Agilent 34970A Specifications Sheet

Digital Input/Ouput
Use the digital outputs with an external power supply to control microwave switches and attenuators, solenoids, power relays, indicators, and more. Use the digital inputs to sense limit switch and digital bus status. There are no complex handshake modes; reads and writes are initiated either from the front panel or the bus.
Totalize Input
Count events from devices like photo interrupters, limit switches, and Hall-effect sensors.
It keeps an updated total which can be read via the front panel or programmatically at any time. With 28 bits of resolution, it can count events at full speed for nearly 45 minutes without an overflow.
Analog Output
Use the two electronically calibrated analog outputs to source bias voltages to your device under test, to control your analog programmable power supplies, or use the outputs as setpoints for your control systems. The outputs are programmed directly in volts, either from the front panel or from the bus.
HP 34907A
Multifunction Module
• 16 bits of digital input and output
• 100 kHz totalizer input
• Two ±12V analog outputs The HP 34907A allows great
flexibility for a variety of sense and control applications. It combines two 8-bit ports of digital input and output, a 100 kHz gated totalizer, and two ±12V analog outputs— all on a single earth-referenced module. The digital inputs and totalizer input may be included in a scan. Alarm limits for the digital and event counter inputs are evaluated continuously, capturing and logging alarm conditions even between scans.
Digital Input/Output
Port 1, 2 8 bit, input or output,
non-isolated Vin(L) < 0.8V (TTL) Vin(H) > 2.0V (TTL) Vout(L) < 0.8V @ Iout = - 400 mA Vout(H) > 2.4V @ Iout = 1 mA Vout(H) max < 42V with external open
drain pull-up Alarming Maskable pattern match
or state change
Speed 4 ms (max) alarm
sampling
Latency 5 ms (typical) to 34970A
alarm output Read/Write Speed 95/s
Totalize Input
Max Count 228- 1 Totalize Input 100 kHz (max)
Rising or falling edge, programmable
Signal level 1 Vp-p (min)
42 Vpk (max)
Threshold 0V or TTL,
jumper selectable Gate Input TTL-Hi, TTL-Lo, or none Count Reset Manual or Read + Reset Read Speed 85/s
Analog Output
D
AC 1, 2 ± 12V, non-isolated Resolution 1 mV I
OUT
10 mA max Settling time 1 ms to 0.01% of output Accuracy ±(% of output + mV)
1 year ±5°C 0.15% + 6 mV 90 day ±5°C 0.1% + 6 mV 24 hour ±1°C 0.04% + 4 mV
Temp. Coefficient ± ( 0.015%
+ 1 mV) / °C
Port 1 Channel 01
Port 2 Channel 02
Bit 0
Bit 7
8
Bit 0
Bit 7
16
8
Channel 03
Channel 04
28 Bits
+IN
-IN
Gate Gate
DIO
TOT
DAC 1
16
Channel 05
DAC 2
HP 34970A Data Acquisition/Switch Unit