No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or
tr ansl at ion int o a fo rei gn lan gu age ) w itho ut
prior agreement and written consent from
Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Agilent Technologies, Deutschland
GmbH
Herrenberger Straße 130
71034 Böblingen, Germany.
Manual Part Number
81623-90B12
Edition
Fourth edition April 2005
Third edition, July 2002
Second edition, January 2002
First edition, July 2001
Subject Matter
The material in this document is
subject to change without notice.
Agilent Technologies makes no
warranty of any kind with regard to this
printed material, including, but not
limited to, the implied warranties of
merchantability and fitness for a
particular purpose.
Agilent Technologies shall not be
liable for errors contained herein or for
incidental or consequential damages
in connection with the furnishing,
performance, or use of this material.
Printing History
New editio ns are complete revisions of
the guide reflecting alterations in the
functionality of the instrument.
Updates are occasionally made to the
guide between editions. The date on
the title page changes when an
updated guide is published. To find out
the current revision of the guide, or to
purchase an updated guide, contact
your Agilent Technologies
representative.
Warranty
This Agilent Technologies instrument
produ ct i s wa rra nte d ag ain st d efe cts in
material and workmanship for a period
of one year from date of shipment.
During the warranty period, Agilent
will, at its option, either repair or
replace products that prove to be
defective.
For warranty service or repair, this
product must be returned to a service
facility designated by Agilent. Buyer
shall prepay shipping charges to
Agilent and Agilent shall pay shipping
charges to return the product to Buyer.
However, Buyer shall pay all shipping
charges, duties, and taxes for products
returned to Agilent from another
country.
Agilent warrants that its software and
firmware designated by Agilent for use
with an instrument will execute its
programming instructions when
properly installed on that instrument.
Agilent does not warrant that the
operation of the instrument, software,
or firmware will be uninterrupted or
error free.
Limitation of Warranty
The foregoing warranty shall not apply
to defects resulting from improper or
inadequate maintenance by Buyer,
Buyer-supplied software or interfacing,
unauthorized modification or misuse,
operation outside of the environmental
specifications for the product, or
improper site preparation or
maintenance.
No other warranty is expressed or
implied. Agilent Technologies
specifically disclaims the implied
warranties of Merchantability and
Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein are
Buyer’s sole and exclusive remedies.
Agilent Technologies shall not be
liable for any direct, indirect, special,
incidental, or consequential damages
whether based on contract, tort, or any
other legal theory.
Assistance
Product maintenance agreements and
other customer assistance agreements
are available for Agilent Technologies
products. For any assistance contact
your nearest Agilent Technologies
Sales and Service Office.
Certification
Agilent Technologies Inc. certifies that
this product met its published
specifications at the time of shipment
from the factory.
Agilent Technologies further certifies
that its calibration measurements are
traceable to the United States National
Institute of Standards and Technology,
NIST (formerly the United States
National Bureau of Standards, NBS) to
the extent allowe d by the Institutes’s
calibration facility, and to the
calibration facilities of other
International Standards Organization
members.
ISO 9001 Certification
Produced to ISO 9001 international
quality system standard as part of our
objective of continually increasing
customer satisfaction through
improved process control.
Agilent Technologies Sales and Service Offices
For more information about Agilent Technologies test and measurement
products, applications, services, and for a current sales office listing, viesit
our web site:
http://www.agilent.com/comms/lightwave
You can also contact one of the following centers and ask for a test and
measurement sales representative.
United States:1 800 829 4444
1 800 829 4433 (FAX)
Canada:1 877 894 4414
1 800 746 4866 (FAX)
Europe:+31 20 547 2111
+31 20 547 2190 (FAX)
Japan:120 421 345
120 421 678 (FAX)
Mexico(52 55) 5081 9469
(52 55) 5081 9467 (FAX)
Australia:800 629 485
800 142 134 (FAX)
Asia-Pacific:+852 800 930 871
+852 800 930 871 (FAX)
Latin America+55 11 4197 3600
+55 11 4197 3800 (FAX)
Ta b l e o f Co ntents
Getting Started with Optical Heads9
Safety Considerations10
Safety Symbols10
Input Power Limitations11
Initial Inspection11
Line Power Requirements11
Operating Environment12
Storage and Shipment12
What is an Optical Head?13
Analog Output15
Optical input15
Heat Sink for 81628B optical Head16
Attaching the heat sink to the 81628B Optical Head16
Applicable adapters16
Mounting Instructions17
Accessories19
Modules and Options20
Connector Adapters - Reference List22
Specifications23
Definition of Terms24
Linearity24
Averaging Time25
Linewidth25
Noise25
Reference conditions25
Power range26
Wavelength range26
Relative uncertainty (spectral ripple) due to interference26
Relative uncertainty due to speckle noise27
Relative uncertainty due to polarization27
Return loss28
Spectral width of optical source28
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
5
Total uncertainty28
Uncertainty at reference conditions28
Optical Head Specifications29
Supplementary Performance Characteristics35
Analog Output35
Performance Tests37
Equipment Required38
Te st R e c ord40
Test Failure41
Instrument Specification42
Functional Tests43
Performance Tests44
For 81628B only45
Accuracy Test46
Linearity Test48
Test S e t u p49
+10 dBm Range (equivalent to +40 dBm range for 81628B)50
Calculation54
Noise Test57
Return Loss Test58
Relative Uncertainty due to Polarization (Optional Test)60
Relative Uncertainty due to Interference (Optional Test)62
Calculation Sheets78
Cleaning Information81
Cleaning Instructions for this Device82
81000xA Optical Head Adapters82
8162xx Optical Power Heads82
Safety Precautions83
Why is it important to clean optical devices?84
What do I need for proper cleaning?85
Standard Cleaning Equipment85
Dust and shutter caps85
Isopropyl alcohol86
Cotton swabs86
Soft tissues87
6Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Pipe cleaner87
Compressed air88
Additional Cleaning Equipment89
Microscope with a magnification range about 50X up to 300X89
Ultrasonic bath89
Warm water and liquid soap90
Preserving Connectors91
Cleaning Instrument Housings92
Which Cleaning Procedure should I use ?93
How to clean connectors94
How to clean connector adapters96
How to clean connector interfaces97
How to clean bare fiber adapters98
How to clean lenses99
How to clean instruments with a fixed connector interface100
How to clean instruments with an optical glass plate100
How to clean instruments with a physical contact interface 101
How to clean instruments with a recessed lens interface102
How to clean optical devices which are sensitive to mechanical
stress and pressure103
How to clean metal filters or attenuator gratings104
Additional Cleaning Information105
How to clean bare fiber ends105
How to clean large area lenses and mirrors106
Other Cleaning Hints108
Index109
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
7
8Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Getting Started with Optical Heads
1
Getting Started with Optical Heads
This chapter introduces the features of the Agilent 81623B/4B/6B/8B
Optical Heads.
Attaching the heat sink to the 81628B Optical Head . . . . . . . .16
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
9
Getting Started with Optical HeadsSafety Considerations
Safety Considerations
The following general safety precautions must be observed during all
phases of operation, service, and repair of this instrument. Failure to
comply with these precautions or with specific warnings elsewhere in this
manual violates safety standards of design, manufacture, and intended
use of the instrument. Agilent Technologies Inc. assumes no liability for
the customer’s failure to comply with these requirements.
Before operation, review the instrument and manual, including the red
safety page, for safety markings and instructions. You must follow these to
ensure safe operation and to maintain the instrument in safe condition.
CAUTIONWARNING
CAUTION
The WARNING sign denotes a hazard. It calls attention to a
procedure, practice or the like, which, if not correctly performed
or adhered to, could result in injury or loss of life. Do not proceed
beyond a WARNING sign until the indicated conditions are fully
understood and met.
A CAUTION notice denotes a hazard. It calls attention to an
operating procedure, practice, or the like that, if not correctly
performed or adhered to, could result in damage to the product or
loss of important data. Do not proceed beyond a CAUTION notice
until the indicated conditions are fully understood and met.
Safety Symbols
The apparatus will be marked with this symbol when it is necessary for the
user to refer to the instruction manual in order to protect the apparatus
against damage.
Caution - hot surface.
Magnetic fields - may interfere with a pacemaker.
10Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Safety ConsiderationsGetting Started with Optical Heads
Input Power Limitations
CAUTIONWARNING
Applicable to Agilent 81628B only -
Operation above 34 dBm (2.51 W) is at your own risk.
Operation at a continuous optical power, or an average optical
power, greater than 34 dBm causes, in the specified operating
temperature range, the metal parts (connector adapters, screws)
to become hotter than the required limit (55°C) of the safety
standard IEC 61010-1.
Agilent Technologies Deutschland GmbH will not be liable for any
damage caused by operation above 34 dBm.
Initial Inspection
Inspect the shipping container for damage. If there is damage to the
container or cushioning, keep them until you have checked the contents of
the shipment for completeness and verified the instrument both
mechanically and electrically.
The Performance Tests give procedures for checking the operation of the
instrument. If the contents are incomplete, mechanical damage or defect
is apparent, or if an instrument does not pass the operator’s checks, notify
the nearest Agilent Technologies Sales/Service Office.
CAUTIONWARNING
To avoid hazardous electrical shock, do not perform electrical
tests when there are signs of shipping damage to any portion of
the outer enclosure (covers, panels, etc.).
Line Power Requirements
The Agilent 81618A and Agilent 81619A Optical Head Interface Modules
operate when installed in the Agilent 8163A Lightwave Multimeter,
Agilent 8164A Lightwave Measurement System, and Agilent 8166A
Lightwave Multichannel System.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
11
Getting Started with Optical HeadsSafety Considerations
Operating Environment
The safety information in the Agilent 8163A Lightwave Multimeter,
Agilent 8164A Lightwave Measurement System, and Agilent 8166A
Lightwave Multichannel System User’s Guide summarizes the operating
ranges for the Agilent 81618A and Agilent 81619A Optical Head Interface
Modules. In order for these modules to meet specifications, the operating
environment must be within the limits specified for the Agilent 8163A
Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
and Agilent 8166A Lightwave Multichannel System.
Storage and Shipment
This module can be stored or shipped at temperatures between
−40°C and +70°C. Prote ct the module from tem perature extremes that ma y
cause condensation within it.
12Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head?Getting Started with Optical Heads
What is an Optical Head?
An optical head measures the power emitted from a connected singlemode or multi-mode fiber or the power applied in an open parallel beam
(with max. 5 mm diameter). The wavelength and power range depends on
the sensor element.
Screw to connector
adapter when fitted
Cable to interface
module
Analog output
Figure 1 Rear view of an Optical Head with Analog Output
Optical interface
connector adapter locking screws
Figure 2 Front View of Optical Head without Adapter
Figure 1 and Figure 2 show views of a typical optical head.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
13
Getting Started with Optical HeadsWhat is an Optical Head?
Figure 3 shows two types of adapter that are available for connecting the
input fiber to an optical head
.
Threaded connector adaptor:
81000xA, 81000BT
Figure 3 Connector adapters
D-type magnetic connector adaptor:
(preferred)
81003xx
81001xx
81000BC
14Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head?Getting Started with Optical Heads
Analog Output
The analog output is the BNC connector on the back of the optical head. It
outputs a voltage directly proportional to the strength of the optical signal
at the optical input in the current range. The analog signal is always in the
range between 0 and 2V, 2V corresponding to a full power input signal in
the current range, 0V corresponding to no input signal.
During autoranging, the level to which 2V corresponds changes. You
should disable autoranging when using the analog output.
Disabling Autoranging
Select the best range for the application using the mainframes user
interface. Select <Menu>, <Range Mode>, <Manual> and then choose
from the list of ranges.
Applications
Possible applications for the analog output are:
• to close the feedback loop controlling the current supplied to a laser
• to aid positioning the system for fiber alignment tasks
• to monitor optical power on an oscilloscope
The analog signal reacts instantaneously to the input signal, whereas the
power shown on the display is subject to averaging.
Optical input
The optical input to the optical head requires a connector adapter to match
the connector type (or bare fiber) to the optical head interface.
The optical head interface can be either threaded or a D-type magnetic
interface depending on the model of optical head (81628B is only available
with a threaded interface). See “Accessories” on page 19 for details of the
connector adapters available.
When a connector adapter is fitted it should be locked in place by
tightening the locking screws located either side of the main optical head
housing.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
15
Getting Started with Optical HeadsWhat is an Optical Head?
Heat Sink for 81628B optical Head
The 81628B optical head has a heat sink that allows an extended operating
power range. This heat sink must be mounted on the integrating sphere for
operation from 34 dBm to 38 dBm.
Attaching the heat sink to the 81628B
Optical Head
NOTE
NOTE
Always attach the heat sink and rubber ring to the 81628B if operating at
an optical power higher than 34dBm!
Applicable adapters
Agilent adapters types 81000xx, 81000BT are applicable to the 81628B
with the heat sink attached (without the heatsink and up to 34dBm the
81000BT bare fiber adapter may be attached). Bare fibers are not intended
to be used in high power applications.
If shipping the 81628B high power optical head please detach the heatsink
from the optical head to avoid damage during transportation.
16Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What is an Optical Head?Getting Started with Optical Heads
Mounting Instructions
1 Check that the blue rubber ring is correctly attached to the integrating
sphere and covers the metallic screws as shown in Figure 4 .
Figure 4 Attaching the Rubber Ring
2
The heatsink consists of two conical metal parts, the bottom part
(smallest diameter at the bottom), and the top part (largest diameter at
top). Slide the bottom part of the heat sink over the throat of the
integrating sphere with the smallest diameter facing to the sphere. Then
screw on the desired adapter (81000xx series) to the sphere as shown in
Figure 5 .
Figure 5 Attaching the Bottom Part of the Heat Sink
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
17
Getting Started with Optical HeadsWhat is an Optical Head?
Attach the upper part of the heat sink to the integrating sphere. Slide the
3
upper part over the connector adapter with the largest diameter facing
opposite to the sphere. Tighten the screws with the Allen key enclosed
in the Heat Sink Kit (see Figure 6 ).
Figure 6 Attaching the Upper Part of the Heat Sink
18Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
2
Accessories
The Agilent 81618A/9A Optical Head Interface Modules and
Agilent 81623B/4B/6B/8B Optical Heads are available in various
configurations for the best possible match to the most common
applications.
This chapter provides information on the available options and
accessories.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
19
AccessoriesModules and Options
Modules and Options
Optical Heads (5mm Sensors)
Accessories
81624CE 4 m extension cable
Interface Module
High Power
Optical Head for:
Connectorized Fiber,
Bare Fiber and
Open beam NA ≤ 0.3
81628B InGaAs
+40 dBm to –60 dBm
Connector Adapters
(threaded).
Interface Modules
81618A Single Head Interface Module
81619A Dual Head Interface Module
Connector Adapters
(threaded)
81000BT Bare Fiber Connectivity Set
(Threaded)
81000FA FC/PC/APC
81000KA SC/PC/APC
81000VA ST
81003LA LC
81000PA E-2000
20Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Modules and OptionsAccessories
High Power Optical Head (with integrating sphere)
High Power Optical Head (with integrating sphere)
Interface M odule
High Power
Optical Head for:
Connectoriz ed Fiber,
Accessories
81624CE 4 m extension cab le
Bare Fiber and
Open beam NA ≤ 0.3
81628B InGaAs
+40 dBm to –60 dBm
Connector Ad apters
(threaded).
OPTIONAL
Interface Modules
81618A Single Head Inte rface Module
81619A Dual Head Interfac e Module
Connector Adapters
(threaded)
81000BT Bare Fibe r Connectivity Set
(Threaded)
81000FA FC/PC/ APC
81000KA SC/PC/APC
81000VA ST
81003LA LC
81000PA E-2000
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
21
AccessoriesModules and Options
Connector Adapters - Reference List
Table 1 Connector Adapters
Preferred Adapters
Integral D-shape for 5 mm
Connector TypeThreaded Version
Barefiber81000BT81000BC
FC/PC81000FA81001FA
SC/PC81000KA81001KA
MU81001MA
E-200081000PA81001PA
ST81000VA
blank81001ZA
LC810003LA810001LA
MTP(ribbon)81003TD
optical heads
22Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
3
Specifications
The Agilent 81618A/9A Optical Head Interface Modules and
Agilent 81623B/4B/6B/8B Optical Heads are produced to the ISO 9001
international quality system standard as part of Agilent Technologies’
commitment to continually increasing customer satisfaction through
improved quality control.
Specifications describe the modules’ and heads’ warranted performance.
Supplementary performance characteristics describe the modules’ and
heads’ non-warranted typical performance.
Because of the modular nature of the instrument, these performance
specifications apply to these modules and heads rather than the
mainframe unit.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
23
SpecificationsDefinition of Terms
Definition of Terms
This section defines terms that are used both in this chapter and
“Performance Tests” on page 37.
Generally, all specifications apply for the given environmental conditions
and after warmup time.
Measurement principles are indicated. Alternative measurement
principles of equal value are also acceptable.
Linearity
The linearity error is defined as the relative difference between the
displayed power ratio, D
caused by changing the displayed power level from the reference level, D
to an arbitrary displayed level, D
if expressed in %
, and the actual (true)power ratio Px/P0
x/D0
. Symbol N.
x
,
0
D
⁄
xD0
N
------ ------- -----1Ð
P
⁄
xP0
100()=
if expressed in dB
D
⁄
xD0
N
dB
10
log=
------- ------ -----
P
⁄
xP0
Conditions: reference level 10 µW, displayed power levels within the
specified range, zero less than specified time prior to measurement.
Note 1: ideally N = 0 %, respectively 0 dB.
Note 2: the power-dependent nonlinearity, N(P
), can alternatively be
x
expressed by the following formula:
rP
()rP0()Ð
()
NP
x
x
------ ------ ---- ----- ------ ---- -----=
()
rP
0
where r(P) is the power-dependent responsivity (for a power meter, the
responsivity is defined as the ratio of displayed power to actual input
power).
24Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Definition of TermsSpecifications
Averaging Time
Time defining the period during which the power meter takes readings for
averaging. At the end of the averaging time the average of the readings is
available (display- or memory-update). Symbol T
avg
.
Linewidth
FWHM spectral bandwidth. The 3 dB width of the optical spectrum,
expressed in Hertz. Symbol: ∆f.
Noise
The peak-to-peak change of displayed power level with zero input power
level (dark).
Conditions: Zero prior to measurement, averaging time and observation
time as specified, lowest power range selected and wavelength range as
specified.
Measurement: the measurement result is obtained by:
− P
Noise = P
max
expressed as peak-to-peak within
min
the given time span. Any offset is
automatically excluded this way.
Reference conditions
The specified conditions during the spectral responsivity calibration, or
conditions which are extrapolated from the conditions during calibration.
Conditions: power level, beam diameter or fiber type, numerical aperture,
wavelength, spectral width, ambient temperature as specified, at the day
of calibration. →Noise and drift observed over a specified observation
time, with a temperature change of not more than±∆T.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
25
SpecificationsDefinition of Terms
Power range
The power range is defined from the highest specifiedinput power level to
the smallest input power level that causes a noticeable change of
displayed power level.
Conditions: wavelength, averaging time as specified.
Wavelength range
The range of wavelengths for which the power meter is calibrated.
Note: Selectable wavelength setting of the power meter for useful power
measurements (operating wavelength range).
Literature
[1]Fiber optic test and measurement, Hewlett Packard Professional
Books, edited by Prentice Hall, ISBN 0-13-534330-5
Relative uncertainty (spectral ripple) due to
interference
Uncertainty of power reading when using a coherent source, due to a
periodic change of the power meter's responsivity caused by optical
interference between reflective interfaces within the power meter's
optical assembly.
Conditions: constant wavelength, constant power level, angled connector
as specified, linewidth of source <100 MHz, temperature as specified.
NOTE
26Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Spectral ripple is measured by stepping the source wavelength over the
wavelength range specified.
Definition of TermsSpecifications
Relative uncertainty due to speckle noise
This is the uncertainty of the power reading when using a coherent source.
This is due to a variation of the power meters responsivity caused by
optical interference
meters optical assembly.
Conditions: constant wavelength, constant power level, angled connector
as specified, linewidth of source as specified, temperature as specified.
