The HP 16533A and HP 16534A Digitizing Oscilloscopes offer basic
oscilloscope functionality. The oscilloscope can be easily correlated
with other instruments in the HP 16600A-series or HP 16700A logic
analysis system.
Getting Started
•“Calibrating the Oscilloscope” on page 10
•“Probing” on page 14
•“Acquiring a Waveform” on page 31
•“Combining the Oscilloscope with a Logic Analyzer” on page 36
Refining Your Measurement
•“Triggering” on page 39
•“Vertical and Horizontal Scaling” on page 53
•“Changing the Sample Rate” on page 55
•“Comparing Channels” on page 57
•“Using Markers” on page 75
Tip s
•“What Do the Display Symbols Mean?” on page 58
•“Changing Waveform Display and Grid” on page 47
•“Automatic Measurements and Algorithms” on page 62
•“Differences from a Standard Digitizing Oscilloscope” on page 77
•“Using Waveform Memories” on page 78
•“Loading and Saving Oscilloscope Configurations” on page 79
2
•“When Something Goes Wrong” on page 80
“Specifications and Characteristics” on page 85
Main System Help (see the HP 16600A/16700A Logic Analysis System
help volume)
Glossary of Terms (see page 97)
3
4
Contents
HP 16533/34A Digitizing Oscilloscope
1 HP 16533/34A Digitizing Oscilloscope
Calibrating the Oscilloscope10
Calibration Reference12
Probing14
Table of Compatible Probes14
Selecting the Proper Probe15
Compensating the Compensated Passive Divider Probe18
Probe Loading19
Descriptions of Probe Types23
Surface Mount Probing30
Acquiring a Waveform31
Autoscale32
Specifying a Measurement33
Combining the Oscilloscope with a Logic Analyzer36
Trigger Concepts 39
Edge Triggering41
Pattern Triggering42
Delayed Triggering43
Getting a Stable Trigger44
The Trigger Setup Window45
5
Contents
Changing Waveform Display and Grid47
Zooming In47
Changing the Persistence of the Waveform47
Viewing Noisy Waveforms with Averaging49
Changing Display Colors 51
Changing the Grid51
Vertical and Horizontal Scaling53
Changing the Sample Rate55
Comparing Channels57
What Do the Display Symbols Mean?58
Display Setup Window60
6
Contents
Automatic Measurements and Algorithms62
How the Scope Makes Measurements 63
Average Voltage (Vavg)64
Period 64
Rise Time 64
Fall Time65
Negative and Positive Pulse Width (±Width)65
Frequency 66
Base Voltage (Vbase)67
Top Voltage (Vtop)67
Preshoot68
Overshoot69
Peak-to-Peak Voltage (Vpp) 69
Minimum Voltage (Vmin)70
Maximum Voltage (Vmax) 70
Time of Minimum Voltage (Tmin)71
Time of Maximum Voltage (Tmax)71
Voltage Amplitude (Vamp)71
Vdcrms (Root Mean Square Voltage, DC)71
About the Measurements72
Increasing the Accuracy of Your Measurements74
Using Markers75
About Automatic Time Markers76
Differences from a Standard Digitizing Oscilloscope77
Using Waveform Memories78
Loading and Saving Oscilloscope Configurations79
When Something Goes Wrong80
Error Messages 80
Calibration Problems80
Triggering Problems80
Other Problems81
7
Contents
Specifications and Characteristics85
What is a Specification89
What is a Characteristic89
What is a Calibration Procedure90
What is a Function Test90
Help - How to Navigate Quickly91
Run/Group Run Function92
Setting a tool for independent or Group Run93
Setting Single or Repetitive Run94
Checking Run Status 94
Demand Driven Data 95
Glossary
Index
8
1
HP 16533/34A Digitizing Oscilloscope
9
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Calibrating the Oscilloscope
Calibrating the Oscilloscope
The oscilloscope requires a full operational accuracy calibration by you
or a service department whenever
•it has been 6 months or 1,000 hours of use since last full calibration.
•the ambient temperature changes more than 10 degrees C from the
temperature at the time of the last full calibration.
•the frame configuration changes.
•you need to optimize measurement accuracy.
You will get more accurate measurements from the oscilloscope if you
perform the operational accuracy calibration at least once a year.
NOTE:Channel skew calibration requires a multi-board oscilloscope. The procedure
cannot be performed on single-board (2-channel) oscilloscopes.
To calibrate the oscilloscope
This is also covered in the Logic Analysis System Installation Guide.
Since this procedure requires you to turn off the system, print this
information if you do not have access to the Installation Guide.
1. If your oscilloscope has more than two channels, disconnect the short
cables on the back of the module that connect the boards.
2. Unprotect the memory.
a. Turn off the HP 16600A or HP 16700A frame.
b. Take the oscilloscope module out of the frame. See the Logic Analysis
System Installation Guide.
c. Set the PROTECT/UNPROTECT switch to UNPROTECT.
10
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Calibrating the Oscilloscope
d. Put the oscilloscope back in the frame.
3. Turn on the HP 16600A or HP 16700A and wait for it to finish booting.
You will get a more accurate calibration if you warm up the system for 30
minutes before calibrating the oscilloscope.
4. Select the oscilloscope icon with the right mouse button, and drag the
cursor to Calibration...
5. Select the procedure ADC through Logic Trigger.
The calibration software will tell you what cables need to be attached.
6. Click Run.
7. Select the procedure Ext Trig Skew and connect the cables as directed.
8. Click Run.
9. Optional - Calibrate the oscilloscope as a multi-board module.
a. Perform the ADC through Logic Trigger and Ext Trig Skew calibrations
on each oscilloscope board first.
b. Click Exit in the system window.
c. Connect the oscilloscopes together with the short board interconnect
cables. Connect the first board’s TRIG OUT to the next board’s TRIG IN
until all boards are connected.
d. Start a session.
e. Select the oscilloscope icon with the right mouse button, and drag the
cursor to Calibration.
f. Select the procedure Channel Skew and connect the cables as
directed.
10. After you have finished calibrating, protect the memory. Follow the steps
given above for unprotecting, setting the switch to PROTECT instead.
See AlsoLogic Analysis System Installation Guide
“Calibration Reference” on page 12
11
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Calibrating the Oscilloscope
Calibration Reference
ADC
The ADC calibration procedure produces a linearization table which is
applied to the data out of the analog-to-digital converters (ADC) to
undo the effects of a non-linear, analog-to-digital conversion.
Gain
The Gain calibration procedure measures the actual attenuation of the
attenuators and measures the actual gain of the preamps.
Offset
The Offset calibration procedure determines the actual offset value
that places a null signal in center screen.
Hysteresis
The Hysteresis calibration procedure determines the hardware setting
which is closest to achieving a hysteresis of 0.28 screen divisions.
Trigger Level
The Trigger Level calibration procedure determines the actual trigger
level values for all possible voltage levels across the screen.
Trigger Delay
The Trigger Delay calibration procedure determines a time delay which
correctly lines up the point at which a trace crosses the trigger level
with the trigger time.
Logic Trigger
The Logic Trigger calibration procedure determines settings which
affect the accuracy of duration trigger measurements.
Ext Trig Skew
The Ext Trig Skew calibration procedure lines up the external trigger
edge with the trigger time when triggering on the external channel.
12
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Calibrating the Oscilloscope
Channel Skew
The Channel Skew calibration procedure is only available for multiboard oscilloscope modules. It deskews the trigger channel and data
channels which are on different boards.
13
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
Probing
The probes covered in the topics below are 1:1 Passive Probes, Active
Probes, Current Probes, Compensated Passive Divider Probes,
Differential Probes, and Resistive Divider Probes.
•“Table of Compatible Probes” on page 14
•“Selecting the Proper Probe” on page 15
•“Compensating the Compensated Passive Divider Probe” on page 18
•“Probe Loading” on page 19
•“Descriptions of Probe Types” on page 23
•“Surface Mount Probing” on page 30
Table of Compatible Probes
* Most frequently used
Model Probe Type Band- Input Div Input R
Input C
Numbers width Z ratio
Applications ECL, CMOS, GaAs probing, analog circuit probing,
transmission line probing, source resistance Š10 kohm, op amp
probing, most accurate for general measurements of circuits of
unknown impedance.
Compensated Passive Divider Probe
Features
Very low resistive loading, accurate amplitude measurements,
16
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
large dynamic range, and low cost.
Tradeoffs Capacitive loading <10 pF, lower bandwidth than active or
50-ohm resistive divider probes.
Applications General purpose probing, probing high-impedance nodes
(Š10 Kohm), op amp probing, CMOS probing (if bandwidth is
adequate), TTL probing (if bandwidth is adequate)
Current Probe
Features
Measures both ac and dc currents on a scope, with minimal
circuit loading.
Tradeoffs Large size.
Applications Power measurements, automotive measurements,
industrial measurements, motors, dynamoes, and alternators.
Differential Probe
Features
High common mode rejection ratio, easy viewing of small
waveforms with large dc offsets, more accurate than subtracting one
channel from another.
Tradeoffs Bigger than a passive probe, high cost, requires power, and
lower bandwidth than other probes.
Applications Measuring waveforms not referenced to the scope ground,
troubleshooting power supplies, and differential amplifier probing.
Resistive Divider Probe
Features
Highest bandwidth, lowest capacitive load, lower cost than
active probes, flat pulse response, good timing measurement accuracy.
Tradeoffs Relatively heavy resistive loading.
Applications ECL probing, GaAs probing, and transmission line probing.
See Also“Descriptions of Probe Types” on page 23
“Table of Compatible Probes” on page 14
17
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
Compensating the Compensated Passive
Divider Probe
Before you can have a flat frequency response when using a
Compensated Passive Divider Probe, the probe’s cable capacitance and
scope input capacitance must be compensated. One of the
compensating capacitors in the probe is adjustable so you can optimize
the step response for flatness.
1. Connect the probe to the BNC Output, labeled AC/DC CAL, on the back of
the oscilloscope.
2. Connect the probe ground lead to ground.
3. Click the oscilloscope icon and select Calibration...
4. At the bottom of the calibration window, set BNC Output to Probe Comp
and close the window.
5. Click the oscilloscope icon and select Setup/Display...
6. Click Autoscale.
7. You should see a waveform similar to one of the following.
8. If necessary, adjust the probe’s compensating capacitor. Set the scope to
keep running by right-clicking Run, selecting repetitive, then clicking Run.
18
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
Probe Loading
There are two major factors influencing probe selection: the load the
probe imposes on your circuit and the required bandwidth of your
circuit with the probe. This is discussed in three sections, below.
Probe Resistance and Capacitance Characteristics (see page 19)
Probe Ground Lead Characteristics (see page 20)
Understanding System Bandwidth at the Probe Tip (see page 21)
Probe Resistance and
Capacitance
Characteristics
The probe load has both resistive and capacitive components. In
addition, the inductance in the probe ground lead causes ringing.
The probe resistance to ground forms a voltage divider network with
the source resistance of your circuit. This reduces the waveform
amplitude and the dc offset. For example, if the probe’s resistance is 9
times the Thevenin equivalent resistance of your circuit, the waveform
amplitude is reduced by about 10 percent. Therefore, if your waveform
has a +5 V to 0.8 V range, the scope probe system shows a 4.5 V to 0.72
V range.
NOTE:At high frequencies, the probe reactance dominates the resistance.
The probe capacitive loading (Cin) to ground forms an RC circuit with
the resistance of your circuit (R
the probe and scope (R
). The time constant of this RC circuit slows
in
) and the resistance looking into
source
the rise time of any transitions, increases the slew rate, and introduces
delay in the actual transition time. The approximate rise time of a
simple RC circuit is:
t
R
RC
Tot al
= 2.2R
= [RinR
Tot alCin
source
where
]/[Rin + R
source
]
Thus, for circuit resistance of 100 ohm, a scope probe system
resistance of 1 Mohm, and a probe capacitance of 8 pF, the real rise
Therefore, the rise time of your circuit cannot be faster than
approximately 1.8 ns, even though it might be faster without the probe.
