6.5-Digit Multimeter, High­Accuracy, C-Size HP E1410A
Technical Specifications
1-Slot, C-size, message based
Vdc/ac, 2- & 4-wire
Quality measurements with high common mode rejection
True RMS from 20 Hz to 1 MHz
Software calibration
Description
The HP E1410A 6.5-Digit Multimeter is a C-size, 1-slot, message-based VXI module. It offers DC voltage and
resistance measurements at rates over 1,450 readings/s in the 3.5-digit mode or normal mode rejection of up to 90 dB in the 6.5-digit mode. You can measure true RMS AC signals from 20 Hz to 1 MHz with programmable settling times. Offset compensated ohms allows for quality resistance measurements by eliminating the effect of small series voltage offsets. Temperature measurements using thermistors and RTDs are supported. Resolution, accuracy, and noise rejection may be set to optimize measurements speed. Extensive triggering is available.
HP E1410A
Refer to the HP Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
PDFINFO H5 56 1- 0 2
Accuracy
Choose the resolution, accuracy, and noise rejection you need. Fast function-range changes allow you to optimize measurement speeds. For low-level signals, long integration times and the system guard yield the most accurate measurements possible. Integration times from 100 power line cycles to 0.0005 power line cycles are selectable for all functions. Resolution from
6.5 to 3.5 digits is selectable as a function of integration time.
Product Specifications
Reading Rate/Resolution
Max. reading rate: 1.45 k
Auto zero off, fixed range, delay 0, AC slow filter on, and offset compensation off.
Flexible Triggering
The DMM has extensive triggering capabilities, including synchronization with external devices. You can access the external trigger and voltmeter complete signals from the front panel or VXIbus (TTL trigger lines).
Typical Reading Rates (rdgs/s)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC voltage 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450 Four-wire resistance 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450 AC voltage 0.17 0.2 0.65 0.7 1.0 1.0 1.0 1.0 1.0 1.0
Aperture
Resolution (bits/digits)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
Binary bits: ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 18.2 ± 18.2 ± 14.9 ± 11.5 Decimal digits: 6.5 6.5 6.5 6.5 6.5 6.5 5.5 5.5 4.5 3.5
Aperture
Noise Rejection (dB)
Noise Rejection Conditions: 1 kimbalance in low lead. NMR is for specified frequencies ± 0.08%.
DC Voltage & Resistance:
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC Common mode rejection 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 50 Hz Power line cycles 100— 10 —1 — —— ——
Normal mode (50 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB 60 Hz Power line cycles 100 10 1 Normal mode (60 Hz) rejection 0 dB 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 400 Hz Power line cycles 800— 80—8 — —— —— Normal mode (400 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB
AC voltage
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC to 60 Hz Common mode rejection >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB
PDFINFO H5 56 1- 0 3
DC Voltage Resolution/Accuracy
Accuracy Conditions: Auto-zero on. One hour warmup. Temperature within ± 5 °C of temperature at calibration (module calibrated at 18-28 °C).
Range Input Resistance Resolution vs Aperture
(Volts)
20/16.7 ms 10 µs 20/16.7 ms 10 µs
30 mV >10 G 10 nV 10 µV 0.0040% + 3.9 µV 0.0040% + 60 µV 300 mV >10 G 100 nV 100 µv 0.0025% + 4.0 µV 0.0025% + 400 µV 3 V >10 G 1 µv 1 mV 0.0017% + 9.0 µV 0.0017% + 4.0 mV 30 V 10 M± 1% 10 µv 10 mV 0.0035% + 200 µV 0.0035% + 40 mV 300 V 10 M± 1% 100 µv 100 mV 0.0063% + 700 µV 0.0050% + 400 mV
DC voltage: 300 V Voltage accuracy (DC): 0.002%
90-Day Accuracy vs Aperture
± (% of Reading + Volts)
Four-Wire Resistance
Source
Maximum Open
Current
Range 20/16.7 ms 10 µs 20/16.7 ms 10 µs
30 1 mA 12 V 10 µΩ 10 m 0.0065% + 4.5 m 0.0065% + 60 m 300 1 mA 12 V 100 µΩ 100 m 0.0045% + 4.5 m 0.0045% + 400 m 3 k 1 mA 12 V 1 m 1 Ω 0.0035% + 7 m 0.0035% + 4 30 k 100 µA 12 V 10 m 10 Ω 0.0035% + 70 m 0.0035% + 40 300 k 10 µA 12 V 100 m 100 Ω 0.0040% + 900 m 0.0040% + 400 3 M 1 µA 12 V 1 1 k 0.0055% + 16 0.0055% + 5 k 30 M 100 nA 8.5 V 10 10 k 0.0250% + 930 0.0250% + 50 k 300 M 100 nA 8.5 V 100 100 k 1.6% + 100 k not specified 3 G 100 nA 8.5 V 1 k 1 M 16% + 1 M not specified
Circuit Voltage
Resolution vs. Aperture
(Volts)
90-Day Accuracy vs. Aperture
± (% of reading = Ω)
Accuracy conditions: Auto-zero on, one hour warmup. On 300 Mand 3 Granges, specification applies to two-wire only, with inputs >10% of full scale and within 24 hrs of internal calibration. Temperature within ± 5 °C of temperature at calibration (module calibrated at
18-28 °C).
