Agilent 83236B Data Sheet

Page 1
Cellular to PCS frequency translator for Agilent Technologies’ successful 800 MHz test sets
• 8920B Option 800 TDMA Mobile Test Set
• 8924E CDMA Mobile Station Service Test Set
TDMA and CDMA Test Solutions
The Agilent Technologies 83236B PCS Interface is a cellular to PCS frequency translator. When com­bined with Agilent Technologies TDMA and CDMA RF test sets, it provides PCS test solutions for your TDMA and CDMA PCS phones and cell sites.
83236B PCS-band test solutions build on the fol­lowing successful 800 MHz test sets.
• 8920B Option 800 TDMA Mobile Test Set
• 8924E CDMA Mobile Station Service Test Set
For configuration information, refer to the Agilent 8920B Configuration Guide, pub. number 5968­5919E.
PCS-band Coverage
The 83236B translates 800 MHz measurement capa­bilities to the 1710 to 1990 MHz frequency range. This frequency range covers the International (1710 to 1880 MHz), Korean (1715 to 1870 MHz), and North American (1850 to 1990 MHz) PCS bands.
Measurement Accuracy
Power measurement accuracy and speed are main­tained at PCS-band frequencies with an internal power meter for measurements on CW, CDMA, and
TDMA (π/4 pulsed or continuous) signals.
Agilent 83236B
PCS Interface
Product Overview
Wireless test solutions for your PCS-band phones
Page 2
2
Specifications describe the instrument’s warranted performance after a 30-minute warm-up period and are valid over the entire operating range unless otherwise noted.
Supplemental Characteristics are intended to provide additional information useful in applying the instrument by giving typical, but non-warranted performance parameters. These are shown in ital­ics or labeled as "typical," "usable to," or "nominal."
Generator Output Path
1
(RF IN/OUT and RF OUT Only connector)
Frequency
Frequency Range:
Through Path:
824 MHz to 849 MHz 869 MHz to 894 MHz
Conversion Path:
1710 MHz to 1785 MHz 1805 MHz to 1910 MHz 1930 MHz to 1990 MHz
Frequency Settling Time: <10 ms
Output
RF IN/OUT Connector: Output Level Range: –130 dBm to –20 dBm
RF OUT Only Connector: Output Level Range: –130 dBm to –10 dBm
Level Accuracy:
±1.8 dB, @ 25 °C ±10 °C ±2.0 dB, @ 0 °C to 55 °C
Typically ±1.0 dB
Output Level Settling Time: <80 ms
Spectral Purity (83236B only)
Spurious
Frequency (MHz)
824 to 849
Type of Spurious 869 to 894 894 < to <1710 1710 to 1990
Harmonic <–30 dBc <–30 dBc
Non-Harmonic <–60 dBc* <–25 dBc** <–60 dBc*
*
Offsets >5 kHz
**
For carrier levels >–100 dBm
SSB Phase Noise: <–100 dBc/Hz at 20 kHz offset from carrier
1. To meet generator output path specifications, the input signal must be from the Test Set with the following characteristics: Frequency Range: 810 MHz to 995 MHz Level Range: –70 dBm to –7 dBm.
Agilent 83236B Specifications
Generator Output Path
Page 3
3
Analyzer Input Path
(RF IN/OUT connector)
Frequency
Frequency Range:
Through Path:
824 MHz to 849 MHz 869 MHz to 894 MHz
Conversion Path:
1710 MHz to 1785 MHz 1805 MHz to 1910 MHz 1930 MHz to 1990 MHz
Max Input Level: The maximum allowable average power depends on the unit under test as follows:
Subscriber Unit Test:
Single carrier TDMA and FM: 40 dBm (10 W) CDMA: 37 dBm (5 W)
Base Station Test:
Single carrier TDMA and FM: 40 dBm (10 W) CDMA and multi-carrier: 30 dBm (1 W)
Max Peak Instantaneous Signal: 30 V
Spectral Purity
Integrated Spurious and Phase Noise: <–57 dBc in a 100 Hz to 32 kHz bandwidth
Spurious Level: <-–60 dBc at 5 kHz and 20 MHz offset from
carrier
Power Measurement
Frequency Range:
824 MHz to 849 MHz 869 MHz to 894 MHz 1710 MHz to 1785 MHz 1805 MHz to 1910 MHz 1930 MHz to 1990 MHz
Measurement Range:
Subscriber Unit Test
Single carrier TDMA and FM: –13 dBm to 40 dBm (50 µW to 10 W) CDMA: –13 dBm to 37 dBm (50 µW to 5 W)
Base Station Test
Single carrier TDMA and FM: –13 dBm to 40 dBm (50 µW to 10 W) CDMA and multi-carrier: –13 dBm to 30 dBm (50 µW to 1 W)
Accuracy: ±5% of reading ±2.5 µW @ 23 °C ±10 °C after power meter zero and calibration ±10% of reading ±2.5 µW
Resolution: 0.01 dB or 10 µW
Analyzer Input Path
Page 4
4
Reference Specifications
(For proper operation, this instrument must be locked to an exter­nal 10 MHz reference.)
