Agilent 8702E Installation Guide

Agilent 8702E Lightwave Component Analyzer Installation Guide
© Copyright Agilent Technologies 2001 All Rights Reserved. Repro­duction, adaptation, or trans­lation without prior written permission is prohibited, except as allowed under copy­right laws.
Agilent Part No. 08702-91031 Printed in USA March 2001
Agilent Technologies Lightwave Division 3910 Brickway Boulevard­Santa Rosa, CA 95403, USA
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Agilent Technologies makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Agilent Technologies shall not be lia­ble for errors contained herein or for incidental or conse­quential damages in connec­tion with the furnishing, performance, or use of this material.
Restricted Rights Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and sub­paragraphs (c) (1) and (c) (2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Warranty.
This Agilent Technologies instrument product is war­ranted against defects in material and workmanship for a period of one year from date of shipment. During the war­ranty period, Agilent Technol­ogies will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility desig­nated by Agilent Technolo­gies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay ship­ping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies war­rants that its software and firmware designated by Agi­lent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Tech­nologies does not warrant that the operation of the instru­ment, or software, or firmware will be uninterrupted or error­free.
Limitation of Warranty.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfac­ing, unauthorized modifica­tion or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specifically dis­claims the implied warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies.
The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technolo­gies shall not be liable for any direct, indirect, special, inci­dental, or consequential dam­ages, whether based on contract, tort, or any other legal theory.
Safety Symbols.
CAUTION
The
caution
sign denotes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WAR NI NG
The
warning
sign denotes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
The instruction man­ual symbol. The prod­uct is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature of the line module input power.
The ON symbols are
|
used to mark the posi­tions of the instrument power line switch.
The OFF symbols
are used to mark the positions of the instru­ment power line switch.
The CE mark is a reg­istered trademark of the European Commu­nity.
The CSA mark is a reg­istered trademark of the Canadian Stan­dards Association.
The C-Tick mark is a registered trademark of the Australian Spec­trum Management Agency.
This text denotes the
ISM1-A
instrument is an Industrial Scientific and Medical Group 1 Class A product.
Typographical Conven­tions.
The following conventions are used in this book:
Key type
for keys or text located on the keyboard or instrument.
Softkey type
for key names that are displayed on the instru­ment’s screen.
Display type
for words or characters displayed on the computer’s screen or instru­ment’s display.
User type
for words or charac-
ters that you type or enter.
Emphasis
type for words or characters that emphasize some point or that are used as place holders for text that you type.
ii

Installation at a Glance

Installation at a Glance
The procedures in this book provide step-by-step instructions for installing the Agilent 8702E.
Be sure to verify the Agilent 8702E’s performance
Chapter 2, “Automated Verification”, Chapter 3, “Manual Verification”, and Chapter 4, “Performance Tests” provide verification procedures and perfor-
mance tests. Verification procedures are intended to provide a high level of confidence that the instrument is operating properly. Two versions of verifica­tion procedures are provided: automated and manual. The performance tests, along with the verification tests, provide the same quality of performance test­ing that is done at the factory. Chapter 5, “Automated Verification – Option
011” and Chapter 6, “Performance Tests – Option 011” provide verification
procedures and performance tests for the Agilent 8702E Option 011.
Agilent Technologies recommends that you verify your analyzer measurement system every six months. Agilent Technologies also suggests that you get your verification kit recertified annually. Refer to
tion Kit Operating and Service Manual
Agilent 85029B 7 mm Verifica-
for more information.
These system verification procedures do (75 ohm analyzers).
not
apply to analyzers with Option 075

General Safety Considerations

This product has been designed and tested in accordance with IEC Publica­tion 61010-1, Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
iii
General Safety Considerations
WARNING
WARNING
WARNING
WARNING
WARNING
If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means for protection are intact) only.
To prevent electrical shock, disconnect the Agilent 8702E from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
This is a Safety Class 1 product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
For continued protection against fire hazard, replace line fuse only with same type and ratings, (type T 0.315A/250V for 100/120V operation and 0.16A/250V for 220/240V operation). The use of other fuses or materials is prohibited. Verify that the value of the line­voltage fuse is correct.
• For 100/120V operation, use an IEC 127 5×20 mm, 0.315 A, 250 V, Agilent part number 2110-0449.
CAUTION
CAUTION
• For 220/240V operation, use an IEC 127 5×20 mm, 0.16 A, 250 V, Agilent Technologies part number 2110-0448.
Before switching on this instrument, make sure that the line voltage selector switch is set to the line voltage of the power supply and the correct fuse is installed. Assure the supply voltage is in the specified range.
This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
iv

Certification

CAUTION
CAUTION
CAUTION
CAUTION
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
connect ac power until you have verified the line voltage is correct.
Do not
Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range.
Certification
Agilent Technologies certifies that this product met its published specifica­tions at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other Interna­tional Standards Organization members.

Assistance

Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Service Office.
v
Assistance
vi

Contents

Installation at a Glance iii General Safety Considerations iii Certification v Assistance v

1 Installing the Agilent 8702E

Step 1. Inspect the Shipment 1-3 Step 2. Set up Static-Safe Workstation 1-4 Step 3. Option 1D5, Connect the Frequency Reference 1-6 Step 4. Check the Fuse and Voltage Selection 1-7 Step 5. Connect the Line-Power Cable 1-9 Step 6. Connect a Keyboard 1-11 Step 7. Turn on the Agilent 8702E 1-12 Step 8. Connect a Printer or Plotter 1-13 Step 9. If You Connect a Printer 1-14 Step 10. If You Connect a Plotter 1-15 Step 11. Set the Clock 1-16 Step 12. Check the Operation 1-17 Step 13. Check the Operation (Option 011) 1-22 Step 14. Copy the EEPROM Disk 1-27

2 Automated Verification

3Manual Verification

4 Performance Tests

5 Automated Verification – Option 011

Agilent 8702E Option 011 and Agilent 85046A/47A System Verification 5-4 Agilent 8702E Option 011 and Agilent 85044A System Verification 5-13

6 Performance Tests – Option 011

7 If You Encounter a Problem

If the display does not light 7-2 If the fan does not run 7-3 If data entry keys don’t respond 7-4
Contents-1
Contents
If there is no RF signal 7-4 Returning the Instrument for Service 7-6 Agilent Technologies Service Offices 7-9
Contents-2
1
Step 1. Inspect the Shipment 1-3 Step 2. Set up Static-Safe Workstation 1-4 Step 3. Option 1D5, Connect the Frequency Reference 1-6 Step 4. Check the Fuse and Voltage Selection 1-7 Step 5. Connect the Line-Power Cable 1-9 Step 6. Connect a Keyboard 1-11 Step 7. Turn on the Agilent 8702E 1-12 Step 8. Connect a Printer or Plotter 1-13 Step 9. If You Connect a Printer 1-14 Step 10. If You Connect a Plotter 1-15 Step 11. Set the Clock 1-16 Step 12. Check the Operation 1-17 Step 13. Check the Operation (Option 011) 1-22 Step 14. Copy the EEPROM Disk 1-27

Installing the Agilent 8702E

Installing the Agilent 8702E
Installing the Agilent 8702E
Installing the Agilent 8702E
The instructions in this chapter show you how to install your Agilent 8702E. You should be able to finish these procedures in about ten to twenty minutes.
Refer to “Specifications and Regulatory Information”, in the
Reference
ture.
If you should ever need to clean the cabinet, use a damp cloth only.
manual, for information on operating conditions such as tempera-
Agilent 8702E
WARNING
CAUTION
CAUTION
CAUTION
Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
This product has autoranging line voltage input. Be sure the supply voltage is within the specified range.
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
1-2
Installing the Agilent 8702E

Step 1. Inspect the Shipment

Step 1. Inspect the Shipment
1
Verify that all components ordered have arrived by comparing the shipping forms to the original purchase order. Inspect all shipping containers.
If your shipment is damaged or incomplete, save the packing materials and no­tify both the shipping carrier and the nearest Agilent Technologies service of­fice. Agilent Technologies will arrange for repair or replacement of damaged or incomplete shipments without waiting for a settlement from the transportation company. Notify the Agilent Technologies customer engineer of any problems.
WARNING
The Agilent 8702E weighs approximately 75 pounds (34 kilograms). Use correct lifting techniques.
The PORT 1 connector moves.
This is accessories can be more easily connected to the instrument.
2
Make sure that the serial number and options listed on the instrument’s rear­panel label match the serial number and options listed on the shipping document.
Figure 1-1. Serial Number Label
a defect. This connector is designed to move so that test sets and other
not
1-3
Installing the Agilent 8702E

Step 2. Set up Static-Safe Workstation

Step 2. Set up Static-Safe Workstation
Electrostatic discharge (ESD) can damage or destroy the input circuits of the Agilent 8702E. ESD can also damage or destroy electronic components that you are measuring. All work should be performed at a static-safe work station. The following figure shows an example of a static-safe work station (without the instrument) using two types of ESD protection:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
1-4
Installing the Agilent 8702E
Step 2. Set up Static-Safe Workstation
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone.
To ensure user safety, the static-safe accessories must provide at least 1 MΩ of isolation from ground. Refer to Ta ble 1- 1 for information on ordering static­safe accessories.
WARNING
These techniques for a static-safe work station should not be used when working on circuitry with a voltage potential greater than 500 volts.
Reducing ESD Damage
The following suggestions may help reduce ESD damage that occurs during testing and servicing operations.
• Personnel should be grounded with a resistor-isolated wrist strap before re­moving any assembly from the unit.
• Be sure all instruments are properly earth-grounded to prevent a buildup of static charge.
Table 1-1. Static-Safe Accessories
Agilent Part Number
9300-0797
9300-0980 Wrist-strap cord 1.5 m (5 ft)
9300-1383 Wrist-strap, color black, stainless steel, without cord, has four adjustable
Description
3M static control mat 0.6 m wire. (The wrist-strap and wrist-strap cord are not included. They must be ordered separately.)
links and a 7 mm post-type connection.
× 1.2 m (2 ft× 4 ft) and 4.6 cm (15 ft) ground
9300-1169 ESD heel-strap (reusable 6 to 12 months).
1-5
Installing the Agilent 8702E

Step 3. Option 1D5, Connect the Frequency Reference

Step 3. Option 1D5, Connect the Frequency Reference
If your instrument has the optional high stability frequency reference installed, connect the jumper cable on the Agilent 8702E rear panel as shown in Figure 1-2.
Figure 1-2. Jumper Cable Connection
1-6
Installing the Agilent 8702E

