Agilent E1411B Data Sheet

Page 1
Agilent E1411B
5.5-Digit Multimeter, C-Size
Data Sheet
• 1-Slot, C-size, register based
• DCV, ACV, 2- & 4-wire Ω, temperature
• 5.5-digit low-noise integrating A/D
• 13 kHz high-speed sampling A/D
• Software calibration
Description
The Agilent Technologies E1411B 5.5-Digit Multimeter is a C-size,1-slot, register-based VXI module. It is identical in electri­cal design to the E1326B, differing only in size. You can use the integrating A/D to make 5.5-digit, low-noise measurements, or switch to the sampling A/D to make 14-bit readings at rates up to 13 kHz.
When combined with any Agilent VXI relay or FET multiplexer, you can create a multichannel scanning multimeter. For example, by sending just one SCPI command to the E1406A, you can program the multimeter and the channels of your multiplexers all at one time. The E1411B provides flexible triggering with built-in timer pacer, also.
Product functions for this DMM include Vdc/ac, 2- and 4-wire
, offset-compensated , thermocouples thermistors,and
RTDs. This autoranging DMM is especially well suited for data acquisition and computer-aided test applications.
Refer to the Agilent Technologies Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
Page 2
2
Reading Rate
Auto zero off, fixed range, default trigger delay, offset comp off, Sample Source “TIMER” for rates >15 readings/s.
Maximum reading rate: 13 K
Typical Reading Rates (rdgs/s)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc voltage: 3 3.5 49 59 365 3125 13000 Four-wire
resistance: 3 3.5 49 59 365 3125 13000 ac voltage: 1.3 1.4 1.9 1.9 1.9 1.9 1.9
Resolution (bits/digit)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
Binary bits: ±22 ±22 ±20 ±20 ±18 ±15 ±14 Decimal
digits: 6.5 6.5 6 6 5.5 4.5 4
Noise Rejection (dB) for DC voltage and Resistance functions.
Noise Rejection Conditions: CMR measured with 1 kin both High and Low leads with a 10% imbalance. Low connted to Common at source,
measured with respect to earth ground. NMR is for specified frequencies ±0.1%
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc: Common mode rejection 150 dB 150 dB 150 dB 150 dB 150 dB 150 dB 150 dB ac (0-400): Common mode rejection* 70 dB 70 dB 70 dB 70 dB 70 dB 70 dB 70 dB
50 Hz: Power line cycles (NPLC) 16 1 ––––
Normal mode rejection 84 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB Effective Common Mode ECMR* 154 dB 70 dB 130 dB 70 dB 70 dB 70 dB 70 dB
60 Hz: Power line cycles (NPLCs) 16 1
Normal mode (60 Hz) rejection 0 dB 84 dB 0 dB 60 dB 0 dB 0 dB 0 dB Effective Common Mode Rejection* 70 dB 154 dB 70 dB 130 dB 70 dB 70 dB 70 dB
400 Hz: Power line cycles (NPLCs) 128 8 1
Normal mode (400 Hz) rejection 84 dB 0 dB 84 dB 0 dB 60 dB 0 dB 0 dB Effective Common Mode ECMR* 154 dB 70 dB 154 dB 70 dB 130 dB 70 dB 70 dB
*
64 & 300 volt ranges reduced by 36 dB
dc Voltage
Accuracy Conditions: Auto zero on,one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
(Volts) ± (% of Reading + Volts)
Range Input Resistance 20/16.7 ms 10 µs 20/16.7 ms 10 µs
125 mV >100 M 120 nV 7.6 µV 0.023% + 5 µV 0.115% + 60 µV 1 V >100 M 1.0 µV 61 µV 0.013% + 15 µV 0.1% + 200 µV 8 V >100 M 7.6 µV 488 µV 0.01% + 50 µV 0.1% + 1.5 µV 64 V 10 M+ 5% 61 µV 3.9 mV 0.015% + 1 mV 0.1% + 20 µV 300 V 10 M+ 5% 488 µV 31 mV 0.015% + 5 mV 0.1% + 80 µV
Product Specifications
Page 3
3
Four-Wire Resistance
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
Source Maximum Open () ± (% of Reading + )
Range Current Circuit Voltage 20/16.7 ms 10 µs 20/16.7 ms 10 µs
256 488 µA 11.5 V 250 µ 15 m 0.035% + 10 m 0.12% + 50 m 2 k 488 µA 11.5 V 2 m 125 m 0.025% + 20 m 0.1% + 200 m 16 k 61 µA 11.5 V 15 m 1 0.025% + 200 m 0.1% + 2 131 k 61 µA 11.5 V 125 m 8 0.025% + 1 0.1% + 16 1 M 7.6 µA 11.5 V 1 64 0.015% + 10 0.1% + 120
NOTE: With offset compensation on, accuracy is the same as for the voltmeter alone.
