) output terminals of the ±25V supplywith an insulated test lead. Repeat steps (3) through (6)after selecting the meter and adjust selection key for the +25V supply.
8Check the current function for the -25V supply.
Disable the outputsby pressing the key and connect a
Output On/Off
short across (-) and (COM) output terminals of +25V supply with an insulated test lead. Repeat steps (3) through (6) after selecting the meter and adjustselection key for the -25V supply.
If anerror has been detectedduringthe output checkout procedures, the
ERROR
annunciator will turnon.See “ErrorMessages”, for more
information starting onpage 113 in chapter5
1
Youcan use the resolution selection keys tomove theblinkingdigit tothe rightor
left when settingthe current.
32
3
3
Front-Panel Operation
Front-Panel Operation
So far you have learned how to install your power supply and performinitial operation. During the initial operation, you were briefly introduced to operatingfrom the front panel as you learned how to check basic voltage and current functions. This chapter willdescribe indetailthe useof thesefront-panel keysand show how they are used to accomplish power supply operation.
•Front-Panel OperationOverview,page35
•Constant Voltage Operation,page36
•Constant Current Operation, page38
•Tracking Operation, page40
•Storing and Recalling Operating States, page41
•Disabling the Outputs,page43
•Knob Locking,page43
•System-Related Operations, page44
•Remote Interface Configuration,page48
•GPIB Interface Configuration,page53
•RS-232 InterfaceConfiguration,page54
•Calibration Overview,page58
Note
Throughout this chapter the key to be pressed is shown in the left margin.
See“ErrorMessages”, starting on page 113 chapter5 if you encounterany errors during front-panel operation
34
Chapter 3 Front-Panel Operation
Front-Panel Operation Overview
Front-Panel OperationOverview
The following section describes an overview of the front-panel keys beforeoperating your power supply.
•The power supply isshippedfromthe factory configured in the
front-panel
automatically set to operate in the front-panel operationmode.Wheninthis mode, thefront-panel keys can be used. When the power supply is in
remote
any time bypressingthe front-panel lockout command. A changebetween front-panel and remote operation modes will
•When you press the key (the blinks), the display of thepower supplygoes to the the present limit values of the selected supply willbe displayed. In this mode, you can also observe the change of the limit values when adjusting the knob. If you press the display time-outafter several seconds, the power supply will return the display to the the actual output voltage and current will be displayed.
•All outputs ofthe power supply can be enabled or disabled from the front panel using the is off, the
•The display provides the present operating status of the power supply with annunciatorsand alsoinforms the user of error codes. For example, the +6V supply is operating in CVmode and controlled fromthe front panel, then theCV and powersupply isremotely controlled, the on, and when the power supply is being addressed over GPIB interface,the
Adrs
for more information.
operation mode. At power-on, the power supply is
operation mode, you can return to front-panel operation mode at
keyif you did not previously send the
Local
not
result inachangein the output parameters.
Lmt
Display Limit
Display Limit
meter
mode (the
Output On/Off
OFF
annunciator turns on and the three outputs are disabled.
+6V
annunciator willturn on. See ‘‘
Lmt
annunciator turns off). In this mode,
key. When the output of the power supply
annunciators willturn on. If, however, the
Rmt
Display Annunciators
annunciator
limit
key again orlet the
annunciator willalso turn
mode and
’’ on page 5
3
35
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Constant Voltage Operation
Toset up the power supply for constant voltage (CV) operation, proceed asfollows.
1Connect a load to the desired output terminals.
With power-off, connect a load to the desired output terminals.
Power
Output
Display
On/Off
Limit
2Turn on the power supply.
The power supply willgo into the disabled (the supply (the
OFF
annunciator turns on); the display is selected for the +6V
+6V
annunciator turns on); and the knob isselected for
power-on / reset
state; all outputs are
control.
3Enable the outputs.
The
OFF
blinking
annunciator turns off and the
digit can be adjusted by turning the knob. Notice that the display is
+6V
and CV annunciators are lit. The
in the meter mode. “Meter mode” means thatthedisplayshows the actual output voltage and current.
Toset up the power supply for +25V supply or -25V supplyoperation, you should press the or key to select the display and adjust
+25V-25V
for +25V supply or -25V supply before proceeding to the nextstep.
4Set the display for the limitmode.
Notice that the limit mode. Whenthe display is inthe
Lmt
annunciator blinks, indicating that the display isin the
limit
mode, you can see the voltage
and current limit values of the selected supply.
In constant voltage mode,thevoltagevalues betweenthemetermodeand limit mode are the same, but the current values arenot. Further if thedisplay is inthe meter mode,youcannot see the change of current limit value whenadjusting the knob. Werecommendthat you should setthe display to “limit” mode to seethechange of current limit value in the constant voltage mode wheneveradjustingthe knob.
voltage
36
Chapter 3 Front-Panel Operation
Constant Voltage Operation
Vol
/
Cur
Vol
/
Cur
Display
Limit
5Adjust the knob for the desired current limit.
Check that the control.The second digitof ammeter will be
Lmt
annunciator still blinks. Set the knob for
blinking
1
current
. Adjust the knob to
the desired current limit.
6Adjust the knob for the desired output voltage.
Set the knob for
blinking
. Adjust the knob to the desired output voltage.
voltage
control. The second digit of the voltmeterwill be
1
7Return to the meter mode.
Press the key or let the display time-out after severalseconds to return to the meter mode. Notice that theturns off and the display returns to the meter mode. In the
Display Limit
Lmt
annunciator
meter
mode, the display shows the actual output voltage and current of theselected supply.
8Verify that the power supply is in the constant voltage mode.
If you operate the +6V supply in the constant voltage (CV) mode, verify that
CV
and
+6V
annunciators are lit. If you operate the power supply for the +25V supply or the -25V supply, the If the CC annunciator is lit, choose a
+25V
higher
or
-25V
annunciator willturn on.
current limit.
3
Note
During actual CV operation, if a load change causes the current limit to beexceeded, the power supplywill automatically crossover tothe constantcurrent modeat the presetcurrent limit and the output voltage willdrop proportionately.
1
Youcan use the resolution selection keys tomove theblinkingdigit tothe rightor
left when settingthe voltage and current.
37
Chapter 3 Front-Panel Operation
Constant Current Operation
Constant Current Operation
Toset up the power supply for constant current (CC)operation, proceed as follows.
1Connect a load to the output terminals of the desired supply.
With power-off, connect a load to the desired output terminals.
Power
Output On/Off
Display Limit
2Turn on the power supply.
The power supply will go into the (the
OFF
annunciator turns on); the display isselected for the
(the
+6V
annunciator turns on); and the knob is selected for
power-on / reset
state; all outputs are disabled
+6V
supply
voltage
3Enable the outputs.
The
OFF
blinking
annunciator turns off and the
digit can be adjusted by turning the knob. Notice that the display is
+6V
and CV annunciators are lit. The
in the meter mode. “Meter mode” means thatthedisplayshows the actual output voltage and current.
Toset up the power supply for +25V supply or -25V supplyoperation, you should press the
or key to select the display and adjust for +25V
+25V-25V
supply or -25V supply before proceeding to the nextstep.
4Set the display for the limitmode.
Notice that the limit mode. When the display is in the
Lmt
annunciator blinks, indicating that the display isin the
limit
mode, you can see the voltage and
current limit values oftheselectedsupply.
In constant current mode, the current values between the meter mode andlimit mode are the same, but the voltage valuesarenot.Furtherifthedisplay is in the meter mode, you cannot see the change of voltage limitvalue whenadjusting the knob. Werecommendthat you should setthe display to“limit” mode tosee the change of voltage limitvalue in theconstant current mode wheneveradjustingthe knob.
control.
38
Chapter 3 Front-Panel Operation
Constant Current Operation
Vol/Cur
Display Limit
Note
5Adjust the knob for the desired voltage limit.
Check that the knob isstill selected for voltage control and the
1
Lmt
annunciator blinks. Adjust the knob for the desired voltage limit.
6Adjust the knob for the desired output current.
Set the knob for
blinking
. Adjust the knob to the desired current output.
current
control. Thesecond digit of the ammeter willbe
1
7Return to the meter mode.
Press thekey orlet the display time-out after several secondsto return the meter mode. Notice that the display returnsto the meter mode. In the
Display Limit
Lmt
annunciator turns off andthe
meter
mode, the display shows the
actual output voltage and current of the selected supply.
8Verify that the power supply is in the constant current mode.
If you operate the +6V supply in the constant current (CC) mode,verify that
CC
and
+6V
annunciators arelit. If you operate the power supply for the +25Vsupply or the -25V supply, the If the CVannunciator is lit, choose a
+25V
or
higher
-25V
annunciator willturnon.
voltage limit.
During actualCCoperation, if aload change causes the voltage limit to be exceeded,the power supply will automatically crossover to constant voltagemode atthe presetvoltagelimit and the outputcurrent will dropproportionately.
3
1
You can use the resolution selection keysto move the blinkingdigit to the
right or left when settingthe voltage and current.
39
Chapter 3 Front-Panel Operation
Tracking Operation
Tracking Operation
Track
The ±25V supplies provide 0 to ±25 V tracking outputs. In the track mode,two voltages ofthe ±25V supplies track each other within ±(0.2% output +20mV) for convenience in varying the symmetrical voltagesneeded by operational amplifiers and other circuits using balanced positive and negativeinputs. The state of track mode isstored intrack isalways off statewhen power has been off or after a remote interfacereset.
Tooperate the power supply in the track mode, proceed as follows:
1Set the +25V supply to the desired voltage as described in previous
section“Constant Voltage Operation”(see information).
2Enable the track mode.
The key must be depressed for
Track
trackmode. When the track mode is first enabled,the -25V supply willbe set to the same voltage level as the +25V supply. Once enabled, any changeof the voltagelevel in either the +25V supply or the -25V supply will be reflected in other supply. The current limit is independentlyset for each ofthe +25V or the -25V supply and isnot affected by the track mode.
3Verify that the ±25V supplies track each other properly.
