Agilent Part No. 83491- 90 01 4
Printed in USA
August, 2004
Agilent Technologies
Digital Signal Analysis
1400 Fountaingrove Parkway
Santa Rosa, CA 95403, USA
Notice.
The information contained in
this document is subject to
change without notice. Companies, names, and data used
in examples herein are fictitious unless otherwise noted.
Agilent Technologies makes
no warranty of any kind with
regard to this material, including but not limited to, the
implied warranties of merchantability and fitness for a
particular purpose. Agilent
Technologies shall not be liable for errors contained herein
or for incidental or consequential damages in connection with the furnishing,
performance, or use of this
material.
Restricte d Ri ghts Legend.
Use, duplication, or disclosure by the U.S. Government
is subject to res tric tion s as se t
forth in subparagraph (c) (1)
(ii) of the Rights in Technical
Data and Computer Software
clause at DFARS 252.227-7013
for DOD agencies, and subparagraphs (c) ( 1) and (c) (2 )
of the Commercial Computer
Software Restricted Rights
clause at FAR 52.227-19 for
other agencies.
Safety Symbols.
CAUTION
The caution sign denotes a
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in damage to
or destruction of the product.
Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met.
WARNING
The warning sign denotes a
hazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in injury or
loss of life. Do not proceed
beyond a warning sign until
the indicated conditions are
fully understood and met.
The instruction manual symbol. The product is marked with this
warning symbol when
it is necessary for the
user to refer to the
instructions in the
manual.
The laser radiation
symbol. This warning
symbol is marked on
products which have a
laser output.
The AC symbol is used
to indicate the
required nature of the
line module input
power.
| The ON symbols are
used to mark the positions of the instrument
power line switch.
❍ The OFF symbols
are used to mark the
positions of the instrument power line
switch.
The CE mark is a registered trademark of
the European Community.
The CSA mark is a registered trademark of
the Canadian Standards Association.
The C-Tick mark is a
registered trademark
of the Australian Spectrum Management
Agency.
This text denotes the
ISM1-A
instrument is an
Industrial Scientific
and Medical Group 1
Class A product.
ii
Contents
1 Getting Started
Introduction 1-2
General Safety Considerations 1-4
Installation 1-5
Returning the Instrument for Service 1-12
2Operation
Introduction 2-2
Front-Panel Features 2-3
Block Diagrams 2-6
To Display a S ignal 2-9
To Compensate for Module Insertion Loss 2-10
Using Probes with an 83491A 2-11
3 Reference
Front-Panel Optical Adapters 3-2
In Case of Difficulty 3-3
Error Messages 3-5
Electrostatic Discharge Infor m ati o n 3-7
Cleaning Connections for Accurate Measurements 3-9
Introduction1-2
General Safety Consider ations1-4
Installation1-5
Step 1. Inspect the shipment1-5
Step 2. Install th e module1-7
Step 3. Connect the front-panel cables1-8
Step 4. Connect a modulated signal1-10
Returning the Instrument for Service1-12
Contacting Agilen t Technologies1-12
Getting Started
Getting Started
Introduction
Introduction
Agilent 83491/2/3/4A clock recovery modules are designed to operate in
86100-series infiniium Digital Communications Analyzers (DCA) and 83480A
DCAs. These modules recover clock and data information at standard telecom
and datacom rates. The resulting trigger signal is made available to the DCA
mainframe via a connector located on the module’s rear-panel. An external
front-pane l cable passes the data signal, with some insertion loss, to the
receiver modu le .
WARNI NGLight energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these
connectors is the slightly attenuated ligh t that is input to the frontpanel INPUT connector.
Online HelpOn 86100-series infiniium D C As, after installing the module as show n in this
chapter, you’ll find all user documentation (except programming) is located in
the instrument’s Help system. To access the Help system , complete th e installation steps, turn on the 86100C, and click Contents on the Help menu.
1-2
Getting Started
Introduction
83494A Auxiliary
Clock Output
The 83494A module’s front-panel auxiliary clock output provides the recovered clock signal, refer to “Auxiliary outpu ts” on page 2-4. To recover the
clock at data rates greater than 2.5 Gb/s, the DCA mainframe must have one of
the following trigger options installed:
• 86100C Option 001, Enhanced Trigger
• 86100A/B Option 001, Divided Trigger
• 83480A Option 100, Divided Tri g ge r
CAUTIONFiber-optic connectors are easily damaged when connected to dirty or
damaged cables and accessories. The 83492A, 83493A, and 83494A front-panel
input connectors are no exce pti o n. Whe n you use improper cleaning and
handling techniques , y o u ri sk expensive instrume nt repairs, damaged cables,
and compromised measurements. Before you connect any fiber-optic cable to
an 83492A, 83493A, or 83494A modu le, re fe r to “Cleaning Connec tions for
Accurate Measurements” on page 3-9.
CAUTIONThe circuits on electrical inputs and outputs can be dama g ed by electrostatic
discharge (ESD). Therefore, avoid applying static discharges to any front or
rear-panel electrical connector. Before connecting any coaxial cable to a frontpanel connector, momentarily short the center and outer conductors of the
cable together. Avoid touching the front-panel connectors without first
touching the frame of the instrument. Be sure that the instrument is properly
earth-grounded to prevent buildup of static charge. Refer to “Electros tatic
Discharge Information” on page 3-7.
Figure 1-1. An Agilent 83493A installed in an Agilent83480A mainframe
1-3
Getting Started
General Safety Considera tions
General Safety Considerations
This product has been designed and tested in accordance with the standards
listed on the Manufacturer's Declaration of Conformity, and has been supplied
in a safe condition. The documentation contains information and warni ng s
that must be followed by the user to ensure safe operation and to maintain the
product in a safe condition.
The 83491/2/3/4A modules work only with Agilent digital communication analyzer mainfram es. Refer to the mainframe documentation to ensure safe operation.
WARNI NGLight energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these
connectors is the slightly attenuated ligh t that is input to the frontpanel INPUT connector.
WARNI NGIf this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must be used in a
normal condition (in w hich all means for protection are intact) only.
WARNI NGTo prevent electrical shock, disconnect the main frame from mains
before cleaning. Use a dry cloth or one slightly dampened with water
to clean the external case part s. Do not attempt to clean internally.
WARNI NGNo operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
CAUTIONThis product i s designed for use in In st allation Category II and Pollutio n
Degree 2 per IEC 10 10 and 664 re s pectively.
CAUTIONElectrostatic discharge (ESD) on or near input connectors can damage circuits
inside the instrument. Repair of damage due to misuse is not covered under
warranty. Before connecting any cable to the electrical input, momentarily
short the center and outer conductors of the cable together. Personnel should
be properly grounded, and should touch the frame of the instrument before
touching any connector.
1-4
Getting Started
Installation
Installation
Although some of the figures in the following procedure show an 83480A DCA
mainframe, the procedures equal ly ap ply to installing th e m odule into an
86100-series DCA mainframe.
