Use the tests in this chapter if you want to check that the Agilent
81110A Pulse Generator Frame with the Agilent 81111A 165
MHz Output Channel(s) is working correctly. Before starting any
testing allow all test equipment to warm up for at least 30 minutes.
Conventions Used
When referring to actions that you perform during the tests, the
following conventions are used:
FUNCTION This indicates that a labelled button must be pressed
[MODE/TRG] This shows that a soft-key must be pressed. A softkey is an unlabelled button whose label is shown on the display,
and which can vary according to the job that the button is doing
CONTINUOUS PULSES This is an option shown on the display, and is selected by use of the vernier keys. It is shown in
upper or lower case to match the case displayed.
Test Results Tables
Tables for entering the results of the tests are included at the end
of this chapter. The tests are numbered and reference numbers for
each Test Result (TR) are given in a small table at the end of each
test. The reference number shows you where the actual results
should be entered in the Test Results Tables.
The Test Results tables at the end of the chapter should be photocopied, and the Test Results entered on the copies. Then, if the
tests need to be repeated, the tables can be copied again.
2Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
If Channel 2 has been fitted to your instrument, make an extra
copy of the Test Results tables for entry of the results of tests on
that channel. In this case, however, it is not necessary to repeat the
Period tests, as these are common to both channels.
Agilent 81110A/’11A Performance Test3
Agilent 81110A/’11A Performance Test
Recommended Test Equipment and Accessories
The following tables list the recommended test equipment you
need to perform all the tests in this chapter. You can use alternative instruments if they meet the critical specifications given. The
test set-ups and procedures assume you are using the recommended equipment.
Test EquipmentModelCritical Specifications
Oscilloscope
or
OscilloscopeAgilent 54750A +
Counter
or
CounterAgilent 53132A
Digital VoltmeterAgilent 3458ADCV up to 20 V
Pulse GeneratorAgilent 8110Aup to 150 MHz
Delay lineAgilent 54008A22 ns
Digitizing Oscilloscopes Accessories
Attenuators
Power Splitter
SMA/SMA (m-m)Adapter
SMA/BNC Adapter
SMA Cable
Agilent 54121T20 GHz, 10 bit vertical resolution, Histogram
Agilent 54751A
Agilent 5334B
#010, 030
#001/010, 030
AccessoriesModelCritical Specifications
20 GHz, 15 bit vertical resolution, Histogram
Period and Time Interval measurements
Oven Osci, 1.3 GHz C-Channel
Frequency measurements > 150 MHz
High-Stability Timebase, 3 GHz Channel
NOTE:When you connect the test equipment for the first time, and
whenever you change the setup during the course of these tests,
use the 8710 - 1582 torque wrench to tighten and loosen SMA
connectors. This will ensure that the connectors are at the correct
tightness and give the best signal transfer.
Agilent 81110A/’11A Performance Test5
Agilent 81110A/’11A Performance Test
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following figure provides a schematic and a parts list except
for the case. The case must provide shielding and maintain
grounding integrity.
The Agilent 81110A is controlled by selecting options in a series
of pages that are displayed on the instrument's screen. These
options vary with the boards that are fitted in the instrument.
When the Agilent 81110A is being tested, therefore, different situations can arise, depending on whether you have a standard instrument or one that has had additional boards fitted. The following
examples illustrate this
Typical Examples of Displayed Screens
Agilent 81110A/’11A Performance Test
Per 1.000µs Normal
OFF
1
Delay
Width
LeadE
TraiE
The OUTPUT Screen in a Standard Agilent 81110A
Agilent 81110A/’11A Performance Test7
0ps
100.0ns
2.00ns
=LeadE
OUTPUTPATTERNLIMITSMODE/TRG
Offset
Amplit
50Ω into 50.0
+0.0mV
1.00V
*OFF
ON
Ω
MODIFY
Agilent 81110A/’11A Performance Test
OFF 1.000µs OFF
1
Delay
Width
LeadE
TraiE
The TIMING Screen in an Agilent 81110A with qty 2 of Agilent
81111A
ON Normal Normal ON
Seperate Outputs
High
Low
50Ω into 50.0
Per
0ps
100.0ns
2.00ns
=LeadE
TIMINGPATTERNLEVELSMODE/TRG
+2.50V
+0.0mV
Ω
2
Delay 0ps
Width
LeadE
TraiE
100.0ns
2.00ns
=LeadE
21
High+2.50V
Low +0.0mV
50Ω into 50.0
Ω
MODIFY
*Period
Frequency
MODIFY
Set TTL
*High-Low
Offs-Ampl
Set ECL
TIMINGPATTERNLEVELSMODE/TRG
The LEVELS Screen in an Agilent 81110A with qty 2 of Agilent
81111A
8Agilent 81110A/’11A Performance Test
Instrument Serial Numbers
You will need to write the serial numbers of the instrument at the
top of the Test Reports. These can be found as follows:
Press HELP, [SERIAL #]
The Agilent 81110A display lists the instrument's products and
serial number.
