Agilent 3325B Calibration Guide

Page 1
T
.O
.
33K3-4-2587-1
TECHNICAL MANUAL
CALIBRATION PROCEDURE
FOR
SYNTHESIZER/FUNCTION GENERATOR
(HEWLETT-PACKARD)
Distribution Statement C - Distribution authorized to U
official use or for administrative or operational purposes only, 8 January 1990
document shall be referred to 562 CBSG/GBHA, 813 Irving-Wick Dr W, Heath, OH 43056-1199. Destruction Notice - For unclassified, limited documents, destroy by any method that will prevent
disclosure of the contents or reconstruction of the document.
Published under Authority of the Secretary of the Air Force
. S
. Government agencies and their contractors for
. Requests for this
30 DECEMBER 2004
CHANGE 3 - 30 AUGUST 2007
Page 2
T.O
.
33K3-4-2587-1
INSERT LATEST CHANGED PAGES
NOTE
I
LIST OF EFFECTIVE PAGES
Date of issue for original and changed pages are:
0
Original
Change
Change
Change
TOTAL NUMBER OF PAGES IN THIS PUBLICATION IS 24, CONSISTING OF THE FOLLOWING:
Page
No
.
Title
A
16
1 -
17
18
19 - 22
1
2
3
* ChangePage
No
.
3 3
0
3
2 0
30 December 2004
30 November 2005
30 August 2007
No
: The portion of the text and illustrations affected by the changes is indicated by a vertical
line in the outer margins of the page.
30 April 2005
* Change
.
No
. DESTROY SUPERSEDED PAGES
Page
.
No
.
.
* Change
No.
A Change 3
*Zero in this column indicates an original page
USAF
Page 3
T.O. 33K3-4-2587-1
SYNTHESIZE R/ FUNCTIO N GENERATOR
3325B, 3325B O PT 001
(HEWLETT-PACKARD)
1 CALIBRATIO N DESCRIPTION:
Table 1.
Test Instrument (TI) Performance Test Characteristics Specifications Method
Frequ ency Accura cy Rang e: 1.0 µHz to Measured with a
20.999 999 999 MHz (Sinewave); Electronic Counter
1.0 µHz to 10.999 999 999 MHz (Squarewave);
1.0 µHz to 10.999 999 999 kHz (Triangle/Ramp Wave)
Accuracy: ±5.0 x 10
Frequency Stability Std TI: ±5.0 X 10 Opt 001: ±5.0 X 10
-6
of setting (20 to 30 °C)
-6
/year; Not Calibrated
-8
/week
(Aging Rate)
Output Amplitude Range: 1.0 mV to 10.0 V p-p; Measured with a
4.0 mV to 40 V p-p (Opt 002 only) Digital Multimeter and Thermal Converter Accuracy:
Sine Wave:
Amplitude
Frequency Limits
13.52 to 23.98 dBm 0.001 Hz to 100.0 kHz ±0.1 dB
100.0 kHz to 20.0 MHz ±0.4 dB
-16.02 to 13.52 dBm 0.001 Hz to 100.0 kHz ±0.2 dB
100.0 kHz to 20 MHz ±0.6 dB
-56.02 to -16.02 dBm 0.001 Hz to 100.0 kHz ±0.2 dB
100.0 kHz to 10.0 MHz ±0.6 dB
10.0 to 20.0 MHz ±0.9 dB
Square Wave:
Amplitude
Frequency Limits
1.0 mV to 3.0 V p-p 0.001 Hz to 100.0 kHz ±2.2%
100.0 kHz to 10.0 MHz ±13.6%
3.0 to 10.0 V p-p 0.001 Hz to 100.0 kHz ±1.0%
100.0 kHz to 10 MHz ±11.1%
1
Page 4
T.O. 33K3-4-2587-1
Table 1. (Cont.)
