Agilent Technologies Probes for
the 16760A Logic Analyzer
(E5378A, E5379A, E5380A, and E5386A)
Probing Solutions for the Agilent 16760A
High Speed State Analyzer — At a Glance
The Agilent 16760A is a high-speed logic analyzer capable of making state
measurements at clock speeds up to 1.5 Gb/s. For high bandwidth measurements,
the 16760A logic analyzer has both a differential and single-ended probes available.
For target systems designed to use the Agilent E5346A 38-Pin Probe, a probe is
available using an Amp “MICTOR 38” connector.
For more information on the 16760A logic analyzer, refer to the online help in the
product.
E5378A 100-pin Single-ended
Probe
Also available as option 010 on the Agilent
16760A.
•34 Channels
•State speeds up to 1.5 Gb/s
•250 mV peak-to-peak sensitivity
•100-pin Samtec connector
•Requires Probing Connector Kit
(see page 44)
E5386A Half-channel Adapter
with E5378A
The E5386A adapter maps the 34 signals
from the 100-pin samtec connector to the
16760A when operating in half-channel
state mode..
16760A module
Logic analyzer
probe cables
E5378A 100-pin single-ended
Two 16760A modules
Four
logic analyzer
probe cables
Two E5386A
half-channel adapters
E5378A 100-pin single-ended probe
2
16760A module
E5379A 100-pin Differential
Probe
Also available as option 011 on the Agilent
16760A.
Two
logic analyzer
probe cables
Two E5379A
differential probes
16760A module
Two probe
cables
E5386A half-channel
adapter
E5379A differential probe
E5380A 38-pin Single-ended
Probe
•17 Channels
•State speeds up to 1.5 Gb/s
•200 mV peak-to-peak sensitivity
•100-pin Samtec connector
•Requires Probing Connector Kit
(see page 44)
E5386A Half-channel Adapter
with E5379A
The E5386A adapter maps the 17
differential signals from the 100-pin
Samtec connector to the 16760A when
operating in half-channel state mode.
16760A module
Also available as option 012 on the Agilent
16760A.
•Compatible with boards designed for
Logic analyser
probe cables
Agilent E5346A 38-pin Probe
•34 Channels
•State speeds up to 600 Mb/s
•300 mV peak-to-peak sensitivity
•38-pin MICTOR connector
E5380A 38-pin
single-ended
probe
•Requires AMP MICTOR 38 Connector
and Agilent Support Shroud
(see page 44)
3
In This Book
In this book, you will find information that helps you understand and implement the
high-bandwidth, high density probing solutions available with the Agilent 16760A
high speed state logic analyzer. Use this information to both evaluate the electrical
and mechanical implications to your target system’s design, and to properly select
and layout the proper components used to connect to the logic analyzer.
Chapter 1 is a description of the available probing options.
Chapter 2 covers the mechanical considerations such as connector/shroud type,
footprint for PC board layout, and probe/connector dimensions.
Chapter 3 covers the electrical considerations such as equivalent probe loads
Chapter 4 provides design theory.
Chapter 5 offers a list of recommended reading for additional information.
Chapter 6 lists connectors and shrouds that may be ordered.
MECL System Design Handbook42
High-speed Digital Design42
Designing High-speed Target Systems for Logic Analyzer Probing42
6
Contents
6 Connectors and Shrouds43
Ordering Probing Connectors and Shrouds44
7
Contents
8
1
Probing Options
Description of the E5378A, E5379A, E5380A, and E5386 probes.
9
Chapter 1: Probing Options
The E5378A 100-pin Single-ended Probe
The E5378A 100-pin Single-ended Probe
The E5378A is a 34-channel single-ended probe capable of capturing data up to 1.5
Gb/s. The probe has the following inputs:
•32 single-ended data inputs, in two groups (pods) of 16.
•Two differential clock or data inputs. Refer to page 37 for a discussion of how to
utilize the clock input. Either or both clock inputs can be acquired as data inputs
if not used as a clock.
•Two data threshold reference inputs, one for each pod (group of 16 data inputs).
Refer to page 37 for guidelines and discussion about how to utilize the threshold
inputs.
The Agilent 16760-68702 or 16760-68703 Probing Connector Kit is required for
connecting the E5378A probe to your target system. The kit contains five mating
connectors and five support shrouds. The connectors and shrouds may be ordered
separately if desired. See the table on page 44 for part numbers.
See AlsoChapter 2 for the mechanical information to design the connector into your target
system board.
E5378A 100-pin single-ended probe and probing connector kit
10
Chapter 1: Probing Options
The E5379A 100-pin Differential Probe
The E5379A 100-pin Differential Probe
The E5379A is a 17-channel differential probe capable of capturing data up to 1.5
Gb/s. Two E5379A probes are required to support all the inputs on one 16760A. The
probe has the following inputs:
•16 differential data inputs.
•One differential clock or data input. Refer to page 38 for a discussion of how to
utilize the clock input. The clock input can be acquired as a data input if not used
as a clock.
The Agilent 16760-68702 or 16760-68703 Probing Connector Kit is required for
connecting the E5379A probe to your target system. The kit contains five mating
connectors and five support shrouds. The connectors and shrouds may be ordered
separately if desired. See the table on page 44 for part numbers.
Differential Input Amplitude
Definition.
difference voltage V - V
than or equal to 200 mV p-p.
For differential signals, the
must be greater
See Also
Chapter 2 for the
mechanical
information to design
the connector into
your target system
boards.
E5379A 100-pin differential probe and probing connector kit
11
Chapter 1: Probing Options
The E5380A 38-pin Single-ended Probe
The E5380A 38-pin Single-ended Probe
The E5380A is designed to be compatible with target systems designed for the
Agilent E5346A 38-pin Probe. If you have a target system designed for connection
to the E5346A 38-pin Probe, the E5380A probe will connect directly to the same
connector. The recommended state speed of this probe is 600 Mbits/second. The
minimum input signal amplitude required by the E5380A is 300 mV.
The probe combines two 17-channel cables into a single-ended 38-pin MICTOR
connector. The probe has the following inputs:
•32 single-ended data inputs, in two groups (pods) of 16.
•Two single-ended clock or data inputs. Refer to page 37 for a discussion of how
to utilize the clock input. Either or both clock inputs can be acquired as data
inputs if not used as a clock.
The Agilent E5346-68701 or E5346-68700 Probing Connector Kit is required for
connecting the E5380A probe to your target system. The kit contains five mating
connectors and five support shrouds. The connectors and shrouds may be ordered
separately if desired. See the table on page 44 for part numbers.
See AlsoChapter 2 for the mechanical information to design the connector into your target
system board
E5380A 38-pin single-ended probe and probing connector kit
12
Chapter 1: Probing Options
The E5386A Half-channel Adapter
The E5386A Half-channel Adapter
The E5386A Half-channel Adapter is intended to be used in half-channel state mode
and works with:
•E5378A 100-pin Single-ended Probe
•E5379A 100-pin Differential Probe
The E5386A Half-channel Adapter has it's own ID code. When using the adapter,
the logic analyzer recognizes its code rather than that of the probe which is attached
to the target. Therefore, the user interface format menu doesn't automatically set
thresholds to the right values. You need to go into the threshold menu and select
(differential, custom, or standard settings).
When using the adapter in half-channel state:
•Clock-bits are not available in half-channel state mode (although JCLK on the
master is still used).
•Be sure to connect Master pod 1 of the logic analyzer to the upper bits, 8-15 +
clk, on the half-channel adapter. This is necessary to connect the clock in the
system under test to the logic analyzer system clock.
•Using the E5386A does not reduce the performance of the 16760A and the
E5378A or E5379A system.
If the E5386A is used in full-channel state mode, the thresholds on the unused (odd)
bits are floating. This could result in spurious activity indicators in the format menu.
13
Chapter 1: Probing Options
The E5386A Half-channel Adapter
14
2
Mechanical Considerations
Once you have decided which probe is required, use the following mechanical
information to design the appropriate connector into your target system board.
.
15
Chapter 2: Mechanical Considerations
E5378A and E5379A Probe Specifications
E5378A and E5379A Probe Specifications
The E5378A and E5379A probes require a probe kit that contains Samtec
connectors and shrouds. Refer to the table in Chapter 6 for the kit part numbers.
Samtec 100-pin connector footprint and support shroud mounting hole
dimensions
CAUTION:The support shrouds are made of conductive metal. Care should be taken to avoid
shorting adjacent boards or components with the shrouds. For this reason it may be
advisable not to connect the shrouds to ground.
16
Samtec 100-pin connector dimensions
Chapter 2: Mechanical Considerations
E5378A and E5379A Probe Specifications
Support shroud dimensions for 100-pin Samtec connector
Ground pins indicated in this table are grounded in the probe. Grounding of specific
ground pins on the target board is optional. However, the following guidelines
should be observed:
Pin
Number
7980Even D16P/Even
8384Even D16N/Even
Pin
Number
Signal
CLKP
CLKN
1) Multiple ground returns are desirable to maintain signal integrity. As many probe
ground pins as possible should be connected to ground in the target system board.
2) The ground pins located between signal pins are particularly important because
they provide improved signal-to-signal isolation. This is particularly important for
differential inputs. Excessive coupling between differential inputs causes the
apparent input capacitance to increase. Capacitance between the two sides of a
differential signal will appear to each side as approximately twice the capacitance to
ground, because the capacitance is connected to a signal of opposite polarity. The
best practice is to ground as many of these pins on the target board as possible.
20
Chapter 2: Mechanical Considerations
E5378A and E5379A Probe Specifications
E5379A 100-pin Differential Probe Connector
Pin Assignments
Signal
Ground12Ground
Do Not Connect34Do Not Connect
Ground56Ground
D0N78D0P
Ground910Ground
D1N1112D1P
Ground1314Ground
D2N1516D2P
Ground1718Ground
D3N1920D3P
Ground2122Ground
D4N2324D4P
Ground2526Ground
D5N2728D5P
Ground2930Ground
D6N3132D6P
Ground3334Ground
D7N3536D7P
Ground3738Ground
D8N3940D8P
Ground4142Ground
D9N4344D9P
Ground4546Ground
D10N4748D10P
Ground4950Ground
D11N5152D11P
Ground5354Ground
D12N5556D12P
Ground5758Ground
D13N5960D13P
Ground6162Ground
D14N6364D14P
Ground6566Ground
Pin
Number
Pin
Number
Signal
21
Chapter 2: Mechanical Considerations
E5378A and E5379A Probe Specifications
E5379A 100-pin Differential Probe Connector
Pin Assignments
Signal
D15N6768D15P
Ground6970Ground
NC7172NC
Ground7374Ground
NC7576NC
Ground7778Ground
D16N/CLKN7980D16P/CLKP
Ground8182Ground
NC8384NC
Ground8586Ground
NC8788NC
Ground8990Ground
NC9192NC
Ground9394Ground
Ground9596Ground
NC9798NC
NC99100NC
Ground pins indicated in this table are grounded in the probe. Grounding of specific
ground pins on the target board is optional. However, the following guidelines
should be observed:
Pin
Number
Pin
Number
Signal
1) Multiple ground returns are desirable to maintain signal integrity. As many probe
ground pins as possible should be connected to ground in the target system board.
2) The ground pins located between signal pins are particularly important because
they provide improved signal-to-signal isolation. This is particularly important for
differential inputs. Excessive coupling between differential inputs causes the
apparent input capacitance to increase. Capacitance between the two sides of a
differential signal will appear to each side as approximately twice the capacitance to
ground, because the capacitance is connected to a signal of opposite polarity. The
best practice is to ground as many of these pins on the target board as possible.
22
Chapter 2: Mechanical Considerations
E5380A 38-pin Single-ended Probe
E5380A 38-pin Single-ended Probe
The E5380A probe is compatible with target systems designed for the Agilent
E5346A 38-pin probe. This probe requires a probe kit that contains MICTOR
connectors and shrouds. Refer to the table in Chapter 6 for the kit part numbers.
38-pin MICTOR connector footprint and support shroud mounting hole
dimensions.