You can verify that the voltage of the -25V supply tracksthat of the +25V supply within ±(0.2%of output + 20 mV) from the front-panel display by comparing thevoltage values of the +25V supply and the -25V supply.
at least 1 second
volatile
page36 for detailed
memory; the
to enable the
Inthe track mode, if theforthe+25V supply, choose a highercurrent limit for the +25Vsupply.If the CCannunciator islit whenthe display isselected for the -25V supply, choose a highercurrentlimit for the -25V supply.
40
CC
annunciator is litwhen the display is selected
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Storing and Recalling Operating States
Store
You can storeup to three different operating states in Thisalso enablesyou torecallthe entireinstrument configuration with justa few key presses fromthe front panel.
Thememory locations are supplied from the factory for front panel operationwith the following states:
output
;
*RST
values ofvoltage and current limits for three outputs; output
disabled
and 0 V and 1 A for the ±25V supplies.The following steps show you how to store and recall an operating state.
1Set up the power supply for the desired operating state.
The storage feature “remembers” the display and knob selection state, the limit values of voltage and current for three outputs, outputon/off state, and track on/off state.
2Turn on the storage mode.
Three memory locations (numbered 1, 2 and 3) are availableto store the operating states. The operating states are stored in are remembered when being recalled.
This messageappears on the display for approximately 3 seconds.
; and
STORE1
track off state
.
*RST
display and knobselection for+6V
values for +6V supply are 0 V and 5 A
non-volatile
non-volatile
memory.
memory and
3
3Store the operating state in memory location “3”.
Turn the knob to the right to specifythememory location 3.
STORE3
To cancel the store operation,let the display time-out after about 3 seconds or pressany other functionkey except the key. The power supply returns to the normal operating modeand to the function pressed.
Store
41
Chapter 3 Front-Panel Operation
Storing and Recalling Operating States
Store
Recall
4Save the operating state.
The operating stateis now stored. To recall the stored state, go to the following steps.
DONE
This messageappears on the display for approximately 1 second.
5Turn on the recall mode.
Memory location “1” willbe displayed in the recallmode.
RECALL1
This messageappears on the display for approximately 3 seconds.
6Recall thestored operating state.
Turn the knob to the right to changethe displayed storage location to “3”.
RECALL3
If thissetting isnot followed within3 seconds with a key
Recall
stroke, the power supply returns to normaloperating modeand willnot recallthe instrument state 3 from memory.
Recall
7Restore the operating state.
The power supply should now be configured inthe samestate aswhen you stored the state on the previous steps.
DONE
This messageappears on the display for approximately 1 second.
42
Chapter 3 Front-Panel Operation
Disabling the Outputs
Disabling the Outputs
The outputs of the power supply can be disabled or enabled from the front panel using the
•When the power supply is in the “Off” state, theOFFand the three outputs are disabled. TheOFFpowersupply returnsto the “On”voltage value is 0 volts and the current valueis 0.05 amps.
•The output state isstored inwhen power has been off or after a remote interfacereset.
While theoutputs are disabled, the control knob and resolutionselection keys are still working. If the display is in the meter mode, you cannot see thechanges of output voltage and current settings onthedisplay whenturning the knob. To see or check the changeswhentheoutputsare disabled,the display should be in thelimitmode.
•Front-panel operation:
You can disable the outputsby pressing the key toggles between the output Off and On states.
•Remote interface operation:
OUTPut {ON|OFF}
The outputs are disabled when the “OFF” parameter isselected and enabled when the “ON” is selected.
Output On/Off
volatile
key.
annunciator turns on
annunciator turns off when the
state. When the outputs are disabled, the
memory; the output is alwaysdisabled
Output On/Off
key. This
3
KnobLocking
The knob locking function can be used to disable the knob,thereby preventing anyunwanted changes during an experiment, or when you leavethe power supply unattended.
To disable the knob, pressthe resolutionselection key until the blinking digit disappears.
43
Chapter 3 Front-Panel Operation
System-Related Operations
System-Related Operations
Thissection gives information on topics such as self-test, error conditions, and front-panel display control.This information is not directlyrelated tosettingup thepower supply but isan important part of operating the power supply.
Self-Test
Apower-onsupply. Thistest assures you that thepower supply is operational. This testdoes not perform the extensive set of tests that are included as part of the complete self-testdescribed below.If the power-on self-test fails, the
ERROR
•A
complete
seconds to execute. If all tests pass, you can have a high confidence that the power supply is operational.
•If the If the self-test fails, “FAIL”on. See the Agilent Technologiesfor service.
•Front-panel operation:
Thecompleteany front panel keys except the key) and the power-line switch simultaneously and then continuing to pressthe key for 5 seconds. The complete self-test willbe finished in2 more seconds.
self-test occurs automatically when youturnon the power
annunciator turns on.
self-test performs a series of tests and takes approximately2
complet
eself-test is successful, “PASS” is displayed on the front panel.
is displayed andthe
Service Guidefor instructionsonreturning the powersupply to
self-test is enabled by pressing the key (actually
Error
ERROR
Recall
Recall
annunciator turns
•Remote interface operation:
*TST?
Returns “0” if the complete self-test passes or “1” if it fails.
44
Chapter 3 Front-Panel Operation
System-Related Operations
Error Conditions
When the front-panelsyntax or hardware errors havebeen detected. A record of up to 20errorscan be stored inthepower supply's error queue.
Messages”, starting on page 113for a completelisting of the errors.
•Errorsare retrieved in first-in-first-out (FIFO) order.The first error returned is the first error that was stored. When you haveread all errors from the queue, the each timean error isgenerated.
ERROR
ERROR
annunciator turnsoff. The power supply beepsonce
annunciator turns on, one or more command
See chapter 5“Error
•If more than 20errors have occurred when you operate the power supplyover the remote interface, the last error storedin the queue (the most recent error) is replaced with stored until you remove errors from the queue. If no errorshave occurred when you read the errorqueue, the powersupply responds with +0, “
error
” over the remote interface or“NO ERRORS” fromthe front panel.
•The error queue iscleared when powerhas been off or after a status) command has been executed. The clear the error queue.
•Front-panel operation:
If the
ERROR
the errorsstored in the queue. All errors are cleared when you read all errors.
•
Remote interface operation:
SYSTem:ERRor?
Errors have the following format (the error string may contain up to 80 characters).
annunciator ison, press the key repeatedlyto read
ERROR -113
-350, “Toomany errors”.
*RST
Error
Readsone error from the errorqueue
No additional errorsare
No
*CLS
(reset) command
(clear
does not
3
-113,"Undefined header"
45
Chapter 3 Front-Panel Operation
System-Related Operations
Display Control
For security reasons, you may want to turn off thefront-panel display.Fromthe remote interface, you can display a12-character message on the front panel.
Thedisplay can beenabled/disabled from the remote interface only.
•When the display isturned off, outputs are not sent to the display and allannunciators are disabledexcept the operation isotherwise unaffected by turning off the display.
ERROR
annunciator. Front-panel
•The display state isstored in
volatile
memory; the display is always enabled when power has been off, after a remote interface reset, or after returning tolocal from remote.
•You can display a messageon thefront panel bysending a command from the remote interface. The power supply can display up to12 characters ofthe message on the front panel; any additional characters are truncated.Commas,periods, and semicolons share a display space with the precedingcharacter, and
are not
considered individual characters. When a message is
displayed, outputs are not sent to the display.
•Sending a message to the display fromthe remoteinterface overrides the display state; this means that you can display a message even if the display isturned off.
The display state isautomatically turned on when you returnto the local(front panel) operation. Pressthe key to return to the local state
Local
from the remote interface
•Remote interface operation:
DISPlay {OFF|ON}
DISPlay:TEXT
<quoted string>
DISPlay:TEXT:CLEar
Disable / enable the display
Display the stringenclosed in quotes
Clear the displayed message
The following statement shows how to displaya message on the front panelfrom a Agilent Technologies controller.
"DISP:TEXT ’HELLO’"
46
Chapter 3 Front-Panel Operation
System-Related Operations
Firmware Revision Query
The power supply has three microprocessors for control ofvarious internalsystems.You can querythe power supply to determine which revision of firmwareis installed for each microprocessor.
Youcan query the firmware revisionfrom the remote interface only.
•The power supply returns four fields separated by commas and the fourth field isarevision code which contains threenumbers. The first number isthe firmware revision number for the mainprocessor; the second isfor theinput/output processor;and the third isfor the front-panel processor.
•Remote interface operation
*IDN?
Besure to dimension a string variable with atleast 40 characters.
Returns
“
HEWLETT-PACKARD,E3631A,0,X.X-X.X-X.X
”
SCPI Language Version
The power supply complieswith the rules and regulationsof the present version ofSCPI (Standard Commands for Programmable Instruments). You can determine theSCPI version with which the power supply isin complianceby sendinga command from the remote interface.
You can query the SCPI version fromthe remote interface only.
•Remote interface operation:
SYSTem:VERSion?
Returns astring intheform “YYYY.V” where the “Y’s” represent the year of the version,and the “V” represents a version numberfor that year (for example, 1995.0).
3
47
Chapter 3 Front-Panel Operation
Remote Interface Configuration
Remote Interface Configuration
Beforeyou can operatethe power supply over the remoteinterface, youmust configure the power supply for the remote interface. This section gives information on configuringthe remote interface. For additional informationon programming the power supply over the remote interface,
“Remote InterfaceReference”,
starting on page 63.
See chapter
4
Remote InterfaceSelection
The power supply is shipped with both an GPIB (IEEE-488) interface and an RS-232 interface on the rear panel. Only oneinterface can beenabled ata time. The
GPIB interface
is selected when the powersupply isshipped
from the factory.
The remote interface canbeselected from the front-panel only.
•The interface selection is stored in
non-volatile
memory, and
does not
changewhen power has been off or after a remote interfacereset.
•If you select the GPIB interface, you must selecta unique address for the power supply. The current address isdisplayed momentarily on the front panel when you turn on the power supply.
1
•Your GPIB bus controller has its own address. Be sure to avoid using the bus controller’s address for any instrument on the interfacebus. Agilent Technologiescontrollers generally use address
“21”
.
