Measurement Characteristics
DC Voltage
Measurement Method: Continuously
integrating multi-slope III A-D Converter
A-D Linearity: 0.00008% of reading + 0.00005%
of range
Input Resistance:
100V (Ch1 only): 10 MΩ +- 1%
1mV through 10V: > 10 GΩ, in parallel with
< 3.6 nF
Input Bias Current: <50 pA at 25 °C
Injected Current: <50 nA pp at 50 or 60 Hz
Input Protection: 150 V peak any input terminal
to Channel 1 LO, continuous
Channel-to-channel switching error (typical):
3 nV
Channel Isolation: Isolation between input
channels >10
10
Ω
Earth Isolation: 350 V peak any input terminal
to earth. Impedance from any input terminal to
earth is >10 GΩ and <400 pF
Maximum Voltage: Channel 1 LO to Channel 2
LO, 150V peak
Resistance
Measurement Method: Selectable 4-wire or
2-wire ohms. Current Source referenced to
Channel 1 LO input
Offset Compensation: Used on all ranges
except 100 kΩ and 1 MΩ. Can be turned off
if desired
Protection: 150 V peak
Open Circuit Voltage: For Resistance and Low
Power Resistance <14 V. 20 mV, 100 mV,
500 mV selectable clamp
Temperature
SPRT: ITS-90 calibrated temperature with the
range of -190°C to +660°C
Thermocouple: ITS-90 conversions of Type B,
E, J, K, N, R, S, T
Thermistor: 5 kΩ
RTD: Type a =.00385 and a =.00392. R
0
from
4.9 Ω to 2.1 kΩ. ITS -90 (IEC-751) Callendar
Van Dusen conversion.
Measurement Noise Rejection 60 (50) Hz
1
dc CMRR: 140 dB
ac CMRR: 70 dB
Integration Time Normal Mode
Rejection
2
200 plc/3.335 s (4 s) 110 dB
3
100 plc/1.675 s (2 s) 105 dB
3
20 plc/334 ms (400 ms) 100 dB
3
10 plc/167 ms (200 ms) 95 dB
3
2 plc/33.3 ms (40 ms) 90 dB
1 plc/16.7 ms (20 ms) 60 dB
<1 plc 0
System Speeds
6
Configuration Rates: 26/s to 50/s
Autorange Rate (Volts): >30/s
ASCII reading to RS-232: 55/s
ASCII reading to GPIB: 250/s
Max. Internal Trigger Rate: 250/s
Max. Ext. Trig. Rate to Memory: 250/s
Triggering and Memory
Reading HOLD Sensitivity: 10%, 1%, 0.1%,
or 0.01% of
range
Samples/Trigger: 1 to 50,000
Trigger Delay: 0 to 3600 s;
10 µs step size
External Trigger Delay: <1 ms
External Trigger Jitter: <500 µs
Memory: 10
24
readings
Math Functions
NULL (Channel 1 dcV, Channel 2 dcV,
Difference, Resistance, Temperature)
STATS (Min, Max, Average, Peak-Peak,
Standard Deviation, Number of readings)
SCALE (Allows linear scaling as y = mx+b)
CHART NULL (Establishes zero for rear panel
output)
Filter (Analog or Digital or Both)
Analog: Low pass 2 pole @ 13Hz, available
for dcV on 1 mV, 10 mV, 100 mV range
Digital: Moving average filter, 10 (fast),
50 (medium), or 100 (slow) reading averages.
Chart Out (Analog Out)
Maximum output: ± 3V
Resolution: 16 bits
Accuracy: ± 0.1% of
output + 1 mV
Output Resistance: 1 kΩ ± 5%
Update rate: once per
reading
Span and Offset: Adjustable
Standard Programming Languages:
SCPI (IEEE 488.2), Keithley 181
Accessories Included
4 ft low thermal cable with copper spade lugs,
4-wire shorting plug, user’s manual, service
manual, test report, contact cleaner, and
power cord.
General Specifications
Front Panel Connection: Shielded, low thermal,
99% copper contacts.
Power Supply: 100V/120V/220V(230V)/ 240V
± 10%.
Power Line Frequency: 45 Hz to 66 Hz and
360 Hz to 440 Hz. Automatically sensed at
power-on.
Power Consumption: 25VA peak (10W average).
Operating Environment: Full accuracy for 0 °C
to 55 °C. Full accuracy to 80% R.H. up to 30 °C.
Storage Environment: -40 °C to 75 °C.
Size: 254.4 mm W x 374.0 mm L x 103.6 mm H
(10.02” W x 14.72” L x 4.08” H)
Weight: 3 kg (6.5 lbs).
Safety: Designed to CSA, UL-1244, IEC-1010.
RFI and ESD: CISPR 11.
Operating Characteristics
4
Function Digits Integration Time Readings/s
5
dcV 71/
2
200 plc .15 (.125)
Thermocouple 7
1
/
2
100 plc .3 (.25)
6
1
/
2
20 plc 1.5 (1.25)
6
1
/
2
10 plc 3 (2.5)
5
1
/
2
1 plc 25 (20.8)
5
1
/
2
0.2 plc 100 (100)
41/
2
0.02 plc 250 (250)
Resistance 7
1
/
2
200 plc .075 (.062)
dcV1/DCV2 7
1
/
2
100 plc .15 (.125)
dcV 1-2 6
1
/
2
20 plc .75 (.625)
RTD 6
1
/
2
10 plc 1.5 (1.25)
Thermistor 5
1
/
2
1 plc 12.5 (10.4)
0.2 plc 50 (50)
4
1
/
2
0.02 plc 125 (125)
1 For 1 kΩunbalanced in LO lead.
2 For power line frequency ± - 0.1%, Filters OFF. For Digital Filter slow add 20 db, for medium or fast add 10 db for NPLC ≥1.
3 For power line frequency ± - 1%, use 80 db, for ± - 3% use 60 db.
4 Speeds are for delay 0, Display OFF, Filters OFF, Offset Compensation OFF.
5 Reading speeds for 60 Hz or (50 Hz), 100 mV through 100 V ranges. 1 mV range 30/s MAX, 10 mV range 170/s MAX, thermocouple 120/s MAX.
6 Speeds are for NPLC 0.02, Delay 0, Display OFF, Chart Out OFF.