2
Ordering Information
Description Product No.
VXI GPIB Command Module, C-Size E1406A
Service manual E1406A 0B3
Taiwan - Chinese localization E1406A AB0
Korea - Korean localization E1406A AB1
China - Chinese localization E1406A AB2
Germany - German localization E1406A ABD
Spain - Spanish localization E1406A ABE
France - French localization E1406A ABF
Japan - Japanese localization E1406A ABJ
Italy - Italian localization E1406A ABZ
Instrument Drivers - See the Agilent Technologies Website
(http://www.agilent.com/find/inst_drivers) for driver availability and
downloading.
Command module firmware: n/a
Command module firmware rev: n/a
I-SCPI Win 3.1: n/a
I-SCPI Series 700: n/a
C-SCPI LynxOS: n/a
C-SCPI Series 700: n/a
Panel Drivers: Yes
VXI
plug&play
Win Framework: Yes
VXI
plug&play
Win95/NT Framework: Yes
VXI
plug&play
HP-UX Framework: No
Module Current
I
PM
I
DM
+5 V: 0.1 0.1
+12 V: 00
–12 V: 00
+24 V: 00
–24 V: 00
–5.2 V: 00
–2 V: 00
Cooling/Slot
Watts/slot: 19.00
∆P mm H
2
O: 0.30
Air Flow liter/s: 1.50
Product Specifications
General Characteristics
Processor: 16 MHz 68000
Interface: IEEE-488
Clock accuracy: 0.005% elapsed time
Memory: 1.75 MB FLASH max.
2 MB RAM max.
Shared memory: 512 KB max.
Agilent IBASIC: Yes
Slot 0 functions: Yes
Resource manager: Yes
MXIbus resource manager: Yes
CLK10: Yes
Temperature coefficient: 0.001%, to 0.012% of time since last
set
(per° C change in temperature)
Resolution: 1.0 s
CLK10
Input: TTL or low-level ac
Minimum input level: 40 mV p-p
Maximum input level: 42.5 V p-p
Output: TTL
Jitter: 0.03% (–55 dB)
Initial accuracy: 50 ppm
Maximum stability: ± 20 ppm/year (0-55° C)
Typical stability: ± 3 ppm/year at 25° C
Trigger Input
Levels: TTL
Input load: 55 kΩ, 50 pF
Maximum rate: 12.5 MHz (TTL), 40 MHz (ECL)
Minimum pulse width: 30 ns (TTL), 12.5 ns (ECL)
Maximum trigger delay: 30 ns
RS-232 Interface (for terminal only)
Baud rate: 300, 1200, 2400, 4800, 9600, 19200
Parity: Even, Odd, One, Zero, None
Character size: 7, 8
Pace: Xon/Xoff, None
Hardware handshake: DTR, RTS
Buffer size: 16 characters
RS-232 terminals supported: HP 700/92, HP 700/94, HP 700/22,
HP 700/43; WYSE WY-30, DEC VT100,
DEC VT220
VXI compliance: Revision 1.4 compliance
General Specifications
VXI device type: Message-based commander
Size: C
Slots: 1
Connectors: P1/2
Shared memory: Yes
VXI buses: Local Bus A
TTL Trigger Bus
ECL Trigger Bus
VXI Characteristics