If the output of the circuit under test is current-limited (as is often the
case for CMOS), the slew rate is limited by the relationship dV/dT = I/C.
Perhaps you have connected a scope to a circuit for troubleshooting
only to have the circuit operate correctly after connecting the probe.
The capacitive loading of the probe can attenuate a glitch, reduce
ringing or overshoot of your waveform, or slow an edge just enough
that a setup or hold time violation no longer occurs.
Probe Ground Lead
Characteristics
NOTE:If you print this page, subscripts and superscripts appear on the main line of
text. If a number seems to be in an odd place in the printed copy, it is probably
a superscript.
The inductance of the probe’s ground lead forms an LC circuit with the
probe’s capacitance and the output capacitance of the circuit under
test, including any parasitic capacitance of PC board traces, and so on.
The ringing frequency (F) of this circuit is:
20
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
Understanding
System Bandwidth at
the Probe Tip
F = (2 (3.14) (LC)
1/2)-1
If the rise time of the waveform is sufficient to stimulate this ringing,
the ringing can appear as part of your captured waveform. To calculate
the ringing frequency, you can assume that the probe’s ground lead has
an inductance of approximately 25 nH per inch. So, a probe with a
capacitance of 8 pF and a 4-inch ground lead has a ringing frequency of
approximately:
1/2)-1
F = (2 (3.14) [(25 nH) (4 inches) (8 pF)]
= 178 MHz
The 178 MHz does not include your circuit capacitance. Therefore, a
waveform with a rise time of less than 1.9 ns can stimulate ringing.
= 0.35/178 MHz = 1.9 ns
t
rise
To minimize the ringing effect, you should use a probe ground lead that
is as short as possible. Some probes add a ferrite bead to the ground
lead to reduce ringing. However, adding the ferrite bead also increases
the ground impedance which reduces the common mode rejection of
the probe.
System bandwidth is the bandwidth of the scope probe system. System
bandwidth affects measurements because the probe becomes part of
the circuit being measured. The rise time that is measured depends on
the actual rise time, the rise time of the scope probe system, and the
rise time of the RC circuit formed by the source resistance and the
scope probe system resistance and capacitance.
t
meas
= [t
act
2
+ t
RC
2
+ t
sys
2
1/2
]
where
= the measured rise time.
t
meas
= the actual rise time of the waveform being measured.
t
act
= the rise time of the RC circuit formed by the source resistance
t
RC
and the scope probe system resistance and capacitance.
= the rise time of the scope probe system.
t
sys
21
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
NOTE:Often the bandwidth of the scope probe system is specified. The rise time is
calculated using the following equation.
= 0.35/SystemBW
t
sys
If the rise time of the scope probe system is not specified, it can be calculated
using the following formula.
probe
2
+ t
= [t
t
sys
scope
1/2
2
]
For example, if the scope probe system rise time is 600 ps, the probe
loading rise time (t
) is 600 ps, and the waveform has a 1-ns rise time,
RC
then the measured rise time is:
= [(1 ns)2 + (600 ps)2 + (600 ps)2]
t
meas
1/2
= 1.3 ns
The answer is in error by 30%.
However, if the scope probe system rise time is 190 ps, the probe
loading rise time is 190 ps, and the waveform has a 1-ns rise time, then
the measured rise time is:
= [(1 ns)2 + (190 ps)2 + (190 ps)2]
t
meas
1/2
= 1.03 ns
Now the error is only 3%.
You may find it useful to memorize three system bandwidth rules:
1. The combined rise time of the scope probe system and the probe loading
should be less than 1/3 of the rise time of the waveform you are measuring
to keep errors below 5%, and less than 1/7 of the rise time of the waveform
you are measuring to keep errors below 1%.
2. Rise time and bandwidth are related by the following approximations: rise
time = 0.35/bandwidth and bandwidth = 0.35/rise time.
3. Rise times add approximately as the square root of the sum of the squares
(for systems with minimal peaking).
NOTE:Because every scope probe has a different loading effect on your circuit, you
should use the equation given for the type of scope probe you are using.
See Also“Descriptions of Probe Types” on page 23
22
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
Descriptions of Probe Types
For each of the probe types listed below, the description gives a
summary of features and tradeoffs and a short text description. Most of
the probe types also give a sample rise time calculation.
•“1:1 Passive Probes” on page 23
•“Active Probes” on page 24
•“Compensated Passive Divider Probes” on page 26
•“Current Probes” on page 27
•“Differential Probes” on page 28
•“Resistive Divider Probes” on page 28
1:1 Passive Probes
Example Rise Time
Calculation
Features No attenuation of waveform.
Tradeoffs High capacitive loading and low bandwidth.
Applications Measuring small, low-bandwidth waveforms when no
attenuation can be tolerated such as power supply ripple.
The 1:1 passive probes provide a way to connect the input impedance
of the scope directly to your circuit with minimum attenuation due to
the resistive loading of the probe. However, 1:1 probes do have very
high capacitive loading which is much larger than that of the scope.
There are two types of 1:1 passive probes. One type is designed to work
with the scope’s input set to high impedance (1 Mohm) and uses a
lossy cable to keep the probe from ringing. The other type is designed
to work with the scope’s input set to low impedance (50 ohm) and uses
a 50-ohm coaxial cable.
Given the following circuit using the HP 1162A probe,
23
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
the input resistance is:
= R
R
in
= 1 Mohm
scope
The total resisitance is:
= (RinR
R
Tot al
= 1 Mohm(50 ohm)/(1 Mohm + 50 ohm) = 50 ohm
R
Tot al
source
)/(Rin + R
source
)
From the Table of Compatible Probes, the probe capacitance is 50 pF.
Therefore, the capacitive load is:
= C
C
in
probe
+ C
= 50 pF + 7 pF = 57 pF
scope
The rise time due to circuit loading is:
= 2.2R
t
RC
Tot alCin
tRC = 2.2(50 ohm)(57 pF) = 6.2 ns
From the Table of Compatible Probes, the scope probe system has a
bandwidth of 25 MHz. Therefore, the rise time of the scope probe
system is: t
= 0.35/25 MHz = 14 ns
t
Sys
The measured rise time is: t
2
ns)
+ (6.2 ns)2 + (14 ns)2]
= 0.35/SystemBW
Sys
= [t
meas
1/2
= 140.8 ns
act
2
+ t
RC
2
+ t
sys
2
1/2
]
t
= [(140
meas
Active Probes
Features Best overall combination of low resistive and capacitive
Applications ECL, CMOS, GaAs probing, analog circuit probing,
transmission line probing, source resistance Š10 kohm, op amp
probing, most accurate for general measurements of circuits of
24
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
unknown impedance.
An active probe has a buffer amplifier at the probe tip. This buffer
amplifier drives a 50-ohm cable terminated in 50 ohms at the scope
input. Active probes offer the best overall combination of resistive
loading, capacitive loading, and bandwidth.
Example Rise Time
Calculation
Given the following circuit using the HP 1152A probe,
the input resistance is:
R
= 100 kohm. The total input resistance is:
in
= (RinR
R
Tot al
= 100 ohm(50 ohm)/(100 ohm + 50 ohm) = 50 ohm
R
Tot al
source
)/(Rin + R
source
)
The rise time due to circuit loading is:
= 2.2R
t
RC
Tot alCtip
tRC = 2.2(50 ohm)(0.6 pF) = 66 ps
Because the rise time of the scope probe system is not given in the
Table of Compatible Probes, we will have to calculate it using the
bandwidth of the probe (2.5 GHz) and the bandwidth of the scope (500
MHz). Therefore, the rise time of the scope probe system is:
= 0.35/ProbeBW = 0.35/2.5 GHz = 140 ps
t
probe
= 0.35/ScopeBW = 0.35/500 MHz = 700 ps
t
scope
t
= [t
sys
t
= [(140 ps)2 + (700 ps)2]
sys
probe
2
+ t
scope
2
1/2
]
1/2
= 714 ps
The measured rise time is:
25
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
t
meas
= [t
act
2
+ t
RC
2
+ t
sys
2
1/2
]
t
= [(2 ns)2 + (66 ps)2 + (714 ps)2]
meas
1/2
= 2.12 ns
Compensated Passive Divider Probes
Features Very low resistive loading, accurate amplitude measurements,
large dynamic range, and low cost.
Tradeoffs Capacitive loading <10 pF, lower bandwidth than active or
50-ohm resistive divider probes.
Applications General purpose probing, probing high-impedance nodes
(Š10 Kohm), op amp probing, CMOS probing (if bandwidth is
adequate), TTL probing (if bandwidth is adequate).
The compensated passive divider probe is the most common type of
scope probe. The 9-Mohm resistor in the tip forms a 10:1 voltage
divider with the 1-Mohm input resistance of the scope.
To have a flat frequency response, the probe tip capacitance is
compensated by the probe's cable capacitance, a compensating
capacitor, and the scope input capacitance. The compensating
capacitor is adjustable so you can optimize the step response for
flatness.
Example Rise Time
Calculation
Not all 9-Mohm divider probes work with all 1-Mohm scope inputs. The
probe data sheet shows the range of scope input capacitance it can
accommodate. You must make sure that the input capacitance of the
scope is within that range.
Given the following circuit using an HP 1160A probe,
26
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
the input resistance is:
= R
R
in
= 9 Mohm + 1 Mohm = 10 Mohm
R
in
probe
+ R
scope
The capacitive load is:
= C
C
in
[C
probe+Ccable+Ccomp+Cscope
tip
+ {[C
probe(Ccable+Ccomp+Cscope
]}
)]/
This number is calculated for the scope and scope probe combination,
and is shown in the Table of Compatible Probes.
The total resistance is: R
= 10 Mohm(50ohm)/(10 Mohm + 50ohm) = 50 ohm
R
Tot al
Total
= (RinR
The rise time due to circuit loading is: t
source
RC
)/(Rin + R
= 2.2R
TotalCin
source
)
tRC = 2.2(50 ohm)(7.5 pF) = 825 ps
From the Table of Compatible Probes, the bandwidth of the scope
probe system is 500 MHz. Therefore, the rise time of the scope probe
system is:
= 0.35/SystemBW
t
sys
= 0.35/500 MHz = 700 ps
t
sys
The measured rise time is:
= [(t
t
meas
t
= [(2 ns)2 + (825 ps)2 + (700 ps)2]
meas
)2 + (tRC)2 + (t
act
sys
)2]
1/2
1/2
= 2.27 ns
Remember that probe input impedance for compensated passive
divider probes is complex. A simple RC network serves only as a firstorder approximation.
Current Probes
Features Measures both ac and dc currents on a scope, with minimal
circuit loading.
Tradeoffs Large size.
Applications Power measurements, automotive measurements,
industrial measurements, motors, dynamoes, and alternators.
Scopes are designed to measure voltage, but by using a current probe
27
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
you can measure current. A current probe measures current in a wire
by enclosing the wire. Therefore, no electrical connection is needed.
Current probes generally use one of two technologies. The simplest
uses the principle of a transformer, with one winding of the
transformer being the measured wire. Because transformers only work
with alternating voltages and currents, current probes of this type
cannot measure direct current.
The other type of current probe uses the Hall effect principle. The Hall
effect produces an electric field in response to an applied magnetic
field. While this technique requires a power supply, it measures both
alternating and direct current.
Differential Probes
Features High common mode rejection ratio, easy viewing of small
waveforms with large dc offsets, more accurate than subtracting one
channel from another.
Tradeoffs Bigger than a passive probe, high cost, requires power, and
lower bandwidth than other probes.
Applications Measuring waveforms not referenced to the scope ground,
troubleshooting power supplies, and differential amplifier probing.
A differential probe is a high-impedance differential amplifier with two
probe tips; a non-inverting input and an inverting input. These two
inputs feed a differential amplifier which in turn drives the 50-ohm
input of the scope. The main advantage of differential probes is their
ability to reject waveforms that are common to both inputs. This type
of probe is often used in floating ground applications.
