Agilent 83494A User Manual

Agilent 83491/2/3/4A Clock Recovery Modules User’s Guide
© Copyright 2000-2004 Agilent Technologies All Rights Reserved. Repro­duction, adaptation, or trans­lation without prior written permission is prohibit ed , except as allowed under copy­right laws.
Agilent Part No. 83491- 90 01 4 Printed in USA August, 2004
Agilent Technologies Digital Signal Analysis 1400 Fountaingrove Parkway Santa Rosa, CA 95403, USA
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Agilent Technologies makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Agilent Technologies shall not be lia­ble for errors contained herein or for incidental or conse­quential damages in connec­tion with the furnishing, performance, or use of this material.
Restricte d Ri ghts Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to res tric tion s as se t forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and sub­paragraphs (c) ( 1) and (c) (2 ) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Safety Symbols.
CAUTION The caution sign denotes a
hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WARNING The warning sign denotes a
hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
The instruction man­ual symbol. The prod­uct is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature of the line module input power.
| The ON symbols are
used to mark the posi­tions of the instrument power line switch.
The OFF symbols
are used to mark the positions of the instru­ment power line switch.
The CE mark is a reg­istered trademark of the European Commu­nity.
The CSA mark is a reg­istered trademark of the Canadian Stan­dards Association.
The C-Tick mark is a registered trademark of the Australian Spec­trum Management Agency.
This text denotes the
ISM1-A
instrument is an Industrial Scientific and Medical Group 1 Class A product.
ii

Contents

1 Getting Started
Introduction 1-2 General Safety Considerations 1-4 Installation 1-5 Returning the Instrument for Service 1-12
2Operation
Introduction 2-2 Front-Panel Features 2-3 Block Diagrams 2-6 To Display a S ignal 2-9 To Compensate for Module Insertion Loss 2-10 Using Probes with an 83491A 2-11
3 Reference
Front-Panel Optical Adapters 3-2 In Case of Difficulty 3-3 Error Messages 3-5 Electrostatic Discharge Infor m ati o n 3-7 Cleaning Connections for Accurate Measurements 3-9
4 Specifications and Regulatory Information
83491A Specifications 4-3 83492A Specifications 4-4 83493A Specifications 4-6 83494A Specifications 4-8 83491/2/3/4A Operating Specifications 4-10 Regulatory Information 4-11
Contents-1
1
Introduction 1-2 General Safety Consider ations 1-4 Installation 1-5
Step 1. Inspect the shipment 1-5 Step 2. Install th e module 1-7 Step 3. Connect the front-panel cables 1-8
Step 4. Connect a modulated signal 1-10 Returning the Instrument for Service 1-12 Contacting Agilen t Technologies 1-12

Getting Started

Getting Started

Introduction

Introduction
Agilent 83491/2/3/4A clock recovery modules are designed to operate in 86100-series infiniium Digital Communications Analyzers (DCA) and 83480A DCAs. These modules recover clock and data information at standard telecom and datacom rates. The resulting trigger signal is made available to the DCA mainframe via a connector located on the module’s rear-panel. An external front-pane l cable passes the data signal, with some insertion loss, to the receiver modu le .
Table 1-1.Module Features
Agilent Module Input Connector Selectable Rates (Mb/s)
83491A 83492A 83493A 83494A 83494A Option 106 83494A Option 107
50
electrical
Multimode fiber (62.5/125 Single-mode fiber (9/125 Single-mode fiber (9/125 Single-mode fiber (9/125 Single-mode fiber (9/125
µm) µm) µm) µm)
155, 622, 1060, 1250, 2120, 2488, 2500 155, 622, 1060, 1250, 2120, 2488, 2500
µm)
155, 622, 1250, 2488, 2500 155, 622, 2488, 9953 155, 622, 2488, 2666, 10664 155, 622, 2488, 2666, 10709
WARNI NG Light energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these connectors is the slightly attenuated ligh t that is input to the front­panel INPUT connector.
Online Help On 86100-series infiniium D C As, after installing the module as show n in this
chapter, you’ll find all user documentation (except programming) is located in the instrument’s Help system. To access the Help system , complete th e instal­lation steps, turn on the 86100C, and click Contents on the Help menu.
1-2
Getting Started
Introduction
83494A Auxiliary Clock Output
The 83494A module’s front-panel auxiliary clock output provides the recov­ered clock signal, refer to “Auxiliary outpu ts” on page 2-4. To recover the clock at data rates greater than 2.5 Gb/s, the DCA mainframe must have one of the following trigger options installed:
• 86100C Option 001, Enhanced Trigger
• 86100A/B Option 001, Divided Trigger
• 83480A Option 100, Divided Tri g ge r
CAUTION Fiber-optic connectors are easily damaged when connected to dirty or
damaged cables and accessories. The 83492A, 83493A, and 83494A front-panel input connectors are no exce pti o n. Whe n you use improper cleaning and handling techniques , y o u ri sk expensive instrume nt repairs, damaged cables, and compromised measurements. Before you connect any fiber-optic cable to an 83492A, 83493A, or 83494A modu le, re fe r to “Cleaning Connec tions for
Accurate Measurements” on page 3-9.
CAUTION The circuits on electrical inputs and outputs can be dama g ed by electrostatic
discharge (ESD). Therefore, avoid applying static discharges to any front or rear-panel electrical connector. Before connecting any coaxial cable to a front­panel connector, momentarily short the center and outer conductors of the cable together. Avoid touching the front-panel connectors without first touching the frame of the instrument. Be sure that the instrument is properly earth-grounded to prevent buildup of static charge. Refer to “Electros tatic
Discharge Information” on page 3-7.
Figure 1-1. An Agilent 83493A installed in an Agilent83480A mainframe
1-3
Getting Started

General Safety Considera tions

General Safety Considerations
This product has been designed and tested in accordance with the standards listed on the Manufacturer's Declaration of Conformity, and has been supplied in a safe condition. The documentation contains information and warni ng s that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
The 83491/2/3/4A modules work only with Agilent digital communication ana­lyzer mainfram es. Refer to the mainframe documentation to ensure safe oper­ation.
WARNI NG Light energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these connectors is the slightly attenuated ligh t that is input to the front­panel INPUT connector.
WARNI NG If this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must be used in a normal condition (in w hich all means for protection are intact) only.
WARNI NG To prevent electrical shock, disconnect the main frame from mains
before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case part s. Do not attempt to clean internally.
WARNI NG No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
CAUTION This product i s designed for use in In st allation Category II and Pollutio n
Degree 2 per IEC 10 10 and 664 re s pectively.
CAUTION Electrostatic discharge (ESD) on or near input connectors can damage circuits
inside the instrument. Repair of damage due to misuse is not covered under warranty. Before connecting any cable to the electrical input, momentarily short the center and outer conductors of the cable together. Personnel should be properly grounded, and should touch the frame of the instrument before touching any connector.
1-4
Getting Started

