This information contained in this document is subject to change without notice.
Agilent Techologies makes no warranty of any kind with regard to this material, including but not
limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent
Techologies shall not be liable for errors contained herein or for incidental or consequential
damages in connection with the furnishing, performance, or use of this material.
Windows is a registered trademark of Microsoft Corporation.
TORX is a registered trademark of TORX Products, division of Camcar/Textron Corporation.
Certification
Agilent Techologies certifies that this product met its published specifications at the time of
shipment from the factory. Agilent Techologies further certifies that its calibration measurements
are tractable to the United States National Institute of Standards and Technology, to the extent
allowed by the Institute’s calibration facility, and to the calibration facilities of other International
Standards Organization members.
Regulatory Information
The regulatory information is in the User’s Guide supplied with the analyzer.
Safety, Warranty, and Assistance
Refer to Chapter 9 for information on safety, warranty, and assistance.
How to Use This Guide
This guide uses the following conventions:
FRONT PANEL KEY
: This indicates a “softkey,” a key whose label is determined by the instruments
firmware, and is displayed on the right side of the instrument’s screen next to the eight
unlabeled keys.
Screen Text: This indicates text displayed on the instrument’s screen.
This guide contains servicing information for the following instruments:
• 8712ET
• 8712ES
: This represents a key physically located on the instrument.
• 8714ET
• 8714ES
iii
8712ET/ES and 8714ET/ES Network Analyzer
Documentation Map
The CDROM provides the contents of all of the documents listed below.
The User’s Guide shows how to make measurements, explains commonly-used
features, and tells you how to get the most performance from the analyzer.
The LAN Interface User’s Guide Supplement shows how to use a local area
network (LAN) for programming and remote operation of the analyzer.
The Automating Measurements User’s Guide Supplement provides
information on how to configure and control test systems for automation of test
processes.
The Programmer’s Guide provides programming information including GPIB
and SCPI command references, as well as short programming examples.
The Example Programs Guide provides a tutorial introduction using
BASIC programming examples to demonstrate the remote operation of
the analyzer.
The Service Guide provides the information needed to adjust, troubleshoot,
repair, and verify analyzer conformance to published specifications.
The HP Instrument BASIC User’s Handbook describes programming and
interfacing techniques using HP Instrument BASIC, and includes a language
reference.
iv
The HP Instrument BASIC User’s Handbook Supplement shows how to use
HP Instrument BASIC to program the analyzer.
The Option 100 Fault Location and Structural Return LossMeasurements User’s Guide Supplement provides theory and measurement
examples for making fault location and SRL measurements. (Shipped only with
Option 100 analyzers.)
The CATV Quick Start Guide provides abbreviated instructions for testing the
quality of coaxial cables. (Shipped only with Option 100 analyzers.)
The Cellular Antenna Quick Start Guide provides abbreviated instructions
for verifying the performance of cellular antenna systems. (Shipped only with
Option 100 analyzers.)
• “Specifications” on page 1-36
The operator's check can be used as a quick 80% confidence test. The only equipment
required for this check is a type-N cable and a good quality 50Ω type-N load (75Ω for
Option 1EC analyzers).
The performance tests will verify that the analyzer meets its published specifications with
greater than a 95% confidence level. The performance tests require an external controller
running HP BASIC and are completely automated.
The specifications provide a complete definition of the warranted and non-warranted
characteristics of the analyzer.
1-2Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersPerforming the Operator's Check
Performing the Operator's Check
The operator's check should be performed when you receive your instrument, and any time
you wish to have confidence that the analyzer is working properly. The operator's check
does not verify conformance to specifications, but should give you a high degree of
confidence that the instrument is performing properly if it passes.
The operator's check consists of making the following measurements with the cable that
was supplied with your analyzer:
8712ET/8714ET8712ES/8714ES
transmissionS21 and S
broadband powerbroadband power
reflectionS11 and S
reflection
1.with a 50Ω or 75Ω load, instead of the cable
1
S11 and S
12
22
22
1
Equipment List
To perform the operator's check, you will need the following:
• A known good cable suc h as the one that was supplied with your analyzer. The cable you
use should have ≤0.5 dB of insertion loss up to 1.3 GHz, and ≤0.75 dB of insertion loss
from 1.3 to 3.0 GHz.
• A known good load (>40 dB return loss) that matches the test port impedance of your
analyzer such as one from calibration kit HP/Agilent 85032B/E (50Ω) or HP/Agilent
85036B/E (75Ω).
NOTEThe illustrations depicting the analyzer’s display in this section were made
using an 8714ET. Other models’ displays will have minor differences in
annotation.
Service Guide1-3
Performance Checks8712ET/ES and 8714ET/ES
Enter
dBm
Default Response
Performing the Operator's CheckRF Network Analyzers
Make a Transmission (or S21 and S12) Measurement(s)
1. Connect the equipment as shown in Figure 1-1. Use a known good cable such as the one
that was supplied with your analyzer.
NOTEThe quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in “Equipment List” on
page 1-3.
Figure 1-1Equipment Setup for Performing the Operator’s Check
2. Press.
3. Press.
4. Press.
PRESETSCALE.1
POWER0
CAL
5. Verify that the data trace falls within ±0.5 dB of 0 dB. See Figure 1-2 for a typical 8714
result. The 8712 should look similar, but will end at 1300 MHz.
1-4Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
S12 Rev Trans
Default Response
RF Network AnalyzersPerforming the Operator's Check
Figure 1-2Verify Transmission (or S
and S12) Measurement
21
Perform the following steps on ES model analyzers only.
6. Press.
7. Press.
MEAS 1
CAL
8. Verify again that the data trace falls within ±0.5 dB of 0 dB. See Figure 1-2.
Service Guide1-5
Performance Checks8712ET/ES and 8714ET/ES
Power
More
Power
Start
MHz
Enter
dBm
Performing the Operator's CheckRF Network Analyzers
Make a Broadband Power Measurement
1. Leave the cable connected to the analyzer as shown in Figure 1-1.
NOTEThe quality of the cable will affect these measurements; make sure you use a
cable with the characteristics described in “Equipment List” on page 1-3.
2.
Press.
3. If is not displayed, press.
MEAS 1
4. Press.
5. Press.
POWER0
FREQ
10
SCALE1
6. Verify that the data trace is within ±2 dB of 0 dBm. See Figure 1-3 for a typical 8714
result. The 8712 should look similar, but will end at 1300 MHz.
Figure 1-3Verify Broadband Power Measurement
1-6Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
S11 Refl Port1
Reflection
Enter
dBm
Default 1-Port
S22 Refl Port2
Default 1-Port
RF Network AnalyzersPerforming the Operator's Check
Make a Reflection (or S11 and S22) Measurement with a Cable
1. Leave the cable connected to the analyzer as shown in Figure 1-1.
NOTEThe quality of the cable will affect these measurements; make sure you use a
cable with the characteristics described in “Equipment List” on page 1-3.
2.
Press.
3. For ES models, press.
For ET models, press.
PRESETMEAS 1
4. Press.
5. Press.
6. Press.
SCALE10
POWER0
CAL
7. Verify that the data trace falls completely below −16 dB. See Figure 1-4 for a typical
8714 result. The 8712 should look similar, but will end at 1300 MHz.
Figure 1-4
Verify Reflection (or S11 and S22) Measurement
Perform the following steps on ES model analyzers only.
8. Press.
9. Press.
MEAS 1
CAL
10.Verify again that the data trace falls completely below −16 dB. See Figure 1-4.
Service Guide1-7
Performance Checks8712ET/ES and 8714ET/ES
S11 Refl Port1
Reference Level
S22 Refl Port2
Reference Level
Performing the Operator's CheckRF Network Analyzers
Make a Reflection (or S11 and S22) Measurement with a Load
1. Disconnect the cable and connect a known good load to the RF OUT/PORT 1 test port as
shown in Figure 1-5.
Figure 1-5Connect the Load
2. For ES models, press.
Verify that the data trace falls below −30 dB. If the data trace is off the screen, press
3.
MEAS 1
SCALE
and the key until the trace moves up onto the screen.
Perform the remaining steps on ES model analyzers only.
4. Disconnect the load from PORT 1 and connect it to PORT 2 as shown in Figure 1-6.
Figure 1-6Connect the Load to Port 2 (ES Models only)
5. Press.
6. Verify that the data trace falls below −30 dB. If the data trace is off the screen, press
1-8Service Guide
SCALE
MEAS 1
and the key until the trace moves up onto the screen.
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performing the Operator's Check
This concludes the operator's check. However, further confidence can be obtained by
performing the following:
•Measure a known filter to verify that its measured response is the same as is expected.
(A 175 MHz filter is supplied with the analyzer.) Verify both the frequency accuracy and
noise floor.
•Check broadband response with the filter using conversion-loss mode (same as B*/R*).
•If the analyzer's frequency accuracy is critical for your application, verify a CW
frequency using a frequency counter. Verify to±.005% accuracy (for example,± 2500 Hz
at 500 MHz). Ensure that the analyzer is placed in trigger-hold mode (press
Trigger Hold
) to measure frequencies.
MENU
If the Analyzer Fails the Operator’s Check
First, repeat the operator's check using a different cable and load to eliminate these as a
possible cause of failure. Second, if the floppy disk from the factory labeled “Correction
Constants” is available, reload the “Correction Constants” (CCs) perChapter 8,
“Correction Constants and Firmware.” (The floppy disk was shipped with the instrument
from the factory.) Then repeat the Operator’s Check.
If your analyzer does not meet the criteria in the operator's check, your analyzer may need
adjustment or servicing. Contact any Agilent Technologies sales or service office for
assistance. Refer toTable 9-1on page9-7 for the nearest office. Before shipping your
analyzer, fill out and attach the blue repair tag, located at the back of this manual.
Service Guide 1-9
Performance Checks8712ET/ES and 8714ET/ES
PRINT RESULTS
Performance Test SoftwareRF Network Analyzers
Performance Test Software
The performance test software provides automated tests for verifying that the analyzer
meets its published specifications with greater than a 95% confidence level. The
performance test software must be run on an external controller running HP BASIC (the
analyzer’s internal IBASIC cannot be used). The performance tests can be considered a
check of the overall calibration of the instrument. Unlike the adjustments described in
Chapter 2, the performance test software only checks the performance of the analyzer. No
changes are made to the calibration constants stored in the instrument. When a
performance test fails, usually one or more adjustments should be performed to attempt to
improve the performance before repairing or replacing hardware.
Results of the automated performance tests can be printed out at any time. They include
all specifications, limits, and uncertainties . Although no manual tests are provided, each of
the automated tests is described in enough detail to allow a knowledgeable technician to
perform all needed tests in an accurate manner. See “Performance Test Notes and
Descriptions” on page 1-23.
The printed test results list uncertainties for each measurement. These uncertainties are
for reference only, and are not taken into consideration when making pass/fail
determinations. The listed uncertainties do not depend on the level of the measured signal;
they assume a signal at the specified level. Generally, uncertainties assume published
specifications for all instruments and devices. For the 110 dB step attenuator and the
20 dB fixed attenuator, device-specific calibration data is required.
Performance test record cards are not provided; the printed results of the automated
performance testing serve as the test record for this product. See “” in
“Using the Program” on page 1-19 for details on pass/fail indications for individual tests.
When a test report contains no “FAIL” indicators, an analyzer is considered to have passed
the performance tests.
Many instrument users require a periodic, documented check of instrument performance.
This is usually referred to as a “periodic calibration.” Successful completion of the
performance tests (no “FAIL” indicators) should suffice for the requirements of a periodic
calibration. Do not confuse the term “periodic calibration” with the terms “user
calibration” or “default calibration” which both refer to measurement calibrations utilized
by operators when measuring devices with the analyzer.
Required Software and Controller
The performance test software is written for an HP 9000 Series 200, 300, or 700
workstation (or equivalent), running HP BASIC 5.1 or greater. There is also a version
available to run on a PC under HP BASIC for Windows with a GPIB interface card.