•If you enablethe RS-232 interface, you must select the baud rate and parityto be used. “RS-232” is displayed momentarily on the front panel when you turn on the power supply ifyou have selected thisinterface.
2
1
Refer to"GPIB Interface Configuration" starting on page 53 for moreinformation on connectingthe power supply toa computer overtheGPIBinterface.
2
Referto "RS-232 Interface Configuration" starting on page 54 formoreinformation on connectingthe power supply toa computer over the RS-232 interface.
48
Chapter 3 Front-Panel Operation
Remote Interface Configuration
GPIB Address
Each device on the GPIB (IEEE-488) interface musthavea unique address.You can set the power supply’s addressto any value between0 and 30.The current address is displayed momentarily on the frontpanel when you turnon thepower supply.The address is set to “05” when the power supply isshipped from the factory.
The GPIBaddresscan be set from the front-panel only.
•The address is stored inpower has been off or after a remote interfacereset.
•Your GPIB bus controller has its own address. Be sure to avoid thebuscontroller’s address for anyinstrument on the interface bus.Agilent Technologiescontrollers generally use address“21”.
non-volatile
memory,and
does not
change when
Baud Rate Selection (RS-232)
You can selectone of six baud rates for RS-232operation. The rate isset to
9600 baudThebaud ratecan be set from the front-panel only.
•Select oneof the following: 300, 600, 1200, 2400, 4800,
factory settingis 9600 baud
•The baud rate selection isstored in changewhen power has been off or after a remote interfacereset.
whenthe power supply isshipped from the factory.
9600
.
non-volatile
memory,and
baud.
does not
The
Parity Selection (RS-232)
You can select the parity for RS-232 operation. The power supply isconfigured for
Theparity can be set from the front-panel only.
•Select oneof the following: Odd (number of data bits.
no parity and 8 data bits
7 data bits
). When youset the parity,you are indirectly setting the
when shipped fromthe factory.
None (8 data bits
), Even (
7 data bits
), or
3
•The parity selection is stored in changewhen power has been off or after a remote interfacereset.
non-volatile
memory,and
does not
49
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the GPIB Address
Toconfigurethe power supply for the GPIB interface, proceed as follows:
I/O Config
I/O Config
I/O Config
1Turn on the remote configuration mode.
GPIB / 488
You will see the above message on the front-panel display if the power supply has not been changed from the defaultsetting. If “RS-232” appears,choose “GPIB / 488” by turning the knob to the right.
2Select the GPIB address.
ADDR05
The address isset to “05” when the power supply isshipped from thefactory. Notice that a different GPIBaddress may appear if the powersupply has been changed from the default setting.
3Turn the knob to change the GPIB address.
The displayed addressis changed when turning the knob to the right or left.
4Save the change and turn off theI/O configuration mode.
CHANGE SAVED
Note
The addressis stored in power has been off or after a remote interfacereset. The power supply
non-volatile
memory,and
does not
change when
displaysamessage to show that the changeis now in effect. If the GPIB address is notchanged, “NO CHANGE” willbe displayed for one second.
To cancel the I/O configuration modewithout any changesduring the GPIB address selection, press the “I/O Config” key untilthe“NO CHANGE” message is displayed.
50
Chapter 3 Front-Panel Operation
Remote Interface Configuration
To Set the Baud Rateand Parity (RS-232)
Toconfigurethe power supply for the RS-232 interface, proceed as follows:
I/O Config
I/O Config
1Turn on the remote configuration mode.
GPIB / 488
You willsee the above message on the display if the power supply has not been changed from the default setting.
Notice that if you changed the remote interface selection to RS-232 before,“RS-232” message will be displayed.
2Choose the RS-232 interface.
RS-232
You can choose the RS-232 interface by turning the knob to the left.
3Select the RS-232 interface and choose the baud rate.
9600 BAUD
The rate isset to factory. Choose fromone ofthe following by turning the knob to the right or left: 300, 600, 1200, 2400, 4800, or
9600
baud when the power supply isshipped from the
9600
baud.
3
I/O Config
4Save the change and choose the parity.
NONE 8 BITS
The power supply isconfigured for 8data bits with no parity when shipped from the factory. Choose fromone of thefollowing byturning the knob tothe right or left: parity,you are indirectly setting the number of the data bits.
None 8 Bits
, Odd 7 Bits, or Even 7Bits. When you set
51
Chapter 3 Front-Panel Operation
Remote Interface Configuration
I/O Config
Note
5Save the change and turn off theI/O configuration mode.
CHANGE SAVED
The RS-232 baud rate and parity selectionsare stored inmemory, and
does not
change when power has been off oraftera remote
non-volatile
interface reset. The power supply displaysa message to show that thechange isnow in effect. If the baud rate and the parity are not changed, “NO CHANGE” willbe displayed for one second.
To cancel the I/O configuration modewithout any changesduring the baud rate and parity selection,pressthe“I/O Config” key until the “NO CHANGE” message is displayed.
52
Chapter 3 Front-Panel Operation
GPIB Interface Configuration
GPIB Interface Configuration
The GPIBconnector on the rear panelconnects your power supply tothe computer and other GPIB devices. Chapter 1 lists the cables that are available from Agilent Technologies. AnGPIB system can be connected together inany configuration (star, linear, or both) as long as the following rules are observed:
•The total number of devices including the computer isno more than 15.
•The total length of all the cables used is no more than 2 meter times the number ofdevices connected together, up to amaximum of 20 meters.
3
Note
IEEE-488statesthat you should exercise caution if your individual cable lengthsexceed 4 meters
Do not stack morethan three connector blocks together on any GPIBconnector. Make sure thatall connectors are fully seated and thatthe lock screwsare firmly finger tightened.
53
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
RS-232 Interface Configuration
You connectthe power supply to the RS-232 interface using the 9-pin (DB-9)serial connector on the rear panel. Thepower supplyis configured as a DTE(Data Terminal Equipment) device. For all communications over the RS-232interface, thepower supply uses two handshake lines:DTR (Data Terminal Ready, on pin 4) and DSR (
The following sectionscontain information to help you use the power supply over the RS-232interface. The programming commands for RS-232areexplained on
page87.
RS-232 ConfigurationOverview
Configure the RS-232 interfaceusing the parameters shown below. Usethe front-panel key to selectthe baud rate, parity, andnumber ofdatabits (see page51 for more information to configure fromthe front panel).
I/O Config
Data Set Ready
, on pin 6).
•Baud Rate: 300, 600, 1200, 2400, 4800, or
•Parity and Data Bits:
•Number of Start Bits:
•Number of Stop Bits:
None / 8 data bits (factory setting
Even /7 data bits, orOdd / 7 data bits
1 bit (fixed
2 bits (fixed
)
)
9600
baud (
factory setting
)
)
RS-232 Data Frame Format
A charactercharacter. The frame isdefinedas the characters from the last
stop bit
number of data bits, and parity type. The power supply uses the followingframe formatsfor seven and eight data bits.
54
frame
consists of all thetransmitted bits that make up a single
start bit
, inclusively. Within the frame, you can select the baud rate,
to the
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
Connectionto a Computer or Terminal
Toconnect the power supply to a computer or terminal, you must havethe proper interface cable.Most computers and terminals are DTE (
Terminal Equipment
) devices.Since thepower supply is also a DTE device,
you mustuse a DTE-to-DTEinterface cable. These cables are also called
null-modem, modem-eliminator
, or
crossover
cables.
The interface cable must also havethe proper connector on each end and the internal wiring must be correct. Connectors typically have 9 pins (DB-9 connector) or 25 pins (DB-25connector) with a “male” or “female” pin configuration. A male connector has pins insidethe connector shell and a femaleconnector has holes inside the connector shell.
If you cannotfind the correct cable for your configuration, you may have to use a
wiring adapter
. If you are using aDTE-to-DTE cable, make sure the adapter isa“straight-through” type. Typical adapters include gender changers, null-modem adapters,and DB-9to DB-25 adapters.
If your configuration isdifferent than thosedescribed, order the Agilent 34399A, 34399AThis kit containsadaptersfor connection to other computers,terminals,and modems. Instructions and pin diagrams are included with the adapter kit.
DB-9 Serial Connection
If your computer or terminal has a 9-pin serial port with a male connector, use the null-modem cableincluded with the
Agilent 34398A Cable Kit
. This cable has a 9-pin female connector on each
end. The cable pin diagram is shown below.
Data
Adapter Kit
3
.
Instrument
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
1
2
3
4
5
6
7
8
9
DB9
Female
5182-4794
Cable
1
2
3
4
5
6
7
8
9
DB9
Female
PC
DCD
RX
TX
DTR
GND
DSR
RTS
CTS
RI
DB9
Male
55
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
DB-25 Serial Connection
If your computeror terminal has a 25-pin serialport with a male connector, use the null-modem cable and 25-pin adapter included withthe
Agilent 34398ACable Kit
. The cableand adapter pin
diagram are shown below.
5182-4794
InstrumentPC
DCD
1
RX
2
TX
3
DTR
4
5
GND
6
DSR
7
RTS
8
CTS
9
RI
DB9
DB9
Female
Male
Cable
1
2
3
4
5
6
7
8
9
DB9
Femal e
1
2
3
4
5
6
7
8
9
DB9
Male
5181-6641
Adapter
2
3
4
5
6
7
8
20
DB25
Female
TX
RX
RTS
CTS
DSR
GND
DCD
DTR
DB25
Male
DTR/DSR Handshake Protocol
The power supply isconfigured as a DTE (device and uses the DTR (
Data Terminal Ready
lines of the RS-232interface to handshake. The power supply uses the DTRline to send a hold-off signal. The DTR line must be TRUE before the power supply will acceptdatafrom the interface. When the power supply sets the DTR line FALSE, the data must ceasewithin 10 characters.
To disable the DTR/DSRhandshake,DSR line to logic TRUE. If you disable the DTR/DSR handshake, also select a slower baud rate to ensure that the data istransmittedcorrectly.