23
Chapter 2: Mechanical Considerations
E5380A 38-pin Single-ended Probe
MICTOR connector dimensions
Support shroud dimensions for the MICTOR connector
Do not connect the following pins. These pins are +5 volt supply and DC return
for analysis probes.
+5 VDC11, 391, 39
GROUND32, 402, 40
Do not connect the following pins. They are used by the Agilent logic analyzer
with an emulator or analysis probe to program or read target information.
SCL2 5
SDA45
27
Chapter 2: Mechanical Considerations
E5386A Half-channel Adapter
E5386A Half-channel Adapter
The E5386A Half-channel Adapter works with the E5378A 100-pin Single-ended
Probe and the E5379A 100-pin Differential Probe.
Half-channel adapter dimensions.
28
Chapter 2: Mechanical Considerations
Signal Name
E5386A Half-channel Adapter
Used with E5378A 100-pin Single-ended Probe
When used with the E5378A 100-pin Single-ended Probe, you need two halfchannel adapters, one adapter for Odd data and one for Even data. The table below
shows the pin assignments.
Electrical considerations such as equivalent probe loads.
31
Chapter 3: Electrical Considerations
Equivalent Probe Loads
Equivalent Probe Loads
The equivalent probe loads for the E5378A, E5379A, and E5380A probes are shown
in the figures below. The equivalent loads include the 100-pin Samtec or 38-pin
MICTOR connector.
E5378A and E5379A Models
The following simple model is accurate up to 1 GHz. Transient analysis with Spice
is fastest with this model.
150
1.6 pF
0.7 pF
20K
+0.75V
The following transmission line model is the most accurate. It is accurate up to 5
GHz. Transient analysis with Spice will be the slowest with this model.
TConnector
Z0=47
Td=75ps
150
0.7 pF
20K
+0.75V
32
Chapter 3: Electrical Considerations
Equivalent Probe Loads
The following lumped LC transmission line model is identical to the transmission
line, but provides faster transient analysis.
1.17nH1.17nH1.17nH
0.27 pF
0.53 pF
0.53 pF
0.27 pF
150
0.7 pF
20K
+0.75V
E5380A Model
The following equivalent probe load for the E5380A includes the target connector.
The model is accurate up to 1 GHz .
180
3 pF
20K
+0.75V
0.7 pF
33
Chapter 3: Electrical Considerations
Equivalent Probe Loads
34
4
Circuit Board Design
Design considerations.
35
Chapter 4: Circuit Board Design
Transmission Line Considerations
Transmission Line Considerations
Stubs connecting signal transmission lines to the connector should be as short as
feasible. Longer stubs will cause more loading and reflections on a transmission
line. If the electrical length of a stub is less than 1/5 of the signal rise time, it can be
modeled as a lumped capacitance. Longer stubs must be treated as transmission
lines.
([DPSOHAssume you are using FR-4 PC board material with a dielectric constant of ~4.3 for
inner-layer traces (microstrip). For example, A 0.28 cm long stub in an inner layer
has a propagation delay of ~20 ps. Therefore, for a signal with a rise time of 100 ps
or greater, a 0.28 cm stub will behave like a capacitor.
The trace capacitance per unit length will depend on the trace width and the spacing
to ground or power planes. If the trace is laid out to have a characteristic impedance
of 50 ohms, it turns out that the capacitance per unit length is ~ 1.2 pF/cm. Therefore
the 0.28 cm stub in the previous example would have an effective capacitance equal
to ~0.34 pF.
This trace capacitance is in addition to the probe load model.
36
Chapter 4: Circuit Board Design
Thresholds
Thresholds
E5378A 100-pin single-ended probe
Data inputs
The E5378A 100-pin single-ended probe has two inputs for a user-supplied
threshold voltage for the data inputs, one for the even pod and one for the odd pod.
The threshold inputs (pins 87 and 88) may be grounded, left open, or connected to a
dc power supply. For each group of data inputs, you may either:
•Supply a threshold voltage between -3V dc and +5V dc to the threshold input.
The logic analyzer will use this threshold to determine when the signal is high or
low.
