PDFINFO H5 56 1- 0 1
6.5-Digit Multimeter, HighAccuracy, C-Size
HP E1410A
Technical Specifications
●
1-Slot, C-size, message based
●
Vdc/ac, 2- & 4-wire Ω
●
Noise rejection with long integration
times/ guarding
●
Quality measurements with high common mode
rejection
●
True RMS from 20 Hz to 1 MHz
●
Software calibration
Description
The HP E1410A 6.5-Digit Multimeter is a C-size, 1-slot,
message-based VXI module. It offers DC voltage and
resistance measurements at rates over 1,450 readings/s
in the 3.5-digit mode or normal mode rejection of up to
90 dB in the 6.5-digit mode. You can measure true RMS
AC signals from 20 Hz to 1 MHz with programmable
settling times. Offset compensated ohms allows for
quality resistance measurements by eliminating the
effect of small series voltage offsets. Temperature
measurements using thermistors and RTDs are
supported. Resolution, accuracy, and noise rejection
may be set to optimize measurements speed. Extensive
triggering is available.
HP E1410A
Refer to the HP Website for instrument driver
availability and downloading instructions, as well as for
recent product updates, if applicable.

PDFINFO H5 56 1- 0 2
Accuracy
Choose the resolution, accuracy, and noise rejection
you need. Fast function-range changes allow you to
optimize measurement speeds. For low-level signals,
long integration times and the system guard yield the
most accurate measurements possible. Integration
times from 100 power line cycles to 0.0005 power line
cycles are selectable for all functions. Resolution from
6.5 to 3.5 digits is selectable as a function of integration
time.
Product Specifications
Reading Rate/Resolution
Max. reading rate: 1.45 k
Auto zero off, fixed range, delay 0, AC slow filter on, and offset compensation off.
Flexible Triggering
The DMM has extensive triggering capabilities,
including synchronization with external devices. You
can access the external trigger and voltmeter complete
signals from the front panel or VXIbus (TTL trigger
lines).
Typical Reading Rates (rdgs/s)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC voltage 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450
Four-wire resistance 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450
AC voltage 0.17 0.2 0.65 0.7 1.0 1.0 1.0 1.0 1.0 1.0
Aperture
Resolution (bits/digits)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
Binary bits: ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 18.2 ± 18.2 ± 14.9 ± 11.5
Decimal digits: 6.5 6.5 6.5 6.5 6.5 6.5 5.5 5.5 4.5 3.5
Aperture
Noise Rejection (dB)
Noise Rejection Conditions: 1 kΩ imbalance in low lead. NMR is for specified frequencies ± 0.08%.
DC Voltage & Resistance:
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC Common mode rejection 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB
50 Hz Power line cycles 100— 10 —1 — —— ——
Normal mode (50 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB
60 Hz Power line cycles — 100 — 10 — 1 — — — —
Normal mode (60 Hz) rejection 0 dB 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB
400 Hz Power line cycles 800— 80—8 — —— ——
Normal mode (400 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB
AC voltage
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC to 60 Hz Common mode rejection >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB