Agilent 33120A Installation Instructions

Installation Instructions for Retrofit Kit
Part Number 33120-90021 October 1998
© Copyright Hewlett-Packard Company 1994, 1998
All Rights Reserved.
HP 33120A / Option 001 Phase-Lock Assembly
HP 33120A / Option 001 Phase-Lock Assembly
Installation Instructions for 33120-80001 Retrofit Kit
HP 33120A. This assembly can be installed only in instruments
the
shipped after August 1994 with serial numbers US34004078 and above.
Refer to the HP 33120A / Option 001 User’s and Service Guide (included with the retrofit kit) for complete details on using the assembly.

Installing the Phase-Lock Assembly

1 Remove the outer cover from the HP 33120A.
Loosen Captive Screws
a
Remove Bottom Screw
Slide Off Outer Cover
b
2 Carefully cut two holes in the rear-panel label.
The Ref Out and Ext Ref In terminals on the Phase-Lock assembly will extend through these holes in the
HP 33120A rear-panel.
Cut Out Holes
3 Connect the coaxial cable to the Phase-Lock circuit board.
Connect one end of the coaxial cable (included with the retrofit kit) to
the connector labeled P2 on the Phase-Lock circuit board. Position the
cable as shown.
Coaxial Cable (33120-61603)
P2
P2
2
4 Install the Phase-Lock assembly in the
Top View
HP 33120A chassis.
Before installing the Phase-Lock assembly, make sure that the red and black wires for the instrument fan are clear of the rear-panel holes.
With the Phase-Lock assembly positioned trace side up, insert the two
BNC connectors through the holes previously cut in the rear-panel label.
Trace Side Up
Lock Washer (2190-0699) Hex Nut (2940-0256)
Side View
5 Secure the Phase-Lock assembly in place.
As shown above, place one lock washer and one hex nut (included with the retrofit kit) on each of the two
BNC connectors. To prevent cross-
threading, be sure to hand-tighten the hex nuts at least one turn before using a tool.
3
6 Connect the coaxial cable to
HP 33120A circuit board.
First, slide the black ferrite bead to the end of the coaxial cable as shown.
Then, connect the coaxial cable to connector J401 on the
HP 33120A
circuit board. Be sure to position the cable away from the heatsinks on the circuit board.
J401
Ferrite Bead
7 Connect the ribbon cable to the
HP 33120A circuit board.
Connect the loose end of the grey ribbon cable to connector J102 on the
HP 33120A circuit board. Press down firmly on the ribbon cable
connector and secure the cable in place with the locking clips.
Ribbon Cable
J102
Snap the Locking Clips over the top of the connector to secure the Ribbon Cable.
4

Calibration /Adjustment Procedure

After installing the Phase-Lock assembly, perform the frequency adjustment procedure (CAL:SETup 0) on page 67 in the HP 33120A Service Guide. Be sure to make the 1 kHz frequency measurement with an accuracy of at least 8 digits (0.01 ppm). You may be required to perform up to initial installation of the Phase-Lock assembly.
For complete details on performing a calibration, see chapter 4 in the HP 33120A Service Guide.
three iterations of the frequency adjustment following
5