Agilent 11713C Operating and Service Manual

Agilent 11713B/C Attenuator/Switch Drivers
Operating and Service Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2007
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agi­lent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extend allowed by the Insti­tute’s calibration facility, and to the calibra­tion facilitiesof the other International Standards Organization members.
Manual Part Number
11713-90024
Edition
First edition, October 2007
Printed in Malaysia
Agilent Technologies, Inc. Phase 3 Bayan Lepas Free Industrial Zone Penang, 11900 Malaysia
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted
General Safety Information
The following general safety precautions must be observed during all phases of operation of this instru­ment. Failure to comply with these precautions or with specific warnings elsewhere in this manual vio­lates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these regulations.
NOTE
The 11713B/C complies with INSTALLATION CATEGORY II as well as POLUTION DEGREE 2 in IEC61010-1. The 11713B/C is an INDOOR USE product.
Before applying power
Verify that all safety precautions are taken. Make all connections to the unit before applying power.
Gound the instrument
This product is provided with protective earth terminals. To minimize shock hazard, the instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly con­nected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
DO NOT operate the product in an explosive atmosphere or in the presence of flammable gases or fumes
DO NOT use repaire fuses or short-circuited fuse holders
For continued protection against fire, repalce the line fuse(s) only with fuse(s) of the same voltage and current rating and type.
Do Not Remove the Instrument Cover
Only qualified, service-trained personnel who are aware of the hazards involved should remove the instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Do Not Modify the Instrument
Do not install substiture parts or perform any unauthorized modification to the instrument. Return the instrument to an Agilent Sales and Service Office for service and repair to ensure that safety features are maintained.
In Case of Damage
Instruments that appear damaged or defective should be made inoperative and secured against unin­tended operation until they can be repaired by qualified, service-trained personnel.
11713B/C Operating and Service Manual 3
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE Directive Annex 1, this product is classed as a “Monitoring and Control Instrumentation” product.
The affixed product label is shown as below:
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent office, or visit:
www.agilent.com/environment/product
for more information.
4 11713B/C Operating and Service Manual
Safety Symbols
The following symbols on the instrument aand in the documentation indicate precautions that must be taken to maintain safe operation of the instrument.
Alternating current
Earth (grounding) terminal
Off (supply)
On (supply)
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
11713B/C Operating and Service Manual 5
Contacting Agilent
For more information, please contact your nearest Agilent office.
Americas
Canada Latin America United States
(877) 894-4414 305 269 7500 (800) 829-4444
Asia Pacific
Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Thailand
1 800 629 485 800 810 0189 800 938 693 1 800 112 929 81 426 56 7832 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008
Europe
Austria Belgium Denmark Finland France Germany Ireland Italy Netherlands Spain Sweden Switzerland(French) Switzerland(German) United Kingdom Other European Countries:
0820 87 44 11 32 (0) 2 404 93 40 45 70 13 15 15 358 (0) 10 855 2100 0825 010 700 01805 24 6333 1890 924 204 39 02 92 60 8484 31 (0) 20 547 2111 34 (91) 631 3300 0200-88 22 55 41 (21) 8113811 (Opt 2) 0800 80 53 53 (Opt 1) 44 (0) 118 9276201 www.agilent.com/find/contactus
Or, go to www.agilent.com/find/assist for more information.
6 11713B/C Operating and Service Manual

Contents

1 Introduction
Key Features of Agilent 11713B/C Attenuator/Switch Drivers 14
Table 1. Key Features of 11713B/11713C 14
Compatible Agilent Attenuators and Switches 15
Table 2. Compatible Agilent Switches 15 Table 3. Compatible Agilent Attenuators 15
Connecting Accessories 16
Table 4. Connecting Accessories for Agilent 11713B/11713C 16
11713B Front and Rear Panels at a Glance 17
Figure 1. 11713B Front Panel Features 17 Figure 2. 11713B Rear Panel Features 19
11713C Front and Rear Panels at a Glance 20
Figure 3. 11713C Front Panel Features 20 Figure 4. 11713C Rear Panel Features 22
2 Installation
Initial Inspection 26
Preparing for Use 27
Bench Operation 27 Figure 5. Handle Positioning for Bench Operation 27 Rack Mounting One Instrument (Option 908) 28 Figure 6. Illustrations of steps 2 and 3 28 Rack Mounting Two Instruments (Option 909) 29 Figure 7. Illustration of completed setup 29
Connecting to Agilent Attenuators and Switches 30
Table 5. Summary of switches and attenuators connections to
11713B/C 30 Table 6. 11713CB/C Front Panel and Back Panel Properties 32 Driving Four-Section Attenuators and Switches 33 Figure 8. Typical connection for a programmable four-section
attenuator 33 Driving Additional Coaxial Switches 34
11713B/C Operating and Service Manual 7
3 Specifications
Figure 9. Typical connection for 8762 and 8765 series coaxial
switches 34 Figure 10. Typical connection for 8761 series coaxial switches 35 Driving Relays 36 Figure 11. Typical connection for relay driving circuit 36
General Specifications 38
Drive Power Supply Specifications 38 Table 7. 11713B/C Drive Power Supply Specifications 38 Supplement Characteristics 38 Table 8. 11713B/C Supplemental Characteristics 38 Physical Specifications 39 Remote Programming Characteristics 39
Environmental Specifications 40
Table 9. 11713B/C Environmental Specifications 40
Safety and Regulatory Information 41
Safety Considerations 41 Compliance With Electromagnetic Compatibility (EMC) 41 Compliance With German Noise Requirement 41
4 Verification
5 Local Operations
Regularity Markings 42
Declaration of Conformity 43
Operator’s Check for Local Operation 46
Recommended Test Equipment 46 Table 10. Recommended Test Equipment 46 Procedure 47 Figure 12. Switching System for Verification 47
Operator’s Check for Remote Operation 48
Getting Started with 11713B/C 50
Step One: Turn on 11713B/C 50 Step Two: Configure 11713B/C settings to drive attenuators and
switches 51 Step Three: Configure 11713B/C connections to attenuators and
switches 55 Table 11. Connection between 11713B/C and programmable
attenuators 55 Table 12. Connection between 11713B/C and switches 55
8 11713B/C Operating and Service Manual
Main Menu of 11713B/C 56
Save/Recall State Menu 62
6 Remote Operations
Configuring Remote Interface 66
Programming Guide (SCPI) 68
SYSTEM Menu 56 ATT E N M enu 57 CYCLE Menu 58 IO Menu 59 INFO Menu 59 UTILITY Menu 60 Preset Menu 61
List of Default Values 63
Interface Selection 66 GPIB Configuration 66 USB Configuration 67 LAN Configuration 67
SCPI Command Syntax 68 Rules for Using a Channel or Scan List 68 Commands Relevant to Attenuators 69 Commands Relevant to Switches 71
Programming Guide (Legacy) 73
Data Message Input Format 73
SCPI and Legacy Commands Compatibility 74
7 Remote Interface Configurations
Connecting the 11713B/C to Your Computer 76
Connecting Over GPIB 76 Connecting Over USB 78 Connecting Over LAN 79
Exploring the 11713B/C Web Interface Over LAN 85
Launching the Web Interface 85 Figure 13. 11713B/C Web Interface’s Welcome Window 86 Navigation Bar 86 Displaying the Browser Web Control Page 87 Figure 14. 11713B/C Brower Web Control Interface 87
8 Replacing Major Assemblies
Replaceable Parts 90
11713B/C Operating and Service Manual 9
Table 13. Major Assemblies Reference Designator and Part
Number 90 Table 14. Front Panel Module Reference Designator and Part
Number 90 Table 15. Rear Panel Module Reference Designator and Part
Number 91 Table 16. PPMC PCA Module Reference Designator and Part
Number 91 Table 17. Fan Module Reference Designator and Part Number 92 Table 19. Power Supply Module Reference Designator and Part
Number 93 Table 18. Motherboard Reference Designator and Part Number 93
Top Module Cover Removal/Replacement 94
To R em ove Top Mod ul e C ove r 94 To Replace Top Module Cover 94
Front Panel Module Removal/Replacement 95
To Remove Front Panel Module 95 To Replace Front Panel Module 96
Rear Panel Module Removal/Replacement 97
To Remove Rear Panel Module 97 To Replace Rear Panel Module 98
PPMC PCA Removal/Replacement 99
To Remove PPMC PCA 99 To Replace PPMC PCA 99
Fan Module Removal/Replacement 100
To R em ove Fa n M od u le 100 To Replace Fan module 100
Motherboard Removal/Replacement 101
To Remove Motherboard 101 To Replace Motherboard 102
Power Supply Module Removal/Replacement 103
To Remove Power Supply Module 103 To Replace Power Supply Module 103
9 Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation 106
Reducing ESD Damage 107
Maintainance and Adjustments 108
Fuse Removal/Replacement 108 Adjustment 108
10 11713B/C Operating and Service Manual
Troubleshooting 109
Isolating a Malfunction 109 Front Panel Troubleshooting 110 Rear Panel Troubleshooting 111 Figure 15. Motherboard with Marked Location of Figure 17 and Figure
18 111 Figure 16. Location of TP1 and TP2 on motherboard 112 Figure 17. Location of TP3, TP4 and TP5 on Motherboard 113
11713B/C Operating and Service Manual 11
12 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
1 Introduction
Key Features of Agilent 11713B/C Attenuator/Switch Drivers 14 Compatible Agilent Attenuators and Switches 15
“Connecting Accessories
11713B Front and Rear Panels at a Glance 17 11713C Front and Rear Panels at a Glance 20
This chapter provides you the overview of Agilent 11713B/C attenuator/switch drivers that includes the instruments’ functions and capabilities, compatibility with Agilent switching components and physical appearances.
Agilent Technologies
13
1 Introduction

Key Features of Agilent 11713B/C Attenuator/Switch Drivers

The 11713B attenuator/switch driver is a GPIB compatible instrument that concurrently drives up to two four- section programmable step attenuators and two microwave coaxial switches, or up to 10 SPDT switches. The 11713B is fully backward compatible with 11713A in terms of functionality and fit. Connectivity using USB and LAN are optional.
The 11713C attenuator/switch driver is a GPIB/USB/LAN compatible instrument that concurrently drives up to four four-section programmable step attenuators and four microwave coaxial switches, or up to 20 SPDT switches. The 11713C comes with tri- voltage selection of +5 V, +15 V and +24 V and also permits user- defined voltage supply capability.
NOTE
The 11713B/C attenuator/switch drivers output continuous current and do not support pulse drive. Please ensure your switching devices can withstand continuous current or have a built-in current interrupt feature.
Tab le 1 Key Features of 11713B/11713C
Key Features 11713B 11713C
Manually-controlled using front panel push buttons Yes Yes
Automatically-control through: GPIB USB LAN
Integrated LCD display Yes Yes
Self-contained power supply with current limiting Yes Yes
Common terminal supplies of: +5 Vdc +15 Vdc +24 Vdc User-defined
Ye s Optional Optional
No No
Ye s
No
Ye s Ye s Ye s
Ye s Ye s Ye s Ye s
TTL control No Yes
14 11713B/C Operating and Service Manual