NOTEIf an 83491/2/3/4A module is installed in an 83480A DCA , the DCA s fi rm war e
revision must be A.06.25 or later. If the module is installed in an Agilent 54750A
digitizing oscilloscope, you must first install the Agilent 83480K
communications firmware upgrade kit. To check the Agilent 83480A’ s firmware
revision code, pr ess the Utility key and then the System config softkey. The
firmware revision number is listed under the Frame sec tion of the display.
Step 1. Inspect the shipment
❒ Inspect the shipping container and instrument for damage. Keep the shipping
container and cushioning material until you have inspected the contents of the
shipment for completeness and have checked the instrument mechanically and
electrically.
❒ Locate the shipping list. Verify that you received all the accessories on this list,
and all the options that you ordered. Table 1-2 on page 1-6 shows the external
front-panel cables that are supplied for each module.
❒ Make sure that the serial number listed on the module’s rear-panel label match-
es the serial numbe r listed on the shipping document.
If your shipment is da m a g ed or incomplete, sav e the packing materi a ls and
notify both the shipping carrier and the ne a re st Agilent Technologies service
office. Agilent Technologies will arrange for repair or replacement of damaged
or incomplete shipments without waiting for a settlement from the transportation company. Notify the Agilent Technologies customer engin eer of any problems.
a. Refer to “Step 3. Connect the front-panel cables” on page1-8 for the use of these c ables.
1-6
Getting Started
Installation
Step 2. Install the module
Up to two modules can be inserted into the 86100-series or 83480A DCA mainframe.
1 Install the 83491/2/3/4A module into the left slot on the mainframe. Install in g
the module into the left slot ensures that the supplied adapter cable will fit.
Finger-tighten the knur le d screw on the front panel of the plu g-in module to
ensure that the module is securely seated in the main f ra me.
Figure 1-2. Position of modules in the mainframe
2 Clean all optical interfaces as described in “Cleaning Connections for Accurate
Measurements” on page 3-9, before making measurements.
1-7
Getting Started
Installation
Step 3. Connect the front-panel cables
83491A ModulePerform this step if you are installing an 83491A m odul e .
❒ Use the RF cable suppl ied with the module to conn ec t the Electrical Output
connector on the 83 491A to the input on the a dja cent measurement module.
Use the short (83491-20001) or long (83491-20002) cable depending on the location of the input on the adjacent measurement module.
Figure 1-3. The adapter cable
83493/4A Modules Perform the following steps if you are installing an 83493/4A module.
1 Because you’ll connect an FC-PC fiber-optic cable in this procedure, confirm
that the receiver module’s optical input connector has an FC-PC adapter.
Fiber-optic adapters can be removed by hand turning them in the counter-
clockwise direction. Refer to “Front-Panel Optical Adapters” on page 3-2 for a
description of available adapters.
2 If you changed the adapter in step one, you can connect the original adapter to
the 83493/4A module’s front-panel In put connector.
3 Use the FC-PC semi-rigid fiber-optic cable suppl ie d wi th the 83493/4 module
(83493-20001) to connect the Optical Output connector on the 83493/4A to the
input on the receiver module.
83492A ModulePerform the following steps if you are instal ling an 83492A module.
1-8
Getting Started
Installation
On 83492A modules, the front-panel fiber-optic connectors reverse input and
output roles depending on the wavelength of the signal. This is shown in
Figure 1-4. Signals in the 750 nm to 860 nm wavelength range are input to the
left connector and output fr om the rig ht connector. Signals in the 1000 nm to
1600 nm wavelength range are input to the ri g ht connector and output from
the left connector.
1 Because you’ll connect an FC-PC fiber-optic cable in this procedure, confirm
that the receiver module’s optical input connector has an FC-PC adapter.
Fiber-optic adapters can be removed by hand turning them in the counter-
clockwise direction. Refer to “Front-Panel Optical Adapters” on page 3-2 for a
description of available adapters.
2 If you changed the adapter in step one, you can connect the original adapter to
the 83493/4A module’s front-panel input connector. See Figure 1-3 to identify
the input connector based on input wavelength.
3 Connect the supplied adapter cable as shown in Figure 1-3. Use the short cable
(83492-20002) when operating at 750 nm to 860 nm wavelengths. Use the long
cable (83492-20001) when operating at 1000 nm to 1600 nm wavelengths.
Figure 1-4. Input and output connections versus wavel ength
1-9
Getting Started
Installation
Step 4. Connect a modulated signal
WARNI NGLight energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these
connectors is the slightly attenuated ligh t that is input to the frontpanel INPUT connector.
CAUTION83491A Modules: T o prevent damage to the module, the maximum safe signal
input level is < ±5V. The input circuits can also be damaged by electrostatic
discharge (ESD). Before connecting any coaxial cable to the connectors,
momentarily short the center and outer conductors of the cable together. A void
touching the front-pa ne l input connectors without fir st touching the frame of
the instrument. Be su re that the instrument is p ro p erly earth-grounded to
prevent buildup of static charge.
CAUTION83492/3/4A Modules: To prevent damage to the module, th e maximum signal
input level is < +3 dBm.
1 Turn on the mainframe and connect a modulated signal to the 83491/2/3/4A
module’s Input connect or.
2 If you are using an 83480A DCA, press the Trigger key and then the Source
softkey. Then, select trigger 2 (the 83491/2/3/4A module) for the trigger source.
3 On the 83491/2/3/4A module, re pe ate dl y press the SELECT key until a front-
panel light indicates the data rate of the signal. Avoid selecting a data rate that
is a multiple of the input signal. For example, don’t select a 622 Mb/s data rate
if the signal is really at 155 Mb/s.
4 Confirm that the Unlocked light is off. If using the 83494A, confirm that the
Trigger Loss light is off. If the UNLOCKED light is on (or TRIGGER LOSS light on the
83494A), clock recovery cannot be establis hed on the signal.
5 Observe the Clock and Data outputs on DCA. Waveforms should be present. The
instrument is now ready for you to begin making measurements. If you cannot
get the clock recovery module to lock on the signal, make sure th at yo u have
selected the correct data rate and that the mainframe trigger level is adjusted
appropriately.
Signals displayed using a data trigger are less reliable than using a recovered
clock. Signals triggered on data can also vary depending upon the trigger level.
1-10
Getting Started
Installation
Green and red data-rate lights
The data-rate indicator lights on the Agilent 83491/2/3A and Agilent 83494A Option 106
change color between red and green to show which data rate is selected. A red light
does not indicate a problem. A red light shows that the adjacent red data rate label is
selected. A green light shows that the adjacent green data rate label is selected.
Repeatedly pressing the SELECT key cycles through the selections in one color before
switching to the opposite color. On Agilent83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second selection cycle shows 1062 Mb/s
selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data
rate is selected. Repeatedly pressing the SELECT key cycles through the available data
rates, as well as the Trigger On Data bypass mode.
Figure 1-5. Front-panel lights
1-11
Getting Started
Returning the Instrument for Servi ce
Returning the Instrument for Service
The instructions in this section show you how to properly return the instrument for repair or calibration. Always call the Agilent Technologies Instrument
Call Center fir st to initiate serv ic e before returning your instrument to a service office. This ensures that the repair (or calibrati on) can be properly
tracked and that your instrument will be returned to you as quickly as possible. Call this number regardless of where you are located.