You could duplicate a differential probe by using two passive probes
and subtracting the two scope channels. However, the electrical paths
of the differential probe are carefully matched to give a high common
mode rejection ratio (CMRR). The higher the CMRR, the smaller the
waveforms you can view in the presence of unwanted noise.
Resistive Divider Probes
Features Highest bandwidth, lowest capacitive load, lower cost than
28
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
active probes, flat pulse response, good timing measurement accuracy.
Tradeoffs Relatively heavy resistive loading.
Applications ECL probing, GaAs probing, and transmission line
probing.
Resistive divider probes are designed for scopes with a 50-ohm input
impedance. The probe tips of the HP 1163A or HP 54006A have either
a 450-ohm or 950-ohm series resistor. The probe cable is a 50-ohm
transmission line. Because the cable is terminated in 50 ohms at the
scope input, it looks like a purely resistive 50-ohm load when viewed
from the probe tip. Therefore, the resistive divider probe is flat over a
wide range of frequencies, limited primarily by the parasitic
capacitance and inductance of the 450-ohm or 950-ohm resistor and
the fixture that holds it. The resistive load of the probe to your circuit
is either 500 ohm or 1 kohm, depending on the probe.
This type of probe has the smallest capacitive load of any probe. The
small capacitance and wide bandwidth make this probe type a good
choice for wide bandwidth measurements or time-critical
measurements.
Example Rise Time
Calculation
Given the following circuit using the HP 1163A probe,
the input resistance is:
R
= R
in
R
in
+ R
tip
= 450 ohm + 50 ohm = 500 ohm
scope
The total resistance is:
R
Tot al
= (RinR
source
)/(Rin + R
source
)
29
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Probing
R
= 500 ohm(50ohm)/(500 ohm + 50ohm) = 45 ohm
Tot al
The rise time due to circuit loading is:
= 2.2R
t
RC
Tot alCtip
tRC = 2.2(45 ohm)(1.5 pF) = 165 ps
From the Table of Compatible Probes, the bandwidth of the scope
probe system is 1.5 GHz. Therefore, the rise time of the scope probe
system is:
= 0.35/SystemBW
t
sys
= 0.35/1.5 GHz = 230 ps
t
sys
The measured rise time is:
= [(t
t
meas
t
= [(2 ns)2 + (165 ps)2 + (230 ps)2]
meas
)2 + (tRC)2 + (t
act
sys
)2]
1/2
1/2
= 2.02 ns
Surface Mount Probing
The HP 10467A 0.5 mm MicroGrabber Accessory Kit is designed for
using the HP 116x family of probes when you are probing fine-pitch
(0.5 mm to 0.8 mm) SMT (Surface Mount Technology) devices. The kit
contains enough parts for two probes.
The HP 116x probe tip plugs into the single-lead end of the dual-lead
adapter. The MicroGrabber connects to the red lead. You can also use a
MicroGrabber on the black lead, which you should connect to your
circuit’s ground. You can also connect the dual-lead end to circuit pins
that are 0.635 mm (0.025 inch) in diameter.
The kit is intended for use with voltages no greater than ±40 V (dc and
ac peak).
30
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Acquiring a Waveform
Acquiring a Waveform
The two ways to acquire a waveform with the oscilloscope are
Autoscale and Run. When you use Run, you can modify settings to
fine-tune your measurement.
You can also save acquired waveforms using waveform memories. (see
page 78)
Autoscale
Autoscale automatically adjusts volts per division and offset so that the
waveform fits into the display. It also attempts to set the seconds per
division so that three periods of the waveform are displayed.
Specifying a Measurement
To set up a measurement, first specify the channel setup then the
trigger. Based on your waveform you may need to change the offset
and scale to get accurate measurements.
A faster way to set up your measurement is to first autoscale, then
adjust only the settings you are interested in.
Running
The default acquisition mode is single-shot. To take another acquisition
immediately after the first one, select Run with the right mouse
button, then choose Repetitive from the menu. The Run button
becomes Run(r).
The scope does not support any modes other than real-time mode. You
can turn on averaging or accumulate under the Display tab. However,
because of the way the oscilloscope samples, this is not the same as the
equivalent time mode of a stand-alone oscilloscope.
31
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Acquiring a Waveform
NOTE:Clicking Run in an instrument window only runs that instrument. To run all
active instruments, select Run All in the System or Workspace window, or
Group Run in the window of any instrument included in a group run. If the
scope is triggered by another instrument, do not change settings while the
scope is waiting for its trigger or it may not trigger.
See Also“Run/Group Run Function” on page 92
“Autoscale” on page 32
“Specifying a Measurement” on page 33
“Using Waveform Memories” on page 78
“Differences from a Standard Digitizing Oscilloscope” on page 77
“Combining the Oscilloscope with a Logic Analyzer” on page 36
Autoscale
Autoscale automatically optimizes the waveform display for each
channel that is turned on. It sets volts per division and offset so that
the waveform fits into the middle of the display, and adjusts the
timebase (horizontal axis) to show three periods. When signals have
different periods, the signal on the lowest-numbered channel is used to
set the horizontal scale. If none of the signals show activity, the
timebase is set to 200 ns per division.
Autoscale also changes the trigger settings. The trigger channel is
channel 1 unless the signal on channel 1 has no detectable voltage
change. Triggering is limited to channels 1 and 2; no higher-numbered
channels can be set, but they will be autoscaled. The trigger mode is
set to the first rising edge and autotriggered. The trigger level is set to
the 60% threshold of the signal. If both channel 1 and channel 2 have
flat signals, the trigger source is set to channel 2 and the trigger level is
set to channel 2’s offset.
The settings changed by autoscale are:
32
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Acquiring a Waveform
Setting Default Algorithm
V/div 200 mV/div Waveform fits within the
middle
(Scale) 6 divisions of the display
Channel Offset 0 Waveform is centered
vertically
Sec/div 200 ns/div Fit three periods on
screen
(Scale)
Time offset 0 Always centers waveform
around trigger
(Delay)
Trigger mode rising edge Always sets trigger to
rising edge
Trigger sweep autotrigger Always sets trigger to
autotrigger
Trigger level not applicable Always sets level to
near 60% threshold
if a non-constant signal
is detected
Trigger occurrence 1 Always sets occurrence
to 1
Trigger source channel 1 Checks channel 1 for an
active signal;
if signal is flat, sets
to channel 2
Specifying a Measurement
1. Connect probes. (see page 14)
2. Set up the channel. (see page 34)
3. Set the display mode. (see page 60)
4. Specify trigger. (see page 39)
33
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Acquiring a Waveform
5. Right-click Run, and specify single or repetitive.
6. Click Run to start the acquisition.
7. Save particular waveforms to waveform memory. (see page 78)
The data is automatically displayed in the oscilloscope window. You can
also connect it to a display tool in order to correlate the oscilloscope
with a logic analyzer.
See Also“Combining the Oscilloscope with a Logic Analyzer” on page 36
“Probing” on page 14
“Channel Setup Window” on page 34
“Display Setup Window” on page 60
“Triggering” on page 39
“Using Waveform Memories” on page 78
Channel Setup Window
To access the Channel Setup window, click Setup... under the
Channels tab. Use this window to specify your probe type and probe
impedance. After the initial setup, you may want to use this window to
adjust channel skew.
On/Off Use this button to turn the channel on or off. You can also do
this from the main oscilloscope window.
Name Channel names can be a maximum of 10 characters long.
Customized names appear anywhere the channel is labeled.
Probe The probe attenuation factor. The arrow keys scroll through the
standard probe attenuation values, or you can enter non-standard
values by typing in the field. Probe attenuation affects the display and
marker measurements.
Input Z / Coupling Probe input impedance. Incorrect impedance will
cause bad measurements. See the “Table of Compatible Probes” on
page 14 for suggested values.
34
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Acquiring a Waveform
Skew Adjust for channel-to-channel skew caused by differing
electrical path lengths of the probes. To deskew the channels for multiboard oscilloscopes, run Channel Skew in the calibration utility.
Preset Select from TTL, ECL, and User. TTL sets the scale to 1 V/div
and the offset to 2.50 V. ECL sets the scale to 250 mV/div and the offset
to -1.3 V. User defaults to TTL values, but if you change the Scale or
Offset settings, the Preset field changes to User.
Scale Scale affects the vertical axis of the waveform display. You can
change it through either the arrow buttons or by typing in the field. It is
the same scale field as in the main oscilloscope window.
Offset Offset moves the waveform vertically in the display window.
Parts of the waveform that go offscreen are clipped, which may affect
any automatic measurements you run. The offset field also appears in
the main oscilloscope window.
You can also display the Channel Setup window by pointing to a
channel in the grid and clicking the right mouse button. Select
Channels... from the menu.
See Also“Calibrating the Oscilloscope” on page 10
“Probing” on page 14
“Vertical and Horizontal Scaling” on page 53
35
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Combining the Oscilloscope with a Logic Analyzer
Combining the Oscilloscope with a Logic
Analyzer
If you want to make a measurement with a logic analyzer and an
oscilloscope, there are three cases:
•“Oscilloscope Triggers Logic Analyzer” on page 36
•“Logic Analyzer Triggers Oscilloscope” on page 37
•“Logic Analyzer and Oscilloscope Correlate Data” on page 37
See Also The Intermodule Window (see the HP 16600A/16700A Logic Analysis
System help volume) for a generic approach.
Oscilloscope Triggers Logic Analyzer
1. Click the Workspace icon.
2. In the workspace window, drag both instruments on to the workspace.
3. In the Correlation Error dialog that appears, select Group Run for the
scope, and Oscilloscope for the logic analyzer.
4. Select the purple Group Run or the green Run All button to start the
acquisition.
5. To view the waveforms together, open the display tool.
•For a Waveform display, right-click one of the labels and select Insert
before... or Insert after.... In the Label Dialog, click the label you want
to insert, then click Apply.
•For the other tools, the oscilloscope labels are already available.
36
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Combining the Oscilloscope with a Logic Analyzer
Logic Analyzer Triggers Oscilloscope
NOTE:When the logic analyzer triggers the oscilloscope, if you are changing
oscilloscope settings when the trigger occurs it may be missed. The message
bar and Run Status window show "Waiting for IMB Arm" when this occurs.
When this happens, click Stop and restart the acquisition.
1. Click the Workspace icon.
2. In the Workspace window, drag both instruments on to the workspace.
3. Connect both to the same display tool.
4. In the Correlation Error dialog that appears, select Group Run for the
logic analyzer and the logic analyzer description for the oscilloscope.
5. A Trigger Problem dialog box may appear. Select Trigger Immediate.
6. Select the purple Group Run or the green Run All button to start the
acquisition.
7. To view the waveforms together, open the display tool.
•For a Waveform display, right-click one of the labels and select Insert
before... or Insert after.... In the Label Dialog, click the label you want
to insert, then click Apply.
•For the other tools, the oscilloscope labels are already available.
Logic Analyzer and Oscilloscope Correlate
Data
1. Click the Workspace icon.
2. In the Workspace window, drag both instruments on to the workspace.
3. Connect both into same display tool.
4. In the Correlation Error dialog that appears, select Group Run for both
instruments.
37
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Combining the Oscilloscope with a Logic Analyzer
5. Select the purple Group Run or the green Run All button to start the
acquisition.
6. To view the waveforms together, open the display tool.
•For a Waveform display, right-click one of the labels and select Insert
before... or Insert after.... In the Label Dialog, click the label you want
to insert, then click Apply.
•For the other tools, the oscilloscope labels are already available.
38
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Trig ge ri ng
Triggering
The default trigger type is auto, which means the oscilloscope will
trigger after 100 milliseconds. The trigger appears in the center of the
acquisition. You can change where the trigger is in the data set by
using the Delay field.
To specify more complicated triggers, click Trigger... at the bottom of
the main oscilloscope window. This brings up the Trigger Setup
Window.
You can also display the Trigger Setup window by pointing to the
trigger marker and clicking the right mouse button. Select Trigge r...
from the menu.
The area to the right of the Trigger button indicates the current trigger.
