Agilent 87050E Data Sheet

Agilent 87050E 50 Ohm Multiport Test Sets
Data Sheet
This document describes the performance and fea­tures of Agilent Technologies 87050E 50 ohm multi­port test sets, both as standalone units and when combined with 50 ohm 8712ET, 8712ES, 8714ET, or 8714ES network analyzers. The following options are available:
• Option 004 (4 ports)
• Option 008 (8 ports)
• Option 012 (12 ports)
For more information about these test sets, please read the following documents:
Agilent 87050E, brochure, literature number, 5968-4763E
Agilent 87050E, configuration guide, literature number, 5968-4765E
Agilent 87050E 3 MHz to 2.2 GHz
2
All specifications and characteristics apply over a 25 °C ± 5 °C range (unless otherwise stated) and 30 minutes after the instrument has been turned on.
Definitions
Specification:
Warranted performance. Specifications include guardbands to account for the expected statistical distribution, measurement uncertainties, and changes in performance due to environmental con­ditions.
Characteristic:
A performance parameter that the product is expected to meet before it leaves the factory, but is not verified in the field and is not covered by the product warranty. A characteristic includes the same guardbands as a specification.
Typical:
Expected performance of an average unit. A typical does not include guardbands. It is not covered by the product warranty.
Nominal:
A general, descriptive term that does not imply a level of performance. It is not covered by the prod­uct warranty.
Supplemental information:
May include typical, nominal, or characteristic values.
Calibration:
The process of measuring known standards from a calibration kit to characterize the systematic (repeatable) errors of a network analyzer.
Corrected (residual) performance:
Indicates performance after error correction (cali­bration). It is determined by the quality of calibra­tion standards and how well “known” they are, plus system repeatability, stability, and noise.
Uncorrected (raw) performance:
Indicates instrument performance without error correction. The uncorrected performance affects the stability of a calibration.
System performance:
Performance of a complete multiport test system, which includes an 87050E test set and an 8712ET/ES or 8714ET/ES network analyzer.
Test Set Cal:
The calibration of a multiport test system, requir­ing the connection of known calibration standards to all of the ports that will be used for measure­ments.
SelfCal:
An automated system calibration that uses calibra­tion standards internal to the test set and the most recent Test Set Cal data to calibrate the test system.
Environmental specifications:
Environmental specifications bound the external conditions for which the specifications are valid. The environmental specifications also bound the external conditions the test set may be subject to without permanently affecting performance or causing physical damage.
Introduction
3 5 8
9 12 13 14 15 16
Table of contents
System performance, two-port calibration System performance, T/R calibration System performance, uncorrected System performance, general Test set input/output performance Test set general information Physical dimensions Block diagrams System features
3
Agilent 85032B/E Type-N calibration kit
Specification1(dB) Typical2(dB)
Parameter 3 MHz 1.3 GHz 2.2 GHz
to 1.3 GHz to 2.2 GHz to 3 GHz
Directivity 47 40 37
Source match 37 35 30
Load match 47 40 37
Reflection tracking ±0.1 ±0.1 ±0.2
Transmission tracking ±0.1 ±0.1 ±0.2
System performance Two-port calibration (user)
Transmission uncertainty (typical)
3
1. These specifications and supplementals apply under the following conditions: measurement uses the “Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibration has been performed; the test set must be used with an 8712ES or 8714ES network ana­lyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on; measurements are made at an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. Corrected system performance is changed to typical when the current temperature has drifted beyond ± 1 °C of the last valid Test Set Cal. Typicals are valid only when the cur­rent temperature is within 25 °C ± 5 °C, and within ± 1 °C of the most recent SelfCal.
3. These uncertainty curves only include the effects of the test port(s) within the measurement path. The effect of the uncorrected match of test ports outside the measurement path is ignored, and is dependent on the isolation between the ports of the DUT that are within the measurement path and ports of the DUT that are outside the measurement path.