Step 4. Check the Fuse and Voltage Selection

Step 4. Check the Fuse and Voltage Selection
1
Locate the line-input connector on the instrument’s rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to open the pull-out fuse drawer.
WARNING
The power cord is connected to internal capacitors that may remain live for 10 seconds after you disconnect the cord from the power supply.
Figure 1-3. Line Fuse Removal and Replacement
4
Verify that the value of the line-voltage fuse in the pull-out drawer is correct. Th e recom mende d fu se is a n IEC 12 7 5×20 mm, 3A, 250 V, Agilent part number 2110-0780.
Notice that an extra fuse is provided in a drawer located on the fuse holder.
1-7
Installing the Agilent 8702E
Step 4. Check the Fuse and Voltage Selection
WARNING
For continued protection against fire hazard, replace line fuse only with same type and ratings, (type T 3A/250V for 100/240V operation). The use of other fuses or materials is prohibited.
5
Set the rear panel line-voltage selector to the position that corresponds to the AC power source. Refer to Figure 1-4.
Figure 1-4. Line Voltage Selector
Table 1-2. Line Power Requirements
Power: 115 VAC: 50 WATTS MAX.
230 VAC: 50 WATTS MAX.
Voltage nominal: 115 VAC / 230 VAC
range 115 VAC: 90 range 230 VAC: 198
Frequency nominals: 50 Hz / 60 Hz
range: 47
1-8
132 V
254 V
63 Hz
Installing the Agilent 8702E

Step 5. Connect the Line-Power Cable

Step 5. Connect the Line-Power Cable
CAUTION
CAUTION
Always use the three-prong AC power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
Do
connect ac power until you have verified the line voltage is correct as
not
described in the following paragraphs. Damage to the equipment could result.
1
Connect the line-power cord to the instrument’s rear-panel connector. Refer to
Figure 1-5.
2
Connect the other end of the line-power cord to the power receptacle. Refer to
Figure 1-5.
Various power cables are available to connect the Agilent 8702E to ac power outlets unique to specific geographic areas. The cable appropriate for the area to which the Agilent 8702E is originally shipped is included with the unit. You can order additional ac power cables for use in different geographic areas. Refer to the cables.
Agilent 8702E Reference
manual for a list of available power
1-9
Installing the Agilent 8702E
Step 5. Connect the Line-Power Cable
Figure 1-5. Power Cord Connection
1-10
Installing the Agilent 8702E

Step 6. Connect a Keyboard

Step 6. Connect a Keyboard
• If you plan to use a keyboard with your Agilent 8702E, connect it now to the instrument’s rear-panel keyboard connector.
Figure 1-6. Rear Panel Peripheral Connections
1-11
Installing the Agilent 8702E

Step 7. Turn on the Agilent 8702E

Step 7. Turn on the Agilent 8702E
1
Press the front-panel
The front-panel ply after the EMC filters and before other parts of the instrument.
2
If the Agilent 8702E fails to turn on properly, consider the following possibilities:
• Is the line fuse good?
• Does the line socket have power?
• Is it plugged into the proper ac power source?
If the instrument still fails, return it to Agilent Technologies for repair. Refer to
“Returning the Instrument for Service” on page 7-6.
LINE
key. Refer to Figure 1-7.
LINE
switch disconnects the mains circuits from the mains sup-
Figure 1-7. Turning on the Instrument
1-12
Installing the Agilent 8702E

Step 8. Connect a Printer or Plotter

Step 8. Connect a Printer or Plotter
1
Connect a printer or plotter to the corresponding interface port. Refer to
Figure 1-6 on page 1-11 for the appropriate rear panel connection.
Printer Interface Recommended Cable
Parallel Agilent 92284A
Serial Agilent 24542G
GPIB Agilent 10833A/B/D
2
If you are using the parallel interface, press
LOCAL
and then
PARALLEL
until
appears.
GPIO
The
selection dedicates the parallel port for general purpose I/O. The Ag­ilent 8702E controls the data input or output through the sequencing capability of the instrument.
3
If you are using a GPIB printer or plotter, press
LOCAL
SYSTEM CONTROLLER
,
set up the Agilent 8702E as the controller.
COPY
to
1-13
Installing the Agilent 8702E

Step 9. If You Connect a Printer

Step 9. If You Connect a Printer
1
2
3
LOCAL
Press
SET ADDRESSES, PRINTER PORT
,
.
Press the key that corresponds to the printer interface: port), or
If you selected
SERIAL
(serial port).
GPIB
, the GPIB address is active so you can then set the address
GPIB, PARAL LEL
of your printer. The default GPIB printer address is 1.
4
If you select
SERIAL
, adjust the Agilent 8702E’s baud rate and handshaking
protocol:
a
PRINTER BAUD RATE
Press
, and use the up and down arrow front-panel keys to
select the baud rate.
b
Set the transmission control,
XON/XOFF
or
DTR/DSR
(equal to the transmission control set on the peripher-
XMIT CNTRL
(handshaking protocol) to either
al).
XON/XOFF
selects software handshaking.
DTR/DSR
selects hardware
handshaking. Consult the printer’s manual for the proper settings.
5
Press
ThinkJet
DeskJet
LaserJet
PaintJet
PRNTR TYPE
(QuietJet)
(except for HP
until the correct printer choice appears:
1
DeskJet 540)
(parallel
Epson-P2
DJ540
1. HP is a registered trademark of Hewlett-Packard Company.
1-14
(printers that conform to the ESC.P2 printer control language)
(converts 100 dpi raster information to 300 dpi raster format)

Step 10. If You Connect a Plotter

Installing the Agilent 8702E
Step 10. If You Connect a Plotter
1
2
3
4
5
LOCAL
Press
Press the key that corresponds to the plotter interface: port), or
If you selected of your plotter. The default GPIB plotter address is 5.
If you select protocol:
a
Press select the baud rate.
b
Set the transmission control,
XON/XOFF
al).
XON/XOFF
handshaking. Consult the printer’s manual for the proper settings.
Press
•Choose
copy.
•Choose
to make your hardcopy.
SET ADDRESSES, PLOTTER PORT
,
SERIAL
(serial port).
GPIB
, the GPIB address is active so you can then set the address
SERIAL
, adjust the Agilent 8702E’s baud rate and handshaking
PLOTTER BAUD RATE
DTR/DSR
or
selects software handshaking.
PLTR TYPE
until the correct plotter selection appears:
PLTR TYPE [PLOTTER]
PLTR TYPE [HPGL PRT]
, and use the up and down arrow front-panel keys to
(equal to the transmission control set on the peripher-
if you will be using a plotter to make your hard-
if you will be using an HPGL compatible printer
.
XMIT CNTRL
GPIB, PARAL LEL
(handshaking protocol) to either
DTR/DSR
selects hardware
(parallel
1-15
Installing the Agilent 8702E

Step 11. Set the Clock

Step 11. Set the Clock
1
2
3
4
SYSTEM
Press so the Agilent 8702E places the time and date on your hardcopies and disk directories.
Press the appropriate softkey to set the time and date.
ROUND SECONDS
Press
TIME STAMP
Press
RETURN
.
SET CLOCK
,
to begin setting and activating the time stamp feature
when the time is exactly as you have set it.
so that ON is displayed on the softkey label. Then, press
Figure 1-8. Setting the Clock
1-16
Installing the Agilent 8702E

Step 12. Check the Operation

Step 12. Check the Operation
If your instrument is an Agilent 8702E Option 011, skip this step and continue with “Step 13. Check the Operation (Option 011)” on page 1-22.
1
Turn off the
2
Locate the serial number and configuration options shown on the display. Compare them to the shipment documents.
3
Press
Figure 1-9. Checking the Operation
LINE
power switch. Then, turn on the
PRESET
, and observe items shown in Figure 1-9.
LINE
power switch.
4
Connect the equipment as shown in Figure 1-10.
1-17
Installing the Agilent 8702E
Step 12. Check the Operation
Figure 1-10. Operation Check Connections
5
6
7
8
PRESET, SYSTEM
Press
SERVICE MENU, TESTS, EXTERNAL TESTS, EXECUTE TEST
,
Follow the prompts shown on the display, and then press
Press the up-arrow key, the display. Then press
EXECUTE TEST
CONTINUE
.
, and then follow the prompts shown on
Connect the equipment as shown in Figure 1-11 and press
CONTINUE
PRESET
.
.
The test port return cable should have low-loss characteristics to avoid a deg­radation in frequency response at higher frequencies.
.
1-18
Figure 1-11. Transmission Mode Connections
Installing the Agilent 8702E
Step 12. Check the Operation
Press
CHAN2
to check the forward transmission mode for channel 2. Look at the
9
measurement trace displayed on the Agilent 8702E. It should be similar to the trace shown in Figure 1-12.
The Agilent 8702E display shown in Figure 1-12 and Figure 1-13 are Agilent 8702E Option 006 displays (300 kHz to 6 GHz span).
Figure 1-12. Forward Transmission Display
Press
MEAS
and then
Trans: E/E S12 REV
to check the reverse transmission mode
10
for channel 2. The measurement trace should be similar to the trace shown in
1-19
Installing the Agilent 8702E
Step 12. Check the Operation
Figure 1-13.
Figure 1-13. Reverse Transmission Display
11
Connect the equipment as shown in Figure 1-14 and press
PRESET
.
Figure 1-14. Transmission Load Match Connections
12
Look at the measurement trace displayed on the Agilent 8702E. It should be similar to the trace displayed in Figure 1-15.
1-20
Figure 1-15. Forward Transmission Mode
Installing the Agilent 8702E
Step 12. Check the Operation
Press
MEAS,
Refl: E S22 REV
to check the reverse reflection mode for channel 1.
13
The measurement trace should be similar to Figure 1-16.
Figure 1-16. Reverse Reflection Mode
14
Continue with “Step 14. Copy the EEPROM Disk” on page 1-27.
1-21
Installing the Agilent 8702E

Step 13. Check the Operation (Option 011)

Step 13. Check the Operation (Option 011)
An Agilent 85047A S-parameter test set must be used when making measure­ments.
1
Turn off the
2
Locate the serial number and configuration options shown on the display. Compare them to the shipment documents.
3
Press
Figure 1-17.
LINE
power switch. Then, turn on the
PRESET
, and observe that the analyzer is operating properly as shown in
LINE
power switch.
Figure 1-17. Checking the Operation (Option 011)
4
Connect the equipment as shown in Figure 1-18.
1-22
Figure 1-18. Operation Check Connections
Installing the Agilent 8702E
Step 13. Check the Operation (Option 011)
5
6
7
8
PRESET, SYSTEM
Press
SERVICE MENU, TESTS, EXTERNAL TESTS, EXECUTE TEST
,
Follow the prompts shown on the display, and then press
Press the up-arrow key, the display. Then press
EXECUTE TEST
CONTINUE
.
, and then follow the prompts shown on
Connect the equipment as shown in Figure 1-19 and press
CONTINUE
PRESET
.
.
The test port return cable should have low-loss characteristics to avoid a deg­radation in frequency response at higher frequencies.
.
1-23
Installing the Agilent 8702E
Step 13. Check the Operation (Option 011)
Figure 1-19. Transmission Mode Connections
Press
CH2
to check the forward transmission mode for channel 2. Look at the
9
measurement trace displayed on the analyzer. It should be similar to the trace shown in Figure 1-20.
The analyzer display shown in Figure 1-20 and Figure 1-21 are Agilent 8702E Option 011, Option 006 displays (30 kHz to 6 GHz span).
Figure 1-20. Forward Transmission Display
1-24
Installing the Agilent 8702E
Step 13. Check the Operation (Option 011)
Press
MEAS
and then
Trans: E/E S12 REV
to check the reverse transmission mode
10
for channel 2. The measurement trace should be similar to the trace shown in
Figure 1-21.
Figure 1-21. Reverse Transmission Display
11
Connect the equipment as shown in Figure 1-22 and press
PRESET
.
Figure 1-22. Transmission Load Match Connections
12
Look at the measurement trace displayed on the analyzer. It should be similar
1-25
Installing the Agilent 8702E
Step 13. Check the Operation (Option 011)
to the trace displayed in Figure 1-23.
Figure 1-23. Forward Transmission Mode
Press
MEAS
Refl: E S22 REV
,
to check the reverse reflection mode for channel 1.
13
The measurement trace should be similar to Figure 1-24.
Figure 1-24. Reverse Reflection Mode
1-26
Installing the Agilent 8702E