Operation is essentially identical to that described for an HP workstation computer.
1-10Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
NOTE To order the performance test software, contact the nearest Agilent
Technologies sales or service office listed inTable 9-1on page9-7. The
workstation version is HP part number 08714-10001 (LIF format). The PC
version is HP part number 08714-10002 (DOS format), and requires HP
BASIC for Windows (HP model number E2060B), and a GPIB interface card.
The performance test software automates all the performance verification tests for the
8712ET/ES and the 8714ET/ES. It will test all option combinations, including both 50Ω
and 75Ω versions.
NOTE This program cannot be used to test an HP/Agilent 8711A, any HP/Agilent
871xB, or any HP/Agilent871xC. To test an HP/Agilent 8711A, order HP part
number 08711-10011. To test an HP/Agilent 871xB, order HP part number
08712-10011. Both of these programs are in LIF format only. To test an
HP/Agilent 871xC, order HP part number 08712-10017 (LIF format) or
08712-10018 (DOS format).
NOTE All the data files and configuration files used in this program are compatible
with the previous 871xB and 871xC versions of this program. It is not
necessary to regenerate or modify any of these files. If you are adding this
program to the same directory as the 871xB/C software, only the main
program file (CAL871XE) need be transferred.
The performance test software is provided on a single disk. You can run the program
directly from the disk, although it is recommended that you run it from your system's hard
disk for best performance. Generally, the program can be stopped, reset and re-run with
little or no loss of data. The LIF version program will operate on a LIF-based system, or an
SRM/HFS environment. The DOS version is limited to DOS environments.
Operation is a simple menu driven system with some softkey selections. For ease of use,
the / / keys on your computer’s keyboard can always be used as
ReturnEnterExecute
softkey #1—the most frequently used key.
NOTE All necessary instructions and test setup diagrams are contained within the
program.
A disk file is generated for each analyzer, and the results of each test are stored in that file.
This file is updated after each test so that results will not be accidentally lost. A printout
can be obtained for any previously tested analyzer.
Estimated Time
Before running the performance tests, turn on the analyzer and allow it to warm up for one
hour. Performance test times are kept to a minimum, allowing analyzers to be fully tested
within one to two hours after the initial warmup.
Service Guide 1-11
Performance Checks8712ET/ES and 8714ET/ES
Performance Test SoftwareRF Network Analyzers
Equipment
In addition to a computer, printer, RF cables, BNC cables, GPIB cables, and adapters,
Table 1-1 lists the equipment that is required to test and adjust the analyzer.
Table 1-1 Required Equipment for Performance Tests
DescriptionHP/Agilent Model #
Function generator8116A or 33120A
Power meter
Power sensor, 50Ω8482A
Power sensor, 75Ω8483A
Cal kit type-N 50Ω85032B Opt 001
Cal kit type-N 75Ω85036B
External 110 dB step
attenuator
Attenuator/switch driver 11713A✓✓
Spectrum analyzer8560 series,
20 dB attenuator 50Ω
20 dB attenuator 75Ω
6 dB attenuator 50Ω
3 dB attenuator 50Ω8491A/B/C Opt 003
3 dB attenuator 75Ω
Power splitter, 50Ω
Power splitter, 75Ω
Minimum loss pad
(quantity: 2)
Precision cable 50Ω8120-8862 or 8120-6469
Needed only if using an 8496G
programmable 110 dB step
attenuator.
Others may be substituted.
Notes
2
1. An 85032B Option 001 is a subset of the standard 85032B. This option eliminates four 7-mm adapters
that are not needed, resulting in a significant cost savings.
2. The spectrum analyzer used must be compatible with the 8566B , both in the GPIB commands for the
basic remote control functions and in frequency accuracy. Any of the 856x series is acceptable, the
8560E is the least expensive. The 859x series is not acceptable because of the marginal frequency
accuracy.
3. The 75Ω 20 dB attenuator can be substituted with two minimum loss pads and a 50Ω 10 dB attenuator.
NOTE: a set of 75Ω attenuators (3,6,10,20 dB) is available as 86213A.
1-12Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersPerformance Test Software
File Structure
The program disk contains several different files in addition to the main program file. The
following list describes the files found on the performance test software disk.
CAL871XEThis is the main program file.
Cfg_xxxxThis is the configuration file. The variable
user-defined name or
Dflt (default). This file contains all the configuration
xxxx represents any
set-ups for the program such as:
• file directories and paths
• test equipment serial numbers
• GPIB addresses
• types of test equipment
• other optional choices
If more than one configuration file exists on the disk, the program will
prompt you to select a specific configuration file to load. If none exists, the
program will create a default file.
P1D_xxxxThe cal factor file for the HP/Agilent 8481D sensor. The variable
xxxx
represents the last four digits of the HP/Agilent 8481D or HP/Agilent
8484A sensor's serial number.
P2A_xxxxThe cal factor file for the HP/Agilent 8482A sensor. The variable xxxx
represents the last four digits of the 8482A sensor's serial number.
P3A_xxxxThe cal factor file for the HP/Agilent 8483A sensor. The variable
xxxx
represents the last four digits of the 8483A sensor's serial number.
Stp_xxxxThe actual attenuation values for the external 110 dB step attenuator. The
variable xxxx represents the last four digits in the attenuator's serial
number. Attenuator values are measured at 30 MHz on an accurate
system (such as an HP/Agilent 8902A Option 050).
Pd5_xxxxThe actual value of the 20 dB 50Ω attenuator as measured by an accurate
system. Values are attenuation versus frequency. The variable
xxxx
represents the last four digits in the attenuator's serial number.
Pd7_xxxxThe actual value of the 20 dB 75Ω attenuator as measured by an accurate
system. Values are attenuation versus frequency. The variable xxxx
represents the last four digits in the attenuator's serial number.
dxxExxxxThe data file that holds the analyzer's test results. The variable
xxxx
represents the last four digits in the analyzer's serial number. The file
“d2TExxxx” holds data for an 8712ET, “d4SExxxx” for an 8714ES.
CurEDUTxThe data file that holds the test results of the most recently tested
analyzer regardless of model or serial number. The variable
x can
represent any character 0 to 9.
Service Guide1-13
Performance Checks8712ET/ES and 8714ET/ES
Performance Test SoftwareRF Network Analyzers
INSTALLThis program is not necessary but can help speed the installation of the
LIF version program. It automatically copies the required files into the
desired location.
NOTESeveral data files are present on the disk with the “xxxx” suffix. These are
default files that allow operation of the program without correction factors.
The values in these file are nominal values. For example: 100 percent for all
sensor cal factors; 20 dB for the 20 dB attenuators; 10.00, 20.00, 30…dB for
the external step attenuator , and so on. This is provided as a convenience only
to demonstrate or learn about the program. The default configuration file will
access these files if no user-defined serial number is specified. A warning
message will be displayed if the program is using these default files.
Configuration File
The configuration (config) file is the file that customizes this program for your usage. It
contains specific information for your setup. Some of the items contained in this file
include:
• the GPIB addresses of all the test equipment
• the GPIB address of the analyzer under test
• the model numbers of the test equipment
• the serial numbers of sensors, fixed attenuators, and the external 110 dB step
attenuator
• the disk drive locations (path and directories)
• other items specific to your setup
Ideally, only one configuration file should exist on the program disk so that the program
will automatically load only that file. However, it is possible to have several configuration
files stored on the same disk. In this case, the program will list all of the available files and
prompt you to select one. This is useful if more than one setup is commonly used or if one
performance test software disk serves several workstations.
1-14Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
SET TIME
RESUME
CONFIG
RF Network AnalyzersPerformance Test Software
Getting Started
CAUTIONMake a copy of the performance test software disk for day-to-day use. The
master copy should remain stored in a safe place. This will decrease the
likelihood of damage to the master disk.
NOTEThe DOS version, which is used in conjunction with HP BASIC for Windows,
requires the proper setup of the GPIB interface card and drivers. This
document assumes that users are familiar with HP BASIC for Windows , that
it is running, and that it has been properly installed and configured.
First, find all serial numbers of the calibration devices required, such as external step
attenuators, sensors, and fixed attenuators for both 50Ω and 75Ω devices (if applicable).
Verify the GPIB addresses you will be using, as well as the mass storage directories and
locations.
If you plan to operate this program only from the supplied floppy disk, insert the disk into
the computer's disk drive and skip step 1 next.
Performance Test Software Installation
1. To install this program to the hard drive, either copy all files (except “INSTALL”) to the
hard drive, or load and run the “INSTALL” program. The INSTALL program will
automatically copy all required files to your specified destination drive or directory
(HFS, SRM or LIF format). The INSTALL program is only contained in the LIF
software version.
NOTEAny directory name used must be less than 80 characters long, and preferably
less than 60 characters, for best visibility.
2. Set the computer's “Mass Storage Is” (MSI) command to the directory or drive location
of the main program file (CAL871XE).
3. GET and RUN the CAL871XE program. The program will display the initial screen
with the software part number, revision date and current time/date. If the current time
or date is incorrect, select the softkey to set it.
4. Press. Remember that the key can also be used as softkey #1.
5. Press to enter the configuration screen.
This menu allows you to enter all the specific information that pertains to your
particular configuration (such as test equipment, addresses, and serial numbers). The
configuration file's selection descriptions and their initial default values are listed
below.
Return
Service Guide1-15
Performance Checks8712ET/ES and 8714ET/ES
Performance Test SoftwareRF Network Analyzers
Configuration Menu
In this section, the default values for the parameters listed in bold type appear in
parenthesis.
The analyzer's GPIB address (716). The first selection is the analyzer’s (DUT) GPIB
address. Please note that this is the only address that has both the Interface Select Code
(ISC) and the device number associated with it. This address can be any valid address with
any valid ISC. The ISC doesnot have to be the same as all the other test equipment (which
must be all on the same ISC). Generally the full address for the DUT is set to either 716 or
816. This also sets the ISC to 7 or 8 respectively. The factory default GPIB device address
for the analyzer is 16.
The ISC for the test equipment (7). This selection is the ISC for all the associated
test equipment. This is generally 7 or 8 and may be the same as that for the DUT.
The power meter model number (437B) and address (13). These two selections
are for the power meter model number and address.
CAUTIONThe program will accept the 436A power meter as a valid selection, but it is
not recommended or supported and may cause the program to halt execution.
The spectrum analyzer (8566B) and address (18). This is the spectrum analyzer
selection. Any of the listed models will suffice, but keep in mind that if you do not use a
frequency counter, then the spectrum analyzer must have high degree of frequency
accuracy since it will be used as the frequency counter.
The function generator model number (8116A or 33120A) and address (14).
This selection is for the function generator and address. The 8116A is the only valid
selection at this time.
The attenuator switch driver model number (11713A) and address (28). Next
is the attenuator switch driver selection and address. At this time, the only two c hoices are
the 11713A and “NONE”. If “NONE” is selected, the program will prompt you to manually
set the external 110 dB step attenuator each time it needs to be changed. Since this is
inconvenient, the use of an 11713A with a 8496G model attenuator is highly
recommended.
The frequency counter model number (5342A) and address (02). The frequency
counter model number and its address are next. If “NONE” is selected, the program will
use the spectrum analyzer to measure frequency accuracy.
The voltmeter model number (3456A) and address (22). This selection is not
currently used by this program.
1-16Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersPerformance Test Software
The printer address (01). The printer address is next. No model number is needed
here; all GPIB printers should work equally well. For the DOS version, entering any valid
number except 10 will direct the output to the attached GPIB printer. Selecting an address
of 10 will redirect the output to the defined Windows default printer. Typically this will be
a printer attached to the computer's parallel port or a networked printer. On the DOS
version (HP BASIC for Windows), the default address is set to 10.
The program directory and the mass storage unit specifier (the current MSI
device).
should point to the location of the CAL871XE program and all of its required data files (all
the files contained on the disk). If you are running this program from a floppy disk, the
program directory entry should be blank. If using an HFS or SRM system, the entire
directory name should be entered here (e.g. /TESTS/ANALYZER/NETWORK/8712E). The
entire length must be less than 80 characters (preferably less than 60 for easier viewing).
The MSUS, which indicates the physical address of the disk drive , also has an 80-character
limit. For the DOS version, this is typically just the drive letter (for example , C:). When the
program is first run, the default will be the current MSI device.
The program directory and Mass Storage Unit Specifier (MSUS) are next. These
The data directory and MSUS. The next two entries are the data directory and
MSUS. These should point to the location where the results of each DUT will be stored (if
applicable). The data directory and MSUS are otherwise similar to the program directory,
including the defaults.
The sensor's, external 110 dB step attenuator's, and fixed attenuator's (pad's)
serial numbers (xxxx).
the serial numbers for the power meter sensors, the step attenuator, and the 20 dB
attenuators. This allows the program to load the proper correction values for each device.
The disk contains files that match these default serial numbers (xxxx), but the correction
data contained within these files is invalid (e.g. 100 percent for all sensor cal factors). This
allows the program to be run for experimental or training purposes only. As a precaution, a
warning message will be displayed if a default file is loaded.
Six of the next seven entries will contain the last four digits of
The step attenuator's 40 dB card selection (3). This selection is choice of the 40 dB
step section. The 8496G step attenuator contains one 10 dB section, one 20 dB section, and
two 40 dB sections. This selection specifies which of the two 40 dB attenuator card sections
of the 8496G step attenuator is to be used; section 3 or 4. Enter the value (3 or 4) for which
you have calibration data.
For example, when the step attenuator was calibrated, one of the two 40 dB sections was
selected for the 40, 50, 60, and 70 dB attenuation value. The data corresponding to the
selected 40 dB section must be input to ensure valid measurement data when using
attenuations of 40 dB or greater. For purposes of identification, section 3 of the attenuator
is the section that is engaged when button #3 or #7 of the 11713A is lit.
This entry is not applicable if you are using a manual attenuator.
The beeper (on). The last selection will suppress or enable the beeper, as desired.
Service Guide1-17
Performance Checks8712ET/ES and 8714ET/ES
CAL DATA ENTRY
Performance Test SoftwareRF Network Analyzers
The current DUT file suffix (1). This file, “CurEDUTx,” contains the test-result data
of the current device under test. Up to ten different file names can be specified. This allows
multiple users to operate from one system storage medium, such as an SRM environment.
This entry can be ignored for single-user applications.
Config file and cal data generation.
1. Once all desired changes have been made, the program will ask if you want to store the
data just entered. A specific 4-character file name suffix can be provided if more than
one configuration is required. Leaving the file name blank will result in the default file
name of “Cfg_dflt.” If the program finds only one config file, that file will be loaded
automatically. If more than one is discovered, the program will list the file names found
and allow you to choose one. At this point the program will also ask you if a hardcopy
printout is desired.
2. After the config file has been generated, the proper data files containing the correction
values must be generated. Press. You will be shown a list
including:
• 3 sensors
• an external 110 dB step attenuator
• two 20 dB attenuators (pads)
You will need to enter the correction data for every piece of test equipment included in
this list that you will be using to test the analyzer. Select the desired item.
Sensors. All sensors require both frequency and cal factor data. Frequencies must always
be in MHz and cal factors in percent. The first entry requested will be the last four digits of
the serial number. The computer will search for a current configuration file containing this
information. If one is present, the program will query you to either view or edit the old
values.
Assuming you are editing the data, the 50 MHz reference cal factor will be requested next.
This value must use percent as its units and be within the range of 50 and 150 percent.
After the cal factor is entered, the program will request that all frequency/cal factor pairs
be entered. Any frequency entered must be greater than zero. You must start with the
lowest frequency for which you have data and sequentially enter higher frequencies until
done. Each entry must consist of both a frequency in MHz and the cal factor in percent,
separated by a comma (e.g. 500, 98.6). Each frequency entered must be higher than the
previous entry. These cal factor values must be between 75 and 125 percent.
TIPIf a mistake is made, you can back up one entry and re-enter a value by
entering a negative frequency. If you accidentally enter only the frequency
(instead of a frequency/cal factor pair), the computer will wait for the cal
factor to be entered; however the only prompt will be a “?”.
Each entry is shown on the CRT. When all entries are completed, enter “0,0” to exit the
data entry portion of the program. At this point the program will redisplay all of the values
entered and ask if you want to store them. (Check to make sure there are no entries for a
zero value frequency.)
1-18Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RESUME
SET TIME
HELP
EXIT
SET TIME
HELP
EXIT
RESUME
CONFIG
LOAD CONFIG
CAL DATA ENTRY
RF Network AnalyzersPerformance Test Software
TIPEven if you find a mistake, it will be easier to answer yes, then go back and
re-edit the values. Otherwise you will have to re-enter all of the values again.
20 dB Attenuators (Pads). A procedure similar to the sensor data entry procedure is
used to enter the frequency/attenuation values for the 20 dB attenuators. Valid entries for
the 20 dB attenuator range from 17.5 to 23.5 dB with all frequency entries being greater
than zero.
External 110 dB Step Attenuator. Step attenuator data is only measured at 30 MHz
so only one value needs to be entered for each nominal attenuation step. All values are
relative to the 0 dB setting, which is defined as zero (i.e. the program does not care about
insertion loss at the 0 dB setting).
Valid entries are within 3 dB of the nominal attenuation setting with all entries being
greater than zero. The program will request data up to 110 dB of attenuation.
Using the Program
NOTERemember that the / / key on your computer’s
ReturnEnterExecute
keyboard can be used as softkey #1.
When the program is first run, the initial information screen is presented. This contains
the program name, revision number, date and other miscellaneous information. You are
presented with four softkey choices:,,, and.
allows you to set the time and date. If the date has never been set, the
program will force you to enter it regardless of whether or not this softkey is pressed.
will present some very brief instructions and other information.
will terminate the program.
will continue execution of the program and proceed to the main menu.
Main Menu
The main menu provides seven softkey choices. An eighth choice may be displayed if you
are returning to this menu from the test selection menu.
allows you to define a particular configuration of test equipment, mass storage,
and GPIB addresses. This selection is described earlier in “Configuration File” on page
1-14.
allows you to select one of several configurations to choose from. If only
one is available, it is automatically loaded without further confirmation. If several are
available, you are instructed to select one.
attenuators. See “Configuration File” on page 1-14.
Service Guide1-19
allows you to enter correction factors for sensors, attenuators and
Performance Checks8712ET/ES and 8714ET/ES
PRINT RESULTS
NEW DUT
GPIB Addr’s
EXIT
RESUME
RESUME
Performance Test SoftwareRF Network Analyzers
allows you to print or view test results from the current DUT or any
previously tested analyzer, assuming the data was archived. You can also add comments or
correct any header information that was previously entered.
Printed test results have a “pass/fail” column for indicating the outcome of each part of
each test. The possible outcomes are
PASSApplies to a warranted specification. The test has passed.
FAILApplies to a warranted specification. The test has failed.
Not TestedIndicates the test was performed, but the result was not checked. This is
usually due to limited capability of the test equipment.
No SpecApplies to a nonwarranted characteristic or typical parameter. The test
result is satisfactory (within the indicated limits).
No Spec*Applies to a nonwarranted characteristic or typical parameter. The test
result is marginal (outside the indicated limits). If the result is
substantially outside the limit, there may be cause for concern about the
proper operation of the analyzer.
will appear only after an analyzer has been tested. This softkey must be
selected before you begin to test another analyzer.
NOTEThis is the only way that the program can tell that the analyzer under test
has changed. Once selected, this softkey selection will disappear, indicating
that the program is ready to search for a new analyzer model/serial number.
will display the current expected addresses of all required test equipment.
In addition it will allow you to check each listed address for the presence of an active
device. It will not indicate which device is set to which address. It only indicates whether
or not the device at a given address is capable of handshaking properly. returns to
the previous menu.
will continue the program and look for an analyzer at the specified address.
Make sure that the power is switched on and that the GPIB cable is connected before
pressing.
Once the program finds the analyzer, the serial number is interrogated and the disk is
searched for any previous test results. If none are found, a message is displayed and you
are then allowed to enter any pertinent data regarding this new DUT such as owner,
technician, repair order, comments etc. The test selection menu is then presented.
1-20Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
Set Serial Number
Step Attenuator Test
Measure Step Attenuator
Measure 20 dB Pad
Measure Step Attenuator
RF Network AnalyzersPerformance Test Software
Test Selection Menu
The test selection menu can be divided into three main categories:
• eleven automated analyzer performance tests
• four utility procedures
• two external broadband detector performance tests (HP/Agilent 86200B or 86201B)
Data for each of the eleven performance tests is archived upon completion.
The four utility programs are described as follows:
1. utility allows you to enter a serial number into the analyzer. This
may be required after some types of repairs.
CAUTIONEnter the number carefully; it cannot be easily changed once it has been
entered. If a mistake is made, it will be necessary to contact an Agilent
Technologies service center for correction.
2. The is a cyclical stress test designed to show any weakness in
the mechanical (and therefore, electrical) integrity of the built-in attenuator. (The
attenuator is standard equipment on ES models and is supplied as Option 1E1 on ET
models.) For each cycle, the attenuator card is switched in and out with each resulting
trace measured. Minimum and maximum excursions are then noted. The test is
performed at a low frequency and a high frequency. The number of cycles is selected by
the user. A minimum of twenty-five is recommended. One hundred cycles will take
about five to ten minutes depending upon computer speed. There is no pass/fail
indication but trace variances of more than 0.1 dB should warrant further investigation
or increased monitoring for signs of wear.
3.
(external) is a utility file that uses a power meter to
measure the actual insertion loss at 30 MHz for the 110 dB step attenuator. This
routine should only be used as a last resort if no other method of obtaining data is
available. Generally the attenuator should be measured using one of the following
methods:
• sending the device to a calibration standards lab
• measuring the device with an 8753D with full two-port correction
• measuring the device with an 8902A Option 050
4. is similar to the utility, but is used for
characterizing the 20 dB pad.
NOTEThe listed uncertainties shown on the test results assume that one of these
three choices was used and not the built-in routine, which has a much higher
degree of uncertainty.
Service Guide1-21
Performance Checks8712ET/ES and 8714ET/ES
Performance Test SoftwareRF Network Analyzers
The detector performance tests consists of an absolute power accuracy test and a frequency
response test as described in the HP/Agilent 86200B/86201B RF Detectors User's andService Guide. These tests are only for testing these two specific detector models.
Archiving of data is not done for these tests.
Additional Information
The program gives full step-by-step instructions along with connection diagrams. For
clarity, some of the required adapters may not be depicted in the connection diagrams.
Generally, solid connection lines between devices indicate a cable connection, where as a
dotted line indicates a direct connection (adapters may be required). All required
equipment should be connected and switched on before proceeding with any test, since the
program will occasionally need to write to a device before it is shown in the connection
diagram. If the device is not connected, the program will time-out before the connection
diagram is drawn. The time-out limit is set to 30 seconds for all devices. If the program
should stop for any reason, the computer may be reset and the program run again with
little or no loss of data (if a
SCRATCH C is not performed).
1-22Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
Performance Test Notes and Descriptions
The test descriptions explain how each test is performed in case it is necessary to perform
them manually without the automated test software. They are also provided to give
automated test users a general idea of what each test measures and how it is performed
under computer control.
Before running or manually performing the performance tests, turn on the analyzer and
allow it to warm up for one hour.
