Agilent 85033E User Manual

User’s and Service Guide
Agilent Technologies 85033E
3.5 mm Calibration Kit
Agilent Part Number: 85033-90028
Printed in USA
Print Date: May 2004
Supersedes: January 2004
Copyright © 2000–2002, 2004 Agilent Technologies, Inc. All rights reserved.
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent products.
For any assistance, contact Agilent Technologies. Refer to Table 5-1 on page 5-3 for contact information.
1. General Information
Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Incoming Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Recording the Device Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Clarifying the Sex of a Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
2. Specifications
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Temperature—What to Watch Out For . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Pin Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Contents
3. Use, Maintenance, and Care of the Devices
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Look for Obvious Defects and Damage First. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect the Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect Female Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Gaging Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Connector Gage Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
When to Gage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
How to Make a Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
How to Separate a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
4. Performance Verification
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
How Agilent Verifies the Devices in This Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
How Often to Recertify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Where to Send a Kit for Recertification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
5. Troubleshooting
Troubleshooting Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Returning a Kit or Device to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
iii
Contents
6. Replaceable Parts
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
A. Standard Definitions
Standard Class Assignments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Blank Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Nominal Standard Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-4
Blank Forms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
iv

1 General Information

85033E 1-1
General Information

Calibration Kit Overview

Calibration Kit Overview
The Agilent 85033E 3.5 mm calibration kit is used to calibrate Agilent network analyzers for measurements of components with 3.5 mm connectors up to 9 GHz.

Kit Contents

The 85033E calibration kit contains the following:
• offset opens and shorts, and broadband load terminations
• a data disk that contains the calibration definitions of the devices in the calibration kit for the PNA series
• a data disk that contains the calibration definitions of the devices in the calibration kit for the 875x and 872x series
• two open-short-load (OSL) holders
• 2.5 mm hex key for use with the (OSL) holder
• torque wrench
Refer to Figure 6-1 on page 6-2 and Table 6-1 on page 6-3 for part numbers of kit contents.
Broadband Loads
The broadband loads are metrology-grade, 50 ohm terminations which have been optimized for performance up to 9 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
Offset Opens and Shorts
The offset opens and shorts are built from parts which are machined to the current state-of-the-art in precision machining. The offset short’s inner conductors have a one-piece construction, common with the shorting plane. The construction provides for extremely repeatable connections. The offset opens have inner conductors which are supported by a strong, low-dielectric-constant plastic to minimize compensation values. Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.
Open-Short-Load Holders
The open-short-load (OSL) holders are included for your convenience. The OSL holders allow you to keep all of the calibration devices in two handy assemblies. Load each OSL holder with calibration devices of the same connector sex so that you can perform a calibration with just one assembly. Use the 2.5 mm hex key included with this kit to secure the calibration devices in the holders. See Figure 1-1 on page 1-3.
1-2 85033E
Figure 1-1 Open-Short-Load Holder
General Information
Calibration Kit Overview
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
• already resident within the analyzer
• loaded from the provided disks
• entered from the front panel
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.
85033E 1-3
General Information

Incoming Inspection

Options

There are several adapter options available for the 85033E. Refer to Table 6-2 on page 6-4. Also, there are calibration certificate options available. Contact the nearest Agilent office for more information. See Table 5-1 on page 5-3 for contact information.

Equipment Required but Not Supplied

Gages, electrostatic discharge (ESD) protection devices, and various connector cleaning supplies are not included with this calibration kit. Gage sets are required for measuring the connector pin depth. Refer to Table 6-3 on page 6-4 for ordering information.
Incoming Inspection
Verify that the shipment is complete by referring to Table on page 6-3.
Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case and its contents are not damaged.
If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent Technologies. See Table 5-1 on page 5-3. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
When you send the kit or device to Agilent, include a service tag (found near the end of this manual) with the following information:
• your company name and address
• the name of a technical contact person within your company, and the person's complete phone number
• the model number and serial number of the kit
• the part number and serial number of the device
• the type of service required
•a detailed description of the problem
1-4 85033E
General Information

Recording the Device Serial Numbers

Recording the Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in
Table . This can help you avoid confusing the devices in this kit with similar devices from
other kits.
The adapters included in the kit are for measurement convenience only and are not serialized.
Table 1-1 Kit and Device Serial Number Record
Device Serial Number
Calibration kit
Male Broadband Load
Female Broadband Load
Male Open
Female Open
Male Short
Female Sho rt
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
85033E 1-5
General Information

Clarifying the Sex of a Connector

Clarifying the Sex of a Connector
In this manual, calibration devices and adapters are referred to in terms of their connector interface. For example, a male open has a male connector.
However, during a measurement calibration, the network analyzer softkey menus label a calibration device with reference to the sex of the analyzer’s test port connector—not the calibration device connector. For example, the label SHORT(F) on the analyzer’s display refers to the short that is to be connected to the female test port. This will be a male short from the calibration kit.
Conversely, connector gages are referred to in terms of the connector that it measures. For instance, a male connector gage has a female connector on the gage so that it can measure male devices.
NOTE This calibration kit has the same standard definition for both male and
female standards. See Appendix A, “Standard Definitions,” for more information.

