Agilent 81619A User Manual

Agilent 81618A/9A Optical Head Interface Modules and Agilent 81623B/4B/6B/8B Optical Heads
User’s Guide
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2002 - 2005
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or tr ansl at ion int o a fo rei gn lan gu age ) w itho ut prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Agilent Technologies, Deutschland GmbH Herrenberger Straße 130 71034 Böblingen, Germany.
Manual Part Number
81623-90B12
Edition
Fourth edition April 2005
Third edition, July 2002
Second edition, January 2002
First edition, July 2001
Subject Matter
The material in this document is subject to change without notice.
Agilent Technologies makes no
warranty of any kind with regard to this printed material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Printing History
New editio ns are complete revisions of the guide reflecting alterations in the functionality of the instrument. Updates are occasionally made to the guide between editions. The date on the title page changes when an updated guide is published. To find out the current revision of the guide, or to
purchase an updated guide, contact your Agilent Technologies representative.
Warranty
This Agilent Technologies instrument produ ct i s wa rra nte d ag ain st d efe cts in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent will, at its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilent. Buyer shall prepay shipping charges to Agilent and Agilent shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent from another country.
Agilent warrants that its software and firmware designated by Agilent for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent does not warrant that the operation of the instrument, software, or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specifically disclaims the implied warranties of Merchantability and Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance contact your nearest Agilent Technologies Sales and Service Office.
Certification
Agilent Technologies Inc. certifies that this product met its published specifications at the time of shipment from the factory.
Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, NIST (formerly the United States National Bureau of Standards, NBS) to the extent allowe d by the Institutes’s calibration facility, and to the calibration facilities of other International Standards Organization members.
ISO 9001 Certification
Produced to ISO 9001 international quality system standard as part of our objective of continually increasing customer satisfaction through improved process control.
Agilent Technologies Sales and Service Offices
For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, viesit our web site:
http://www.agilent.com/comms/lightwave
You can also contact one of the following centers and ask for a test and measurement sales representative.
United States: 1 800 829 4444
1 800 829 4433 (FAX)
Canada: 1 877 894 4414
1 800 746 4866 (FAX)
Europe: +31 20 547 2111
+31 20 547 2190 (FAX)
Japan: 120 421 345
120 421 678 (FAX)
Mexico (52 55) 5081 9469
(52 55) 5081 9467 (FAX)
Australia: 800 629 485
800 142 134 (FAX)
Asia-Pacific: +852 800 930 871
+852 800 930 871 (FAX)
Latin America +55 11 4197 3600
+55 11 4197 3800 (FAX)
Ta b l e o f Co ntents
Getting Started with Optical Heads 9
Safety Considerations 10
Safety Symbols 10
Input Power Limitations 11
Initial Inspection 11 Line Power Requirements 11 Operating Environment 12 Storage and Shipment 12
What is an Optical Head? 13
Analog Output 15 Optical input 15 Heat Sink for 81628B optical Head 16 Attaching the heat sink to the 81628B Optical Head 16
Applicable adapters 16
Mounting Instructions 17
Accessories 19
Modules and Options 20
Connector Adapters - Reference List 22
Specifications 23
Definition of Terms 24
Linearity 24 Averaging Time 25 Linewidth 25 Noise 25 Reference conditions 25 Power range 26 Wavelength range 26 Relative uncertainty (spectral ripple) due to interference 26 Relative uncertainty due to speckle noise 27 Relative uncertainty due to polarization 27 Return loss 28 Spectral width of optical source 28
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
5
Total uncertainty 28 Uncertainty at reference conditions 28
Optical Head Specifications 29 Supplementary Performance Characteristics 35
Analog Output 35
Performance Tests 37
Equipment Required 38 Te st R e c ord 40 Test Failure 41 Instrument Specification 42 Functional Tests 43 Performance Tests 44
For 81628B only 45 Accuracy Test 46 Linearity Test 48
Test S e t u p 49 +10 dBm Range (equivalent to +40 dBm range for 81628B) 50
Calculation 54
Noise Test 57 Return Loss Test 58 Relative Uncertainty due to Polarization (Optional Test) 60 Relative Uncertainty due to Interference (Optional Test) 62 Calculation Sheets 78
Cleaning Information 81
Cleaning Instructions for this Device 82
81000xA Optical Head Adapters 82 8162xx Optical Power Heads 82
Safety Precautions 83 Why is it important to clean optical devices? 84 What do I need for proper cleaning? 85
Standard Cleaning Equipment 85
Dust and shutter caps 85
Isopropyl alcohol 86 Cotton swabs 86 Soft tissues 87
6 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Pipe cleaner 87
Compressed air 88
Additional Cleaning Equipment 89
Microscope with a magnification range about 50X up to 300X 89 Ultrasonic bath 89 Warm water and liquid soap 90
Premoistened cleaning wipes 90 Polymer film 90 Infrared Sensor Card 90
Preserving Connectors 91 Cleaning Instrument Housings 92 Which Cleaning Procedure should I use ? 93
How to clean connectors 94 How to clean connector adapters 96 How to clean connector interfaces 97 How to clean bare fiber adapters 98 How to clean lenses 99 How to clean instruments with a fixed connector interface 100 How to clean instruments with an optical glass plate 100 How to clean instruments with a physical contact interface 101 How to clean instruments with a recessed lens interface 102 How to clean optical devices which are sensitive to mechanical
stress and pressure 103 How to clean metal filters or attenuator gratings 104
Additional Cleaning Information 105
How to clean bare fiber ends 105 How to clean large area lenses and mirrors 106
Other Cleaning Hints 108
Index 109
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
7
8 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Getting Started with Optical Heads
1
Getting Started with Optical Heads
This chapter introduces the features of the Agilent 81623B/4B/6B/8B Optical Heads.
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Input Power Limitations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Line Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Storage and Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
What is an Optical Head? . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Optical input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Heat Sink for 81628B optical Head . . . . . . . . . . . . . . . . . . . . . . . 16
Attaching the heat sink to the 81628B Optical Head . . . . . . . .16
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
9
Getting Started with Optical Heads Safety Considerations
Safety Considerations
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no liability for the customer’s failure to comply with these requirements.
Before operation, review the instrument and manual, including the red safety page, for safety markings and instructions. You must follow these to ensure safe operation and to maintain the instrument in safe condition.
CAUTIONWARNING
CAUTION
The WARNING sign denotes a hazard. It calls attention to a procedure, practice or the like, which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Safety Symbols
The apparatus will be marked with this symbol when it is necessary for the user to refer to the instruction manual in order to protect the apparatus against damage.
Caution - hot surface.
Magnetic fields - may interfere with a pacemaker.
10 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Safety Considerations Getting Started with Optical Heads
Input Power Limitations
CAUTIONWARNING
Applicable to Agilent 81628B only -
Operation above 34 dBm (2.51 W) is at your own risk.
Operation at a continuous optical power, or an average optical power, greater than 34 dBm causes, in the specified operating temperature range, the metal parts (connector adapters, screws) to become hotter than the required limit (55°C) of the safety standard IEC 61010-1.
Agilent Technologies Deutschland GmbH will not be liable for any damage caused by operation above 34 dBm.
Initial Inspection
Inspect the shipping container for damage. If there is damage to the container or cushioning, keep them until you have checked the contents of the shipment for completeness and verified the instrument both mechanically and electrically.
The Performance Tests give procedures for checking the operation of the instrument. If the contents are incomplete, mechanical damage or defect is apparent, or if an instrument does not pass the operator’s checks, notify the nearest Agilent Technologies Sales/Service Office.
CAUTIONWARNING
To avoid hazardous electrical shock, do not perform electrical tests when there are signs of shipping damage to any portion of the outer enclosure (covers, panels, etc.).
Line Power Requirements
The Agilent 81618A and Agilent 81619A Optical Head Interface Modules operate when installed in the Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, and Agilent 8166A Lightwave Multichannel System.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
11
Getting Started with Optical Heads Safety Considerations
Operating Environment
The safety information in the Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, and Agilent 8166A Lightwave Multichannel System User’s Guide summarizes the operating ranges for the Agilent 81618A and Agilent 81619A Optical Head Interface Modules. In order for these modules to meet specifications, the operating environment must be within the limits specified for the Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, and Agilent 8166A Lightwave Multichannel System.
Storage and Shipment
This module can be stored or shipped at temperatures between
40°C and +70°C. Prote ct the module from tem perature extremes that ma y cause condensation within it.
12 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head? Getting Started with Optical Heads
What is an Optical Head?
An optical head measures the power emitted from a connected single­mode or multi-mode fiber or the power applied in an open parallel beam (with max. 5 mm diameter). The wavelength and power range depends on the sensor element.
Screw to connector adapter when fitted
Cable to interface module
Analog output
Figure 1 Rear view of an Optical Head with Analog Output
Optical interface
connector adapter locking screws
Figure 2 Front View of Optical Head without Adapter
Figure 1 and Figure 2 show views of a typical optical head.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
13
Getting Started with Optical Heads What is an Optical Head?
Figure 3 shows two types of adapter that are available for connecting the input fiber to an optical head
.
Threaded connector adaptor:
81000xA, 81000BT
Figure 3 Connector adapters
D-type magnetic connector adaptor:
(preferred) 81003xx 81001xx
81000BC
14 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head? Getting Started with Optical Heads
Analog Output
The analog output is the BNC connector on the back of the optical head. It outputs a voltage directly proportional to the strength of the optical signal at the optical input in the current range. The analog signal is always in the range between 0 and 2V, 2V corresponding to a full power input signal in the current range, 0V corresponding to no input signal.
During autoranging, the level to which 2V corresponds changes. You should disable autoranging when using the analog output.
Disabling Autoranging
Select the best range for the application using the mainframes user interface. Select <Menu>, <Range Mode>, <Manual> and then choose from the list of ranges.
Applications
Possible applications for the analog output are:
• to close the feedback loop controlling the current supplied to a laser
• to aid positioning the system for fiber alignment tasks
• to monitor optical power on an oscilloscope
The analog signal reacts instantaneously to the input signal, whereas the power shown on the display is subject to averaging.
Optical input
The optical input to the optical head requires a connector adapter to match the connector type (or bare fiber) to the optical head interface.
The optical head interface can be either threaded or a D-type magnetic interface depending on the model of optical head (81628B is only available with a threaded interface). See “Accessories” on page 19 for details of the connector adapters available.
When a connector adapter is fitted it should be locked in place by tightening the locking screws located either side of the main optical head housing.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
15
Getting Started with Optical Heads What is an Optical Head?
Heat Sink for 81628B optical Head
The 81628B optical head has a heat sink that allows an extended operating power range. This heat sink must be mounted on the integrating sphere for operation from 34 dBm to 38 dBm.
Attaching the heat sink to the 81628B Optical Head
NOTE
NOTE
Always attach the heat sink and rubber ring to the 81628B if operating at an optical power higher than 34dBm!
Applicable adapters
Agilent adapters types 81000xx, 81000BT are applicable to the 81628B with the heat sink attached (without the heatsink and up to 34dBm the 81000BT bare fiber adapter may be attached). Bare fibers are not intended to be used in high power applications.
If shipping the 81628B high power optical head please detach the heatsink from the optical head to avoid damage during transportation.
16 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head? Getting Started with Optical Heads
Mounting Instructions
1 Check that the blue rubber ring is correctly attached to the integrating
sphere and covers the metallic screws as shown in Figure 4 .
Figure 4 Attaching the Rubber Ring
2
The heatsink consists of two conical metal parts, the bottom part (smallest diameter at the bottom), and the top part (largest diameter at top). Slide the bottom part of the heat sink over the throat of the integrating sphere with the smallest diameter facing to the sphere. Then screw on the desired adapter (81000xx series) to the sphere as shown in Figure 5 .