Installation

Installation
Although some of the figures in the following procedure show an 83480A DCA mainframe, the procedures equal ly ap ply to installing th e m odule into an 86100-series DCA mainframe.
NOTE If an 83491/2/3/4A module is installed in an 83480A DCA , the DCA s fi rm war e
revision must be A.06.25 or later. If the module is installed in an Agilent 54750A digitizing oscilloscope, you must first install the Agilent 83480K communications firmware upgrade kit. To check the Agilent 83480A’ s firmware revision code, pr ess the Utility key and then the System config softkey. The firmware revision number is listed under the Frame sec tion of the display.
Step 1. Inspect the shipment
Inspect the shipping container and instrument for damage. Keep the shipping
container and cushioning material until you have inspected the contents of the shipment for completeness and have checked the instrument mechanically and electrically.
Locate the shipping list. Verify that you received all the accessories on this list,
and all the options that you ordered. Table 1-2 on page 1-6 shows the external front-panel cables that are supplied for each module.
Make sure that the serial number listed on the module’s rear-panel label match-
es the serial numbe r listed on the shipping document. If your shipment is da m a g ed or incomplete, sav e the packing materi a ls and
notify both the shipping carrier and the ne a re st Agilent Technologies service office. Agilent Technologies will arrange for repair or replacement of damaged or incomplete shipments without waiting for a settlement from the transporta­tion company. Notify the Agilent Technologies customer engin eer of any prob­lems.
1-5
Getting Started
Installation
Table 1-2.Supplied External Front-Panel Cables
Supplied Cable
a
Agilent Part Number
83491A
RF cable, SMA-SMA, 9.75 cm 83491-20001 RF cable, SMA-SMA, 15.2 cm 83491-20002
83492A
Fiber-optic cable (semi-rigid), multimode, 14 cm 83492-20001 Fiber-optic cable (semi-rigid), multimode, 18.5 cm 83492-20002
83493A
Fiber-optic cable (semi-rigid), single-mode, 14 cm 83493-20001
83494A
Fiber-optic cable (semi-rigid), single-mode, 14 cm 83493-20001
a. Refer to “Step 3. Connect the front-panel cables” on page1-8 for the use of these c ables.
1-6
Getting Started
Installation
Step 2. Install the module
Up to two modules can be inserted into the 86100-series or 83480A DCA main­frame.
1 Install the 83491/2/3/4A module into the left slot on the mainframe. Install in g
the module into the left slot ensures that the supplied adapter cable will fit. Finger-tighten the knur le d screw on the front panel of the plu g-in module to ensure that the module is securely seated in the main f ra me.
Figure 1-2. Position of modules in the mainframe
2 Clean all optical interfaces as described in “Cleaning Connections for Accurate
Measurements” on page 3-9, before making measurements.
1-7
Getting Started
Installation
Step 3. Connect the front-panel cables
83491A Module Perform this step if you are installing an 83491A m odul e .
Use the RF cable suppl ied with the module to conn ec t the Electrical Output
connector on the 83 491A to the input on the a dja cent measurement module. Use the short (83491-20001) or long (83491-20002) cable depending on the lo­cation of the input on the adjacent measurement module.
Figure 1-3. The adapter cable
83493/4A Modules Perform the following steps if you are installing an 83493/4A module.
1 Because you’ll connect an FC-PC fiber-optic cable in this procedure, confirm
that the receiver module’s optical input connector has an FC-PC adapter. Fiber-optic adapters can be removed by hand turning them in the counter-
clockwise direction. Refer to “Front-Panel Optical Adapters” on page 3-2 for a description of available adapters.
2 If you changed the adapter in step one, you can connect the original adapter to
the 83493/4A module’s front-panel In put connector.
3 Use the FC-PC semi-rigid fiber-optic cable suppl ie d wi th the 83493/4 module
(83493-20001) to connect the Optical Output connector on the 83493/4A to the input on the receiver module.
83492A Module Perform the following steps if you are instal ling an 83492A module.
1-8
Getting Started
Installation
On 83492A modules, the front-panel fiber-optic connectors reverse input and output roles depending on the wavelength of the signal. This is shown in
Figure 1-4. Signals in the 750 nm to 860 nm wavelength range are input to the
left connector and output fr om the rig ht connector. Signals in the 1000 nm to 1600 nm wavelength range are input to the ri g ht connector and output from the left connector.
1 Because you’ll connect an FC-PC fiber-optic cable in this procedure, confirm
that the receiver module’s optical input connector has an FC-PC adapter. Fiber-optic adapters can be removed by hand turning them in the counter-
clockwise direction. Refer to “Front-Panel Optical Adapters” on page 3-2 for a description of available adapters.
2 If you changed the adapter in step one, you can connect the original adapter to
the 83493/4A module’s front-panel input connector. See Figure 1-3 to identify the input connector based on input wavelength.
3 Connect the supplied adapter cable as shown in Figure 1-3. Use the short cable
(83492-20002) when operating at 750 nm to 860 nm wavelengths. Use the long cable (83492-20001) when operating at 1000 nm to 1600 nm wavelengths.
Figure 1-4. Input and output connections versus wavel ength
1-9
Getting Started
Installation
Step 4. Connect a modulated signal
WARNI NG Light energy can radiate from the front panel OUTPUT connectors on
83492A, 83493A, and 83494A modules. The light emitted from these connectors is the slightly attenuated ligh t that is input to the front­panel INPUT connector.
CAUTION 83491A Modules: T o prevent damage to the module, the maximum safe signal
input level is < ±5V. The input circuits can also be damaged by electrostatic discharge (ESD). Before connecting any coaxial cable to the connectors, momentarily short the center and outer conductors of the cable together. A void touching the front-pa ne l input connectors without fir st touching the frame of the instrument. Be su re that the instrument is p ro p erly earth-grounded to prevent buildup of static charge.
CAUTION 83492/3/4A Modules: To prevent damage to the module, th e maximum signal
input level is < +3 dBm.
1 Turn on the mainframe and connect a modulated signal to the 83491/2/3/4A
module’s Input connect or.
2 If you are using an 83480A DCA, press the Trigger key and then the Source
softkey. Then, select trigger 2 (the 83491/2/3/4A module) for the trigger source.
3 On the 83491/2/3/4A module, re pe ate dl y press the SELECT key until a front-
panel light indicates the data rate of the signal. Avoid selecting a data rate that is a multiple of the input signal. For example, don’t select a 622 Mb/s data rate if the signal is really at 155 Mb/s.
4 Confirm that the Unlocked light is off. If using the 83494A, confirm that the
Trigger Loss light is off. If the UNLOCKED light is on (or TRIGGER LOSS light on the 83494A), clock recovery cannot be establis hed on the signal.
5 Observe the Clock and Data outputs on DCA. Waveforms should be present. The
instrument is now ready for you to begin making measurements. If you cannot get the clock recovery module to lock on the signal, make sure th at yo u have selected the correct data rate and that the mainframe trigger level is adjusted appropriately.
Signals displayed using a data trigger are less reliable than using a recovered clock. Signals triggered on data can also vary depending upon the trigger level.
1-10
Getting Started
Installation
Green and red data-rate lights
The data-rate indicator lights on the Agilent 83491/2/3A and Agilent 83494A Option 106 change color between red and green to show which data rate is selected. A red light does not indicate a problem. A red light shows that the adjacent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On Agilent83491A modules for example, the first selec­tion cycle shows 155 Mb/s selected. The second selection cycle shows 1062 Mb/s selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data rate is selected. Repeatedly pressing the SELECT key cycles through the available data rates, as well as the Trigger On Data bypass mode.
Figure 1-5. Front-panel lights
1-11
Getting Started

