The HP 34420A is a 7
meter. Its combination of bench-top and system features makes this
meter a versatile solution for your testing requirements now and in the
future.
Convenient bench-top features
• Built-in math operations including thermistor, thermocouple and
RTD temperature measurements
• Two channel input allows ratio and difference functions for voltage
measurements
• Highly visible vacuum-fluorescent display
• Portable, ruggedized case with non-skid feet
Flexible system features
• HP-IB (IEEE-488) interface and RS-232 interface are standard
• SCPI (Standard Commands for Programmable Instruments) and
Keithley 181 compatibility
• Readings at up to 250 per second
1
digit, high performance nanovolt, micro-ohm
/
2
HP 34420A
Nano Volt/Micro Ohm Meter
The Front-Panel at a Glance
1 Input channel select5 Menu operation keys
2 Measurement function keys6 Range/number of dig its displayed keys
3 Math operation key s7 Single trigger/autot r ig ger/a ut o hol d key
4 Filter select key8 Shift / local key
2
The Front-Panel Menu at a Glance
The menu is organized in a top-down tree structure with three levels.
NOTE: The commands enclosed in square brackets ([ ]) in the CAL MENU a re "hidden" unless th e meter is UNSECURE D for calibration.
Ø
3
Display Annunciators
*
AdrsMeter is addressed to listen or talk over the HP-IB interface.
RmtMeter is in remote mode (using remote interface).
ManMeter is using manual ranging (autorange is disabled).
TrigMeter is waiting for a single trigger or external trigger.
OC OffOffset compensation is turned off.
NullA null value is being used.
StatsMath statistics operations are being used.
ScaleMath scaling operations are being used.
ERRORHardware or remote interface command error(s) detected.
Shift"Shift" key has been pressed.
Ch1Meter input is on Channel 1.
Ch2Meter input is on Channel 2.
Ch1 - Ch2Meter is indicati ng the difference between inputs on channel 1 and channe l 2.
Ch1 / Ch2Meter is indicating the ratio of the inputs on channel 1 and channel 2.
ΩMeter is measuring resistance (Ohms).
LPMeter is using low power Ohms.
2WMeter is using 2-wire Ohms (annunciator off indicates a 4-wire measurement).
LoVMeter is using voltage li mited Ohms.
HoldAutomatic reading hold is enabled.
FiltThe analog and/or the digital filter is enabled.
MemTurns on when reading memory is enabled.
Turns on during a measurement.
To review the display annunciators, hold down the Shift key as you
turn on the meter.
4
The Rear Panel at a Glance
1 Chassis ground5 Voltmeter complete output terminal
2 Power-line fuse-holder assembly6 External trigger input te rminal
3 Power-line voltage setting7 HP-IB (IEEE-488) interface connector
4 Chart recorder out pu t ter minal
(Analog out)
8 RS-232 interface connector
Use the front-panel Input / Out put Menu to :
• Select the HP-IB or RS- 232 int er fa ce
• Set the HP-IB bus address
• Set the RS-232 baud rate and parity
5
In This Book
Quick Start Chapter 1 prepares the meter for use and helps you get
familiar with a few of its front-panel features.
Front-Panel Operation Chapter 2 introduces you to the front-panel menu
and describes some of the meter’s menu features.
Features and Functions Chapter 3 gives a detailed description of the
meter’s capabilities and operation. You will find this chapter useful
whether you are operating the meter from the front panel or from a
remote interface.
Remote Interface Reference Chapter 4 contains reference
information to help you program the meter using a remote interface.
Error Messages Chapter 5 lists the error messages that may appear as
you are working with the meter. Each listing contains enough
information to help you diagnose and solve the problem.
Application Programs Chapter 6 contains several remote interface
application programs to help you develop programs for your
measurement application.
Measurement Tutorial Chapter 7 discusses measurement
considerations and techniques to help you obtain the best accuracy and
reduce sources of measurement error.
Specifications Chapter 8 lists the meter’s specifications and describes
how to interpret these specifications.
If you have questions relating to the operation of the meter, call
1-800-452-4844 in the United States, or contact your nearest
Hewlett-Packard Sales Office.
6
Contents
Chapter 1 Quick Start
To Prepare the Meter For Use 13
If the Meter Does Not Turn On 14
To Adjust the Carrying Handle 16
To Make Input Connections 17
To Measure Voltage 19
To Measure Resistance 20
To Measure Temperature With Thermistors 21
To Measure Temperature With RTDs 21
To Measure Temperature With Thermocouples 22
To Select a Range 23
To Set the Number of Digits 24
To Set the Integration Time 25
Front-Panel Display Formats 26
To Rack Mount the Meter 27
Chapter 2 Front-Panel Operation
Front-Panel Menu Reference 31
A Front-Panel Menu Tutorial 33
To Select an Input Channel 41
To Select a Range 42
To Set or Change the Number of Digits 43
To Set the Integration Time 44
To Make Null (Relative) Measurements 45
To Store Minimum and Maximum Readings (Stats) 46
To Trigger the Meter 47
To Use Reading Hold 48
To Make Voltage Ratio and Difference Measurements 49
To Use Reading Memory 50
Chapter 3 Features and Functions
General Measurement Configuration 55
Input Filters 55
Integration Time 58
Reducing Measurement Noise 60
Number of Digits Displayed 62
Ranging 64
Overload Detection 65
Math Operations 74
Null (Relative) Operation 77
Reading Hold 79
Triggering 80
Trigger Source Choices 82
The Wait-for-Trigger State 85
Halting a Measurement in Progress 85
Number of Samples 86
Number of Triggers 86
Trigger Delay 87
Automatic Trigger Delays 89
Using the INITiate and FETCh? Commands 131
The MEASure? and CONFigure Commands 134
Setting the Function, Range, and Resolution 137
Selecting the Input Channel 140
Special Resistance Measurement Commands 141
Temperature Measurement Commands 142
The Wait-for-Trigger State 153
Triggering Commands 154
Chart Output (Analog Out) Commands 156
System-Related Commands 157
The SCPI Status Model 159
What is an Event Register? 159
What is an Enable Register? 159
What is a Condition Register? 159
SCPI Status System 160
The Status Byte 161
Using Service Request (
Using
*STB? to Read the Status Byte 163
To Interrupt Your Bus Controller Using
To Determine When a Command Sequence is Completed 163
How to Use the Message Available Bit (MAV) 164
*OPC to Signal When Data is in the Output Buffer 164
Using
The Standard Event Register 165
The Questionable Data Register 167
The Operational Status and Condition Registers 169
Status Reporting Commands 171
Calibration Commands 174
RS-232 Interface Configuration 176
RS-232 Configuration Overview 176
RS-232 Data Frame Format 177
Connection to a Computer or Terminal 177
Connection to a Printer 179
DTR / DSR Handshake Protocol 179
RS-232 Troubleshooting 181
RS-232 Interface Commands 182
An Introduction to the SCPI Language 183
SCPI Data Types 185
Input Message Terminators 186
Output Data Formats 186
Using Device Clear to Halt Measurements 187
TALK ONLY for Printers 187
To Set the HP-IB Address 188
To Select the Remote Interface 189
To Set the Baud Rate 190
To Set the Parity 191
To Select the Programming Language 192
Alternate Programming Language Compatibility 193
Specific Issues 193
SCPI Compliance Information 194
IEEE-488 Compliance Information 197
HP BASIC / HP-IB Program Example 1 215
HP BASIC / HP-IB Program Example 2 216
HP BASIC / HP-IB Program Example 3 218
HP BASIC / HP-IB Program Example 4 221
RS-232 Operations with QuickBASIC 223
Microsoft
®
Visual Basic Language Program 224
C Language Programs 231
Compiling and Linking a C Program 231
C / HP-IB Example 1 233
C / HP-IB Example 2 235
C / HP-IB Program Example 3 239
®
Microsoft
Excel Macro Example 243
Macro Comments 244
Chapter 7 Measurement Tutorial
Measurement Techniques and Sources of Error 249
Voltage Measurements 249
Resistance Measurements 258
Temperature Measurements 263
Cleaning the Connector Contacts 271
Input Connector 272
Obtaining Parts For the Input Connector 273
10
Chapter 8 Specifications
Accuracy Specifications
Measurement Characteristics
Operating Characteristics
Dimensions
To Calculate Total Measurement Error 281
Interpreting Meter Specifications 285
Number of Digits and Overrange 285
Sensitivity 285
Resolution 286
Accuracy 286
24-Hour Accuracy 287
90-Day and 1-Year Accuracy 287
Temperature Coefficients 287
Configuring for Highest Accuracy Measurements 287
Index 289
Declaration of Conformity 295
1
Quick Start
Quick Start
This chapter helps you prepare the meter for use and contains exercises
designed to get you started with the meter, its menus, and the front
panel.
The front panel has two rows of keys to select various functions and
operations. Most keys have a shifted function printed in blue above
the key. To perform a shifted function, press
annunciator will turn on). Then, press the key that has the desired label
above it. For example, to select the temperature measurement function,
Shift DCV .
press
Shift (the Shift display
If you accidentally press
annunciator.
The rear cover of this book is a fold-out Quick Reference Guide. On this
cover you will find a quick summary of various meter features. Inside
the rear cover is a diagram of the front panel menu options.
Shift , just press it again to turn off the Shift
12
Chapter 1 Quick Start
To Prepare the Meter For Use
To Prepare the Meter For Use
The following steps help you verify that the meter is ready for use.
1Check the list of supplied items.
Verify that you have received the following items with your meter. If any
item is missing, contact your nearest Hewlett-Packard Sales Office.
One low thermal input cable.
One low thermal four-wire shorting plug.
One power cord.
This User’s Guide.
One Service Guide.
One folded Quick Reference card.
Certificate of Calibration with removeable calibration label.
One bottle of DeoxIT contact cleaner.
2Connect the power cord and turn the meter on.
The front panel display will light up while the meter performs its
power-on self-test. The HP-IB bus address is displayed. The meter is left
1
digit resolution, channel 1 input, and digital filter on.
in 6
/
2
1
To view the display with all annunciators turned on,
hold down
3Perform a complete self-test.
The complete self-test performs a more extensive series of tests than
those performed at power-on. Hold down
power switch to turn on the meter; continue to hold 5 seconds. The self-test will begin when you release the key.
If the self-test is successful, "PASS" is displayed on the front panel. If the
test is not successful, "FAIL" is displayed and the ERROR annunciator
turns on. See the Service Guide for instructions on returning the meter to
Hewlett-Packard for service.
DeoxIT is a trademark of CAIG Laboratories, Inc., San Diego, California.
Shift as you turn on the meter.
Shift as you press the
Shift for more than
13
Chapter 1 Quick Start
If the Meter Does Not Turn On
If the Meter Does Not Turn On
Use the following steps to help solve problems you might experience
when turning on the meter. If you need more help, see the Service Guide
for instructions on returning the meter to Hewlett-Packard for service.
1Verify there is ac power to the meter.
First verify that the meter’s power switch is in the "On" position. Make
sure the power cord is firmly plugged into the power module on the
meter’s rear panel. Verify that the power source is energized.
2Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the
meter is shipped from the factory. Change the voltage setting if it is not
correct. The settings are: 100, 120, 220, or 240 Vac (for 230 Vac operation,
use the 220 Vac setting).
See the diagram on the next page to verify or change the setting.
To replace the 250 mAT 250 V fuse, order HP part number 2110-0817.
This fuse is used for all power line voltage settings.