Figure 5 Attaching the Bottom Part of the Heat Sink
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
17
Getting Started with Optical Heads What is an Optical Head?
Attach the upper part of the heat sink to the integrating sphere. Slide the
3
upper part over the connector adapter with the largest diameter facing opposite to the sphere. Tighten the screws with the Allen key enclosed in the Heat Sink Kit (see Figure 6 ).
Figure 6 Attaching the Upper Part of the Heat Sink
18 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
2
Accessories
The Agilent 81618A/9A Optical Head Interface Modules and Agilent 81623B/4B/6B/8B Optical Heads are available in various configurations for the best possible match to the most common applications.
This chapter provides information on the available options and accessories.
Modules and Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Optical Heads (5mm Sensors) . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Connector Adapters - Reference List. . . . . . . . . . . . . . . . . . . . . .22
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
19
Accessories Modules and Options
Modules and Options
Optical Heads (5mm Sensors)
Accessories
81624CE 4 m extension cable
Interface Module
High Power
Optical Head for:
Connectorized Fiber,
Bare Fiber and
Open beam NA 0.3
81628B InGaAs
+40 dBm to –60 dBm
Connector Adapters
(threaded).
Interface Modules
81618A Single Head Interface Module
81619A Dual Head Interface Module
Connector Adapters (threaded)
81000BT Bare Fiber Connectivity Set (Threaded) 81000FA FC/PC/APC 81000KA SC/PC/APC 81000VA ST 81003LA LC 81000PA E-2000
20 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Modules and Options Accessories
High Power Optical Head (with integrating sphere)
High Power Optical Head (with integrating sphere)
Interface M odule
High Power
Optical Head for:
Connectoriz ed Fiber,
Accessories
81624CE 4 m extension cab le
Bare Fiber and
Open beam NA 0.3
81628B InGaAs
+40 dBm to –60 dBm
Connector Ad apters
(threaded).
OPTIONAL
Interface Modules
81618A Single Head Inte rface Module
81619A Dual Head Interfac e Module
Connector Adapters (threaded)
81000BT Bare Fibe r Connectivity Set (Threaded) 81000FA FC/PC/ APC 81000KA SC/PC/APC 81000VA ST 81003LA LC 81000PA E-2000
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
21
Accessories Modules and Options
Connector Adapters - Reference List
Table 1 Connector Adapters
Preferred Adapters
Integral D-shape for 5 mm
Connector Type Threaded Version
Barefiber 81000BT 81000BC
FC/PC 81000FA 81001FA
SC/PC 81000KA 81001KA
MU 81001MA
E-2000 81000PA 81001PA
ST 81000VA
blank 81001ZA
LC 810003LA 810001LA
MTP(ribbon) 81003TD
optical heads
22 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
3
Specifications
The Agilent 81618A/9A Optical Head Interface Modules and Agilent 81623B/4B/6B/8B Optical Heads are produced to the ISO 9001 international quality system standard as part of Agilent Technologies’ commitment to continually increasing customer satisfaction through improved quality control.
Specifications describe the modules’ and heads’ warranted performance. Supplementary performance characteristics describe the modules’ and heads’ non-warranted typical performance.
Because of the modular nature of the instrument, these performance specifications apply to these modules and heads rather than the mainframe unit.
Definition of Terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Linearity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Averaging Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Linewidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Reference conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Power range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Wavelength range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Relative uncertainty (spectral ripple) due to interference . . . . 26
Relative uncertainty due to speckle noise . . . . . . . . . . . . . . . . .27
Relative uncertainty due to polarization . . . . . . . . . . . . . . . . . . .27
Return loss. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Spectral width of optical source. . . . . . . . . . . . . . . . . . . . . . . . . .28
Total uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Uncertainty at reference conditions . . . . . . . . . . . . . . . . . . . . . . 28
Optical Head Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 29
Supplementary Performance Characteristics . . . . . . . . . .35
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
23
Specifications Definition of Terms
Definition of Terms
This section defines terms that are used both in this chapter and “Performance Tests” on page 37.
Generally, all specifications apply for the given environmental conditions and after warmup time.
Measurement principles are indicated. Alternative measurement principles of equal value are also acceptable.
Linearity
The linearity error is defined as the relative difference between the displayed power ratio, D caused by changing the displayed power level from the reference level, D to an arbitrary displayed level, D
if expressed in %
, and the actual (true) power ratio Px/P0
x/D0
. Symbol N.
x
,
0
D
xD0

N
------ ------- ----- 1Ð

P
xP0
100()=
if expressed in dB
D
xD0
N
dB

10
log=
------- ------ -----

P
xP0
Conditions: reference level 10 µW, displayed power levels within the
specified range, zero less than specified time prior to measurement.
Note 1: ideally N = 0 %, respectively 0 dB.
Note 2: the power-dependent nonlinearity, N(P
), can alternatively be
x
expressed by the following formula:
rP
()rP0()Ð
()
NP
x
x
------ ------ ---- ----- ------ ---- -----=
()
rP
0
where r(P) is the power-dependent responsivity (for a power meter, the responsivity is defined as the ratio of displayed power to actual input power).
24 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Definition of Terms Specifications
Averaging Time
Time defining the period during which the power meter takes readings for averaging. At the end of the averaging time the average of the readings is available (display- or memory-update). Symbol T
avg
.
Linewidth
FWHM spectral bandwidth. The 3 dB width of the optical spectrum, expressed in Hertz. Symbol: f.
Noise
The peak-to-peak change of displayed power level with zero input power level (dark).
Conditions: Zero prior to measurement, averaging time and observation time as specified, lowest power range selected and wavelength range as specified.
Measurement: the measurement result is obtained by:
P
Noise = P
max
expressed as peak-to-peak within
min
the given time span. Any offset is automatically excluded this way.
Reference conditions
The specified conditions during the spectral responsivity calibration, or conditions which are extrapolated from the conditions during calibration.
Conditions: power level, beam diameter or fiber type, numerical aperture, wavelength, spectral width, ambient temperature as specified, at the day of calibration. Noise and drift observed over a specified observation time, with a temperature change of not more than ±T.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
25
Specifications Definition of Terms
Power range
The power range is defined from the highest specified input power level to the smallest input power level that causes a noticeable change of displayed power level.
Conditions: wavelength, averaging time as specified.
Wavelength range
The range of wavelengths for which the power meter is calibrated.
Note: Selectable wavelength setting of the power meter for useful power measurements (operating wavelength range).
Literature
[1]Fiber optic test and measurement, Hewlett Packard Professional Books, edited by Prentice Hall, ISBN 0-13-534330-5
Relative uncertainty (spectral ripple) due to interference
Uncertainty of power reading when using a coherent source, due to a periodic change of the power meter's responsivity caused by optical interference between reflective interfaces within the power meter's optical assembly.
Conditions: constant wavelength, constant power level, angled connector as specified, linewidth of source <100 MHz, temperature as specified.
NOTE
26 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Spectral ripple is measured by stepping the source wavelength over the wavelength range specified.
Definition of Terms Specifications
Relative uncertainty due to speckle noise
This is the uncertainty of the power reading when using a coherent source. This is due to a variation of the power meters responsivity caused by optical interference meters optical assembly.
Conditions: constant wavelength, constant power level, angled connector as specified, linewidth of source as specified, temperature as specified.
1
Incoming light to the integrating sphere undergoes numerous internal reflections (as many as several hundred) prior to hitting the detector. If the source is sufficiently coherent, a complicated interference pattern (speckle pattern) appears spread over the whole sphere volume. Slight changes of the ambient conditions (temperature, vibration, shock) affect particular optical paths which consequently changes the speckle pattern and results in an instability of the head power reading (speckle noise).
1
between different optical paths within the power
NOTE
Changes in the source wavelength have a similar effect on speckle noise as environmental changes, because they cause fluctuations in the interference pattern. If the source coherence length is small compared to the effective path length, the interference pattern disappears and relative uncertainty due to speckle noise becomes negligible.
Measurement: In contrast to the spectral ripple definition the source wavelength isn’t stepped when measuring speckle noise.
Relative uncertainty due to polarization
Also termed polarization-dependent responsivity (PDR), the relative uncertainty due to polarization is the uncertainty of the displayed power level on the input polarization state, expressed as the difference between the highest and the lowest displayed power. Uncertainty figures are based upon a 95% confidence level.
Conditions: laser source with variable polarization state, generation of all possible polarization states (covering the entire Poincaré sphere), constant wavelength, constant power level, angled connector as specified, temperature as specified.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
27
Specifications Definition of Terms
Return loss
The ratio of the incident power to the reflected power expressed in dB. Symbol: RL.
P
in

RL 10
Conditions: the return loss excludes any reflections from the fiber end
used as radiation source.
log=
------ ------ --

P
back
Spectral width of optical source
Full width at half maximum. The 3 dB width of the optical spectrum, expressed in nm. Symbol: FWHM.
To ta l u nc e rt ai nt y
The uncertainty for a specified set of operating conditions, including noise and drift.
Conditions: power level, beam diameter or fiber type, numerical aperture, wavelength, spectral width, ambient temperature, re-calibration period as specified. →Noise and drift observed over a specified observation time, with a temperature change of not more than ±T.
Uncertainty at reference conditions
The uncertainty for the specified set of reference conditions, including all uncertainties in the calibration chain from the national laboratory to the test meter.
28 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head Specifications Specifications
Optical Head Specifications
All optical heads have to be operated with the single (Agilent 81618A) or dual
(Agilent 81619A) Interface Modules.
Table 2 Optical Head Specifications
Agilent 81623B
Agilent 81623B
Sensor element Ge, 5 mm
Wavelength range 750 nm to 1800 nm
Power range
Applicable fiber type
Open beam
Uncertainty at refer­ence conditions [1]
Total uncertainty [2], [9]
Relative uncertainty:
- due to polarization [3]
- spectral ripple
± 2.2 % (1000 nm to 1650 nm)
± 3.0 % (800 nm to 1000 nm)
± 3.5% ± 100 pW (1000 nm to 1650 nm)
± 4.0% ± 250 pW (800 nm to 1000 nm)
Standard SM and MM max 100 µm core size, NA 0.3
Calibration option C85 / C86
80 dBm to +10 dBm
Parallel beam max 4 mm
± 2.2 % (1000 nm to 1650 nm)
± 2.5 % (800 nm to 1000 nm)
± 3.5% ± 100 pW (1000 nm to 1650 nm)
± 3.5% ± 250 pW (800 nm to 1000 nm)
[10]
< ± 0.01 dB
< ± 0.006 dB (typ. < ± 0.003 dB)
(typ. < ± 0.005 dB)
Agilent 81623B
Calibration option C01 / C02
± 1.7 % (1000 nm to 1650 nm)
± 3.0 % (800 nm to 1000 nm)
± 3.0% ± 100 pW (1000 nm to 1650 nm)
± 4.0% ± 250 pW (800 nm to 1000 nm)
(due to interference) [4]
Linearity (power): [5]
at 23°C ±5°C
at operating temp. range
Return loss [7] > 50 dB, typ. > 55 dB
Noise (peak to peak) [6]
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
(CW –60 dBm to +10 dBm)
< ± 0.025 dB
[8]
< 100 pW (1200 nm to 1630 nm)
< 400 pW (800 nm to 1200 nm)
[9]
< ± 0.05 dB
> 56 dB
29
Specifications Optical Head Specifications
Averaging time (mini-
100 µs
mal)
Analog Output included
Dimensions 57 mm x 66 mm x 156 mm
Weight 0.5 kg
Recommended recali-
2 years
bration period
Operating temperature 0°C to 40°C
Humidity Non-condensing
Warm-up time 40 min
[1]
.