Test Instrument (TI) Performance Test Characteristics Specifications Method
Output Amplitude
Triangle:
Amplitude
1.0 mV to 3.0 V p-p 0.001 Hz to 2.0 kHz ±2.7%
2.0 to 10.0 kHz ±6.2%
3.0 to 10.0 V p-p 0.001 Hz to 2.0 kHz ±1.5%
2.0 to 10.0 kHz ±5.0%
Ramps:
Amplitude
1.0 mV to 3.0 V p-p 0.001 to 500.0 Hz ±2.7%
500.0 Hz to 10.0 kHz ±11.2%
3.0 to 10.0 V p-p 0.001 to 500.0 Hz ±1.5%
500.0 Hz to 10.0 kHz ±10.0%
High Amplitude (Opt 002) Accuracy: ±2.0% of full output for each ra nge at 2.0 kHz
Spectral Purity
Harmonic Distortion Range: 0.1 Hz to 20.0 MHz Measured with a Spectrum Analyzer Accuracy:
Frequency Range
0.1 Hz to 50.0 kHz -65.0 dBc
50.0 to 200.0 kHz -60.0 dBc
200.0 kHz to 2.0 MHz -40.0 dBc
2.0 to 15.0 MHz -30.0 dBc
15.0 to 20.0 MHz -25.0 dBc
Non-harmonic sign a ls Range: 0.1 Hz to 20.0 MHz
Accuracy: 70.0 d B bel ow the car rier or -90.0 dBm, whichever is greater
Square Wave Range: 0.1 to 11.0 MHz Measured with an Oscilloscope Rise/Fall time: ≤20.0 ns from 10 to 90% of Square Wave, at full output
Symmetry: ≤0. 02% of period + 3 ns
Frequency Limits
Frequency Limits
Harmonic Level
2
Page 5
T.O. 33K3-4-2587-1
Table 1. (Cont.)
Test Instrument (TI) Performance Test Characteristics Specifications Method
DC Offset Rang e: 0 to ±5.0 V Measured with a Digital Multimeter Accuracy: DC only: ±0.02 to ±20.0 mV, depen d s on offset chosen
Amplitude Modulation Range: To 80% modulation Measured with a Distortion at 1 kHz, 0 VDC offset Measurin g Receiver a nd Distortion Analyzer
Accuracy:
-30 dB
Phase Offset Range: +719.9° to -719.9° Measured with a Frequency Counter and Frequency Accuracy: ±0.2° Synthesizer
2 EQUIPMENT REQUIREMENTS:
Minimum Use Calibration Sub­ Noun Specifications Equipment Item
2.1 ELECTRONIC Range: 1 to 45 MHz Agilent COUNTER 53132A OPT012-124 As Available Accuracy: <1.2 5 X 10-6
2.2 DIGITAL Range: 0.0 to 5.0 VDC Hewlett-Packard As Available MULTIMETER 0.0 to 40.0 V rms 3458A
Accuracy: VDC: ±0.05% of reading; VAC: ±0.25% of reading 100 Hz to 90 kHz
2.3 THERMAL Range: 3.0 V Hewlett-Packard As Available CONVERTER 11049A Accuracy: ±0.5%
2.4 HIGH FREQUENCY Range: 2 to 80 MHz, Hewlett-Packard Hewlett-Packard SPECTRUM -80 to 0 dBm (Scale Fidelity) 8563E 8566B ANALYZER Accuracy: ±1.6 dB
2.5 OSCILLOSCOPE Range: DC to 20.0 MHz, Tektronix As Available 100 MHz BW TDS430A
Accuracy: Risetime: ≤5.0 ns
3
Page 6
T.O. 33K3-4-2587-1
Minimum Use Calibration Sub­ Noun Specifications Equipment Item
2.6 MEASURING Range: 1.0 MHz Hewlett-Packard As Available RECEIVER 8902A Accuracy: N/A
2.7 DISTORTION Range: 1.0 kHz Hewlett-Packard As Available ANALYZER 334A Accuracy: ±3.0% of range setting
2.8 FEEDTHROUGH Range: 50 Hewlett-Packard As Available TERMINATION 11048C Accuracy: ±0.2%
2.9 FREQUENCY Range: 0.1 MHz Fluke As Available SYNTHESIZER 6011A/AJ Accuracy: ±0.5%
2.10 LOW FREQUENCY Range: 100 Hz to 2 MHz Hewlett-Packard As Available SPECTRUM -80 to 0 dBm (Scale Fidelity) 3585A ANALYZER Accuracy: ±1.0 dB
2.11 THERMAL Range: 1.0 V Hewlett-Packard CONVERTER 11050A Accuracy: ±0.5%
2.12 DC VOLTAGE Range: 0 to 3 V Fluke STANDARD 5440B Accuracy: ±0.01%
3 PRELIMINARY O PERATIONS:
3.1 Review and become familiar with the entire procedure before beginning Calibration Process.
WARNING
Unless otherwise designated, and prior to beginning the Calibration Process, ensure that all test equipment voltage and/or current outputs are set to zero (0) or turned off, where applicable. Ensure that all equipment switches are set to the proper position before making connections or applying power.
3.2 The Opt 001 TI has a high stability time base. Connect a jumper cable from the 10 MHz OVEN OUTPUT to the EXT REF IN Connector for proper operation of the Opt 001.
