Agilent E1440A Technical Specifications

PDFINFO H5535-01
Synthesized Function/Sweep Generator VXI Module HP E1440A
Technical Specifications
Five different waveforms
Multi-interval sweep and multimarker mode
1 µHz - 60 MHz TTL clock
High-voltage output option, isolated (floating) outputs
Description
The HP E1440A function/sweep generator is a C-size, 2-slot, message-based VXI module. It provides low-
distortion sinewaves and a variety of waveforms for applications requiring high frequency stability and resolution (11 digits). A programmable relative phase output synchronized to a companion module is also available. Two or more HP E1440A modules can be used for generating multi-phase related signals.
HP E1440A
With this module, you can use the modulation source as an arbitrary function generator via HP-IB to provide user-defined waveforms, or use the save-recall memory that includes nonvolative memory locations for simple and rapid access to frequently used test setups. Additionally, you can produce five different waveforms including: sine, square, triangle, negative, and positive ramps.
Refer to the HP Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
PDFINFO H5535-02
Product Specifications
Those specifications indicated as “typical” describe the instrument’s typical performance; all others describe the instrument’s warranted performance.
Waveforms
Standard waveforms: Sine, square, triangle, negative and positive
Arbitrary waveform function: No
Frequency Range
Frequency bandwidth: 21 MHz (sine) Sine: 1 µHz-21 MHz Square: 1 µHz-11 MHz Triangle: 1 µHz-11 kHz Ramps: 1 µHz-11 kHz TTL clock: 1 µHz-60 MHz Resolution: 11 digits Accuracy: ± 5 ppm of selected value, 20 to 30 °C, at
Stability: ± 5 ppm/year, 20 to 30 °C Modulation: AM, PM
Main Signal Output
(Typical)
Impedance: 50 ± 1 Max. external voltages: 42 V pk (ac+dc) max. (chassis ground to
Connector: BNC
Amplitude
(50 all waveforms except TTL clock, without dc offset)
Range: 1 mV to 10 Vp-p in 8 amplitude ranges,
Resolution: 4 digits (0.03% of full range) Amplitude accuracy (AC): ± 2% FS Amplitude accuracy (DC): n/a
Sinewave Spectral Purity
Phase noise: –55 dB for a 30 kHz band centered on a
Spurious: –60 dBc or –85 dBm, whichever is greater Sinewave harmonic distortion:
Frequency Range Harmonic
0.1 Hz - 199 kHz –60 dBc 200 kHz - 1.99 MHz –40 dBc 2 MHz - 14.9 MHz –30 dBc 15 MHz - 20 MHz –25 dBc
Squarewave Characteristics
(Typical)
Rise/fall time: 20 ns Overshoot: 5% of p-p amplitude at full output Triangle/ramp linearity: ± 0.05% of full p-p output voltage for each
Ramps, dc, TTL clock
time of calibration with standard frequency reference
circuit ground, 0-10 kHz) ± 10 V max. (floating ground to signal output)
1–3–10 sequence, amplitude can also be set up in Vrms and dBm
20 MHz carrier (excluding ± 1 Hz about the carrier)
Level
(10% to 90% of p-p output voltage)
(10% - 90%, 10 kHz)
range
DC Offset Range (50 )
dc only: 0 to ± 5 V dc + ac: max. ± 4.5 V Resolution: 4 digits
Phase Offset
(Related to another HP E1440A or equivalent)
Range: ± 719.9° Resolution: 0.1° Increment accuracy: ± 0.5° Stability: ± 1° of phase/° C
Sinewave Amplitude Modulation
(Typical)
Modulation depth: 0-98% Modulation frequency
range: Envelope distortion: –30 dB for modulation to 80% at 1kHz
Sensitivity: ± 5 V peak for maximum modulation
dc to 350 kHz
frequency)
(0 Vdc Offset)
(1 µHz-21 MHz carrier
Phase Modulation
(Typical)
Sinewave range: ± 900°, ± 5 V input Sinewave linearity: ± 0.5%, best fit straight line up to ± 720° of
Squarewave range: ± 450° Triangle range: ± 45° Positive and negative
ramps range: ± 90° Modulation frequency
range: DC to 5 kHz
modulation range
Sweep
Sweep: Frequency Sweep sequence modes: Single, continuous Sweep function modes:
Multi-Interval: (Up to 50 different intervals can be
Linear or logarithmic sweep: (Can be set for each interval)
Sweep time:
Linear: 0.01 s to 105 s Logarithmic: 0.1 s to 105 s
Minimum sweep width:
Linear: 0 Hz Logarithmic: 1 decade Maximum sweep width: Full frequency range
Minimum sweep rate:
Linear: 0.2 Hz/s Phase continuity: Sweep is phase continuous over the full
Multi-marker:
Linear sweep only: Up to nine markers can be set in this one
Sweep time: 0.01 s to 105 s Sweep width: From 0 Hz to full frequency range
sequenced and repeated in any order in a sequence that can contain up to 100 intervals)
frequency range of the main output for all sweep modes
dedicated interval
Loading...