Agilent 8498A Operations Guide

Page 1
8498A Fixed Attenuator
Operating and Service Manual
Agilent Part Number: 08498-90008 Printed in USA Print Date: April 2001 Supersedes: February 1999
Page 2

Notice

The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Agilent Technologies assumes no responsibility for the use or reliability of its software on equipment that is not furnished by Agilent Technologies.
This document contains proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without prior written consent of Agilent Technologies.
Restricted Rights Legend
Use, duplication, or disclosure by the U.S. Government is subject to restrictions as set forth in subparagraph (c)(1)(ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and subparagraphs (c)(1) and (c)(2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Agilent Technologies, Inc Santa Rosa Systems Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, U.S.A.
© Copyright 1997, 1999, 2001 Agilent Technologies, Inc.
ii Agilent Technologies 8498 Operating and Service Manual
Page 3

Warranty

Certification Agilent Technologies certifies that this product met its published specifications at the time of

shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and
Technology (NIST, formerly NBS), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Warranty This Agilent Technologies system product is warranted against defects in
materials and workmanship for a period corresponding to the individual warranty periods of its component products. Instruments are warranted for a period of one year . During the w arranty period, Agile nt T echnol ogies will, at its option, either repair or replace products that prove to be defective.
Warranty service for products installed by Agilent Technologies and certain other products designated by Agilent Technologies will be performed at
Buyer’s facility at no charge within Agilent Technologies service travel areas. Outside Agilent Technologies service travel areas, warranty service will be performed at Buyer’s fac ility onl y upon Agil ent Tec hnologi es’s prior agreement and Buyer shall pay Agilent Technologies’s round trip travel expenses. In all other areas, products must be returned to a service facility designated by Agilent Technologies.
For products returned to Agilent Technologies for warranty service, Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. Howev er , Buyer sh all pay all ship ping char ges, duti es, and taxe s for products returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error free.
LIMITATION OF WARRANTY. The foregoing warrant y shall not a pply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Agilent Technologies 8498 Operating and Service Manual iii
Page 4
EXCLUSIVE REMEDIES. THE REMEDIES PRO VIDED HEREIN ARE
BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

Assistance Product maintenance agreements and other customer assistance agreements

are available for Agilent Technologies products. For assistan ce, ca ll your local Agi lent Technolog ies Sale s and S ervice Of f ice
(refer to “Service and Support” on page v).
iv Agilent Technologies 8498 Operating and Service Manual
Page 5

Service and Support

By internet, phone, or fax, get assistanc e with all your test and meas urement needs.
Online assistance: www.agilent.com/find/assist United States
(tel) 1 800 452 4844
New Zealand
(tel) 0 800 738 378 (fax) (+64) 4 495 8950
Latin America
(tel) (305) 269 7500 (fax) (305) 269 7599
Japan
(tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840
Canada
(tel) 1 877 894 4414 (fax) (905) 282-6495
Australia
(tel) 1 800 629 485 (fax) (+61) 3 9210 59 47
Europe
(tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390
Asia Call Center Numbers
Country Phone Number Fax Number
Singapore 1-800-375-8100 (65) 836-0252 Malaysia 1-800-828-848 1-800-801664 Philippines (632) 8426802
1-800-16510170 (PLDT Subscriber Only)
Thailand (088) 226-008 (outside Bangkok)
(662) 661-3999 (within Bangkok) Hong Kong 800-930-871 (852) 2506 9233 Taiwan 0800-047-866 (886) 2 25456723
(632) 8426809 1-800-16510288 (PLDT Subscriber Only)
(66) 1-661-3714
People’s Republic of China
India 1-600-11-2929 000-800-650-1101
800-810-0189 (preferred)
10800-650-0021
10800-650-0121
Agilent Technologies 8498 Operating and Service Manual v
Page 6

Safety and Regulatory Information

Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a procedure,
practice, or the like, that, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practice , or t he like, which, if not correctly performed or a dher ed to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Instrument Markings
When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information.
!
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
1SM1-A This text indicates that the instrument is an Industrial Scientific and Medical Group 1
Class A product (CISPER 11, Clause 4). This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
vi Agilent Technologies 8498 Operating and Service Manual
Page 7
Safety and Regulatory Information

Safety Earth Ground

This is a Safety Class I product (provided with a protective earthing terminal). An uninterrup ti ble saf et y ear th gr ound must be pr o vided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.

