Agilent 33120A User's Guide

User’s Guide
Part Number 33120-90005 August 1997
For Safety information, Warranties, and Regulatory information, see the pages behind the Index.
© Copyright Hewlett-Packard Company 1994, 1997
All Rights Reserved.
HP 33120A Function Generator /
Arbitrary Waveform Generator
Note: Unless otherwise indicated, this manual applies to all Serial Numbers.
The HP 33120A is a high-performance 15 MHz synthesized function generator with built-in arbitrary waveform capability. Its combination of bench-top and system features makes this function generator a versatile solution for your testing requirements now and in the future.
Convenient bench-top features
10 standard waveforms
Built-in 12-bit 40 MSa/s arbitrary waveform capability
Easy-to-use knob input
Highly visible vacuum-fluorescent display
Instrument state storage
Portable, ruggedized case with non-skid feet
Flexible system features
Four downloadable 16,000-point arbitrary waveform memories
HP-IB (IEEE-488) interface and RS-232 interface are standard
SCPI (Standard Commands for Programmable Instruments) compatibility
Optional HP 34811A BenchLink/Arb Waveform Generation Software
®
for Microsoft
Windows
TM
HP 33120A Function Generator /
Arbitrary Waveform Generator

The Front Panel at a Glance

1 Function / Modulation keys 2 Menu operation keys 3 Waveform modify keys 4 Single / Internal Trigger key
Burst and Sweep only)
(
2
5 Recall / Store instrument state key 6 Enter Number key 7 Shift / Local key 8 Enter Number “units” keys

Front-Panel Number Entry

You can enter numbers from the front-panel using one of three methods.
Use the knob and the arrow keys to modify the displayed number.
Use the arrow keys to edit individual digits.
Increments the flashing digit.
Decrements the flashing digit.
Moves the flashing digit
Moves the flashing digit
to the right.
to the left.
Use the “Enter Number” mode to enter a number with the appropri ate units.
Use “Enter” for those operations that do not require units to be specified (AM Level, Offset, % Duty, and Store/Recall State).
3

The Front-Panel Menu at a Glance

The menu is organized in a top-down tree structure with three levels.
A: MODulation MENU
1: AM SHAPE 2: AM SOURCE 3: FM SHAPE 4: BURST CNT 5: BURST RATE
6: BURST PHAS 7: BURST SRC 8: FSK FREQ 9: FSK RATE 10: FSK SRC
B: SWP (Sweep) MENU
1: START F 2: STOP F 3: SWP TIME 4: SWP MODE
C: EDIT MENU
1: NEW ARB [ 2: POINTS ] [ 3: LINE EDIT ] [ 4: POINT EDIT ] [ 5: INVERT ] [ 6: SAVE AS ] 7: DELETE
The commands enclosed in square brackets ( [ ] ) are “hidden” until you make a selection from the
*
NEW ARB command to initiate a new edit session.
D: SYStem MENU
1: OUT TERM 2: POWER ON 3: ERROR 4: TEST 5: COMMA 6: REVISION
E: Input / Output MENU
1: HPIB ADDR 2: INTERFACE 3: BAUD RATE 4: PARITY 5: LANGUAGE
F: CALibration MENU
1: SECURED [ 1: UNSECURED ] [ 2: CALIBRATE ] 3: CAL COUNT 4: MESSAGE
The commands enclosed in square brackets ( [ ] ) are “hidden” unless the function generator
*
is UNSECURED for calibration.
*
*
4

Display Annunciators

Adrs Rmt Trig AM FM Ext FSK Burst Swp
ERROR
Offset Shift Num Arb
Function generator is addressed to listen or talk over a remote interface. Function generator is in remote mode (remote interface). Function generator is waiting for a single trigger or external trigger (Burst, Sweep). AM modulation is enabled. FM modulation is enabled. Function generator is set for an external modulation source (AM, FSK, Burst). FSK (frequency-shift keying) modulation is enabled. Burst modulation is enabled. Sweep mode is enabled. Hardware or remote interface command errors are detected. The waveform is being output with an offset voltage.
“Shift” key has been pressed. “Enter Number” mode is enabled.
Arbitrary waveform function is enabled. Sine waveform function is enabled. Square waveform function is enabled. Triangle waveform function is enabled. Ramp waveform function is enabled.
Press “Shift” again to turn off.
Press “Shift-Cancel” to disable.
To review the display annunciators, hold down the Shift key as you turn on the function generator.
5

The Rear Panel at a Glance

1 Chassis ground 2 Power-line fuse-holder assembly 3 Power-line voltage setting 4 AM modulation input terminal
Use the front-panel Input / Output Menu to:
5 External Trigger / FSK / Burst modulation
input terminal
6 HP-IB (IEEE-488) interface connector 7 RS-232 interface connector
Select the HP-IB or RS-232 interface (see chapter 4).
Set the HP-IB bus address (see chapter 4).
Set the RS-232 baud rate and parity (see chapter 4).
6

In This Book

Quick Start Chapter 1 prepares the function generator for use and helps you get familiar with a few of its front-panel features.
Front-Panel Menu Operation Chapter 2 introduces you to the
front-panel menu and describes some of the function generator’s menu features.
Features and Functions Chapter 3 gives a detailed description of the function generator’s capabilities and operation. You will find this chapter useful whether you are operating the function generator from the front panel or over the remote interface.
Remote Interface Reference Chapter 4 contains reference information to help you program the function generator over the remote interface.
Error Messages Chapter 5 lists the error messages that may appear as you are working with the function generator. Each listing contains enough information to help you diagnose and solve the problem.
Application Programs Chapter 6 contains several remote interface application programs to help you develop programs for your application.
Tutorial Chapter 7 discusses the fundamentals of signal generation and modulation techniques.
Specifications Chapter 8 lists the function generator’s specifications.
For information on using the Phase-Lock Option for the HP 33120A,
refer to the User’s and Service Guide included with the Option 001.
If you have questions relating to the operation of the HP 33120A, call 1-800-452-4844 in the United States, or contact your nearest Hewlett-Packard Sales Office.
If your HP 33120A fails within three years of purchase, HP will repair or replace it free of charge. Call 1-800-258-5165 (“Express Exchange”) in the United States, or contact your nearest Hewlett-Packard Sales Office.
7

Contents

Chapter 1 Quick Start
To prepare the function generator for use 15 If the function generator does not turn on 16 To adjust the carrying handle 18 To set the output frequency 19 To set the output amplitude 20 To set a dc offset voltage 21 To set the duty cycle 22 To output a stored arbitrary waveform 23 To output a dc voltage 24 To store the instrument state 25 To rack mount the function generator 27
Chapter 2 Front-Panel Menu Operation
Front-panel menu reference 31 A front-panel menu tutorial 33 To select the output termination 40 To output a modulated waveform 41 To output an FSK waveform 44 To output a burst waveform 47 To output a frequency sweep 49 To trigger a burst or sweep 51 To turn off the comma separator 52
Contents
Chapter 3 Features and Functions
Output Configuration 55 Amplitude Modulation (AM) 71 Frequency Modulation (FM) 76 Burst Modulation 81 Frequency-Shift Keying (FSK) Modulation 90 Frequency Sweep 94 Triggering 98 Arbitrary Waveforms 103 System-Related Operations 109 Remote Interface Configuration 114 Calibration Overview 118 Power-On and Reset State 123
9
Contents
Contents
Chapter 4 Remote Interface Reference
SCPI Command Summary 127 Simplified Programming Overview 136 Using the APPLy Command 138 Output Configuration Commands 145 AM Modulation Commands 154 FM Modulation Commands 157 Burst Modulation Commands 160 Frequency-Shift Keying (FSK) Commands 167 Frequency Sweep Commands 170 Arbitrary Waveform Commands 174 Triggering 186 System-Related Commands 188 Calibration Commands 193 RS-232 Interface Configuration 195 RS-232 Interface Commands 200 The SCPI Status Registers 201 Status Reporting Commands 209 An Introduction to the SCPI Language 211 Halting an Output in Progress 216 To set the HP-IB address 217 To select the remote interface 218 To set the baud rate 219 To set the parity 220 SCPI Conformance Information 221 IEEE-488 Conformance Information 225
Chapter 5 Error Messages
Execution Errors 229 Self-Test Errors 237 Calibration Errors 238 Arbitrary Waveform Errors 240
10
Contents
Chapter 6 Application Programs
HP BASIC Programs 244 C Language Programs 244 QuickBASIC Language Programs 247 Using the APPLy Command 248 Using the Low-Level Commands 252 Downloading an Arbitrary Waveform over HP-IB 255 Using the Status Registers 261 Downloading an Arbitrary Waveform over RS-232 267
Chapter 7 Tutorial
Direct Digital Synthesis 273 Signal Imperfections 276 Creating Arbitrary Waveforms 278 Output Amplitude Control 280 Floating Signal Generators 282 Attributes of AC Signals 283 Modulation 287
Contents
Chapter 8 Specifications
Frequency Characteristics 298 Sinewave Spectral Purity 298 Signal Characteristics 298 Output Characteristics 298 Modulation Characteristics 299 Frequency Sweep 299 System Characteristics 299 General Specifications 300 Product Dimensions 301
Index 303
Declaration of Conformity 309
11
1
1
Quick Start
Quick Start
One of the first things you will want to do with your function generator is to become acquainted with its front panel. We have written the exercises in this chapter to prepare the function generator for use and help you get familiar with some of the front-panel operations.
The front panel has two rows of keys to select various functions and operations. Most keys have a shifted function printed in blue above the key. To perform a shifted function, press ciator will turn on). Then, press the key that has the desired label above it. For example, to select the function, press
Shift AM (the shifted version of the key).
AM (amplitude modulation)
Shift (the Shift annun-
If you accidentally press
Shift annunciator.
Shift , just press it again to turn off the
Most keys also have a number printed in green next to the key. To enable the number mode, press
Enter Number (the Num annunciator
will turn on). Then, press the keys that have the desired numbers
printed next to them. For example, to select the number “10”,
Enter Number 1 0 (next to the and Recall keys).
press
If you accidentally press
Enter Number , just press Shift Cancel
to turn off the Num annunciator.
14
Chapter 1 Quick Start

To prepare the function generator for use

To prepare the function generator for use
The following steps help you verify that the function generator is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your function generator. If anything is missing, contact your nearest Hewlett-Packard Sales Office.
One power cord.
This User’s Guide.
One Service Guide.
One folded Quick Reference card.
Certificate of Calibration.
2 Connect the power cord and turn on the function generator.
1
The front-panel display will light up while the function generator performs its power-on self-test. The Notice that the function generator powers up in the sine wave function
at 1 kHz with an amplitude of 100 mV peak-to-peak (into a 50
termination).
To review the power-on display with all annunciators turned on, hold down
3Perform a complete self test.
The complete self-test performs a more extensive series of tests than those performed at power-on. Hold down switch to turn on the function generator; hold down the key for more than 5 seconds. The self-test will begin when you release the key.
If the self-test is successful, “ If the self-test is not successful, “ annunciator turns on. See the Service Guide for instructions on returning the function generator to Hewlett-Packard for service.
Shift as you turn on the function generator.
PASS” is displayed on the front panel.
HP-IB bus address is displayed.
Shift as you press the Power
FAIL” is displayed and the ERROR
15
Chapter 1 Quick Start

If the function generator does not turn on

If the function generator does not turn on
Use the following steps to help solve problems you might experience when turning on the function generator. If you need more help, see the Service Guide for instructions on returning the function generator to Hewlett-Packard for service.
1 Verify that there is ac power to the function generator.
First, verify that the function generator’s Power switch is in the “On” position. Also, make sure that the power cord is firmly plugged into to the power module on the rear panel. You should also make sure that the power source you plugged the function generator into is energized.
2 Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the function generator is shipped from the factory. Change the voltage setting if it is not correct. The settings are: 100, 120, 220, or 240 Vac (for 230 Vac operation, use the 220 Vac setting).
See the next page if you need to change the line-voltage setting.
3 Verify that the power-line fuse is good.
The function generator is shipped from the factory with a 500 mAT fuse installed. This is the correct fuse for all line voltages.
See the next page if you need to change the power-line fuse.
To replace the 500 mAT fuse, order HP part number 2110-0458.
16
Chapter 1 Quick Start
If the function generator does not turn on
1
1 Remove the power cord. Remove the
fuse-holder assembly from the rear panel.
3 Rotate the line-voltage selector until the
correct voltage appears in the window.
2 Remove the line-voltage selector from
the assembly.
Fuse: 500 mAT (for all line voltages) HP Part Number: 2110-0458
4 Replace the fuse-holder assembly in the rear panel.
100, 120, 220 (230), or 240 Vac
Verify that the correct line voltage is selected and the power-line fuse is good.
17
Chapter 1 Quick Start

