The Agilent Technologies 33220A is a 20 MHz synthesized function
generator with built-in arbitrary waveform and pulse capabilities.
Its combination of bench-top and system features makes this function
generator a versatile solution for your testing requirements now and in
the future.
5 USB Interface Connector
6 LAN Interface Connecto r
7 GPIB Interface Connector
8 Chassis Ground
Use themenu to:
• Select the GPIB address (see chapter 3).
• Set the network parameter s for the LAN interface (see chapter 3).
• Display the current network parameters (see chapter 3).
Note: The External and Internal 10 MHz Reference Terminals (1 and 2,
above) are present only if Option 001, External Timebase Reference, is
installed. Otherwise, the holes for these connectors are plugged.
WARNINGFor protection from electrical shock, the power cord ground must not be
defeated. If only a two-contact electrical outlet is available, connect the
instrument’s chassis ground screw (see above) to a good earth ground.
6
In This Book
Specifications Chapter 1 lists the function generator’s specifications.
Quick Start Chapter 2 prepares the function generator for use and
helps you get familiar with a few of its front-panel features.
Front-Panel Menu Operation Chapter 3 introduces you to the frontpanel menu and describes some of the function generator’s menu features.
Calibration Procedures
and adjustment procedures for the
Theory of Operation Chapter 5 describes block and circuit level
theory related to the operation of the function generator.
Service Chapter 6 provides guidelines for returning your function
generator to Agilent Technologies for servicing, or for servicing it
yourself.
Replaceable Parts Chapter 7 contains a detailed parts list of the
function generator.
Backdating Chapter 8 describes the differences between this manual
and older issues of this manual.
Schematics Chapter 9 contains the function generator’s schematics and
component locator drawings.
You can contact Agilent T echnologies at one of the following telephone
numbers for warranty, service, or technical support information.
Chapter 4 provides calibration, verification,
function generator.
In the United States: (800) 829-4444
In Europe: 31 20 547 2111
In Japan: 0120-421-345
Or use our Web link for information on contacting Agilent worldwide.
www.agilent.com/find/assist
Or contact your Agilent Technologies Representative.
To Prepare the Function Generator for Use 21
To Adjust the Carrying Handle 22
To Set the Output Frequency 23
To Set the Output Amplitude 24
To Set a DC Offset Voltage 26
To Set the High-Level and Low-Level Values 27
To Select “DC Volts” 28
To Set the Duty Cycle of a Square Wave 29
To Configure a Pulse Waveform 30
To View a Waveform Graph 31
To Output a Stored Arbitrary Waveform 32
To Use the Built-In Help System 33
To Rack Mount the Function Generator 35
Chapter 3 Front-Panel Menu Operation 37
Front-Panel Menu Reference 39
To Select the Output Termination 41
To Reset the Function Generator 41
To Read the Calibration Information 42
To Unsecure and Secure for Calibration 43
To Store the Instrument State 46
To Configure the Remote Int erface 47
Contents
9
Contents
Contents
Chapter 4 Calibration Procedures 53
Agilent Technologies Calibration Services 55
Calibration Interval 55
Adjustment is Recommended 55
Time Required for Calibration 56
Automating Calibration Procedures 57
Recommended Test Equipment 58
Test Considerations 59
Performance Verification Tests 60
Internal Timebase Verification 64
AC Amplitude (high-impedance) Verification 65
Low Frequency Flatness Verification 66
0 dB Range Flatness Verification 67
+10 dB Range Flatness Verification 69
+20 dB Range Flatness Verification 71
Calibration Security 73
Calibration Message 75
Calibration Count 75
General Calibration/Adjus tment Procedure 76
Aborting a Calibration in Progress 77
Sequence of Adjustments 77
Self-Test 78
Frequency (Internal Timebase) Adjustment 79