4
Page 7
T.O. 33K3-4-2587-1
3.3 The Opt 002 TI has a high voltage output capability of 4 mV to 40 V p-p.
3.4 Connect TI and all test equipment to the appropriate power source, set POWER switches to ON, and allow a warm-up period as required by manufacturer. Ensure that a TI with an Opt 001 T im e Base has been connected to power source for a minimum of 72 hours.
3.5 The Std TI an d Opt 001 TI both h a ve a Frequency Accuracy of ±5.0 X 10
-6
of setting. The frequency stability is not calibrated due to the time requirements in checking the drift rate of the oscillators. The drift rate for the standard T I i s per year and can not be divided by 365 for a per day specification. The drift err or associated with the oscillator will reflect in the frequency accuracy. If the frequency accuracy in paragraph 4.1 is out of tolerance, the TI will have to be adjusted in accordance with the maintenance T.O.
4 CALIBRATIO N PROC ESS:
NOTE
Unless ot herwi se s p ecified, verify the results of each test and ta ke corr ective acti on when ever th e test r eq u irem ent is not met , before proceedin g .
4.1 FREQ UE NCY ACCURACY CALIBRAT IO N:
4.1.1 C on nect TI MAIN SIGNAL connector th rough the Feedthr ough Term i nation to the Electronic Counter CHANNEL 1 input .
4.1.2 Press the TI keys, as necessary, to obtain the functions as follows:
Instr Preset
HIGH VOLTAGE OUTPUT (Opt-002) OFF
MAIN FUNCTION
ENTRY FREQ
DATA 20 MHz
ENTRY AMPTD
DATA 1 V RMS
ENTRY DC OFFSET
DATA 0 V
4.1.3 Set the Electronic Counter controls as required for a repeatable frequency indication.
4.1.4 The Electronic Counter m ust i ndicate between 19.999 90 to 20.000 10 MHz.
4.1.5 Set TI output to minimum and disconnect the test setup.
5
Page 8
T.O. 33K3-4-2587-1
4.2 OUTPUT AM PLIT UDE CAL IBRATION:
4.2.1 C on nect TI MAIN SIGNAL connector th rough the Feedthr ough Term i nation t o the Digita l Multimeter Input connector. Set the Digital Multimeter for a VAC measurement in the SYNC mode.
4.2.2 Press the TI keys, as necessary, to obtain the functions as follows:
Instr Preset
MAIN FUNCTION
ENTRY FREQ
DATA 100 Hz
ENTRY AMPTD
DATA 3.536 V rms
4.2.3 Press the TI AMPTD CAL key.
4.2.4 The Digital Multimeter must indicate between the corresponding values listed in the Limits column of Table 2.
4.2.5 Set TI Amplitude to each of the values listed in the Applied column of Table 2 and repeat step 4.2.4.
Applied Limits
3.536 V rms 3.495 to 3.577 V rms
1.061 V rms 1.048 to 1.073 V rms
353.0 mV rms 344.9 to 361.2 mV rms
106.0 mV rms 103.5 to 108.5 mV rms
30.0 mV rms 29.3 to 30.7 mV rms
9.8 mV rms 9.57 to 10.03 mV rms
3.0 mV rms 2.93 to 3.07 mV rms
4.2.6 Repeat steps 4.2.2 through 4.2.5 except in step 4.2.2 set TI FREQ to 90.0 kHz.
Table 2.