To adjust the carrying handle

To adjust the carrying handle
To adjust the position, grasp the handle by the sides and pull outward. Then, rotate the handle to the desired position.
Bench-top viewing positions Carrying position
18
Chapter 1 Quick Start

To set the output frequency

To set the output frequency
At power-on, the function generator outputs a sine wave at 1 kHz with
an amplitude of 100 mV peak-to-peak (into a 50 termination).
The following steps show you how to change the frequency to 1.2 MHz.
1
Freq
Enter Number
.
1
2
MHz m Vpp
1 Enable the frequency modify mode.
The displayed frequency is either the power-on value or the previous frequency selected. When you change functions, the same frequency is used if the present value is valid for the new function.
1.000,000,0 KHz
2 Enter the magnitude of the desired frequency.
1
Notice that the Num annunciator turns on and “ENTER NUM” flashes on
the display, indicating that the number mode is enabled.
1.2
To cancel the number mode, press
Shift Cancel .
3 Set the units to the desired value.
The units are selected using the arrow keys on the right side of the front panel. As soon as you select the units, the function generator outputs the waveform with the displayed frequency. To turn off the
flashing digit, move the cursor to the left of the display using the arrow keys.
1.200,000,0 MHz
1
You can also use the knob and arrow keys to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
19
Chapter 1 Quick Start

To set the output amplitude

To set the output amplitude
At power-on, the function generator outputs a sine wave with an
amplitude of 100 mV peak-to-peak (into a 50Ω termination).
The following steps show you how to change the amplitude to 50 mVrms.
Ampl
Enter Number 5
0
Shift
kHz
m Vrms
1 Enable the amplitude modify mode.
The displayed amplitude is either the power-on value or the previous amplitude selected. When you change functions, the same amplitude is used if the present value is valid for the new function.
100.0 mVPP
2 Enter the magnitude of the desired amplitude.
1
Notice that the Num annunciator turns on and “ENTER NUM” flashes on
the display, indicating that the number mode is enabled.
50
To cancel the number mode, press
Shift Cancel .
3 Set the units to the desired value.
The units are selected using the arrow keys on the right side of the front panel. As soon as you select the units, the function generator outputs the waveform with the displayed amplitude. To turn off the
flashing digit, move the cursor to the left of the display using the arrow keys.
50.00 mVRMS
1
You can also use the knob and arrow keys to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
20
Chapter 1 Quick Start

To set a dc offset voltage

To set a dc offset voltage
At power-on, the function generator outputs a sine wave with a dc offset
voltage of 0 volts (into a 50Ω termination). The following steps show you
how to change the offset to –1.5 mVdc.
1
Offset
Enter Number ±
Shift
.
1
kHz m Vrms
1 Enable the offset modify mode.
The displayed offset voltage is either the power-on value or the previous offset selected. When you change functions, the same offset is used if the present value is valid for the new function.
+0.000 VDC
2 Enter the magnitude of the desired offset.
5
Notice that the Num annunciator turns on and “ENTER NUM” flashes on the display, indicating that the number mode is enabled. Notice that
±
toggles the displayed value between + and – .
-1.5
To cancel the number mode, press
Shift Cancel .
3 Set the units to the desired value.
At this point, the function generator outputs the waveform with the displayed offset. Notice that the
Offset annunciator turns on, indicating
that the waveform is being output with an offset. The annunciator will turn on when the offset is any value other than 0 volts. To turn off the
flashing digit, move the cursor to the left of the display using the arrow keys.
1
-01.50 mVDC
1
You can also use the knob and arrow keys to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
21
Chapter 1 Quick Start

To set the duty cycle

To set the duty cycle
Applies only to square waves. At power-on, the duty cycle for square waves is 50%. You can adjust the duty cycle for a square waveform from 20% to 80%, in increments of 1% (for frequencies above 5 MHz, the range is 40% to 60%). The following steps show you how to change the
duty cycle to 45%.
% Duty Shift
Enter Number
5
4
Enter
1 Select the square wave function.
Notice that the annunciator turns on, indicating that the square wave function is enabled.
2 Enable the duty cycle modify mode.
The displayed duty cycle is either the power-on value or the previous value selected.
50 % DUTY
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
3 Enter the desired duty cycle.
1
Notice that the Num annunciator turns on and “ENTER NUM” flashes on
the display, indicating that the number mode is enabled.
45
To cancel the number mode, press
Shift Cancel .
4 Output the waveform with the displayed duty cycle.
45 % DUTY
1
You can also use the knob and arrow keys to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
22
Chapter 1 Quick Start

To output a stored arbitrary waveform

To output a stored arbitrary waveform
There are five built-in arbitrary waveforms stored in non-volatile memory for your use. You can output these waveforms directly from non-volatile memory. The following steps show you how to output an
“exponential rise” waveform from memory.
1
Shift
>
Enter
>
Arb List
1 Display the list of arbitrary waveforms.
The list contains the five built-in arbitrary waveforms (sinc, negative ramp, exponential rise, exponential fall, and cardiac). The list may also contain up to four user-defined arbitrary waveform names. The first
choice on this level is “
SINC”.
SINC
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
2 Move across to the
EXP_RISE choice.
1
EXP_RISE
3 Select and output the displayed arbitrary waveform.
Notice that the
Arb annunciator turns on, indicating that the output is
an arbitrary waveform. The waveform is output using the present settings for frequency, amplitude, and offset unless you change them.
The selected waveform is now assigned to the
Arb key. Whenever
you press this key, the selected arbitrary waveform is output.
1
You can also use the knob to scroll left or right through the choices in the list.
See “Front-Panel Number Entry” on page 3 for more information.
23
Chapter 1 Quick Start

To output a dc voltage

To output a dc voltage
In addition to generating waveforms, you can also output a dc voltage in
the range ± 5 Vdc (into a 50Ω termination). The following steps show
you how to output +155 mVdc.
Enter Number
5
5 1
Shift
kHz
m Vrms
1 Press the key and hold it down for more than 2 seconds.
To enter the dc voltage mode, press the top row of function keys and
Offset
Offset key or any key in the
hold it down for more than 2 seconds.
The displayed voltage is either the power-on value or the previous offset voltage selected.
DCV
+0.000 VDC
2 Enter the magnitude of the desired voltage.
1
Notice that the Num annunciator turns on and “ENTER NUM” flashes on
the display, indicating that the number mode is enabled.
155
To cancel the number mode, press
Shift Cancel .
3 Set the units to the desired value.
At this point, the function generator outputs the displayed dc voltage. Notice that the
Offset annunciator turns on (all other annunciators
are off), indicating that a dc voltage is being output. The annunciator will turn on when the offset is any value other than 0 volts.
+155.0 mVDC
1
You can also use the knob and arrow keys to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
24
Chapter 1 Quick Start

To store the instrument state

To store the instrument state
You can store up to three different instrument states in non-volatile memory. This enables you to recall the entire instrument configuration with just a few key presses from the front panel. The following steps
show you how to store and recall a state.
1 Set up the function generator to the desired configuration.
The state storage feature “remembers” the function, frequency, amplitude, dc offset, duty cycle, as well as any modulation parameters.
1
Enter
Store Shift
2 Turn on the state storage mode.
Three memory locations (numbered 1, 2, and 3) are available to store instrument configurations. The instrument configuration is stored in non-volatile memory and is remembered when power has been off.
STORE 1
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
3 Store the instrument state in memory location “2”.
1
Use the up and down arrow keys to select the memory location.
STORE 2
To cancel the store operation, press
Shift Store again or let the
display time-out after 10 seconds.
4 Save the instrument state.
The instrument state is now stored. To recall the stored state, turn to the next page.
1
You can also use the knob or “enter number” mode to enter a memory location.
See “Front-Panel Number Entry” on page 3 for more information.
25
Chapter 1 Quick Start
To store the instrument state
To verify that the state was stored properly, you can turn the power off before recalling the state.
Recall
Enter
5 Recall the stored instrument state.
To recall the stored state, you must use the same memory location used previously to store the state. Use the up and down arrow keys to change the displayed storage location.
RECALL 2
To cancel the restore operation, press
Recall again.
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
6 Restore the instrument state.
The function generator should now be configured in the same state as when you stored the setup on the previous page.
When power is turned off, the function generator automatically stores
its state in memory location “0”. You can recall the power-down state, but you cannot store the state to location “0” from the front panel.
Use the POWER ON LAST STATE command in the SYS MENU to automatically recall the power-down state when power is turned on. See chapter 2 for more information on using the front-panel menus.
26
Chapter 1 Quick Start

To rack mount the function generator

To rack mount the function generator
You can mount the function generator in a standard 19-inch rack cabinet using one of three optional kits available. Instructions and mounting hardware are included with each rack-mounting kit.
HP System II instrument of the same size can be rack-mounted
Any beside the HP 33120A Function Generator.
Remove the carrying handle, and the front and rear rubber bumpers, before rack-mounting the function generator.
1
To remove the handle, rotate it to the vertical position and pull the ends outward.
Front Rear (bottom view)
To remove the rubber bumper, stretch a corner and then slide it off.
27
Chapter 1 Quick Start
To rack mount the function generator
To rack mount a single instrument, order adapter kit 5063- 9240.
To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and flange kit 5063- 9212.
To instal l one or two instr uments in a slidin g support shelf, or der shelf 5063-9 255, and slide kit 1494-0015 (for a single instrument, also order filler panel 5002-3999).
28
2
2
Front-Panel Menu Operation
Front-Panel Menu Operation
By now you should be familiar with some of the basic features of the front panel. Chapter 1 shows you how to prepare the function generator for use and describes a few of the front-panel features. If you are not familiar with this information, we recommend that you read chapter 1,
“Quick Start,” starting on page 13.
Chapter 2 introduces you to the use of the front-panel menu. This chapter does not give a detailed description of every front-panel key or menu operation. It does, however, give you a good overview of the front-panel menu and many front-panel operations. See chapter 3 “Features and Functions,” starting on page 53, for a complete discussion of the function generator’s capabilities and operation.
If you purchased the Phase-Lock Option for the HP 33120A, an additional menu
For information on using the Phase-Lock Option, refer to the User’s and Service Guide included with Option 001.
(G: PHASE MENU) is available from the front panel.
30
Chapter 2 Front-Panel Menu Operation

Front-panel menu reference

Front-panel menu reference
A: MODulation MENU
1: AM SHAPE 2: AM SOURCE 3: FM SHAPE 4: BURST CNT 5: BURST RATE
2
6: BURST PHAS 7: BURST SRC 8: FSK FREQ 9: FSK RATE 10: FSK SRC
1: AM SHAPE 2: AM SOURCE 3: FM SHAPE 4: BURST CNT 5: BURST RATE 6: BURST PHAS 7: BURST SRC 8: FSK FREQ 9: FSK RATE 10: FSK SRC
Selects the shape of the AM modulating waveform. Enables or disables the internal AM modulating source. Selects the shape of the FM modulating waveform. Sets the number of cycles per burst (1 to 50,000 cycles). Sets the burst rate in Hz for an internal burst source. Sets the starting phase angle of a burst (-360 to +360 degrees). Selects an internal or external gate source for burst modulation.
Sets the FSK “hop” frequency. Selects the internal FSK rate between the carrier and FSK frequency. Selects an internal or external source for the FSK rate.
B: SWP (Sweep) MENU
1: START F 2: STOP F 3: SWP TIME 4: SWP MODE
1: START F 2: STOP F 3: SWP TIME 4: SWP MODE
C: EDIT MENU
1: NEW ARB [ 2: POINTS ] [ 3: LINE EDIT ] [ 4: POINT EDIT ] [ 5: INVERT ] [ 6: SAVE AS ] 7: DELETE
*
Sets the start frequency in Hz for sweeping. Sets the stop frequency in Hz for sweeping. Sets the repetition rate in seconds for sweeping. Selects linear or logarithmic sweeping.
1: NEW ARB 2: POINTS 3: LINE EDIT 4: POINT EDIT 5: INVERT 6: SAVE AS 7: DELETE
The commands enclosed in square brackets ( [ ] ) are “hidden” until you make a selection from the NEW ARB command
*
to initiate a new edit session.
Initiates a new arb waveform or loads the selected arb waveform. Sets the number of points in a new arb waveform (8 to 16,000 points). Performs a linear interpolation between two points in the arb waveform. Edits the individual points of the selected arb waveform. Inverts the selected arb waveform by changing the sign of each point. Saves the current arb waveform in non-volatile memory. Deletes the selected arb waveform from non-volatile memory.
31
Chapter 2 Front-Panel Menu Operation
Front-panel menu reference
D: SYStem MENU
1: OUT TERM 2: POWER ON 3: ERROR 4: TEST 5: COMMA 6: REVISION
1: OUT TERM 2: POWER ON 3: ERROR 4: TEST 5: COMMA 6: REVISION
Selects the output termination (50 Enables or disables automatic recall of the power-down state. Retrieves errors from the error queue (up to 20 errors). Performs a complete self-test. Enables or disables a comma separator between digits on the display.
Displays the function generator’s firmware revision codes.
or high impedance).
E: Input / Output MENU
1: HPIB ADDR 2: INTERFACE 3: BAUD RATE 4: PARITY 5: LANGUAGE
1: HPIB ADDR 2: INTERFACE 3: BAUD RATE 4: PARITY 5: LANGUAGE
F: CALibration MENU
1: SECURED [ 1: UNSECURED ] [ 2: CALIBRATE ] 3: CAL COUNT 4: MESSAGE
*
1: SECURED 1: UNSECURED 2: CALIBRATE 3: CAL COUNT 4: MESSAGE
Sets the HP-IB bus address (0 to 30). Selects the HP-IB or RS-232 interface. Selects the baud rate for RS-232 operation. Selects even, odd, or no parity for RS-232 operation. Verifies the interface language: SCPI.
The function generator is secured against calibration; enter code to unsecure. The function generator is unsecured for calibration; enter code to secure. Performs individual calibrations; must be UNSE CURED. Reads the total number of times the function generator has been calibrated. Reads the calibration string (up to 11 characters) entered from remote.
The commands enclosed in square brackets ( [ ] ) are “hidden” unless the function generator is UNSECURED for calibration.
*
32
Chapter 2 Front-Panel Menu Operation