True RMS ac Voltage (ac coupled)
Crest Factor: 7 at 0% full scale; 5 at full scale. Accuracy Conditions: Sine wave inputs >0% of full scale. dc component <0% of ac component. Auto-zero on, one hour warmup. Temperature within ± 5°C of calibration temperature (module calibrated at 8-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
Range Input (Volts) ± (% of Reading + Volts) (RMS) Impedance Frequency 320/267 ms 10 µs 320/267 ms All other apertures
87.5 mV >100 M, 20-50 Hz 30 nV 7.6 µV 2.175% +200 µV 2.175% +1 mV
<100 pF 50 Hz-1 kHz 0.675% +200 µV 0.675% +200 µV
5-10 kHz 3.175% +200 µV 3.175% +200 µV
700 mV >100 M, 20-50 Hz 0.24 µV 61 µV 2.125% +1.5 mV 2.125% +8 mV
<100 pF 50 Hz-1 kHz 0.625% +1.5 mV 0.625% +1.5 mV
1-5 kHz 0.625% +1.5 mV 0.625% +1.5 mV
5-10 kHz 3.125% +1.5 mV 3.125% +1.5 mV
5.6 V >100 M, 20-50 Hz 2.0 µV 488 µV 2.125% +15 mV 2.125% +80 mV
<100 pF 50 Hz-1 kHz 0.625% +15 mV 0.625% +15 mV
1-5 kHz 1.125% +15 mV 1.125% +15 mV
5-10 kHz 10.125% +15 mV 10.125% +15 mV
44.8 V 10 M± 5%, 20-50 Hz 15 µV 3.9 mV 2.125% +100 mV 2.125% + 500 mV
<100 pF 50 Hz-1 kHz 0.625% +100 mV 0.625% +100 mV
1-5 kHz 1.125% +100 mV 1.125% +100 mV
5-10 kHz 10.125% +100 mV 10.125% +100 mV
300 V 10 M± 5%, 20-50 Hz 122 µV 31 mV 2.125% + 500 mV 2.125% + 2.5 V
<100 pF 50 Hz-1 kHz 0.625% + 500 mV 0.625% + 500 mV
1-5 kHz 1.125% + 500 mV 1.125% + 500 mV
5-10 kHz 10.125% + 500 mV 10.125% + 500 mV
Timing/Synchronization
Timer/pacer
Timer range: 76 µs to 65.5 ms Resolution: 2 µs
Programmable delay
Delay range: 40 µs to 16 s Resolution: 2 µs
External trigger
Minimum pulse width: 100 ns Maximum trigger rate: 5 kHz (Trigger Condition, negative
edge; Fixed range,10 µs aperture)
Isolation 450 Vpk between any terminal
and chassis.
dc Voltage Accuracy with Relay Multiplexers
Range 20/16.7 ms 10 µs 20/16.7 ms 10 µs
125 mV 0.023%+9 µV 0.115%+64 µV 0.023%+55 µV 0.115%+110 µV 1 V 0.013%+19 µV 0.1%+204 µV 0.013%+65 µV 0.1%+250 µV 8 V 0.01%+54 µV 0.1%+1.5 mV 0.01%+100 µV 0.1%+1.55 mV 64 V 0.015%+1 mV 0.1%+20 mV 0.015%+1.05 mV 0.1%+20 mV 300 V 0.015%+5 mV 0.1%+80 mV 0.015%+5.05 mV 0.1%+80 mV
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 8-28°C).
Page 4
4
True RMS ac Voltage (ac coupled)with Relay Multiplexers
1-5 kHz and 5-10 kHz frequencies (all apertures) when using Relay Multiplexers (E1343A, E1345A, E1346A, or E1347A). Add 0.2% to the ac Voltage specifications.
Strain Measurements with Strain Relay Multiplexers
All measurements are made using the MEAS command.
NOTE: The Agilent E1406A command module and embedded controllers
provide units conversion; if the E1411B is register programmed, your program must make the units conversion.
Vs = 5 V Power Supply/Gage Factor = 2
18-20°C Temp. Coefficient
µe% %e µe% %e
Relays Quarter 25.8 0.023 1.96 0.006
Half 33.2 0.023 0.23 0.006 Full 16 0.023 0.053 0.006
FETs Quarter 353 0.023 103 0.006
Half 169 0.023 50.7 0.006 Full 83.8 0.023 25.3 0.006
Vs = 1 V Power Supply/Gage Factor = 2
18-20°C Temp. Coefficient
µe% %e µe% %e
Relays Quarter 25.8 0.023 1.96 0.006
Half 5.39 0.023 0.23 0.006 Full 2.07 0.023 0.053 0.006
FETs Quarter 52.9 0.023 12 0.006
Half 18.9 0.023 5.27 0.006 Full 8.85 0.023 2.57 0.006
Vs = 0.1 V Power Supply/Gage Factor = 2
18-20°C Temp. Coefficient
µe% %e µe% %e
Relays Quarter 81.3 0.023 1.96 0.006
Half 33.2 0.023 0.23 0.006 Full 16 0.023 0.053 0.006
FETs Quarter 353 0.023 103 0.006
Half 169 0.023 50.7 0.006 Full 83.8 0.023 25.3 0.006
Four-Wire Resistance with Relay Multiplexers
Accuracy Conditions: Auto zero on, one hour warmup, temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
NOTE: With offset compensation on, accuracy is the same as for
the voltmeter alone.