6
Page 9
T.O. 33K3-4-2587-1
4.2.7 Press TI keys in the following order:
MAIN FUNCTION
ENTRY FREQ
DATA 100 Hz
ENTRY AMPTD
DATA 5.0 V r m s
4.2.8 Press TI AMPTD CAL key.
4.2.9 The Digital Multimeter must indicate between 4.950 to 5.050 V rms.
4.2.10 Enter 1. 49 V r m s on TI DATA keys, then press the AMPTD CAL key.
4.2.11 The Digital Multimeter must indicate between 1.457 to 1.523 V rms.
4.2.12 Repeat steps 4.2.7 through 4.2.11 except set TI Frequency to 90.0 kHz.
4.2.13 Press TI keys in the following order:
MAIN FUNCTION
ENTRY FREQ
DATA 100 Hz
ENTRY AMPTD
DATA 2.88 V r ms
4.2.14 Press TI AMPTD CAL key.
4.2.15 The Digital Multimeter must indicate between 2.836 to 2.923 V rms.
4.2.16 Enter 837. 1 m V r m s on TI DATA keys, then press the AMPTD CAL key.
4.2.17 The Digital Multimeter must indicate between 814.5 to 859.7 mV rms.
4.2.18 Repeat steps 4.2.13 through 4.2.17 except set TI Frequency to 1.9 kHz.
4.2.19 Press TI keys in the following order:
ENTRY FREQ
DATA 10 kHz
ENTRY AMPTD
DATA 2.88 V r ms
7
Page 10
T.O. 33K3-4-2587-1
4.2.20 Press TI AMPTD CAL key.
4.2.21 The Digital Multimeter must indicate between 2.736 to 3.024 V rms.
4.2.22 Enter 837. 1 m V r m s on TI DATA keys, then press the AMPTD CAL key.
4.2.23 The Digital Multimeter must indicate between 785.2 to 889.0 mV rms.
4.2.24 Press TI keys in the following order:
MAIN FUNCTION
ENTRY FREQ
DATA 100 Hz
ENTRY AMPTD
DATA 2.88 V r ms
4.2.25 Press TI AMPTD CAL key.
4.2.26 The Digital Multimeter must indicate between 2.837 to 2.923 V rms.
4.2.27 Enter 837. 1 m V r m s on TI DATA keys, then press the AMPTD CAL key.
4.2.28 The Digital Multimeter must indicate between 814.5 to 859.7 mV rms.
4.2.29 Press TI keys in the following order:
ENTRY FREQ
DATA 10 kHz
ENTRY AMPTD
DATA 2.88 V r ms
4.2.30 Press TI AMPTD CAL key.
4.2.31 The Digital Multimeter must indicate between 2.592 to 3.168 V rms.
4.2.32 Enter 837. 1 m V r m s on TI DATA keys, then press the AMPTD CAL key.
4.2.33 The Digital Multimeter must indicate between 743.3 to 930.9 mV rms.
4.2.34 Repeat steps 4.2.24 through 4.2.33 except set TI FUNCTION to
4.2.35 Set the TI for minimum output and disconnect the test setup.
.
8
Page 11
T.O. 33K3-4-2587-1
CAUTION
Set the source amplitude to minimum before making or removing connections from the Thermal Converter (2.3 or 2.11). Take several seconds to increase the source a mpli tude to t he desired level and several s econds to decrease t he sour ce amplitude to minimum. E nsure that TI OUT P UT d oes not exceed 3 .0 V r ms while it is connected to the Thermal Converter (2.3) (1.0 V rms for Thermal Converter (2.11)) or damage to the Thermal Converter (2.3 or 2.11) will occur.
4.2.36 Connect the DC Voltage Standard to the input of the Thermal Converter (2.3). Connect the output of the Thermal Converter (2.3) to the Digital Multimeter Input. Set the Digital Multimeter for an autoranged DCV measurement.
4.2.37 Set the DC Voltage Standard to each of the following voltages and record the Digital Multimeter indication. (Do not use divided outputs on the DC Standard.)
DC Voltage Digital Multimeter Standard Indication
1.4325 VDC
1.5 VDC (Ref)
1.5707 VDC
4.2.38 Set the DC Voltage Standard for min i mum output and disconnect the test setup.
4.2.39 Connect equipment as shown in Figure 1.
TI DIGITAL MULTIMETE R
INPUT
MAIN SIGNAL
THERMAL
CONVERTER
(2.3 OR 2.11)
Figure 1.
4.2.40 Press TI keys in the following order:
MAIN FUNCTION
ENTRY FREQ
DATA 1.0 kHz
ENTRY AMPTD
9
Page 12
T.O. 33K3-4-2587-1
4.2.41 Press TI AMPTD CAL key. Slowly adjust the TI Amplitude using the DATA keys for a Digital Multimeter ind icati on equa l to th e value recorded in st ep 4 .2.3 7 for th e 1 .5 VDC reference.
4.2.42 Set TI Frequency to 2.0 MHz and incr ease TI Frequency in 2.0 MHz steps up to 20.0 MHz. The Digital Multimeter must indicate between the values recorded in step 4.2.37 for 1.4325 to 1.5707 VDC.
4.2.43 Set the TI for minimum output and disconnect the test setup.
4.2.44 Connect the DC Voltage Standard to the input of the Thermal Converter (2.11). Connect the output of the Thermal Converter (2.11) to the Digital Multimeter Input.
4.2.45 Set the DC Voltage Standard to each of the following voltages and record the Digital Multimeter indication.
DC Voltage Digital Multimeter Standard Indication
0.8399 VDC
0.9 VDC (Ref)
0.9644 VDC
4.2.46 Set the DC Voltage Standard for min i mum output and disconnect the test setup.
4.2.47 Connect equipment as shown in Figure 1.
4.2.48 Press the TI keys, as necessary, to obtain the functions as follows:
MAIN FUNCTION
ENTRY FREQ
DATA 1.0 kHz
ENTRY AMPTD
4.2.49 Press the TI AMPTD CAL key. Slowly adjust the TI Amplitude using the DATA keys for a Digital Multimeter indication equal to the value recorded in step 4.2.45 for the 0.9 VDC Reference.
