Agilent 8501A Data Sheet

High-end Analyzer at a Low-end Price
R2590 Communications System Analyzer
2
The R2590 system analyzer sets a new standard for value in two-way communications testing. The unit’s light weight (28 lbs.), small footprint (7” x 13.5” x 12”) and optional battery with two hours of run time combine to make it as valuable in the field as on the bench.
Here are just a few of the features that are standard on every R2590:
1GHz RF Signal Generator with AM/FM/PM
modulation, 1Hz resolution, and output range from –10dBm to –125dBm
Duplex Generator with a fully adjustable
(0 Hz to 1GHz) frequency offset
1GHz RF Receiver with 2µV sensitivity and 1Hz
resolution, frequency error meter with actual fre­quency display, and AM/FM demodulation meters
Broadband and Selective Power Meters
1GHz Spectrum Analyzer with powerful
analytical tools usually found only in expensive stand-alone instruments
50kHz Oscilloscope with voltage and frequency
measurements
Audio Modulation Synthesizer, including:
Dual tone generator, sub-audible tone generator, DTMF generator, and external modulation input
Signal Strength Indicator
SINAD Meter with AUTO Mode
CTCSS, DCS, and DTMF Decoder
6.4” color LCD Display with external VGA output
RS232 Port for computer control and “flash”
software updates
Additional optional features include:
CLEARCHANNEL
LTR
Paging encoder: 2 tone, 5/6 tone, programmable
sequence, POCSAG
Tracking GeneratorCable Fault/Antenna Analyzer with Return
Loss Bridge
Internal battery with up to 2 hours of run time
Powerful, portable, and priced to fit within your budget, the R2590 is the new standard for two-way communica­tions service monitors.
Big solutions on a small budget.
R2590 Communications System Analyzer
3
Standard System Features
Feature Description Benefits
Tracking Generator (option)
View the performance characteristics of many RF devices with this convenient tracking generator. A sweep range from 5 KHz to 1 GHz, with a resolution of 1 Hz, permits accurate device alignment.
Allows the operator to quickly and easily diagnose critical problems with receiver front ends, IFs, helical filters, cavities, combiners and duplexers.
Full Featured Spectrum Analyzer
The R2590 comes complete with a powerful, yet very easy to use spectrum analyzer. Independent Start, Stop, and Span settings permit ready adjustment of the displayed spectrum from a 5 KHz window to a full 1 GHz span width. A fully adjustable reference level, and a set of four individual markers allows for precise measurements of levels and frequencies.
Stop, Average, Peak Hold functions and one-button measurements such as Occupied Bandwidth make for a powerful analyzer that can handle almost any situation.
Our unique ‘Scanning Spectrum Audio’ function replaces the typical ‘look and listen’ capability by tuning the receiver (and the recovered audio) to an adjustable marker frequency, rather than the center frequency of the spectrum display.
The R2590 comes packed with features typically found on stand-alone spectrum analyzers, and much more.
This feature is handy when attempting to identify a variety of signals without the need to continually change the center frequency of the spectrum display.
AM/FM Receiver
With a typical sensitivity of less than 2 microvolts at the antenna port, the R2590 can be used for the off-the-air monitoring of remote transmitters. Variable squelch aids in picking up weak signals, but can be set tighter to ensure the proper S/N ratio for measurement accuracy.
This feature reduces service costs by enabling frequent preventative maintenance checks for system degradation or interference identifica­tion without ever leaving the shop.
AM/FM/PM Generator
The R2590 includes a 250 KHz to 1 GHz AM/FM/PM signal generator that has an adjustable output level from –125 dBm to –10 dBm.
The generator output port is fuse-protected, assuring against damage by the inadvertent application of power, therefore saving a trip to the repair shop.
Selectable metering includes a ‘One Button’ SINAD measurement with adjustable target setting. The R2590 will automatically determine the RF level needed to produce a specified SINAD reading.
This flexible signal generator reduces receiver test time by displaying all the necessary control information simultaneously. The wide range of output levels cover all manner of troubleshooting situations.
The R2590 simplifies routine measurements with ‘One Button’ convenience.
Full Band Duplex Offset Generator
Full output level control from –125 dBm to –10 dBm over the entire frequency range of the instrument is available from the duplex offset generator.
Duplex frequency offsets are also fully variable, from 0 Hz to 1 GHz in 1 Hz steps.
The duplex generator provides enhanced capability to service equipment such as repeaters and full duplex radios.
The 1 Hz frequency offset resolution extends functionality to cross-band repeaters and component level receiver and transmitter troubleshooting.