It does not show details such as occurence count.
•“Trigger Concepts” on page 39
•“Edge Triggering” on page 41
•“Pattern Triggering” on page 42
•“Delayed Triggering” on page 43
•“Getting a Stable Trigger” on page 44
•“The Trigger Setup Window” on page 45
Trigger Concepts
Trigger B a s i cs
The scope trigger circuitry helps you locate the waveform you want to
view. There are several types of triggering, but the one that is used
most often is edge triggering. Edge triggering identifies a trigger
39
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Tr ig ge ri ng
condition by looking for the slope (rising or falling) and voltage level
(trigger level) on the source you select. The trigger source is restricted
to channel 1, channel 2, and the external trigger. If you have more
channels on your oscilloscope, they cannot be used as trigger sources.
This figure shows the trigger circuit diagram.
Your waveform enters the positive input to the trigger comparator
where it is compared to the trigger level voltage on the other input. The
trigger comparator has a rising edge and a falling edge output. When a
rising edge of your waveform crosses the trigger level, the rising edge
comparator output goes high and the falling edge output goes low.
When a falling edge of your waveform crosses the trigger level, the
rising edge output goes low and the falling edge output goes high. The
scope uses the output you have selected as the trigger output.
Aliasing and Triggering
While aliasing does not cause unstable triggering, it does make it
difficult to tell when the scope is triggered. An aliased waveform can
appear as a lower frequency waveform that drifts across the display. To
ensure that your waveform is not aliased, you should decrease the
horizontal scale to its minimum value (maximum sampling rate), then
increase it to view your waveform.
40
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Trig ge ri ng
Edge Triggering
Edge trigger is the default trigger setting. Edge mode sets the
oscilloscope to trigger on an edge. You can set the source, trigger level,
and slope in the oscilloscope main window. Clicking Trigg e r... brings
up the Trigger Setup window, which lets you set the number of edges.
You can also display the Trigger Setup window by pointing to the
trigger marker and clicking the right mouse button. Select Trigge r...
from the menu.
The oscilloscope identifies an edge trigger by looking for the specified
slope (rising edge or falling edge) of your waveform. Once the slope is
found, the oscilloscope will trigger when your waveform crosses the
trigger level.
If you set the source to External, the trigger level is fixed at -1.30 V.
NOTE:The oscilloscope always fills a certain amount of acquisition memory before
looking for a trigger. When counting edge occurrences, you may see more
edges before the trigger than the number you specified. This happens because
some edges were already in memory but are not included in the occurrence
count.
When you set the trigger level on your waveform, it is usually best to
set it to a voltage near the middle of your waveform. The middle range
is best because there may be ringing or noise at the high and low ends
which can cause false triggers.
When you adjust the arm level control, a horizontal dashed line with a
T on the right-hand side appears, showing you where the arm level is
with respect to your waveform. After a period of time the dashed line
will disappear. You can get the line back by adjusting the arm level
control again.
41
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Tr ig ge ri ng
Pattern Triggering
Pattern triggering is similar to the way that a logic analyzer captures
data. This mode is useful when you are looking for a particular set of
ones and zeros on a computer bus or control lines. You can use channel
1 and 2 and the external trigger to form the trigger pattern. Because
you can set the voltage level that determines a logic 1 or a logic 0, any
logic family that you are probing can be captured. Channels 3 through
8, available in multi-board oscilloscopes, cannot be used in the pattern.
You can display the Trigger Setup window by pointing to the trigger
marker and clicking the right mouse button. Select Trigger... from the
menu.
When Pattern
There are five ways you can use to further qualify the pattern that you
want to view. They are:
EnteredWhen the scope finds the pattern, it triggers on the edge of
the pulse that makes the pattern valid.
ExitedThe scope arms the trigger circuitry when it has found the
pattern and triggers on the edge of the pulse that ends the
pattern.
Present >The scope triggers when the pattern is found and is present
for greater than the time value that you specify.
Present <The scope triggers when the pattern is found and is present
for less than the time value that you specify.
Range >The scope triggers when the pattern is present within the
time range that you specify.
NOTE:For Present >, Present <, and Range >, the oscilloscope does not trigger until
the pattern is exited.
42
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Trig ge ri ng
1. Set up a pattern by clicking the X after each channel name.
X means the channel is not part of the pattern.
Low and High let you set the threshold voltages for channel 1 and channel
2.
2. Select the When Pattern that you want.
3. If you have selected Present >, Present <, or Range >, set the time values.
The minimum is 20 ns, and the maximum is 160 milliseconds.
4. Close the Trigger Setup dialog box.
The area to the right of Trigger.. . shows the pattern you set up.
Delayed Triggering
You can delay the trigger by setting the Delay field. The Delay field
changes the acquisiton delay. Acquisition delay is the amount of time
between the trigger event and the center of the acquisition. It is the
only way to change the pre-trigger and post-trigger amounts in the
HP 16533A or HP 16534A Digitizing Oscilloscope.
To Store Mostly PostTrigger Data
The value shown in the Delay field is the sum of the acquisition delay
and the display delay. The display delay is controlled by the scrollbar,
and indicates which portion of the acquisition is currently being
displayed.
1. Calculate 16,350 (about half the acquisition memory) divided by the
sample rate. You should get a value in seconds.
2. Enter that value in the delay field.
Use n for nanoseconds, u for microseconds, and m for milliseconds.
3. If the value is correct, the scrollbar will move to one end of its range and
the current signal will not cross the entire display.
4. Click Run. The scrollbar returns to the middle.
If you adjust the scrollbar before clicking Run, the oscilloscope treats the
value as a display delay only.
To store mostly pre-trigger data, calculate the same value and enter it
43
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Tr ig ge ri ng
as a negative number.
To Re-center the
Trigger in the
Acquisition
See Also“Vertical and Horizontal Scaling” on page 53
1. Drag the scrollbar to the center of the scroll area. There is a slight delay in
movement when the bar is at the center.
2. Set Delay at the bottom of the window to 0 seconds, and press enter.
The scrollbar may jump away from the center. Do not reset it.
3. Click Run to get a new acquisition.
Getting a Stable Trigger
For most waveforms, the easiest way for you to get a stable trigger is to
use Autoscale. Autoscale analyzes your waveform and sets the trigger
mode to edge and the vertical scale, horizontal scale, and trigger level
to best display your waveform.
Manual Triggering
While Autoscale is the easiest way to obtain a stable trigger, there are
times when you may need to set the trigger manually to capture more
complex waveforms. To stabilize these waveforms:
•Set the Trigger Level to the proper point on the waveform.
The proper point is usually somewhere around 50% to avoid possible
ringing and noise at the top and base voltages.
•Increase the sampling rate to avoid aliasing.
The sampling rate is controlled by the horizontal scale at the bottom of the
screen. The maximum sampling rates are 1 gigasample per second for the
HP 16533A and 2 gigasamples per second for the HP 16534A.
•Set the Trigger Sweep to Triggered for low-frequency waveforms.
The Trigger Sweep field is in the Trigger Setup dialog.
•Remove noise from your waveform.
44
Chapter 1: HP 16533/34A Digitizing Oscilloscope
You can display the Trigger Setup window by pointing to the trigger
marker and clicking the right mouse button. Select Trigger... from the
menu.
See Also“The Trigger Setup Window” on page 45
“Autoscale” on page 32
“Changing the Sample Rate” on page 55
The Trigger Setup Window
The Trigger Setup window is for setting up complex triggers. You
access it by clicking Tri g g e r... at the bottom of the main oscilloscope
window.
Trig ge ri ng
Edge
Pattern
The two selections that are always availabe in the window are Mode
and Sweep. Mode specifies the type of condition you want to trigger
on. Sweep indicates whether the oscilloscope should wait for the
condition (Tri ggered ) or trigger immediately if the condition doesn’t
show up in 100 milliseconds (Auto).
Edge mode sets the oscilloscope to trigger on an edge. You can specify
the source, trigger level, slope, and occurrence.
The oscilloscope identifies an edge trigger by looking for the specified
slope (rising edge or falling edge) of your waveform. Once the slope is
found, the oscilloscope will trigger when your waveform crosses the
trigger level.
If you set the source to External, the trigger level is fixed at -1.30 V.
Use pattern mode for triggering on glitches or unusually long pulses, or
for a trigger involving 2 channels.
45
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Tr ig ge ri ng
To Trigger on a Glitch
1. Click Trigger.. .
2. Set the mode to Pattern and the sweep to Triggered.
3. Specify the glitch source by setting it to high or low. An X means the
channel is not part of the pattern.
4. Click the button under When Pattern and choose Present <.
5. Set the duration field to less than your clock’s pulse width.
Immediate
Use immediate mode when the oscilloscope is triggered by another
instrument in the measurement, or to acquire data as soon as you click
Run. No other levels or settings may be specified for this mode.
You can also display the Trigger Setup window by pointing to the
trigger marker and clicking the right mouse button. Select Trigge r...
from the menu.
See Also“Edge Triggering” on page 41
“Pattern Triggering” on page 42
“Trigger Concepts” on page 39
46
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
Changing Waveform Display and Grid
•“Zooming In” on page 47
•“Changing the Persistence of the Waveform” on page 47
•“Viewing Noisy Waveforms with Averaging” on page 49
•“Changing Display Colors” on page 51
•“Changing the Grid” on page 51
Zooming In
To zoom in on a particular area of your waveform, move the mouse to
one corner of the area, press the left mouse button, drag the mouse to
the diagonally opposite corner, and release the button.
To undo zoom, right-click in the display area and select Undo Zoom.
You can also select Setup->Undo Zoom from the menu bar.
Zoom may change your vertical scale (V/div), offset value, horizontal
scale (timebase or s/div), and scrollbar position to match the current
section of the waveform as though you had acquired it in that state.
When the new settings exceed limits, the display change does not
occur. This is most likely to happen with extreme negative delays and
detailed vertical scaling (V/div). You may be able to zoom in if you
enclose a larger area in the zoom.
If you Run then Undo Zoom, the original settings will be restored, but
your waveform may look wrong. The gaps are due to clipping; the
HP 16533A or HP 16534A oscilloscopes treat clipped data by leaving it
at the top and bottom edges of the display.
Changing the Persistence of the Waveform
Normally, a waveform is displayed only for one acquisition. When the
47
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
next run occurs, the previous waveform is erased and the newly
acquired waveform is drawn on the display.
By using accumulate, you can see a visual history of a waveform’s
acquisitions over time. For example, you can see the accumulated
peak-to-peak noise of a waveform over time which may appear
significantly different than in only one acquisition. You can see timing
jitter, the variance of the waveform from the trigger event, by
accumulating acquisitions on the display. By using accumulate, viewing
a waveform’s extremes over time is much easier.
Waveform mode sets the amount of time a waveform sample appears
on the display. Automated measurements cannot be performed on
accumulated waveforms but will be performed on the most recent
waveform in acquisition memory. Waveform accumulation does not
occur beyond the display area boundary.
The HP 16533A or HP 16534A Digitizing Oscilloscope have three
waveform modes: Normal, Accumulate, and Average.
Normal
In the normal waveform mode, a waveform data point is displayed for
at least 10 ms or one trigger cycle then erased. If no further triggers
occur, the last acquisition is left on the display. This is the default
setting. Use this mode for the fastest display update rate.
Accumulate
Accumulate is most like infinite persistence. In the accumulate
waveform mode, a waveform sample point is displayed until settings
are changed. All sample points are shown at full intensity. Use
accumulate to measure jitter or eye diagrams, see a waveform’s
envelope, look for timing violations, and find infrequent events.
Average
When Averaging is enabled, the # Avgs control tells the oscilloscope
the number of waveforms you want to use in calculating the average
value for each sample point. The HP 16533A or HP 16534A
oscilloscopes can average from 2 to 512 waveform acquisitions but the
larger the number of acquisitions, the more time it will take to
48
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
accumulate all the waveforms you have requested.