Preventive Maintenance

The best techniques for maintaining the integrity of the devices in the kit include:
• routine visual inspection
• cleaning
• proper gaging
• proper connection techniques
All of these are described in Chapter 3, “Use, Maintenance, and Care of the Devices.” Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-3
on page 2-4) or from bad connection techniques, can also damage these devices.
1-6 85033E

2 Specifications

85033E 2-1
Specifications

Environmental Requirements

Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Operating Temperature
Error-Corrected Temperature Range
Storage Temperature −40 °C to +75 °C (−40 °F to +167 °F)
Altitude
Operation < 4,500 meters (15,000 feet)
Storage < 15,000 meters (50,000 feet)
Relative Humidity Always Non-Condensing
Operation 0 to 80% (26 °C maximum dry bulb)
Storage 0 to 95%
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.
a
b
+15 °C to +35 °C (+59 °F to +95 °F)
±1 °C of measurement calibration temperature

Temperature—What to Watch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in
Table 2-1.
IMPORTANT Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2-2 85033E
Specifications

Mechanical Characteristics

Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
"Gaging Connectors," on page 3-7 explains how to use gages to determine if the kit devices
have maintained their mechanical integrity. (Refer to Table 2-3 on page 2-4 for typical and observed pin depth limits.)
Table 2-2 Mechanical Characteristics
Dimension Typical Value
Inside diameter of outer conductor 3.5 ±0.0025 mm
Outside diameter of center conductor 1.5199 ±0.002 mm

Pin depth

Pin depth
a. See
a
: male devices
a
: female devices
Figure 2-1 on page 2-4.
0 to 0.0127 mm
0.0025 to 0.0254 mm
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. See Figure 2-1. The pin depth of a connector can be in one of two states: either protruding or recessed. Protrusion is the condition when the center conductor extends beyond the outer conductor mating plane, and will measure a positive value on the connector gage. Recession is when the center conductor is set back from the outer conductor mating plane and will measure negative.
The pin depth value of each calibration device in this kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in this kit take into account the effect of pin depth on the device’s performance. Table 2-3 lists the typical pin depths and measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for a visual representation of proper pin depth (slightly recessed).
85033E 2-3
Specifications
Mechanical Characteristics
Figure 2-1 Connector Pin Depth
Table 2-3 Pin Depth Limits
Device T ypical Pin Depth
Opens 0 to 0.0127 mm
0 to 0.0005 in.
Shorts 0 to 0.0127 mm
0 to 0.0005 in.
Fixed Loads 0.0025 to 0.0254 mm
0.0001 to 0.001 in.
Adapters
(3.5 mm end)
Adapters (APC-7 end)
Adapters (type-N end)
a. Approximately +2 sigma to 2 sigma of gage uncertainty based on studies done at the factory
according to recommended procedures.
b. Observed pin depth limits are the range of observation limits seen on the gage reading due to
measurement uncertainty. The depth could still be within specifications.
0.0025 to 0.0508 mm
0.0001 to 0.002 in.
0 to -0.0508 mm 0 to -0.0020 in.
0 to -0.0127 mm 0 to -0.0005 in.
Measurement Uncertainty
+0.0064 to 0.0064 mm +0.00025 to 0.00025 in.
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in.
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in.
+0.0041 to 0.0041 mm +0.00016 to 0.00016 in.
+0.0038 to -0.0038 mm
-0.00015 to -0.00015 in.
+0.0038 to -0.0038 mm +0.00015 to -0.00015 in.
a
Observed Pin Depth Limit s
+0.0064 to 0.0191 mm +0.00025 to 0.00075 in.
+0.0041 to 0.0168 mm +0.0016 to 0.00066 in.
+0.0016 to 0.02953 mm +0.00006 to 0.00116 in.
+0.0016 to 0.0549 mm +0.00006 to 0.00216 in.
+0.0038 to -0.0546 mm +0.00015 to -0.00215 in.
+0.0038 to -0.0165 mm +0.00015 to -0.00065 in.
b
2-4 85033E
Specifications

Electrical Specifications

Electrical Specifications
The electrical specifications in Table 2-4 apply to the devices in your calibration kit when connected with an Agilent precision interface.
Table 2-4 Electrical Specifications for 3.5 mm Devices
Device Specification Frequency (GHz)
Broadband Loads Return Loss ≥46 dΒ (ρ ≤0.005) DC to ≤2
(male and female) Return Loss ≥44 dΒ (ρ ≤0.006) >2 to ≤3
Return Loss ≥38 dΒ (ρ ≤0.013) >3 to ≤9
Offset Opens
(male and female) ±0.65 ° from Nominal >2 to ≤3
Offset Shorts
(male and female) ±0.50 ° from Nominal >2 to ≤3
a. The specifications for the opens and shorts are given as allowed deviation from the
a
a
nominal model as defined in the standard definitions (see Table A-3 in Appendix A).
±0.55 ° from Nominal
±0.85 ° from Nominal >3 to 6
±1.00 ° from Nominal >6 to 9
±0.48 ° from Nominal
±0.55 ° from Nominal >3 to 6
±0.65 ° from Nominal >6 to 9
DC to ≤2
DC to ≤2
85033E 2-5
Specifications
Electrical Specifications