The performance tests relate to the analyzer’s specifications, listed inTable1-6,Table1-7,
Table1-8 andTable1-9 in“Specifications”starting on page 1-36. Each value listed in these
tables is classified as a “specification,” a “characteristic,” a “nominal,” or a “typical.” If you
are using the automated software, you can print the test results to indicate the outcome of
each test. See “” in“Using the Program” on page 1-19 for an
explanation of pass/fail information.
NOTE The numbers preceding each of the tests in this section correspond to the test
PRINT RESULTS
number in the automated performance tests.
Service Guide 1-23
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
PT01: Frequency Accuracy
This test measures the frequency accuracy of the output of the analyzer.
Test Process.Set the analyzer to several CW frequencies while in trigger-hold mode.
Connect a frequency measuring device to the output of the analyzer. The device used to
measure the frequency can either be a frequency counter or a spectrum analyzer with a
high degree of frequency accuracy. If the network analyzer under test has Option 1EC
(75Ω system impedance and connectors) a minimum loss pad will be required in order to
connect a 50Ω frequency counter or spectrum analyzer.
Specification Verified.This test verifies the “Test Port Output: CW Accuracy”
specification inTable1-8 starting on page1-57.
If This Test Fails.Failures of this test are rare. If this test should fail by a small amount,
adjust the reference crystal oscillator by performing the“Frequency Accuracy Adjustment”
on page 2-5. If it fails by a significant amount, and the A3 fractional-N/reference
assembly's 10 MHz reference output is accurate, suspect excessive noise on the RF output
of the source.
1-24 Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT02: Gain Compression
This test measures narrowband gain compression at or near the maximum rated input
levels for the analyzer.
Test Process.Set the power output to the lesser of +10 dBm or the maximum value for
the analyzer’s model and option configuration. Set the system bandwidth for narrow
(250Hz). Perform an enhanced-response calibration for a transmission measurement.
Measure transmission on a through cable over the full frequency range of the analyzer.
Insert a 20dB fixed attenuator and measure transmission again. The difference is
compared to the known insertion loss of the calibrated 20 dB fixed attenuator to determine
gain compression. Gain compression must be less than the limit stated in the specification.
This test process is performed once for an ET model analyzer and twice for an ES model
(both forward and reverse directions).
Specification Verified.This test verifies the “Test Port Input: Maximum Input Level”
specification inTable1-9 starting on page1-63. In some cases, the test must be performed
at a power level less than the +10 dBm level indicated in the specification.
If This Test Fails.If this test fails, replace the PC board on the A5 receiver assembly.
Service Guide 1-25
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
PT03: Noise Floor
This test makes transmission (ratioed) measurements to determine the noise floor for both
narrowband and broadband measurements.
Test Process.Preset the analyzer and set the power level to 0 dBm. Set the system
bandwidth to Fine (15 Hz). Set the averaging factor to 4 and enable averaging. Connect
loads to both test ports. Perform a Response and Isolation calibration. The peak value for a
given measurement trace is considered to be the level of noise floor. For ES models repeat
the test for an S12 measurement.
Repeat the test process above for a conversion loss measurement (without the Response
and Isolation calibration). This will measure the broadband noise floor.
Specification Verified.This test verifies the “Test Port Input: Noise Floor” specification
inTable1-9 starting on page1-63. (The measurement process defined above is considered
to be equivalent to the measurement process described in the specification’s footnote.)
If This Test Fails.First verify that all cable connections and screws are secure, then
perform the“Switched Gain Correction, Adjustment #102” on page 2-10.
1-26 Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT04: Dynamic Accuracy
This test measures narrowband ratio measurement accuracy over a range of power levels
at 30 MHz. A 110 dB step attenuator is needed—typically an 8496A or 8496G is used. Two
minimum loss pads must be used with a 75Ω analyzer (Option 1EC).
Test Process.Connect the step attenuator between the test ports on the analyzer. Set the
CW frequency to 30 MHz and select fine system bandwidth (15 Hz). Set the power to 0
dBm for 50Ω analyzers and to 13 dBm for 75Ω analyzers. Record the transmission (or S21)
readings for attenuator settings from 0 dB through 90 dB, in 10 dB increments. Apply the
correction factors for the attenuator setting for each reading before checking against the
specification. The reading at−20 dB should be used as the reference power. Use averaging,
if necessary, to improve the test results. For ES models, repeat the test for S12
measurements.
Specification Verified.This test verifies the “Test Port Input: Dynamic Accuracy”
specification inTable1-9 starting on page1-63
the phase accuracy.
If This Test Fails.First, perform the“Switched Gain Correction, Adjustment #102” on
page 2-10. If the problem persists, suspect calibration errors in the step attenuator used to
perform this adjustment; see“The step attenuator's 40 dB card selection (3)” on page 1-17.
Crosstalk can also affect this test; verify that all cables and screws are secure.
. It checks the magnitude accuracy, but not
Service Guide 1-27
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
PT05: Power Flatness
This test checks the maximum variation in output power across the frequency range of the
analyzer.
Test Process.Connect a calibrated power sensor to the RF OUT/PORT 1 test port.
Determine the power levels to test:
1.Refer to “Test Port Output: Power Sweep Ranges” inTable1-8 starting on page1-57.
2.Find the entry for the model and option configuration of the analyzer under test.
3.In the “Attn = 0 dB” column of the table, the power sweep range is defined in the form
“ntoPmax”. Usen as one power level to measure, use Pmax as another power level to
measure, and use the midpoint between the two as the third power level to measure.
4.If the analyzer under test is an ET model with Option 1E1, or an ES model, then also
use a midpoint power level from each of the following three columns: “Attn = 10 dB”,
“Attn = 20 dB” and “Attn = 30 dB”.
5.For each separate power level, record the reading on the power meter at 27 different
frequencies, across the frequency range of the analyzer. Correct the power readings
using the frequency flatness data for the power sensor. The difference between the max
and min values for the 27 readings should be within the specified flatness limit as
stated in “Test Port Output: Level Accuracy inTable1-8 starting on page1-57. Note
that the specifications are listed in the form of±n dB. This equates to an allowable 2n
variation (max−min). Measurement of power levels below−30 dBm are outside the
specified range of the power sensor and should not be included in the test.
6.For ES models, make the same set of measurements in the reverse transmission
direction (S12) with the power sensor connected to PORT 2.
NOTE Be sure to zero the power meter before making any series of measurements
below−19 dBm.
A second test is performed by the test software to check for obvious power level
inconsistencies that might not be detected by the specification check. It is a precaution to
ensure that no power holes (drop-outs) exist. A through cable is connected and the
analyzer is swept from 10 MHz to its high-end frequency using 1601 points per sweep. The
resulting B* trace is checked for any significant power holes that could have been missed
during the previous section of this test. For all ES models and for ET models with
Option1E1, repeat the full-sweep test for the three power levels tested in step 4 above.
When power holes are detected by this test, it is considered to be a failure of the
specification.
Specification Verified.This test verifies the “Test Port Output: Level Accuracy”
specification inTable1-8 starting on page1-57. Notice that values for 75Ω instruments
above 2GHz are nonwarranted characteristics,not warranted specifications.
If This Test Fails.Repeat the test, verifying that all connections are secure. If the test
still fails, perform the“Source Power Correction, Adjustment #104” on page 2-15.
1-28 Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT06: Absolute Power Accuracy
This test verifies the accuracy of the B* broadband detector from the maximum input level
of +16dBm down to the specified noise floor. The test is designed to be performed at
30MHz, however 15 MHz may be used if the frequency range of the signal source does not
extend to 30 MHz. The 8116A is the preferred signal source, however the 33120A may be
used as an alternative. An external 110 dB step attenuator is used to vary the test signal
over the power range to be tested.
Test Process.Connect the step attenuator to the output of the signal generator. Set the
frequency to 30 MHz (preferred) or 15 MHz. Connect a 6 dB pad to the output of the step
attenuator. For 75Ω analyzers, use a minimum loss pad in place of the 6 dB attenuator.
Calibrate the power meter and use it to measure the output from the 6 dB pad. Set the
step attenuator to 0 dB. Adjust the output from the signal generator so that the power
output from the 6 dB pad is +16.0 dBm. Record the exact reading. Use calibration data
from the power sensor to compensate for the frequency difference between the power
meter calibration and the signal generator. Disconnect the power sensor and connect the
6dB pad to the RF IN/PORT 2 test port on the analyzer. Set the system bandwidth to
narrow (250 Hz). Repeat the previously used steps for the signal generator output, and
record the readings on B*. Switch the step attenuator as necessary to provide levels from
+16 dBm to−44 dBm to the analyzer. Record the B* reading for each level and correct the
readings using the calibration data for the step attenuator. Averaging may be used as
needed. Each power reading on B* must meet the specified accuracy until the noise floor is
reached.
Specification Verified.This test verifies the “Test Port Input: Power Accuracy”
characteristic inTable1-9 starting on page1-63. In some cases, it may be performed at 15
MHz instead of the specified 30 MHz.
If This Test Fails.If you are using a programmable attenuator, first verify that the
correct 40 dB section of the step attenuator is being used. See“The step attenuator's 40 dB
card selection (3)” on page 1-17. If the correct attenuator is being used, perform the“B*
Amplitude Correction, Adjustment #110” on page 2-21.
Service Guide 1-29
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
PT07: Broadband Frequency Response
This test checks the frequency response flatness of the broadband B* detector. It is checked
at a nominal−6 dBm signal level.
Test Process.Connect the output of the analyzer to a power splitter. Connect the power
splitter outputs to the analyzer and to a calibrated power meter. Preset the analyzer and
make sure the power level is set for 0 dBm. Step the analyzer through a cycle of
approximately 12 CW frequencies while the B* and power meter readings are recorded.
(The CW frequencies must include 10 MHz and 1.3 or 3.0 GHz, as appropriate.) Swap the
output connections on the splitter and repeat the process. Average the two sets of readings
to eliminate differences within the power splitter. Calculate the difference between the B*
reading and the power meter reading for each frequency. Use a reading at 30 MHz as a
reference. Subtract the reference from each of the other differences, and check the results
against the specification.
Adapters are generally required for this test. A 75Ω analyzer requires a 75Ω power splitter
and power sensor. A power divider may be used in place of the power splitter.
Specification Verified.This test verifies the “Test Port Input: Broadband Flatness”
characteristic inTable1-9 starting on page1-63. This test also verifies the “Test Port
Input: Frequency Range” specification for broadband inTable 1-9.
If This Test Fails.Repeat the test after verifying that all connections are secure. There is
no corresponding adjustment for this test.
1-30 Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT08: Directivity
This test checks directivity, source match, and load match. The uncorrected (raw)
directivity and source match values are derived from the measurement calibration arrays.
Load match is checked using a calibration performed at the end of a through cable.
These tests require an 85032B calibration kit (standard or Option 001) for a 50Ω analyzer,
or an 85036B calibration kit for a 75Ω analyzer. The standard economy (E) versions of
these kits do not provide the necessary components to perform these procedures.
NOTE The three tests described below assume that an ES model is operating in the
forward (S21) direction. These tests should be repeated for the reverse
direction (S12) on ES model analyzers.
Test Process for Directivity.Preset the analyzer and set the power level to 0 dBm.
Measure reflection on channel 1. Perform a “1-port” calibration on ET models, or a “user
1-port” calibration on ES models. Leave the load attached. Disable measurement
calibration by pressing the following keys
Service Utilities Meas Cal Options Meas Cal
the reflection tracking array, press . Combine the
traces by pressing . The resulting trace should be completely
DISPLAY
View Array Master Tracking
Data/Mem
SYSTEM OPTIONS
(OFF should be in caps). Access
Service
below the values for the specification (not the typical).
Test Process for Source Match.(These instructions assume that the directivity test has
been completed and that the calibration is still valid.) Access the source match array. To
meet the specification, the trace must be below the limit indicated in the specification (not
the typical). For example, if the specification for 300 kHz to 1.3 GHz is 18 dB, the trace
must be below−18 dB for that frequency range.
