3
87405B Product Specifications
Frequency range 10-4000 MHz
Small signal gain 22-27 dB min-max
Gain flatness ±1.5 dB 10-4000 MHz
Output power (1 dB compression pt.) 8 dBm at 4 GHz
Noise figure 5 dB 10-4000 MHz
Third order intercept (TOI) 15 dBm typical
Harmonics 30 dBc typical
(@ +4 dBm output power)
Input VSWR 1.5:1 10-1000 MHz
1.7:1 1000-2000 MHz
2.1:1 2000-3000 MHz
2.2:1 3000-4000 MHz
Output VSWR 1.9:1 10-4000 MHz
Impedance 50 ohms nominal
Reverse isolation 40 dB typical
Survival input power (max) +10 dBm typical
Power dissipation 1.6 W nominal
Temperature (operating) 0 to +55 °C
(storage) -40 to +70 °C
EMC
IEC 61326:1997/
EN 61326:1997
CISPR 11:1997
This ISM device complies with Canadian ICES-001.
AC line harmonic current emissions IEC/EN 61000-3-2
Line voltage interrupt (1 cycle, 100%) IEC/EN 61000-4-11
Surge test (1.2 x 50us, 0.5/1 kV) IEC/EN 61000-4-5
Electrical fast transients IEC/EN 61000-4-4
Radiated emissions CISPR 11, Class A
Radiated immunity (3 V/m, 80-1000 MHz) IEC/EN 61000-4-3
Conducted emissions CISPR 11, Class A
Conducted immunity (3 V, 0.15-80 MHz) IEC/EN 61000-4-6
ESD (4 kV contact, 8 kV air discharge) IEC/EN 61000-4-2
General Specifications
Bias voltage and current 15 ±6% Vdc @ 105 mA nominal
Connectors RF Type N(f) in, N(m) out
DC Probe Power Connector (f)
Weight net 233 g, shipping 546 g
Environmental Information
Moisture resistance 65 °C at 95% RH for 10 days per
Mil-Std-883C method 1004.5
Altitude, non-operating 15,000 meters per Mil-Std-883C
method 1001-C
Reliability
Random vibration 5.2 G (rms) to 2000 Hz per Mil-Std-883C
method 2026-11A
Shock 1500 G (peak), 0.5 ms per Mil-Std-883C method 2002.3-B
Operating temperature (0 to +55 °C) Class B2/B1 HPETM 757
Storage temperature (-40 to +70 °C) Class B HPETM 757
5 Day Op. humidity cycle (40C/95% RH) Class A2, B1, B2 HPETM
758, 4.6
Product safety evaluation CSA 61010-1
Specifications describe the instrument’s warranted performance
over the temperature range 0 °C to +55 °C (unless otherwise
noted). All specifications apply after the instrument’s temperature has been stabilized after one hour continuous operation.
Typical characteristics are intended to provide information useful
in applying the instrument by giving typical but nonwarranted
performance parameters. These are denoted as “typical” or
“nominal” and apply over the temperature range 0 °C to +55 °C.
Caution: Electrostatic Discharge (ESD) can damage or destroy
electronic compnents. It is recommended that this preamplifier,
like other electronic components, be installed and operated at a
static-free workstation or in an environment where precautions
against ESD have been implemented.