If the instrument is still unde r warranty or is covered by an Agilen t Technologies maintenance contract, it will be repaired under the terms of the warranty
or contract (the warranty is at the front of this manual). If the instrument is
no longer under warranty or is not covered by an Agilent Technologies maintenance plan, Agilent Technologies will notify you of the cost of the repair after
examining the unit.
When an instrument is returned to a Agilent Technologies service office for
servicing, it must be adequately packaged and have a complete des cri pti on of
the failure sympt o m s a ttached. When describing the failure, please be as sp ecific as possible about the nature of the problem. Include copies of additional
failure information (such as the instrument failure settings, data related to
instrument failure, and error messages) along with the instrument being
returned.
Contacting Agilent Technologies
Call CenterFor technical assistance, you can contact your local Agilent Call Center.
• In the Americas, call 1 (800) 829-4444
• In other regions, visit http://www.agilent .co m/fin d/assist
Service CenterBefore returning an instrument for service, you must first call the Agilent
Technologies Instrument Support Center.
• In all regions, call (800) 829-4444
1-12
Getting Started
Returning the Instrument for Service
Preparing the module for shipping
1 Write a complete description of the failure and attach it to the instrument.
Include any specific performance details related to the problem. The following
information should be returned with the instrument.
• Type of ser v ice required.
• Date instrument was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether instrument is temperature-sensitive.
• Wheth er instrument is vibr a tion-sensitive.
• Instrument settings require d to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact perso n.
• Model number of returned instrument.
• Full serial number of returned instrument.
• List of any accessories returned with instrument.
2 Cover all front or rear-panel connectors that were originally covered when you
first received the instrument.
CAUTIONCover electrical connectors to protect sensitive components from electrostatic
damage. Cover optical connectors to protect them from damage due to physical
contact or dust.
CAUTIONInstrument damage can result from using packaging materials other than the
original materials. Never use styrene pellets as packaging material. They do not
adequately cus hion the instrument or prevent it from shift ing in the carton.
They may also cause instrument damage by generating static electricity.
3 Pack the instrument in the original shipping containers. Original materials are
available through any Agilent Technologies office. Or, use the following
guidelines:
• Wrap the instrument in antistatic plastic to reduce the possibility of damage
caused by electrostatic discharge.
• For instruments weighing less than 54 kg (120 lb), use a double-walled, cor-
rugated cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on
1-13
Getting Started
Returning the Instrument for Servi ce
all sides of the instrument for packing material, and strong enough to accommodate the weight of the instrument.
• Surround the equipment with approximately 7 cm (3 inches) of packing material, to protect the instrument and prevent it from moving in the carton. If
packing foam is not available, the best alternative is S.D-240 Air Cap™ from
Sealed Air Corporation (Commerce, Califor nia 900 01). Air Cap look s like a
plastic sheet filled with air bubbl es . U se the pink (antistati c ) Air Cap™ to
reduce static ele ctricity. Wrapping the instrument sev eral times in this material will protect the instrument and prevent it from moving in the carton.
4 Seal the carton with strong nylon adhesive tape.
5 Mark the carton “FRAGILE, HANDLE WITH CARE”.
6 Retain copies of all shipping papers.
1-14
2
Introduction2-2
Front-Panel Features2-3
Block Diagra m s 2-6
To Display a S ignal2-9
To Compensate for Module Insertion Loss 2-10
Using Probes with an 83491A2-11
Operation
Operation
Introduction
Introduction
If you use an 83491/2/3/4A module in an 861 00- se ri e s inf in ii um DC A , information on using the clock recovery modules, along with specifi cati ons , is
included in the instrument’s help system. To acce ss the built-in infor matio n
system, referred to as Help, simply click Contents on the Help menu. GPIB
programming commands for the mo dule s are documented in the 86100-serie s
infiniium DCA p ro grammer’s guide.
If you use an 83491/2/3/4A module in an 83480A DCA, you’ll notice that, unlike
most modules designed to be used with the 834 80A, the clock recovery modules do not include Channel keys or menus. Also, there are no GPIB programming commands for these modu les when used in an 83480A mainframe.
Multimode module and single-mode reference receivers
Agilent Technologies does not recommend using the Agilent 83492A mul ti mode module with si ng le-mode referenc e re c eivers such as the
Agilent 83481A, 83482A, or 83485A,B modules. Connecti ng multi m ode to single-mode fibers cau ses large reflection s a nd insertion loss because of the
reduction of the optical fiber’s core from 62.5 µm to 9 µm.
Single-mode module and multimode reference receivers
It is acceptable to use an Agilent 834 93A or an 83494A si ng le-mo de module
with a multimode reference receiver such as the Agilent 83486A module. This
is true provided that single- m ode fiber is con nected to the Agilent 83493A or
83494A module’s front-panel INPUT connector.
2-2
Front-Panel Features
Operation
Front-Panel Features
Figure 2-1. Agilent 83491/2/3/4A front panels
2-3
Operation
Front-Panel Fe atures
SELECT keyPressing this key changes the data rate selection. The recovered and retimed
clock trigger is sent to the mainframe. The Trigger On Data selection is a bypass
mode where the data stream directly triggers the mainframe. Refer to “Blo c k
Diagrams” on page 2- 6 to view a schematic of the normal and bypass paths.
Green and red data-rate lights
The data-rate indicator lights on the Agilent 83491/2/3A and Agilent83494A Option 106
change color between red and green to show which data rate is selected. A red light
does not indicate a problem. A red light shows that the adjacent red data rate label is
selected. A green light shows that the adjacent green data rate label is selected.
Repeatedly pressing the SELECT key cycles through the selections in one color before
switching to the opposite color. On Agilent83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second selection cycle shows 1062 Mb/s
selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data
rate is selected. Repeatedly pressing the SELECT key cycles through the available data
rates, as well as the Trigger On Data bypass mode.
UNLOCKED
indicator
83491/2/3A
This light shows when clock recovery cannot be established on the signal. If a
clock rate is selected, the trigger output to the mainfram e is disabled to prevent free-run triggering . Howe ver in bypass mode (Trigger On Data selected),
triggering is not disabled. When the UNLOCKED light is on, you can establish a
trigger on the data inpu t to the reference receiver.
Trigger Loss
indicator
83494A
This light shows when clock recovery cannot be established on the signal. If a
clock rate is selected, the trigger output to the mainfram e is disabled to prevent free-run triggering . Howe ver in bypass mode (Trigger On Data selected),
triggering is not disabled. When the Trigger Loss light is on, you can estab lish a
trigger on the data inpu t to the reference receiver.
Auxiliary outputsDATA connector: This co nnector provides a fully regenerated ver sion of the
input signal. It is intended for monitoring purposes only and not for rigorous
eye mask compl iance tes ting. Th e frequenc y respons e does no t confo rm to the
requirements for eye mask testing as described in ITU-T G.957 and Bellcore
GR-253-CORE. On Agilent 83492A and 83493A modules, this port is amplitude stabilized fo r input signals greate r than approximatel y –23 dBm. On
Agilent 83494A modules, this port is amplitude stabilized for input signals
greater than approximately –10 dBm.