Reference conditions:
• Power level 10 µW ( -20 dBm), continuous wave (CW)
• Parallel beam, 3 mm spot diameter on the center of the detector
• Ambient temperature 23°C ± 5°C
• On day of calibration (add ± 0.3% for aging over one year, add ± 0.6% over two years)
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[2]
Operating Conditions:
• P aral lel beam , 3m m sp ot d iame te r on the c ent er o f th e de tect or o r co nnec tor ized fib er w ith NA 0 .2 ( stra igh t
connector)
•Averaging time 1s
• For NA > 0.2: add 1%
• Within one year after calibration, add 0.3 % for second year
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[3]
All states of polarization at constant wavelength (1550 nm ± 30 nm) and straight connector, T = 23°C ± 5°C.
For angled connector (8 û ) add 0.01 dB typ.
[4] Conditions:
• Wavelength 1550 nm ± 30 nm, fixed state of polarization
• Temperature 23°C± 5°C
•Linewidth of source ÷ 100 M Hz
• Angled connector 8°
[5]
Does not include noise; for wavelength < 1000 nm applies for Ð 50dBm to + 10 dBm
[6]
Averaging time 1 s, T = 23°C ± 5°C, T± 1°C, observa tion tim e 300 s.
[7]
Conditions:
Wavelengths 1550 nm ± 30 nm. Standard single mode fiber, angled connector min 8°
[8]
With D-shape adapter 81001x x
Return Loss: > 60 dB typical
[9]
For input power > 2 mW
add ± 0.004 dB / mW; zeroing required
[10]
Specif ication vali d for opti cal heads with
S/N starting with “DE413…” (shipping began April 1, 2001)
30 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head Specifications Specifications
Table 3 High Power Optical Head Specifications
Sensor element
Wavelength range
Power range
Applicable fiber type
Open beam
Uncertainty at refer
ence conditions
Total uncertainty
[2] ± 3.5% ± 5 pW
[1]
Agilent 81624B
Calibration option
Agilent 81624B
C01 / C02
InGaAs, 5 mm InGaAs, 5mm
800 nm to 1700 nm 850 nm to 1650nm
90 dBm to +10 dBm 70 to +27 dBm (1250 nm to 1650 nm)
Standard SM and MM max 100 µm core size,
NA 0.3
Parallel beam max 4 mm
± 2.2 %
(1000nm to 1630 nm)
± 1.5 %
(970 nm to 1630 nm)
± 2.8% ± 5 pW
(1000nm to 1630 nm)
(970 nm to 1630 nm)
Agilent 81626B
Calibration option
Agilent 81626B
C01 / C02
70 to +23 dBm (850 nm to 1650 nm)
Standard SM and MM max 100 µm core size,
NA 0.3
Parallel beam max 4 mm
± 3.0 %
(950 nm to 1630 nm)
± 5.0% ± 500 pW
[8]
(950 nm to 1630 nm)
± 2.5 %
(950 nm to 1630 nm)
± 4.5% ± 500 pW
(950 to 1630 nm max
23 dBm)
(1250 to 1630 nm
max 27 dBm)
[8]
Relative uncertainty:
- due to polarization
[3]
- spectral ripple (due to interference)
[4]
Linearity (power):
- at 23 ûC ±5ûC
- at operat. temp. range
Return loss
Noise (peak to peak)
Averaging time (min.)
Analog Output
Dimensions
Weight
Recommended recali
bration period
[5]
[7]
[7]
± 0.005 dB (typ. ± 0.002 dB)
± 0.005 dB (typ < ± 0.002 dB)
CW –70 dBm to +10 dBm, 1000 nm to 1630 nm
< ± 0.02 dB
< ± 0.05 dB
CW – 50 dBm to + 27dBm, 950 nm to 1630 nm
± 0.005 dB (typ. ± 0.002 dB)
± 0.005 dB (typ. < ± 0.002 dB)
< ± 0.04 dB
< ± 0.15 dB
typ. 60 dB > 45 dB > 47 dB
[6] < 5 pW < 500 pW
100 µs 100 µs
included Included
57 mm x 66 mm x 156 mm 57 mm x 66 mm x 156 mm
0.5 kg 0.5 kg
2 years 2 years
[8]
[8]
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
31
Specifications Optical Head Specifications
Operating temperature
Humidity
Warm-up time
0°C to 40°C 0°C to +35°C
[9]
Non-condensing Non-condensing
40 min 40 min
[1]
Reference conditions:
• Power level 10 µW (-20 dBm), continuous wave (CW)
• Parallel beam, 3 mm spot diameter on the center of the detector
• Ambient temperature 23°C ± 5°C
• On day of calibration (add ± 0.3% foraging over one year, add ± 0.6% over two years)
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[2]
Operating Conditions:
• Parallel beam, 3mm spot di ameter on the center of the detector or connectorized fiber with NA 0.2 (straight connector)
•Averaging time 1s
• For NA > 0.2: add 1%.
• Within one year after calibration, add 0.3 % for second year.
•Zeroing required
[3]
All state s of polar ization at constant wavelength (155 0 nm ± 30 nm),
straight connector, T = 23°C ± 5°C. For angled connector (8 û ) add 0.01 dB typ.
[4]
Conditions:
• Wavelength 1550 nm ± 30 nm, fixed state of pol arization
• Temperature 23°C ± 5°C
•Linewidth of source ≥100 MHz
• Angled connector 8°
[5]
Does not include noise;
zeroing required
[6]
Averagin g time 1s, T = 23°C ± 5°C, T ± 1°C, observation time 300 s.
Wavelength range 1200 nm to 1630nm
[7]
Conditions:
• Wavelengths 1550nm ± 30 nm
• Standard single mode fiber, angled connector min 8°
• With D-shape adapter 81001xx Return Loss > 60 dB typica l
[8]
For input power > + 10 mW add: add typ. ± 0.0016 dB/mW In case of negative power change > 50 dB allow additional recovery time of 3 min
[9]
Max 30°C for > + 20 dBm input power
32 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head Specifications Specifications
Table 4 Agilent 81628B Specification
Agilent 81628B
with integrating sphere
Sensor element InGaAs
Wavelength range 800 nm to 1700 nm
Power range
60 dBm to +40 dBm (800 nm to 1700 nm)
For operation higher than 34 dBm see safety note
Damage Power 40.5 dBm
Applicable fiber type
Open beam
Uncertainty at reference conditions
Total uncertainty
[1] [8]
[2] [8]
10 dBm
>10 dBm to 20 dBm
>20 dBm to 38 dBm
Single mode NA 0.2, Multimode NA 0.4 ∅ ≤ 3mm center of sphere
± 3.0 % (970 nm to 1630nm)
(970 nm to 1630nm)
± 4.0% ± 5 nW
± 4.5%
± 5%
Relative uncertainty:
[3]
- due to polarization
typ. ± 0.006 dB
- due to speckle noise at
source linewidth:
[4]
typ. ± 0.02 dB
0.1pm to 100pm >100pm
Linearity (power):
[5] [8]
10 dBm
>10 dBm to 20 dBm
>20 dBm to 37 dBm
>37 dBm to 38 dBm
Return loss
Noise (peak to peak)
[6]
Averaging time (minimal) 100 µs
Analog Output Included
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
typ. ± 0.002 dB
(CW –40 dBm to + 38 dBm), (970 nm to 1630 nm)
± 0.03 dB
± 0.06 dB
± 0.09 dB
± 0.10 dB
at 23°C ± 5°C, for operating temperature range add ±0.03 dB
typ. > 75 dB
< 5 nW
33
Specifications Optical Head Specifications
Dimensions 55mm x 80 mm x 250 mm
Weight 0.9 kg (without heat sink)
Recommended Recalibra-
2 years
tion period
Operating temperature
[7]
0°C to +40°C
Humidity Non-condensing
Warm-up time 40 min
Reference conditi ons:
[1]
Power lev el 10 µW (-20 dBm), continuous wave (CW)
•Averaging time 1s
• Parallel beam, 3 mm, center of sphere input
• Ambient tempe rature 23 °C ± 5 °C
• On day of calibration (add ± 0.3 % for aging over one year, add ± 0.6 % over two years)
• Spectral width of source <10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4 nm
• Humidity 50 % ± 10 %
[2]
Operating Conditions:
• Parallel beam, 3mm , center of sphere input, or connectorized fiber with NA 0.2 (straight connector)
• For NA > 0.2: add 1%
• Within one year after calibration, add ± 0.3% for second year
• Operating temperature range as specified, humidity <80% and non-condensing
•Zeroing required
[3]
All stat es of polar ization at constant wavelength (1550 nm ± 30 nm) and co nstant power
[4]
Conditions:
• Wavelength 1550 nm ± 30 nm, fixed state of polarization, constant power
• Temperature 23°C ± 5°C Measurement time ≤ 3 min
[5]
Does not include noise;
zeroing required
[6]
Averagin g time 1s, T = 23°C ± 5 °C,
T ± 1°C, observa tion tim e 300 s, wavelength ran ge 970 nm to 1630 nm
Thermal drift at 38 dBm, exposure time 30 min: Recovery time 10 min: 30nW 30 m in: 10nW
[7]
For optical power > 30 dBm the maximal opera ting temperature is limited to 35°C
[8]
Wavelength must not be equal to any water absorption line
Safety Note:
For optical power higher than 34 dBm the attached heat sink MUST be used! For continuous optical power or average optical power higher than 38 dBm the connector adapters will get warmer than permitted according to the safety standard IEC 61010-1. The 81628B Optical Head can handle optical power up to 40 dBm, however, operation above 38 dBm is at the operators own risk. Agilent Technologies Deutschland GmbH will not be liable for any damages caused by an operation above 38 dBm.
34 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Supplementary Performance Characteristics Specifications
Supplementary Performance Characteristics
Analog Output
Bandwith DC, 300 to 5000 Hz depending on range and optical head. Output voltage 0 to 2 V non-terminated Output impedance 600 Ohm typ. Max. input voltage ±10V
Table 5 3dB-bandwidth of the Analog Output
Bandwidth
Range
+40 dBm N/A N/A 3.5 kHz
+30 dBm N/A N/A 3.5 kHz
+20 dBm N/A N/A 3.5 kHz
+10dBm 5.0kHz 5.0kHz 3.5kHz
0dBm 5.0kHz 5.0kHz 1.8kHz
-10dBm 5.0kHz 5.0kHz 1.8kHz
-20dBm 5.0kHz 5.0kHz 0.12kHz
-30dBm 0.5kHz 1.8kHz 0.12kHz
-40dBm 0.5kHz
-50 dBm 0.05 kHz
-60 dBm 0.05 kHz
[1]
Not applicable for 81626B
[2]
Not applicable for 81622B
81622B/3B
[2]
[2]
[2]
Bandwidth
81624B/6B/7B
1.8 kHz N/A
0.12 kHz
0.12 kHz
[1]
[1]
Bandwidth
81628B
N/A
N/A
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
35
Specifications Supplementary Performance Characteristics
36 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
4
Performance Tests
The procedures in this section test the performance of the instrument. The complete specifications to which Agilent 81623B/4B/6B/8B Optical Heads are tested are given in “Specifications” on page 23. All tests can be performed without access to the interior of the instrument. The test equipment given corresponds to tests carried out with Diamond HMS - 10 connectors.
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Tes t R e c ord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
Test Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Instrument Specification. . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
For 81628B only. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Accuracy Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Linearity Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Noise Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Return Loss Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Relative Uncertainty due to Polarization (Optional Test) . . . . . 60
Relative Uncertainty due to Interference (Optional Test) . . . . . 62
Calculation Sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
37
Performance Tests Equipment Required
Equipment Required
Equipment required for the performance test is listed in the table below. Any equipment that satisfies the critical specifications of the equipment given in the table may be substituted for the recommended models.