NOTE:If you are using Accumulate or Average and you change the vertical or
horizontal scaling, position, offset, trigger source or level, zoom, or drag the
waveform then the display is redrawn and any accumulated waveforms are
cleared. Only the last acquisition is displayed. Set up markers and any
measurements before using accumulate or averaging. Adding markers or
clearing measurements later can erase acquired waveforms.
See Also“Viewing Noisy Waveforms with Averaging” on page 49
“Display Setup Window” on page 60
Viewing Noisy Waveforms with Averaging
The Waveform Average mode under Display tells the oscilloscope to
acquire waveforms from several acquisitions and average them all
together, point by point. The greater the number of averages, the less
impact each new waveform has on the composite averaged waveform.
The perceived display update rate is slowed down as the number of
averages is increased because the averaged waveform doesn’t change
as much.
Sometimes, a waveform consists of a signal along with some random or
asynchronous noise. By using Waveform Average, these noise sources
can average to zero over time while the underlying waveform is
preserved. This will improve the accuracy of waveform measurements
because measurements are made on a more stable waveform and
measurement variances are reduced. The effective resolution of the
displayed waveform also improves as more acquisitions are averaged
together, providing the input waveform is repetitive and has a stable
trigger point.
Incidentally, if Waveform Average is enabled but the scope is not
properly triggering (perhaps the scope is set to Auto trigger and the
wrong trigger channel is selected), you may not see the waveform you
expect on the display. In this case, the input waveform is asynchronous
to the scope and will average to zero over time even though a non-zero
input waveform is being measured.
49
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
When Waveform Average is enabled, the # Avgs control sets the
number of waveforms you want to use in calculating the average value
for each sample point. The HP 16533A or HP 16534A can average from
2 to 512 waveform acquisitions but the larger the number of
acquisitions, the more time it will take to accumulate all the waveforms
you have requested.
The following formula is used to calculate the average for each data
point:
For n between 1 and M. After terminal count is reached (n greater than
or equal to M),
where:
= the average sample value
Ave
n
n = the current average number
M = setting of # Avgs control (terminal count)
= the ith sample.
S
i
See Also“Changing the Persistence of the Waveform” on page 47
“Trigger Concepts” on page 39
50
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
“Getting a Stable Trigger” on page 44
Changing Display Colors
The display colors which indicate channels, memories, and markers are
editable. These colors are also used by the display tools in the rest of
the logic analysis system.
To Change Colors
1. In the menu bar, select Setup.
2. Select Display...
The Display Setup dialog appears.
3. Under Colors, select the channel to modify.
4. Click the color you want it to be.
5. Click Edit Colors... to change a color's value.
NOTE:If you Close the Color Edit box, the new color values will be used in this
session only. If you Apply the color values, they will be used in this session and following sessions. To restore the factory colors, click Reset Defaults.
Changing the Grid
The HP 16533A or HP 16534A Digitizing Oscilloscope has a 10 by 8
display graticule grid which you can turn on or off. When on, a grid line
is place on each vertical and horizontal division. When the grid is off, a
frame with tic marks surrounds the graticule edges.
You can dim the grid’s intensity or turn the grid off to better view
waveforms which the graticule lines might obscure. Otherwise, you can
use the grid to estimate waveform measurements such as amplitude
and period. The grid intensity control doesn’t affect printing. You must
explicitly turn the grid off to remove the grid from a hardcopy.
51
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing Waveform Display and Grid
1. In the menu bar, select Setup.
2. Select Display...
The Display Setup dialog appears.
3. Click the Grid Type button to change the grid to axes-only scales, frameonly scale, or a background grid. The intensity field controls the
brightness. You cannot change the grid color.
52
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Vertical and Horizontal Scaling
Vertical and Horizontal Scaling
The vertical scale is volts per division (V/div). Changing the vertical
scale affects the height of the waveform. Extreme changes to the
vertical scale can affect your offset values. If the waveform extends
beyond the top or bottom of the display, data will be clipped. You
cannot measure clipped data, and when you adjust the offset or vertical
scale, clipped data stays at the top or bottom edge with a break in the
waveform.
The horizontal scale is seconds per division (s/div). Changing the
horizontal scale compresses and expands a waveform, and changes the
sampling rate. The automatic measurements only measure what is
currently shown in the display window, however.
Compressing the waveform may cause your sample rate to slow down.
Similarly, expanding your waveform may cause your sample rate to
increase, up to 1 gigasample per second for the HP 16533A or 2
gigasamples per second for the HP 16534A. See the table in “Changing
the Sample Rate” on page 55 for timebase and sampling rates.
The vertical (V/div) scale control is located under the Channels tab.
The horizontal (s/div) scale control is located at the bottom left corner
of the oscilloscope window.
Scrolling
The scrollbar below the display indicates what portion of the current
data set you are viewing. Its size shows the percentage of the data you
are looking at, and its location indicates the location of the data within
the data set.
You can scroll through your data set by dragging the scrollbar. You can
also use the Delay field, but this may change your acquisition delay as
described in “Delayed Triggering” on page 43.
To scroll short distances, drag the waveforms or trigger reference
marker. Individual waveforms can also be dragged vertically. Dragging
waveforms does change the delay and offset fields and will affect your
next acquisition.
53
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Vertical and Horizontal Scaling
When you click Run after having moved the scrollbar, the display
shows the same section of the data set that you were viewing before.
For example, if you had the scrollbar at the right end, you were viewing
the last part of the data set. When you click Run, the oscilloscope
acquires more data and again displays the last portion.
Sometimes you may not be able to move the scrollbar through the
entire scrolling area. This is because you have increased the sample
rate. The scrolling area indicates the size of the next acquisition, but
you can only move the scrollbar through the area filled with the current
data set. You can use the Delay arrows to move the scrollbar past its
dragging limits.
See Also“Delayed Triggering” on page 43
“Changing the Sample Rate” on page 55
“Channel Setup Window” on page 34
54
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing the Sample Rate
Changing the Sample Rate
The s/div scale controls the sample rate. The relationship is shown in a
table at the end of this topic.
The sample rate is displayed in the bottom left corner of the display
area. The maximum sample rate is 1 gigasamples per second for the
HP 16533A and 2 gigasamples per second for the HP 16534A. The
minimum sample rate is 500 samples per second.
Aliasing
Aliasing occurs when the sample rate is not at least four times as fast as
the high frequencies of your waveform. If you cannot see why the
oscilloscope triggered, or if the waveform moves around on screen, or if
the waveform looks slower than it should, suspect aliasing. To increase
your sample rate, set the s/div scale to a higher number, then run again.
See Also“Trigger Concepts” on page 39
s/div Sample Rate
< 200 ns 2 GSa/s
500 ns1 GSa/s
1 us500 MSa/s
2 us250 MSa/s
5 us100 MSa/s
10 us50 MSa/s
20 us25 MSa/s
50 us10 MSa/s
100 us5 MSa/s
200 us2.5 MSa/s
500 us1 MSa/s
1 ms500 KSa/s
55
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Changing the Sample Rate
2 ms250 KSa/s
5 ms - 20 ms100 KSa/s
50 ms50 KSa/s
100 ms25 KSa/s
200 ms10 KSa/s
500 ms5 KSa/s
1 s2.5 KSa/s
2 s1 KSa/s
5 s500 Sa/s
56
Comparing Channels
The HP 16533A or HP 16534A Digitizing Oscilloscope do not support
waveform math (A+B or A-B). However, you can easily overlay
waveforms by putting the 0 V indicators () on top of each other, and
making sure the waveforms have the same scale. The automatic
measurements are done on only one waveform, however.
Using waveform memories, you can also compare a waveform from a
previous acquisition to the current display. The waveform must be
loaded into memory when it is captured. The captured waveform can
be displayed either with the current scale settings or with the ones
used when it was captured.
See Also“Using Markers” on page 75
“Using Waveform Memories” on page 78
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Comparing Channels
57
Chapter 1: HP 16533/34A Digitizing Oscilloscope
What Do the Display Symbols Mean?
What Do the Display Symbols Mean?
All the indicators around the edge of the grid are draggable.
Local voltage marker. The color indicates which channel it is
measuring.
Local time marker. Time markers are channel independent.
Trigger event indicator.
Global marker. The global markers measure time and retain their
position within the total acquisition of all instruments in a Group Run.
58
Trigger level indicator. The color indicates which channel it is set on.
0 V (ground) indicator. The color indicates which channel it is set on.
The 0 V indicator is controlled by the offset setting. When offset is
negative, the 0 V indicator is above the center line. When offset is
positive, the 0 V indicator is below the center line. Also referred to as
offset indicator.
Offscreen indicator. The color indicates which channel it is set on. The
offscreen indicator appears when the 0 V indicator moves offscreen.
See Also“Using Markers” on page 75
Chapter 1: HP 16533/34A Digitizing Oscilloscope
What Do the Display Symbols Mean?
59
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Display Setup Window
Display Setup Window
The settings under Display control how waveforms are displayed. Only
"Acquisition Memory to Display" affects acquisitions.
Waveform Mode
Normal
Accumulate draws subsequent waveforms in the same area, without
is the default setting. It shows the current acquisition only.
erasing previous waveforms. The accumulated waveforms are erased if
any settings are changed, however.
Average averages the current acquisition with the specified number of
prior acquisitions. All acquisitions are equally weighted. The averaged
waveforms are replaced with the current acquisition if any settings are
changed.
Acquisition Memory to Display
For lower time resolutions, these settings allow you to optimize the
oscilloscope for greater detail or longer duration. All gives greater
detail by sampling more frequently. Partial stretches the acquisition
memory over a longer duration by slowing down the sample rate. These
settings do not make a difference when the horizontal scale is finer
than 1.00 microsecond/division for an HP 16533A, or 500 nanoseconds/
division for an HP 16534A.
Setup...
The Setup... button opens the Display Setup window. This window
contains the same controls under the Display tab, and also lets you
change the graticule and waveform colors.
Clear Display
The Clear Display button removes all channels from the graticule. It
does not affect waveform memories or markers.
See Also“Changing the Persistence of the Waveform” on page 47
60
Chapter 1: HP 16533/34A Digitizing Oscilloscope
“Changing Display Colors” on page 51
“Changing the Grid” on page 51
Display Setup Window
61
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
Automatic Measurements and Algorithms
Automatic measurements are simpler and usually more accurate to
make than the corresponding measurement done manually. (Manual
measurements and simple statistics are done using markers; see “Using
Markers” on page 75.) Blank values in the automatic measurement field
means that requirements for that measurement were not met. For
specific measurements, see the list below.
•“How the Scope Makes Measurements” on page 63
•“Average Voltage (Vavg)” on page 64
•“Base Voltage (Vbase)” on page 67
•“Fall Time” on page 65
•“Frequency” on page 66
•“Maximum Voltage (Vmax)” on page 70
•“Minimum Voltage (Vmin)” on page 70
•“Negative and Positive Pulse Width (±Width)” on page 65
•“Overshoot” on page 69
•“Peak-to-Peak Voltage (Vpp)” on page 69
•“Period” on page 64
•“Preshoot” on page 68
•“Rise Time” on page 64
•“Time of Maximum Voltage (Tmax)” on page 71
•“Time of Minimum Voltage (Tmin)” on page 71
•“Top Voltage (Vtop)” on page 67
•“Voltage Amplitude (Vamp)” on page 71
•“Vdcrms (Root Mean Square Voltage, DC)” on page 71
See Also “About the Measurements” on page 72
62
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
“Increasing the Accuracy of Your Measurements” on page 74
“Autoscale” on page 32
How the Scope Makes Measurements
Automatic parametric measurements are calculated from a histogram.
Measurements are done as soon as valid data is available. The absolute
minimum and maximum are derived from the histogram.
Next, the statistical top and base values are calculated. The top 40% of
the histogram is scanned for the top value and the bottom 40% is
scanned for the base value. The center 20% of the histogram is not
scanned to prevent selecting the middle of a tri-state waveform.
The measurement algorithm decides whether the absolute maximum
and minimum values should be used, as in the case of triangle
waveforms, or the statistical top and base should be used, as in the case
of square waveforms.