Thepowersupply sets the DTR lineFALSE in the following cases:
Data Terminal Equipment
) and DSR(
do not
connect the DTR lineand tie the
Data Set Ready
)
)
1
When the power supply’s input buffer isfull (when approximately 100 charactershave been received), it sets the DTR line FALSE (pin4 on the RS-232 connector). When enough characters have been removed to makespace in the input buffer, the power supply sets the DTR line TRUE, unless the second case (see next) prevents this.
56
Chapter 3 Front-Panel Operation
RS-232 Interface Configuration
2
When the power supply wants to “talk” over the interface (which means thatit has processeda query) and hasreceived a <terminator,it will set the DTR line FALSE. This implies that once a query hasbeen sentto the power supply, the bus controller should read the responsebefore attempting to send more data. It alsomeans that a <terminate the command string. After the response has been output, the powersupply sets the DTR line TRUE again, unless the first case(seeabove) prevents this.
The power supply monitorsthe DSR line to determine when the bus controller is ready to accept dataover the interface. The power supplymonitors theDSR line (pin 6 on the RS-232connector) before eachcharacter is sent. The output is suspended if the DSR line isFALSE. Whenthe DSR line goes TRUE, transmission will resume.
The power supply holds the DTR line FALSEwhile output issuspended. A form of interfaceline TRUE to allow the power supply to complete the transmission. You can break the interface deadlock by sending the <clears the operation in progress and discards pending output (this isequivalentto the IEEE-488 device clear action).
deadlock
exists untilthe bus controller asserts the DSR
new line>
Ctrl-C>
character, which
message
new line>
must
3
For the <Ctrl-C> character toberecognized reliably bythe powersupplywhileit holds DTRFALSE,the bus controller must first set DSRFALSE.
RS-232 Troubleshooting
Here are a few things to check if you arehaving problems communicating over the RS-232interface. Ifyou need additional help, refer to the documentation that camewith your computer.
•Verify that the power supply and your computer are configured for the same baud rate, parity, and number of data bits. Make sure that yourcomputer isset up for on the power supply).
•Make sure to execute the power supply in the remote mode.
•Verify that you haveconnected the correct interfacecable and adapters.Even if the cable hasthe proper connectors for yoursystem,the internalwiring may be incorrect.Theconnect the power supply to most computers or terminals.
•Verify that you haveconnected the interface cable to thecorrect serialport on your computer (COM1, COM2, etc.).
1 startbit
and
2 stop bits
SYSTem:REMote
Agilent 34398A Cable Kit
(these values are fixed
command to place the
can be used to
57
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Overview
Thissection gives an overview of the calibration features of the power supply. For more detailed discussion of the calibration procedures, seethe
Service Guide
.
Calibration Security
This feature allowsyou to enter a security code to prevent accidental or unauthorizedcalibrations of the power supply. When you first receiveyour power supply, itis secured. Before you can calibrate thepower supply, you must unsecure it by entering thecorrect security code.
•
The security code is set to “HP003631” when the powersupply isshipped from the factorydoes not
•To secure the powersupply fromthemay contain up to 12alphanumeric characters as shown below. Thefirst character numbers. You do not haveto use all 12 characters but thefirst character must always be a
•To secure the powersupply fromtheunsecured from the front panel,use the eight-character format shown below. The first two charactersmust be “HP”and the remaining characters mustbe numbers.Onlythe last six characters are recognizedfrom the front panel, but all eight characters are required.
the power supply fromthefront panel,omit the “H P”and enter the remaining numbers as shown onthe following
change when power has been off or after a remoteinterface reset.
A _ _ _ _ _ _ _ _ _ _ _
. Thesecurity code is stored in
mustbe a letter
letter
, but the remaining characters can be lettersor
.
(12 characters)
non-volatile
remote interface
remote interface
memory, and
, the security code
so that it can be
To unsecure
H P _ _ _ _ _ _ (
If you forget your securitycode,you can disable the security featureby adding a jumper insidethepower supply,and thenentering a new code.See theService Guide for more information.
58
8 characters
)
Chapter 3 Front-Panel Operation
Calibration Overview
To Unsecure for Calibration
calibration either from the front panel or over the remote interface.
power supply issecured when shipped from the factory, and the security codeis set to “HP003631”
•Front-Panel Operation
You can unsecure the power supply for
.
The
SECURED
If the power supply issecured, you will see theabove message for one second by holding thethe power supply.To unsecure the power supply, press the key after the “CALMODE” message isdisplayed in the calibration mode,enter the security codeusing the knob and resolution selection keys,and then press the key.
Secure
Calibrate
key for 5 secondswhen you turn on
Secure
000000 CODE
When you press the key to save the change,you willsee the message below for one second if the security code iscorrect. The unsecured setting is stored in when power has been off or after a remote interfacereset.
calibration mode, turn the power off and on.Noticethat if the security isincorrect, thepowersupply returns to the
code entering mode for you to enter the correct code.
Secure
non-volatile
memory, and
does not
To exit the
change
3
UNSECURED
•Remote Interface Operation:
CALibrate:SECure:STATe, {OFF|ON},
Tounsecurethe power supply, send the abovecommand with the same code used to secure. For example,
<code>
"CAL:SEC:STAT OFF, HP003631"
59
Chapter 3 Front-Panel Operation
Calibration Overview
To Secure Against Calibration
calibration either from the front panel or over the remote interface.
You can secure the powersupply against
Thepower supply issecured when shipped from the factory, and the security codeis set to “HP003631”.
Besure to read the securitycode rules onpage58 beforeattempting to secure the power supply.
•Front-Panel Operation:
UNSECURED
If the power supply isunsecured, you will seethe above message for onesecond by holding the
Calibrate
the power supply.To secure the power supply, press the key after the “CAL MODE” message is displayed inthecalibration mode, enter the security codeusing the knob and resolution selection keys,and then press key.
Secure
key for 5 seconds when you turn on
Secure
Noticethat you shouldomit the “H P”and enter the remaining numbersas shown below.
000000 CODE
When you press the key to save the change,you willsee the message below. The secured setting isstored in and
does not
change when power has been off or after a remote
Secure
non-volatile
memory,
interface reset. To exit the calibration mode, turn the power off and on.
SECURED
•Remote Interface Operation:
CALibrate:SECure:STATe {OFF|ON},
<code>
Tosecure thepower supply, send the above command with the same codeas used tounsecure. For example,
"CAL:SEC:STAT ON, HP003631"
60
Chapter 3 Front-Panel Operation
Calibration Overview
To Change the Security Code
first unsecure the power supply, and then enter a new code.
Besure to read the securitycode rules onpage58 beforeattempting to secure the power supply.
•Front-Panel Operation:
Tochangethe security code, first make sure thatthepower supplyis
unsecured
displayed in the calibration mode, enter the new security codeusing the knob and resolution selection keys, then press the key.
Changingthecode from the front panelalso changesthecode requiredfrom the remoteinterface.
•Remote Interface Operation:
CALibrate:SECure:CODE
To change the security code,first unsecure the powersupply using the old security code. Then, enter the new code. For example,
"CAL:SEC:STAT OFF, HP003631"
"CAL:SEC:STAT ON, ZZ00143"
. Press the
"CAL:SEC:CODE ZZ001443"
Secure
To change the security code,you must
key after the “CAL MODE” messageis
Secure
<new code>
Unsecurewith old codeEnter new codeSecurewith new code
3
61
Chapter 3 Front-Panel Operation
Calibration Overview
Calibration Count
You can determine the number of times that your power supply hasbeen calibrated. Your power supply was calibrated before it left thefactory.When you receiveyour power supply,read the countto determineits initial value.
The calibration countfeature canbeperformed from the remoteinterface only.
•The calibration count is stored in
changewhen power has been off or after a remote interfacereset.
•The calibration count increments up to amaximum of 32,767 after which
it wraps-around to 0. Since the value increments by one for each calibration point, acomplete calibration willincrease the valueby 6 counts.
•Remote Interface Operation:
CALibrate:COUNt?
non-volatile
memory, and
doesnot
Calibration Message
You can use the calibrationmessage feature to record calibration information about your power supply. For example, you can store such information as the last calibration date, the nextcalibration due date, the power supply's serial number, or even the name and phone number of the person to contact for a new calibration.
Youcan recordand read informationin the calibration messagefrom the remote interface only.
•The power supply should be unsecured before sending a calibration
message.
•The calibration message may contain up to40 characters.
•The calibration message is stored in
changewhen power has been off or after a remote interfacereset.
•Remote Interface Operation:
CALibrate:STRing
The following command stringshows how to store a calibration message.
"CAL:STR 'CAL 05-1-95'"
62
<quoted string>
non-volatile
Store the calmessage
memory, and
does not
4
4
Remote Interface Reference
Remote Interface Reference
• SCPICommand Summary, page 65
• Simplified ProgrammingOverview, page 70
•Using the APPLy Command, page 73
•Output Setting and Operation Commands, page74
•Triggering Commands, page 79
•System-Related Commands, page 82
•Calibration Commands, page 85
•RS-232 Interface Commands, page 87
•The SCPI StatusRegisters, page 88
•Status Reporting Commands, page 98
•An Introduction to the SCPI Language, page 102
•Halting an Output in Progress, page107
•SCPI Conformance Information, page108
•IEEE-488 Conformance Information, page111
If youarea first-time userof the SCPI language, you may wantto refer to these sections to become familiar withthelanguage beforeattempting toprogram the power supply.
64
Chapter 4 Remote Interface Reference
SCPI Command Summary
SCPI CommandSummary
This section summarizes the SCPI (
Instruments
remote interface. Refer to the later sections in thischapter for more complete details on each command.
Throughout this manual, the following conventions areused for SCPIcommand syntax.
•Square brackets ([ ]) indicate optional keywords or parameters.
•Braces ({ }) enclose parameters within a command string.
•Triangle brackets (< >) indicate that youmust substitute a value or a code for the enclosed parameter.
•A vertical bar ( | ) separates oneof two or more alternativeparameters.
First-time SCPI users,see
) commandsavailable to programthe power supply overthe
Thissection gives an overview of the basic techniques used to programthepower supply over the remote interface. This section is only an overview and does not give all of the details you will need to write your own application programs.Refer to the remainder of this chapter and also chapter 6, Application Programs, for more details and examples. Also refer to the programming reference manual thatcame with your computer for detailson outputting command strings and entering data.