Step 14. Copy the EEPROM Disk

Step 14. Copy the EEPROM Disk
The Agilent 8702E is shipped from the factory with an EEPROM backup disk which is unique to each instrument. The backup disk contains a copy of cor­rection constants which were stored in the instrument’s internal memory at the factory. If your Agilent 8702E should ever be damaged, these correction constants may need to be reinstalled from the disk. Use the following proce­dure to create a backup copy of the disk.
1
Insert a 3.5-inch disk into the Agilent 8702E’s disk drive.
2
3
4
5
6
PRESET
Press
If the disk is not formatted, press
• To format a LIF disk, select
is LIF. The Agilent 8702E does not support LIF-HFS format.)
• To format a DOS disk, select
Press
Press
SAVE RCL MENU, SELECT DISK, INTERNAL DISK, RETURN, SAVE STATE
correction-constants data onto floppy disk.
A default file “FILE0” is created. The file name appears in the upper left-hand corner of the display. The file type “ISTATE(E)” describes the file as an instru­ment-state with EEPROM backup.
Press softkey to rename the file “FILE0” to “N12345” where 12345 represents the last 5 digits of the instrument’s serial number. (The first character in the file name must be a letter.) When finished, press
Label the disk with the serial number of the instrument and the words “EEPROM Backup Disk”.
Whenever the Agilent 8702E is returned to Agilent Technologies for servicing and/or calibration, the EEPROM backup disk should be returned with the Agi­lent 8702E. This will significantly reduce the instrument repair time. The EE­PROM backup disk procedure is now complete.
so that the preset state is stored.
SAVE/RECALL
FORMAT:LIF
FORMAT:DOS
FORMAT INT DISK,
SYSTEM
FILE UTILITIES, RENAME FILE, ERASE TITLE
SERVICE MENU, SERVICE MODES, MORE, STORE EEPR ON
,
and answer
YES
at the query.
DONE
FILE UTILITIES, FORMAT DISK
,
(The supplied EEPROM backup disk
.
SAVE/RECALL,
,
to store the
. Use the RPG and the
.
SELECT LETTER
.
1-27
Installing the Agilent 8702E
Step 14. Copy the EEPROM Disk
1-28
2
Step 1. Initialization 2-3 Step 2. Measurement Calibration 2-4 Step 3. Device Verification 2-6 In Case of Difficulty 2-9

Automated Verification

Automated Verification
Automated Performance Verification
Automated Performance Verification
The automated performance verification procedure in this chapter provide a high level of confidence that the Agilent 8702E is working properly. Although this procedure is automated, it does need to locate the form the manual version located in Chapter 3, “Manual Verification”. If you are using an Agilent 8702E Option 011, refer to Chapter 5, “Automated Verifica-
tion – Option 011”.
Allow the Agilent 8702E to warm up for one hour before starting this proce­dure.
In order to run this procedure, the following equipment is required:
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
Verification Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . Agilent 85029B Option 001
Test Port Extension Cable Set, 7 mm . . . . . . . . . . . . . . . . . . . . .Agilent 11857D
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP ThinkJet/DeskJet/LaserJet
Verification Data Disk
require a computer. However, you’ll
not
. If you cannot locate the disk, per-
Agilent Technologies recommends that you verify your analyzer measurement system every six months. Agilent Technologies also suggests that you get your verification kit recertified annually. Refer to the
7 mm Verification Kit Operating and Service Manual
tion.
NOTE
The system verification procedures do
2-2
apply to analyzers with Option 075 (75 Ω).
not
Agilent 85029B Option 001
for more informa-
Automated Verification
Step 1. Initialization
How to select the system verification procedure
Check to see how the verification kit floppy disk is labeled:
• If the disk is labeled mated Mode System Verification in this chapter.
• If the disk is labeled Mode System Verification procedure, located in Chapter 3, “Manual Verifica-
tion”.
NOTE
If your verification disk is older than your Agilent 8702E, you may send your Agilent 85029B Option 001 7 mm verification kit to the nearest service center for recerti­fication, which includes a data disk that you can use with the Agilent 8702E.
Verification Data Disk
Verification Data Disc
, proceed with the Agilent 8702E Auto-
, proceed with the Agilent 8702E Manual
Step 1. Initialization
1
Connect the equipment as shown in Figure 2-1. Let the Agilent 8702E warm up for one hour.
Figure 2-1. System Verification Test Setup
2
While the equipment is warming up, review the connector care information in
2-3
Automated Verification
Step 2. Measurement Calibration
the
Agilent 8702E User’s Guide
3
Insert the verification kit disk into the Agilent 8702E’s disk drive.
4
5
6
7
8
PRESET, SAVE/RECALL
Press
If you want a printout of the verification data for all the devices, press
SERVICE MENU, TEST OPTIONS, RECORD ON
If you switch on the record function, you cation procedure.
Position the paper in the printer so that printing starts at the top of the page.
SYSTEM
Press
When the Agilent 8702E displays, procedure is complete.
Do
press
not
strument state that you loaded during the initialization procedure.
SERVICE MENU
,
PRESET
,
or recall another instrument state. You must use the in-
.
SELECT DISK, INTERNAL DISK
.
cannot
TESTS
,
SYS VER TESTS, EXECUTE TEST
,
Sys Ver Init DONE
switch it off during the verifi-
.
SYSTEM
.
, the initialization
Step 2. Measurement Calibration
1
CAL
Press
MENU, FULL 2-PORT
CAL KIT & STDS
,
SELECT CAL KIT
,
.
CAL KIT:7mm
,
RETURN, RETURN, CALIBRATE
,
,
2
3
4
ISOLATION
Press
REFLECTION
Press
Connect the “open” end of the open/short combination (supplied in the calibration kit) to reference test port 1, as shown in Figure 2-2.
2-4
OMIT ISOLATION
,
.
ISOLATION DONE
,
.
Step 2. Measurement Calibration
Figure 2-2. Connections for Measurement Calibration Standards
Automated Verification
5
6
FORWARD:OPEN
Press
.
When the Agilent 8702E finishes measuring the standard, connect the “short” end of the open/short combination to reference test port 1.
7
8
FORWARD:SHORT
Press
.
When the Agilent 8702E finishes measuring the standard, connect the 50 ohm termination (supplied in the calibration kit) to reference test port 1.
9
10
FORWARD:LOAD
Press
.
When the Agilent 8702E finishes measuring the standard, connect the “open” end of the open/short combination to reference test port 2.
11
12
REVERSE:OPEN
Press
.
When the Agilent 8702E finishes measuring the standard, connect the “short” end of the open/short combination to reference test port 2.
13
14
REVERSE:SHORT
Press
.
When the Agilent 8702E finishes measuring the standard, connect the 50 ohm termination to reference test port 2.
15
16
REVERSE:LOAD
Press
.
When the Agilent 8702E finishes measuring the standard, press
STANDARDS
2-5
Automated Verification
Step 3. Device Verification
DONE
. The Agilent 8702E briefly displays,
17
Connect the test port cables as shown Figure 2-3.
COMPUTING CAL COEFFICIENTS
Figure 2-3. Transmission Calibration Setup
18
19
20
TRANSMISSION, FWD TRANS THRU
Press
.
When the Agilent 8702E finishes the measurement, press
When the Agilent 8702E finishes the measurement, press
FWD MATCH THRU
REV TRANS THRU
.
.
.
21
When the Agilent 8702E finishes the measurement, press
22
When the Agilent 8702E displays,
STANDARDS DONE, DONE 2-PORT CAL
SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE
,
23
STD(s)
SAVE/RECALL
Press
, press
PRESS ’DONE’ IF FINISHED WITH
.
REV MATCH THRU
the calibration into the Agilent 8702E internal memory.
24
When the Agilent 8702E finishes saving the instrument state, press
INTERNAL DISK
.
SELECT DISK
Step 3. Device Verification
Press
SYSTEM
SERVICE MENU, TESTS
,
At the prompt, connect the 20 dB attenuator (supplied in the verification kit)
1
2
as shown in Figure 2-4.
2-6
28, x1
,
EXECUTE TEST
,
.
.
to save
,
Automated Verification
Step 3. Device Verification
3
CONTINUE
Press
to run the test:
• If you switched the record function
, you have to press
off
CONTINUE
each S-parameter measurement.
• If you switched the record function on, the Agilent 8702E measures all S-
parameters (magnitude and phase) without pausing. Also, the Agilent 8702E only displays and prints the pass/fail information for the S-parameter measurements that are valid for system verification.
Figure 2-4. Connections for the 20 dB Verification Device
4
When the Agilent 8702E finishes all the measurements, connect the 50 dB attenuator (supplied in the verification kit), as shown in Figure 2-5.
after
Figure 2-5. Connections for the 50 dB Verification Device
5
Press ⇑, 29, x1,
EXECUTE TEST
CONTINUE
,
.
2-7
Automated Verification
Step 3. Device Verification
6
When all measurements are complete, replace the verification device with the verification mismatch, as shown in Figure 2-6. Be sure that you connect Port A of the verification mismatch to reference test port 1.
Figure 2-6. Mismatch Device Verification Setup
7
8
Press
RETURN, TESTS
30, x1,
,
EXECUTE TEST, CONTINUE
.
When the Agilent 8702E finishes all the measurements, connect the mismatch verification device, as shown in Figure 2-7. Notice that Port B is now connected to reference test port 1.
2-8
Figure 2-7. Mismatch Device Verification Setup
Automated Verification
In Case of Difficulty
9
10
Press
RETURN, TESTS
,
EXECUTE TEST, CONTINUE
31, x1,
.
You have completed the system verification procedure when the Agilent 8702E displays,
Ver Def 4 DONE
.
In Case of Difficulty
1
Inspect all connections. Do test ports. Doing so will invalidate the calibration that you have done earlier.
2
3
PRESET, SAVE/RECALL
Press
Using the front panel knob, highlight the title of the full 2-port calibration that you have done earlier, then press
4
Repeat “Step 3. Device Verification” on page 2-6.
5
If the Agilent 8702E still fails the test, check the measurement calibration as follows:
a
b
c
PRESET
Press
.
Recall the calibration by pressing
RY
RETURN
,
.
Use the front panel knob to highlight the calibration you want to recall and
RECALL STATE
press
.
disconnect the cables from the Agilent 8702E
not
SELECT DISK
,
INTERNAL MEMORY, RETURN
,
RECALL STATE
SAVE/RECALL
.
SELECT DISK
,
INTERNAL MEMO-
,
.
2-9
Automated Verification
In Case of Difficulty
d
Connect the short to reference test port 1.
e
f
g
h
i
j
k
MEAS
Press
Press
Refl: E S11 FWD
,
SCALE REF
SCALE/DIV
,
Check that the trace response is 0.00 ±0.05 dB.
Disconnect the short and connect it reference test port 2.
MEAS
Press
Refl: E S22 REV
,
Check that the trace response is 0.00 ± 0.05 dB.
If any of the trace responses are out of the specified limits, repeat the “Mea-
MENU
,
, .05, x1.
.
TRIGGER MENU
,
CONTINUOUS
,
.
surement Calibration” and “Device Verification” procedures.
2-10
3
Step 1. Initialization 3-2 Step 2. Measurement Calibration 3-4 Step 3. Device Verification 3-5 In Case of Difficulty 3-9