Table 6 Equipment Required for Performance Test
Interface Optical Head
Instrument/Accessory Recommended Model
81618A
81619A
81623B
81624B
81626B
81628B
Multimeter Mainframe Agilent 8163A/B 1 ea 1 ea 2 ea 2 ea 2 ea 2 ea Agilent
Interface Module Agilent 81618A ––xxxx Agilent
Optical Head 8162x 1 ea 2 ea – CW Laser Module Agilent 81657A ––xxxx CW Laser Module Agilent 81654A xxxx Agilent
Power Meter Standard Agilent 81618A Optical
Head Interface Module with Agilent 81623B #C01
Power Meter Standard Agilent 81618A Optical
Head Interface Module with
Agilent 81624B #C01 Power Sensor ModuleAgilent 81634B ––xxxx Optical Attenuator Agilent 8156A #221 ––xxxx Optical Attenuator Agilent 8156A #101 ––xxxx Return Loss
Reference Cable Return Loss ModuleAgilent 81613A ––xxxx Kit, Performance Test 81628-68705 x Tunable Laser Source Agilent 8164A and
Polarization Controller Agilent 11896A #025 o o o Wavelength Indepen­dent Coupler (3 dB) Best IF Adapter Special Tool o o o Worst IF Adapter Special Tool o o o
Agilent 81610CC ––xxxx
Agilent 81680A #022
Special Tool o o o
––x–––
–––xxx
–––ooo
Alterna-
tive
Models
Required
Characteristics
8164A/B
81619A
81657A
Singlemode Fiber Agilent 81101AC (2/1* ea)––xxxx
Legend : – not applicable; x necessary; o optional.
38 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Equipment Required Performance Tests
Table 6 Equipment Required for Performance Test
Interface Optical Head
Instrument/Accessory Recommended Model
81618A
81619A
81623B
81624B
81626B
81628B
Agilent 81101PC (1 ea) –––––– Agilent 81102SC (1 ea) x Agilent 81113PC (3 ea) ––xxxx Agilent 81113SC (1 ea) ––xxxx
Connector Adapters Agilent 81000AA (2 ea) ––xxxx
Agilent 81000SA (1 ea) ––xxxx Agilent 81001SA (1 ea) –––––– Agilent 81000FA (1/2* ea)––xxxx Agilent 81001FA (1 ea) ––––––
Connector InterfacesAgilent 81000AI (3 ea) ––xxxx
Agilent 81000FI (3 ea) ––xxxx Agilent 81000SI (4 ea) ––xxxx
Plastic Cap PN 5040-9351 ––xxxx
Legend : – not applicable; x necessary; o optional.
Alterna-
tive
Models
Required
Characteristics
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
39
Performance Tests Tes t R ec o r d
Test Record
Results of the performance test may be tabulated on the Test Record provided at the end of the test procedures. It is recommended that you fill out the Test Record and refer to it while doing the test. Since the test limits and setup information are printed on the Test Record for easy reference, the record can also be used as an abbreviated test procedure (if you are already familiar with the test procedures). The Test Record can also be used as a permanent record and may be reproduced without written permission from Agilent Technologies.
40 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Te s t F ai lu r e Performance Tests
Test Failure
If the Agilent 81618A/19A Interface module or Agilent 81623B/4B/6B/8B Optical Head fails any performance test, return the instrument to the nearest Agilent Technologies Sales / Service Office for repair.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
41
Performance Tests Instrument Specification
Instrument Specification
Specifications are the performance characteristics of the instrument that is certified. These specifications, listed in “Specifications” on page 23, are the performance standards or limits against which the Agilent 81623B/4B/6B/8B Optical Head can be tested. “Specifications” on page 23 also lists some supplemental characteristics of the Agilent 81623B/4B/6B/8B Optical Head. Supplemental characteristics should be considered as additional information.
Agilent 81618A/19A Interface Modules are tested for functionality only.
Any changes in the specifications due to manufacturing changes, design, or traceability to the National Institute of Standards and Technology (NIST), will be covered in a manual change supplement, or revised manual. Such specifications supercede any that were previously published.
42 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Functional Tests Performance Tests
Functional Tests
The functional test applies to the Agilent 81618A/19A Interface modules.
Mainframe 8163A/B with
8161xA Interface Module as DUT
Figure 7 Functional Test Setup
1
Set up the equipment as shown in Figure 7
2 If you are using an Agilent 81619A Interface module connect one optical
head to channel 1 and the other to channel 2
3 Power up the mainframe.
If the Agilent 81618A/19A passes all self-tests, the module is considered fully functional.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
43
Performance Tests Performance Tests
Performance Tests
The performance tests given in this section includes the Accuracy Test, the Linearity Test, the Return Loss Test (for the 81623B/26B) and the Noise Test. The performance tests for the Agilent 81624B/6B also include – as optional tests – the Relative Polarization Uncertainty and the Relative Interference Uncertainty Test. Perform each step in the order given, using the corresponding test equipment.
NOTE
Make sure that all optical connections are dry and clean. DO NOT USE INDEX MATCHING OIL. For cleaning, use the cleaning instructions given in “Cleaning Information” on page 81.
The optical cables from the laser source to the 8156A Attenuator and from there to the power meter must be fixed to the table to ensure minimum cable movement during the tests.
44 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
For 81628B only
The performance test for the 81628B High Power Head does not differ from the test procedure for an optical head without an integrating sphere, except for the linearity test. The integrating sphere of the 81628B has to be disconnected for testing the linearity of the head alone. A special adapter and a disconnecting tool are necessary for disconnection. These tools are included in the Performance Test Kit (order P/N 81628-68705).
If the integrating sphere is disconnected the performance of the 81628B has to be verified. If performing a performance test on the 81628B that includes the linearity test, it is necessary to start the procedure with the linearity test and then perform the accuracy and noise tests afterwards.
The non-linearity of the 81628B is not normally tested up to the specified power level of +38 dBm. The linearity test only verifies the linearity of the head without the integrating sphere up to 10 dBm (equivalent to 40 dBm with the integrating sphere). The attenuation of the integrating sphere is typically 31 ± 1 dB, hence the maximum input power for the photodiode is in the range 6 - 8 dBm and so the test described below covers this input power range of the photodiode. The contribution of the sphere to the non­linearity is caused by the heating of the sphere due to applied power > 30 dBm. These linearity characteristics are determined by sample testing of the 81628B in OCMD’s metrology lab.
The thermal characteristics of the sphere do not change from device to device, or during the product lifetime. If a head without the sphere passes the linearity test within the specifications given on the Test Record Sheet, then the whole system - sphere and head - is within the product specification.
The head and the integrating sphere of the 81628B are adjusted and serialized together, therefore the specifications of the 81628B are only valid for that given combination of head and integrating sphere.
The disconnecting and connecting of the integrating sphere may change the performance verification data slightly.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
45
Performance Tests Performance Tests
Accuracy Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical Heads.
NOTE
The linearity test must only be performed at either 1310 nm or 1550 nm. The accuracy test must be performed in the -20 dBm range at 10.0 µW at both 1310 nm and 1550 nm.
Tes t Se tu p
1 Make sure that cable connector, detectors and adapters are clean.
2 Connect the equipment as shown in Figure 8 .
Figure 8 Accuracy Test Setup
NOTE
Make sure that the cables to and from the attenuator are fixed on the table and that both the optical head and the DUT are close together so that minimum cable movement is required when connecting the cable to the head or to the DUT.
Use 81101PC patchcord for measurement path and 81001FA connector adapter (for the DUT) and 81000FA connector adapter (for reference power meter) instead.
46 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
3 Move to the Laser Source channel, move to the wavelength parameter,
[λ], press Enter, select the lower wavelength source, and press Enter.
4 If you are using an Agilent 81657A Laser Source make sure you initialize
the Agilent 8156A Optical Attenuator with 30 dB attenuation.
5 Turn the instruments on, enable the laser source and allow the
instruments to warm up for at least 20 minutes.
6 Perform the following sub-procedure for both Power Meters:
a Move to the Power Meter channel. b Move to the wavelength parameter, [λ], press Enter, enter the
wavelength of the laser source, and press Enter.
c Move to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.
d Move to the averaging time parameter, [AvgTime], press Enter,
move to <500 ms>, and press Enter.
e Move to the power parameter, [P], press [Pwr unit], move to
<Watt>, and press Enter.
7 Make sure the optical input of the Device Under Test (DUT), 8162xA/B,
is not receiving any light by placing a plastic cap over the input. Move to the DUT Power Meter channel, press [Menu], move to <Zero>, press Enter.
8 Ensure that the Agilent 8156A output is disabled. Move to the reference
Power Meter channel that uses the Agilent 8162xB Optical Head and Agilent 81618A Optical Head Interface module, press [Menu], move to <Zero>, press Enter.
9 Enable the Agilent 8156A output and change the attenuation until the
reference Power Meter displays 10.00 mW.
10 Connect the attenuator output cable to the DUT Power Meter. Note the
power value returned from the DUT, [P], from the display and note the result in the test record.
11 Move to the Laser Source channel, move to the wavelength parameter,
[λ], press Enter, select the longer wavelength source, and press Enter.
12 Repeat steps 4 to 10 at the second wavelength with the corresponding
source.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
47
Performance Tests Performance Tests
Linearity Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical Heads
NOTE
81628B only:
You have to disconnect the integrating sphere of the 81628B head to verify the linearity. Use the 81102SC high return loss patchcord to connect the DUT to the attenuator.
Figure 9 Measurement Setup for Power Linearity
48 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Test Setup
NOTE
NOTE
• Do not turn the laser off during the measurement!
• Clean all connectors carefully before you start with the measurement!
The linearity test checks the in-range linearity as well as the range discontinuity. Therefore, two values are measured at the same power but within different power ranges (range discontinuity) and two different power values , one on the lower and the other on the upper limit of the power range, are measured within the same power range (in-range linearity).
Perform the test until the power reading of the DUT shows:
for 81623B: -47.x dBm for 81624B: -57.x dBm for 81626B: -37.x dBm for 81628B: -57.x dBm (without sphere)
1 Make sure that cable connector, detectors and adapters are clean.
2 Make sure that you perform this test in a temperature-controlled
environment with temperature fluctuations less than ±1°C.
3 Setup the equipment as shown in Figure 9 . Disable both attenuators
and enable the laser source, where the source wavelength is chosen to 1550 nm. If you are using the 81657A, move to the wavelength parameter, [λ], press Enter, select the longer wavelength source (1550 nm nominally), and press Enter.
4 Set the wavelength of both attenuators to the same wavelength as the
laser source.
5 Perform the following sub-procedure for both Power Meters:
a Move to the Power Meter channel.
b Move to the wavelength parameter, [λ], press Enter, enter the
wavelength of the laser source, and press Enter.
c Move to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.
d Move to the averaging time parameter, [AvgTime], press Enter,
move to <100 ms>, and press Enter.
e Move to the power parameter, [P], press [Pwr unit], move to <dBm>,
and press Enter.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
49
Performance Tests Performance Tests
Press the [Menu] softkey and move to <Number of digits>, press
f
Enter, move to <3>, press Enter and press [Close].
NOTE
Always include at least three digits after the decimal point when you note a power reading.
6 Initialize the two attenuators as follows:
a Set the attenuation of the 8156A #221 with Monitor Output (referred
to as Atty1) to 0 dB.
b Set the attenuation of the other 8156A (referred to as Atty2) to
35 dB.
