Agilent E1410A Technical Specifications

PDFINFO H5 56 1- 0 1
6.5-Digit Multimeter, High­Accuracy, C-Size HP E1410A
Technical Specifications
1-Slot, C-size, message based
Vdc/ac, 2- & 4-wire
Quality measurements with high common mode rejection
True RMS from 20 Hz to 1 MHz
Software calibration
Description
The HP E1410A 6.5-Digit Multimeter is a C-size, 1-slot, message-based VXI module. It offers DC voltage and
resistance measurements at rates over 1,450 readings/s in the 3.5-digit mode or normal mode rejection of up to 90 dB in the 6.5-digit mode. You can measure true RMS AC signals from 20 Hz to 1 MHz with programmable settling times. Offset compensated ohms allows for quality resistance measurements by eliminating the effect of small series voltage offsets. Temperature measurements using thermistors and RTDs are supported. Resolution, accuracy, and noise rejection may be set to optimize measurements speed. Extensive triggering is available.
HP E1410A
Refer to the HP Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
PDFINFO H5 56 1- 0 2
Accuracy
Choose the resolution, accuracy, and noise rejection you need. Fast function-range changes allow you to optimize measurement speeds. For low-level signals, long integration times and the system guard yield the most accurate measurements possible. Integration times from 100 power line cycles to 0.0005 power line cycles are selectable for all functions. Resolution from
6.5 to 3.5 digits is selectable as a function of integration time.
Product Specifications
Reading Rate/Resolution
Max. reading rate: 1.45 k
Auto zero off, fixed range, delay 0, AC slow filter on, and offset compensation off.
Flexible Triggering
The DMM has extensive triggering capabilities, including synchronization with external devices. You can access the external trigger and voltmeter complete signals from the front panel or VXIbus (TTL trigger lines).
Typical Reading Rates (rdgs/s)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC voltage 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450 Four-wire resistance 0.4 0.49 4.0 4.9 47 56 312 360 1250 1450 AC voltage 0.17 0.2 0.65 0.7 1.0 1.0 1.0 1.0 1.0 1.0
Aperture
Resolution (bits/digits)
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
Binary bits: ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 21.5 ± 18.2 ± 18.2 ± 14.9 ± 11.5 Decimal digits: 6.5 6.5 6.5 6.5 6.5 6.5 5.5 5.5 4.5 3.5
Aperture
Noise Rejection (dB)
Noise Rejection Conditions: 1 kimbalance in low lead. NMR is for specified frequencies ± 0.08%.
DC Voltage & Resistance:
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC Common mode rejection 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 140 dB 50 Hz Power line cycles 100— 10 —1 — —— ——
Normal mode (50 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB 60 Hz Power line cycles 100 10 1 Normal mode (60 Hz) rejection 0 dB 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 400 Hz Power line cycles 800— 80—8 — —— —— Normal mode (400 Hz) rejection 90 dB 0 dB 80 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB 0 dB
AC voltage
2.0 s 1.67 s 200 ms 167 ms 20 ms 16.7 ms 2.0 ms 1.67 ms 100 µs 10 µs
DC to 60 Hz Common mode rejection >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB >86 dB
Loading...
+ 3 hidden pages