Returning the Instrument for Servi ce

Returning the Instrument for Service
The instructions in this section show you how to properly return the instru­ment for repair or calibration. Always call the Agilent Technologies Instrument Call Center fir st to initiate serv ic e before returning your instrument to a ser­vice office. This ensures that the repair (or calibrati on) can be properly tracked and that your instrument will be returned to you as quickly as possi­ble. Call this number regardless of where you are located.
If the instrument is still unde r warranty or is covered by an Agilen t Technolo­gies maintenance contract, it will be repaired under the terms of the warranty or contract (the warranty is at the front of this manual). If the instrument is no longer under warranty or is not covered by an Agilent Technologies mainte­nance plan, Agilent Technologies will notify you of the cost of the repair after examining the unit.
When an instrument is returned to a Agilent Technologies service office for servicing, it must be adequately packaged and have a complete des cri pti on of the failure sympt o m s a ttached. When describing the failure, please be as sp e­cific as possible about the nature of the problem. Include copies of additional failure information (such as the instrument failure settings, data related to instrument failure, and error messages) along with the instrument being returned.
Contacting Agilent Technologies
Call Center For technical assistance, you can contact your local Agilent Call Center.
• In the Americas, call 1 (800) 829-4444
• In other regions, visit http://www.agilent .co m/fin d/assist
Service Center Before returning an instrument for service, you must first call the Agilent
Technologies Instrument Support Center.
• In all regions, call (800) 829-4444
1-12
Getting Started
Returning the Instrument for Service
Preparing the module for shipping
1 Write a complete description of the failure and attach it to the instrument.
Include any specific performance details related to the problem. The following information should be returned with the instrument.
• Type of ser v ice required.
• Date instrument was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether instrument is temperature-sensitive.
• Wheth er instrument is vibr a tion-sensitive.
• Instrument settings require d to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact perso n.
• Model number of returned instrument.
• Full serial number of returned instrument.
• List of any accessories returned with instrument.
2 Cover all front or rear-panel connectors that were originally covered when you
first received the instrument.
CAUTION Cover electrical connectors to protect sensitive components from electrostatic
damage. Cover optical connectors to protect them from damage due to physical contact or dust.
CAUTION Instrument damage can result from using packaging materials other than the
original materials. Never use styrene pellets as packaging material. They do not adequately cus hion the instrument or prevent it from shift ing in the carton. They may also cause instrument damage by generating static electricity.
3 Pack the instrument in the original shipping containers. Original materials are
available through any Agilent Technologies office. Or, use the following guidelines:
• Wrap the instrument in antistatic plastic to reduce the possibility of damage
caused by electrostatic discharge.
• For instruments weighing less than 54 kg (120 lb), use a double-walled, cor-
rugated cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on
1-13
Getting Started
Returning the Instrument for Servi ce
all sides of the instrument for packing material, and strong enough to accom­modate the weight of the instrument.
• Surround the equipment with approximately 7 cm (3 inches) of packing ma­terial, to protect the instrument and prevent it from moving in the carton. If packing foam is not available, the best alternative is S.D-240 Air Cap™ from Sealed Air Corporation (Commerce, Califor nia 900 01). Air Cap look s like a plastic sheet filled with air bubbl es . U se the pink (antistati c ) Air Cap™ to reduce static ele ctricity. Wrapping the instrument sev eral times in this ma­terial will protect the instrument and prevent it from moving in the carton.
4 Seal the carton with strong nylon adhesive tape. 5 Mark the carton “FRAGILE, HANDLE WITH CARE”. 6 Retain copies of all shipping papers.
1-14
2
Introduction 2-2 Front-Panel Features 2-3 Block Diagra m s 2-6 To Display a S ignal 2-9 To Compensate for Module Insertion Loss 2-10 Using Probes with an 83491A 2-11

Operation

Operation

Introduction

Introduction
If you use an 83491/2/3/4A module in an 861 00- se ri e s inf in ii um DC A , informa­tion on using the clock recovery modules, along with specifi cati ons , is included in the instrument’s help system. To acce ss the built-in infor matio n system, referred to as Help, simply click Contents on the Help menu. GPIB programming commands for the mo dule s are documented in the 86100-serie s infiniium DCA p ro grammer’s guide.
If you use an 83491/2/3/4A module in an 83480A DCA, you’ll notice that, unlike most modules designed to be used with the 834 80A, the clock recovery mod­ules do not include Channel keys or menus. Also, there are no GPIB program­ming commands for these modu les when used in an 83480A mainframe.
Multimode module and single-mode reference receivers
Agilent Technologies does not recommend using the Agilent 83492A mul ti ­mode module with si ng le-mode referenc e re c eivers such as the Agilent 83481A, 83482A, or 83485A,B modules. Connecti ng multi m ode to sin­gle-mode fibers cau ses large reflection s a nd insertion loss because of the reduction of the optical fiber’s core from 62.5 µm to 9 µm.
Single-mode module and multimode reference receivers
It is acceptable to use an Agilent 834 93A or an 83494A si ng le-mo de module with a multimode reference receiver such as the Agilent 83486A module. This is true provided that single- m ode fiber is con nected to the Agilent 83493A or 83494A module’s front-panel INPUT connector.
2-2

Front-Panel Features

Operation
Front-Panel Features
Figure 2-1. Agilent 83491/2/3/4A front panels
2-3
Operation
Front-Panel Fe atures
SELECT key Pressing this key changes the data rate selection. The recovered and retimed
clock trigger is sent to the mainframe. The Trigger On Data selection is a bypass mode where the data stream directly triggers the mainframe. Refer to “Blo c k
Diagrams” on page 2- 6 to view a schematic of the normal and bypass paths.
Green and red data-rate lights
The data-rate indicator lights on the Agilent 83491/2/3A and Agilent83494A Option 106 change color between red and green to show which data rate is selected. A red light does not indicate a problem. A red light shows that the adjacent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On Agilent83491A modules for example, the first selec­tion cycle shows 155 Mb/s selected. The second selection cycle shows 1062 Mb/s selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data rate is selected. Repeatedly pressing the SELECT key cycles through the available data rates, as well as the Trigger On Data bypass mode.
UNLOCKED indicator 83491/2/3A
This light shows when clock recovery cannot be established on the signal. If a clock rate is selected, the trigger output to the mainfram e is disabled to pre­vent free-run triggering . Howe ver in bypass mode (Trigger On Data selected), triggering is not disabled. When the UNLOCKED light is on, you can establish a trigger on the data inpu t to the reference receiver.
Trigger Loss indicator 83494A
This light shows when clock recovery cannot be established on the signal. If a clock rate is selected, the trigger output to the mainfram e is disabled to pre­vent free-run triggering . Howe ver in bypass mode (Trigger On Data selected), triggering is not disabled. When the Trigger Loss light is on, you can estab lish a trigger on the data inpu t to the reference receiver.
Auxiliary outputs DATA connector: This co nnector provides a fully regenerated ver sion of the
input signal. It is intended for monitoring purposes only and not for rigorous eye mask compl iance tes ting. Th e frequenc y respons e does no t confo rm to the requirements for eye mask testing as described in ITU-T G.957 and Bellcore GR-253-CORE. On Agilent 83492A and 83493A modules, this port is ampli­tude stabilized fo r input signals greate r than approximatel y –23 dBm. On Agilent 83494A modules, this port is amplitude stabilized for input signals greater than approximately –10 dBm.
2-4
Operation
Front-Panel Features
CLOCK connector: This connector provides the recovered clock signal. You can
use this signal to measure jitter transfer, because this output can track and fol­low input data with very fast jitter; it has a wide bandwidth jitter transfer func­tion when compared to the recovered clock signal which is routed through a rear-panel conne ctor to the mainframe for triggering. Note that the CLOCK Aux- iliary Output remains synchronized to input signals several dB below the onset of errors at the DATA Auxiliary Output.
Input and Output connectors
The input connectors pass the digitally modulated signal to the receiver mod­ule. The input signal, slightly attenuated and available at the OUTPUT connec­tor, is connected to the input of the receiver module. The connectors on optical modules include adapters which can e asily be chang e d to mat ch the type of connectors that are used on your fiber-optic cables. Refer to “Fr ont-
Panel Optical Adapters” on page 3 -2 for a description of the available adapt-
ers.
Multimode and single-mode connections
Agilent 83492A modules use multimode fiber. Connecting the output to the Optical INPUT connector on Agilent 83481/2/5 single-mode modules results in large reflections and insertion loss.
Agilent 83493A and 83494A modules use 9/125 µm single-mode fiber. Connecting mul­timode fiber to the Optical Input connector results in large reflections and insertion loss.
Recovered Clock The recovered clock signal is routed directly to the 86100-series or 83480A
mainframes through the module’s rear panel. This output has a lower jitter modulation bandwidth than the front-panel CLOCK Auxiliary Output. Because of the reduced jitter modulation bandwidth on the mainframe trigger signal, a more complete view of the jitter on the waveform data is obtained.
2-5
Operation