Manual Verification

Manual Verification
Manual Performance Verification
Manual Performance Verification
The manual performance verification procedure in this chapter provides a high level of confidence that the Agilent 8702E is working properly. If you can locate the version instead, which is documented in Chapter 2, “Automated Verification”.
Allow the Agilent 8702E to warm up for one hour before starting this proce­dure.
In order to run this procedure, the following equipment is required:
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
Verification Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85029B
Test Port Extension Cable Set, 7 mm . . . . . . . . . . . . . . . . . . . . .Agilent 11857D
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP ThinkJet/DeskJet/LaserJet
Agilent 8702E Verification Data Disk
, perform the automated
Step 1. Initialization
1
Connect the equipment as shown in Figure 3-1. Let the Agilent 8702E warm up for one hour.
3-2
Manual Verification
Step 1. Initialization
Figure 3-1. System Verification Test Setup
2
While the equipment is warming up, review the connector care information in the
Agilent 8702E User’s Guide
.
3
Set up the system to print the verification data:
a
b
c
d
LOCAL
Press
Press
SYSTEM CONTROLLER
,
.
SET ADDRESSES, PRINTER PORT
.
Press the softkey that corresponds to your printer port.
If you have difficulty with the printer:
• If the interface on your printer is GPIB, verify that the printer address is set
to 1.
• If the interface on your printer is serial or parallel, be sure that you correctly
selected the printer port
er’s Guide
4
Insert the verification kit disk into the Agilent 8702E disk drive.
5
6
7
PRESET, SAVE/RECALL
Press
SYSTEM
Press
When the Agilent 8702E displays,
for more information on how to do these tasks).
SERVICE MENU, TESTS, SYS VER TESTS, EXECUTE TEST
,
printer type (refer to the
and
SELECT DISK, INTERNAL DISK
,
Sys Ver Init DONE
Agilent 8702E Us-
.
.
, the test is complete.
3-3
Manual Verification
Step 2. Measurement Calibration
CAUTION
press
PRESET
or recall another instrument state. You must use the
Do
not
instrument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1
2
3
4
5
6
7
8
9
SWEEP SETUP
Press dBm.
CAL
Press
MENU, FULL 2-PORT
Press
Press
Connect the “open” end of the open/short combination (supplied in the calibration kit) to reference test port 1, as shown in Figure 3-1.
Press
When the Agilent 8702E finishes measuring the standard, connect the “short” end of the open/short combination to reference test port 1.
Press
When the Agilent 8702E finishes measuring the standard, connect the 50 ohm termination (supplied in the calibration kit) to reference test port 1.
CAL KITS & STDS, SELECT CAL KIT, CAL KIT:7mm, RETURN, RETURN, CALIBRATE
,
ISOLATION, OMIT ISOLATION, ISOLATION DONE
REFLECTION
FORWARD:OPEN
FORWARD:SHORT
,
.
.
POWER
, –2, x1 to set the Agilent 8702E test port power to –2
.
.
.
10
11
12
13
14
15
16
17
REFLECTION, FORWARD:LOAD
Press
When the Agilent 8702E finishes measuring the standard, connect the “open” end of the open/short combination to reference test port 2, as shown in
Figure 3-1.
REVERSE:OPEN
Press
When the Agilent 8702E finishes measuring the standard, connect the “short” end of the open/short calibration standard to reference test port 2.
REVERSE:SHORT
Press
When the Agilent 8702E finishes measuring the standard, connect the 50 ohm termination to reference test port 2.
REVERSE:LOAD
Press
When the Agilent 8702E finishes measuring the standard, press
DONE
.
3-4
.
.
.
.
STANDARDS
18
Connect the test port cables together as shown in Figure 3-2.
Figure 3-2. Transmission Calibration Setup
Manual Verification
Step 3. Device Verification
19
20
21
22
23
24
25
TRANSMISSION, FWD TRANS THRU
Press
When the Agilent 8702E finishes measuring the standard, press
THRU
.
When the Agilent 8702E finishes measuring the standard, press
THRU
.
When the Agilent 8702E finishes measuring the standard, press
THRU
.
When the Agilent 8702E finishes measuring the standard, press
DONE, DONE 2-PORT CAL
SAVE/RECALL
Press the measurement calibration into the Agilent 8702E internal memory.
SELECT DISK, INTERNAL DISK
Press
.
SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE
,
.
FWD MATCH
REV TRANS
REV MATCH
STANDARDS
.
Step 3. Device Verification
1
2
SYSTEM
Press
Connect the 20 dB attenuator (supplied in the verification kit) as shown in
SERVICE MENU, TESTS
,
, 28, x1,
EXECUTE TEST
.
to save
3-5
Manual Verification
Step 3. Device Verification
Figure 3-3.
Figure 3-3. Connections for the 20 dB Verification Device
3
4
5
Press
Press
Press
CONTINUE
to run the test.
SWEEP SETUP
COPY
PRINT MONOCHROME
,
POWER
,
CAL
, –2, x1,
CORRECTION ON
,
.
to obtain the graphical representation of the
measurement.
Press
COPY
LIST, PRINT ALL MONOCHROME
,
if you prefer a tabular representation
6
of the measurement.
Press
SYSTEM
SERVICE MENU, TEST OPTIONS, CONTINUE TEST
,
Ignore the error message,
Press
SWEEP SETUP
POWER
,
.
CAUTION: CALIBRATION REQUIRED
CAL
, –2, x1,
CORRECTION ON
,
.
.
7
8
For those S-parameter measurements with upper and lower point limits shown on the Agilent 8702E display, you will notice that when you press the the measurement test status changes from
If the test status remains
FAIL
after you press
FAIL
to
PASS
CORRECTION ON
CORRECTION ON
.
, refer to “In
softkey,
Case of Difficulty” on page 3-9.
9
Repeat Step 5 through Step 8 until you have made all the S-parameter measurements for the verification device.
For each verification device, there are a total of eight measurements (magni­tude and phase for the four S-parameters).
3-6
Manual Verification
Step 3. Device Verification
10
When the Agilent 8702E displays, only if
the tested S-parameter measurements show a
all
Ver Dev 1 FAIL
, proceed to the next step
PASS
status after you activate error correction in Step 8. Otherwise, refer to “In Case of Difficulty” on
page 3-9.
11
Connect the 50 dB attenuator (supplied in the verification kit) as shown in
Figure 3-4.
Figure 3-4. Connections for the 50 dB Verification Device
12
Press
TESTS
29, x1,
,
EXECUTE TEST, CONTINUE
.
Ignore the error message.
Press
SWEEP SETUP
POWER
,
, –2, x1,
Repeat Step 5 through Step 8 until you have made all the S-parameter
13
14
CAL
CORRECTION ON
,
.
measurements for the verification device.
15
When the Agilent 8702E displays, only if
S-parameter measurements show a
all
Ver Dev 2 FAIL,
PASS
proceed to the next step
status under the same
condition mentioned in Step 10.
16
Connect the verification mismatch (supplied in the kit) as shown in Figure 3-5. Be sure port A of the verification mismatch is connected to reference test port 1.
3-7
Manual Verification
Step 3. Device Verification
Figure 3-5. Mismatch Device Verification Setup
17
Press
TESTS
, 30, x1,
EXECUTE TEST, CONTINUE
.
Ignore the error message.
Press
SWEEP SETUP
POWER
,
,
Repeat Step 5 through Step 8 until you have made all the S-parameter
18
19
–2, x1,
CAL
CORRECTION ON
,
.
measurements for the verification device.
20
When the Agilent 8702E displays, only if
S-parameter measurements show a
all
Ver Dev 3 FAIL
PASS
, proceed to the next step
status under the same
condition mentioned in Step 10.
21
Connect the verification mismatch as shown in Figure 3-6. Notice that port B of the verification mismatch is now connected to the reference test port 1.
3-8
Figure 3-6. Mismatch Device Verification Setup
Manual Verification
In Case of Difficulty
22
Press
TESTS
31, x1,
,
EXECUTE TEST, CONTINUE
.
Ignore the displayed error message.
Press
SWEEP SETUP
POWER
,
,
Repeat Step 5 through Step 8 until you have made all the S-parameter
23
24
–2, x1,
CAL
CORRECTION ON
,
.
measurements for the verification device.
25
The test is complete when the Agilent 8702E displays,
S-parameter measurements show a
all
PASS
status under the same condition
Ver Def 4 FAIL
mentioned in Step 10.
In Case of Difficulty
1
Inspect all connections. Do test ports. Doing so will invalidate the calibration done earlier.
2
PRESET, SAVE/RECALL
Press front panel knob, highlight the full 2-port cal done earlier and press
STATE
.
3
Repeat “Step 3. Device Verification” on page 3-5.
4
If the test still fails, do the following to check the calibration:
disconnect the cables from the Agilent 8702E
not
SELECT DISK, INTERNAL MEMORY, RETURN
,
, and
. Using the
RECALL
3-9
Manual Verification
In Case of Difficulty
Press
PRESET
SAVE/RECALL, SELECT DISK, INTERNAL MEMORY, RETURN
,
.
Use the front panel knob to highlight the calibration you want to recall and
RECALL STATE
press
Press
SWEEP SETUP
.
TRIGGER MENU, CONTINUOUS
,
to trigger the continuous
a
b
c
sweeps.
d
Connect the short to reference test port 1.
e
f
g
h
i
j
k
MEAS
Press
Press
SCALE REF
Refl: E S11 FWD
,
SCALE/DIV
,
Check to be sure the trace response is 0.00 ±0.05 dB.
Disconnect the short and connect the reference to test port 2.
MEAS
Press
Refl: E S22 REV
,
Check to be sure the trace response is 0.00 ±0.05 dB.
If any of the trace responses are out of the specified limits, repeat the pro-
SWEEP SETUP
,
.05, x1.
,
.
TRIGGER MENU, CONTINUOUS
,
cedures in this chapter.
.
3-10
4
1. Test Port Output Frequency Range and Accuracy 4-3
2. External Source Mode Frequency Range 4-5
3. Test Port Output Power Accuracy 4-6
4. Test Port Output Power Range and Linearity 4-8
5. Minimum R Channel Level 4-10
6. Test Port Input Noise Floor Level 4-15
7. Test Port Input Frequency Response 4-18
8. Test Port Crosstalk 4-29
9. Calibration Coefficients 4-34
10. System Trace Noise (Only for Analyzers without Option 006) 4-40
11. System Trace Noise (Only for Analyzers with Option 006) 4-42
12. Test Port Input Impedance 4-44
13. Test Port Receiver Magnitude Dynamic Accuracy 4-48
14. Test Port Receiver Phase Dynamic Accuracy 4-55
15. Test Port Receiver Magnitude Compression 4-56
16. Test Port Receiver Phase Compression 4-59
17. Test Port Output/Input Harmonics (Option 002 Analyzers without Option 006 Only) 4-62
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only) 4-66 Performance Test Records 4-70

