Use the tests in this chapter if you want to check that the Agilent
81110A Pulse Generator Frame with the Agilent 81111A 165
MHz Output Channel(s) is working correctly. Before starting any
testing allow all test equipment to warm up for at least 30 minutes.
Conventions Used
When referring to actions that you perform during the tests, the
following conventions are used:
FUNCTION This indicates that a labelled button must be pressed
[MODE/TRG] This shows that a soft-key must be pressed. A softkey is an unlabelled button whose label is shown on the display,
and which can vary according to the job that the button is doing
CONTINUOUS PULSESThis is an option shown on the display, and is selected by use of the vernier keys. It is shown in
upper or lower case to match the case displayed.
Test Results Tables
Tables for entering the results of the tests are included at the end
of this chapter. The tests are numbered and reference numbers for
each Test Result (TR) are given in a small table at the end of each
test. The reference number shows you where the actual results
should be entered in the Test Results Tables.
The Test Results tables at the end of the chapter should be photocopied, and the Test Results entered on the copies. Then, if the
tests need to be repeated, the tables can be copied again.
2Agilent 81110A/’11A Performance Test
Page 3
Agilent 81110A/’11A Performance Test
If Channel 2 has been fitted to your instrument, make an extra
copy of the Test Results tables for entry of the results of tests on
that channel. In this case, however, it is not necessary to repeat the
Period tests, as these are common to both channels.
Agilent 81110A/’11A Performance Test3
Page 4
Agilent 81110A/’11A Performance Test
Recommended Test Equipment and Accessories
The following tables list the recommended test equipment you
need to perform all the tests in this chapter. You can use alternative
instruments if they meet the critical specifications given. The test
set-ups and procedures assume you are using the recommended
equipment.
Test EquipmentModelCritical Specifications
Oscilloscope
or
OscilloscopeAgilent 54750A +
Counter
or
CounterAgilent 53132A
Digital VoltmeterAgilent 3458ADCV up to 20 V
Pulse GeneratorAgilent 8110Aup to 150 MHz
Delay lineAgilent 54008A22 ns
Digitizing Oscilloscopes Accessories
Attenuators
Power Splitter
SMA/SMA (m-m)Adapter
SMA/BNC Adapter
SMA Cable
Agilent 54121T20 GHz, 10 bit vertical resolution, Histogram
Agilent 54751A
Agilent 5334B
#010, 030
#001/010, 030
AccessoriesModelCritical Specifications
20 GHz, 15 bit vertical resolution, Histogram
Period and Time Interval measurements
Oven Osci, 1.3 GHz C-Channel
Frequency measurements > 150 MHz
High-Stability Timebase, 3 GHz Channel
NOTE:When you connect the test equipment for the first time, and
whenever you change the setup during the course of these tests,
use the 8710 - 1582 torque wrench to tighten and loosen SMA
connectors. This will ensure that the connectors are at the correct
tightness and give the best signal transfer.
Agilent 81110A/’11A Performance Test5
Page 6
Agilent 81110A/’11A Performance Test
50 Ohm, 0.1%, 10 W Feedthrough Termination
The following figure provides a schematic and a parts list except
for the case. The case must provide shielding and maintain
grounding integrity.
The Agilent 81110A is controlled by selecting options in a series
of pages that are displayed on the instrument's screen. These
options vary with the boards that are fitted in the instrument.
When the Agilent 81110A is being tested, therefore, different situations can arise, depending on whether you have a standard instrument or one that has had additional boards fitted. The following
examples illustrate this
Typical Examples of Displayed Screens
Agilent 81110A/’11A Performance Test
Per 1.000
µµµµ
s Normal
OFF
1
Delay
Width
LeadE
TraiE
The OUTPUT Screen in a Standard Agilent 81110A
Agilent 81110A/’11A Performance Test7
0ps
100.0ns
2.00ns
=LeadE
OUTPUTPATTERNLIMITSMODE/TRG
Offset
Amplit
50
ΩΩΩΩ
into 50.0
+0.0mV
1.00V
ΩΩΩΩ
*OFF
ON
MODIFY
Page 8
Agilent 81110A/’11A Performance Test
OFF1.000
Per
1
Delay
Width
LeadE
TraiE
The TIMING Screen in an Agilent 81110A with qty 2 of Agilent
81111A
ONNormalNormalON
Seperate Outputs
High
Low
50
ΩΩΩΩ
+2.50V
into 50.0
0ps
100.0ns
2.00ns
=LeadE
TIMINGPATTERNLEVELSMODE/TRG
+0.0mV
ΩΩΩΩ
High+2.50V
Low+0.0mV
50
µµµµ
sOFF
2
Delay0ps
Width
LeadE
TraiE
100.0ns
2.00ns
=LeadE
21
ΩΩΩΩ
into 50.0
ΩΩΩΩ
MODIFY
*Period
Frequency
MODIFY
Set TTL
*High-Low
Offs-Ampl
Set ECL
TIMINGPATTERNLEVELSMODE/TRG
The LEVELS Screen in an Agilent 81110A with qty 2 of Agilent
81111A
8Agilent 81110A/’11A Performance Test
Page 9
Instrument Serial Numbers
You will need to write the serial numbers of the instrument at the
top of the Test Reports. These can be found as follows:
Press HELP, [SERIAL #]
The Agilent 81110A display lists the instrument's products and
serial number.