1
Incoming light to the integrating sphere undergoes numerous internal
reflections (as many as several hundred) prior to hitting the detector. If the
source is sufficiently coherent, a complicated interference pattern (speckle
pattern) appears spread over the whole sphere volume. Slight changes of
the ambient conditions (temperature, vibration, shock) affect particular
optical paths which consequently changes the speckle pattern and results
in an instability of the head power reading (speckle noise).
1
between different optical paths within the power
NOTE
Changes in the source wavelength have a similar effect on speckle noise
as environmental changes, because they cause fluctuations in the
interference pattern. If the source coherence length is small compared to
the effective path length, the interference pattern disappears and relative
uncertainty due to speckle noise becomes negligible.
Measurement: In contrast to the spectral ripple definition the source
wavelength isn’t stepped when measuring speckle noise.
Relative uncertainty due to polarization
Also termed polarization-dependent responsivity (PDR), the relative
uncertainty due to polarization is the uncertainty of the displayed power
level on the input polarization state, expressed as the difference between
the highest and the lowest displayed power. Uncertainty figures are based
upon a 95% confidence level.
Conditions: laser source with variable polarization state, generation of all
possible polarization states (covering the entire Poincaré sphere),
constant wavelength, constant power level, angled connector as specified,
temperature as specified.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
27
SpecificationsDefinition of Terms
Return loss
The ratio of the incident power to the reflected power expressed in dB.
Symbol: RL.
P
in
RL10
Conditions: the return loss excludes any reflections from the fiber end
used as radiation source.
log=
------ ------ --
P
back
Spectral width of optical source
Full width at half maximum. The 3 dB width of the optical spectrum,
expressed in nm. Symbol: FWHM.
To ta l u nc e rt ai nt y
The uncertainty for a specified set of operating conditions, including noise
and drift.
Conditions: power level, beam diameter or fiber type, numerical aperture,
wavelength, spectral width, ambient temperature, re-calibration period as
specified. →Noise and drift observed over a specified observation time,
with a temperature change of not more than ±∆T.
Uncertainty at reference conditions
The uncertainty for the specified set of reference conditions, including all
uncertainties in the calibration chain from the national laboratory to the
test meter.
28Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head SpecificationsSpecifications
Optical Head Specifications
All optical heads have to be operated with the single (Agilent 81618A) or dual
(Agilent 81619A) Interface Modules.
Table 2 Optical Head Specifications
Agilent 81623B
Agilent 81623B
Sensor elementGe, ∅ 5 mm
Wavelength range750 nm to 1800 nm
Power range–
Applicable fiber type
Open beam
Uncertainty at reference conditions [1]
Total uncertainty [2],
[9]
Relative uncertainty:
- due to polarization [3]
- spectral ripple
± 2.2 % (1000 nm to 1650 nm)
± 3.0 % (800 nm to 1000 nm)
± 3.5% ± 100 pW
(1000 nm to 1650 nm)
± 4.0% ± 250 pW
(800 nm to 1000 nm)
Standard SM and MM max 100 µm core size, NA ≤ 0.3
Calibration option C85 / C86
80 dBm to +10 dBm
Parallel beam max ∅ 4 mm
± 2.2 % (1000 nm to 1650 nm)
± 2.5 % (800 nm to 1000 nm)
± 3.5% ± 100 pW
(1000 nm to 1650 nm)
± 3.5% ± 250 pW
(800 nm to 1000 nm)
[10]
< ± 0.01 dB
< ± 0.006 dB (typ. < ± 0.003 dB)
(typ. < ± 0.005 dB)
Agilent 81623B
Calibration option C01 / C02
± 1.7 % (1000 nm to 1650 nm)
± 3.0 % (800 nm to 1000 nm)
± 3.0% ± 100 pW
(1000 nm to 1650 nm)
± 4.0% ± 250 pW
(800 nm to 1000 nm)
(due to interference)
[4]
Linearity (power): [5]
at 23°C ±5°C
at operating temp.
range
Return loss [7]> 50 dB, typ. > 55 dB
Noise (peak to peak)
[6]
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
• Parallel beam, 3 mm spot diameter on the center of the detector
• Ambient temperature 23°C ± 5°C
• On day of calibration (add ± 0.3% for aging over one year, add ± 0.6% over two years)
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[2]
Operating Conditions:
• P aral lel beam , 3m m sp ot d iame te r on the c ent er o f th e de tect or o r co nnec tor ized fib er w ith NA ≤ 0 .2 ( stra igh t
connector)
•Averaging time 1s
• For NA > 0.2: add 1%
• Within one year after calibration, add 0.3 % for second year
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[3]
All states of polarization at constant wavelength (1550 nm ± 30 nm) and straight connector, T = 23°C ± 5°C.
For angled connector (8 û ) add 0.01 dB typ.
[4] Conditions:
• Wavelength 1550 nm ± 30 nm, fixed state of polarization
• Temperature 23°C± 5°C
•Linewidth of source ÷ 100 M Hz
• Angled connector 8°
[5]
Does not include noise;
for wavelength < 1000 nm
applies for Ð 50dBm to + 10 dBm
[6]
Averaging time 1 s, T = 23°C ± 5°C, ∆T± 1°C, observa tion tim e 300 s.
[7]
Conditions:
Wavelengths 1550 nm ± 30 nm.
Standard single mode fiber, angled connector min 8°
[8]
With D-shape adapter 81001x x
Return Loss: > 60 dB typical
[9]
For input power > 2 mW
add ± 0.004 dB / mW;
zeroing required
[10]
Specif ication vali d for opti cal heads with
S/N starting with “DE413…”
(shipping began April 1, 2001)
30Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head SpecificationsSpecifications
Table 3 High Power Optical Head Specifications
Sensor element
Wavelength range
Power range
Applicable fiber type
Open beam
Uncertainty at refer
ence conditions
Total uncertainty
[2]± 3.5% ± 5 pW
[1]
Agilent 81624B
Calibration option
Agilent 81624B
C01 / C02
InGaAs, ∅ 5 mmInGaAs, ∅ 5mm
800 nm to 1700 nm850 nm to 1650nm
–90 dBm to +10 dBm–70 to +27 dBm (1250 nm to 1650 nm)
Standard SM and MM max 100 µm core size,
NA ≤ 0.3
Parallel beam max ∅ 4 mm
± 2.2 %
(1000nm to 1630 nm)
± 1.5 %
(970 nm to 1630 nm)
± 2.8% ± 5 pW
(1000nm to 1630 nm)
(970 nm to 1630 nm)
Agilent 81626B
Calibration option
Agilent 81626B
C01 / C02
–70 to +23 dBm (850 nm to 1650 nm)
Standard SM and MM max 100 µm core size,
NA ≤ 0.3
Parallel beam max ∅ 4 mm
± 3.0 %
(950 nm to 1630 nm)
± 5.0% ± 500 pW
[8]
(950 nm to 1630 nm)
± 2.5 %
(950 nm to 1630 nm)
± 4.5% ± 500 pW
(950 to 1630 nm max
23 dBm)
(1250 to 1630 nm
max 27 dBm)
[8]
Relative uncertainty:
- due to polarization
[3]
- spectral ripple
(due to interference)
[4]
Linearity (power):
- at 23 ûC ±5ûC
- at operat. temp.
range
Return loss
Noise (peak to peak)
Averaging time (min.)
Analog Output
Dimensions
Weight
Recommended recali
bration period
[5]
[7]
[7]
≤ ± 0.005 dB (typ. ± 0.002 dB)
≤ ± 0.005 dB (typ < ± 0.002 dB)
CW –70 dBm to +10 dBm, 1000 nm to 1630 nm
< ± 0.02 dB
< ± 0.05 dB
CW – 50 dBm to + 27dBm, 950 nm to 1630 nm
≤ ± 0.005 dB (typ. ± 0.002 dB)
≤ ± 0.005 dB (typ. < ± 0.002 dB)
< ± 0.04 dB
< ± 0.15 dB
typ. 60 dB> 45 dB> 47 dB
[6]< 5 pW< 500 pW
100 µs100 µs
includedIncluded
57 mm x 66 mm x 156 mm 57 mm x 66 mm x 156 mm
0.5 kg0.5 kg
2 years2 years
[8]
[8]
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
31
SpecificationsOptical Head Specifications
Operating temperature
Humidity
Warm-up time
0°C to 40°C0°C to +35°C
[9]
Non-condensingNon-condensing
40 min40 min
[1]
Reference conditions:
• Power level 10 µW
(-20 dBm), continuous wave (CW)
• Parallel beam, 3 mm spot diameter on the center of the detector
• Ambient temperature 23°C ± 5°C
• On day of calibration (add ± 0.3% foraging over one year, add ± 0.6% over two years)
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4nm
[2]
Operating Conditions:
• Parallel beam, 3mm spot di ameter on the center of the detector or connectorized fiber with NA ≤0.2 (straight
connector)
•Averaging time 1s
• For NA > 0.2: add 1%.
• Within one year after calibration,
add 0.3 % for second year.
•Zeroing required
[3]
All state s of polar ization at constant wavelength (155 0 nm ± 30 nm),
straight connector, T = 23°C ± 5°C.
For angled connector (8 û )
add 0.01 dB typ.
[4]
Conditions:
• Wavelength 1550 nm ± 30 nm,
fixed state of pol arization
• Temperature 23°C ± 5°C
•Linewidth of source ≥100 MHz
• Angled connector 8°
[5]
Does not include noise;
zeroing required
[6]
Averagin g time 1s, T = 23°C ± 5°C, ∆T ± 1°C, observation time 300 s.