2-4
Operation
Front-Panel Features
CLOCK connector: This connector provides the recovered clock signal. You can
use this signal to measure jitter transfer, because this output can track and follow input data with very fast jitter; it has a wide bandwidth jitter transfer function when compared to the recovered clock signal which is routed through a
rear-panel conne ctor to the mainframe for triggering. Note that the CLOCK Aux-iliary Output remains synchronized to input signals several dB below the onset of
errors at the DATA Auxiliary Output.
Input and Output
connectors
The input connectors pass the digitally modulated signal to the receiver module. The input signal, slightly attenuated and available at the OUTPUT connector, is connected to the input of the receiver module. The connectors on
optical modules include adapters which can e asily be chang e d to mat ch the
type of connectors that are used on your fiber-optic cables. Refer to “Fr ont-
Panel Optical Adapters” on page 3 -2 for a description of the available adapt-
ers.
Multimode and single-mode connections
Agilent 83492A modules use multimode fiber. Connecting the output to the Optical
INPUT connector on Agilent 83481/2/5 single-mode modules results in large reflections
and insertion loss.
Agilent 83493A and 83494A modules use 9/125 µm single-mode fiber. Connecting multimode fiber to the Optical Input connector results in large reflections and insertion loss.
Recovered ClockThe recovered clock signal is routed directly to the 86100-series or 83480A
mainframes through the module’s rear panel. This output has a lower jitter
modulation bandwidth than the front-panel CLOCK Auxiliary Output. Because of
the reduced jitter modulation bandwidth on the mainframe trigger signal, a
more complete view of the jitter on the waveform data is obtained.
2-5
Operation
Block Diagrams
Block Diagrams
Figure 2-2. 83491A Block Diagram
Figure 2-3. 83492A and 83493A Block Diagram
2-6
Operation
Block Diagrams
Figure 2-4. 83494A Block Diagram
2-7
Operation
Block Diagrams
Figure 2-5. 83494A Option 106 Block Diagram
2-8
Operation
T o Display a Signal
To Display a Signal
1 Install the module as described in “Installation” on page 1-5. Be sure to connect
all of the cables as described in the procedure.
2 Repeatedly press the SELECT key on the clock reco v ery module until the front-
panel light indicates the proper data rate of the signal.
Green and red data-rate lights
The data-rate indicator lights on the 83491/2/3A and 83494A Option 106 change color
between red and green to show which data rate is selected. A red light does not indicate
a problem. A red light shows that the adjacent red data rate label is selected. A green
light shows that the adjacent green data rate label is selected. Repeatedly pressing the
SELECT key cycles through the selections in one color before switching to the opposite
color. On 83491A modules for example, the first selection cycle shows 155 Mb/s
selected. The second selection cycle shows 1062 Mb/s selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data
rate is selected. Repeatedly pressing the SELECT key cycles through the available data
rates, as well as the Trigger On Data bypass mode.
• If the UNLOCKED light is on (or TRIGGER LO SS light on the 83494A), clock recovery cannot be establishe d on the si gnal.
• Avoid selecting a data rate that is a multiple of the input signal. For example,
don’t select a 622 Mb/s data rate if the sign al is reall y at 155 Mb/s.
• If you cannot get the clock recovery module to lock on the signal, make sure
that you have sel ect ed th e co rrect dat a rat e and t hat the ma inf ram e tri gger
level is adju sted appropria t ely.
• Signa ls displayed us ing a data trig ger are less re liable than usin g a recovered
clock. Signals triggered on data can also vary depending upon the trigger
level.
2-9
Operation
To Compensate for Module Insertion Loss
To Compensate for Module Insertion Loss
If you are using an 86100-series in fin ii um DCA , re fe r to the instrument’s help
system for information on compensating for module insertion loss. Use the following steps on an 83480A DCA to allow you to enter an offset to compensate
for the insertion loss of the clock recovery module. This provides accurate
amplitude measu r e ments at the input to t he clock recovery module.
1 Disconnect the cable from the clock recovery module’s Input connector.
2 Measure the signal using a power meter. You can use either the 83480A’s built-
in power meter or an external power meter.
3 Reconnect the cable to the clock recovery module.
4 Disconnect the cable from the reference receiver module’s input connector.
5 Measure the signal using a power meter. You can use either the 83480A’s built-
in power meter or an external power meter.
6 Subtract the two measurements to determine the insertion loss of the module.
7 On the reference receiver module, pres s the front-panel channel SETUP key.
8 Press External scale, and set the Atten units to “decibel”.
9 Press Attenuation, and enter the value calculated in Step 6.
2-10
Operation
Using Probes with an 83491A
Using Probes with an 83491A
You can use external passive and active probes with the 83491A electrical
clock recovery module. If you are using an 86100-series infiniium DCA, refer
to the infiniium DC A’s help system; the pro c e dures in this section ar e fo r use
on an 83480A DCA.
The following probes are available for use with 83491A clock recovery modules:
• Agilent N1020A TDR probe. This passive probe (1:1, 50Ω) provides a fixture
for positioning an d holding the probe tip on the device being tested .
• Agilent 54701A 2.5 GHz active probe. This is a 100kΩ, 10:1, probe.
• Agilent 54006A 6 GHz handheld low-impedanc e probe. This passive probe
(10:1, 500Ω, 20:1, 1kΩ) has an input capacitance of 0.25 pf.
• Agilent 1163A 1 GHz resistive-divider probe. This passive 500Ω probe has an
input capacitance of 1.5 pf.
If you are using an 83480A DCA, use the procedures in this section to generate vertical scale factors. These factors are applie d to the calibr ati on of the
reference receiver module’s electrical channel. When selecting a probe, keep
in mind that the input impedance of the 83491A is 50Ω.
If the probe being calibrated has an attenu ati on factor that allows the instrument to adjust the gain to produce even steps in the vertical scale factors, the
instrument will do so. Typically, probes have standard attenuation factors such
as divide by 10, divide by 20, or divide by 100. Because the following pr ocedures include com pensation for inser tion loss of the clock re c ov ery module,
do not perform the procedur e “To Compensate for Module Insertion Loss” on
page 2-10.
To compensate for a passive probe
1 Connect the probe to the Input connector on the 83491A clock recovery module.
2 Attach the pro be ti p to th e CAL hook that is located near the floppy disk drive.
3 Press the reference receiver module’s front-panel channel SETUP key.
2-11
Operation
Using Probes with an 83491A
4 Press Calibrate and then Calibrate probe.
To compensate an 54701A active probe
1 Connect the 83491A output to the electrical measurement channel input.
2 Connect the probe to the Input connector on the 83491A clock recovery module.
3 Connect the probe po wer cable to the Probe Power connect o r on the reference
receiver modu le .