Performance Tests

Performance Tests
Manual Performance Tests
Manual Performance Tests
This chapter contains procedures which test the electrical performance of the Agilent 8702E. If you are using an Agilent 8702E Option 011, perform the tests in Chapter 6, “Performance Tests – Option 011”. Before starting these proce­dures, allow the instrument to warm up for one hour.
Each procedure has a “Performance Test Record” that is located at the end of this chapter.
To obtain the same quality of performance testing that Agilent Technologies has administered at the factory, you must perform:
• the system verification procedures in Chapter 2, “Automated Verification” or
Chapter 3, “Manual Verification”.
of the performance test procedures in this chapter.
all
This quality of performance testing guarantees that the analyzer is performing within Certificate of Calibration for your analyzer if two conditions are met.
1
Your analyzer passes all the performed tests.
2
The equipment and standards that you used to perform the tests are traceable to a national standards institute.
NOTE
If you have a particular type of measurement application that does not use all of the ana­lyzer’s measurement capabilities, you may ask your nearest Agilent Technologies Service Office for a subset of specifications that you want verified. However, this may create a potential for making incorrect measurements, by using a different application than what was specified in the procedures.
4-2
of the published specifications. Agilent Technologies will issue a
all
Performance Tests
1. Test Port Output Frequency Range and Accuracy
WARNING
Required Equipment
Procedure
Any servicing, adjustment, maintenance, or repair of this product must be performed only by qualified personnel. Repair information provided in the provided for qualified service personnel and are intended to be used with the
Agilent 8702E Service Guide
In Case of Difficulty
sections of this chapter are
.
1. Test Port Output Frequency Range and Accuracy
Perform this test to verify the frequency accuracy of the Agilent 8702E over its entire operating frequency range.
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 5343A
Cable, 50Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
Adapter, APC-3.5 (f) to Type-N (f) . . . . . . . . . . . . . . . . Agilent P/N 1250-1745
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 11524A
Adapter, Type-N (f) to BNC (m) . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1477
Additional Equipment for Agilent 8702E Option 075:
Minimum Loss Pad, 50Ω to 75Ω . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11852B
1
Connect the equipment as shown in Figure 4-1.
4-3
Performance Tests
1. Test Port Output Frequency Range and Accuracy
In Case of Difficulty
Figure 4-1. Test Port Output Frequency Range and Accuracy Test Setup
2
3
PRESET, SWEEP SETUP
Press
Press 30,
k/m
CW FREQ
,
.
and write the frequency counter reading on the “Performance
Test Record.”
4
Repeat Step 3 for each instrument frequency listed in the “Performance Test Record.”
1
If any measured frequency is close to the specification limits, check the time base accuracy of the counter used.
2
If the analyzer fails by a significant margin at
frequencies (especially if the
all
deviation increases with frequency), the master time base probably needs
4-4
Performance Tests
2. External Source Mode Frequency Range
Required Equipment
Procedure
adjustment. In this case, refer to the
Agilent 8753D Service Guide
. The
“Fractional-N Frequency Range Adjustment” also affects frequency accuracy.
2. External Source Mode Frequency Range
Perform this test to verify that the analyzer’s reference channel, input R, is capable of phase locking to an external CW signal.
External Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 83640A
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Adapter, APC-3.5 (f) to APC-7 . . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1747
Adapter, APC-3.5 (m) to APC-7 . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1746
1
On the external source, press
GHz/dB(m)
2
Connect the equipment as shown in Figure 4-2.
.
PRESET, SWEEP SETUP, CW Freq, 10, M/u, POWER, 20
,
Figure 4-2. External Source Mode Frequency Range Test Setup
3
On the analyzer, press
Press
µ
.
SYSTEM
INSTRUMENT MODE
,
4
PRESET, MEAS
EXT SOURCE AUTO
,
INPUT PORTS
,
R
,
.
SWEEP SETUP
,
CW FREQ
,
, 10,
M/
4-5
Performance Tests
3. Test Port Output Power Accuracy
5
Check to see if the analyzer is phase locking to the external CW signal:
• If the analyzer displays any phase lock error messages, write “unlock” in the
“Performance Test Record” for the set CW signal.
In Case of Difficulty
• If the analyzer does
in the “Performance Test Record” for the set CW signal.
6
On the external source, press CW, 50,
7
On the analyzer, press 50,
8
Repeat Step 5 through Step 7 for the other external source CW frequencies listed in the “Performance Test Record.” For analyzers with Option 006, change the power sensor to the Agilent 8481A for frequencies above 4 GHz.
1
Be sure the external source power is set within 0 to –25 dBm.
2
Make sure the analyzer’s “Ext Source Auto" feature is selected. In addition, verify that the analyzer is set to measure its input channel R.
3
Verify that all connections are tight.
display any phase lock error messages, write “lock”
not
MHz/
µ
.
M/
µ
.
3. Test Port Output Power Accuracy
Perform this test to confirm the accuracy of the Agilent 8702E source output power.
Required Equipment
Equipment Required for 50 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11524A
Additional Equipment Required for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
Equipment Required for 75 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8483A Option H03
4-6
Procedure
Performance Tests
3. Test Port Output Power Accuracy
1
Zero and calibrate the power meter. For more information of how to perform this task, refer to the power meter operating manual.
2
Connect the equipment as shown in Figure 4-3.
Figure 4-3. Source Output Power Accuracy Test Setup
Press
PRESET
.
3
4-7
Performance Tests
4. Test Port Output Power Range and Linearity
NOTE
The factory preset test port power is 0 dBm.
In Case of Difficulty
Required Equipment
4
5
6
1
2
SWEEP SETUP
Press meter for this CW frequency.
Write the power meter reading on the "Performance Test Record."
Repeat Step 4 and Step 5 for each CW frequency listed in the "Performance Test Record."
Be sure the source power is switched on. Press
SOURCE PWR
to switch on the source power.
Refer to the information.
softkey, ON
CW FREQ
,
Agilent 8753D Service Guide
300, k/m
,
should
. Set the calibration factor on the power
SWEEP SETUP
be highlighted. Otherwise, press
for more troubleshooting
POWER
,
SOURCE PWR
. Check the
4. Test Port Output Power Range and Linearity
Perform this test to verify the analyzer’s test port output power range and power level linearity at selected CW frequencies.
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11524A
Procedure
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
Additional Required Equipment for Analyzers with Option 075
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8483A Option H03
1
Zero and calibrate the power meter. Refer to the power meter operating and service manual for more information on how to do this task.
4-8
Performance Tests
4. Test Port Output Power Range and Linearity
2
On the analyzer, press
PRESET, SWEEP SETUP
meter calibration factor for this CW frequency.
3
Connect the equipment as shown in Figure 4-4.
CW FREQ
,
300, k/m
,
. Set the power
Figure 4-4. Test Port Output Power Range and Accuracy Test Setup
4
On the Agilent 438A, press
REL
. This sets the current power level for relative
power measurement.
5
On the analyzer, press
6
Write the power meter reading in the "Results Measured" column on the
SWEEP SETUP
POWER
,
PWR RANGE MAN
,
.
4-9
Performance Tests
5. Minimum R Channel Level
"Performance Test Record."
7
Calculate the difference between the analyzer test port power (which appears on the analyzer’s display) and the power meter reading. Write the result in the "Power Level Linearity" column on the "Performance Test Record."
8
Repeat Step 5 through Step 7 for the other power levels listed in the "Performance Test Record."
In Case of Difficulty
9
10
11
12
13
14
1
2
3
CW FREQ
Press
Set the power meter calibration factor for this CW frequency and press set the reference at this new frequency.
SWEEP SETUP
Press
Write the power meter reading in the "Results Measured" column on the "Performance Test Record."
Calculate the difference between the analyzer test port power and the power meter reading. Write the result in the "Power Level Linearity" column of the "Performance Test Record."
Repeat Step 11 through Step 13 for the other power levels listed in the "Performance Test Record."
Ensure that the power meter and power sensor(s) are operating to specifications. Be sure you set the power meter calibration factor for the CW frequency that you are testing.
Verify that there is power coming out of the analyzer’s test port 1. Be sure you did
SWEEP SETUP
Repeat this performance test.
accidently switch off the analyzer’s internal source. If you did so, press
not
, 3,
POWER
,
G/n
.
POWER
,
,
SOURCE PWR ON
,
–15, x1
.
.
REL
to
5. Minimum R Channel Level
Perform this test to determine the minimum R channel input power level at which phase lock can be accomplished.
4-10
Performance Tests
5. Minimum R Channel Level
Required Equipment
Procedure
For 50 Analyzers
Adapter, APC-3.5 (m) to APC-7 . . . . . . . . . . . . . . . . . . Agilent P/N 1250-1746
Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75 Analyzers (Option 075)
Minimum Loss Pad, 50Ω to 75Ω . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11852B
Cable, 50Ω Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
Adapter, APC-3.5 (m) to Type-N (f) . . . . . . . . . . . . . . . Agilent P/N 1250-1750
1
Connect the equipment as shown in Figure 4-5.
Figure 4-5. Minimum R Channel Level Test Setup
2
3
4
5
6
7
8
PRESET, MEAS
Press
SWEEP SETUP
Press
SCALE REF
Press
SWEEP SETUP
Press
SWEEP SETUP
Press
The analyzer displays the message,
INPUT LEVEL.
INPUT PORTS
,
POWER
,
REFERENCE VALUE
,
CW FREQ
,
POWER
,
PWR RANGE MAN
,
,
–65, x1
,
R
,
.
POWER RANGES, RANGE 4 –55 to –30
,
–70, x1
,
300, k/m
.
.
.
CAUTION: NO IF FOUND: CHECK R
Press ⇑ to increase the test port power by 1 dBm.