4.2.50 Set the TI Frequency to 2.0 MHz and increase the TI Frequency in 2.0 MHz steps up to 20.0 MHz. The Digital Multimeter must indicate between the values recorded in step 4.2.45 for 0.8399 to 0.9644 VDC.
4.2.51 Set the TI for minimum output and disconnect the test setup.
4.3 HIGH AM PLIT UDE CAL IBRATION:
4.3.1 C on nect TI MAIN SIGNAL connector to th e Dig i tal Multimeter VOLTS INPUT connector (do not use the Feedthrough Termination). Set the Digital Multimeter controls as required to measure 15.0 V rms.
4.3.2 Press TI MAIN SIGNAL 40 V p- p key to enable the High Ampl it ude Mode.
(Opt 002 only)
10
Page 13
T.O. 33K3-4-2587-1
4.3.3 Press the TI keys, as necessary, to obtain the functions as follows:
MAIN FUNCTION
ENTRY FREQ
DATA 2.0 kHz
ENTRY AMPTD
DATA 14.14 V rms
4.3.4 Press TI AMPTD CAL key.
4.3.5 The Digital Multimeter must indicate between 13.857 to 14.423 V rms.
4.3.6 Press TI keys in the following order:
MAIN FUNCTION
DATA 20.0 V r ms
4.3.7 Press TI AMPTD CAL key.
4.3.8 The Digital Multimeter must indicate between 19.6 to 20.4 V rms.
4.3.9 Press TI keys in the following order:
MAIN FUNCTION
DATA 11.54 V rms
4.3.10 Press TI AMPTD CAL key.
4.3.11 The Digital Multimeter must indicate between 11.309 to 11.771 V rms.
4.3.12 Press TI MAIN SIGNAL 40 V p- p k ey (to disable t he High Ampli tude Mode) and disconnect the test setup.
4.4 SPECTRAL PURITY CAL IBRATION:
4.4.1 C on nect TI MAIN SIGNAL connector to th e Low Frequen cy Spectrum Analyzer 50- 7 5 Inpu t conn ector and select 50 imp ed ance.
4.4.2 Press TI keys in the following order:
Instr Preset
ENTRY FREQ
DATA 40 kHz
ENTRY AMPTD
DATA 0.0 dBm
11
Page 14
T.O. 33K3-4-2587-1
4.4.3 Set the Spectrum Analyzer controls, as required, to display the fundamental frequency and several harmonics.
4.4.4 Adjus t the Sp ectrum Ana lyzer controls to set the peak of the fun d amental signal at a convenient reference point.
4.4.5 The peaks of all harmonics must be more than 65.0 dB below the peak of the fundamental signal.
4.4.6 The peaks of all non-harmonics must be more than 70.0 dB below the peak of the fundamental signal.
4.4.7 Press TI keys in the following order:
ENTRY FREQ
DATA 180.0 kHz
4.4.8 Repeat steps 4.4.3 and 4.4.4. The peaks of all harmonics must be more than 60.0 dB below the peak of the fundamental signal. and the peaks of all non-harmonics must be more than 70.0 dB below the peak of the fundamental signal.
4.4.9 Press TI keys in the following order:
ENTRY FREQ
DATA 1.8 MHz
4.4.10 Repeat steps 4.4.3 and 4.4.4. The peaks of all harmonics must be more than 40.0 dB below the peak of the fundamental signal. and the peaks of all non-harmonics must be more than 70.0 dB below the peak of the fundamental signal.
4.4.11 Set the TI for minimum output and disconnect the test setup.
4.4.12 C on nect the T I MAI N SI GNAL outp ut to the High Frequency Spectrum Analyzer RF INPUT conn ector.
4.4.13 Press the TI keys in the following order:
ENTRY AMPTD
DATA 0.0 dBm
ENTRY FREQ
DATA 14.8 MHz
4.4.14 Repeat steps 4.4.3 and 4.4.4. The peaks of all harmonics must be more than 30.0 dB below the peak of the fundamental signal and the peaks of all non-harmonics must be more than 70.0 dB below the peak of the fundamental signal.
4.4.15 Press TI keys in the following order:
ENTRY FREQ
DATA 20.0 MHz
12
Page 15
T.O. 33K3-4-2587-1
4.4.16 Repeat steps 4.4.3 and 4.4.4. The peaks of all harmonics must be more than 25.0 dB below the peak of the fundamental signal and the peaks of all non-harmonics must be more than 70.0 dB below the peak of the fundamental signal.
4.4.17 Set the TI for minimum output and disconnect the test setup.
4.5 SQ UARE WAVE CAL IBRATION:
4.5.1 C on nect TI MAIN SIGNAL connector th rough the Feedthr ough Term i nation t o the Oscilloscope Channel 1 input.
4.5.2 Press TI keys in the following order:
Instr Preset
MAIN FUNCTION
ENTRY FREQ
DATA 1 MHz
ENTRY AMPTD
DATA 10.0 V p-p
4.5.3 Adjust the Oscilloscope Vertical and Horizontal controls, as required, to measure the Rise Time between the 10 and 90% points of the displayed Square Wave. The Rise Time must be 20.0 ns.
4.5.4 Adjust the Oscilloscope Vertical and Horizontal controls, as required, to measure the Fall Time between the 10 and 90% points of the displayed Square Wave. The Fall Time must be 20.0 ns.
4.5.5 Set TI to minimum output and disconnect the test setup.
4.5.6 C on nect TI MAIN SIGNAL connector to th e Electronic Counter CHANNEL 1 input.
4.5.7 Set TI controls as follows:
MAIN FUNCTION
ENTRY FREQ
DATA 1 MHz
ENTRY AMPTD
DATA 1.0 V r m s
ENTRY DC OFFSET
DATA 0 VOLT
13
Page 16
T.O. 33K3-4-2587-1
4.5.8 Set the Electronic Counter controls as follows:
Time and Period PERIOD 1
CHANNEL 1
50Ω, 1MΩ 50
4.5.9 Set the Electronic Counter controls as necessary to display a stable and reliable indication. Record the Elect roni c C ou nter indicati on as P.
4.5.10 Set the Electronic Counter controls as follows:
Time and Period TI 1 TO 2
CHANNEL 2
50Ω, 1MΩ 50
Trigger, Sensitivity
SLOPE NEG
COMMON 1 ON
4.5.11 Set the Electronic Counter controls as necessary to display a stable and reliable indication. Record the Elect roni c C ou nter indicati on as W.
4.5.12 Calculate the Squar e Wave Symmetry by using the following formula:
P
TS
Where TS = TI Square Wave Symmetry
P = Valu e recorded in s tep 4.5 .9
W = Value recorded in step 4 .5.11
4.5.13 Calculate the TI limits by using the following formula:
TSL P ns=+0 0002 3. Χ
Where TSL = TI Square Wave Symmetry Limits
P = Valu e recorded in s tep 4.5 .9
4.5.14 Verify the absolute value of TS calculated in step 4.5.12 is the TSL value cal cu lated in st ep 4.5.13.
4.5.15 Repeat steps 4.5.7 through 4.5.14 except in step 4.5.7 set TI FREQ to 10 MHz.
4.5.16 Set TI to minimum output and disconnect the test setup.
()
W=
2
14
Page 17
T.O. 33K3-4-2587-1
4.6 DC OFFSET CALIBRATION:
4.6.1 C on nect TI MAIN SIGNAL outpu t con nector directly to the Feedthrough Termin a t ion and t hen with a cable to the Digital Multimeter Input connectors.
4.6.2 Press TI MAIN FUNCTION key that is active (indicated by an illuminated indicator).
4.6.3 Set the Digital Multimeter controls for a DC measurement.
4.6.4 Press TI keys in the following order:
ENTRY DC OFFSET
DATA 5.0 VOLT
4.6.5 Press TI AMPTD CAL key. The Digital Multimeter must indicate between 4.980 and 5.020 VDC.
4.6.6 Set TI DC OFFSET to -5.0 V.
4.6.7 The Digital Multimeter must indicate between -4.980 and -5.020 VDC.
4.6.8 Set the TI DC OFFSET to each of the values listed in the Applied column of Table 3.
Range (VDC) Applied (VDC) Limits (VDC)
-5.0 to 5.0 1.499 1.49300 to 1.50499
-1.499 -1.49300 to -1.50499
499.9 m 497.90 to 501.90 m
-499.9 m -497.90 to -501.90 m
149.9 m 149.30 to 150.50 m
-149.9 m -149.30 to -150.50 m
49.99 m 49.79 to 50.19 m
-49.99 m -49.79 to -50.19 m
14.99 m 14.93 to 15.05 m
-14.99 m -14.93 to -15.05 m
Table 3.
15
Page 18
T.O. 33K3-4-2587-1
Table 3. (Cont.)
Range (VDC) Applied (VDC) Limits (VDC)
4.999 m 4.979 to 5.019 m
-4.999 m -4.979 to -5.019 m
1.499 m 1.479 to 1.519 m
-1.499 m -1.479 to -1.519 m
4.6.9 The Digital Multimeter must indicate within the corresponding values listed in the Limits column of Table 3.
4.6.10 Set TI for minimum output and disconnect the test setup.
4.7 AMPLITUDE MODULATION DISTO RTION CALIBRATION
:
4.7.1 Press the TI INSTR PRESET key.
4.7.2 Connect the equipment as sh own in Figure 2.
4.7.3 Set the Distortion Analyzer FUNCTION to SET LEVEL and the METER RANGE to 0 dB.
4.7.4 Set t he Measu ring Receiver contr ol s as required to mak e an AM measuremen t.
(REAR PANEL)
MOD
TI
SOURCE OUT
AMPTD MOD IN
MAIN SIGNAL
DISTORTION ANALYZER
INPUT
MEASURING RECEIVER
MODULATION OUT
INPUT
Figure 2.
4.7.5 Press TI keys in the following order:
ENTRY FREQ
DATA 1.0 MHz
16
Page 19
T
.O, 33K3-4-2587-1
ENTRY
DATA
DATA
MOD SOURCE
ENTRY
DATA
ENTRY
DATA
4.7.6 Set the Distortion Analyzer FUNCTION to DISTORTION . Make a distortion measurement setting the Distortion Analyzer controls as required.
4.7.7 The Distortion Analyzer must indicate <_-30
4.7.8 Set the Distortion Analyzer METER RANGE to 0 dB and FUNCTION to SET LEVEL OFF (Shift Recall).
.0 dB.
AMPTD
1
.0Vrms
AM ON (SHIFT, STORE)
SHIFT, FREQ
1
.0 kHz
SHIFT, AMPTD
As required to produce
80% modulation as indicated
on the Measuring Receiver
. Set TI AM MOD to
.9 Set TI for minimum output and disconnect the test setup.
4
.7
.8 PHASE OFFSET CALIBRATION:
4
4.8.1 Connect equipment as shown in Figure 3.
TI
EXT
REF IN
MAIN SIGNAL
0
FREQUENCY SYNTHESIZER
10 MHz REF
Figure 3
OUTPUT
0
.
ELECTRONIC COUNTER
CHANNEL CHANNEL
1
0 0
2
Change 3 17
Page 20
T.O. 33K3-4-2587-1
4 .8
.2 Press the TI keys, as necessary, to obtain the functions as follows:
HIGH VOLTAGE OUTPUT (Opt-002)
MAIN FUNCTION
ENTRY DATA ENTRY
DATA
DATA
4.8.3 Set the Frequency Synthesizer controls for a 100 kHz at 13 dBm output. .4 Set the Electronic Counter controls as necessary for the following:
4
.8
Other Meas Stats
Show
:
N
1
:
OFF
FREQ
100
.0 kHz
AMPTD
13 dBm
(1)M ON(SHIFT, CLEAR)
Phase 1 to 2
MEAN
100
STATS
USE ON SINGLE
RUN CHANNEL 1 & CHANNEL 2
4 .8
.5 Press the TI PHASE key to display phase. Using the TI MODIFY keys, adjust the TI PHASE until the
Electronic Counter reads about 0 Electronic Counter display to two decimal places.
4 .8
.6 Set the TI phase to -1°
4 .8
.7 Determine the difference between the Electronic Counter recorded values in step 4.8.6 and 4.8.5
result is within 0 4 .8
.8 Set the TI phase to -10°. Record the Electronic Counter display to two decimal places.
.8 to 1
.2 deg.
:
:
:
DC/AC Sensitivity
100 kHz Filter
.0 DEG
. Record the Electronic Counter display to two decimal places.
. Press the TI Blue (entry prefix) key, then Asgn Zero (I)
ON ALL MEAS N
enabled
DC LO enabled
. Record the
. Verify the
.9 Determine the difference between the Electronic Counter recorded values in step 4
4
.8
result is within 9 .8 to 10
18 Change 2
.2 deg.
.8
.8 and 4.8.5
. Verify the
Page 21
T.O. 33K3-4-2587-1
4.8.10 Set the TI phase to -100°. Record the Electronic Counter display to two decimal places.
4.8.11 Determine th e difference between the Electronic Counter recorded values in step 4.8.10 and 4. 8. 5. Verify the result is within 99. 8 t o 100.2 deg.