Using the APPLy Command
The
APPLy
command provides the most straightforward method toprogram the power supply over the remote interface. For example, the following statement executed fromyour computer will setthe +6V supply toan output of 3 V rated at1 A:
"APPL P6V, 3.0, 1.0"
Using the Low-LevelCommands
Although the program the power supply, the low-level commands give you more flexibility to change individual parameters. For example, the following statementsexecuted from your computer willset the +6V supply to an output of 3 Vrated at 1 A:
APPLy
command provides the most straightforward method to
"INST P6V"
"VOLT 3.0"
"CURR 1.0"
70
Select +6V output
Set output voltage to 3.0 V
Set output current to 1.0 A
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Reading a Query Response
Onlythe query commands (commands that end with “?”) will instruct the power supply to send a response message. Queries return either outputvalues or internalinstrument settings. For example, the following statements executed from your computer willread the power supply's error queue and print the most recent error:
dimension statement
"SYST:ERR?"
bus enter statement
print statement
Dimension string array (80 elements)
Read error queu
Enter error string into computer
Print error string
Selecting a Trigger Source
The power supply willaccept a “bus” (software)trigger or animmediate internal trigger as a trigger source. By default, the “selected. If you want the power supply to use an immediate internal trigger, you must select“executed from your computer willset the +6V supply to an output of 3 V/1 A immediately:
"INST P6V"
"VOLT:TRIG 3.0"S
"CURR:TRIG 1.0"
"TRIG:SOUR IMM"
"INIT"
IMMediate
”. For example, the followingstatements
Select the +6V output
etthetriggered voltage level to 3.0 V
Set the triggered current level to 1.0 A
Select the immediate trigger as asource
Cause the trigger system toinitiate
BUS
” trigger source is
4
71
Chapter 4 Remote Interface Reference
Simplified Programming Overview
Programming Ranges and Output Identifiers
Output setting commands requirea parameter for programming ranges and an output nameor an output number as the identifier ofeach output and most queries will return a parameter. The programmingrange for aparameter varies according to the selected output of the power supply. The following table lists the programming ranges, output names,and output numbers for each output.
Refer to this table to identify parameters whenprogrammingthe power supply.
Table 4-1. Agilent E3631A Programming Ranges and Output Identifiers
Output
+6V output+25V output-25V output
VoltageProgramming
CurrentProgramming
Output identifierP6VP25VN25V
Output number123
Range
MAX value6.18 V25.75 V-25.75 V
MIN value0 V0 V0 V
*RST value
(DEFault value)
Range
MAX value5.15 A1.03 A1.03 A
MIN value0 A0 A0 A
*RST value
(DEFault value)
0 to 6.18 V0 to +25.75 V0 to -25.75 V
0 V0 V0 V
0 to 5.15 A0 to 1.03 A0 to 1.03 A
5 A1 A1 A
72
Chapter 4 Remote Interface Reference
Using the APPLy Command
Usingthe APPLy Command
The
APPLy
command provides the most straightforward method to program the power supply over the remote interface. You can select the specific output, output voltage, and output current all in one command.
APPLy
{P6V | P25V | N25V}[,{<
This command iscombination of
VOLTage
the current of the specified outputare changed as soon as the command isexecuted.
You can identify each outputby the output name (P6V, P25V or N25V) as described in Table4-1. For the command, the ranges depend on the output currently selected. You can substitute “MINimum”, “MAXimum”,or“DEFault” in place of a specific value for the and current values allowed for the selectedoutput. MAX selects the highest voltage and current values allowed. The for all outputs. The ±25V outputs. The default voltage and current values areexactlythe same asthe
If you specify only one value for the parameter, the powersupply regards it as voltage setting value. If you do not specify any value for the parameter,the
INSTrument
, and
[SOURce:]CURRent
voltage
*RST
values.
APPLy
command only selects the output specified and acts as the
command.
voltage
>| DEF | MIN | MAX}[,{<
INSTrument:SELect
commands. Thevalues of voltageand
voltage
and
current
default current
See Table 4-1 for details of parameters.
and
current
parameters.MIN selectsthe lowest voltage
default voltage
values are 5 A for +6V output and 1 A for
current
parameters of the
values are 0 volts
>| DEF | MIN | MAX}]]
, [
SOURce:
]
APPLy
4
APPLy? [{P6V | P25V | N25V}]
Thiscommand queries the power supply's presentvalues for each output and returnsa quoted string. The voltage and current are returned in sequence as shown in the sample string below (the quotation marks are returned aspart of the string). If any output identifier isnotspecified, the voltage and the currentof the currently selected output are returned.
"5.000000,1.000000"
In the above string, thefirst number 5.000000 is the voltage limit valueand the secondnumber 1.000000 is the current limit value for the specified output.
voltage
and
current
73
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output Setting and OperationCommands
Thissection describes the low-level commandsused to program the power supply. Although themethod to program the power supply, the low-level commands give you more flexibility to change individual parameters.
See page102 forprogramming ranges, output identifiers, andMIN / MAXvaluesin the following commands.
Output Selection Commands
INSTrument[:SELect] {P6V | P25V | N25V}
This command selects the output to be programmed among three outputs by the output identifier.The outputsof the powersupply are considered three logical instruments. The identify and select an output.When one output isselected, the other outputs are unavailablefor programming untilselected. Thecommands which are affected by the(
SOURce
(
CALibration
output and “N25V” isfor -25V output.
), measurementcommands (
APPLy
INSTrument
). “P6V” isthe identifier for +6V output, “P25V” is for +25V
command provides the most straightforward
INSTrument
command are output setting commands
command provides a mechanism to
MEASure
), and calibration commands
INSTrument[:SELect]?
Thisquery returns the currently selected output by the[
:SELect
parameter is“P6V”,“P25V”, or “N25V”.
INSTrument:NSELect {1 | 2 | 3}
This command selects the output to be programmed among three outputs by a
numeric value
[
:SELect
“3” selects -25V output.
INSTrument:NSELect?
Thisquery returns the currently selected output by the
:NSELect
is “1” for +6V output, “2” for +25V outputor “3” for -25V output.
74
] or
INSTrument:NSELect
instead of the output identifier used inthe
] command.“1” selects+6V output,“2” selects +25V output, and
or
INSTrument[:SELect
command. The returned
]command. The returned parameter
INSTrument
INSTrument
INSTrument
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INST:COUP
Example (1)
INSTrument:COUPle[:TRIGger] {ALL | NONE |<
Thiscommand defines a coupling between various logical outputs of the powersupply. The couple command consists of an optional subsystemnode followedby a single parameter. The subsystem node iscommand,
The parameter indicates to which logical outputs the specified couplingisto apply. “ALL” indicates thatspecified coupling isto apply to all outputs.“NONE” indicates that specified coupling is to be removed. A listof outputs specifies a particular set of logicaloutputs to be coupled.Atoutputs are uncoupled.
suppliescan be coupled.
The following program segment showshowto use the
COUPle
outputs with voltageand current triggered levels.The power supply isset to the newly programmed values as set by the
command to couple two outputs between the +6V and the+25V
TRIGger
subsystem isassumed to be coupled.
subsystem. If no node follows the couple
Noticethat
commands.
only
TRACk
list
>}
validparameter for the optional
*RST
, all
must be off beforethe ±25V
INSTrument:
VOLTage:TRIGgered
Select the +6VoutputSet triggered level to 5 VSet triggered level to 3 ASelect the +25VoutputSet triggered levelto 20 VSet triggered level to 0.5 ACouple the +6V and +25V supplySet trigger to immediateTrigger the power supplytooutput the trigger valuesforthe+6V and the +25Vsupplies
and
4
Note If you select the bus trigger source in the above program (see page 79 for
the detailed information), you must send the *TRG or Group Execute
Trigger (GET) command to start the trigger action after sending the
INITiate command.
75
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
INSTrument:COUPle[:TRIGger]?
Thisquery returns the currently coupled output. Returns “ALL”, “NONE”,or a list. Ifany outputis not coupled, “NONE” isreturned. If all of three outputs are coupled, “ALL” is returned. Ifa listof outputs is coupled, the listis returned.
Measurement Commands
MEASure:CURRent[:DC]? [{P6V | P25V | N25V}]
This command queriesthe current measured at thepower supply.The physical outputs of measurement are specified by theoutput identifier. Ifany outputidentifier is not specified, the current of the currently selected outputis returned.
MEASure[:VOLTage][:DC]? [{P6V | P25V | N25V}]
Thiscommand queries the voltage measured at the powersupply. If any output identifier isnot specified,the voltage of the currently selected outputis returned.
output terminals
output terminals
of the
of the
76
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
Output On/Offand Tracking OperationCommands
OUTPut[:STATe] {OFF | ON}
Thiscommand enables or disables all state of thedisabled outputs isacondition of less than 0.6 volts ofopposite polarity with no load and less than 60 mA of opposite direction with a short circuit. At
OUTPut[:STATe]?
Thiscommand queries the output state ofthe power supply.The returned value is“0” (OFF) or “1” (ON).
OUTPut:TRACk[:STATe] {OFF | ON}
This commandenables or disablesthe powersupply to operate in the track mode. When the track modeis first enabled, the -25V supply willbe set tothe same voltage level as the +25V supply. Once enabled, anychange of the programmed voltage level in either +25V supply or -25V supply willbe reflected in theother supply. Thesupply to the non-track mode. The ±25V supplies must not be coupled toenable “Track”. At
*RST
, the output state is off.
*RST
, the track mode isdisabled.
three
outputs of the power supply. The
TRACk OFF
command returns the power
4
OUTPut:TRACk[:STATe]?
This command queriesthe track mode state of the powersupply. The returned value is“0” (OFF) or “1” (ON).
Thiscommand directly programs the immediate current level of the power supply. The immediate level is the current limitvalue ofthe outputselected with the
Thisquery returns the presently programmed current limit level ofthe selected output. the maximum and minimum output.
|MINimum | MAXimum}
INSTrument
command.