Block Diagrams

Block Diagrams
Figure 2-2. 83491A Block Diagram
Figure 2-3. 83492A and 83493A Block Diagram
2-6
Operation
Block Diagrams
Figure 2-4. 83494A Block Diagram
2-7
Operation
Block Diagrams
Figure 2-5. 83494A Option 106 Block Diagram
2-8
Operation

T o Display a Signal

To Display a Signal
1 Install the module as described in “Installation” on page 1-5. Be sure to connect
all of the cables as described in the procedure.
2 Repeatedly press the SELECT key on the clock reco v ery module until the front-
panel light indicates the proper data rate of the signal.
Green and red data-rate lights
The data-rate indicator lights on the 83491/2/3A and 83494A Option 106 change color between red and green to show which data rate is selected. A red light does not indicate a problem. A red light shows that the adjacent red data rate label is selected. A green light shows that the adjacent green data rate label is selected. Repeatedly pressing the SELECT key cycles through the selections in one color before switching to the opposite color. On 83491A modules for example, the first selection cycle shows 155 Mb/s selected. The second selection cycle shows 1062 Mb/s selected.
Agilent 83494A modules use only the green data-rate indicator light to show which data rate is selected. Repeatedly pressing the SELECT key cycles through the available data rates, as well as the Trigger On Data bypass mode.
• If the UNLOCKED light is on (or TRIGGER LO SS light on the 83494A), clock recov­ery cannot be establishe d on the si gnal.
• Avoid selecting a data rate that is a multiple of the input signal. For example, don’t select a 622 Mb/s data rate if the sign al is reall y at 155 Mb/s.
• If you cannot get the clock recovery module to lock on the signal, make sure that you have sel ect ed th e co rrect dat a rat e and t hat the ma inf ram e tri gger level is adju sted appropria t ely.
• Signa ls displayed us ing a data trig ger are less re liable than usin g a recovered clock. Signals triggered on data can also vary depending upon the trigger level.
2-9
Operation

To Compensate for Module Insertion Loss

To Compensate for Module Insertion Loss
If you are using an 86100-series in fin ii um DCA , re fe r to the instrument’s help system for information on compensating for module insertion loss. Use the fol­lowing steps on an 83480A DCA to allow you to enter an offset to compensate for the insertion loss of the clock recovery module. This provides accurate amplitude measu r e ments at the input to t he clock recovery module.
1 Disconnect the cable from the clock recovery module’s Input connector. 2 Measure the signal using a power meter. You can use either the 83480A’s built-
in power meter or an external power meter.
3 Reconnect the cable to the clock recovery module. 4 Disconnect the cable from the reference receiver module’s input connector. 5 Measure the signal using a power meter. You can use either the 83480A’s built-
in power meter or an external power meter.
6 Subtract the two measurements to determine the insertion loss of the module.
Insertion loss: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________
7 On the reference receiver module, pres s the front-panel channel SETUP key. 8 Press External scale, and set the Atten units to “decibel”. 9 Press Attenuation, and enter the value calculated in Step 6.
2-10
Operation

Using Probes with an 83491A

Using Probes with an 83491A
You can use external passive and active probes with the 83491A electrical clock recovery module. If you are using an 86100-series infiniium DCA, refer to the infiniium DC A’s help system; the pro c e dures in this section ar e fo r use on an 83480A DCA.
The following probes are available for use with 83491A clock recovery mod­ules:
• Agilent N1020A TDR probe. This passive probe (1:1, 50) provides a fixture for positioning an d holding the probe tip on the device being tested .
• Agilent 54701A 2.5 GHz active probe. This is a 100kΩ, 10:1, probe.
• Agilent 54006A 6 GHz handheld low-impedanc e probe. This passive probe (10:1, 500, 20:1, 1k) has an input capacitance of 0.25 pf.
• Agilent 1163A 1 GHz resistive-divider probe. This passive 500 probe has an input capacitance of 1.5 pf.
If you are using an 83480A DCA, use the procedures in this section to gener­ate vertical scale factors. These factors are applie d to the calibr ati on of the reference receiver module’s electrical channel. When selecting a probe, keep in mind that the input impedance of the 83491A is 50Ω.
If the probe being calibrated has an attenu ati on factor that allows the instru­ment to adjust the gain to produce even steps in the vertical scale factors, the instrument will do so. Typically, probes have standard attenuation factors such as divide by 10, divide by 20, or divide by 100. Because the following pr oce­dures include com pensation for inser tion loss of the clock re c ov ery module, do not perform the procedur e “To Compensate for Module Insertion Loss” on
page 2-10.
To compensate for a passive probe
1 Connect the probe to the Input connector on the 83491A clock recovery module. 2 Attach the pro be ti p to th e CAL hook that is located near the floppy disk drive. 3 Press the reference receiver module’s front-panel channel SETUP key.
2-11
Operation
Using Probes with an 83491A
4 Press Calibrate and then Calibrate probe.
To compensate an 54701A active probe
1 Connect the 83491A output to the electrical measurement channel input. 2 Connect the probe to the Input connector on the 83491A clock recovery module. 3 Connect the probe po wer cable to the Probe Power connect o r on the reference
receiver modu le .
4 Attach the pro be ti p to th e CAL hook that is located near the floppy disk drive. 5 Press the reference receiver module’s front-panel channel SETUP key. 6 Press Calibrate and then Calibrate probe.
To compensate for other devices
The information in this section applies to both optical and electr ical measure­ments. Since the mainframe’s CAL signal is a voltage source, it cannot be used to calibrate to the probe tip when the units are set to Ampere, Watt, or Unknown. Instead , s et t he ext ern al gai n an d ex te rnal o ff set t o com pens ate for the actual characteristics of the devic e. If you do not know the actual charac­teristics, you can refer to the typical specifications that came with the device.
1 Press the reference receiver module’s front-panel channel SETUP key. 2 Press External scale. 3 Press Atten units Ratio
, Attenuation 1:1, and then Units Ampere (Volt, Watt, or
Unknown).
4 Press Ext gain, and enter the actual gain characteristics of the device. 5 Press Ext offset, and enter the offset introduced by the device.
2-12
3
Front-Panel Optical Adapters 3-2 In Case of Difficulty 3-3 Error Messages 3-5 Electrostatic Discharge Infor m ati on 3-7 Cleaning Connections for Accurate Measurements 3-9

Reference

Reference

Front-Panel O ptical Adapter s

Front-Panel Optical Adapters
Front Panel Fiber-Optic Adapter
Description Agilent Part Number
Diamond HMS-10 81000AI
a
FC/PC
D4 81000GI
SC 81000KI
DIN 81000SI
ST 81000VI
Biconic 81000WI
Dust Covers FC connector 1005-0594 Diamond HMS-10 connector 1005-0593 DIN connector 1005-0595 ST connector 1005-0596 SC connector 1005-0597
81000FI
a. The FC/PC adapter is the standard adapter supplied with the instrument.
3-2