After the top and base are calculated, the IEEE 10%, 50%, and 90%
thresholds are calculated. These thresholds determine edges and are
used by all timing measurements. For example, rise time is measured
from the lower threshold to the 90% threshold of a rising edge. Period,
frequency, and pulse width measurements use the 50% threshold.
Once the thresholds have been calculated, the edges can be
determined. A rising edge is defined as a transition that passes through
the 10%, 50%, and 90% threshold levels. A falling edge is defined as a
transition that passes through the 90%, 50%, and 10% threshold levels.
For an edge to be detected, it must complete the transition through all
63
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
three threshold levels.
Once the oscilloscope locates rising and falling edges, it calculates rise
time, fall time, and frequency. If too few sample points fall along an
edge, the measurement is not made. The oscilloscope ignores
incomplete transitions.
Average Voltage (Vavg)
Vavg is the average voltage of waveform data over the display. It does
not include data that is offscreen. The value is calculated by summing
all the data points on the screen and dividing by the number of them.
Period
Period is defined as the time between the 50% threshold crossings of
two consecutive, like-polarity edges.
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the leftmost edge of the display.
Rise Time
Rise time is defined as the time at the 90% threshold minus the time at
the 10% threshold on the edge you are measuring.
64
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the leftmost edge of the display.
Fall Time
Fall time is defined as the time at the 10% threshold minus the time at
the 90% threshold on the edge you are measuring.
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the leftmost edge of the display.
Negative and Positive Pulse Width (±Width)
Negative pulse width is defined as the time from the 50% threshold of
the first falling edge to the 50% threshold of the next rising edge.
Positive pulse width is defined as the time from the 50% threshold of
the first rising edge to the 50% threshold of the next falling edge.
65
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
Negative Pulse Width
Positive Pulse Width
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the leftmost edge of the display.
Frequency
Frequency is defined as 1/Period. Period is the time between the 50%
threshold crossings of two consecutive, like-polarity edges.
66
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement on the first edge of the leftmost portion of the display.
Base Voltage (Vbase)
Vbase is the voltage of the statistical minimum level of the waveform
display, which is defined as the most frequently occurring voltage in
the histogram of the bottom 40% of the waveform.
Vbase may be equal to Vmin for many waveforms, such as triangle
waveforms. Similarly, Vtop may be equal to Vmax.
This measurement is position-independent and the entire display is
used for the measurement.
Top Voltage (Vtop)
Vtop is the voltage of the statistical maximum level of the waveform
display, which is defined as the most frequently occurring voltage in
67
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
the histogram of the top 40% of the waveform.
Vtop may be equal to Vmax for many waveforms, such as triangle
waveforms. Similarly, Vbase may be equal to Vmin.
This measurement is position-independent and the entire display is
used for the measurement.
Preshoot
Preshoot is a waveform distortion that precedes an edge transition.
If the edge is rising, preshoot will be 100*(base - local minimum)/ (top
- base). The local minimum is found half way from the 10% threshold
level to the 10% threshold level at the previous falling edge.
If the edge is falling, preshoot will be 100*(local maximum - top)/(top base). The local maximum is found half way from the 90% threshold
level to the previous 90% threshold level at the previous rising edge.
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the first edge of the leftmost portion of the display.
68
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
Overshoot
Overshoot is a waveform distortion that follows a major edge transition.
If the edge is rising, the overshoot will be 100*(local maximum - top)/
(top - base). The local maximum is found half way from the 90%
threshold level to the next 90% threshold level at the falling edge.
If the edge is falling, the overshoot will be 100*(base - local minimum)/
(top - base). The local minimum is found half way from the 10%
threshold level to the next 10% threshold level at the next rising edge.
The HP 16533A or HP 16534A Digitizing Oscilloscope starts the
measurement at the first edge on the leftmost portion of the display.
Peak-to-Peak Voltage (Vpp)
Peak-to-peak voltage is defined as Vmax - Vmin. Vmax is the absolute
maximum voltage of the display. Vmin is the absolute minimum voltage
of the display.
69
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
This measurement is position-independent and the entire display is
used for the measurement.
Minimum Voltage (Vmin)
Vmin is the absolute minimum voltage of the waveform display.
This measurement is position-independent and the entire display is
used for the measurement.
Maximum Voltage (Vmax)
Vmax is the absolute maximum voltage of the waveform display.
This measurement is position-independent and the entire display is
used for the measurement.
70
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
Time of Minimum Voltage (Tmin)
Tmin is the first time that the minimum voltage occurs on the display.
This measurement is position-independent and the entire display is
used for the measurement.
Time of Maximum Voltage (Tmax)
Tmax is the first time that the maximum voltage occurs on the display.
This measurement is position-independent and the entire display is
used for the measurement.
Voltage Amplitude (Vamp)
Vamp is the amplitude of the waveform display. Vamp = Vtop - Vbase.
This measurement is position-independent and the entire display is
used for the measurement.
Vdcrms (Root Mean Square Voltage, DC)
Vdcrms is the root-mean-square voltage of the waveform. The equation
71
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
used is:
This measurement is position-independent and the entire display is
used for the measurement. Data points offscreen are not included.
About the Measurements
The HP 16533A or HP 16534A Digitizing Oscilloscope makes
measurements after every trigger event, always maintaining continuity
between the measurement results and the oscilloscope display. This
makes sure that no aberration in the waveform under observation is
missed.
If the waveform is clipped, the oscilloscope cannot make some
automatic measurements. These measurements will show clipped
where a value would normally appear. Other indicators you may see
are:
?Value is questionable. This can occur because the signal is
clipped, there are not enough points, or the amplitude is
too small.
<The result is less than or equal to the value shown. This can
occur when the waveform is clipped low or not enough
points are available.
>The result is greater than or equal to the value shown. This
can occur when the waveform is clipped high or not enough
points are available.
(blank)No value could be calculated. The most common reason is
missing edges, such as when the display shows less than a
full period of a waveform.
The HP 16533A or HP 16534A will interpolate sample points if
necessary to determine pulse parameters for automatic measurements.
Excessive interpolation can lead to jitter on measurements; if this
72
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
occurs you may have to increase the sample rate.
By default, the oscilloscope uses the IEEE thresholds of 10, 50, and 90
percent for pulse measurements. A rising or falling edge is only
recognized after passing through all three thresholds. These thresholds
appear on the example pulse waveform as shown below:
Period and Frequency Measurements
At least one full cycle of the waveform with at least two like edges must
be displayed for period and frequency measurements. Automatic
waveform measurements use a single pulse and may have significant
errors introduced by interpolation and trigger inaccuracies. The
leftmost cycle is used for making measurements.
Pulse Width Measurements
For either the -Width or +Width measurements, a complete pulse must
be displayed to make a valid measurement. Remember that an edge
must pass through all three thresholds to be recognized as an edge.
Therefore, it is important that the pulse be positioned so that both
pulse edges transition through all three thresholds and are displayed
on the screen. Pulse width is measured from the leftmost valid edge to
the next valid edge.
Rise Time, Fall Time, Preshoot, and Overshoot Measurements
The leading, rising edge of the waveform must be displayed for rise
time and rising edge preshoot and overshoot measurements. The
trailing, falling edge of the waveform must be displayed for fall time
and falling edge preshoot and overshoot measurements. The leftmost
edge is used for measurements.
Remember that an edge must pass through all three thresholds to be
73
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Automatic Measurements and Algorithms
recognized as an edge. Therefore, it is important that the pulse be
positioned so that all three thresholds are displayed on the screen. Rise
time, fall time, preshoot, and overshoot measurements will be more
accurate if you expand the edge of the waveform by choosing a faster
sweep speed. Expanding the waveform will provide more data points
on an edge, reduce interpolation, and thus provide a more accurate
measurement.
Increasing the Accuracy of Your Measurements
Things you can do to make your measurements more accurate:
•Deskew the oscilloscope channels.
To deskew the oscilloscope channels, perform the Channel Skew
procedure as part of calibrating the oscilloscope (see page 10).
•Use automatic measurements where possible.
•For positive and negative pulse width, make sure enough top and bottom
voltages are showing to accurately calculate the top and base voltages.
•When using markers, increase the sampling rate.
•Minimize the effect of DC errors on time measurements by
•measuring between identical edges (same slew rate, amplitude and
offset).
•using the same DC level to reference each endpoint of the interval.
•performing the measurement on the fastest-slewing portion of each
edge.
•making the waveform as large as possible.
•Calibrate (see page 10) your oscilloscope after it has warmed up.
See Also“Getting a Stable Trigger” on page 44
74
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Using Markers
Using Markers
The HP 16533A or HP 16534A Digitizing Oscilloscope has both local
and global markers. The local markers can only be used within the
oscilloscope window. The global markers retain their position across
the data sets of all instruments that are part of the same group run as
the oscilloscope.
The global markers are G1 and G2 and only measure time. The local
markers consist of four voltage markers, V1 - V4, and two time
markers, T1 and T2.
To access the markers, click the Markers tab then click Setup.... The
Marker Setup dialog appears, from which you can turn on any of the
markers by clicking on the appropriate button. From this dialog, you
can also change the channel of the voltage markers.
The information after this does not apply to automatic time markers.
Those are covered in a separate topic listed below.
When the markers are turned on, you can set their values by dragging
them. If a marker moves offscreen because of scrolling or from typing
in a value in the Marker Setup window, you can return it to the edge of
the display by clicking the arrow buttons in the Marker Setup window.
You can also place markers by pointing to the area you want the marker
on and clicking the right mouse button. Select Place Marker and the
marker you want from the menu. If your cursor is on a marker already,
select Markers... to display the Marker Setup dialog.
See Also“What Do the Display Symbols Mean?” on page 58
“About Automatic Time Markers” on page 76
Working with Global Markers in Correlated Displays (see the Markers help
volume)
75
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Using Markers
About Automatic Time Markers
Automatic time markers indicate the time at which a specified voltage
crossing occurs. For instance, if you scroll the display, an automatic
time marker defined for the third edge will automatically reposition
itself on the edge third from the left. A regular time marker would
remain placed on the waveform and scroll with it off the display.
When you turn on an automatic time marker, Min T2-T1, Max T2-T1,
and Mean T2-T1 appear at the bottom of the Markers area. You can
gather statistics as long as both time markers are on, and one is an
automatic marker. Changing any part of the display will clear statistics.
To turn on automatic time markers
1. Under Markers, click Setup...
2. Click the time marker.
3. Click Marker [OFF] to turn on the marker.
4. Click Automatic [OFF] to put it in automatic mode.
5. Click Define Automatic Marker...
If you did not put the marker in automatic mode, the area to the right of
T1 or T2 is a time from trigger setting.
6. Specify the automatic setting. Percentage level is based on the
automatically calculated top and base voltage.
Automatic time markers cannot be dragged when they are on their
specified edge. If the specified edge does not exist, they can be
dragged like regular time markers.
See Also“Top Voltage (Vtop)” on page 67
“Base Voltage (Vbase)” on page 67
“How the Scope Makes Measurements” on page 63
76
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Differences from a Standard Digitizing Oscilloscope
Differences from a Standard Digitizing
Oscilloscope
There are some differences between an HP 16533A or HP 16534A
Digitizing Oscilloscope and a standard stand-alone digitizing
oscilloscope.
•Clipped data (data that is offscreen at acquisition) stays at the top or
bottom edge of the display area when the clipped waveform is resized or
has the offset moved. The waveform appears to have a gap in it where the
clipped data was.
•Repetitive run is repetitive real-time acquisitions, not what is also referred
to as equivalent-time mode.
•When there are not enough data points to map at least one sample to each
column of pixels in the display, the sin(x)/x interpolation filter is on. You
cannot turn it off.
•There is only one graticule area. All oscilloscope channels are sampled and
displayed at the same rate.
•The HP 16533A or HP 16534A oscilloscope is easier to use in conjunction
with a logic analyzer because arming is handled by the HP 16600A-series
or HP 16700A frame. You do not need to connect any wires between the
two modules.