4.8.12 Set all outputs for minimum and disconnect and secure all test equipment.
CALIBRATION PERFORMANCE TABLE
4.1 FREQUENCY ACCURACY CALIBRATION:
Range
20 MHz 1.0 V rms
4.2 AMPLITUDE ACCURACY CALIBRATION:
Range
100 Hz & 90.0 kHz
1.061 1.048 to 1.073
353.0 m 344.9 to 361.2 m
106.0 m 103.5 to 108.5 m
30.0 m 29.3 to 30.7 m
9.8 m 9.57 to 10.03 m
3.0 m 2.93 to 3.07 m
100 Hz & 90.0 kHz
1.49 1.457 to 1.523
100 Hz & 1.9 kHz
837.1 m 814.5 to 859.7 m
10 kHz
837.1 m 785.2 to 889.0 m
Applied Limits
Std: 19.99990 to 20.00010 MHz
Applied (V rm s) Limits (V rms)
3.536 3.495 to 3.577
5.0 4.950 to 5.050
2.88 2.837 to 2.923
2.88 2.736 t o 3.024
19
Page 22
T.O. 33K3-4-2587-1
CALIBRATION PERFORMANCE TABLE (Cont.)
4.2 AMPLITUDE ACCURACY CALIBRATION: (Cont.)
Range
100 Hz
837.1 m 814.5 to 859.7 m
10 kHz
837.1 m 743.3 to 930.9 m
100 Hz
837.1 m 814.5 to 859.7 m
10 kHz
837.1 m 743.3 to 930.9 m
2.0 to 20.0 MHz
0.9 0.8399 to 0.9644
4.3 HIGH AMPLITUDE CALIBRATION:
Range
2.0 kHz
2.0 kHz
2.0 kHz
4.4 SPE CTRAL P URI TY CA L IBRATI O N:
HARMONICS
Range
0.1 Hz to 20.0 MHz 40 kHz -65
180.0 kHz -60
1.8 MHz -40
14.8 MHz -30
20.0 MHz -25
Applied (V rm s) Limits (V rms)
2.88 2.837 to 2.923
2.88 2.592 to 3.168
2.88 2. 837 t o 2.923
2.88 2. 592 t o 3.168
1.5 1.4325 to 1.5707
Applied (V rm s) Limits (V rms)
14.14 13.857 to 14.423
20.0 19.6 to 20.4
11.54 11. 309 t o 11.771
Applied (at 0. 0 dBm) Limits (dBc)
20
Page 23
T.O. 33K3-4-2587-1
CALIBRATION PERFORMANCE TABLE (Cont.)
4.4 SPE CTRAL P URI TY CA L IBRATI O N: (Cont.)
NON-HARMONICS
Range
0.1 Hz to 20.0 MHz 0.0 -70
4.5 SQUARE WAVE CALIBRATION:
RISE/FALL TIME
Range (MHz)
0.1 to 11.0 1 MHz 20.0
SYMMETRY
Applied (MHz)
1 0.02% of period + 3 ns
10 0.02% of period + 3 ns
4.6 DC OFFSET C ALIBRATION:
Range (VDC)
-5.0 to 5.0 1.499 1.49300 to 1.50499
-1.499 -1.49300 to -1.50499
499.9 m 497.90 to 501.90 m
-499.9 m -497.90 to -501.90 m
149.9 m 149.30 to 150.50 m
-149.9 m -149.30 to -150.50 m
49.99 m 49.79 to 50.19 m
-49.99 m -49.79 to -50.19 m
14.99 m 14.93 to 15.05 m
Applied (dBm) Limits (dBc)
Appli ed (at 10. 0 V) Limits (ns)
Limits
Applied (VDC) Limits (VDC)
21
Page 24
T.O. 33K3-4-2587-1
CALIBRATION PERFORMANCE TABLE (Cont.)
4.6 DC OFFSET C ALIBRATION: (Cont.)
Range (VDC)
-14.99 m -14.93 to -15.05 m
4.999 m 4.979 to 5.019 m
-4.999 m -4.979 to -5.019 m
1.499 m 1.479 to 1.519 m
-1.499 m -1.479 to -1.519 m
4.7 AM DISTORTI ON CALIBRATION:
Applied (at 1.0 kHz)
80% mod -30
4.8 PHASE OFFSET CALIBRATION:
Range Applied Limits
-1° 100 kHz @ 13 dBm -1.2 to -0.8 deg
-10° 100 kHz @ 13 dBm -10.2 to -9.8 deg
-100° 100 kHz @ 13 dBm -100.2 to -99.8 deg
Applied (VDC) Limits (VDC)
Limits (dB)
22
Loading...