4 Attach the pro be ti p to th e CAL hook that is located near the floppy disk drive.
5 Press the reference receiver module’s front-panel channel SETUP key.
6 Press Calibrate and then Calibrate probe.
To compensate for other devices
The information in this section applies to both optical and electr ical measurements. Since the mainframe’s CAL signal is a voltage source, it cannot be used
to calibrate to the probe tip when the units are set to Ampere, Watt, or
Unknown. Instead , s et t he ext ern al gai n an d ex te rnal o ff set t o com pens ate for
the actual characteristics of the devic e. If you do not know the actual characteristics, you can refer to the typical specifications that came with the device.
1 Press the reference receiver module’s front-panel channel SETUP key.
2 Press External scale.
3 Press Atten units Ratio
, Attenuation1:1, and then Units Ampere (Volt, Watt, or
Unknown).
4 Press Ext gain, and enter the actual gain characteristics of the device.
5 Press Ext offset, and enter the offset introduced by the device.
2-12
3
Front-Panel Optical Adapters3-2
In Case of Difficulty3-3
Error Messages 3-5
Electrostatic Discharge Infor m ati on3-7
Cleaning Connections for Accurate Measurements3-9
Reference
Reference
Front-Panel O ptical Adapter s
Front-Panel Optical Adapters
Front Panel
Fiber-Optic
Adapter
DescriptionAgilent Part Number
Diamond HMS-1081000AI
a
FC/PC
D481000GI
SC81000KI
DIN81000SI
ST81000VI
Biconic81000WI
Dust Covers
FC connector1005-0594
Diamond HMS-10 connector1005-0593
DIN connector1005-0595
ST connector1005-0596
SC connector1005-0597
81000FI
a. The FC/PC adapter is the standard adapter supplied with the instrument.
3-2
In Case of Difficulty
In Case of Difficulty
This section provides a list of suggestions for you to follow if the plug-in module fails to operate. A list of messages that may be displayed is also included in
this chapter. Before calling Agilent Technologies or returning th e un it for service, a few minutes spent performing some simple checks may save waiting for
your instrument to be repaired.
If the mainframe does not operate
❒ Is the line fuse good?
❒ Does the line socket have power?
Reference
❒ Is the unit plugged in to the prop er ac power source?
❒ Is the mainframe tur ned on?
❒ Is the rear-panel line switch set to on?
❒ Will the mainframe po w er u p without the plug-in module installed?
If the plug-in does not operate
❒ Is the plug-in module firmly seated in the mainframe slot?
❒ Are the knurled sc rews at the bottom of the plug-in module finger-tight?
❒ Is the clock recovery module set to the modulation rate of the input signal?
❒ If other equipment, cables, and connectors are being used with the plug-in
module, are they co nnected properly and o perating correc tly?
❒ Review the procedure for the test being performed when the problem ap-
peared. Are all the settings correct? Can the problem be reproduced?
❒ Are the connectors clean? See “Clean ing Connections fo r Accurate Measur e-
3-3
Reference
In Case of Difficulty
ments” on page 3-9 for more information.
❒ Make sure that the instrument is ready to acquire data.
❒ Find any signals on the channel inputs by using the Autoscale function.
❒ See if any signals are present at the channel inputs by using Freerun triggering.
❒ Make sure Channel Display is on.
❒ Make sure the channel offset is adjusted so the waveform is not clipped off the
display.
❒ Make sure the mainframe identifies the plug-in module:
• One the 86100-series DCA, select the All Calibrati ons command on the Calibrate menu. Confirm that the Model number and Seri al Number are listed
in the displayed dialog box.
• On the 83480A DCA, press Utility, then System config. Confirm that the Model
number is listed in the box labeled “Plug-ins”. If “~known” is displayed instead of the model number of the plug-in module, remove and reinsert the
plug-in module in the same slot. If “~known” is still displayed, the mainframe
may need to have the latest operating system firmware i n stalled. Options
001 and 002 provide this firmware on a 3.5 inch diskette. To load new firmware, follow the instructions provided with this diskette. If you do not have
the optional diskette, contact your local Agilent Technologies service office
(refer to “Contacting Agilent Technologies” on page 1-12).
If the model number of the plug-in module is list ed, the mainframe ha s identified the plug-in.
3-4
Reference
Error Messages
Error Messages
This section pr o v ides description of error messages that ar e displayed if you
are using the module in an 83480A DCA. If you are using an 86100-series infiniium DCA, refer to the instrument’s help system for information on error
messages displayed on that mainframe.
Memory error occurred in plug-in_:Try
reinstalling plug-in
The mainframe could not correctly read the contents of the memory in the
plug-in.
❒ Remove and reinstall the plu g- i n mo dule. Each time a plug-in is instal le d, the
mainframe re-reads the memory in the plug-in module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knur led scr ews at the botto m of the pl ug-in mo dule ar e finger -tight .
❒ Install the plug-in in a different slot in the mainframe.
Busy timeout occurred with plug-in_:Try
reinstalling plug-in
The mainframe is having trouble communicating with the plug-in module .
Make sure there is a goo d connection between the mainframe and the plug-in
module.
❒ Remove and reinstall the plu g- i n module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knurled screws at the bottom of the plug-in module are finger-tight.
❒ Install the plug-in in a different slot in the mainframe.
3-5
Reference
Communications failure exists at slot_:Service is required
Communications failure exists at slot_:Service
is required
An illegal hardware state is detected at the mainframe-to-plug-in module
interface of the specifie d sl ot.
• If the slot is empty, there is a mainframe hardware problem. Refer to the
Agilent 83480A, Agilent 54750A Service Guide.
• If a plug-in is installed in the slot, there is a plug-in module hardware problem.
Return the plug-in module to a qualifie d service departmen t.
ID error occurred in plug-in_:Service is
required
The informatio n read from the memory of the plug-in module do e s no t m a tc h
the hardware in the plug-in module. This can be caused by a communication
problem between the mainframe and the plug-in module. Make sure there is a
good connection between the mainframe and the plug-in.
❒ Remove and re-install the plug-in module.
❒ Verify the plug-in module is firmly seated in the mainframe slot.
❒ Verify the knurled screws at the botto m of the plug-in module ar e fi nger tight.
❒ The standard Agilent 54750A mainframe does not accept the Agilent 83491/2/
3/4A module. To use the module, a firm ware upgrade must first be installed.
Order the Agilent 8348 0K communi cati ons firmware kit and install according
to the instruct ions.
❒ The Agilent 83480A, Agilent 54750A mainframes do not accept plug-in mod-
ules designed for use with the Agilent 54710A, 54720A.
Plug-in is not supported:System firmware
upgrade is needed
The mainframe ma y need to have the latest o perating system firmware
installed. Options 001 and 002 provide this firmware on a 3.5 inch diskette. T o
load the new firmware, follow the instructions provided with the diskette. If
you do not have the optional diskette, contact your local Agilent Technologies
service office.
3-6
Reference
Electrostatic Discharge Information
Electrostatic Discharge Information
Electrostatic dis c h arge (ESD) can damage or destro y electronic components .
All work on electronic assemblies should be performed at a static-safe work
station. The following figure shows an example of a static-safe work station
using two types of ESD protecti on:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Figure 3-1. Static-safe work station
3-7
Reference
Electrostatic Discharge Informa tio n
Both types, when used together, provide a significant level of ESD protection.