Internal ADC Adjustment 80
Output Impedance Adjustment 81
AC Amplitude (high-impedance) Adjustment 83
Low Frequency Flatness Adjustment 85
0 dB Range Flatness Adjustments 86
+10 dB Range Flatness Adjustments 88
+20 dB Range Flatness Adjustment 90
Calibration Errors 93
10
Chapter 5 Theory of Operation 95
Block Diagram 97
Power Supplies 100
Main Power Supply 100
Earth Referenced Power Supplies 101
Floating Power Supplies 102
Analog Circuitry 103
Waveform DAC and Filters 103
Squarewave Comparator 104
Square and P ulse Level Translator 104
Main Output Circuitry 106
System ADC 107
System DAC 108
Digital Circuitry 110
Synthesis IC and Waveform Memory 110
Timebase, Sync Output, and Relay Drivers 111
Main Processor 112
Front Panel 114
External Timebase (Option 001) 115
Contents
4
Contents
Chapter 6 Service 117
Operating Checklist 118
Types of Service Available 119
Repackaging for Shipment 120
Cleaning 120
Electrostatic Discharge (ESD) Precauti ons 121
Surface Mount Repair 121
Troubleshooting Hi nts 122
Self-Test Procedures 124
Disassembly 127
Chapter 7 Replaceable Parts 139
33220-66511 – Main PC Assembly 141
33220-66502 – Front-Panel PC Assembly 155
33220-66503 – External Timebase PC Assembly 156
33220A Chassis Assembly 158
Manufacturer List 159
11
Contents
Contents
Chapter 8 Backdating 161
Chapter 9 Schematics 163
A1 Clocks, IRQ, RAM, ROM, and USB Schematic 165
A1 Front Panel Interface, LAN, GPIB, and Beeper Schematic 166
A1 Cross Guard, Serial Communications, Non-Volatile Memory, and
Trigger Schematic 167
A1 Power Distribution Schematic 168
A1 Synthesis IC and Wa veform RAM Schematic 169
A1 Timebase, Sync, and Relay Drivers Schematic 170
A1 System ADC Schematic 171
A1 System DAC Schematic 172
A1 Waveform DAC and Filters and Square Wave Comparator Schematic
173
A1 Square / Pulse Level Translation Schematic 174
A1 Gain Switching and Output Amplifier Schematic 175
A1 Earth Referenced Power Supply Schematic 176
A1 Isolated Power Supply S chematic 177
A2 Keyboard Scanner and Display Connector Schematic 178
A2 Key Control Schematic 179
A3 External Timebase Schematic 180
A1 Component Locator (top) 181
A1 Component Locator (bottom) 182
A2 Compone nt Locator 183
A3 Compone nt Locator 184
Output Impedance:50 Ω, typical
Maximum Rate:1 MHz
Fanout:<
selectable
>
1 kΩ
4 Agilent 33220As
Programming Times (typical)
Configuration Times
USB 2.0
Function
Change
Frequency
Change
Amplitude
Change
Select User
Arb
111 m s111 ms111 ms
1.5 ms2.7 ms1.2 ms
30 ms30 ms30 ms
124 ms124 ms123 ms
Arb Download Times (binary transfer)
USB 2.0
64 K points96.9 ms191.7 ms336.5 ms
16 K points24.5 ms48.4 ms80.7 ms
4 K points7.3 ms14.6 ms19.8 ms
Download times do not include setup or output time.
LAN
(VXI-11)
LAN
(VXI-11)
GPIB
GPIB
16
Chapter 1 Specifications
Agilent 33220A Function /Arbitrary Waveform Generator
1
General
Power Supply:CAT II
100 to 240 V
50/60 Hz (-5%, +10%)
100 to 120 V @
400 Hz (± 10%)
Power Consumption:50 VA maximum
Operating Environment:IEC 61010
Pollution Degree 2
Indoor Location
Operating Temperature:0 °C to 55 °C
Operating Humidity:5% to 80% RH,
non-condensing
Operating Altitude: Up to 3000 meters
Storage Temperature:-30 °C to 70 °C
State Storage Memory:Power off state
automatically sav ed .
Four user-configurabl e
stored states.
Interface:GPIB, USB, and LAN
standard
Language:SCPI - 1993,
IEEE-488.2
Dimensions (W x H x D):
Bench T op: 261.1 mm by 103.8 mm
by 303.2 mm
Rack Mount: 212.8 mm by 88.3 mm
by 272.3 mm
Weight:3.4 kg (7.5 lbs)
Safety Designed to:UL-1244, CSA 1010,
EN61010
EMC Tested to:MIL-461C, EN55011,
EN50082-1
Vibration and Shock:MIL-T-28800, Type III,
Class 5
Acoustic Noise:30 dBa
Warm-up Time:1 hour
@
Note: Specifications are subject to change
without notice. For the latest specifications,
go to the Agilent 33220A product page and
find the 33220A Datasheet.
www.agilent.com/find/33220A
This ISM device complie s w ith C ana di an ICES-001.