The display on your instrument should look similar to this:
Agilent 81110A/’11A Performance Test
FRAME :
Serial No :
81110A 165 MHz
DE38700135
OUTPUTS
Ch1-Bd. :
Ch2-Bd. :
81111A
81111A
The serial number given for the FRAME applies to the Mainframe,
the Power Supply, the Microprocessor Board, and the Timing
Board. The number(s) available of the Output Channel(s) applies
to the installed numbers of outputs and Model Number.
Agilent 81110A/’11A Performance Test9
Page 10
Agilent 81110A/’11A Performance Test
Initial Setup of the Agilent 81110A
In the majority of these tests the initial setting up of the instrument
is identical. Therefore, it is described once here, and then referredto where appropriate. In cases where the initial setup differs, an
illustration of the settings is shown.
Set up the Agilent 81110A as follows:
1. Select [MODE/TRG]
• CONTINUOUS PULSES
•Single-Pulses at Out 1 (plus Single-Pulses at Out 2, if
second channel is installed
• Pulse-Period:internal Osc
2. If a second output channel is installed, select MORE [CONFIG]
screen and set up as follows:
GPIB Address:10
Perform: Timing Calibration
Group Params by:OUTPUT1/2
PLL-Ref : Internal
TRG-LEVCONFIGMEMCARDLIMITS
CONFIG Screen, Parameters grouped by OUTPUT
10Agilent 81110A/’11A Performance Test
MODIFY
Tim/Lev
*Out 1/2
Page 11
Agilent 81110A/’11A Performance Test
NOTE:Set-ups are given in all the tests for [OUTPUT 1] and [OUTPUT 2].
If you are testing a single channel instrument set up the [
OUTPUT]
screen with the settings given for [OUTPUT 1].
Agilent 81110A/’11A Performance Test11
Page 12
Agilent 81110A/’11A Performance Test
Test 1: Period (PLL not active)
Test Specifications
Range 6.06 ns to 999.5 s
Resolution 3.5 digits, best case 5 ps
Accuracy +3%
typical +0.5% after selfcal
Equipment Needed
Counter
Cable, 50 Ω, coaxial, BNC
Procedure
1. Connect the Agilent 81110A to the Counter as shown:
81110A UNDER TEST
81110A
165/330 MHz
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
TRIG OUT
5334B Counter
Connecting the Agilent 81110A to the Counter
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
12Agilent 81110A/’11A Performance Test
INPUT AC
Page 13
Agilent 81110A/’11A Performance Test
On the Agilent 81110A press MORE and set up [OUTPUT 1] and
[OUTPUT 2] pages as shown in the following illustrations:
PerNormalON
6.060ns
1
MODIFY
Delay
DtyCyc
LeadE
TraiE
Configuring Output 1
PerNormalOFF
6.060ns
Delay
DtyCyc
LeadE
TraiE
0ps
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
0ps
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50
ΩΩΩΩ
into 50.0
Offset
Amplit
50
ΩΩΩΩ
into 50.0
Separate Out2
+0.0mV
1.00V
+0.0mV
1.00V
6.060
ΩΩΩΩ
2
6.060
ΩΩΩΩ
ns
MODIFY
ns
Configuring Output 2
Agilent 81110A/’11A Performance Test13
Page 14
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. For the Period test you can switch OFF the channels that are not
being tested.
3. Set the Counter to:
FUNCTIONPeriod A / Freq C
INPUT A
SENSE
50 ΩΟn
4. Check the Agilent 81110A period at the following settings:
PeriodAcceptable RangeTR entry
6.060 ns
9.990 ns
10.00 ns
50.00 ns
99.90 ns
100 ns
500 ns
1 µs
500 µs
500 ms
without selfcal!
5.878 ns to 6.242 ns
9.690 ns to 10.290 ns
9.7 ns to 10.3 ns
48.5 ns to 51.5 ns
96.903 ns to 102.897 ns
97 ns to 103 ns
485 ns to 515 ns
970 ns to 1030 ns
485µs to 515 µs
485 ms to 515 ms
1 - 1
1 - 2
1 - 3
1 - 4
1 - 5
1 - 6
1 - 7
1 - 8
1 - 9
1 - 10
14Agilent 81110A/’11A Performance Test
Page 15
Agilent 81110A/’11A Performance Test
5. To perform the Timing Calibration (shown as selfcal) set up
[CONFIG] page as shown in the following illustration:
HP-IB Address:10
Perform: Timing Calibration
Group Params by:OUTPUT1/2
PLL-Ref : Internal
TRG-LEVCONFIGMEMCARDLIMITS
*Calibrate
6. Press ENTER and wait till the display shows
TIMING CALIBRATION PASSED
and gets back to the above shown display .
Test 2: PLL Period
NOTE:This test is only performed if PLL is switched on.
Test Specifications
Range 6.06 ns to 999.5 s
Resolution 3.5 digits, best case 5 ps
Accuracy + 0.01%
MODIFY
Selftest
Equipment Needed
Counter Agilent 53132A
Agilent 81110A/’11A Performance Test15
Page 16
Agilent 81110A/’11A Performance Test
Cable, 50 Ω, coaxial, BNC
NOTE:The Agilent 53132A counter is used in frequency mode to meet
the MIL CAL A uncertainty requirements for TAR (Test
Accuracy Ratio) > 4:1.