Compatible Agilent Attenuators and Switches

The 11713B/C attenuator/switch drivers are designed to drive the following Agilent attenuators and switches. If you are using attenuators and switches made by another supplier, check the switching characteristics against those specified in Chapter 3, “Specifications. Refer to Agilent 11713B/C Configuration Guide, literature number 5989- 7277EN, for the most updated list.
Tab le 2 Compatible Agilent Switches
Introduction 1
Agilent Model Number Description
8761A/B, 8765A/B/C/D/F (33314A/B/D), N1810UL SPDT, unterminated
8762A/B/C/F (33311A/B/C), N1810TL SPDT, terminated
8763A/B/C (33312A/B/C), N1811TL Bypass, 4-port, terminated
8764A/B/C (33313A/B/C), N1812UL Bypass, 5-port, unterminated
8766K (33366K) SP3T, unterminated
8767K (33367K), 8767M, L7204A/B/C SP4T, unterminated
87104A/B/C/D/E, 87204A/B/C/D/E, L7104A/B/C SP4T, terminated
8768K (33368K), 8768M SP5T, unterminated
8769K (33369K), 8769M, L7206A/B/C SP6T, unterminated
87106A/B/C/D/E, 87206A/B/C/D/E, L7106A/B/C SP6T, terminated
87222C/D/E, L7222C DPDT (transfer), unterminated
87406B Matrix, 4-port, terminated
87606B Matrix, 6-port, terminated
U9397A/C SPDT, terminated, solid state
* Electromechanical switches unless specified
*
Tab le 3 Compatible Agilent Attenuators
Agilent Model Number Description
8494G/H (33320G/H), 84904K/L/M (33324K/L) 11 dB, 1 dB steps
8495G/H/K (33321 G/H/K), 84907K/L (33327K/L) 70 dB, 10 dB steps
8496G/H (33322G/H) 110 dB, 10 dB steps
8497K ( 33323K), 84906K/L ( 33326K/L) 90 dB, 10 dB steps
84905M 60 dB, 10 dB steps
84908M 65 dB, 5 dB steps
11713B/C Operating and Service Manual 15
1 Introduction

Connecting Accessories

Various types of connecting accessories are available to drive Agilent attenuators and switches using the 11713B/C attenuator/switch drivers. Table 4 lists available accessories.
Tab le 4 Connecting Accessories for Agilent 11713B/11713C
Connecting
Accessories
11716A Interconnect kit (Type-N connectors)
11716C Interconnect kit (SMA connectors)
11713B-001 11713C-001
11713B-101 11713C-101
11713B-201 11713C-201
11713B-301 11713C-301
11713B-401 11713C-401
11713B-501 11713C-501
11713B-601 11713C-601
11713B-701 11713C-701
11764-60004, Viking connector to 10-pin DIP connector, 60 inches long
8120-2703, Viking connector to Viking connector, 60 inches long
5061-0969, Viking connector to 12-Pin conductor cable, bare wire (for five switches), 60 inches long
11761-60001, Viking connector to (4) ribbon cables, connect up to four switches, 60 inches long
11713-60042, Dual Viking connector to 16-Pin DIP connector, 60 inches long
11713-60043, Viking connector to (4) 9-Pin Dsub connectors, connect up to four switches, 60 inches long
11713-60044, Viking connector to 16-Pin DIP connector, 60 inches long
5064-7848, Viking connector to 14-pin DIP connector, 60 inches long
Description
*
*
11713B-801 11713C-801
11713B-908 11713C-908
11713B-909 11713C-909
* Order this kit to connect two programmable step attenuators in series
11713-60047, Viking connector to (4) 10-pin DIP connectors, connect up to 4 switches, 50 inches long
5063-9240, Rack mount kit for one instrument
5061-9694 & 5063-9212, Rack mount kit for two instruments
Refer to Agilent 11713A/B/C Attenuator/Switch Driver Configuration Guide, literature number 5989- 3703EN, for configuration details.
16 11713B/C Operating and Service Manual

11713B Front and Rear Panels at a Glance

This section briefly describes the function of the front panel keys of 11713B.
Introduction 1
2
9
3
4
6
8
7
12
1
1
1
10

Figure 1 11713B Front Panel Features

1 LCD screen.
2 Softkeys. These unmarked keys are referred to by the text on
display next to them.
5
3 Menu/Enter. Press this key to select the highlighted parameter
On/Off or select the highlighted field or go back to the main menu.
4 Preset. Press this key to preset the driver.
5 Config. Press this key to access the configuration menu. You can
set the atttenuator type, supply voltage and TTL condition through this menu.
6 Save/Recall. Press this key to save current settings or recall saved
settings.
7 Navigation buttons. The arrow keys are used to navigate
parameters displayed on the LCD screen or change parameters such as GPIB address.
11713B/C Operating and Service Manual 17
1 Introduction
8
Switches. In the local mode, pushbutton switches 9 and 0 change the position of a coaxial switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively.
9 Attenuator Y. In the local mode, pushbuttons 5, 6, 7, and 8
change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel.
10 Attenuator X. In the local mode, pushbuttons 1, 2, 3, and 4
change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel.
11 On/Standby. Press this key to switch between on and standby.
When power is supplied, the background LED is red. Pressing the key once, switches the driver on and the background LED turns to green.
12 Local. Press this key to control the driver from the front panel
when it is operating via the remote interfaces.
18 11713B/C Operating and Service Manual
Introduction 1
This section briefly describes the function of the rear panel connectors of 11713B.
3
1
11
2
10
4
9
5
6
8

Figure 2 11713B Rear Panel Features

1 ATTEN X. Viking connector for connection to attenuator or
switch(es).
7
2 ATTEN Y. Viking connector for connection to attenuator or
switch(es).
3 S9 A/B. Banana jack connectors for connection to coaxial switch.
4 24 VDC COM. Banana jack connector to provide common +24 Vdc
in driving the coaxial switches connected to S9 and/or S0.
5 S0 A/B. Banana jack connectors for connection to coaxial switch.
6 Receptacle. Matches transformer primary to line voltage via power
cable.
7 Alert symbol. This symbol is used to point out a necessary
reference for the user.
8 GPIB Connector. The interface connector from a source device to
a listening device for the remote mode of operation.
9 LAN Connector. The interface connector for LAN cable (option
LXI only).
10 USB Connector. The interface connector for Type mini B 5- pin
USB cable (option LXI only).
11 Instrument Markings.
11713B/C Operating and Service Manual 19
1 Introduction

11713C Front and Rear Panels at a Glance

This section briefly describes the function of the front panel keys of 11713C.
17
16
1
15
14
2
12
13
11
3
4
5
6
7
10
8
9

Figure 3 11713C Front Panel Features

1 LCD screen.
2 Softkeys. These unmarked keys are referred to by the text on
display next to them.
3 Navigation buttons. The arrow keys are used to navigate
parameters displayed on the LCD screen or change parameters such as GPIB address.
4 Menu/Enter. Press this key to select the highlighted parameter
On/Off or select the highlighted field or go back to the main menu.
5 Preset. Press this key to preset the driver.
6 Config. Press this key to access the configuration menu. You can
set the atttenuator type, supply voltage and TTL condition through this menu.
7 Save/Recall. Press this key to save current settings or recall saved
settings.
8 Supply Voltage for Bank 1. Indicates supply voltage setting
(background LED in red) for bank 1.
20 11713B/C Operating and Service Manual
Introduction 1
Supply Voltage for Bank 2. Indicates supply voltage setting
9
(background LED in red) for bank 2.
10 Switches for Bank 1. In the local mode, pushbutton switches 9
and 0 change the position of a coaxial switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively, for bank 1.
11 Switches for Bank 2. In the local mode, pushbutton switches 9
and 0 change the position of a coaxial switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively for, bank 2.
12 Attenuator Y for Bank 1. In the local mode, pushbuttons 5, 6, 7,
and 8 change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel, for bank 1.
13 Attenuator Y for Bank 2. In the local mode, pushbuttons 5, 6, 7,
and 8 change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel, for bank 2.
14 Attenuator X for Bank 1. In the local mode, pushbuttons 1, 2, 3,
and 4 change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel, for bank 1.
15 Attenuator X for Bank 2. In the local mode, pushbuttons 1, 2, 3,
and 4 change the attenuation setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel, for bank 2.
16 On/Standby. Press this key to switch between on and standby.
When power is supplied, the background LED is red. Pressing the key once, switches the driver on and the background LED turns to green.
17 Local. Press this key to control the driver from the front panel
when it is operating via the remote interfaces.
11713B/C Operating and Service Manual 21
1 Introduction
This section briefly describes the function of the rear panel connectors of 11713C.
3
1
2
18
5
4
17
7
6
8
16
9
11
10
12
13
15
14

Figure 4 11713C Rear Panel Features

1 ATTEN X Bank 1. Viking connector for connection to attenuator
or switch(es), for bank 1.
2 ATTEN X Bank 2. Viking connector for connection to attenuator
or switch(es), for bank 2.
3 ATTEN Y Bank 1. Viking connector for connection to attenuator
or switch(es), for bank 1.
4 ATTEN Y Bank 2. Viking connector for connection to attenuator
or switch(es), for bank 2.
5 S9 A/B Bank 1. Banana jack connectors for connection to coaxial
switch, for bank 1.
6 S9 A/B Bank 2. Banana jack connectors for connection to coaxial
switch, for bank 2.
7 VDC COM Bank 1. Banana jack connector to provide common
Vdc in driving the coaxial switches connected to S9 and/or S0, for bank 1.
8 VDC COM Bank 2. Banana jack connector to provide common
Vdc in driving the coaxial switches connected to S9 and/or S0, for bank 2.
22 11713B/C Operating and Service Manual
Introduction 1
S0 A/B Bank 1. Banana jack connectors for connection to coaxial
9
switch, for bank 1.
10 S0 A/B Bank 2. Banana jack connectors for connection to coaxial
switch, for bank 2.
11 External VDC. Banana jack connector to provide user-defined
Vdc, for both banks.
12 Ground. Banana jack connector to provide grounding, for both
banks.
13 Receptacle. Matches transformer primary to line voltage via power
cable.
14 Alert symbol. This symbol is used to point out a necessary
reference for the user.
15 GPIB Connector. The interface connector from a source device to
a listening device for the remote mode of operation.
16 LAN Connector. The interface connector for LAN cable.
17 USB Connector. The interface connector for Type mini B 5- pin
USB cable.
18 Instrument Markings.
11713B/C Operating and Service Manual 23
1 Introduction
24 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
2 Installation
Initial Inspection 26 Preparing for Use 27
“Bench Operation “Rack Mounting One Instrument (Option 908) “Rack Mounting Two Instruments (Option 909)
Connecting to Agilent Attenuators and Switches 30 Driving Four-Section Attenuators and Switches 33 Driving Additional Coaxial Switches 34 Driving Relays 36
This chapter provides you important information on how to unpack and check your instrument, how to prepare your instrument for bench operation and tips on configuring 11713B/C with Agilent attenuators and switches.
Agilent Technologies
25
2 Installation

