Agilent 8702D Reference

Page 1
Reference
HP 8702D Lightwave Component Analyzer
Page 2
© Copyright Hewlett-Packard Company 1998 All Rights Reserved. Repro­duction, adaptation, or trans­lation without prior written permission is prohibited, except as allowed under copy­right laws.
HP Part No. 08702-90050 Printed in USA Februrary 1998
Hewlett-Packard Company Lightwave Operations 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, USA (707) 577-1400
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Hewlett-Packard makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Hewlett­Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Restricted Rights Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and sub­paragraphs (c) (1) and (c) (2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
War ran ty.
This Hewlett-Packard instru­ment product is warranted against defects in material and workmanship for a period of one year from date of ship­ment. During the warranty period, Hewlett-Packard Com­pany will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by Hewlett-Pack­ard. Buyer shall prepay ship­ping charges to Hewlett­Packard and Hewlett-Packard shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Hewlett-Packard from another country.
Hewlett-Packard warrants that its software and firmware designated by Hewlett-Pack­ard for use with an instrument will execute its programming instructions when properly installed on that instrument. Hewlett-Packard does not warrant that the operation of the instrument, or software, or firmware will be uninter­rupted or error-free.
Limitation of Warranty.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfac­ing, unauthorized modifica­tion or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Hewlett-Packard specifically
disclaims the implied warran­ties of merchantability and fit­ness for a particular purpose.
Exclusive Remedies.
The remedies provided herein are buyer's sole and exclusive remedies. Hewlett­Packard shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.
Safety Symbols.
CAUTION
The hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WARNING
The hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
warning symbol when it is necessary for the user to refer to the instructions in the man­ual.
products which have a laser output.
sign denotes a
caution
warning
sign denotes a
The instruction man­ual symbol. The prod­uct is marked with this
The laser radiation symbol. This warning symbol is marked on
The AC symbol is used to indicate the required nature of
the line module input power.
The ON symbols are
|
used to mark the posi­tions of the instrument
power line switch.
The OFF symbols
are used to mark the positions of the instrument power line switch.
The CE mark is a reg­istered trademark of the European Com-
munity.
The CSA mark is a reg­istered trademark of the Canadian Stan-
dards Association.
ISM1-A This text denotes that the instrument is an Industrial Scientific and Medical Group 1 Class A product.
Typographical Conven­tions.
The following conventions are used in this book:
key type
for keys or text located on the keyboard or instrument.
softkey type
are displayed on the instru­ment’s screen.
display type
characters displayed on the computer’s screen or instru­ment’s display.
user type
ters that you type or enter.
emphasis
characters that emphasize some point or that are used as place holders for text that you type.
for key names that
for words or
for words or charac-
type for words or
ii
Page 3

Contents

1 Menu Maps
2 Reference
Instrument Options 2-4 Available Accessories 2-6 Front Panel Features 2-12 Analyzer Display 2-14 Rear Panel Features and Connectors 2-16 Reference Documents 2-19 Location of Softkeys 2-21 Connectors, Adjustments, and Display Annotation 2-45 Preset Conditions 2-48 Power-on Conditions 2-56
3 Error Messages
Error Messages in Alphabetical Order 3-3 Error Messages in Numerical Order 3-27
4 Specifications and Regulatory Information
Instrument Specifications 4-4 Instrument Specifications (Option 011) 4-12 Instrument General Characteristics 4-21 Environmental Characteristics 4-25 System E/E Characteristics 4-27 System O/E Characteristics 4-56 System E/O Characteristics 4-69 System O/O Characteristics 4-82 Lightwave Receiver Measurement Repeatability 4-91 HP-IB Interface 4-93 Regulatory Information 4-94
5Concepts
How the HP 8702D Works 5-3 How the HP 8702D Processes Data 5-7 Using the Response Functions 5-13 What is Measurement Calibration? 5-19 Understanding and Using Time Domain 5-73
Contents-1
Page 4
Instrument Preset State and Memory Allocation 5-101
Contents-2
Page 5
1

Menu Maps

Page 6
Menu Maps
Menu Maps
Menu Maps
The menu maps that are in this chapter graphically represent the softkey menus. Maps for each softkey menu are shown in alphabetical order.
AVG
AVG
MENU
AVERAGING
RESTART
AVERAGING
FACTOR
AVERAGING
on OFF
SMOOTHING
APERTURE
SMOOTHING
on OFF
AVG
menu map
1-2
IF BW
[ ]
Page 7
Menu Maps
Menu Maps
CAL
menu map, 1 of 2
1-3
Page 8
Menu Maps
Menu Maps
CAL
menu map, 2 of 2
1-4
Page 9
Menu Maps
Menu Maps
COPY
menu map
1-5
Page 10
Menu Maps
Menu Maps
DISPLAY
FORMAT
1-6
menu map
menu map
Page 11
Menu Maps
Menu Maps
LOCAL
menu map
1-7
Page 12
Menu Maps
Menu Maps
MARKER
1-8
menu map
Page 13
Menu Maps
Menu Maps
MARKER FCTN
menu map
1-9
Page 14
Menu Maps
Menu Maps
MEAS
menu map, standard and option 011 with test set
MEAS
menu map, option 011, no test set
1-10
Page 15
Menu Maps
Menu Maps
MENU
menu map
1-11
Page 16
Menu Maps
Menu Maps
PRESET
menu map
1-12
Page 17
Menu Maps
Menu Maps
SAVE/RECALL
menu map
1-13
Page 18
Menu Maps
Menu Maps
SCALE REF
menu map
1-14
Page 19
Menu Maps
Menu Maps
SEQ
menu map
1-15
Page 20
Menu Maps
Menu Maps
SYSTEM
1-16
menu map
Page 21
2
Reference
Page 22
Reference
Reference
Reference
This chapter is a dictionary reference table of front and rear-panel connectors, front-panel keys (hardkeys), softkeys, and display annotations. With the exception of a few front-panel keys, softkeys control all instrument functions.
This chapter is designed for quick access of information. For example, during operation you may find a softkey or hardkey whose function is unfamiliar to you. Note the key name, find the key in this chapter, and note the applicable function. Some keys will have more than one function. Keys that begin with a symbol are listed at the front of the table.
2-2
Page 23
What you’ll find in this chapter
Instrument Options 2-4 Available Accessories 2-6
Calibration Kits 2-6 Verification Kit 2-6 Test Port Return Cables 2-7 Adapter Kits 2-7 Transistor Test Fixtures 2-8 System Accessories Available 2-9
AC Line-Power Cords 2-11 Front Panel Features 2-12 Analyzer Display 2-14 Rear Panel Features and Connectors 2-16 Reference Documents 2-19
General Measurement and Calibration Techniques 2-19
Fixtures and Non-coaxial Measurements 2-19
On-Wafer Measurements 2-20 Location of Softkeys 2-21 Connectors, Adjustments, and Display Annotation 2-45 Preset Conditions 2-48 Power-on Conditions 2-56
Reference
Reference
2-3
Page 24
Reference

Instrument Options

Instrument Options
Option 1D5
Option 002
Option 006
Option 011
Option 075
Option 110
High Stability Frequency Reference. Option 1D5 offers ±0.05 ppm tempera­ture stability from 0 to 60°C (referenced to 25°C).
Harmonic Mode. Provides measurement of second or third harmonics of the test device’s fundamental output signal. Frequency and power sweep are sup­ported in this mode. Harmonic frequencies can be measured up to the maxi­mum frequency of the receiver. However, the fundamental frequency may not be lower than 16 MHz.
Extends the maximum source and receiver frequency of the analyzer to 6GHz.
Receiver Configuration. Option 011 allows front panel access to the R, A, and B samplers and receivers. The transfer switch, couplers, and bias tees have been removed. Therefore, external accessories are required to make most measurements.
75Ω Impedance. Option 075 offers 75 ohm impedance bridges with type-N test port connectors.
Deletes Time Domain. This option displays the time domain response of a network by computing the inverse Fourier transform of the frequency domain response. It shows the response of a test device as a function of time and distance. Displaying the reflection coef­ficient of a network versus time determines the magnitude and location of each discontinuity. Displaying the transmission coefficient of a network versus time determines the characteristics of individual transmission paths. Time domain operation retains all accuracy inherent with the correction that is active in the frequency domain. The time domain capability is useful for the design and characterization of such devices as SAW filters, SAW delay lines, RF cables, and RF antennas.
removes
the time domain capability, which
2-4
Page 25
Reference
Instrument Options
Option 1CM
Option 1CP
Rack Mount Flange Kit Without Handles. Option 1CM is a rack mount kit con­taining a pair of flanges and the necessary hardware to mount the instrument, with handles detached, in an equipment rack with 482.6 mm (19 inches) hori­zontal spacing.
Rack Mount Flange Kit With Handles. Option 1CP is a rack mount kit contain­ing a pair of flanges and the necessary hardware to mount the instrument with handles attached in an equipment rack with 482.6 mm (19 inches) spacing.
2-5
Page 26
Reference

Available Accessories

Available Accessories

Calibration Kits

The following calibration kits contain prevision standards and required adapt­ers of the indicated connector type. The standards (known devices) facilitate measurement calibration, also called vector error correction. Refer to the data sheet and ordering guide for additional information. Part numbers for the standards are in their manuals.
• HP 85031B 7 mm Calibration Kit
• HP 85032B 50 Ohm Type-N Calibration Kit
• HP 85033D 3.5 mm Calibration Kit
• HP 85036B 75 Ohm Type-N Calibration Kit
• HP 85039A 75 Ohm Type-F Calibration Kit

Verification Kit

Accurate operation of the analyzer system can be verified by measuring known devices other than the standards used in calibration, and comparing the results with recorded data.
HP 85029B 7 mm Verification Kit
This kit contains traceable precision 7 mm devices used to confirm the sys­tem’s error-corrected measurement uncertainty performance. Also included is verification data on a 3.5 inch disk, together with a hard-copy listing. A system verification procedure is provided with this kit and is also located in the
HP 8702D Installation Guide
2-6
.
Page 27
Reference
Available Accessories

Test Port Return Cables

The following RF cables are used to connect a two-port device between the test ports. These cables provide shielding for high dynamic range measure­ments.
HP 11857D 7 mm Test Port Return Cable Set
This set consists of a pair of test port return cables that can be used in mea­surements of 7 mm devices. They can also be used with connectors other than 7 mm, by using the appropriate precision adapters.
HP 11857B 75 Ohm Type-N Test Port Return Cable Set
This set consists of test port return cables for use with the HP 8702D Option 075.

Adapter Kits

HP 11852B 50 to 75 Ohm Minimum Loss Pad
This device converts impedance from 50 ohms to 75 ohms or from 75 ohms to 50 ohms. It is used to provide a low SWR impedance match between a 75 ohm device under test and the HP 8702D analyzer (without Option 075) or between the 50Ω lightwave source and receiver and the HP 8702D Option 075.
HP 11853A 50 Ohm Type-N Adapter Kit HP 11854A 50 Ohm BNC Adapter Kit HP 11855A 75 Ohm Type-N Adapter Kit HP 11856A 75 Ohm BNC Adapter Kit
These adapter kits contain the connection hardware required for making mea­surements on devices of the indicated connector type.
2-7
Page 28
Reference
Available Accessories

Transistor Test Fixtures

The following Hewlett-Packard transistor test fixture is compatible with the HP 8702D. Additional test fixtures for transistors and other devices are avail­able from Inter-Continental Microwave. To order their catalog, request HP lit­erature number 5091–4254E, or contact Inter–Continental Microwave as follows:
1515 Wyatt Drive Santa Clara, CA 95054-1524 (tel) 408 727-1596 (fax) 408 727-0105
HP 11608A Option 003 Transistor Fixture
This fixture is designed to be user-milled to hold stripline transistors for S­parameter measurements. Option 003 is pre-milled for 0.205 inch diameter disk packages, such as the HP HPAC-200.
2-8
Page 29

