Manual part number: 00536-90013
Printed in USA
March 2001
Supersedes: May 2000
Page 2
Documentation Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT
DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO
THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT
NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR
ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE
MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the per formance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “co mmercial item” as defined in FAR
2.101(a) or as “Restricted computer sof tware” as defined in FAR 52.227-19 (J une 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disc losure of
Software is subject to Agilent Technologies’ standard commercial license terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users
will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Agilent
web site:
•Go to http://www.agilent.com.
• Enter the document’s pa rt number (located on the title page) in the Quick Search box.
• Click GO.
• Click on the hyperlink for the document.
• Click the printer icon located in the tool bar.
Page 3
Contacting Agilent
This information supersedes all prior HP contact information.
Review this product and related documentation to familiarize yourself with
safety markings and instructions before you operate the instrument. This
product has been designed and tested in accordance with international
standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a procedure,
practice, or the like, that, if not correctly performed or adhered to, could result
in personal injury. Do not proceed beyond a WARNING notice until the
indicated conditions are fully understood and met.
CAUTION The CAUTION notice denotes a hazard. It calls attention to an operating
procedure, practic e, o r the like, which, i f not correctly performe d or a dhe red
to, could result in damage to the product or loss of important data. Do not
proceed beyond a CAUTION notice until the indicated conditions are fully
understood and met.
Instrument Markings
When you see this symbol on your instrument, you should refer to the instrument’s
!
1SM1-AThis text indicates that the instrument is an Industrial Scientific and Medical Group 1
instruction manual for important information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is
accompanied by a year, it indicates the year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
Class A product (CISPER 11, Clause 4).
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
Agilent 537A Operating And Service Manual
Page 6
Safety Earth
Ground
This is a Safety Class I product (provided with a protective earthing
terminal). An uninterrup ti ble saf et y ear th gr ound must be pr o vided from the
main power source to the product input wiring terminals, power cord, or
supplied power cord set. Whenever it is likely that the protection has been
impaired, the product must be made inoperative and secured against any
unintended operation.
Befor e Applying P o werVerify that the product is c onfigured to match the available main power
source as d escribed in the input power configuration instructions in this
manual. If this product is to be powered by autotransformer, make sure the
common terminal is connected to the neutral (grounded) side of the ac power
supply.
Agilent 537A Operating And Service Manual
Page 7
Overview
DescriptionThe Agilent 537A coaxial frequency meter is used for measuring frequency
in the 0.96 to 12.4 GHz range. Accuracy is obtained from the spiral scale.
The meter contains a coaxial line coupled to a quarter-wavelength coaxial
resonant cavity loaded with capacitance at the bottom of the band. At
resonance, power is absorbed by the cavity, producing a dip in the coaxial
line power. The frequency at which this dip occurs is read d irectly from the
dial in gigahertz.
Tunin g is accomplished b y moving a pis ton. A precision l ead scre w , whic h is
spring-loaded to pre v ent backl ash, is use d to po sition t he piston in the ca v ity.
Due to the design of the cavity, its electrical length is extended at lower
frequencies, and there are no spurious modes.
SpecificationsThe instrument specifications are provided in Table 1 on page 1.
Table 1Specifications
CharacteristicValue
Frequency range3.7 to 12.4 GHz
Reflection coefficient off-resonance0.33 (2.0 SWR, 9.5 dB return lost)
Calibration increments10 MHz
Dimensions
Height
Diameter
Weight3-1/2 lb (1.6 kg)
Dip at resonance1 dB (minimum)
Dial accuracy
1. Width (including connector): 4-5/8in (118 mm).
2. Overall Accuracy: ±0.17% which includes allowance of ±0.02% for 0 to 100% relative humidity,
±0.0016 per ×C from 13 to 33×C and 0.03% backlash.
1
2
5-3/4 in (146 mm)
3-1/2 in (89 mm)
±0.1%
MeasurementA detector and indicator are req uired to indicat e the di p in coaxi al line power
at resonance. It is r ecommended tha t 10 dB of att enuation be placed be tween
the signal source and the frequency meter. If insufficient power is available
to use a 10 dB attenuator, an attenuator offering less attenuation may be
used, to a minimum of 3 dB. The attenuator pr events a shift in the f re quen c y
of the signal source due to line impedance changes at resonance.