Supplemental Characteristics

Supplemental characteristics are provided as additional information that may be helpful in applying the devices. These characteristics are typical of most devices but are not warranted. Table 2-5 lists the typical characteristics of the adapters.
Table 2-5 Supplemental Electrical Characteristics of the Adapters
Adapter
Option 100
3.5 mm female to female R
Option 200
3.5 mm male to male R
Option 300
3.5 mm male to female R
Option 400
Type-N male to 3.5 mm male Return Loss ≥28 dΒ (ρ ≤0.040) DC to 18
Type-N male to 3.5 mm female Return Loss ≥28 dΒ (ρ ≤0.040) DC to 18
Type-N female to 3.5 mm female Return Loss ≥28 dΒ (ρ ≤0.040) DC to 18
Type-N female to 3.5 mm male Return Loss ≥24 dΒ (ρ ≤0.060) DC to 18
Option 500
3.5 mm male to APC-7 Return Loss ≥34 dΒ (ρ ≤0.20) DC to 18
3.5 mm female to APC-7 R
Return Loss, Typical
eturn Loss ≥32 dΒ (ρ ≤0.025) DC to 26.5
eturn Loss ≥32 dΒ (ρ ≤0.025) DC to 26.5
eturn Loss ≥32 dΒ (ρ ≤0.025) DC to 26.5
eturn Loss ≥28 dΒ (ρ ≤0.040) DC to 18
Frequency (GHz)
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See "How Agilent Verifies the
Devices in This Kit," on page 4-2 for more information.
2-6 85033E
3 Use, Maintenance, and Care of the
Devices
85033E 3-1
Use, Maintenance, and Care of the Devices

Electrostatic Discharge

Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
always wear a grounded wrist strap having a 1 M resistor in series with it when handling components and devices or when making connections to the test set.
always use a grounded, conductive table mat while making connections.
always wear a heel strap when working in an area with a conductive floor. If you are uncertain about the conductivity of your floor, wear a heel strap.
always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
always ground the center conductor of a test cable before making a connection to the analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 6, “Replaceable Parts,” for part numbers and instructions for ordering ESD protection devices.
Figure 3-1 ESD Protection Setup
3-2 85033E
Use, Maintenance, and Care of the Devices

Visual Inspection

Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
In some cases, magnification is necessary to see damage to a connector; a magnifying device with a magnification of ≥10× is recommended. However, not all defects that are visible only under magnification will affect the electrical performance of the connector. Use the following guidelines when evaluating the integrity of a connector.

Look for Obvious Defects and Damage First

Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are not properly connected.
Connector wear eventually degrades the performance of the device. Calibration devices should have a long life if their use in on the order of a few times per week.
The test port connectors on the network analyzer test set may have many connections each day, and are therefore more subject to wear. It is recommend that an adapter be used as a test port saver to minimize the wear on the test set’s connectors. Replace devices with worn connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. See Figure 3-2. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance. If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again.
85033E 3-3
Use, Maintenance, and Care of the Devices
Visual Inspection
Figure 3-2 Mating Plane Surfaces

Inspect Female Connectors

When using female connectors, pay special attention to the contact fingers in the female center conductor. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will not make good electrical contact and must be replaced.
NOTE This is particularly important when you are mating nonprecision to precision
devices.
3-4 85033E
Use, Maintenance, and Care of the Devices

Cleaning Connectors

Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections. Refer to Table 6-3 on page 6-4 for part numbers for isopropyl alcohol and cleaning swabs.
WARNING Always use protective eyewear when using compressed air or
nitrogen.
1. Use Compressed Air or Nitrogen
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. Clean air cannot damage a connector or leave particles or residues behind.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” on
page 3-2 for additional information.
WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a
tightly closed container. Isopropyl alcohol is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide: water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Keep away from heat, sparks, and flame. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
85033E 3-5
Use, Maintenance, and Care of the Devices
Cleaning Connectors
2. Clean the Connector Threads
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-3 on page 6-4 for isopropyl alcohol and cleaning swab part numbers.
a. Apply a small amount of isopropyl alcohol to the lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a new swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 3-2 on
page 3-4. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Inspect Each Connector
Inspect the connector again to make sure that no particles or residue are present.
3-6 85033E
Use, Maintenance, and Care of the Devices