In Case of Difficulty

In Case of Difficulty
This section provides a list of suggestions for you to follow if the plug-in mod­ule fails to operate. A list of messages that may be displayed is also included in this chapter. Before calling Agilent Technologies or returning th e un it for ser­vice, a few minutes spent performing some simple checks may save waiting for your instrument to be repaired.
If the mainframe does not operate
Is the line fuse good? Does the line socket have power?
Reference
Is the unit plugged in to the prop er ac power source? Is the mainframe tur ned on? Is the rear-panel line switch set to on? Will the mainframe po w er u p without the plug-in module installed?
If the plug-in does not operate
Is the plug-in module firmly seated in the mainframe slot? Are the knurled sc rews at the bottom of the plug-in module finger-tight? Is the clock recovery module set to the modulation rate of the input signal? If other equipment, cables, and connectors are being used with the plug-in
module, are they co nnected properly and o perating correc tly?
Review the procedure for the test being performed when the problem ap-
peared. Are all the settings correct? Can the problem be reproduced?
Are the connectors clean? See “Clean ing Connections fo r Accurate Measur e-
3-3
Reference
In Case of Difficulty
ments” on page 3-9 for more information.
Make sure that the instrument is ready to acquire data. Find any signals on the channel inputs by using the Autoscale function.See if any signals are present at the channel inputs by using Freerun triggering.Make sure Channel Display is on. Make sure the channel offset is adjusted so the waveform is not clipped off the
display.
Make sure the mainframe identifies the plug-in module:
• One the 86100-series DCA, select the All Calibrati ons command on the Cal­ibrate menu. Confirm that the Model number and Seri al Number are listed in the displayed dialog box.
• On the 83480A DCA, press Utility, then System config. Confirm that the Model number is listed in the box labeled “Plug-ins”. If “~known” is displayed in­stead of the model number of the plug-in module, remove and reinsert the plug-in module in the same slot. If “~known” is still displayed, the mainframe may need to have the latest operating system firmware i n stalled. Options 001 and 002 provide this firmware on a 3.5 inch diskette. To load new firm­ware, follow the instructions provided with this diskette. If you do not have the optional diskette, contact your local Agilent Technologies service office (refer to “Contacting Agilent Technologies” on page 1-12).
If the model number of the plug-in module is list ed, the mainframe ha s identi­fied the plug-in.
3-4
Reference

Error Messages

Error Messages
This section pr o v ides description of error messages that ar e displayed if you are using the module in an 83480A DCA. If you are using an 86100-series infi­niium DCA, refer to the instrument’s help system for information on error messages displayed on that mainframe.
Memory error occurred in plug-in_:Try reinstalling plug-in
The mainframe could not correctly read the contents of the memory in the plug-in.
Remove and reinstall the plu g- i n mo dule. Each time a plug-in is instal le d, the
mainframe re-reads the memory in the plug-in module.
Verify the plug-in module is firmly seated in the mainframe slot. Verify the knur led scr ews at the botto m of the pl ug-in mo dule ar e finger -tight . Install the plug-in in a different slot in the mainframe.
Busy timeout occurred with plug-in_:Try reinstalling plug-in
The mainframe is having trouble communicating with the plug-in module . Make sure there is a goo d connection between the mainframe and the plug-in module.
Remove and reinstall the plu g- i n module. Verify the plug-in module is firmly seated in the mainframe slot. Verify the knurled screws at the bottom of the plug-in module are finger-tight.Install the plug-in in a different slot in the mainframe.
3-5
Reference
Communications failure exists at slot_:Service is required
Communications failure exists at slot_:Service is required
An illegal hardware state is detected at the mainframe-to-plug-in module interface of the specifie d sl ot.
• If the slot is empty, there is a mainframe hardware problem. Refer to the Agilent 83480A, Agilent 54750A Service Guide.
• If a plug-in is installed in the slot, there is a plug-in module hardware problem. Return the plug-in module to a qualifie d service departmen t.
ID error occurred in plug-in_:Service is required
The informatio n read from the memory of the plug-in module do e s no t m a tc h the hardware in the plug-in module. This can be caused by a communication problem between the mainframe and the plug-in module. Make sure there is a good connection between the mainframe and the plug-in.
Remove and re-install the plug-in module.Verify the plug-in module is firmly seated in the mainframe slot.Verify the knurled screws at the botto m of the plug-in module ar e fi nger tight.The standard Agilent 54750A mainframe does not accept the Agilent 83491/2/
3/4A module. To use the module, a firm ware upgrade must first be installed. Order the Agilent 8348 0K communi cati ons firmware kit and install according to the instruct ions.
The Agilent 83480A, Agilent 54750A mainframes do not accept plug-in mod-
ules designed for use with the Agilent 54710A, 54720A.
Plug-in is not supported:System firmware upgrade is needed
The mainframe ma y need to have the latest o perating system firmware installed. Options 001 and 002 provide this firmware on a 3.5 inch diskette. T o load the new firmware, follow the instructions provided with the diskette. If you do not have the optional diskette, contact your local Agilent Technologies service office.
3-6
Reference

Electrostatic Discharge Information

Electrostatic Discharge Information
Electrostatic dis c h arge (ESD) can damage or destro y electronic components . All work on electronic assemblies should be performed at a static-safe work station. The following figure shows an example of a static-safe work station using two types of ESD protecti on:
• Conductive table-mat and wrist-strap combination.
• Conductive floor-mat and heel-strap combination.
Figure 3-1. Static-safe work station
3-7
Reference
Electrostatic Discharge Informa tio n
Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone.
T o ensure user safety, the static-safe accessories must provide at least 1 M of isolation from ground. Refer to Table 3-1 for information on ordering static­safe accessories.
WARNI NG These techniques for a static-safe work station should not be used
when working on circuitry with a voltage potential greater than 500 volts.
Reducing ESD Damage
The following suggestions ma y he lp reduce ESD damag e that occurs during testing and servic ing operations.
• Personnel shoul d be grounded with a resi stor-isolated wrist strap before re­moving any assembly from the unit.
• Be sure all instrument s are properly earth-grounded to prevent a buildup of static charge.
Table 3-1.Static-Safe Accessories
Agilent Part Number
9300-0797
9300-0980 Wrist-strap cord 1.5 m (5 ft) 9300-1383 Wrist-strap, color black, st ainless steel, without cord, has fo ur adjustable
9300-1169 ESD heel-strap (reusable 6 to 12 months).
3-8
Description
Set includes: 3M static control mat 0.6 m ft) ground wire. (The wrist-strap and wrist-strap cord are not included. They must be ordered separately.)
links and a 7 mm post-type connection.
× 1.2 m (2 ft× 4 ft) and 4.6 cm (15
Reference