Wavelength range 1200 nm to 1630nm
[7]
Conditions:
• Wavelengths 1550nm ± 30 nm
• Standard single mode fiber, angled connector min 8°
• With D-shape adapter 81001xx
Return Loss > 60 dB typica l
[8]
For input power > + 10 mW add:
add typ. ± 0.0016 dB/mW
In case of negative power change > 50 dB allow additional recovery time of 3 min
[9]
Max 30°C for > + 20 dBm input power
32Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Optical Head SpecificationsSpecifications
Table 4 Agilent 81628B Specification
Agilent 81628B
with integrating sphere
Sensor elementInGaAs
Wavelength range800 nm to 1700 nm
Power range–
60 dBm to +40 dBm (800 nm to 1700 nm)
For operation higher than 34 dBm
see safety note
Damage Power40.5 dBm
Applicable fiber type
Open beam
Uncertainty at reference
conditions
Total uncertainty
[1] [8]
[2] [8]
≤ 10 dBm
>10 dBm to ≤ 20 dBm
>20 dBm to ≤ 38 dBm
Single mode NA ≤ 0.2, Multimode NA ≤ 0.4
∅ ≤ 3mm center of sphere
± 3.0 % (970 nm to 1630nm)
(970 nm to 1630nm)
± 4.0% ± 5 nW
± 4.5%
± 5%
Relative uncertainty:
[3]
- due to polarization
typ. ≤ ± 0.006 dB
- due to speckle noise at
source linewidth:
[4]
typ. ≤ ± 0.02 dB
0.1pm to 100pm >100pm
Linearity (power):
[5] [8]
≤ 10 dBm
>10 dBm to ≤ 20 dBm
>20 dBm to ≤ 37 dBm
>37 dBm to ≤ 38 dBm
Return loss
Noise (peak to peak)
[6]
Averaging time (minimal)100 µs
Analog OutputIncluded
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
typ. ≤ ± 0.002 dB
(CW –40 dBm to + 38 dBm), (970 nm to 1630 nm)
≤ ± 0.03 dB
≤ ± 0.06 dB
≤ ± 0.09 dB
≤ ± 0.10 dB
at 23°C ± 5°C,
for operating temperature range add ±0.03 dB
typ. > 75 dB
< 5 nW
33
SpecificationsOptical Head Specifications
Dimensions 55mm x 80 mm x 250 mm
Weight0.9 kg (without heat sink)
Recommended Recalibra-
2 years
tion period
Operating temperature
[7]
0°C to +40°C
HumidityNon-condensing
Warm-up time40 min
Reference conditi ons:
[1]
Power lev el 10 µW (-20 dBm), continuous wave (CW)
•Averaging time 1s
• Parallel beam, 3 mm,
center of sphere input
• Ambient tempe rature 23 °C ± 5 °C
• On day of calibration (add ± 0.3 % for aging over one year, add ± 0.6 % over two years)
• Spectral width of source
<10 nm (FWHM)
• Wavelength setting at power sensor must correspond to source wavelength ± 0.4 nm
• Humidity 50 % ± 10 %
[2]
Operating Conditions:
• Parallel beam, ∅ 3mm , center of sphere input, or connectorized fiber with NA ≤0.2 (straight connector)
• For NA > 0.2: add 1%
• Within one year after calibration, add
± 0.3% for second year
• Operating temperature range as specified, humidity <80% and non-condensing
•Zeroing required
[3]
All stat es of polar ization at constant wavelength (1550 nm ± 30 nm) and co nstant power
[4]
Conditions:
• Wavelength 1550 nm ± 30 nm, fixed state of polarization, constant power
• Temperature 23°C ± 5°C
Measurement time ≤ 3 min
[5]
Does not include noise;
zeroing required
[6]
Averagin g time 1s, T = 23°C ± 5 °C,
∆T ± 1°C, observa tion tim e 300 s,
wavelength ran ge 970 nm to 1630 nm
Thermal drift at 38 dBm,
exposure time 30 min:
Recovery time 10 min: ≤ 30nW
30 m in: ≤ 10nW
[7]
For optical power > 30 dBm the maximal opera ting temperature is limited to 35°C
[8]
Wavelength must not be equal to any water absorption line
Safety Note:
For optical power higher than 34 dBm the attached heat sink MUST be used!
For continuous optical power or average optical power higher than 38 dBm the connector
adapters will get warmer than permitted according to the safety standard IEC 61010-1.
The 81628B Optical Head can handle optical power up to 40 dBm, however, operation
above 38 dBm is at the operators own risk.
Agilent Technologies Deutschland GmbH will not be liable for any damages caused by an
operation above 38 dBm.
34Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Bandwith ≥DC, ≤300 to 5000 Hz depending on range and optical head.
Output voltage 0 to 2 V non-terminated
Output impedance 600 Ohm typ.
Max. input voltage ±10V
Table 5 3dB-bandwidth of the Analog Output
Bandwidth
Range
+40 dBmN/AN/A3.5 kHz
+30 dBmN/AN/A3.5 kHz
+20 dBmN/AN/A3.5 kHz
+10dBm5.0kHz5.0kHz3.5kHz
0dBm5.0kHz5.0kHz1.8kHz
-10dBm5.0kHz5.0kHz1.8kHz
-20dBm5.0kHz5.0kHz0.12kHz
-30dBm0.5kHz1.8kHz0.12kHz
-40dBm0.5kHz
-50 dBm0.05 kHz
-60 dBm0.05 kHz
[1]
Not applicable for 81626B
[2]
Not applicable for 81622B
81622B/3B
[2]
[2]
[2]
Bandwidth
81624B/6B/7B
1.8 kHzN/A
0.12 kHz
0.12 kHz
[1]
[1]
Bandwidth
81628B
N/A
N/A
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
36Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
4
Performance Tests
The procedures in this section test the performance of the instrument. The
complete specifications to which Agilent 81623B/4B/6B/8B Optical
Heads are tested are given in “Specifications” on page 23. All tests can be
performed without access to the interior of the instrument. The test
equipment given corresponds to tests carried out with Diamond HMS - 10
connectors.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
37
Performance TestsEquipment Required
Equipment Required
Equipment required for the performance test is listed in the table below.
Any equipment that satisfies the critical specifications of the equipment
given in the table may be substituted for the recommended models.
Table 6 Equipment Required for Performance Test
InterfaceOptical Head
Instrument/AccessoryRecommended Model
81618A
81619A
81623B
81624B
81626B
81628B
Multimeter MainframeAgilent 8163A/B1 ea 1 ea 2 ea 2 ea 2 ea 2 eaAgilent
Legend : – not applicable; x necessary; o optional.
Alterna-
tive
Models
Required
Characteristics
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
39
Performance TestsTes t R ec o r d
Test Record
Results of the performance test may be tabulated on the Test Record
provided at the end of the test procedures. It is recommended that you fill
out the Test Record and refer to it while doing the test. Since the test limits
and setup information are printed on the Test Record for easy reference,
the record can also be used as an abbreviated test procedure (if you are
already familiar with the test procedures). The Test Record can also be
used as a permanent record and may be reproduced without written
permission from Agilent Technologies.
40Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Te s t F ai lu r ePerformance Tests
Test Failure
If the Agilent 81618A/19A Interface module or Agilent 81623B/4B/6B/8B
Optical Head fails any performance test, return the instrument to the
nearest Agilent Technologies Sales / Service Office for repair.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
41
Performance TestsInstrument Specification
Instrument Specification
Specifications are the performance characteristics of the instrument that
is certified. These specifications, listed in “Specifications” on page 23, are
the performance standards or limits against which the Agilent
81623B/4B/6B/8B Optical Head can be tested. “Specifications” on page 23 also lists some supplemental characteristics of the Agilent
81623B/4B/6B/8B Optical Head. Supplemental characteristics should be
considered as additional information.
Agilent 81618A/19A Interface Modules are tested for functionality only.
Any changes in the specifications due to manufacturing changes, design,
or traceability to the National Institute of Standards and Technology
(NIST), will be covered in a manual change supplement, or revised manual.
Such specifications supercede any that were previously published.
42Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Functional TestsPerformance Tests
Functional Tests
The functional test applies to the Agilent 81618A/19A Interface modules.
Mainframe 8163A/B with
8161xA Interface Module as DUT
Figure 7 Functional Test Setup
1
Set up the equipment as shown in Figure 7
2 If you are using an Agilent 81619A Interface module connect one optical
head to channel 1 and the other to channel 2
3 Power up the mainframe.
If the Agilent 81618A/19A passes all self-tests, the module is considered
fully functional.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
43
Performance TestsPerformance Tests
Performance Tests
The performance tests given in this section includes the Accuracy Test,
the Linearity Test, the Return Loss Test (for the 81623B/26B) and the
Noise Test. The performance tests for the Agilent 81624B/6B also include
– as optional tests – the Relative Polarization Uncertainty and the Relative
Interference Uncertainty Test. Perform each step in the order given, using
the corresponding test equipment.
NOTE
Make sure that all optical connections are dry and clean. DO NOT USE
INDEX MATCHING OIL. For cleaning, use the cleaning instructions given
in “Cleaning Information” on page 81.
The optical cables from the laser source to the 8156A Attenuator and from
there to the power meter must be fixed to the table to ensure minimum
cable movement during the tests.
44Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
For 81628B only
The performance test for the 81628B High Power Head does not differ from
the test procedure for an optical head without an integrating sphere,
except for the linearity test. The integrating sphere of the 81628B has to be
disconnected for testing the linearity of the head alone. A special adapter
and a disconnecting tool are necessary for disconnection. These tools are
included in the Performance Test Kit (order P/N 81628-68705).
If the integrating sphere is disconnected the performance of the 81628B
has to be verified. If performing a performance test on the 81628B that
includes the linearity test, it is necessary to start the procedure with the
linearity test and then perform the accuracy and noise tests afterwards.
The non-linearity of the 81628B is not normally tested up to the specified
power level of +38 dBm. The linearity test only verifies the linearity of the
head without the integrating sphere up to 10 dBm (equivalent to 40 dBm
with the integrating sphere). The attenuation of the integrating sphere is
typically 31 ± 1 dB, hence the maximum input power for the photodiode is
in the range 6 - 8 dBm and so the test described below covers this input
power range of the photodiode. The contribution of the sphere to the nonlinearity is caused by the heating of the sphere due to applied power > 30
dBm. These linearity characteristics are determined by sample testing of
the 81628B in OCMD’s metrology lab.
The thermal characteristics of the sphere do not change from device to
device, or during the product lifetime. If a head without the sphere passes
the linearity test within the specifications given on the Test Record Sheet,
then the whole system - sphere and head - is within the product
specification.
The head and the integrating sphere of the 81628B are adjusted and
serialized together, therefore the specifications of the 81628B are only
valid for that given combination of head and integrating sphere.
The disconnecting and connecting of the integrating sphere may change
the performance verification data slightly.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
45
Performance TestsPerformance Tests
Accuracy Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical
Heads.