77
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Using Waveform Memories
Using Waveform Memories
The four waveform memories store copies of waveforms for display
within the oscilloscope tool. The memories cannot be exported to other
display tools.
To Use Waveform Memory
This procedure assumes you already have acquired a waveform that
you want to store.
1. Under Memories, clickSetup...
2. In the Load Waveform area, select the waveform source.
The button text depends on your setup, but the default is channel 1.
3. Click Load.
Sources can be any channel or another waveform memory. If the source
does not contain any data (for example, the channel was off during the last
acquisition), "Waveform data is not valid!" appears briefly at the top of the
Waveform Memory Setup window. If the display is set to accumulate or
average, only the last acquired waveform is loaded.
4. To view the memory, click the Off radio button.
5. To make the memory display independent of the main display controls,
click the box to the left of Horizontal. This enables the independent
horizontal scale controls.
Waveform memories are erased between sessions. They are not
cleared when the display is cleared. In all other ways, waveform
memories are treated like channels by the display functions.
78
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Loading and Saving Oscilloscope Configurations
Loading and Saving Oscilloscope
Configurations
Oscilloscope settings and data can be saved to a configuration file. You
can also save any tools connected to the oscilloscope. Later, you can
restore your data and settings by loading the configuration file into the
oscilloscope.
•Loading Configuration Files (see the HP 16600A/16700A Logic Analysis
System help volume)
•Saving Configuration Files (see the HP 16600A/16700A Logic Analysis
System help volume)
79
Chapter 1: HP 16533/34A Digitizing Oscilloscope
When Something Goes Wrong
When Something Goes Wrong
•“Error Messages” on page 80
•“Calibration Problems” on page 80
•“Triggering Problems” on page 80
•“Other Problems” on page 81
Error Messages
•No valid signals have been found. (see page 82)
•Incorrect Calibration Factors for this software revision. (see page 83)
•Waiting for IMB Arm. (see page 83)
Calibration Problems
Calibration is not possible because NV RAM is protected
Unprotect the NV RAM and try again.
Calibration Procedure did not complete successfully
The Calibration window shows which tests passed and failed. If any test
failed and all cables were correctly hooked up, you should contact your
Hewlett-Packard Service Center.
Triggering Problems
Scope loses trigger when changes made to offset
The trigger level may be too high or too low to reliably trigger before
autotrigger. Sometimes this over-sensitivity to offset is caused by too
80
Chapter 1: HP 16533/34A Digitizing Oscilloscope
When Something Goes Wrong
high an attenuation factor in the channel setup.
Low frequency waveform appears to have unstable trigger
See “Getting a Stable Trigger” on page 44.
Nothing happening
If the scope is being triggered by another instrument, and you were
changing some settings, the scope may have missed its trigger signal.
The other common causes when the trigger is set to Triggered are
•trigger level occurs outside normal signal
•the oscilloscope is waiting for a rare event
You can either stop the oscilloscope or set the trigger to Auto to look
for these conditions.
Other Problems
Autoscale failed to find a waveform
Check that either channel 1 or channel 2 is on. Autoscale autoscales all
channels, but can only trigger on channel 1 or 2.
Also check for correct channel setup. The wrong setup can attenuate a
proper signal into an apparent DC signal.
Waveform has gaps when offset is changed or V/div increased
If a waveform is clipped, when you move the offset the clipped portion
of the waveform will stay at the top or bottom of the display, with a
break in the rest of the waveform. This also happens if you increase V/
div so that more of the waveform fits in the display. Those portions that
were off the display during acquisition contain uncertain data and so
are not displayable.
To get good data, make the changes to your settings and run again.
81
Chapter 1: HP 16533/34A Digitizing Oscilloscope
When Something Goes Wrong
Trigger won’t return to center
The trigger delay is affected by both the scrollbar position and the
value in the delay field. To re-center the trigger in the acquisition,
1. Drag the scrollbar to the center of the scroll area. There is a slight delay in
movement when the bar is at the center.
2. Set Delay at the bottom of the window to 0 seconds, and press enter.
The scrollbar may jump away from the center. Do not reset it.
3. Click Run to get a new acquisition.
The trigger is now in the center of the acquisition as well as the display
window.
Scope keeps stopping
The HP 16533A or HP 16534A Digitizing Oscilloscope default to singleshot acquisition. To have the oscilloscope keep running, right-click the
Run button and select Repetitive. Then click Run.
NOTE:The HP 16533A or HP 16534A Digitizing Oscilloscope does not do equivalent-
time sampling. All runs in the repetitive run mode are single acquisitions.
Scope locked up
The logic analysis system may be busy. Wait a few minutes and try
again. If the problem persists, exit the session and cycle power on the frame.
If you can find a sequence of steps that always or frequently causes this
to happen, please contact the HP Sales Office to report this bug.
Error Message: No Valid Signals
This message only comes up when autoscale is run and the oscilloscope
is unable to detect any line activity. Possible causes are:
•No channels are turned on. Autoscale does not check channels that are
turned off.
•The oscilloscope board is damaged. Run calibration to check for signal
82
Chapter 1: HP 16533/34A Digitizing Oscilloscope
When Something Goes Wrong
detection.
Autoscale does detect flat, dc signals so that is not a cause.
Error Message: Incorrect Calibration Factors
This message appears when you have upgraded your oscilloscope
software, but not re-calibrated the HP 16533A or HP 16534A
oscilloscope.
See the Logic Analysis System Installation Guide or “Calibrating the
Oscilloscope” on page 10 for instructions on calibrating the
oscilloscope.
Status Message: Waiting for IMB Arm
This message appears when the oscilloscope is being triggered by
another instrument in the logic analysis system. The instrument that
will trigger the oscilloscope has not yet found its own trigger, and
therefore hasn't sent the IMB Arm signal to the oscilloscope.
If the other instrument has already triggered, perform the Self Test
(see page 83) on the oscilloscope. If any of the tests fail, contact your
HP Sales Office for service.
Performing the Self Tests. To verify that the oscilloscope hardware
is operational, run the Self Test utility. The Self Tst function of the logic
analysis system performs functional tests on both the system and any
installed modules.
NOTE:The operational accuracy calibration requires that the oscilloscope hardware
meets specifications. The self test only requires that the hardware function.
An oscilloscope can pass self-test and still fail calibration.
To Run the Self-Test Utility
1. If you have any work in progress, save it to a configuration file. (see the HP
16600A/16700A Logic Analysis System help volume)
2. From the system window, click the System Admin icon.
3. Click the Admin tab, then Self Test....
83
Chapter 1: HP 16533/34A Digitizing Oscilloscope
When Something Goes Wrong
The system closes all windows before starting up Self Test.
4. Click Master Frame.
If the module is in an expansion frame, click Expansion Frame.
5. Click the oscilloscope.
6. In the Self Test dialog box, click Te s t Al l .
You can also run individual tests by clicking on them. Tests that require
you to do something must be run this way.
If any test fails, contact your local Hewlett-Packard Sales Office or
Service Center for assistance.
84
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
Specifications and Characteristics
NOTE:Specifications are valid after a 30 minute warm-up period, and within 10 °C
from the firmware calibration temperature.
NOTE:Definition of Terms To understand the difference between specifications (see
page 89) and characteristics (see page 89), and what gets a calibration
procedure (see page 90) and what gets a function test (see page 90), refer to
appropriate links within this note.
•Specifications (see page 85)
•Operating Environment (see page 86)
•Characteristics (see page 86)
Specifications
Note: Specifications refer to the input to the BNC connector.
Bandwidth
HP 16533A dc to 250 MHz
HP 16534A dc to 500 MHz
dc offset accuracy ±(1% of offset + 2% of
full scale)
dc voltage measurement accuracy ±(1.25% of full scale
+ offset accuracy
+ 0.016 div)
Time interval measurement accuracy ±[(0.005% of *T) +
(2e-6 x delay setting)
at maximum sampling rate, on a + 100 ps]
single card, on a single acquisition
Trigger sensitivity from 10 mV/div to 10 V/div
dc to 50 MHz 0.25 div
50 MHz to 500 MHz 0.5 div
Trigger sensitivity at 4 mV/div
dc to 50 MHz 0.63 div
85
Operating
Environment
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
50 MHz to 500 MHz 1.25 div
Input resistance 1 Mohm ± 1%
50 ohm ± 1%
Power Requirements
All power supplies required for operating the
oscilloscope are
supplied through the backplane connector in the logic
analysis system.
Operating Environment Characteristics
The oscilloscope module's reliability is enhanced when
operating
the module within the following ranges:
- Indoor use only.
- Temperature: +20 degrees C to +35 degrees C
(+68 degrees F to +95 degrees F)
- Humidity: 20% to 80% noncondensing
Characteristics
General
Maximum sampling rate
HP 16533A 1 GSa/S
HP 16534A 2 GSa/S
Number of channels 2 to 8 channels using
the same timebase
and trigger.
Waveform record length 32768 points
Vertical (Voltage)
(characteristics refer to the input at the BNC connector)
Vertical sensitivity range 4 mV/div to 10 V/div
in 1:2:4 steps
86
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
Vertical resolution 8 bits over 4 vertical
divisions
Rise time (calculated from
bandwidth)
HP 16533A 1.4 ns
HP 16534A 700 ps
dc gain accuracy ±(1.25% of full scale
+ 0.08% per °C
difference from
calibration temperature)
dc offset range (1:1 probe)
Vertical sensitivity Offset range
4 mV/div - 100 mV/div ±2 V
100 mV/div - 400 mV/div ±10 V
400 mV/div - 2.5 V/div ±50 V
2.5 V/div - 10 V/div ±250 V
Probe attenuation Any ratio from 1:1 to
1000:1 factor
Channel-to-channel isolation (with channel sensitivities
equal)
dc to 50 MHz 40 dB
50 MHz to 250 MHz (HP 16533A) 30 dB
50 MHz to 500 MHz (HP 16534A) 30 dB
Maximum safe input voltage
1 Mohm ±250 Vdc + peak ac
(<10KHz), CAT I
50 ohm 5 Vrms, CAT I
Input coupling
1 Mohm ac, dc
50 ohm dc only
Input C approximately 7 pF
Number of channels: 2,4,6, or 8 simultaneous channels
using the same
trigger OR up to 10 channels with independent triggers
for each pair
of channels. Maximum of 20 channels with HP 16701A
expansion frame.
87
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
Horizontal (Time)
Timebase ranges 0.5 ns/div to 5 s/div
Timebase resolution 10 ps
Delay range, pre-trigger 81.8 s, 5 divisions
Delay range, post-trigger 2.5e3 seconds
Time interval measurement accuracy ±[(0.005% of *T) +
(2e-6 x delay setting)
for sampling rates other than + (0.15/sample rate)]
maximum, for bandwidth-limited
signals (signal rise time
> 1.4/sample rate) on a single card,
on a single acquisition
Time interval measurement accuracy ±[(0.005% of *T) +
(2e-6 x delay setting)
for 2, 3, or 4 cards operation on + 550 ps]
a single timebase, for measurements
made between channels on different
cards, at maximum sampling rate
Tr ig ge r
Trigger level range Within display window
(vertical offset
± 2 divisions)
Immediate trigger mode Triggers immediately
after arming
condition is met
Edge trigger mode Triggers on rising or
falling edge on
channel 1 or channel 2
Pattern trigger mode Triggers on entering
or exiting a
specified pattern across
both channels
Auto condition trigger mode Self-triggers if
trigger is not satisfied
within approximately
88
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
100 ms after arming
Events delay trigger mode The trigger can be set
to occur on the
nth occurrence of an
edge or pattern,
n <= 32000
Intermodule trigger mode Arms another
measurement module or acti vates a trigger output
on the rear panel
BNC connector when the
trigger condition
is met
What is a Specification
A Specification is a numeric value, or range of values, that bounds the
performance of a product parameter. The product warranty covers the
performance of parameters described by specifications. Products
shipped from the factory meet all specifications. Additionally, the
products sent to HP Customer Service Centers for calibration and
returned to the customer meet all specifications.