Cleaning Connections for Accurate Measurements

Cleaning Connections for Accurate Measurements
Today, advances in measurement capabilities make connectors and connec­tion techniques more important than ever. Damage to the connectors on cali­bration and verification devices, test ports, cables , and ot her device s can degrade measurement accuracy and damage instruments. Replacing a dam­aged connector can cost thousands of dollars, not to mention lost time! This expense can be avoided by observing the simple precautions presented in this book. This book also contains a brief list of tips for caring for electrical connec­tors.
Choosing the Right Connector
A critical but often overlooked factor in making a good lightwave measure ­ment is the selection of the fiber-optic connector. The differences in connec­tor types are mainly in the mechanic a l a ssembly that holds th e fe rrule in position against another identical ferrule. Connectors also vary in the polish, curve, and conce ntricity of the core within the cladding. Mating one style of cable to another requires an adapter. Agilent T echnologies offers adapters for most instruments to allow testin g w i th many different c ables. Figure 3-2 on
page 3-10 shows the basic componen ts of a ty pica l conn ect ors .
The system tole ranc e for reflection and in se rtion loss must be kn ow n w he n selecting a connector from the wide variety of currently available connectors. Some items to consider when selecting a connector are:
• How much insertion loss can be allowed?
• Will the connector need to make multiple connections? Some connectors are better than others, and some are very poor for making repeate d connections.
• What is the reflection tolerance? Can the system take reflection degradation?
• Is an instrument-grade connector with a preci sion core alig nmen t re qu ir ed ?
• Is repeatability tolerance for reflection and loss important? Do your specifica-
3-9
Reference
Cleaning Connections for Accurate Measurements
tions take repeatability uncertain ty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be re­quired? For example, many DFB lasers cannot operate with reflections from connectors. Often as much as 90 dB isolation is needed.
Figure 3-2. Basic components of a connector.
Over the last few ye a rs, the FC/PC style c onnector has emerge d as the most popular connector for fiber-optic applications. While not the highest perform­ing connector, it represents a good compromise between performance, reli­ability, and cost. If properly maintained and cleaned, this connector can withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most connectors, including the FC/PC style, can deliver. These instruments cannot tolerate connectors with the l arge non-concentricities of the fiber common with ceramic style ferrules. When tighter alignme n t is required, Agi lent Technologies instru ments typi cally use a connector such as the Diamond HMS-10, which has concentric tolerances within a few tenths of a micron. Agi­lent Technologies then uses a special unive rsal adapter, which allows oth er cable types to mate with this precision connector. See Figure 3-3.
3-10
Reference
Cleaning Connections for Accurate Measurements
Figure 3-3. Universal adapters
The HMS-10 enca se s the fiber within a soft nickel silver (C u/Ni/Zn) center which is surrounded by a tough tungsten carbide casing, as shown in
Figure 3-4.
Figure 3-4. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber to be moved to the desired positi on. This process first stakes the soft n ickel silver to fix the fiber in a near-center location, then uses a post-active staking to shift the fiber into the desired position within 0.2 µm. This process, plus the keyed axis, allo ws very precise core-to-core alignments. This conne c to r is found on most Agilent Technologies lightwave instruments.
3-11
Reference
Cleaning Connections for Accurate Measurements
The soft core, wh ile allowing precise c entering, is also th e c h ief liability of the connector. The soft material is easily damaged. Care must be taken to mini­mize excessive scratching and wear. While minor wear is not a problem if the glass face is not affected, scratches or grit can cause the glass fiber to move out of alig nment. Also, if unkeye d co nnecto rs ar e used , the nicke l sil ver ca n be pushed onto the glass surface. Scratches, fiber movement, or glass contamina­tion will cause lo s s of sign al and increa sed re flect ion s, resu ltin g in p oor retu rn loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immedi­ately obvious to the naked eye, poor measurements result without the user being aware. Microscopic examination and return loss measure men ts are the best way to ensure good measurements. Good cleaning practices can help ensure that optim u m c onnector performanc e is maintained. With glass - to ­glass interface s, any degradatio n of a ferrule or the end of the fi ber, any stray particles, or finger oil can have a significant effect on connector performance. Where many repeat connections are required, use of a connector saver or patch cable is recommended.
Figure 3-5 shows the end of a clean fiber-optic cable. The dark circle in the
center of the micrograph is the fiber’s 125 µm core and cladding which carries the light. The surrounding area is the soft nickel-silver ferrule. Figure 3-6 shows a dirty fiber end from neglect or perhaps improper cleaning. Material is smeared and gr ou nd i nto t he end o f t he f ib er cau si ng l igh t sc att erin g an d po or reflection. Not only is the precision polish lost, but this action can grind off the glass face and destroy the connector.
Figure 3-7 shows physical damage to the glass fibe r end caused by either
repeated connections made without removing loose particles or using improper cleaning tools. When severe, the damage of one connector end can be transferred to another good connector en dface that comes in contact with the damaged one. Periodic checks of fiber ends, and replacing connecting cables after many connections is a wise practice.
The cure for these p roblems is discipline d c o nnector care as described in the following list and in “Cleaning Connectors” on page3-16.
3-12
Reference
Cleaning Connections for Accurate Measurements
Use the following guidelines to achieve the best possible performance when making measurements on a fiber-optic system:
• Never use metal or sharp objects to clean a connector and never scrape the connector.
• Avoid matching gel and oils.
Figure 3-5. Clean, problem-free fiber en d and f errule.
Figure 3-6. Dirty fiber end and ferrule from poor cleaning.
3-13
Reference
Cleaning Connections for Accurate Measurements
Figure 3-7. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The oil or gel grabs and holds grit that is then ground into the end of the fiber. Also, some early ge ls were designed fo r use with the FC, non-con t ac ting con­nectors, using small glass spheres. When used with contacting connectors, these glass balls can scratch and pit the fiber. If an index matching gel or oil must be used, apply it to a freshly cleaned connector, make the measurement, and then immed ia tely clean it off. Never us e a gel for longer- term connections and never use it to improve a damaged connector. The gel can mask the extent of damage and continued use of a damaged fiber can trans fer damage to the instrument.
• When inserting a fiber-optic c able into a connector, gently inse rt i t in a s straight a line as possible. Tipping and inserting at an angle can scrape material off the inside of the connector or even break the inside sleeve of connectors made with ceramic material.
• When inserting a fiber-optic connector into a connector, make sure that the fi­ber end does not touch the outside of the mating connec tor or adapter.
• Avoid over tightening connections. Unlike common electrical connections, tighte r is not better. The purpose of
the connector is to bring two fiber ends together. Once they touch, tightening only causes a greater force to be applied to the delicate fibers. With connec­tors that have a convex fiber end, the end can be pushed off-axis resulting in misalignmen t a nd e xc essive return lo ss. Many measurem ents are actually improved by backing off the connector pressu re . Also, if a piece of grit does happen to get by the cleaning procedure, the tighter connection is more likely to damage the glass. Tighten the connectors just until the two fibers touc h.
3-14
Reference
Cleaning Connections for Accurate Measurements
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best con­nector possible. Replace connecting cables regularly. Frequently measure the return loss of the connector to check for degradation, and clean every connec­tor, every time.
All connectors should be treated like the high-quality lens of a good camera. The weak link in instrument and syste m reliability is of te n the inappropriate use and care of the connector. Because current connectors are so easy to use, there tends to be reduced vigilance in connector care and cleaning. It takes only one missed cle a ning for a piece of grit to permanently dama g e the glass and ruin the conne c to r.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipme nt are a good indica­tion that you have good connections. S ince return loss and insertion loss are key factors in determining optical connector performance they can be used to determine connector degradation. A smooth, polished fiber end should pro­duce a good return- loss measurement. The quality of the polish establishes the difference between the “PC” (physical contact) and the “S upe r PC” con­nectors. Most connectors today are physical contact which make glass-to-glass connections, therefore it is critical that the area around the glass core be clean and free of scratches. Altho ug h the major area of a connector, excluding the glass, may show scra tc hes and wear, if the glass has maintained its polished smoothness, the connector can st ill provide a good low lev el return loss con­nection.
If you test your cables and acces so ries for insertion loss and re turn loss upon receipt, and retain the measured data for comparison, you will be able to tell in the future if any degradation has occurred. Typical values are less than 0.5 dB of loss, and somet im es as little as 0.1 dB of lo ss with high performanc e con­nectors. Return loss is a measure of reflection: the less reflection the bett er (the larger the return loss, the smaller the reflection). The best physically contacting connectors have return losses better than 50 dB, although 30 to 40 dB is more common.
3-15
Reference
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections on the cable end face can be detected as well as cracks or chips in the fiber itself. Use a microscope (100X to 200X magni fi cati on) to in spect the entire end face for contam ina tion, raised meta l , o r de nts in the metal as well a s an y other imperfections. Inspect the fiber for cracks and chips. Visible imperfec­tions not touching the fibe r cor e m ay no t affe ct pe rformance (unless the imperfections keep the fibers from contacting).
WARNI NG Always remove both ends of fiber-optic cables from any instrument,
system, or device before visually inspecting the fiber ends. Disable all optical sources before disconnecting fiber-optic cables. Failure to do so may result in permanent injury to your eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber­optic cables and Agilent Technologies universal adapter s. The initi al cleani ng , using the alcohol as a solvent, gently removes any grit and oil. If a caked-on layer of material is still present, (this can happen if the beryllium-copper sides of the ferrule retainer get scraped and deposited on the end of the fiber during insertion of the cable), a second cleaning should be performed. It is not uncommon for a cable or connector to require more than one cleanin g.
CAUTION Agilent Technologies strongly recommends that index matching compounds
not be applied to their instruments and accessories. Some compounds, such as gels, may be difficult to remove and can contain damaging particulates. If you think the use of such compounds is necessary, refer to the compound manufacturer for infor mat ion on application and cleaning procedures.
Table 3-2.Cleaning Accessories
Item AgilentPart Number
Any commercially available denatured alcohol — Cotton swabs 8520-0023 Small foam swabs 9300-1223 Compressed dust remover (non-resi due) 8500-5262
3-16
Reference
Cleaning Connections for Accurate Measurements
Table 3-3. Dust Caps Provided with Lightwave Instruments
Item AgilentPart Number
Laser shutter cap 08145-64521 FC/PC dust cap 08154-44102 Biconic dust cap 08154-44105 DIN dust cap 5040-9364 HMS10/dust cap 5040-9361 ST dust cap 5040-9366
To clean a non-lensed connector
CAUTION Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber e nds that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end after cleaning.
2 Clean the ferrules and o the r parts of the connector while avoidi ng the end of
the fiber.
3 Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper. 4 Clean the fiber end with the swab or lens paper.
Do not scrub durin g this initial clea ning because grit c a n be c a ught in the swab and become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint-free cott on swab or len s
paper.
6 Blow across the connector end face from a distance of 6 to 8 inches using
filtered, dry, compressed air. Aim the compressed air at a shallow angle to the fiber end face. Nitrogen gas or compressed dust remover can also be used.
3-17
Reference
Cleaning Connections for Accurate Measurements
CAUTION Do not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the compressed air canister.
7 As soon as the connec tor is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the con­nector again. Often a second cleaning will restore proper performance. The second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many Agilent Technologies instruments employ a universal adapter such as those shown in the following picture. These adapters allow you to connect the instrument to different types of fiber-optic cables.
Figure 3-8. Universal adapters.
1 Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The foam swabs listed in this section’s int roduction are smal l en ough to fit into adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and the risk of other conta m ination buildup on the inside of adapte rs greatly out­weighs the risk of contamination by foam swabs.
2 Clean the adapter with the foam swab. 3 Dry the inside of the adapter with a clean, dry, foam swab. 4 Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into the air. Refer to instructions provided on the compressed air canister.
3-18
4
83491A Specifications 4-3 83492A Specifications 4-4 83493A Specifications 4-6 83494A Specifications 4-8 83491/2/3/4A Operating Specifications 4-10 Regulatory Information 4-11