Initial Inspection

1 Unpack and inspect the shipping container and its contents
thoroughly to ensure that nothing was damaged during shipment. If the shipping container or cushioning material is damaged, the contents should be checked both mechanically and electrically.
2 If the contents are damaged or defective, contact your nearest
Agilent Technologies Service and Support Office. Refer to the Service and Support information in the front matter of this manual. Agilent Technologies will arrange for repair or replacement of the damaged or defective equipment. Keep the shipping materials for the carrier's inspection.
3 If you are returning your instrument for service, repackaging the
attenuator/switch driver requires original shipping containers and materials or their equivalents. Agilent Technologies can provide packaging materials identical to the original materials. Refer to "Service and Support" on page v for the Agilent Technologies nearest you.
26 11713B/C Operating and Service Manual

Preparing for Use

Bench Operation

Installation 2
Pull the handle outwards, and adjust it into one of the two positions illustrated in Figure 5.
(Top) Handle placed underneath the instrument to assure self- alignment of the instruments when stacked
(Bottom) Handle tilted to raise the front of the instrument for easier viewing of the front panel
Figure 5 Handle Positioning for Bench Operation
11713B/C Operating and Service Manual 27
2 Installation

Rack Mounting One Instrument (Option 908)

1 Remove the handle and front/rear bumpers
2 Fit the rack mount flanges (part number 5063- 9240)
3 Ready for installation
Figure 6 Illustrations of steps 2 and 3
CAUTION
When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts forced convection must be used.
28 11713B/C Operating and Service Manual

Rack Mounting Two Instruments (Option 909)

1 Remove the handle and front/rear bumpers
2 Fit one small mounting flange to opposite sides of each
instrument (part numbers 5061- 9694 and 5063- 9212)
3 Fit two front linking plates to each instrument
4 Engage the linking plates at the front of the instruments
5 Attach the rear linking brackets
6 Ready for installation
Installation 2
Figure 7 Illustration of completed setup
CAUTION
When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts forced convection must be used.
11713B/C Operating and Service Manual 29
2 Installation

Connecting to Agilent Attenuators and Switches

The 11713B/C attenuator/switch drivers can be used to drive various switches and attenuators. Table 5 shows the summary of switches and attenuators connections to 11713B/C, with various interface cables for point- to- point connection. Table 6 shows the properties of 11713B/C’s front panel and rear panel.
For configuration details, please refer to Agilent 11713B/C Configuration Guide, literature number 5989- 7277EN.
Tab le 5 Summary of switches and attenuators connections to 11713B/C
Switches/Attenuators 11713B/C
Cable
*
Option
Switches
8761A/B
Switches
8762A/B/C/F
8763A/B/C
201 1 No Yes
201 2 Yes Yes
8764A/B/C
Switches
8765A/B/C/D/F
Switches
8766K
Switches
N1810UL/TL
N1811TL
N1812UL
Switches
87222C/D/E
201, 301 2 Yes Yes
001, 101
201, 501 2 Yes Yes
201, 801 2 Yes No
L7222C
# of channels required
2
Control by
ATTEN X (1-4)
ATTEN Y (5-8)
Control by
SWITCHES
(9/0)
Ye s Ye s
Switches
8767K
Attenuators
001, 101 3 Yes Yes
8495G/H
Switches
8767M
Attenuators
001 3 Yes Yes
84905M
84907K/L
30 11713B/C Operating and Service Manual
Installation 2
Tab le 5 Summary of switches and attenuators connections to 11713B/C
Switches
8768K
Attenuators
8494G/H
001, 101 4 Yes Yes
8496G/H
8495K, 8497K
Switches
8768M
Attenuators
84904K/L/M
001 4 Yes Yes
84906K/L
84908M
Switches
87104A/B/C/D
87204A/B/C
201, 601 4
**
Ye s
L7104A/B/C
L7204A/B/C
Ye s
Switches
8769K
Switches
8769M
Switches
101 5 Yes Yes
701 5 Yes Yes
87106A/B/C/D
87206A/B/C
L7106A/B/C
201, 401 6
††
Ye s Ye s
L7206A/B/C
87406B, 87606B
* Type of interface cable required depends on the DC connector on the switching device
† One channel represent control with one pushbutton
‡ Ensure switch is equiped with current interrupt (option 403) to protect switch from overhetaing and
destruction as this switch cannot withstand continuous current
**If option 601 is used, number of channels required is 5, to cater for open all path function controlled via
S9/S0
††If option 401 is used, number of channels required is 7, to cater for open all path function controlled via
S9/S0
11713B/C Operating and Service Manual 31
2 Installation
Tab le 6 11713CB/C Front Panel and Back Panel Properties
Front Panel Pushbuttons
*
Pushbutton Number Pushbutton
LEDs
SWITCHES
ON ATTEN X-3 (S9-A) Gray
9
OFF ATTEN X-4 (S9-B) White/Red
ON ATTEN Y-3 (S0-A) Gray
0
OFF ATTEN Y-4 (S0-B) White/Red
ATTENUATORS
XY
OFF 5 Violet
15
ON 6 Yellow
OFF 7 Black
26
ON 8 Green
Rear Panel Connectors
Pin Numbers Wire Color Code
1 Red (Vcc)
2 White/Brown (Gnd)
37
OFF 9 Orange
ON 10 Blue
OFF 11 Brown
48
ON 12 White
* The ON/OFF status of the pushbutton LEDs indicates which cable wire or pin on the rear panel
connector is grounded. As an example, if ATTENUATOR X pushbutton 3 is illuminated, pin 10 of the ATTEN X connector (blue wire from cable) is grounded and pin 9 floats at a high impedance. For SWITCHES, if pin 3 is grounded (LED on), pin 4 is connected to +Vcc.
† With reference to interface cable option 201 (Viking connector to 12-pin bare), consists of 12
color-coded wires
32 11713B/C Operating and Service Manual
11713B
Installation 2

Driving Four-Section Attenuators and Switches

To use one four- section attenuator assembly, connect an attenuator interface cable either to the ATTEN X output (A6J1) or ATTEN Y output (A6J2). Connect all outputs (two for 11713B and four for 11713C) to have more than four attenuator segments.
A typical connection for a programmable four- section attenuator to 11713B is illustrated in Figure 8, together with pin number for each connector.
Using these same connections to Agilent 8762 or 8765 series coaxial switches, control can be extended to number of switches in multiple of four. If S9 and S0 outputs are utilized, 11713B and 11713C can drive up to 10 switches and 20 switches respectively.
11713B
Vcc
Vcc
TWO ATTENUATOR CABLES ARE SUPPLIED WITH 11713B
Figure 8 Typical connection for a programmable four-section attenuator
11713B/C Operating and Service Manual 33
2 Installation

Driving Additional Coaxial Switches

Make switch connections to S0 outputs, S9 outputs, or to rear panel ATTEN X output or ATTEN Y output.
Figure 9 below shows the rear panel connections to S0 outputs
and the corresponding switch positions reflected by pushbutton indicators.
Connections to Agilent 8762 or 8765 series coaxial switches can also be made to the ATTEN X output or ATTEN Y output as illustrated in Figure 8.
Figure 9 Typical connection for 8762 and 8765 series coaxial switches
34 11713B/C Operating and Service Manual
Installation 2
Connections to Agilent 8761 series coaxial switches can only be made to S0 outputs and S9 outputs.
Figure 10 illustrates the rear panel connections to S9 outputs and
the corresponding switch positions reflected by pushbutton indicators.
Figure 10 Typical connection for 8761 series coaxial switches
11713B/C Operating and Service Manual 35
2 Installation

Driving Relays

To drive ten devices for 11713B, connect attenuator cables at ATTEN X and Y and switch cables to S9 and S0.
A total of 10 relays may be on at one time if the total curent is less than 1.7 A. However, since there are dual transistor and relay drivers, where one driver is on while the other is off, a total of 20 relays may be controlled.
NOTE
CAUTION
11713C is capable of driving double the amount of devices that 11713B can. However, the total load current can be consumed is still 1.7 A.
If the total load current of 1.7 A is exceeded, damage may result.
Figure 11 below shows the connections for a simplified relay
driving circuit. The circuit is adaptable for simple non-latching relays.
Figure 11 Typical connection for relay driving circuit
NOTE
36 11713B/C Operating and Service Manual
It is also recommended that two 28.7 V zener diodes be connected back-to-back across the relay coils to reduce voltage transients.
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
3 Specifications
General Specifications 38
“Drive Power Supply Specifications “Supplement Characteristics “Physical Specifications “Remote Programming Characteristics
Environmental Specifications 40 Safety and Regulatory Information 41
“Safety Considerations “Compliance With Electromagnetic Compatibility (EMC) “Compliance With German Noise Requirement
Regularity Markings 42 Declaration of Conformity 43
This chapter provides you the specifications of Agilent 11713B/C attenuator/switch drivers.
Agilent Technologies
37
3 Specifications

General Specifications

Drive Power Supply Specifications

Specifications below describe warranted performance over the temperature range 0 to +50 operation, unless otherwise noted.
Tab le 7 11713B/C Drive Power Supply Specifications
°C after one hour of continuous
Voltage +24 ± 8% Vdc
Current 1.7 A maximum continuous current
* For 11713C only

