Agilent E1440A Technical Specifications

PDFINFO H5535-01
Synthesized Function/Sweep Generator VXI Module HP E1440A
Technical Specifications
Five different waveforms
Multi-interval sweep and multimarker mode
1 µHz - 60 MHz TTL clock
High-voltage output option, isolated (floating) outputs
Description
The HP E1440A function/sweep generator is a C-size, 2-slot, message-based VXI module. It provides low-
distortion sinewaves and a variety of waveforms for applications requiring high frequency stability and resolution (11 digits). A programmable relative phase output synchronized to a companion module is also available. Two or more HP E1440A modules can be used for generating multi-phase related signals.
HP E1440A
With this module, you can use the modulation source as an arbitrary function generator via HP-IB to provide user-defined waveforms, or use the save-recall memory that includes nonvolative memory locations for simple and rapid access to frequently used test setups. Additionally, you can produce five different waveforms including: sine, square, triangle, negative, and positive ramps.
Refer to the HP Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
PDFINFO H5535-02
Product Specifications
Those specifications indicated as “typical” describe the instrument’s typical performance; all others describe the instrument’s warranted performance.
Waveforms
Standard waveforms: Sine, square, triangle, negative and positive
Arbitrary waveform function: No
Frequency Range
Frequency bandwidth: 21 MHz (sine) Sine: 1 µHz-21 MHz Square: 1 µHz-11 MHz Triangle: 1 µHz-11 kHz Ramps: 1 µHz-11 kHz TTL clock: 1 µHz-60 MHz Resolution: 11 digits Accuracy: ± 5 ppm of selected value, 20 to 30 °C, at
Stability: ± 5 ppm/year, 20 to 30 °C Modulation: AM, PM
Main Signal Output
(Typical)
Impedance: 50 ± 1 Max. external voltages: 42 V pk (ac+dc) max. (chassis ground to
Connector: BNC
Amplitude
(50 all waveforms except TTL clock, without dc offset)
Range: 1 mV to 10 Vp-p in 8 amplitude ranges,
Resolution: 4 digits (0.03% of full range) Amplitude accuracy (AC): ± 2% FS Amplitude accuracy (DC): n/a
Sinewave Spectral Purity
Phase noise: –55 dB for a 30 kHz band centered on a
Spurious: –60 dBc or –85 dBm, whichever is greater Sinewave harmonic distortion:
Frequency Range Harmonic
0.1 Hz - 199 kHz –60 dBc 200 kHz - 1.99 MHz –40 dBc 2 MHz - 14.9 MHz –30 dBc 15 MHz - 20 MHz –25 dBc
Squarewave Characteristics
(Typical)
Rise/fall time: 20 ns Overshoot: 5% of p-p amplitude at full output Triangle/ramp linearity: ± 0.05% of full p-p output voltage for each
Ramps, dc, TTL clock
time of calibration with standard frequency reference
circuit ground, 0-10 kHz) ± 10 V max. (floating ground to signal output)
1–3–10 sequence, amplitude can also be set up in Vrms and dBm
20 MHz carrier (excluding ± 1 Hz about the carrier)
Level
(10% to 90% of p-p output voltage)
(10% - 90%, 10 kHz)
range
DC Offset Range (50 )
dc only: 0 to ± 5 V dc + ac: max. ± 4.5 V Resolution: 4 digits
Phase Offset
(Related to another HP E1440A or equivalent)
Range: ± 719.9° Resolution: 0.1° Increment accuracy: ± 0.5° Stability: ± 1° of phase/° C
Sinewave Amplitude Modulation
(Typical)
Modulation depth: 0-98% Modulation frequency
range: Envelope distortion: –30 dB for modulation to 80% at 1kHz
Sensitivity: ± 5 V peak for maximum modulation
dc to 350 kHz
frequency)
(0 Vdc Offset)
(1 µHz-21 MHz carrier
Phase Modulation
(Typical)
Sinewave range: ± 900°, ± 5 V input Sinewave linearity: ± 0.5%, best fit straight line up to ± 720° of
Squarewave range: ± 450° Triangle range: ± 45° Positive and negative
ramps range: ± 90° Modulation frequency
range: DC to 5 kHz
modulation range
Sweep
Sweep: Frequency Sweep sequence modes: Single, continuous Sweep function modes:
Multi-Interval: (Up to 50 different intervals can be
Linear or logarithmic sweep: (Can be set for each interval)
Sweep time:
Linear: 0.01 s to 105 s Logarithmic: 0.1 s to 105 s
Minimum sweep width:
Linear: 0 Hz Logarithmic: 1 decade Maximum sweep width: Full frequency range
Minimum sweep rate:
Linear: 0.2 Hz/s Phase continuity: Sweep is phase continuous over the full
Multi-marker:
Linear sweep only: Up to nine markers can be set in this one
Sweep time: 0.01 s to 105 s Sweep width: From 0 Hz to full frequency range
sequenced and repeated in any order in a sequence that can contain up to 100 intervals)
frequency range of the main output for all sweep modes
dedicated interval
PDFINFO H5535-03
Auxiliary Outputs
(Typical)
SYNC-OUT TTL:
Signal: Phase synchronous squarewave with same
Output impedance: 50 Connector: BNC and TTL trigger bus
X-Drive 0 to 10 V:
Signal: 0 - 100 s sweeps only (proportional ramp to Output impedance: 650
Output level: 0 to + 10 V (into open circuit) Connector: BNC
Pen lift:
Signal: TTL-compatible voltage levels capable of
Connector: BNC
Marker TTL:
Signal: High-to-low transitions at selected marker
Pulsewidth in multimarker mode: 1 ms
Connector: BNC & TTL trigger bus Fan out: 4
REF out 10 MHz:
Signal: 10 MHz squarewave for phase-locking Output impedance: 50
Output levels (into 50 ): High level >2 V, low level <0.2 V AC-coupled output
levels: 10 dBm Connector: BNC
frequency as the main signal output, or 1 µHz to 60 MHz TTL clock (main signal output switched off)
the entire sweep time)
sinking current from a positive source. Current 200 mA, voltage 45 V
frequencies. TTL-and CMOS_ compatible output levels
additional instruments to the HP E1440A.