The display on your instrument should look similar to this:
Agilent 81110A/’11A Performance Test
FRAME :
Serial No :
81110A 165 MHz
DE38700135
OUTPUTS
Ch1-Bd. :
Ch2-Bd. :
81111A
81111A
The serial number given for the FRAME applies to the Mainframe,
the Power Supply, the Microprocessor Board, and the Timing
Board. The number(s) available of the Output Channel(s) applies
to the installed numbers of outputs and Model Number.
Agilent 81110A/’11A Performance Test9
Agilent 81110A/’11A Performance Test
Initial Setup of the Agilent 81110A
In the majority of these tests the initial setting up of the instrument
is identical. Therefore, it is described once here, and then referredto where appropriate. In cases where the initial setup differs, an
illustration of the settings is shown.
Set up the Agilent 81110A as follows:
1. Select [MODE/TRG]
• CONTINUOUS PULSES
•Single-Pulses at Out 1 (plus Single-Pulses at Out 2, if
second channel is installed
• Pulse-Period:internal Osc
2. If a second output channel is installed, select MORE [CONFIG]
screen and set up as follows:
GPIB Address: 10
Perform: Timing Calibration
Group Params by: OUTPUT 1 / 2
PLL-Ref : Internal
TRG-LEVCONFIGMEMCARDLIMITS
CONFIG Screen, Parameters grouped by OUTPUT
10Agilent 81110A/’11A Performance Test
MODIFY
Tim/Lev
*Out 1/2
Agilent 81110A/’11A Performance Test
NOTE:Set-ups are given in all the tests for [OUTPUT 1] and [OUTPUT 2].
If you are testing a single channel instrument set up the [OUTPUT]
screen with the settings given for [OUTPUT 1].
Agilent 81110A/’11A Performance Test11
Agilent 81110A/’11A Performance Test
Test 1: Period (PLL not active)
Test Specifications
Range 6.06 ns to 999.5 s
Resolution 3.5 digits, best case 5 ps
Accuracy +3%
typical +0.5% after selfcal
Equipment Needed
Counter
Cable, 50 Ω, coaxial, BNC
Procedure
1. Connect the Agilent 81110A to the Counter as shown:
81110A UNDER TEST
81110A
165/330 MHz
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
5334B Counter
Connecting the Agilent 81110A to the Counter
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
12Agilent 81110A/’11A Performance Test
INPUT AC
Agilent 81110A/’11A Performance Test
On the Agilent 81110A press MORE and set up [OUTPUT 1] and
[OUTPUT 2] pages as shown in the following illustrations:
Per Normal ON
Delay
DtyCyc
LeadE
TraiE
Configuring Output 1
Per Normal OFF
Delay2 0ps
DtyCyc
LeadE
TraiE
6.060ns
0ps
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
6.060ns
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50Ω into 50.0
Offset
Amplit
50Ω into 50.0
Separate Out2
1
+0.0mV
1.00V
Ω
+0.0mV
1.00V
Ω
MODIFY
6.060
ns
MODIFY
6.060
ns
Configuring Output 2
Agilent 81110A/’11A Performance Test13
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. For the Period test you can switch OFF the channels that are not
being tested.