Magnitude
Magnitude
Phase
Phase
4
System performance Two-port calibration (user)
Specification1(dB) Typical2(dB)
Parameter 3 MHz 1.3 GHz 2.2 GHz
to 1.3 GHz to 2.2 GHz to 3 GHz
Directivity 40 38 35
Source match 38 35 30
Load match 40 38 35
Reflection tracking ±0.1 ±0.1 ±0.2
Transmission tracking ±0.1 ±0.1 ±0.2
Transmission uncertainty (typical)
3
Magnitude
Magnitude
Phase
Phase
1. These specifications and supplementals apply under the following conditions: measurement uses the “Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibration has been performed; the test set must be used with an 8712ES or 8714ES network analyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on; measurements are made at an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. Corrected system performance is changed to typical when the current temperature has drifted beyond ± 1 °C of the last valid Test Set Cal. Typicals are valid only when the cur­rent temperature is within 25 °C ± 5 °C, and within ± 1 °C of the most recent SelfCal.
3. These uncertainty curves only include the effects of the test port(s) within the measurement path. The effect of the uncorrected match of test ports outside the measurement path is ignored, and is dependent on the isolation between the ports of the DUT that are within the measurement path and ports of the DUT that are outside the measurement path.
Agilent 85033D 3.5mm calibration kit
5
Agilent 85032B/E Type-N calibration kit
1. These specifications and supplementals are valid for reflection (one-port) and transmission (enhanced-response) Test Set Cals, and apply under the following conditions: meas­urement uses the “Fine” (15 Hz) bandwidth, no averaging, and isolation cal “on”; Test Set Cals use the “Fine” (15 Hz) bandwidth, 16 averages, and assume an isolation calibra­tion has been performed; the test set must be used with an 8712ET, 8712ES, 8714ET, or 8714ES network analyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on; measurements are made at an environmental temperature of 25 °C ± 5 °C and within ± 1 °C of the last valid Test Set Cal.
2. Corrected system performance is changed to typical when the current temperature has drifted beyond ± 1 °C of the last valid Test Set Cal. Typicals are valid only when the cur­rent temperature is within 25 °C ± 5 °C, and within ± 1 °C of the most recent SelfCal.
System performance Transmission/reflection (T/R) calibration (user)
Specification1( dB) Typical2(dB)
Parameter 3 MHz 1.3 GHz 3 MHz 1.3 GHz 2.2 GHz
to 1.3 GHz to 2.2 GHz to 1.3 GHz to 2.2 GHz to 3 GHz
Directivity 47 40 40
Source match 37 32 30
Load match (reflection calibration) 18 18 23 24 20
Load match (transmission calibration) 14 14 18 18 14
Reflection tracking ±0.1 ±0.1 ±0.2
Transmission tracking ±0.1 ±0.1 ±0.2
6
System performance Transmission/reflection (T/R) calibration (user)
1. These uncertainty curves only include the effects of the test port(s) within the measurement path. The effect of the uncorrected match of test ports outside the measurement path is ignored, and is dependent on the isolation between the ports of the DUT that are within the measurement path and ports of the DUT that are outside the measurement path.
Magnitude, ET models with attenuator
Magnitude, ET models without attenuator
Phase, ES models
Phase, ET models with attenuator
Phase, ET models without attenuator
Agilent 85032B/E Type-N calibration kit (continued)
Transmission uncertainty, enhanced response calibration (typical)
1
Magnitude, ES models
7
Agilent 85032B/E Type-N calibration kit
1. These uncertainty curves only include the effects of the test port(s) within the measurement path. The effect of the uncorrected match of test ports outside the measurement path is ignored, and is dependent on the isolation between the ports of the DUT that are within the measurement path and ports of the DUT that are outside the measurement path.