2/4-wire Ω: 3 G
True RMS AC Voltage (AC coupled)
Crest Factor: 3.5 at full scale. Accuracy Conditions: Sine wave inputs >10% of full scale. DC component <10% of AC component. AC slow fil-
ter on. Auto-zero on. One hour warmup. Temperature within ± 5 °C of temperature at calibration (module calibrated at 18-28 °C).
Range (RMS)
30 mV 1 M± 1%,
300 mV 1 M± 1%,
3 V 1 M± 1%,
30 V 1 M± 1%,
300 V 1 M± 1%,
Input Impedance
< 90 pF
< 90 pF
< 90 pF
< 90 pF
< 90 pF
Resolution Aperture = 20/16.7 ms
10 nV 20 Hz-45 Hz
100 nV 20-45 Hz
1 µV 20-45 Hz
10 µV 20-45 Hz
100 µV 20-45 Hz
Frequency 90-Day Accuracy
45-100 Hz 100 Hz-20 kHz 20-100 kHz 100-300 kHz 300 kHz-1 MHz
45-100 Hz 100 Hz-20 kHz 20-100 kHz 100-300 kHz 300 kHz-1 MHz
45-100 Hz 100 Hz-20 kHz 20-100 kHz 100-300 kHz 300 kHz-1 MHz
45-100 Hz 100 Hz-20 kHz 20-100 kHz 100-300 kHz 300 kHz-1 MHz
45-100 Hz 100 Hz-20 kHz 20-100 kHz 100 kHz-1 MHz
± (% of reading + Volts) Aperture = 20/16.7 ms
0.58% + 37.3 µV
0.23% + 37.3 µV
0.15% + 37.3 µV
0.68% + 47.1 µV
3.35% + 123 µV
10.35% + 691 µV
0.58% + 133 µV
0.23% + 133 µV
0.15% + 133 µV
0.68% + 231 µV
3.35% + 991 µV
10.35% + 6.67 mV
0.58% + 1.33 mV
0.23% + 1.33 mV
0.15% + 1.33 mV
0.68% + 2.31 mV
3.35% + 9.91 mV
10.35% + 66.7 mV
0.58% + 13.3 mV
0.23% + 13.3 mV
0.15% + 13.3 mV
0.68% + 23.1 mV
3.35% + 99.1 mV
10.35% + 667 mV
0.64% + 133 mV
0.29% + 133 mV
0.21% + 133 mV
1.08% + 390 mV not specified
PDFINFO H5 56 1- 0 4
AC voltage: 300 V Voltage accuracy (AC): 0.194%
Frequency and Period
Sensitivity (sinewave): 10 mV rms Trigger level: Triggers and counts on zero crossings Conditions: 0-55 °C
Frequency Range Period Range 1 Year Accuracy ±
10-400 Hz 0.1-0.025 s 0.05% 400 Hz-1.5 MHz 0.025 s-667 ns 0.01%
(% of Reading)
Timing/Synchronization
Timer/pacer:
Timer range: 600 µs to 2100 s Resolution: 1.0 µs
Programmable delay:
Delay range: 0 to 2100 s
Resolution: 1.0 µs
External trigger:
Trigger condition (programmable): Negative or positive edge
Minimum pulse width: 10 ns
Memory
Reading storage: 4,096 readings Multimeter state memory: 10 states
Instrument Drivers
See the HP Website (http://www.hp.com/go/inst_drivers) for driver availability and downloading.