Gaging Connectors

Gaging Connectors
The gages available from Agilent Technologies (see Table 6-3 on page 6-4 for part number information) are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices. Do not use the
gages for precise pin depth measurements.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the repeatability uncertainties that are associated with the measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteristics of the devices.
With proper technique, however, the gages are useful in detecting gross pin depth errors on device connectors. To achieve maximum accuracy, random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as ± 0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-3 on page 2-4 assumes new gages and gage masters. Therefore, these systematic errors were not included in the uncertainty analysis. As the gages endure more use, the systematic errors can become more significant in the accuracy of the measurement.
The measurement uncertainties are primarily a function of the assembly materials and design, and the unique interaction each device type has with the gage. Therefore, these uncertainties can vary among the different devices. For example, note the difference between the uncertainties of the opens and shorts in Table 2-3 on page 2-4.
The observed pin depth limits add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages. See“Pin
Depth” on page 2-3. Refer to "Kit Contents," on page 1-2 for more information on the design
of the calibration devices in this kit.
When measuring pin depth, the measured value (resultant average of three or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-3 on page 2-4 to evaluate the condition of device connectors.
85033E 3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors

When to Gage Connectors

Gage a connector at the following times:
• Prior to using a device for the first time: record the pin depth measurement so that it can be compared with future readings. This serves as a good troubleshooting tool when you suspect damage may have occurred to the device.
• If either visual inspection or electrical performance suggests that the connector interface may be out of typical range (due to wear or damage, for example).
• If a verification device is used by someone else or on another system or piece of equipment.
Initially after every 100 connections, and after that as often as experience indicates.
3-8 85033E
This page intentionally left blank.
Use, Maintenance, and Care of the Devices
Gaging Connectors
85033E 3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors

Gaging Procedures

Gaging Male 3.5 mm Connectors
NOTE Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
Select the proper gage for your connector. Refer to Table 6-3 on page 6-4 for gage part numbers.
5. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” on page 3-3 and “Cleaning Connectors” on page 3-5.
6. Zero the connector gage (refer to Figure 3-3 on page 3-11):
a. While holding the gage by the barrel, and without turning the gage or the device,
screw the male gage master connecting nut onto the male gage, just until you meet resistance. Connect the nut finger tight. Do not overtighten.
b. Use the torque wrench recommended for use with the kit to tighten the connecting
nut to 90 N-cm (8 in-lb). Refer to “Connections” on page 3-14 for more information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with zero.
d. Remove the gage master.
7. Gage the device connector (refer to Figure 3-3 on page 3-11):
a. While holding the gage by the barrel, and without turning the gage or the device,
screw the connecting nut of the male device being measured onto the male gage, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with the kit to tighten the connecting
nut to 90 N-cm (8 in-lb). Refer to “Connections” on page 3-14 for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read only the black ± signs; not the red ± signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
Compare the average reading with the observed pin depth limits in Table 2-3 on page 2-4.
3-10 85033E
Figure 3-3 Gaging Male 3.5 mm Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85033E 3-11
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Female 3.5 mm Connectors
NOTE Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
Select the proper gage for your connector. Refer to Table 6-3 on page 6-4 for gage part numbers.
8. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” on page 3-3and “Cleaning Connectors” on page 3-5 earlier in this chapter.
9. Zero the connector gage (refer to Figure 3-4 on page 3-13):
a. While holding the gage by the barrel, and without turning the gage or the device,
screw the female gage connecting nut onto the female gage master, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with the kit to tighten the connecting
nut to 90 N-cm (8 in-lb). Refer to “Connections” on page 3-14 for more information.
c. As you watch the gage pointer, gently tap the barrel of the gage to settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with zero.
d. Remove the gage master.
10.Gage the device connector (refer to Figure 3-4 on page 3-13):
a. While holding the gage by the barrel, and without turning the gage or the device,
screw the female gage connecting nut onto the female device being measured, just until you meet resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with the kit to tighten the connecting
nut to 90 N-cm (8 in-lb). Refer to “Connections” on page 3-14 for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. Read only the black ± signs; not the red ± signs.
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. Use different orientations of the gage within the connector. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
Compare the average reading with the observed pin depth limits in Table 2-3 on page 2-4.
3-12 85033E
Figure 3-4 Gaging Female 3.5 mm Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85033E 3- 13
Use, Maintenance, and Care of the Devices