NOTE
The linearity test must only be performed at either 1310 nm or 1550 nm.
The accuracy test must be performed in the -20 dBm range at 10.0 µW at
both 1310 nm and 1550 nm.
Tes t Se tu p
1 Make sure that cable connector, detectors and adapters are clean.
2 Connect the equipment as shown in Figure 8 .
Figure 8 Accuracy Test Setup
NOTE
Make sure that the cables to and from the attenuator are fixed on the table
and that both the optical head and the DUT are close together so that
minimum cable movement is required when connecting the cable to the
head or to the DUT.
Use 81101PC patchcord for measurement path and 81001FA connector
adapter (for the DUT) and 81000FA connector adapter (for reference power
meter) instead.
46Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
3 Move to the Laser Source channel, move to the wavelength parameter,
[λ], press Enter, select the lower wavelength source, and press Enter.
4 If you are using an Agilent 81657A Laser Source make sure you initialize
the Agilent 8156A Optical Attenuator with 30 dB attenuation.
5 Turn the instruments on, enable the laser source and allow the
instruments to warm up for at least 20 minutes.
6 Perform the following sub-procedure for both Power Meters:
aMove to the Power Meter channel.
bMove to the wavelength parameter, [λ], press Enter, enter the
wavelength of the laser source, and press Enter.
cMove to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.
dMove to the averaging time parameter, [AvgTime], press Enter,
move to <500 ms>, and press Enter.
eMove to the power parameter, [P], press [Pwr unit], move to
<Watt>, and press Enter.
7 Make sure the optical input of the Device Under Test (DUT), 8162xA/B,
is not receiving any light by placing a plastic cap over the input. Move to
the DUT Power Meter channel, press [Menu], move to <Zero>, press
Enter.
8 Ensure that the Agilent 8156A output is disabled. Move to the reference
Power Meter channel that uses the Agilent 8162xB Optical Head and
Agilent 81618A Optical Head Interface module, press [Menu], move to
<Zero>, press Enter.
9 Enable the Agilent 8156A output and change the attenuation until the
reference Power Meter displays 10.00 mW.
10 Connect the attenuator output cable to the DUT Power Meter. Note the
power value returned from the DUT, [P], from the display and note the
result in the test record.
11 Move to the Laser Source channel, move to the wavelength parameter,
[λ], press Enter, select the longer wavelength source, and press Enter.
12 Repeat steps 4 to 10 at the second wavelength with the corresponding
source.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
47
Performance TestsPerformance Tests
Linearity Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical Heads
NOTE
81628B only:
You have to disconnect the integrating sphere of the 81628B head to verify
the linearity. Use the 81102SC high return loss patchcord to connect the
DUT to the attenuator.
Figure 9 Measurement Setup for Power Linearity
48Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Test Setup
NOTE
NOTE
• Do not turn the laser off during the measurement!
• Clean all connectors carefully before you start with the measurement!
The linearity test checks the in-range linearity as well as the range
discontinuity. Therefore, two values are measured at the same power but
within different power ranges (range discontinuity) and two different
power values , one on the lower and the other on the upper limit of the
power range, are measured within the same power range (in-range
linearity).
Perform the test until the power reading of the DUT shows:
for 81623B: -47.x dBm
for 81624B: -57.x dBm
for 81626B: -37.x dBm
for 81628B: -57.x dBm (without sphere)
1 Make sure that cable connector, detectors and adapters are clean.
2 Make sure that you perform this test in a temperature-controlled
environment with temperature fluctuations less than ±1°C.
3 Setup the equipment as shown in Figure 9 . Disable both attenuators
and enable the laser source, where the source wavelength is chosen to
1550 nm. If you are using the 81657A, move to the wavelength
parameter, [λ], press Enter, select the longer wavelength source
(1550 nm nominally), and press Enter.
4 Set the wavelength of both attenuators to the same wavelength as the
laser source.
5 Perform the following sub-procedure for both Power Meters:
aMove to the Power Meter channel.
bMove to the wavelength parameter, [λ], press Enter, enter the
wavelength of the laser source, and press Enter.
cMove to the calibration parameter, [CAL], press Enter, set the
calibration parameter to zero, and press Enter.
dMove to the averaging time parameter, [AvgTime], press Enter,
move to <100 ms>, and press Enter.
eMove to the power parameter, [P], press [Pwr unit], move to <dBm>,
and press Enter.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
49
Performance TestsPerformance Tests
Press the [Menu] softkey and move to <Number of digits>, press
f
Enter, move to <3>, press Enter and press [Close].
NOTE
Always include at least three digits after the decimal point when you note
a power reading.
6 Initialize the two attenuators as follows:
aSet the attenuation of the 8156A #221 with Monitor Output (referred
to as Atty1) to 0 dB.
bSet the attenuation of the other 8156A (referred to as Atty2) to
35 dB.
7 Wait at least 15 minutes until the laser source has stabilized.
8 Perform the following sub-procedure for the reference Power Meter,
81634B:
aPress [Menu], move to <Range mode>, move to <Manual>, and
press Enter.
bMove to <Range>, press Enter, move to <-50 dBm>, press Enter, and
press [Close].
9 Perform the following sub-procedure for the DUT, 8162xB:
aPress [Menu], move to <Range mode>, move to <Manual>, and
press Enter.
bMove to <Range>, press Enter, move to <10 dBm>(40 dBm for
81628B), and press Enter.
cZero both Power Meters. Move to <Zero all> and press Enter.
10 Enable both attenuators.
11 Adjust the attenuation of Atty2 in order to achieve -47.2 dBm on the
reference Power Meter.
NOTE
For 81628B:
Before starting the measurement move to the DUT channel and enter a
calibration factor of 31 dB. This simulates the disconnected integrating
sphere. If the power at the DUT overflows when switching to the lower
range decrease the calibration factor (e.g. to 30 dB) and start the linearity
test again.
+10 dBm Range (equivalent to +40 dBm range for 81628B)
1 Applies to the 81626B only:
50Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Switch to the 20 dBm range and note both power readings as n=0 in the
test record, which is given at the end of the test descriptions (#0, that is
n=0). Switch back to the 10 dBm range.
2 Note both power readings as the first value (n=1) in the test record,
which is given at the end of the test descriptions (#1).
3 Increase the attenuation of Atty1 until the power reading of the DUT
shows about +2.8 dBm.
4 Note the InRange-values in the test record (#2)
5 At the DUT, 8162xB, move to <Range>, press Enter, move to <0 dBm> (30
dBm for 81628B), and press Enter.
6 Adjust the attenuation of Atty2 in order to be on the upper limit of the -
50 dBm range at the reference power meter (i.e. -47.2 dBm).
0 dBm Range (equivalent to +30 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the +10 dBm range (+40dBm for 81628).
5 Note both power readings (#3).
6 Switch down to the previous range (0 dBm, 30 dBm for the 81628B) and
note the values again (#4).
7 Increase Atty1 by 10 dB and note the results in the test record (#5).
8 At the DUT, 8162xB, press [Menu], move to <Range>, press Enter, move
to <-10 dBm>(20 dBm for the 81628B), press Enter, and press [Close].
9 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
10 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50 dBm range at the reference power meter.
-10 dBm Range (equivalent to +20 dBm range for 81628B)
1 Disable Atty1
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zeroall>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the 0 dBm range (30 dBm for the 81628B) and
note the power readings (#6).
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
51
Performance TestsPerformance Tests
Switch down to the previous range (-10 dBm, 20 dBm for the 81628B) and
5
note the values again (#7).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#8).
7 At the DUT, 8162xB, press [Menu], move to <Range>, press Enter, move
to <-20 dBm>(10 dBm for the 81628B), press Enter and press [Close].
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 to 10 dB in order to be on the upper
limit of the -50 dBm range at the reference power meter.
-20 dBm Range (equivalent to +10 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -10 dBm (20 dBm for the 81628B) range and
note the power readings (#9).
5 Switch down to the previous (-20 dBm, 10 dBm for the 81628B) and note
the power readings again (#10).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#11).
7 On the DUT switch one range down to the -30 dBm (0 dBm for the
81628B) range.
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50 dBm range.
-30 dBm Range (equivalent to 0 dBm range for 81628B)
1 Disable Atty1.
2 Zero both power meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -20 dBm range (10 dBm for the 81628B) and
note the power readings (#12).
5 Switch down to the previous range (-30 dBm, 0 dBm for the 81628B) and
note the values again (#13).
52Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#14).
7 On the DUT switch one range down to the -40 dBm (-10 for 81628B)
range.
Change Setup
1 Disable Atty1 and switch the output with the monitor output.
2 Set the attenuation of Atty1 to 35 dB and of Atty2 to 25 dB.
3 Enable Atty1 again.
4 Adjust the attenuation of both attenuators in the following order:
• Atty1: DUT Power Meter shows a reading of -37.2 dBm.
• Atty2: the REF Power Meter shows a reading of -47.2 dBm.
-40 dBm Range (equivalent to -10 dBm range for 81628B)
1 Disable Atty1.
2 Zero both Power Meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -30 dBm (0 dBm for the 81628B) range and
note the power readings (#15).
5 Switch one range down (-40 dBm, -10 dBm for the 81628B) and note the
power readings again (#16).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#17).
7 O n t he DUT s witch one range down to the - 50 dBm ran ge (- 20 dB m f or the
81628B).
8 If necessary, adjust the attenuation of Atty1 in order to be on the upper
limit of the range (i.e. -x7.y dBm).
9 Decrease the attenuation of Atty2 by 10 dB in order to be on the upper
limit of the -50dBm range at the reference power meter.
-50 dBm Range (equivalent to -20 dBm range for 81628B)
1 Disable Atty1.
2 Zero both power meters. On the 8163A/B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
3 Enable Atty1.
4 Switch one range up to the -40 dBm (-10 dBm for the 81628B) range and
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
53
Performance TestsPerformance Tests
note the power readings (#18).
5 Switch one range down (-50 dBm, -20 dBm for the 81628B)) and note the
power readings (#19).
6 Increase the attenuation of Atty1 by 10 dB and note the results in the test
record (#20).