Supplement Characteristics

Supplement characteristics are intended to provide useful information and are typical but non- warranted performance parameters.
Tab le 8 11713B/C Supplemental Characteristics
Power 100 to 240 Vac, automatic selection, 50/60 Hz
Response Time 100 μs maximum for contact pairs 1 through 8
+5 ± 5% Vdc +15 ± 12% Vdc
Contact pairs 1 through 8, 9, and 0, maximum current of 1.7 A continuous through all contacts (< 0.7 A per contact)
100 VA maximum Mains supply voltage fluctuations are not to exceed 10 percent of the nominal supply voltage
20 ms maximum for contact pairs 9 and 0
*
*
Driver Life > 2,000,000 switchings at 0.7 A for contact pairs 9 and 0
Maximum Load Inductance
Maximum Load Capacitance
38 11713B/C Operating and Service Manual
500 mH
< 0.01 μF for contact pairs 9 and 0

Physical Specifications

Net Weight 3.2 kg (7.1 lbs)
Specifications 3
Dimensions (H x W x D) with handle and rubber bumper
Dimensions (H x W x D) without handle and rubber bumper
103.8 mm x 232.2 mm x 378.7 mm (4.1 inches x 9.1 inches x 14.9 inches)
88.3 mm x 212.7 mm x 346.0 mm (3.5 inches x 8.4 inches x 14.3 inches)

Remote Programming Characteristics

Interface GPIB interface operates to IEEE 488.2 and IEC65
10/100 BaseT LAN interface USB 2.0 interface
Command Language SCPI standard interface commands (Agilent 11713A backward
compatible)
GPIB Compatibility SH0, AH1, T0, TE0, L2, LE0, SR0, RL1, PP0, DC0, DT0, C0
11713B/C Operating and Service Manual 39
3 Specifications

Environmental Specifications

Agilent 11713B/C attenuator/switch drivers are designed to fully comply with Agilent Technologies’ product operating environmental specifications shows in table below.
Tab le 9 11713B/C Environmental Specifications
Temperature
Operating
Storage
Humidity
Operating
Storage
Condensing
Shock
End-user handling
Bench Handling
Functional
Transportation
0
°C to +50°C
°C to +70°C
-40
95% RH at 40 90% RH at 65 95% RH at 40
Half-sine: 2-3 ms duration, 60 in/s (1.6 ms) delta-V Half-sine: 11 ms duration, 30 grms Trapezoital: 18-22 ms duration, 337 in/s (8.56 ms) delta-V Per MIL-PRF-28800F
°C, 5 days cyclic °C, 24 hours °C, 5 hours (condensation 15 minutes)
Vibration
Operating
Survival
Altitude
Operating
Non-operating
Random: 0.21G rms, 5 to 500 Hz, 10 min/axis Random: 2.09G rms, 5 to 500 Hz, 10 min/axis Swept-sine: 0.5 G rms, 5 to 500 Hz, 10 min/axis
< 2,000 meters (6,540 feet) < 15,300 meters (50,000 feet)
40 11713B/C Operating and Service Manual

Safety and Regulatory Information

Safety Considerations

This product has been designed and tested in accordance with IEC 61010- 1:2001, Safety Requirements for Electronic Measuring Apparatus, and has been supplied In a safe condition. The Instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product In safe condition.

Compliance With Electromagnetic Compatibility (EMC)

This product conforms with the protection requirements of EMC Directive 2004/108/EC for Electromagnetic Compatibility.
The conformity assessment requirements have been met using the technical Construction file route to compliance, using EMC test specifications EN 55011:1990 (Group 1, Class A).
Specifications 3
In order to preserve the EMC performance of the product, any cable which becomes worn or damaged must be replaced with the same type and specifications.
Refer to the “Declaration of Conformity" on page 43.

Compliance With German Noise Requirement

Acoustic Noise Emission This is to declare that this instrument is in
conformance with the German Regulation on Noise Declaration for Machines.
LpA < 70 dB
Operator position
Normal position
per ISO 7779
Gerauschemission Laermangabe nach der
Maschinenlaermrerordnung- 3.GSGV Deutschland.
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 t.19
11713B/C Operating and Service Manual 41
3 Specifications

Regularity Markings

The following markings can be found on the rear panel.
The CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with Canadian ICES-001. Cet appareill ISM est conforme a la norme NMB-001 du Canada.
The CSA mark is a registered trademark of the Canadian Standards Association. A CSA mark with the indicators “C” and “US” means that the product is certified for both the U.S. and Canadian markets, to the applicable American and Canadian standards.
This is the symbol of an Industrial Scientific and Medical Group 1 Class A product.
The C-Tick mark is a registered trademark of the Spectrum Management Agency of Australia. This signifies compliance with the Australian EMC Framework Regulations under the terms of the Radio communications Act of 1992.
42 11713B/C Operating and Service Manual

Declaration of Conformity

Specifications 3
11713B/C Operating and Service Manual 43
3 Specifications
44 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
4 Verification
Operator’s Check for Local Operation 46
“Recommended Test Equipment “Procedure
Operator’s Check for Remote Operation 48
This chapter provides you simply instructions to verify Agilent 11713B/C attenuator/switch drivers’ functionality in both local operation and remote (GPIB/USB/LAN) operation.
Agilent Technologies
45
4 Verification

Operator’s Check for Local Operation

Recommended Test Equipment

Table 10 lists the test equipment required for performance test
verification and equipment troubleshooting. Equipment other than the recommended models can be used, provided minimum specifications are satisfied.
Tab le 10 Recommended Test Equipment
Instrument Type Critical Specifications Use
Digital Voltmeter 0 to +30 Vdc T
Attenuators (2 required) Programmable, 4 sections P, T
Switches (2 required) +5 Vdc, +15 Vdc or +24 Vdc
drive source
* P = Performance, T = Troubleshooting
P, T
*
46 11713B/C Operating and Service Manual

Procedure

Verification 4
1 Configure switching system illustrated in Figure 12 by following
all steps found in Chapter 5, “Local Operations.
Figure 12 Switching System for Verification
2 Once configuration is completed, press and depress the 10
numbered pushbuttons on the driver front panel. Each LED should alternate between off and on as each key is pressed.
3 In addition, if any switching devices is connected (attenuators,
relays or switches), an audible click should be heard from the unit actuated. Pressing any numbered pushbutton should not cause any other push button to change state.
11713B/C Operating and Service Manual 47
4 Verification

Operator’s Check for Remote Operation

These procedure verify that the driver can be controlled remotely using GPIB, USB and/or LAN.
1 Refer to Chapter 7, “Remote Interface Configurations, to connect
11713B/C to your computer through GPIB, USB and/or LAN.
2 Once remote connection is available, send the following SCPI
commands to the driver and note the changes on front panel LEDs.
Description Command Observation
Close switching paths from channel 1 to channel 4 (bank1)
Open switching paths from channel 5 to channel 8 (bank 1)
Query status on channel 2 (bank 1)ROUTe:CLOSe? (@102) Return value “1” (LED for
Query status on channel 7 (bank 1)ROUTe:CLOSe? (@107) Return value “0” (LED for
ROUTe:CLOSe (@101:104) LEDs for pushbuttons 1 to
4 light ON
ROUTe:OPEn (@105:108) LEDs for pushbuttons 5 to
8 light OFF
pushbutton 2 light ON)
pushbutton 7 light OFF)
If the above checks are successful, the driver’s remote operation is working correctly. These procedures do not check all of the drivers’s program codes that can be executed. However, if the driver work correctly from the front panel, there is a high probability that the driver will respond to all the program codes.
48 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
5 Local Operations
Getting Started with 11713B/C 50
“Step One: Turn on 11713B/C “Step Two: Configure 11713B/C settings to drive attenuators and switches “Step Three: Configure 11713B/C connections to attenuators and
switches
Main Menu of 11713B/C 56
“SYSTEM Menu “ATTEN Menu “CYCLE Menu “IO Menu “INFO Menu “UTILITY Menu
Preset Menu 61 Save/Recall State Menu 62 List of Default Values 63
This chapter outlines the some simple steps to begin using the 11713B/C in local operations. Also, functionality of all menus are described to assist operations using 11713B/C.
Agilent Technologies
49
5 Local Operations

Getting Started with 11713B/C

Agilent 11713B/C attenuator/switch driver can be easily configured to drive programmable attenuators and/or switches through the front panel operations. The following three simple steps will guide you through the configuration of 11713B/C.

Step One: Turn on 11713B/C

NOTE
NOTE
For Step One, all details on 11713C is applicable to 11713B
1 Connect AC power supply to 11713B/C. You should see:
background LED of the power button is red which indicates
that 11713B/C is in standby mode
Allow 10 seconds before pressing the power button to turn on 11713B/C. This duration is needed to power up the processor within.
2 Press power button once to turn on the 11713B/C. You should see:
background LED turns green
six menus displayed on the LCD screen
all numbered buttons are lighted up
1
3 Ensure Local button is lighted up to enable operation through
front panel. If not, press button once.
4 Now, you are ready to configure the 11713B/C.
1 At factory default setting. Last state is followed if instrument was configured before shut
down.
50 11713B/C Operating and Service Manual
Local Operations 5

Step Two: Configure 11713B/C settings to drive attenuators and switches

To drive programmable attenuators
NOTE
Step Two - to drive programmable attenuators, only item 2 is applicable to 11713B. When Config button is pressed, the next screen is CONFIGURATION MENU.
Item Action Illustration
1 Select bank
(only for 11713C)
2 Select
attenuator type
a Press Config button b At SELECT BANK
screen, you can see two softkey selections: BANK1 and BANK2
c Press BANK1 softkey
to select bank 1
d Press BANK2 softkey
to select bank 2
a At CONFIGURATION
MENU screen, press
ATTEN TYPE softkey
b At ATTENUATOR
MODEL screen, scroll to highlight the desired attenuator model number using the navigation keys.
c Once the attenuator
model is determined, press ATTEN X or ATTEN Y softkey to assign the selected attenuator model
d Repeat step b and c if
needed
e Model assigned to
ATTEN X and ATTEN Y
is marked <X> (e.g. 8494G/H) and <Y> (e.g. 8496G/H)
respectively
f To return to the
previous screen, press
BACK softkey
11713B/C Operating and Service Manual 51
5 Local Operations
Item Action Illustration
3 Select
voltage type (only for 11713C)
a At CONFIGURATION
MENU screen, press
Voltage LEVEL softkey
b At OUTPUT VOLTAGE
screen, scroll to highlight the desired voltage (e.g. +5V)
c Press SET softkey to
assign highlighted voltage
d At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel
e Ouput voltage
assigned is marked <*>
f To exit this screen,
press BACK softkey
52 11713B/C Operating and Service Manual
To drive switches
Local Operations 5
NOTE
Step Two - to drive switches, only applicable to 11713C as 11713B is predefined with +24 Vdc supply and no TTL drive
Item Action Illustration
1 Select bank
(only for 11713C)
2 Select
voltage type (only for 11713C)
a Press Config button b At SELECT BANK
screen, you can see two softkey selections: BANK1 and BANK2
c Press BANK1 softkey
to select bank 1
d Press BANK2 softkey
to select bank 2
a At CONFIGURATION
MENU screen, press
TTL ON/OFF softkey
b At TTL MODE ON/OFF
screen, scroll to highlight the TTL OPTION or Non TTL
(e.g. TTL OPTION)
c Press SET softkey to
assign highlighted voltage
d At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel
e Selection is marked
<*>
f To exit this screen,
press BACK softkey
\
11713B/C Operating and Service Manual 53
5 Local Operations
Item Action Illustration
3 Select
voltage type (only for 11713C)
a At CONFIGURATION
MENU screen, press
Voltage LEVEL softkey
b At OUTPUT VOLTAGE
screen, scroll to highlight the desired voltage (e.g. +5V)
c Press SET softkey to
assign highlighted voltage
d At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel
e Ouput voltage
assigned is marked <*>
f To exit this screen,
press BACK softkey
54 11713B/C Operating and Service Manual
Local Operations 5