Procedure
Connect the Agilent 81110A to the counter as follows:
81110A UNDER TEST
165/330 MHz
81110A
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
TRIG OUT
53132A Counter
3
CHANNEL 12
Connecting Agilent 81110A to the Counter
7. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
8. Select the [MODE/TRG] screen on the Agilent 81110A and set
up as follows:
16Agilent 81110A/’11A Performance Test
Page 17
Agilent 81110A/’11A Performance Test
CONTINUOUS PULSES
MODIFY
Single-Pulses at Out1
Single-Pulses at Out2
Pulse-Period: internal PLL
int. OSC
*int. PLL
CLK-IN
TIMINGPATTERNLEVELSMODE/TRG
The MODE/TRG Screen Setup
9. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the test before!
Agilent 81110A/’11A Performance Test17
Page 18
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. For the Period test you can switch OFF the channels that are not
being tested.
10. Set the Counter to measure the frequency at the choosen input
1 / 3
11. Check the Agilent 81110A PLL pulse period at the following
settings:
14. The RMS-jitter for period of 500 ns is 65 ps. Enter the result in
the Test Report as TR entry 6.1a - 2
Agilent 81110A/’11A Performance Test43
Page 44
Agilent 81110A/’11A Performance Test
Test 6.1b: Period Jitter, Internal PLL
Test Specifications
RMS-Jitter 0.001% + 15 ps
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
Cable, 50 Ω, coaxial, BNC
Cable, SMA
Procedure
1. Connect Agilent 81110A to the Scope as shown.
54750A + 54751A
81110A UNDER TEST
81110A
165/330 MHz
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
6 dB Attenuator
54008A Delay Line
INPUT
BNC/SMA
Adapter
OUT 1
SMA Cable
Equipment Set-up for Jitter Test using the Agilent 54750A +
54751A
Using the Agilent 54121T the Set-up is the same as before.
44Agilent 81110A/’11A Performance Test
OUTPUT
POWER SPLITTER
11667B
TRIG
SMA Cable
Page 45
Agilent 81110A/’11A Performance Test
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
3. Select the [MODE/TRG] screen on the Agilent 81110A and set
up as follows:
CONTINUOUS PULSES
Single-Pulses at Out1
Single-Pulses at Out2
Pulse-Period: internal PLL
TIMINGPATTERNLEVELSMODE/TRG
The TRG MODE Screen Setup
4. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the following illustrations:
PerNormalON
20.00ns
1
Delay
Width
LeadE
TraiE
0ps
10.00ns
2.00ns
=LeadE
Offset
Amplit
50
ΩΩΩΩ
into 50.0
+500mV
1.00V
ΩΩΩΩ
MODIFY
int. OSC
*int. PLL
CLK-IN
MODIFY
20.00
ns
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Agilent 81110A/’11A Performance Test45
Page 46
Agilent 81110A/’11A Performance Test
Configuring Output Screen 1
PerNormalOFF
20.00ns
2
MODIFY
Delay
Width
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the
other channel.
0ps
10.00ns
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50
ΩΩΩΩ
into 50.0
Separate Out2
+500mV
1.00V
20.00
ΩΩΩΩ
ns
5. Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Select the Display menu and set the Number of Averages to 64
•Select the Channel menu and set the Attenuation factor of channel
2 to 2
46Agilent 81110A/’11A Performance Test
Page 47
Agilent 81110A/’11A Performance Test
•Set the VOLTS/DIV of channel 2 to 10 mV/div
•Set OFFSET to 500 mV
•Select the Timebase menu and set the TIME/DIV to 100 ps/div
•Center the first positive-going edge of the signal (approximate Delay
= 29 ns)
•Select the Delta V menu and turn the V markers On
•Set the Marker 1 Position to 490 mV and the Marker 2 Position
to 500 mV
•Select the Delta t menu and turn the T Markers On
•Set START ON EDGE = POS1 and STOP ON EDGE = POS1
•Press the PRECISE EDGE FIND key
6. RECORD the delta t reading. This is the rise time of the refer-
ence signal within a 1% amplitude window of the signal connected to Input 2. This value is needed later to calculate the
correct jitter. (delta.t.up)
7. Select the Timebase menu and center the second positive-
going edge of the signal (approximate Delay = 49 ns)
8. Press MORE and HISTOGRAM
•Select the Window submenu and set:
•Source is channel 2
•Choose the Time Histogram
•Press WINDOW MARKER 1 and set it to 490 mV
•Press WINDOW MARKER 2 and set it to 500 mV
Agilent 81110A/’11A Performance Test47
Page 48
Agilent 81110A/’11A Performance Test
9. Select the Acquire submenu, set the Number of Samples to
1000 and press START ACQUIRING
10. After the data for the time histogram has been acquired (#
Samples = 100%), select the Result submenu.
11. Press MEAN and SIGMA. RECORD the values of sigma
12. The RMS-jitter is calculated as follows:
RMS - jitter =
6sigma-delta.t.up
6
13. The RMS-jitter for period of 20 ns is 15.2 ps. Enter the result
in the Test Report as TR entry 6.1b - 1
NOTE:See the Agilent54750A User’s Guide / Service Guide to get the
info needed to do the Jitter Test using this scope.