Magnitude, ES models
Phase, ES models
Magnitude, ET models with attenuator Phase, ET models with attenuator
Magnitude, ET models without attenuator
Phase, ET models without attenuator
8
Type-N
Specification1 ( dB) Typical (dB)
Parameter 3 MHz 1.3 GHz 3 MHz 1.3 GHz 2.2 GHz
to 1.3 GHz to 2.2 GHz to 1.3 GHz to 2.2 GHz to 3 GHz
Source match, ratioed 12 10 17 14 11
Load match, test port unselected
2
18 18 23 24 20
Load match, test port selected
3
12 12 16 17 13
Transmission tracking
4
±3 ±2.5 ±3
Reflection tracking
4
±3 ±2.5 ±3
Crosstalk5, 8712ET/8714ET 93 89 81
Crosstalk5, 8712ES/8714ES 83 79 71
System performance Uncorrected
1. These specifications and supplementals apply under the following conditions: measurement uses the “Fine” (15 Hz) bandwidth, with narrowband detection and no averaging; the test set must be used with an 8712ET, 8712ES, 8714ET, or 8714ES network analyzer with firmware revision E.06.00 or later; the test set and the analyzer must have had their performance verified within the last year; both instruments must have warmed up for at least 30 minutes after turn-on.
2. This is the match of any test port that is unselected (not in the measurement path). If the network analyzer is performing a reflection measurement with one-port calibration, then only one port on the test set is selected (the source port). If the network analyzer is performing a transmission measurement or a reflection measurement with two-port calibration, then only two ports on the test set are selected (the source and load ports).
3. This is the match of the test set port that has been selected as the load port. The network analyzer must be making a transmission measurement or a reflection measurement with two-port calibration for a test port to be selected as the load port.
4. The uncorrected tracking terms are defined as the deviation over the defined frequency band, ignoring offset loss.
5. Crosstalk is computed by normalizing the result of an isolation measurement to a through measurement between the two ports such that the path losses are taken into account. Isolation is defined as the transmission signal measured between any two ports of the test system when those two ports are terminated with shorts.
9
Specification Typical
System bandwidth
Fine Fine Med Wide Wide 15 Hz 15 Hz 4000 Hz 6500 Hz
Network analyzer 3 MHz 1.3 GHz 3 MHz 1.3 GHz 3 MHz 1.3 GHz 3 MHz 1.3 GHz
to to to to to to to to
1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz 1.3 GHz 2.2 GHz
System noise floor1(dBm)
8712ET/8714ET –97 –95 –106 –103 –82 –79 –42 –39
8712ES/8714ES –88 –86 –100 –97 –75 –72 –42 –39
System dynamic range2(dB)
8712ET, no attenuator 101 97 112 108 87 83 49 45
8712ET, with attenuator 100 96 111 107 86 82 48 44
8712ES 89 85 103 99 78 74 45 41
8714ET, no attenuator 99 95 109 105 84 80 46 42
8714ET, with attenuator 97 93 108 104 83 79 45 41
8714ES 86 82 100 96 81 77 42 38
1. Noise floor is defined as the rms value of the trace (in linear format) for a transmission measurement in CW mode, using the “Fine” bandwidth (15 Hz), the test ports terminated in loads, 0 dBm at the test set source port, and no averaging. This measurement ignores the effects of crosstalk. This is a system specification–the test set increases the network analyzer noise floor by adding loss to the network analyzer measurement.
2. The system dynamic range is calculated as the difference between the receiver minimum input (noise floor plus calibrated crosstalk) and the system’s maximum output power. System dynamic range applies to transmission measurements only, since reflection measurements are limited by directivity.
System performance General
10
Specification Supplemental information
System bandwidth
Network analyzer Fine Fine Med Wide Wide
15 Hz 15 Hz 4000 Hz 6500 Hz
Receiver dynamic range1(dB)
8712ET/8714ET 115 121 96 58
8712ES/8714ES 106 115 90 57
System performance General (continued)
1. The receiver dynamic range is calculated as the difference between the receiver minimum input (noise floor) and the receiver maximum input. Receiver dynamic range applies to transmission measurements only, since reflection measurements are limited by directivity.
