Agilent 34405A Multimeter
5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price
Data Sheet
Features
120000 counts resolution
16 built-in measurement functions including temperature and capacitance
USB 2.0
SCPI compatible
Agilent IO Library Suite and DMM Intuilink connectivity software included
Affordable and Feature-Rich Measurement Tool
The 34405A represents the latest member in the Agilent’s DMM family and this expands Agilent’s offerings in the electronics measurement tools. It provides a broad range of features and measurement functions such as DC voltage, DC current, true-RMS AC voltage and AC current, 2-wire resistance, frequency, diode test and continuity which are designed to meet general industrial needs. Furthermore, its built-in thermistor sensor is able to measure temperature ranging from -80
o
C to 150oC. The true value is more evident with its capability to measure capacitance ranging from 1000pF to 10000mF. Agilent 34405A also improves efficiency and accuracy with its 6 built-in math operations: Null, dBm, dB, MinMax, Limit and Hold.
Quick Connection to the PC with USB 2.0 Interface
For those with a need to control and take preset measurements with a PC, the built­in USB 2.0 interface provides an easy and robust connection between the PC and DMM. The USB interface that is compliant with the TMC-488.2 Standards, works seamlessly with Agilent Connectivity software and can be controlled remotely via industry standard SCPI commands or through DMM Intuilink Connectivity
software. IVI-COM and LabVIEW drivers are included to ensure an easy integration with different programming environments.
Bright Display, Fast Reading Speed and Configuration Storage
When high throughput and productivity are the priority, Agilent 34405A VFD dual display feature allows users to take more than one measurement and display them simultaneously on the front panel. For speed critical applications, Agilent 34405A can take up to 19 readings/sec at 4.5 digits resolution directly to the PC. In addition, the user can configure and store complete instrument setups and recall them at anytime from any of the four built­in storing states.
Rugged and Reliable
The 34405A is designed and tested according to major Safety and Regulatory Standards. In addition, the shock absorbing bumpers is designed to prevent physical damage from your day-to-day use.
Go to the WEB for more information on Agilent’s DMM. Visit www
.agilent.com
2
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE
[2]
TEST CURRENT OR BURDEN VOLTAGE INPUT IMPEDENCE
[3]
1 Year
23
o
C ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55 oC
VOLTAGE 100.000 mV -
10.0 MW\2%
0.025+0.008 0.0015+0.0005
1.00000 V -
10.0 MW\2%
0.025+0.006 0.0010+0.0005
10.0000 V -
10.1 MW\2%
0.025+0.005 0.0020+0.0005
100.000 V -
10.1 MW\2%
0.025+0.005 0.0020+0.0005
1000.00 V -
10.0 MW\2%
0.025+0.005 0.0015+0.0005
RESISTANCE
100.000 W
1.0 mA -
0.05+0.008
[3]
0.0060+0.0008
1.00000 kW
0.83 mA -
0.005+0.005
[3]
0.0060+0.0005
10.0000 kW 100 mA
-