Connections

Connections
Good connections require a skilled operator. The most common cause of measurement error is poor connections.The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on an antistatic mat. Refer to “Electrostatic Discharge,” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-5.
Use a connector gage to verify that all center conductors are within observed pin depth values in Table 2-3 on page 2-4.
4. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
5. Push the connectors straight together. Do not twist or screw them together. As the center conductors mate, there is usually a slight resistance.
CAUTION Do not turn the device body. Only turn the connector nut. Damage to the
center conductor can occur if the device body is twisted.
6. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough.
7. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
3-14 85033E
Use, Maintenance, and Care of the Devices
Connections
Final Connection Using a Torque Wrench
NOTE Using a torque wrench guarantees that the connection is not too tight,
preventing possible connector damage. It also guarantees that all connections are equally tight.
Table 3-1 Torque Wrench Information
Connector Type Torque Setting Torque Tolerance
3.5 mm 90 N-cm (8 in-lb) ±9.0 N-cm (±0.8 in-lb)
1. Turn the connector nut. This may be possible to do by hand if one of the connectors is fixed (as on a test port). In all situations, the use of an open-end wrench to keep the body of the device from turning is recommended.
2. Position both wrenches within 90 degrees of each other before applying force. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action that can misalign and stress the connections of the device involved This is especially true when several devices are connected together. See Figure 3-5.
Figure 3-5 Wrench Positions
85033E 3- 15
Use, Maintenance, and Care of the Devices
Connections
3. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
Figure 3-6.
Figure 3-6 Using the Torque Wrench
4. Apply force downward to the wrench handle. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.
CAUTION You don’t have to fully break the handle of the torque wrench to reach the
specified torque; doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.
5. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-6. Do not tighten the connection further.
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an unknown amount of torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more than the recommended torque.

How to Separate a Connection

To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.
CAUTION Turn the connector nut, not the device body. Major damage to the center
conductor can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending.
3-16 85033E
Use, Maintenance, and Care of the Devices

Handling and Storage

Handling and Storage
• Use the plastic end caps and store the calibration devices in the foam-lined storage case when not in use.
• Never store connectors loose in a box, in a desk, or in a bench drawer. This is the most common cause of connector damage during storage.
• Keep connectors clean.
• Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily transferred to a connector interface and are very difficult to remove.
• Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
85033E 3- 17
Use, Maintenance, and Care of the Devices
Handling and Storage
3-18 85033E

4 Performance Verification

85033E 4-1
Performance Verification

Introduction

Introduction
The performance of your calibration kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices inside the kit has been specially manufactured and is not commercially available.

How Agilent Verifies the Devices in This Kit

Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to NIST (National Institute of Standards and Technology). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced back to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Agilent includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass/fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.
4-2 85033E
Performance Verification

Recertification

Recertification
The following will be provided with a recertified kit:
• new calibration sticker affixed to the case
• certificate of calibration
• a calibration report for each device in the kit listing measured values, specification, and uncertainties
NOTE A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. For contact information, see Table 5-1 on
page 5-3.

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.
NOTE The recertification interval should begin on the date the kit is first used after
the recertification date.

Where to Send a Kit for Recertification

Contact Agilent Technologies for information on where to send your kit for recertification. For contact information, see Table 5-1 on page 5-3.
When you return the kit, fill out and attach a service tag. Refer to "Returning a Kit or
Device to Agilent," on page 5-3 for details.
85033E 4-3
Performance Verification
Recertification
4-4 85033E

5 Troubleshooting

85033E 5-1
Troubleshooting

Troubleshooting Process

Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5-2 85033E
Troubleshooting

Returning a Kit or Device to Agilent

Returning a Kit or Device to Agilent
If your kit or device requires service, contact Agilent Technologies for information on where to send it. See Table 5-1. Include a service tag (found at the end of this manual) on which you provide the following information:
• your company name and address
• a technical contact person within your company, and the person's complete phone number
• the model number and serial number of the kit
• the part number and serial number of each device
• the type of service required
•a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)