Of the two, only the table-mat and wrist-strap combination provides adequate
ESD protection when used alone.
T o ensure user safety, the static-safe accessories must provide at least 1 MΩ of
isolation from ground. Refer to Table 3-1 for information on ordering staticsafe accessories.
WARNI NGThese techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than 500
volts.
Reducing ESD Damage
The following suggestions ma y he lp reduce ESD damag e that occurs during
testing and servic ing operations.
• Personnel shoul d be grounded with a resi stor-isolated wrist strap before removing any assembly from the unit.
• Be sure all instrument s are properly earth-grounded to prevent a buildup of
static charge.
Table 3-1.Static-Safe Accessories
Agilent Part
Number
9300-0797
9300-0980Wrist-strap cord 1.5 m (5 ft)
9300-1383Wrist-strap, color black, st ainless steel, without cord, has fo ur adjustable
9300-1169ESD heel-strap (reusable 6 to 12 months).
3-8
Description
Set includes: 3M static control mat 0.6 m
ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They
must be ordered separately.)
links and a 7 mm post-type connection.
× 1.2 m (2 ft× 4 ft) and 4.6 cm (15
Reference
Cleaning Connections for Accurate Measurements
Cleaning Connections for Accurate
Measurements
Today, advances in measurement capabilities make connectors and connection techniques more important than ever. Damage to the connectors on calibration and verification devices, test ports, cables , and ot her device s can
degrade measurement accuracy and damage instruments. Replacing a damaged connector can cost thousands of dollars, not to mention lost time! This
expense can be avoided by observing the simple precautions presented in this
book. This book also contains a brief list of tips for caring for electrical connectors.
Choosing the Right Connector
A critical but often overlooked factor in making a good lightwave measure ment is the selection of the fiber-optic connector. The differences in connector types are mainly in the mechanic a l a ssembly that holds th e fe rrule in
position against another identical ferrule. Connectors also vary in the polish,
curve, and conce ntricity of the core within the cladding. Mating one style of
cable to another requires an adapter. Agilent T echnologies offers adapters for
most instruments to allow testin g w i th many different c ables. Figure 3-2 on
page 3-10 shows the basic componen ts of a ty pica l conn ect ors .
The system tole ranc e for reflection and in se rtion loss must be kn ow n w he n
selecting a connector from the wide variety of currently available connectors.
Some items to consider when selecting a connector are:
• How much insertion loss can be allowed?
• Will the connector need to make multiple connections? Some connectors are
better than others, and some are very poor for making repeate d connections.
• What is the reflection tolerance? Can the system take reflection degradation?
• Is an instrument-grade connector with a preci sion core alig nmen t re qu ir ed ?
• Is repeatability tolerance for reflection and loss important? Do your specifica-
3-9
Reference
Cleaning Connections for Accurate Measurements
tions take repeatability uncertain ty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be required? For example, many DFB lasers cannot operate with reflections from
connectors. Often as much as 90 dB isolation is needed.
Figure 3-2. Basic components of a connector.
Over the last few ye a rs, the FC/PC style c onnector has emerge d as the most
popular connector for fiber-optic applications. While not the highest performing connector, it represents a good compromise between performance, reliability, and cost. If properly maintained and cleaned, this connector can
withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most
connectors, including the FC/PC style, can deliver. These instruments cannot
tolerate connectors with the l arge non-concentricities of the fiber common
with ceramic style ferrules. When tighter alignme n t is required, Agi lent
Technologies instru ments typi cally use a connector such as the Diamond
HMS-10, which has concentric tolerances within a few tenths of a micron. Agilent Technologies then uses a special unive rsal adapter, which allows oth er
cable types to mate with this precision connector. See Figure 3-3.
3-10
Reference
Cleaning Connections for Accurate Measurements
Figure 3-3. Universal adapters
The HMS-10 enca se s the fiber within a soft nickel silver (C u/Ni/Zn) center
which is surrounded by a tough tungsten carbide casing, as shown in
Figure 3-4.
Figure 3-4. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber
to be moved to the desired positi on. This process first stakes the soft n ickel
silver to fix the fiber in a near-center location, then uses a post-active staking
to shift the fiber into the desired position within 0.2 µm. This process, plus the
keyed axis, allo ws very precise core-to-core alignments. This conne c to r is
found on most Agilent Technologies lightwave instruments.
3-11
Reference
Cleaning Connections for Accurate Measurements
The soft core, wh ile allowing precise c entering, is also th e c h ief liability of the
connector. The soft material is easily damaged. Care must be taken to minimize excessive scratching and wear. While minor wear is not a problem if the
glass face is not affected, scratches or grit can cause the glass fiber to move
out of alig nment. Also, if unkeye d co nnecto rs ar e used , the nicke l sil ver ca n be
pushed onto the glass surface. Scratches, fiber movement, or glass contamination will cause lo s s of sign al and increa sed re flect ion s, resu ltin g in p oor retu rn
loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immediately obvious to the naked eye, poor measurements result without the user
being aware. Microscopic examination and return loss measure men ts are the
best way to ensure good measurements. Good cleaning practices can help
ensure that optim u m c onnector performanc e is maintained. With glass - to glass interface s, any degradatio n of a ferrule or the end of the fi ber, any stray
particles, or finger oil can have a significant effect on connector performance.
Where many repeat connections are required, use of a connector saver or
patch cable is recommended.
Figure 3-5 shows the end of a clean fiber-optic cable. The dark circle in the
center of the micrograph is the fiber’s 125 µm core and cladding which carries
the light. The surrounding area is the soft nickel-silver ferrule. Figure 3-6
shows a dirty fiber end from neglect or perhaps improper cleaning. Material is
smeared and gr ou nd i nto t he end o f t he f ib er cau si ng l igh t sc att erin g an d po or
reflection. Not only is the precision polish lost, but this action can grind off the
glass face and destroy the connector.
Figure 3-7 shows physical damage to the glass fibe r end caused by either
repeated connections made without removing loose particles or using
improper cleaning tools. When severe, the damage of one connector end can
be transferred to another good connector en dface that comes in contact with
the damaged one. Periodic checks of fiber ends, and replacing connecting
cables after many connections is a wise practice.
The cure for these p roblems is discipline d c o nnector care as described in the
following list and in “Cleaning Connectors” on page3-16.
3-12
Reference
Cleaning Connections for Accurate Measurements
Use the following guidelines to achieve the best possible performance when
making measurements on a fiber-optic system:
• Never use metal or sharp objects to clean a connector and never scrape the
connector.
• Avoid matching gel and oils.
Figure 3-5. Clean, problem-free fiber en d and f errule.
Figure 3-6. Dirty fiber end and ferrule from poor cleaning.
3-13
Reference
Cleaning Connections for Accurate Measurements
Figure 3-7. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The
oil or gel grabs and holds grit that is then ground into the end of the fiber.