Or
•Adjust the logic threshold in the user interface to between -3V dc and +5V dc.
The advantages of supplying a threshold voltage via the threshold input on the probe
are:
•A threshold supplied from the source will typically track changes in supply
voltage, temperature, etc.
•A threshold supplied from the target is typically the same threshold that the
target system's logic uses to evaluate the signals. Therefore the data captured by
the logic analyzer will be congruent with the data as interpreted by the target
system.
Clock input
The clock input to the E5378A probe is differential. If you supply a differential
clock, you should select the "differential" option in the clock threshold user
interface.
If your system uses a single-ended clock signal, the clock
grounded or connected to a dc power supply. You may:
37
input should be either
Chapter 4: Circuit Board Design
Thresholds
•Ground the clock input and adjust the clock threshold from the user interface to
between -3V dc and +5V dc.
Or
•Supply a threshold reference voltage between -3V dc and +5V dc to the clock
input.
If your circuit uses a resistive divider to provide a threshold reference, be sure to
consider the equivalent circuit consisting of the 20k
as shown on page 32 and 33.
The threshold for the clock input has a separate adjustment in the user interface,
independent of the data inputs.
Ω resistor connected to +0.75V
E5379A 100-pin differential probe
Data inputs
If you are using the E5379A 100-pin differential probe to acquire differential
signals, you would normally allow the logic analyzer to discriminate between high
and low states based on the crossover of the data and data
You may also use the E5379A 100-pin differential probe to acquire single-ended
signals. If you are using the E5379A probe to acquire single-ended signals, you
should either ground the data
•Ground the data
Or
•Supply a threshold reference voltage to the data
inputs and adjust the threshold in the user interface.
inputs or connect them to a dc power supply. You may:
inputs.
inputs.
If your circuit uses a resistive divider to provide a threshold reference, be sure to
consider the equivalent circuit consisting of the 20k
as shown on page 32 and 33.
Ω resistor connected to +0.75V
Clock input
The same choices exist for the clock input on the E5379A 100-pin differential probe
as outlined above for the data inputs. The clock input has a separate, independent
threshold adjustment.
38
Chapter 4: Circuit Board Design
Thresholds
E5380A 38-pin single-ended probe
All inputs on the E5380A 38-pin probe are single-ended. The E5380A probe does
not have a threshold reference input. When you use the E5380A, you adjust the logic
threshold in the user interface.
The clock input on the E5380A is single-ended. The clock threshold may be
adjusted independent of the data.
Signal Access
Labels split across probes
If a label is split across more than one pod, this leads to restrictions in triggering.
Refer to "Triggering with the Agilent 16760A" for more details.
Reordered bits
If bits need to be reordered within a label, this leads to additional restrictions in
triggering. Specifically, equalities can be used to evaluate the value of a label with
reordered bits, but inequalities cannot be used. You may be able to avoid the need to
reorder bits in a label by routing signals to appropriate pins on the probe connector.
39
Chapter 4: Circuit Board Design
Thresholds
Half-channel 1.5 Gb/s mode
In the half-channel 1.5 Gb/s mode, the analyzer accesses only the even channels
(0,2,4, etc.). In the Format user interface, the connections within a pod (16-signal
group) are mapped as follows:
Connector
pins
7,8D0Bit0
15,16D2Bit1
23,24D4Bit2
31,32D6Bit3
39,40D8Bit4
47,48D10Bit5
55,56D12Bit6
63,64D14Bit7
Note that in the 1.5 Gb/s half-channel mode, the clock inputs cannot be assigned as
bits in a label.
Connection name in this
document (pages 19- 22)
Reference in format
window
E5386A Half-channel Adapter. The E5386A can be used with the E5378A
100-pin Single-ended Probe or the E5379A 100-pin Differential Probe to map the
signals from the 100-pin Samtec connector to the 16760A when operating in halfchannel state mode.
40
5
Recommended Reading
A list of recommended reading for more information about systems and high-speed
digital design.