2. The reference power for dynamic accuracy is –20 dBm.
Magnitude, ES models
Magnitude, ET models
Phase, ES models
Phase, ET models
11
Network analyzer Specification Characteristic (dBm)
Output power (system maximum)
<1 GHz >1 GHz
8712ET, no attenuator 10.5 7
8712ET, with attenuator 9.5 6
8712ES 7.5 4
8714ET, no attenuator 5.5 2
8714ET, with attenuator 4.5 1
8714ES 3.5 –1
Parameter Specification Supplemental information
Port switching time1(sec)
Reflection calibration 0.7
Transmission enhanced calibration 0.7
Two-port calibration 0.7
SelfCal time2(sec)
Reflection calibration 1 (typical)
Transmission enhanced calibration 3 (typical)
Two-port calibration 8 (typical)
1. This is the time required to switch to any new port configuration, and requires that the Test Set Cal has been performed for 201 points, the new measurement configuration is 201 points, and no new SelfCal occurs during switching.
2. This is the time required to perform the SelfCal for any single port configuration, assuming that the Test Set Cal was performed for 201 points, the new measurement configu­ration is 201 points, and the measurement bandwidth is “Med Wide” (4000 Hz). SelfCal times for other settings can be found in Table 4-1 of the Agilent 87050E/87075C Users and Service Guide (87050-90026).
12
Test set input/output performance
Parameter Specification Supplemental
Frequency range 3 MHz to 2.2 GHz
RF input power
Maximum input power 16 dBm (nominal) at 0.1 dB compression
1
Input damage power 20 dBm (characteristic)
Parameter Specification (dB) Typical (dB)
3 MHz 1.3 GHz 3 MHz 1.3 GHz 2.2 GHz to 1.3 GHz to 2.2 GHz to 1.3 GHz to 2.2 GHz to 3 GHz
Source match, test port
2
14 11 18 15 12
Load match, test port unselected
3
18 18 23 24 20
Load match, test port selected
4
14 14 18 18 14
Interconnect match, reflection port
5
12 10 17 15 10
Interconnect match, transmission port
5
11 8 14 13 11
Insertion loss, reflection port to port-n
6
7.5 9.5 6 8 12
Insertion loss, transmission port to port-n
6
7.5 9.5 6 8 12
Tracking, reflection port to port-n
6
1.5 1.2 1.5
Tracking, transmission port to port-n
6
1.5 1.2 1.5
Crosstalk, uncalibrated, adjacent ports
7
90 90 100 95 85
Crosstalk, uncalibrated, non-adjacent ports790 90 105 100 90
1. Compression is defined for the test set, independent of the network analyzer.
2. This is the match of the test set port that has been selected as the source port. The test set interconnect reflection port must be terminated with a load standard from an
Agilent 85032B/E calibration kit.
3. This is the match of any test port that is unselected (not in the measurement path). Only the source port of the test set is selected when you make a reflection measurement
with one-port calibration. Only the source and load ports of the test set are selected when you make a transmission measurement, or a reflection measurement with two-port calibration.
4. This is the match of the test set port that has been selected as the load port. A transmission or two-port measurement is required for a test port to be selected as the load port.
The test set interconnect transmission port must be terminated with a load standard from an 85032B/E calibration kit.
5. This is the match of the test set interconnect ports (transmission and reflection ports) with the test set in transmission, or two-port mode. The selected test set test ports must
be terminated with load standards from 85032B/E calibration kits.
6. The reflection and transmission ports of the test set are connected to the corresponding ports of the network analyzer. Port-n is any one of the test ports used to connect to
the device-under-test.
7. This is crosstalk of the test set measured between the test set’s interconnect ports, with shorts on the selected test ports. Crosstalk is computed by normalizing the result of
an isolation measurement to a through measurement between the two ports such that the path losses are taken into account. Isolation is defined as the transmission signal measured between any two ports of the test system when these two ports are terminated with shorts.