48Agilent 81110A/’11A Performance Test
Page 49
Test 6.2: Width Jitter (PPL not active)
Test Specifications
RMS-Jitter 0.01% + 15 ps
Equipment Needed
Digitizing Oscilloscope with Accessories
Delay Line (22 ns)
Power Splitter
Cable, 50 Ω, coaxial, BNC
Cable, SMA
Procedure
1. Connect Agilent 81110A to the Scope as shown:
Agilent 81110A/’11A Performance Test
81110A UNDER TEST
165/330 MHz
81110A
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
BNC - SMA
Adapter
OUT 1
INPUT 1
54121T Frontend
SMA - SMA
Adapter
SMA Cable
2
SMA Cable
54008A Delay Line
4
3
INPUT
TRIG
POWER SPLITTER
OUTPUT
11667B
Equipment Set-up for Jitter Test
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
Agilent 81110A/’11A Performance Test49
Page 50
Agilent 81110A/’11A Performance Test
3. On the Agilent 81110A set up [OUTPUT 1] and [OUTPUT 2]
pages as shown in the following illustrations:
Per 1.000
Delay
Width
LeadE
TraiE
Configuring Output Screen 1
Per 1.000
Delay
Width
LeadE
TraiE
3.030ns
µµµµ
s NormalON
0ps
3.030ns
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
µµµµ
s NormalOFF
0ps
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50
ΩΩΩΩ
Offset
Amplit
50
ΩΩΩΩ
Separate Out2
+500mV
1.00V
into 50.0
+500mV
1.00V
into 50.0
1
3.030
ΩΩΩΩ
2
3.030
ΩΩΩΩ
MODIFY
ns
MODIFY
ns
Configuring Output Screen 2
50Agilent 81110A/’11A Performance Test
Page 51
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the
other channel.
Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Select the Display menu and set the Number of Averages to 128
•Select the Channel menu and set the Attenuation factor of channel 2
to 2
•Set the VOLTS/DIV of channel 2 to 10 mV/div
•Set OFFSET to 500 mV
•Select the Timebase menu and set the TIME/DIV to 10 ps/div
•Center the first negative-going edge of the signal (approximate Delay
= 33.8 ns)
•Select the Delta V menu and turn the V markers On
•Set the Marker 1 Position to 500 mV and the Marker 2 Position to
490 mV
•Select the Delta t menu and turn the T Markers On
•Set START ON EDGE = NEG1 and STOP ON EDGE = NEG1
•Press the PRECISE EDGE FIND
Agilent 81110A/’11A Performance Test51
key
Page 52
Agilent 81110A/’11A Performance Test
4. RECORD the delta t reading. This is the fall time of the referencesignal within a 1% amplitude window of the signal connected to Input 2. This value isneeded later to calculate the
correct jitter. (delta.t.dn)
5. Set the Agilent 81110A Pulse Width to 50 ns
6. Select the Timebase menu and center the first negative-going
edge of the signal (approximate Delay = 80.5 ns)
7. Press MORE and HISTOGRAM
8. Select the Window submenu and set:
•Source is channel 4
•Choose the Time Histogram
•Press WINDOW MARKER 1
•Press WINDOW MARKER 2
and set it to 500 mV
and set it to 490 mV
9. Select the Acquire submenu, set the Number of Samples to
1000 and press START ACQUIRING
10. After the data for the time histogram has been acquired (#
Samples = 100%), select the Result submenu.
11. Press MEAN and SIGMA. RECORD the value of sigma
12. The RMS-jitter is calculated as follows:
52Agilent 81110A/’11A Performance Test
Page 53
Agilent 81110A/’11A Performance Test
RMS - jitter =
6 sigma - delta.t.dn
6
13. The RMS-jitter for pulse width of 50 ns is 20 ps. Enter the
result in the Test Report as TR entry 6.2 - 1
14. Set the Agilent 81110A for pulse width of 500ns
16. The RMS-jitter for pulse width of 500 ns is 65 ps. Enter the
result in the Test Report as TR entry 6.2 - 2
17. Repeat steps 1. to 17. for Width Jitter - PLL active.
Test Specifications
RMS-Jitter 0.001% + 15 ps
18. Enter the results in the Test Report as TR entry 6.2a - 1 and
TR entry 6.2a - 2
The RMS-jitter for pulse width of 50 ns is 15.5 ps
The RMS-jitter for pulse width of 500 ns is 20 ps
NOTE:Repeat the entire test for the second channel, if it is installed.
Agilent 81110A/’11A Performance Test53
Page 54
Agilent 81110A/’11A Performance Test
Test 6.3: Delay Jitter (PLL not active)
Test Specifications
RMS-Jitter 0.01% + 15 ps
Equipment Needed
Digitizing Oscilloscope with Accessories
Procedure
1. Connect Agilent 81110A to the Scope as shown:
81110A UNDER TEST
81110A
165/330 MHz
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
TRIG OUT
OUT 1
54121T Frontend
INPUT 1234TRIG
Equipment Set-up for Delay Jitter Test
2. For calculating the RMS-jitter, the rise time of the reference
signal within a 1% amplitude window is required. If this value
54Agilent 81110A/’11A Performance Test
6 dB
Attenuator
with SMA/BNC
Adapter
20 dB
Page 55
Agilent 81110A/’11A Performance Test
is not already measured in the Period Jitter test, then perform
the first 6 steps of the Period Jitter test.
3. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
4. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Per 1.000
Delay
LeadE
TraiE
Configuring Output Screen 1
Per 1.000
Delay
LeadE
TraiE
Configuring Output Screen 2
µµµµ
s NormalON
50.00ns
50.00nsWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
µµµµ
s NormalOFF
50.00ns
50.00nsWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Offset
Amplit
50
Offset
Amplit
50
Separate Out2
ΩΩΩΩ
into 50.0
ΩΩΩΩ
into 50.0
1
+500mV
1.00V
ΩΩΩΩ
2
+500mV
1.00V
ΩΩΩΩ
MODIFY
50.00
ns
MODIFY
50.00
ns
Agilent 81110A/’11A Performance Test55
Page 56
Agilent 81110A/’11A Performance Test
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested.
If you then test the other channel:
c. Switch ON the channel your are testing, and switch OFF the
other channel.
Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Select the Display menu and set the Number of Averages to 64
•Set the VOLTS/DIV = 10 mV/div
•Set OFFSET to 500 mV
•Select the Timebase menu and set the TIME/DIV to 100 ps/div
•Center the first positive-going edge of the signal (approximate Delay
= 64 ns)
5. Press MORE and HISTOGRAM
6. Select the Window submenu and press WINDOW MARKER
1 and set it to 490 mV
7. Press WINDOW MARKER 2
and set it to 500 mV
8. Select the Acquire submenu, set the Number of Samples to
1000 and press START ACQUIRING
9. After the delta for the time histogram has been acquired (#
Samples = 100%), select the Result submenu.
10. Press MEAN and SIGMA. RECORD the values of sigma!
11. The RMS-jitter is calculated as follows:
56Agilent 81110A/’11A Performance Test
Page 57
Agilent 81110A/’11A Performance Test
RMS - jitter =
6
12. The RMS-jitter for delay of 50 ns is 20 ps. Enter the result in
the Test Report as TR entry 6.3 - 1
15. The RMS jitter for delay of 500 ns is 65 ps. Enter the result in
the Test Report as TR entry 6.3 - 2
16. Repeat steps 1. to 16 . for Delay Jitter - PLL active.
Test Specifications
RMS-Jitter 0.001% + 15 ps
17. Enter the results in the Test Report as TR entry 6.3a - 1 and
TR entry 6.3a - 2
The RMS-jitter for pulse width of 50 ns is 15.5 ps
The RMS-jitter for pulse width of 500 ns is 20 ps
NOTE:Repeat the entire test for the second channel, if it is installed.
Agilent 81110A/’11A Performance Test57
Page 58
Agilent 81110A/’11A Performance Test
Test 7: High and Low Levels
The following tests are required:
1. High level from 50Ω into 50Ω
2. Low level from 50Ω into 50Ω
3. High level from 1KΩ into 50Ω
4. Low level from 1KΩ into 50Ω
Test Specifications
Source Impedance50 Ω1 KΩ
High Level-9.90 V to +10.0 V-19.8 V to +20.0 V
Low Level-10.0 V to +9.9 V-20.0 V to +19.8 V
Amplitude0.10 Vpp to 10.0 Vpp0.20 Vpp to 20.0 Vpp
Load Impedance 50 Ω
Level Resolution10 mV20 mV
Level Accuracy+
1% of ampl + 50 mV+ 1% of ampl + 100 mV
Equipment Needed
1. Digitizing Voltmeter (DVM)
2. 50 Ω Feedthrough Termination, 0.1%, 10 W Adapter.
3. BNC to dual banana plug ( 1251-2277)
4. Cable, 50 Ω, coaxial, BNC
Procedure
Connect Agilent 81110A to the DVM as shown:
58Agilent 81110A/’11A Performance Test
for amlitude ≤ 19V
Page 59
81110A UNDER TEST
81110A
165/330 MHz
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
Agilent 81110A/’11A Performance Test
EXT.
3458A DVM
TRIGGER
INPUT
BNC-DUAL
BANANA
TRIG OUT
OUT 1
Connecting the DVM for High and Low Levels Tests
Test 7.1: High Level, 50 Ohms into 50 Ohms
1. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
2. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Per 100.0ms NormalON
Delay
LeadE
TraiE
25.00ms
50.00msWidth
2.00ns
=LeadE
High
Low
50
ΩΩΩΩ
+10.0V
+0.0mV
into 50.0
50 OHM
Feedthrough
1
+10.0
ΩΩΩΩ
0.1%
10W
MODIFY
V
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 1
Agilent 81110A/’11A Performance Test59
Page 60
Agilent 81110A/’11A Performance Test
Per 100.0ms NormalOFF
2
Delay
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a.Configure both channels.
b. Switch OFF the channel that is not being tested
If you then test the other channel:
c.Switch ON the channel you are testing, and switch OFF the other
channel.
25.00ms
50.00msWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
High
Low
50
ΩΩΩΩ
Separate Out2
+10.0V
+0.0mV
into 50.0
ΩΩΩΩ
MODIFY
+10.0
V
3. Set the DVM Agilent 3458A to:
Function: DCV
Trigger: TRIG EXT
AD-Converter integration time NPLC: 0.1
(Number of Power Line Cycles)
60Agilent 81110A/’11A Performance Test
Page 61
Agilent 81110A/’11A Performance Test
4. Check the Agilent 81110A high level at the following high
level settings with the low level set to 0.0 V.