14
Chapter 1 Quick Start
If the Meter Does Not Turn On
1 Remove the power cord. Remove
the fuse holder assembly from the rear
3 Rotate the line-voltage selector until
the correct voltage appears in the window.
2Install the fuse. Remove the
line-voltage selector from the assembly.
250 mAT fuse
(250V)
4 Replace the fuse-holder assembly in the
rear panel.
1
100, 120, 220 (230), or 240 Vac
Install the correct fuse and verify that the correct line voltage appears in the window.
15
Chapter 1 Quick Start
To Adjust the Carrying Handle
To Adjust the Carrying Handle
To adjust the position, grasp the handle by the sides and pull outward.
Then, rotate the handle to the desired position.
Bench-top viewing positions
16
Carrying Position
Chapter 1 Quick Start
To Make Input Connections
To Make Input Connections
Using the Cable Provided
Align the front panel connector and cable conductors, press in.
Tighten coupling nut.
Torque to 20 in-lb
maximum
CH1 BLACK LO
CH2GREEN
1
RED HI
HI
WHITE
LO
Warning
The connector and cable are an integral part of the measurement system.
For the highest accuracy, use the copper cable and connectors supplied
by Hewlett-Packard.
The conductors may require occasional cleaning to remove oxides.
Cleaning the conductors is described on page 271.
The connector body and cable shield are connected to earth ground.
17
Chapter 1 Quick Start
To Make Input Connections
Using a Custom Cable
Channel 1
HILO
DCV+–
4-WireΩ
2-WireΩ
ThermistorSource and Sense +Source and Sense –
Source +Source –
Source and Sense +Source and Sense –
Earth Ground (shell)
Channel 2
HI LO
DCV+ –
4-WireΩ
Thermocouple+ –
Sense + Sense –
Refer to page 272 for additional information about building custom input
cables.
Thermocouple types: B, E, J, K, N, R, S, T
Reference: external thermistor, external fixed value, or internal thermistor.
Channel 2 only.
ISOTHERMAL BLOCK
RED
CH2
OR
INTERNAL
CH1 BLACK THERMISTOR
GREEN
CH2
RED
CH1OPEN
BLACK
GREEN
CH2
WHITE
WHITE
OR
FIXED VALUE
ICE BATH
REFERENCE
EXTERNAL
When using the internal thermistor as the temperature reference, you
should make a custom cable to connect the thermocouple wire directly
to the input terminals (see page 272).
22
Chapter 1 Quick Start
To Select a Range
To Select a Range
You can let the meter automatically select the range using autoranging
or you can select a fixed range using manual ranging.
Toggles between autoranging
and manual ranging
Selects a higher range
and disables autoranging
Selects a lower range
and disables autoranging
• Autoranging is selected at power-on and after a remote interface reset.
Man annunciator is on when
manual range is enables
1
• Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
• If the input signal is greater than the present range can measure, the
meter will give an overload indication ("OVLD").
• When making difference or ratio measurements, an overload on either
channel will give an overload indication ("OVLD").
For voltage measurements, ranging is local to the selected channel. This
means that you can select the ranging method (auto or manual) for each
channel independently. When manually ranging, the selected range is
local to the active channel; the meter remembers the range when you
switch between channels.
23
Chapter 1 Quick Start
To Set the Number of Digits
To Set the Number of Digits
1
1
You can set the display to show 4
⁄
, 5
2
1
⁄
, 6
⁄
2
2
book, the most significant digit (leftmost on the display) is referred to as
1
⁄
the “
” digit, since it can only be a “0” or “1”. The number of digits
2
displayed also depends upon the integration time set, see page 25.
, or 7
1
⁄
digits. In this
2
Press the Shift key
1
Selects 5
(1 PLC)
⁄2 digits
• The number of digits is set to 6
Selects 7
(100 PLC)
Selects 6
(10 PLC)
1
⁄2 digits
1
⁄
digits at power-on and after a
2
1
⁄2 digits
remote interface reset.
• The number of digits shown is dependent upon integration time and
filter settings. The meter will not allow you to show more digits than
the practical measurement capability of the meter. You can, however,
reduce the number of digits shown.
Fewer More
Digits Digits
24
Chapter 1 Quick Start
To Set the Integration Ti me
To Set the Integration Time
Integration time is specified in Number of Power Line Cycles (NPLC).
You can set the NPLC to 0.02, 0.2, 1, 2, 10, 20, 100, or 200.
• You can set the integration time to one of three fixed values by
choosing the number of digits displayed, see page 24.
• You can set the integration time in the MEASure menu using the
INTEGRATE command.
• The Integration Time is directly related to the maximum number of
digits the meter will display.
NPLC
0.02
0.2
1
2
10
20
100
200
Filter Off
Max Digits Shown
41⁄2 5
51⁄2 6
61⁄2 7
61⁄2 7
1
⁄2 7
6
71⁄2 7
71⁄2 7
71⁄2 7
Filter On
Max Digits Shown
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
• You can always show fewer digits than the maximum allowed
1
⁄
(the minimum number of digits shown is 4
).
2
Integration Time is local to the selected function. This means that you
can select the integration time for each function independently. The
meter remembers integration time when you switch between functions.
25
6 digits
Chapter 1 Quick Start
Front-Panel Display Formats
Front-Panel Display Formats
-H.DDD,DDD,D EFF
Front-panel display format.
-Negative sign or blank (positive)
H "1/2" digit (0 or 1)
D Numeric Digits
EExponent (m,k,M)
FMeasurement units
"1/2" digit
5 digits
"1/2" digit
7 digits
"1/2" digit
1.021,651 V This is the 1 V range,
-045.230 mV This is the 100 mV range,
0.133,256,7 ΩThis is the 1 ohm range,
1
⁄2 digits are displayed.
6
1
⁄2 digits are displayed.
5
1
⁄2 digits are displayed.
7
O.VLD mV
This is an overload indication on the 1 mV range.
26
Chapter 1 Quick Start
To Rack Mount the Meter
To Rack Mount the Meter
You can mount the meter in a standard 19-inch rack cabinet using one of
three optional kits available. Instructions and mounting hardware are
included with each rack-mounting kit. Any HP System II instrument of the
same size can be rack-mounted beside the HP 34420A meter.
Remove the carrying handle, and the front and rear rubber bumpers,
before rack-mounting the meter.
To remove the handle, rotate it to the vertical position and pull the ends o utward.
1
To remove the rubber bumper, stretch a corner and then slide it off.
27
Chapter 1 Quick Start
To Rack Mount the Meter
To rack mount a single instrument, order adapter kit 5063-9240.
To rack mount two instruments side-by-side, order lock-link kit 5061-9694
and flange kit 5063-9212.
To install one or two instruments in a sliding support shelf,
order shelf 5063-9255, and slide kit 1494-0015 (for a single instrument,
also order filler panel 5002-3999).
28
2
Front-Panel Operation
Front-Panel Operation
This chapter assumes you are familiar with the meter and menu
operations. You should also understand how to make connections for the
various types of measurements. If you are not familiar with this
information, please read chapter 1, "Quick Start", starting on page 11.
This chapter does not give a detailed description of every front-panel key
or menu operation. It does, however, give you a good overview of the
front-panel menu and the most common front-panel operations. See
chapter 3 “Features and Functions,” starting on page 53, for a complete
discussion of the meter’s capabilities and operation.
1: DIG FILTERSelects the digital filter speed. Can be set to FAST, MEDIUM, or SLOW.
2: INTEGRATESets the measurement integration time.
3: OCOMP ΩEnables or disables offset compensation for resistance measurements.
4: LOW POWER ΩEnables or disables low power ohms measurements.
5: LOW VOLT ΩEnables or disables voltage limited resistance measurements.
6: LoV LIMIT Ω Sets the voltage limit for voltage limited resistance measurements.
B: TEMPerature MENU
1:PROBE TYPE Ø 2: UNITS Ø 3: RTD TYPE Ø 4:RTD Ro Ø 5:T/C TYPE Ø 6: COLD JUNCT Ø 7:JUNCT TEMP
1: PROBE TYPESelects thermocouples, thermistors, or RTDs as the temperature probe.
2: UNITSSets the measurement units to either °C, °F, or Kelvins.
3: RTD TYPESets the type of RTD in use.
4: RTD RoSets the nominal (0°C) value for the RTD in use.
5: T/C TYPESelects the type of thermocouple in use.
6: COLD JUNCTS ets the source of the temperature reference junction.
7: JUNCT TEMPAllows direct entry of reference junction temperature.
C: MATH MENU
1: STATS Ø 2: NULL VALUE Ø 3: SCALE GAIN Ø 4 : SCALE OFST
1:STATSRecalls the values of minimum, maximum, standard deviation, average,
peak-to-peak, and number of readings.
2: NULL VALUESelects independent null settings for voltage on channels 1 and 2, resistance,
and temperature functions.
3: SCALE GAINSets the gain for linear scaling of the data.
4: SCALE OFSTSets the offset for linear scaling of data.
D: TRIGger MENU
1: READ HOLD Ø 2: TRIG DELAY Ø 3: N SAMPLES
2
1: READ HOLDS elects the reading hold sensiti vity band.
2: TRIG DELAYSpecifies a time delay between the trigger and the start of the measurement.
3: N SAMPLESSets the number of samples taken per trigger.
1: RDGS STOREEnables or disables reading memory.
2: SAVED RDGSRecalls readings stored in memory (up to 1024 readings).
3: ERRORRetrieves errors from the error queue (up to 20 errors).
4: TESTPerforms a complete self-test or individual self-tests.
5: CHART OUTEnables or disables the chart recorder output (Analog Out).
6: CHART SPANSets the chart recorder output range (Analog Out).
7: CHART NULLSets the chart recorder output offset (Analog Out).
8: DISPLAYEnables or disables the front panel display.
9: COMMAEnables or disables a comma separator between digits in the display.
10: PRESETReturns the meter to factory default settings.
11: REVISIONDisplays the meter’s firmware revision code.
1: HP-IB ADDRSets the HP-IB bus address (0 to 30, default 22).
2: INTERFACESelects either the HP-IB or RS-232 remote interface.
3: BAUD RATESets the baud rate for RS-232 operation.
4: PARITYSelects even, odd, or no parity for RS-232 operation.
5: LANGUAGESelects interface language, either SCPI or Keithley 181.
* The commands enclosed in square brackets ( [ ] ) are "hidden" unless the meter is UNSECURED for calibration.
Ø6:CAL COUNT Ø 7:MESSAGE
1: SECUREDThe meter is secured against calibration; enter code to unsecure.
1: UNSECUREDThe meter is unsecured for calibration; enter code to secure.
2: CALIBRATEPerforms a calibration of meter; must be UNSECURED.
3: CHART ZEROPerforms a calibration of the chart recorder output zero level.
4: CHART GAINPerforms a calibration of the chart recorder output gain.
5: INJECTED IPerforms a calibration to minimize the injected current.
6: CAL COUNTReads the total number of times the meter has been calibrated.
7: MESSAGEReads the first 11 characters of a calibration string, if any,
entered from the remote interface.
32
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
A Front-Panel Menu Tutorial
This section is a step-by-step tutorial which shows how to use the
front-panel menu. We recommend that you spend a few minutes with this
tutorial to get comfortable with the structure and operation of the menu.
The menu is organized in a top-down tree structure with three levels
(menus, commands, and parameters). You move down
the menu tree to get from one level to the next. Each of the three levels
has several horizontal choices which you can view by moving left
right
Commands
Parameters
• To turn on the menu, press
• To
∨ or up
> .