CURRent?MAXimum
programmable
and
CURRent? MINimum
currentlevels of the selected
return
77
Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{
<current>
| MINimum | MAXimum}
This command programs the pending triggeredcurrent level of the power supply. The pending triggered currentlevel is a transferred to the output terminals when a trigger occurs. A pendingtriggered level is not affected by subsequent
Thisquery returns the presently programmed triggered current level.If no triggered level is programmed, the
:TRIGgered? MAXimum
and
CURRent
level isreturned.
CURRent
CURRent:TRIGgered? MINimum
return the
maximum and minimum programmable triggered current levels.
VOLTage[:LEVel][:IMMediate][:AMPLitude]
{
<voltage>
| MINimum | MAXimum}
Thiscommand directly programs the immediate voltage level of the power supply. The immediate level isthe voltage limit value of the selected output with the
Thisquery returns the presently programmed voltage limit level of theselected output. the maximum and minimum
VOLTage? MAXimum
programmable
and
VOLTage? MINimum
voltage levels of the selected
return
output.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
{
<voltage>
| MINimum | MAXimum}
Thiscommand programsthe pending triggered voltage level of the powersupply. The pending triggered voltage level is a transferred to the output terminals when a trigger occurs. A pending triggered level is not affected by subsequent
Thisquery returns the presently programmed triggered voltage level.Ifno triggered level is programmed, the
:TRIGgered? MAXimum
and
VOLTage
level isreturned.
VOLTage
VOLTage:TRIGgered? MINimum
return
the maximum and minimum programmable triggered voltage levels.
78
Chapter 4 Remote Interface Reference
Triggering Commands
Triggering Commands
The power supply's triggering system allows a change in voltage and current when receivinga trigger, to select a trigger source, and to insert a trigger.Triggering the power supply isa multi-step process.
•First, you must select an output with the
command and then configure the powersupply for the triggered outputlevel by usingcommands.
•Then,you must specify the source fromwhich the power supply will
accept the trigger. The power supply will accept a bus (software) trigger or an immediate trigger from the remoteinterface.
•Then,youcan set the time delay betweenthe detection of the trigger on
the specified trigger source and the start of any correspondingoutput change.
source
.
•Finally, youmust providean
IMMediate
level immediately. But if the trigger source isthebus, thepower supply isset to the triggered level after receivingthe Group Execute Trigger (GET) or
*TRG
CURRent:TRIGgered
Notice that thetime delay is valid foronly the bustrigger
INITiate[:IMMediate]
source isselected, the selected output is set to the triggered
command.
INSTrument:SELect
and
VOLTage:TRIGgered
command. If the
Trigger Source Choices
You mustspecify the source from which the power supply willaccept a trigger. The trigger is stored inwhen the power supplyhas been off or after a remote interfacereset.
volatile
memory; the source isset tobus
4
Bus (Software) Triggering
•To select the bustrigger source,send the following command.
TRIGger:SOURce BUS
•To trigger thepower supply from the remote interface (GPIB or RS-232)
after selecting the bus source, send the the
*TRG
issent, the trigger action starts after the specified time delay if
any delay is given.
*TRG
(trigger) command.When
79
Chapter 4 Remote Interface Reference
Triggering Commands
•Youcan also trigger the power supply from the GPIB interface by sending theIEEE-488 Group Execute Trigger (GET) message. Thefollowing statement shows how to send a GET from a Agilent Technologiescontroller.
TRIGGER 705 (group execute trigger)
•To ensure synchronization when the bussource is selected,sendthe
*WAI
(wait) command. When thesupply waits for all pending operations to complete before executing any additional commands. For example, the followingcommand string guarantees that the first trigger is accepted and isexecuted before the second trigger isrecognized.
*WAI
command isexecuted, the power
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
•You can use the
*OPC
(operationcomplete) command tosignal when the operation iscomplete. The operation iscomplete. Thethe Standard Event register when the operation is complete.
Immediate Triggering
•To select the immediate triggersource, send the following command.
This command causes thetrigger system toinitiate. Thiscommandcompletes one full trigger cycle when the triggersource is animmediate and initiatesthe triggersubsystem when the trigger source isbus.
TRIGger[:SEQuence]:DELay{
This command sets the time delay between the detection of an event on the specified trigger source and the start of any corresponding trigger action on the power supply output. Selectfrom 0 to 3600 seconds. MIN = 0seconds. MAX= 3600 seconds. At
TRIGger[:SEQuence]:DELay?
Thiscommand queries the trigger delay.
TRIGger[:SEQuence]:SOURce {BUS | IMMediate}
Thiscommand selects the source from which thepower supply will accept a trigger. The power supply will accept a bus (software) trigger or an internal immediate trigger. At
TRIGger[:SEQuence]:SOURce?
Thiscommand queries the present trigger source. Returns “BUS” or “IMM”.
*TRG
This commandgenerates a triggerto the trigger subsystem that hasselecteda bus (software) trigger as its source (command has the sameeffect asthe Group Execute Trigger (GET) command. For RS-232operation, make sure the power supply is in the remoteinterface mode bysending the
<seconds>
*RST
*RST
, the bus trigger source isselected.
| MINimum | MAXimum}
, thisvalue isset to 0 seconds.
TRIGger:SOURce BUS
SYSTem:REMote
). The
command first.
4
81
Chapter 4 Remote Interface Reference
System-Related Commands
System-Related Commands
DISPlay[:WINDow][:STATe] {OFF | ON}
Thiscommand turns the front-panel display off or on. When the display isturned off, outputs are not sent to the display and all annunciators are disabledexcept the
The display state isautomatically turned on when you returnto the localmode. Press the key to return to the local state from theremoteinterface.
DISPlay[:WINDow][:STATe]?
Thiscommand queries the front-panel display setting. Returns “0” (OFF) or “1” (ON).
ERROR
annunciator.
Local
DISPlay[:WINDow]:TEXT[:DATA]
<quoted string>
Thiscommand displaysamessage on the frontpanel. The power supply willdisplay up to12 characters in a message; any additionalcharacters aretruncated. Commas, periods, and semicolons share a display spacewith the preceding character,and are not considered individual characters.
DISPlay[:WINDow]:TEXT[:DATA]?
This command queriesthe message sent to the front panel and returns a quoted string.
DISPlay[:WINDow]:TEXT:CLEar
This command clears the message displayed on the front panel.
SYSTem:BEEPer[:IMMediate]
Thiscommand issues a single beep immediately.
82
Chapter 4 Remote Interface Reference
System-Related Commands
SYSTem:ERRor?
Thiscommand queries the power supply's error queue. When the front-panel
ERROR
errors have been detected. Up to 20 errors can be stored in the error queue.
See “Error Messages” in chapter5.
annunciatorturnson, one or morecommand syntax or hardware
•Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the first error thatwas stored. When you haveread all errors from the queue, the beeps once each time an error is generated.
•If more than 20errors have occurred, the last error stored in the queue(the mostrecent error) isreplaced with additionalerrors are stored untilyou remove errors from the queue.If no errors have occurred when you read the error queue, the power supply responds with
•The error queue iscleared when powerhas been off or after a (clear status) command has been executed.The
does not
SYSTem:VERSion?
Thiscommand queries the power supply to determine the present SCPIversion. The returned value isof a string in the form YYYY.V where the “Y’s” represent the year of the version, and the “V” represents a version number for that year(for example, 1995.0).
*IDN?
Thisquery command reads the power supply's identification string. Thepowersupply returnsfour fields separated by commas.The manufacturer's name, the not used (always “0”), and the three numbers. The firstnumber isthe firmware revision number for the
main
and the third is for the The command returns a string withthe following format (be sure to
dimension a string variable with at least
clear the error queue.
power supply processor; the second is for the
ERROR
+0, “Noerror”
second
front-panel
annunciator turns off. The power supply
-350, “Too manyerrors”
.
*RST
(reset) command
first
field is the model number, the
fourth
field isa revision code which contains
input/output
processor.
40 characters
):
. No
*CLS
field is the
third
field is
processor;
4
HEWLETT-PACKARD,E3631A,0,X.X-X.X-X.X
83
Chapter 4 Remote Interface Reference
System-Related Commands
*RST
This command resets the power supply to its power-on state as follows:CommandState
The reset operation sets the current of +6Voutput to 5 A and the current of
+25V and -25Voutputs to 1 A
.
*TST?
Thisquery performs a
complete
self-test ofthe power supply. Returns “0” ifthe self-test passes or “1”or any non-zerovalue if itfails. If the self-test fails, an error message is also generated with additional information on why the test failed.
*SAV { 1 | 2 | 3 }
Thiscommand stores the present state of the power supply to the specified location in
non-volatile
memory. Three memory locations (numbered 1, 2 and 3) are available to storeoperatingstates of the power supply. The state storage feature “remembers” the states or values of
. Torecalla stored state, you must usethe same memory location
INST[:SEL],
, and
used previously to store the state.
*RCL {1 | 2 | 3 }
Thiscommand recalls a previously stored state. Torecall a stored state, you must use the samememory location used previously to store the state. You recall
*RST
states or values of the power supply from a memory location
that wasnot previously specified asa storage location.
84
Chapter 4 Remote Interface Reference
Calibration Commands
Calibration Commands
Seechapter 3 “Calibration Overview”, starting on page 58 for an overviewof the calibration features ofthe power supply. For moredetailed discussionof the calibration procedures, see the Service Guide.
CALibration:COUNt?
Thiscommand queries the power supply to determine the number oftimes ithas beencalibrated. Your power supply was calibratedbefore itleft the factory. When you receive your powersupply,read the count to determine its initial value. Since the value increments by one for each calibration point, a complete calibration for three outputs will increase the value by six counts.
CALibration:CURRent[:DATA]
This command canonly be used after calibration isunsecured. It enters acurrent value of a selected output that you obtained byreadingan external meter. You mustfirst select a calibration level (value being entered. Two successive values (one for each end ofthe calibration range) must be selected and entered. The power supply then computes new calibration constants. These constants are then stored in
non-volatile
CALibration:CURRent:LEVel {MINimum | MAXimum}
Beforeusing this command, you must select the output which is to be calibrated by usingused after calibration is unsecured. It sets the power supply to a calibration point that isentered withDuring calibration, two points must be entered and the low-end point (MIN) must beselected and entered first.