If the analyzer displays a phase lock error message, continue increasing the test
.
4-11
In Case of Difficulty
Performance Tests
5. Minimum R Channel Level
port power until phase lock is achieved.
9
Write the test port power, that is displayed on the analyzer, on the "Performance Test Record."
10
Repeat Step 5 through Step 9 for the other CW frequencies listed in the "Performance Test Record."
WARNING
WARNING
WARNING
These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
Disconnect the instrument from all voltage sources while it is being opened. The opening or removal of the covers or parts is likely to expose dangerous voltages.
The power cord is connected to internal capacitors that may remain live for five seconds after disconnecting the plug from its power supply.
1
Check the flexible RF cable (W8, as shown in Figure 4-6) between the R sampler assembly (A4) and the A11 phase lock assembly. Make sure it is connected between A11J1 (PL IF IN) and 1st IF Out.
4-12
Performance Tests
5. Minimum R Channel Level
CAUTION
Loop the W8 cable around the A5 sampler as shown in Figure 4-6. the cable down next to the A11 phase lock assembly.
Do not
push
Figure 4-6. Component Locations
2
Using an ohmmeter, verify that the RF cable is
open. In addition, examine
not
both the cable connectors - measure the resistance between the cable center pin and the cable connector and make sure it is
3
Check the R sampler by substituting it with the B sampler (A6).
a
Move cable W8 to the B sampler (A6), as shown in Figure 4-7.
close to zero.
not
4-13
Performance Tests
5. Minimum R Channel Level
Figure 4-7. Connections for Substituting the R Sampler (A4)
4
Connect the equipment as shown in Figure 4-8.
Figure 4-8. Setup for Checking the R Sampler (A4)
5
Repeat the test, but select the B sampler (A6) by pressing
4-14
MEAS
INPUT PORTS
,
,
Performance Tests
6. Test Port Input Noise Floor Level
B
in Step 2. Use the following specifications:
300 kHz to 3 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –27 dBm
3 GHz to 6 GHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . < –22 dBm
6
If the analyzer fails the test, replace the A11 assembly.
7
Verify that the high/low band adjustments are still within specifications. For more information on how to perform this task, refer to the
Service Guide
8
Refer to the
.
Agilent 8753D Service Guide
for more troubleshooting
Agilent 8753D
information.
6. Test Port Input Noise Floor Level
Perform this test to determine the Agilent 8702E port 1 and port 2 noise floor levels at the input test ports.
Required Equipment
Procedure
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
For 75 Analyzers
Calibration Kit, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85036B
Port 1 Noise Floor Level from 50 kHz to 3 GHz (IF BW= 3 kHz)
1
Connect the equipment as shown in Figure 4-9.
Figure 4-9. Source Input Noise Floor Test Setup
4-15
Performance Tests
6. Test Port Input Noise Floor Level
2
3
4
5
PRESET, SWEEP SETUP
Press
MEAS
Press
Press
SINGLE
,
MARKER FCTN
.
INPUT PORTS
MARKER MODE MENU, STATS ON
,
POWER
,
A
TESTPORT 2
,
,
When the analyzer finishes the sweep, notice the mean value (which appears
–85, x1, STOP, 3, G/n
,
FORMAT
,
LIN MAG
,
,
.
SCALE REF
,
SWEEP SETUP
AUTO SCALE
,
TRIGGER MENU
,
on the analyzer display).
6
Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm
) = 20 * [log
(
linear magnitude mean value
10
)]
NOTE
Notice that the mean value that is displayed on the analyzer is in example, if the displayed value is 62
µµµµ
U, the value that you would put in the
µµµµ
Units. For
equation is
–6
××××
(62
10
).
7
Write this calculated value on the "Performance Test Record."
Port 1 Noise Floor Level from 50 kHz to 3 GHz (IF BW= 10 Hz)
8
Press
AVG
,
IF BW
10, x1
,
to change the IF bandwidth to 10 Hz.
.
,
Press
SWEEP SETUP
TRIGGER MENU
,
SINGLE
,
.
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using
9
10
11
this equation.
Power (dBm
12
Write this calculated value on the "Performance Test Record."
) = 20 * [log
(
linear magnitude mean value
10
)]
Port 2 Noise Floor Level from 50 kHz to 3 GHz (IF BW= 10 Hz)
13
14
15
16
MEAS
Press
Press
INPUT PORTS
,
SWEEP SETUP
,
B
TESTPORT 1
,
,
TRIGGER MENU
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using
SINGLE
,
FORMAT
,
.
LIN MAG
,
.
this equation.
4-16
Performance Tests
6. Test Port Input Noise Floor Level
Power (dBm
17
Write this calculated value on the "Performance Test Record."
) = 20 * [log
(
linear magnitude mean value
10
)]
Port 2 Noise Floor Level from 50 kHz to 3 GHz (IF BW= 3 kHz)
18
19
20
21
AVG
Press
Press
IF BW
,
SWEEP SETUP
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using
k/m
, 3,
to change the IF bandwidth to 3 kHz.
TRIGGER MENU
,
SINGLE
,
.
this equation.
Power (dBm
22
Write this calculated value on the "Performance Test Record."
23
This completes the "Test Port Input Noise Floor Level" procedure if your analyzer does
) = 20 * [log
have Option 006. Otherwise, continue with the next section.
not
(
linear magnitude mean value
10
)]
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF BW= 3 kHz)
24
25
26
27
START, 3, G/n, STOP, 6, G/n
Press
Press
SWEEP SETUP
TRIGGER MENU
,
.
SINGLE
,
.
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm
28
Write this calculated value on the "Performance Test Record."
) = 20 * [log
(
linear magnitude mean value
10
)]
Port 2 Noise Floor Level from 3 GHz to 6 GHz (IF BW= 10 Hz)
29
30
31
32
AVG
Press
Press
IF BW
,
, 10, x1 to change the IF bandwidth to 10 Hz.
SWEEP SETUP
TRIGGER MENU
,
SINGLE
,
.
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using this equation.
Power (dBm
33
Write this calculated value on the "Performance Test Record."
) = 20 * [log
(
linear magnitude mean value
10
)]
Port 1 Noise Floor Level for 3 GHz to 6 GHz (IF BW= 10 Hz)
Press
MEAS
INPUT PORTS
,
A
TESTPORT 2
,
,
.
34
4-17
Performance Tests
7. Test Port Input Frequency Response
In Case of Difficulty
Press
SWEEP SETUP
TRIGGER MENU
,
SINGLE
,
.
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using
35
36
37
this equation.
Power (dBm
38
Write this calculated value on the "Performance Test Record."
) = 20 * [log
(
linear magnitude mean value
10
)]
Port 1 Noise Floor Level from 3 GHz to 6 GHz (IF BW= 3 kHz)
39
40
41
42
AVG
Press
Press
IF BW
,
SWEEP SETUP
When the analyzer finishes the sweep, notice the mean value.
Convert the measured linear magnitude mean value to log magnitude, using
k/m
, 3,
.
TRIGGER MENU
,
SINGLE
,
.
this equation.
Power (dBm
43
Write this calculated value on the "Performance Test Record."
1
Perform the "ADC Linearity Correction Constants (Test 52)," located in the
) = 20 * [log
(
linear magnitude mean value
10
"Adjustments and Correction Constants" chapter of the
Service Guide
.
)]
Agilent 8753D
2
Repeat the "Test Port Input Noise Floor Level" procedure.
3
Check the A10 Digital IF assembly if the analyzer fails both test port input noise floor tests.
4
Refer to the
Agilent 8753D Service Guide
for more troubleshooting
information.
7. Test Port Input Frequency Response
Perform this test to examine the vector sum of all test setup error vectors in both magnitude and phase change as a function of frequency.
4-18
Performance Tests
7. Test Port Input Frequency Response
Required Equipment
Procedure
For 50 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11524A
Additional Equipment Required for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
For 75 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8483A Option H03
Cable, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Power Meter Calibration for Test Port 1 from 300 kHz to 3 GHz
1
Zero and calibrate the power meter.
2
Connect the equipment as shown in Figure 4-10.
4-19
Performance Tests
7. Test Port Input Frequency Response
Figure 4-10. Setup for Power Meter Calibration on Test Port 1
3
4
5
6
PRESET, START, 300, k/m
Press
.
Only for Analyzers with Option 006:
LOCAL
Press
Press
SYSTEM CONTROLLER
,
SET ADDRESSES
and
.
POWER MTR
until the analyzer shows the correct power
Press
STOP, 3, G/n
.
meter model.
7
ADDRESS: P MTR/HPIB
Press
. The default power meter GPIB address is 13. Make sure it is the same as your power meter GPIB address. Otherwise, use the analyzer front panel keypad to enter the correct GPIB address for your power meter.
4-20
Performance Tests
7. Test Port Input Frequency Response
Press
Press
SWEEP SETUP
POWER
PWR RANGE MAN
,
NUMBER of POINTS
,
to turn the auto power range off.
, 51, x1.
8
9
NOTE
The analyzer displays the
PRm
annotation, indicating that the analyzer power
range is set to MANUAL.
10
11
12
PORT POWER
Press
CAL
Press
LOSS/SENSR LISTS
Press
to uncouple the test port output power.
PWRMTR CAL
,
.
CAL FACTOR SENSOR A
,
. Refer to the back of the power sensor to locate the different calibration factor values along with their corresponding frequencies.
NOTE
The analyzer’s calibration factor sensor table can hold a MAXIMUM of 12 calibration factor data points.
The following softkeys are included in the sensor calibration factor entries menu:
SEGMENT
EDIT
DELETE
press to select a point where you can use the front panel knob or entry keys to enter a value
press to edit or change a previously entered value
press to delete a point from the sensor calibration factor table
ADD
select this key to add a point into the sensor calibration factor table
CLEAR LIST
select this key to erase the entire sensor calibration
factor table
DONE
select this key when done entering points to the sensor calibration factor table
As an example, the following are the keystrokes for entering the first two cali­bration factor data points for the Agilent 8482A power sensor (assuming CF%= 96.4 at 100 kHz and CF%= 98.4 at 300 kHz):
4-21
Performance Tests
7. Test Port Input Frequency Response
a
From the sensor calibration factor entries menu, press
b
Press
FREQUENCY
100, k/m
,
. If you make an entry error, press ← and re-enter
ADD
.
the correct value again.
c
d
e
f
g
h
i
CAL FACTOR
Press
DONE
Press
To enter the second cal factor data point, press
FREQUENCY
Press
CAL FACTOR
Press
To terminate the second calibration factor data point entry, press
SEGMENT
Press
96.4, x1
,
.
to terminate the first calibration factor data point entry.
ADD
.
300, k/m
,
98.4, x1
,
.
.
and use the front panel knob to scroll through the sensor cal­ibration factors table. Check to be sure all values are entered correctly. If you spot an error, use the front panel knob to point to the data point you want to modify and press