41
Chapter 5: Recommended Reading
For More Information
For More Information
MECL System Design Handbook
Blood, William R. Jr., "MECL System Design Handbook," 4th edition, 1988,
published by Motorola. This handbook can be obtained from ON Semiconductor on
the web. Go to <http://onsemi.com>. Click on "On-line ordering" under
"Documentation." Click on the link "General search." Type in "HB205" in the
"Document number" field. Click "Submit." To view the document online, click on
"PDF" in the right-hand column titled "PDF MFAX." Or order a hardcopy of the
handbook on-line.
High-speed Digital Design
Johnson, Howard W., and Martin Graham, "High-speed Digital Design," PrenticeHall, 1993, ISBN 0-13-395724-1
Designing High-speed Target Systems for Logic
Analyzer Probing
“Designing High-speed Target Systems for Logic Analyzer Probing” Agilent
Technologies application note publication number 5988-2989EN.
42
6
Connectors and Shrouds
A table of part numbers for ordering connectors, shrouds, and kits.
43
Chapter 6: Connectors and Shrouds
Ordering Probing Connectors and Shrouds
Ordering Probing Connectors and Shrouds
Connectors and shrouds may be ordered in kits or ordered separately. Select a
support shroud appropriate for the thickness of your PC board. The following table
lists the Agilent part numbers for each.
CAUTION:The support shrouds marked with an asterisk in the following table are made of
conductive metal. Care should be taken to avoid shorting adjacent boards or
components with the shrouds. For this reason it may be advisable not to connect the
shrouds to ground.
For Probe
Model #
E5378A &
E5379A
E5380AE5346-687015 MICTOR Connectors
Agilent Part NumberConsists of
16760-687025 Mating Connectors &
16760-68703up to 3.05 mm (0.120 in.)
1253-3620 (or Samtec
#ASP-65067-01)
16760-023021 Support Shroud*up to 1.57 mm (0.062 in.)
16760-023031 Support Shroud*up to 3.05 mm (0.120 in.)
E5346-687005 MICTOR Connectors
1252-74311 MICTOR Connectorn/a
AMP part #2-767004-21 MICTOR Connectorn/a
E5346-447011 Support Shroudup to 1.57 mm (0.062”)
E5346-447041 Support Shroud1.575 to 3.175 mm
E5346-447031 Support Shroud3.175 to 4.318 mm
5 Support Shrouds*
1 100-pin Mating Connectorn/a
& 5 Support Shrouds
& 5 Support Shrouds
For Target PC Board
Thickness
up to 1.57 mm (0.062 in.)
up to 1.57 mm (0.062 in.)
1.575 to 3.175 mm
(0.062 to 0.125 in.)
(0.062 to 0.125 in.)
(0.125 to 0.70 in.)
44
Index
A
adapter, E5386A half-channel
C
circuit board design
connector part numbers
connector specifications
support shroud
shroud part numbers
signal access
single-ended probe, 100-pin
single-ended probe, 38-pin
specifications
E5378A 100-pin single-ended
probe
E5379A 100-pin differential
probe
E5380A 38-pin single-ended
probe
E5386A Half-channel Adapter
T
thresholds
transmission line considerations
, 17
, 43
, 39
, 10
, 12
, 16
, 16
, 23
, 37
, 28
, 36
45
Index
46
Safety
Notices
This apparatus has been
designed and tested in accordance with IEC Publication 1010,
Safety Requirements for Measuring Apparatus, and has been
supplied in a safe condition.
This is a Safety Class I instrument (provided with terminal for
protective earthing). Before
applying power, verify that the
correct safety precautions are
taken (see the following warnings). In addition, note the
external markings on the instrument that are described under
"Safety Symbols."
Warnings
• Before turning on the instrument, you must connect the protective earth terminal of the
instrument to the protective conductor of the (mains) power
cord. The mains plug shall only
be inserted in a socket outlet
provided with a protective earth
contact. You must not negate
the protective action by using an
extension cord (power cable)
without a protective conductor
(grounding). Grounding one
conductor of a two-conductor
outlet is not sufficient protection.
• Only fuses with the required
rated current, voltage, and specified type (normal blow, time
delay, etc.) should be used. Do
not use repaired fuses or shortcircuited fuseholders. To do so
could cause a shock or fire hazard.
