Agilent 5361B Data Sheet

Agilent Technologies 5361B Pulse/CW Microwave Counter
Data Sheet
Product Specifications
Input Characteristics
Input 1 Input 2 Input 2 (50 Ù) (1 MÙ) (50 Ù) Frequency range 500 MHz to 20, 10 Hz to 10 MHz to
0.5 to 12.4 GHz -28 dBm
12.4 to 20 GHz -23 dBm
0.5 to 26.5 GHz -20 dBm (Opt 026, 040)
26.5 to 40 GHz 0.37 x f (in GHz) (Opt 040) -29.8 dBm
Frequency (Input 1)
Automatic and Manual Acquisition: 500 MHz to 20 GHz; 500 MHz to
26.5 GHz (Option 026); 500 MHz to 40 GHz (Option 040)
Least Significant Digit: 1 MHz to 1 Hz for frequency, 0.001 Hz for PRF
CW Frequency Measurements FM tolerance: 55 MHz peak-to-peak Tracking speed (fast acquisition): 800 MHz/s Acquisition time: Manual mode, <40 ms; automatic mode, fast acq., <100 ms Gate times (1 Hz resolution): 200 to 1000 ms Measurement time: >8.5 ms (in Dump Mode) Accuracy: See datasheet
Pulse Parameters (Input 1)
> ) 100 nx Accuracy: ±(20 nx + timebase uncertainty x ± (20 nx) x (PRF)
2
(100 ave.) measurement ±LSD ±LSD ±timebase uncertainty
Profile (Input 1)
Frequency Range (min/max for Y axis): 500 MHz/40 GHz FM Chirp Tolerance (max span for Y axis): 50 MHz peak-to-peak Time Range (min/max span for X axis): 100 ns/10 ms; 1 ns resolution Internal Gate Width: Minimum 11 ns to 23 ns; typical minimum: 14 ns External Gate Width: Minimum: Manual acquisition 20 ns;auto acquisition 60 ns Number of Data Points: Up to 100
Profile Frequency Measurements
Printers supported: HP 2225A, HP 227B, HP 3630A Opt 002 Profile phase measurements: See Application Note 377-4 for details. Computer required.
Frequency (Input 2)
Range: 10 Hz to 525 MHz Accuracy: 0.001 Hz to 1 Hz Resolution/LSD: 0.001 to 1 Hz