2
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE
[2]
TEST CURRENT OR
BURDEN VOLTAGE INPUT IMPEDENCE
[3]
1 Year
23
o
C ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55 oC
VOLTAGE 100.000 mV -
10.0 MW\2%
0.025+0.008 0.0015+0.0005
1.00000 V -
10.0 MW\2%
0.025+0.006 0.0010+0.0005
10.0000 V -
10.1 MW\2%
0.025+0.005 0.0020+0.0005
100.000 V -
10.1 MW\2%
0.025+0.005 0.0020+0.0005
1000.00 V -
10.0 MW\2%
0.025+0.005 0.0015+0.0005
RESISTANCE
100.000 W
1.0 mA -
0.05+0.008
[3]
0.0060+0.0008
1.00000 kW
0.83 mA -
0.005+0.005
[3]
0.0060+0.0005
10.0000 kW 100 mA
-
0.005+0.006
[3]
0.0060+0.0005
100.000 kW 10.0 mA
- 0.05+0.007 0.0060+0.0005
1.00000 MW
900 nA - 0.06+0.007 0.0060+0.0005
10.0000 MW
205 nA - 0.25+.005 0.0250+0.0005
100.000 MW 205 nA||10MW
- 2.00+0.005 0.3000+0.0005
CURRENT 10.0000 mA < 0.2 V - 0.05+0.015 0.0055+0.0005
100.000 mA < 0.2 V - 0.05+0.005 0.0055+0.0005
1.00000 A < 0.5 V - 0.20+0.007 0.0100+0.0005
10.0000 A < 0.6 V - 0.25+0.007 0.0150+0.0005
CONTINUITY
1000 W
0.83 mA - 0.05+0.005 0.0050+0.0005
DIODE TEST
[4]
1.0000 V 0.83 mA - 0.05+0.005 0.0050+0.0005
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE
[5]
FREQUENCY
1 Year
23 oC ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55 oC
TRUE-RMS 100.000 mV 20 Hz - 45 Hz 1.0+0.1 0.02+0.02
AC VOLTAGE
[6]
45 Hz - 10 kHz 0.2+0.1 0.02+0.02
10 kHz - 30 kHz 1.5+0.3 0.05+0.02
30 kHz - 100 kHz
[7]
5.0+0.3 0.10+0.02
1.00000 V to 750.00 V 20 Hz - 45 Hz
1.0+0.1
[14]
0.02+0.02
45 Hz - 10 kHz 0.2+0.1 0.02+0.02
10 kHz - 30 kHz 1.0+0.1 0.05+0.02
30 kHz - 100 kHz
[7]
3.0+0.2
[15]
0.10+0.02
TRUE-RMS 10.0000 mA 20 Hz - 45 Hz 1.5+0.1 0.02+0.02
AC CURRENT
[8]
100.000 mA 45 Hz - 1 kHz 0.5+0.1 0.02+0.02
10.0000 A
1 kHz - 10 kHz
[9]
2.0+0.2 0.02+0.02
FREQUENCY
[10]
100 mV to 750 V < 2 Hz 0.18+0.003 0.005
< 20 Hz 0.04+0.003 0.005
20 Hz - 100 kHz
[11]
0.02+0.003 0.005
100 kHz ~ 300 kHz
[12]
0.02+0.003 0.005
10 mA to 10 A < 2 Hz 0.18+0.003 0.005
< 20 Hz 0.04+0.003 0.005
20 Hz ~ 10 kHz
[11]
0.02+0.003 0.005
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE TEST CURRENT, etc.
1 Year
23
o
C ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55oC
TEMPERATURE
-80 oC - 150 oC
5 kW thermistor probe
Probe accuracy + 0.2 oC 0.002 oC
- 110.0 oF - 300.0 oF
5 kW thermistor probe
Probe accuracy + 0.4 oF 0.0036 oF
CAPACITANCE 1.000 nF
0.75 mA
2.0+0.8 0.02+0.001
10.00 nF
0.75 mA
1.0+0.5 0.02+0.001
100.0 nF
8.3 mA
1.0+0.5 0.02+0.001
1.000 mF - 100.0 mF 83 mA
1.0+0.5 0.02+0.001
1000 mF
0.83 mA 1.0+0.5 0.02+0.001
10,000 mF
0.83 mA 2.0+0.5 0.02+0.001
[1] Specifications are for 30 minutes warm-up, 5 1/2 digit resolution and calibration temperature 18 oC - 28 oC.
[2] 20% over range on all ranges except 1000Vdc.
[3] Specifications are 2-wire ohms using Math Null. If without Math Null, add 0.2W additional error.
[4] Specifications are for the voltage measured at the input terminals only.
[5] 20% over range on all range except 750 Vac
[6] Specifications are for sinewave inputs > 5% of range. Maximum crest factor : 3 at full scale.
[7] Additional error to be added as frequency > 30kHz and signal input < 10% of range. 30kHz ~ 100kHz: 0.003% of full scale per kHz.
[8] For 12A terminal, 10A dc or ac rms continuous, > 10A dc or ac rms for 30 seconds ON and 30 seconds OFF.
[9] For 1A and 10A ranges, the frequency is verified for less than 5kHz.
[10] Specifications are for half-hour warm-up, using 0.1 second aperture. The frequency can be measured up 1Mhz as 0.5V signal to 100mV/1V ranges.
[11] For 20Hz ~ 10kHz, the sensitivity is AC input current from 10% to 120% of range except where noted.
[12] For 100kHz ~ 300kHz, the sensitivity will be 12% ~ 120% of range except 750V range.
[13] Input Impedence is in paralleled with capacitance < 120 pF.
[14] For input < 200V rms
[15] For input < 300V rms
DC CHARACTERISTICS
[1]
AC CHARACTERISTICS
[1]
TEMPERATURE and CAPACITANCE CHARACTERISTICS
[1]