3. Set the Counter to:
FUNCTIONPeriod A / Freq C
INPUT A
SENSE
50 ΩΟn
4. Check the Agilent 81110A period at the following settings:
PeriodAcceptable RangeTR entry
6.060 ns
9.990 ns
10.00 ns
50.00 ns
99.90 ns
100 ns
500 ns
1 µs500 µs
500 ms
without selfcal!
5.878 ns to 6.242 ns
9.690 ns to 10.290 ns
9.7 ns to 10.3 ns
48.5 ns to 51.5 ns
96.903 ns to 102.897 ns
97 ns to 103 ns
485 ns to 515 ns
970 ns to 1030 ns
485µs to 515 µs
485 ms to 515 ms
1 - 1
1 - 2
1 - 3
1 - 4
1 - 5
1 - 6
1 - 7
1 - 8
1 - 9
1 - 10
14Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
5. To perform the Timing Calibration (shown as selfcal) set up
[CONFIG] page as shown in the following illustration:
HP-IB Address: 10
Perform: Timing Calibration
Group Params by: OUTPUT 1 / 2
PLL-Ref : Internal
TRG-LEVCONFIGMEMCARDLIMITS
Selftest
*Calibrate
6. Press ENTER and wait till the display shows
TIMING CALIBRATION PASSED
and gets back to the above shown display .
Test 2: PLL Period
NOTE:This test is only performed if PLL is switched on.
Test Specifications
Range 6.06 ns to 999.5 s
Resolution 4 digits, best case 1 ps
Accuracy + 0.01%
MODIFY
Equipment Needed
Counter Agilent 53132A
Agilent 81110A/’11A Performance Test15
Agilent 81110A/’11A Performance Test
Cable, 50 Ω, coaxial, BNC
NOTE:The Agilent 53132A counter is used in frequency mode to meet
the MIL CAL A uncertainty requirements for TAR (Test
Accuracy Ratio) > 4:1.
Procedure
Connect the Agilent 81110A to the counter as follows:
81110A UNDER TEST
165/330 MHz
81110A
S
Agilent
PULSE-/ PATTERN GENERATOR
TRIG OUT
53132A Counter
3
CHANNEL 12
Connecting Agilent 81110A to the Counter
7. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
8. Select the [MODE/TRG] screen on the Agilent 81110A and set
up as follows:
16Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
CONTINUOUS PULSES
MODIFY
Single-Pulses at Out1
Single-Pulses at Out2
Pulse-Period: internal PLL
int. OSC
*int. PLL
CLK-IN
TIMINGPATTERNLEVELSMODE/TRG
The MODE/TRG Screen Setup
9. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the test before!
Agilent 81110A/’11A Performance Test17
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. For the Period test you can switch OFF the channels that are not
being tested.
10. Set the Counter to measure the frequency at the choosen input
1 / 3
11. Check the Agilent 81110A PLL pulse period at the following
settings:
14. The RMS-jitter for period of 500 ns is 65 ps. Enter the result in
the Test Report as TR entry 6.1a - 2
Agilent 81110A/’11A Performance Test43
Agilent 81110A/’11A Performance Test
Test 6.1b: Period Jitter, Internal PLL
Test Specifications
RMS-Jitter 0.001% + 15 ps
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
Cable, 50 Ω, coaxial, BNC
Cable, SMA
Procedure
1. Connect Agilent 81110A to the Scope as shown.
54750A + 54751A
81110A UNDER TEST
81110A
165/330 MHz
S
Agilent
PULSE-/ PATTERN GENERATOR
6 dB Attenuator
54008A Delay Line
INPUT
BNC/SMA
Adapter
OUT 1
SMA Cable
Equipment Set-up for Jitter Test using the Agilent 54750A +
54751A
Using the Agilent 54121T the Set-up is the same as before.