Step Three: Configure 11713B/C connections to attenuators and switches

1 Determine interface cable option of the 11713B/C
2 Determine DC connector option of the attenuator(s) or switch(es)
3 Check for compatibility using Table 11 for programmable
attenuators and Table 12 for switches:
Tab le 11 Connection between 11713B/C and programmable attenuators
11713B/C Attenuators (Option)
Option 001 8494G/H, 8495G/H, 8496G/H, 8495K, 8497K (Option 016)
84904K/L/M, 84905M, 84906K/L, 84907K/L, 84908M (No option)
Option 101 8494G/H, 8495G/H, 8496G/H, 8495K, 8497K (Option 060)
Tab le 12 Connection between 11713B/C and switches
11713B/C Switches (Option)
Option 001 8766K, 8767K/M, 8768K/M (Option 016)
Option 101 8766K, 8767K/M, 8768K/M, 8769K/M (Option 060)
Option 201 8763A/B/C, 8764A/B/C, 8761A/B, 8762A/B/C/F (No option)
8765A/B/C/D/F (Option 305/310/315/324) N1810UL/TL, N1811TL, N1812UL (Option 87104A/B/C/D, 87204A/B/C, 87106A/B/C/D, 87206A/B/C (Option 100) L7104A/B/C, L7204A/B/C, L7106A/B/C, L7206A/B/C (Option 100)
87406B, 87606B (Option 100) 87222C/D/E, L7222C (Option 100)
Option 301 8765A/B/C/D/F (Option 005/010/015/024)
Option 401 87106A/B/C/D, 87206A/B/C (Option 161)
L7106A/B/C, L7206A/B/C (Option 161) 87406B, 87606B (Option 161)
Option 501 N1810UL/TL, N1811TL, N1812UL (Option
Option 601 87104A/B/C/D, 87204A/B/C (Option 161)
L7104A/B/C, L7204A/B/C (Option 161)
Option 701 8769K/M (Option 016)
Option 801 87222C/D/E, L7222C (Option 161)
* If TTL is required, include Option 401. Must order switch with Option 403 (current interrupt) as switch
cannot withstand continuous current supplied by 11713B/C
*
202)
*
201)
4 For more details, refer to 11713B/C Configuration Guide,
literature number 5989- 7277EN.
11713B/C Operating and Service Manual 55
5 Local Operations

Main Menu of 11713B/C

The main menu can be displayed on the LCD screen by pressing the Menu/Enter button. The six submenus, each with their own functionality, are described in the subsequent sections.

SYSTEM Menu

Function Action Illustration
Display attenuator type(s) configured for 11713B/C
a Navigate to the
SYSTEM icon using
the navigation keys
b Press Menu/Enter
button when SYSTEM icon is highlighted
c On display are
attenuator models assigned to ATTEN X and ATTEN Y for each bank (e.g. 8494G/H,
8496G/H, N/A, N/A)
d Press Menu/Enter
button again to return to main menu
56 11713B/C Operating and Service Manual

ATTEN Menu

Function Action Illustration
Local Operations 5
Display attenuation levels of each bank
a Navigate to the ATT EN
icon using the navigation keys
b Press Menu/Enter
button when ATTEN icon is highlighted
c On display are
attenuation level for ATTEN X and ATTEN Y (e.g. 11 dB, 110 dB,
0, 0)
d Press Menu/Enter
button again to return to main menu
Attenuation value on display changes according to input from front panel pushbuttons or through the virtual web interface:
Pushbutton LED ON - attenuation card selected (attenuation applied) on corresponding attenuator section
Pushbutton LED OFF - thru path selected (attenuation lifted) on corresponding attenuator section
11713B/C Operating and Service Manual 57
5 Local Operations

CYCLE Menu

Function Action Illustration
Display number of cycles of switching path for each channel
10 channels
for 11713B
(CH1 to CH9, CH0)
20 channels
for 11713C
(B1-1 to B1-10, B2-1 to B2-10)
Clear cycle for selected channel(s)
a Navigate to the CYCLE
icon using the navigation keys
b Press Menu/Enter
button when CYCLE icon is highlighted
c On display are number
of relay cycles for each channel
d Press Menu/Enter
button again to return to main menu
a Scroll to the desired
bank-channel (e.g. B1-1) using the navigation keys
b Press CLEAR CYCLE c At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel
d Note that relay cycle
for B1-1 is 0
58 11713B/C Operating and Service Manual
Function Action Illustration
Local Operations 5
Clear cycle for all channels
a Press CLEAR ALL b At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel.
c Note that relay cycle
for all channels is 0

IO Menu

Refer to Chapter 7, “Remote Interface Configurations.

INFO Menu

Function Action Illustration
Display system info for 11713B/C
Model
Serial
Firmware
GPIB address
LAN IP
USB adress
MAC address
number
revision
a Navigate to the INFO
icon using the navigation keys
b Press Menu/Enter
button when INFO icon is highlighted
c Press Menu/Enter
button again to return to main menu
11713B/C Operating and Service Manual 59
5 Local Operations

UTILITY Menu

Function Action Illustration
Display system utility settings for 11713B/C and adjust beep volume
Beep volume range from 0 (mute) to 8 (loudest)
Set factory default
For more
information:
“List of Default Values" on
page 63
a Navigate to the
UTILITY icon using the
navigation keys
b Press Menu/Enter
button when UTILITY icon is highlighted
c Scroll to highlight
Beep VOL using the
navigation keys
d Adjust beep volume
using Volume up and
Volume down softkey
e Press Menu/Enter
button again to return to main menu
a Scroll to highlight
Factory default using
the navigation keys, then press Ye s softkey
b At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel.
60 11713B/C Operating and Service Manual
Function Action Illustration
Reset LAN a Scroll to highlight LAN
RESET using the
navigation keys, then press Ye s softkey
b At the next screen,
press OK softkey to confirm decision or press Cancel softkey to cancel.
Local Operations 5

Preset Menu

The preset function is used to apply full attenuation or open all switching paths (all LEDs light ON). For more information: “List of
Default Values" on page 63.
11713B/C Operating and Service Manual 61
5 Local Operations

Save/Recall State Menu

The save state function is important to enable user to store up to four different configurations and the recall state function allows user to reinstate one of the four saved states for use. This greatly helps to minimize the amount of works required to make “standard” configuration changes.
Function Action Illustration
To save state a Press Save/Recall
button
b Scroll to the desired
state to store configuration (e.g. STATE 0) using the navigation keys
c Press Save State
softkey to store configuration
d Press Menu/Enter
button again to return to main menu
To r ec al l s ta te a Press Save/Recall
button
b Scroll to the desired
state to recall configuration (e.g. STATE 2) using the navigation keys
c Press Recall State
softkey to reinstate configuration
a Press Menu/Enter
button again to return to main menu
62 11713B/C Operating and Service Manual

List of Default Values

Local Operations 5
Key Operation Factory-shipped
setting
Factory
Default
Key
Preset
(*RST)
Backup Save/
Recall
Available means of
defining a
setting
Channel All Lights ON
*
<<
<< * * K/C
Attenuator Type N/A << * * K/C
Supply Voltage +24V << * * K/C
TTL OFF << * * K/C
Beep Volume 8 << * K
GPIB Address 28 << * K
IP Config AUTO << * K
Manual IP Addr “192.168.1.101” << * K
Manual Gateway “0.0.0.0” << * K
Manual Subnet Mask “255.255.255.0” << * K
* When channel light is ON, it means attenuation is being applied (when attenuator is connected) or
switch path is open (when switch is connected)
† “<<“ symbol shows that the setup is the same as that in the box to the left.
‡ K = Using front panel. C = Using SCPI command.
Definition
Factory Default All data that the user can set are cleared. Relay cycle count is not cleared.
Preset Status when you press Preset key
*RST Status when you execute *RST in your program
Backup Settings that are backed up (set state not affected by turning power
ON/OFF) In the table, a setting that is automatically backed up is denoted in the
following manner: *: Backup operation performed Blank: Backup operation not performed
Save/Recall Settings that permit Save/Recall of a setup state
In the table, states that can be saved/recalled are denoted in the following manner:
*: Save/Recall can be performed Blank: Save/Recall cannot be performed
11713B/C Operating and Service Manual 63
5 Local Operations
64 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
6 Remote Operations
Configuring Remote Interface 66
“Interface Selection “GPIB Configuration “USB Configuration “LAN Configuration
Programming Guide (SCPI) 68
“SCPI Command Syntax “Rules for Using a Channel or Scan List “Commands Relevant to Attenuators “Commands Relevant to Switches
Programming Guide (Legacy) 73
“Data Message Input Format
SCPI and Legacy Commands Compatibility 74
This chapter provides the programming guide for 11713B/C in SCPI commands and legacy command.
Agilent Technologies
65
6 Remote Operations

Configuring Remote Interface

This section briefly describes how to configure the GPIB, USB and LAN remote interfaces.
NOTE
For more detailed information on remote interface connectivity configuration refer to the Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. If you have installed the IO Libraries Suite, you can access the Connectivity Guide via the Agilent IO Libraries Control icon. Alternatively, you can access the Connectivity Guide via the Web at www.agilent.com/find/connectivity.

Interface Selection

You can choose to control the 11713B/C remotely using the GPIB, USB or LAN interfaces.
For information on selecting and configuring the remote interface manually from the driver front panel or remotely via virtual front panel, refer to Chapter 5, “Local Operations.
NOTE
It is expected that most users will use the front panel keys to set up the remote interfaces. The remote interface commands are provided for completeness (for front panel operation).