Test Process for Load Match.Preset the analyzer. Connect a type-N cable to the
RFOUT/PORT 1 test port. Perform a 1-port calibration at the free end of the cable. Be
sure to specify a male calibration kit (“Type-N (m)”). Note that during the open portion of
the calibration, the extender pin must be placed over the center conductor of the cable.
Connect the cable to RF IN/PORT 2. The reflection trace must be below the limits given in
the load match specification (not the typical).
Specifications Verified.These tests verify performance of the following items in the
“System Performance, Uncorrected” category ofTable1-6on page1-55 andTable1-7on
page1-56:
•Directivity
•Source Match (ratio)
•Load Match
Service Guide 1-31
Performance Checks8712ET/ES and 8714ET/ES
Performance Test SoftwareRF Network Analyzers
If This Test Fails. If directivity (uncorrected) fails, the most likely cause is a defective
coupler: the A coupler for the forward direction, or the B coupler for the reverse direction.
Source match depends mostly on hardware in the A5 receiver assembly. Load match is
solely dependent upon the A5 receiver assembly. If either of these two tests fail, first
inspect the two type-N connectors for damage, especially on 75Ω instruments. If load
match fails, also suspect the A5 receiver assembly limiter or transfer switch.
1-32Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT09: Harmonics
This test uses a spectrum analyzer to measure harmonics on the network analyzer output
at 18 different frequencies. Harmonics above 2.9 GHz are not checked.
Test Process.Connect the output of the analyzer to the spectrum analyzer with a high
quality type-N cable. Step the analyzer through 18 CW frequencies, ranging from 0.3 MHz
to 1440 MHz. Set the analyzer power to the maximum specified level for each frequency.
Check all harmonics below 2.9 GHz for conformance to the specification. When this test is
performed under computer control, the time bases of the two instruments are connected
and the network analyzer’s Ext Ref function is turned on.
Specification Verified.This test verifies the “Test Port Output: Signal Purity:
Harmonics” specification inTable1-8 starting on page1-57.
If This Test Fails.If the harmonics test fails, perform the“Source Power Correction,
Adjustment #104” on page 2-15. The most likely cause of a harmonics test failure is the A4
source assembly.
Service Guide 1-33
Performance Checks 8712ET/ES and 8714ET/ES
Performance Test Software RF Network Analyzers
PT10: Trace Noise
This test checks trace noise magnitude on narrowband and broadband transmission
measurements. The measurement trace is normalized to remove the effects of frequency
response on the baseline. The marker statistics for standard deviation (SDev) are
considered to be equivalent to an RMS calculation.
Test Process.Connect a through cable. Preset the analyzer and set the power level to
0dBm. Set the system bandwidth to narrow (250 Hz). Set the averaging factor to 8 and
enable averaging. Normalize the trace, then turn off the averaging. Position marker 1 at
the beginning of the trace and position marker 2 at the end of the trace. Use the marker
math statistics feature to display the SDev value. SDev is equal to the RMS value of the
trace noise.
Repeat this test for a conversion loss measurement, and set the start frequency to 10 MHz.
For ES models, repeat this test for a reverse transmission (S12) measurement.
Specification Verified.This test verifies the “Test Port Input: Trace Noise” specification
inTable1-9 starting on page1-63.
If This Test Fails.When this test fails by less than 1/2 the specified value, try using an
averaging factor of 16 or 32. Almost anything within the entire analyzer system can
contribute to excessive trace noise. The most likely assemblies to fail are the A5, A4, and
A3, in that order. If trace noise can be isolated to only one input (A, B or R), then the
problem is most likely in the A5 receiver assembly. Otherwise, suspect A4 or A3.
1-34 Service Guide
8712ET/ES and 8714ET/ES Performance Checks
RF Network Analyzers Performance Test Software
PT11: Crosstalk
This test measures the crosstalk between the output signal path and the input signal path
inside the analyzer at maximum specified output power level. The test makes a
narrowband transmission measurement with shorts connected to both test ports.
Test Process.Attach a short to each test port on the analyzer. Preset the analyzer and set
the power level to the maximum power specified for the particular model/option
configuration of the instrument. Set the system bandwidth to fine (15 Hz) and the
averaging factor to 4. Enable averaging. Check the trace; it must be below the crosstalk
limit given in the specification. For example, if the specification is 80 dB, the trace must be
below−80 dB.
For ES models, repeat the test for a reverse transmission (S12) measurement.
Specification Verified.This test verifies the “System Performance, Uncorrected:
Crosstalk” specification inTable1-6on page1-55 andTable1-7on page1-56.
If This Test Fails.Check cabling and shielding on the A4 source board and the A5
receiver assembly.
Service Guide 1-35
Performance Checks8712ET/ES and 8714ET/ES
SpecificationsRF Network Analyzers
Specifications
Definitions
All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise
stated) and sixty minutes after the instrument has been turned on.
Specification (spec.): Warranted performance. Specifications include guardbands to
account for the expected statistical distribution, measurement uncertainties, and changes
in performance due to environmental conditions.
Characteristic (char.): A performance parameter that the product is expected to meet
before it leaves the factory, but is not verified in the field and is not covered by the product
warranty. A characteristic includes the same guardbands as a specification.
Typical (typ.): Expected performance of an average unit which does not include
guardbands. It is not covered by the product warranty.
Nominal (nom.): A general descriptive term that does not imply a level of performance. It
is not covered by the product warranty.
Calibration: The process of measuring known standards from a calibration kit to
characterize the systematic (repeatable) errors of a network analyzer.
Corrected (residual) Performance: Indicates performance after error correction
(calibration). It is determined by the quality of calibration standards and how well
“known” they are, plus system repeatability, stability, and noise.
Uncorrected (raw) Performance: Indicates instrument performance without error
correction. The uncorrected performance affects the stability of a calibration.
1-36Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
System Performance
Table 1-2 System Performance, Corrected, 50Ω, 2-Port Calibration
8712ES/8714ES (50 Ω)
85031B (7-mm, 50 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine” bandwidth, no averaging, and at an environmental temperature of 25°±5 °C,
with less than 1 °C deviation from the calibration temperature.
2. For transmission measurements, the effect of crosstalk is neglected and
S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
Service Guide1-37
1
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-2 System Performance, Corrected, 50Ω, 2-Port Calibration (continued)
8712ES/8714ES (50 Ω)
85032B/E (Type-N, 50 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-38Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-2 System Performance, Corrected, 50Ω, 2-Port Calibration (continued)
8712ES/8714ES (50 Ω)
85033D (3.5 mm, 50 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
Service Guide1-39
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-2 System Performance, Corrected, 50Ω, 2-Port Calibration (continued)
8712ES/8714ES (50 Ω)
85038A (7-16, 50 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-40Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-3 System Performance, Corrected, 75Ω, 2-Port Calibration
8712ES/8714ES (75 Ω)
85036B/E (Type-N, 75 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
Service Guide1-41
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-3 System Performance, Corrected, 75Ω, 2-Port Calibration (continued)
8712ES/8714ES (75 Ω)
85039B (Type-F, 75 Ω) Cal Kit, User 2-Port Calibration
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. Assumes using an 85039B cal kit, and a DUT with a center pin
conforming to the 0.77 to 0.86 mm limits.
3. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-42Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation
from the calibration temperature.
Service Guide1-43
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-44Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET
and 871414ET.
2. These specifications apply for measurements made
using “fine” bandwidth, no averaging, and at an environmental temperature of 25°±5 °C, with less than 1 °C
deviation from the calibration temperature.
Service Guide1-45
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
2. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
3. For transmission measurements, the effect of crosstalk is neglected
and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-46Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation from
the calibration temperature.
Service Guide1-47
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-4 System Performance, Corrected, 50Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-48Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation from
the calibration temperature.
Service Guide1-49
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-50Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the
8712ET/8714ET.
2. These specifications apply for measurements made
using “fine” bandwidth, no averaging, and at an environmental temperature of 25°±5 °C, with less than 1 °C
deviation from the calibration temperature.
Service Guide1-51
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
2. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature of
25°±5 °C, with less than 1 °C deviation from the calibration temperature.
3. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-52Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using
“fine” bandwidth, no averaging, and at an environmental
temperature of 25°±5 °C, with less than 1 °C deviation from
the calibration temperature.
Service Guide1-53
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-5 System Performance, Corrected, 75Ω, T/R Calibration (continued)
1. These specifications apply for measurements made using “fine”
bandwidth, no averaging, and at an environmental temperature
of 25°±5 °C, with less than 1 °C deviation from the calibration
temperature.
2. For transmission measurements, the effect of crosstalk is
neglected and S12 = S21 for S21 <1.0, S12 = 1 / S21 for S21 > 1.0.
1-54Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersSystem Performance
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
2. These numbers apply for a measurement made using the “Fine” bandwidth at an
environmental temperature of 25°±5 °C.
3. The uncorrected directivity of a network analyzer is calculated in linear terms by
dividing the reflection measurement of an ideal load by the average of the reflection measurements of an ideal short and an ideal open.
4. The uncorrected source match is the source match of the network analyzer when
making a ratioed, uncalibrated measurement.
5. The uncorrected load match is the match of the network analyzer port used on
the load side of a measurement.
6. Measured by setting output power to the maximum specified setting, connecting
shorts to both ports, and measuring transmission. Typical and specified crosstalk
values are 5 dB worse than those shown in the table below 1 MHz (for all models)
and above 2.2 GHz (for 8714ET/ES models).
Service Guide1-55
Performance Checks8712ET/ES and 8714ET/ES
System PerformanceRF Network Analyzers
Table 1-7 System Performance, Uncorrected, 75Ω
8712ET/ES and 8714ET/ES (75 Ω)
8712ET/8714ET with
Description8712ES/8714ES
300 kHz
to 1.3
GHz
Specification2(in dB)
Directivity
Source Match
(Ratio)
Load Match
Crosstalk
Typical2(in dB)
Directivity403040304030
Source Match (Ratio) 232024202823
Load Match242224222422
Reflection Tracking±1.5±1.5±1.0±1.5±1.0±1.0
Transmission
Tracking
Crosstalk
3
4
5
6
6
271927192718
181521152319
181518151815
888897979797
±1.5±1.5±1.5±1.5±1.0±1.0
9595105105105105
1.3 GHz
to 3 GHz
Attenuator Option
300 kHz
to 1.3
GHz
1E1
1
1.3 GHz
to 3 GHz
8712ET/8714ET
(without Attenuator)
300 kHz
to 1.3
GHz
1.3 GHz
to 3 GHz
1. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
2. These numbers apply for a measurement made using the “Fine” bandwidth at an
environmental temperature of 25°±5 °C.
3. The uncorrected directivity of a network analyzer is calculated in linear terms by
dividing the reflection measurement of a perfect load by the average of the reflection measurements of an ideal short and an ideal open.
4. The uncorrected source match is the source match of the network analyzer when
making a ratioed, uncalibrated measurement.
5. The uncorrected load match is the match of the network analyzer port used on the
load side of a measurement.
6. Measured by setting output power to the maximum specified setting, connecting
shorts to both ports, and measuring transmission. Typical and specified crosstalk
values are 5 dB worse than those shown in the table below 1 MHz (for all models)
and above 2.2 GHz (for 8714ET/ES models).
1-56Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersTest Port Output
Test Port Output
Table 1-8 Test Port Output
8712ET/ES and 8714ET/ES Test Port Output
DescriptionSpecification
Frequency
Range:
8712ET/ES300 kHz to 1.3 GHz
8714ET/ES300 kHz to 3.0 GHz
Resolution1 Hz
Stability±5 ppm, 0° to 55 °C, typ.
CW Accuracy±5 ppm at 25°±5 °C<1Hz at 10% change in line
voltage, typ.
Signal Purity
Harmonics:
8712ET/ES<−20 dBc at <1 MHz
<−30 dBc at >1 MHz
8714ET/ES<−30 dBc
Nonharmonic Spurious:
8712ET/ES, <50 kHz from carrier<−25 dBc, char.