Cet appareil ISM est conforme à la norme NMB-001
One of the first things yo u will want to do w ith your function ge nerator is
to become acquainted with the fro nt panel. We have w ri tten the exercises
in this chapter to prepare the instrument for use and help you get
familiar with some of its front-panel operations. This chapter is divided
into the following sections:
2
• To Prepare the Function Generator for Use, on page 21
• To Adjust the Carrying Handle, on page 22
• To Set the Output Frequency, on page 23
• To Set the Output Amplitude, on page 24
• To Set a DC Offset Voltage, on page 26
• To Set the High-Level and Low-Level Values, on page 21
• To Select “DC Volts”, on page 22
• To Set the Duty Cycle of a Square Wave, on page 29
• To Configure a Pulse Waveform, on page 30
• To View a Waveform Graph, on page 31
• To Output a Stored Arbitrary Waveform, on page 32
• To Use the Built-In Help System, on page 33
• To Rack Mount the Function Generator, on page 35
20
Chapter 2 Quick Start
To Prepare the Function Generator for Use
To Prepare the Function Generator for Use
Power
Switch
1 Check the list of supplied items.
Verify that you have received the following items with your instrument.
If anything is missing, please contact your nearest Agilent Sales Office.
One power cord.
One User’s Guide.
This Service Guide.
One folded Quick Start Tutorial.
One folded Quick Reference Guide.
Certificate of Calibration.
Connectivity software on CD-ROM.
One USB 2.0 cable.
2 Connect the power cord and turn on the function generator.
The instrument runs a short power-on self test, which takes a few
seconds. When the instrument is ready for use it displays a message
about how to obtain help, along with the current GPIB address. The
function generator powers up in the sine wave function at 1 kHz with an
amplitude of 100 mV peak-to-peak (into a 50Ω termination).
the
Output
thekey
connector
.
is disabled. To enable the Output
At power-on,
connector, press
2
4
If the function generator does not turn on, verify that the power cord is
firmly connected to the power receptacle on the rear p a ne l (t he po w er -line
voltage is automatically sensed at power-on). You should also make sure
that the function generator is connected to a power source that is energized
Then, verify that the function generator is turned on.
If the power-on self test fails, “Self-T est Failed” is displayed along with an
error code. See Chapter 6 for information on self-test error codes, and
for
instructions on returning the function generator to Agilent for service.
21
.
Chapter 2 Quick Start
To Adjust the Carrying Handle
To Adjust the Carrying Handle
2
To adjust the position, grasp the handle by the sides and pull outward.
Then, rotate the handle to the desired position.
Retracted
Carrying
Position
Extended
22
To Set the Output Frequency
Chapter 2 Quick Start
To Set the Output Frequency
At power-on, the function generator outputs a sine wave at 1 kHz with
an amplitude of 100 mV peak-to-peak (into a 50Ω termination).
The following steps show you how to change the frequency to 1.2 MHz.
1Press the “Freq” softkey.
The displayed frequency is either the power-on value or the frequency
previously selected. When you change functions, the same frequency is
used if the present value is valid for the new function. To set the
waveform period instead, press the Freq softkey again to toggle to the
Period softkey (the current selection is highlighted).
2 Enter the magnitude of the desired frequency.
Using the numeric keypad, enter the value “1.2”.
2
4
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the function generator outputs a waveform with the displayed
frequency (if the output is enabled). For this example, press MHz
Note: Y
keys.
ou can also enter the desired value using the knob and cursor
.
23
Chapter 2 Quick Start
To Set the Output Amplitude
To Set the Output Amplitude
2
At power-on, the function generator outputs a sine wave with an
amplitude of 100 mV peak-to-peak (into a 50Ω termination) .
The following steps show you how to change the amplitude to 50 mVrms.
1Press the “Ampl” softkey.
The displayed amplitude is either the power-on value or the amplitude
previously selected. When you change functions, the same amplitude is
used if the present value is valid for the new function. To set the amplitude
using a high level and low level, press the Ampl softkey again to toggle to
the HiLevel and LoLevel softkeys (the current selection is highlighted).
2 Enter the magnitude of the desired amplitude.
Using the numeric keypad, enter the value “50”.
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the functio n ge ner ator out puts the wave form w ith the disp layed
amplitude (if the output is enabled). For this example, press mV
Note: Y
keys.
24
ou can also enter the desired value using the knob and cursor
RMS
.
Chapter 2 Quick Start
To Set the Output Amplitude
You can easily convert the displayed amplitude from one unit to another.
For example, the following steps show you how to convert the amplitude
from Vrms to Vpp.
4 Enter the numeric entry mode.
Press thekey to enter the numeric entry mode.
5 Select the new units.
Press the softkey that corresponds to the desired units. The displayed
value is converted to the new units. For this example, press the Vpp
softkey to convert 50 mVrms to its equivalent in volts peak-to-peak.