0.005+0.006
[3]
0.0060+0.0005
100.000 kW 10.0 mA
- 0.05+0.007 0.0060+0.0005
1.00000 MW
900 nA - 0.06+0.007 0.0060+0.0005
10.0000 MW
205 nA - 0.25+.005 0.0250+0.0005
100.000 MW 205 nA||10MW
- 2.00+0.005 0.3000+0.0005
CURRENT 10.0000 mA < 0.2 V - 0.05+0.015 0.0055+0.0005
100.000 mA < 0.2 V - 0.05+0.005 0.0055+0.0005
1.00000 A < 0.5 V - 0.20+0.007 0.0100+0.0005
10.0000 A < 0.6 V - 0.25+0.007 0.0150+0.0005
CONTINUITY
1000 W
0.83 mA - 0.05+0.005 0.0050+0.0005
DIODE TEST
[4]
1.0000 V 0.83 mA - 0.05+0.005 0.0050+0.0005
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE
[5]
FREQUENCY
1 Year
23 oC ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55 oC
TRUE-RMS 100.000 mV 20 Hz - 45 Hz 1.0+0.1 0.02+0.02 AC VOLTAGE
[6]
45 Hz - 10 kHz 0.2+0.1 0.02+0.02 10 kHz - 30 kHz 1.5+0.3 0.05+0.02 30 kHz - 100 kHz
[7]
5.0+0.3 0.10+0.02
1.00000 V to 750.00 V 20 Hz - 45 Hz
1.0+0.1
[14]
0.02+0.02 45 Hz - 10 kHz 0.2+0.1 0.02+0.02 10 kHz - 30 kHz 1.0+0.1 0.05+0.02 30 kHz - 100 kHz
[7]
3.0+0.2
[15]
0.10+0.02
TRUE-RMS 10.0000 mA 20 Hz - 45 Hz 1.5+0.1 0.02+0.02 AC CURRENT
[8]
100.000 mA 45 Hz - 1 kHz 0.5+0.1 0.02+0.02
10.0000 A
1 kHz - 10 kHz
[9]
2.0+0.2 0.02+0.02
FREQUENCY
[10]
100 mV to 750 V < 2 Hz 0.18+0.003 0.005
< 20 Hz 0.04+0.003 0.005 20 Hz - 100 kHz
[11]
0.02+0.003 0.005
100 kHz ~ 300 kHz
[12]
0.02+0.003 0.005
10 mA to 10 A < 2 Hz 0.18+0.003 0.005
< 20 Hz 0.04+0.003 0.005 20 Hz ~ 10 kHz
[11]
0.02+0.003 0.005
ACCURACY ± (% of reading + % of range)
FUNCTION RANGE TEST CURRENT, etc.
1 Year
23
o
C ± 5 oC
Temperature Cefficient
0
o
C - 18 oC
28
o
C - 55oC
TEMPERATURE
-80 oC - 150 oC
5 kW thermistor probe
Probe accuracy + 0.2 oC 0.002 oC
- 110.0 oF - 300.0 oF
5 kW thermistor probe
Probe accuracy + 0.4 oF 0.0036 oF
CAPACITANCE 1.000 nF
0.75 mA
2.0+0.8 0.02+0.001
10.00 nF
0.75 mA
1.0+0.5 0.02+0.001
100.0 nF
8.3 mA
1.0+0.5 0.02+0.001
1.000 mF - 100.0 mF 83 mA
1.0+0.5 0.02+0.001
1000 mF
0.83 mA 1.0+0.5 0.02+0.001
10,000 mF
0.83 mA 2.0+0.5 0.02+0.001
[1] Specifications are for 30 minutes warm-up, 5 1/2 digit resolution and calibration temperature 18 oC - 28 oC. [2] 20% over range on all ranges except 1000Vdc. [3] Specifications are 2-wire ohms using Math Null. If without Math Null, add 0.2W additional error. [4] Specifications are for the voltage measured at the input terminals only. [5] 20% over range on all range except 750 Vac [6] Specifications are for sinewave inputs > 5% of range. Maximum crest factor : 3 at full scale. [7] Additional error to be added as frequency > 30kHz and signal input < 10% of range. 30kHz ~ 100kHz: 0.003% of full scale per kHz. [8] For 12A terminal, 10A dc or ac rms continuous, > 10A dc or ac rms for 30 seconds ON and 30 seconds OFF. [9] For 1A and 10A ranges, the frequency is verified for less than 5kHz. [10] Specifications are for half-hour warm-up, using 0.1 second aperture. The frequency can be measured up 1Mhz as 0.5V signal to 100mV/1V ranges. [11] For 20Hz ~ 10kHz, the sensitivity is AC input current from 10% to 120% of range except where noted. [12] For 100kHz ~ 300kHz, the sensitivity will be 12% ~ 120% of range except 750V range. [13] Input Impedence is in paralleled with capacitance < 120 pF. [14] For input < 200V rms [15] For input < 300V rms