CALibration:SECure:CODE
This command entersa new security code.Tochange the security code, firstunsecurethe power supply using theold security code. Then, enter the new code. The calibration code may contain up to 12 charactersover the remoteinterface but the first character must alwaysbe a
memory.
INSTrument
<numeric value>
CAL:CURR:LEV
command. This command can only be
CALibration:CURRent[:DATA
<new code>
letter
.
) for the
] command.
4
85
Chapter 4 Remote Interface Reference
Calibration Commands
CALibration:SECure:STATe {OFF | ON>},
<code>
Thiscommand unsecures or secures the power supply for calibration. The calibration code may contain up to 12 characters over the remote interface.
CALibration:SECure:STATe?
Thiscommand queries the secured state for calibration of the power supply.The returned parameter is“0” (OFF) or “1” (ON).
CALibration:STRing
Thiscommand records calibration information about your power supply.
<quoted string>
For example, you can store such information as the last calibration date, the next calibration duedate, or the power supply’s serial number. Thecalibration message may contain up to 40 characters. Thepower supplyshould be unsecured before sending a calibration message.
CALibration:STRing?
Thiscommand queries the calibration messageand returns a quoted string.
CALibration:VOLTage[:DATA]
<numeric value>
This command canonly be used after calibration isunsecured. It enters avoltage value of a selected output that you obtained by reading an externalmeter. You mustfirst select a calibration level (value being entered. Two successive values (one for each end ofthe
CAL:VOLT:LEV
) for the
calibration range) must be selected and entered. The power supply then computes new voltage calibration constants. These constants arethen stored in
non-volatile
memory.
CALibration:VOLTage:LEVel {MINimum | MAXimum}
Beforeusing this command, you must select the output which is to be calibrated by using
INSTrument
command. This command can only be used after calibration is unsecured. It sets the power supply to a calibration point that isentered with
CALibration:VOLTage[:DATA]
command.During calibration, two points must be entered and the low-end point (MIN) must beselected
86
Chapter 4 Remote Interface Reference
RS-232 Interface Commands
RS-232 Interface Commands
Use the front-panel “I/O configuration” key toselectthe baud rate, parity, and the number of data bits (See chapter 3 “Remote Interface Configuration”,starting on page 48).
SYSTem:LOCal
Thiscommand places the power supply in the operation. All keys on the front panelare fully functional.
SYSTem:REMote
Thiscommand places the power supply in the remote mode for RS-232operation. All keys on the front panel, except the “Local” key,are disabled.
It isvery important that you sendtheplace the power supply inthe remote mode. Sending or receiving dataover the RS-232 interface when not configuredfor remote operation can cause unpredictable results.
SYSTem:RWLock
Thiscommand places the power supply in the remote mode for RS-232operation. Thiscommand is the same as the exceptthat all keys on the front panel are disabled, including the “Local”key.
Ctrl-C
This command clears the operation in progressover the RS-232 interfaceand discard any pending output data.
device clear actionover the GPIB interface.
This is equivalent to the IEEE-488
local
mode during RS-232
SYSTem:REMote
SYSTem:REMote
commandto
command
4
87
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The SCPIStatus Registers
All SCPIinstruments implement statusregisters in the sameway. The statussystem records various instrument conditions in three register groups: the Status Byte register, theStandard Event register, and the Questionable Statusregister group. The status byte register records high-level summary information reported in the other register groups. The diagrams on the subsequent pages illustrate the SCPIstatus system used by the power supply.
An example program is included in chapter 6,“ApplicationPrograms,” which shows the use of the status registers. You mayfind itusefultoreferto theprogram afterreading the following section inthis chapter.
What is an
An
event register
within the power supply. Bits in an event register bitis set, subsequent statechanges are ignored. Bits in an event register are automaticallycleared bya query of thatregister (such as
STAT:QUES:EVEN?
reset (
*RST
an event register returns adecimal value which corresponds to thebinary-weighted sum of all bits setin the register.
What is an
An
enable register
logicallyORed together to forma single summary bit.Enableregistersare both readable and writable. Querying an enable register(clear status) command does not clear enableregisters but it does clear the bits in the event registers.Toenable bits in an enable register, you must writea decimal value which corresponds to the binary-weighted sum of the bits youwish to enable in the register.
What is a
The three-logical outputsof the powersupply include an INSTrument summary status register and an individual instrument ISUMmary register for each logical output. The ISUMmary registersreport to the INSTrumentregister, which in turn reports to bit 13 ofthe Questionable status register. Thisis shown pictorially on the next page.
Event
) or device clear willnot clear bits in event registers. Querying
Enable
Multiple
Register?
is a read-only register that reports defined conditions
) or by sending the
are latched
*CLS
(clear status) command.A
. Once an event
*ESR?
or
Register?
defines which bits in the correspondingevent register are
will not
clear it. The
*CLS
LogicalOutput?
88
Chapter 4 Remote Interface Reference
The SCPI Status Registers
Using such a status register configuration allows a status event to be cross-referenced by output and type of event. The INSTrument register indicates which output(s) have generated an event. The ISUMmary register is a pseudo-questionable status registerfor a particular logicaloutput.
4
89
Chapter 4 Remote Interface Reference
The SCPI Status Registers
SCPI Status System
90
Binary Weights
20 = 1
1
= 2
2
2
2
= 4
3
= 8
2
4
2
= 16
5
= 32
2
6
2
= 64
7
= 128
2
8
2
= 256
9
= 512
2
10
2
= 1024
11
= 2048
2
12
2
= 4096
13
= 8192
2
14
2
= 16384
15
= 32768
2
Chapter 4 Remote Interface Reference
The SCPI Status Registers
TheQuestionable Status Register
The Questionable Statusregister provides information about unexpectedoperation of thepower supply.Bit 4 reports a fault with the fan, and bit 13summarizes questionable outputs for any of the three supplies. Forexample if one of thethree supplies is inconstant voltage modeand due to an overloadlooses regulation, bit 13 isset (latched). Send the command
STAT:QUES?
enable registersyou wish to summarize withbit 13. Send
:INST:ENAB14
STAT:QUES:INST:ISUM<n>:ENAB 3
to readthe register. Tomake use ofbit13 you must first
STAT:QUES
to enable the Questionable Instrument register.Thensend
for each supply to enable the
Questionable Instrument Summary register, where n is 1, 2,or 3.
Table 4-2. Bit Definitions - Questionable Status Register
Bit
0-3 Not used0Always set to 0.
4 FAN16The fan has a fault condition.
5-12 Not Used0Always set to 0.
13 ISUM8192Summary of QUES:INST and QUES:INST:ISUM registers.
14-15 Not Used0Always set to 0.
Decimal
Value
Definition
4
The Questionable Instrument StatusRegister
The Questionable Instrument register providesinformationabout unexpected operations for each of the three supplies. For example if the +6Vsupply is in the constant voltage mode and looses regulation, then bit 1 set indicating a possibleoverload in the +6V supply.The +25V supplyisreported as bit 2, and the-25V supply as bit 3. Send the command
QUES:INST?
to readthe register. The
STAT:QUES:INST:ISUM<n>
registers must be enabled to makeuse of the Questionable Instrumentregister. Send
STAT:QUES:INST:ISUM<n>:ENAB 3
to enable outputn.
STAT
The Questionable Instrument Summary Register
There arethree Questionable Instrument Summaryregisters, one for each supply output. These registers provide information about voltage andcurrent regulation. Bit 0 isset when the voltagebecomes unregulated, and bit1 isset if the currentbecomes unregulated. For example if a supply which isoperatingas a voltage source (constant voltage mode) momentarily goes to constant current mode, bit 0 isset to indicatethat the voltage output isnot regulated.Toread the register for each supply, send
:INST:ISUM<n>?
, where n is 1, 2, or 3.
STAT:QUES
91
Chapter 4 Remote Interface Reference
The SCPI Status Registers
To determine the operating mode (CV or CC) for the power supply send
STAT:QUES:INST:ISUM<n>:COND?
the output.Bit 1 true indicatesthe output is in constant voltage mode, bit 0true indicates constant current mode, both bits true indicates neither the voltage nor the currentis regulated, and both bits false indicates the outputs of the power supply are off.
The Questionable Status Event register is cleared when:
, where n is 1, 2, or 3 depending on
•You executethe
•You query the event register using
[:EVENt]?
For example, 16 is returned when you havequeried the status of the questionable event register, the FAN condition is questionable.
The Questionable Status Enable registeris cleared when:
•You execute
For example, you must send the
*CLS
(clear status)command.
STATus:QUEStionable
(Status Questionable Event register) command.
STATus:QUEStionable:ENABle 0
STAT:QUES:ENAB 16
command.
to enable the FAN bit.
92
Chapter 4 Remote Interface Reference
The SCPI Status Registers
TheStandard Event Register
The Standard Event register reports the following types of instrument events: power-on detected, command syntax errors,commandexecutionerrors, self-test or calibration errors, query errors, or when an
*OPC
command is executed. Anyor all of these conditions can be reported in the Standard Event Summary bit (ESB, bit 5) of Status Byte register through theenable register.Toset the enable register mask, you write a decimal value to the register usingthe
*ESE
(Event Status Enable) command.
An error condition (Standard Eventregisterbits 2,3, 4, or 5)willalways record one or more errors in the power supply's error queue.Read the error queue using the
Table 4-3. Bit Definitions - Standard Event Register
Bit
0OPC1
1Not Used0Always set to 0.
2QYE4
3DDE8
4EXE16
5CME32
6Not Used0Always set to 0.
7PON128
Decimal
Value
SYSTem:ERRor?
Definition
Operation Complete. All commands prior to and including an
*OPC command have been executed.
Query Error. The power supply tried to read the output buffer
but it was empty. Or, new command line was received before
a previous query had been read. Or, both the input and output
buffers are full.