7 Wait at least 15 minutes until the laser source has stabilized.
8 Perform the following sub-procedure for the reference Power Meter,
81634B:
a Press [Menu], move to <Range mode>, move to <Manual>, and
press Enter.
b Move to <Range>, press Enter, move to <-50 dBm>, press Enter, and
press [Close].
9 Perform the following sub-procedure for the DUT, 8162xB:
a Press [Menu], move to <Range mode>, move to <Manual>, and
press Enter.
b Move to <Range>, press Enter, move to <10 dBm>(40 dBm for
81628B), and press Enter.
c Zero both Power Meters. Move to <Zero all> and press Enter.
10 Enable both attenuators.
11 Adjust the attenuation of Atty2 in order to achieve -47.2 dBm on the
reference Power Meter.
NOTE
For 81628B:
Before starting the measurement move to the DUT channel and enter a calibration factor of 31 dB. This simulates the disconnected integrating sphere. If the power at the DUT overflows when switching to the lower range decrease the calibration factor (e.g. to 30 dB) and start the linearity test again.
+10 dBm Range (equivalent to +40 dBm range for 81628B)
1 Applies to the 81626B only:
50 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Switch to the 20 dBm range and note both power readings as n=0 in the test record, which is given at the end of the test descriptions (#0, that is n=0). Switch back to the 10 dBm range.
2 Note both power readings as the first value (n=1) in the test record,
which is given at the end of the test descriptions (#1).
3 Increase the attenuation of Atty1 until the power reading of the DUT
shows about +2.8 dBm.
4 Note the InRange-values in the test record (#2)
5 At the DUT, 8162xB, move to <Range>, press Enter, move to <0 dBm> (30
dBm for 81628B), and press Enter.
6 Adjust the attenuation of Atty2 in order to be on the upper limit of the -
50 dBm range at the reference power meter (i.e. -47.2 dBm).
0 dBm Range (equivalent to +30 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the +10 dBm range (+40dBm for 81628).
5 Note both power readings (#3).
6 Switch down to the previous range (0 dBm, 30 dBm for the 81628B) and
note the values again (#4).
7 Increase Atty1 by 10 dB and note the results in the test record (#5).
8 At the DUT, 8162xB, press [Menu], move to <Range>, press Enter, move
to <-10 dBm>(20 dBm for the 81628B), press Enter, and press [Close].
9 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
10 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50 dBm range at the reference power meter.
-10 dBm Range (equivalent to +20 dBm range for 81628B)
1 Disable Atty1
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the 0 dBm range (30 dBm for the 81628B) and
note the power readings (#6).
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
51
Performance Tests Performance Tests
Switch down to the previous range (-10 dBm, 20 dBm for the 81628B) and
5
note the values again (#7).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#8).
7 At the DUT, 8162xB, press [Menu], move to <Range>, press Enter, move
to <-20 dBm>(10 dBm for the 81628B), press Enter and press [Close].
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 to 10 dB in order to be on the upper
limit of the -50 dBm range at the reference power meter.
-20 dBm Range (equivalent to +10 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -10 dBm (20 dBm for the 81628B) range and
note the power readings (#9).
5 Switch down to the previous (-20 dBm, 10 dBm for the 81628B) and note
the power readings again (#10).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#11).
7 On the DUT switch one range down to the -30 dBm (0 dBm for the
81628B) range.
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50 dBm range.
-30 dBm Range (equivalent to 0 dBm range for 81628B)
1 Disable Atty1.
2 Zero both power meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -20 dBm range (10 dBm for the 81628B) and
note the power readings (#12).
5 Switch down to the previous range (-30 dBm, 0 dBm for the 81628B) and
note the values again (#13).
52 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#14).
7 On the DUT switch one range down to the -40 dBm (-10 for 81628B)
range.
Change Setup
1 Disable Atty1 and switch the output with the monitor output.
2 Set the attenuation of Atty1 to 35 dB and of Atty2 to 25 dB.
3 Enable Atty1 again.
4 Adjust the attenuation of both attenuators in the following order:
• Atty1: DUT Power Meter shows a reading of -37.2 dBm.
• Atty2: the REF Power Meter shows a reading of -47.2 dBm.
-40 dBm Range (equivalent to -10 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -30 dBm (0 dBm for the 81628B) range and
note the power readings (#15).
5 Switch one range down (-40 dBm, -10 dBm for the 81628B) and note the
power readings again (#16).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#17).
7 O n t he DUT s witch one range down to the - 50 dBm ran ge (- 20 dB m f or the
81628B).
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50dBm range at the reference power meter.
-50 dBm Range (equivalent to -20 dBm range for 81628B)
1 Disable Atty1.
2 Zero both power meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -40 dBm (-10 dBm for the 81628B) range and
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
53
Performance Tests Performance Tests
note the power readings (#18).
5 Switch one range down (-50 dBm, -20 dBm for the 81628B)) and note the
power readings (#19).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#20).
Calculation
12 Calculate the non-linearity using the formulas given in the test record.
Example: Measurement Results
Information only Your Entries
Atty1 / #221
n
1 0 42 10 -47,1915 9,4273 1. Value
2 6,6 42 10 -53,769 2,8467 InRange
3 6,6 35,5 10 -47,262 2,8524 RangeDisc / lower limit
4 6,6 35,5 0 -47,2605 2,8543 RangeDisc / upper limit
5 16,6 35,5 0 -57,2304 -7,1141 InRange
6 16,6 25,5 0 -47,2363 -7,1118 RangeDisc / lower limit
7 16,6 25,5 -10 -47,2347 -7,1097 RangeDisc / upper limit
8 26,6 25,5 -10 -57,2345 -17,111 InRange
9 26,6 15,5 -10 -47,2411 ,17,106 RangeDisc / lower limit
10 26,6 15,5 -20 -47,2382 -17,1054 RangeDisc / upper limit
11 36,6 15,5 -20 -57,228 -27,0942 InRange
12 36,6 5,5 -20 -47,2317 -27,0864 RangeDisc / lower limit
13 36,6 5,5 -30 -47,2318 -27,0873 RangeDisc / upper limit
14 46,6 5,5 -30 -57,2336 -37,089 InRange
15 35,5 17,6 -30 -47,219 -37,253 RangeDisc / lower limit
16 35,5 17,6 -40 -47,2195 -37,22538 RangeDisc / upper limit
17 45,5 17,6 -40 -57,1333 -47,166 InRange
18 45,5 7,6 -40 -47,1411 -47,164 RangeDisc / lower limit
19 41,2 11,3 -50 -47,1355 -47,1664 RangeDisc / upper limit
20 51,2 11,3 -50 -57,134 -57,163 InRange
[dB]
Atty2 / #100
[dB]
DUT Range
[dBm]
REF
Power [dBm]
DUT Power
[dBm]
Notes
54 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Calculations
Conversion [dBm] [mW]
Calculation as given Calculation as given
n Ref / R DUT / D Relation1 / A Relation2 / B Non-Linearity
[mW] [mW]
= An*Bn (NL
n±1
+1) - 1
[%]
1 1,90919E-05 8,76455761 0,2199125
4,5505092
0,1
2 4,19856E-06 1,926060834
3 1,87845E-05 1,928590397 1,0003454 0,9995626 0,03
4 1,8791E-05 1,929434323 0,1006955 9,9275024 0,04
5 1,89217E-06 0,194352441
6 1,8896E-05 0,194455396 1,0003685 0,9995166 0,08
n
/R
n+1
= R
n+1
/D
n
= D
7 1,8903E-05 0,194549447 0,1000046 10,002994 0,09
8 1,89038E-06 0,019449122
9 1,88751E-05 0,019471526 1,000668 0,9998619 0,05
10 1,88877E-05 0,019474217 Reference Level 0,00
11 1,89322E-06 0,00195245 9,9765412
0,1002582
0,02
12 1,8916E-05 0,00195596
13 1,89156E-05 0,001955555 1,000023 0,9997928 0,01
14 1,89078E-06 0,000195479 10,004146 0,0999609 0,01
15 1,89714E-05 0,000188235
16 1,89692E-05 0,0001882 1,0001151 0,99981581 0,00
n
/R
n-1
= R
n-1
/D
n
= D
17 1,93495E-06 1,92044E-05 9,803474 0,1013767 0,03
18 1,93593E-05 1,92132E-05
19 1,93997E-05 1,92026E-05 1,001013651 0,9994475 0,08
20 1,93464E-06 1,92176E-06 9,996546719 0,1000783 0,13
NOTE
: The Nonlinearity is calculated recursively using n=10 as reference
point. The formula is:
where NL n=1, 4, 7 the values are NL
n+1
or NL
is the previously calculated non-linearity value. For
n-1
, while they are NL
n+2
for n=13, 16, 19.
n-2
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
55
Performance Tests Performance Tests
Figure 10 Example of Linearity Test Result.
56 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Noise Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical Heads.
NOTE
You must insert a module or a blank panel in the second channel position of the Agilent 8163A/B Lightwave Multimeter before you perform the noise measurement.
1 Insert an Agilent 81618A or an Agilent 81619A Optical Head Interface
module into one slot of the Agilent 8163A/B Lightwave Multimeter and connect the device under test, an 8162xA/B Optical Head, to the Optical Head Interface module.
2 Make sure the optical input of the Device Under Test (DUT), an
8162xA/B Optical Head, is not receiving any light by placing a plastic cap over the input adapter. Move to the DUT Power Meter channel, press [Menu], move to <Zero>, and press Enter.
3 Press Appl, move to <Stability>, and press Enter. The Stability Setup
Screen appears.
4 Ensure that the correct channel is selected in the upcoming Module
Selection box.
5 Press [Menu] to access the Logging application menu screen.
6 Move to <Pwr unit>, press Enter, move to <W>, and press Enter.
7 Move to <AvgTime>, press Enter, move to <1 s>, and press Enter.
8 Move to <Range mode>, press Enter, move to <Auto>, and press Enter,
and press [Close].
9 Press the [Parameter] softkey, move to [TotalTime], press Enter, set the
total time to 00:05:00, which is 5 minutes, and press Enter.
10 Press the [Measure] softkey to start the measurement.
11 After the stability application has finished, press the [Analysis] softkey,
press the [more] softkey, and note [P] as the noise value in the test record.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
57
Performance Tests Performance Tests
Return Loss Test
This performance test applies to Agilent 81623B/4B/6B/Optical Heads.
1 Make sure that all connectors are clean.
2 Connect the equipment as shown in Figure 11 .
3 Press [Preset] on the mainframe.
NOTE
CAUTION
Figure 11 Return Loss Reference Setup
To ensure traceability, use the 81610CC Reference Cable for calibration measurements. Do not use the 81610CC Reference Cable for measurements on a Device Under Test. Instead, use a measurement patchcord.
It is important to maintain the quality of the straight connector end of the 81610CC Reference Cable. Never add another connector to the straight end of the 81610CC Reference Cable, since a physical connection is made. When the straight end of the 81610CC Reference Cable is connected to the 81634B Power Sensor module, no physical connection is made so there should be no degradation of connector quality.
Zero the Return Loss Module and the Optical Head
4
5 Enable the internal laser source of the Return Loss Module and allow 20
minutes for it to stabilize.
58 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
6 At the Power Meter:
a Set the averaging time [AvTime] to 1s
b Set the wavelength [λ] to the wavelength of the RTL source.
c Set [Pwr unit] to <db>.
7 At the Return Loss Module:
a Set the averaging time [AvTime] to 1s.
b Enter the Return Loss Reference value RLref of the 81610CC
reference cable for this wavelength.
c Press [RefCal to calibrate the Return Loss module at reference
condition.
8 At the Power Meter press [Disp -> Ref]. The Power Meter should now
read 0.0 dB.
9 Disconnect the 81610CC Reference Cable from the setup and connect a
81113SC user cable to the Return Loss Module and leave the other end open as shown in Figure 8.