Menus
Shift < (Menu On/Off).
turn off the menu, press Shift < (Menu On/Off), or press any of
the function or math keys on the top row of front-panel keys.
2
∧
< or
• To execute a menu command, press
• To recall the last menu command that was executed, press
(Menu Recall).
• To turn off the menu at any time without saving changes, press any
function key
If you become confused or get lost at any point during the tutorial, simply
turn off the menu and start over again with step 1 for that example.
Auto/Man .
Shift >
33
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
Messages Displayed During Menu Use
TOP OF MENU -You pressed
∧ while on the menus level; this is
the top level of the menu and you cannot go any higher.
To turn off the menu, press
the choices on a level, press
MENUS - You are on the menus level. Press
COMMANDS - You are on the commands level. Press
Shift <
< or >
(Menu On/Off). To move across
. To move down a level, press
< or > view the choices.
< or > to view
∨ .
the command choices within the selected menu group.
PARAMETER - You are on the parameter level. Press
< or > to view
and edit the parameter for the selected command.
MENU BOTTOM - You pressed
∨ while on the parameter level; this is
the bottom level of the menu and you cannot go any lower.
To turn off the menu, press
press
∧ .
Shift <
(Menu On/Off). To move up a level,
ENTERED - The change made on the parameter level is saved. This is
displayed after you press
Auto/Man (Menu Enter) to execute the command.
MIN VALUE - The value you specified on the parameter level is too small for the
selected command. Th e minimum value allo wed is displayed for you to edit.
MAX VALUE - The value you specified on the parameter level is too large for the
selected command. The maximum value allowed is displayed for you to edit.
EXITING - You will see th is m essa ge if yo u tu rn of f th e me nu b y
pressing
Auto/Man (Menu On/Off) or a front-panel function/math key. You
did not edit an y va lues on th e pa ram et er leve l a nd ch an ge s w ere NO T sa ved .
NOT ENTERED - You will see this message if you turn off the menu by
pressing
Shift < (Menu On/Off) or a front-panel function/math key.
You did some editing of parameters but the changes were NOT saved.
Press
Auto/Man ( M en u E nt er ) to sa v e c ha ng es made on the paramet er
level.
VOLTS ONLY - The requested function is only available for voltage
measurements.
UNAVAILABLE - The action requested is NOT allowed in the present configuration.
34
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
Menu Example 1The following steps show you how to turn on the menu, move up or down
between levels, move across the choices on each level, and turn off the
menu.
In this example, you will turn off the display comma separator.
The meter can display readings on the front panel with or without a
comma separator. The following steps show how to disable the comma.
- 0.824,153 V - 0.824153 V
With comma separatorWithout co mma separator
(factory setting)
On/Off
Shift < 1Turn on the menu.
You enter the menu on the menus level. The MEAS MENU is your first
choice on this level.
A: MEAS MENU
> > > > 2Move across to the SYS MENU choice on this level.
There are six menu group choices available on the menus level. Each choice
has a letter prefix for easy identification (
A: , B: , etc.).
2
E: SYS MENU
∨ 3Move down to the commands level within the SYS MENU.
The RDGS STORE command is your first choice on this level.
1: RDGS STORE
35
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
> > > > 4Move across to the COMMA command on the commands level.
> > > > There are eleven command choices available in the SYS MENU. Each choice
on this level has a number prefix for easy identification (
1: , 2: , etc.).
9: COMMA
∨ 5Move down a level to the COMMA parameter choices.
The first parameter choice is ON for the COMMA command (the comma
setting is stored in non-volatile memory and ON is the factory setting).
ON
> 6Move across to the OFF choice.
There are two parameter choices for COMMA.
OFF
Auto/Man
ENTER7Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now in
effect. You are then exited from the menu. The meter will no longer show
commas in the display.
ENTERED
36
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
Menu Example 2The following exercise demonstrates how to use the menu recall feature
as a shortcut to set the COMMA command back to its original setting.
You must perform the steps in Menu Example 1 before you start this
example.
Recall
Shift > 1Use menu recall to return to the COMMA command.
This returns you to the COMMA command, which was the last command
used before you exited the menu in the Example 1.
9: COMMA
v 2Move down to the COMMA parameter choices.
The first parameter choice is OFF (the current setting from Example 1).
OFF
2
> 3Move across to the ON choice.
Set the parameter back to its original value.
ON
Auto/Man
ENTER4Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now in
effect. You are then exited from the menu.
ENTERED
37
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
Menu Example 3Some commands in the menu require that you enter a numeric
parameter value. The following steps show you how to enter a number in
the menu. For this example, you will set the null value for channel 1 to
–30 millivolts.
Make sure the meter has channel 1 selected, is in the dc voltage function,
1
⁄
and has 6
digits displayed. Connect the four-wire short to the meter
2
input.
On/Off
Shift < 1Turn on the menu.
You enter the menu on the menus level. The MEAS MENU is your first
choice on this level.
A: MEAS MENU
> > 2Move across to the MATH MENU choice on this level.
There are six menu group choices available on this level.
C: MATH MENU
v 3Move down to the commands level within the MATH MENU.
The STATS command is your first choice on this level.
1: STATS
> 4Move across to the NULL VALUE command.
2: NULL VALUE
38
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
v 5Move down to edit the NULL VALUE parameter.
The null value should be 0.000000 V when you come to this point in the
menu for the first time. For this example, you will first set the null value
to –0.300000 volts.
Ù0.000,000,0 V
When you see the flashing “∧” on the left side of the display, you can abort
the edit and return to the “commands” level by pressing
v v 6Make the number negative.
∧
The left most character on the display toggles between +, – , and ^.
– 0.000,000,0 V
> > 7Move the flashing cursor over to edit the second digit.
Notice that the digit to the right of the decimal point is flashing.
– 0.000,000,0 V
^ ^ ^ 8Increment the digit until “3” is displayed.
You decrement or increment each digit independently. Neighboring digits
are not affected.
2
.
– 0.300,000,0 V
39
Chapter 2 Front-Panel Operation
A Front-Panel Menu Tutorial
> > > > 9Move the flashing cursor over to the units location.
> > > Notice that the units are flashing on the right side of the display.
– 0.300,000,0 V
v 10 Decrease the displayed number by a factor of 10.
Notice that the position of the decimal point changes and the displayed
number increases by a factor of 10 and the mV annunciator is on.
– 030.000,00 mV
Auto/Man
ENTER11 Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now
in effect. You are then exited from the menu.
ENTERED
You will be returned to the normal measurement display, and, if you
have installed the four-wire short, the display will show a positive
measurement in the approximate amount of the null value just entered.
The Null annunciator is on in the display.
Keep in mind that null is turned on and - 0.03 volts is used as the null
value for measurements. To clear the null value, press
Null .
Hint You can use the menu recall feature to save steps when entering the null
value. First, press
Null on the front panel and then press Shift >
(Menu Recall). The meter will jump to the 2:NULL VALUE command.
v . The null value in the display is the first reading taken after
Press
you turned NULL on. You can then edit this number as described.
This is the end of the front-panel menu tutorial. The remainder of the
chapter discusses several of the most common front-panel operations.
40
Chapter 2 Front-Panel Operation
To Select an Input Channel
To Select an Input Channel
For voltage measurements, you have two independent input channels.
TOGGLES MEASUREMENT CHANNEL
(SCAN) ALTERNATING MEASUREMENT CHANNELS
• Pressing
Channel toggles the display between Channel 1 and
Channel 2.
• Pressing
Shift Channel (Scan) creates an alternating Channel 1 and
Channel 2 display. Press
RED
BLACK
GREEN
WHITE
Channel again to cancel.
DC Voltage
DC Voltage
2
Channel 1
Channel 2
Scan mode alternates between input channels every 3 seconds or every
measurement, whichever is longer. If you are using the scan mode, turn
off all the input filters (both digital and analog), see page 55.
41
Chapter 2 Front-Panel Operation
To Select a Range
To Select a Range
You can let the meter automatically select the range using autoranging
or you can select a fixed range using manual ranging.
Toggles between autoranging
and manual ranging
Selects a higher range
and disables autoranging
Selects a lower range
and disables autoranging
Man annunciator is on when
manual range is enables
• Autoranging is selected at power-on and after a remote interface reset.
• Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
• If the input signal is greater than the present range can measure, the
meter will give an overload indication (“OVLD”).
• When making difference or ratio measurements, an overload on either
channel will give an overload indication (“OVLD”).
For voltage measurements, ranging is local to the selected channel. This
means that you can select the ranging method (auto or manual) for each
channel independently. When manually ranging, the selected range is
local to the active channel; the meter remembers the range when you
switch between channels.
42
Chapter 2 Front-Panel Operation
To Set or Change the Number of Digits
To Set or Change the Number of Digits
1
1
You can set the display to show 4
⁄
, 5
2
⁄
2
, 6
1
⁄
2
book, the most significant digit (leftmost on the display) is referred to as
1
⁄
the “
” digit, since it can only be a “0” or “1”. The number of digits
2
displayed also depends upon the integration time set, see page 25.
, or 7
1
⁄
digits. In this
2
2
Press the Shift key
Selects 6
(10 PLC)
Selects 5
(1 PLC)
1
⁄
2
digits
• The number of digits is set to 6
Selects 7
(100 PLC)
1
⁄
digits
2
1
⁄
digits
2
1
⁄
digits at power-on and after a
2
remote interface reset.
• The number of digits shown is dependent upon integration time and
filter settings. The meter will not allow you to show more digits than
the practical measurement capability of the meter. You can, however,
reduce the number of digits shown.
Fewer More
Digits Digits
43
Chapter 2 Front-Panel Operation
To Set the Integration Ti me
To Set the Integration Time
Integration time is specified in Number of Power Line Cycles (NPLC).
You can set the NPLC to 0.02, 0.2, 1, 2, 10, 20, 100, or 200. Also see
"Integration Time" on page 58.
• You can set the integration time to one of three fixed values by
choosing the number of digits displayed, see page 43.
• You can set the integration time in the MEASure menu using the
INTEGRATE command.
• The Integration Time is directly related to the maximum number of
digits the meter will display.
NPLC
0.02
0.2
1
2
10
20
100
200
Digital Filter Off
Max Digits Shown
41⁄2 5
51⁄2 6
61⁄2 7
61⁄2 7
1
⁄2 7
6
71⁄2 7
71⁄2 7
71⁄2 7
Digital Filter On
Max Digits Shown
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
1
⁄2
• You can always select to show fewer digits than the maximum allowed
1
⁄
(the minimum number of digits shown is 4
).
2
Integration Time is local to the selected function. This means that you
can select the integration time for each function independently. The
meter remembers integration time when you switch between functions.
44
Chapter 2 Front-Panel Operation
To Make Null (Relative) Measurements
To Make Null (Relative) Measurements
Each null measurement, also called relative, is the difference between a
stored null value and the input signal. See page 77.
2
Enables null operation;
Press again to disable
Reading = measurement – null value
• You can make null measurements on any function or channel. You can
set a null for each channel and function independently. The meter
remembers the null value when you switch between input channels or
functions.
• To null the test lead resistance for more accurate two-wire ohms
measurements, short the ends of the test leads together and then
Null .
press
• The first reading taken after you press
in the Null Register. Any previously stored value is replaced with the
new value.