8712ET/ES, >50 kHz from carrier<−20 dBc at <1MHz, char.
<−30 dBc at >1MHz, char.
8714ET/ES, <50 kHz from carrier<−25 dBc, char.
8714ET/ES, >50 kHz from carrier<−30 dBc, char.
1. Noise floor is defined as the RMS value of the trace (in linear
format) for a transmission measurement in CW Frequency
mode, with the RF connectors terminated in loads, output
power set to 0 dBm, and no averaging. The noise floor specifications and typicals for narrowband detection measurements
assume that an isolation calibration has been performed
using an average factor of 16. F or the 8712ES/8714ES, external broadband detectors will provide a lower noise floor than
the internal broadband detectors.
Wide
(6500 Hz)
1-64Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersTest Port Input
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dB)
System Bandwidths:
Typical
(in dB)
Fine
(15 Hz)
System Dynamic Range
8712ET:
50 Ω:
With Attenuator Option 1E12:
Narrowband1151219658
Broadband (Internal)60676044
Without Attenuator Option 1E1
Narrowband1151219658
Broadband (Internal)62686145
75 Ω:
With Attenuator Option 1E12:
Narrowband1101189355
Broadband (Internal)53615439
Without Attenuator Option 1E1
Narrowband1131199456
Broadband (Internal)56625540
1
Fine
(15 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
1. The System Dynamic Range is calculated as the difference between
the receiver noise floor and the minimum of either the source maximum output (maximum power setting minus output power level
accuracy) or the receiver maximum input. System Dynamic Range
applies to transmission measurements only, since reflection measurements are limited by directivity. The System Dynamic Range for
8714ET/ES 75 Ω analyzers is not a specification for frequencies
>2 GHz; it is a characteristic. For the 8712ES/8714ES, external
broadband detectors will provide more dynamic range than the
internal broadband detectors.
2. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
Service Guide1-65
Performance Checks8712ET/ES and 8714ET/ES
Test Port InputRF Network Analyzers
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dB)
System Bandwidths:
Typical
(in dB)
Fine
(15 Hz)
System Dynamic Range1(continued)
8714ET:
50 Ω:
With Attenuator Option 1E12:
Narrowband1121209557
Broadband (Internal)57645741
Without Attenuator Option 1E1
Narrowband1141219658
Broadband (Internal)59655842
75 Ω:
With Attenuator Option 1E12:
Narrowband1071159052
Broadband (Internal)50585136
Without Attenuator Option 1E1
Narrowband1101169153
Broadband (Internal)53595237
Fine
(15 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
1. The System Dynamic Range is calculated as the difference between
the receiver noise floor and the minimum of either the source maximum output (maximum power setting minus output power level
accuracy) or the receiver maximum input. System Dynamic Range
applies to transmission measurements only, since reflection measurements are limited by directivity. The System Dynamic Range for
8714ET/ES 75 Ω analyzers is not a specification for frequencies
>2 GHz; it is a characteristic. For the 8712ES/8714ES, external
broadband detectors will provide more dynamic range than the internal broadband detectors.
2. Option 1E1 adds a 60 dB step attenuator to the 8712ET/8714ET.
1-66Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersTest Port Input
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dB)
System Bandwidths:
Typical
(in dB)
Fine
(15 Hz)
System Dynamic Range1(continued)
8712ES:
50 Ω:
Narrowband1041159057
Broadband (Internal)46534633
75 Ω:
Narrowband991118754
Broadband (Internal)39474027
8714ES:
50 Ω:
Narrowband1011128754
Broadband (Internal)43504330
75 Ω:
Narrowband961088451
Broadband (Internal)36443724
Fine
(15 Hz)
Med Wide
(4000 Hz)
1. The System Dynamic Range is calculated as the difference between
the receiver noise floor and the minimum of either the source maximum output (maximum power setting minus output power level
accuracy) or the receiver maximum input. System Dynamic Range
applies to transmission measurements only, since reflection measurements are limited by directivity. The System Dynamic Range
for 8714ET/ES 75 Ω analyzers is not a specification for frequencies
>2 GHz; it is a characteristic. For the 8712ES/8714ES, external
broadband detectors will provide more dynamic range than the
internal broadband detectors.
Wide
(6500 Hz)
Service Guide1-67
Performance Checks8712ET/ES and 8714ET/ES
Test Port InputRF Network Analyzers
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Description
Spec.
(in dB)
System Bandwidths:
Typical
(in dB)
Fine
(15 Hz)
Receiver Dynamic Range
8712ET/8714ET:
50 Ω:
Narrowband1151219658
Broadband (Internal)66716448
75 Ω:
Narrowband1141199456
Broadband (Internal)63686146
8712ES/8714ES:
50 Ω:
Narrowband1061159057
Broadband (Internal)54595239
75 Ω:
Narrowband1051149057
Broadband (Internal)51564936
1
Fine
(15 Hz)
Med Wide
(4000 Hz)
1. The Receiver Dynamic Range is calculated as the difference
between the receiver noise floor and the receiver maximum input.
Receiver Dynamic Range applies to transmission measurements
only, since reflection measurements are limited by directivity. The
Receiver Dynamic Range for 8714ET/ES 75 Ω analyzers is not a
specification for frequencies >2 GHz; it is a characteristic. For the
8712ES/8714ES, external broadband detectors will provide more
dynamic range than the internal broadband detectors.
Wide
(6500 Hz)
1-68Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersTest Port Input
Magnitude 0.01 dB rms 0.01 dB-pp 0.03 dB-pp 0.16 dB-pp
1
Narrow
(250 Hz)
Med Wide
(4000 Hz)
Wide
(6500 Hz)
1. Trace noise is defined for a transmission measurement in CW
mode, using a “through” cable having 0 dB loss, with the source
set to 0 dBm, and the analyzer’s averaging function toggled off.
Service Guide1-69
Performance Checks8712ET/ES and 8714ET/ES
Test Port InputRF Network Analyzers
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Dynamic Accuracy (Specification)
8712ET/8714ET:
8712ES/8714ES:
1,2
1. Narrowband detection mode
2. The reference power for dynamic accuracy is −20 dBm.
1-70Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersTest Port Input
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Power Accuracy: (Characteristic)
8712ET/8714ET1:
Magnitude
8712ES/8714ES1:
Magnitude
1. At 30 MHz, broadband mode, internal detectors
Service Guide1-71
Performance Checks8712ET/ES and 8714ET/ES
Test Port InputRF Network Analyzers
Table 1-9 Test Port Input (continued)
8712ET/ES and 8714ET/ES Test Port Input
Group Delay Accuracy1 (Characteristic)
8712ET/ES and 8714ET/ES
1. Valid for 85032B/E (type-N 50 Ω) and 85036B/E (type-N 75 Ω) cal kits using
either a two-port or enhanced response calibration.
1-72Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersGeneral Information
General Information
Table 1-10 General Information
8712ET/ES and 8714ET/ES General Information
DescriptionSpecificationSupplemental Information
Display Range
Magnitude200 dB (at 20 dB/div), max
Phase1800° (at 180°/div), max
Polar1 MUnit, max
Display Resolution
Magnitude0.01 dB/div, min
Phase0.1°/div, min
Polar10 µUnit Full Scale, min
Reference Level Range
Magnitude500 dB, max
Phase360°, max
Reference Level Resolution
Magnitude0.01 dB, min
Phase0.01°, min
Marker Resolution
Magnitude0.001 dB, min
Phase0.01°, min
Polar0.01 µUnit, min; 0.01°, min
Group Delay Aperture
Minimum20% of frequency span
Maximumfrequency span ÷ (# of points − 1)
Group Delay Range
1 ÷ (2 x minimum aperture)The maximum delay is limited
to measuring no more than
180° of phase change within
the minimum aperture.
Service Guide1-73
Performance Checks8712ET/ES and 8714ET/ES
General InformationRF Network Analyzers
Table 1-10 General Information (continued)
8712ET/ES and 8714ET/ES General Information
DescriptionSpecificationSupplemental Information
System Bandwidths
Wide (6500 Hz)6500 Hz, nom
Medium Wide (4000 Hz)4000 Hz, nom
Medium (3700 Hz)3700 Hz, nom
Medium Narrow (1200 Hz)1200 Hz, nom
Narrow (250 Hz)250 Hz, nom
Fine (15 Hz)15 Hz, nom
Frequency47 Hz to 63 Hz
Voltage at 115 V setting90 V to 132 V115 V, nom.
Voltage at 220 V setting198 V to 264 VAC230 V, nom.
Power300 VA, max230 W, nom.
Front Panel
RF ConnectorsType-N female; 50 Ω, nom.
Probe Power:3-pin connector; male
Positive Supply200 mA, max+15 V, nom; 0.75 A fuse, nom;
Negative Supply250 mA, max−12.6 V, nom; 0.75 A fuse, nom.
12-pin circular; female
6-pin mini DIN (PS/2); female
(With Option 1EC only:Type-Nfemale; 75 Ω, nom.)
1. A third-wire ground is required.
Service Guide1-75
Performance Checks8712ET/ES and 8714ET/ES
General InformationRF Network Analyzers
Table 1-10 General Information (continued)
8712ET/ES and 8714ET/ES General Information
DescriptionSpecificationSupplemental Information
General Environmental
ESDMinimize using static-safe
work procedures and an
antistatic bench mat (part
number 9300-0797).
DustMinimize for optimum
reliability.
Operating Environment
Temperature0 °C to +55 °C
Humidity5% to 95% at +40 °C
Altitude0 to 4.5 km (15,000 ft)
Storage Conditions
Temperature−40° C to +70 °C
Humidity0% to 95% RH at +65 °C
(noncondensing)
Altitude0 to 15.24 km (50,000 ft)
Cabinet Dimensions
Height × Width × Depth179 × 425 × 514 mm, nom.
(7.0 × 16.75 × 20.25 in, nom.)
Cabinet dimensions exclude
front and rear protrusions.
Weight
Shipping34.8 kg (77 lb), nom.
Net24.4 kg (54 lb), nom.
1-76Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersGeneral Information
Table 1-10 General Information (continued)
8712ET/ES and 8714ET/ES General Information
Measurement Speed Conditions
1
Typical
Cal
Type
Number
of Chans
MeasBWNumber
of Points
Freq
Span
2
Cycle
Time
Recall
3
State & Cal
4
Data
Xfer
5
Meas
Cycle
1-Port16500201100 MHz72 ms
1-Port165002012 GHz0.16 s
1-Port1400011100 MHz37 ms
1-Port1400021100 MHz42 ms
1-Port1400051100 MHz55 ms0.47 s26 ms0.63 s
1-Port14000101100 MHz76 ms
1-Port14000201100 MHz0.12 s0.58 s38 ms0.76 s
1-Port140002012 GHz0.18 s
1-Port14000401100 MHz0.21 s
1-Port14000801100 MHz0.38 s
1-Port140001601100 MHz0.73 s1.6 s0.16 s2.5 s
1-Port13700201100 MHz0.16 s
1-Port137002012 GHz0.22 s
1-Port11200201100 MHz0.33 s
1-Port112002012 GHz0.40 s
1-Port1250201100 MHz1.5 s
1-Port12502012 GHz1.6 s
1-Port115201100 MHz12.3 s
1-Port1152012 GHz12.3 s
1-Port2400051100 MHz56 ms0.63 s58 ms0.84 s
1-Port24000201100 MHz0.12 s0.84 s80 ms1.1 s
1-Port240001601100 MHz0.74 s2.6 s0.31 s3.7 s
2-Port1400051100 MHz0.11 s0.5 s26 ms0.72 s
2-Port14000201100 MHz0.24 s0.67 s38 ms1.04 s
2-Port140001601100 MHz1.5 s2.2 s0.16 s3.95 s
2-Port2400051100 MHz0.11 s0.71 s60 ms1.13 s
2-Port24000201100 MHz0.25 s0.94 s78 ms1.47 s
2-Port240001601100 MHz1.5 s3.5 s0.31 s5.48 s
6
See footnotes on the following page.