Calculation
12 Calculate the non-linearity using the formulas given in the test record.
54Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Calculations
Conversion [dBm] → [mW]
Calculation as givenCalculation as given
nRef / RDUT / D Relation1 / ARelation2 / BNon-Linearity
[mW][mW]
= An*Bn (NL
n±1
+1) - 1
[%]
11,90919E-058,764557610,2199125
4,5505092
0,1
24,19856E-061,926060834
31,87845E-051,9285903971,00034540,99956260,03
41,8791E-051,9294343230,10069559,92750240,04
51,89217E-060,194352441
61,8896E-050,1944553961,00036850,99951660,08
n
/R
n+1
= R
n+1
/D
n
= D
71,8903E-050,1945494470,100004610,0029940,09
81,89038E-060,019449122
91,88751E-050,0194715261,0006680,99986190,05
101,88877E-050,019474217Reference Level0,00
111,89322E-060,001952459,9765412
0,1002582
0,02
121,8916E-050,00195596
131,89156E-050,0019555551,0000230,99979280,01
141,89078E-060,00019547910,0041460,09996090,01
151,89714E-050,000188235
161,89692E-050,00018821,00011510,999815810,00
n
/R
n-1
= R
n-1
/D
n
= D
171,93495E-061,92044E-059,8034740,10137670,03
181,93593E-051,92132E-05
191,93997E-051,92026E-051,0010136510,99944750,08
201,93464E-061,92176E-069,9965467190,10007830,13
NOTE
: The Nonlinearity is calculated recursively using n=10 as reference
point. The formula is:
where NL
n=1, 4, 7 the values are NL
n+1
or NL
is the previously calculated non-linearity value. For
n-1
, while they are NL
n+2
for n=13, 16, 19.
n-2
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
55
Performance TestsPerformance Tests
Figure 10 Example of Linearity Test Result.
56Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Noise Test
This performance test applies to Agilent 81623B/4B/6B/8B Optical
Heads.
NOTE
You must insert a module or a blank panel in the second channel position
of the Agilent 8163A/B Lightwave Multimeter before you perform the
noise measurement.
1 Insert an Agilent 81618A or an Agilent 81619A Optical Head Interface
module into one slot of the Agilent 8163A/B Lightwave Multimeter and
connect the device under test, an 8162xA/B Optical Head, to the Optical
Head Interface module.
2 Make sure the optical input of the Device Under Test (DUT), an
8162xA/B Optical Head, is not receiving any light by placing a plastic
cap over the input adapter. Move to the DUT Power Meter channel,
press [Menu], move to <Zero>, and press Enter.
3 Press Appl, move to <Stability>, and press Enter. The Stability Setup
Screen appears.
4 Ensure that the correct channel is selected in the upcoming Module
Selection box.
5 Press [Menu] to access the Logging application menu screen.
6 Move to <Pwr unit>, press Enter, move to <W>, and press Enter.
7 Move to <AvgTime>, press Enter, move to <1 s>, and press Enter.
8 Move to <Range mode>, press Enter, move to <Auto>, and press Enter,
and press [Close].
9 Press the [Parameter] softkey, move to [TotalTime], press Enter, set the
total time to 00:05:00, which is 5 minutes, and press Enter.
10 Press the [Measure] softkey to start the measurement.
11 After the stability application has finished, press the [Analysis] softkey,
press the [more] softkey, and note [∆P] as the noise value in the test
record.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
57
Performance TestsPerformance Tests
Return Loss Test
This performance test applies to Agilent 81623B/4B/6B/Optical Heads.
1 Make sure that all connectors are clean.
2 Connect the equipment as shown in Figure 11 .
3 Press [Preset] on the mainframe.
NOTE
CAUTION
Figure 11 Return Loss Reference Setup
To ensure traceability, use the 81610CC Reference Cable for calibration
measurements. Do not use the 81610CC Reference Cable for
measurements on a Device Under Test. Instead, use a measurement
patchcord.
It is important to maintain the quality of the straight connector end
of the 81610CC Reference Cable. Never add another connector to
the straight end of the 81610CC Reference Cable, since a physical
connection is made. When the straight end of the 81610CC
Reference Cable is connected to the 81634B Power Sensor module,
no physical connection is made so there should be no degradation
of connector quality.
Zero the Return Loss Module and the Optical Head
4
5 Enable the internal laser source of the Return Loss Module and allow 20
minutes for it to stabilize.
58Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
6 At the Power Meter:
aSet the averaging time [AvTime] to 1s
bSet the wavelength [λ] to the wavelength of the RTL source.
cSet [Pwr unit] to <db>.
7 At the Return Loss Module:
aSet the averaging time [AvTime] to 1s.
bEnter the Return Loss Reference value RLref of the 81610CC
reference cable for this wavelength.
cPress [RefCal to calibrate the Return Loss module at reference
condition.
8 At the Power Meter press [Disp -> Ref]. The Power Meter should now
read 0.0 dB.
9 Disconnect the 81610CC Reference Cable from the setup and connect a
81113SC user cable to the Return Loss Module and leave the other end
open as shown in Figure 8.
Figure 12 Return Loss Measurement Setup
At the Return Loss Module, press [TermCal] in order to calibrate the
10
Return Loss Module at termination condition.
Do not wrap the cable. The termination condition is realized with the open
NOTE
end of the patchcord.
11 Now connect the open end to the optical head.
12 Enter the Power reading [P] on the optical head as [FPDelta] on the
Return Loss Meter. Make sure you enter the sign of this value correctly.
13 Note the Return Loss reading as the Return Loss result into the test
record.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
59
Performance TestsPerformance Tests
Relative Uncertainty due to Polarization
(Optional Test)
NOTE
The performance test "Relative Uncertainty due to Polarization" is
optional, since the polarization is given with the production of the unit by
mechanical and optical cavities and is unchanged by normal use of the
sensor module.
Refer to Figure 13 for a setup to verify the relative uncertainty due to
polarization of the sensor module.
Generally, during this measurement procedure the tunable laser source is
swept through a predefined wavelength range. After every wavelength
step, a single PDL-measurement is made, where the polarization controller
generates all the different polarization states. The highest PDL value is
taken as the "relative uncertainty due to polarization". The low output
power path of the coupler is used to monitor the power stability of the
setup.
60Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Figure 13 Measurement Setup for PDL Test
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
61
Performance TestsPerformance Tests
Relative Uncertainty due to Interference
(Optional Test)
NOTE
The performance test "Relative Uncertainty due to Interference" is
optional, since the interference is given with the production of the unit by
mechanical and optical cavities and is unchanged by normal use of the
sensor module.
Refer to Figure 14 for a test setup to verify the relative uncertainty due to
interference within the optical head optical assembly. In order to perform
the relative uncertainty due to interference test, it is mandatory to use two
mainframes, since the time difference between measurement A und B for
a specific wavelength point has to be no greater than 2 ms. Due to this
short measurement interval, the performance test of the Relative
Uncertainty Due to Interference can only be performed under computer
control.
62Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Figure 14 Setup for Relative Uncertainty due to Interference Measurement
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
63
Performance TestsPerformance Tests
Theoretically, both Power Meters are monitoring the power ratio over the
variable wavelength in a predefined range as shown in Figure 15 . Ensure
that the tunable laser source is mode-hop free in the tested wavelength
range.
Figure 15 Interference Ripple
64Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Performance Test for the Agilent 81623B
Page 1 of 3
ModelAgilent 81623B Optical Head_Date ___________
Serial No._________________________Ambient Temperature ___________°C
72Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Performance Test for the Agilent 81626B
Page 3 of 3
Model Agilent 81626B Optical HeadReport No.________Date ________________
Te s t
No.Test Description
IAccuracy Test
measured at _______ nm
(1310nm) Output Power
measured at _______ nm
(1550nm) Output Power
IILinearity Test
RangeP
[dBm]P
DUT
For Calculations you may want to use the appropriate sheet
[dBm]Loss [%]
DUT
+20+9 _______________<± 1.29%
+10+9 _______________<± 0.93 %
+10+3 _______________<± 0.93 %
0+3_______________<± 0.93 %
0-7_______________<± 0.93 %
-10-7_______________<± 0.93 %
-10-17_______________<± 0.93 %
Min.
Spec.Result
Max.
Spec.
[µW]
9.64 µW_______10.36 µW
9.64 µW_______10.36 µW
[1]
Measurement
Uncertainty
-20-17_______________
-20*-17*Reference
-20-27_______________
-30-27_______________
-30-37_______________
-40-37_______________
-40-47_______________
IIINoise Test
0.0
[pW]
<± 0.93 %
<± 0.93 %
<± 0.95 %
<± 0.95 %
<± 1.18 %
<± 0.18 %
<± 3.43 %
_______< 500 pW
IVReturn Loss Test
[dB]
60 dB <_______
[1] The 20 dB range measurement is only performed to check the functionality of the current booster. The published specification does not apply to operation in non-automode as used here to check the functionality.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
73
Performance TestsPerformance Tests
NOTE
The nonlinearity of the 81626B is not usually tested to the specified power
level of +27 dBm. Instead, limited testing up to +9 dBm is used to test the
electronic circuitry of the 81626B. Above +9 dBm the largest contribution
to nonlinearity is from the absorbing glass filter (which is tested on a
sample basis) that does not change its linearity with time. Thus it is
guaranteed by design that testing the 81626B up to +9 dBm ensures
specification compliance to +27 dBm
74Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Performance Test for the Agilent 81628B
Page 1 of 3
ModelAgilent 81628B Optical Head_Date ___________
Serial No._________________________Ambient Temperature ___________°C
76Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Performance Test for the Agilent 81628B
Page 3 of 3
Model Agilent 81628B Optical HeadReport No.________Date ________________
Test No.Test Description
ILinearity Test (without sphere)
RangeP
+40+9 _______________<± 0.46 %
+40+3 _______________<± 0.46 %
+30+3_______________<± 0.46 %
+30-7_______________<± 0.46 %
+20-7_______________<± 0.46 %
+20-17_______________<± 0.46 %
+10*-17*Reference
+10-27_______________
0-27_______________
0-37_______________
-10-37_______________
[dBm]P
DUT
Min.
Spec.
For Calculations you may want to use the appropriate sheet
[dBm]Loss [%]
DUT
Result Max.
Spec.