13
Description Specification Supplemental information
Front panel ports
87050E, Option 004 4 Type-N, female 50Ω (nominal)
87050E, Option 008 8 Type-N, female 50Ω (nominal)
87050E, Option 012 12 Type-N, female 50Ω (nominal)
Real panel
Parallel in connector 25-pin D-subminiature female (DB-25)
Parallel out connector 25-pin D-subminiature female (DB-25)
Line power
1
Frequency 47 to 63 Hz
Input voltage, operating 90 to 264 V
Input power <45 W (typical)
General environment
ESD Minimize using static-safe work procedures and
an antistatic bench mat
(Part number 9300-0797)
Dust Minimize for optimum reliability
Operating environment
General Indoor use only
Temperature 0 °C to + 55 °C
Humidity (relative) less than 80% RH up to 31 °C,
decreasing linearly to 50% RH at 40 °C
Altitude 0 to 4.5 km (15,000 ft)
Storage conditions
Temperature –40 °C to + 70 °C
Humidity (relative) 0% to 90% RH at + 65 °C (noncondensing)
Altitude 0 to 15 km (50,000 ft)
Cabinet dimensions
Height x width x depth 132.8 x 425 x 497 mm (nominal)
5.2 x 16.7 x 19.6 in (nominal)
Cabinet dimensions exclude front and
rear protrusions
Weight
Net 8 kg (18 lb) (nominal)
Shipping 11 kg (24 lb) (nominal)
Test set general information
1. A third-wire ground is required.
14
Physical dimensions
Physical dimensions of the Agilent 87050E Option 004 multiport test set
1
Physical dimensions of the Agilent 87050E Option 008 multiport test set
1
Physical dimensions of the Agilent 87050E Option 012 multiport test set
1
1. These dimensions exclude rear protrusions.
15
Block diagrams
Block diagram for the Agilent 87050E (only one test set port pair is shown)
Block diagram for the Agilent 8712ES and 8714ES
Block diagram for the Agilent 8712ET and 8714ET
REAR PANEL
ADC and Processor
CRT
AUX Input
Y
Input B
Input B*
X
Input R
Input R*
Input A
External Detectors
X
Reference
Reflected
Y
RF
Source
Incident
Transmission
FRONT PANEL Reflection
With Attenuator Option 1E1
Narrowband Detector
Broadband Detector
(RF Out)
Device
Under
Test
External Detectors
Y
Input B
Input B*
Input R
Input R*
Input A
X
Reference
Port 1
RF
Source
Device
Under
Test
REAR PANEL
ADC and Processor
CRT
FRONT PANEL
Narrowband Detector
Broadband Detector
AUX Input
Y
X
Transmission
(RF In)
Port 2
16
Test set control
Control of the switches inside the test set and cali­bration of the test system can be accomplished from the front panel of the network analyzer—an external computer is not required. However, the analyzers are fully programmable for use in auto­mated test environments.
Measurement
Number of display measurements
Two measurement displays are available, with independent control of display parameters includ­ing format type, scale per division, reference level, reference position, and averaging. The displays can share network analyzer sweep parameters, or, by using alternate sweep, each measurement can have independent sweep parameters including frequen­cy settings, IF bandwidth, power level, and number of trace points. The instrument can display a single measurement, or dual measurements on a split (two graticules) or overlaid (one graticule) screen.
Measurement choices
• Narrowband ET models: reflection (A/R), transmission (B/R), A, B, R ES models: S11(A/R), S22(B/R), S21(B/R), S12(A/R), A, B, R
• Broadband X, Y, Y/X, X/Y, Y/R*, power (B*, R*), conversion loss (B*/R*)
Note: X and Y denote external broadband-detec­tor inputs; * denotes internal broadband detec­tors.
