HP 11722A Sensor Module
HP 11792A Sensor Module
HP 11793A Microwave Converter
HP 11812A Verification Kit
he HP 8902A measuring receiver delivers the
T
accuracy and resolution of a high perfor-
mance power meter at frequencies from 150 kHz to
1.3 GHz (50 MHz to 26.5 GHz with the HP 11793A
microwave converter) and levels from +30 dBm to
–127 dBm. It accurately measures AM, FM and φM,
including residuals and incidentals, with a single
keystroke. The HP 8902A measuring receiver, with
the HP 11793A, counts RF signals to 26.5 GHz with
10 Hz resolution and excellent long-term frequency
stability. The HP 8902A measuring receiver with
Option 050 offers increased power measurement
accuracy. This option specifies Tuned RF Level on
the HP 8902A measuring receiver to an accuracy of
±(0.015 dB + 0.005 dB/10 dB step).
Specifications
HP 8902A Measuring Receiver*
Technical Specifications
Specifications describe the test set’s warranted performance
and are valid over the entire operation and environmental ranges
unless otherwise noted. All specifications are valid after a
30-minute warm up period of continuous operation, and within
the frequency ranges defined below.
Supplemental characteristics are intended to provide
additional information useful in applying the instrument by
giving typical, but non-warranted performance parameters.
These characteristics are shown in Italics and labeled as
“nominal,” “typical,” or “supplemental.”
AM REJECTION (50 Hz to 3 kHz BW)3:
AM RejectionFrequency RangeRatesAM Depths
<20 Hz peak
150 kHz – 1300 MHz400 Hz or 1 kHz≤50%
deviation
<20 Hz peak
150 kHz – 26.5 GHz400 Hz or 1 kHz≤50%
deviation
RESIDUAL FM (50 Hz to 3 kHz BW):
<8 Hz
frequency to <1 Hz
<17 Hz
<33 Hz
<49 Hz
<65 Hz
at 1300 MHz, decreasing linearly with
rms
1300 MHz <f
rms'
6.2 GHz <f
rms'
12.4 GHz <f
rms'
18.6 GHz <f
rms'
for 100 MHz and below.
rms
≤6.2 GHz.
c
≤12.4 GHz.
c
≤18.6 GHz.
c
≤26.5 GHz.
c
* Shaded text signifies measurements made with the HP 8902A
measuring receiver using the HP 11793A microwave converter
and HP 11792A sensor module.
With this configuration, all standard HP 8902A specifications
apply except where changes are shown as shaded text.
Frequency Modulation
RATES1:
20 Hz to 10 kHz, 150 kHz ≤ f
20 Hz to 200 kHz, 10 MHz ≤ f
20 Hz to 200 kHz, 10 MHz ≤ f
DEVIATIONS1:
40 kHz
400 kHz
400 kHz
ACCURACY
FM Accuracy
±2% of reading
±1 digit
±1% of reading
±1 digit
±5% of reading
±1 digit
±1% of reading
±1 digit
±5% of reading
±1 digit
maximum, 150 kHz ≤ f
peak
maximum, 10 MHz ≤ fc ≤1300 MHz.
peak
maximum, 10 MHz ≤ fc ≤26.5 GHz.
peak
1, 2, 3
:
Frequency Range
250 kHz – 10 MHz
10 MHz – 1300 MHz
10 MHz – 1300 MHz
10 MHz – 26.5 GHz50 Hz – 100 kHz≤400 kHz
10 MHz – 26.5 GHz
For rms detector add ±3% of reading.
DEMODULATED OUTPUT DISTORTION
THDFrequency RangeRatesDeviations
<0.1%400 kHz – 10 MHz20 Hz – 10 kHz<10 kHz
<0.1%10 MHz – 1300 MHz20 Hz – 100 kHz<100 kHz
<0.1%10 MHz – 26.5 GHz20 Hz – 100 kHz<100 kHz
<10 MHz.
c
≤1300 MHz.
c
≤26.5 GHz.
c
<10 MHz.
c
RatesDeviations
20 Hz – 10 kHz
50 Hz – 100 kHz
20 Hz – 200 kHz
≤40 kHz
≤400 kHz
≤400 kHz
20 Hz – 200 kHz≤400 kHz
1, 4
:
peak
peak
peak
peak
peak
Supplemental Characteristics:
MAXIMUM FM DEVIATION, RESOLUTION, AND
MAXIMUM DEMODULATED OUTPUT SENSITIVITY
ACROSS AN OPEN CIRCUIT (600W output impedance)
Maximum
Resolution
100 Hz
10 Hz
1 Hz1.0 mV/Hz
0.1 Hz
(rms detector only)
Maximum Demodulated
Output Sensitivity
0.01 mV/Hz
0.1 mV/Hz
Deviations
(
D
F)
∆
F
≥ 40 kHz
peak
4.0 kHz ≤
∆
F
<40 kHz
peak
∆
F
< 4 kHz
peak
∆
F
< 0.3 kHz1.0 mV/ Hz
rms
Resolution is increased one digit with 750 µs de-emphasis
and pre-display on.