• After enabling null, you can edit the stored null value by
pressing
VALUE” command in the MATH MENU (only if null is enabled). Go
down to the “parameter” level, and then edit the displayed value.
Shift > (Menu Recall). This takes you to the “NULL
Null annunciator is on when
null operation is enabled
Null is stored as the null value
45
Chapter 2 Front-Panel Operation
To Store Minimum and Maximum Readings (Stats )
To Store Minimum and Maximum Readings (Stats)
You can store the minimum and maximum readings during a series of
measurements. The following discussion shows how to read the minimum,
maximum, average, peak-to-peak, standard deviation, and reading count.
Enables stats operation;
Press again to disable
Stats annunciator is on when
stats operation is enabled
• You can use stats with any function. When you change functions, stats
operation is disabled.
• Changing input channels or selecting Scan does not disable stats. The
stored readings, however, are not marked as to the input channel.
• After enabling stats, you can read the stored minimum, maximum,
average, standard deviation, peak-to-peak, and count by pressing
Shift > (Menu Recall). This takes you to the “STATS” command in
the MATH MENU (only if stats is enabled). Go down to the
“parameter” level
pressing
< or > .
• After reading the stored value, pressing
∨
, and then read the values by
Shift < (Menu On/Off)
returns to the measurement display and continues stats operations.
• The stored values are cleared when you turn stats on, turn off the
power, or perform either a remote interface reset or preset.
• The average and standard deviation are of all readings taken since
stats was enabled (not just the average of the stored minimum and
maximum). The count is the total number of readings taken since
stats was enabled.
46
Chapter 2 Front-Panel Operation
To Trigger the Meter
To Trigger the Meter
You can trigger the meter from the front panel using single trigger
or autotrigger.
2
Enables single trigger
and triggers the meter
• Autotriggering is enabled when you turn on the meter. Notice that the
* (sample) annunciator turns on during each measurement.
• Single triggering takes one reading each time you press
then waits for the next trigger. Continue pressing this key to trigger
the meter.
Using an External Trigger
The external trigger mode is also enabled by pressing
It is like the single trigger mode except that you apply a trigger pulse
to the rear-panel
negative edge of a TTL pulse.
The front-panel
Toggles between autotrigger
and reading hold
Ext Trig
Single key is disabled when in remote.
terminal. The meter is triggered on the
Trig annunciator is on when the meter
is waiting for single
trigger (autotrigger disabled)
* (sample) annunciator is on
during each measurement
Single and
Single .
47
Chapter 2 Front-Panel Operation
To Use Reading Hold
To Use Reading Hold
The reading hold feature allows you to capture and hold a stable reading
on the display. When a stable reading is detected, the meter emits a beep
and holds the value on the display.
Toggles between autotrigger
and reading hold
• Reading hold has an adjustable sensitivity band to allow you to select
which readings are considered stable enough to be displayed. The
band is expressed as a percent of reading on the selected range. The
meter will capture and display a new value only after three
consecutive readings are within the band.
• The default band is 0.10% of reading. After enabling reading hold,
you can choose a different band by pressing
This takes you to the “READ HOLD” command in the TRIG MENU
(only if reading hold is enabled).
Go down to the “parameter” level, and then select a value: 0.01%,
0.10%, 1.00%, or 10.00% of reading.
• The sensitivity band is stored in volatile memory; the meter sets the
band to 0.10% of reading when power has been off or after a remote
interface reset.
* (sample) annunciator is on
during each measurement
Shift > (Menu Recall).
48
Chapter 2 Front-Panel Operation
To Make Voltage Ratio and Difference Measurements
To Make Voltage Ratio and Difference Measurements
To calculate a voltage ratio, the meter measures the voltages applied
to the Channel 1 input terminals and Channel 2 input terminals. The
• Maximum input signal on Channel 1 is 120 Vdc, the meter will
autorange on the actual input or can be manually set.
• Maximum input signal on Channel 2 is 12 Vdc, the meter will
autorange on the actual input or can be manually set.
2
• Maximum voltage difference between Channel 1 LO and
Channel 2 LO is 150 Vdc.
• Null, if used, is independent for each channel. An additional null* of
the difference can also be made. If null is used on either input channel,
the ‘Null’ annunciator will flash when the difference function is
selected. If you null the difference, the ‘Null’ annunciator will be
steady.
*The difference null is only available from the front panel.
49
Chapter 2 Front-Panel Operation
To Use Reading Memory
To Use Reading Memory
The meter can store up to 1024 readings in internal memory. The
following steps demonstrate how to store readings and retrieve them.
1Select the function.
Select any measurement function. You can change the function at any
time during reading memory.
Single 2Select the single trigger mode.
Notice that the Trig annunciator turns on. When reading memory is
enabled, readings are stored when you trigger the meter.
For this example, single triggering is used to store readings. You can also
use auto triggering or reading hold.
On/Off
Shift < 3 Turn on the menu.
A: MEAS MENU
< < < 4Move across to the SYS MENU choice on this level.
E: SYS MENU
v 5Move down to a level to the RDGS STORE command.
1: RDGS STORE
50
Chapter 2 Front-Panel Operation
To Use Reading Memory
v > 6Move down a level and then across to the “ON” choice.
ON
Auto/Man
ENTER7Save the change and exit the menu.
Notice that the Mem (memory) annunciator turns on to indicate that the
meter is ready to store readings. Up to 1024 readings can be stored in
first-in-first-out (FIFO) order. When memory is full, the Mem
annunciator will turn off.
Readings are preserved until you re-enable reading memory at another
time, turn off the power, or perform a remote interface reset.
Single
Single Single 8Trigger the meter three times.
This stores three readings in memory.
2
Recall
Shift > 9Use menu recall to retrieve the stored readings.
This takes you to the “SAVED RDGS” command in the SYS MENU.
2: SAVED RDGS
51
Chapter 2 Front-Panel Operation
To Use Reading Memory
v 10 Move down a level to view the first stored reading.
Reading memory is automatically turned off when you go to the
“parameter” level in the menu.
The first reading displayed is the first reading that was stored (FIFO). If no
readings are stored in memory, “EMPTY” is displayed. The first five
digits of the reading and the reading number are displayed. After a brief
pause, the reading number is removed and all reading digits are shown.
The stored readings are displayed with their units ( µ, m, k, etc.) when
appropriate. For example:
Reading
number
10.31607K: 11
Units
10.316076 K
> > 11 Move across to view the two remaining stored readings.
The readings are stored horizontally on the “parameter” level.
If you press
last reading and you will know how many readings were stored.
On/Off
Shift < 12 Turn off the menu.
EXITING
< when you first get to the “parameter” level, you will see the
52
3
Features and Functions
Features and Functions
You will find that this chapter makes it easy to look up all the details
about a particular feature of the meter. Whether you are operating the
meter from the front panel or from the remote interface, this chapter will
be useful.
This chapter is divided into the following sections:
• General Measurement Configuration, page 55
• Voltage Measurement Configuration, page 66
• Resistance Measurement Configuration, page 68
• Temperature Measurement Configuration, page 71
• Math Operations, page 74
• Null (Relative) Operation, page 77
• Reading Hold, page 79
• Triggering, page 80
• System-Related Operations, page 90
• Voltmeter Complete Terminal, page 97
• External Trigger Terminal, page 97
• Chart Output (Analog Output) Operation, page 98
• Remote Interface Configuration, page 101
• Calibration, page 107
• Defaults, Power-On and Reset States, page 112
Some knowledge of the front-panel menu will be helpful before you read this
chapter. If you have not already read chapter 2, “Front-Panel Menu
Operation,” starting on page 29, you may want to read it now. Chapter 4,
“Remote Interface Reference,” starting on page 115, lists the syntax for the
SCPI (Standard Commands for Programmable Instruments) commands
available to program the meter.
Throughout this manual, the following conventions are used for SCPI
command syntax. Square brackets ( [] ) indicate optional keywords or
parameters. Braces ( {} ) enclose a list of parameters separated with a
vertical bar (|), use only one of the values . Triangle brackets ( <> )
indicate that you must substitute a value for the enclosed parameter.
The actual command string you send to the meter should not include
any of the conventions indicated above.
54
Chapter 3 Features and Functions
General Measurement Configuration
General Measurement Configuration
This section contains information to help you configure the meter for
making measurements. You may never have to change any of the
measurement parameters discussed here, yet they are provided to give
you the flexibility you might need.
Input Filters
Two input filters are available; a digital filter and an analog filter. You
can enable the digital filter, the analog filter, or both. The digital filter
may be used for any measurement type. The analog filter is only
available for thermocouple temperature measurements and voltage
measurements on the 1 mV, 10 mV, or 100 mV ranges.
3
TURN FILTERS ON OR OFF
DIGITAL Ø ANALOG Ø ANA + DIG
55
Chapter 3 Features and Functions
General Measurement Configuration
• Front Panel operation: Press Shift Filter and scroll to
one of DIGITAL, ANALOG, or ANA + DIG . Press
filter type. Press the
Filter key to turn the filter on or off.
Enter to enable the
To set the digital filter averaging, use the MEASurement menu
1: DIG FILTER command.
• Remote operation: the following commands turn the filters off, on,
Using the filters in remote operation is not recommended. Use:
INPut:FILTer:STATe OFF
to disable the filters. The digital filter is also disabled following a
remote reset (*RST) or preset (SYStem:PREset).
Use the math operation (AVERage) commands to obtain a single
reading that is the average of a number of readings (see page 148). An
example program is shown on page 221 in chapter 6.
If you attempt to use the digital filter in remote operations, be aware
of the following special conditions:
The digital filter requires a number of readings to settle (10, 50 or
100). The digital filter readings only occur when the meter is triggered
to take a reading. In remote operation a reading is only taken when
requested. For example, if you have set the digital filter to MEDIUM
(50 readings), you will need to request (trigger) 50 readings from the
remote interface before the filter can settle. It is also possible that
after 50 readings the digital filter will still not be settled as described
in the next paragraph.
The STATus:OPERational:CONDition register contains a bit that
indicates the settled/unsettled state of the digital filter on a per
reading basis. See page 169 in Chapter 4. You should check the state
of this bit for each reading to ensure the digital filter count has not
reset (filter reset operation is described in the next section).
56
Chapter 3 Features and Functions
General Measurement Configuration
Digital Filter
• The digital filter is a moving average (boxcar) filter. Equal weighting
is applied to all readings to calculate the displayed reading.
• Three digital filters are available:
SLOW (average last 100 readings),
MEDIUM (average last 50 readings), or
FAST (average last 10 readings)
• When the digital filter is enabled, the ‘Filt’ annunciator flashes until
the required number of readings are obtained. The measurement
displayed before the ‘Filt’ annunciator is steady is the average of the
number of readings taken so far. For example, with the filter set to
MEDIUM (average of 50 readings), the measurement displayed after
20 readings will be the average those 20 readings. The ‘Filt’
annunciator will flash until 50 readings have been taken.
• The filter count resets (and the ‘Filt’ annunciator will flash) if the
function, range, integration time, or channel is changed.
• The filter count resets if a reading is outside a set range of the
average. The amount of input reading change required to reset the
filter count is:
3
1mV RangeAll Other Ranges
FAST
MEDIUM
SLOW
Analog Filter
The analog filter is an 11 Hz, two-pole low pass filter. Use this filter to remove
line frequency noise from the input measurement.
• The filter requires approximately 300 milliseconds after the input is
present to settle. Using the filter, therefore, slows the measurement
rate significantly.
