Agilent E1428A Data Sheet

Agilent E1428A
1 GSa/s VXI Oscilloscope
Data Sheet
1-Slot, C-size, message based
250 MHz bandwidth, 2 channels
1GSa/s maximum sample rate
Automatic pulse parameter measurements
Recommended for transient signals
The SCPI language is available for users who are more familiar with SCPI. Not all of the complex triggering capabilities of the 54510A and E1428A are implemented in the SCPI language.
Refer to the Agilent Technologies Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
Agilent E1428A
Description
The Agilent E1428A Digitizing Oscilloscope is a C-size, 1- slot, message-based VXI module. It has two channels
with each channel containing an 8-bit A/D, and 8,000 point memory to simultaneously capture at up to 1 GSa/s. This preserves the timing correlation of both channels without a reduction in sample rate.
For fast capture of many waveforms, the E1428A offers sequential single-shot mode which internally stores successive waveforms rapidly. The memory can be optimized for acquisition speed or capacity. There are 100 K-words of internal RAM or 500 K-words of shared RAM for segment storage.
The E1428A comes with both the SCPI command set and an Agilent 54510A-compatible command set. The compatible language provides a use model where programs can be developed using the 54510A bench oscilloscope. Using the 54510A provides direct visual feedback to the programmer during program development. Once programs have been developed on the 54510A, they can be run on the E1428A with only minor modifications.
E1428A Features
Built-in Automatic Test Features
The user can choose from a variety of powerful measurement features that are built into the E1428A, thus simplifying test program development and improving test times.
Nineteen automatic pulse parameter measurements are available in the box, freeing the controller from these calculations. These measurements conform to IEEE definitions. Maximum and minimum statistics on the values of these measurements are also accumulated.
The following measurements are made at user-defined thresholds or at the 10/90% or 50% voltage thresholds, as defined by IEEE Std 194-1977: rise time, fall time, frequency, period, +pulse width, –pulse width, duty cycle, delay, volts ampl, volts base, volts top, volts p-p, volts avg, volts max, overshoot, preshoot, volts min, Vac rms, Vdc rms.
GO/NO-GO testing is provided by another built-in feature: Measurement Limit Test lets the user set upper and lower limits on any of the module’s automatic measurements. If limits are exceeded, the violating waveform and its data can be stored or transferred to the controller.
In addition to the measurement limit test for GO/NO-GO testing, the E1428A provides built-in waveform compare test. With this feature,
stored template.
a live signal can be compared against a
PDFINFO H 5 5 8 1 - 0 1
Shared RAM On Board Eliminates Bus Transfer Time
1 MB of shared RAM allows storage and retrieval of waveform data as well as direct access to that data by the VXI controller or other instrument modules. Since the controller operates directly on the data in shared RAM, bus transfer of waveforms can be eliminated. Up to 45 milliseconds of bus transfer time can be saved.
Choice of Two Command Sets
SCPI Command Set—Based on the industry standard SCPI definitions for instrument programming.
54500-Compatible Command Set—Program the E1428A like the 54510B bench oscilloscopes.
Advanced Logic Trigger
Agilent’s advanced logic triggering can be used to trigger on glitches as well as a wide variety of user specified conditions. You can trigger on edge, pattern, time-qualified trigger, state, or trigger-after-delay to capture such elusive events as timing violations or infrequent bus phenomena. Using the built-in TV trigger, you can select line and field for a variety of video waveforms. And you can use the E1428A’s ability to send and receive VXI backplane triggers to combine the power of multiple VXI instruments.
Probes
The E1428A is compatible with the 54510B benchtop oscilloscope. Therefore, probes for the benchtop oscilloscope work with the VXI oscilloscope. Refer to the Agilent Test and Measurement catalog for probe ordering information.
E1428A Product Specifications
Specifications valid for temperature range ± 10 °C from software calibra­tion temperature with eight or more averages selected.
Bandwidth (–3 dB), repetitive: DC to 250 MHz Bandwidth (–3 dB), single­shot: DC to 250 MHz Maximum sample rate: 1 GSa/s (simultaneously on both ch.) Waveform record length: 8,000 points (real time)
Number of inputs: 2 simultaneous acquisition inputs,
Output BNC(s): Probe comp, Cal, Trigger out Vertical sensitivity range: 8 mV to 40 V full scale Vertical gain accuracy (DC): ± 1.25% of full scale Vertical resolution: ± 0.4% (8 bit A/D)
Input R (selectable): 1 Mor 50 Input C: 7 pF nominal Input coupling: AC, DC Maximum input voltage: 1 M: ± 250 V
Offset accuracy: ± (1% of channel offset, + 2% of full scale) Offset range: Up to ± 250 V Time base range: 10 ns to 50 s full scale Delta-t accuracy, repetitive: Delta-t accuracy, single­shot: Time base resolution: 20 ps Minimum trigger pulse width: VXI backplane triggering: ECL Number of instrument setups: 48 VXI shared memory, amount: 1 MB
500 points (repetitive)
1 external trigger input
(Note: E1428A has 2 acquisition inputs, both with 1 GSa/s A/D converters, so both repetitive and single-shot acquisitions are taken on both input channels simultaneously.)
± 0.1% (10 bits with averaging)
50 : 5 V rms
-6
± (0.005% x delta-t + 2 150 ps) for 8 averages
± (0.005% x delta-t + 2 150 ps)
Internal: 1.75 ns External: 2.8 ns
x delay setting +
-6
x delay setting +
2
PDFINFO H 5 5 8 1 - 0 2