Figure 12 Return Loss Measurement Setup
At the Return Loss Module, press [TermCal] in order to calibrate the
10
Return Loss Module at termination condition.
Do not wrap the cable. The termination condition is realized with the open
NOTE
end of the patchcord.
11 Now connect the open end to the optical head.
12 Enter the Power reading [P] on the optical head as [FPDelta] on the
Return Loss Meter. Make sure you enter the sign of this value correctly.
13 Note the Return Loss reading as the Return Loss result into the test
record.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
59
Performance Tests Performance Tests
Relative Uncertainty due to Polarization (Optional Test)
NOTE
The performance test "Relative Uncertainty due to Polarization" is optional, since the polarization is given with the production of the unit by mechanical and optical cavities and is unchanged by normal use of the sensor module.
Refer to Figure 13 for a setup to verify the relative uncertainty due to polarization of the sensor module.
Generally, during this measurement procedure the tunable laser source is swept through a predefined wavelength range. After every wavelength step, a single PDL-measurement is made, where the polarization controller generates all the different polarization states. The highest PDL value is taken as the "relative uncertainty due to polarization". The low output power path of the coupler is used to monitor the power stability of the setup.
60 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Figure 13 Measurement Setup for PDL Test
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
61
Performance Tests Performance Tests
Relative Uncertainty due to Interference (Optional Test)
NOTE
The performance test "Relative Uncertainty due to Interference" is optional, since the interference is given with the production of the unit by mechanical and optical cavities and is unchanged by normal use of the sensor module.
Refer to Figure 14 for a test setup to verify the relative uncertainty due to interference within the optical head optical assembly. In order to perform the relative uncertainty due to interference test, it is mandatory to use two mainframes, since the time difference between measurement A und B for a specific wavelength point has to be no greater than 2 ms. Due to this short measurement interval, the performance test of the Relative Uncertainty Due to Interference can only be performed under computer control.
62 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Figure 14 Setup for Relative Uncertainty due to Interference Measurement
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
63
Performance Tests Performance Tests
Theoretically, both Power Meters are monitoring the power ratio over the variable wavelength in a predefined range as shown in Figure 15 . Ensure that the tunable laser source is mode-hop free in the tested wavelength range.
Figure 15 Interference Ripple
64 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81623B
Page 1 of 3
Model Agilent 81623B Optical Head_ Date ___________
Serial No. _________________________ Ambient Temperature ___________°C
Options _________________________ Relative Humidity ___________ %
Firmware Rev. _________________________ Line Frequency ___________ Hz
Test Facility _________________________ Customer _________________________
Performed by _________________________ Report No _________________________
Special Notes ______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
65
Performance Tests Performance Tests
Performance Test for the Agilent 81623B Test Equipment Used
Page 2 of 3
Description Model No. Trace No Cal. Due Date
1a1 Lightwave Multimeter (Std.) Agilent 8163A/B ____________ ___________
1a2 Lightwave Multimeter (DUT) Agilent 8163A/B ____________ ___________
1b TLS Mainframe Agilent 8164A/B ____________ ___________
2a Interface Module Agilent 81618A ____________ ___________
2b Interface Module Agilent 81619A ____________ ___________
3a CW Laser Source Agilent 81656A ____________ ___________
3b CW Laser Source Agilent 81657A ____________ ___________
4 CW Laser Source Agilent 81654A ____________ ___________
5a1 Opt. Head Interface Module Agilent 81618A ____________ ___________
5a2 Optical Head, Reference Agilent 81623A ____________ ___________
6 Sensor Module Agilent 81634A ____________ ___________
7 Optical Attenuator Agilent 8156A #221 ____________ ___________
8a Optical Attenuator Agilent 8156A #101 ____________ ___________
9 Return Loss Module Agilent 81613A ____________ ___________
10 ____________________________ ____________ ____________ ___________
11 ____________________________ ____________ ____________ ___________
12 ____________________________ ____________ ____________ ___________
Accessories # Product # Product # Product
Singlemode Fibers 2 81101AC 1 81113SC 3 81113PC
1 81610CC
Connector Interfaces 3 81000AI 3 81000FI 4 81000SI
Connector Adapters 2 81000AA 1 81000SA 1 81000FA
66 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81623B
Page 3 of 3
Model Agilent 81623B Optical Head Report No.________ Date ________________
Te s t
No. Test Description
IAccuracy Test
measured at _______ nm (1310nm) Output Power
measured at _______ nm (1550nm) Output Power
II Linearity Test
Range P
+10 +9 ________ _______ <± 1.13 %
+10 +3 ________ _______ <± 0.58 %
0 +3 ________ _______ <± 0.58 %
0 -7 ________ _______ <± 0.58 %
-10 -7 ________ _______ <± 0.58 %
-10 -17 ________ _______ <± 0.58 %
-20* -17* Reference
[dBm] P
DUT
For Calculations you may want to use the appropriate sheet
[dBm] Loss [%]
DUT
Min.
Spec.
9.72 µW _______ 10.28 µW
9.72 µW _______ 10.28 µW
Result Max.
[µW]
0.0
Spec.
<± 0.58 %
Measurement
Uncertainty
-20 -27 ________ _______
-30 -27 ________ _______
-30 -37 ________ _______ <± 0.63 %
-40 -37 ________ _______ <± 0.63 %
-40 -47 ________ _______ <± 1.08 %
-50 -47 ________ _______ <± 1.08%
-50 -57 ________ _______ <± 5.59 %
III Noise Test
IV Return Loss Test
50 dB < _______
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
<± 0.58 %
<± 0.58 %
[pW]
_______ < 100 pW
[dB]
67
Performance Tests Performance Tests
Performance Test for the Agilent 81624B
Page 1 of 3
Model Agilent 81624B Optical Head_ Date ___________
Serial No. _________________________ Ambient Temperature ___________°C
Options _________________________ Relative Humidity ___________ %
Firmware Rev. _________________________ Line Frequency ___________ Hz
Test Facility _________________________ Customer _________________________
Performed by _________________________ Report No _________________________
Special Notes ______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
68 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81624B Test Equipment Used
Page 2 of 3
Description Model No. Trace No Cal. Due Date
1a1 Lightwave Multimeter (Std.) Agilent 8163A/B ____________ ___________
1a2 Lightwave Multimeter (DUT) Agilent 8163A/B ____________ ___________
1b TLS Mainframe Agilent 8164A/B ____________ ___________
2a Interface Module Agilent 81618A ____________ ___________
2b Interface Module Agilent 81619A ____________ ___________
3a CW Laser Source Agilent 81656A ____________ ___________
3b CW Laser Source Agilent 81657A ____________ ___________
4 CW Laser Source Agilent 81654A ____________ ___________
5a1 Opt. Head Interface Module Agilent 81618A ____________ ___________
5a2 Optical Head, Reference Agilent 81624A ____________ ___________
6 Sensor Module Agilent 81634A ____________ ___________
7 Optical Attenuator Agilent 8156A #221 ____________ ___________
8a Optical Attenuator Agilent 8156A #101 ____________ ___________
9 Return Loss Module Agilent 81613A ____________ ___________
10 ____________________________ ____________ ____________ ___________
11 ____________________________ ____________ ____________ ___________
12 ____________________________ ____________ ____________ ___________
Accessories # Product # Product # Product
Singlemode Fibers 2 81101AC 1 81113SC 3 81113PC
1 81610CC
Connector Interfaces 3 81000AI 3 81000FI 4 81000SI
Connector Adapters 2 81000AA 1 81000SA 1 81000FA
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
69
Performance Tests Performance Tests
Performance Test for the Agilent 81624B
Page 3 of 3
Model Agilent 81624B Optical Head Report No.________ Date ________________
Tes t
No. Test Description
I Accuracy Test
measured at _______ nm (1310nm) Output Power
measured at _______ nm (1550nm) Output Power
II Linearity Test
Range P
+10 +9 ________ _______ <± 0.46 %
+10 +3 ________ _______ <± 0.46 %
0 +3 ________ _______ <± 0.46 %
0 -7 ________ _______ <± 0.46 %
-10 -7 ________ _______ <± 0.46 %
-10 -17 ________ _______ <± 0.46 %
-20* -17* Reference
[dBm] P
DUT
Min.
Spec.
9.72 µW _______ 10.28 µW
9.72 µW _______ 10.28 µW
For Calculations you may want to use the appropriate sheet
[dBm] Loss [%]
DUT
Result Max.
[µW]
0.0
Spec.
<± 0.46 %
Measurement
Uncertainty
-20 -27 ________ _______
-30 -27 ________ _______
-30 -37 ________ _______
-40 -37 ________ _______
-40 -47 ________ _______
-50 -47 ________ _______
-50 -57 ________ _______ <± 2.97 %
III Noise Test
IV Return Loss Test
[pW]
_______ < 5 pW
[dB]
typ. 60 dB _______
<± 0.46 %
<± 0.46 %
<± 0.49 %
<± 0.49 %
<± 0.71 %
<± 0.71 %
70 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81626B
Page 1 of 3
Model Agilent 81626B Optical Head_ Date ___________
Serial No. _________________________ Ambient Temperature ___________°C
Options _________________________ Relative Humidity ___________ %
Firmware Rev. _________________________ Line Frequency ___________ Hz
Test Facility _________________________ Customer _________________________
Performed by _________________________ Report No _________________________
Special Notes ______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
71
Performance Tests Performance Tests
Performance Test for the Agilent 81626B Test Equipment Used
Page 2 of 3
Description Model No. Trace No Cal. Due Date
1a1 Lightwave Multimeter (Std.) Agilent 8163A/B ____________ ___________
1a2 Lightwave Multimeter (DUT) Agilent 8163A/B ____________ ___________
1b TLS Mainframe Agilent 8164A/B ____________ ___________
2a Interface Module Agilent 81618A ____________ ___________
2b Interface Module Agilent 81619A ____________ ___________
3a CW Laser Source Agilent 81656A ____________ ___________
3b CW Laser Source Agilent 81657A ____________ ___________
4 CW Laser Source Agilent 81654A ____________ ___________
5a1 Opt. Head Interface Module Agilent 81618A ____________ ___________
5a2 Optical Head, Reference Agilent 81624A ____________ ___________
6 Sensor Module Agilent 81634A ____________ ___________
7 Optical Attenuator Agilent 8156A #221 ____________ ___________
8a Optical Attenuator Agilent 8156A #101 ____________ ___________
9 Return Loss Module Agilent 81613A ____________ ___________
10 ____________________________ ____________ ____________ ___________
11 ____________________________ ____________ ____________ ___________
12 ____________________________ ____________ ____________ ___________
Accessories # Product # Product # Product
Singlemode Fibers 2 81101AC 1 81113CC 3 81113PC
1 81610CC
Connector Interfaces 3 81000AI 3 81000FI 4 81000SI
Connector Adapters 2 81000AA 1 81000SA 1 81000FA
72 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81626B
Page 3 of 3
Model Agilent 81626B Optical Head Report No.________ Date ________________
Te s t
No. Test Description
IAccuracy Test
measured at _______ nm (1310nm) Output Power
measured at _______ nm (1550nm) Output Power
II Linearity Test
Range P
[dBm] P
DUT
For Calculations you may want to use the appropriate sheet
[dBm] Loss [%]
DUT
+20 +9 ________ _______ <± 1.29%
+10 +9 ________ _______ <± 0.93 %
+10 +3 ________ _______ <± 0.93 %
0 +3 ________ _______ <± 0.93 %
0 -7 ________ _______ <± 0.93 %
-10 -7 ________ _______ <± 0.93 %
-10 -17 ________ _______ <± 0.93 %
Min.
Spec. Result
Max.
Spec.