GPIB Configuration

Each device on the GPIB (IEEE- 488) interface must have a unique address. You can set the 11713B/C’s address to any value between 0 an 30. The attenuator/switch driver is shipped with a default address set to 28. The GPIB address is stored in non- volatile memory and does not change when the driver is switched off or after a remote interface reset.
For information on setting the GPIB address manually from the front panel, refer to “Connecting Over GPIB" on page 76.
66 11713B/C Operating and Service Manual

USB Configuration

The USB interface requires no front panel or remote configuration. The USB cannot be changed - it is set at the factory and is unique for each 11713B/C.
Remote Operations 6
NOTE
NOTE
NOTE
For further information on USB configuration, refer to “Connecting Over
USB" on page 78.
Before connecting the USB cable, make sure that I/O software is installed on your computer.
For more information about Agilent IO Libraries software refer to the Connectivity Guide. If you have installed other I/O software refer to documentation that accompanies the software.

LAN Configuration

The 11713B/C has three LAN operating modes:
Dynamic mode(Dynamic Host Configuration Protocol or DHCP)
Auto IP mode (Local PC Control or isolated LAN)
Static mode (Manual mode)
Refer to “Connecting Over LAN" on page 79 to configure the above LAN operating modes.
11713B/C Operating and Service Manual 67
6 Remote Operations

Programming Guide (SCPI)

SCPI Command Syntax

The following conventions are used for SCPI command syntax for remote interface programming.
Square brackets ([]) indicate optional keywords or parameters
Braces ({}) enclose paramter choices within a command string
Angle brackets (<>) enclose parameters for which you must specify
a value
A vertical bar (|) separates multiple parameters

Rules for Using a Channel or Scan List

Many of the SCPI commands include a channel list or scan list parameter which allows you to specify one or more channels. The channel number has the form (@bnn), where b is the bank number and nn is the channel number.
NOTE
Channel list is from 101 to 110 for bank 1 and 201 ro 210 for bank
2.
11713B has only one bank. Therefore, by default, b = 1.
You can specify a bank, a single channel, multiple channels or a range of channels as describe below.
The following command closes a single channel (channel 4) on bank 1:
ROUTe:CLOSe (@104)
The following command closes multiple channels onmodules in banks 1 and 2:
ROUTe:CLOSe (@104,107, 201, 206)
The following command closes a range of channels. When you specify a range of channels, the range may contain invalid channels (they are ignored), but the first and last channel in the list must be valid:
ROUTe:CLOSe (@101:109)
68 11713B/C Operating and Service Manual

Commands Relevant to Attenuators

To Configure Attenuator Type
The commands below set the attenuator type (model number) for corresponding attenuators.
:CONFigure:[BANK1]:X {NA|AG8494g|AG8495g|AG8495k|AG8496g|AG8497k|AG84904k|A G84905m|AG84906k|AG84907k|AG84908m}
:CONFigure:[BANK1]:Y {NA|AG8494g|AG8495g|AG8495k|AG8496g|AG8497k|AG84904k|A G84905m|AG84906k|AG84907k|AG84908m}
:CONFigure:BANK2:X {NA|AG8494g|AG8495g|AG8495k|AG8496g|AG8497k|AG84904k|A G84905m|AG84906k|AG84907k|AG84908m}
:CONFigure:BANK2:Y {NA|AG8494g|AG8495g|AG8495k|AG8496g|AG8497k|AG84904k|A G84905m|AG84906k|AG84907k|AG84908m}
Remote Operations 6
NOTE
For AG849xh, use AG849xg. For AG84904l/m, use AG84904k. For 84906l, use AG84906k. For AG84907l, use AG849907k.
To Set Attenuation Level
The commands below are used to set the attenuation level of corresponding attenuator(s).
ATTenuator:[BANK1]:X <attenuation value in dB>
ATTenuator:[BANK1]:Y <attenuation value in dB>
ATTenuator:BANK2:X <attenuation value in dB>
ATTenuator:BANK2:Y <attenuation value in dB>
To Query Attenuation Level
The commands below return the current attenuation value for corresponding attenuator(s).
ATTenuator:[BANK1]:X?
ATTenuator:[BANK1]:Y?
ATTenuator:BANK2:X?
ATTenuator:BANK2:Y?
11713B/C Operating and Service Manual 69
6 Remote Operations
To Set Supply Voltage
The commands below set the supply voltage for each bank.
CONFigure:BANK1 {OFF|P5v|P15v|P24v|USER}
CONFigure:BANK2 {OFF|P5v|P15v|P24v|USER}
To Query Supply Voltage
The commands below query the supply voltage set for each bank. Value to be returned: OFF, P5, P15, P24 or USER.
CONFigure:BANK1?
CONFigure:BANK2?
To Query Number of Relay Cycles
The command below query the number of relay cycles for each channel. Value to be returned: number of cycles.
DIAGnostic:RELay:CYCles? <channel list>
To Clear Relay Cycles
The commands below clear relay cycle for selected or all channels.
DIAGnostic:RELay:CLEAr <channel list>
DIAGnostic:RELay:CLEAr:ALL
70 11713B/C Operating and Service Manual

Commands Relevant to Switches

To Open/Close Switch
The commands below close/open switching path(s).
[:ROUTe]:CLOSe <channel list>
[:ROUTe]:OPEn <channel list>
[:ROUTe]:CLOSe:ALL
[:ROUTe]:OPEn:ALL
To Query Switch Status
The commands below query the status of switching path(s). Value to be returned: “1” if true and “0” if false.
[:ROUTe]:CLOSe? <channel list>
[:ROUTe]:OPEn? <channel list>
Remote Operations 6
To Set Supply Voltage
The commands below set the supply voltage for each bank.
CONFigure:BANK1 {OFF|P5v|P15v|P24v|USER}
CONFigure:BANK2 {OFF|P5v|P15v|P24v|USER}
To Query Supply Voltage
The commands below query the supply voltage set for each bank. Value to be returned: OFF, P5, P15, P24 or USER.
CONFigure:BANK1?
CONFigure:BANK2?
To Set TTL ON/OFF
The commands below set TTL ON/OFF for each bank.
CONFigure:BANK1:TTL {OFF|ON}
CONFigure:BANK2:TTL {OFF|ON}
To Query TTL status
The commands below query the supply voltage status for each bank. Value to be returned: “1” if true and “0” if false.
CONFigure:BANK1:TTL?
CONFigure:BANK2:TTL?
11713B/C Operating and Service Manual 71
6 Remote Operations
To Query Number of Relay Cycles
The command below query the number of relay cycles for each channel. Value to be returned: number of cycles.
DIAGnostic:RELay:CYCles? <channel list>
To Clear Relay Cycles
The commands below clear relay cycle for selected or all channels.
DIAGnostic:RELay:CLEAr <channel list>
DIAGnostic:RELay:CLEAr:ALL
72 11713B/C Operating and Service Manual

Programming Guide (Legacy)

Data Message Input Format

Remote Operations 6
NOTE
NOTE
Legacy commands cannot be utilize to control Bank 2 of 11713C.
The programming data string for the driver consists of a string of ASCII coded characters composed of one or both of the following fields:
[Adm] [Bdn]
where: A or a = general ON command; B or b = general OFF command
dm = any digits 0 through 9 (excluding dn)
dn = any digits 0 through 9 (excluding dm)
The driver responds to an alphanumeric data string such as “A12B34” when the driver is in the remote state and addressed to listen.
The alpha characters in the data string can be either upper or lower case.
NOTE
NOTE
11713B/C Operating and Service Manual 73
The maximum number of character in a data string is 12.
The data string is further defiined as follows:
Ax Bx = data string for ATTEN X (x can be none, one or more of digits 1, 2, 3 or 4)
Ay By = data string for ATTEN Y (y can be none, one or more of digits 5, 6, 7 or 8)
If Ax (Ay) uses a digit, then that digit may not be used in Bx (By).
A/B 9 = legacy command for switch S9
A/B 0 = legacy command for swtich S0
6 Remote Operations

SCPI and Legacy Commands Compatibility

Front Panel Rear
Pushbutton
Number
1
2
3
4
5
6
7
8
9ON
Pushbutton
LED
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
OFF
ON
OFF
Connector
*
ATTEN X 5
ATTEN Y 5
Connector
Panel
Number
S9 A (3)
Pin
6 7 8
9 10 11 12
6
7
8
9 10 11 12
B (4)
Command Result
(Legacy) A
(SCPI) [:ROUTe]:CLOSe: <channel list>
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
G
+Vcc
(Legacy) B
(SCPI) [:ROUTe]:OPEn: <channel list>
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
+Vcc
G
0ON
OFF
ON
OFF
*Note:
LED lights on = attenuation applied (for attenuators)/switching path opened (for switches).
LED lights off = no attenuation (for attenuators)/switching path closed.
S0 A (3)
B (4)
G
+Vcc
+Vcc
G
Legends
H = Open (Low)
L = Closed (High)
G = Connected to ground
74 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
7 Remote Interface Configurations
Connecting the 11713B/C to Your Computer 76
“Connecting Over GPIB “Connecting Over USB “Connecting Over LAN
Exploring the 11713B/C Web Interface Over LAN 85
“Launching the Web Interface “Displaying the Browser Web Control Page
This chapter provides you information and steps to achieve remote operations of Agilent 11713B/C attenuator/switch drivers via GPIB, USB and LAN. You will also be introduced to 11713B/C web interface over LAN.
Agilent Technologies
75
7 Remote Interface Configurations

Connecting the 11713B/C to Your Computer

NOTE
NOTE
To easily connect the 11713B/C to your PC, configure and verify your connection, you can use the Agilent IO Library Suite, the E2094M Agilent IO Libraries for Windows, or and equivalent.
Agilent IO Libraries Suite for Windows 98/2000/ME/XP. For information and to install, use the Automation Ready CD with the Agilent IO Libraries Suite, which is shipped with the 11713B/C.
E2094M Agilent IO Libraries Suite for Windows 98/NT/2000/ME/XP. You can find information and install this application via the Web at www.agilent.com/find/iolib.
You can also access other information about Agilent IO Libraries at www.agilent.com/find/iolib.
The procedures in this section refer to the Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. If you have installed the IO Libraries Suite, you can
access the Connectivity Guide via the Agilent IO Libraries Control icon. Or, you can access the Connectivity Guide via the Web at www.agilent.com/find/connectivity.