The demodulated output signal present at the
Modulation Out/Audio In connector is increased in
But not to exceed: 20 kHz rates and 40 kHz peak deviations with 750 µs
de-emphasis filter.
2
Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with
rms detector.
3
Peak residuals must be accounted for in peak readings.
4
With 750 µs de-emphasis and pre-display "off," distortion is not specified for
modulation outputs >4V peak. This condition can occur near maximum deviation for a measurement range, at rates <2 kHz.
5
For optimum flatness, cables should be terminated with their
characteristic impedance.
5
:
2
Amplitude Modulation
RATES:
20 Hz to 10 kHz, 150 kHz ≤f
20 Hz to 100 kHz, 10 MHz ≤f
DEPTH: to 99%
ACCURACY
AM AccuracyFrequency RangeRatesDepths
±2% of reading
±1 digit
±3% of reading
±1 digit
±1% of reading
±1 digit
±3% of reading
±1 digit
±1.5% of reading
±1 digit
±3% of reading
±1 digit
2, 3, 6
:
150 kHz – 10 MHz
150 kHz – 10 MHz
10 MHz – 1300 MHz
10 MHz – 1300 MHz
1300 MHz – 26.5 GHz50 Hz – 50 kHz5% – 99%
10 MHz – 26.5 GHz20 Hz – 100 kHz to 99%
For rms detector add ±3%of reading.
FLATNESS
FlatnessFrequency RangeRatesDepths
±0.3% of reading
±1 digit
±0.3% of reading
±1 digit
5, 7
:
10 MHz – 1300 MHz90 Hz – 10 kHz20% – 80%
10 MHz – 26.5 GHz90 Hz – 10 kHz20% – 80%
<10 MHz.
c
≤1300 MHz.
c
50 Hz – 10 kHz5% – 99%
20 Hz – 10 kHz
50 Hz – 50 kHz
20 Hz – 100 kHz
to 99%
5% – 99%
to 99%
Phase Modulation
RATES:
200 Hz to 10 kHz, 150 kHz ≤f
200 Hz to 20 kHz, 10 MHz ≤f
200 Hz to 20 kHz, 10 MHz ≤f
ACCURACY3:
±4% of reading ±1 digit, 150 kHz ≤f
±3% of reading ±1 digit, 10 MHz ≤f
±3% of reading ±1 digit, 10 MHz ≤f
For rms detector add ±3% of reading.
DEMODULATED OUTPUT DISTORTION: <0.1% THD.
AM REJECTION (for 50% AM at 1 kHz rate)
<0.03 radians peak (50 Hz to 3 kHz BW).
MAXIMUM DEVIATION, RESOLUTION, AND MAXIMUM
DEMODULATED OUTPUT SENSITIVITY ACROSS AN
OPEN CIRCUIT (600 W output impedance)
<10 MHz.
c
≤1300 MHz.
c
≤26.5 GHz.
c
<10 MHz.
c
≤1300 MHz.
c
≤26.5 GHz.
c
150 kHz £fc <10 MHz
3
:
5
:
DEMODULATED OUTPUT DISTORTION:
<0.3% THD for ≤50% depth.
<0.6% THD for ≤95% depth.
>1300 MHz add 0.4% THD.
For f
c
FM REJECTION (50 Hz to 3 kHz BW)3:
FM RejectionFrequency RangeRatesDeviations
<0.2% AM250 kHz – 10 MHz400 Hz or 1 kHz<5 kHz
<0.2% AM10 MHz – 1300 MHz400 Hz or 1 kHz<50 kHz
<0.2% AM10 MHz – 26.5 GHz400 Hz or 1 kHz<50 kHz
RESIDUAL AM (50 Hz to 3 kHz BW): <0.01%
rms
peak
peak
peak
.
Supplemental Characteristics:
DETECTORS: +peak, –peak, ±peak/2, peak hold, average
(rms sinewave calibrated), rms.
MAXIMUM DEPTH, RESOLUTION, AND MAXIMUM
DEMODULATED OUTPUT SENSITIVITY ACROSS AN
OPEN CIRCUIT (600W output impedance)
Not to exceed for stated accuracy: 50 Hz to 40 kHz rates with
rms detector.