Formats
Log or linear magnitude, SWR, phase, group delay, real and imaginary, Smith chart, polar, and imped­ance magnitude
Trace functions
Current data, memory data, memory with current data, division of data by memory
Display annotations
Start/stop, center/span, or CW frequency, scale per division, reference level, marker data, softkey labels, warning and caution messages, screen titles, time and date, and pass/fail indication
Limits
Measurement data can be compared to any combi­nation of line or point limits for pass/fail testing. User-defined limits can also be applied to an amplitude- or frequency-reference marker. A limit­test TTL output is available on the rear panel for external control or indication. Limits are only available with rectilinear formats.
Data markers
Each measurement channel has eight markers. Markers are coupled between channels. Any one of eight markers can be the reference marker for delta-marker operation. Annotation for up to four markers can be displayed at one time.
Marker functions
Markers can be used in absolute or delta modes. Other marker functions include marker to center frequency, marker to reference level, marker to electrical delay, searches, tracking, and statistics. Marker searches include marker to maximum, marker to minimum, marker to target value, band­width, notch, multi-peak and multi-notch. The marker-tracking function enables continuous update of marker search values on each sweep. Marker statistics enable measurement of the mean, peak-to-peak and standard deviation of the data between two markers. For rapid tuning and testing of cable-TV broadband amplifiers, slope and flat­ness functions are also available.
Storage
Internal memory
1.5 Mbytes (ET models) or 1 Mbyte (ES models) of nonvolatile storage is available to store instrument states, measurement data, screen images, and IBA­SIC programs. Instrument states can include all control settings, limit lines, memory data, calibra­tion coefficients, and custom display titles. If no other data files are saved in nonvolatile memory, between approximately 20 and 150 instrument states can be saved (depending on the model type and on instrument parameters). Approximately 14 Mbytes of volatile memory are also available for temporary storage of instrument states, measure­ment data, screen images, and IBASIC programs.
System features
17
Disk drive
Trace data, instrument states (including calibra­tion data), and IBASIC programs can be saved on floppy disks using the built-in 3.5 inch disk drive. All files are stored in MS-DOS®-compatible format. Instrument data can be saved in binary or ASCII format (including Touchstone/.s2p format), and screen graphics can be saved as PCX (bit-mapped), HP-GL (vector), or PCL5 (printer) files.
NFS: See description under Control via LAN
Data hard copy
Hard copy prints can be made using PCL and PCL5 printers (such as HP DeskJet or LaserJet series printers), or Epson-compatible graphics printers. Single color and multicolor formats are supported. Hard copy plots can be automatically produced with HP-GL-compatible plotters such as the Agilent 7475A, or with printers that support HP-GL. The analyzer provides Centronics (parallel), RS-232C, GPIB, and LAN interfaces.
Automation
Controlling via GPIB Interface
The GPIB interface operates to IEEE 488.2 and SCPI standard-interface commands.
Control
The analyzer can either be the system controller, or pass bus control to another active controller.
Data transfer formats
• ASCII
• 32- or 64-bit IEEE 754 floating-point format
• Mass-memory-transfer commands allow file transfer between external controller and analyzer.
Control via LAN
The built-in LAN interface and firmware support data transfer and control via direct connection to a 10 Base-T (Ethertwist) network. A variety of stan­dard protocols are supported, including TCP/IP, sockets, ftp, http, telnet, bootp, and NFS. The LAN interface is standard.
SCPI
The analyzer can be controlled by sending Standard Commands for Programmable Instruments (SCPI) within a telnet session or via a socket connection and TCP/IP (the default socket port is 5025). The analyzer’s socket applications programming interface (API) is compatible with Berkeley sockets, Winsock and other standard socket APIs. Socket programming can be done in a variety of environments including C programs, Agilent VEE, SICL/LAN, or a Java™ applet. A stan­dard web browser and the analyzer’s built-in web page can be used to remotely enter SCPI com­mands via a Java applet.