13
Press the appropriate softkeys to create a power sensor calibration factors
EDIT
.
table.
14
15
Press
Press
DONE
to exit the sensor calibration factor entries menu.
RETURN
ONE SWEEP, TAKE CAL SWEEP
,
to start the power meter calibration.
Wait until the analyzer finishes the sweep, then continue with this procedure.
DONE
.
NOTE
The analyzer displays the PC annotation, indicating the power meter calibra­tion is done and the error correction is active.
Test Port 2 Input Frequency Response from 300 kHz to 3 GHz
16
Connect the equipment as shown in Figure 4-11.
4-22
7. Test Port Input Frequency Response
Figure 4-11. Test Port 2 Input Frequency Response Test Setup
Performance Tests
17
18
19
20
MEAS
Press
SCALE REF
Press
MARKER
Press
SEARCH:MIN
MARKER
Press
SEARCH:MAX
INPUT PORTS
,
SCALE/DIV
,
MARKER 1
,
B
,
.
, 1, x1.
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
to put marker 1 at the minimum magnitude location of the trace.
MARKER 2
,
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
to position marker 2 at the maximum magnitude location of the
trace.
21
Write the marker 1 or marker 2 value (which appears on the analyzer display), whichever has the larger absolute magnitude, in the “Performance Test Record.”
Power Meter Calibration on Port 2 from 300 kHz to 3 GHz
22
Connect the equipment as shown Figure 4-12.
,
,
4-23
Performance Tests
7. Test Port Input Frequency Response
Figure 4-12. Setup for Power Meter Calibration on Test Port 2
Press
Press
MEAS
INPUT PORTS
,
CAL
PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP
,
TESTPORT 2
,
.
23
24
calibration for test port 2.
25
When the analyzer displays the message,
SWEEP DONE
4-24
, connect the equipment as shown as in Figure 4-13.
P
OWER METER CALIBRATION
to start the power meter
Performance Tests
7. Test Port Input Frequency Response
Figure 4-13. Test Port 1 Input Frequency Response Test Setup
Test Port 1 Input Frequency Response from 300 kHz to 3 GHz
26
27
28
29
MEAS
Press
MARKER
Press
SEARCH:MIN
MARKER
Press
SEARCH:MAX
INPUT PORTS
,
MARKER 1
,
.
MARKER 2
,
.
A
,
.
MARKER FCTN
,
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
MARKER MODE MENU, MKR SEARCH
,
Write the marker 1 or marker 2 reading, whichever has the larger absolute magnitude, in the “Performance Test Record.”
30
This completes the “Test Port Input Frequency Response” procedure if your analyzer does
have Option 006. Otherwise, continue with the next sections.
not
Power Meter Calibration for Test Port 2 from 3 GHz to 6 GHz
31
Replace the power sensor with the Agilent 8481A, and then setup the power meter:
• If the power meter is an Agilent 438A, press
• If the power meter is an Agilent 437B, press
LCL
.
PRESET/LOCAL
.
• If the power meter is an Agilent 436A, cycle the line power.
32
Connect the equipment as shown in Figure 4-14.
,
,
4-25
Performance Tests
7. Test Port Input Frequency Response
Figure 4-14. Setup for Power Meter Calibration on Test Port 2
33
34
35
START, 3, G/n, STOP, 6, G/n
Press
CAL
Press
Press
PWRMTR CAL
,
LOSS/SENSR LISTS
.
.
CAL FACTOR SENSOR B
,
. Repeat Step 10 to build a
calibration factor sensor table for the Agilent 8481A power sensor.
36
37
38
DONE
Press
to exit the sensor calibration factor entries menu.
To select the Agilent 8481A power sensor, press
Press
4-26
RETURN
TAKE CAL SWEEP
,
to start the power meter calibration.
USE SENSOR B
.
Performance Tests
7. Test Port Input Frequency Response
Test Port 1 Input Frequency Response from 3 GHz to 6 GHz
39
When the analyzer finishes the calibration sweep, connect the equipment as shown in Figure 4-15.
Figure 4-15. Setup for Test Port 1 Input Frequency Response
40
41
42
MEAS
Press
MARKER
Press
SEARCH:MIN
MARKER
Press
SEARCH:MAX
INPUT PORTS
,
MARKER 1
,
A
,
.
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
to put marker 1 at the minimum magnitude location of the trace.
MARKER 2
,
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
to position marker 2 at the maximum magnitude location of the
trace.
43
Write the marker 1 or marker 2 reading, whichever has the largest absolute magnitude, in the “Performance Test Record.”
Power Meter Calibration on Test Port 1 from 3 GHz to 6 GHz
44
Connect the equipment as shown in Figure 4-16.
,
,
4-27
Performance Tests
7. Test Port Input Frequency Response
Figure 4-16. Setup for Power Meter Calibration on Test Port 1
Press
Press
MEAS
INPUT PORTS
,
CAL
PWRMTR CAL, ONE SWEEP, TAKE CAL SWEEP
,
TESTPORT 1
,
.
to start the power meter
45
46
calibration for output test port 1.
Test Port 2 Input Frequency Response from 3 GHz to 6 GHz
47
When the analyzer displays the message,
SWEEP DONE
4-28
, connect the equipment as shown as in Figure 4-17.
POWER METER CALIBRATION
Figure 4-17. Test Port 2 Input Frequency Response Test Setup
Performance Tests
8. Test Port Crosstalk
In Case of Difficulty
48
49
50
51
MEAS
Press
MARKER
Press
SEARCH:MIN
MARKER
Press
SEARCH:MAX
INPUT PORTS
,
MARKER 1
,
.
MARKER 2
,
.
B
,
.
MARKER FCTN
,
MARKER FCTN
,
MARKER MODE MENU, MKR SEARCH
,
MARKER MODE MENU, MKR SEARCH
,
Write the marker 1 or marker 2 reading, whichever has the largest magnitude, in the “Performance Test Record.”
1
Be sure you have used the correct power sensor for the frequency range.
2
Verify that the calibration factors that you have entered for the power sensors are correct.
3
Repeat this test with a “known good” thru cable.
8. Test Port Crosstalk
Perform this test to verify the signal leakage between the analyzer’s test ports.
,
,
4-29
Performance Tests
8. Test Port Crosstalk
Required Equipment
Procedure
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75 Analyzers
Calibration Kit, 75Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85036B
1
Connect the equipment as shown in Figure 4-18.
Figure 4-18. Test Port Crosstalk Test Setup
2
3
PRESET, SWEEP SETUP
Press
AVG
Press
4-30
IF BW
,
, 10, x1.
POWER
,
, 10, x1.
Crosstalk to Test Port 2 from 300 kHz to 3 GHz
Performance Tests
8. Test Port Crosstalk
4
5
6
7
8
9
START, 300, k/m, STOP, 3, G/n
Press
Press
Press
Press
Press
MEAS
Trans: E/E S21 FWD
,
SCALE REF
REFERENCE VALUE
,
SWEEP SETUP
MARKER FCTN
TRIGGER MENU
,
MARKER MODE MENU, MKR SEARCH
,
.
.
–100, x1
,
SINGLE
,
.
.
SEARCH: MAX
,
Write the marker value (which appears on the analyzer display) in the “Performance Test Record.”
Crosstalk to Test Port 1 from 300 kHz to 3 GHz
10
11
12
13
MEAS
Press
Press
Press
Trans: E/E S12 REV
,
SWEEP SETUP
MARKER FCTN
.
TRIGGER MENU
,
MARKER MODE MENU, MKR SEARCH
,
SINGLE
,
.
SEARCH: MAX
,
Write the marker value (which appears on the analyzer display) in the “Performance Test Record.”
14
This completes the “Test Port Crosstalk” performance test if your analyzer does
have Option 006. Otherwise, proceed to the next section.
not
Crosstalk to Test Port 1 from 3 GHz to 6 GHz
15
16
START, 3, G/n, STOP, 6, G/n
Press
SWEEP SETUP
Press
TRIGGER MENU
,
.
SINGLE
,
.
.
.
Press
MARKER FCTN
MARKER MODE MENU, MKR SEARCH
,
SEARCH: MAX
,
Write the marker value (which appears on the analyzer display) in the
17
18
“Performance Test Record.”
Crosstalk to Test Port 2 from 3 GHz to 6 GHz
1
2
3
4
MEAS
Press
Press
Press
Trans: E/E S21 FWD
,
SWEEP SETUP
MARKER FCTN
.
TRIGGER MENU
,
MARKER MODE MENU, MKR SEARCH
,
SINGLE
,
.
SEARCH: MAX
,
Write the marker value (which appears on the analyzer display) in the “Performance Test Record.”
.
.
4-31
In Case of Difficulty
Performance Tests
8. Test Port Crosstalk
WARNING
WARNING
WARNING
These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
Disconnect the instrument from all voltage sources while it is being opened. The opening or removal of the covers or parts is likely to expose dangerous voltages.
The power cord is connected to internal capacitors that may remain live for five seconds after disconnecting the plug from its power supply.
1
Remove the instrument top cover. Using an 8 lb-inch torque wrench, verify that
semi-rigid cables connected to the sampler/mixer assemblies are tight. In
all
addition tighten any loose screws on the sampler/mixer assemblies (A4/5/6) and the pulse generator assembly (A7).
2
Remove the instrument bottom cover. Refer to Figure 4-19. Verify that cables W1, W31 and W32 are tight.
3
Repeat this test.
4-32
Performance Tests
8. Test Port Crosstalk
Figure 4-19. Agilent 8702E Bottom View
4-33
Performance Tests
9. Calibration Coefficients
9. Calibration Coefficients
Perform this procedure to verify the analyzer uncorrected test port character­istics.
Required Equipment
Procedure
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75 Analyzers
Calibration Kit, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85036B
Cable, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
NOTE
The crosstalk calibration coefficients are omitted in this procedure. They are covered in the “Test Port Crosstalk” performance test.
First Full 2-Port Calibration
1
Connect the equipment as shown in Figure 4-20.
Figure 4-20. First Full 2-Port Calibration Test Setup
4-34
Performance Tests
9. Calibration Coefficients
2
3
4
5
PRESET, START, 300, k/m
Press
CAL
Press
CAL KITS & STDS
,
CALIBRATE MENU, FULL 2-PORT
Press
ISOLATION
OMIT ISOLATION
,
.
SELECT CAL KIT, CAL KIT: 7mm
,
.
ISOLATION DONE
,
.
RETURN
,
RETURN
,
Connect the “open” end of the open/short combination (supplied in the calibration kit) to analyzer test port 1.
6
7
8
9
Press
REFLECTION
FORWARD:OPEN
,
.
Connect the “short” end of the open/short combination to analyzer test port 1.
FORWARD:SHORT
Press
.
Replace the open/short combination with the 50 ohm termination (supplied in the calibration kit).
10
11
FORWARD:LOAD
Press
.
Connect the “open” end of the open/short combination to the reference test port 2.
12
13
REVERSE:OPEN
Press
.
Connect the “short” end of the open/short combination to the reference test port 2.
14
REVERSE: SHORT
Press
.
,
15
Connect the 50 ohm termination to the reference test port 2.
16
17
REVERSE: LOAD
Press
At the prompt,
STANDARDS DONE
.
PRESS ’DONE’ IF FINISHED WITH STD(s)
.
Wait for the message,
COMPUTING CAL COEFFICIENTS,
the analyzer display before proceeding to the next step.
18
Connect the equipment as shown in Figure 4-21.
, press
to disappear from
4-35
Performance Tests
9. Calibration Coefficients
Figure 4-21. Transmission Calibration Test Setup
19
20
21
22
23
Press
TRANSMISSION
FWD TRANS THRU
,
.
When the analyzer finishes the measurement, press
When the analyzer finishes the measurement, press
After the analyzer finishes the measurement, press