System Accessories Available

Reference
Available Accessories
System Cabinet
Plotters and Printers
The HP 85043D system cabinet is designed to rack mount the analyzer in a system configuration. The 132 cm (52 in) system cabinet includes a bookcase, a drawer, and a convenient work surface.
The analyzer is capable of plotting or printing displayed measurement results directly (without the use of an external computer) to a compatible peripheral. The analyzer supports HP-IB, serial, and parallel peripherals. Most Hewlett­Packard desktop printers and plotters are compatible with the analyzer. Some common compatible peripherals are listed here (some are no longer available for purchase, but are listed here for your reference).
These plotters are compatible:
• HP 7440A ColorPro Eight-Pen Color Graphics Plotter
• HP 7470A Two-Pen Graphics Plotter
• HP 7475A Six-Pen Graphics Plotter
• HP 7550A/B High-Speed Eight-Pen Graphics Plotter
These printers are compatible:
• HP DeskJet 1200C (can also be used to plot)
• HP DeskJet 500
• HP C2170A, DeskJet 520
• HP DeskJet 500C
• HP DeskJet 540
• HP DeskJet 550C
• HP DeskJet 560C
• HP DeskJet 600, 660C, 682C, 690C, 850C, 870C, 1600C
• All LaserJets (LaserJet III and above can also be used to plot)
• HP C2621A DeskJet Portable InkJet
• PaintJet 3630A PaintJet Color Graphics Printer
• Epson printers which are compatible with the Epson ESC/P2 printer control language, such as the LQ570, are also supported by the analyzer. Older Epson printers, however, such as the FX-80, will not work with the analyzer.
Mass Storage
The analyzer has the capability of storing instrument states directly to its internal memory, to an internal disk, or to an external disk. The internal
3.5 inch floppy disk can be initialized in both LIF and DOS formats and is capa­ble of reading and writing data in both formats. Using the internal disk drive is
2-9
Page 30
Reference
Available Accessories
the preferred method, but all the capability of previous generation analyzers, to use external disk drives, still exists with the HP 8702D. Most external disks using CS80 protocol are compatible.
The analyzer does not support the LIF-HFS (hierarchy file system) directory format.
CAUTION
HP-IB Cables
Interface Cables
Keyboards
External Monitors
Do not use the older single-sided disks in the analyzer’s internal drive.
An HP-IB cable is required for interfacing the analyzer with a plotter, printer, external disk drive, or computer. The cables available are:
• HP 10833A HP-IB Cable, 1.0 m (3.3 ft.)
• HP 10833B HP-IB Cable, 2.0 m (6.6 ft.)
• HP 10833D HP-IB Cable, 0.5 m (1.6 ft.)
• HP C2951A Centronics (Parallel) Interface Cable, 3.0 m (9.9 ft.)
• HP C2913A RS-232C Interface Cable, 1.2 m (3.9 ft.)
• HP 24542G Serial Interface Cable, 3 m (9.9 ft.)
• HP C2950A Parallel Interface Cable, 2 m (6 ft.)
A keyboard can be connected to the analyzer for data input, such as titling files. The HP C1405B Option ABA keyboard, with the HP part number C1405­60015 adapter, is suitable for this purpose. Or, the analyzer is designed to accept most PC-AT-compatible keyboards with a standard DIN connector. Keyboards with a mini-DIN connector are compatible with the HP part num­ber C1405-60015 adapter.
The analyzer can drive both its internal display and an external monitor simul­taneously. One compatible color monitor is the HP 35741A/B. (It is no longer available for purchase, but is listed here for your reference.)
External Monitor Requirements:
• 60 Hz vertical refresh rate
• 25.5 kHz horizontal refresh rate
• RGB with synchronization on green
• 75 ohm video input impedance
• video amplitude 1 V p-p (0.7 V= white, 0 V= black, –0.3 V= synchronization)
2-10
Page 31

AC Line-Power Cords

Reference
Available Accessories
2-11
Page 32
Reference

Front Panel Features

Front Panel Features
Figure 2-1. HP 8702D Front Panel
1
LINE switch. This switch controls ac power to the analyzer. 1 is on, 0 is off.
2
Display. This shows the measurement data traces, measurement annotation, and softkey labels. The display is divided into specific information areas, illustrated in Figure 2-2.
3
Softkeys. These keys provide access to menus that are shown on the display.
4
STIMULUS function block. The keys in this block allow you to control the analyzer source’s frequency, power, and other stimulus functions.
5
RESPONSE function block. The keys in this block allow you to control the measurement and display functions of the active display channel.
6
ACTIVE CHANNEL keys. The analyzer has two independent display channels.
2-12
Page 33
Reference
Front Panel Features
These keys allow you to select the active channel. Then any function you enter applies to this active channel.
7
The ENTRY block. This block includes the knob, the step (⇑, ⇓) keys, and the number pad. These allow you to enter numerical data and control the markers.
8
INSTRUMENT STATE function block. These keys allow you to control channel-independent system functions such as the following:
• copying, save/recall, and HP-IB controller mode
• limit testing
• external source mode
• tuned receiver mode
• frequency offset mode
• test sequence function
• harmonic measurements (Option 002)
• time domain transform (Option 010)
HP-IB STATUS indicators are also included in this block.
PRESET
9
key. This key returns the instrument to either a known factory preset
state, or a user preset state that can be defined.
10
PORT 1 and PORT 2. These ports output a signal from the source and receive input signals from a device under test. PORT 1 allows you to measure S S
. PORT 2 allows you to measure S21 and S22.
11
11
PROBE POWER connector. This connector (fused inside the instrument) supplies power to an active probe for in-circuit measurements of ac circuits.
12
R CHANNEL connectors. These connectors allow you to apply an input signal to the analyzer’s R channel, for frequency offset mode.
13
Disk drive. This 3.5 inch drive allows you to store and recall instrument states and measurement results for later analysis.
12
and
2-13
Page 34
Reference

Analyzer Display

Analyzer Display
Figure 2-2. Analyzer Display (Single Channel, Cartesian Format)
The analyzer display shows various measurement information:
• the grid where the analyzer plots the measurement data.
• the currently selected measurement parameters.
• the measurement data traces.
1
Stimulus Start Value. This value could be any one of the following:
• the start frequency of the source in frequency domain measurements.
• the start time in CW mode (0 seconds) or time domain measurements.
• the lower power value in power sweep.
When the stimulus is in center/span mode, the center stimulus value is shown in this space.
2-14
Page 35
Reference
Analyzer Display
2
Stimulus Stop Value. This value could be any one of the following:
• the stop frequency of the source in frequency domain measurements.
• the stop time in time domain measurements or CW sweeps.
• the upper limit of a power sweep.
When the stimulus is in center/span mode, the span is shown in this space. The stimulus values can be blanked.
3
Status Notations. This area shows the current status of various functions for the active channel.
4
Active Entry Area. This displays the active function and its current value.
5
Message Area. This displays prompts or error messages.
6
Title. This is a descriptive alpha-numeric string title that you define and enter through an attached keyboard.
7
Active Channel. This is the number of the current active channel, selected with
ACTIVE CHANNEL
the
keys. If dual channel is on with an overlaid display, both
channel 1 and channel 2 appear in this area.
8
Measured Input(s). This shows the S-parameter, input, or ratio of inputs currently measured, as selected using the
MEAS
key. Also indicated in this area
is the current display memory status.
9
Format. This is the display format that you selected using the
10
Scale/Div. This is the scale that you selected using the
SCALE/REF
FORMAT
key, in units
appropriate to the current measurement.
11
Reference Level. This value is the reference line in Cartesian formats or the outer circle in polar formats, whichever you selected using the
SCALE/REF
The reference level is also indicated by a small triangle adjacent to the graticule, at the left for channel 1 and at the right for channel 2.
12
Marker Values. These are the values of the active marker, in units appropriate to the current measurement.
13
Marker Stats, Bandwidth. These are statistical marker values that the analyzer calculates when you access the menus with the
14
Softkey Labels. These menu labels redefine the function of the softkeys that are
MKR FCTN
key.
located to the right of the analyzer display.
15
Pass/Fail. During limit testing, the result will be annunciated as “PASS” if the limits are not exceeded, and “FAIL” if any points exceed the limits.
key.
key.
2-15
Page 36
Reference

Rear Panel Features and Connectors

Rear Panel Features and Connectors
Figure 2-3. HP 8702D Rear Panel
1
Serial number plate.
2
External Monitor. Red, green, and blue video output connectors provide analog red, green, and blue video signals which you can use to drive an external monitor such as the HP 3571A/B or monochrome monitor such as the HP 35731A/B. You can use other analog multi-sync monitors if they are compatible with the analyzer’s 25.5 kHz scan rate and video levels: 1 V p-p,
0.7 V=white, 0 V=black, –0.3 V sync, sync on green.
3
HP-IB connector. This allows you to connect the analyzer to an external controller, compatible peripherals, and other instruments for an automated system.
4
PARALLEL connector. This connector allows the analyzer to output to a peripheral with a parallel input. Also included, is a general purpose input/ output (GPIO) bus that can control eight output bits and read five input bits through test sequencing.
2-16
Page 37
Rear Panel Features and Connectors
5
RS-232 connector. This connector allows the analyzer to output to a peripheral with an RS-232 (serial) input.
6
KEYBOARD input (DIN) connector. This connector allows you to connect an external keyboard. This provides a more convenient means to enter a title for storage files, as well as substitute for the analyzer’s front panel keyboard. The keyboard must be connected to the analyzer before the power is switched on.
7
Power cord receptacle, with fuse.
8
Line voltage selector switch.
9
10 MHZ REFERENCE ADJUST. (Option 1D5)
10
10 MHZ PRECISION REFERENCE OUTPUT. (Option 1D5)
11
EXTERNAL REFERENCE INPUT connector. This allows for a frequency reference signal input that can phaselock the analyzer to an external frequency standard for increased frequency accuracy.
12
AUXILIARY INPUT connector. This allows for a dc or ac voltage input from an external signal source, such as a detector or function generator, which you can then measure, using the S-parameter menu.
Reference
13
EXTERNAL AM connector. This allows for an external analog signal input that is applied to the ALC circuitry of the analyzer’s source. This input analog signal amplitude modulates the RF output signal.
14
EXTERNAL TRIGGER connector. This allows connection of an external negative-going TTL-compatible signal that will trigger a measurement sweep. The trigger can be set to external through softkey functions.
15
TEST SEQUENCE. Outputs a TTL signal that can be programmed in a test sequence to be high or low, or pulse (10 µseconds) high or low at the end of a sweep for robotic part handler interface.
16
LIMIT TEST. Outputs a TTL signal of the limit test results as follows:
• Pass: TTL high
• Fail: TTL low
17
BIAS INPUTS AND FUSES. These connector bias devices connected to port 1 and port 2. The fuses (1 A, 125 V) protect the port 1 and port 2 bias lines.
18
TEST SET INTERCONNECT. This allows you to connect an HP 8702D Option 011 analyzer to an HP 85046A/B or 85047A S-parameter test set using the interconnect cable supplied with the test set. The S-parameter test set is then fully controlled by the analyzer.
2-17
Page 38
Reference
Rear Panel Features and Connectors
Figure 2-4. Rear Panel Connectors
2-18
Page 39

Reference Documents

Reference
Reference Documents
Hewlett-Packard Company, HP publication number 5956-4363
Hewlett-Packard Company,
Seminar Papers
Hewlett-Packard Company,
HP 8720 Network Analyzer
5091-1942E
Hewlett-Packard Company,
work Analyzer
, HP publication number 5091-8804E
, Product Note 8753-2A, HP publication number 5952-2771
Simplify Your Amplifier and Mixer Testing
RF and Microwave Device Test for the ’90s -
Testing Amplifiers and Active Devices with the
, Product Note 8720-1, HP publication number
Mixer Measurements using the HP 8753 Net-
,

General Measurement and Calibration Techniques

Blacka, Robert J., Reprint from March/April 1989
Montgomery, David,
Computer Design
from
Rytting, Doug, Hewlett-Packard RF and Microwave Measurement Symposium paper, HP pub­lication number 5954-8378, June 1987
Rytting, Doug,
work Analyzers
sium paper, 1991
TDR Gated Measurements of Stripline Terminations
Microwave Product Digest
Borrowing RF Techniques for Digital Design
, HP publication number 5952-9335, May 1982
Advances in Microwave Error Correction Techniques
Improved RF Hardware and Calibration Methods for Net-
Hewlett-Packard RF and Microwave Measurement Sympo-
, HP publication number 5952-0359,
, Reprint
,
,

Fixtures and Non-coaxial Measurements

Hewlett-Packard Company,
Coaxial Measurements
5091-3645E, February 1992
Applying the HP 8510 TRL Calibration for Non-
, Product Note 8510-8A, HP publication number
2-19
Page 40
Reference
Reference Documents
Hewlett-Packard Company,
Network Analyzers and Inter-Continental Microwave Test Fixtures
uct Note 8510-17, HP publication number 5091-5674E, September 1992
Hewlett-Packard Company,
Using TRL* Calibration
5091-1943E, August 1991
Hewlett-Packard Company,
Modular Microcircuit Package
ber 5952-1907, May 1990
“Test Fixtures and Calibration Standards,“ Inter-Continental Microwave Prod­uct Catalog, HP publication number 5091-4254E
Curran, Jim, and Microwave Measurement Symposium paper, HP publication number 5954-8346, September 1986
Curran, Jim, Packard Semiconductor Test Symposium paper
Elmore, Glenn and Louis Salz, “Quality Microwave Measurement of Packaged Active Devices,”
Techniques for Fixtured Devices
ber 5952-2766, June 1990
Network Analysis of Fixtured Devices
TRL Calibration for Non-Coaxial Measurements
Hewlett-Packard Journal
Measuring Chip Capacitors with the HP 8520C
, Prod-
In-Fixture Microstrip Device Measurement
, Product Note 8720-2, HP publication number
Calibration and Modeling Using the HP 83040
, Product Note 83040-2, HP publication num-
, Hewlett-Packard RF
, Hewlett-
, February 1987,
, HP 8510/8720 News, HP publication num-
Measurement

On-Wafer Measurements

Hewlett-Packard Company,
Network Analyzer and Cascade Microtech Wafer Probes
8510-6, HP publication number 5954-1579
Barr, J.T., T. Burcham, A.C. Davidson, E.W. Strid,
Wafer Probing Calibration Techniques
Measurement Symposium paper, 1991
Lautzenhiser, S., A. Davidson, D. Jones,
Via LRM Calibration
number 5952-1286, January 1990,
erations
1993
2-20
, HP 8510/8720 News, HP publication number 5091-6837, February
On-Wafer Measurements Using the HP 8510
Advancements in On-
, Hewlett-Packard RF and Microwave
Improve Accuracy of On-Wafer Tests
, Reprinting from
Microwaves and RF
On-Wafer Calibration: Practical Consid-
, Product Note
, HP publication
Page 41