3
Peak residuals must be accounted for in peak readings.
5
For optimum flatness, cables should be terminated with their
characteristic impedance.
6
For peak measurements only: AM accuracy may be affected by distortion
generated by the measuring receiver. In the worst case this distortion can
decrease accuracy by 0.1% of reading for each 0.1% of distortion.
7
Flatness is the variation in indicated AM depth for constant depth on
input signal.
3
Modulation Reference
Internal Time Base Reference
AM CALIBRATOR DEPTH AND ACCURACY:
33.33% depth nominal, internally calibrated to an
accuracy of ±0.1%.
FM CALIBRATOR DEVIATION AND ACCURACY:
34 kHz
deviation nominal, internally calibrated
peak
to an accuracy of ±0.1%.
Supplemental Characteristics:
CARRIER FREQUENCY: 10.1 MHz.
MODULATION RATE: 10 kHz.
OUTPUT LEVEL: – 25 dBm.
Frequency Counter
RANGE:
150 kHz to 1300 MHz.
150 kHz to 26.5 GHz.
SENSITIVITY:
12 mV
22 mV
40 mV
71 mV
40 mV
MAXIMUM RESOLUTION:
1 Hz.
10 Hz.
ACCURACY:
± reference accuracy ± 3 counts of least-significant
digit, f
± reference accuracy ± 3 counts of least-significant
digit, or 30 Hz, whichever is larger, f
Supplemental Characteristics:
MODES: Frequency and Frequency Error (displays the
difference between the frequency entered via the keyboard
and the actual RF input frequency).
SENSITIVITY IN MANUAL TUNING MODE:
Approximate frequency must be entered from keyboard.
0.22 mV
0.71 mV
Using the RF amplifier and the IF amplifiers,
sensitivity can be increased to approximately:
–100 dBm.
–90 dBm, f
–75 dBm, 1300 MHz <f
8
After 30-day warm-up.
9
The HP 8902A fundamental RF power measurement units are watts. Further
internal processing is done on this number to display all other units.
10
When using a power sensor, the noise specification may mask the linearity
specification and become the predominant error. When operating on the top
RF power range, add the power sensor's linearity percentages found in the
power sensor's specifications.
(–25 dBm), 150 kHz ≤fc ≤650 MHz.
rms
(–20 dBm), 650 MHz ≤fc ≤1300 MHz.
rms
(–15 dBm), 150 kHz ≤fc ≤650 MHz.
rms
(–10 dBm), 650 MHz <fc ≤1300 MHz.
rms
(–15 dBm), 1300 MHz <fc ≤26.5 GHz.
rms
<100 MHz.
c
≥100 MHz.
c
(–60 dBm).
rms
(–50 dBm).
rms
≤1300 MHz.
c
≤26.5 GHz.
c
FREQUENCY: 10 MHz.
AGING RATE:
<1 ´ 10–6/month.
–9
<1 ´ 10
/day (Option 002)8.
Supplemental Characteristics:
INTERNAL REFERENCE ACCURACY:
Overall accuracy is a function of timebase calibration,
aging rate, temperature effects, line voltage effects and
short-term stability.
Aging Rate
Temperature Effects
Line Voltage Effects
(+5%, –10% Line
Voltage Change)
Short-Term Stability
Standard
<1 x 10–6/mo.
<2 x 10–7/°C
–6
<1 x 10
—
Option 002
<1 x 10–9/day
–10
<2 x 10
<6 x 10
<1 x 10–9 for
1 second average
RF Power
The HP 8902A measuring receiver, with HP 11722A sensor
module, performs RF power measurements from –20 dBm
(10 µW) to +30 dBm (1 W) at frequencies from 100 kHz to
2.6 GHz.
The HP 8902A measuring receiver, with HP 11792A sensor
module, performs RF power measurements from –20 dBm
(10 µW) to +30 dBm (1 W) at frequencies from 50 MHz to
26.5 GHz.
9
RF POWER RESOLUTION
0.01% of full scale in watts or volts mode.
0.01 dB in dBm or dB
LINEARITY (includes sensor non-linearity):
RF range linearity ± RF range-to-range
change error.
RF RANGE LINEARITY (using recorder output)
±0.02 dB, RF ranges 2 through 5.
±0.03 dB, RF range 1.
Using front-panel display add ±1 count of
least-significant digit.
RF RANGE-TO-RANGE CHANGE ERROR (using recorder
output): ±0.02 dB/RF range change from reference range.
Using front-panel display add ±1 count of least-significant digit.