• If you energize this instrument
by an auto transformer (for voltage reduction or mains isolation), the common terminal must
be connected to the earth terminal of the power source.
• Whenever it is likely that the
ground protection is impaired,
you must make the instrument
inoperative and secure it against
any unintended operation.
• Service instructions are for
trained service personnel. To
avoid dangerous electric shock,
do not perform any service
unless qualified to do so. Do not
attempt internal service or
adjustment unless another person, capable of rendering first
aid and resuscitation, is present.
• Do not install substitute parts
or perform any unauthorized
modification to the instrument.
• Capacitors inside the instrument may retain a charge even if
the instrument is disconnected
from its source of supply.
• Do not operate the instrument
in the presence of flammable
gasses or fumes. Operation of
any electrical instrument in such
an environment constitutes a
definite safety hazard.
• Do not use the instrument in a
manner not specified by the
manufacturer.
To clean the instrument
If the instrument requires cleaning: (1) Remove power from the
instrument. (2) Clean the external surfaces of the instrument
with a soft cloth dampened with
a mixture of mild detergent and
water. (3) Make sure that the
instrument is completely dry
before reconnecting it to a
power source.
Safety Symbols
!
Instruction manual symbol: the
product is marked with this symbol when it is necessary for you
to refer to the instruction manual in order to protect against
damage to the product..
Hazardous voltage symbol.
Earth terminal symbol: Used to
indicate a circuit common connected to grounded chassis.
Agilent Technologies Inc.
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901-2197, U.S.A.
No part of this manual may be
reproduced in any form or by
any means (including electronic
storage and retrieval or translation into a foreign language)
without prior agreement and
written consent from Agilent
Technologies, Inc. as governed
by United States and international copyright laws.
Manual Part Number
16760-97007, February 2002
Print History
16760-97005, January 2002
16760-97003, May 2001
16760-97002, April 2001
16760-97001, February 2001
16760-97000, December 2000
Agilent Technologies, Inc.
1601 California Street
Palo Alto, CA 94304 USA
Restricted Rights Legend
If software is for use in the performance of a U.S. Government
prime contract or subcontract,
Software is delivered and
licensed as “Commercial computer software” as defined in
DFAR 252.227-7014 (June 1995),
or as a “commercial item” as
defined in FAR 2.101(a) or as
“Restricted computer software”
as defined in FAR 52.227-19
(June 1987) or any equivalent
agency regulation or contract
clause. Use, duplication or disclosure of Software is subject to
Agilent Technologies’ standard
commercial license terms, and
non-DOD Departments and
Agencies of the U.S. Government will receive no greater
than Restricted Rights as
defined in FAR 52.227-19(c)(1-2)
(June 1987). U.S. Government
users will receive no greater
than Limited Rights as defined in
FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2)
(November 1995), as applicable
in any technical data.
Document Warranty
The material contained in
this document is provided
“as is,” and is subject to
being changed, without
notice, in future editions.
Further, to the maximum
extent permitted by applicable law, Agilent disclaims
all warranties, either
express or implied, with
regard to this manual and
any information contained
herein, including but not
limited to the implied warranties of merchantability
and fitness for a particular
purpose. Agilent shall not be
liable for errors or for incidental or consequential
damages in connection with
the furnishing, use, or performance of this document
or of any information contained herein. Should Agilent and the user have a
separate written agreement
with warranty terms covering the material in this document that conflict with these
terms, the warranty terms in
the separate agreement
shall control.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and
may be used or copied only in
accordance with the terms of
such license.
WARNING
A WARNING notice
denotes a hazard. It calls
attention to an operating
procedure, practice, or
the like that, if not
correctly performed or
adhered to, could result
in personal injury or
death. Do not proceed
beyond a WARNING
notice until the indicated
conditions are fully
understood and met.
CAUTION
A CAUTION notice
denotes a hazard. It calls
attention to an operating
procedure, practice, or
the like that, if not
correctly performed or
adhered to, could result in
damage to the product or
loss of important data. Do
not proceed beyond a
CAUTION notice until the
indicated conditions are
fully understood and met.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.