• Use the analog filter only when line power noise (50/60 Hz) is causing
an overload (OVLD) condition. See page 65.
• The analog filter is available for voltage and thermocouple
temperature measurements. The analog filter is not available for
resistance measurements.
• The analog filter is available for the 1 mV, 10 mV, and 100 mV ranges.
± 400 ppm of range± 100 ppm of range
± 700 ppm of range± 300 ppm of range
± 2000 ppm of range± 1000 ppm of range
57
Chapter 3 Features and Functions
General Measurement Configuration
Integration Time
Integration time is the period during which the meter’s analog-to-digital
(A/D) converter samples the input signal for a measurement. Integration
time affects the measurement resolution (for better resolution, use a
longer integration time), and measurement speed (for faster
measurements, use a shorter integration time).
• Integration time is specified in number of power line cycles (
NPLCs).
The choices are 0.02, 0.2, 1, 2, 10, 20, 100, or 200 power line cycles.
The default is 10 NPLCs.
• The integration time is stored in volatile memory; the meter selects 10
NPLC when power has been off or after an interface reset.
• Only the integer number of power line cycles (1, 2, 10, 20, 100 or 200
NPLC) provide normal mode (line frequency noise) rejection.
• Integration time is related to the maximum number of digits the
meter will display. You can always choose to show fewer digits than
1
⁄
the maximum allowed (the minimum number of digits shown is 4
NPLC
0.02
0.2
1
2
10
20
100
200
Digital Filter Off
Max Digits Shown
1
4
⁄
5
2
1
5
⁄
6
2
1
6
⁄
7
2
1
⁄
7
6
2
1
6
⁄
7
2
1
7
⁄
7
2
1
7
⁄
7
2
1
⁄
7
7
2
Digital Filter On
Max Digits Shown
1
⁄
2
1
⁄
2
1
⁄
2
1
⁄
2
1
⁄
2
1
⁄
2
1
⁄
2
1
⁄
2
).
2
58
Chapter 3 Features and Functions
General Measurement Configuration
• Front Panel operation: Integration time can be set indirectly when
you select the number of digits (See page 43).
You can also set the integration time in the MEASurement menu with
the 2: INTEGRATE command.
• Remote operation: Refer to the table on page 129. Use one of the
Integration time and resolution are related. The CONFigure and
MEASure? commands can also set the integration time, see page 129.
Integration Time is local to the selected function. This means that you
can select the integration time for each function independently. The
meter remembers the integration time when you switch between
functions.
3
59
Chapter 3 Features and Functions
General Measurement Configuration
Reducing Measurement Noise
At high resolutions and low measurement levels, measurement noise can
become an important factor in the accuracy of your measurements. One
possible source of measurement noise is the wiring and cabling in the test
setup. These noise sources and their effects are discussed in Chapter 7,
"Measurement Tutorial". A smaller level of noise is also inherent in the
meter itself.
To minimize the effects of noise on your measurements, use the highest
possible integration time (NPLC) and average the readings. Faster
reading rates are more susceptible to noise. Choosing the best NPLC and
averaging method for your measurement requires a tradeoff between the
reading rate and the level of acceptable noise in the resulting
measurement.
The Integration Time (NPLC) and averaging can be multiplied to create
an equivalent NPLC. The equivalent NPLC defines the reading rate and
characterizes the expected noise level. Averaging can be performed in the
meter by using either the digital filter or by using the statistics (Stats) mode
of operation.
Measurement noise characteristics of the meter closely match theoretical
gaussian noise distributions. The graph on the next page shows typical
input noise measurements for the 1 mV range with the low thermal
shorting plug applied to the input terminals. Other functions and ranges
will exhibit similar noise reduction characteristics versus NPLC.
For example, if you set the NPLC to 20 and enable the digital filter
set to FAST (10 reading average), the equivalent NPLC is 200. At an
equivalent NPLC of 200, the reading rate is approximately 6 readings per
minute (digital filter fully settled) and the typical noise level expected is
1.6 nVrms.
Alternately, if you take readings every 5 minutes, you can set an
equivalent NPLC of 5000 to minimize the noise in the measurement.
An equivalent NPLC of 5000 could be obtained by setting the integration
time to 100 NPLC and enabling the digital filter set to MEDIUM (50
reading average). At this equivalent NPLC, a typical noise level
of 0.35 nVrms can be expected.
60
nV rms Noise
Chapter 3 Features and Functions
General Measurement Configuration
30 minute measurement, ±0.5° C, zero input, rms noise (typical)
Equivalent NPLC
Approximate
Readings per second
Readings per minute
Minutes per reading
Reading Rate
3
• Front Panel operation: Integration time can be set indirectly when
you select the number of digits (See page 43).
You can also set the integration time in the MEASurement menu with
the 2: INTEGRATE command.
Enable the digital filter and set the filter reading average as described
on page 55.
• Remote operation: Set the integration time for the measurement
function. If possible, use NPLC 200. If you use the CONFigure or
MEASure commands, refer to the table on page 129.
Use the math operations (AVERage) to average the readings or
average the readings in your program. An example program that uses
the math operations to average readings is shown on page 221 in
Chapter 6. Do not use the digital filter to average readings, see page 56.
61
Chapter 3 Features and Functions
General Measurement Configuration
Number of Digits Displayed
Front Panel operation only. See also “Integration Time,” on page 58.
• You can set the number of digits shown in the display to 4, 5, 6, or 7
1
⁄
full digits, plus a “
” digit which can only be a “0” or “1”.
2
• You can set both the number of digits displayed and the Integration
Time in a single operation by pressing digits keys (see page 43).
• You can change the number of digits displayed without disturbing
the Integration Time settings by pressing the
< or > keys
(see page 43).
1
⁄
• The number of digits is set to 6
digits at power-on and after either a
2
remote interface reset or preset.
• The number of digits shown is dependent upon integration time and
filter settings. The meter will not allow you to show more digits than
the practical measurement capability of the meter. You can, however,
reduce the number of digits shown. The minimum number of digits
1
⁄
available is 4
. (See the table on page 58).
2
62
7 digits
Chapter 3 Features and Functions
General Measurement Configuration
1
⁄
" digit
"
2
6 digits
1
⁄
" digit
"
2
5 digits
1
⁄
"
2
"digit
0.003,256,4 ΩThis is the 1 ohm range,
1
⁄
digits are displayed.
7
2
0.216,569 mVThis is the 1 mV range,
1
⁄
digits are displayed.
6
2
-045.231 mVThis is the 100 mV range,
1
⁄
digits are displayed.
5
2
3
63
Chapter 3 Features and Functions
General Measurement Configuration
Ranging
You can let the meter automatically select the range using autoranging
or you can select a fixed range using manual ranging. Autoranging is
convenient because the meter automatically selects the appropriate
range for each measurement. However, you can use manual ranging for
faster measurements since the meter does not have to determine which
range to use for each measurement.
• The selected mode (auto or manual range) is stored in volatile
memory; the meter returns to autoranging when power has been off or
after a remote interface reset.
• Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
• If the input signal is greater than the present range can measure, the
meter gives an overload indication: “OVLD” from the front panel or
“9.90000000E+37” from the remote interface.
• For ratio and difference measurements, the specified range is
independent for both Channel 1 and Channel 2.
^ v Auto/ Man • Front Panel operation: Press a front-panel RANGE key to set
manual ranging and to up or down range. Press the Auto/Man key to
return to autoranging. See also “To Select a Range,” on page 42.
• Remote operation: Range is set with parameters in the MEASure
and CONFigure subsystems. For example:
CONF:VOLT:DC 10.0
Range can be set directly with the SENSe subsystem. For example:
SENS:VOLT:DC:RANGe 0.001
SENS:RES:RANGE:AUTO
Sets the 10 Volt DC ra nge
Sets the 1 mV DC range
Sets auto ranging on
2-wire resistance
measurements
64
Chapter 3 Features and Functions
General Measurement Configuration
Overload Detection
The HP 34420A uses both an analog and digital method to generate an
overload condition and display the OVLD message. An understanding of
the cause of the overload can help you make more accurate
measurements.
Digital Overload Detection
The digital overload detect occurs when the magnitude of the
measurement is greater than 120% of the Range. For example,
±
on the 1 volt Range, signal levels of greater than
an OVLD in the display.
Analog Overload Detection
Another type of overload can result in measurement errors that are hard
to identify. For example if an overload signal only occurs for a portion of
the measurement period, the input amplifier could become non-linear
resulting in an inaccurate reading. The displayed result may appear to be
correct because the overload signal only occurred during a small portion
of the measurement cycle. The HP 34420A detects this problem by
providing comparators on the input amplifier. If signals large enough to
overload the input amplifier are detected, the display will read OVLD.
The comparators detect signals greater than 140% of the range setting.
1.2 volts will cause
3
If an overload occurs, two solutions are possible.
• Setting the meter to the next higher range will provide 10 times
improvement in the overload signal handling capability. For
example,if you are measuring on the 1 mV range and have 2 mV rms
of power line (50 or 60 Hz) pickup, the noise overload comparators
would sense the peaks of the signals and an OVLD message would be
displayed. If you set the range to 10 mV, the resolution (with 7 digits)
is still 1 nanovolt, and the reading rate is faster than if ANALOG
FILTER had been turned on. Additionally, the 10 mV range noise is
almost as low as the 1 mV range.
• Alternately, if using the 1 mV, 10 mV, or 100 mV ranges, turn on the
ANALOG FILTER. The reading rate slows down with the filter on. In
the case of a larger overloads (for example, 20 mV rms noise on the
1 mV range) using the ANALOG FILTER would give a lower noise
result than changing to the 100 mV range.
65
Chapter 3 Features and Functions
Voltage Measurement Configuration
Voltage Measurement Configuration
Input Channels
The meter has two independent input channels for measuring dc Volts.
You can make measurements on either or both channels, measure the
difference between the channels, or measure the ratio between the
channels. Each channel has an independent settings for null and range.
The figure on page 19 shows connections for two input channels.
• Channel 1 is the default input channel. Both channels use the same
integration time setting.
• The maximum measurable input on Channel 1 is 120V, on
Channel 2 is 12 V.
• For two channel operation, the maximum voltage difference betweenChannel 1 LO and Channel 2 LO is ±150 V.
• For either channel, the maximum non-destructive input is ±150 V.
• If you enable input filters, the filters are active for both input
channels. Switching between the channels will reset the digital filter
reading count (see page 57).
Channel • Front Panel operation: Toggle the input channel by pressing the
Channel key on the front panel. The CH1 or CH2 annunciator will
light in the display. The meter keeps range and null settings
independently for each channel.
Scan
Shift Channel Choose the scan mode to alternately show the CH1 and CH2 readings.
Disable the digital filter when using the scan mode. In scan mode, the
meter alternates between the CH1 and CH2 readings every 3 seconds
or for each measurement (whichever is longer).
66
Chapter 3 Features and Functions
Voltage Measurement Configuration
• Remote operation: The CONFigure and MEASure subsystems use
an optional parameter to specify the input channel as either 1 or 2:
*Difference Null is only available from the Front Panel.
67
Chapter 3 Features and Functions
Resistance Measurement Configuration
Resistance Measurement Configuration
The meter can make 2-wire or 4-wire ohms measurements. The meter
can compensate for voltages in the resistive circuit being measured using
offset compensation. You can also choose to either limit the power applied
or limit the open circuit voltage applied during a resistance
measurement.