Service Guide1-77
Performance Checks8712ET/ES and 8714ET/ES
General InformationRF Network Analyzers
1. Measurements are always made with error correction enabled.
2. Center frequency is set to 1 GHz.
3. “Cycle Time” is the time required for the analyzer to finish one complete
sweep cycle including the forward sweep, retrace, bandcrossings, and calculation time when in the “Continuous Sweep” mode.
4. This is the time to recall both the system state and calibration data.
5. “Data Transfer” is performed using an HP S700 workstation. The GPIB port is
used to transfer “corrected” 64-bit, floating point numbers (real and imaginary).
6. A “Measurement Cycle” is defined as the time required for an HP S700 workstation to control the analyzer to: (1) recall the state and calibration (analyzer
is now in “sweep hold” mode), (2) sweep (using the “:INIT1;*OPC?” command),
and (3) transfer data. This may be less than the sum of the other columns
since a complete “Cycle Time” doesn’t need to be done for the controller to
transfer data.
1-78Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersFeatures
Features
Measurement
Number of Display Measurements
Two measurement displays are available. Each measurement can have independent
instrument sweep and control parameters (including frequency settings, IF bandwidth,
power level, number of trace points, averaging, format, and scale). The instrument can
display a single measurement or dual measurements, on a split (two graticules) or overlaid
(one graticule) screen.
Measurement Choices
• Narrowband:
ET models: reflection (A/R), transmission (B/R), A, B, R
ES models: S
• Broadband: X, Y, Y/X, X/Y, Y/R*, power (B*, R*), conversion loss (B*/R*)
NOTE: X and Y denote external broadband-detector inputs; * denotes internal broadband
detectors.
(A/R), S22 (B/R), S21 (B/R), S12 (A/R), A, B, R
11
Formats
• rectilinear: log or linear magnitude, SWR
• phase, group delay, real and imaginary, Smith chart, and polar
Trace Functions
The analyzer can display current-measurement data, memory data or
current-measurement with memory data simultaneously. Division of
current-measurement data by memory data is also available.
Display Annotations
Analyzer display annotations include start/stop, center/span, or CW frequency,
scale/division, reference level, marker data, softkey labels, warning and caution messages ,
screen titles, time and date, and pass/fail indication.
Limits
Measurement data can be compared to any combination of line or point limits for pass/fail
testing. User-defined limits can also be applied to an amplitude- or frequency-reference
marker. A limit-test TTL output is available on the rear panel for external control or
indiction. Limits are only available in rectilinear formats.
Service Guide1-79
Performance Checks8712ET/ES and 8714ET/ES
FeaturesRF Network Analyzers
Data Markers
Each measurement channel has eight markers. Markers are coupled between channels.
Any one of eight markers can be the reference marker for delta-marker operation.
Annotation for up to four markers can be displayed at one time.
Marker Functions
Markers can be used in absolute or delta modes. Other marker functions include:
mkr→center, mkr→reference, mkr→electrical delay, searches, tracking, and statistics.
Marker searches include: mkr to max, mkr to min, mkr→target, mkr bandwidth, notch,
multi-peak, and multi-notch. The marker-tracking function enables continuous update of
marker search values on each sweep. Marker statistics enable measurement of the mean,
peak-to-peak, and standard deviation of the data between two markers. For rapid tuning
and testing of cable-TV broadband amplifiers, slope and flatness functions are also
available.
Storage
Internal Memory
1.5 Mbytes of nonvolatile storage is available to store instrument states, measurement
data, screen images, and IBASIC programs. Instrument states can include all control
settings, limit lines, memory data, calibration coefficients, and custom display titles. If no
other data files are saved in nonvolatile memory, between about 30 and 150 instrument
states (depending on the instrument parameters) can be saved. Approximately 32 Mbytes
of volatile memory is also available for temporary storage of instrument states,
measurement data, screen images, and IBASIC programs.
Disk Drive
Trace data, instrument states (including calibration data), and IBASIC programs can also
be saved on floppy disks using the built-in 3.5” disk drive. All files are stored in MS-DOS®
-compatible format. Instrument data can be stored in binary or ASCII format (including
Touchstone/.s1p format), and screen graphics can be saved as PCX (bit mapped), HP-GL
(vector), or PCL5 (printer) files.
Network File System (NFS)
Remote files and directories, such as those stored on a computer, can behave like local files
and directories stored within the analyzer through the use of NFS and the LAN interface.
This allows test data from the analyzer to be saved directly to a remote directory—the
remote machine does not have to initiate an ftp session to retrieve the data. The analyzer
has a built-in NFS client. For Windows-based applications, third-party NFS-server
software must be installed on the PC. NFS is fully supported in most versions of UNIX.
1-80Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersFeatures
Data Hardcopy
Hardcopy prints can be made using PCL and PCL5 printers (such as HP DeskJet or
LaserJet series of printers), or Epson-compatible graphics printers. Single color and
multicolored formats are supported. Hardcopy plots can be automatically produced with
HP-GL-compatible plotters such as the HP 7475A, or with printers that support HP-GL.
The analyzer provides Centronics (parallel), RS-232C, GPIB, and LAN interfaces.
Automation
Controlling via GPIB
Interface: GPIB interface operates to IEEE 488.2 and SCPI standard interface
commands.
Control: The analyzer can either be the system controller, or pass bus control to
another active controller.
Data Transfer Formats:
• ASCII
• 32- or 64-bit IEEE 754 floating-point format
• mass memory-transfer commands allow file transfer between external controller and
analyzer
Controlling via LAN
The built-in LAN interface and firmware support data transfer and control via direct
connection to a 10 Base-T (Ethertwist) network. A variety of standard protocols are
supported, including TCP/IP, sockets, ftp, http, telnet, bootp, and NFS. The LAN interface
is a standard feature.
SCPI: The analyzer can be controlled by sending Standard Commands for
Programmable Instruments (SCPI) commands within a telnet session or via a socket
connection and TCP/IP (the default socket port is 5025). The analyzer’s socket
applications programming interface (API) is compatible with Berkeley sockets, Winsock
and other standard socket APIs. Socket programming can be done in a variety of
environments, including C programs, HP VEE, SICL LAN, or Java applets. A standard
web browser and the analyzer’s built-in web page can be used to remotely enter SCPI
commands via a Java applet.
FTP: Instrument state and data files can be transferred via file-transfer protocol (ftp).
An internal dynamic data disk, which is an ftp directory, provides direct access to
instrument states, screen dumps, trace data, and operating parameters..
HTTP: The instrument’s built-in web page can be accessed with any standard web
browser using hypertext transfer protocol (http) and the analyzer’s IP address. The
built-in web page can be used to control the network analyzer, view screen images,
download documentation, and link to other sites for firmware upgrades and
VXIplug&play drivers. Some word processor and spreadsheet programs, such as
Microsoft Word 97 and Excel 97, provide methods to directly import graphics and
data with http via a LAN connection.
Service Guide1-81
Performance Checks8712ET/ES and 8714ET/ES
FeaturesRF Network Analyzers
SICL LAN: SICL LAN is a LAN protocol using the Standard Instrument Control
Library (SICL). It provides control of your analyzer over the LAN using a variety of
computing platforms, I/O interfaces, and operating systems. With SICL LAN, a remote
analyzer can be controlled over the LAN with the same methods used for a local
analyzer connected directly to the controller with GPIB. SICL LAN can also be used
with Windows 95/98/NT, or HP-UX.
Bootp: Bootstrap protocol (bootp) allows a network analyzer to automatically configure
itself at power-on with the necessary information to operate on the network. After a
bootp request is sent by the analyzer, the host server downloads an IP address, a
gateway address, and a subnet mask. In addition, the analyzer can request an IBASIC
file, which automatically executes after the transfer is complete. For Windows-based
applications, third-party bootp-server software must be installed on the PC. Bootp is
fully supported in most versions of UNIX.
Programming using IBASIC
As a standard feature, all 8712ET/ES and 8714ET/ES network analyzers come with the
Instrument BASIC programming language (IBASIC). IBASIC facilitates automated
measurements and control of other test equipment, improving productivity. For simpler
applications, you can use IBASIC as a keystroke recorder to easily automate manual
measurements. Or, you can use an optional, standard PC keyboard to write custom test
applications that include:
• special softkey labels
• tailored user prompts
• graphical setup diagrams
• barcode-reading capability
• control of other test instruments via the GPIB, serial or parallel interfaces
Measurement Calibration
Measurement calibration significantly reduces measurement uncertainty due to errors
caused by system directivity, source match, load match, frequency response, and crosstalk.
The analyzer features factory-installed default calibrations that use vector error
correction, so that measurements can be made on many devices without performing a user
calibration.
For greater accuracy, especially for special test setups with significant loss or reflection,
user calibrations should be performed. For reflection measurements, both one-port and
two-port calibrations are available (two-port calibration requires an ES model). For
transmission measurements, the following calibrations are available: response, response
and isolation, enhanced response, and two-port (two-port calibration requires an ES
model).
1-82Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
Isolation on OFF
RF Network AnalyzersFeatures
Calibration Interpolation
Calibration interpolation is always active. The analyzer automatically recalculates the
error coefficients when the test frequencies or the number of points have changed. The
resulting frequency range must be within the frequency range used during the user
calibration. If this is not the case, the analyzer reverts to the factory default calibration.
When calibration interpolation is used, the analyzer displays the C? annotation. System
performance is not specified for measurements using calibration interpolation.
Available Calibrations
ES models only
• Two-Port Calibration
Compensates for frequency response, load match, source match, and directivity
errors while making S-parameter measurements of transmission (S
reflection (S
, S22). Compensates for crosstalk when the softkey
11
is toggled ON. Requires open, short, load, and through standards.
ET and ES models: transmission measurements
• Normalization
Simultaneous magnitude and phase correction of frequency response errors.
Requires a through connection. Used for both narrowband and broadband detection.
Does not support calibration interpolation.
, S12) and
21
• Response
Simultaneous magnitude and phase correction of frequency response errors.
Requires a through connection.
• Response and Isolation
Compensates for frequency response and crosstalk errors. Requires a load
termination on both test ports and a through standard.
• Enhanced Response
Compensates for frequency response and source match errors. Requires open, short,
load and through standards.
ET and ES models: reflection measurements
• One-Port calibration
Compensates for frequency response, directivity, and source match errors. Requires
an open, short, and load standards.
Service Guide1-83
Performance Checks8712ET/ES and 8714ET/ES
FeaturesRF Network Analyzers
Calibration Kits
Data for several standard calibration kits are stored in the instrument for use by
calibration routines. They include:
• 3.5 mm (85033D with Option 001
• type-F 75 ohm (85039B)
• type-N 50 ohm (85032B with Option 001
2
• type-N 75 ohm (85036B/E
)
• 7 mm (APC-7) (85031B)
• 7-16 (85038A
3/F4/M5
)
In addition, you can also describe the standards for a user defined kit (for example,
open-circuit capacitance coefficients, offset short length, or through-standard loss).
For more information about calibration kits available from Agilent Technologies, consult
the RF Economy Network Analyzers Configuration Guide, literature number 5967-6315E.
1
)
1
, or 85032E2)
1. Option 001 doesn’t include 7 mm adaptors.
2. “E” version doesn’t include adaptors or cal standards with female connectors.
3. Contains cal standards with female connectors and male connectors.
4. Contains cal standards with female connectors only.
5. Contains cal standards with male connectors only.
1-84Service Guide
8712ET/ES and 8714ET/ESPerformance Checks
RF Network AnalyzersFeatures
Options
75 Ohms (Option 1EC)
Provides 75 ohm system impedance.