Befor e Applying P o wer Verify that the produ ct is configured to match the available main power

source as d escribed in the input power configuration instructions in this manual. If this product is to be powered by autotransformer, make sure the common terminal is connected to the neutral (grounded) side of the ac power supply.
Agilent Technologies 8498 Operating and Service Manual vii
Page 8

Typeface Conventions

Typeface Conventions
Italics
Used to emphasize important information:
only
Use this software
with the xxxxxX system.
Used for the title of a publication:
Refer to the
xxxxxX System-Level User’s Guide.
Used to indicate a variable:
LOAD BIN filename.
Type
Instrument Display Used to show on-screen prompts and messages that you will see on the
display of an instrument:
The xxxxxX will display the message
[Keycap] Used for labeled keys on the front panel of an instrument or on a
computer keyboard:
[Return].
Press
{Softkey} Used for simulated keys that appea r on an instrum ent display:
{Prior Menu}.
Press
User Entry Used to indicate text that you will enter using the computer keyboard;
text shown in this typeface must be typed
LOAD PARMFILE
Type
CAL1 SAVED.
exactly
as printed:
Used for examples of programming code:
#endif // ifndef NO_CLASS
Path Name
Computer Display Used to show messages, prompts, and window labels that appear on a
Used for a subdirectory name or file path:
Edit the file
computer monitor:
Edit Parameters window will appear on the screen.
The
usr/local/bin/sample.txt
Used for menus, lists, dialog boxes, and button boxes on a computer
monitor from which you make selections using the mouse or keyboard:
Double-click
EXIT to quit the program.
viii Agilent Technologies 8498 Operating and Service Manual
Page 9
Contents
Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
Service and Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Safety and Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Safety Earth Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vii
Typeface Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Attenuator Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Table 1. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Table 2. Supplemental Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . 2
Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Storage and Shipping Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Mating Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Figure 1. Type N Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Preparation For Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operator’s Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Figure 2. Attenuator Exploded View. . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Table 3. Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Table 4. Attenuator Cartridge Assembly Parts (Item 16 in Figure 2) . 9
Agilent Technologies 8498A Operating and Service Manual Contents 1
Page 10

General Information

Attenuator Overview The 8498A is a 30 dB, 30 W (25 W between +35° ×C and + 55° C) fixed

attenuator with Type N connectors. Either connector may be used as the input or output connector.
The attenuator is designed to work in a 50 ohm system, and operates
over the frequency range from dc to 18.0 GHz.
The attenuator may be used (1) to reduce a high power signal to a level
that is compatible with sensit ive equipment, (2) to reduce reflections, or
(3) as a known attenuation for substitution.
Instruments Covered By Manual
The attenuators covered by this manual have a two-part serial number. The first fo ur char acte rs con sti tute the s erial nu mber pr ef i x. The r emaini ng d igits form the sequential suffix that is unique to each attenuator.
The contents of this manual apply di rectly to t he attenuator s having the serial prefix US91. An attenuator manufactured after the printing of this manual may have a serial prefix that is not noted above. This unlisted serial prefix indicates that the attenuator differs in some respect from the information in this manual.
Warranty
The attenuator is warranted and certified as indicated on the warranty page of this manual. For further information, contact the nearest Agilent Technologies Sales and Service office. Refer to “Service and Support” on
page v.
The attenuator is warranted only when it is operated within its specifications. This is especiall y t rue of power handling c apa bi li ty. Any attenuator returned to Agilent Technologies under warranty will be examined carefully to determine if the failure was possibly due to improper use.

Safety Considerations The warning that follows is related to possible personal injury.

WARNING The high power attenuator card has a substrate of beryllium oxide. Beryllium
oxide in a powder form is a hazardous material and may be injurious to your health if inhaled. Do not perform any operation on the beryllium oxide that might generate dust. Defective attenuator cards should be returned to Agilent Technologies for proper disposal.
Agilent Technologies 8498A Operating and Service Manual 1
Page 11

Specifications

Specifications
Instrument specifications are listed in Table 1. These specifications are the performance standa rds, or limits against which the attenuator may be tested.
Table 1 Specifications
Frequency Range dc to 18.0 GHz Attenuation 30 ± 1.0 dB Input Impedance 50 ohm, nominal Maximum SWR (reflection coefficient) 1.15 (0.07) dc to 8.0 GHz
Power Range Up to 25 W CW Maximum Input Power Average
Connectors Type N (male and female)
1.25 (0.11) 8.0 to12.4 GHz
1.30 (0.13) 12.4 to 18.0 GHz
1
30 W at 0 to ≤+35° C 25 W at > +35° C up to +55° C Peak 500 W dc to ≤ 5.8 GHz 125 W > 5.8 GHz up to 18.0 GHz Energy per pulse 500 W/µs
2
1. For pulses greater than 30W the maximum average power (Pa) is limited by the energy per pulse (E) in W
2. Compatible with IEE 287.
s according to Pa = 30 0.02E.
Supplemental characteristics in Table 2 are not specifications but are typical characteristics supplied as additional information for the user.
Table 2 Supplemental Characteristics
Temperature Stability < 0.003 dB/° C Power Sensitivity < 0.006 dB/W Dimensions 83 x 114 x 152 mm (approx. 3.25 x 4.5 x 6.0 in.) Weight 0.6 kg (1.25 lbs.)
2 Agilent Technologies 8498A Operating and Service Manual
Page 12
Specifications