The test current sourced by the meter depends upon the range and
conditions as shown below:
RangeTest current
2-wire or 4-wire
measurements
1Ω
10 Ω
100 Ω
1 kΩ
10 kΩ100 µA10 µA
100 kΩ*10 µA5 µA
1 MΩ*5 µA5 µA
* Offset Compensation is not applied to measurements on these ranges
10 mA10 mANot Available
10 mA10 mA1 mA
10 mA1 mA
1 mA
4-wire measurements
Low PowerVoltage Limited
100 µA
100 µA
Not Available
Not Available
Not Available
Not Available
68
Chapter 3 Features and Functions
Resistance Measurement Configuration
Offset Compensated Measurements
Offset compensation removes the effect of any voltages in the circuit
being measured. The technique involves taking two measurements, one
with the current source turned on, and one with the current source
turned off and computing the difference. A description of this operation isgiven on page 259.
• Front Panel operation: Offset compensation is on at power on.
Turn offset compensation off under the 3: OCOMPΩ command in
the MEASure menu.
• Remote operation: Offset compensation is off following a remote
reset or preset. Use the following command to change the state of
offset compensation:
Low Power measurements use a reduced output measurement current to
limit the power dissipation in the circuit being measured.
3
• Low Power measurements apply only to 4-wire ohms measurements.
• When the low power measurement mode is enabled, the ‘LP’
annunciator is on in the display.
• Setting the Voltage Limited measurement mode will cancel the
Low Power mode.
• Front Panel operation: set low power mode under the
4: LOW POWER Ω command in the MEASure menu.
• Remote operation: use the following command to set the low power
ohms mode:
SENSe:FRESistance:POWer:LIMit {OFF|ON}
69
Chapter 3 Features and Functions
Resistance Measurement Configuration
Voltage Limited Measurements
Voltage limited resistance measurements clamp the open circuit voltage
to predefined limits.
• Voltage limits apply only to 4-wire ohms measurements.
• Voltage limited resistance measurements are only available on
the 10 Ω and 100 Ω ranges. If a manual range is selected in either
2-wire or 4-wire ohms, and then Voltage Limited measurements are
enabled, the range will be set to the highest valid range (either 10 Ω
or 100 Ω ).
• The open circuit voltage can be limited to 20 mV, 100 mV, or 500 mV.
• Setting the Low Power measurement mode will cancel the
Voltage Limited mode.
• Front panel operation: Enable or disable the voltage limit under the5: LOW VOLTΩ command in the MEASure menu.
Set the voltage limit under the 6: LoV LIMITΩ command in the
MEASure menu.
• Remote operation: the following command enables or disable the
low voltage limit mode:
SENSe:FRESistance:VOLTage:LIMit {OFF|ON}
Set the limit with the following command:
value
SENSe:FRESis tanc e:VO LTag e: LIMi t:VA Lue {<
70
>|MIN|MAX}
Chapter 3 Features and Functions
Temperature Measurement Configuration
Temperature Measurement Configuration
This section contains information to help you configure the meter for
making temperature measurements. To measure temperature, you will
need a temperature transducer. Transducer descriptions and specific
comments about their usage are given on page 263. The meter supports
three general categories of transducer:
• RTD (Resistive Temperature Device)
• THERM (Thermistor)
• T/C (Thermocouple)
The meter supports specific types of transducers in each category:
3
RTD’s
R
= 4.9 Ω to 2.1 kΩ
0
α = 0.00385 (DIN/IEC 751)
α = 0.00391
Once you set the measurement units, probe type, and transducer type,
the setting is stored in non-volatile memory. This makes the last probe
type used the default probe type.
ThermistorThermocouples
5k Ω
B, E, J, K, N, R, S, T
Measurement Units
• The meter can report temperature in °C (Celsius), °F (Fahrenheit),
or K (Kelvins).
• The measurement units set are stored in non-volatile memory.
• Front Panel Operation: Set the units under the 2: UNITS command
in the TEMPerature menu.
• Remote operation: Set the units using the following command:
UNIT:TEMPerature {C|F|K}
71
Chapter 3 Features and Functions
Temperature Measurement Configuration
RTD Measurements
• The meter supports RTDs with α = 0.00385 (DIN/IEC 751) or
α = 0.00391. The nominal value of the RTD (R
of 4.9 Ω to 2.1 kΩ. RTD Connections are shown on page 21.
• The meter makes a 4-wire measurement for RTD’s. All four input
leads must be connected.
• Front Panel operation: Set RTD as the probe type under the 1: PROBE TYPE command in the TEMPerature menu.
Select the RTD type under the 3: RTD TYPE command in the
TEMPerature menu.
) must be in the range
0
Enter the 0° C resistance of the RTD (R
in the TEMPerature menu.
• Remote operation: Use the following commands for temperature
measurements with RTD’s:
• The thermocouple must be connected to the Channel 2 inputs. Be sure
to observe the correct polarity. Thermocouple connections are shown on
page 22.
• Thermocouple measurements require a reference junction
temperature. You can input a known fixed junction temperature, use
an external thermistor temperature, or use the internal thermistor
temperature as the reference junction temperature.
The accuracy of the measurement is very dependent upon the
thermocouple connections and the reference junction temperature.
Use a fixed temperature reference for highest accuracy measurements.
The internal thermistor reference produces the lowest accuracy
measurements. See page 265 for a discussion of the reference junction
temperature measurements.
• Front Panel operation: Set T/C as the probe type under the
1: PROBE TYPE command in the TEMPerature menu.
3
Select the thermocouple type under the 4: T/C TYPE command in the
Temperature menu.
Set the reference junction temperature source under the
5: COLD JUNCT command in the TEMPerature menu.
If you used a FIXed reference junction temperature, set the reference
temperature under the 6: JUNCT TEMP command in the TEMPerature
menu. The number entered must be in the range of –1° C to 55° C.
• Remote operation: use the following commands for temperature
There are two math operations available, only one of which can be
enabled at a time. You can choose to either scale the readings as they are
taken, or keep statistics on a group of readings. The selected math
operation remains in effect until you disable it, change functions, turn off
the power, or perform a remote interface reset.
Scaling
• The scaling is applied as:
reading = gain ∗ (measurement − offset)
Scale
Shift Stats • Front Panel operation: Set the values of gain and offset in the
MATH menu. Enable scaling by pressing the scale key. The Stats
annunciator will light in the display.
• Remote operation: the math operations and registers are controlled
using commands within the CALCulate command subsystem. First,
select the math operation you want to use
CALCulate:FUNCtion SCALe
Then, set the values for gain and offset.
gain
CALCulate:SCALe:GAIN <
CALCulate:SCALe:OFFSet <
Finally, enable the selected math function by turning the math state on:
CALCulate:STATe ON
74
>
offset
>
Chapter 3 Features and Functions
Math Operations
Statistics
• After you enable statistics, the first reading that the meter takes is
stored as both the minimum and maximum value. The minimum is
replaced with any subsequent value that is less. The maximum is
replaced with any subsequent value that is greater.
• All values are stored in volatile memory; the meter clears the values
when stats is turned on, when power has been off, or after a remote
interface reset or preset.
• When enabled, the following statistics are available:
MINminimum reading
MAXmaximum reading
AVERarithmetic average of all readings
STD DEVstandard deviation of all readings
PEAK – PEAK peak to peak value of minimum and maximum
COUNTthe total number of readings
• Front Panel operation: Enable the statistics by pressing
Stats key. Read the statistics gathered in the MATH menu
the
under the STATS command.
3
After reading the stored statistics, press
return to the measurement and continue taking statistics.
Shift < (Menu On/Off) to
75
Chapter 3 Features and Functions
Math Operations
• Remote operation: the math operations and registers are controlled
using math operation commands. First, select the math operation you
want to use
CALCulate:FUNCtion AVERage
Then, enable the selected math function by turning the math state on:
CALCulate:STATe ON
When you have finished collecting readings, obtain the desired statistics:
After reading the statistics, supply another trigger to continue taking
measurements and collecting statistics. Measurements are not taken
while reading the results.
If you are interested in the statistics on a number of readings, but not
the readings themselves, you can use the following command:
DATA:FEED RDG_STORE, ""
The null value parameter in the command tells the meter not to store
the individual readings. The meter does, however, accumulate
statistics on the readings. This method removes the 1024 reading
limitation and you can calculate statistics on any number of readings.
See page 150 for more information.
76
Chapter 3 Features and Functions
Null (Relative) Operation
Null (Relative) Operation
When making null measurements, also called relative, each reading is the
difference between the input signal and a stored null value. You could,
for example, make a more accurate two-wire ohms measurement by
shorting the test leads and pressing
resistance. See also “To Make Null (Relative) Measurements,” on page 45.
Reading = measurement – null value
• The null value is adjustable and you can set it to any value between
0 and ± 120% of the highest range, for the present function.
Null to remove the test lead
• The null value is stored in volatile memory; the value is cleared when
power has been off or after a remote interface reset.
• Each function and channel has an independent null value. For voltage
measurements, each input channel has an independant null.
Temperature measurement null is applied to all temperature
measurements (if you change the probe type, null is still enabled and
applied). The null for resistance measurements applies to both 2-wire
and 4-wire measurements.
• The null value is stored in the meter’s Null Register. There are two
ways you can specify the null value. First, you can enter a specific
number into the register from the front-panel menu or from the remote
interface. Any previously stored value is replaced with the new value.
If you are operating the meter from the front panel, entering a null
value also turns on the null function.
The second way to enter the null value is to let the meter store the
first reading in the register. Press the
displayed measurement the null value and enable the Null operation.
After you enable null, the first reading displayed will be zero. If you
entered a number into the register, as described in the paragraph
above, the first reading does not overwrite the stored value.
• Since an independent null value is stored for each input channel,
difference or ratio functions also use these null values. Additionally,
when using the difference function, an additional difference null can
be applied (the difference null is only available from the front panel).
See page 67 for a description of how null is applied to difference and
ratio functions.
Null key to make the
3
77
Chapter 3 Features and Functions
Null (Relative) Operation
• Front-panel operation: After enabling null, you can edit the stored
null value by pressing
value is replaced with the new value. Turning on the menu does not
disable the null operation; the meter will resume taking
measurements when you turn off the menu.
When null is enabled, the Null annunciator lights in the display. For
ratio or difference measurement, if either input channel has a null
enabled, the Null annunciator will flash. If you apply the difference
null, the Null annunciator will light (not flash).
• Remote interface operation: You can use the following commands
to make null measurements. Null must be enabled before you can store
a value in the Null Register.
SENSe1 | SENSe2
:VOLTage:DC
:NULL {OFF|ON}
:NULL {<
[SENSe:]
FRESistance | RESistance
:NULL { OFF | ON }
:NULL {<
TEMPerature
:NULL { OFF | ON }
:NULL {<
Shift > (Menu Recall). Any previously stored
value
value
value
>|MIN|MAX}
>|MIN|MAX}
>|MIN|MAX}
To enable null and set a null value (0.10) for the 2-wire resistance
function execute the following commands in the order shown.
SENS:RES:NULL ON
SENS:RES:NULL 0.10
You can use the first reading as a null value from the remote interface
by first setting the function, range, and resolution and then executing
the following command:
[SENSe:]NULL [{OFF | ON | ONCE}]
78
Chapter 3 Features and Functions
Reading Hold
Reading Hold
The reading hold feature allows you to capture and hold a stable reading
on the front-panel display. This is especially useful in situations where
you want to take a reading, remove the test probes, and have the reading
remain on the display. When a stable reading is detected, the meter
beeps and holds the reading on the display.
