This manual applies directly to 85029B verification kits that have
serial number prefix 2818A. Verification kits with a higher serial
prefix number are different than the kits documented in this
manual, and are supplied with a manual-change supplement which
documents the differences.
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS.
FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW,
AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT
SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD
AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent products.
For any assistance, contact Agilent Technologies. Refer to page 6 for a list of Agilent
contacts.
ii85029B
85029B 7 mm Verification Kit
Agilent 85029B Verification Kit1
85029B 7 mm Verification Kit
Product Description
Product Description
Intended Uses
System Verification
Measuring the devices of this kit and comparing the results with supplied data provides a
tops-down verification of an Agilent 8753A/B network analyzer with a 7-mm 50 ohm
test-set. The frequency range covered by the Agilent 85029B verification kit is 300 kHz to
6 GHz. The system must be calibrated before performing a system verification.
A tops-down verification is a procedure in which the devices in a verification kit are
measured on the 8753A/B system. These measured values are then compared with the
NBS traceable data that Agilent supplies with each verification kit. The 85029B
verification kit contains three verification devices for this purpose. The kit contains a
7-mm 20 dB attenuator, a 7-mm 50 dB attenuator, and a 7 mm mismatch attenuator. NIST
traceable magnitude and phase data is provided with the 85029B verification kit at
specific frequencies from 300 kHz to 6 GHz for all four S-parameters.
The uncertainty limits provided with these verification devices are the worst case sum of
the measurement uncertainties of the factory 85029B production test system, plus the
measurement uncertainties of the system being verified. If the measured data for the
verification devices falls within the supplied uncertainty limits, then the measurement
system passes its performance verification.
Residual errors of the measurement system are, in part, due to the characteristics of the
calibration devices used during system calibration. You should use the same calibration kit
that will be used when calibrating the system for actual measurements. Using a different
calibration kit can cause inconsistent performance.
As an example, a user calibrates the system with a calibration kit that is in excellent
condition and the system passes the verification procedure easily. The system is then sent
to production and is calibrated using a worn or damaged calibration kit. In this case, the
residual errors of the system could fall outside the allowable uncertainty limits.
Each 85029B verification kit is measured at the factory in a tightly controlled environment
using an 8753 network analyzer. The test system is calibrated with special NIST-traceable
standards, allowing the accuracy of the test system to exceed that of a standard 8753. This
system is verified daily, using a bottoms-up procedure that quantifies individual sources of
measurement error.
Refer to Table 1 on page 3 to see which measurements are the basis for system verification.
2Agilent 85029B Verification Kit
85029B 7 mm Verification Kit
Recommended Practices
Table 1 Measurements Used for System Verification
Verification DeviceS11/S22S12/S21
MagnitudePhaseMagnitudePhase
20 dB attenuatorxxx
60 dB attenuatorxx
mismatch attenuatorxx
PASS/FAIL information is displayed only for the most useful measurements, as indicated
in Table 1. However, all four S-parameters are measured. Additionally, the uncertainties
calculated for each device are included in printed form.
Statistical quality control of instrument performance
The 85029B allows the user to periodically record measured data. This history data is
useful for statistical quality control of instrument performance. Also, unusual deviations
from historical data can warn the operator of possible degradation in instrument or
verification kit performance.
Application Systems
The 85029B 7-mm verification kit has been designed to check the performance of:
• 8753, or 8720D and its test sets
• 85044A, 85046A, and 85047A
Recommended Practices
To maintain optimum performance from this verification kit, observe these simple
precautions:
• Always place the verification devices back in the protective case between uses.
• Make connections carefully to avoid misalignment, leading to inaccurate measurements
of connector damage.
• Keep the connectors free of dirt and metallic particles.
• If you must clean the connectors, use clean compressed air first. NEVER use abrasives.
When cleaning connectors, use liquid Freon exclusively and apply with a plastic swab.
For more information, refer to Principles of Microwave Connector Care Quick Reference Card (part number 08510-90360).
Agilent 85029B Verification Kit3
85029B 7 mm Verification Kit
General Information
General Information
Weight
Net: 0.9 kg (2 lb)
Shipping: 1.4 kg (3 lb)
Recommended Recertification Period
Agilent recommends that the verification kit be recertified annually; contact Agilent
Technologies for details (refer to “Contacting Agilent” on page 6).
What This Kit Contains
Verification devices
The kit contains precision standard devices. The storage case is supplied to protect the
devices from damage that will degrade accuracy. Place the devices in the case immediately
after use.
Measured data and uncertainty limits are supplied on disc. In order to get a hardcopy, you
must order Option UK6.
Operation
System Verification Procedure
The system verification procedure is located in the on-site system service manual of your
particular instrument.
Making a Copy of the Master Data Disk
This disk contains data files which are unique to the verification standards shipped with
your 85029B verification kit. Agilent recommends that you make a copy of the master data
disk for everyday use, and store the master disk in a safe place.
Making a backup disk using the 8753 is time consuming, as many files must be copied
individually.
4Agilent 85029B Verification Kit
85029B 7 mm Verification Kit
Replaceable Parts
Replaceable Parts
The three attenuators are separately available and should be ordered by the numbers
given below. Each of these devices has a serial number and the kit has a serial number. All
four serial numbers appear on the verification disk label.
Table 2 Replaceable Parts
DescriptionPart Number
7 mm mismatch attenuator
7-mm 20 dB attenuator
7-mm 50 dB attenuator
Operating and Service Manual85029-90010
Instrument case85029-80010
Foam pads85029-80007
a. Replacement part number, includes disk with
device-specific data.
85029-60004
85029-60005
85029-60006
a
a
a
Replacing the Master Disk
The 85029B master disk contains unique data that applies to the individual verification
devices. No two devices have the same performance data. It is not a trivial matter to
replace a lost or damaged master disk, so it is important to make one or more backup
copies of the master disk.
If your master data disc is lost or damaged, and you have no backup copies, take one of the
following actions:
• If Recertification is not required in the near future. Contact your nearest Agilent
service office for a replacement. Please specify the serial number of the verification kit,
and if it has been recertified. If a recertification has been performed, please indicate
which service office performed it.
• If recertification will be required soon. Agilent recommends that you have the
verification kit recertified early. A new master disk will be generated during the
recertification process.