A front-panel menu tutorial

A front-panel menu tutorial
This section is a step-by-step tutorial which shows you how to use the front-panel menu. We recommend that you spend a few minutes with this tutorial to get comfortable with the structure and operation of the menu.
The menu is organized in a top-down tree structure with three levels (menus, commands, and parameters). You move down or up
the menu tree to get from one level to the next. Each of the
three levels has several horizontal choices which you can view by moving left
< or right > .
2
Commands
Parameters
Menus
The menu is organized in a top-down tree structure with three levels.
To turn on the menu, press
To turn off the menu, press
To execute a menu command, press
Shift Menu On/Off .
Shift Menu On/Off .
Enter .
To recall the last menu command that was executed,
Shift Recall Menu .
press
To turn off the menu at any time without saving changes,
Shift Cancel .
press
33
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial
Messages Displayed During Menu Use
TOP OF MENU You pressed while on the “MENUS” level; this is the top level of the
menu and you cann ot go an y hi gh er.
Shift
To turn off the menu, press . To move across the choices on a level, pre ss or . To move down a level, press .
MENUS You are on the “MENUS” level. Press or to view the choices. COMMANDS You are on the “COMMANDS” level. Press or to view the command
<
>
Menu On/Off
<
>
<
>
choices within the selected menu group.
>
PARAMETER You are on the “PARAMETER” level. Press or to view and edit the
<
parameter for the selected command.
MENU BOTTOM You pressed while on the “PARAMETER ” level; this is the bottom
level of the menu and you cannot go any lower.
Menu On/Off
To turn off the menu, press . To move up a level, press .
ENTERED The change made on the “PARAMETER” level is saved. This is displayed after
you press (Menu Enter) to execute the command.
MIN VALUE The value you specified on the “PARAMETER” level is too small for the
Enter
Shift
selected command. The minimum value allowed is displayed for you to edit.
MAX VALUE The value you specified on the “PARAMETER” level is too larg e fo r the
selected comma nd. The maximum value allowed is displayed for you to edit.
EXITING You will see this message if you turn off the menu by pressing
Menu On/Off Shift Cancel
Shift
or . You did not edit any values on the “PARAMETER” level and changes were NOT sa ve d.
NOT ENTERED You will see this message if you turn off the menu by pressing
Shift
Menu On/Off
or . You did some editing of parameters but the changes were NOT saved. Press the key (Menu Enter) to save changes made on the “
PARAMETER” level.
Shift
Cancel
Enter
34
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial

Menu Example 1 The following steps show you how to turn on the menu, move up and

down between levels, move across the choices on each level, and turn off the menu. In this example, you will enable the function generator to
automatically recall the power-down state when power is turned on.
2
Shift Menu On/Off
>
>
>
>
1 Turn on the menu.
You enter the menu on the “
MENUS” level. The MOD MENU is your first
choice on this level.
A: MOD MENU
2
Move across to the
SYS MENU choice on this level.
1
There are six menu group choices available on the “MENUS” level. Each choice has a letter prefix for easy identification (
A: , B: , etc.).
D: SYS MENU
3 Move down to the “
OUT TERM command is your first choice on this level.
The
COMMANDS” level within the SYS MENU.
1: OUT TERM
4 Move across to the
POWER ON command on this level.
1
There are six command choices available in the SYS MENU. Each choice on this level has a number prefix for easy identification (
1: , 2: , etc.).
2: POWER ON
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
35
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial
>
Enter
5 Move down a level to the “PARAMETER” choices.
The first parameter choice is “
DEFAULT” is the factory setting and is stored in non-volatile memory).
(“
DEFAULT” for the POWER ON command
DEFAULT
6 Move across to the
“LAST STATE” choice.
1
There are two parameter choices for POWER ON.
LAST STATE
7 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
ENTERED
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
36
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial

Menu Example 2 The following example shows how to use the recall menu feature as a

shortcut to set the
POWER ON command back to its original setting.
You must perform the steps in Example 1 before you start this example.
Shift <
Recall Menu
1 Use recall menu to return to the
This returns you to the
POWER ON command, which was the last
POWER ON command.
command used before you exited the menu in the Example 1.
2: POWER ON
2
>
Enter
2 Move down to the “
The first parameter choice is “
PARAMETER” choices.
LAST STATE” (the current setting from
Example 1).
LAST STATE
Move across to the
3
“DEFAULT” choice.
1
Set the parameter back to its original value.
DEFAULT
4 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
ENTERED
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
37
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial

Menu Example 3 Some commands in the menu require that you enter a numeric

parameter value. The following steps show you how to enter a number in the menu. For this example, you will set the number of cycles for the
burst modulation mode.
Shift Menu On/Off
> > >
1 Turn on the menu.
You enter the menu on the “
MENUS” level. The MOD MENU is your first
choice on this level.
A: MOD MENU
2 Move down to the “
AM SHAPE command is your first choice on this level.
The
COMMANDS” level within the MOD MENU.
1: AM SHAPE
3 Move across to the
BURST CNT command on this level.
There are ten command choices available in the MOD MENU.
4: BURST CNT
1
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
38
Chapter 2 Front-Panel Menu Operation
A front-panel menu tutorial
<
4 Move down a level to edit the BURST CNT parameter.
The number of cycles should be “1” when you come to this point in the menu for the first time. For this example, you will set the number of cycles to “4”. Notice that the
Burst annunciator flashes, indicating that
the displayed value is for the burst mode.
00001 CYC
When you see the flashing “” on the left side of the display, you can
press
to abort the edit and return to the “COMMANDS” level.
5 Move the flashing cursor over to edit the last digit.
Notice that the rightmost digit is flashing (the digit wraps around).
00001 CYC
6 Increment the last digit until “4” is displayed.
1
You decrement or increment each digit independently.
2
Enter
00004 CYC
7 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
ENTERED
1
You can also use the knob or “enter number” mode to enter a number.
See “Front-Panel Number Entry” on page 3 for more information.
39
Chapter 2 Front-Panel Menu Operation

To select the output termination

To select the output termination
The function generator has a fixed output impedance of 50 ohms on the OUTPUT terminal. You can specify whether you are terminating the output into a 50 between the source and load will result in an output amplitude or dc offset which does not match the specified value.
load or an open circuit. Incorrect impedance matching
Shift Menu On/Off
> > >
>
1 Turn on the menu.
A: MOD MENU
2 Move across to the SYS MENU choice on this level.
D: SYS MENU
3 Move down a level to the
OUT TERM command.
1: OUT TERM
4 Move down a level and then across to the
HIGH Z choice.
With the output termination set to “HIGH Z”, the function generator allows you to set the unloaded (open circuit) output voltage.
HIGH Z
1
1
Enter
5 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
40
Chapter 2 Front-Panel Menu Operation

To output a modulated waveform

To output a modulated waveform
A modulated waveform consists of a carrier and a modulating waveform. In is varied by the amplitude of the modulating waveform. For this
example, you will output an AM waveform with 80% modulation depth. The carrier will be a 5 kHz sine wave and the modulating waveform will be a 200 Hz sine wave.
AM (amplitude modulation), the amplitude of the carrier
2
Shift <
Recall Menu
1 Select the function, frequency, and amplitude of the carrier.
For the carrier waveform, you can select a sine, square, triangle, ramp, or arbitrary waveform. For this example, select a
an amplitude of
AM Shift
2 Select AM.
Notice that the
5 Vpp.
AM annunciator turns on.
5 kHz sine wave with
3 Use the menu to select the shape of the modulating waveform.
After you enable the
automatically take you to the
AM function, the “recall menu” key will
AM SHAPE command in the MOD MENU.
1: AM SHAPE
41
Chapter 2 Front-Panel Menu Operation
To output a modulated waveform
Enter
4 Move down a level and verify that “SINE” is selected.
For the modulating waveform, you can select a sine, square, triangle, ramp, noise, or arbitrary waveform. For this example, you will modulate the carrier with a sine waveform. Notice that the flashes, indicating that the displayed parameter is for
AM annunciator
AM.
SINE
5 Save the change and turn off the menu.
The modulating waveform is now a sine waveform.
ENTERED
Freq Shift
6 Set the modulating frequency to 200 Hz.
Notice that the frequency is the modulating frequency for
AM annunciator flashes, indicating that the displayed
AM. Also notice that the
modulating frequency is displayed with fewer digits than the carrier frequency. For more information on editing numbers in the menu,
refer to “Menu Example 3” earlier in this chapter.
MOD 200.0 Hz
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
42
Chapter 2 Front-Panel Menu Operation
To output a modulated waveform
Shift
Level
7 Set the modulation depth to 80%.
Notice that the percentage is the
080 % DEPTH
This message appears on the display for approximately 10 seconds. Repeat this step as needed.
At this point, the function generator outputs the specified modulation parameters.
AM annunciator flashes, indicating that the displayed
AM depth (also called percent modulation).
AM waveform with the
2
43
Chapter 2 Front-Panel Menu Operation

To output an FSK waveform

To output an FSK waveform
You can configure the function generator to “shift” its output frequency between two preset values using The rate at which the output shifts between the two frequencies (called the “carrier frequency” and the “hop frequency”) is determined by the internal rate generator or the signal level on the rear-panel FSK terminal. For this example, you will set the “carrier” frequency to 3 kHz and
the “hop” frequency to 500 Hz, with an FSK rate of 100 Hz.
FSK (frequency-shift keying) modulation.
Shift
Shift
<
Recall Menu
FSK
1 Select the function, frequency, and amplitude of the carrier.
For the carrier waveform, you can select a sine, square, triangle, ramp, or arbitrary waveform. For this example, select a
an amplitude of
5 Vpp.
3 kHz sine wave with
2 Enable FSK.
Notice that the
FSK annunciator turns on.
3 Use the menu to set the “hop” frequency.
After you enable the
take you to the
8: FSK FREQ
FSK mode, the “recall menu” key will automatically
FSK FREQ command in the MOD MENU.
44
Chapter 2 Front-Panel Menu Operation
To output an FSK waveform
Enter
Shift <
Recall Menu
4 Move down a level and set the “hop” frequency to 500 Hz.
Notice that the parameter is for the
FSK annunciator flashes, indicating that the displayed
FSK mode. Also notice that the hop frequency is
displayed with fewer digits than the carrier frequency. For more
information on editing numbers in the menu, refer to “Menu Example 3” earlier in this chapter.
500.0 Hz
5 Save the change and turn off the menu.
At this point, the carrier and hop frequencies are set and output.
Now, we will go back into the menu to set the FSK “shift” rate — this is the rate at which the function generator shifts between the carrier frequency and hop frequency.
6 Use the menu to set the
The “recall menu” key returns you to the
FSK “shift” rate.
FSK FREQ command, which
was the last command used before you exited the menu.
2
8: FSK FREQ
45
Chapter 2 Front-Panel Menu Operation
To output an FSK waveform
>
Enter
7 Move across to the FSK RATE command.
9: FSK RATE
8 Move down a level and set the
Notice that the parameter is for the
FSK annunciator flashes, indicating that the displayed
FSK mode. For more information on editing
FSK “shift” rate to 100 Hz.
numbers in the menu, refer to “Menu Example 3” earlier in this chapter.
100.0 Hz
9 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
ENTERED
At this point, the function generator outputs the
FSK waveform.
You can also use an external signal to shift between the carrier frequency and hop frequency. For more information, see “Frequency-Shift Keying (FSK) Modulation” in chapter 3.
46
Chapter 2 Front-Panel Menu Operation