DC CHARACTERISTICS
[1]
AC CHARACTERISTICS
[1]
TEMPERATURE and CAPACITANCE CHARACTERISTICS
[1]
3
SYSTEM SPEED
FUNCTION DIGITS READING SPEED
[1]
FUNCTION CHANGE
(sec)
[2]
RANGE CHANGE
(sec)
[3]
AUTO RANGE
(sec)
[4]
READING SPEED
OVER USB
[5]
DCV, DCI 5 1/2 15/s
0.6 0.7 2.2
8/s
4 1/2 70/s 19/s
ACV, ACI 5 1/2 2.5/s
5.0 2.2 6.1 1/s
4 1/2 2.5/s
FREQUENCY
[6]
5 1/2 9/s
7.0 2.5 6.1 1/s
4 1/2 9/s
[1] Reading rate of the A/D converter. [2] Time to change from 2-wire resistance to this specified functions and to take at least one reading in 4.5 digit using the SCPI “FUNC” and “READ?” commands. [3] Time to change one range to the next higher range and to take at least one reading in 4.5 digit using the SCPI “FUNC” and “READ?” commands. [4] Time to automatically change one range and to take at least one reading in 4.5 digit using SCPI “CONF AUTO” and “READ?” commands. [5] Number of measurements per second that can be read through USB using SCPI “ READ?” command. [6] Reading rate depend on signal frequency > 10Hz.
OPERATING CHARACTERISTICS
SUPPLEMENTAL MEASUREMENT CHARACTERISTICS
DC VOLTAGE
Measuring Method:
Sigma Delta A-to-D converter
Input Resistance:
10MW\2% range (typical)
Input Protection:
1000V on all ranges
RESISTANCE
Measuring Method:
2-wire Ohms
Open-circuit voltage:
Limited to < 5V
Input Protection:
1000V on all ranges
DC CURRENT
Shunt Resistance:
0.1W to 10W for 10mA to 1.2A ranges
0.01W for 12A range
Input Protection:
Front Panel 1.25A, 500V fuse for I terminal
Internal 15A, 600V fuse for 12A terminal
CONTINUITY/DIODE TEST
Measuring Method:
Uses 0.83mA\0.2% constant current source, < 5V open circuit voltage
Response Time:
70 samples/second with audible tone
Continuity Threshold:
10W fixed
Input Protection:
1000V
TEMPERATURE
Measurement Method:
2-wire Ohms measurement of 5kW thermistor sensor (YSI 4407) with
computer conversion
Auto-ranging measurement, no manual range selection
Input Protection:
1000V
MEASUREMENT NOISE REJECTION
CMRR (Common Mode Rejection) for 1kW unbalance LO lead
DC 120 dB
AC 70 dB
NMR (Normal Mode Rejection) For 60 Hz (50Hz) \ 0.1%
5 1/2 digit 65 dB (55 dB)
4 1/2 digit 0 dB
AC VOLTAGE
Measurement Method:
AC coupled true-RMS - measure the ac component with up to 400 Vdc bias any
range
Crest Factor:
Maximum 5:1 at full scale
Input Impedence:
1MW\2% in parallel with < 100pF of all ranges
Input Protection:
750Vrms on all ranges
AC CURRENT
Measuring Method:
DC coupled to the fuse and current shunt, AC coupled true-RMS measurement
(measure the AC component only)
Shunt Resistance:
0.1W to 10W for 10mA to 1.2A range
0.1W for 12A range
Input Protection:
Externally accessible 1.25A, 500V fuse for I terminal
Internally replaceable 15A, 600V fuse for 12A terminal
FREQUENCY
Measurement Method:
Reciprocal counting technique. AC coupled input using AC voltage function.
Signal Level:
10% of range to full scale input on all ranges
Auto or manual range selection
Gate Time:
0.1 second or 1 period of the input signal, whichever is longer.
Input Protection:
750Vrms on all ranges
MATH FUNCTIONS
Null, dBm, dB, Min/Max/Avg, Hold, Limit Test
TRIGGER and MEMORY
Single trigger, 1 reading memory
REMOTE INTERFACE
USB 2.0 full speed, USBTMC class device (GPIB over USB)
PROGRAMMING LANGUAGE
SCPI, IEEE-488.1, IEEE-488.2
4
POWER SUPPLY
100V/120V(127V)/220V(230V)/240V \ 10%
AC line frequency 45Hz - 66Hz and (360Hz - 440Hz, 100/120V operation)
POWER CONSUMPTION
16VA maximum, <11W average
OPERATING ENVIRONMENT
Full accuracy at 0oC to 55 oC
Full accuracy to 80% RH at 30oC (non-condensing) Altitute up to 3000 meters
STORAGE COMPLIANCE
- 40 oC to 70 oC
SAFETY COMPLIANCE
Certified to CSA for IEC/EN/CSA/UL 61010-1 2ndEdition
MEASUREMENT CATEGORY
CAT II, 300V: CAT I 1000Vdc, 750Vac rms, 2500Vpk transient over voltage
Pollution degree 2
EMC COMPLIANCE
Certified to IEC/EN 61326:2002, CISPR 11, and equivalents for Group 1, Class A
SHOCK and VIBRATION
Tested to IEC/EN 60086-2
DIMENSION (HxWxD)
Rack: 88.5mm x 212.6mm x 272.3mm
Bench: 103.8mm x 261.2mm x 303.2mm
WEIGHT
3.75 kg, 8.27 lb
WARM UP TIME
30 minutes
WARRANTY
1 year
GENERAL CHARACTERISTICS
Accessories included:
Test lead kit Test report
Power cord USB interface cable Quick Start Guide User’s and Service Guide Product Reference CD-ROM Agilent IO Library Suite CD-ROM
Options:
Opt. 1CM - Rack mount adapter kit
DIMENSIONS
Experience the new 34405A digital multimeter for yourself.
Watch the 34405A in action on your PC by downloading the interactive demo from the 34405A homepage at www.agilent.com/find/34405a
Agilent Optional Accessories
34132A Deluxe Test Lead Kits
34133A Precision Electronics Test Leads
34330A 30A Current Shunt
303.2
261.2
103.8
5
Agilent 34405A Multimeter: Versatile and low cost solution for benchtop testing.
Math functions and utility menu that allow users to take reference measurements (ie. Min/Max value and etc.) and store the measurement setups from the front panel.
Selecting the secondary display measurements.
Connect the supplied test leads to the Input Terminals to start your measurements.
5.5 digit dual display increases productivity and throughput in troubleshooting.
Use the Up-Down keys to select the desired measurement range. Just press Shift -> Auto key to switch measurement range automatically.
Superior value with a broad range of functions, which includes the temperature and capacitance measurements.
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Two concepts underlie Agilent’s overall support policy: “Our Promise” and “Your Advantage.”
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you receive your new Agilent equipment, we can help verify that it works properly and help with initial product operation.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products.
www.agilent.com
For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at:
www.agilent.com/find/contactus
Phone or Fax
United States:
(tel) 800 829 4444 (fax) 800 829 4433
Canada:
(tel) 877 894 4414 (fax) 800 746 4866
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) 31 20 547 2111
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (080) 769 0800 (fax) (080) 769 0900
Latin America:
(tel) (305) 269 7500
Taiwan: (tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6755 0042 Email: tm_ap@agilent.com Contacts revised: 09/26/05
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2006 Printed in USA, April 15, 2006 5989-4906EN
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Agilent Email Updates
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test equipment solutions with confidence.
Loading...