Connecting Over GPIB

1 Make sure you have installed the required I/O software on your
computer.
2 Follow the instructions included with your GPIB interface card to
install the GPIB hardware in your computer.
3 Connect the GPIB cable (not provided) between the computer and
the 11713B/C.
4 Make sure power is applied to your computer and verify that the
operating system is fully booted. Then apply power to 11713B/C.
5 Configure your GPIB hardware as instructed by the documentation
provided by the hardware venfor.
6 Use the Connection Expert utility of the Agilent IO Libraries Suite
to add the 11713B/C and verify a connection.
NOTE
If you have installed any other I/O software, refer to documentation included with that software.
76 11713B/C Operating and Service Manual
Remote Interface Configurations 7
The 11713B/C is shipped from the factory with a default GPIB
7
address of 28. To change the address, refer to table below:
Step Action Illustration
1 Changing
GPIB address (integer between 0 to
30)
a At IO SETUP screen,
press GPIB ADDR softkey
b Current GPIB address
is display on screen (default address is 28)
c Use navigation key to
change value
d Press SAVE ADDR to
save and GPIB ADDR SAVED is displayed at the bottom of LCD
e Press BACK soffkey to
return to IO SETUP
8 Now you can use various programming environments to control
the 11713B/C. For an overview on programming instruments via GPIB, refer to the Connectivity Guide.
11713B/C Operating and Service Manual 77
7 Remote Interface Configurations

Connecting Over USB

The USB interface requires no front panel configuration.
NOTE
NOTE
NOTE
Before connecting the USB cable, make sure that I/O software is installed on your computer. See “Connecting the 11713B/C to Your Computer" on page 76 for more information on installing the Agilent IO Libraries software.
1 After I/O software is installed on your computer, connect a
11713B/C to your computer using a Type Mini- B 5 pin USB cable.
2 Make sure power is applied to your computer and verify that the
operating system is fully booted. Then apply power to the 11713B/C.
3 The Found New Hardware Wizard will automatically start and
guide you through configuring the 11713B/C as a USB device. To install the software automatically, accept all defaults.
If you have installed the Agilent I/O Libraries Software, this also install the required low-level software drivers. Therefore you do not need to insert the CD when the Found New Hardware Wizard instructs you to do so.
If you have installed any other I/O interface software, refer to documentation included in that software.
4 Use the Connectivity Expert utility of the Agilent I/O Libraries
Suite to verify that the 11713B/C is displayed under the USB interface.
5 Now you can use various programming environments to control
the 11713B/C. For an overview on programming instructions via USB, refer to the Connectivity Guide.
78 11713B/C Operating and Service Manual

Connecting Over LAN

The 11713B/C has two LAN operating modes:
Auto mode
Manual mode
Selecting the LAN Network Type
You can connect and configure your 11713B/C for Site LAN or isolated (non- site) LAN operation.
A Site LAN network is defined as a local area network (LAN) in which computers and LAN- enabled instruments are connected to a site LAN (workgroup LAN, Intranet, or enterprise LAN) via optional routers, hubs, and/or switches.
An Isolated LAN network is defined as a local area network (LAN) in which computers and LAN- enabled instruments are not connected to a site LAN.
Remote Interface Configurations 7
11713B/C Operating and Service Manual 79
7 Remote Interface Configurations
Auto Mode
This mode is used when you are connecting 11713B/C via site LAN. In auto mode, the IP Adress, Subnet Mask, and Default Gateway values are obtained from the DHCP server or automatically assigned and the values cannot be configured from the front panel.
To prepare 11713B/C for auto mode,
Step Action Illustration
1 Ensure your
computer and 11713B/C are turned on and connected to LAN
2 Verif y
11713B/C LAN configuration is ready for auto mode
a Connect your computer and 11713B/C to LAN outlets using standard
LAN cable
b Make sure power is applied to your computer and verify that the
operating system is fully booted
c Apply power to 11713B/C and background LED color turns red,
indicating standby mode.
d Allow 11713B/C to warm up for 15 seconds, then press the ON
button (LED turn from red to green)
a Press Menu/Enter
button on the front panel of 11713B/C
b Select IO using
navigation keys and press Menu/Enter
c At IO SETUP screen,
press LAN CONFIG softkey
d At LAN
CONFIGURATION screen, ensure IP Configuration is AUTO
e If needed, press
Auto/Manual softkey
until AUTO is displayed and SAVE
f Press BACK softkey to
go back to IO SETUP screen
80 11713B/C Operating and Service Manual
Remote Interface Configurations 7
Step Action Illustration
3 Verify LAN
status on 11713B/C
a At IO SETUP screen,
select IP ADDR softkey
b At IP ADDRESS
screen, select LAN
STAT softkey
c Verify values
CURRENT IP ADDRESS, CURRENT SUBNET and CURRENT GATEWAY are valid
d Ensure LAN STATUS
is NORMAL
e If LAN STATUS is
FAULT, check your
LAN connection and then press Restart
Netw softkey and wait
f Press BACK softkey to
go back to IP ADDRESS screen
NOTE
Now you can use the 11713B/C Web Browser Interface to access and control the instrument. See “Exploring the 11713B/C Web Interface
Over LAN" on page 85.
If you plan to program the instrument over LAN or use such programs, make sure you have installed the required I/O software on your computer.
Use the Connectivity Expert utility of the Agilent IO Libraries Suite to add the 11713B/C and verify a connection. When identifying the instrument, it is easiest if you use the IP address noted in step 3 above.
If you have installed any other I/O software, refer to documentation included with that software.
Now you can use various programming environments to control the 11713B/C. For an overview on programming instructions via USB, refer to the Connectivity Guide.
11713B/C Operating and Service Manual 81
7 Remote Interface Configurations
Manual Mode
In a manual mode, you must set up the IP Address, Subnet Mask and Default Gateway that is compatible with your network infrastructure. If it is not correctly setup, 11713B/C will not be visible on your network.
To prepare 11713B/C for manual mode,
Step Action Illustration
1 Ensure your
computer is connected to 11713B/C and turned on.
2 Verif y
11713B/C LAN configuration is ready for manual mode
a Connect your computer to 11713B/C using CAT5 crossover cable b Make sure power is applied to your computer and verify that the
operating system is fully booted
c Apply power to 11713B/C and background LED color turns red,
indicating standby mode.
d Allow 11713B/C to warm up for 15 seconds, then press the ON
button (LED turn from red to green)
a Press Menu/Enter
button on the front panel of 11713B/C
b Select IO using
navigation keys and press Menu/Enter
c At IO SETUP screen,
press LAN CONFIG softkey
d At LAN
CONFIGURATION screen, ensure IP Configuration is
Manual
e If needed, press
Auto/Manual softkey
until Manual is displayed and SAVE
f Press BACK softkey to
go back to IO SETUP screen
82 11713B/C Operating and Service Manual
Remote Interface Configurations 7
Step Action Illustration
3 Manually
enter IP ADDRESS, SUBNET and GATEWAY
4 Verify LAN
status on 11713B/C.
a At IO SETUP screen,
press IP ADDR softkey
b At IP ADDRESS
screen, press
MANUAL IP softkey
c At MANUAL IP
ADDRESS screen, enter IP ADDRESS,
SUBNET and GATEWAY (scroll
using NEXT ITEM softkey)
d Then SAVE. e Press BACK softkey to
go back to IP ADDRESS screen
a At IO SETUP screen,
select IP ADDR softkey
b At IP ADDRESS
screen, select LAN
STAT softkey
c Verify values
CURRENT IP ADDRESS, CURRENT SUBNET and CURRENT GATEWAY are valid
d Ensure LAN STATUS
is NORMAL
e If LAN STATUS is
FAULT, check your
LAN connection and then press Restart
Netw softkey and wait
f Press BACK softkey to
go back to IP ADDRESS screen
11713B/C Operating and Service Manual 83
7 Remote Interface Configurations
Now you can use the 11713B/C Web Browser Interface to access and control the instrument. See “Exploring the 11713B/C Web Interface
Over LAN" on page 85.
If you plan to program the instrument over LAN or use such programs, make sure you have installed the required I/O software on your computer.
Use the Connectivity Expert utility of the Agilent IO Libraries Suite to add the 11713B/C and verify a connection. When identifying the instrument, it is easiest if you use the IP address noted in step 3 above.
NOTE
If you have installed any other I/O software, refer to documentation included with that software.
Now you can use various programming environments to control the 11713B/C. For an overview on programming instructions via USB, refer to the Connectivity Guide.
84 11713B/C Operating and Service Manual

Exploring the 11713B/C Web Interface Over LAN

Remote Interface Configurations 7
You can use the 11713B/C’s Web Browser Interface for remote access and control of the instrument via a Java as Microsoft® Internet Explorer. Using the Web Interface, you can configure, troubleshoot and monitor your system remotely. This section provides an overview of the 11713B/C Web Interface.
NOTE
The following tasks assume you have configured the 11713B/C for LAN communication and have verified connection to a LAN network. See “Connecting
Over LAN" on page 79 for more information.

Launching the Web Interface

1 Open your Internet browser from your computer
2 From the Tools>Internet Options menu, navigate to Connections
(exact navigation depends on your browser), and then select LAN Settings
3 From the LAN Settings dialog, select/activate bypass proxy server
for local addresses (exact terminology depends on your browser)
4 Exit the Options window
5 Enter the IP address of the 11713B/C in the Address field and
press return. Use the 11713B/C’s front panel utility menu to read the IP address
- enabled Web browser, such
NOTE
6 After entering the appropriate IP address, the 11713B/C Web
Interface’s Welcome Window should appear
The procedure in this section will help you understand tasks commonly performed using the 11713B/C Web Interface. For additional help on using the interface, click the ?Help with the Page tab on the lower-left cornner of the Web Interface window.
11713B/C Operating and Service Manual 85
7 Remote Interface Configurations
Figure 13 11713B/C Web Interface’s Welcome Window

Navigation Bar

Welcome page Display various setting information
Browser Web Control Virtual front panel and sending/reading
SCPI commands
View & Modify Configuration Displays and modifies various setting information
Get Image Receives image
Print Display Calls the printing function of the browser
Help with this Page Displays the Help file
86 11713B/C Operating and Service Manual