Operating Environment

The operating environment of the attenuator should be within the following limitations:
Temperature.............................................................................0 to +55
Humidity................................................................<95% relative at 40
Altitude..........................................................<4,600 meters (15,000 feet)

Storage and Shipping Environment

The attenuator should be stored in a clean, dry environment. The following environmental limitations apply to both storage and shipment:
Temperature......................................................................... -40 to +75
Humidity............................................................... <95% relative at 40
Altitude........................................................<15,300 meters (50,000 feet)

Mating Connectors The attenuator has Type N connectors, one male and one female. These

connectors mate with all Type N connectors whose dimensions are compatible with IEEE 287.
° C ° C
° C
° C
Type-N Male Connector Type-N Female Connector
Center Conductor
Mating Plane
0.207 inch (5.258 mm)
Figure 1 Type N Connectors
minimum
Center Conductor
Mating Plane
Outer Conductor Mating Plane
Outer Conductor Mating Plane
0.207 inch (5.258 mm) maximum
Agilent Technologies 8498A Operating and Service Manual 3
Page 13

Installation

Installation

Initial Inspection 1. Inspect the shipping container for damage. If the shipping container or

cushioning material is damaged, it should be kept until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
2. If the contents are incomplete, if there is mechanical damage or defect, or if the inst rument doe s not pass the e lec tric al pe rforman ce te st , conta ct the nearest Agilent Technologies Sales and Service office. Refer to the Service and Support information in the front matter of this manual. Agilent Technologies will arrange for repair or replacement of the damaged or defective equipment. Keep the shipping materials for the
carrier’s inspection.
3. If you are returning the instrument for service, repack aging the instrument requires original shipping containers and materials or their equivalents. Agilent Technologies can provide packaging materials identical to the original materials. Refe r to Service and Support information in the front matter of this manual for the Agilent Technologies nearest you. Attach a tag indicating the type of service required, re turn address, model number, and serial number. Mark the container FRAGILE to insure careful handling. In any correspondence, refer to the instrument by model number and serial number.

Preparation For Use When you install the attenuator, make sure that the connectors do not

support weight or bear torque. Take care not to drop or mechanically damaged the attenuator.
4 Agilent Technologies 8498A Operating and Service Manual
Page 14

Operating Information

Operating Information

Principles of Operation

The attenuation is achieved by a resistive film deposition on a beryllium oxide substrate. The high po wer hea t is dissip ated by f ins in the cooling cage that surrounds the attenuator cartridge assembly.
WARNING The high power attenuator card has a substrate of beryllium oxide. Beryllium
oxide in a powder form is a hazardous material and may be injurious to your health if inhaled. Do not perform any operation on the beryllium oxide that might generate dust. Defective attenuator cards should be returned to Agilent Technologies for proper disposal.

Operator’s Check The most accurate check of the attenua tor is made usi ng a netwo rk analyz er.

However, a simplified operator’s check of the attenuator can be made by using either the substitution or insertion loss method.
If a similar attenu ator of kno wn acc ura cy is a vailable, a refer en ce can be
established and then the unknown attenuator can be substituted in place of the reference attenuator. If a known good attenuator is not available, establish a reference on the measuring equipment, insert the device under test, and check for the proper attenuated signal level.
Use an RF source in the proper frequency range and a detector with
readout device. Equipment should have sufficient accuracy to show that the attenua t or is working .

Performance Tests The best method f or testing the attenuator is with an automatic networ k

analyzer or equivalent equipment of suitable accuracy. This method will allow greater accuracy of measurement since system ambiguities can be
corrected. Simple quick-check techniques are mentioned under Operator’s Check above. These techniques will provide a reasonable degree of confidence that the attenuator meets its specifications.
Agilent Technologies 8498A Operating and Service Manual 5
Page 15