To output a burst waveform

To output a burst waveform
You can configure the function generator to output a waveform with a specified number of cycles, called a burst. You can output the burst at a rate determined by the internal rate generator or the signal level on the rear-panel Ext Trig terminal. For this example, you will output a
three-cycle sine wave. You will not change the other parameters from their default settings: internal burst source, 0 degree starting phase, and 100 Hz burst rate.
2
Shift
Shift <
Recall Menu
Burst
1 Select the function, frequency, and amplitude for the burst.
For bursts, you can select sine, square, triangle, ramp, or arbitrary waveforms. For this example, select a
amplitude of
5 Vpp.
1 kHz sine wave with an output
2 Enable the burst mode.
Notice that the
Burst annunciator turns on.
3 Use the menu to set the burst count.
After you enable the burst mode, the “recall menu” key will automatically take you to the
BURST CNT command in the MOD MENU.
4: BURST CNT
47
Chapter 2 Front-Panel Menu Operation
To output a burst waveform
Enter
4 Move down to the parameter level and set the count to “3”.
Notice that the
Burst annunciator flashes, indicating that the displayed
parameter is for the burst mode. For more information on editing
numbers in the menu, refer to “Menu Example 3” earlier in this chapter.
00003 CYC
You can also select an infinite burst count. Press the right or left arrow keys until the “ to display “
CYC” units are flashing. Then, press the down arrow key
INFINITE”.
5 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the change is now in effect. You are then exited from the menu.
ENTERED
At this point, the function generator outputs a continuous three-cycle burst.
You can generate a single burst (with the specified burst count) by pressing the “Single” key from the front panel or applying a trigger pulse to the rear-panel Ext Trig terminal. For more information, see “To trigger a burst or sweep” on page 51.
You can also use an external gate signal to turn the output signal on and off. For more information, see “Burst Modulation” in chapter 3.
48
Chapter 2 Front-Panel Menu Operation

To output a frequency sweep

To output a frequency sweep
In the frequency sweep mode, the function generator “steps” from the start frequency to the stop frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with either linear or logarithmic spacing. For this example, you will output a swept sine wave
from 50 Hz to 5 kHz. You will not change the other parameters from their default settings: internal sweep trigger, linear spacing, and 1 second sweep time.
2
Shift
Shift <
Recall Menu
Sweep
1 Select the function and amplitude for the sweep.
For sweeps, you can select sine, square, triangle, ramp, or arbitrary waveforms. For this example, select a sine wave with a
5 Vpp amplitude.
2 Enable the sweep mode.
Notice that the
Swp annunciator turns on.
3 Use the menu to select the start frequency.
After you enable the sweep mode, the “recall menu” key will automatically take you to the
START F command in the SWP MENU.
1: START F
49
Chapter 2 Front-Panel Menu Operation
To output a frequency sweep
Enter
Shift <
Recall Menu
>
4 Move down a level and set the start frequency to 50 Hz.
Notice that the
Swp annunciator flashes, indicating that the displayed
parameter is for the sweep mode. For more information on editing
numbers in the menu, refer to “Menu Example 3” earlier in this chapter.
50.00 Hz
5 Save the change and turn off the menu.
The start frequency is now set to 50 Hz.
6 Use recall menu to return to the
The “recall menu” key returns you to the
START F command.
START F command, which was
the last command used before you exited the menu.
1: START F
7 Move across to the
STOP F command.
2: STOP F
Enter
8 Move down a level and set the stop frequency to 5 kHz.
5.000 KHz
9 Save the change and turn off the menu.
At this point, the function generator outputs a continuous sweep from 50 Hz to 5 kHz.
You can generate one frequency sweep by pressing the “Single” key or applying a trigger pulse to the rear-panel Ext Trig terminal. For more information, see “To trigger a burst or sweep” on the next page.
50
Chapter 2 Front-Panel Menu Operation

To trigger a burst or sweep

To trigger a burst or sweep
You can issue triggers from the front-panel for burst modulation and frequency sweeps using single trigger or internal trigger.
Enables single trigger and triggers the generator.
2
Enables internal trigger.
Trig annunciator is on when the generator is waiting for single trigger (internal trigger disabled).
Internal or “automatic” triggering is enabled when you turn on the
function generator. In this mode, the function generator outputs continuously when burst modulation or sweep is selected.
Single triggering outputs one burst or initiates one frequency sweep
each time you press
Single . Continue pressing this key to re-trigger
the function generator.
Pressing
Single to enable the single trigger mode also enables the
external trigger mode. External triggering is like the single trigger
mode except that you apply a trigger signal to the rear-panel Ext Trig terminal. The function generator is triggered on the rising edge of a
TTL pulse.
The
Single key is disabled when in remote (the Rmt annunciator
turns on when in remote) and when a function other than burst modulation or sweep is selected.
51
Chapter 2 Front-Panel Menu Operation

To turn off the comma separator

To turn off the comma separator
The function generator can display values on the front panel with or without a comma separator. The following steps show how to turn off the comma separator.
Shift Menu On/Off
>
> >
>
15.000,000 MHz
With comma separator Without comma separator
(factory setting)
15.000000 MHz
1 Turn on the menu.
A: MOD MENU
2 Move across to the
SYS MENU choice on the “MENUS” level.
D: SYS MENU
< <
3 Move down a level and then across to the
COMMA command.
5: COMMA
4 Move down a level and then across to the “
OFF” choice.
Enter
OFF
5 Save the change and turn off the menu.
The comma separator setting is stored in non-volatile memory, and does not change when power has been off or after a remote interface reset.
52
3
3
Features and Functions
Features and Functions
You will find that this chapter makes it easy to look up all the details about a particular feature of the function generator. Whether you are operating the function generator from the front panel or over the remote interface, this chapter will be useful. This chapter is divided into the following sections:
Output Configuration, starting on page 55
Amplitude Modulation (AM), starting on page 71
Frequency Modulation (FM), starting on page 76
Burst Modulation, starting on page 81
Frequency-Shift Keying (FSK) Modulation, starting on page 90
Frequency Sweep, starting on page 94
Triggering, starting on page 98
Arbitrary Waveforms, starting on page 103
System-Related Operations, starting on page 109
Remote Interface Configuration, starting on page 114
Calibration Overview, starting on page 118
Power-On and Reset State, on page 123
Some knowledge of the front-panel menu will be helpful before you read
this chapter. If you have not already read chapter 2, “Front-Panel Menu Operation,” starting on page 29, you may want to read it now. Chapter 4, “Remote Interface Reference,” starting on page 125, lists the syntax for
SCPI commands available to program the function generator.
the
Throughout this manual, the following conventions are used for
SCPI command syntax for remote interface programming.
Square brackets ( [ ] ) indicate optional keywords or parameters.
Braces ( { } ) enclose parameters within a command string.
Triangle brackets ( < > ) indicate that you must substitute a value
for the enclosed parameter.
A vertical bar (
54
|
) separates multiple parameter choices.
Chapter 3 Features and Functions

Output Configuration

Output Configuration
This section contains information to help you configure the function generator for outputting waveforms. You may never have to change some of the parameters discussed here, but they are provided to give you the flexibility you might need.

Output Function

The function generator can output five standard waveforms including sine, square, triangle, ramp, and noise. You can also select one of five predefined arbitrary waveforms or download your own custom waveforms. You can internally modulate any of the standard waveforms (including arbitrary) using frequency sweeping is available for any of the standard waveforms (except noise) and arbitrary waveforms. The default function is sine wave.
Possible Conflict with Output Frequency: The output frequency is
automatically adjusted if you select a function whose maximum frequency is less than that of the currently active function. For example, if you output a 1 MHz sine wave and then change the function to triangle wave, the function generator will adjust the output to 100 kHz (the upper limit for triangle waves).
AM, FM, FSK, or burst modulation. Linear or logarithmic
3
From the front panel, “ adjusted. From the remote interface, a -221, “Settings conflict” error
is generated and the frequency is adjusted.
Possible Conflict with Output Amplitude: The output amplitude is
automatically adjusted if you select a function whose maximum amplitude is less than that of the currently active function. This conflict may arise when the output units are Vrms or dBm due to the differences in crest factor for the output functions. For example, if you output a 5 Vrms square wave (into 50 ohms) and then change the function to sine wave, the function generator will adjust the output amplitude to 3.535 Vrms (the upper limit for sine waves in Vrms).
From the front panel, “ is adjusted. From the remote interface, a -221, “Settings conflict”
error is generated and the amplitude is adjusted.
FREQ LIMIT” is displayed and the frequency is
AMPL LIMIT” is displayed and the amplitude
55
Chapter 3 Features and Functions
Output Configuration
Output Function
(continued)
The following matrix shows which output functions are allowed with
each modulation mode. Each “
X” indicates a valid combination. If you
change to a function that is not allowed with the selected modulation, the modulation mode is turned off.
AM Carrier AM Modulating Wav e FM Carrier FM Modulating Wave FSK Modulation Burst Modulation Frequency Sweep
Sine
X X X X X X X
Square
X X X X X X X
Triangle
X X X X X X X
Ramp
X X X X X X X
Noise
X X
Arb
X X X X X X X
Front-Panel Operation: To select a function, press any key in the top
row of function keys. Press
Arb to output the arbitrary waveform
currently selected (to scroll through the waveform choices and make a selection, press
To select
dc volts from the front panel, press Offset and hold it down
for more than 2 seconds
Arb List ).
.
Remote Interface Operation:
FUNCtion:SH APe {SIN usoid|SQ Uare|T RIangl e|RAMP |NOISe|U SER|DC }
You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command.
56
Chapter 3 Features and Functions
Output Configuration

Output Frequency

As shown below, the output frequency range depends on the function currently selected. The default frequency is 1 kHz for all functions.
Function
Minimum Frequency
Maximum Frequency
Sine Square Triangle Ramp Built-In Arbs
1 There are five built-in arbitrary waveforms stored in non-volatile memory: sinc, negative ramp, exponential rise, exponential fall, and cardiac.
1
100 100 100 100 100
µHz µHz µHz µHz µHz
15 MHz 15 MHz 100 kHz 100 kHz 5 MHz
For arbitrary waveforms that you create and download to memory, the maximum frequency depends on the number of points specified in the waveform. As shown below, the maximum output frequency decreases as you specify more points in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
Minimum Frequency
100 100 100
µHz µHz µHz
Maximum Frequency
5 MHz
2.5 MHz 200 kHz
3
57
Chapter 3 Features and Functions
Output Configuration
Output Frequency
(continued)
Possible Conflict with Function Change: The output frequency is
automatically adjusted if you select a function whose maximum frequency is less than that of the currently active function. For example, if you output a 1 MHz sine wave and then change the function to triangle wave, the function generator will adjust the output to 100 kHz (the upper limit for triangle waves).
From the front panel, “
FREQ LIMIT” is displayed and the frequency is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the frequency is adjusted.
Possible Conflict with Duty Cycle (square wave only): For output
frequencies above 5 MHz, the duty cycle is limited to values between 40% and 60% (below 5 MHz, the range is 20% to 80%). The duty cycle is automatically adjusted if you select a frequency that is not valid with the present duty cycle. For example, if you set the duty cycle to 70% and then change the frequency to 8 MHz, the function generator will automatically adjust the duty cycle to 60% (the upper limit for this frequency).
From the front panel, “
% DUTY LIMIT” is displayed and the duty cycle is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the duty cycle is adjusted.
Front-Panel Operation: To set the output frequency, press
Freq .
Then use the knob, arrow keys, or “Enter Number” mode to set the frequency.
Remote Interface Operation:
FREQuency {<
frequency>|MINimum|MAXimum}
You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command.
58
Chapter 3 Features and Functions
Output Configuration