After the analyzer finishes the measurement, press
PORT CAL
.
FWD MATCH THRU
REV TRANS THRU
REV MATCH THRU
.
.
STANDARDS DONE, DONE 2-
Directivity (Forward) Calibration Coefficient
24
25
26
SYSTEM
Press
When the analyzer finishes the test, press
SERVICE MENU
,
TESTS
,
, 32, x1,
EXECUTE TEST
MARKER
.
.
Using the front panel knob, locate the maximum value of the data trace for the 300 kHz to 1.3 GHz frequency range.
27
Write the maximum value in the “Performance Test Record.”
28
Repeat the previous two steps for the other frequency range(s) listed on the “Performance Test Record.”
Source Match (Forward) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 33, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 25 through Step 28.
29
30
.
4-36
Performance Tests
9. Calibration Coefficients
Transmission Tracking (Forward) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 37, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 25 through Step 28.
31
32
Reflection Tracking (Forward) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 34, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 25 through Step 28.
33
34
Load Match (Reverse) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 42, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 25 through Step 28.
35
36
Transmission Tracking (Reverse) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 43, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 25 through Step 28.
37
38
Second Full 2-Port Calibration
39
Connect the equipment as shown in Figure 4-22.
Figure 4-22. Second Full 2-Port Calibration Test Setup
40
41
PRESET, START, 300, k/m
Press
CAL
Press
CAL KITS & STDS
,
.
SELECT CAL KIT, CAL KIT: 7mm
,
RETURN, RETURN
,
,
4-37
Performance Tests
9. Calibration Coefficients
CALIBRATE MENU, FULL 2-PORT
42
43
ISOLATION, OMIT ISOLATION
Press
Connect the “open” end of the open/short combination (supplied in the
.
ISOLATION DONE
,
.
calibration kit) to reference test port 1.
44
45
46
47
Press
REFLECTION
FORWARD:OPEN
,
.
Connect the “short” end of the open/short combination to reference test port 1.
FORWARD:SHORT
Press
.
Replace the open/short combination with the 50 ohm termination (supplied in the calibration kit).
48
49
FORWARD:LOAD
Press
.
Connect the “open” end of the open/short combination to the analyzer test port 2.
50
51
REVERSE:OPEN
Press
.
Connect the “short” end of the open/short combination to the analyzer test port 2.
52
53
54
REVERSE: SHORT
Press
.
Connect the 50 ohm termination to the analyzer test port 2.
REVERSE: LOAD
Press
.
55
When the analyzer displays,
STANDARDS DONE
press
Wait for the message,
PRESS ’DONE’ IF FINISHED WITH STD(s)
.
COMPUTING CAL COEFFICIENTS
the analyzer display before proceeding to the next step.
56
Connect the equipment as shown in Figure 4-23.
4-38
,
, to disappear from
Figure 4-23. Transmission Calibration Test Setup
Performance Tests
9. Calibration Coefficients
57
58
59
60
61
TRANSMISSION
Press
FWD TRANS THRU
,
.
When the analyzer finishes the measurement, press
When the analyzer finishes the measurement, press
When the analyzer finishes the measurement, press
When the analyzer displays,
STANDARDS DONE
press
PRESS ’DONE’ IF FINISHED WITH STD(s)
DONE 2-PORT CAL
,
.
FWD MATCH THRU
REV TRANS THRU
REV MATCH THRU
.
.
Load Match (Forward) Calibration Coefficient
62
63
64
SYSTEM
Press
When the test is done, press
SERVICE MENU
,
TESTS
,
MARKER
, 36, x1,
MARKER 1
,
EXECUTE TEST
.
.
Using the front panel knob, locate the maximum value of the data trace for the 300 kHz to 1.3 GHz frequency range.
65
Write the maximum value on the “Performance Test Record.”
66
Repeat the previous three steps for the other frequency range(s) listed on the “Performance Test Record.”
Directivity (Reverse) Calibration Coefficient
Press
SYSTEM
SERVICE MENU
,
TESTS
,
, 38, x1,
EXECUTE TEST
.
When the analyzer finishes the test, repeat Step 63 through Step 66.
67
68
.
,
4-39
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
Source Match (Reverse) Calibration Coefficient
Required Equipment
Procedure
69
70
71
72
SYSTEM
Press
When the analyzer finishes the test, repeat Step 63 through Step 66.
Reflection Tracking (Reverse) Calibration Coefficient
SYSTEM
Press
When the analyzer finishes the test, repeat Step 63 through Step 66.
SERVICE MENU
,
SERVICE MENU
,
TESTS
,
TESTS
,
, 39, x1,
, 40, x1,
EXECUTE TEST
EXECUTE TEST
.
.
10. System Trace Noise (Only for Analyzers without Option 006)
Perform this test to measure the system trace noise at a designated frequency in both the A/R and B/R ratioed measurements.
For 50 Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75 Analyzers
Cable, 75Ω, Type-N 24-inch . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
1
Connect the equipment as shown in Figure 4-24.
Figure 4-24. System Trace Noise Test Setup
4-40
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
2
3
4
PRESET, SWEEP SETUP
Press
RETURN, CW FREQ
Press
MARKER FCTN
Press
POWER
,
G/n
, 3,
MARKER MODE MENU, STATS ON
,
, 5, x1.
NUMBER of POINTS
,
1601, x1
,
.
to activate the instrument’s
statistic feature.
System Trace Noise for A/R Magnitude
5
6
7
8
MEAS
Press
Press
When the analyzer displays the “
Trans: E/E S12 REV
,
SWEEP SETUP
.
TRIGGER MENU
,
NUMBER of GROUPS
,
Hld
” annotation, press
, 5, x1.
SCALE REF
AUTO SCALE
,
Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.”
System Trace Noise for A/R Phase
9
10
11
12
FORMAT
Press
SWEEP SETUP
Press
When the analyzer finishes the number of sweeps, press
PHASE
,
.
TRIGGER MENU, NUMBER of GROUPS
,
, 5, x1.
SCALE REF
AUTO SCALE
,
Write the s.dev (standard deviation) value on the “Performance Test Record.”
System Trace Noise for B/R Magnitude
Press
MEAS
Trans: E/E S21 FWD
,
.
13
.
.
In Case of Difficulty
14
15
16
SWEEP SETUP
Press
When the analyzer finishes the number of sweeps, press
TRIGGER MENU
,
NUMBER of GROUPS
,
, 5, x1.
SCALE REF
AUTO SCALE
,
Write the s.dev (standard deviation) value on the “Performance Test Record.”
System Trace Noise for B/R Phase
17
18
19
20
1
FORMAT
Press
SWEEP SETUP
Press
When the analyzer finishes the number of sweeps, press
PHASE
,
,
.
TRIGGER MENU
NUMBER of GROUPS
,
, 5, x1.
SCALE REF
AUTO SCALE
,
Write the s.dev (standard deviation) value on the “Performance Test Record.”
Perform the “ADC Offset Correction Constants” procedure, located in the “Adjustments and Correction Constants” chapter of the
Service Guide
.
Agilent 8753D
.
.
4-41
Performance Tests
11. System Trace Noise (Only for Analyzers with Option 006)
2
Repeat this performance test.
3
Check the A10 Digital IF board assembly if the analyzer still fails the test.
11. System Trace Noise (Only for Analyzers with Option
006)
Perform this test to measure the system trace noise at designated CW fre­quencies in both the A/R and B/R ratioed measurements.
Required Equipment
Procedure
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
1
Connect the equipment as shown in Figure 4-25.
Figure 4-25. System Trace Noise Test Setup
2
PRESET, SWEEP SETUP
Press
POWER
,
, 5, x1,
RETURN
NUMBER of POINTS
,
1601, x1
,
.
3
4
MARKER FCTN
Press statistic feature.
System Trace Noise for A/R Magnitude from 300 kHz to 3 GHz
MEAS
Press
4-42
Trans: E/E S12 REV
,
MARKER MODE MENU, STATS ON
,
.
to activate the instrument’s
Performance Tests
11. System Trace Noise (Only for Analyzers with Option 006)
Press
SWEEP SETUP
CW FREQ
,
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value shown, which appears on the
5
6
7
, 3,
G/n
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
analyzer display, on the “Performance Test Record.”
System Trace Noise for A/R Magnitude from 3 GHz to 6 GHz
Press
SWEEP SETUP
CW FREQ
,
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value, which appears on the analyzer
8
9
10
, 6,
G/n
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
display, on the “Performance Test Record.”
System Trace Noise for A/R Phase from 3 GHz to 6 GHz
11
12
13
14
FORMAT
Press
SWEEP SETUP
Press
When the analyzer finishes the number of sweeps, press
PHASE
,
,
.
TRIGGER MENU
NUMBER of GROUPS
,
, 5, x1.
SCALE REF
Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.”
System Trace Noise for A/R Phase from 300 kHz to 3 GHz
Press
SWEEP SETUP
CW FREQ
,
When the analyzer finishes the number of sweeps, press
15
16
, 3,
G/n
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
, 5, x1.
AUTO SCALE
,
, 5, x1.
AUTO SCALE
,
AUTO SCALE
,
, 5, x1.
AUTO SCALE
,
.
.
.
.
17
Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.”
System Trace Noise for B/R Magnitude from 300 kHz to 3 GHz
18
19
20
MEAS
Press
5, x1.
Trans: E/E S21 FWD
,
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value, which appears on the analyzer
SWEEP SETUP
,
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
display, on the “Performance Test Record.”
System Trace Noise for B/R Magnitude from 3 GHz to 6 GHz
Press
SWEEP SETUP
CW FREQ
,
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value, which appears on the analyzer
21
22
23
, 6,
G/n
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
display, on the “Performance Test Record.”
AUTO SCALE
,
, 5, x1.
AUTO SCALE
,
,
.
.
4-43
Performance Tests
12. Test Port Input Impedance
System Trace Noise for B/R Phase from 3 GHz to 6 GHz
In Case of Difficulty
24
25
26
27
28
29
1
2
3
FORMAT
Press
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.”
System Trace Noise for B/R Phase from 300 kHz to 3 GHz
SWEEP SETUP
Press
When the analyzer finishes the number of sweeps, press
Write the s.dev (standard deviation) value, which appears on the analyzer display, on the “Performance Test Record.”
Perform the “ADC Offset Correction Constants” procedure, located in the “Adjustments and Correction Constants” chapter of the
Service Guide
Repeat this performance test.
Suspect the A10 Digital IF board assembly if the analyzer still fails the test.
,
PHASE
,
.
SWEEP SETUP
,
CW FREQ
, 3,
TRIGGER MENU
,
G/n
TRIGGER MENU
,
NUMBER of GROUPS
,
SCALE REF
NUMBER of GROUPS
,
SCALE REF
Agilent 8753D
,
,
, 5, x1.
AUTO SCALE
, 5, x1.
AUTO SCALE
.
.
Required Equipment
Procedure
12. Test Port Input Impedance
Perform this test to measure the return loss of each input test port.
For 50 Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85031B
For 75 Analyzers
Cable, 75Ω, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Calibration Kit, 75Ω, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85036B
1
Connect the equipment as shown in Figure 4-26.
4-44
Loading...