0.0
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
Measurement
Uncertainty
-10-47_______________
-20-47_______________
-20-57_______________<± 0.71 %
IIAccuracy Test (with sphere)
measured at _______ nm
(1310nm) Output Power
measured at _______ nm
(1550nm) Output Power
IIINoise Test
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
9.64 µW_______10.36 µW
9.64 µW_______10.36 µW
<± 0.49 %
<± 0.49 %
[µW]
[pW]
_______< 5 nW
77
Performance TestsPerformance Tests
Calculation Sheets
Table 7 Calculation Sheet for Linearity Measurement (81623B, 81624B, 81628B)
Your E n trie s
nREF Power
[dBm]
DUT Power
[dBm]
Conversion [dBm] → [mW]
Ref / RDUT / DRelation1 / A
[mW][mW]= An*Bn (NL
Calculation as givenCalculation as given
= R
n+1/Rn
Relation2 / B
= D
n/Dn+1
Non-Linearity
n±1
[%]
1
2
3
4
5
6
n
/R
n+1
=R
n+1
/R
n
=D
7
8
9
10Reference Level0.00
11
12
13
14
15
16
17
n
/R
n-1
=R
n-1
/R
n
=D
18
19
20
+1) - 1
78Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Performance TestsPerformance Tests
Table 8 Calculation Sheet for Linearity Measurement (81626B)
Your E nt ries
nREF Power
[dBm]
DUT Power
[dBm]
Conversion [dBm] → [mW]
Ref / RDUT / DRelation1 / A
[mW][mW]
Calculation as givenCalculation as given
= R
n+1/Rn
Relation2 / B
= D
n/Dn+1
Non-Linearity
= An*Bn (NL
n±1
[%]
1
2
3
4
5
6
n
/R
n+1
=R
n+1
/R
n
=D
7
8
9
10Reference Level0.00
11
12
13
14
15
16
17
/R
=R
n
n-1
n-1
/R
n
=D
18
19
20
+1) - 1
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
79
Performance TestsPerformance Tests
80Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
5
Cleaning Information
The following Cleaning Information contains some general safety
precautions, which must be observed during all phases of cleaning.
Consult your specific optical device manuals or guides for full information
on safety matters.
Please try, whenever possible, to use physically contacting connectors,
and dry connections. Clean the connectors, interfaces, and bushings
carefully after use.
If you are unsure of the correct cleaning procedure for your optical device,
we recommend that you first try cleaning a dummy or test device.
Agilent Technologies assume no liability for the customer’s failure to
comply with these requirements.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
81
Cleaning InformationCleaning Instructions for this Device
Cleaning Instructions for this
Device
This Cleaning Information applies to a number of different types of Optical
Equipment.
Sections of particularly relevance to the following devices are crossreferenced below.
81000xA Optical Head Adapters
When using optical head adapters, periodically inspect the optical head's
front window (see “How to clean connector adapters” on page 96 for
cleaning procedures). Dust and metal particles can be propelled through
the adapter's pinhole while inserting the connector ferrule into the
receptacle. These dirt particles collect on the head's front window and, if
not removed, can lead to incorrect measurement results.
8162xx Optical Power Heads
Periodically inspect the optical head's front window for dust and other
particles (see “How to clean instruments with an optical glass plate” on page 100 for cleaning procedures). These can be propelled through the
optical head adapter's pinhole while inserting a connector ferrule into the
receptacle. Particles on the optical head's front window can significantly
impair measurement results.
82Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Safety PrecautionsCleaning Information
Safety Precautions
Please follow the following safety rules:
• Do not remove instrument covers when operating.
• Ensure that the instrument is switched off throughout the cleaning
procedures.
• Use of controls or adjustments or performance of procedures other than
those specified may result in hazardous radiation exposure.
• Make sure that you disable all sources when you are cleaning any
optical interfaces.
• Under no circumstances look into the end of an optical device attached
to optical outputs when the device is operational. The laser radiation is
not visible to the human eye, but it can seriously damage your eyesight.
• To prevent electrical shock, disconnect the instrument from the mains
before cleaning. Use a dry cloth, or one slightly dampened with water, to
clean the external case parts. Do not attempt to clean internally.
• Do not install parts or perform any unauthorized modification to optical
devices.
• Refer servicing only to qualified and authorized personnel.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
83
Cleaning InformationWhy is it important to clean optical devices?
Why is it important to clean
optical devices?
In transmission links optical fiber cores are about 9 µm (0.00035") in
diameter. Dust and other particles, however, can range from tenths to
hundredths of microns in diameter. Their comparative size means that they
can cover a part of the end of a fiber core, and as a result will reduce the
performance of your system.
Furthermore, the power density may burn dust into the fiber and cause
additional damage (for example, 0 dBm optical power in a single mode
fiber causes a power density of approximately 16 million W/m
happens, measurements become inaccurate and non-repeatable.
Cleaning is, therefore, an essential yet difficult task. Unfortunately, when
comparing most published cleaning recommendations, you will discover
that they contain several inconsistencies. In this section, we want to
suggest ways to help you clean your various optical devices, and thus
significantly improve the accuracy and repeatability of your lightwave
measurements.
2
). If this
84Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning?Cleaning Information
What do I need for proper
cleaning?
Some Standard Cleaning Equipment is necessary for cleaning your
instrument. For certain cleaning procedures, you may also require certain
Additional Cleaning Equipment.
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following
standard equipment:
• Dust and shutter caps
• Isopropyl alcohol
• Cotton swabs
• Soft tissues
• Pipe cleaner
• Compressed air
Dust and shutter caps
All of Agilent Technologies’ lightwave instruments are delivered with
either laser shutter caps or dust caps on the lightwave adapter. Any cables
come with covers to protect the cable ends from damage or
contamination.
We suggest these protective coverings should be kept on the equipment at
all times, except when your optical device is in use. Be careful when
replacing dust caps after use. Do not press the bottom of the cap onto the
fiber too hard, as any dust in the cap can scratch or pollute your fiber
surface.
If you need further dust caps, please contact your nearest Agilent
Technologies sales office.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
85
Cleaning InformationWhat do I need for proper cleaning?
Isopropyl alcohol
This solvent is usually available from any local pharmaceutical supplier or
chemist's shop.
If you use isopropyl alcohol to clean your optical device, do not
immediately dry the surface with compressed air (except when you are
cleaning very sensitive optical devices). This is because the dust and the
dirt is solved and will leave behind filmy deposits after the alcohol is
evaporated. You should therefore first remove the alcohol and the dust
with a soft tissue, and then use compressed air to blow away any
remaining filaments.
If possible avoid using denatured alcohol containing additives. Instead,
apply alcohol used for medical purposes.
Never drink this alcohol, as it may seriously damage your health.
Do not use any other solvents, as some may damage plastic materials and
claddings. Acetone, for example, will dissolve the epoxy used with fiber
optic connectors. To avoid damage, only use isopropyl alcohol.
Cotton swabs
We recommend that you use swabs such as Q-tips or other cotton swabs
normally available from local distributors of medical and hygiene products
(for example, a supermarket or a chemist's shop). You may be able to
obtain various sizes of swab. If this is the case, select the smallest size for
your smallest devices.
Ensure that you use natural cotton swabs. Foam swabs will often leave
behind filmy deposits after cleaning.
Use care when cleaning, and avoid pressing too hard onto your optical
device with the swab. Too much pressure may scratch the surface, and
could cause your device to become misaligned. It is advisable to rub gently
over the surface using only a small circular movement.
Swabs should be used straight out of the packet, and never used twice.
This is because dust and dirt in the atmosphere, or from a first cleaning,
may collect on your swab and scratch the surface of your optical device.
86Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning?Cleaning Information
Soft tissues
These are available from most stores and distributors of medical and
hygiene products such as supermarkets or chemists' shops.
We recommend that you do not use normal cotton tissues, but multilayered soft tissues made from non-recycled cellulose. Cellulose tissues
are very absorbent and softer. Consequently, they will not scratch the
surface of your device over time.
Use care when cleaning, and avoid pressing on your optical device with
the tissue. Pressing too hard may lead to scratches on the surface or
misalignment of your device. Just rub gently over the surface using a small
circular movement.
Use only clean, fresh soft tissues and never apply them twice. Any dust
and dirt from the air which collects on your tissue, or which has gathered
after initial cleaning, may scratch and pollute your optical device.
Pipe cleaner
Pipe cleaners can be purchased from tobacconists, and come in various
shapes and sizes.The most suitable one to select for cleaning purposes
has soft bristles, which will not produces scratches.
There are many different kinds of pipe cleaner available from tobacco
nists.
The best way to use a pipe cleaner is to push it in and out of the device
opening (for example, when cleaning an interface). While you are cleaning,
you should slowly rotate the pipe cleaner.
Only use pipe cleaners on connector interfaces or on feed through
adapters. Do not use them on optical head adapters, as the center of a pipe
cleaner is hard metal and can damage the bottom of the adapter.
Your pipe cleaner should be new when you use it. If it has collected any
dust or dirt, this can scratch or contaminate your device.
The tip and center of the pipe cleaner are made of metal. Avoid
accidentally pressing these metal parts against the inside of the device, as
this can cause scratches.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
87
Cleaning InformationWhat do I need for proper cleaning?
Compressed air
Compressed air can be purchased from any laboratory supplier.
It is essential that your compressed air is free of dust, water and oil. Only
use clean, dry air. If not, this can lead to filmy deposits or scratches on the
surface of your connector. This will reduce the performance of your
transmission system.
When spraying compressed air, hold the can upright. If the can is held at a
slant, propellant could escape and dirty your optical device. First spray into
the air, as the initial stream of compressed air could contain some
condensation or propellant. Such condensation leaves behind a filmy
deposit.
Please be friendly to your environment and use a CFC-free aerosol.
88Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
What do I need for proper cleaning?Cleaning Information
Additional Cleaning Equipment
Some Cleaning Procedures need the following equipment, which is not
required to clean each instrument:
• Microscope with a magnification range about 50X up to 300X
• Ultrasonic bath
• Warm water and liquid soap
• Premoistened cleaning wipes
• Polymer film
• Infrared Sensor Card
Microscope with a magnification range about 50X up to
300X
A microscope can be found in most photography stores, or can be obtained
through or specialist mail order companies. Special fiber-scopes are
available from suppliers of splicing equipment.