Also, some early ge ls were designed fo r use with the FC, non-con t ac ting connectors, using small glass spheres. When used with contacting connectors,
these glass balls can scratch and pit the fiber. If an index matching gel or oil
must be used, apply it to a freshly cleaned connector, make the measurement,
and then immed ia tely clean it off. Never us e a gel for longer- term connections
and never use it to improve a damaged connector. The gel can mask the extent
of damage and continued use of a damaged fiber can trans fer damage to the
instrument.
• When inserting a fiber-optic c able into a connector, gently inse rt i t in a s
straight a line as possible. Tipping and inserting at an angle can scrape material
off the inside of the connector or even break the inside sleeve of connectors
made with ceramic material.
• When inserting a fiber-optic connector into a connector, make sure that the fiber end does not touch the outside of the mating connec tor or adapter.
• Avoid over tightening connections.
Unlike common electrical connections, tighte r is not better. The purpose of
the connector is to bring two fiber ends together. Once they touch, tightening
only causes a greater force to be applied to the delicate fibers. With connectors that have a convex fiber end, the end can be pushed off-axis resulting in
misalignmen t a nd e xc essive return lo ss. Many measurem ents are actually
improved by backing off the connector pressu re . Also, if a piece of grit does
happen to get by the cleaning procedure, the tighter connection is more likely
to damage the glass. Tighten the connectors just until the two fibers touc h.
3-14
Reference
Cleaning Connections for Accurate Measurements
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best connector possible. Replace connecting cables regularly. Frequently measure the
return loss of the connector to check for degradation, and clean every connector, every time.
All connectors should be treated like the high-quality lens of a good camera.
The weak link in instrument and syste m reliability is of te n the inappropriate
use and care of the connector. Because current connectors are so easy to use,
there tends to be reduced vigilance in connector care and cleaning. It takes
only one missed cle a ning for a piece of grit to permanently dama g e the glass
and ruin the conne c to r.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipme nt are a good indication that you have good connections. S ince return loss and insertion loss are
key factors in determining optical connector performance they can be used to
determine connector degradation. A smooth, polished fiber end should produce a good return- loss measurement. The quality of the polish establishes
the difference between the “PC” (physical contact) and the “S upe r PC” connectors. Most connectors today are physical contact which make glass-to-glass
connections, therefore it is critical that the area around the glass core be clean
and free of scratches. Altho ug h the major area of a connector, excluding the
glass, may show scra tc hes and wear, if the glass has maintained its polished
smoothness, the connector can st ill provide a good low lev el return loss connection.
If you test your cables and acces so ries for insertion loss and re turn loss upon
receipt, and retain the measured data for comparison, you will be able to tell in
the future if any degradation has occurred. Typical values are less than 0.5 dB
of loss, and somet im es as little as 0.1 dB of lo ss with high performanc e connectors. Return loss is a measure of reflection: the less reflection the bett er
(the larger the return loss, the smaller the reflection). The best physically
contacting connectors have return losses better than 50 dB, although 30 to
40 dB is more common.
3-15
Reference
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections
on the cable end face can be detected as well as cracks or chips in the fiber
itself. Use a microscope (100X to 200X magni fi cati on) to in spect the entire
end face for contam ina tion, raised meta l , o r de nts in the metal as well a s an y
other imperfections. Inspect the fiber for cracks and chips. Visible imperfections not touching the fibe r cor e m ay no t affe ct pe rformance (unless the
imperfections keep the fibers from contacting).
WARNI NGAlways remove both ends of fiber-optic cables from any instrument,
system, or device before visually inspecting the fiber ends. Disable all
optical sources before disconnecting fiber-optic cables. Failure to do
so may result in permanent injury to your eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiberoptic cables and Agilent Technologies universal adapter s. The initi al cleani ng ,
using the alcohol as a solvent, gently removes any grit and oil. If a caked-on
layer of material is still present, (this can happen if the beryllium-copper sides
of the ferrule retainer get scraped and deposited on the end of the fiber during
insertion of the cable), a second cleaning should be performed. It is not
uncommon for a cable or connector to require more than one cleanin g.
CAUTIONAgilent Technologies strongly recommends that index matching compounds
not be applied to their instruments and accessories. Some compounds, such as
gels, may be difficult to remove and can contain damaging particulates. If you
think the use of such compounds is necessary, refer to the compound
manufacturer for infor mat ion on application and cleaning procedures.
Table 3-2.Cleaning Accessories
Item AgilentPart Number
Any commercially available denatured alcohol—
Cotton swabs8520-0023
Small foam swabs9300-1223
Compressed dust remover (non-resi due)8500-5262
3-16
Reference
Cleaning Connections for Accurate Measurements
Table 3-3. Dust Caps Provided with Lightwave Instruments
Item AgilentPart Number
Laser shutter cap08145-64521
FC/PC dust cap08154-44102
Biconic dust cap08154-44105
DIN dust cap5040-9364
HMS10/dust cap5040-9361
ST dust cap5040-9366
To clean a non-lensed connector
CAUTIONDo not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber e nds that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end
after cleaning.
2 Clean the ferrules and o the r parts of the connector while avoidi ng the end of
the fiber.
3 Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4 Clean the fiber end with the swab or lens paper.
Do not scrub durin g this initial clea ning because grit c a n be c a ught in the
swab and become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint-free cott on swab or len s
paper.
6 Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the
fiber end face. Nitrogen gas or compressed dust remover can also be used.
3-17
Reference
Cleaning Connections for Accurate Measurements
CAUTIONDo not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the
compressed air canister.
7 As soon as the connec tor is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the connector again. Often a second cleaning will restore proper performance. The
second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many Agilent Technologies
instruments employ a universal adapter such as those shown in the following
picture. These adapters allow you to connect the instrument to different types
of fiber-optic cables.
Figure 3-8. Universal adapters.
1 Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The
foam swabs listed in this section’s int roduction are smal l en ough to fit into
adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and
the risk of other conta m ination buildup on the inside of adapte rs greatly outweighs the risk of contamination by foam swabs.
2 Clean the adapter with the foam swab.
3 Dry the inside of the adapter with a clean, dry, foam swab.
4 Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip,
or invert compressed air canisters, because this releases particles in the can
into the air. Refer to instructions provided on the compressed air canister.
This chapter lists specifications and characteristics of the clock recovery modules. Specifications apply over the temperature range +15°C to +35°C (unless
otherwise noted) after the instrument’s temperature has been stabilized after
60 minutes of continuous operation.
SpecificationsSpecifications described warranted performance.
CharacteristicsCharacteristics provide useful, nonwarranted, information about the func-
tions and performance of the instrument. Characteristics are printed in
italics.
Calibration cycleAgilent Technologies warrants instrument specifications over the recom-
mended calibrat ion interv al. T o ma intain speci fications , periodic re calibrati ons
are necessary. We recommend that the Agilent 83491/2/3/4A be calibrated at
an Agilent Technologies service facility eve ry 24 months.