The reading hold feature is available only from the front panel. If you go
to remote when reading hold is enabled, the meter ignores it; when you
return to local (front panel), reading hold is enabled again.
• Reading hold has an adjustable sensitivity band (adjustable only from
the front panel) to allow you to select which readings are considered
stable enough to be displayed. The band is expressed as a percent of
reading, on the selected range. The meter will capture and display a
new value only after three consecutive readings are within the band.
Select one of these values: 0.01%, 0.10% (default), 1.00%, or 10.0%
of reading.
For example, assume that the 1.00% band is selected and a 5 volt
signal is applied to the meter. If three consecutive readings are
between 4.95 volts and 5.05 volts, the meter will beep and display a
new reading.
• The sensitivity band is stored in volatile memory; the meter sets the
band to 0.10% when power has been off or after an interface reset.
• If the meter is in autorange when you enable reading hold, it will
autorange to the correct range. If the meter is in the manual range
mode, the same fixed range will be used for reading hold.
• For certain applications, it may be useful to use reading hold with
reading memory. See also “Reading Memory,” on page 90.
• Front Panel operation: After enabling reading hold, you can select
a different sensitivity band by pressing
Shift > (Menu Recall).
3
1: READ HOLD
79
Chapter 3 Features and Functions
Triggering
Triggering
The meter’s triggering system allows you to generate triggers either
manually or automatically, take multiple readings per trigger, and insert
a delay before each reading. Normally, the meter will take one reading
each time it receives a trigger, but you can specify multiple readings (up
to 50,000) per trigger.
• You can trigger the meter from the front panel, with an external
trigger, or use autotriggering. Single triggering takes one reading each
time you press the
triggering, but the meter waits for a pulse on the rear-panel Ext Trig
(external trigger) terminal before taking a reading. Autotriggering
takes continuous readings at the fastest rate possible for the present
configuration. See also “To Trigger the Meter,” on page 47.
Single key. External triggering is like single
• Press
Triggering the meter from the remote interface is a multi-step process
that offers triggering flexibility.
• First, you must configure the meter for the measurement by selecting
the function, range, integration time, etc.
• Then, you must specify the source from which the meter will accept
the trigger. The meter will accept a software (bus) trigger from the
remote interface, a hardware trigger from the Ext Trig terminal, or an
immediate internal trigger.
• Then, you must make sure that the meter is ready to accept a trigger
from the specified trigger source (this is called the wait-for-trigger
state).
• Finally, to make the measurement, you must provide an INITiate,
READ?, or MEASure command.
The diagram on the next page shows the meter’s triggering system.
Shift Auto/Hold to return the meter to auto triggering.
80
Initiate Triggering
MEASure?
READ?
INITiate
Trigger Source
TRIGger:SOURce IMMediate
TRIGger:SOURce EXTernal
TRIGger:SOURce BUS
front-panel "Single" key
Trigger Delay
TRIGger:DELay
Chapter 3 Features and Functions
Triggering
Idle
State
Wait-for Trigger
State
3
Delay
Sample (*)
Annunciator
Measurement
Sample
Triggering the meter is a multi-step process.
YesYes
Sample
Count >1
No
Trigger
Count >1
No
81
Chapter 3 Features and Functions
Triggering
Trigger Source Choices
You must specify the source from which the meter will accept a trigger.
• The trigger source is stored in volatile memory; the source is set to
autotrigger (front panel) or immediate (remote interface) when power
has been off or after a remote interface reset.
• Front Panel operation: the meter will accept a either a front panel
single trigger or a hardware trigger from the Ext Trig terminal, or
continuously take readings using autotrigger. At power-on,
autotriggering is used.
• Remote operation: the meter will accept a software (bus) trigger, a
hardware trigger from the Ext Trig terminal, or an immediate internal
trigger. See "Triggering" on page 80 for more information.
Autotriggering In the autotrigger mode (front panel only), the meter
continuously takes readings at the fastest rate possible for the present
configuration. This is the power-on trigger source for front-panel
operation.
Single Triggering In the single trigger mode (front panel only), you
can manually trigger the meter by pressing
reading, or the specified number of readings (sample count), each time
you press the key. The
‘Trig’ annunciator turns on when the meter is
waiting for a trigger.
Single . The meter takes one
Pressing the
Single key also allows the meter to accept triggers on the
Ext Trig terminal.
The front-panel
Single key is disabled when in remote.
82
Chapter 3 Features and Functions
Triggering
External Triggering In the external trigger mode, the meter will
accept a hardware trigger applied to the Ext Trig terminal. The meter
takes one reading, or the specified number of readings (sample count),
each time Ext Trig receives a low-true pulse.
See also “External Trigger Terminal,” on page 97.
• The meter buffers one external trigger. This means that if the meter is
taking a reading and another external trigger occurs, that trigger is
accepted (a “Trigger ignored” error is not reported). After the reading
in progress is complete, the stored trigger satisfies the trigger source
and then the trigger is issued.
• Front-panel operation: The external trigger mode is like the single
trigger mode except that you apply the trigger to the Ext Trig
terminal. Pressing
enables the external trigger mode. The
when the meter is waiting for an external trigger.
Single to enable the single trigger mode also
‘Trig’ annunciator turns on
3
The front-panel
• Remote operation: Use the following command to enable external
trigger.
TRIGger:SOURce EXTernal
Single key is disabled when in remote.
83
Chapter 3 Features and Functions
Triggering
Internal Triggering In the internal trigger mode (remote interface
only), the trigger signal is always present. When you place the meter in
the wait-for-trigger state, the trigger is issued immediately. This is the
power-on trigger source for remote interface operation.
To select the internal trigger source, send the following command. The
CONFigure and MEASure? commands automatically set the trigger
source to IMMediate.
TRIGger:SOURce IMMediate
After setting the source to IMMediate, an INITitate or a READ?
command will trigger the measurement.
Software (Bus) Triggering The bus trigger mode is available only
from the remote interface. This mode is similar to the single trigger mode
from the front panel, but you trigger the meter by sending a bus trigger
command.
• To select the bus trigger source, send the following command.
TRIGger:SOURce BUS
• To trigger the meter from the remote interface (HP-IB or RS-232),
send the *TRG (trigger) command. The *TRG command will not be
accepted unless the meter is in the wait-for-trigger state.
• You can also trigger the meter from the
HP-IB interface by sending the
IEEE-488 Group Execute Trigger (GET) message. The meter must be
in the wait-for-trigger state. The following statement shows how to
send a GET from a Hewlett-Packard controller.
TRIGGER 722 (group execute trigger)
84
Chapter 3 Features and Functions
Triggering
The Wait-for-Trigger State
After you have configured the meter and selected a trigger source,
you must place the meter in the wait-for-trigger state. A trigger will not
be accepted until the meter is in this state. If a trigger signal is present,
and if meter is in the “wait-for-trigger” state, the measurement sequence
begins and readings are taken.
The “wait-for-trigger” state is a term used primarily for remote interface
operation. From the front panel, the meter is always in the “wait-fortrigger” state and will accept triggers at any time, unless a measurement
is already in progress.
You can place the meter in the “wait-for-trigger” state by executing any of
the following commands from the remote interface.
MEASure?
READ?
INITiate
Halting a Measurement in Progress
You can send a device clear at any time to halt a measurement in
progress and place the meter in the “idle state.” The following statement
shows how to send a device clear from a Hewlett-Packard controller if you
are using the HP-IB interface. See also page 187.
CLEAR 722 (device clear)
For operation with the RS-232, you must establish a means to send a
device clear (for example, by pressing ^C (Control C)).
A device clear does not affect the configuration of the triggering system.
The trigger source, sample count, trigger delay, and number of triggers
are not changed.
If you have set the meter to infinite sampling (TRIGger:COUNt
INFinity), the device clear is the only way to halt the measurements.
3
85
Chapter 3 Features and Functions
Triggering
Number of Samples
Normally, the meter takes one reading (or sample) each time it receives
a trigger from the selected trigger source (if the meter is in the
wait-for-trigger state). You can, however, instruct the meter to take
multiple readings for each trigger received.
• Number of samples: 1 to 50,000. The default is 1 sample per trigger.
• The sample count is stored in volatile memory; the meter sets the
sample count to 1 when power has been off or after a remote interface
reset. The CONFigure and MEASure? commands automatically set
the sample count to 1.
• Front Panel operation: set the number of samples per trigger in the 3: N SAMPLES command under the TRIGger menu.
• Remote operation: use the following command:
value
SAMPle:COUNt {<
>|MIN|MAX}
Number of Triggers
Normally, the meter will accept only one trigger before returning to the
“idle” trigger state. You can, however, instruct the meter to accept
multiple triggers.
This feature is available only from the remote interface. If you set the
trigger count and then go to local (front panel), the meter ignores the
trigger count setting; when you return to remote, the trigger count returns
to the value you selected.
• The selected number of triggers is stored in volatile memory; the
meter sets the trigger count to 1 when power has been off or after a
remote interface reset or preset. The CONFigure and MEASure?
commands automatically set the trigger count to 1.
• The number of triggers can be set to any number between 1 (MIN) and
50000 (MAX) or can be set to be INFinity. When the trigger count is
set to be infinite, use the READ? command to collect the readings.
Send a device clear to stop the measurements. See page 187 for
information about device clear.
value
TRIGger:COUNt {<
>|MIN|MAX|INFinity}
86
Chapter 3 Features and Functions
Triggering
Trigger Delay
You can insert a delay between the trigger signal and each sample that
follows. This may be useful in applications where you want to allow the
input to settle before taking a reading, or for pacing a burst of readings.
If you do not specify a trigger delay, the meter automatically selects a
delay for you.
• Delay range: 0 to 3600 seconds. The default trigger delay is
automatic; the delay is determined by function, range, and integration
time (see also “Automatic Trigger Delays,” on page 89).
• The trigger delay is stored in volatile memory; the meter selects an
automatic trigger delay when power has been off or after a remote
interface reset. The CONFigure and MEASure? commands set the
trigger delay to automatic.
• If you specify a delay other than automatic, that same delay is used for
all functions and ranges.
• If you have configured the meter to take more than one reading per
trigger (sample count > 1), the specified trigger delay is inserted
between the trigger and each reading.
3
• Front Panel operation: You can use an automatic trigger delay or
you can specify a delay in seconds. Set the delay time in the
2: TRIG DELAY command of the TRIGger menu.
If an automatic trigger delay is enabled, “AUTO” is displayed
momentarily before the actual number of seconds is displayed.
AUTO
87
Chapter 3 Features and Functions
Triggering
• Front Panel operation (continued)
To set the delay to 0 seconds, select the “parameter” level of the
TRIG DELAY command. Move the flashing cursor to the “units”
location on the right side of the display. Press
ZERO DELAY is reached, then press
Enter .
∨ until
ZERO DELAY
• To select the automatic trigger delay, select the “parameter” level of
the TRIG DELAY command. Move the flashing cursor to the “units”
location on the right side of the display. Press
AUTO DELAY is reached, then press
Enter .
∨ until
AUTO DELAY
• Remote operation:
You can use the following command to set the trigger delay.
TRIGger:DELay {<
seconds
>|MIN|MAX}
You can use the following command to set an automatic trigger delay.