Location of Softkeys

Table 2-1. Location of Softkeys (1 of 24)
Softkey Menu Location
∆ MODE MENU Front-panel access key: MARKER ∆ MODE OFF Front-panel access key: MARKER ∆ REF= 1 Front-panel access key: MARKER ∆ REF= 2 Front-panel access key: MARKER ∆ REF= 3 Front-panel access key: MARKER ∆ REF= 4 Front-panel access key: MARKER
Reference
Location of Softkeys
∆ REF= 5 Front-panel access key: MARKER ∆ REF= ∆ FIXED MKR Front-panel access key: MARKER
1/S Front-panel access key: MEAS
2.4 mm Front-panel access key: CAL
2.92* Front-panel access key: CAL
2.92 mm Front-panel access key: CAL
3.5mmC Front-panel access key: CAL
3.5mmD Front-panel access key: CAL 7 mm Front-panel access key: CAL A Front-panel access key: MEAS A/B Front-panel access key: MEAS A/R Front-panel access key: MEAS ACTIVE ENTRY Front-panel access key: DISPLAY ACTIVE MKR MAGNITUDE Front-panel access key: DISPLAY ADD Front-panel access key: CAL and MENU
2-21
Page 42
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (2 of 24)
Softkey Menu Location
ADDRESS: 8702 Front-panel access key: LOCAL ADDRESS: CONTROLLER Front-panel access key: LOCAL ADDRESS: DISK Front-panel access key: LOCAL ADDRESS: DISK Front-panel access key: SAVE/RECALL ADDRESS: P MTR/HPIB Front-panel access key: LOCAL ADJUST DISPLAY Front-panel access key: DISPLAY ALIAS SPAN on OFF Front-panel access key: SYSTEM ALL OFF Front-panel access key: MARKER ALL SEGS SWEEP Front-panel access key: MENU ALTERNATE A AND B Front-panel access key: CAL AMPLITUDE Front-panel access key: SYSTEM AMPLITUDE OFFSET Front-panel access key: SYSTEM ANALOG IN Aux Input Front-panel access key: MEAS ARBITRARY IMPEDANCE Front-panel access key: CAL ASCII Front-panel access key: SAVE/RECALL ASSERT SRQ Front-panel access key: SEQ AUTO-FEED ON off Front-panel access key: COPY AUTO SCALE Front-panel access key: SCALE REF AVERAGING FACTOR Front-panel access key: AVG AVERAGING on OFF Front-panel access key: AVG AVERAGING RESTART Front-panel access key: AVG B Front-panel access key: MEAS B/R Front-panel access key: MEAS BACKGROUND INTENSITY Front-panel access key: DISPLAY BACKSPACE Front-panel access key: CAL BANDPASS Front-panel access key: SYSTEM BEEP DONE ON off Front-panel access key: DISPLAY
2-22
Page 43
Table 2-1. Location of Softkeys (3 of 24)
Softkey Menu Location
BEEP FAIL on OFF Front-panel access key: SYSTEM BEEP WARN on OFF Front-panel access key: DISPLAY BLANK DISPLAY Front-panel access key: DISPLAY BRIGHTNESS Front-panel access key: DISPLAY C0 Front-panel access key: CAL C1 Front-panel access key: CAL C2 Front-panel access key: CAL C3 Front-panel access key: CAL CAL FACTOR Front-panel access key: CAL CAL FACTOR SENSOR A Front-panel access key: CAL CAL FACTOR SENSOR B Front-panel access key: CAL CAL KITS & STDS Front-panel access key: CAL CAL KIT: 2.4 mm Front-panel access key: CAL
Reference
Location of Softkeys
CAL KIT: 2.92* Front-panel access key: CAL CAL KIT: 2.92 mm Front-panel access key: CAL CAL KIT: 3.5mmC Front-panel access key: CAL CAL KIT: 3.5mmD Front-panel access key: CAL CAL KIT: 7mm Front-panel access key: CAL CAL KIT: N 50 Front-panel access key: CAL CAL KIT: N 75 Front-panel access key: CAL CAL KIT: TRL 3.5 mm Front-panel access key: CAL CALIBRATE MENU Front-panel access key: CAL CALIBRATE: NONE Front-panel access key: CAL CAL Z0 LINE Z0 Front-panel access key: CAL CENTER Front-panel access key: MENU CENTER Front-panel access key: SYSTEM CH1 DATA [ ] Front-panel access key: COPY
2-23
Page 44
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (4 of 24)
Softkey Menu Location
CH1 DATA LIMIT LN Front-panel access key: DISPLAY CH1 MEM Front-panel access key: DISPLAY CH1 MEM [ ] Front-panel access key: COPY CH2 DATA [ ] Front-panel access key: COPY CH2 DATA LIMIT LN Front-panel access key: DISPLAY CH2 MEM [ ] Front-panel access key: COPY CH2 MEM REF LINE Front-panel access key: DISPLAY CHAN POWER [COUPLED] Front-panel access key: MENU CHAN POWER [UNCOUPLED] Front-panel access key: MENU CHOP A AND B Front-panel access key: CAL CLEAR BIT Front-panel access key: SEQ CLEAR LIST Front-panel access key: CAL and MENU CLEAR SEQUENCE Front-panel access key: SEQ COAX Front-panel access key: CAL COAXIAL DELAY Front-panel access key: SCALE REF COLOR Front-panel access key: DISPLAY COMPONENT ANALYZER Front-panel access key: SYSTEM CONFIGURE EXT DISK Front-panel access key: SAVE/RECALL CONTINUE SEQUENCE Front-panel access key: SEQ CONTINUOUS Front-panel access key: MENU and MARKER FCTN CONVERSION [ ] Front-panel access key: MEAS CONVERSION OFF Front-panel access key: MEAS CORRECTION on OFF Front-panel access key: CAL COUPLED Front-panel access key: MARKER FCTN COUPLED CH ON off Front-panel access key: MENU CW FREQ Front-panel access key: MENU CW TIME Front-panel access key: MENU
2-24
Page 45
Table 2-1. Location of Softkeys (5 of 24)
Softkey Menu Location
D2/D1 TO D2 on OFF Front-panel access key: DISPLAY DATA AND MEMORY Front-panel access key: DISPLAY DATA ARRAY on off Front-panel access key: SAVE/RECALL DATA/MEM Front-panel access key: DISPLAY DATA – MEM Front-panel access key: DISPLAY
Reference
Location of Softkeys
DATA
MEMORY
DATA ONLY on off Front-panel access key: SAVE/RECALL DECISION MAKING Front-panel access key: SEQ DECR LOOP COUNTER Front-panel access key: SEQ DEFAULT COLORS Front-panel access key: DISPLAY DEFAULT PLOT SETUP Front-panel access key: COPY DEFAULT PRNT SETUP Front-panel access key: COPY DEFAULT RCVR COEFF Front-panel access key: CAL DEFAULT SRC COEFF Front-panel access key: CAL DEFAULT STANDARDS Front-panel access key: CAL DEFINE DISK-SAVE Front-panel access key: SAVE/RECALL DEFINE PLOT Front-panel access key: COPY DEFINE PRINT Front-panel access key: COPY DEFINE STANDARD Front-panel access key: CAL DELAY Front-panel access key: FORMAT DELAY/THRU Front-panel access key: CAL DELETE Front-panel access key: CAL, MENU and SYSTEM DELETE ALL FILES Front-panel access key: SAVE/RECALL
Front-panel access key: DISPLAY
DELETE FILE Front-panel access key: SAVE/RECALL DELTA LIMITS Front-panel access key: SYSTEM DEMOD: OFF Front-panel access key: SYSTEM DIRECTORY SIZE (LIF) Front-panel access key: SAVE/RECALL
2-25
Page 46
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (6 of 24)
Softkey Menu Location
DISCRETE Front-panel access key: MARKER FCTN DISK Front-panel access key: LOCAL DISK UNIT NUMBER Front-panel access key: SAVE/RECALL DISPLAY: DATA Front-panel access key: DISPLAY DISP MKRS ON off Front-panel access key: MARKER FCTN DO BOTH FWD + REV Front-panel access key: CAL DONE 1-PORT CAL Front-panel access key: CAL DONE 2-PORT CAL Front-panel access key: CAL DONE LINE/MATCH Front-panel access key: CAL DONE RESPONSE Front-panel access key: CAL DONE RESP ISOL’N CAL Front-panel access key: CAL DONE SEQ MODIFY Front-panel access key: SEQ DONE TRL/LRM Front-panel access key: CAL DO SEQUENCE Front-panel access key: SEQ DOS Front-panel access key: SAVE/RECALL DOWN CONVERTER Front-panel access key: SYSTEM DUAL CHAN on OFF Front-panel access key: DISPLAY DUPLICATE SEQUENCE Front-panel access key: SEQ EACH SWEEP Front-panel access key: CAL EDIT Front-panel access key: CAL, MENU and SYSTEM EDIT LIMIT LINE Front-panel access key: SYSTEM EDIT LIST Front-panel access key: MENU ELECTRICAL DELAY Front-panel access key: SCALE REF EMIT BEEP Front-panel access key: SEQ END OF LABEL Front-panel access key: DISPLAY END SWEEP HIGH PULSE Front-panel access key: SEQ END SWEEP LOW PULSE Front-panel access key: SEQ
2-26
Page 47
Table 2-1. Location of Softkeys (7 of 24)
Softkey Menu Location
ENTER RCVR COEFF MENU Front-panel access key: CAL ENTER SRC COEFF MENU Front-panel access key: CAL ERASE TITLE Front-panel access key: CAL ERASE TITLE Front-panel access key: DISPLAY ERASE TITLE Front-panel access key: SAVE/RECALL EXT SOURCE AUTO Front-panel access key: SYSTEM EXT SOURCE MANUAL Front-panel access key: SYSTEM EXT TRIG ON POINT Front-panel access key: MENU EXT TRIG ON SWEEP Front-panel access key: MENU EXTENSION INPUT A Front-panel access key: CAL EXTENSION INPUT B Front-panel access key: CAL EXTENSION PORT 1 Front-panel access key: CAL EXTENSION PORT 2 Front-panel access key: CAL
Reference
Location of Softkeys
EXTENSIONS on OFF Front-panel access key: CAL EXTERNAL DISK Front-panel access key: SAVE/RECALL FACTORY Front-panel access key: PRESET FILENAME FILE0 Front-panel access key: SAVE/RECALL FILE UTILITIES Front-panel access key: SAVE/RECALL FIXED Front-panel access key: CAL FIXED MKR AUX VALUE Front-panel access key: MARKER FIXED MKR POSITION Front-panel access key: MARKER FIXED MKR STIMULUS Front-panel access key: MARKER FIXED MKR VALUE Front-panel access key: MARKER FLAT LINE Front-panel access key: SYSTEM FORM FEED Front-panel access key: DISPLAY FORMAT ARY on off Front-panel access key: SAVE/RECALL FORMAT DISK Front-panel access key: SAVE/RECALL
2-27
Page 48
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (8 of 24)
Softkey Menu Location
FORMAT: DOS Front-panel access key: SAVE/RECALL FORMAT EXT DISK Front-panel access key: SAVE/RECALL FORMAT INT DISK Front-panel access key: SAVE/RECALL FORMAT INT MEMORY Front-panel access key: SAVE/RECALL FORMAT: LIF Front-panel access key: SAVE/RECALL FORWARD: Front-panel access key: CAL FREQ OFFS MENU Front-panel access key: SYSTEM FREQ OFFS on OFF Front-panel access key: SYSTEM FREQUENCY Front-panel access key: CAL FREQUENCY BLANK Front-panel access key: DISPLAY FREQUENCY: CW Front-panel access key: SYSTEM FREQUENCY: SWEEP Front-panel access key: SYSTEM FULL 2-PORT Front-panel access key: CAL FULL PAGE Front-panel access key: COPY FWD ISOL’N ISOL’N STD Front-panel access key: CAL FWD MATCH Front-panel access key: CAL FWD MATCH THRU Front-panel access key: CAL FWD TRANS Front-panel access key: CAL FWD TRANS THRU Front-panel access key: CAL G+jB MKR Front-panel access key: MARKER FCTN GATE: CENTER Front-panel access key: SYSTEM GATE on OFF Front-panel access key: SYSTEM GATE SHAPE Front-panel access key: SYSTEM GATE SHAPE MAXIMUM Front-panel access key: SYSTEM GATE SHAPE MINIMUM Front-panel access key: SYSTEM GATE SHAPE NORMAL Front-panel access key: SYSTEM GATE SHAPE WIDE Front-panel access key: SYSTEM
2-28
Page 49
Table 2-1. Location of Softkeys (9 of 24)
Softkey Menu Location
GATE: SPAN Front-panel access key: SYSTEM GATE: START Front-panel access key: SYSTEM GATE: STOP Front-panel access key: SYSTEM GOSUB SEQUENCE Front-panel access key: SEQ GRAPHICS on off Front-panel access key: SAVE/RECALL GRATICULE [ ] Front-panel access key: COPY GRATICULE TEXT Front-panel access key: DISPLAY GUIDED SETUP Front-panel access key: PRESET and SYSTEM HARMONIC MEAS Front-panel access key: SYSTEM HARMONIC OFF Front-panel access key: SYSTEM HARMONIC SECOND Front-panel access key: SYSTEM HARMONIC THIRD Front-panel access key: SYSTEM HOLD Front-panel access key: MENU
Reference
Location of Softkeys
HP-IB DIAG on OFF Front-panel access key: LOCAL I/O FWD Front-panel access key: SEQ I/O REV Front-panel access key: SEQ IF BIT H Front-panel access key: SEQ IF BIT L Front-panel access key: SEQ IF BW [ ] Front-panel access key: AVG and SYSTEM IF LIMIT TEST FAIL Front-panel access key: SEQ IF LIMIT TEST PASS Front-panel access key: SEQ IF LOOP COUNTER=0 Front-panel access key: SEQ IF LOOP COUNTER < > 0 Front-panel access key: SEQ IMAGINARY Front-panel access key: FORMAT INCR LOOP COUNTER Front-panel access key: SEQ INDEX of REFRACTION Front-panel access key: CAL and SYSTEM INPUT PORT CAL Front-panel access key: CAL
2-29
Page 50
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (10 of 24)
Softkey Menu Location
INPUT PORTS Front-panel access key: MEAS INSTRUMENT MODE Front-panel access key: SYSTEM INTENSITY Front-panel access key: DISPLAY INTERNAL DISK Front-panel access key: SAVE/RECALL INTERNAL MEMORY Front-panel access key: SAVE/RECALL INTERPOL on OFF Front-panel access key: CAL I/O REV Front-panel access key: SEQ I/S Front-panel access key: MEAS ISOLATION Front-panel access key: CAL ISOLATION DONE Front-panel access key: CAL ISOL’N STD Front-panel access key: CAL ISTATE CONTENTS Front-panel access key: SAVE/RECALL KIT DONE (MODIFIED) Front-panel access key: CAL LABEL CLASS Front-panel access key: CAL LABEL CLASS DONE Front-panel access key: CAL LABEL KIT Front-panel access key: CAL LABEL STD Front-panel access key: CAL LEFT LOWER Front-panel access key: COPY LEFT UPPER Front-panel access key: COPY LIF Front-panel access key: SAVE/RECALL LIGHTWAVE PARAMETERS Front-panel access key: MEAS LIMIT LINE OFFSETS Front-panel access key: SYSTEM LIMIT LINE on OFF Front-panel access key: SYSTEM LIMIT MENU Front-panel access key: SYSTEM LIMIT TEST on OFF Front-panel access key: SYSTEM LIMIT TEST RESULT Front-panel access key: DISPLAY LIMIT TYPE Front-panel access key: SYSTEM
2-30
Page 51
Table 2-1. Location of Softkeys (11 of 24)
Softkey Menu Location
LINE/MATCH Front-panel access key: CAL LIN FREQ Front-panel access key: MENU LIN MAG Front-panel access key: FORMAT LIN MKR Front-panel access key: MARKER FCTN LINE/MATCH Front-panel access key: CAL LINE Z0: Front-panel access key: CAL LINE TYPE DATA Front-panel access key: COPY LINE TYPE MEMORY Front-panel access key: COPY LIST Front-panel access key: COPY LIST FREQ Front-panel access key: MENU LIST VALUES Front-panel access key: COPY LN/MATCH 1 LOAD Front-panel access key: CAL LN/MATCH 2 LOAD Front-panel access key: CAL
Reference
Location of Softkeys
LO CONTROL on OFF Front-panel access key: SYSTEM LO MENU Front-panel access key: SYSTEM LO SOURCE ADDRESS Front-panel access key: SYSTEM LOAD Front-panel access key: CAL LOAD RCVR DISK MENU Front-panel access key: CAL LOAD SEQ FROM DISK Front-panel access key: SEQ LOAD SRC DISK MENU Front-panel access key: CAL LOG FREQ Front-panel access key: MENU LOG MAG Front-panel access key: FORMAT LOG MKR Front-panel access key: MARKER FCTN LOOP COUNTER Front-panel access key: DISPLAY and SEQ LOSS Front-panel access key: CAL LOSS/SENSR LISTS Front-panel access key: CAL LOWER LIMIT Front-panel access key: SYSTEM
2-31
Page 52
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (12 of 24)
Softkey Menu Location
LOW PASS IMPULSE Front-panel access key: SYSTEM LOW PASS STEP Front-panel access key: SYSTEM MANUAL TRG ON POINT Front-panel access key: MENU
MARKER
AMP. OFS.
Front-panel access key: SYSTEM
MARKER CENTER MARKER CW MARKER DELAY MARKER MIDDLE MARKER REFERENCE
MARKER
MARKER START MARKER STIMULUS MARKER STOP MARKER 1, 2, 3, 4, and 5 Front-panel access key: MARKER MARKER MODE MENU Front-panel access key: MARKER FCTN MARKERS: CONTINUOUS Front-panel access key: MARKER FCTN
MARKERS: COUPLED Front-panel access key: MARKER FCTN MARKERS: DISCRETE Front-panel access key: MARKER FCTN MARKERS: UNCOUPLED Front-panel access key: MARKER FCTN MAX Front-panel access key: MARKER FCTN MAXIMUM FREQUENCY Front-panel access key: CAL
SPAN
Front-panel access key: MARKER FCTN Front-panel access key: SEQ Front-panel access key: MARKER FCTN Front-panel access key: SYSTEM Front-panel access key: MARKER FCTN and SCALE REF Front-panel access key: MARKER FCTN Front-panel access key: MARKER FCTN Front-panel access key: SYSTEM Front-panel access key: MARKER FCTN