Data triggering (character istic)50 Mb/s to 2500 Mb/s
Operatin g in put power level
a b
Triggering operati on, all rates
–10
10
BER, all rates
c
50 to 70 kHz
4 to 5 MHz
±0.1%
±0.1%
±0.1%
±0.1%
±0.1%
±0.1%
±0.1%
–10 dBm to 3 dB m
–10 dBm to 3 dB m
Insertion loss (through path)
DC through 2500 MHz
Output jitter, all rates
d
≤ 7 dB
0.0125 UI
Maximum continuous electrical power bef or e damage (characteristic)1W peak
DATA and CLO CK output amplitude, all rates (characteristic)0.5Vp- p
INPUT electrical return loss
DC through 1250 MHz (characteristic)
1250 MHz through 2500 MHz (character ist ic)
≥ 20 dB
≥ 15 dB
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (charact er istic)
2000 MHz through 2500 MHz (character ist ic)
a. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
b. Operating power level applies over temperature range 25
c. Better than 10
d. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°.
≥ 10 dB
≥ 6 dB
rms
4-3
Specifications and Regulatory Information
83492A Specification s
83492A Specifications
Table 4-2.83492A Specifications (1 of 2)
Wavelength range (characterist ic)750 nm to 860 nm and
1000 nm to 1600 nm
Optical INPUT and OUTPUT fiber (characteristic)62.5/125 multimode
Data triggering (character istic)50 Mb/s to 2500 Mb/s
Operating input power level
c d
750 nm to 860 nm
Triggering op e r a t io n, all rates
–10
BER, all rates
10
e
–10 dBm to 3 dB m
–10 dBm to 3 dB m
1000 nm to 1600 nm
Triggering op e r a t io n, all rates
–10
BER, all rates
10
Output jitter, all rates
f
g
–13 dBm to 3 dB m
–13 dBm to 3 dB m
0.0125 UI
rms
4-4
Specifications and Regulat ory Information
83492A Specifications
Table 4-2.83492A Specifications (2 of 2)
Maximum continuous optical power bef or e damage (characteristic)10 mW peak
DATA and CLO CK output amplitude, all rates (characteristic)0.5Vp- p
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (charact er istic)
2000 MHz through 2500 MHz (character ist ic)
a. Minimum loss in 850 nm window.
b. Single-mode backreflection tested with FC/PC adapter and single-mode fiber. Optical output terminated
with > 33 dB return lo ss. R etur n loss with fully filled 62.5
c. Source extinction ratio
d. Operating power level applies over temperature range 25
e. Better than 10
f. Better than 10
g. Measured on an oscilloscope eye diagram with PRBS 2
≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
–10
BER when tested with PR BS 223–1 pattern.
–10
BER when tested with PR BS 223–1 pattern.
µm core multimode fiber may be sl igh tly lower .
°C ± 5°.
23
–1 test pattern.
≥ 10 dB
≥ 6 dB
4-5
Specifications and Regulatory Information
83493A Specification s
83493A Specifications
Table 4-3.83493A Specifications
Wavelength range (characteristic)1000 nm to 1600 nm
Optical INPUT fiber (characteristic)9/125 s ingle mode
Optical insertion loss (through path)
Optical return loss
Data triggering (character istic)50 Mb/s to 2500 Mb/s
Operating input power level
Triggering operation, all rates
–10
10
BER, all rates
Output jitter, all rates
b c
–20 dBm to 3 dB m
d
e
–17 dBm to 3 dB m
0.0125 UI
rms
Maximum continuous optical power bef or e damage (characteristic)10 mW peak
DATA and CLO CK output amplitude, all rates (characteristic)0.5Vp- p
DATA and CLOCK output electrical return loss
50 MHz through 2500 MHz (characteristic)
2.5 GHz through 10 GHz (characteristic)
≥ 10 dB
≥ 8 dB
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss.
4-6
Specifications and Regulat ory Information
b. Source extinction ratio ≥ 8.2 dB when measured per TIA/EIA OFSTP-4A.
c. Operating power level applies over temperature ra nge 25
d. Better than 10
–10
BER when tested with PR BS 223–1 pattern.
e. Measured on an oscilloscope eye diagram with PRBS 2
°C ± 5°.
23
–1 test pattern.
83493A Specifications
4-7
Specifications and Regulatory Information
83494A Specification s
83494A Specifications
Table 4-4.83494A Specifications (1 of 2)
Wavelength range (characteristic)1200 nm to 1600 nm
Optical INPUT fiber (charact er istic)9/125 single mode
Optical insertion loss (through path)1.5 dB
Optical return loss
Clock Recovery Phase Locked Loop Ban dwidth (characteristic)
Data triggering (charact eristic)50 Mb/s to 2500 Mb/s
Operating input power level
b c
Triggering operation
9953.00 Mb/s, 10312.500 Mb /s, 10664.228 Mb/s, and
–8 dBm to 3 dBm
10709.225316 Mb/s
Triggering operation, all other rates
-12 dBm to 3 dBm
-14 dBm to 3 dBm (t y pical)
–10
10
BER, all rates
d
0 dBm
4-8
Specifications and Regulat ory Information
83494A Specifications
Table 4-4.83494A Specifications (2 of 2)
Output jitter e
all rates
155.52 Mb/s
622.08 Mb/s
2488.32 Mb/s
Maximum continuous optical power before damage (characteristic)10 mW peak
DATA and CLOCK output amplitude, all rates (characteristic)0 .5Vp-p
DATA and CLOCK output electrical return loss
50 MHz through 2500 MHz (characteristic)
2.5 GHz through 10 GHz (characteristic)
0.02 UI
0.01 UI
0.01 UI
0.01 UI
≥ 10 dB
≥ 8 dB
rms
(typical)
rms
(typical)
rms
(typical)
rms
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss.
b. Source extinction ratio
c. Operating power level applies over temperature range 25
d. Better than 10
e. Measured on an oscilloscope eye diagram with P RBS 2
≥ 8.2 dB when mea sured per TIA/EI A OFSTP-4A.
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°.
23
–1 test pattern.
4-9
Specifications and Regulatory Information
83491/2/3/4A Operating S pecifications
83491/2/3/4A Operating Specifications
Table 4-5. Operating Specifications
UseIndoor
Temperature
Operating
Non-operating
Altitude
Operating
Non-operating
0°C to +55°C
–40
°C to +70°C
4600 m (15,000 ft)
15,300 m (50,000 ft)
Humidity
Operating
Non-operating
Net weightappr ox i mately 1.2 kg (2.6 lb.)
Shipping weightapproximately 2.1 kg (4.6 lb.)
Dimensions H x W x D131.66 mm x 80.14 mm x 266.9 mm
Power RequirementsSupplied by mainframe
4-10
up to 90% relative humidity at <35
up to 90% relative humidity at <35
(5.18 in x 3.15 in x 10.5 i n )
°C
°C
Specifications and Regulat ory Information
Regulatory Information
Notice for Germany: Noise Declaration
Acoustic Noise EmissionGeraeuschemission
LpA < 70 dBLpA < 70 dB
Operator positionam Arbeitsplatz
Normal positionnormaler Betrieb
per ISO 7779nach DIN 45635 t.19
Regulatory Information
Notice for Canada: Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du Canada.
4-11
Specifications and Regulatory Information
Regulatory Information
4-12
Index
Numerics
83494A, 4-8
A
adapter cable, 1-9
Agilent 54701A active probe, 2-12
Agilent 83492A