PDFINFO H5535-01
Synthesized Function/Sweep
Generator VXI Module
HP E1440A
Technical Specifications
●
Five different waveforms
●
Multi-interval sweep and multimarker mode
●
Amplitude and phase modulation
●
1 µHz - 60 MHz TTL clock
●
High-voltage output option, isolated (floating)
outputs
Description
The HP E1440A function/sweep generator is a C-size,
2-slot, message-based VXI module. It provides low-
distortion sinewaves and a variety of waveforms for
applications requiring high frequency stability and
resolution (11 digits). A programmable relative phase
output synchronized to a companion module is also
available. Two or more HP E1440A modules can be
used for generating multi-phase related signals.
HP E1440A
With this module, you can use the modulation source as
an arbitrary function generator via HP-IB to provide
user-defined waveforms, or use the save-recall memory
that includes nonvolative memory locations for simple
and rapid access to frequently used test setups.
Additionally, you can produce five different waveforms
including: sine, square, triangle, negative, and positive
ramps.
Refer to the HP Website for instrument driver
availability and downloading instructions, as well as for
recent product updates, if applicable.

PDFINFO H5535-02
Product Specifications
Those specifications indicated as “typical” describe the
instrument’s typical performance; all others describe the
instrument’s warranted performance.
Waveforms
Standard waveforms: Sine, square, triangle, negative and positive
Arbitrary waveform function: No
Frequency Range
Frequency bandwidth: 21 MHz (sine)
Sine: 1 µHz-21 MHz
Square: 1 µHz-11 MHz
Triangle: 1 µHz-11 kHz
Ramps: 1 µHz-11 kHz
TTL clock: 1 µHz-60 MHz
Resolution: 11 digits
Accuracy: ± 5 ppm of selected value, 20 to 30 °C, at
Stability: ± 5 ppm/year, 20 to 30 °C
Modulation: AM, PM
Main Signal Output
(Typical)
Impedance: 50 Ω ± 1 Ω
Max. external voltages: 42 V pk (ac+dc) max. (chassis ground to
Connector: BNC
Amplitude
(50 Ωall waveforms except TTL clock, without dc offset)
Range: 1 mV to 10 Vp-p in 8 amplitude ranges,
Resolution: 4 digits (0.03% of full range)
Amplitude accuracy (AC): ± 2% FS
Amplitude accuracy (DC): n/a
Sinewave Spectral Purity
Phase noise: –55 dB for a 30 kHz band centered on a
Spurious: –60 dBc or –85 dBm, whichever is greater
Sinewave harmonic distortion:
Frequency Range Harmonic
0.1 Hz - 199 kHz –60 dBc
200 kHz - 1.99 MHz –40 dBc
2 MHz - 14.9 MHz –30 dBc
15 MHz - 20 MHz –25 dBc
Squarewave Characteristics
(Typical)
Rise/fall time: ≤20 ns
Overshoot: 5% of p-p amplitude at full output
Triangle/ramp linearity: ± 0.05% of full p-p output voltage for each
Ramps, dc, TTL clock
time of calibration with standard frequency
reference
circuit ground, 0-10 kHz)
± 10 V max. (floating ground to signal
output)
1–3–10 sequence, amplitude can also be
set up in Vrms and dBm
20 MHz carrier (excluding ± 1 Hz about the
carrier)
Level
(10% to 90% of p-p output voltage)
(10% - 90%, 10 kHz)
range
DC Offset Range (50 Ω)
dc only: 0 to ± 5 V
dc + ac: max. ± 4.5 V
Resolution: 4 digits
Phase Offset
(Related to another HP E1440A or equivalent)
Range: ± 719.9°
Resolution: 0.1°
Increment accuracy: ± 0.5°
Stability: ± 1° of phase/° C
Sinewave Amplitude Modulation
(Typical)
Modulation depth: 0-98%
Modulation frequency
range:
Envelope distortion: –30 dB for modulation to 80% at 1kHz
Sensitivity: ± 5 V peak for maximum modulation
dc to 350 kHz
frequency)
(0 Vdc Offset)
(1 µHz-21 MHz carrier
Phase Modulation
(Typical)
Sinewave range: ± 900°, ± 5 V input
Sinewave linearity: ± 0.5%, best fit straight line up to ± 720° of
Squarewave range: ± 450°
Triangle range: ± 45°
Positive and negative
ramps range: ± 90°
Modulation frequency
range: DC to 5 kHz
modulation range
Sweep
Sweep: Frequency
Sweep sequence modes: Single, continuous
Sweep function modes:
Multi-Interval: (Up to 50 different intervals can be
Linear or logarithmic
sweep: (Can be set for each interval)
Sweep time:
Linear: 0.01 s to 105 s
Logarithmic: 0.1 s to 105 s
Minimum sweep width:
Linear: 0 Hz
Logarithmic: 1 decade
Maximum sweep width: Full frequency range
Minimum sweep rate:
Linear: 0.2 Hz/s
Phase continuity: Sweep is phase continuous over the full
Multi-marker:
Linear sweep only: Up to nine markers can be set in this one
Sweep time: 0.01 s to 105 s
Sweep width: From 0 Hz to full frequency range
sequenced and repeated in any order in a
sequence that can contain up to 100
intervals)
frequency range of the main output for all
sweep modes
dedicated interval