Output Amplitude

As shown below, the output amplitude range depends on the function currently selected and the output termination. The default amplitude is
100 mVpp (into 50 ohms) for all functions.
Function
Output Termination
Minimum Amplitude
Maximum Amplitude
Sine Square Triangle Ramp Noise Built-In Arbs
Sine Square Triangle Ramp Noise Built-In Arbs
1 There are five built-in arbitrary waveforms stored in non-volatile memory: sinc, negative ramp, exponential rise, exponential fall, and cardiac.
1
1
50 50 50 50 50 50
Open Circuit Open Circuit Open Circuit Open Circuit Open Circuit Open Circuit
50 mVpp 50 mVpp 50 mVpp 50 mVpp 50 mVpp 50 mVpp
100 mVpp 100 mVpp 100 mVpp 100 mVpp 100 mVpp 100 mVpp
10 Vpp 10 Vpp 10 Vpp 10 Vpp 10 Vpp 10 Vpp
20 Vpp 20 Vpp 20 Vpp 20 Vpp 20 Vpp 20 Vpp
For arbitrary waveforms, the maximum amplitude will be limited if
the data points do not span the full range of the output
to-Analog Converter). For example, the built-in “ not use the full range of values between
±1 and therefore its
SINC” waveform does
DAC (Digital-
maximum amplitude is 6.084 Vpp (into 50 ohms).
Possible Conflict with Function Change: The output amplitude is
automatically adjusted if you select a function whose maximum amplitude is less than that of the currently active function. This conflict may arise when the output units are Vrms or dBm due to the differences in crest factor for the output functions. For example, if you output a 5 Vrms square wave (into 50 ohms) and then change the function to sine wave, the function generator will adjust the output amplitude to 3.535 Vrms (the upper limit for sine waves in Vrms).
3
From the front panel, “
AMPL LIMIT” is displayed and the amplitude is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the amplitude is adjusted.
59
Chapter 3 Features and Functions
Output Configuration
Output Amplitude
(continued)
Output Amplitude and Output Termination: The output amplitude is
automatically adjusted (and no error is generated) if you change the output termination. For example, if you set the amplitude to 10 Vpp
and then change the termination from 50 ohms to “high impedance”, the displayed amplitude will double to 20 Vpp. If you change from “high impedance” to 50 ohms, the displayed amplitude will drop in half. See “Output Termination” on page 65 for more information.
Offset Voltage Restrictions: The output amplitude (in Vpp) and the
dc offset voltage must obey the following restrictions. If the specified amplitude is not valid, the function generator will automatically adjust it to the maximum value allowed with the present offset voltage. (Vmax is either 10 volts for a high impedance termination or 5 volts for a 50 ohm termination; Vpp is the output amplitude in volts peak-to-peak.)
V
offset
From the front panel, “
+
V
pp
V
2
MAX VALUE” is displayed and the amplitude
max
and
V
offset
2 x V
pp
is adjusted. From the remote interface, a -221, “Settings conflict”
error is generated and the amplitude is adjusted.
A momentary glitch occurs in the output waveform at certain
voltages due to output attenuator switching. This positive-going glitch occurs when the output voltage crosses the break-point voltage either from a lower voltage or a higher voltage. The voltages are shown below (for a 0 volt dc offset):
252 mVpp, 399 mVpp, 502 mVpp, 796 mVpp, 1 Vpp, 1.59 Vpp,
2.0 Vpp, 3.17 Vpp, 3.99 Vpp, 6.32 Vpp, 7.96 Vpp
The output voltage will momentarily drop to 0 volts at certain voltages
due to output relay switching. This occurs when the output voltage crosses the break-point voltage either from a lower voltage or a higher voltage. The voltages are shown below (for a 0 volt dc offset):
317 mVpp, 632 mVpp, 1.26 Vpp, 2.52 Vpp, 5.02 Vpp
60
Chapter 3 Features and Functions
Output Configuration
You can set the units for output amplitude to Vpp, Vrms, or dBm.
See “Output Units” on page 64 for more information.
For dc volts, the output level is actually controlled by setting the
offset voltage. You can set the dc voltage to any value between into 50 ohms or
on page 62 for more information.
±10 Vdc into an open circuit. See “DC Offset Voltage”
±5 Vdc
To select
for more than 2 seconds
Front-Panel Operation: To set the output amplitude, press
Then use the knob, arrow keys, or “Enter Number” mode to set the amplitude.
Remote Interface Operation:
VOLTage {<
You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command.
dc volts from the front panel, press Offset and hold it down
.
Ampl .
amplitude>|MINimum|MAXimum}
3
61
Chapter 3 Features and Functions
Output Configuration

DC Offset Voltage

At power-on, the dc offset is set to 0 volts. You can set the offset to a positive or negative number with the restrictions shown below. If the specified offset voltage is not valid, the function generator will automatically adjust it to the maximum dc voltage allowed with the present amplitude. (Vmax is either 10 volts for a high impedance termination or 5 volts for a 50 ohm termination; Vpp is the output amplitude in volts peak-to-peak.)
V
V
offset
+
pp
2
V
max
and
V
offset
2 x V
pp
From the front panel, “
MAX VALUE” is displayed and the offset
is adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the offset is adjusted.
DC Offset and Output Termination: The offset voltage is automatically
adjusted (and no error is generated) if you change the output termination. For example, if you set the offset to 100 mVdc and then
change the termination from 50 ohms to “high impedance”, the displayed offset will double to 200 mVdc. If you change from “high impedance” to 50 ohms, the displayed offset will drop in half.
See “Output Termination” on page 65 for more information.
For dc volts, the output level is actually controlled by setting the
offset voltage. You can set the dc voltage to any value between into 50 ohms or
To select
dc volts from the front panel, press Offset and hold it down
for more than 2 seconds
±10 Vdc into an open circuit.
.
±5 Vdc
62
Chapter 3 Features and Functions
Output Configuration
For arbitrary waveforms, the Offset annunciator will turn on if the
waveform data has an inherent offset present (if the average is not equal to zero). The function generator calculates the average of the data points and compares the average to zero volts. If the average is not within two
Offset annunciator turns on.
the
DAC (Digital-to-Analog Converter) counts of zero volts,
Front-Panel Operation: To set the dc offset, press
the knob, arrow keys, or “Enter Number” mode to set the offset.
Remote Interface Operation:
VOLTage:OFFSet {<
You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command.
offset>|MINimum|MAXimum}
Offset . Then use
3
63
Chapter 3 Features and Functions
Output Configuration

Output Units

Applies only to output amplitude (does not affect offset). At power-on, the units for output amplitude are volts peak-to-peak.
Output units:
Vpp, Vrms, or dBm. The default is Vpp.
The unit setting is stored in volatile memory; the units are set to
“Vpp” when power has been off or after a remote interface reset.
When switching from remote programming over the interface back to
front-panel (local) operation, the output units are remembered. For example, if you select “Vrms” from the remote interface, the units are shown as “Vrms” on the front-panel display.
Front-Panel Operation: Use the “Enter Number” units keys
(the arrow keys located on the right side of the front panel) to select the output units. You can also convert from one unit to another from the front panel. For example, to convert 2 Vpp to its equivalent value in Vrms, press the following keys:
Enter Number
kHz m Vrms
The converted value is 707.1 mVrms.
Remote Interface Operation:
VOLTage:UNIT {VPP|VRMS|DBM|DEFault}
64
Chapter 3 Features and Functions
Output Configuration

Output Termination

Applies only to output amplitude and offset voltage. The function generator has a fixed output impedance of 50 ohms on the OUTPUT terminal. You can specify whether a 50 ohm load or an open circuit. Incorrect impedance matching between the function generator and your load will result in an amplitude or offset which does not match the specified signal level.
you are terminating the output into
Output termination:
See the table on page 59 for a list of amplitude limits for all functions.
The output termination setting is stored in volatile memory; 50
selected when power has been off or after a remote interface reset.
The amplitude (or dc offset) is automatically adjusted (and no error is
generated) if you change the output termination. For example, if you set the amplitude to 10 Vpp and then change the termination from
50 ohms to “high impedance”, the displayed amplitude will double to 20 Vpp. If you change from “high impedance” to 50 ohms, the displayed amplitude will drop in half.
If you specify a 50 ohm termination but are actually terminating into
an open circuit, the displayed output will be twice the value specified. For example, if you set the offset to 100 mVdc (and specify a 50 ohm termination) but are actually terminating the output into an open circuit, the actual displayed offset will be 200 mVdc.
Front-Panel Operation: Choose “
front-panel menu.
1: OUT TERM (SYS MENU)
Remote Interface Operation:
OUTPut:LOAD {50|INFinity|MINimum|MAXimum}
50or High impedance. The default is 50Ω.
is
50 OHM” or “HIGH Z” from the
3
65
Chapter 3 Features and Functions
Output Configuration

Duty Cycle

Applies only to square waves. Duty cycle is specified as a percentage and represents the amount of time per cycle that the square wave is high.
T
D
T
T
Duty Cycle =
D
(where T =
T
Duty cycle: 20% to 80%, in 1% increments (frequency 5 MHz).
40% to 60%, in 1% increments (frequency > 5 MHz). The default is 50%.
1
frequency
)
The duty cycle is stored in volatile memory; the duty cycle is set to
50% when power has been off or after a remote interface reset. The APPLy command automatically sets the duty cycle to 50% for square waves.
Before attempting to set the duty cycle, you must enable the square
wave function. From the front panel, “
SQUARE ONLY” is displayed if
square wave is not currently active. From the remote interface, no error is generated, but the specified duty cycle is remembered when you change to the square wave function.
The duty cycle setting is remembered when you change from square
wave to another function. When you return to the square wave function, the previous duty cycle is used.
66
Chapter 3 Features and Functions
Output Configuration
Possible Conflict with Output Frequency: The duty cycle is auto-
matically adjusted if you select a frequency that is not valid with the present duty cycle. For example, if you set the duty cycle to 70% and then change the frequency to 8 MHz, the function generator will automatically adjust the duty cycle to 60% (the upper limit for this frequency).
From the front panel, “ adjusted. From the remote interface, a -221, “Settings conflict” error
is generated and the duty cycle is adjusted.
Front-Panel Operation: To set the duty cycle, press
(the value appears on the display for approximately 10 seconds).
Then use the knob, arrow keys, or “Enter Number” mode to set the duty cycle.
Remote Interface Operation:
PULSe:DCYCle {<
The APPLy command automatically sets the duty cycle to 50% for square waves.
% DUTY LIMIT” is displayed and the duty cycle is
Shift % Duty
percent>|MINimum|MAXimum}
3
67
Chapter 3 Features and Functions
Output Configuration

SYNC Signal

A sync signal output is provided on the front-panel SYNC terminal. All of the standard output functions (except dc and noise) have an associated sync signal. For certain applications where you may not want to output the sync signal, you can disable the SYNC terminal.
The SYNC terminal can be enabled/disabled from the remote interface only.
By default, the sync signal is routed to the SYNC terminal (enabled).
When the sync signal is disabled, the output level on the SYNC
terminal is indeterminate (it might be a
For sine, square, triangle, and ramp waveforms, the sync signal is a
TTL “high” when the waveform’s output is positive, relative to zero
volts (or the dc offset value). The signal is a output is negative, relative to zero volts (or the dc offset value).
TTL “high” or a TTL “low”).
TTL “low” when the
For arbitrary waveforms, a momentary
TTL “high” pulse (> 200 ns) is
output which corresponds to the first downloaded point in the waveform.
For AM and FM, the sync signal is referenced to the modulating
signal (not the carrier). A momentary
TTL “high” pulse (> 200 ns) is
output at each zero-crossing point of the modulating signal.
For the triggered burst mode, a
TTL “low” signal is output while the
specified number of cycles is output (for the duration of the burst). After the specified number of cycles has been output, the sync signal goes “high” until the next burst.
For the external gated burst mode, the sync signal is a
TTL “high”
when the output is positive, relative to zero volts (or the dc offset value). The signal is a
TTL “low” when the output is negative, relative
to zero volts (or the dc offset value).
For FSK, a momentary
TTL “high” pulse (> 200 ns) is output on the
transition to the “hop” frequency.
For frequency sweeps, the sync signal is a
sweep (when the start frequency is output) and is a
TTL “low” at the start of the
TTL “high” at the
end of the sweep (when the stop frequency is output).
Remote Interface Operation:
OUTPut:SYNC {OFF|ON}
Setting is stored in volatile memory.
68
Chapter 3 Features and Functions
Output Configuration