44Agilent 81110A/’11A Performance Test
OUTPUT
POWER SPLITTER
11667B
TRIG
SMA Cable
Agilent 81110A/’11A Performance Test
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
3. Select the [MODE/TRG] screen on the Agilent 81110A and set
up as follows:
CONTINUOUS PULSES
Single-Pulses at Out1
Single-Pulses at Out2
Pulse-Period: internal PLL
TIMINGPATTERNLEVELSMODE/TRG
The TRG MODE Screen Setup
4. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the following illustrations:
Per Normal ON
20.00ns
1
Delay
Width
LeadE
TraiE
0ps
10.00ns
2.00ns
=LeadE
Offset
Amplit
50Ω into 50.0
+500mV
1.00V
Ω
MODIFY
int. OSC
*int. PLL
CLK-IN
MODIFY
20.00
ns
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 1
Agilent 81110A/’11A Performance Test45
Agilent 81110A/’11A Performance Test
Per Normal OFF
Delay
Width
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
20.00ns
2
0ps
10.00ns
2.00ns
=LeadE
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the
other channel.
Offset
Amplit
50Ω into 50.0
Separate Out2
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
+500mV
1.00V
Ω
MODIFY
20.00
ns
5. Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Select the Display menu and set the Number of Averages to 64
•Select the Channel menu and set the Attenuation factor of channel
2 to 2
46Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
•Set the VOLTS/DIV of channel 2 to 10 mV/div
•Set OFFSET to 500 mV
•Select the Timebase menu and set the TIME/DIV to 100 ps/div
•Center the first positive-going edge of the signal (approximate Delay
= 29 ns)
•Select the Delta V menu and turn the V markers On
•Set the Marker 1 Position to 490 mV and the Marker 2 Position
to 500 mV
•Select the Delta t menu and turn the T Markers On
•Set START ON EDGE = POS1 and STOP ON EDGE = POS1
•Press the PRECISE EDGE FIND key
6. RECORD the delta t reading. This is the rise time of the refer-
ence signal within a 1% amplitude window of the signal connected to Input 2. This value is needed later to calculate the
correct jitter. (delta.t.up)
7. Select the Timebase menu and center the second positive-
going edge of the signal (approximate Delay = 49 ns)
8. Press MORE and HISTOGRAM
•Select the Window submenu and set:
•Source is channel 2
•Choose the Time Histogram
•Press WINDOW MARKER 1 and set it to 490 mV
•Press WINDOW MARKER 2 and set it to 500 mV
Agilent 81110A/’11A Performance Test47
Agilent 81110A/’11A Performance Test
9. Select the Acquire submenu, set the Number of Samples to
1000 and press START ACQUIRING
10. After the data for the time histogram has been acquired (#
Samples = 100%), select the Result submenu.
11. Press MEAN and SIGMA. RECORD the values of sigma
12. The RMS-jitter is calculated as follows:
RMS - jitter =
6sigma-delta.t.up
6
13. The RMS-jitter for period of 20 ns is 15.2 ps. Enter the result
in the Test Report as TR entry 6.1b - 1
NOTE:See the Agilent54750A User’s Guide / Service Guide to get the
info needed to do the Jitter Test using this scope.
48Agilent 81110A/’11A Performance Test
Test 6.2: Width Jitter (PLL not active)
Test Specifications
RMS-Jitter 0.01% + 15 ps
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
Cable, 50 Ω, coaxial, BNC
Cable, SMA
Procedure
1. Connect Agilent 81110A to the Scope as shown:
Agilent 81110A/’11A Performance Test
81110A UNDER TEST
165/330 MHz
81110A
S
Agilent
PULSE-/ PATTERN GENERATOR
BNC - SMA
Adapter
OUT 1
INPUT 1
54121T Frontend
SMA - SMA
Adapter
SMA Cable
2
SMA Cable
54008A Delay Line
4
3
INPUT
TRIG
POWER SPLITTER
OUTPUT
11667B
Equipment Set-up for Jitter Test
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
Agilent 81110A/’11A Performance Test49
Agilent 81110A/’11A Performance Test
3. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the following illustrations:
Per 1.000µs Normal ON
Delay
Width
LeadE
TraiE
Configuring Output Screen 1
Per 1.000µs Normal OFF
Delay
Width
LeadE
TraiE
0ps
3.030ns
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
0ps
3.030ns
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50Ω into 50.0
Offset
Amplit
50Ω into 50.0
Separate Out2
+500mV
1.00V
1
+500mV
1.00V
Ω
2
Ω
MODIFY
3.030
ns
MODIFY
3.030
ns
Configuring Output Screen 2
50Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the
other channel.
Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Select the Display menu and set the Number of Averages to 128
•Select the Channel menu and set the Attenuation factor of channel 2
to 2
•Set the VOLTS/DIV of channel 2 to 10 mV/div
•Set OFFSET to 500 mV
•Select the Timebase menu and set the TIME/DIV to 10 ps/div
•Center the first negative-going edge of the signal (approximate Delay
= 33.8 ns)
•Select the Delta V menu and turn the V markers On
•Set the Marker 1 Position to 500 mV and the Marker 2 Position to
490 mV
•Select the Delta t menu and turn the T Markers On
•Set START ON EDGE = NEG1 and STOP ON EDGE = NEG1
•Press the PRECISE EDGE FIND key
Agilent 81110A/’11A Performance Test51
Agilent 81110A/’11A Performance Test
4. RECORD the delta t reading. This is the fall time of the referencesignal within a 1% amplitude window of the signal connected to Input 2. This value isneeded later to calculate the
correct jitter. (delta.t.dn)
5. Set the Agilent 81110A Pulse Width to 50 ns
6. Select the Timebase menu and center the first negative-going
edge of the signal (approximate Delay = 80.5 ns)
7. Press MORE and HISTOGRAM
8. Select the Window submenu and set:
•Source is channel 2
•Choose the Time Histogram
•Press WINDOW MARKER 1 and set it to 500 mV
•Press WINDOW MARKER 2 and set it to 490 mV
9. Select the Acquire submenu, set the Number of Samples to
1000 and press START ACQUIRING
10. After the data for the time histogram has been acquired (#
Samples = 100%), select the Result submenu.
11. Press MEAN and SIGMA. RECORD the value of sigma
12. The RMS-jitter is calculated as follows:
52Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
RMS - jitter =
6 sigma - delta.t.dn
6
13. The RMS-jitter for pulse width of 50 ns is 20 ps. Enter the
result in the Test Report as TR entry 6.2 - 1
14. Set the Agilent 81110A for pulse width of 500ns
When you connect the test equipment the first time, and whenever
you change the setup during the following tests, use the torque
wrench (8170-1582) to tighten and loosen the SMA connectors.
This will ensure that the connectors are at the correct tightness and
give the best signal transfer!
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
TRIG OUT
BNC - SMA Adaptor
OUT 1
Cable, SMA
54121T Frontend
INPUT 1234TRIG
20dB
Attenuator
3. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Agilent 81110A/’11A Performance Test69
Agilent 81110A/’11A Performance Test
Per 500.0µs Normal ON
Delay
DtyCyc
LeadE
TraiE
Configuring Output Screen 1
Per 500.0µs Normal OFF
Delay
DtyCyc
LeadE
TraiE
0ps
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
0ps
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50Ω into 50.0
Offset
Amplit
50Ω into 50.0
Separate Out2
+0.0mV
5.00V
1
Ω
2
+0.0mV
5.00V
Ω
MODIFY
2.00
ns
MODIFY
2.00
ns
Configuring Output Screen 2
NOTE:
When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested
70Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
If you then test the other channel:
c. Switch ON the channel you are testing, and switch OFF the
other channel.
4. Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Center one pulse on screen, e.g.:
•TIME/DIV = 50 µs/div, DELAY = 380 µs,
•Select the Display menu and set the Number of Averages to 32
•Select the Channel menu and set the Attenuation factor to 10
•Select the Delta V menu and turn the voltage markers On
•Set the Preset Levels = 10-90% and press AUTO LEVEL SET
•Select the Timebase menu and set TIME/DIV = 1 ns/div, DELAY =
19.5 ns
•Select the Delta t menu and turn the markers On
•Set START ON EDGE = POS1 and STOP ON EDGE = POS1
5. Set period of Agilent 81110A to: Period = 1 µs and
change the Agilent 81110A Delay to center the leading edge
of the first pulse on the screen
6. After the averaging, while the oscilloscope is in the Delta t
menu, Press the PRECISE EDGE FIND key
Agilent 81110A/’11A Performance Test71
Agilent 81110A/’11A Performance Test
7. Check the Agilent 81110A rise times at the following leading
edge settings:
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
1-1 6.06ns 5.878 ns ______ 6.242 ns ____ ____
1-2 9.99ns 9.690 ns ______ 10.290 ns ____ ____
1-3 10.0ns 9.7 ns ______ 10.3 ns ____ ____
1-4 50.0ns 48.5 ns ______ 51.5 ns ____ ____
1-5 99.9ns 96.903 ns ______ 102.897 ns ____ ____
Agilent 81110A/’11A Performance Test83
Agilent 81110A/’11A Performance Test
Counter Uncertainty factor ___________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
PLL Period
(Results measured as frequency by counter)
Counter Uncertainty factor ____________
_____________________________________________________
TR Test Limit Actual Limit Pass Fail
Entry Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.1a-1 50 ns ______ 20 ps ____ ____
6.1a-2 500 ns ______ 65 ps ____ ____
6.1b-1 20 ns _______ 15.2 ps ____ ____
86Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
Test Results for Agilent 81111A Output Channel ______
Serial No. ________________
Width
Scope Uncertainty factor ________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.2-1 50 ns ________ 20 ps ____ ____
6.2-2 500 ns ________ 65 ps ____ ____
88Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
Delay
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.3-1 50 ns ________ 20 ps ____ ____
6.3-2 500 ns ________ 65 ps ____ ____
90Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
Double Pulse Delay
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
5-1 6.06 ns 5.628 ns _______ 6.392 ns ____ ____
5-2 10.0 ns 9.550 ns _______ 10.45 ns ____ ____
5-3 50.0ns 48.35 ns _______ 51.65 ns ____ ____
5-4 100ns 96.85 ns _______ 103.15 ns ____ ____
Counter Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
5-5 500 ms 485 ms _______ 515 ms ____ ____
5-6 1 s 970.0 ms ________ 1030.0 ms ____ ____
Agilent 81110A/’11A Performance Test91
Agilent 81110A/’11A Performance Test
High Level 50Ω-50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.1-1 10.0 V 9.85 V __________ 10.15 V ____ ____
7.1-2 5.0 V 4.90 V __________ 5.10 V ____ ____
7.1-3 3.0V 2.92 V __________ 3.08 V ____ ____
7.1-4 1.0 V 0.94 V __________ 1.06 V ____ ____
7.1-5 0.5 V 445 mV __________ 555 mV ____ ____
7.1-6 0.1 V 49 mV __________ 151 mV ____ ____
High Level 1KΩ−50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.3-1 19.0 V 18.71V __________ 19.29 V ____ ____
7.3-2 10.0 V 9.80 V __________ 10.20 V ____ ____
7.3-3 5.0 V 4.85 V __________ 5.15 V ____ ____
7.3-4 1.0 V 0.89 V __________ 1.11V ____ ____
7.3-5 0.2 V 98 mV __________ 302mV ____ ____
92Agilent 81110A/’11A Performance Test
Agilent 81110A/’11A Performance Test
Low Level 50Ω-50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.4-1 -0.2V -98 mV __________-302 mV ____ ____
7.4-2 -1.0V -0.89 V __________ -1.11 V ____ ____
7.4-3 -5.0V -4.85V __________ -5.15 V ____ ____
7.4-4 -10.0V -9.80 V _________ -10.20 V ____ ____
7.4-5 -19.0V -18.71 V __________-19.29 V ____ ____
Agilent 81110A/’11A Performance Test93
Agilent 81110A/’11A Performance Test
Leading Edge
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
8.1b-1 2.0 ns <2 ns __________ 2.4 ns ____ ____
8.1b-2 10 ns 8.8 ns __________ 11.2 ns ____ ____
8.1b-3 50 ns 44.8 ns __________ 55.2ns ____ ____
8.1b-4 500 ns 449.8 n __________ 550.2 ns ____ ____
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
9-1 5V __________ + 5% of ampl.____ ____
+20mV
9-2 500 mV _________ +5% of ampl. ____ ____
+20mV
Preshoot
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
9-3 0 V _________ +5% of ampl. ____ ____
+20mV
96Agilent 81110A/’11A Performance Test
Publication Number: 5988-4853EN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.