High LevelAcceptable RangeTR Entry
10.0 V
5.0 V
3.0 V
1.0 V
0.5 V
0.1 V
9.85 V to 10.15 V
4.90 V to 5.10 V
2.92 V to 3.08 V
0.94 V to 1.06 V
445 mV to 555 mV
49 mV to 151 mV
The low level may vary within + 1% of amplitude+ 50 mV
Test 7.2: Low Level, 50 Ohms into 50 Ohms
1. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
2. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Per 100.0ms NormalON
Delay
LeadE
TraiE
75.00ms+0.0mV
50.00msWidth
2.00ns
=LeadE
High
Low
50
ΩΩΩΩ
7.1 - 1
7.1 - 2
7.1 - 3
7.1 - 4
7.1 - 5
7.1 - 6
-100mV
into 50.0
1
-100
ΩΩΩΩ
MODIFY
mV
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 1
Agilent 81110A/’11A Performance Test61
Page 62
Agilent 81110A/’11A Performance Test
Per 100.0ms NormalOFF
2
Delay
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested
If you then test the other channel:
75.00ms+0.0mV
50.00msWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
High
Low
50
ΩΩΩΩ
Separate Out2
-100mV
into 50.0
ΩΩΩΩ
MODIFY
-100
mV
c. Switch ON the channel you are testing, and switch OFF the
other channel.
3. Check the Agilent 81110A low level at the following low level
settings with the high level set to 0.0 V
62Agilent 81110A/’11A Performance Test
Page 63
Agilent 81110A/’11A Performance Test
Low LevelAcceptable RangeTR Entry
-0.1 V
-0.5 V
-1.0 V
-3.0 V
-5.0 V
-10.0 V
-49 mV to -151 mV
-445 mV to -555 mV
-0.94 V to -1.06 V
-2.92 V to -3.08 V
-4.90 V to -5.10 V
-9.85 V to -10.15 V
The high level 0.0 V may vary + 1% of amplitude +50 mV.
Test 7.3: High Level, 1K Ohms into 50 Ohms
1. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
2. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Per 100.0ms NormalON
Delay
LeadE
TraiE
25.00ms
50.00msWidth
2.00ns
=LeadE
High
Low
1k
ΩΩΩΩ
7.2 - 1
7.2 - 2
7.2 - 3
7.2 - 4
7.2 - 5
7.2 - 6
+20.0V
+0.0mV
into 50.0
1
+20.0
ΩΩΩΩ
MODIFY
V
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 1
Agilent 81110A/’11A Performance Test63
Page 64
Agilent 81110A/’11A Performance Test
Per 100.0ms NormalOFF
2
Delay
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested
If you then test the other channel:
c. Switch ON the channel you are testing, and switch OFF the
other channel.
25.00ms
50.00msWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
High
Low
1k
ΩΩΩΩ
Separate Out2
+20.0V
+0.0mV
into 50.0
ΩΩΩΩ
MODIFY
+20.0
V
3. Check the Agilent 81110A high level at the following high
level settings with the low level set to 0.0 V.
64Agilent 81110A/’11A Performance Test
Page 65
Agilent 81110A/’11A Performance Test
High LevelAcceptable RangeTR Entry
19.0 V
10.0 V
5.0 V
1.0 V
0.2 V
18.71 V to 19.29 V
9.80 V to 10.20 V
4.85 V to 5.15 V
0.89 V to 1.11 V
98 mV to 302 mV
The low level 0.0 V may vary + 1% of amplitude + 100 mV.
Test 7.4: Low Level, 1K Ohms into 50 Ohms
1. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
2. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Per 100.0ms NormalON
Delay
LeadE
TraiE
75.00ms+0.0mV
50.00msWidth
2.00ns
=LeadE
High
Low
1k
ΩΩΩΩ
7.3 - 1
7.3 - 2
7.3 - 3
7.3 - 4
7.3 - 5
-200mV
into 50.0
1
-200
ΩΩΩΩ
MODIFY
mV
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
Configuring Output Screen 1
Agilent 81110A/’11A Performance Test65
Page 66
Agilent 81110A/’11A Performance Test
Per 100.0ms NormalOFF
2
Delay
LeadE
TraiE
Configuring Output Screen 2
NOTE:When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested
If you then test the other channel:
c. Switch ON the channel you are testing, and switch OFF the
other channel.
75.00ms+0.0mV
50.00msWidth
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
High
Low
1k
ΩΩΩΩ
Separate Out2
-200mV
into 50.0
ΩΩΩΩ
MODIFY
-200
mV
3. Check the Agilent 81110A low level at the following low level
settings with the high level set to 0.0 V.
66Agilent 81110A/’11A Performance Test
Page 67
Agilent 81110A/’11A Performance Test
Low LevelAcceptable RangeTR Entry
-0.2 V
-1.0 V
-5.0 V
-10.0 V
-19.0 V
-98 mV to -302 mV
-0.89 V to -1.11 V
-4.85 V to -5.15 V
-9.80 V to -10.20 V
-18.71 V to -19.29 V
7.4 - 1
7.4 - 2
7.4 - 3
7.4 - 4
7.4 - 5
The high level 0.0 V may vary + 1% of amplitude + 100 mV
NOTE:Repeat the High and Low Level tests for the second channel, if it
is installed.