Device Error. A self-test or calibration error occurred (see error
numbers 601 through 748 in
Execution Error. An execution error occurred (se e error numbers
-211 through -224 in
Command Error. A command syntax error occurred (see error
number -101 through -178 in
Power On. Power has been turned off and on since the last
time the event register was read or cleared
command.
chapter 5).
chapter 5).
chapter 5).
4
93
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The Standard Event register is cleared when:
•You executethe
•You query the event register using the
*CLS
(clear status)command.
*ESR?
(Event Status register)
command.
For example, 28 (4 + 8 + 16) is returnedwhen you have queried the status of the StandardEventregister, QYE, DDE,and EXE conditions have occurred.
The Standard Event Enable register is cleared when:
•You executethe
*ESE 0
command.
•You turn on the power and have previously configured the power supply
using the
•The enableregister
configured the power supply using the
For example, you must send the
*PSC 1
command.
will not
be cleared atpower-on if you havepreviously
*PSC 0
*ESE 24
command.
(8 +16)to enable DDE andEXE
bits.
The Status Byte Register
The Status Byte summary register reports conditions fromthe other status registers. Query data that iswaiting in the power supply's output bufferisimmediately reported through the “Message Available” bit (bit 4) of StatusByte register. Bits in the summary register are not latched. Clearing an event register willclear the corresponding bits in the Status Byte summary register. Reading all messages in the output buffer,including any pending queries, will clear the message availablebit.
Table 4-4. Bit Definitions - Status Byte Summary Register
Bit
0-2Not Used0Always set to 0.
3QUES8
4MAV16Data is available in the pow er supply output buffer.
5ESB32
6RQS64The power supply is requesting service (serial poll).
7Not Used0Always set to 0.
Decimal
Value
Definition
One or more bits are set in the questionable status
register (bits must be “enabled” in the enable register).
One or more bits are set in the standard event register
(bits must be “enabled” in the enable register).
94
Chapter 4 Remote Interface Reference
The SCPI Status Registers
The Status Byte Summary register is cleared when:
•You executethe
•Querying the Standard Event register (bit5 in the Status Bytesummary register.
For example, 24 (8 + 16) is returned when you havequeried the status of theStatus Byte register, QUES and MAVconditions have occurred.
The Status Byte Enable register (Request Service) is cleared when:
•You executethe
•You turn on the power and have previously configured the power supply using the
•The enable register configured the power supply using
Forexample, you must send thebits.
*PSC1
*CLS
(clear status)command.
*SRE0
command.
command.
will not
be cleared at power-on if you have previously
*SRE96
*ESR?
*PSC0
(32 + 64) to enable ESB and RQS
command) will clear only
.
Using Service Request (SRQ) and SerialPOLL
You must configure your bus controller to respond to theIEEE-488 service request (SRQ) interrupt to usethis capability. Use the Status Byte enableregister (low-level IEEE-488 service request signal.When bit 6 (request service) issetin the Status Byte, an IEEE-488 service request interrupt message isautomatically sent tothe bus controller. The bus controller may then poll the instruments on thebus to identifywhich onerequested service (theinstrument with bit 6 set in itsStatus Byte).
*SRE
command) to select which summary bitswill set the
4
The request service bitiscleared onlyby readingtheStatus Byte using anIEEE-488 serial pollorby readingtheevent register whose summarybitis causing the service request.
To read the StatusByte summary register, send the IEEE-488 serial pollmessage. Querying the summary register will return a decimal value whichcorresponds to the binary-weighted sum of the bits set in the register. Serial poll will automatically clear the “requestservice” bit in the Status Byte summary register. No other bits are affected. Performing aserial poll willnot affectinstrument throughput.
95
Chapter 4 Remote Interface Reference
The SCPI Status Registers
Caution The IEEE-488 standard does not ensure synchronization between your
bus controller program and the instrument. Use the *OPC? command to
guarantee that commands previously sent to the instrument have
completed. Executing a serial poll before a *RST,*CLS, or other
commands have completed can cause previous conditions to be reported.
Using *STB? toRead the Status Byte
The
*STB?
(Status Bytequery) command is similarto a serial poll but itisprocessed likeany other instrument command. The returns the same result as a serial poll but the “request service” bit (bit6)
not
cleared.
The
*STB?
command is not handled automaticallyby the IEEE-488 businterface hardwareand will be executed completed. Polling isnot possible using the
*STB?
command does not clear the Status Byte summary register.
only
*STB?
*STB?
after previous commands have
command. Executingthe
command
Using the Message Available Bit (MAV)
You can use the Status Byte “message available” bit (bit 4) to determine when data isavailable to read into your bus controller. The power supply subsequently clears bit 4 output buffer.
only
after all messages havebeen read fromthe
To Interrupt Your Bus Controller Using SRQ
is
1
Send a device clear message to clear the power supply's output buffer (e.g.,
CLEAR 705
2
Clear the event registerswith the
3
Set up the enableregister masks. Execute the Standard Event register and the
4
Send theto ensure synchronization.
5
Enable your bus controller's IEEE-488SRQ interrupt.
96
).
*CLS
(clear status) command.
*ESE
command to set up the
*SRE
command for the StatusByte.
*OPC?
(operation complete query) command and enter the result
Chapter 4 Remote Interface Reference
The SCPI Status Registers
To Determine When a Command Sequence is Completed
1
Send a device clear message to clear the power supply's output buffer (e.g.,
CLEAR 705
2
Clear the event registerswith the
3
Enable the “operation complete” bit (bit0) in the Standard Event register by executing the
4
Send theto ensure synchronization.
5
Execute your command stringto program the desired configuration, and then execute thecommand.When the command sequence is completed, the “operation complete” bit (bit 0) is set in the Standard Event register.
6
Use a serial poll tocheck to see when bit 5 (standard event) isset inthe StatusByte summary register. You could also configure the power supply for an SRQ interrupt by sending
).
*ESE 1
*OPC?
*CLS
(clear status) command.
command.
(operationcomplete query)command and enter the result
*OPC
(operation complete) command as the last
*SRE 32
(Status Byte enableregister, bit 5).
4
Using *OPC to Signal When Data is intheOutput Buffer
Generally, it isbest tousethe “operationcomplete” bit (bit 0)intheStandard Event register to signal when a command sequence iscompleted. Thisbitis set in the register after anyou send supply's output buffer (query data), you can use the “operation complete” bit to determinewhen the message isavailable. However, if too manymessages are generated before the buffer will fill and the power supply willstopprocessing commands.
*OPC
after a command which loadsa message in the power
*OPC
*OPC
command has been executed. If
command executes (sequentially), the output
97
Chapter 4 Remote Interface Reference
Status Reporting Commands
Status Reporting Commands
Seediagram “SCPI Status System”, on page90 in thischapter fordetailedinformation of the status register structure of the power supply.
SYSTem:ERRor?
Thisquery command reads one error from the error queue. When the front-panel
ERROR
annunciator turns on, one or more command syntax orhardwareerrors have been detected. A record of up to 20 errors can be stored in the power supply’s error queue.
See “ErrorMessages” in chapter 5
•Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error thatwas stored. When you haveread all errors from the queue, the
ERROR
annunciator turns off. The power supply
beeps once each time an error is generated.
•If more than 20errors have occurred, the last error stored in the queue
(the mostrecent error) isreplaced with
-350, “Too manyerrors”
. No additionalerrors are stored until you remove errors fromthe queue. If no errors have occurred when you read the error queue, the power supply responds with
•The error queue iscleared when powerhas been off or after a (clear status)command has been executed. The
does not
STATus:QUEStionable[:EVENt]?
+0, “Noerror”
clear the error queue.
.
*CLS
*RST
(reset) command
Thiscommand queries the Questionable Status event register. The power supply returns a decimal value which corresponds to the binary-weighted sumof all bits in the register.
STATus:QUEStionable:ENABle
<enable value>
This command enablesbits in the Questionable Status enable register.Theselected bits are then reported to the Status Byte.
STATus:QUEStionable:ENABle?
Thiscommand queries the Questionable Status enable register. The power supply returns a binary-weighted decimal representingthe bits set in the enable register.
98
Chapter 4 Remote Interface Reference
Status Reporting Commands
STATus:QUEStionable:INSTrument[:EVENt]?
Thiscommand queries the Questionable Instrument event register. The power supply returns a decimal value which corresponds to the binary-weighted sum of all bitsinthe register and clears the register.
STATus:QUEStionable:INSTrument:ENABle
<enable value>
This command sets the value of the Questionable Instrument enable register. Thisregister isamask for enabling specific bits from the Questionable Instrument event register to set theInstrument Summary bit (ISUM, bit 13) of the Questionable Status register. The “ISUM” bit of theQuestionableStatusregister is the logical ORof all the Questionable Instrument event register bits thatareenabled by the Questionable Instrument enable register.
STATus:QUEStionable:INSTrument:ENABle?
Thisquery returns the value ofthe QuestionableInstrument enable register.
Thisquery returns the value oftheQuestionableInstrument Isummary event register for aspecific output of thethree-output power supply.Theparticular output must be specified by a numeric value. n is 1, 2, or 3. See Table 4-1 on page72for the output number. Theevent register is a read-only register which holds (latches) all events. Reading theQuestionableInstrument Isummary event register clears it.
Thisquery returns the CVor CC condition of the specified instrument. If “2” isreturned, the queried instrument is in the CV operating mode. If “1” is returned,the queried instrument isin the CC operating mode. If “0” isreturned, the outputs of the instrument are off or unregulated. If ‘3” is returned, the instrument isin the hardware failure. n is 1, 2, or 3.
4
STATus:QUEStionable:INSTrument:ISUMmary<n>:ENABle
Thiscommand sets the value of the Questionable Instrument Isummary
<enable value>
enable register for a specific output of the three-output powersupply. The particular output must be specified by a numeric value. n is 1, 2, or 3. See Table 4-1 on page72 for the output number. This register isa maskfor enabling specific bitsfrom the Questionable Instrument Isummary event register to set the Instrument Summarybit (bit 1, 2, and 3) ofthe QuestionableInstrument register. These bits 1, 2, and bit 3are the logical ORof all the Questionable Instrument Isummary event register bitsthat areenabled by the Questionable Instrument Isummary enable register.
99
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.