Instrument State Storage

You can store up to three different instrument states in non-volatile memory. This enables you to recall the entire instrument configuration with a single command from the remote interface or with just a few key presses from the front panel.
Three memory locations (numbered 1, 2, and 3) are available to store
instrument configurations. The state storage feature “remembers” the function (including arbitrary waveforms), frequency, amplitude, dc offset, duty cycle, as well as any modulation parameters. To recall a stored state, you must use the same memory location used previously to store the state.
You cannot recall the instrument state from a memory location that
was not previously specified as a storage location. For example, an error is generated if you attempt to recall from memory location “2” but have never stored to that location.
3
From the front panel, “ in the specified memory location. From the remote interface, a
+810, “State has not been stored” error is generated if nothing is stored in the specified memory location.
Any arbitrary waveforms downloaded to “
remembered. However, if an arbitrary waveform is being output from non-volatile memory when the state is stored, the waveform data is stored. The stored waveform is output when the instrument state is recalled.
If you delete an arbitrary waveform after storing the state, the
waveform data is lost and the function generator will not output the waveform when the state is recalled. The “ in place of the deleted waveform.
NOT STORED” is displayed if nothing is stored
VOLATILE” memory are not
SINC” waveform is output
69
Chapter 3 Features and Functions
Output Configuration
State Storage
(continued)
When power is turned off, the function generator automatically stores
its state in memory location “0”. You can configure the function generator to automatically recall the power-down state when power is restored. The recall mode is disabled when the function generator is shipped from the factory.
Select the enable the power-down recall mode. Select
POWER ON LAST STATE command from the SYS MENU to
POWER ON DEFAULT to
disable the power-down recall mode. See “Power-Down Recall Mode” on page 109 for more information.
From the remote interface, you can use memory location “0” to store
a fourth instrument configuration (you cannot store to this memory location from the front panel). However, keep in mind that memory location “0” is automatically overwritten when the power is turned off.
Front-Panel Operation:
To store the state, press
Shift Store . Then use the knob, arrow
keys, or “Enter Number” mode to select the memory location.
To restore the state, press
Recall . Then use the knob, arrow keys,
or “Enter Number” mode to select the memory location.
Remote Interface Operation:
*SAV {0|1|2|3}
State 0 is the instrument state at power down.
*RCL {0|1|2|3} States 1, 2, and 3 are user-defined states.
From the remote interface, you can delete individual stored states
and clear the memory location. If nothing is stored in the specified memory location, a +810, “State has not been stored” error is generated.
MEMory:STATe:DELete {0|1|2|3}
70
Chapter 3 Features and Functions

Amplitude Modulation (AM)

Amplitude Modulation (AM)
A modulated waveform consists of a carrier waveform and a modulating waveform. In of the modulating waveform. The function generator will accept an internal modulating signal, an external modulating signal, or both.
For more information on the fundamentals of Amplitude Modulation,
refer to chapter 7, “Tutorial”.
AM, the amplitude of the carrier is varied by the amplitude

To Select AM Modulation

The AM annunciator turns on when AM is enabled.
Only one modulation mode can be enabled at a time. When you
enable
Front-Panel Operation: Enable
modulation parameters. Press using the present settings for the carrier frequency, modulating frequency, output amplitude, and offset voltage.
Remote Interface Operation: To ensure proper operation, you should
enable
AM:STATe {OFF|ON}
AM, the previous modulation mode is turned off.
AM before setting up the other
Shift AM to output an AM waveform
AM after you have set up the other modulation parameters.
3
71
Chapter 3 Features and Functions
Amplitude Modulation (AM)

Carrier Waveform Shape

AM carrier shape: Sine, Square, Triangle, Ramp, or Arbitrary
waveform. The default is Sine.
You cannot use the noise function or dc volts as the carrier waveform.
Front-Panel Operation: Press any of the front-panel function keys
except
Noise . For arbitrary waveforms, the Arb key selects the
waveform that is currently downloaded.
Remote Interface Operation:
FUNCtion:SHAPe {SINusoid|SQUare|TRIangle|RAMP|USER}
You can also use the APPLy command to select the function, frequency, amplitude, and offset with a single command.

Carrier Frequency

Carrier frequency: 100 µHz to 15 MHz (100 kHz for triangle and ramp).
The default is 1 kHz.
For arbitrary waveforms, the maximum carrier frequency depends on
the number of points specified in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
Minimum Frequency
100 100 100
Front-Panel Operation: Press the
µHz µHz µHz
Freq key to select any valid
Maximum Frequency
5 MHz
2.5 MHz 200 kHz
frequency for the selected function (the present amplitude and offset voltage are used).
Remote Interface Operation:
FREQuency {<
frequency>|MINimum|MAXimum}
72
Chapter 3 Features and Functions
Amplitude Modulation (AM)

Modulating Waveform Shape

The function generator will accept an internal modulating signal, an external modulating signal, or both.
Modulating waveform shape (internal source):
Triangle, Ramp, Noise, or Arbitrary waveform. The default is Sine.
You can use the noise function as the modulating waveform.
However, you cannot use the noise function or dc volts as the
carrier waveform.
Front-Panel Operation: After enabling
directly to the
1: AM SHAPE (MOD MENU)
Remote Interface Operation:
AM:INTernal :FUNCtion {SIN usoid|SQUare |TRIangle| RAMP|NOISe|U SER}
AM SHAPE command in the menu.
AM, press Recall Menu to go
Sine, Square,
3
73
Chapter 3 Features and Functions
Amplitude Modulation (AM)

Modulating Waveform Frequency

The function generator will accept an internal modulating signal, an external modulating signal, or both.
Modulating frequency (internal source): 10 mHz to 20 kHz.
The default is 100 Hz.
The sync signal for
carrier). A momentary
AM is referenced to the modulating signal (not the
TTL “high” pulse (> 200 ns) is output at each
zero-crossing point of the modulating signal. The signal is output from the front-panel SYNC terminal.
Front-Panel Operation: Press
Shift Freq to set the modulating
frequency (the value is displayed for approximately 10 seconds).
Before setting the modulating frequency from the front panel, you must have
AM is not enabled, “SELECT AM, FM” is displayed.
AM enabled. If you attempt to set the frequency when
Remote Interface Operation:
AM:INTernal:FREQuency {<
frequency>|MINimum|MAXimum}

Modulation Depth

The modulation depth is expressed as a percentage and represents the extent of the amplitude variation. At 0% modulation, the output amplitude is half of the selected value. At 100% modulation, the output amplitude equals the selected value.
Modulation depth: 0% to 120%. The default is 100%.
Front-Panel Operation: Press
(the value is displayed for approximately 10 seconds).
Shift Level to set the depth in percent
Before setting the modulation depth from the front panel, you must
AM enabled. If you attempt to set the depth when AM is not
have enabled, “
SELECT AM, FM” is displayed.
Remote Interface Operation:
AM:DEPTh {<
depth in percent>|MINimum|MAXimum}
74
Chapter 3 Features and Functions
Amplitude Modulation (AM)

Modulating Source

The function generator will accept an internal modulating signal, an external modulating signal, or both.
Modulating source: Internal-External (both) or External only.
The default is Both (internal-external).
The External modulating source is always enabled.
When both sources are enabled (internal-external), the function
generator adds the internal and external modulating signals
(the carrier waveform is actually modulated with two waveforms).
When the internal source is disabled (external only), the carrier
waveform is modulated with the external waveform. The
annunciator turns on to indicate that the function generator is
expecting a modulating signal on the rear-panel AM Modulation
terminal.
You apply the external modulating waveform to the AM Modulation
terminal. The modulation depth is controlled by the signal level
present (5 volts
peak corresponds to 100% modulation).
Ext
3
Front-Panel Operation:
2: AM SOURCE (MOD MENU)
Remote Interface Operation:
AM:SOURce {BOTH|EXTernal}
5 V
0 V
-5 V
75
Chapter 3 Features and Functions

Frequency Modulation (FM)

Frequency Modulation (FM)
A modulated waveform consists of a carrier waveform and a modulating waveform. In of the modulating waveform. The function generator will accept only an internal
For more information on the fundamentals of Frequency Modulation,
refer to chapter 7, “Tutorial”.

To Select FM Modulation

The FM annunciator turns on when FM is enabled.
Only one modulation mode can be enabled at a time. When you
enable
FM, the frequency of the carrier is varied by the amplitude
FM modulating signal (no external source is available).
FM, the previous modulation mode is turned off.
Front-Panel Operation: Enable
modulation parameters. Press
FM before setting up the other
Shift FM to output an FM waveform
using the present settings for the carrier frequency, modulating frequency, output amplitude, and offset voltage.
Remote Interface Operation: To ensure proper operation, you should
enable
FM after you have set up the other modulation parameters.
FM:STATe {OFF|ON}
76
Chapter 3 Features and Functions
Frequency Modulation (FM)

Carrier Waveform Shape

FM carrier shape: Sine, Square, Triangle, Ramp, or Arbitrary
waveform. The default is Sine.
You cannot use the noise function or dc volts as the carrier waveform.
Front-Panel Operation: Press any of the front-panel function keys
except waveform that is currently downloaded.
Remote Interface Operation:
FUNCtion:SHAPe {SINusoid|SQUare|TRIangle|RAMP|USER}
You can also use the APPLy command to select the function,
frequency, amplitude, and offset with a single command.
Noise . For arbitrary waveforms, the Arb key selects the
3
77
Chapter 3 Features and Functions
Frequency Modulation (FM)

Carrier Frequency

Carrier frequency: 10 mHz to 15 MHz (100 kHz for triangle
and ramp). The default is 1 kHz.
For arbitrary waveforms, the maximum carrier frequency depends on
the number of points specified in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
Minimum Frequency
Maximum Frequency
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
10 mHz 10 mHz 10 mHz
5 MHz
2.5 MHz 200 kHz
The carrier frequency must always be greater than or equal to the
peak frequency deviation. If you attempt to set the carrier frequency to a value less than the deviation, the function generator will auto­matically adjust the carrier frequency to equal the present deviation.
From the front panel, “
MIN VALUE” is displayed and the carrier
frequency is adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the carrier frequency is adjusted.
The sum of the carrier frequency and peak frequency deviation must
be less than or equal to the maximum frequency for the selected function
plus 100 kHz (15.1 MHz for sine and square, 200 kHz for
triangle and ramp, and 5.1 MHz for arbitrary waveforms). If you attempt to set the carrier frequency to a value that is not valid, the function generator will automatically adjust the carrier frequency to equal the present deviation.
From the front panel, “
MAX VALUE” is displayed and the deviation is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the deviation is adjusted.
Front-Panel Operation: Press the
Freq key to select any valid
frequency for the selected function (the present amplitude and offset voltage are used).
Remote Interface Operation:
FREQuency {<
frequency>|MINimum|MAXimum}
78
Chapter 3 Features and Functions
Frequency Modulation (FM)

Modulating Waveform Shape

The function generator will accept only an internal modulating signal.
Modulating waveform shape (internal source):
Sine, Square,
Triangle, Ramp, Noise, or Arbitrary waveform. The default is Sine.
You can use the noise function as the modulating waveform.
However, you cannot use the noise function or dc volts as the carrier
waveform.
Front-Panel Operation: After enabling
directly to the
FM SHAPE command in the menu.
FM, press Recall Menu to go
3: FM SHAPE (MOD MENU)
Remote Interface Operation:
FM:INTernal :FUNCtion {SIN usoid|SQUare |TRIangle| RAMP|NOISe|U SER}

Modulating Waveform Frequency

The function generator will accept only an internal modulating signal.
Modulating frequency (internal source): 10 mHz to 10 kHz.
The default is 10 Hz.
The sync signal for
carrier). A momentary
zero-crossing point of the modulating signal. The signal is output
from the front-panel SYNC terminal.
FM is referenced to the modulating signal (not the
TTL “high” pulse (> 200 ns) is output at each
3
Front-Panel Operation: Press
Shift Freq to set the modulating
frequency (the value is displayed for approximately 10 seconds).
Before setting the modulating frequency from the front panel, you must have
FM is not enabled, “SELECT AM, FM” is displayed.
FM enabled. If you attempt to set the frequency when
Remote Interface Operation:
FM:INTernal:FREQuency {<
frequency>|MINimum|MAXimum}
79
Chapter 3 Features and Functions
Frequency Modulation (FM)

Peak Frequency Deviation

The peak frequency deviation represents the variation in frequency of the modulating waveform from the carrier frequency.
Peak frequency deviation: 10 mHz to 7.5 MHz. The default is 100 Hz.
The carrier frequency must always be greater than or equal to the
peak frequency deviation. If you attempt to set the deviation to a value greater than the carrier frequency (with function generator will automatically adjust the deviation to equal the present carrier frequency.
FM enabled), the
From the front panel, “
MAX VALUE” is displayed and the deviation is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the deviation is adjusted.
The sum of the carrier frequency and peak frequency deviation must
be less than or equal to the maximum frequency for the selected function
plus 100 kHz (15.1 MHz for sine and square, 200 kHz for
triangle and ramp, and 5.1 MHz for arbitrary waveforms). If you attempt to set the deviation to a value that is not valid, the function generator will automatically adjust the deviation to the maximum value allowed with the present carrier frequency.
From the front panel, “
MAX VALUE” is displayed and the deviation is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the deviation is adjusted.
Front-Panel Operation: Press
Shift Level to set the peak deviation
(the value is displayed for approximately 10 seconds).
Before setting the deviation from the front panel, you must have
FM enabled. If you attempt to set the deviation when FM is not
enabled, “
SELECT AM, FM” is displayed.
Remote Interface Operation:
FM:DEViation {<
peak deviation in Hz>|MINimum|MAXimum}
80
Chapter 3 Features and Functions