Agilent 81110A/’11A Performance Test67
Page 68
Agilent 81110A/’11A Performance Test
Test 8: Transition Time
Test Specifications
Range
Minimum Transitions
Accuracy
Linearity
Equipment Needed
Digitizing Oscilloscope with Accessories
Cable, SMA
Procedure
Perform the tests as shown in the following sections:
2.0 ns to 200 ms
(measured between 10% and 90% of amplitude)
< 2.0 ns
(typical 1.4 ns for ECL levels
measured between 20% and 80% of amplitude
typical 5 ns for 1kOhm source impedance)
± 10% ± 200 ps
typical ± 3% for transitions > 100 ns
68Agilent 81110A/’11A Performance Test
Page 69
Test 8.1a: Leading Edge Test
Minimum Leading Edge and Leading Edge ranges .
1. Connect Agilent 81110A to the Scope as shown:
Agilent 81110A/’11A Performance Test
81110A UNDER TEST
330 MHz
81110A
5555
#IKNGPV
PULSE-/ PATTERN GENERATOR
Connecting Agilent 81110A to the Scope
NOTE:
When you connect the test equipment the first time, and whenever
you change the setup during the following tests, use the torque
wrench (8170-1582) to tighten and loosen the SMA connectors.
This will ensure that the connectors are at the correct tightness and
give the best signal transfer!
2. Set up the Agilent 81110A as described in "Initial Setup of the
Agilent 81110A"
TRIG OUT
BNC - SMA Adaptor
OUT 1
Cable, SMA
54121T Frontend
INPUT 1234TRIG
20dB
Attenuator
3. On the Agilent 81110A press MORE and set up [OUTPUT 1]
and [OUTPUT 2] pages as shown in the following illustrations:
Agilent 81110A/’11A Performance Test69
Page 70
Agilent 81110A/’11A Performance Test
Per 500.0
Delay
DtyCyc
LeadE
TraiE
Configuring Output Screen 1
Per 500.0
Delay
DtyCyc
LeadE
TraiE
µµµµ
s NormalON
0ps
Offset
50.00%
2.00ns
=LeadE
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
50.00%
2.00ns
=LeadE
Amplit
50
µµµµ
s NormalOFF
0ps
Offset
Amplit
50
Separate Out2
OUTPUT 1PATTERNOUTPUT 2MODE/TRG
ΩΩΩΩ
into 50.0
ΩΩΩΩ
into 50.0
1
+0.0mV
5.00V
ΩΩΩΩ
2
+0.0mV
5.00V
ΩΩΩΩ
MODIFY
2.00
ns
MODIFY
2.00
ns
Configuring Output Screen 2
NOTE:
When you are testing instruments with 2 output channels it is
necessary to:
a. Configure both channels.
b. Switch OFF the channel that is not being tested
70Agilent 81110A/’11A Performance Test
Page 71
Agilent 81110A/’11A Performance Test
If you then test the other channel:
c. Switch ON the channel you are testing, and switch OFF the
other channel.
4. Set the Digitizing Oscilloscope Agilent 54121T:
•Press AUTOSCALE
•Center one pulse on screen, e.g.:
•TIME/DIV = 50
•Select the Display menu and set the Number of Averages to 32
•Select the Channel menu and set the Attenuation factor to 10
•Select the Delta V menu and turn the voltage markers On
•Set the Preset Levels = 10-90% and press AUTO LEVEL SET
•Select the Timebase menu and set TIME/DIV = 1 ns/div, DELAY =
19.5 ns
•Select the Delta t menu and turn the markers On
•Set START ON EDGE = POS1 and STOP ON EDGE = POS1
µs/div, DELAY = 380 µs,
5. Set period of Agilent 81110A to: Period = 1 µs and
change the Agilent 81110A Delay to center the leading edge of
the first pulse on the screen
6. After the averaging, while the oscilloscope is in the Delta t
menu, Press the PRECISE EDGE FIND
key
Agilent 81110A/’11A Performance Test71
Page 72
Agilent 81110A/’11A Performance Test
7. Check the Agilent 81110A rise times at the following leading
edge settings:
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
1-1 6.06ns 5.878 ns ______ 6.242 ns ____ ____
1-2 9.99ns 9.690 ns ______ 10.290 ns ____ ____
1-3 10.0ns 9.7 ns ______ 10.3 ns ____ ____
1-4 50.0ns 48.5 ns ______ 51.5 ns ____ ____
1-5 99.9ns 96.903 ns ______ 102.897 ns ____ ____
Agilent 81110A/’11A Performance Test83
Page 84
Agilent 81110A/’11A Performance Test
Counter Uncertainty factor ___________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
1-6 100 ns 97.0ns ________ 103.0 ns ____ ____
1-7 500 ns 485.0 ns _______ 515.0 ns ____ ____
1-8 1 µs 970.0 ns _______ 1030.0 ns ____ ____
1-9 5 00µs 485 µs _______ 5 15 µs ____ ____
1-10 500 ms 485 ms _______ 515 ms ____ ____
84Agilent 81110A/’11A Performance Test
Page 85
Agilent 81110A/’11A Performance Test
PLL Period
(Results measured as frequency by counter)
Counter Uncertainty factor ____________
_____________________________________________________
TR Test Limit Actual Limit Pass Fail
Entry Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.1a-1 50 ns ______ 20 ps ____ ____
6.1a-2 500 ns ______ 65 ps ____ ____