[µW]
9.64 µW _______ 10.36 µW
9.64 µW _______ 10.36 µW
[1]
Measurement
Uncertainty
-20 -17 ________ _______
-20* -17* Reference
-20 -27 ________ _______
-30 -27 ________ _______
-30 -37 ________ _______
-40 -37 ________ _______
-40 -47 ________ _______
III Noise Test
0.0
[pW]
<± 0.93 %
<± 0.93 %
<± 0.95 %
<± 0.95 %
<± 1.18 %
<± 0.18 %
<± 3.43 %
_______ < 500 pW
IV Return Loss Test
[dB]
60 dB < _______
[1] The 20 dB range measurement is only performed to check the functionality of the current booster. The published specifi­cation does not apply to operation in non-automode as used here to check the functionality.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
73
Performance Tests Performance Tests
NOTE
The nonlinearity of the 81626B is not usually tested to the specified power level of +27 dBm. Instead, limited testing up to +9 dBm is used to test the electronic circuitry of the 81626B. Above +9 dBm the largest contribution to nonlinearity is from the absorbing glass filter (which is tested on a sample basis) that does not change its linearity with time. Thus it is guaranteed by design that testing the 81626B up to +9 dBm ensures specification compliance to +27 dBm
74 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81628B
Page 1 of 3
Model Agilent 81628B Optical Head_ Date ___________
Serial No. _________________________ Ambient Temperature ___________°C
Options _________________________ Relative Humidity ___________ %
Firmware Rev. _________________________ Line Frequency ___________ Hz
Test Facility _________________________ Customer _________________________
Performed by _________________________ Report No _________________________
Special Notes ______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
75
Performance Tests Performance Tests
Performance Test for the Agilent 81628B Test Equipment Used
Page 2 of 3
Description Model No. Trace No Cal. Due Date
1a1 Lightwave Multimeter (Std.) Agilent 8163A/B ____________ ___________
1a2 Lightwave Multimeter (DUT) Agilent 8163A/B ____________ ___________
1b TLS Mainframe Agilent 8164A/B ____________ ___________
2a Interface Module Agilent 81618A ____________ ___________
2b Interface Module Agilent 81619A ____________ ___________
3a CW Laser Source Agilent 81656A ____________ ___________
3b CW Laser Source Agilent 81657A ____________ ___________
4 CW Laser Source Agilent 81654A ____________ ___________
5a1 Opt. Head Interface Module Agilent 81618A ____________ ___________
5a2 Optical Head, Reference Agilent 81624A ____________ ___________
6 Sensor Module Agilent 81634A ____________ ___________
7 Optical Attenuator Agilent 8156A #221 ____________ ___________
8a Optical Attenuator Agilent 8156A #101 ____________ ___________
9 Return Loss Module Agilent 81613A ____________ ___________
10 ____________________________ ____________ ____________ ___________
11 ____________________________ ____________ ____________ ___________
12 ____________________________ ____________ ____________ ___________
Accessories # Product # Product # Product
Singlemode Fibers 2 81101AC 1 81113SC 3 81113PC
1 81610CC
Connector Interfaces 3 81000AI 3 81000FI 4 81000SI
Connector Adapters 2 81000AA 1 81000SA 1 81000FA
76 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Performance Test for the Agilent 81628B
Page 3 of 3
Model Agilent 81628B Optical Head Report No.________ Date ________________
Test No. Test Description
I Linearity Test (without sphere)
Range P
+40 +9 ________ _______ <± 0.46 %
+40 +3 ________ _______ <± 0.46 %
+30 +3 ________ _______ <± 0.46 %
+30 -7 ________ _______ <± 0.46 %
+20 -7 ________ _______ <± 0.46 %
+20 -17 ________ _______ <± 0.46 %
+10* -17* Reference
+10 -27 ________ _______
0 -27 ________ _______
0 -37 ________ _______
-10 -37 ________ _______
[dBm] P
DUT
Min.
Spec.
For Calculations you may want to use the appropriate sheet
[dBm] Loss [%]
DUT
Result Max.
Spec.
0.0
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
Measurement Uncertainty
-10 -47 ________ _______
-20 -47 ________ _______
-20 -57 ________ _______ <± 0.71 %
II Accuracy Test (with sphere)
measured at _______ nm (1310nm) Output Power
measured at _______ nm (1550nm) Output Power
III Noise Test
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
9.64 µW _______ 10.36 µW
9.64 µW _______ 10.36 µW
<± 0.49 %
<± 0.49 %
[µW]
[pW]
_______ < 5 nW
77
Performance Tests Performance Tests
Calculation Sheets
Table 7 Calculation Sheet for Linearity Measurement (81623B, 81624B, 81628B)
Your E n trie s
n REF Power
[dBm]
DUT Power
[dBm]
Conversion [dBm] [mW]
Ref / R DUT / D Relation1 / A
[mW] [mW] = An*Bn (NL
Calculation as given Calculation as given
= R
n+1/Rn
Relation2 / B
= D
n/Dn+1
Non-Linearity
n±1
[%]
1 2 3 4 5 6
n
/R
n+1
=R
n+1
/R
n
=D
7 8 9
10 Reference Level 0.00 11 12 13 14 15 16 17
n
/R
n-1
=R
n-1
/R
n
=D
18 19 20
+1) - 1
78 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance Tests Performance Tests
Table 8 Calculation Sheet for Linearity Measurement (81626B)
Your E nt ries
n REF Power
[dBm]
DUT Power
[dBm]
Conversion [dBm] [mW]
Ref / R DUT / D Relation1 / A
[mW] [mW]
Calculation as given Calculation as given
= R
n+1/Rn
Relation2 / B
= D
n/Dn+1
Non-Linearity
= An*Bn (NL
n±1
[%]
1 2 3 4 5 6
n
/R
n+1
=R
n+1
/R
n
=D
7 8 9
10 Reference Level 0.00 11 12 13 14 15 16 17
/R
=R
n
n-1
n-1
/R
n
=D
18 19 20
+1) - 1
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
79
Performance Tests Performance Tests
80 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
5
Cleaning Information
The following Cleaning Information contains some general safety precautions, which must be observed during all phases of cleaning. Consult your specific optical device manuals or guides for full information on safety matters.
Please try, whenever possible, to use physically contacting connectors, and dry connections. Clean the connectors, interfaces, and bushings carefully after use.
If you are unsure of the correct cleaning procedure for your optical device, we recommend that you first try cleaning a dummy or test device.
Agilent Technologies assume no liability for the customer’s failure to comply with these requirements.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
81
Cleaning Information Cleaning Instructions for this Device
Cleaning Instructions for this Device
This Cleaning Information applies to a number of different types of Optical Equipment.
Sections of particularly relevance to the following devices are cross­referenced below.
81000xA Optical Head Adapters
When using optical head adapters, periodically inspect the optical head's front window (see “How to clean connector adapters” on page 96 for cleaning procedures). Dust and metal particles can be propelled through the adapter's pinhole while inserting the connector ferrule into the receptacle. These dirt particles collect on the head's front window and, if not removed, can lead to incorrect measurement results.
8162xx Optical Power Heads
Periodically inspect the optical head's front window for dust and other particles (see “How to clean instruments with an optical glass plate” on page 100 for cleaning procedures). These can be propelled through the optical head adapter's pinhole while inserting a connector ferrule into the receptacle. Particles on the optical head's front window can significantly impair measurement results.
82 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Safety Precautions Cleaning Information
Safety Precautions
Please follow the following safety rules:
• Do not remove instrument covers when operating.
• Ensure that the instrument is switched off throughout the cleaning procedures.
• Use of controls or adjustments or performance of procedures other than those specified may result in hazardous radiation exposure.
• Make sure that you disable all sources when you are cleaning any optical interfaces.
• Under no circumstances look into the end of an optical device attached to optical outputs when the device is operational. The laser radiation is not visible to the human eye, but it can seriously damage your eyesight.
• To prevent electrical shock, disconnect the instrument from the mains before cleaning. Use a dry cloth, or one slightly dampened with water, to clean the external case parts. Do not attempt to clean internally.
• Do not install parts or perform any unauthorized modification to optical devices.
• Refer servicing only to qualified and authorized personnel.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
83
Cleaning Information Why is it important to clean optical devices?
Why is it important to clean optical devices?
In transmission links optical fiber cores are about 9 µm (0.00035") in diameter. Dust and other particles, however, can range from tenths to hundredths of microns in diameter. Their comparative size means that they can cover a part of the end of a fiber core, and as a result will reduce the performance of your system.
Furthermore, the power density may burn dust into the fiber and cause additional damage (for example, 0 dBm optical power in a single mode fiber causes a power density of approximately 16 million W/m happens, measurements become inaccurate and non-repeatable.
Cleaning is, therefore, an essential yet difficult task. Unfortunately, when comparing most published cleaning recommendations, you will discover that they contain several inconsistencies. In this section, we want to suggest ways to help you clean your various optical devices, and thus significantly improve the accuracy and repeatability of your lightwave measurements.
2
). If this
84 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning? Cleaning Information
What do I need for proper cleaning?
Some Standard Cleaning Equipment is necessary for cleaning your instrument. For certain cleaning procedures, you may also require certain Additional Cleaning Equipment.
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following standard equipment:
Dust and shutter caps
Isopropyl alcohol
Cotton swabs
Soft tissues
Pipe cleaner
Compressed air
Dust and shutter caps
All of Agilent Technologies’ lightwave instruments are delivered with either laser shutter caps or dust caps on the lightwave adapter. Any cables come with covers to protect the cable ends from damage or contamination.
We suggest these protective coverings should be kept on the equipment at all times, except when your optical device is in use. Be careful when replacing dust caps after use. Do not press the bottom of the cap onto the fiber too hard, as any dust in the cap can scratch or pollute your fiber surface.
If you need further dust caps, please contact your nearest Agilent Technologies sales office.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
85
Cleaning Information What do I need for proper cleaning?
Isopropyl alcohol
This solvent is usually available from any local pharmaceutical supplier or chemist's shop.
If you use isopropyl alcohol to clean your optical device, do not immediately dry the surface with compressed air (except when you are cleaning very sensitive optical devices). This is because the dust and the dirt is solved and will leave behind filmy deposits after the alcohol is evaporated. You should therefore first remove the alcohol and the dust with a soft tissue, and then use compressed air to blow away any remaining filaments.
If possible avoid using denatured alcohol containing additives. Instead, apply alcohol used for medical purposes.
Never drink this alcohol, as it may seriously damage your health.
Do not use any other solvents, as some may damage plastic materials and claddings. Acetone, for example, will dissolve the epoxy used with fiber optic connectors. To avoid damage, only use isopropyl alcohol.
Cotton swabs
We recommend that you use swabs such as Q-tips or other cotton swabs normally available from local distributors of medical and hygiene products (for example, a supermarket or a chemist's shop). You may be able to obtain various sizes of swab. If this is the case, select the smallest size for your smallest devices.
Ensure that you use natural cotton swabs. Foam swabs will often leave behind filmy deposits after cleaning.
Use care when cleaning, and avoid pressing too hard onto your optical device with the swab. Too much pressure may scratch the surface, and could cause your device to become misaligned. It is advisable to rub gently over the surface using only a small circular movement.
Swabs should be used straight out of the packet, and never used twice. This is because dust and dirt in the atmosphere, or from a first cleaning, may collect on your swab and scratch the surface of your optical device.
86 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning? Cleaning Information
Soft tissues
These are available from most stores and distributors of medical and hygiene products such as supermarkets or chemists' shops.
We recommend that you do not use normal cotton tissues, but multi­layered soft tissues made from non-recycled cellulose. Cellulose tissues are very absorbent and softer. Consequently, they will not scratch the surface of your device over time.