Specifications and Regulatory Information

Specifications and Regulatory Information
Specifications and Regulatory Information
Specifications and Regulatory Information
This chapter lists specifications and characteristics of the clock recovery mod­ules. Specifications apply over the temperature range +15°C to +35°C (unless otherwise noted) after the instrument’s temperature has been stabilized after 60 minutes of continuous operation.
Specifications Specifications described warranted performance.
Characteristics Characteristics provide useful, nonwarranted, information about the func-
tions and performance of the instrument. Characteristics are printed in italics.
Calibration cycle Agilent Technologies warrants instrument specifications over the recom-
mended calibrat ion interv al. T o ma intain speci fications , periodic re calibrati ons are necessary. We recommend that the Agilent 83491/2/3/4A be calibrated at an Agilent Technologies service facility eve ry 24 months.
4-2
Specifications and Regulat ory Information

83491A Specifications

83491A Specifications
Table 4-1.83491A Specifications
Clock Recovery Phase Locked Loop Bandwidth (characteristic)
Internal Path Triggering External Output
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s 1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
Data triggering (character istic) 50 Mb/s to 2500 Mb/s
Operatin g in put power level
a b
Triggering operati on, all rates
–10
10
BER, all rates
c
50 to 70 kHz 4 to 5 MHz
±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
–10 dBm to 3 dB m –10 dBm to 3 dB m
Insertion loss (through path)
DC through 2500 MHz
Output jitter, all rates
d
7 dB
0.0125 UI
Maximum continuous electrical power bef or e damage (characteristic) 1W peak DATA and CLO CK output amplitude, all rates (characteristic) 0.5Vp- p INPUT electrical return loss
DC through 1250 MHz (characteristic) 1250 MHz through 2500 MHz (character ist ic)
20 dB 15 dB
DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (charact er istic) 2000 MHz through 2500 MHz (character ist ic)
a. Source extinction ratio 8.2 dB when measured per TIA/EIA OFSTP-4A. b. Operating power level applies over temperature range 25 c. Better than 10 d. Measured on an oscilloscope eye diagram with PRBS 223–1 test pattern.
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°.
10 dB 6 dB
rms
4-3
Specifications and Regulatory Information

83492A Specification s

83492A Specifications
Table 4-2.83492A Specifications (1 of 2)
Wavelength range (characterist ic) 750 nm to 860 nm and
1000 nm to 1600 nm
Optical INPUT and OUTPUT fiber (characteristic) 62.5/125 multimode
Optical insertion loss (th rough path)
750 nm to 860 nm 1000 nm to 1600 nm
Optical return loss
b
a
5.0 dB 5.0 dB
28 dB
Clock Recovery Phase Locked Loop Bandwidth (characteristic)
Internal Path Triggering External Output
50 to 70 kHz 4 to 5 MHz
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
1062.50 Mb/s
1250 Mb/s
2125.00 Mb/s
2488.32 Mb/s
2500.00 Mb/s
±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
Data triggering (character istic) 50 Mb/s to 2500 Mb/s
Operating input power level
c d
750 nm to 860 nm
Triggering op e r a t io n, all rates
–10
BER, all rates
10
e
–10 dBm to 3 dB m –10 dBm to 3 dB m
1000 nm to 1600 nm
Triggering op e r a t io n, all rates
–10
BER, all rates
10
Output jitter, all rates
f
g
–13 dBm to 3 dB m –13 dBm to 3 dB m
0.0125 UI
rms
4-4
Specifications and Regulat ory Information
83492A Specifications
Table 4-2.83492A Specifications (2 of 2)
Maximum continuous optical power bef or e damage (characteristic) 10 mW peak DATA and CLO CK output amplitude, all rates (characteristic) 0.5Vp- p DATA and CLOCK electrical return loss
50 MHz through 2000 MHz (charact er istic) 2000 MHz through 2500 MHz (character ist ic)
a. Minimum loss in 850 nm window. b. Single-mode backreflection tested with FC/PC adapter and single-mode fiber. Optical output terminated
with > 33 dB return lo ss. R etur n loss with fully filled 62.5 c. Source extinction ratio d. Operating power level applies over temperature range 25 e. Better than 10 f. Better than 10 g. Measured on an oscilloscope eye diagram with PRBS 2
8.2 dB when measured per TIA/EIA OFSTP-4A.
–10
BER when tested with PR BS 223–1 pattern.
–10
BER when tested with PR BS 223–1 pattern.
µm core multimode fiber may be sl igh tly lower .
°C ± 5°.
23
–1 test pattern.
10 dB6 dB
4-5
Specifications and Regulatory Information