TRIGger:DELay:AUTO {OFF|ON}
88
Chapter 3 Features and Functions
Triggering
Automatic Trigger Delays
If you do not specify a trigger delay, the meter selects an automatic delay
for you. The delay is determined by function, range, and integration time.
• DC voltage :
Range
m
V
1
m
10
V to 120 V
NPLC < 1
Trigger Delay
15 ms
1.0 ms
• Resistance (2-wire and 4-wire):
Range NPLC < 1
Trigger Delay
Ω
1
10 Ω
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
1.0 ms
1.0 ms
1.0 ms
1.0 ms
1.0 ms
4 ms
40 ms
• Temperature:
NPLC < 1
Trigger Delay
RTD
THERmistor
ThermoCouple
1.0 ms
1.0 ms
1.0 ms
NPLC ³ 1
Trigger Delay
15 ms
1.5 ms
NPLC ³ 1
Trigger Delay
1.5 ms
1.5 ms
1.5 ms
1.5 ms
1.5 ms
6 ms
60 ms
NPLC ³ 1
Trigger Delay
1.5 ms
1.5 ms
1.5 ms
3
89
Chapter 3 Features and Functions
System-Related Operations
System-Related Operations
This section gives information on topics such as reading memory, errors,
self-test, and front-panel display control. This information is not directly
related to making measurements but is an important part of operating
the meter.
Reading Memory
The meter can store up to 1024 readings in internal memory. Readings
are stored in first-in-first-out (
the first reading stored. See also “To Use Reading Memory,” on page 50.
• You can use reading memory with all functions, math operations, and
also reading hold. After you have enabled reading memory, you can
change the function or input channel. Be aware, however, that the
function labels and channel labels (
reading.
• Readings taken while reading memory is enabled are stored in volatile
memory; the meter clears the stored readings when reading memory is
turned on again, when power has been off, after a self-test, or after a
remote interface reset.
FIFO) order. The first reading returned is
V, OHM, etc.) are not stored with the
• You can use reading memory with autotrigger, single trigger, external
trigger, and reading hold. If you configure the meter for multiple
readings per trigger, the specified number of readings are stored in
memory each time a trigger is received.
90
Chapter 3 Features and Functions
System-Related Operations
• Front-panel operation: enable readings storage under the
1: RDGS STORE command in the SYStem menu. The ‘Mem’
annunciator lights in the display.
Recall the stored readings under the 2: SAVED RDGS command in the
SYStem menu. Press
memory to quickly go to the Saved Readings command.
Reading memory is automatically turned off when you go to the
“parameter” level in the menu to recall the readings.
• Remote interface operation: The INITiate command uses reading
memory to store readings prior to a FETCh? command. You can query
the number of stored readings in memory by sending the
DATA:POINts? command from the remote interface.
Shift > (Menu Recall) after enabling reading
3
91
Chapter 3 Features and Functions
System-Related Operations
Error Conditions
When the front-panel ‘ERROR’ annunciator turns on, one or more
command syntax or hardware errors have been detected. A record of up
to 20 errors is stored in the meter’s error queue. See chapter 5, “Error
Messages,” for a complete listing of the errors.
• Errors are retrieved in first-in-first-out (FIFO) order. The first error
returned is the first error that was stored. When you have read all
errors from the queue, the ERROR annunciator turns off. The meter
beeps once for each error generated.
• If more than 20 errors have occurred, the last error stored in the
queue (the most recent error) is replaced with –350 “Queue Overflow”.
No additional errors are stored until you remove errors from the
queue. If no errors have occurred and you read the error queue, the
meter responds with +0, “No error”.
• The error queue is cleared when power has been off or after a *CLS
(clear status) command has been executed.
• Front-panel operation: read the errors under the 3: ERROR
command in the SYStem menu.
ERROR annunciator is on, press Shift > (Menu Recall) to read
If the
the errors stored in the queue. The errors are listed horizontally on the
“parameter” level. The error queue is cleared when you go to the
“parameter” level and then turn off the menu.
ERR 1: -113
First error in queueError code
• Remote interface operation: use the following command to read
the errors.
SYSTem:ERRor?
Errors have the following format (the error string may contain up to
80 characters).
–113, "Undefined header"
92
Chapter 3 Features and Functions
System-Related Operations
Self-Test
A power-on self-test occurs automatically when you turn on the
multimeter. This limited test assures you that the meter is operational.
This self-test does not perform the extensive set of tests that are included
as part of the complete self-test described below.
A complete self-test runs a series of tests and takes approximately
15 seconds to execute. If all tests pass, you can have a high confidence
that the meter is operational.
• The results of the complete self-test are stored in internal reading
memory (see page 50). Memory is cleared as the self-test stores this
information. Other than clearing memory, the complete self-test
does not alter the state of the meter.
• If the complete self-test is successful, “PASS” is displayed on the front
panel. If the self-test fails, “FAIL” is displayed and the
annunciator turns on. See the HP 34420A Service Guide for
instructions on returning the meter to Hewlett-Packard for service.
• Front Panel operation: You can perform some of the tests
individually or you can perform all tests together at once. Scroll
through the parameters under the 4: TEST command in the SYStem
menu.
ERROR
3
Another way to perform the complete self-test from the front panel is
as follows: Hold down
the meter; hold down the key for more than 5 seconds. The self-test will
begin when you release the key.
• Remote operation: use the following command to initiate a self-test
and return the results:
*TST?
Returns “0” if the self-test is successful, or “1” if it fails.
Shift as you press the Power switch to turn on
93
Chapter 3 Features and Functions
System-Related Operations
Display Control
To speed up your measurement rate, or for security reasons, you may
want to turn off the front-panel display. From the remote interface, you
can also display a 11-character message on the front panel. See page 43
for information about the number of digits displayed.
• When the display is turned off, readings are not sent to the display
and all display annunciators except ‘ERROR’ and ‘Shift’ are disabled.
Front-panel operation is otherwise unaffected by turning off the
display.
• The display state is stored in volatile memory; the display is enabled
when power has been off or after a remote interface reset.
• You can display a message on the front panel by sending a command
from the remote interface. The meter can display up to 11 characters
of the message on the front panel; any additional characters are
truncated. Commas, periods, and semicolons share a display space
with the preceding character, and are not considered individual
characters. When a message is displayed, readings are not sent to the
display.
• Sending a message to the display from the remote interface overrides
the display state; this means that you can display a message even if
the display is turned off.
• Front-panel operation: set the display state under the 5: DISPLAY
command in the SYStem menu.
The display always turns on for menu operation; this means that even
when the display is turned off, you can still operate the menu.
• Remote operation: use the following commands to control the
display:
DISPlay {OFF|ON}
DISPlay:TEXT <
quoted string
disable/enab le t he d is play
display the string
>
enclosed in quotes
DISPlay:TEXT:CLEar
clear the message in the
display
The following statement shows how to display a message on the front
panel from a Hewlett-Packard controller.
OUTPUT 722; "DISP:TEXT ’HELLO’"
94
Chapter 3 Features and Functions
System-Related Operations
Comma Separators
The meter can display readings on the front panel with or without a
comma separator. This feature is available only from the front panel. See
“Menu Example 1” on page 35.
08.241,53 V 08.24153 V
With comma separator Without comma separator
(factory setting)
• The display format is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset. The
comma separator is enabled when the meter is shipped from the
factory.
• Front-panel operation: change the state of the comma display under
the 9: COMMA command in the SYStem menu.
Firmware Revision Query
The meter has three microprocessors for control of various internal
systems. You can query the meter to determine which revision of
firmware is installed for each microprocessor.
• The meter returns three numbers. The first number is the firmware
revision number for the measurement processor; the second is for the
input/output processor; and the third is for the front-panel processor.
• Front-panel operation: read the revisions under the 11: REVISION
command in the SYStem menu.
X.X-X.X-X.X
• Remote operation: use the following command to read the revision:
3
*IDN? returns “HEWLETT-PACKARD,34420A,0,X.X-X.X-X.X”
Be sure to dimension a string variable with at least 40 characters.
95
Chapter 3 Features and Functions
System-Related Operations
SCPI Language Version
The meter complies with the rules and regulations of the present version
SCPI (Standard Commands for Programmable Instruments). You can
of
determine the
sending a command from the remote interface.
SCPI version with which the meter is in compliance by
You cannot query the
• The following command returns the
SCPI version from the front panel.
SCPI version.
SYSTem:VERSion?
Returns a string in the form “YYYY.V” where the “Y’s” represent the year
of the version, and the “V” represents a version number for that year
(for example, 1994.0).
96
Chapter 3 Features and Functions
Voltmeter Complete Terminal
Voltmeter Complete Terminal
The rear-panel VM Comp (voltmeter complete) terminal provides a low-true
pulse after the completion of each measurement. Voltmeter complete and
external trigger (see below) implement a standard hardware handshake
sequence between measurement and switching devices.
Output
Approximately 2µs
External Trigger Terminal
You can trigger the meter by applying a low-true pulse to the rear-panel
Ext Trig (external trigger) terminal. To use this terminal from the remote
interface, you must select the external trigger source
(TRIGger:SOURce EXTernal).
Input
>1µs
You can use a simple switch to generate an external trigger using the
Ext Trig input as shown below.
3
97
Chapter 3 Features and Functions
Chart Output (Analog Output)
Chart Output (Analog Output)
You can use the Chart Output (Analog Output) connector on the rear
panel to run a strip chart recorder or similar instrument. The connector
provides an output voltage proportional to the measured voltage.
• The output can range from –3.00 V to +3.00 V.
• When chart output is disabled (Off), the output is set to 0 V.
• The output impedance is 1 kΩ.
• The output voltage is related to the input signal as follows:
(reading − chartnull)
span
= chart output voltage
Note that reading includes the results of measurement unit
conversion, filtering, null, and any math operation performed on the
measurement.
• The chart output state (including the span and null values) is stored
in non-volatile memory.
• span cannot have a negative value.
• The chart output has a voltage "rollover" to prevent the loss of data,
see page 100.
Examples:
See page 38 for a menu procedure to enter a number.
1) To set the chart span so that a ±1 mV input signal produces a ±1.0
Volt chart output:
(0.001 − 0.0)
span
span =
0.001
1.00
= 1.00
= 0.001
When chart null is set to 0, the value
chart output = gain ∗ reading.
98
1
can be considered the gain and
span
Chapter 3 Features and Functions
Chart Output (Analog Output)
2) To set the chart output voltage to ±3.0 V to correspond to an input
temperature range of 5° C to 45° C with a mid temperature of 25° C
(chart output is 0 V at 25° C).
(25− chart null )
span
= 0chart null = 25
and
(45− 25)
span
= 3.0span=
20
3.0
= 6.66
• Front Panel operation: Enable the chart output under the
5: CHART OUT command in the SYStem menu.
Set the chart span under the 6: CHART SPAN command in the
SYStem menu.
Set the chart offset under the 7: CHART NULL command in the
SYStem menu.
You can also press
Shift Null (Chart Null) to null the chart output
voltage based upon the displayed reading. This action places a null
value in the Chart Offset register. Pressing
Shift > (Menu Recall)
will take you to the 7: CHART NULL command where you can manually
edit the null value.
• Remote operation: use the following commands to enable and set
the chart output values.
OUTPut {OFF|ON}
OUTPut:REFerence
:OFFSet {<
number
>|MIN|MAX}
:OFFSet:NULL
:SPAN { <
number
>|MIN|MAX}
:SPAN?
:VALue {<
number
>|MIN|MAX}
:GAIN?
3
99
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.