Service

Service

Repair Repair of the attenuator is limited to the replacement of the attenuator

cartridge assembly (16) shown in Figure 2. Refer to Figu re 2 to identify the numbered items in the fo llowing procedur e.
Attenuator Disassembly Procedure
1. Remove the 8 screws (10) that hold the two attenuator housings (9) and (30) and remove the housings.
2. Remove the three screws (36) and lock washers (35) from both ends of the attenuator assembly.
3. Remove the end fins (31) and (34), the RFI seals (11) and (19), spacers (33), and the framed attenuator fins (32).
4. Remove the three cap screws (12) from both ends of the attenuator assembly. It may be necessary to move the housing slightly to remove the screws.
5. Slide the inner connector bodies (14) and (18) from the attenuator cartridge assembl y (16) . Be care ful not to le t t he sli ding co ntact s (8 ) and (29) and compressi on spri ngs (7 ) a nd (28) slid e from t he cont act h old ers (6) and (27). Under normal conditions they will not slide out.
6. At this point the attenuator cartridge assembly (16) and the shim washers are free from the overall assembly and can be returned to Agilent Technologies. The attenuator cartridge assembly is the housing for the attenuator card, the card half sections, and the shims.
Attenuator Reassembly Procedure
1. Place the framed attenuat or f ins (32), “D” ring spacers (33 ), and end f ins (31) and (34) on the attenuator cartridge assembly (16).
2. Position the RFI seals (11) and (19) and install end fins (31) and (34) with screws (36) and lock washers (35).
3. Install the connector assemblies with the six scr ews (12). Use t he e nd of a toothpick to apply two smal l drops of Loct ite 262 for eac h scre w to the threads in the atten uator body (16). T i ghten the scre ws e venly to a torque specification of 0.90 Nm (8 in-lbs).
CAUTION Overtighteni ng sc rews (12) beyo nd th e specified torq ue may di stort the card
clamp half sections and fracture the attenuator card.
6 Agilent Technologies 8498A Operating and Service Manual
Page 16
Service
WARNING Under no circumstances is the attenuator card to be machined, ground or
scraped. Beryllium oxide in a powder form is a hazardous material may be injurious to your health if inhaled. Defective attenuator cards should be returned to Agilent Technologies for proper disposal.
4. Install the eight screws (10) that hold the two housings (9) and (30).

Troubleshooting To troubleshoot the attenuator perform the operator’s check. If the

instrument does not perform within limits, check the connectors. If the connectors are not damaged, replace the attenuator cartridge assembly.
37
37
37
Figure 2 Attenuator Exploded View
Agilent Technologies 8498A Operating and Service Manual 7
Page 17

Replacement Parts

Replacement Parts
To order a part listed in the replacement parts table, contact the nearest Agilent Technologies office. Refer to “Service and Support” on page v.
Table 3 Replaceable P a rts
Reference Designation
1 08498-80001 1 Outer connector body Type N, female 2 1250-0915 1 Contact-RF connector series 7 mm-N; female 3 5040-0306 2 Insulator 4 08498-20014 2 Center conductor 5 08498-20019 2 Damper 6 08498-20016 2 Contact holder 7 1460-1618 2 Spring-CPRSN .054-IN-OD .258-IN-0A-LG 8 5020-3297 2 Contact, sliding 9 08498-40002 2 Attenuator housing 10 2200-0145 8 Screw-MACH 4-40 .438-IN-LG PAN-HD-POZI 11 08498-20024 2 RFI seal 12 3030-0070 1 Screw-SKT HD CAP 4-40 .625-IN-LG ALY STL
14 08498-20013 2 Inner connector body
Part Number QTY Description
0470-1590 A/R Loctite 262**
15 08498-20018 2 Insulator 16 08498-60002 1 Attenuator cartridge assembly 17 08498-20018 Insulator 18 08498-20013 Inner connector body 19 08498-20024 RFI seal 20 1250-0918 1 Component-RF connector series 7 mm-N; SST 21 1250-0016 1 Component-RF connector series N; .75 IN 22 1250-0916 1 Body-RF connector series 7 mm-N straight 23 1250-0917 1 Contact-RF connector series 7 mm-N; male 24 5040-0306 Insulator 25 08498-20014 Center conductor 26 08498-20019 Damper
8 Agilent Technologies 8498A Operating and Service Manual
Page 18
Table 3 Replaceable P a rts (Continued)
Replacement Parts
Reference Designation
27 08498-20016 Contact holder 28 1460-1618 Spring-CPRSN .054-IN-OD .258-IN-0A-LG 29 5020-3297 Contact, sliding 30 08498-40002 Attenuator housing 31 08498-00002 2 End fin 32 08498-40001 6 Framed attenuator fin 33 08498-00005 1 D ring spacer* 34 08498-00002 End fin 35 2190-0014 1 Washer-LK INTL T NO. 2 .089-IN-ID 36 0520-0129 6 Screw-MACH 2-56 .312-IN-LG PAN-HD-POZI 37 08498-60006 1 Attenuator assembly
* NOTE: Some attenuators may have two spacers instead of one; 08498-00005 is the direct
replacement.
** Not shown
Part Number QTY Description
7120-7127** 1 Label, identification
Table 4 Attenuator Cartridge Assembly Parts (Item 16 in Figure 2)
Part Qty. Part Number
30 dB, 25W attenuator card 1 1GT1-2674 Attenuator body 1 08498-20011 Card clamp half section 2 08498-20025 Washer 2 08498-20017 Lock 2 08498-20022 Tuning screw 2 08498-20023
Agilent Technologies 8498A Operating and Service Manual 9