Command module firmware: n/a
Command module firmware rev: n/a
I-SCPI Win 3.1: n/a I-SCPI Series 700: n/a C-SCPI LynxOS: n/a C-SCPI Series 700: n/a HP Panel Drivers: Yes
plug&play
VXI Framework: Yes
plug&play
VXI Framework: Yes
plug&play
VXI Framework: No
Win
Win 95/NT
HP-UX
Module Current
I
PM
+5 V: 1 0.1 +12 V: 0.5 0.15 –12 V: 00 +24 V: 00 –24 V: 00 –5.2 V 00 –2 V: 00
I
DM
Functions
Idc: Iac: Frequency: 1.5 MHz
Period: 1 µs
Temp.: Tm, RTD
General Specifications
VXI Characteristics
VXI device type: Message based Size: C Slots: 1 Connectors: P1/2 Shared memory: Yes VXI busses: TTL Trigger Bus C-size compatibility: n/a
Cooling/Slot
Watts/slot: 11.00
P mm H
Air Flow liter/s: 0.92
O: 0.15
2
Ordering Information
Description Product No.
6.5-Digit Multimeter, High Accuracy HP E1410A Service Manual HP E1410A 0B3 Mil Std 45662A Calibration w/Test Data HP E1410A 1BP Japan - Japanese Localization HP E1410A ABJ 3 Yr. Retn. to HP to 1 Yr. OnSite Warr. HP E1410A W01
PDFINFO H5 56 1- 0 5
Related Literature
1998 Test System and VXI Products Data Book,
HP Pub. No. 5966-2812E
1999 Test System and VXI Products Catalog,
HP Pub. No. 5968-3698
Warranty
Standard Hewlett-Packard VXIbus hardware products are warranted against defects in materials and workmanship for a period of three years unless otherwise noted. HP software and firmware products that are designated by HP for use with a hardware product, when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in materials and workmanship.
For more information about Hewlett­Packard test & measurement products, applications, services, and for a current sales office listing, visit our website,
http://www.hp.com/go/tmdir
. You can also
contact one of the following centers and ask for a test & measurement sales representative.
United States:
Hewlett-Packard Company Test and Measurement Call Center P.O. Box 4026 Englewood, CO 80155-4026 1 800 452 4844
Canada:
Hewlett-Packard Canada Ltd. 5150 Spectrum Way Mississauga, Ontario L4W 5G1 (905) 206 4725
For a complete and detailed warranty statement please see
Test System and VXI Products Data Book
the HP
or visit the
HP Website at http://www.hp.com/go/vxi.
Website Directory
HP VXI Product Information
http://www.hp.com/go/vxi
HP VXI Channel Partners
http://www.hp.com/go/vxichanpart
HP VEE Application Website
http://www.hp.com/go/hpvee
Data Acquisition and Control Website
http://www.hp.com/go/data_acq
HP Instrument Driver Downloads
http://www.hp.com/go/inst_drivers
Electronics Manufacturing Test Solutions
http://www.hp.com/go/manufacturing
Europe:
Hewlett-Packard European Marketing Centre P.O. Box 999 1180 AZ Amstelveen The Netherlands (31 20) 547 9900
Japan:
Hewlett-Packard Japan Ltd. Measurement Assistance Center 9-1, Takakura-Cho, Hachioji-Shi, Tokyo 192, Japan Tel: (81) 426 56 7832 Fax: (81) 426 56 7840
Latin America:
Hewlett-Packard Latin American Region Headquarters 5200 Blue Lagoon Drive, 9th Floor Miami, Florida 33126 U.S.A. Tel: (305) 267-4245
(305) 267-4220
Fax: (305) 267-4288
Australia/New Zealand:
Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackburn, Victoria 3130 Australia 1 800 629 485
Asia Pacific:
Hewlett-Packard Asia Pacific Ltd. 17-21/F Shell Tower, Times Square, 1 Matheson Street, Causeway Bay, Hong Kong Tel: (852) 2599 7777 Fax: (852) 2506 9285
Data Subject to Change Copyright © April 1999 Hewlett-Packard Company HP Publication No.: 5965-5561E
Loading...