Fault location and structural return loss (Option 100)
For fully characterizing cable performance and antenna-feedline systems, this option
provides both fault location and structural return loss capability. Fault-location
measurements help identify where cable or system faults, such as bends, shorts, or
corroded or damaged connectors, occur. In addition to displaying faults in terms of
distance into the cable or feedline, the magnitude of the fault is also displayed.
Structural return loss is a special case of return loss (reflection) measurements,
optimized for measuring periodic reflections of small magnitude. These periodic
reflections can occur from physical damage to the cable caused by rough handling, or
from minor imperfections imparted during the manufacturing process. Structural
return loss problems occur when these periodic reflections sum at half-wavelength
intervals, causing high signal reflection (and low transmission) at the corresponding
frequency.
Step Attenuator (Option 1E1): 8712ET/8714ET only
This option adds a built-in 60 dB step attenuator, extending the source output power
low-end range to −60 dBm. This attenuator is a standard feature on the
8712ES/8714ES.
Transport case and fault location and structural return loss (Option 101)
Combines a rugged transport and operation case (part number 08712-60059) with
Option 100 for field measurements of fault location and structural return loss.
Test Sets
87050E multiport test sets
Provides a complete solution for testing 50 ohm multiport devices like duplexers, power
dividers, or multi-couplers. Test sets can be configured with four, eight, or twelve test
ports (for more information, please consult the product brochure, literature number
5968-4763E).
87075C multiport test sets
Provides a complete solution for testing 75 ohm multiport devices like CATV
distribution amplifiers or multi-taps. Test sets can be configured with six or twelve test
ports (for more information, please consult the product brochure, literature number
5968-4766E).
Switching test sets
Switching test sets enhance productivity by allowing multiple measurements with a
single connection to the device under test. They are available in 50 and 75 ohm
versions, in a variety of configurations and connector types. Please contact your sales
representative for more information.
Service Guide1-85
8712ET/ES and 8714ET/ES
RF Network Analyzers
2Adjustments
Service Guide2-1
Adjustments 8712ET/ES and 8714ET/ES
Introduction RF Network Analyzers
Introduction
This chapter contains procedures to adjust the analyzer. Generally these adjustment
procedures are only used when hardware has been replaced in the instrument, or if there
is a failure in the operators check or the performance tests (seeChapter 1). Refer to
Table3-1, “Post Repair Procedures,” on page3-23 to determine which adjustment
procedures you need to do when a particular part has been replaced.
Adjustments #102 through #111 affect the correction constants that are stored in the
non-volatile memory of the instrument. As adjustments are performed, new correction
constants are created.The user controls when the new correction constants are stored in
the non-volatile memory. The frequency accuracy adjustment, which could require a
potentiometer adjustment, and the serial number (Adjustment #100) do not affect the
correction constants.
Before performing adjustments, turn on the analyzer and allow it to warm up for one hour.
Most adjustments are accessed through the analyzer's service menus.
You should perform the adjustment procedures in the order given here:
1.”Frequency Accuracy Adjustment”on page 2-5
2.”Switched Gain Correction, Adjustment #102”on page 2-10
3.”External Detector Gain Correction, Adjustment #103”on page 2-12
11.”R* Frequency Response Correction, Adjustment #109”on page 2-25
CAUTION Procedures 2 through 11 all generate correction constants. For efficient operation,
these correction constants must be handled properly. Please read the“Correction
Constants (CCs) Storage” on page2-4 carefully.
CAUTION The preset power level is user-definable. After pressing in any
adjustment procedure, change the power level to 0 dBm if necessary.
PRESET
NOTE During the adjustments, the analyzer will display a list of instructions in which the
names for the connectors may vary. For user clarification, “REFLECTION,” “RF
OUT,” and “PORT 1” all refer to the connector closest to the analyzer’s display, while
“TRANSMISSION,” “RF OUT,” and “PORT 2” all refer to the connector farthest from
the display.
2-2 Service Guide
8712ET/ES and 8714ET/ESAdjustments
RF Network AnalyzersIntroduction
Required Equipment
Table 2-1 Required Equipment for Adjustments
DescriptionHP/Agilent Model #
Power meter
Power sensor, 50Ω8482A
Low power sensor
Cal kit type-N 50Ω
Cal kit type-N 75Ω
6 dB attenuator 50Ω8491A/B/C Opt 006
10 dB attenuator 50Ω
10 dB attenuator 75Ω
20 dB attenuator 50Ω8491A/B/C Opt 020
20 dB attenuator 75Ω
Minimum loss pad11852B—✓
Frequency counter535xB or equivalent
Precision cable 50Ω8120-8862 or 8120-6469✓—
Precision cable 75Ω8120-8898 or 8120-6468—✓
Voltage reference source 08712-60031
10 dB attenuator 50Ω8491A, Option 010
BNC cable, 50Ω8120-1839 or equivalent
Adapter, type-N (m) to
BNC (f)
1. An 85032B Option 001 is a subset of the standard 85032B. This option eliminates four 7-mm
adapters that are not needed, resulting in a significant cost savings.
2. The required attenuators depend on the firmware revision of the analyzer. The attenuators
listed above apply to all versions of firmware.
3. The frequency counter is useful when very accurate frequency measurements are needed. However, if you are using a synthesized spectrum analyzer, the frequency can be measured by using
the spectrum analyzer. Note that the frequency measurement using a spectrum analyzer will
not be quite as accurate as a frequency counter, but is sufficient in most cases. (An 8560 series
spectrum analyzer may be used.)
437B, 438A
E4418A, E4419A
8481D or 8484A
85032B Opt 001
85036B
2
8491A/B/C Opt 010
2
0955-0767
2
0955-0768—✓
1250-0780 or equivalent
1
50Ω
75ΩNotes
STD
✓✓
Do not use an 436A.
✓✓
✓✓
For the ES models and
ET Option 1E1.
✓—
—✓
✓—
✓—
—✓
✓—
——
✓✓
✓
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
Required for adjustments only if
analyzer does not have an internal
attenuator installed. (May substitute
2 min loss pads.)
Required for adjustments only if
analyzer does not have an internal
attenuator installed.
Required for adjustments only if
analyzer does not have an internal
attenuator installed. (May substitute
2 min loss pads and a 50Ω 10 dB pad.)
May not be required.
Included in the service kit, part
number 08712-60012.
During normal operation, the network analyzer uses correction constants that are stored
in the RAM buffer. When the analyzer's line power is switched on, the analyzer copies the
current correction constants from the CPU EPROM into the RAM buffer.
During the adjustment tests, the analyzer generates correction constants and saves them
to the RAM buffer only. If the power is switched off without performing the steps below , the
newly generated correction constants will be lost.
To transfer the newly generated correction constants from the RAM buffer to the CPU
EPROM, press the following keys:
SYSTEM OPTIONS
To make a backup copy of the correction constants to a 3.5" disk, perform the following
steps:
1. Insert a formatted 3.5" disk into the analyzer’s internal disk drive.
2. Press. This
creates a file (or writes over an existing file) where the correction constants are stored.
This file allows you to quickly restore the adjustment data if you replace the CPU
board.
When performing the adjustment procedures, it is beneficial to store the correction
constants to the 3.5" disk. This allows you to restore data quickly in case you need to
switch off the analyzer's line power before completing the adjustments.
CAUTIONIf the analyzer's line power is switched off at any time during the
SYSTEM OPTIONS
adjustments, you must reload the previously stored correction constants (if
they have been saved to disk) before proceeding. Failure to do so will result in
the loss of previously performed adjustment data at the time of the next
correction constant data save. The adjustments will have to be performed
again.
If the correction constants were never saved to disk, the adjustments will
have to be performed again as well.
.
2-4Service Guide
8712ET/ES and 8714ET/ESAdjustments
RF Network AnalyzersFrequency Accuracy Adjustment
Frequency Accuracy Adjustment
In this adjustment, the frequency accuracy of the analyzer's source is calibrated by
adjusting the 10 MHz internal reference clock.
Required Equipment
Recommended HP/Agilent
Equipment Description
For 50Ω AnalyzersFor 75Ω Analyzers
Model or Part Number
Frequency counter
BNC cable
Minimum loss pad
Adapter, type-N(m) to BNC(f)
535xB
8120-1839
N/A
1250-0780
Warmup Time
Allow the analyzer to warm up for one hour.
Estimated Adjustment Time
This adjustment takes approximately five minutes to perform.
Procedure
1. Connect the equipment as shown in Figure 2-1.
Figure 2-1 Setup for Frequency Accuracy Adjustment
535xB
8120-1839
11852B
1250-0780
Service Guide2-5
Adjustments 8712ET/ES and 8714ET/ES
Frequency Accuracy Adjustment RF Network Analyzers
2.Set the frequency counter input switches to the “10 Hz – 500 MHz” and 50Ω positions.
3.On the analyzer, press (seecaution on page 2-2)
MENU
Trigger Hold
PRESETFREQ
. If the frequency counter reading is 500 MHz±2500 Hz, you
CW
500
MHz
do not need to make this adjustment. However, you can still make this adjustment to
improve the frequency accuracy.
4.To proceed with the adjustment, disconnect the frequency counter and adapter(s) from
the analyzer’s RF OUT/ PORT 1 port (seeFigure 2-1on page2-5).
5.Remove the handles and front panel by following these steps and referring to
Figure2-2.
a.Remove the trim strip from the handles.
b.Remove the screws attaching each handle to the analyzer.
c.Pull the analyzer toward you until it extends about two inches over the edge of the
table top.
d.Grasp and pull the front panel with two hands: one on the top-middle of the panel,
and the other on the bottom-middle of the panel.
Figure 2-2 Removing the Handles and Front Panel
6.Reconnect the frequency counter and adapter(s) to the RF OUT/PORT 1 port (see
Figure 2-1on page2-5).
7.To obtain a counter reading of 500 MHz±2500 Hz or better, adjust R204 on the
fractional-N/reference assembly (A3) (accessible through a hole in shield of the A3
assembly).
8.Reassemble the analyzer.
2-6 Service Guide
8712ET/ES and 8714ET/ESAdjustments
Service
Instrument Info
RF Network AnalyzersSet Serial Number, Adjustment #100
Set Serial Number, Adjustment #100
Adjustment #100 cannot be used to set an established serial number. The procedure
outlined below can only be used to enter the serial number if the analyzer displays a
default serial number value of
If it is necessary to modify a serial number that is other than the default value, contact the
nearest Agilent Technologie sales or service office (see Chapter 9).
NOTEWhere XXXXXXXXXX appears throughout this procedure, replace those
characters with the serial number of your analyzer, but maintain the leading
and following apostrophes (both are ASCII character 39). For example, if the
serial number of your analyzer is US34405555, the HP BASIC line of code
would be:
OUTPUT 800;"DIAG:SNUM 'US34405555'"
IMPORTANT: Make sure there is a space between SNUM and the first
apostrophe.
?????????? (ten question marks).
Required Equipment
A PC-compatible keyboard is required.
If a keyboard is not available, this adjustment can be done from the front panel. (Refer to
the note in the following procedure.)
Estimated Adjustment Time
This adjustment takes approximately two minutes to perform.
Procedure
1. Write down the 10-character serial number, exactly as shown on the analyzer rear-panel label.
2. With an external keyboard connected, press to view the IBASIC command line. Then
type the following:
OUTPUT 800;"DIAG:SNUM 'XXXXXXXXXX'"
then press.
ENTER
NOTEIf you are using the front panel to set the serial number, enter the following
IBASIC code, then “run” the program.
OUTPUT 800;"DIAG:SNUM 'XXXXXXXXXX'"
END
ESC
3.
Service Guide2-7
Refer to Chapter 5 of the HP Instrument Basic User’s Handbook Supplement
for information on programming from the front panel of the instrument.
Press to verify that you correctly
installed the serial number. The serial number is displayed in the dialogue box.
SYSTEM OPTIONS
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.