90-DAY ACCURACY vs APERTURE
±(% of reading + Ω)
E1326B & E1345A/47A
Range 20/16.7 ms 10 µs
256 0.035%+18.2 m 0.12%+58.2 m 2 k 0.025%+28.2 m 0.1%+208 m 16 k 0.025%+266 m 0.1%+2.1 131 k 0.025%+1.1 0.1%+16.1 1 M 0.025%+10.5 0.1%+121
NOTE: Accuracy data includes all errors contributed by the multimeter, analog
bus ribbon cables, multiplexer, and transducer linearization (if applicable). The accuracies do not include transducer accuracy errors.
Page 5
5
Temperature
The temperature accuracy graphs (below) include instrument and firmware linearization errors. The linearization algorithm used is based on the ITS-90 standard transducer curves. Add your transducer accuracy to determine total measurement error.
NOTE: The E1406A command modules and Agilent embedded VXI controllers
provide units conversion; if the E1411B is register-programmed, your program must make the necessary units conversion.
Thermocouple: (E1411B Multimeters and E1347A / E1476A TC MUX)
16 ms aperture (1 PLC): 100 µs aperture:
Thermistors: (E1411B Multimeters and E1345A/E1347A/E1476A MUXs)
4-wire Ω:
16 ms aperture (1 PLC): 100 µs aperture:
Page 6
6
RTDs: (E1411B Multimeters and E1345A / E1476A MUXs)
4-wire Ω:
16 ms aperture (1 PLC): 100 µs aperture:
Functions
Idc: Iac: Frequency: Period: Temp.: TmTcRTD
General Specifications
VXI Characteristics
VXI device type: Register based Data transfer bus: Not specified Size: C Slots: 1 Connectors: P1/2 Shared memory: Yes, shared memory available
with E1406A SCPI driver
VXI buses: TTL Trigger Bus
Instrument Drivers
See the Agilent Technologies Website
(http://www.agilent.com/find/inst_drivers)
for driver availability and downloading
Command module firmware: Downloadable Command module firmware rev: A.02 I-SCPI Win 3.1: Yes I-SCPI Series 700: Ye s C-SCPI LynxOS: Ye s C-SCPI Series 700: Ye s Panel Drivers: Ye s VXIplug&play Win Framework: Yes VXIplug&play Win 95/NT Framework: Ye s VXIplug&play HP-UX Framework: No
Module Current
I
pm
I
dm
+5 V: 0.2 0.01 +12 V: 0.55 0.01
-12 V: 00 +24 V: 00
-24 V: 00
-5.2 V: 00
-2 V: 00
Cooling Slot
Watts/slot: 8.50
P mm H
2
O: 0.14
Air Flow liter/s: 0.71
Ordering Information
Description Product Number
5.5-Digit Multimeter, High-Accuracy, C-Size E1411B
Service Manual E1411B 0B3
Japan - Japanese Localization E1411B ABJ
ANSI Z540 Compliant Calibration E1411B A6J
Page 7
By internet, phone, or fax, get assistance with all your test & measurement needs.
Online assistance:
www.agilent.com/find/assist
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent’s overall support policy: "Our Promise" and "Your Advantage."
Our Promise
"Our Promise" means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you’re choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engi­neers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
"Your Advantage" means that Agilent offers a wide range of additional expert test and measurement serv­ices, which you can purchase according to your unique technical and business needs. Solve problems effi­ciently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-war­ranty repairs, and on-site education and training, as well as design, system integration, project manage­ment, and other professional services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and sys­tems, and obtain dependable measurement accuracy for the life of those products.
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
For more assistance with your test & measurement needs or to find your local Agilent office go to
:
www.agilent.com/find/assist
Phone or Fax
United States:
(tel) 1 800 829 4444
Canada:
(tel) 1 877 894 4414 (fax) (905) 282 6495
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115
Latin America:
(tel) (305) 269 7500 (fax) (305) 269 7599
Taiwan:
(tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6836 0252 (e-mail) tm_asia@agilent.com
Data Subject to Change © Agilent Technologies, Inc. 2004 Printed in the U.S.A. May 1, 2004 5965-5562E