2
4
To change the displayed amplitude by decades, press the right-cursor
key to move the cursor to the units on the right side of the display.
Then, rotate the knob to increase or decrease the displayed amplitude
by decades.
25
Chapter 2 Quick Start
T o Set a DC Offset Voltage
To Set a DC Offset Voltage
2
At power-on, the function generator outputs a sine wave with a dc offset
of 0 volts (into a 50Ω termination). The following steps show you how to
change the offset to –1.5 mVdc.
1Press the “Offset” softkey.
The displayed off set voltage is either the power-on value or the offset
previously selected. When you change functions, the same offset is used
if the present value is valid for the new function.
2 Enter the magnitude of the desired offset.
Using the numeric keypad, enter the value “–1.5”.
3 Select the desired units.
Press the softkey that corresponds to the desired units. When you select
the units, the functio n ge ner ator out puts the wave form w ith the disp layed
offset (if the output is enabled). For this example, press mV
Note: Y
keys.
26
ou can also enter the desired value using the knob and cursor
DC
.
Chapter 2 Quick Start
To Set the High-Level and Low-Level Values
To Set the High-Level and Lo w-Level Values
You can specify a signal by setting its amplitude and dc offset values, as
described previously. Another way to set the limits of a signal is to
specify its high-level (maximum) and low-level (minimum) values. This
is typically convenient for digital applications. In the following example,
let's set the high-level to 1.0 V and the low-level to 0.0 V.
1 Press the "Ampl" softkey to select "Ampl".
2 Press the softkey again to toggle to "HiLevel".
Note that both the Ampl and Offset softkeys toggle together, to HiLevel
and LoLevel, respectively.
3 Set the "HiLevel" value.
Using the numeric keypad or the knob, select a value of "1.0 V". (If you
are using the keypad, you will need to select the unit, "V", to enter the
value.)
2
4
4 Press the "LoLevel" softkey and set the value.
Again, use the numeric keyp ad or the knob to enter a value of "0.0 V".
Note that these settings (high-level = "1.0 V" and low-level = "0.0 V") are
equivalent to setting an amplitude of "1.0 Vpp" and an offset of "500
mVdc".
27
Chapter 2 Quick Start
T o Select “DC Volts”
To Select “DC Volts”
2
You can select the "DC Volts" feature from the “Utility” menu, and then
set a constant dc voltage as an "Offset" value. Let's set "DC Volts" = 1.0
Vdc.
1 Press and then select the DC On softkey.
The Offset value becomes selected.
2 Enter the desired voltage level as an "Offset".
Enter 1.0 Vdc with the numeric keypad or knob.
You can enter any dc voltage from -5 Vdc to +5 Vdc.
28
Chapter 2 Quick Start
T o Set the Duty Cycle of a Square Wave
To Set the Duty Cycle of a Square Wave
At power-on, the duty cycle for square waves is 50%. You can adjust th e
duty cycle from 20% to 80% for output frequencies up to 10 MHz. The
following steps show you how to change the duty cycle to 30%.
1 Select the square wave function.
Press thekey and then set the desired output frequency to any
value up to 10 MHz.
2Press the “Duty Cycle” softkey.
The displayed duty cycle is either the power-on value or the percentage
previously selected. Th e duty cycle represents the amount of time per
cycle that the square wave is at a high level (note the icon on the right
side of the display).
3 Enter the desired duty cycle.
Using the numeric keypad or the knob, select a duty cycle value of “30”.
The function generator adjusts the duty cy cle immediately and outputs a
square wave with the specified value (if the output is enabled).
2
4
29
Chapter 2 Quick Start
T o Configure a Pulse Waveform
To Configure a Pulse Waveform
2
You can configure the function generator to output a pulse waveform
with variable pulse width and edge time. The following steps show you
how to configure a 500 ms pulse waveform with a pulse width of 10 ms
and edge times of 50 ns.
1 Select the pulse function.
Press thekey to select the pulse function and output a pulse
waveform with the default parameters.
2 Set the pulse period.
Press the Period softkey and then set the pulse period to 500 ms.
3 Set the pulse width.
Press the Width softkey and then set the pulse width to 10 ms. The pulse
width represents the time from the 50% threshold of the rising edge to
the 50% threshold of the next falling edge (note the display icon).
4 Set the edge time for both edges.
Press the Edge Time softkey and then set the edge time for both the
rising and falling edges to 50 ns. The edge time represents the time from
the 10% threshold to the 90% threshold of each edge (note the display icon
30
).
Loading...
+ 155 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.