Burst Modulation

Burst Modulation
You can configure the function generator to output a waveform with a specified number of cycles, called a burst. You can output the burst at a rate determined by the internal rate generator or an external signal applied to the rear-panel connector. The function generator can produce
a burst using sine, square, triangle, ramp, and arbitrary waveforms.
For more information on the fundamentals of Burst Modulation,
refer to chapter 7, “Tutorial”.
You can use burst modulation in the following two modes. The function generator enables one burst mode at a time based on the trigger source and burst source that you select (see the table below).
Triggered Burst Mode: In this mode, the function generator outputs
a waveform with a specified number of cycles (burst count) each time
a trigger is received. After the specified number of cycles has been
output, the function generator waits for the next trigger. You can
configure the function generator to accept an internal trigger or you
can send the trigger externally by pressing the front-panel
Single key
or by applying a trigger signal to the rear-panel Ext Trig terminal.
At power-on, the internally-triggered burst mode is enabled.
External Gated Burst Mode: In this mode, the output waveform is
either “on” or “off ” based on the level of the external signal applied to
the rear-panel Ext Trig terminal. When the gate signal is true,
the function generator outputs a continuous waveform. When the
gate signal is false, the output is turned off.
Trigger Source
(TRIG:SOUR)
External Trigger e d Mode External Internal Available Not Used Available
Burst Source
(BM:SOUR)
Burst Count
(BM:NCYC)
Burst Rate
(BM:INT:RATE)
Burst Phase
(BM:PHAS)
3
Internal Triggered Mode Internal Internal Available Available Available
External Gated Mode Internal External Not Used Not Used Not Used
81
Chapter 3 Features and Functions
Burst Modulation

Burst Trigger Source

In the triggered burst mode, the function generator outputs a waveform with the specified number of cycles (burst count) each time a trigger is received. After the specified number of cycles has been output, the function generator waits for the next trigger while outputting no signal (zero volts or the dc offset level). At power-on, the internally-triggered
burst mode is enabled.
Trigger source:
Internal, Single, or External. The default is Internal.
In the triggered burst mode, the “burst source” is Internal.
When the internal trigger source is selected, the frequency at which
the burst is generated is determined by the burst rate parameter. The burst rate frequency defines the interval between bursts.
Pressing
Single to enable the single trigger mode also enables the
external trigger mode. External triggering is like the front-panel
single trigger mode except that you apply a trigger signal to the rear-panel Ext Trig terminal. The function generator is triggered on the rising edge of a
The
Trig annunciator turns on when the function generator is waiting
TTL pulse.
for a single or external trigger.
When the single or external trigger source is selected, the burst count
and burst phase remain in effect but the burst rate is ignored.
When the external gate source is selected (“gated” burst mode), the
specified trigger source is ignored. The external gated mode overrides the triggered mode source.
Front-Panel Operation: Pressing
Single to enable the single
trigger mode also enables the external trigger mode.
Shift Internal to enable the internal trigger mode.
Press
Remote Interface Operation:
TRIGger:SOURce {IMMediate|EXTernal|BUS}
See “Triggering” on page 98 for more information.
82
Chapter 3 Features and Functions
Burst Modulation

Burst Source

In the external gated burst mode, the output waveform is either
“on” or “off ” based on the level of the external signal applied to the rear-panel Ext Trig terminal. When the gate signal is true, the function generator outputs a continuous waveform. When the gate signal is false, the output is turned off.
Burst source:
When the internal burst source is selected, the external gated mode
is disabled.
When the external gate source is selected, the output is enabled or
disabled based on the logic level of the gate signal applied to the
Ext Trig terminal. When the gate signal is true (
the function generator outputs a continuous waveform. When the
gate signal is false (
the dc offset level).
When the external gate source is selected, the burst count, burst rate,
burst phase, and burst trigger source are ignored (these parameters
are used for the triggered burst mode only).
The
Front-Panel Operation:
Remote Interface Operation:
Ext annunciator turns on when the function generator is waiting
for the external gate signal.
7: BURST SRC (MOD MENU)
BM:SOURce {INTernal|EXTernal}
Internal or External Gate. The default is Internal.
TTL high),
TTL low), the output is turned off (zero volts or
3
83
Chapter 3 Features and Functions
Burst Modulation

To Select Burst Modulation

The Burst annunciator turns on when burst modulation is enabled.
Only one modulation mode can be enabled at a time. When you
enable the burst mode, the previous modulation mode is turned off.
Front-Panel Operation: Enable burst modulation before you set up
the other modulation parameters. Press
Shift Burst to output the
burst using the present settings for the carrier frequency, output amplitude, and offset voltage.
Remote Interface Operation: To ensure proper operation, you should
enable the burst mode after you have set up the other modulation parameters.
BM:STATe {OFF|ON}

Burst Carrier Frequency

The carrier frequency defines the repetition rate of the burst waveform in the triggered and external gated modes. In the triggered mode, the number of cycles specified by the burst count are output at the frequency of the carrier signal. In the external gated mode, the carrier frequency is output when the external gate signal is true (
TTL high).
Keep in mind that the carrier frequency is different than the “burst rate” which specifies the interval between bursts (triggered mode only).
Carrier frequency: 10 mHz to 5 MHz (100 kHz for triangle and ramp).
The default is 1 kHz. You can use sine, square, ramp, triangle, or arbitrary waveforms for the carrier waveshape.
Be sure to note the restrictions for carrier frequency and
burst count shown on the following pages.
84
Chapter 3 Features and Functions
Burst Modulation
For arbitrary waveforms used as the carrier waveform, the maximum
frequency depends on the number of points specified in the waveform.
The five built-in arbitrary waveforms can be output at a maximum
of 5 MHz (be sure to note the restrictions below).
Number of Arb Points
Minimum Frequency
Maximum Frequency
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
10 mHz 10 mHz 10 mHz
5 MHz
2.5 MHz 200 kHz
For sine, square, and arbitrary waveforms (does not apply to ramp
and triangle waveforms), the relationship between the carrier
frequency and the minimum burst count is shown below.
Carrier Frequency
10 mHz to 1 MHz >1 MHz to 2 MHz >2 MHz to 3 MHz >3 MHz to 4 MHz >4 MHz to 5 MHz
Minimum Burst Count
1 2 3 4 5
For sine, square, and arbitrary waveforms only.
If you attempt to set the carrier frequency to a value that is not valid,
the function generator will automatically adjust the frequency to the
maximum value allowed with the present burst count.
From the front panel, “
MAX VALUE” is displayed and the carrier
frequency is adjusted. From the remote interface, a -221, “Settings
conflict” error is generated and the carrier frequency is adjusted.
3
85
Chapter 3 Features and Functions
Burst Modulation
Carrier Frequency
(continued)
For all waveforms used with burst, if the carrier frequency is set
less than or equal to 100 Hz, the following relationship applies.
Burst Count
Carrier Frequency
500 seconds For Carrier ≤ 100 Hz
If you attempt to set the carrier frequency to a value that is not valid, the function generator will automatically adjust the frequency to the minimum value allowed with the present burst count.
From the front panel, “
MIN VALUE” is displayed and the frequency is
adjusted. From the remote interface, a -221, “Settings conflict” error is generated and the frequency is adjusted.
For the triggered burst mode, the sync signal is a
TTL “low” while the
specified number of cycles is output (for the duration of the burst). After the specified number of cycles has been output, the sync signal goes “high” until the next burst. The sync signal is output from the front-panel SYNC terminal.
For the external gated burst mode, the sync signal is a
TTL “high”
when the output is positive, relative to zero volts (or the dc offset value). The signal is a
TTL “low” when the output is negative, relative to
zero volts (or the dc offset value). The sync signal is output from the front-panel SYNC terminal.
Front-Panel Operation: Press the
Freq key and select any valid
frequency for the selected function (the present amplitude and offset voltage are used).
Remote Interface Operation:
FREQuency {<
frequency>|MINimum|MAXimum}
You can also use the APPLy command to select the function, frequency, amplitude, and offset of the carrier with a single command.
86
Chapter 3 Features and Functions
Burst Modulation

Burst Count

The burst count defines the number of cycles to be output per burst. Used only in the triggered burst mode (internal or external source).
Certain combinations of burst count and carrier frequency are not allowed. If you attempt to specify a burst count that is not valid, the function generator will automatically adjust the count to the maximum value allowed with the present carrier frequency.
Make sure you note the restrictions in “Burst Carrier Frequency” starting on page 85 before setting the burst count.
Burst count: 1 to 50,000 cycles, in 1 cycle increments. You can also
select an infinite burst count. The default is 1 cycle.
When the internal trigger source is selected, the specified number of
cycles is output continuously at a rate determined by the burst rate
setting. The burst rate frequency defines the interval between bursts.
When the single or external trigger source is selected, the burst count
and burst phase remain in effect but the burst rate is ignored. The
specified number of cycles is output each time when a trigger signal is applied to the Ext Trig terminal. The function generator is triggered on the rising edge of a
When the external gate burst source is selected, the burst count,
burst rate, burst phase, and burst trigger source are ignored
(these parameters are used for the triggered burst mode only).
Front-Panel Operation:
4: BURST CNT (MOD MENU)
Single is pressed or
TTL pulse.
3
Remote Interface Operation:
BM:NCYCles {<
# cycles>|INFinity|MINimum|MAXimum}
87
Chapter 3 Features and Functions
Burst Modulation

Burst Rate

The burst rate defines the frequency at which internally triggered bursts are generated. The burst rate frequency defines the interval between bursts. Used only in the internal triggered burst mode.
Keep in mind that the burst rate is different than the “carrier frequency” which specifies the frequency of the bursted signal.
Burst rate: 10 mHz to 50 kHz. The default is 100 Hz.
The burst rate setting is used only when internal triggering is
enabled (triggered mode). The burst rate is ignored when single triggering or external triggering is enabled.
When the single or external trigger source is selected (triggered mode),
the burst count and burst phase remain in effect but the burst rate is ignored.
When the external gate burst source is selected (“gated” burst mode),
the burst count, burst rate, burst phase, and burst trigger source are ignored (these parameters are used for the triggered burst mode only).
It is possible to specify a burst rate which is too fast for the function
generator to output with the specified carrier frequency and burst count. If the burst rate is too high, the function generator will internally adjust it as needed to continuously re-trigger the burst. The adjustment is handled internally by the function generator (the burst rate displayed or queried will be the same as specified).
Front-Panel Operation:
5: BURST RATE (MOD MENU)
Remote Interface Operation:
BM:INTernal:RATE {<
frequency>|MINimum|MAXimum}
88
Chapter 3 Features and Functions
Burst Modulation

Burst Phase

The burst phase defines the starting phase of the burst.
Burst phase: -360 degrees to +360 degrees, in 0.001 degree increments.
The default is 0 degrees.
• For sine, square, triangle, and ramp waveforms, 0 degrees is the point
at which the waveform crosses zero volts (or the dc offset value), in a
positive-going direction.
For arbitrary waveforms, 0 degrees is the first data point downloaded
to memory.
When the external gate source is selected (“gated” burst mode),
the burst count, burst rate, burst phase, and burst trigger source are
ignored (these parameters are used for the triggered burst mode only).
Front-Panel Operation:
6: BURST PHAS (MOD MENU)
3
Remote Interface Operation:
BM:PHASe {<
degrees>|MINimum|MAXimum}
89
Chapter 3 Features and Functions

Frequency-Shift Keying (FSK) Modulation

Frequency-Shift Keying (FSK) Modulation
You can configure the function generator to “shift” its output frequency between two preset values using output shifts between the two frequencies (called the “carrier frequency” and the “hop frequency”) is determined by the internal rate generator or the signal level on the rear-panel FSK terminal. The function generator
can produce an FSK waveform using sine, square, triangle, ramp, and arbitrary waveforms.
For more information on the fundamentals of FSK Modulation,
refer to chapter 7, “Tutorial”.