REF IN
Input Frequency: 10 MHz
Input Level Range: – 5 dBm to +10 dBm
REF OUT
Output Frequency: 10 MHz Output Level: –1 dBm
Accuracy: Buffered signal from REF IN, or 10 MHz ±10 ppm (if no external reference is connected to REF IN)
Remote Control
GPIB: Agilent Technologies’ implementation of IEEE Standard
488.2
Serial Port:
Connector type: D–SUB15(F) Interface: RS–232C
General Specifications
Isolation between "RF IN/OUT" and "RF OUT Only": >40 dB
Size: 84 H x 340 W x 500 D mm
Weight: 5.6 kg
Operating Temperature: 0 °C to 55 °C
Operating Humidity: 15 to 95% RH @ 40 °C
Operating Altitude: 0 to 2000 meters
Non-operating Temperature: –55 °C to 70 °C
Non-operating Humidity: To 90% RH @ 65 °C
Non-operating Altitude: 0 to 4572 meters
Power: AC 90 to 132 V, 198 V to 264 V, 47 to 63 Hz, 100 VA max
Others
Safety: Complies with IEC 1010-1:1990 + A1:1992/EN 61010-
1:1993 Certified by CSA-C22.2 No. 231-M89
EMC:
Radiated Emission: Complies with EN 55011:1991/CISPR
11:1990- Group 1, Class A
Flicker: Complies with EN 61000-3-3:1995/ IEC 1000-3-3:1994
Electro-Static Discharge: EN 50082-1:1992/IEC 801-2:1991
–4 kV CD, 8 kV AD
Radiated Immunity: EN 50082-1: 1992/IEC 801-3: 1984 –3 V/m
Note: When tested at 3 V/m according to IEC 801-3/1984, the output level accuracy will be within specifications over the full immunity test frequency range of 26 to 1000 MHz except when the analyzer frequency is identical to the transmitted interference signal test frequency.
Electrical Fast Transient/Burst: EN 50082-1:1992/IEC 801-4:1988
- 0.5 kV Signal Lines, 1kV Power Lines
Page 5
5
Connectors
1. FROM DUPLEX OUT: (Input from 892NX)
Input Impedance: 50 (nominal)
Connector Type: Type N (F) SWR: <1.3:1 Input Frequency Range: 810 MHz to 995 MHz Input Level Range: –70 dBm to –7 dBm
2. To ANT IN: (Output to 892NX) Frequency Range: 650 MHz to 940 MHz
Output Impedance: 50 (nominal)
Connector Type: Type N (F) SWR: <2.3:1
3. RF IN/OUT: (Input/Output to UUT)
Input/Output Impedance: 50 (nominal)
Connector Type: Type N (F) SWR: <1.2:1 Applied Power:
Subscriber Unit Test
Single carrier TDMA and FM: 40 dBm (10 W) CDMA: 37 dBm (5 W)
Base Station Test
Single carrier TDMA and FM: 40 dBm (10 W) CDMA and multi-carrier: 30 dBm (1 W)
Max Peak Instantaneous Signal: 30 V
4. RF OUT Only: (Output to UUT)
Output Impedance: 50 (nominal)
Connector Type: Type N (F) SWR: <1.6:1 Reverse Power Protection: 200 mW CW max
5. REF IN: (Input from 892NX)
Input Frequency: 10 MHz
Input Level Range: –5 dBm to +10 dBm
Input Impedance: 50 (nominal)
Connector Type: BNC (F)
6. REF OUT: (Output)
Output Frequency: 10 MHz Output Level: –1 dBm
Output Impedance: 50 (nominal)
Connector Type: BNC (F)
7. EXT TRIG IN: (Input for TDMA) External Trigger Signal: (Required for TDMA RF input level
range –13 dBm to –5 dBm)
Input Level: TTL
Input Impedance: >1 kat 1 MHz (nominal)
Triggering Type: Positive edge Connector Type: SMC (M)
8. DET OUT: (Output for CDMA)
Output Impedance: 50 (nominal)
Connector Type: SMC (M)
1 2 3 4
5 6 7 8
Page 6
Ordering Information
83236B PCS Interface
Options
AX4 Rack flange kit without handles
1AB Benchtop cabinet kit (for use with the Agilent 8924E)
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the produc­tion life of the product. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
“Our Promise” means your Agilent test and measurement equip­ment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifica­tions and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabili­ties, at no extra cost upon request. Many self-help tools are available.
Your Advantage
“Your Advantage” means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out­of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional services. Experienced Agilent engineers and tech­nicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
Get assistance with all your test and measurement needs at:
www.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
Copyright © 1996, 2000 Agilent Technologies Printed in U.S.A. 9/00 5965-5625E