FTP
Instrument state and data files can be transferred via ftp (file-transfer protocol). An internal, dynamic­data disk provides direct access to instrument states, screen dumps, trace data, and operating parameters.
HTTP
The instrument’s built-in web page can be accessed with any standard web browser using http (hyper­text transfer protocol) and the network analyzer’s IP address. The built-in web page can be used to control the network analyzer, view screen images, download documentation, and link to other sites for firmware upgrades and VXIplug&play drivers. Some word processor and spreadsheet programs, such as Microsoft® Word 97 and Excel 97, provide methods to directly import graphics and data via a LAN connection using http and the network ana­lyzer’s IP address.
SICL/LAN
The analyzer’s support for SICL (standard instru­ment control library) over the LAN provides con­trol of the network analyzer using a variety of com­puting platforms, I/O interfaces, and operating sys­tems. With SICL/LAN, the analyzer is controlled remotely over the LAN with the same methods used for a local analyzer connected directly to the computer via a GPIB interface. SICL/LAN protocol also allows the use of Agilent’s free VXI plug&play driver to communicate with the multiport test sys­tem over a LAN. SICL/LAN can be used with Windows® 95/98/NT, or HP-UX.
18
NFS
The analyzer’s built-in NFS (network file system) client provides access to remote files and directo­ries using the LAN. With NFS, remote files and directories (stored remotely on a computer) behave like local files and directories (stored locally within the analyzer). Test data taken by the net­work analyzer can be saved directly to a remote PC or UNIX® directory, eliminating the need for a remotely initiated ftp session. For Windows-based applications, third-party NFS-server software must be installed on the PC. NFS is fully supported in most versions of UNIX.
Bootp
Bootstrap protocol (bootp) allows a network ana­lyzer to automatically configure itself at power-on with the necessary information to operate on the network. After a bootp request is sent by the ana­lyzer, the host server downloads an IP and gateway address, and a subnet mask. In addition, the ana­lyzer can request an IBASIC file, which automati­cally executes after the transfer is complete. For Windows-based applications, third-party bootp­server software must be installed on the PC. Bootp is fully supported in most versions of UNIX.
Programming with IBASIC
As a standard feature, all Agilent 8712ET/ES and 8714ET/ES network analyzers come with the Instrument BASIC programming language (IBA­SIC). IBASIC facilitates automated measurements and control of other test equipment, improving productivity. For simpler applications, you can use IBASIC as a keystroke recorder to easily automate manual measurements. Or, you can use an optional, standard PC keyboard to write custom test appli­cations that include:
• Special softkey labels
• Tailored user prompts
• Graphical setup diagrams
• Barcode-reading capability
• Control of other test instruments via the GPIB, serial, or parallel interfaces
Measurement calibration
Measurement calibration is a process that improves measurement accuracy by using error­correction arrays to remove systematic measure­ment errors.
The Test Set Cal and SelfCal features on your mul­tiport test system increase the accuracy of your measurements and significantly increase the test efficiency of your work stations by eliminating fre­quent and lengthy calibration procedures.
Test Set Cal
A Test Set Cal is a calibration that should be per­formed on a regular, but relatively infrequent basis (at least once a month is recommended). A Test Set Cal requires connection of mechanical calibra­tion standards to all of the ports you will be using for your measurements.
Test Set Cal for the Agilent 8712ET and Agilent 8714ET analyzers
The data collected by the analyzer during a Test Set Cal always includes both transmission (enhanced response) calibration data and reflec­tion (one-port) calibration data. When making measurements after calibration, the analyzer auto­matically recalls and uses the correct set of cali­bration data for the type of measurement chosen.
Test Set Cal for the Agilent 8712ES and 8714ES analyzers
The S-parameter network analyzers perform either two-port calibration or enhanced response/one­port calibration. Choosing a two-port calibration for the Test Set Cal removes the most systematic errors, giving you the greatest measurement accu­racy. Choosing the enhanced response/one-port calibration allows faster measurement speeds, but is not as accurate as full two-port calibration.