MEASURE RESTART Front-panel access key: MENU MEMORY Front-panel access key: DISPLAY MIDDLE VALUE Front-panel access key: SYSTEM MIN Front-panel access key: MARKER FCTN MINIMUM FREQUENCY Front-panel access key: CAL MKR SEARCH [ ] Front-panel access key: MARKER FCTN
2-32
Page 53
Table 2-1. Location of Softkeys (13 of 24)
Softkey Menu Location
MKR ZERO Front-panel access key: MARKER MODIFY [ ] Front-panel access key: CAL MODIFY COLORS Front-panel access key: DISPLAY MODIFY STANDARDS Front-panel access key: CAL MODIFY THRU/RCVR Front-panel access key: CAL N 50 Front-panel access key: CAL N 75 Front-panel access key: CAL NEW SEQ/MODIFY SEQ Front-panel access key: SEQ NEWLINE Front-panel access key: DISPLAY NUMBER OF GROUPS Front-panel access key: MENU NUMBER of POINTS Front-panel access key: SYSTEM NUMBER OF POINTS Front-panel access key: MENU NUMBER OF READINGS Front-panel access key: CAL
Reference
Location of Softkeys
OFFSET DELAY Front-panel access key: CAL OFFSET LOSS Front-panel access key: CAL OFFSET Z0 Front-panel access key: CAL OMIT ISOLATION Front-panel access key: CAL ONE-PATH 2-PORT Front-panel access key: CAL ONE SWEEP Front-panel access key: CAL OP PARMS (MKRS etc.) Front-panel access key: COPY OPEN Front-panel access key: CAL OPTICAL STANDARDS Front-panel access key: CAL P MTR/HPIB TO TITLE Front-panel access key: SEQ PAGE Front-panel access key: COPY PARALLEL Front-panel access key: LOCAL PARALLEL [ ] Front-panel access key: LOCAL PARALLEL OUT ALL Front-panel access key: SEQ
2-33
Page 54
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (14 of 24)
Softkey Menu Location
PARALL IN BIT NUMBER Front-panel access key: SEQ PARALL IN IF BIT H Front-panel access key: SEQ PARALL IN IF BIT L Front-panel access key: SEQ PAUSE Front-panel access key: SEQ PEN NUM DATA Front-panel access key: COPY PEN NUM GRATICULE Front-panel access key: COPY PEN NUM MARKER Front-panel access key: COPY PEN NUM MEMORY Front-panel access key: COPY PEN NUM TEXT Front-panel access key: COPY PERIPHERAL HPIB ADDR Front-panel access key: SEQ PHASE Front-panel access key: FORMAT and SYSTEM PHASE OFFSET Front-panel access key: SCALE REF PLOT Front-panel access key: COPY and SYSTEM PLOT DATA ON off Front-panel access key: COPY PLOT GRAT ON off Front-panel access key: COPY PLOT MEM ON off Front-panel access key: COPY PLOT MKR ON off Front-panel access key: COPY PLOT SPEED [ ] Front-panel access key: COPY PLOT TEXT ON off Front-panel access key: COPY PLOTTER BAUD RATE Front-panel access key: LOCAL PLOTTER FORM FEED Front-panel access key: COPY PLOTTER PORT Front-panel access key: LOCAL PLTR PORT HPIB Front-panel access key: LOCAL PLTR TYPE [ ] Front-panel access key: LOCAL POLAR Front-panel access key: FORMAT POLAR MKR MENU Front-panel access key: MARKER FCTN PORT EXTENSIONS Front-panel access key: CAL
2-34
Page 55
Table 2-1. Location of Softkeys (15 of 24)
Softkey Menu Location
PORT POWER [COUPLED] Front-panel access key: MENU PORT POWER [UNCOUPLED] Front-panel access key: MENU POWER Front-panel access key: MENU and SYSTEM POWER: FIXED Front-panel access key: SYSTEM POWER LOSS Front-panel access key: CAL POWER MTR: [ ] Front-panel access key: LOCAL POWER RANGES Front-panel access key: MENU POWER: SWEEP Front-panel access key: MENU and SYSTEM PRESET: FACTORY Front-panel access key: PRESET PRESET: USER Front-panel access key: PRESET PRINT: COLOR Front-panel access key: COPY PRINT COLORS Front-panel access key: COPY PRINT: MONOCHROME Front-panel access key: COPY
Reference
Location of Softkeys
PRINT ALL MONOCHROME Front-panel access key: COPY PRINT MONOCHROME Front-panel access key: COPY and SYSTEM PRINT SEQUENCE Front-panel access key: SEQ PRINTER BAUD RATE Front-panel access key: LOCAL PRINTER FORM FEED Front-panel access key: COPY PRINTER PORT Front-panel access key: LOCAL
PRm = display annotation Power range is in manual mode. PRNTR PORT HPIB Front-panel access key: LOCAL PRNTR TYPE [ ] Front-panel access key: LOCAL PWR LOSS on OFF Front-panel access key: CAL PWR RANGE AUTO man Front-panel access key: MENU PWRMTR CAL [ ] Front-panel access key: CAL PWRMTR CAL OFF Front-panel access key: CAL R Front-panel access key: MEAS
2-35
Page 56
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (16 of 24)
Softkey Menu Location
R+jX MKR Front-panel access key: MARKER FCTN RANGE 0 –15 TO +10 Front-panel access key: MENU RANGE 1 –25 TO 0 Front-panel access key: MENU RANGE 2 –35 TO –10 Front-panel access key: MENU RANGE 3 –45 TO –20 Front-panel access key: MENU RANGE 4 –55 TO –30 Front-panel access key: MENU RANGE 5 –65 TO –40 Front-panel access key: MENU RANGE 6 –75 TO –50 Front-panel access key: MENU RANGE 7 –85 TO –60 Front-panel access key: MENU RAW ARRAY on off Front-panel access key: SAVE/RECALL Re/Im MKR Front-panel access key: MARKER FCTN REAL Front-panel access key: FORMAT RECALL COLORS Front-panel access key: DISPLAY RECALL KEYS MENU Front-panel access key: SAVE/RECALL RECALL KEYS on OFF Front-panel access key: SAVE/RECALL RECALL REG1 – 7 Front-panel access key: SAVE/RECALL RECALL STATE Front-panel access key: SAVE/RECALL RECEIVER STANDARDS Front-panel access key: CAL REFERENCE POSITION Front-panel access key: SCALE REF REFERENCE VALUE Front-panel access key: SCALE REF Refl: E S11 FWD Front-panel access key: MEAS Refl: E S22 (B/R) Front-panel access key: MEAS Refl: E S22 REV Front-panel access key: MEAS
Refl: O (PORT 1→2) REFLECT Front-panel access key: CAL REFLECT’N Front-panel access key: CAL RENAME FILE Front-panel access key: SAVE/RECALL
2-36
Front-panel access key: MEAS
Page 57
Table 2-1. Location of Softkeys (17 of 24)
Softkey Menu Location
RE-SAVE STATE Front-panel access key: SAVE/RECALL RESET COLOR Front-panel access key: DISPLAY RESPONSE Front-panel access key: CAL RESPONSE & ISOL’N Front-panel access key: CAL RESPONSE & MATCH Front-panel access key: CAL RESTORE DISPLAY Front-panel access key: COPY RESUME CAL SEQUENCE Front-panel access key: CAL REV ISOL’N ISOL’N STD Front-panel access key: CAL REV MATCH Front-panel access key: CAL REV MATCH THRU Front-panel access key: CAL REV TRANS Front-panel access key: CAL REV TRANS THRU Front-panel access key: CAL REVERSE: Front-panel access key: CAL
Reference
Location of Softkeys
RF > LO Front-panel access key: SYSTEM RF < LO Front-panel access key: SYSTEM RIGHT LOWER Front-panel access key: COPY RIGHT UPPER Front-panel access key: COPY ROUND SECONDS Front-panel access key: SYSTEM S11 1-PORT Front-panel access key: CAL S11 REFL OPEN Front-panel access key: CAL S11A Front-panel access key: CAL S11B Front-panel access key: CAL S11C Front-panel access key: CAL S22 1-PORT Front-panel access key: CAL S22 REFL OPEN Front-panel access key: CAL S22A Front-panel access key: CAL S22B Front-panel access key: CAL
2-37
Page 58
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (18 of 24)
Softkey Menu Location
S22C Front-panel access key: CAL SAVE/RECALL MENU Front-panel access key: SAVE/RECALL SAVE COLORS Front-panel access key: DISPLAY SAVE RCVR COEFF Front-panel access key: CAL SAVE SRC COEFF Front-panel access key: CAL SAVE STANDARDS Front-panel access key: CAL SAVE STATE Front-panel access key: SAVE/RECALL SAVE USER KIT Front-panel access key: CAL SAVE USING BINARY Front-panel access key: SAVE/RECALL SCALE/DIV Front-panel access key: SCALE REF SCALE PLOT [ ] Front-panel access key: COPY SEARCH LEFT Front-panel access key: MARKER FC TN SEARCH: MAX Front-panel access key: MARK ER FCTN SEARCH: MIN Front-panel access key: MA RKER FCTN SEARCH: OFF Front-panel access key: MARKER FCTN SEARCH RIGHT Front-panel access key: MA RKER FCTN SECOND Front-panel access key: SYSTEM SEGMENT Front-panel access key: CAL, MENU and SYSTEM SEGMENT: CENTER Front-panel access key: MENU SEGMENT: SPAN Front-panel access key: MENU SEGMENT: START Front-panel access key: MENU SEGMENT: STOP Front-panel access key: MENU SEL QUAD [ ] Front-panel access key: COPY SELECT CAL KEY Front-panel access key: CAL SELECT CAL KIT Front-panel access key: CAL SELECT DISK Front-panel access key: SAVE/RECALL SELECT LETTER Front-panel access key: CAL, DISPLAY and SAVE/RECALL
2-38
Page 59
Table 2-1. Location of Softkeys (19 of 24)
Softkey Menu Location
SEQUENCE 1 SEQ1 Front-panel access key: SEQ SEQUENCE 2 SEQ2 Front-panel access key: SEQ SEQUENCE 3 SEQ3 Front-panel access key: SEQ SEQUENCE 4 SEQ4 Front-panel access key: SEQ SEQUENCE 5 SEQ5 Front-panel access key: SEQ SEQUENCE 6 SEQ6 Front-panel access key: SEQ SERIAL Front-panel access key: LOCAL SERVICE MENU Front-panel access key: SYSTEM SET ADDRESSES Front-panel access key: LOCAL SET BIT Front-panel access key: SEQ SET CLOCK Front-panel access key: SYSTEM SET DAY Front-panel access key: SYSTEM SET FREQ LOW PASS Front-panel access key: SYSTEM
Reference
Location of Softkeys
SET HOUR Front-panel access key: SYSTEM SET MINUTES Front-panel access key: SYSTEM SET MONTH Front-panel access key: SYSTEM SET REF Front-panel access key: CAL SET YEAR Front-panel access key: SYSTEM SET Z0 Front-panel access key: CAL SHORT Front-panel access key: CAL SHOW MENUS Front-panel access key: SEQ SINGLE Front-panel access key: MENU SINGLE POINT Front-panel access key: SYSTEM SINGLE SEG SWEEP Front-panel access key: MENU SLIDING OFFSET Front-panel access key: CAL SLOPE Front-panel access key: MENU SLOPE ON off Front-panel access key: MENU
2-39
Page 60
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (20 of 24)
Softkey Menu Location
SLOPING LINE Front-panel access key: SYSTEM SMITH CHART Front-panel access key: FORMAT SMITH MKR MENU Front-panel access key: MARKER FCTN SMOOTHING APERTURE Front-panel access key: AVG SMOOTHING on OFF Front-panel access key: AVG SOURCE PWR ON off Front-panel access key: MENU SOURCE STANDARDS Front-panel access key: CAL SPAN Front-panel access key: MENU SPAN Front-panel access key: SYSTEM S PARAMETERS Front-panel access key: MEAS SPECIAL FUNCTIONS Front-panel access key: SEQ SPECIFY CLASS Front-panel access key: CAL SPECIFY CLASS DONE Front-panel access key: CAL SPECIFY GATE Front-panel access key: SYSTEM SPECIFY OFFSET Front-panel access key: CAL SPECIFY TRL THRU Front-panel access key: CAL SPLIT DISP ON off Front-panel access key: DISPLAY SP MKRS ON off Front-panel access key: MARKER FCTN STANDARDS DONE Front-panel access key: CAL START Front-panel access key: MENU START FREQUENCY Front-panel access key: SYSTEM STATS on OFF Front-panel access key: MARKER FCTN STD DONE (DEFINED) Front-panel access key: CAL STD OFFSET DONE Front-panel access key: CAL STD TYPE:LOAD Front-panel access key: CAL STD TYPE: OPEN Front-panel access key: CAL STD TYPE: SHORT Front-panel access key: CAL
2-40
Page 61
Table 2-1. Location of Softkeys (21 of 24)
Softkey Menu Location
STEP SIZE Front-panel access key: MENU STIMULUS VALUE Front-panel access key: SYSTEM STIMULUS OFFSET Front-panel access key: SYSTEM STOP Front-panel access key: MENU STOP FREQUENCY Front-panel access key: SYSTEM STORE SEQ TO DISK Front-panel access key: SEQ SWEEP TIME [ ] Front-panel access key: MENU and SYSTEM SWEEP TYPE MENU Front-panel access key: MENU SWR Front-panel access key: FORMAT SYSTEM CONTROLLER Front-panel access key: LOCAL SYSTEM Z0 Front-panel access key: CAL TAKE CAL SWEEP Front-panel access key: CAL TALKER/LISTENER Front-panel access key: LOCAL
Reference
Location of Softkeys
TARGET Front-panel access key: MARKER FCTN TERMINAL IMPEDANCE Front-panel access key: CAL TESTPORT 1 2 Front-panel access key: MEAS TESTSET I/O FWD Front-panel access key: SEQ TESTSET I/O REV Front-panel access key: SEQ TESTSET SW CONT hld Front-panel access key: CAL TEXT Front-panel access key: DISPLAY TEXT [ ] Front-panel access key: COPY THIRD Front-panel access key: SYSTEM THRU Front-panel access key: CAL THRU THRU Front-panel access key: CAL TIME STAMP ON off Front-panel access key: SYSTEM TINT Front-panel access key: DISPLAY TITLE Front-panel access key: DISPLAY
2-41
Page 62
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (22 of 24)
Softkey Menu Location
TITLE SEQUENCE Front-panel access key: SEQ TITLE TO MEMORY Front-panel access key: SEQ TITLE TO P MTR/HPIB Front-panel access key: SEQ TITLE TO PERIPHERAL Front-panel access key: SEQ TITLE TO PRNTR/HPIB Front-panel access key: SEQ TRACKING on OFF Front-panel access key: MARKER FCTN Trans: E/E S11 (A/R) Front-panel access key: MEAS Trans: E/E S12 REV Front-panel access key: MEAS Trans: E/E S21 (B/R) Front-panel access key: MEAS Trans: E/E S21 FWD Front-panel access key: MEAS
Trans: E/O (PORT 1
Trans: O/E (PORT 1
Trans: O/O (PORT 1 TRANSFORM MENU Front-panel access key: SYSTEM TRANSFORM on OFF Front-panel access key: SYSTEM TRANSFORM PARAMETERS Front-panel access key: SYSTEM TRANSFORM SPAN Front-panel access key: SYSTEM TRANSMISSION Front-panel access key: CAL TRIGGER MENU Front-panel access key: MENU TRIGGER: TRIG OFF Front-panel access key: MENU TRL/LRM OPTION Front-panel access key: CAL TRL*/LRM* 2-PORT Front-panel access key: CAL TRL 3.5 mm Front-panel access key: CAL TRL LINE OR MATCH Front-panel access key: CAL TRL REFLECT Front-panel access key: CAL TTL I/O Front-panel access key: SEQ TTL OUT Front-panel access key: SEQ
2)2)
2)
Front-panel access key: MEAS Front-panel access key: MEAS Front-panel access key: MEAS
2-42
Page 63
Table 2-1. Location of Softkeys (23 of 24)
Softkey Menu Location
TTL OUT HIGH Front-panel access key: SEQ TTL OUT LOW Front-panel access key: SEQ TUNED RECEIVER Front-panel access key: SYSTEM UNCOUPLED Front-panel access key: MARKER FC TN UP CONVERTER Front-panel access key: SYSTEM UPPER LIMIT Front-panel access key: SYSTEM USE MEMORY on OFF Front-panel access key: SYSTEM USE PASS CONTROL Front-panel access key: LOCAL USER Front-panel access key: PRESET USER KIT Front-panel access key: CAL USE SENSOR A / B Front-panel access key: CAL VELOCITY FACTOR Front-panel access key: CAL VIEW MEASURE Front-panel access key: SYSTEM
Reference
Location of Softkeys
VOLUME NUMBER Front-panel access key: LOCAL VOLUME NUMBER Front-panel access key: SAVE/RECALL WAIT X Front-panel access key: SEQ WARNING Front-panel access key: DISPLAY WARNING [ ] Front-panel access key: COPY WAVEGUIDE Front-panel access key: CAL WAVEGUIDE DELAY Front-panel access key: SCALE REF WIDE Front-panel access key: SYSTEM WIDTH VALUE Front-panel access key: MARKER F CTN WIDTHS on OFF Front-panel access key: MARKER FCTN WINDOW Front-panel access key: SYSTEM WINDOW: MAXIMUM Front-panel access key: SYSTEM WINDOW: MINIMUM Front-panel access key: SYSTEM WINDOW: NORMAL Front-panel access key: SYSTEM
2-43
Page 64
Reference
Location of Softkeys
Table 2-1. Location of Softkeys (24 of 24)
Softkey Menu Location
XMIT CNTRL [ ] Front-panel access key: LOCAL Y: Refl Front-panel access key: MEAS Y: Trans Front-panel access key: MEAS Z: Refl Front-panel access key: MEAS Z: Trans Front-panel access key: MEAS
2-44
Page 65