Agilent 537A Operating Manual 1
Page 8
Overview
The detector-indicator combination can be a thermistor mount and power
meter such as the Agilent 478A and 432A, or a crystal detector and SWR
meter. If a SWR meter is used as the indicator, the signal source must be
modulated at 1000 Hz (sine or square wave modulation). For visual display,
an oscilloscope c an be used as the indicat or. When using an oscilloscope for
swept frequency displ ays , be sur e to sweep the oscilloscope at the same ra te
as the signal source. Resonance will be noted as a dip in the oscilloscope
pattern
CAUTION Do NOT run dial into stops at either end of dial
If frequency measureme nt is one part of the measurement pro cedure, be sure
to move meter off resonance before proceeding with other measurements.
Figure 1Type-N Connector Dimensions
ConnectorsThe input and output connect ors are type-N, f emale and are compatibl e with
connectors whose dimensions conform to MIL-C-39012.
N-female Pin Depth: 0.207 to 0.201 in (effective 0 to –0.006)
2 Agilent 537A Operating Manual
Page 9
Performance Test
Performance Test
PurposeThe procedures listed in Table 2 and Figure 3 on page 7 check performance
for incoming inspection, periodic evaluation, and calibration. The tests are
performed without access to the instrument interior. The performance
standards are contained in Table 1 on page 1.
Recommended Test
Equipment
Table 2Recommended Test Equipment
Test instruments required to make the performance checks are listed in
Table 2. Instruments other than those listed may be used provided their
performance equals or exceeds the critical specifications given.
modulator, swept amplitude analyzer, and a dual directional coupler in a
reflectometer test setup. The reflectometer is calibrated using a short. The
537A coaxial frequency meter under test is connected and the return
measured.
Agilent 537A Operating Manual 3
Page 10
Performance Test
Figure 2Return Loss Test Setup
Table 3Equipment
Instrument TypeSuggested Agilent ModelGHz or Connector type
1. Two single directional couplers connected as a dual directional coupler can also be used.
–
appropriate RF plugins)
1
778D
779D
110 MHz to 2 GHz
1.7 to 12.4 GHz
Type -N male
APC-7
1.2 GHz cutoff
2.2 GHz cutoff
4.1 GHz cutoff
Type-N male
4 Agilent 537A Operating Manual
Page 11
Procedure1. Connect the equipment as shown in Figure 2.
2. Set the analyzer control to 0 dB.
3. Set CHANNEL A OFFSET CAL switch to OFF
4. Press CHANNEL A DISPLAY POSITION push button.
5. Adjust CHANNEL A screwdriver adjustment marked POSITION to
place trace on the center graticule. Increase resolution and make fine
adjustment.
6. Place short on reflectometer.
7. Press CHANNEL A push buttons marked DISPLAY R and 10 dB/DIV
8. Set the sweep oscillator to sweep the band of interest.
9. Adjust the sweep oscillator output power level to place the trace on the
second graticule below the center graticule. (Power input is –20 dBm to
the R detector.)
10. Press CHANNEL A push button marked DISPLAY A/R.
Performance Test
11. Turn CHANNEL A OFFSET CAL control to ON and adjust the
OFFSET CAL control to place the trace on the center graticule.
12. Increase the dB/DIV resolution by steps to 0.25 dB/DIV if necessary,
adjust the OFFSET CAL control to return the trace to the center
graticule.
13. Disconnect the short from the reflectometer test port.
14. Connect the 537A under test to the reflectometer test port by
disconnecting the short and inserting the properly terminated 537A
under test.
15. Use CHANNEL A OFFSET dB thumbwheel to return the trace to the
center of the screen, or as close as possible.
16. Read the return loss at the CHANNEL A OFFSET dB window and on
the oscilloscope. Return loss is the total indication of both the
CHANNEL A OFFSET dB window and the trace on the oscilloscope.
Add the trace indication to the window reading if the trace is below the
center graticule, subtract if above.
Agilent 537A Operating Manual 5
Page 12
Performance Test
17. Return loss measured should be equal to or greater than the following
limits:
Table 4Return Loss Limits
Specification26 dB30 dB40 dB
Lower Limit of Reading
Coupler Directivity
20.8 dB
9.5 dB
27.8 dB
11.0 dB
24.6 dB
10.5 dB
21.9 dB
9.8 dB
18. If the above limits are not met, test the 537A using a slotted line at the
frequency in question.