83493A Specification s

83493A Specifications
Table 4-3.83493A Specifications
Wavelength range (characteristic) 1000 nm to 1600 nm Optical INPUT fiber (characteristic) 9/125 s ingle mode
Optical insertion loss (through path) Optical return loss
a
1.5 dB 28 dB
Clock Recovery Phase Locked Loop Bandwidth (characteristic)
Internal Path Triggering External Output
50 to 70 kHz 4 to 5 MHz
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s 1250 Mb/s
2488.32 Mb/s
2500.00 Mb/s
±0.1% ±0.1% ±0.1% ±0.1% ±0.1%
Data triggering (character istic) 50 Mb/s to 2500 Mb/s
Operating input power level
Triggering operation, all rates
–10
10
BER, all rates
Output jitter, all rates
b c
–20 dBm to 3 dB m
d
e
–17 dBm to 3 dB m
0.0125 UI
rms
Maximum continuous optical power bef or e damage (characteristic) 10 mW peak DATA and CLO CK output amplitude, all rates (characteristic) 0.5Vp- p DATA and CLOCK output electrical return loss
50 MHz through 2500 MHz (characteristic)
2.5 GHz through 10 GHz (characteristic)
10 dB 8 dB
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss.
4-6
Specifications and Regulat ory Information
b. Source extinction ratio 8.2 dB when measured per TIA/EIA OFSTP-4A. c. Operating power level applies over temperature ra nge 25 d. Better than 10
–10
BER when tested with PR BS 223–1 pattern.
e. Measured on an oscilloscope eye diagram with PRBS 2
°C ± 5°.
23
–1 test pattern.
83493A Specifications
4-7
Specifications and Regulatory Information

83494A Specification s

83494A Specifications
Table 4-4.83494A Specifications (1 of 2)
Wavelength range (characteristic) 1200 nm to 1600 nm Optical INPUT fiber (charact er istic) 9/125 single mode
Optical insertion loss (through path) 1.5 dB Optical return loss
Clock Recovery Phase Locked Loop Ban dwidth (characteristic)
Internal Path Triggering External Output
a
28 dB
90 kHz 4 to 5 MHz
Clock recovery rates (NRZ coding)
155.52 Mb/s
622.08 Mb/s
2488.32 Mb/s
2666.06 Mb/s (Options 106 and 107)
9953.00 Mb/s (Standard)
10312.500 Mb/s (Option 103)
10664.228 Mb/s (Option 106)
10709.225316 Mb/s (Option 107)
±0.1% ±0.1% ±0.1% ±0.1% ±0.03% ±0.03% ±0.03% ±0.03%
Data triggering (charact eristic) 50 Mb/s to 2500 Mb/s
Operating input power level
b c
Triggering operation
9953.00 Mb/s, 10312.500 Mb /s, 10664.228 Mb/s, and
–8 dBm to 3 dBm
10709.225316 Mb/s Triggering operation, all other rates
-12 dBm to 3 dBm
-14 dBm to 3 dBm (t y pical)
–10
10
BER, all rates
d
0 dBm
4-8
Specifications and Regulat ory Information
83494A Specifications
Table 4-4.83494A Specifications (2 of 2)
Output jitter e
all rates
155.52 Mb/s
622.08 Mb/s
2488.32 Mb/s
Maximum continuous optical power before damage (characteristic) 10 mW peak DATA and CLOCK output amplitude, all rates (characteristic) 0 .5Vp-p DATA and CLOCK output electrical return loss
50 MHz through 2500 MHz (characteristic)
2.5 GHz through 10 GHz (characteristic)
0.02 UI
0.01 UI
0.01 UI
0.01 UI
10 dB 8 dB
rms
(typical)
rms
(typical)
rms
(typical)
rms
a. Tested with FC/PC adapter. Optical output terminated without > 33 dB return loss. b. Source extinction ratio c. Operating power level applies over temperature range 25 d. Better than 10 e. Measured on an oscilloscope eye diagram with P RBS 2
8.2 dB when mea sured per TIA/EI A OFSTP-4A.
–10
BER when tested with PRBS 223–1 pattern.
°C ± 5°.
23
–1 test pattern.
4-9
Specifications and Regulatory Information

83491/2/3/4A Operating S pecifications

83491/2/3/4A Operating Specifications
Table 4-5. Operating Specifications
Use Indoor Temperature
Operating Non-operating
Altitude
Operating Non-operating
0°C to +55°C –40
°C to +70°C
4600 m (15,000 ft) 15,300 m (50,000 ft)
Humidity
Operating Non-operating
Net weight appr ox i mately 1.2 kg (2.6 lb.) Shipping weight approximately 2.1 kg (4.6 lb.) Dimensions H x W x D 131.66 mm x 80.14 mm x 266.9 mm
Power Requirements Supplied by mainframe
4-10
up to 90% relative humidity at <35 up to 90% relative humidity at <35
(5.18 in x 3.15 in x 10.5 i n )
°C °C
Specifications and Regulat ory Information

Regulatory Information

Notice for Germany: Noise Declaration
Acoustic Noise Emission Geraeuschemission LpA < 70 dB LpA < 70 dB Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t.19
Regulatory Information
Notice for Canada: Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada.
4-11
Specifications and Regulatory Information
Regulatory Information
4-12

Index

Numerics
83494A, 4-8
A
adapter cable, 1-9 Agilent 54701A active probe, 2-12 Agilent 83492A
maximum input level, 1-10
Agilent Technologies, contacting, 1-12
B
Bellcore GR-253-CORE, 2-4 block diagram, 2-6 built-in information system, 2-2 bypass mode, 2-4
C
cabinet, cleaning, 1-4 calibration cycle, 4-2 care
of cabinet, 1-4 care of fiber optics, 1-3 channel
key, 1-3
setup, 1-3 Channel key, 2-2 characteristics, defined, 4-2 cleaning
adapters, 3-18
cabinet, 1-4
fiber-optic connections, 3-9, 3-17
non-lensed connectors, 3-17 CLOCK Auxiliary Output connector, 2-5 compensation
Agilent 54701A active probe, 2-12
insertion loss, 2-10
other devices, 2-12
passive probe, 2-11 compressed dust remover, 3-16 connector
care, 3-9 Contents topic, 2-2 cotton swabs, 3-16 customer assistance, 1-4
D
damaged shipment, 1-5 DATA Auxiliary Output connector, 2-4, 2-5 data rate
indicator lights, 1-11
data rates, 1-2
multiple of, 1-10, 2-9 declaration of conformity, 1-4 dust caps, 3-17
E
electrostatic discharge, 1-3, 3-7 EMC requirements, 4-11 ESD
reducing damage caused by ESD, 3-8
static-safe work station, 3-8 extender cables, 1-7
F
fiber
adapters, 3-2
care of, 1-3
multimode, 1-2
reflections, 2-5
single-mode, 1-2
single-mode connections, 2-5 fiber optics
cleaning connections, 3-9
connectors, covering, 1-13 firmware
upgrade kit, 1-5
version required, 1-5 foam swabs, 3-16 front panel, 2-3
adapters, 3-2
features, 2-3
G
GPIB programming, 2-2 green light, 1-11
H
Help, 2-2
Index-1
Index
I
input
connector, 3-9 Input connector, 1-3, 1-9 input connector, 1-2 input signal, maximum safe, 1-10 insertion loss, 2-2, 2-5
compensation, 2-10 instrument
returning for service, 1-12 ITU-T G.957, 2-4
M
mainframe troubleshooting, 3-3 maximum input level, 1-10 module
installing, 1-7
N
noise declaration, 4-11
O
online help, 2-2 Output connector, 2-5
P
packaging for shipment, 1-13 plug-in module
serial number, 1-5 programming, 2-2
returning for, 1-12
shipping
procedure, 1-13 single-mode fiber, 2-5 specifications, 4-2
Agilent 83491A , 4-3
Agilent 83492A , 4-4
Agilent 83493A , 4-6
Agilent 83494A , 4-8
defined, 4-2
operating, 4-10
temperature range, 4-2 swabs, 3-16
T
technical assistance, 1-4 testing, responsi v ity , 1-7 trigger bypass mode, 2-4 Trigger Loss light, 1-10, 2-9 Trigger On Data light, 2-4 troubleshooting, 3-3
U
Unlocked light, 1-10, 2-4, 2-9
R
red light, 1-11 regulatory information, 4-2 responsivity, testing, 1-7 returning for service, 1-12
S
safety information, 1-4, 1-7 SELECT key, 1-10, 2-4 serial number, 1-5 service, 1-12
Index-2
Loading...