6.1b-1 20 ns _______ 15.2 ps ____ ____
86Agilent 81110A/’11A Performance Test
Page 87
Agilent 81110A/’11A Performance Test
Test Results for Agilent 81111A Output Channel ______
Serial No. ________________
Width
Scope Uncertainty factor ________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
3-1 3.030 ns 2.689 ns ________ 3.371 ns ____ ____
3-2 6.06ns 5.528 ns ________ 6.492 ns ____ ____
3-3 10.0 ns 9.450ns ________ 10.550 ns ____ ____
3-4 50.0 ns 48.25 ns ________ 51.75 ns ____ ____
3-5 100 ns 96.75 ns ________ 103.25 ns ____ ____
3-6 500 ns 484.75 ns ________ 515.25 ns ____ ____
3-7 50 µs 48.5 µs _________ 51.5 µs ____ ____
3-8 5 ms 4.85 ms ________ 5.15 ms ____ ____
3-9 500 ms 485 ms ________ 515 ms ____ ____
Agilent 81110A/’11A Performance Test87
Page 88
Agilent 81110A/’11A Performance Test
Width Jitter
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.2-1 50 ns ________ 20 ps ____ ____
6.2-2 500 ns ________ 65 ps ____ ____
6.2a-1 50 ns ________ 15.5 ps ____ ____
6.2a-2 500 ns ________ 20 ps ____ ____
88Agilent 81110A/’11A Performance Test
Page 89
Agilent 81110A/’11A Performance Test
Delay
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
4-1 0.00 ns _________Fixed Delay ____ ____
4-2 5.00 ns 4.35 ns _________ 5.65 ns ____ ____
4-3 10 ns 9.20 ns _________ 10.80 ns ____ ____
4-4 50.0 ns 48.0 ns _________ 52.0 ns ____ ____
4-5 100 ns 96.5 ns __________ 103.5 ns ____ ____
4-6 500 ns 484.5 ns __________ 515.5 ns ____ ____
4-7 50 µs 48.5 µs _________ 51.5 µs ____ ____
4-8 5 ms 4.85 ms _________ 5.15 ms ____ ____
4-9 500 ms 485 ms _________ 515 ms ____ ____
Agilent 81110A/’11A Performance Test89
Page 90
Agilent 81110A/’11A Performance Test
Delay Jitter
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
6.3-1 50 ns ________ 20 ps ____ ____
6.3-2 500 ns ________ 65 ps ____ ____
6.3a-1 50 ns ________ 15.5 ps ____ ____
6.3a-2 500 ns ________ 20 ps ____ ____
90Agilent 81110A/’11A Performance Test
Page 91
Agilent 81110A/’11A Performance Test
Double Pulse Delay
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
5-1 6.06 ns 5.628 ns _______ 6.392 ns ____ ____
5-2 10.0 ns 9.550 ns _______ 10.45 ns ____ ____
5-3 50.0ns 48.35 ns _______ 51.65 ns ____ ____
5-4 100ns 96.85 ns _______ 103.15 ns ____ ____
Counter Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
5-5 500 ms 485 ms _______ 515 ms ____ ____
5-6 1 s 970.0 ms ________ 1030.0 ms ____ ____
Agilent 81110A/’11A Performance Test91
Page 92
Agilent 81110A/’11A Performance Test
High Level 50Ω-50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.1-1 10.0 V 9.85 V __________ 10.15 V ____ ____
7.1-2 5.0 V 4.90 V __________ 5.10 V ____ ____
7.1-3 3.0V 2.92 V __________ 3.08 V ____ ____
7.1-4 1.0 V 0.94 V __________ 1.06 V ____ ____
7.1-5 0.5 V 445 mV __________ 555 mV ____ ____
7.1-6 0.1 V 49 mV __________ 151 mV ____ ____
High Level 1KΩ−50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.3-1 19.0 V 18.71V __________ 19.29 V ____ ____
7.3-2 10.0 V 9.80 V __________ 10.20 V ____ ____
7.3-3 5.0 V 4.85 V __________ 5.15 V ____ ____
7.3-4 1.0 V 0.89 V __________ 1.11V ____ ____
7.3-5 0.2 V 98 mV __________ 302mV ____ ____
92Agilent 81110A/’11A Performance Test
Page 93
Agilent 81110A/’11A Performance Test
Low Level 50Ω-50Ω
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
7.4-1 -0.2V -98 mV __________-302 mV ____ ____
7.4-2 -1.0V -0.89 V __________ -1.11 V ____ ____
7.4-3 -5.0V -4.85V __________ -5.15 V ____ ____
7.4-4 -10.0V -9.80 V _________ -10.20 V ____ ____
7.4-5 -19.0V -18.71 V __________-19.29 V ____ ____
Agilent 81110A/’11A Performance Test93
Page 94
Agilent 81110A/’11A Performance Test
Leading Edge
Scope Uncertainty factor __________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
8.1b-1 2.0 ns <2 ns __________ 2.4 ns ____ ____
8.1b-2 10 ns 8.8 ns __________ 11.2 ns ____ ____
8.1b-3 50 ns 44.8 ns __________ 55.2ns ____ ____
8.1b-4 500 ns 449.8 n __________ 550.2 ns ____ ____
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
9-1 5V __________ + 5% of ampl.____ ____+20mV
9-2 500 mV _________ +5% of ampl. ____ ____
+20mV
Preshoot
_____________________________________________________
TR Entry Test Limit Actual Limit Pass Fail
Min Result Max
_____________________________________________________
9-3 0 V _________
96Agilent 81110A/’11A Performance Test
+5% of ampl. ____ ____
+20mV
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.