To Select FSK Modulation

The FSK annunciator turns on when FSK is enabled.
Only one modulation mode can be enabled at a time. When you
enable
FSK, the previous modulation mode is turned off.
FSK modulation. The rate at which the
Front-Panel Operation: Enable
modulation parameters. Press
FSK before you set up the other
Shift FSK to output the waveform
using the present settings for frequency, amplitude, and offset voltage.
Remote Interface Operation: To ensure proper operation, you should
enable
FSK after you have set up the other modulation parameters.
FSKey:STATe {OFF|ON}
90
Chapter 3 Features and Functions
Frequency-Shift Keying (FSK) Modulation

FSK Carrier Frequency

Carrier frequency: 10 mHz to 15 MHz (100 kHz for triangle
and ramp). The default is 1 kHz. You can use sine, square, ramp,
triangle, or arbitrary waveforms for the carrier waveform.
For arbitrary waveforms, the maximum carrier frequency depends on
the number of points specified in the waveform. The five built-in
arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
Minimum Frequency
10 mHz 10 mHz 10 mHz
Maximum Frequency
5 MHz
2.5 MHz 200 kHz
3
For an externally-controlled
output when a “low”
TTL level is applied to the FSK terminal. The hop
frequency is output when a “high”
For
FSK, the sync signal is referenced to the FSK “hop” signal (not the
carrier). A momentary
FSK waveform, the carrier frequency is
TTL level is applied.
TTL “high” pulse (> 200 ns) is output on the
transition to the “hop” frequency. The signal is output from the front-
panel SYNC terminal.
Front-Panel Operation: Press the
Freq key and select any valid
frequency for the selected function (the present amplitude and offset voltage are used).
Remote Interface Operation:
FREQuency {<
frequency>|MINimum|MAXimum}
You can also use the APPLy command to select the function, frequency,
amplitude, and offset of the carrier with a single command.
91
Chapter 3 Features and Functions
Frequency-Shift Keying (FSK) Modulation

FSK “Hop” Frequency

Hop frequency: 10 mHz to 15 MHz (100 kHz for triangle and ramp).
The default is 100 Hz. You can use sine, square, ramp, triangle, or arbitrary waveforms for the hop frequency waveshape.
For arbitrary waveforms, the maximum hop frequency depends on
the number of points specified in the waveform. The five built-in arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
Minimum Frequency
Maximum Frequency
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
For an externally-controlled
output when a “low” frequency is output when a “high”
For
FSK, the sync signal is referenced to the FSK “hop” signal (not the
carrier). A momentary
10 mHz 10 mHz 10 mHz
FSK waveform, the carrier frequency is
TTL level is applied to the FSK terminal. The hop
TTL level is applied.
TTL “high” pulse (> 200 ns) is output on the
5 MHz
2.5 MHz 200 kHz
transition to the “hop” frequency. The signal is output from the front­panel SYNC terminal.
Front-Panel Operation: After enabling
directly to the
FSK FREQ command in the menu.
FSK, press Recall Menu to go
8: FSK FREQ (MOD MENU)
Remote Interface Operation:
FSKey:FREQuency {<
frequency>|MINimum|MAXimum}
92
Chapter 3 Features and Functions
Frequency-Shift Keying (FSK) Modulation

FSK Rate

The FSK rate is the rate at which the output frequency “shifts” between the carrier frequency and the hop frequency when you select the internal
FSK rate (internal source): 10 mHz to 50 kHz. The default is 10 Hz.
FSK source.
The
Front-Panel Operation:
Remote Interface Operation:
FSK rate is ignored when the external source is selected.
9: FSK RATE (MOD MENU)
FSKey:INTernal:RATE {<
rate in Hz>|MINimum|MAXimum}

FSK Source

FSK source: Internal or External. The default is Internal.
When the internal source is selected, the rate at which the output
frequency “shifts” between the carrier frequency and hop frequency
is determined by the FSK rate specified.
When the external source is selected, the output frequency is
determined by the signal level on the rear-panel FSK terminal.
When a “low”
When a “high”
The maximum external
The
Ext annunciator turns on when the function generator is waiting
for an external trigger signal.
TTL level is present, the carrier frequency is output.
TTL level is present, the hop frequency is output.
FSK rate is 1 MHz.
3
Front-Panel Operation:
10: FSK SRC (MOD MENU)
Remote Interface Operation:
FSKey:SOURce {INTernal|EXTernal}
93
Chapter 3 Features and Functions

Frequency Sweep

Frequency Sweep
In the frequency sweep mode, the function generator “steps” from the start frequency to the stop frequency at a sweep rate which you specify. You can sweep up or down in frequency, and with either linear or logarithmic spacing. You can also configure the function generator to output a single sweep (one pass from start frequency to stop frequency) by applying an external trigger. The function generator can produce a
frequency sweep for sine, square, triangle, ramp, or arbitrary waveforms.
For more information on the fundamentals of frequency sweeps, refer to
chapter 7, “Tutorial”.

To Select Sweep Modulation

The Swp annunciator turns on when the sweep mode is enabled.
Only one modulation mode can be enabled at a time. When you
enable the sweep mode, the previous modulation mode is turned off.
Front-Panel Operation: Enable the sweep mode before you set up the
other sweep parameters. Press
the present settings for amplitude and offset voltage.
Remote Interface Operation: To ensure proper operation, you should
enable the sweep mode after you have set up the other modulation parameters.
SWEep:STATe {OFF|ON}
Shift Sweep to output a sweep using
94
Chapter 3 Features and Functions
Frequency Sweep

Start Frequency and Stop Frequency

The start frequency and stop frequency set the upper and lower frequency bounds for the sweep. The function generator begins at the start frequency, sweeps to the stop frequency, and then resets back to the start frequency.
Start and Stop frequencies: 10 mHz to 15 MHz (100 kHz for triangle
and ramp). The sweep is phase continuous over the full frequency
range. The default start frequency is 100 Hz. The default stop
frequency is 1 kHz. The *RST command sets the start frequency to
10 mHz (minimum) and the stop frequency to 15 MHz (maximum).
For arbitrary waveforms, the maximum start or stop frequency
depends on the number of points specified in the waveform. The five
built-in arbitrary waveforms can be output at a maximum of 5 MHz.
Number of Arb Points
Minimum Frequency
Maximum Frequency
3
8 to 8,192 (8k) 8,193 to 12,287 (12k) 12,288 to 16,000
To sweep
To sweep
up in frequency, set the start frequency < stop frequency. down in frequency, set the start frequency > stop frequency.
The sync signal is a
start frequency is output) and is a
10 mHz 10 mHz 10 mHz
TTL “low” at the start of the sweep (when the
TTL “high” at the end of the sweep
5 MHz
2.5 MHz 200 kHz
(when the stop frequency is output). The signal is output from the
front-panel SYNC terminal.
Front-Panel Operation: After enabling sweeps, press Recall Menu to
go directly to the
START F command in the menu.
1: START F (SWP MENU)
2: STOP F (SWP MENU)
Remote Interface Operation:
FREQuency:STARt {<
FREQuency:STOP {<
frequency>|MINimum|MAXimum}
frequency>|MINimum|MAXimum}
95
Chapter 3 Features and Functions
Frequency Sweep

Sweep Time

The sweep time specifies the number of seconds required to sweep from the start frequency to the stop frequency. The number of frequency points in the sweep depends on the sweep time you select and is automatically calculated by the function generator.
Sweep time: 1 ms to 500 seconds. The default is 1 second.
The function generator computes between 2,048 and 4,096 frequency
points from the start frequency to the stop frequency.
Front-Panel Operation:
3: SWP TIME (SWP MENU)
Remote Interface Operation:
SWEep:TIME {<
seconds>|MINimum|MAXimum}

Sweep Mode

You can sweep with either linear or logarithmic spacing. The output frequency of a linear sweep changes linearly during the duration of the sweep. For logarithmic spacing, the output frequency changes exponentially during the duration of the sweep.
Sweep mode:
Front-Panel Operation:
4: SWP MODE (SWP MENU)
Remote Interface Operation:
SWEep:SPACing {LINear|LOGarithmic}
Linear or Logarithmic. The default is Linear.
96
Chapter 3 Features and Functions
Frequency Sweep

Sweep Trigger Source

In the triggered sweep mode, the function generator outputs a single sweep each time a trigger is received. After one sweep from the start frequency to the stop frequency, the function generator waits for the next trigger while outputting the start frequency.
Sweep trigger source:
is Internal.
When the internal trigger source is enabled (default), the function
generator outputs a continuous sweep at a rate determined by the
sweep rate specified.
Pressing
external trigger mode. External triggering is like the front-panel single trigger mode except that you apply a trigger signal to the rear-panel Ext Trig terminal. The function generator is triggered on the rising edge of a
The
for a single or external trigger.
Front-Panel Operation: Pressing
trigger mode also enables the external trigger mode. Press
Remote Interface Operation:
TRIGger:SOURce {IMMediate|EXTernal|BUS}
See “Triggering” on page 98 for more information.
Single to enable the single trigger mode also enables the
Trig annunciator turns on when the function generator is waiting
Shift Internal to enable the internal trigger mode.
Internal, Single, or External. The default
TTL pulse.
Single to enable the single
3
97
Chapter 3 Features and Functions

Triggering

Triggering
Applies only to bursts and frequency sweep. You can issue triggers for bursts and sweeps using internal triggering, single triggering, or external triggering.
Internal or “automatic” triggering is enabled when you turn on the
function generator. In this mode, the function generator outputs continuously when the burst mode or sweep mode is selected.
Single triggering outputs one burst or initiates one sweep each time
you press
re-trigger the function generator.
Single from the front-panel. Continue pressing this key to
Pressing
Single to enable the single trigger mode also enables the
external trigger mode. External triggering is like the front-panel
single trigger mode except that you apply a trigger signal to the rear-panel Ext Trig terminal. The function generator is triggered on the rising edge of a
The
Single key is disabled when in remote (the Rmt annunciator
TTL pulse.
turns on when in remote) and when a function other than burst or sweep is selected.
98
Chapter 3 Features and Functions
Triggering

Trigger Source Choices

Applies only to burst and sweep. You must specify the source from which the function generator will accept a trigger.
From the front panel, the function generator will accept a single
trigger, a hardware trigger from the Ext Trig terminal, or
continuously output bursts or sweeps using the internal trigger.
At power-on, internal trigger is selected.
From the remote interface, the function generator will accept
a software (bus) trigger, a hardware trigger from the Ext Trig
terminal, or an immediate internal trigger.
The trigger source setting is stored in volatile memory; the source is
set to internal trigger (front panel) or immediate (remote interface)
when power has been off or after a remote interface reset.
To select the trigger source from the remote interface, use the
following command. The APPLy command automatically sets the
trigger source to IMMediate .
3
TRIGger:SOURce {IMMediate|EXTernal|BUS}
Internal Triggering In this trigger mode, the function generator continuously outputs the burst or sweep (as specified by the burst rate or sweep time). This is the power-on trigger source for both front-panel and remote interface use.
To select the internal trigger source from the remote interface, use the following command. The APPLy command automatically sets the trigger source to IMMediate.
"TRIGGER:SOURCE IMMEDIATE"
99
Chapter 3 Features and Functions
Triggering
Single Triggering In the single trigger mode (front panel only), you can manually trigger the function generator by pressing the front-panel initiates one sweep each time you press the key. The
Single key. The function generator outputs one burst or
Trig
annunciator turns on when the function generator is waiting for a single trigger.
The front-panel
Single key is disabled when in remote.
External Triggering In this trigger mode, the function generator will accept a hardware trigger applied to the Ext Trig terminal. The function generator outputs one burst or initiates one sweep each time Ext Trig receives the rising edge of a
TTL pulse.
See also “Ext Trig/FSK/Burst Input Terminal,” on page 102.
Front-Panel Operation: The external trigger mode is like the single
trigger mode except that you apply the trigger to the Ext Trig terminal. Pressing
enables the external trigger mode. The
Single to enable the single trigger mode also
Trig annunciator turns on
when the function generator is waiting for an external trigger.
The front-panel
Single key is disabled when in remote.
Remote Interface Operation:
TRIGger:SOURce EXTernal
100
Chapter 3 Features and Functions
Triggering
Software (Bus) Triggering The bus trigger mode is available only from the remote interface. This mode is similar to the single trigger mode from the front panel, but you trigger the function generator by sending a bus trigger command. The function generator outputs one burst or initiates one sweep each time the trigger command is received.
To select the bus trigger source, send the following command.
"TRIGGER:SOURCE BUS"
To trigger the function generator from the remote interface (
RS-232), send the *TRG (trigger) command.
You can also trigger the function generator from the
by sending the
The following statement shows how to send a
TRIGGER 710
The following statement shows how to send a trigger from the HP-IB
interface using the HP-IB Command Library for C or QuickBASIC.
IOTRIGGER (710)
IEEE-488 Group Execute Trigger (GET) message.
Group Execute Trigger
HP-IB interface
GET using HP BASIC.
HP-IB or
3
101
Loading...