Contacting Agilent

By internet, phone, or fax, get assistance with all your test and measurement needs.
Table 5-1 Contacting Agilent
Online assistance: www.agilent.com/find/assist
Americas
Brazil
(tel) (+55) 11 4197 3600 (fax) (+55) 11 4197 3800
Australia
(tel) 1800 629 485 (alt) 1800 143 243 (fax) 1800 142 134
Japan
(tel) 0120 421 345 (alt) (+81) 426 56 7832 (fax) 0120 421 678
Canada
(tel) 877 894 4414 (fax) (+1) 905 282-6495
Asia Pacific and Japan
China
(tel) 800 810 0189 (alt) (+86) 10800 650 0021 (fax) 800 820 2816
Malaysia
(tel) 1800 888 848 (alt) 1800 828 848 (fax) 1800 801 664
Mexico
(tel) (+52) 55 5081 9469 (alt) 01800 5064 800 (fax) (+52) 55 5081 9467
Hong Kong (tel) 800 930 871 (alt) (+852) 3197 7889 (fax) (+852) 2 506 9233
Singapore
(tel) 1800 375 8100 (alt) (+65) 6 375 8100 (fax) (+65) 6836 0252
United States
(tel) 800 829 4444 (alt) (+1) 303 662 3998 (fax) 800 829 4433
India
(tel) 1600 112 929 (fax) 000800 650 1101
South Korea
(tel) 080 769 0800 (alt) (+82) 2 2004 5004 (fax) (+82) 2 2004 5115
Taiwan
(tel) 0800 047 866 (alt) 00801 651 317 (fax) 0800 286 331
Thailand
(tel) 1800 226 008 (alt) (+66) 2 268 1345 (fax) (+66) 2 661 3714
85033E 5-3
Troubleshooting
Returning a Kit or Device to Agilent
Table 5-1 Contacting Agilent
Europe
Austria
(tel) 0820 87 44 11* (fax) 0820 87 44 22
France
(tel) 0825 010 700* (alt) (+33) (0)1 6453 5623 (fax) 0825 010 701*
Italy
(tel) (+39) (0)2 9260 8484 (fax) (+39) (0)2 9544 1175
Spain
(tel) (+34) 91 631 3300 (alt) (+34) 91 631 3000 (fax) (+34) 91 631 3301
Switzerland (Italian)
(tel) 0800 80 5353 opt. 3* (alt) (+39) (0)2 9260 8484 (fax) (+41) (0)22 567 5314
Belgium
(tel) (+32) (0)2 404 9340 (alt) (+32) (0)2 404 9000 (fax) (+32) (0)2 404 93 95
Germany
(tel) 01805 24 6333* (alt) 01805 24 6330* (fax) 01805 24 6336*
Luxemburg
(tel) (+32) (0)2 404 9340 (alt) (+32) (0)2 404 9000 (fax) (+32) (0)2 404 93 95
Sweden
(tel) 0200 88 22 55* (alt) (+46) (0)8 5064 8686 (fax) 020 120 2266*
United Kingdom
(tel) (+44) (0)7004 666666 (alt) (+44) (0)7004 123123 (fax) (+44) (0)7004 44 4555
Denmark
(tel) (+45) 7013 1515 (alt) (+45) 7013 7313 (fax) (+45) 7013 1555
Ireland
(tel) (+353) (0)1 890 924 204 (alt) (+353) (0)1 890 924 206 (fax)(+353) (0)1 890 924 024
Netherlands
(tel) (+31) (0)20 547 2111 (alt) (+31) (0)20 547 2000 (fax) (+31) (0)20 547 2190
Switzerland (Frenc h)
(tel) 0800 80 5353 opt. 2* (alt) (+33) (0)1 6453 5623 (fax) (+41) (0)22 567 5313
Finland
(tel) (+358) 10 855 2100 (fax) (+358) 10 855 2923
Israel
(tel) (+972) 3 9288 500 (fax) (+972) 3 9 288 501
Russia
(tel) (+7) 095 797 3963 (alt) (+7) 095 797 3900 (fax) (+7) 095 797 3901
Switzerland (German)
(tel) 0800 80 5353 opt. 1* (alt) (+49) (0)7031 464 6333 (fax) (+41) (0)1 272 7373
(tel) = primary telephone nu mber; (alt) = alternate telephone number; (fax) = FAX nu mber; * = in country number
04/04
5-4 85033E

6 Replaceable Parts

85033E 6-1
Replaceable Parts

Introduction

Introduction
Table 6-1 lists the replacement part numbers for items included in the 85033E calibration
kit and Figure 6-1 illustrates each of these items.
Table 6-2 lists the replacement part numbers for adapters that are available as options.
Table 6-3 lists the replacement part numbers for items not included in the calibration kit
that are either required or recommended for successful operation of the kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Agilent Technologies. For contact information, see Table
5-1 on page 5-3.
Figure 6-1 Replaceable Parts for the 85033E Calibration Kit
6-2 85033E
Replaceable Parts
Table 6-1 Replaceable Parts
Item No. Description Qty Agilent Part Number
Introduction
Calibration Devices
1 3.5 mm male broadband load 1 85033-60016
2 3.5 mm female broadband load 1 85033-60017
3 3.5 mm male offset open 1 85033-60018
4 3.5 mm female offset open 1 85033-60019
5 3.5 mm male offset short 1 85033-60020
6 3.5 mm female offset short 1 85033-60021
Protective End Caps for Connectors
7 3.5 mm male connectors as
8 3.5 mm female connectors as
Calibration Kit Storage Case
9 Box (without foam pads) 1 5180-8446
10 Box pad (top) 1 85032-80028
11 Box pad (bottom) 1 85033-80025
a
a
1401-0208
required
1401-0202
required
Disk holder (not shown in
Connector care-quick reference card (not shown in
12 5/16 in, 90 N-cm (8 in-lb) torque wrench 1 8710-1765
13 Open-short-load holder 2 85033-40001
14 2.5 mm hex key 1 8710-1181
15 Calibration definitions disk (PNA series) 1 85033-10011
16 Calibration definitions disk (875x and 872x series) 1 85033-10012
17 User’s and service guide 1 85033-90028
a. See “Clarifying the Sex of a Connector” on page 1-6.
Figure 6-1)
Figure 6-1)
Miscellaneous Items
1 5180-8491
1 08510-90360
85033E 6-3
Replaceable Parts
Introduction
Table 6-2 Adapters
Description Qty Agilent Part Number
Option 100
3.5 mm female to female
Option 200
3.5 mm male to male 1 85027-60007
Option 300
3.5 mm female to male 1 85027-60006
Option 400
Type-N male to 3.5 mm male 1 1250-1743
Type-N male to 3.5 mm female 1 1250-1744
Type-N female to 3.5 mm female 1 1250-1745
Type-N female to 3.5 mm male 1 1250-1750
Option 500
3.5 mm male to APC7 2 1250-1746
3.5 mm female to APC7 2 1250-1747
a. See “Clarifying the Sex of a Connector” on page 1-6.
a
1 85027-60005
a
a
a
a
Table 6-3 Items Not Included in Kit
Description Agilent Part Number
3.5 mm gage set (for female connectors) 85052-60042
3.5 mm gage set (for male connectors) 85052-60043
Isopropyl alcohol (30 ml) 8500-5344
Cleaning swabs (100) 9301-1243
Grounding wrist strap 9300-1367
5 ft. grounding cord for wrist strap 9300-0980
2 x 4 ft. conductive table mat and 15 ft. ground wire 9300-0797
ESD heel strap 9300-1308
6-4 85033E

