Agilent 1670G Specifications Sheet

Installation Instructions
for
E2460GS
Oscilloscope Upgrade Kit
Agilent 1670G-Series Logic Analyzers
This kit upgrades either the Agilent Technologies 1670G, Agilent 1671G, Agilent 1672G, or the Agilent 1673G logic analyzer to an analyzer that contains the 2 GSa/s oscilloscope.
Note Your logic analyzer and this upgrade kit must be shipped together to an Agilent Technologies
TOOLS REQUIRED:
T10 TORX screwdriver
T15 TORX screwdriver
T20 TORX screwdriver
#1 pozidrive screwdriver
5/8-inch deep-well nut driver
April 2000 Part Number E2460-92009
© Copyright Agilent Technologies, Inc. 2000
1900 Garden of the Gods Road, Colorado Springs, Colorado U.S.A.

PARTS SUPPLIED

Most parts are not labeled. Refer to figure used in the instructions for easy identification.
Part Number Description QTY.
01660-41904 Elastomeric Keypad 1 01670-61602 Cable, 60 Conductor (Analyzer Board to Scope Board) 1 01660-68721 2 Chan., 2GSa/s Digital Oscilloscope Board Assy. 1 01670-68714 SW Disk Pouch (Latest Version) 1 01670-94305 Label-Line Switch 1 E2460-68705 Rear Panel Assembly 1 0515-0430 Machine Screw 1 01160-60001 1160A 10:1 Passive Probe 2 2190-0068 Washer, Lock 0.505 0.630 0.02 2 2950-0054 Nut, Hex 1/2-28 0.125 2 54503-25701 Nut, Hex, Tapered 2 5090-4383 Serial Number Tag 1

INTRODUCTION

Refer to figure 1, Agilent Technologies 167xG Logic Analyzer Exploded View, when performing this upgrade. Note that A5 is the oscilloscope card.
2
Figure 1 (Tear out and use as reference)
Agilent Technologies 167xG Logic Analyzer Exploded View
3
This page intentionally left blank
4

REMOVAL PROCEDURES

WARNING
Hazardous voltages exist on the power supply. To avoid electrical shock, disconnect the power from the instrument before performing the following procedures. After disconnecting the power, wait at least three minutes for the capacitors on the power supply board to discharge before servicing the instrument.
CAUTION Electrostatic discharge can damage electronic components. Use grounded wrist
straps and mats when performing any service to the logic analyzer.
CAUTION Do not remove or replace any circuit assemblies in this instrument while power
is applied. The assemblies contain components which may be damaged if the assembly is removed or replaced while the instrument is powered.
Remove the Probe Mounting Plate and the Probe Cables (See Figure 1)
1. Turn off the power and unplug the logic analyzer.
2. Using a #1 pozidrive screwdriver, loosen the five screws and remove the probe mounting plate (MP23).
3. Remove the probe cables (W1) by pulling them out of their connectors on the rear panel.
Remove the Cover Assembly (See Figure 1)
1. Using a T10 TORX screwdriver, remove the seven screws holding the trim strip (MP6) and the cover assembly (MP22) to the chassis assembly (MP29). Remove the trim strip (MP6).
2. Using a T10 TORX screwdriver, remove the four screws to remove the four rear feet (MP7).
3. Using a T10 TORX screwdriver, remove the two screws that attach the handle assembly (MP34) to the chassis assembly (MP29). Remove the handle assembly (MP34).
4. To remove the cover assembly (MP22) , set the instrument’s front panel facing toward you. Prepare to remove the cover by using a T10 TORX to turn back three revolutions each the four screws fastening the equipment pouch to the cover assembly. Slide the chassis toward the front, out of the cover. Set the instrument on a static-safe work area.
Remove the Disk Drive (See Figures 1)
1. Disconnect the flexible disk drive cable (W5) from the CPU board (A2). Disconnect the hard disk drive cable (W4) from the CPU board (A2).
2 . Using a T10 TORX screwdriver, remove the two screws holding the disk drive bracket (MP3) to the support
tray (MP24).
3 . Slide the disk drive bracket assembly toward the rear of the instrument, then lift it up and out.
5
Remove the Acquisition Board (See Figures 1 and 2)
1. Remove the PCB locking pins (MP1) from the chassis assembly (MP29). Slide the support tray (MP24) out of the chassis assembly.
2. On the bottom of the acquisition board (A4), disconnect the following cables:
Power Sense
Fan
Power Supply
CPU Board
3. On the top of the acquisition board, disconnect the following cables:
Trigger input (white wire to J6)
Trigger output (orange wire to J5)
Figure 2
Acquisition Board Connector Location Diagram
3. Remove the acquisition board (A4).
a. Using a T10 TORX screwdriver, remove six screws that secure the acquisition board to the deck (MP17). b. Angle the front edge of the acquisition board up out of the chassis assembly. c. Lift the acquisition board up and out of the chassis.
6