Displaying the Browser Web Control Page

1 From the Welcome Window, click the Browser Web Control tab
on the left side of the window
2 You will notice a pop- up window, requesting for password
3 Enter password (default password is “agilent”) and click submit
4 You can view the virtual front panel of 11713B/C
Remote Interface Configurations 7
5 Now, you can control the switching devices remotely, similar to
the front panel of the actual instrument
Figure 14 11713B/C Brower Web Control Interface
11713B/C Operating and Service Manual 87
7 Remote Interface Configurations
88 11713B/C Operating and Service Manual
Agilent 11713B/C Attenuator/Switch Drivers Operating and Service Manual
8 Replacing Major Assemblies
Replaceable Parts 90 Top Module Cover Removal/Replacement 94 Front Panel Module Removal/Replacement 95 Rear Panel Module Removal/Replacement 97 PPMC PCA Removal/Replacement 99 Fan Module Removal/Replacement 100 Motherboard Removal/Replacement 101 Power Supply Module Removal/Replacement 103
In this chapter, you will procedures for removal and replacement of major assemblies in 11713B/C and list of replaceable parts.
Agilent Technologies
89
8 Replacing Major Assemblies

Replaceable Parts

Tab le 13 Major Assemblies Reference Designator and Part Number
Reference Designation Agilent Part Number Description
M1 34401-86020 Bumper kit (front and rear)
M2 5041-7717 Top clamshell
M3 5041-7718 Bottom clamshell
M4-1 11713-60021 Front panel module (11713C)
M4-2 11713-60027 Front panel module (11713B-STD)
M4-3 11713-60035 Front panel module (11713B-LXI)
M5-1 11713-60027 Rear panel module (11713C)
M5-2 11713-60034 Rear panel module (11713B-STD)
M5-3 11713-60036 Rear panel module (11713B-LXI)
M6 11713-60038 8-pin cable assembly
Tab le 14 Front Panel Module Reference Designator and Part Number
Reference Designation Agilent Part Number Description
A1 2090-0964 LCD module
A2 11713-60020 Front panel board
A3-1 11713-40001 Keypad (11713C)
A3-2 11713-40002 Keypad (11713B)
A4-1 11713-20017 Window
A5-1 11713-00025 Front panel bezel (11713C)
A5-2 11713-00028 Front panel bezel (11713B)
A6-1 11713-80005 Front panel label (11713C)
A6-2 11713-80006 Front panel label (11713B-STD)
A6-3 11713-80008 Front panel label (11713B-LXI)
A7 0515-0430 Screw M3
A8 11713-00034 EMI gasket for front panel
90 11713B/C Operating and Service Manual
Replacing Major Assemblies 8
Tab le 15 Rear Panel Module Reference Designator and Part Number
Reference Designation Agilent Part Number Description
B1-1 11713-60023 Rear panel plate (11713C)
B1-2 11713-60030 Rear panel plate (11713B-STD)
B1-3 11713-60031 Rear panel plate (11713B-LXI)
B2 0960-2776 Line filter module
B3 E5810-61605 GPIB cable assembly
B4-1 E3631-20011 Red top binding post
B4-2 E3631-20012 Green top binding post (11713C)
B4-3 E3631-20013 Black top binding post
B8 0380-4870 Hex screw standoff
B7 0510-0079 Retaining ring
B9 0515-0390 Screws
B6 2360-0198 Flathead screw
B5 2950-0144 Plastic hex nut
B10 11713-20018 EMI absorber
B11 11713-00032 GPIB gasket
B12-1 11713-60039 10-pin cable assembly (11713C)
B12-2 11713-60041 10-pin cable assembly (11713B)
B13 11713-60040 2-pin cable assembly (11713C)
B14-1 11713-60022 Rear panel board (11713C)
B14-2 11713-60028 Rear panel board (11713B)
Tab le 16 PPMC PCA Module Reference Designator and Part Number
Reference Designation Agilent Part Number Description
C1 0515-1940 Screw M2.5
C2 11713-60048 PPMC PCA - programmed
C3 11713-00035 EMI bracket
C4 0636-0170 RFI strip
11713B/C Operating and Service Manual 91
8 Replacing Major Assemblies
Tab le 17 Fan Module Reference Designator and Part Number
Reference Designation Agilent Part Number Description
D1 11713-60025 Fan assembly
92 11713B/C Operating and Service Manual
Replacing Major Assemblies 8
Tab le 18 Motherboard Reference Designator and Part Number
Reference Designation Agilent Part Number Description
E1 0515-0433 Screw M4
E2 11713-60046 Motherboard - programmed
Tab le 19 Power Supply Module Reference Designator and Part Number
Reference Designation Agilent Part Number Description
F1 3030-1382 Screw
F2 3050-2287 Spring washer
F3 3050-2326 Flat washer
F4 11713-00033 PSU cover
F5 0515-0372 Screws M3
F6 11713-60032 Power supply
11713B/C Operating and Service Manual 93
8 Replacing Major Assemblies

Top Module Cover Removal/Replacement

CAUTION

To Re m o v e Top Mod u l e C over

NOTE
This module contains components that can be damaged or destroyed by electrostatic discharge. It should be serviced at a static-safe workstation. Refer to “preparing a static-safe workstation” in chapter 9.
1 Remove bumper kit (M1)
2 Turn the driver over with bottom facing up
3 Loosen captive screws with T20 Torx bit
Captive screws are torqued with 21 in-lbs.
4 Turn the driver over with top facing up
5 Gently, lift top clamshell (M2)
6 Disconnect:
8- pin cable (M6) from J5 connector on motherboard
Line module cables (green/blue/brown) from power supply
module
7 Now you can remove the top module cover

To Replace Top Module Cover

1 Hold top clamshell (M2) near the driver
2 Connect:
8- pin cable (M6) from J5 connector on motherboard
Line module cables (green/blue/brown) from power supply
module
3 Gently, place and secure the top clamshell (M2) onto the bottom
clamshell (M3)
4 Turn the driver over with bottom facing up
5 Tighten captive screws using 21 in- lbs torque driver/wrench with
T20 Torx bit
6 Fix bumper kit (M1) onto driver, as needed
94 11713B/C Operating and Service Manual

Front Panel Module Removal/Replacement

Replacing Major Assemblies 8
CAUTION

To Remove Front Panel Module

This module contains components that can be damaged or destroyed by electrostatic discharge. It should be serviced at a static-safe workstation. Refer to “preparing a static-safe workstation” in chapter 9.
1 Remove top clamshell (M2), refer “To Remove Top Module
Cover" on page 94
2 Disconnect:
50- pin cable from J6 connector on motherboard
LCD flex cable from J8 connector on motherboard
LCD power cable (red/black) from J11 connector on
motherboard
Front Panel Module
50-Pin Cable
LCD Flex Cable
LCD Power Cable
3 Slide front panel module (M4) upwards to detach from bottom
clamshell (M3)
4 Front panel module (M4) can be further disassembled to the
following parts/modules:
LCD module (A1)
Front panel board (A2)
Keypad (A3)
Window (A4)
Front panel bezel (A5)
Front panel label (A6)
11713B/C Operating and Service Manual 95
8 Replacing Major Assemblies
A5
A5
A6

To Replace Front Panel Module

1 Slot front panel module (M4) into groove at the bottom clamshell
(M3)
2 Connect the following cables to motherboard:
50- pin cable from J6 connector on motherboard
A1
A7
A2
A3
A4
LCD flex cable from J8 connector on motherboard
LCD power cable (red/black) from J11 connector on
motherboard
3 Replace top clamshell (M2), refer “To Replace Top Module
Cover" on page 94
96 11713B/C Operating and Service Manual

Rear Panel Module Removal/Replacement

Replacing Major Assemblies 8
CAUTION

To Remove Rear Panel Module

This module contains components that can be damaged or destroyed by electrostatic discharge. It should be serviced at a static-safe workstation. Refer to “preparing a static-safe workstation” in chapter 9.
1 Remove top clamshell (M2), refer “To Remove Top Module
Cover" on page 94
2 Disconnect:
50- pin cable from J7 connector on motherboard
10- pin cable (B12) from J12 connector on motherboard
GPIB cable assembly from PPMC PCA
2- pin (B13) cable from J9 connector on motherboard (11713C
only)
Rear panel module
50-pin cable
GPIB cable assembly
10-pin cable (B12)
3 Slide rear panel module (M5) upwards to detach from bottom
clamshell (M3)
4 Rear panel module (M3) can be further disassembled to the
following parts/modules:
Rear panel plate (B1)
Line filter module (B2)
GPIB assembly (B3)
Binding posts (B4)
Rear panel board (B14)
11713B/C Operating and Service Manual 97
8 Replacing Major Assemblies

To Replace Rear Panel Module

1 Slot rear panel module (M5) into groove at the bottom clamshell
2 Connect:
3 Replace top clamshell (M2), refer “To Replace Top Module
(M3)
50- pin cable from J7 connector on motherboard
10- pin cable (B12) from J12 connector on motherboard
GPIB cable assembly from PPMC PCA
2- pin (B13) cable from J9 connector on motherboard (11713C
only)
Cover" on page 94
98 11713B/C Operating and Service Manual

PPMC PCA Removal/Replacement

Replacing Major Assemblies 8
CAUTION

To Re m o v e PPMC P C A

This module contains components that can be damaged or destroyed by electrostatic discharge. It should be serviced at a static-safe workstation. Refer to “preparing a static-safe workstation” in chapter 9.
1 Remove top clamshell (M2), refer “To Remove Top Module
Cover" on page 94
2 Remove rear panel module (M5), refer to “To Remove Rear Panel
Module" on page 97
3 Remove screws (C1) that secure the PPMC PCA (C2) onto the
motherboard
4 Remove EMI bracket (C3) and RFI strip (C4)
5 PPMC PCA (C2) can now be removed
PPMC PCAPPMC PCA (C2)
EMI bracket (C3)
RFI strip (C4)

To Replace PPMC PCA

1 Press fit the PPMC PCA (C2) onto the motherboard carefully.
2 Place EMI bracket (C3) on top of the PPMC PCA
3 Secure the parts with screws (C1)
4 Stick RFI strip (C4) on EMI bracket (C3)
5 Replace rear panel module, refer to “To Replace Rear Panel
Module" on page 98
6 Replace to clamshell, refer “To Replace Top Module Cover" on
page 94
11713B/C Operating and Service Manual 99
8 Replacing Major Assemblies

Fan Module Removal/Replacement

CAUTION

To Re m o v e Fan Mo d u l e

This module contains components that can be damaged or destroyed by electrostatic discharge. It should be serviced at a static-safe workstation. Refer to “preparing a static-safe workstation” in chapter 9.
1 Remove top clamshell (M2), refer “To Remove Top Module
Cover" on page 94
2 Disconnect cable from J10 connector on motherboard
3 Fan module (D1) can now be removed
Fan assembly (D1)

To Re p l a c e Fan m o d u l e

1 Press fit fan module (D1) into slot
2 Connect cable to J10 connector in motherboard
3 Replace top clamshell (M2), refer “To Replace Top Module
Cover" on page 94
100 11713B/C Operating and Service Manual
Loading...