Auxiliary Inputs
(Typical)
External REF in 1/10 MHz: (For phase locking the HP E1440A to an external frequency reference)
Signal: From 0 dBm to 20 dBm into 50
Connector: BNC or VXI-system clock
AM:
Input impedance: 10 k Connector: BNC Max. external voltage: ± 15 V
PM:
Input impedance: >40 k Connector: BNC Max. external voltage: ± 15 V
signal must be a subharmonic of 10 MHz from 1 MHz to 10 MHz)
(reference
Option 001 High-Voltage Output
Frequency range: 1 µHz to 1 MHz Ampitude:
Range: 4 mV to 40 V p-p in eight ranges, 4-12-40
Accuracy: ± 2% of full output for each range at 2 kHz Flatness: ± 10% relative to programmed amplitude
Sinewave harmonic distortion: Frequency Range Harmonic Level
10 Hz - 199 kHz –60 dBc 200 kHz - 1 MHz –40 dBc
Output impedance: <3 at DC, <10 at 1 MHz
DC offset range: Four times the specified range of the
sequence into 500 , <500 pF load; ranges are four times the standard instrument ranges, without dc offset
impedance 500 , 500 pF, max. output current 40 mA peak)
standard instrument
(load
General Specifications
VXI Characteristics
VXI device type: Message based Data transfer bus: A16/A24, D16 Master, A16/D16 Slave Size: C Slots: 2 Connectors: P1/2 Shared memory: n/a VXI busses: TTL Trigger Bus (T) C-size compatibility: n/a
Instrument Drivers
See the HP Website (http://www.hp.com/go/inst_drivers) for driver availability and downloading.
Command module firmware: n/a
Command module firmware rev: n/a
I-SCPI Win 3.1: n/a I-SCPI Series 700: n/a C-SCPI LynxOS: n/a C-SCPI Series 700: n/a HP Panel Drivers: Yes
plug&play
VXI Framework: Yes
plug&play
VXI Framework: Planned 1998
plug&play
VXI Framework: No
Win
Win 95/NT
HP-UX
Module Current
I
PM
+5 V: 1 0.01 +12 V: 00 –12 V: 00 +24 V: 0.55 0.05 –24 V: 0.6 0.05 –5.2 V: 0.14 0.03 –2 V: 00
I
DM
Cooling/Slot
Watts/slot: 18.00 P mm H Air Flow liter/s: 2.00
O: 0.40
2
Ordering Information
Description Product No.
Synth. Funct/Sweep Generator VXI-Module HP E1440A High-Voltage Output HP E1440A 001 Operation Manual HP E1440A 0B2 Service Manual HP E1440A 0B3 Refurbished Equipment HP E1440A 8ZE
PDFINFO H5535-04
Related Literature
1998 Test System and VXI Products Data Book,
HP Pub. No. 5966-2812E
1998 Test System and VXI Products Catalog,
HP Pub. No. 5966-2815
Warranty
Standard Hewlett-Packard VXIbus hardware products are warranted against defects in materials and workmanship for a period of three years unless otherwise noted. HP software and firmware products that are designated by HP for use with a hardware product, when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in materials and workmanship.
For more information about Hewlett­Packard test & measurement products, applications, services, and for a current sales office listing, visit our website,
http://www.hp.com/go/tmdir
. You can also
contact one of the following centers and ask for a test and measurement sales representative.
United States:
Hewlett-Packard Company Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844
Canada:
Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206 4725
For a complete and detailed warranty statement please see the HP
Test System and VXI Products Data Book
or visit the
HP Website at http://www.hp.com/go/vxi.
HP Website Directory
Main HP Website
http://www.hp.com
HP Test and Measurement
http://www.hp.com/go/tmdir
HP VXI Product Information
http://www.hp.com/go/vxi
HP VXI Channel Partners
http://www.hp.com/go/vxichanpart
HP VEE Application Website
http://www.hp.com/go/hpvee
Data Acquisition and Control Website
http://www.hp.com/go/data_acq
HP Instrument Driver Downloads
http://www.hp.com/go/inst_drivers
Europe:
Hewlett-Packard European Marketing Centre P.O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900
Japan:
Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192, Japan Tel: (81) 426 56 7832 Fax: (81) 426 56 7840
Latin America:
Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive, 9th Floor Miami, Florida 33126 U.S.A. Tel: (305) 267-4245
(305) 267-4220
Fax: (305) 267-4288
Australia/New Zealand:
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 Australia 1 800 629 485
Asia Pacific:
Hewlett-Packard Asia Pacific Ltd. 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Tel: (852) 2599 7777 Fax: (852) 2506 9285
Data Subject to Change Copyright © May 1998 Hewlett-Packard Company HP Publication No.: 5965-5535E
Loading...