Connectors, Adjustments, and Display Annotation

Connectors, Adjustments, and Display Annotation
Table 2-2. Connectors, Adjustments, and Display Annotation (1 of 3)
Key Description
Reference
= display annotation
* = display annotation Source parameters changed: measured data in doubt until a complete fresh sweep
10 MHZ PRECISION REFERENCE connector
ADJ adjustment This adjustment adjusts the frequency of the Option 1D5 10 MHz reference.
AUX INPUT connector This allows for a dc or ac voltage input from an external signal source, such as a
Avg = display annotation Sweep-to-sweep averaging is on. The averaging count is shown immediately
BIAS CONNECT connector These connect bias devices connected to port 1 and port 2. The fuses (1 A, 125 V)
C? = display annotation Stimulus parameters have changed from the error-corrected state, or interpolated
C2 = display annotation Full two-port error-correction is active when either the power range for each port is
Cor = display annotation Error correction is on.
Del = display annotation Electrical delay has been added or subtracted, or port extensions are active.
Fast sweep indicator. This symbol is displayed in the status notation block when sweep time is less than 1.0 second. When sweep time is greater than 1.0 second, this symbol moves along the displayed trace.
has been taken.
Option 1D5
detector or function generator, which you can then measure, using the S-parameter menu.
below.
protect the port 1 and port 2 bias lines.
error correction is on.
different (uncoupled), or the
TESTSET HOLD
is activated.
DIN KYBD connector This connector is used for the optional AT-compatible keyboard for titles and
remote front-panel operation. This provides a more convenient means to enter a title for storage files, as well as substitute for the analyzer’s front panel keyboard. The keyboard must be connected to the analyzer before the power is switched on.
2-45
Page 66
Reference
Connectors, Adjustments, and Display Annotation
Table 2-2. Connectors, Adjustments, and Display Annotation (2 of 3)
Key Description
Disk drive This 3.5 inch drive allows you to store and recall instrument states and
measurement results for later analysis.
ext = display annotation Waiting for an external trigger.
EXT AM connector This allows for an external analog signal input that is applied to the ALC circuitry
of the analyzer’s source. This input analog signal amplitude modulates the RF output signal.
EXT MON connector RED, GREEN, and BLUE video output connectors provide analog red, green, and
blue video signals which you can use to drive an external monitor, such as the HP 3571A/B, or monochrome monitor, such as the HP 35731A/B. You can use other analog multi-sync monitors if they are compatible with the analyzer’s 25.5 kHz scan rate and video levels: 1 V p-p, 0.7 V=white, 0 V=black, –0.3 V sync, sync on green.
EXT REF connector This allows for a frequency reference signal input that can phaselock the analyzer
to an external frequency standard for increased frequency accuracy.
EXT TRIG connector This allows connection of an external negative-going TTL-compatible signal that
will trigger a measurement sweep. The trigger can be set to external through softkey functions.
Gat = display annotation Gating is on.
H=2 = display annotation Harmonic mode is on, and the second harmonic is being measured. (Harmonics
Option 002 only.)
H=3 = display annotation Harmonic mode is on, and the third harmonic is being measured. (Harmonics
Option 002 only.)
Hld = display annotation Hold sweep.
HP-IB connector This connector allows communication with compatible devices including external
controllers, printers, plotters, disk drives, and power meters.
LIMIT TEST connector Outputs a TTL signal of the limit test results. A TTL high state indicates a ‘pass’
condition. A TTL low state indicates a “fail” condition.
LINE key This switch controls ac power to the analyzer. 1 is on, 0 is off. Refer to Table 2-5,
“Power-on Conditions (versus Preset),” on page 2-56, for more information.
man = display annotation Waiting for manual trigger.
Of? = display annotation Frequency offset mode error, the IF frequency is not within 10 MHz of expected
frequency. LO inaccuracy is the most likely cause.
Ofs = display annotation Frequency offset mode is on.
2-46
Page 67
Reference
Connectors, Adjustments, and Display Annotation
Table 2-2. Connectors, Adjustments, and Display Annotation (3 of 3)
Key Description
P? = display annotation Source power is unleveled at start or stop of sweep.
PØ = display annotation Source power has been automatically set to minimum, due to receiver overload.
PARALLEL PORT connector This connector is used with parallel (or Centronics interface) peripherals such as
printers and plotters. It can also be used as a general purpose I/O port, with control provided by test sequencing functions.
PC = display annotation Power meter calibration is on.
PC? = display annotation The analyzer’s source could not be set to the desired level, following a power
meter calibration.
PORT 1 and PORT 2 connectors These ports output a signal from the source and receive input signals from a device
under test. PORT 1 allows you to measure S12 and S11. PORT 2 allows you to measure S21 and S22.
PRESET key This key returns the instrument to either a known factory preset state, or a user
preset state that can be defined. Refer to “Preset Conditions” on page 2-48 for more information.
PRm = display annotation Power range is in manual mode.
PROBE POWER connector This connector (fused inside the instrument) supplies power to an active probe for
in-circuit measurements of ac circuits.
R CHANNEL connectors These connectors allow you to apply an input signal to the analyzer’s R channel, for
frequency offset mode.
RS-232 connector This connector allows the analyzer to output to a peripheral with an RS-232 (serial)
input. This includes printers and plotters.
Smo = display annotation Trace smoothing is on.
TEST SEQ connector This connector outputs a TTL signal which can be programmed by the user in a test
sequence to be high or low. By default, this output provides an end-of-sweep TTL signal. (For use with part handlers).
TEST SET I/O INTERCONNECT connector This allows you to connect an HP 8702D Option 011 analyzer to an HP 85046A/B or
85047A S-parameter test set using the interconnect cable supplied with the test set. The S-parameter test set is then fully controlled by the analyzer.
tsH = display annotation Indicates that the test set hold mode is engage. That is, a mode of operation is
selected which would cause repeated switching of the step attenuator. This hold mode may be overridden.
2-47
Page 68
Reference