6 Agilent 537A Operating Manual
Page 13
Overall Accuracy
Overall Accuracy
Specification for the 537A: ±0.17% overall accuracy.
DescriptionA sweep oscillator, 537A under test, detector, and oscilloscope are
connected in series. The sweep oscillator is set to AF to sweep the band in
question. The 537A is set to the frequency in question and the notch located
on the trace. The sweep oscillator is set to CW and adjusted to the center of
the absorption notch of the 537A. The frequency of the sweep oscillator is
measured with a counter and co mp are d wi t h the specificati ons in Table 1 on
page 1. The sweep oscillator is set to automatic sweep and the notch depth
across the b and is checked.
Figure 3Overall Accuracy Test Setup
Table 5Test Instruments
Instrument TypeSuggested Agilent ModelFrequency Range (GHz)
Procedure1. Connect the equipment as shown in Figure 3 on page 7.
2. Set oscilloscope for a vertical sensitivity of about .02 V/cm (ac) and
horizontal sensitivity of about 1 V/cm.
3. Set sweep oscillator for ^F automat ic sweep of about 3.8 GHz + 12 MHz
for 537A.
4. Set the 537A to 3.8 GHz.
5. Locate the frequency meter notch and center it on the display by
adjusting the sweep limits on the sweep oscillator.
6. Set the sweep oscillator to CW and adjust the frequency to the center of
the notch (lowest indication on the oscilloscope).
7. Read the frequency on the counter. Frequency should be within the
specifications given in Table 1 on page 1.
8. Repeat above procedure every 200 MHz across the band.
9. Set sweep oscillator to automatic sweep and tune the 537A across
frequency range of sweep oscillator. The notch travel across the
frequency ran ge should cause at least the notch depth given in Table 1-1.
Stop Gear Replacement
Procedure
1. Upon completion of the stop gear replacement it will be necessary to
check the frequency calibration of the instrument.
2. The followin g special tool is required for t his procedure. A W a ldes No. 3
Truarc pliers with tips bent 45 degrees.
3. For the following procedures refer to Figure 4 on page 12 and Figure 1
on page 1.
Disassembly Procedure(Numbers in parentheses r efer to numbers on drawing in Fi gure 4. To replace
the stop gear, proceed as follows:
1. Remove knob (1).
2. Scribe dial position to give approximate location of dial on dial holder.
Do not use pencil as this may be inadvertently erased.
3. Unscrew dial-window retaining nut (3).
4. Remove nut and window
5. Loosen two No. 6 Allen se t-scre ws (9) holdi ng dial. These are l ocated on
the inside or dial holder (10).
6. Turn dial holder until cap (7) is approximately flush with top of dial
holder (near high frequency stop position).
7. Scribe a line from lead screw on plunger to cap for later alignment.
8 Agilent 537A Operating Manual
Page 15
8. Loosen both No. 6 Allen set screws (9) holding cap in place on lead
screw. These are internal and may be found by looking through the two
inspection holes in dial holder. These are spaced approximately 90
degrees apart.
9. While holding cap in place, back out both screws sufficiently to allow
removal of cap. This should be approximately three full turns of the
Allen screw. If spring and cap jump out, replace them with big-diameter
end of spring at bottom.
CAUTION Do not turn or remove plunger.
10. Turn the 537A upside down and remove the Truarc retaining-ring
holding dial holder (3). (This requires special Truarc pliers. It may be
necessary to adapt some of your own tools to the job of removing the
retaining ring.).
11. Lift 537A off dial holder and place base assembly upright again.
Examine both sections for pieces of broken stop gear.
Overall Accuracy
12. Inspect bearing and gear rack to make sure that these hasn’t been
damaged by portions of the broken stop gear. If any of the internal
workings other than sto p gear hav e been damaged, it will be necessary to
return to the factory for a major rebuilding.
Reassembly Procedure1. Replace dial holder , t aking ca re not to damage gears when e ngaging stop
gear with gear rack
2. Check rotation of the dial holder. There should be approximately 11-1/4
turns of the dial. If not , dise ngage and tr y ag ain with a n e w gear posit ion.
3. When the stop gear is positioned to give approximately 11-1/4 turns
from stop to stop, secure d ial holder with th e T ruarc retaini ng ring. Make
sure ring is seated properly on inside groove of dial holder.