A Standard Definitions

85033E A-1
Standard Definitions

Standard Class Assignments

Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Table A-1 lists the classes of the devices in this calibration kit. This information resides on the calibration constants disk that accompany this kit.
Table A-1 Standard Class Assignments
Disk File Name for 8752A/B: A5235E Disk File Name for 8753B/C: C5335E
Class A B C D E F G Standard Class
S
A2 Open
11
S
B1 Short
11
S
C3 Loads
11
S
A2 Open
22
S
B1 Short
22
S
C3 Loads
22
Forward Transmission 4 Thru
Reverse Transmission 4 Thru
Forward M atc h 4 Thru
Reverse Match 4 Thru
Response 1 2 4 Response
Response and Isolation 1 2 4 Response
Calibration Kit Label:
3.5 mmE
Label
A-2 85033E
Standard Definitions
Standard Class Assignments

Blank Form

The blank form (Table A-2) is provided for use in modifying the standard class assignments to meet your individual requirements.
Table A-2 Standard Class Assignments Blank Form
Disk File Name:________________ ____ Calibration Kit Label:________ _______
Class A B C D E F G Standard Class
Label
S
A
11
S
B
11
S
C
11
A
S
22
S
B
22
S
C
22
Forward Transmission
Reverse Transmission
Forward M atc h
Reverse Match
Response
Response and Isolation
85033E A-3
Standard Definitions

Nominal Standard Definitions

Nominal Standard Definitions
Standard definitions provide the constants needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these constants are theoretically derived from the physical dimensions and material of each calibration standard, or from actual measured response. These values are used to determine the measurement uncertainties of the network analyzer. The standard definitions in Table A-3 lists typical calibration kit parameters used to specify the mathematical model of each device. This information must be loaded into the network analyzer in order to perform valid calibrations. Refer to your analyzer user’s guide for information on how to load calibration constants.
NOTE The values in the standard definitions table are valid only over the specified
operating temperature range.
Table A-3 Standard Definitions
Standard
Number
Short
Thru
a
= 50
0
b
F
15
C0 ×10
H
12
Type
L0 ×10
d
2.0765 108.54 2.1705 0.01 31.088 50 2.36 0 999 Coax Short
F/Hz
27
C1 ×10
H/Hz
24
L1 ×10
2
36
2
33
F/Hz
C2 ×10
H/Hz
L2 ×10
3
F/Hz
45
C3 ×10
3
H/Hz
42
L3 ×10
Calibration Kit Label:
Offset Frequency
c
Fixed or Sliding
Delay in ps
0 50 0 0 999 Coax Thru
0
Z
3.5 mmE
Loss in GΩ/s
in GHz
Min
Max
Coax or Waveguide
Standard Label
band
System Z
1
2 Open 49.43 310.13 23.17 0.16 29.243 50 2.2 0 999 Coax Open
3 Load Fixed 0 50 0 0 999 Coax Broad-
4Delay/
5
6
7
8
a. Ensure system Z b. Open, short, load, delay/thru, or arbitrary impedance.
c. Load or arbitrary impedance only. d. L terms apply to PNA and ENA series of network analyzers.
of network analyzer is set to this value.
0
A-4 85033E
Standard Definitions
Nominal Standard Definitions