Preset Conditions

Preset Conditions
PRESET
PRESET
PRESET
PRESET
key is pressed, the analyzer reverts to a known state called
PRESET
key pressed, the analyzer performs a
SAVE/RECALL
PRESET: USER
,
key is now toggled to the
and selecting
menu).
.
FACTORY
USER
selection and your defined instru-
PRESET
and when you turn
.
When the the factory preset state. This state is defined in Table 2-4 on page 2-50. There are subtle differences between the preset state and the power-up state. These differences are documented in Table 2-5 on page 2-56. If power to battery-pro­tected memory is lost, the analyzer will have certain parameters set to default settings.
When line power is cycled, or the self-test routine. Upon successful completion of that routine, the instrument state is set to the conditions shown in Table 2-4 on page 2-50. The same condi­tions are true following a “PRES;” or “RST;” command over HP-IB, although the self-test routines are not executed.
You also can configure an instrument state and define it as your user preset state:
a
Set the instrument state to your desired preset conditions.
b
Save the state (
c
Rename that register to “UPRESET”.
d
Press
The ment state will be recalled each time you press power on. You can toggle back to the factory preset instrument state by press­ing
NOTE
When you send a preset over HP-IB, you will always get the factory preset. You can, however, activate the user-defined preset over HP-IB by recalling the register in which it is stored.
2-48
Page 69
Table 2-3. Preset Display Formats
Reference
Format Scale
Position Value
Log Magnitude (dB) 10.0 5.0 0.0
Phase (degree) 90.0 5.0 0.0
Group Delay (ns) 10.0 5.0 0.0
Smith Chart 1.00 1.0
Polar 1.00 1.0
Linear Magnitude 0.1 0.0 0.0
Real 0.2 5.0 0.0
Imaginary 0.2 5.0 0.0
SWR 1.00 0.0 1.0
Reference
Preset Conditions
2-49
Page 70
Reference
Preset Conditions
Table 2-4. Preset Conditions (1 of 6)
Preset Conditions Preset Value
Analyzer Mode
Analyzer Mode Component Analyzer Mode
Frequency Offset Off
Operation Offset Value 0
Harmonic Operation Off
Stimulus Conditions
Sweep Type Linear Frequency
Display Mode Start/Stop
Trigger Type Continuous
External Trigger Off
Sweep Time 100 ms, Auto Mode
Start Frequency 300 kHz
Frequency Span (std.) 2999.97 MHz
Frequency Span (Opt. 006) 5999.97 MHz
Start Time 0
Time Span 100 ms
CW Frequency 1000 MHz
Source Power 0 dBm
Power Slope 0 dB/GHz; Off
Start Power –15.0 dBm
Power Span 25 dB
Coupled Power On
Source Power On
Coupled Channels On
Coupled Port Power On
Power Range Auto; Range 0
Number of Points 201
Frequency List
Frequency List Empty
Edit Mode Start/Stop, Number of Points
Response Conditions
2-50
Page 71
Preset Conditions
Table 2-4. Preset Conditions (2 of 6)
Preset Conditions Preset Value
Parameter Channel 1: S11;
Channel 2: S21
Conversion Off
Format Log Magnitude (all inputs)
Display Data
Color Selections Same as before PRESET
Dual Channel Off
Active Channel Channel 1
Frequency Blank Disabled
Split Display On
Intensity If set to 15%, PRESET has no effect. If set to < 15%,
PRESET increases intensity to 15%.
Beeper: Done On
Beeper: Warning Off
D2/D1 to D2 Off
Title Channel 1 = [hp]
Channel 2 = Empty
IF Bandwidth 3000 Hz
IF Averaging Factor 16; Off
Smoothing Aperture 1% SPAN; Off
Phase Offset 0 Degrees
Electrical Delay 0 ns
Scale/Division 10 dB/Division
Calibration
Correction Off
Calibration Type None
Calibration Kit 3.5 mm
System Z0 50 ohms
Velocity Factor 1
Reference
2-51
Page 72
Reference
Preset Conditions
Table 2-4. Preset Conditions (3 of 6)
Preset Conditions Preset Value
Extensions
Port 1 Port 2 Input A Input B
Chop A and B On
Power Meter Calibration
Number of Readings Power Loss Correction Sensor A/B
Interpolated Error Correction Off
Markers (coupled)
Markers 1, 2, 3, 4, 5 1 GHz; All Markers Off
Last Active Marker 1
Reference Marker None
Marker Mode Continuous
Display Markers On
Delta Marker Mode Off
Coupling On
Marker Search Off
Marker Target Value –3 dB
Marker Width Value –3 dB; Off
Marker Tracking Off
Marker Stimulus Offset 0 Hz
Marker Value Offset 0 dB
Marker Aux Offset (Phase) 0 Degrees
Marker Statistics Off
Polar Marker Lin Mkr
Smith Marker R + jX Mkr
Limit Lines
Limit Lines Off
Limit Testing Off
Limit List Empty
Off
0 s 0 s 0 s 0 s
Off 1 Off A
2-52
Page 73
Table 2-4. Preset Conditions (4 of 6)
Preset Conditions Preset Value
Edit Mode Upper/Lower Limits
Stimulus Offset 0 Hz
Amplitude Offset 0 dB
Limit Type Sloping Line
Beep Fail Off
Time Domain
Transform Off
Transform Type Bandpass
Start Transform –20 nanoseconds
Transform Span 40 nanoseconds
Gating Off
Gate Shape Normal
Gate Start –10 nanoseconds
Gate Span 20 nanoseconds
Demodulation Off
Window Normal
Use Memory Off
System Parameters
HP-IB Addresses Last Active State
HP-IB Mode Last Active State
Focus Last Active State
Clock Time Stamp On
Preset: Factory/User Last Selected State
Copy Configuration
Parallel Port Last Active State
Plotter Type Last Active State
Plotter Port Last Active State
Plotter Baud Rate Last Active State
Plotter Handshake Last Active State
HP-IB Address Last Active State
Printer Type Last Active State
Reference
Preset Conditions
2-53
Page 74
Reference
Preset Conditions
Table 2-4. Preset Conditions (5 of 6)
Preset Conditions Preset Value
Printer Port Last Active State
Printer Baud Rate Last Active State
Printer Handshake Last Active State
Printer HP-IB Address Last Active State
Disk Save Configuration (Define Store)
Data Array Off
Raw Data Array Off
Formatted Data Array Off
Graphics Off
Data Only Off
Directory Size Default
Save Binary Binary
Select Disk Internal Memory
Disk Format LIF
Sequencing
b
Loop Counter 0
TTL OUT High
Service Modes
HP-IB Diagnostic Off
Source Phase Lock Loop On
Sampler Correction On
Spur Avoidance On
Aux Input Resolution Low
Analog Bus Mode 11 (Aux Input)
Plot
Plot Data On
Plot Memory On
Plot Graticule On
Plot Text On
Plot Marker On
a
2-54
Page 75
Preset Conditions
Table 2-4. Preset Conditions (6 of 6)
Preset Conditions Preset Value
Autofeed On
Plot Quadrant Full Page
Scale Plot Full
Plot Speed Fast
Pen Number:
Ch1 Data Ch2 Data Ch1 Memory Ch2 Memory Ch1 Graticule Ch2 Graticule Ch1 Text Ch2 Text Ch1 Marker Ch2 Marker
Line Type:
Ch1 Data Ch2 Data Ch1 Memory Ch2 Memory
Print
Printer Mode Last Active State
Auto-Feed On
Printer Colors
Ch1 Data Ch1 Mem Ch2 Data Ch2 Mem Graticule Warning Text
a. The directory size is calculated as 0.013% of the floppy disk size (which is approximately 256) or 0.005%
of the hard disk size.
b. Pressing PRESET turns off sequencing modify (edit) mode and stops any running sequence.
2 3 5 6 1 1 7 7 7 7
7 7 7 7
Magenta Green Blue Red Cyan Black Black
Reference
2-55
Page 76
Reference

Power-on Conditions

Power-on Conditions
Table 2-5. Power-on Conditions (versus Preset)
HP-IB MODE Talker/listener.
SAVE REGISTERS Power meter calibration data and calibration data not
associated with an instrument state are cleared.
COLOR DISPLAY Default color values.
INTENSITY Factory stored values. The factory values can be changed by
running the appropriate service routine. Refer to the HP 8753D Service Guide.
SEQUENCES Sequence 1 through 5 are erased.
DISK DIRECTORY Cleared.
2-56
Page 77
3

Error Messages

Page 78
Error Messages
Error Messages
Error Messages
This chapter contains the information to help you interpret any error mes­sages that may be displayed on the HP 8702D or transmitted by the instru­ment over HP-IB.
Some messages described in this chapter are for information only and do not indicate an error condition. These messages are not numbered and so they will not appear in the numerical listing.
What you’ll find in this chapter
Error Messages in Alphabetical Order 3-3 Error Messages in Numerical Order 3-27
3-2
Page 79