Ideally, the light source on your microscope should be very flexible. This
will allow you to examine your device closely and from different angles.
A microscope helps you to estimate the type and degree of dirt on your
device. You can use a microscope to choose an appropriate cleaning
method, and then to examine the results. You can also use your
microscope to judge whether your optical device (such as a connector) is
severely scratched and is, therefore, causing inaccurate measurements.
Ultrasonic bath
Ultrasonic baths are also available from photography or laboratory
suppliers or specialist mail order companies.
An ultrasonic bath will gently remove fat and other stubborn dirt from your
optical devices. This helps increase the life span of the optical devices.
Only use isopropyl alcohol in your ultrasonic bath, as other solvents may
cause damage.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
89
Cleaning InformationWhat do I need for proper cleaning?
Warm water and liquid soap
Only use water if you are sure that there is no other way of cleaning your
optical device without causing corrosion or damage. Do not use hot water,
as this may cause mechanical stress, which can damage your optical
device.
Ensure that your liquid soap has no abrasive properties or perfume in it.
You should also avoid normal washing up liquid, as it can cover your device
in an iridescent film after it has been air dried.
Some lenses and mirrors also have a special coating, which may be
sensitive to mechanical stress, or to fat and liquids. For this reason we
recommend you do not touch them.
If you are not sure how sensitive your device is to cleaning, please contact
the manufacturer or your sales distributor.
Premoistened cleaning wipes
Use pre-moistened cleaning wipes as described in each individual cleaning
procedure. Cleaning wipes may be used in every instance where a
moistened soft tissue or cotton swab is applied.
Polymer film
Polymer film is available from laboratory suppliers or specialist mail order
companies.
Using polymer film is a gentle method of cleaning extremely sensitive
devices, such as reference reflectors and mirrors.
Infrared Sensor Card
Infrared sensor cards are available from laboratory suppliers or specialist
mail order companies.
With this card you are able to control the shape of laser light emitted. The
invisible laser beam is projected onto the sensor card, then becomes
visible to the normal eye as a round spot.
Take care never to look into the end of a fiber or any other optical
component, when they are in use. This is because the laser can seriously
damage your eyes.
90Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Preserving ConnectorsCleaning Information
Preserving Connectors
Listed below are some hints on how best to keep your connectors in the
best possible condition.
Making Connections
Before you make any connection you must ensure that all cables and
connectors are clean. If they are dirty, use the appropriate cleaning
procedure.
When inserting the ferrule of a patchcord into a connector or an adapter,
make sure that the fiber end does not touch the outside of the mating
connector or adapter. Otherwise you will rub the fiber end against an
unsuitable surface, producing scratches and dirt deposits on the surface of
your fiber.
Dust Caps and Shutter Caps
Be careful when replacing dust caps after use. Do not press the bottom of
the cap onto the fiber as any dust in the cap can scratch or dirty your fiber
surface.
When you have finished cleaning, put the dust cap back on, or close the
shutter cap if the equipment is not going to be used immediately.
Always keep the caps on the equipment when it is not in use.
All of Agilent Technologies’ lightwave instruments and accessories are
shipped with either laser shutter caps or dust caps. If you need additional
or replacement dust caps, contact your nearest Agilent Technologies
Sales/Service Office.
Immersion Oil and Other Index Matching Compounds
Wherever possible, do not use immersion oil or other index matching
compounds with your device. They are liable to impair and dirty the surface
of the device. In addition, the characteristics of your device can be
changed and your measurement results affected.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
91
Cleaning InformationCleaning Instrument Housings
Cleaning Instrument Housings
Use a dry and very soft cotton tissue to clean the instrument housing and
the keypad. Do not open the instruments as there is a danger of electric
shock, or electrostatic discharge. Opening the instrument can cause
damage to sensitive components, and in addition your warranty will be
voided.
92Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ?Cleaning Information
Which Cleaning Procedure
should I use ?
Light dirt
If you just want to clean away light dirt, observe the following procedure
for all devices:
• Use compressed air to blow away large particles.
• Clean the device with a dry cotton swab.
• Use compressed air to blow away any remaining filament left by the
swab.
Heavy dirt
If the above procedure is not enough to clean your instrument, follow one
of the procedures below. Please consult “Cleaning Instructions for this Device” on page 82 for the procedure relevant for this instrument.
If you are unsure of how sensitive your device is to cleaning, please
contact the manufacturer or your sales distributor
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
93
Cleaning InformationWhich Cleaning Procedure should I use ?
How to clean connectors
Cleaning connectors is difficult as the core diameter of a single-mode fiber
is only about 9
scratches on the surface. To be certain of the condition of the surface of
your connector and to check it after cleaning, you need a microscope.
In the case of scratches, or of dust that has been burnt onto the surface of
the connector, you may have no option but to polish the connector. This
depends on the degree of dirtiness, or the depth of the scratches. This is a
difficult procedure and should only be performed by a skilled person, and
as a last resort as it wears out your connector.
µm. This generally means you cannot see streaks or
CAUTIONWARNING
Never look into the end of an optical cable that is connected to an
active source.
To assess the projection of the emitted light beam you can use an infrared
sensor card. Hold the card approximately 5 cm from the output of the
connector. The invisible emitted light is projected onto the card and
becomes visible as a small circular spot.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the connector by rubbing a new, dry cotton swab over the surface
using a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the connector:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the connector by rubbing the cotton swab over the surface using
a small circular movement.
3 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4 Blow away any remaining lint with compressed air.
94Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ?Cleaning Information
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an
ultrasonic bath with isopropyl alcohol.
1 Hold the tip of the connector in the bath for at least three minutes.
2 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
3 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
95
Cleaning InformationWhich Cleaning Procedure should I use ?
How to clean connector adapters
CAUTION
Some adapters have an anti-reflection coating on the back to reduce
back reflection. This coating is extremely sensitive to solvents and
mechanical abrasion. Extra care is needed when cleaning these
adapters.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the adapter by rubbing a new, dry cotton swab over the surface
using a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the adapter:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the adapter by rubbing the cotton swab over the surface using a
small circular movement.
3 Take a new, dry soft tissue and remove the alcohol, dissolved sediment
and dust, by rubbing gently over the surface using a small circular
movement.
4 Blow away any remaining lint with compressed air.
96Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ?Cleaning Information
How to clean connector interfaces
CAUTION
Be careful when using pipe cleaners, as the core and the bristles of
the pipe cleaner are hard and can damage the interface.
Do not use pipe cleaners on optical head adapters, as the hard core
of normal pipe cleaners can damage the bottom of an adapter.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the interface by pushing and pulling a new, dry pipe cleaner into
the opening. Rotate the pipe cleaner slowly as you do this.
2 Then clean the interface by rubbing a new, dry cotton swab over the
surface using a small circular movement.
3 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the interface:
1 Moisten a new pipe cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe cleaner into the
opening. Rotate the pipe cleaner slowly as you do this.
3 Moisten a new cotton swab with isopropyl alcohol.
4 Clean the interface by rubbing the cotton swab over the surface using a
small circular movement.
5 Using a new, dry pipe cleaner, and a new, dry cotton swab remove the
alcohol, any dissolved sediment and dust.
6 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
97
Cleaning InformationWhich Cleaning Procedure should I use ?
How to clean bare fiber adapters
Bare fiber adapters are difficult to clean. Protect from dust unless they are
in use.
CAUTION
CAUTION
Never use any kind of solvent when cleaning a bare fiber adapter as
solvents can:
• Damage the foam inside some adapters.
• Deposit dissolved dirt in the groove, which can then dirty the surface of
an inserted fiber.
Preferred Procedure
Use the following procedure on most occasions.
1 Blow away any dust or dirt with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the adapter:
1 Clean the adapter by pushing and pulling a new, dry pipe cleaner into
the opening. Rotate the pipe cleaner slowly as you do this.
Be careful when using pipe cleaners, as the core and the bristles of
the pipe cleaner are hard and can damage the adapter.
2 Clean the adapter by rubbing a new, dry cotton swab over the surface
using a small circular movement.
3 Blow away any remaining lint with compressed air.
98Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
Which Cleaning Procedure should I use ?Cleaning Information
How to clean lenses
Some lenses have special coatings that are sensitive to solvents, grease,
liquid and mechanical abrasion. Take extra care when cleaning lenses with
these coatings.
Lens assemblies consisting of several lenses are not normally sealed.
Therefore, use as little alcohol as possible, as it can get between the
lenses and in doing so can change the properties of projection.
Preferred Procedure
Use the following procedure on most occasions.
1 Clean the lens by rubbing a new, dry cotton swab over the surface using
a small circular movement.
2 Blow away any remaining lint with compressed air.
Procedure for Stubborn Dirt
Use this procedure when there is greasy dirt on the lens:
1 Moisten a new cotton swab with isopropyl alcohol.
2 Clean the lens by rubbing the cotton swab over the surface using a
small circular movement.
3 Using a new, dry cotton swab remove the alcohol, any dissolved
sediment and dust.
4 Blow away any remaining lint with compressed air.
Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition
99
Cleaning InformationWhich Cleaning Procedure should I use ?
How to clean instruments with a fixed
connector interface
You should only clean instruments with a fixed connector interface when it
is absolutely necessary. This is because it is difficult to remove any used
alcohol or filaments from the input of the optical block.
It is important, therefore, to keep dust caps on the equipment at all times,
except when your optical device is in use.
If you do discover filaments or particles, the only way to clean a fixed
connector interface and the input of the optical block is to use compressed
air.
If there are fluids or fat in the connector, please refer the instrument to the
skilled personnel of Agilent’s service team.
CAUTION
Only use clean, dry compressed air. Make sure that the air is free of
dust, water, and oil. If the air that you use is not clean and dry, this
can lead to filmy deposits or scratches on the surface of your
connector interface. This will degrade the performance of your
transmission system.
Never try to open the instrument and clean the optical block by
yourself, because it is easy to scratch optical components, and
cause them to become misaligned.
How to clean instruments with an optical
glass plate
Some instruments, for example, the optical heads from Agilent
Technologies have an optical glass plate to protect the sensor. Clean this
glass plate in the same way as optical lenses (see “How to clean lenses” on page 99).
100Agilent 81618A/9A and Agilent 81623B/4B/6B/8B Optical Heads, Fourth Edition