Use care when cleaning, and avoid pressing on your optical device with the tissue. Pressing too hard may lead to scratches on the surface or misalignment of your device. Just rub gently over the surface using a small circular movement.
Use only clean, fresh soft tissues and never apply them twice. Any dust and dirt from the air which collects on your tissue, or which has gathered after initial cleaning, may scratch and pollute your optical device.
Pipe cleaner
Pipe cleaners can be purchased from tobacconists, and come in various shapes and sizes.The most suitable one to select for cleaning purposes has soft bristles, which will not produces scratches.
There are many different kinds of pipe cleaner available from tobacco nists.
The best way to use a pipe cleaner is to push it in and out of the device opening (for example, when cleaning an interface). While you are cleaning, you should slowly rotate the pipe cleaner.
Only use pipe cleaners on connector interfaces or on feed through adapters. Do not use them on optical head adapters, as the center of a pipe cleaner is hard metal and can damage the bottom of the adapter.
Your pipe cleaner should be new when you use it. If it has collected any dust or dirt, this can scratch or contaminate your device.
The tip and center of the pipe cleaner are made of metal. Avoid accidentally pressing these metal parts against the inside of the device, as this can cause scratches.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
87
Cleaning Information What do I need for proper cleaning?
Compressed air
Compressed air can be purchased from any laboratory supplier.
It is essential that your compressed air is free of dust, water and oil. Only use clean, dry air. If not, this can lead to filmy deposits or scratches on the surface of your connector. This will reduce the performance of your transmission system.
When spraying compressed air, hold the can upright. If the can is held at a slant, propellant could escape and dirty your optical device. First spray into the air, as the initial stream of compressed air could contain some condensation or propellant. Such condensation leaves behind a filmy deposit.
Please be friendly to your environment and use a CFC-free aerosol.
88 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning? Cleaning Information
Additional Cleaning Equipment
Some Cleaning Procedures need the following equipment, which is not required to clean each instrument:
Microscope with a magnification range about 50X up to 300X
Ultrasonic bath
Warm water and liquid soap
Premoistened cleaning wipes
Polymer film
Infrared Sensor Card
Microscope with a magnification range about 50X up to 300X
A microscope can be found in most photography stores, or can be obtained through or specialist mail order companies. Special fiber-scopes are available from suppliers of splicing equipment.
Ideally, the light source on your microscope should be very flexible. This will allow you to examine your device closely and from different angles.
A microscope helps you to estimate the type and degree of dirt on your device. You can use a microscope to choose an appropriate cleaning method, and then to examine the results. You can also use your microscope to judge whether your optical device (such as a connector) is severely scratched and is, therefore, causing inaccurate measurements.
Ultrasonic bath
Ultrasonic baths are also available from photography or laboratory suppliers or specialist mail order companies.
An ultrasonic bath will gently remove fat and other stubborn dirt from your optical devices. This helps increase the life span of the optical devices.
Only use isopropyl alcohol in your ultrasonic bath, as other solvents may cause damage.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
89
Cleaning Information What do I need for proper cleaning?
Warm water and liquid soap
Only use water if you are sure that there is no other way of cleaning your optical device without causing corrosion or damage. Do not use hot water, as this may cause mechanical stress, which can damage your optical device.
Ensure that your liquid soap has no abrasive properties or perfume in it. You should also avoid normal washing up liquid, as it can cover your device in an iridescent film after it has been air dried.
Some lenses and mirrors also have a special coating, which may be sensitive to mechanical stress, or to fat and liquids. For this reason we recommend you do not touch them.
If you are not sure how sensitive your device is to cleaning, please contact the manufacturer or your sales distributor.
Premoistened cleaning wipes
Use pre-moistened cleaning wipes as described in each individual cleaning procedure. Cleaning wipes may be used in every instance where a moistened soft tissue or cotton swab is applied.
Polymer film
Polymer film is available from laboratory suppliers or specialist mail order companies.
Using polymer film is a gentle method of cleaning extremely sensitive devices, such as reference reflectors and mirrors.
Infrared Sensor Card
Infrared sensor cards are available from laboratory suppliers or specialist mail order companies.
With this card you are able to control the shape of laser light emitted. The invisible laser beam is projected onto the sensor card, then becomes visible to the normal eye as a round spot.
Take care never to look into the end of a fiber or any other optical component, when they are in use. This is because the laser can seriously damage your eyes.
90 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Preserving Connectors Cleaning Information
Preserving Connectors
Listed below are some hints on how best to keep your connectors in the best possible condition.
Making Connections
Before you make any connection you must ensure that all cables and connectors are clean. If they are dirty, use the appropriate cleaning procedure.
When inserting the ferrule of a patchcord into a connector or an adapter, make sure that the fiber end does not touch the outside of the mating connector or adapter. Otherwise you will rub the fiber end against an unsuitable surface, producing scratches and dirt deposits on the surface of your fiber.
Dust Caps and Shutter Caps
Be careful when replacing dust caps after use. Do not press the bottom of the cap onto the fiber as any dust in the cap can scratch or dirty your fiber surface.
When you have finished cleaning, put the dust cap back on, or close the shutter cap if the equipment is not going to be used immediately.
Always keep the caps on the equipment when it is not in use.
All of Agilent Technologies’ lightwave instruments and accessories are shipped with either laser shutter caps or dust caps. If you need additional or replacement dust caps, contact your nearest Agilent Technologies Sales/Service Office.
Immersion Oil and Other Index Matching Compounds
Wherever possible, do not use immersion oil or other index matching compounds with your device. They are liable to impair and dirty the surface of the device. In addition, the characteristics of your device can be changed and your measurement results affected.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
91
Cleaning Information Cleaning Instrument Housings
Cleaning Instrument Housings
Use a dry and very soft cotton tissue to clean the instrument housing and the keypad. Do not open the instruments as there is a danger of electric shock, or electrostatic discharge. Opening the instrument can cause damage to sensitive components, and in addition your warranty will be voided.
92 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ? Cleaning Information
Which Cleaning Procedure should I use ?
Light dirt
If you just want to clean away light dirt, observe the following procedure for all devices:
• Use compressed air to blow away large particles.
• Clean the device with a dry cotton swab.
• Use compressed air to blow away any remaining filament left by the
swab.
Heavy dirt
If the above procedure is not enough to clean your instrument, follow one of the procedures below. Please consult Cleaning Instructions for this Device” on page 82 for the procedure relevant for this instrument.
If you are unsure of how sensitive your device is to cleaning, please contact the manufacturer or your sales distributor
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
93
Cleaning Information Which Cleaning Procedure should I use ?
How to clean connectors
Cleaning connectors is difficult as the core diameter of a single-mode fiber is only about 9 scratches on the surface. To be certain of the condition of the surface of your connector and to check it after cleaning, you need a microscope.
In the case of scratches, or of dust that has been burnt onto the surface of the connector, you may have no option but to polish the connector. This depends on the degree of dirtiness, or the depth of the scratches. This is a difficult procedure and should only be performed by a skilled person, and as a last resort as it wears out your connector.
µm. This generally means you cannot see streaks or
CAUTIONWARNING
Never look into the end of an optical cable that is connected to an active source.
To assess the projection of the emitted light beam you can use an infrared sensor card. Hold the card approximately 5 cm from the output of the connector. The invisible emitted light is projected onto the card and becomes visible as a small circular spot.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the connector by rubbing a new, dry cotton swab over the surface
using a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the connector:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the connector by rubbing the cotton swab over the surface using
a small circular movement.
3 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular movement.
4 Blow away any remaining lint with compressed air.
94 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ? Cleaning Information
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an ultrasonic bath with isopropyl alcohol.
1 Hold the tip of the connector in the bath for at least three minutes.
2 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular movement.
3 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
95
Cleaning Information Which Cleaning Procedure should I use ?
How to clean connector adapters
CAUTION
Some adapters have an anti-reflection coating on the back to reduce back reflection. This coating is extremely sensitive to solvents and mechanical abrasion. Extra care is needed when cleaning these adapters.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the adapter by rubbing a new, dry cotton swab over the surface
using a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the adapter:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the adapter by rubbing the cotton swab over the surface using a
small circular movement.
3 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular movement.
4 Blow away any remaining lint with compressed air.
96 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ? Cleaning Information
How to clean connector interfaces
CAUTION
Be careful when using pipe cleaners, as the core and the bristles of the pipe cleaner are hard and can damage the interface.
Do not use pipe cleaners on optical head adapters, as the hard core of normal pipe cleaners can damage the bottom of an adapter.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the interface by pushing and pulling a new, dry pipe cleaner into
the opening. Rotate the pipe cleaner slowly as you do this.
2 Then clean the interface by rubbing a new, dry cotton swab over the
surface using a small circular movement.
3 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the interface:
1 Moisten a new pipe cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe cleaner into the
opening. Rotate the pipe cleaner slowly as you do this.
3 Moisten a new cotton swab with isopropyl alcohol.
4 Clean the interface by rubbing the cotton swab over the surface using a
small circular movement.
5 Using a new, dry pipe cleaner, and a new, dry cotton swab remove the
alcohol, any dissolved sediment and dust.
6 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
97
Cleaning Information Which Cleaning Procedure should I use ?
How to clean bare fiber adapters
Bare fiber adapters are difficult to clean. Protect from dust unless they are in use.
CAUTION
CAUTION
Never use any kind of solvent when cleaning a bare fiber adapter as solvents can:
• Damage the foam inside some adapters.
• Deposit dissolved dirt in the groove, which can then dirty the surface of an inserted fiber.
Preferred Procedure
Use the following procedure on most occasions.
1 Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the adapter:
1 Clean the adapter by pushing and pulling a new, dry pipe cleaner into
the opening. Rotate the pipe cleaner slowly as you do this.
Be careful when using pipe cleaners, as the core and the bristles of the pipe cleaner are hard and can damage the adapter.
2 Clean the adapter by rubbing a new, dry cotton swab over the surface
using a small circular movement.
3 Blow away any remaining lint with compressed air.
98 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ? Cleaning Information
How to clean lenses
Some lenses have special coatings that are sensitive to solvents, grease, liquid and mechanical abrasion. Take extra care when cleaning lenses with these coatings.
Lens assemblies consisting of several lenses are not normally sealed. Therefore, use as little alcohol as possible, as it can get between the lenses and in doing so can change the properties of projection.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the lens by rubbing a new, dry cotton swab over the surface using
a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the lens:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the lens by rubbing the cotton swab over the surface using a
small circular movement.
3 Using a new, dry cotton swab remove the alcohol, any dissolved
sediment and dust.
4 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
99
Cleaning Information Which Cleaning Procedure should I use ?
How to clean instruments with a fixed connector interface
You should only clean instruments with a fixed connector interface when it is absolutely necessary. This is because it is difficult to remove any used alcohol or filaments from the input of the optical block.
It is important, therefore, to keep dust caps on the equipment at all times, except when your optical device is in use.
If you do discover filaments or particles, the only way to clean a fixed connector interface and the input of the optical block is to use compressed air.
If there are fluids or fat in the connector, please refer the instrument to the skilled personnel of Agilent’s service team.
CAUTION
Only use clean, dry compressed air. Make sure that the air is free of dust, water, and oil. If the air that you use is not clean and dry, this can lead to filmy deposits or scratches on the surface of your connector interface. This will degrade the performance of your transmission system.
Never try to open the instrument and clean the optical block by yourself, because it is easy to scratch optical components, and cause them to become misaligned.
How to clean instruments with an optical glass plate
Some instruments, for example, the optical heads from Agilent Technologies have an optical glass plate to protect the sensor. Clean this glass plate in the same way as optical lenses (see “How to clean lenses” on page 99).
100 Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition