For Safety information, Warranties, and Regulatory information,
see the pages following the Index.
Agilent 33250A
80 MHz Function /
Arbitrary Waveform Generator
Agilent 33250A at a Glance
The Agilent Technologies 33250A is a high-performance 80 MHz
synthesized function generator with built-in arbitrary waveform and
pulse capabilities. Its combination of bench-top and system features
makes this function generator a versatile solution for your testing
requirements now and in the future.
WARNINGFor protection from electrical shock, the power cord ground must not be
defeated. If only a two-contact electrical outlet is available, connect the
instrument’s chassis ground screw (see above) to a good earth ground.
6
In This Book
Specifications Chapter 1 lists the function generator’s specifications.
Quick Start Chapter 2 prepares the function generator for use and
helps you get familiar with a few of its front-panel features.
Front-Panel Menu Operation Chapter 3 introduces you to the frontpanel menu and describes some of the function generator’s menu features.
Calibration Procedures Chapter 4 provides calibration, verification,
and adjustment procedures for the function generator.
Theory of Operation Chapter 5 describes the block diagram and
circuit-level theory related to the operation of the function generator.
Service Chapter 6 provides guidelines for returning your function
generator to Agilent Technologies for servicing, or for servicing it
yourself.
Replaceable Parts Chapter 7 contains a detailed parts list of the
function generator.
Backdating Chapter 8 describes the differences between this manual
and older issues of this manual.
Schematics Chapter 9 contains the function generator’s schematics
and component locator drawings.
If you have questions relating to the operation of the Agilent 33250A,
call 1-800-452-4844 in the United States, or contact your nearest Agilent
Technologies Office.
If your 33250A fails within three years of purchase, Agilent will either
repair or replace it free of charge. Call 1-877-447-7278 in the United
States (and ask for “Agilent Express”) or contact your local Agilent
Technologies Office.
7
8
Contents
Chapter 1 Specifications 13
Chapter 2 Quick Start
To Prepare the Function Generator for Use 21
To Adjust the Carrying Handle 22
To Set the Output Frequency 23
To Set the Output Amplitude 24
To Set a DC Offset Voltage 26
To Set the Duty Cycle 27
To Configure a Pulse Waveform 28
To View a Waveform Graph 29
To Output a Stored Arbitrary Waveform 30
To Use the Built-In Help System 31
To Rack Mount the Function Generator 33
Chapter 3 Front-Panel Menu Operation
4
Front-Panel Menu Reference 37
To Reset the Function Generator 39
To Select the Output Termination 39
To Read the Calibration Information 40
To Unsecure and Secure for Calibration 41
To Store the Instrument State 44
To Configure the Remote Interface 45
Chapter 4 Calibration Procedures
Agilent Technologies Calibration Services 49
Calibration Interval 50
Adjustment is Recommended 50
Time Required for Calibration 51
Automating Calibration Procedures 52
Recommended Test Equipment 53
Test Considerations 54
Performance Verification Tests 55
Internal Timebase Verification 60
AC Amplitude (high-impedance) Verification 61
Low Frequency Flatness Verification 62
0 dB Range Flatness Verification 63
+10 dB Range Flatness Verification 65
+20 dB Range Flatness Verification 66
Contents
9
Contents
Chapter 4 Calibration Procedures (continued)
Calibration Security 68
Calibration Message 70
Calibration Count 70
General Calibration/Adjustment Procedure 71
Aborting a Calibration in Progress 72
Sequence of Adjustments 72
Self-Test 73
Frequency (Internal Timebase) Adjustment 74
Internal ADC Adjustment 75
Output Impedance Adjustment 76
AC Amplitude (high-impedance) Adjustment 78
Low Frequency Flatness Adjustment 80
0 dB Range Flatness Adjustments 81
+10 dB Range Flatness Adjustments 83
+20 dB Range Flatness Adjustment 85
Pulse Width (Trailing Edge Delay) Adjustment 87
Pulse Edge Time Adjustment 88
Duty Cycle Adjustment 89
Output Amplifier Adjustment (Optional) 90
Calibration Errors 91
Contents
Chapter 5 Theory of Operation
Block Diagram 95
Main Power Supply 97
On-Board Power Supplies 98
Waveform DAC and Filters 100
Digital Waveform, Pulse, and Sync 101
Digital Waveform Translator 104
Amplitude Multiplier 106
Main Output Circuitry 107
System ADC 110
System DAC 112
10
Contents
Chapter 5 Theory of Operation (continued)
Synthesis IC 113
Timebase 115
Phase Locked Loops 116
Clock Divider 118
Trigger and Delay 120
Waveform RAM 122
Synchronous Multiplexer 123
Main Processor 124
Main Gate Array 125
DSP and Gateway 126
Earth-Referenced Logic 126
Front Panel 127
Chapter 6 Service
Operating Checklist 130
Types of Service Available 131
Repackaging for Shipment 132
Cleaning 132
Electrostatic Discharge (ESD) Precautions 133
Surface Mount Repair 133
Troubleshooting Hints 134
Self-Test Procedures 136
Disassembly 140
4
Contents
Chapter 7 Replaceable Parts
33250-66511 Main PC Assembly (A1) 153
33205-66502 Front-Panel PC Assembly (A2) 176
33250A Chassis Assembly 177
33250A Front-Panel Assembly 178
Manufacturer’s List 179
Lock Range:10 MHz ± 35 kHz
Level:100 mVpp to 5 Vpp
Impedance:
Lock Time:< 2 s
Internal Reference Output
Frequency:10 MHz
3
Level:632 mVpp (0 dBm),
Impedance:
° to
°
1 kΩ nominal, ac coupled
nominal
50Ω nominal, ac coupled
SYNC OUTPUT
Level:TTL-compatible
Impedance:50Ω nominal
_______________
1
Time to change parameter and output new signal.
Ω
2
Modulation or sweep off.
3
Times for 5-digit integer and 12-digit real numbers.
into > 1 kΩ
+360
°
16
Agilent 33250A Function /Arbitrary Waveform Generator
GENERAL SPECIFICATIONS
Power Supply:100-240 V (±10%)
Power Consumption:140 VA
Operating Environment:
Operating Altitude:3000 meters
Pollution Degree:Indoor or Sheltered Use,
Storage Temperature:
Stored States:Four (4) named user
Power-On State:Default or Last
for 50-60 Hz operation,
100-127 V (±10%)
for 50-400 Hz operation.
IEC 60664 CAT II
0 °C to
55
70
°C
°C
80% R.H. to 40 °C
IEC 60664 Degree 2
-30 °C to
configurations
Chapter 1 Specifications
Safety Designed to:
1
EMC Tested to:
Acoustic Noise:40 dBA
Warm-Up Time:1 hour
Calibration Interval:1 year
Warranty:3 years standard
Accessories Included:
IEC-61326-1
EN61010-1, CSA1010.1
UL-3111-1
IEC-61000-4-3 criteria B
IEC-61000-4-6 criteria B
User’s Guide,
Service Guide,
Quick Reference
Test Data,
Connectivity Software,
RS-232 Cable,
Power Cord
1
,
4
Guide,
Interface:IEEE-488 and RS-232
standard
Language:SCPI-1997, IEEE-488.2
Dimensions (WxHxD)
Bench Top:254 x 104 x 374 mm
Rack Mount:213 x 89 x 348 mm
Weight:4.6 kg
_______________
1
Radiated and conducted immunity testing:
When the product is tested at 3 V/m according
to IEC/EN 61000-4-3:1995 or tested at 3 Vrms
according to IEC/EN 61000-4-6:1996, the product
may not meet criteria A, but does meet criteria B.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001
du Canada.
N10149
17
1
Chapter 1 Specifications
Agilent 33250A Function / Arbitrary Waveform Generator
PRODUCT DIMENSIONS
103.6 mm
254.4 mm
88.5 mm
216.6 mm348.3 mm
374.0 mm
All dimensions are
shown in millimeters.
18
2
2
Quick Start
2
Quick Start
One of the first things you will want to do with your function generator
is to become acquainted with the front panel. We have written the
exercises in this chapter to prepare the instrument for use and help
you get familiar with some of its front-panel operations. This chapter
is divided into the following sections:
• To Prepare the Function Generator for Use, on page 21
• To Adjust the Carrying Handle, on page 22
• To Set the Output Frequency, on page 23
• To Set the Output Amplitude, on page 24
• To Set a DC Offset Voltage, on page 26
• To Set the Duty Cycle, on page 27
• To Configure a Pulse Waveform, on page 28
• To View a Waveform Graph, on page 29
• To Output a Stored Arbitrary Waveform, on page 30
• To Use the Built-In Help System, on page 31
• To Rack Mount the Function Generator, on page 33
20
Chapter 2 Quick Start
Power
Switch
To Prepare the Function Generator for Use
To Prepare the Function Generator for Use
1 Check the list of supplied items.
Verify that you have received the following items with your instrument.
If anything is missing, please contact your nearest Agilent Sales Office.
❑ One power cord.
❑ One User’s Guide.
❑ This Service Guide.
❑ One folded Quick Reference Guide.
❑ Certificate of Calibration.
❑ Connectivity software on CD-ROM.
❑ One RS-232 cable.
2 Connect the power cord and turn on the function generator.
Several power-on information messages are displayed after the function
generator performs its power-on self-test. The GPIB address is displayed.
The function generator powers up in the sine wave function at 1 kHz
with an amplitude of 100 mV peak-to-peak (into a 50Ω termination).
At power-on, the Output connector is disabled. To enable the Output
connector, press thekey.
If the function generator does not turn on, verify that the power cord is
firmly connected to the power receptacle on the rear panel (the powerline voltage is automatically sensed at power-on). You should also make
sure that the function generator is connected to a power source that is
energized. Then, verify that the function generator is turned on.
2
4
If you need further assistance, refer to chapter 6 for instructions on
returning the function generator to Agilent for service.
21
Chapter 2 Quick Start
To Adjust the Carrying Handle
To Adjust the Carrying Handle
2
To adjust the position, grasp the handle by the sides and pull outward.
Then, rotate the handle to the desired position.
Bench-top viewing positionsCarrying position
22
To Set the Output Frequency
Chapter 2 Quick Start
To Set the Output Frequency
At power-on, the function generator outputs a sine wave at 1 kHz with
an amplitude of 100 mV peak-to-peak (into a 50Ω termination).
The following steps show you how to change the frequency to 1.2 MHz.
1Press the “Freq” softkey.
The displayed frequency is either the power-on value or the frequency
previously selected. When you change functions, the same frequency is
used if the present value is valid for the new function. To set the
waveform period instead, press the Freq softkey again to toggle to the
Period softkey (the current selection is highlighted).
2 Enter the magnitude of the desired frequency.
Using the numeric keypad, enter the value “1.2”.
2
4
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the function generator outputs a waveform with the displayed
frequency (if the output is enabled). For this example, press MHz
Note: You can also enter the desired value using the knob and arrow keys.
.
23
Chapter 2 Quick Start
To Set the Output Amplitude
To Set the Output Amplitude
2
At power-on, the function generator outputs a sine wave with an
amplitude of 100 mV peak-to-peak (into a 50Ω termination).
The following steps show you how to change the amplitude to 50 mVrms.
1Press the “Ampl” softkey.
The displayed amplitude is either the power-on value or the amplitude
previously selected. When you change functions, the same amplitude is
used if the present value is valid for the new function. To set the amplitude
using a high level and low level, press the Ampl softkey again to toggle to
the HiLevel and LoLevel softkeys (the current selection is highlighted).
2 Enter the magnitude of the desired amplitude.
Using the numeric keypad, enter the value “50”.
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the function generator outputs the waveform with the displayed
amplitude (if the output is enabled). For this example, press mV
Note: You can also enter the desired value using the knob and arrow keys.
24
RMS
.
Chapter 2 Quick Start
To Set the Output Amplitude
You can easily convert the displayed amplitude from one unit to another.
For example, the following steps show you how to convert the amplitude
from Vrms to Vpp.
4 Enter the numeric entry mode.
Press thekey to enter the numeric entry mode.
5 Select the new units.
Press the softkey that corresponds to the desired units. The displayed
value is converted to the new units. For this example, press the Vpp
softkey to convert 50 mVrms to its equivalent in volts peak-to-peak.
2
4
To change the displayed amplitude by decades, press the right-arrow
key to move the cursor to the units on the right side of the display.
Then, rotate the knob to increase or decrease the displayed amplitude
by decades.
25
Chapter 2 Quick Start
To Set a DC Offset Voltage
To Set a DC Offset Voltage
2
At power-on, the function generator outputs a sine wave with a dc offset
of 0 volts (into a 50Ω termination). The following steps show you how to
change the offset to –1.5 mVdc.