4. Turn dial holder to stop at high-frequency end of dial.
5. Ins ert spring into dial holder.
6. Replace cap by compressing spring. Note alignment of cap and lead
screw. See steps 6 thru 9 of disassembly procedure.
7. Tighten both No. 6 Allen set-screws holding cap to lead screws.
8. Turn from stop to stop. Check for binding or other erratic behavior. It
may be necessary to make slight adjustments in cap position so that no
binding occurs during rotation.
9. Turn dial holder counterclockwise (toward lower frequency) to expose
dial retaining set screws.
Agilent 537A Operating Manual 9
Page 16
Overall Accuracy
10. Place dial on dial holder and check alignment wit h scri be mark of st ep 2.
11. Tighten both No. 6 Allen set-screws to hold dial in place.
CAUTION Do not tighten excessively as knob may not slip into place over the dial/dial
holder.
12. Turn dial clockwise near the stop at the high-frequency end and place
both cursors (6) into the dial grooves to indicate approximately 12.4
GHz.
13. Carefully replace window making sure that cursors ride in the window
grooves.
14. Align window with ba se. Window keys must be in the window retention
grooves.
15. Replace window-retaining nut.
16. Turn from stop to stop. Check for binding or other erratic behavior. It
may be necessary to make slight adjustments in screw tightness or dial
position so that no binding occurs during dial rotation.
17. When assured of smooth operation, check calibration. Calibration
should be reasonably close if ca re wa s taken i n alignment of the cap le ad
screw and dial. Replace knob.
10 Agilent 537A Operating Manual
Page 17
Calibration Procedure
Calibration Procedure
Test equipment recommended for testing the 537A are listed in Table 2 on
page 3. Test equipment other than that recommended may be substituted if
their performance equals or exceeds the minimum required specifications
given. Figure 2 on page 4 and Figure 3 on page 7 list procedures for
checking instrument specifications. If dial accuracy is not within listed
specifications, adjust 537A as detailed in the Calibration Alignment
Procedure, as follows.
Calibration Alignment
Procedure
If dial accuracy of the 537A is not within specifications. The following
procedure can be used for readjustment.
1. Remove knob (1).
2. Remove base (11) retaining screws. With base retaining screws
removed, pla ce the 537A in up right posit ion and mak e electr ical hookup
for OVERALL ACCURACY TEST.
3. In the original high-frequency end setting, 12.4 GHz is set almost
one-half turn (about 150 degrees) counterclockwise from the stop. The
dial (5) is held to the dial holder (10) by two set screws (9). Check
frequency accuracy at 12.4 GHz, and if not within overall accuracy
specification, loosen set screws (9) and slip dial. T he set screws ar e
loosened from inside the dial holder; the dial must be rotated about 6
turns before the access holes become visible.
4. The low-frequency end adjustment is made at 3.7 GHz by moving
capacitiv e load with two cap screws. This adjustmen t sh oul d not chan ge
the high frequency setting.
5. Check dial accuracy in vicinity of 3.7 to 4.1 and 8.2, 12.0, and 12.4
GHz. Reposition the dial, or capacitive load if a compromise is
necessary
6. If accuracy specifications cannot be met, further disassembly is not
recommended. The unit should be returned to your local Agilent
Technologies Sales and Service office.
7. Replace base and knob.
Agilent 537A Operating Manual 11
Page 18
Calibration Procedure
Figure 4537A Repair and Recalibration.
12 Agilent 537A Operating Manual
Page 19
Replaceable Parts
Table 6Replaceable P a rts
Replaceable Parts
Reference
Designator
1Knob5040–69501
2Screw, set 8-32 x 3/163030–00012
3Nut, dial window retainer00532–400051
4WindowK532A–181
5Dial00537–200311
6CursorK532A-202
7CapR532A–161
8Spring, lead screw1460–00441
9Screw, set 6-32 x 1/43030–01062
10Holder, dialP532A–41
11Base00537–200291
12Ring retaining0510–00691
13Coupling loop assembly00537–601
14Bearing, ball1410–00051
DescriptionPart NumberQuantity
1
15Window support00537–206
16Screw, fill Hd 2-56 x 1/40520–0012
17Shaft stop gearP532A–51
18Gear stopP532A–191
19Ring, retaining0510–00701
20Cork ring2190–03451
1. See Figure 4 for parts location.
Agilent 537A Operating Manual 13
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.