Blank Forms

The blank form (Table A-4) is provided for use in modifying the standard definitions to meet your individual requirements.
Table A-4 Standard Definitions Blank Form
System Z
a
= ________________________
0
Disk File Name:_______________________
Type
Short
b
F
15
C0 ×10
H
12
L0 ×10
d
F/Hz
27
C1 ×10
H/Hz
24
L1 ×10
Standard
Number
1
2Open
3 Load
4Delay/
Thru
5
6
2
36
2
33
F/Hz
C2 ×10
H/Hz
L2 ×10
Calibration Kit Label: Disk File Name:_______________________
3
F/Hz
45
Offset Frequency
in GHz
C3 ×10
3
c
H/Hz
42
0
L3 ×10
Fixed or Sliding
Delay in ps
Loss in GΩ/s
Z
Min
Max
Coax or Waveguide
STandard Label
7
8
a. Ensure system Z
of network analyzer is set to this value.
0
b. Open, short, load, delay/thru, or arbitrary impedance. c. Load or arbitrary impedance only. d. L terms apply to PNA and ENA series of network analyzers.
85033E A-5
Standard Definitions
Nominal Standard Definitions
A-6 85033E

Index

A
adapters Agilent Technologies, contacting, 5-3 altitude, 2-2
B
blank form
standard class assignments standard definitions, A-5
broadband loads, 1-2
C
calibration
certificate of temperature, 2-2
calibration definitions
permanently stored
calibration kit
contents overview, 1-2
calibration report, 4-2 certificate of calibration, 4-2 characteristics
electrical, 2-6 mechanical, 2-3 supplemental, 2-6
class assignments
blank form, A-3
class assignments, standard, A-2 cleaning connectors, 3-5 compressed air or nitrogen, 3-5 connections, 3-2, 3-14, 3-15, 3-16 connector
cleaning female, 3-12 gaging, 3-7
male, 3-10 sex, 1-6
slotted, 3-4 threads, 3-5 visual inspection, 3-3 wear, 3-3
connector gage
handling zeroing, 3-10, 3-12
contacting Agilent Technologies, 5-3 contents
incomplete
, 6-3
, 4-2
, 1-2
, 3-5
when to do, 3-8
clarifying, 1-6
, 3-10, 3-12
what to do
, 1-4
, A-3
, 1-3
entering, 1-3 permanently stored, 1-3
standard, A-4
deviation from nominal phase, 2-5 device
connecting, 3-14 disconnecting, 3-16 handling, 3-17 maintenance, 1-6 storage, 3-17 temperature, 2-2 visual inspection, 3-3
dimensions, 2-3 disconnections, 3-16 documentation warranty, -ii
E
electrical characteristics electrical specifications, 2-5 electrostatic discharge, 3-2 environmental requirements, 2-2 equipment required, 1-4 ESD, 3-2
precautions, 3-2
G
gage
connector
handling
zeroing, 3-10, 3-12
gaging
connectors
when to do
female connectors, 3-12 male connectors, 3-10
gaging connectors, 3-7
H
handling humidity, 2-2
I
incoming inspection inspection
damage defects, 3-3 incoming, 1-4 mating plane, 3-3 slotted connectors, 3-4 visual, 3-3
, 3-10, 3-12
, 3-8
, 3-17
, 3-3
, 2-6
, 1-4
D
damage
what to do
damaged connectors, 3-3 data, recertification, 4-2 defective connectors, 3-3 definitions
calibration
85058B/E
, 1-4
, 1-3
K
kit contents, calibration kit overview, contents, 1-2
L
loads, broadband, 1-2
, 1-2
Index-1
Index
M
maintenance
preventive, 1-6
maintenance of devices, 1-6 making connections, 3-14 mating plane inspection, 3-3 mechanical characteristics, 2-3
N
nitrogen nominal standard definitions, A-4
O
offsets opens and shorts, 1-2 options, 1-4 ordering parts, 6-1 overview of kit, 1-2
P
part numbers parts, replacing, 6-1 performance verification, 4-2 permanently stored calibration definitions, 1-3 pin depth, 2-3 preventive maintenance, 1-6
, 3-2
, 3-5
, 1-2
, 6-2
calibration device, 2-2 error-corrected temperature range, 2-2 measurement, 2-2 operating range, 2-2
verification and measurement, 2-2 test data, 4-2 torque wrench, 3-15 traceability, 4-2 troubleshooting, 5-2
V
verification
performance
temperature, 2-2 visual inspection, 3-3
W
warranty, documentation, -ii
Z
zeroing
connector gage
, 2-2
, 4-2
, 3-10, 3-12
R
recertification
interval, 4-3
where it's done, 4-3 replaceable parts, 6-1, 6-2 requirements, environmental, 2-2 return loss, 2-5 return, kit or device, 5-3
S
serial numbers service
, 1-4
tag slotted connectors, 3-4
specifications, 2-2
electrical, 2-5
environmental, 2-2
temperature, 2-2
verifying, 4-2 standard class assignments, A-2
blank form, A-3 standard definitions, A-4, A-5
blank form, A-5
nominal, A-4 static discharge, 3-2
storage, 3-17 supplemental characteristics, 2-6
, 4-2
, 1-5
T
tag
service temperature
Index-2
, 1-4
85058B/E
Loading...