Error Messages in Alphabetical Order

Error Messages in Alphabetical Order
Error Messages
ABORTING COPY OUTPUT
Information Message
This message is displayed briefly if you have pressed operation. If the message is not subsequently replaced by error message num­ber 25, more to exit the abort process and verify the status of the copy device. At this point, the copy device will probably have an error condition which must be fixed. (For example: out of paper or paper jam.)
PRINT ABORTED
LOCAL
to abort a copy
, the copy device may be hung. Press
LOCAL
once
ADDITIONAL STANDARDS NEEDED TO COVER SPAN
Error Number 68
Error correction for the selected calibration class cannot be computed until you have measured all the necessary standards to cover the current frequency span.
ADDRESSED TO TALK WITH NOTHING TO SAY
Error Number 31
You have sent a read command to the analyzer (such as ENTER 716) without first requesting data with an appropriate output command (such as OUTP­DATA). The analyzer has no data in the output queue to satisfy the request.
ALL REGISTERS HAVE BEEN USED
Error Number 200
You have used all of the available registers; you can store no more instrument states even though you may still have sufficient memory. There are 31 regis­ters available, plus the present instrument state.
ASCII: MISSING ’BEGIN’ STATEMENT
Error Number 193
The citifile you just downloaded over the HP-IB or via disk was not properly organized. The analyzer is unable to read the “BEGIN” statement.
3-3
Page 80
Error Messages
Error Messages in Alphabetical Order
ASCII: MISSING ’CITIFILE’ STATEMENT
Error Number 194
The citifile you just downloaded over the HP-IB or via disk was not properly organized. The analyzer is unable to read the “CITIFILE” statement.
ASCII: MISSING ’DATA’ STATEMENT
Error Number 195
The citifile you just downloaded over the HP-IB or via disk was not properly organized. The analyzer is unable to read the “DATA” statement.
ASCII: MISSING ’VAR’ STATEMENT
Error Number 196
The citifile you just downloaded over the HP-IB or via disk was not properly organized. The analyzer is unable to read the “VAR” statement.
AVERAGING INVALID ON NON-RATIO MEASURE
Error Number 13
You cannot use sweep-to-sweep averaging in single-input measurements. Sweep-sweep averaging is valid only for ratioed measurements (A/R, B/R, A/B, and S-parameters). You can use noise reduction techniques, such as narrower IF bandwidth, for single input measurements.
BAD FREQ FOR HARMONIC OR FREQ OFFSET
Error Number 181
You turned on time domain or recalled a calibration that resulted in start and stop frequencies that are beyond the allowable limits.
BATTERY FAILED. STATE MEMORY CLEARED
Error Number 183
The battery protection of the battery-operated memory has failed. The mem­ory has been cleared. Refer to the
HP 8753D Service Guide
for battery
replacement instructions.
BATTERY LOW! STORE SAVE REGS TO DISK
Error Number 184
The battery protection of the memory is in danger of failing. If this occurs, all of the instrument state registers stored in memory will be lost. Save these states to a disk and refer to the
HP 8702D User’s Guide
for battery replace-
ment instructions.
3-4
Page 81
Error Messages
Error Messages in Alphabetical Order
BLOCK INPUT ERROR
Error Number 34
The analyzer did not receive a complete data transmission. This is usually caused by an interruption of the bus transaction. Clear by pressing the
LOCAL
key or aborting the I/O process at the controller.
BLOCK INPUT LENGTH ERROR
Error Number 35
The length of the header received by the analyzer did not agree with the size of the internal array block. Refer to the
HP 8702D Programmer’s Guide
for
instructions on using analyzer input commands.
CALIBRATION ABORTED
Error Number 74
You have changed the active channel during a calibration so the calibration in progress was terminated. Make sure the appropriate channel is active and restart the calibration.
CALIBRATION REQUIRED
Error Number 63
A calibration set could not be found that matched the current stimulus state or measurement parameter. You will have to perform a new calibration.
CANNOT FORMAT DOS DISKS ON THIS DRIVE
Error Number 185
You have attempted to initialize a floppy disk to DOS format on an external disk drive that does not support writing to all 80 tracks of the double density (gray) and high density disks (usually black). The older single-sided disks (blue) had only 66 tracks and some disk drives were limited to accessing that number of tracks. To format the disk, either choose another external disk drive or use the analyzer’s internal disk drive.
CANNOT MODIFY FACTORY PRESET
Error Number 199
You have attempted to rename, delete, or otherwise alter the factory preset state. The factory preset state is permanently stored in memory and cannot be altered. If your intent was to create a user preset state, you must create a new instrument state, save it, and then rename it to “UPRESET”.
3-5
Page 82
Error Messages
Error Messages in Alphabetical Order
CANNOT READ/WRITE HFS FILE SYSTEM
Error Number 203
The disk is being accessed by the analyzer and is found to contain an HFS (hierarchical file system). The analyzer does not support HFS.
CAN’T CHANGE-ANOTHER CONTROLLER ON BUS
Error Number 37
You must remove the active controller from the bus or the controller must relinquish the bus before the analyzer can assume the system controller mode.
CAN’T STORE/LOAD SEQUENCE, INSUFFICIENT MEMORY
Error Number 127
Your sequence transfer to or from a disk could not be completed due to insuf­ficient memory.
CH1 (CH2) TARGET VALUE NOT FOUND
Error Number 159
Your target value for the marker search function does not exist on the current data trace.
CONTINUOUS SWITCHING NOT ALLOWED
Error Number 10
Your current measurement requires the power range to switch between chan­nel 1 and channel 2. To protect the attenuator from undue mechanical wear, it will not be switched continuously. The “tsH” (test set hold) indicator in the left margin of the display indicates that the inactive channel has been put in the sweep hold mode.
COPY: device not responding; copy aborted
Error Number 170
The printer or plotter is not accepting data. Verify the cable connections, HP­IB addresses, and otherwise ensure that the copy device is ready.
COPY OUTPUT COMPLETED
Information Message
The analyzer has completed outputting data to the printer or plotter. The ana­lyzer can now accept another copy command.
3-6
Page 83
Error Messages
Error Messages in Alphabetical Order
CORRECTION AND DOMAIN RESET
Error Number 65
When you change the frequency range, sweep type, or number of points, error-correction is switched off and the time domain transform is recalculated, without error-correction. You can either correct the frequency range, sweep type, or number of points to match the calibration, or perform a new calibra­tion. Then perform a new time domain transform.
CORRECTION CONSTANTS NOT STORED
Error Number 3
A store operation to the EEPROM was not successful. You must change the position of the jumper on the A9 CPU assembly.
CORRECTION TURNED OFF
Error Number 66
Critical parameters in your current instrument state do not match the param­eters for the calibration set, therefore correction has been turned off. The crit­ical instrument state parameters are sweep type, start frequency, frequency span, and number of points.
CURRENT PARAMETER NOT IN CAL SET
Error Number 64
Correction is not valid for your selected measurement parameter. Either change the measurement parameters or perform a new calibration.
D2/D1 INVALID WITH SINGLE CHANNEL
Error Number 130
You can only make a D2/D1 measurement if both channels are on.
D2/D1 INVALID. CH1 CH2 NUM PTS DIFFERENT
Error Number 152
You can only make a D2/D1 measurement if both channels have the same num­ber of points.
DEADLOCK
Error Number 111
A fatal firmware error occurred before instrument preset completed. Call your local Hewlett-Packard sales and service office.
3-7
Page 84
Error Messages
Error Messages in Alphabetical Order
DEMODULATION NOT VALID
Error Number 17
Demodulation is only valid for the CW time mode.
DEVICE: not on, not connect, wrong addrs
Error Number 119
The device at the selected address cannot be accessed by the analyzer. Verify that the device is switched on, and check the HP-IB connection between the analyzer and the device. Ensure that the device address recognized by the analyzer matches the HP-IB address set on the device itself.
DIRECTORY FULL
Error Number 188
There is no room left in the directory to add files. Either delete files or get a new disk.
DISK HARDWARE PROBLEM
Error Number 39
The disk drive is not responding correctly. Refer to the disk drive operating manual.
DISK IS WRITE PROTECTED
Error Number 48
The store operation cannot write to a write-protected disk. Slide the write­protect tab over the write-protect opening in order to write data on the disk.
DISK MEDIUM NOT INITIALIZED
Error Number 40
You must initialize the disk before it can be used.
DISK MESSAGE LENGTH ERROR
Error Number 190
The analyzer and the external disk drive aren’t communicating properly. Check the HP-IB connection and then try substituting another disk drive to isolate the problem instrument.
3-8
Page 85
Error Messages
Error Messages in Alphabetical Order
DISK: not on, not connected, wrong addrs
Error Number 38
The disk cannot be accessed by the analyzer. Verify power to the disk drive, and check the HP-IB connection between the analyzer and the disk drive. Ensure that the disk drive address recognized by the analyzer matches the HP-IB address set on the disk drive itself.
DISK READ/WRITE ERROR
Error Number 189
There may be a problem with your disk. Try a new floppy disk. If a new floppy disk does not eliminate the error, suspect hardware problems.
DISK WEAR - REPLACE DISK SOON
Error Number 49
Cumulative use of the disk is approaching the maximum. Copy files as neces­sary using an external controller. If no controller is available, load instrument states from the old disk and store them to a newly initialized disk using the save/recall features of the analyzer. Discard the old disk.
DOMAIN RESET
Error Number 67
Time domain calculations were reset due to a change in the frequency range, sweep type, or number of points. Perform a new time domain transform on the new state.
DOS NAME LIMITED TO 8 CHARS + 3 CHAR EXTENSION
Error Number 180
A DOS file name must meet the following criteria:
• minimum of 1 character
• first character must be alpha; the remainder must be alphanumeric or underscore
•format is
filename.ext
• maximum of 8 characters in the filename
• maximum of 3 characters in the extension field
• a dot separates the filename from the extension field (the dot is not part of
the name on the disk)
3-9
Page 86
Error Messages
Error Messages in Alphabetical Order
DUPLICATING TO THIS SEQUENCE NOT ALLOWED
Error Number 125
A sequence cannot be duplicated to itself.
EXCEEDED 7 STANDARDS PER CLASS
Error Number 72
When modifying calibration kits, you can define a maximum of seven stan­dards for any class.
EXTERNAL SOURCE MODE REQUIRES CW TIME
Error Number 148
An external source can only be phase locked and measured in the CW time sweep mode.
EXT SOURCE NOT READY FOR TRIGGER
Error Number 191
There is a hardware problem with the external source. Verify the connections between the analyzer and the external source. If the connections are correct, refer to the source operating manual.
EXT SRC: NOT ON/CONNECTED OR WRONG ADDR
Error Number 162
The analyzer is unable to communicate with the external source. Check the connections and the HP-IB address on the source.
FILE NOT COMPATIBLE WITH INSTRUMENT
Information Message
You cannot recall user graphics that had been saved on an earlier model of analyzer with a monochrome display. These files cannot be used with the HP 8702D.
FILE NOT FOUND
Error Number 192
The requested file was not found on the current disk medium.
3-10
Page 87
Error Messages
Error Messages in Alphabetical Order
FILE NOT FOUND OR WRONG TYPE
Error Number 197
During a resave operation, either the file was not found or the type of file was not an instrument state file.
FIRST CHARACTER MUST BE A LETTER
Error Number 42
The first character of a disk file title or an internal save register title must be an alpha character.
FORMAT NOT VALID FOR MEASUREMENT
Error Number 75
Conversion measurements (Z or Y reflection and transmission) are not valid with Smith chart and SWR formats.
FORMATTING DATA
Information Message
The list information is being processed for a list data output to a copy device and stored in the copy spool buffer. During this time, the analyzer’s resources are dedicated to this task (which takes less than a few seconds.)
FREQ OFFSET ONLY VALID IN COMPONENT ANALYZER MODE
Error Number 140
You can only make frequency offset measurements in the component analyzer mode.
FUNCTION NOT VALID
Error Number 14
The function you requested is incompatible with the current instrument state.
FUNCTION NOT VALID DURING MOD SEQUENCE
Error Number 131
You cannot perform sequencing operations while a sequence is being modi­fied.
3-11
Page 88
Error Messages
Error Messages in Alphabetical Order
FUNCTION NOT VALID FOR INTERNAL MEMORY
Error Number 201
The function you selected only works with disk files.
FUNCTION ONLY VALID DURING MOD SEQUENCE
Error Number 164
You can only use the
GOSUB SEQUENCE
capability when you are building a sequence. Attempting to use this softkey at any other time returns an error message and no action is taken.
HPIB COPY IN PROGRESS, ABORT WITH LOCAL
Error Number 169
An HP-IB copy was already in progress when you requested the HP-IB for another function. To abort the first copy, press
LOCAL
, otherwise the HP-IB is
unavailable until the first copy is completed.
ILLEGAL UNIT OR VOLUME NUMBER
Error Number 46
The disk unit or volume number set in the analyzer is not valid. Refer to the disk drive operating manual.
INIT DISK removes all data from disk
Information Message
Continuing with the initialize operation will
destroy
any data currently on the
disk.
INITIALIZATION FAILED
Error Number 47
The disk initialization failed, probably because the disk is damaged.
INSTRUMENT STATE MEMORY CLEARED
Error Number 56
All instrument state registers have been cleared from memory along with any saved calibration data, memory traces, and calibration kit definitions. Addi­tionally, all user-settable selections, (such as HP-IB addresses) are set to their defaults.
3-12
Page 89
Error Messages
Error Messages in Alphabetical Order
INSUFFICIENT MEMORY
Error Number 51
Your last front panel or HP-IB request could not be implemented due to insuf­ficient memory space. In some cases, this is a fatal error from which you can escape only by presetting the instrument.
INSUFFICIENT MEMORY FOR PRINT/PLOT
Error Number 168
There is not enough memory available for the print or plot function. Increase the available memory by changing or eliminating a memory-intensive opera­tion such as reducing the number of points in the sweep.
INSUFFICIENT MEMORY, PWR MTR CAL OFF
Error Number 154
There is not enough memory space for the power meter calibration array. Increase the available memory by clearing one or more save/recall registers, or by reducing the number of points.
INVALID CAL DATA FILE
Error Number 142
The calibration file you are attempting to read does not contain valid data.
INVALID KEY
Error Number 2
You pressed an undefined softkey.
LIST MODE OFF: INVALID WITH LO FREQ
Error Number 182
List mode has been turned off in the frequency offset mode because it is incompatible with your selected LO frequency.
LIST TABLE EMPTY
Error Number 9
The frequency list is empty. To implement list frequency mode, add segments to the list table.
3-13
Page 90
Error Messages
Error Messages in Alphabetical Order
LOAD RECEIVER CAL DATA FROM DISK FILE
Error Number 134
You are attempting to measure a receiver cal standard that has not been defined. Define by loading cal data from the appropriate file.
LOAD SOURCE CAL DATA FROM DISK FILE
Error Number 135
You are attempting to measure a source cal standard that has not been defined. Define by loading cal data from the appropriate file.
LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN
Error Number 150
A logarithmic sweep is only valid if the stop frequency is greater than four times the start frequency. For frequency spans of less than two octaves, the sweep type automatically reverts to linear sweep.
LOW PASS: FREQ LIMITS CHANGED
Information Message
The frequency domain data points must be harmonically related from dc to the stop frequency. That is, stop = n × start, where n = number of points. If this condition is not true when a low pass mode (step or impulse) is selected and transform is turned on, the analyzer resets the start and stop frequencies. The stop frequency is set close to the entered stop frequency, and the start frequency is set equal to stop/n.
MEMORY FOR CURRENT SEQUENCE IS FULL
Error Number 132
All the memory in the sequence you are modifying is filled with instrument commands.
MORE SLIDES NEEDED
Error Number 71
When you use a sliding load (in a user-defined calibration kit), you must set at least three slide positions to complete the calibration.
NO CALIBRATION CURRENTLY IN PROGRESS
Error Number 69
RESUME CAL SEQUENCE
The
softkey is not valid unless a calibration is already in
progress. Start a new calibration.
3-14
Page 91
Error Messages
Error Messages in Alphabetical Order
NO DISK MEDIUM IN DRIVE
Error Number 41
You have no disk in the current disk unit. Insert a disk, or check the disk unit number stored in the analyzer.
NO FAIL FOUND
Service Error Number 114
The self-diagnose function of the instrument operates on an internal test fail­ure. At this time, no failure has been detected.
NO FILE(S) FOUND ON DISK
Error Number 45
No files of the type created by an analyzer store operation were found on the disk. If you requested a specific file title, that file was not found on the disk.
NO IF FOUND: CHECK R INPUT LEVEL
Error Number 5
The first IF signal was not detected during pre-tune. Check the front panel R channel jumper. If there is no visible problem with the jumper, refer to the
HP 8753D Service Guide
NO LIMIT LINES DISPLAYED
for troubleshooting.
Error Number 144
You can turn limit lines on but they cannot be displayed on polar or Smith chart display formats.
NO MARKER DELTA -SPAN NOT SET
Error Number 15
You must turn the delta marker mode on, with at least two markers displayed, in order to use the
NO MEMORY AVAILABLE FOR INTERPOLATION
MARKER SPAN
softkey function.
Error Number 123
You cannot perform interpolated error correction due to insufficient memory.
NO MEMORY AVAILABLE FOR SEQUENCING
Error Number 126
You cannot modify the sequence due to insufficient memory.
3-15
Page 92
Error Messages
Error Messages in Alphabetical Order
NO PHASE LOCK: CHECK R INPUT LEVEL
Error Number 7
The first IF signal was detected at pre-tune, but phase lock could not be acquired. Refer to the
NO SPACE FOR NEW CAL. CLEAR REGISTERS
HP 8753D Service Guide
for troubleshooting.
Error Number 70
You cannot store a calibration set due to insufficient memory. You can free more memory by clearing a saved instrument state from an internal register (which may also delete an associated calibration set, if all the instrument states using the calibration kit have been deleted.) You can store the saved instrument state and calibration set to a disk before clearing them. After delet­ing the instrument states, press
NOT ALLOWED DURING POWER METER CAL
PRESET
to run the memory packer.
Error Number 198
When the analyzer is performing a power meter calibration, the HP-IB bus is unavailable for other functions such as printing or plotting.
NOT ENOUGH SPACE ON DISK FOR STORE
Error Number 44
The store operation will overflow the available disk space. Insert a new disk or purge files to create free disk space.
NOT VALID FOR CURRENT PARAMETER
Error Number 143
You are attempting a response and match calibration on a parameter other than E/O or O/E.
NO VALID MEMORY TRACE
Error Number 54
If you are going to display or otherwise use a memory trace, you must first store a data trace to memory.
NO VALID STATE IN REGISTER
Error Number 55
You have requested the analyzer, over HP-IB (or by sequencing), to load an instrument state from an
3-16
empty
internal register.
Page 93
Error Messages
Error Messages in Alphabetical Order
ONLY LETTERS AND NUMBERS ARE ALLOWED
Error Number 43
You can only use alpha-numeric characters (and underscores) in disk file titles or internal save register titles. Other symbols are not allowed, except for the “underscore” symbol.
OPTIONAL FUNCTION; NOT INSTALLED
Error Number 1
The function you requested requires a capability provided by an option to the standard analyzer. That option is not currently installed. (Option 002 har­monic measurement capability, Option 006 6 GHz operation, Option 110 deletes time domain transform, and Option 075 75 ohm impedance.)
OVERLOAD ON INPUT A, POWER REDUCED
Error Number 58
See error number 57.
OVERLOAD ON INPUT B, POWER REDUCED
Error Number 59
See error number 57.
OVERLOAD ON INPUT R, POWER REDUCED
Error Number 57
You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to –85 dBm. The annotation P appears in the left margin of the display to indicate that the power trip func­tion has been activated. When this occurs, reset the power to a lower level, then toggle the
SOURCE PWR on OFF
softkey to switch on the power again.
PARALLEL PORT NOT AVAILABLE FOR GPIO
Error Number 165
You have defined the parallel port as
COPY
for sequencing in the HP-IB menu.
To access the parallel port for general purpose I/O (GPIO), set the selection to
GPIO
.
PARALLEL PORT NOT AVAILABLE FOR COPY
Error Number 167
3-17
Page 94
Error Messages
Error Messages in Alphabetical Order
You have defined the parallel port as general purpose I/O (GPIO) for sequenc­ing. The definition was made under the lel port for copy, set the selection to
LOCAL
key menus. To access the paral-
PARALLEL [COPY]
.
PHASE LOCK CAL FAILED
Error Number 4
An internal phase lock calibration routine is automatically executed at power­on, preset, and any time a loss of phase lock is detected. This message indi­cates that phase lock calibration was initiated and the first IF detected, but a problem prevented the calibration from completing successfully. Refer to the
HP 8753D Service Guide
and execute pre-tune correction test 48.
This message may appear if you connect a mixer between the RF output and R input before turning on frequency offset mode. Ignore it: it will go away when you turn on frequency offset. This message may also appear if you turn on fre­quency offset mode before you define the offset.
PHASE LOCK LOST
Error Number 8
Phase lock was acquired but then lost. Refer to the
HP 8753D Service Guide
for troubleshooting information.
PLOT ABORTED
Error Number 27
When you press the
PLOTTER: not on, not connect, wrong addrs
LOCAL
key, the analyzer aborts the plot in progress.
Error Number 26
The plotter does not respond to control. Verify power to the plotter, and check the HP-IB connection between the analyzer and the plotter. Ensure that the plotter address recognized by the analyzer matches the HP-IB address set on the plotter itself.
PLOTTER NOT READY-PINCH WHEELS UP
Error Number 28
The plotter pinch wheels clamp the paper in place. If you raise the pinch wheels, the plotter indicates a “not ready” status on the bus.
3-18
Page 95
Error Messages
Error Messages in Alphabetical Order
POSSIBLE FALSE LOCK
Error Number 6
Phase lock has been achieved, but the source may be phase locked to the wrong harmonic of the synthesizer. Perform the source pre-tune correction routine documented in the “Adjustments and Correction Constants” chapter
HP 8753D Service Guide
in the
.
POWER UNLEVELED
Error Number 179
There is either a hardware failure in the source or you have attempted to set the power level too high. Check to see if the power level you set is within spec­ifications. If it is, refer to the
HP 8753D Service Guide
for troubleshooting. You will only receive this message over the HP-IB. On the analyzer, P? is dis­played.
POW MET INVALID
Error Number 116
The power meter indicates an out-of-range condition. Check the test setup.
POW MET NOT SETTLED
Error Number 118
Sequential power meter readings are not consistent. Verify that the equipment is set up correctly. If so, preset the instrument and restart the operation.
POW MET: not on, not connected, wrong addrs
Error Number 117
The power meter cannot be accessed by the analyzer. Verify that the power meter address and model number set in the analyzer match the address and model number of the actual power meter.
POWER SUPPLY HOT!
Error Number 21
The temperature sensors on the A8 post-regulator assembly have detected an over-temperature condition. The power supplies regulated on the post-regula­tor have been shut down.
POWER SUPPLY SHUT DOWN!
Error Number 22
3-19
Page 96
Error Messages
Error Messages in Alphabetical Order
One or more supplies on the A8 post-regulator assembly have been shut down due to an over-current, over-voltage, or under-voltage condition.
PRESS [MENU], SELECT CW (IF) FREQ, THEN SWEPT LO
Error Number 161
When you are sweeping the RF and LO, the IF must be fixed.
PRINT ABORTED
Error Number 25
When you press the
print color not supported with EPSON
LOCAL
key, the analyzer aborts output to the printer.
Error Number 178
You have defined the printer type as EPSON-P2. Color print is not supported with this printer. The print will abort.
PRINTER: busy
Error Number 176
The parallel port printer is not accepting data.
PRINTER: error
Error Number 175
The parallel port printer is malfunctioning. The analyzer cannot complete the copy function.
PRINTER: not connected
Error Number 173
There is no printer connected to the parallel port.
PRINTER: not handshaking
Error Number 177
The printer at the parallel port is not responding.
PRINTER: not on line
Error Number 172
The printer at the parallel port is not set on line.
3-20
Page 97
Error Messages
Error Messages in Alphabetical Order
PRINTER: not on, not connected, wrong addrs
Error Number 24
The printer does not respond to control. Verify power to the printer, and check the HP-IB connection between the analyzer and the printer. Ensure that the printer address recognized by the analyzer matches the HP-IB address set on the printer itself.
PRINTER: paper error
Error Number 171
There is a paper-related problem with the parallel port printer such as a paper jam or out-of-paper condition.
PRINTER: power off
Error Number 174
The power to the printer at the parallel port is off.
PRINT/PLOT IN PROGRESS, ABORT WITH LOCAL
Error Number 166
If a print or plot is in progress and you attempt a second print or plot, this message is displayed and the second attempt is ignored. To abort a print or plot in progress, press
PROBE POWER SHUT DOWN!
LOCAL
.
Error Number 23
The analyzer biasing supplies to the HP 85024A external probe are shut down due to excessive current. Troubleshoot the probe, and refer to the power sup­ply troubleshooting section of the
HP 8753D Service Guide
.
PROCESSING DISPLAY LIST
Information Message
The display information is being processed for a screen print to a copy device and stored in the copy spool buffer. During this time, the analyzer’s resources are dedicated to this task (which takes less than a few seconds.)
REQUESTED DATA NOT CURRENTLY AVAILABLE
Error Number 30
The analyzer does not currently contain the data you have requested. For example, this condition occurs when you request error term arrays and no cal­ibration is active.
3-21
Page 98
Error Messages
Error Messages in Alphabetical Order
SAVE FAILED. INSUFFICIENT MEMORY
Error Number 151
You cannot store an instrument state in an internal register due to insufficient memory. Increase the available memory by clearing one or more save/recall registers and pressing
SELECTED SEQUENCE IS EMPTY
PRESET
, or by storing files to a disk.
Error Number 124
The sequence you attempted to run does not contain instrument commands.
SELF TEST #n FAILED
Service Error Number 112
Internal test #n has failed. Several internal test routines are executed at instrument preset. The analyzer reports the first failure detected. Refer to the
HP 8753D Service Guide
for troubleshooting information on internal tests
and the self-diagnose feature.
SEQUENCE ABORTED
Error Number 157
The sequence running was stopped prematurely when you pressed the
LOCAL
key.
SEQUENCE MAY HAVE CHANGED, CAN’T CONTINUE
Error Number 153
When you pause a sequence, you cannot continue it if you have modified it. You must start the sequence again.
SLIDES ABORTED (MEMORY REALLOCATION)
Error Number 73
You cannot perform sliding load measurements due to insufficient memory. Reduce memory usage by clearing save/recall registers, then repeat the sliding load measurements.
SOURCE PARAMETERS CHANGED
Error Number 61
Some of the stimulus parameters of the instrument state have been changed, because you have turned correction on. A calibration set for the current mea­surement parameter was found and activated. The instrument state was updated to match the stimulus parameters of the calibration state.
3-22
Page 99
Error Messages
Error Messages in Alphabetical Order
This message also appears when you have turned on harmonic mode or fre­quency offset and the present frequency range cannot be used with one of these modes.
SOURCE POWER TRIPPED, RESET UNDER POWER MENU
Information Message
You have exceeded the maximum power level at one of the inputs and power has been automatically reduced. The annotation P⇓ indicates that power trip has been activated. When this occurs, reset the power and then press
POWER
SOURCE PWR on OFF
,
, to switch on the power. This message follows error
MENU
,
numbers 57, 58, and 59.
STARTING COPY SPOOLER
Information Message
The analyzer is beginning to output data from the spool buffer to the copy device. The analyzer resumes normal operation; the data is being output to the copy device in the background.
STOP/CW FREQ + OFFSET MUST BE < 3 GHz
Error Number 141
The output frequency of the mixer cannot violate the minimum/maximum fre­quency of the analyzer.
SWEEP MODE CHANGED TO CW TIME SWEEP
Error Number 187
If you select external source auto or manual instrument mode and you do not also select CW mode, the analyzer is automatically switched to CW.
SWEEP TIME INCREASED
Error Number 11
You have made instrument changes that cause the analyzer sweep time to be automatically increased. Some parameter changes that cause an increase in sweep time are narrower IF bandwidth, an increase in the number of points, and a change in sweep type.
SWEEP TIME TOO FAST
Error Number 12
The fractional-N and digital IF circuits have lost synchronization. Refer to the
HP 8753D Service Guide
for troubleshooting information.
3-23
Page 100
Error Messages
Error Messages in Alphabetical Order
SWEEP TRIGGER SET TO HOLD
Information Message
The instrument is in a hold state and is no longer sweeping.
SWEEP TYPE CHANGED TO LINEAR SWEEP
Error Number 145
If you have the frequency list mode active when you change the instrument mode to harmonic measurements, and the list frequencies do not fall in the allowable frequency range of these modes, the list mode automatically is turned off.
SYNTAX ERROR
Error Number 33
You have improperly formatted an HP-IB command. Refer to the
Programmer’s Guide
SYST CTRL OR PASS CTRL IN LOCAL MENU
for proper command syntax.
HP 8702D
Error Number 36
The analyzer is in talker/listener mode. In this mode, the analyzer cannot con­trol a peripheral device on the bus. Use the local menu to change to system controller or pass control mode.
TEST ABORTED
Error Number 113
You have prematurely stopped a service test.
THIS LIST FREQ INVALID IN HARM/3 GHZ RNG
Error Number 133
You have set frequencies in the list that are outside of the allowable frequency range for harmonic measurements, or are greater than 3 GHz on instruments without Option 006. Reduce the frequency range of the list.
TOO MANY NESTED SEQUENCES. SEQ ABORTED
Error Number 164
You can only nest sequences to a maximum level of six. The sequence will abort if you nest more than six.
3-24