SelfCal
A SelfCal is an internal system calibration that automatically executes in just a few seconds on a regular, frequent basis (once per hour is recom­mended). A SelfCal does not require that you remove your DUT or that you make any connec­tions of external calibration standards. The SelfCal uses the results of the most recent Test Set Cal, along with current measurements of internal, elec­tronically switched, open, short, load and through standards. SelfCal removes the drift of the network analyzer and multiport test set due to environmen­tal variations.
System features (continued)
19
Other calibrations
Besides using a Test Set Cal, individual instrument states and their corresponding calibrations can be saved and recalled for use with specific measure­ment paths. For example, to improve measurement throughput, one signal path could be tested using a response calibration, while all other paths are tested with a Test Set Cal using two-port calibration. Note: the SelfCal feature is only supported with Test Set Cals.
A variety of calibration types are available and described below:
ES models only
Two-port calibration
Compensates for frequency response, source and load match, and directivity errors while making S-parameter measurements of transmis­sion (S21, S12) and reflection (S11, S22). Compensates for transmission crosstalk when the Isolation on OFF softkey is toggled to ON. Requires short, open, load, and through stan­dards.
ET and ES models: transmission measurements
Normalization
Provides simultaneous magnitude and phase correction of transmission frequency-response errors. Requires a through connection. Used for both narrowband and broadband detection. Does not support calibration interpolation.
Response
Simultaneous magnitude and phase correction of frequency response errors for transmission measurements. Requires a through standard.
Response and isolation
Compensates for frequency response and crosstalk errors. Requires a load termination on both test ports and a through standard.
Enhanced response
Compensates for frequency response and source match errors. Requires short, open, load, and through standards.
ET and ES models: reflection measurements
One-port calibration
Compensates for frequency response, directivity, and source-match errors. Requires short, open, and load standards.
Calibration kits
Data for several standard calibration kits are stored in the instrument for use by the calibration routines. They include:
• 3.5 mm (85033D)
• Type-N (85032B/E)
• 7 mm (85031B)
• 7-16 (85038A)
In addition, you can also describe the standards for a user-defined calibration kit (for example, open-circuit capacitance coefficients, offset-short length, or through-standard loss).
For more information about calibration kits avail­able from Agilent Technologies, consult the Agilent 87050E Configuration Guide, Agilent literature number 5968-4765E.
Key network analyzer options
Step attenuator (Option 1E1)
Adds a built-in 60 dB step attenuator to transmis­sion/reflection (ET) models to extend the output power range to –60 dBm. The attenuator is stan­dard in S-parameter (ES) models.
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc­tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equip­ment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifica­tions and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabili­ties, at no extra cost upon request. Many self-help tools are available.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out­of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional services. Experienced Agilent engineers and tech­nicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
By internet, phone, or fax, get assistance with all your test & measurement needs
Online assistance:
www.agilent.com/find/assist
Phone or Fax
United States: (tel) 1 800 452 4844
Canada: (tel) 1 877 894 4414 (fax) (905) 282-6495
China: (tel) 800-810-0189 (fax) 1-0800-650-0121
Europe: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea: (tel) (82-2) 2004-5004 (fax) (82-2) 2004-5115
Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599
Taiwan: (tel) 080-004-7866 (fax) (886-2) 2545-6723
Other Asia Pacific Countries: (tel) (65) 375-8100 (fax) (65) 836-0252 Email: tm_asia@agilent.com
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2001 Printed in USA November 14, 2001 5968-4764E
Java is a U.S. trademark of Sun Microsystems, Inc.
Microsoft, MS-DOS, Windows, Windows 95, Windows 98, Windows
NT, Word 97, and Excel 97 are U.S. registered trademarks of Microsoft
Corporation.
UNIX is a registered trademark of the Open Group.
Loading...