Specifications are verified by Calibration Procedures.
What is a Characteristic
Characteristics describe product performance that is useful in the
application of the product, but that is not covered by the product
warranty. Characteristics describe performance that is typical of the
majority of a given product, but not subject to the same rigor
associated with specifications.
Characteristics are verified by Function Tests.
89
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Specifications and Characteristics
What is a Calibration Procedure
Calibration procedures verify that products or systems operate within
the specifications. Parameters covered by specifications have a
corresponding calibration procedure. Calibration procedures include
both performance tests and system verification procedure. Calibration
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance, which is
referred to as an "operational accuracy calibration".
What is a Function Test
Function tests are quick tests designed to verify basic operation of a
product. Function tests include operator’s checks and operation
verification procedures. An operator’s check is normally a fast test
used to verify basic operation of a product. An operation verification
procedure verifies some, but not all, specifications, and often at a lower
confidence level than a calibration procedure.
90
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Help - How to Navigate Quickly
Help - How to Navigate Quickly
1. Place mouse cursor anywhere in a help window.
2. Press the right mouse button.
3. Select desired destination.
You can also access all navigation and search commands from the help
window menu bar.
91
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Run/Group Run Function
Run/Group Run Function
•Setting a tool for independent or Group Run (see page 93)
•Setting Single or Repetitive Run (see page 94)
•“Checking Run Status” on page 94
Understanding Run/Run All/Group Run
The Run/Run All/Group Run buttons initiate data capture in the
instrument tools you have configured. When an instrument tool is
connected to analysis or display tools, any of the tools can initiate a
run. When two or more instrument tools are configured, you can run
them independently or as a group. Two or more instruments running as
a group is called an Intermodule measurement.
Use the Intermodule Window (see the HP 16600A/16700A Logic Analysis System help volume) to coordinate the run function of
multiple instruments as a "Group Run". A common "Group Run"
configuration is to run the instrument tools at the same time. A more
advanced measurement is to configure one instrument to arm another
instrument, each with their own trigger conditions.
•Run appears in the setup dialog and icon menu of an instrument if it is not
part of an Intermodule measurement.
•Group Run appears in the setup dialog and icon menu of each tool if two or
more instruments are configured for an Intermodule measurement.
•Run All always appears in the System, Workspace and Run Status
windows, and initiates a run in all configured instruments, whether they
are run independently or are part of a Group Run.
Intermodule measurements are configured between individual
instruments. Arming between two machines that belong to one
analyzer is configured in the Arming Info... dialog found in the Tri gg e r
window of the analyzer.
Understanding Stop/Stop_All/Cancel
•Stop will terminate an individual instrument measurement that is running.
92
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Run/Group Run Function
(perhaps waiting for a trigger condition)
•Stop All, when selected from the Workspace, will terminate running
measurements from all instruments currently on the Workspace.
•Cancel will terminate the processing of trace data from an instrument to an
analysis or display tool connected to its output.
See Also “Demand Driven Data” on page 95
Setting a tool for independent or Group Run
Use the Intermodule Window to change between Group Run and
independent Run.
•Click the Intermodule icon in the System Window, OR
•Use Navigate->System->Intermodule
In the Intermodule window, move instruments between independent
Run and Group Run by clicking the icon and selecting the desired
arming source. All instruments in "Group Run" will run simultaneously.
93
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Run/Group Run Function
Setting Single or Repetitive Run
A single measurement will stop after memory is full or a store
qualification is met. A repetitive measurement executes successive
Single measurements until Stop is selected.
When a single or repetitive measurement is stopped, only data that has
been captured to that point is available for viewing.
Select single or repetitive by right-clicking on the Run button in the
tool’s setup window.
If you have problems displaying trace data when running Repetitive
measurements, see “Demand Driven Data” on page 95.
Checking Run Status
The Run Status dialog provides status information about the currently
configured instruments, and the status of the run with respect to the
trigger specification.
To access the Run Status dialog:
•The Run Status icon in the System Window, OR
•Navigate->System->Run Status
94
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Run/Group Run Function
Demand Driven Data
When an analyzer measurement occurs, acquisition memory is filled
with data that is then transferred to the display memory of the analysis
or display tools you are using, as needed by those tools. In normal use,
this demand driven data approach saves time by not transferring
unnecessary data.
Since acquisition memory is cleared at the beginning of a
measurement, stopping a run may create a discrepancy between
acquisition memory and the memory buffer of connected tools. Without
a complete trace of acquisition memory, the display memory will
appear to have ’holes’ in it which appear as filtered data.
This situation will occur in these cases:
•If you stop a repetitive measurement after analyzer data has been cleared
and before the measurement is complete.
•If a trigger is not found by the analyzer and the run must be stopped to
regain control.
To make sure all of the data in a repetitive run is available for viewing:
•In the workspace, attach a Filter tool to the output of the analyzer.
•In the Filter, select "Pass Matching Data"
•In the filter terms, assure the default pattern of all "Don't Cares" (Xs).
95
Chapter 1: HP 16533/34A Digitizing Oscilloscope
Run/Group Run Function
This configuration will always transfer all data from acquisition
memory. While this configuration will increase the time of each run, it
will guarantee that repetitive run data is available regardless of when it
is stopped.
96
Glossary
absolute Denotes the time period
or count of states between a captured
state and the trigger state. An
absolute count of -10 indicates the
state was captured ten states before
the trigger state was captured.
acquisition Denotes one complete
cycle of data gathering by a
measurement module. For example,
if you are using an analyzer with
128K memory depth, one complete
acquisition will capture and store
128K states in acquisition memory.
analysis probe A probe connected
to the target microprocessor. It
provides an interface between the
signals of the target microprocessor
and the inputs of the logic analyzer.
Also called a "preprocessor".
analyzer 1 In a logic analyzer with
two machines, refers to the machine
that is on by default. The default
name is Analyzer<N>, where N is
the slot letter.
analyzer 2 In a logic analyzer with
two machines, refers to the machine
that is off by default. The default
name is Analyzer<N2>, where N is
the slot letter.
instruments are armed immediately
when Run or Group Run is selected.
You can set up one instrument to arm
another using the Intermodule Window. In these setups, the second
instrument cannot search for its
trigger condition until it receives the
arming signal from the first
instrument. In some analyzer
instruments, you can set up one
analyzer machine to arm the other
analyzer machine in the Trigger Window.
asterisk (*) See edge terms,
glitch, and labels.
bits Bits represent the physical logic
analyzer channels. A bit is a channel
that has or can be assigned to a label.
A bit is also a position in a label.
card This refers to a single
instrument intended for use in the
HP 16600A-series or HP 16700A
mainframe. One card fills one slot in
the mainframe. A module may
comprise a single card or multiple
cards cabled together.
channel The entire signal path from
the probe tip, through the cable and
module, up to the label grouping.
arming An instrument tool must be
armed before it can search for its
trigger condition. Typically,
97
click To click an item, position the
cursor over the item. Then quickly
press and release the left mouse
Glossary
button.
clock channel A logic analyzer
channel that can be used to carry the
clock signal. When it is not needed
for clock signals, it can be used as a
data channel, except in the
HP 16517A.
context record A context record is
a small segment of analyzer memory
that stores an event of interest along
with the states that immediately
preceded it and the states that
immediately followed it.
context store If your analyzer can
perform context store
measurements, you will see a button
labeled Context Store under the
Trigger tab. Typical context store
measurements are used to capture
writes to a variable or calls to a
subroutine, along with the activity
preceding and following the events. A
context store measurement divides
analyzer memory into a series of
context records. If you have a 64K
analyzer memory and select a 16state context, the analyzer memory is
divided into 4K 16-state context
records. If you have a 64K analyzer
memory and select a 64-state
context, the analyzer memory will be
divided into 1K 64-state records.
count The count function records
periods of time or numbers of state
transactions between states stored in
memory. You can set up the analyzer
count function to count occurrences
of a selected event during the trace,
such as counting how many times a
variable is read between each of the
writes to the variable. The analyzer
can also be set up to count elapsed
time, such as counting the time spent
executing within a particular function
during a run of your target program.
cross triggering Using intermodule
capabilities to have measurement
modules trigger each other. For
example, you can have an external
instrument arm a logic analyzer,
which subsequently triggers an
oscilloscope when it finds the trigger
state.
data channel A channel that
carries data. Data channels cannot be
used to clock logic analyzers.
data field A data field in the pattern
generator is the data value associated
with a single label within a particular
data vector.
data set A data set is made up of all
labels and data stored in memory of
any single analyzer machine or
instrument tool. Multiple data sets
can be displayed together when
sourced into a single display tool. The
98
Glossary
Filter tool is used to pass on partial
data sets to analysis or display tools.
debug mode See monitor.
delay The delay function sets the
horizontal position of the waveform
on the screen for the oscilloscope and
timing analyzer. Delay time is
measured from the trigger point in
seconds or states.
demo mode An emulation control
session which is not connected to a
real target system. All windows can
be viewed, but the data displayed is
simulated. To start demo mode,
select Start User Session from the
Emulation Control Interface and
enter the demo name in the
Processor Probe LAN Name field.
Click Help in the Start User Session
window for details.
deskewing To cancel or nullify the
effects of differences between two
different internal delay paths for a
signal. Deskewing is normally done
by routing a single test signal to the
inputs of two different modules, then
adjusting the Intermodule Skew so
that both modules recognize the
signal at the same time.
don’t care For terms, a "don’t care"
means that the state of the signal
(high or low) is not relevant to the
measurement. The analyzer ignores
the state of this signal when
determining whether a match occurs
on an input label. "Don’t care" signals
are still sampled and their values can
be displayed with the rest of the data.
Don’t cares are represented by the X
character in numeric values and the
dot (.) in timing edge specifications.
dot (.) See edge terms, glitch,
labels, and don’t care.
double-click To double-click an
item, position the cursor over the
item, and then quickly press and
release the left mouse button twice.
drag and drop To drag and drop an
item, position the cursor over the
item, and then press and hold the left mouse button. While holding the left
mouse button down, move the mouse
to drag the item to a new location.
When the item is positioned where
you want it, release the mouse
button.
edge mode In an oscilloscope, this
is the trigger mode that causes a
trigger based on a single channel
edge, either rising or falling.
edge terms Logic analyzer trigger
resources that allow detection of
transitions on a signal. An edge term
can be set to detect a rising edge,
99
Glossary
falling edge, or either edge. Some
logic analyzers can also detect no
edge or a glitch on an input signal.
Edges are specified by selecting
arrows. The dot (.) ignores the bit.
The asterisk (*) specifies a glitch on
the bit.
emulation module A module
within the logic analysis system
mainframe that provides an
emulation connection to the debug
port of a microprocessor. An E5901A
emulation module is used with a
target interface module (TIM) or an
analysis probe. An E5901B emulation
module is used with an E5900A
emulation probe.
emulation probe The stand-alone
equivalent of an emulation module.
Most of the tasks which can be
performed using an emulation
module can also be performed using
an emulation probe connected to
your logic analysis system via a LAN.
emulator An emulation module or
an emulation probe.
Ethernet address See link-level
address.
events Events are the things you
are looking for in your target system.
In the logic analyzer interface, they
take a single line. Examples of events
are Label1 = XX and Timer 1 > 400 ns.
filter expression The filter
expression is the logical OR
combination of all of the filter terms.
States in your data that match the
filter expression can be filtered out or
passed through the Pattern Filter.
filter term A variable that you
define in order to specify which
states to filter out or pass through.
Filter terms are logically OR’ed
together to create the filter
expression.
Format The selections under the
logic analyzer Format tab tell the
logic analyzer what data you want to
collect, such as which channels
represent buses (labels) and what
logic threshold your signals use.
frame The HP 16600A-series or
HP 16700A logic analysis system
mainframe. See also logic analysis system.
gateway address An IP address
entered in integer dot notation. The
default gateway address is 0.0.0.0,
which allows all connections on the
local network or subnet. If
connections are to be made across
networks or subnets, this address
must be set to the address of the
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.