1Press the “Offset” softkey.
The displayed offset voltage is either the power-on value or the offset
previously selected. When you change functions, the same offset is used
if the present value is valid for the new function.
2 Enter the magnitude of the desired offset.
Using the numeric keypad, enter the value “–1.5”.
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the function generator outputs the waveform with the displayed
offset (if the output is enabled). For this example, press mVdc.
Note: You can also enter the desired value using the knob and arrow keys.
Note: To select dc voltsfrom the front panel, pressand then select
theDC Onsoftkey. Press the Offset softkey to enter the desired voltage level.
26
To Set the Duty Cycle
Chapter 2 Quick Start
To Set the Duty Cycle
Applies only to square waves. At power-on, the duty cycle for square
waves if 50%. You can adjust the duty cycle from 20% to 80% for output
frequencies up to 25 MHz. The following steps show you how to change
the duty cycle to 30%.
1 Select the square wave function.
Press thekey and then set the desired output frequency to any
value less than 25 MHz.
2Press the “Duty Cycle” softkey.
The displayed duty cycle is either the power-on value or the percentage
previously selected. The duty cycle represents the amount of time per
cycle that the square wave is at a high level (note the icon on the right
side of the display).
3 Enter the desired duty cycle.
Using the numeric keypad or the knob, select a duty cycle value of “30”.
The function generator adjusts the duty cycle immediately and outputs a
square wave with the specified value (if the output is enabled).
2
4
27
Chapter 2 Quick Start
To Configure a Pulse Waveform
To Configure a Pulse Waveform
2
You can configure the function generator to output a pulse waveform
with variable pulse width and edge time. The following steps show you
how to configure a 500 ms pulse waveform with a pulse width of 10 ms
and edge times of 50 µs.
1 Select the pulse function.
Press thekey to select the pulse function and output a pulse
waveform with the default parameters.
2 Set the pulse period.
Press the Period softkey and then set the pulse period to 500 ms.
3 Set the pulse width.
Press the Pulse Width softkey and then set the pulse width to 10 ms.
The pulse width represents the time from the 50% threshold of the rising
edge to the 50% threshold of the next falling edge (note the display icon).
4 Set the edge time for both edges.
Press the Edge Time softkey and then set the edge time for both the rising
and falling edges to 50 µs. The edge time represents the time from the
10% threshold to the 90% threshold of each edge (note the display icon).
28
To View a Waveform Graph
Chapter 2 Quick Start
To View a Waveform Graph
In the Graph Mode, you can view a graphical representation of the
current waveform parameters. Each softkey parameter is shown in a
different color corresponding to the lines above the softkeys at the
bottom of the display. Note that the softkeys are listed in the same order
as in the normal display mode.
1 Enable the Graph Mode.
Press thekey to enable the Graph Mode. The name of the parameter
currently selected is shown in the upper-left corner of the display and the
numeric value is highlighted.
2 Select the desired parameter.
To select a specific parameter, note the colored bars above the softkeys at
the bottom of the display and select the corresponding color. For example,
to select amplitude, press the softkey below the magenta-colored bar.
• As in the normal display mode, you can edit numbers using the
numeric keypad or the knob and arrow keys.
2
4
• Parameters which normally toggle when you press a key a second time
(e.g., Freq / Period) also toggle in the Graph Mode.
• To exit the Graph Mode, pressagain.
Thekey also serves as akey to restore front-panel control
after remote interface operations.
29
Chapter 2 Quick Start
To Output a Stored Arbitrary Waveform
To Output a Stored Arbitrary Waveform
2
There are five built-in arbitrary waveforms stored in non-volatile memory.
The following steps show you how to output the built-in “exponential fall”
waveform from the front panel.
1 Select the arbitrary waveform function.
When you press thekey to select the arbitrary waveform function,
a temporary message is displayed indicating which waveform is currently
selected (the default is “exponential rise”).
2 Select the active waveform.
Press the Select Wform softkey and then press the Built-In softkey to
select from the five built-in waveforms. Then press the Exp Fall softkey.
The waveform is output using the present settings for frequency,
amplitude, and offset unless you change them.
The selected waveform is now assigned to thekey. Whenever you
press this key, the selected arbitrary waveform is output. To quickly
determine which arbitrary waveform is currently selected, press.
30
Loading...
+ 181 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.