Agilent 11713B Operating and Service Manual KT

Keysight 11713B/C Attenuator/Switch Driver
Operating and Service Manual
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 2007-2020 No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation into a for­eign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Manual Part Number
11713-90024
Edition
Edition 4, December 4, 2020
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight products may be down­loaded from the Web. Go to http://www.key-
sight.com/go/conformity. You can then search
by product number to find the latest Declara­tion of Conformity.
U.S. Government Rights
The Software is “commercial computer soft­ware,” as defined by Federal Acquisition Regu­lation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.key-
sight.com/find/sweula. The license set forth in
the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documenta­tion that is not customarily provided to the public; or (2) Relinquish to, or otherwise pro­vide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, dis­play, or disclose commercial computer soft­ware or commercial computer software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCU­MENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXI­MUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRAN­TIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFOR­MATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRAN­TIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCU­MENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CON­FLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indi­cated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight 11713B/C Operating and Service Manual

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST, formerly NBS), to the extend allowed by the Institute’s calibration facility, and to the calibration facilities of the other International Standards Organization members.
Keysight 11713B/C Operating and Service Manual 3

Safety Considerations

WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer's failure to comply with these requirements.
BEFORE APPLYING POWER
– Verify that the correct fuse is installed.
– Ensure the mains supply voltage fluctuation do not exceed ±10% of the nominal supply
voltage.
GROUND THE INSTRUMENT
This product is a Safety Class I instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the AC power supply mains through a three-conductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed pole) of the AC power lines (supply mains).
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE OR WET ENVIRONMENTS
Do not operate the device around flammable gases or fumes, vapor, or wet environments.
DO NOT OPERATE DAMAGED OR DEFECTIVE INSTRUMENTS
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
USE THE POWER CORD PROVIDED
Use the device with the power cord provided with the shipment.
4 Keysight 11713B/C Operating and Service Manual
WARNING
USE THE DEVICE AS SPECIFIED
WARNING
CAUTION
CAUTION
If the device is used in a manner not specified by manufacturer, the device protection may be impaired.
Keysight 11713B/C attenuator/switch drivers are designed for indoor use and in an area with low condensation.
CLEAN WITH SLIGHTLY DAMPENED CLOTH
Clean the outside of the instrument with a soft, lint-free, slightly dampened cloth. Do not use detergent, volatile liquids, or chemical solvents.
The instrument is designed for use in Overvoltage Category II and Pollution Degree 2.

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
When you see this symbol on your instrument, you should refer to the instrument’s instruction manual for important information.
This symbol indicates the protective earth (ground) terminal.
This symbol indicates that the instrument requires alternating current (ac) input.
This symbol indicates the Fuse.
Keysight 11713B/C Operating and Service Manual 5

Regulatory Markings

KCC-REM-ATi-
WN11713B
The CE mark is a registered trademark of the European Community. If it is accompanied by a year, it indicates the year the design was proven.
This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product (CISPR 11, Clause 4).
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
This symbol is a South Korean Class A EMC Declaration. This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.

South Korean Class A EMC Declaration

Information to the user:
This instrument has been conformity assessed for use in business environments. In a residential environment, this equipment may cause radio interference.
This EMC statement applies to the equipment only for use in business environment.
The CSA mark is a registered trademark of the Canadian Standards Association.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.

Safety and EMC Requirements

6 Keysight 11713B/C Operating and Service Manual
This instrument is designed to comply with the following safety and EMC (Electromagnetic Compatibility) requirements:
– Low Voltage Directive 2014/35/EU
– EMC Directive 2014/30/EU

Waste Electrical and Electronic Equipment (WEEE) Directive

This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit
http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/11713
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight 11713B/C Operating and Service Manual 7
8 Keysight 11713B/C Operating and Service Manual
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK

Table of Contents

Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
South Korean Class A EMC Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety and EMC Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . . . . . . . . . . . . . . 7
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1Introduction
Key Features of Keysight 11713B/C Attenuator/Switch Drivers . . . . . . . . . . . . . . . . . . 16
Compatible Keysight Attenuators and Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Connecting Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
11713B Front and Rear Panels at a Glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
11713C Front and Rear Panels at a Glance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
2Installation
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Preparing for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Connecting to Keysight Attenuators and Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3 Specifications
General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4 Verification
Operator’s Check for Local Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operator’s Check for Remote Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5 Local Operations
Getting Started with the 11713B/C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Main Menu of the 11713B/C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Save/Recall State Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6 Remote Operations
Configuring Remote Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Programming Guide (SCPI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Programming Guide (Legacy) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
SCPI and Legacy Commands Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7 Remote Interface Configurations
Keysight 11713B/C Operating and Service Manual 9
Connecting the 11713B/C to your Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Exploring the 11713B/C Web Interface over LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
8 Servicing the Attenuator/Switch Driver
Preparing a Static-Safe Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Maintenance and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
10 Keysight 11713B/C Operating and Service Manual

List of Figures

Figure 1-1 11713B front panel features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Figure 1-2 11713B rear panel features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Figure 1-3 11713C front panel features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Figure 1-4 11713C rear panel features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Figure 2-1 Handle positioning for bench operation . . . . . . . . . . . . . . . . . . . . . . . 27
Figure 2-2 Illustrations of steps 2 and 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 2-3 Illustration of completed setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Figure 2-4 Typical connection for a programmable four-section attenuator . . . . 33
Figure 2-5 Typical connection for 8762 and 8765 series coaxial switches . . . . . 34
Figure 2-6 Typical connection for 8761 series coaxial switches . . . . . . . . . . . . . . 35
Figure 2-7 Typical connection for relay driving circuit . . . . . . . . . . . . . . . . . . . . . 36
Figure 4-1 Switching system for verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Figure 6-1 Example of control using Keysight IO Libraries Suite . . . . . . . . . . . . . 64
Figure 7-1 11713B/C Web Interface’s Welcome Window . . . . . . . . . . . . . . . . . . 79
Figure 7-2 11713B/C Browser Web Control Interface . . . . . . . . . . . . . . . . . . . . . 81
Keysight 11713B/C Operating and Service Manual 11
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
12 Keysight 11713B/C Operating and Service Manual

List of Tables

Table 1-1 Key features of 11713B/11713C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Table 1-2 Compatible Keysight switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Table 1-3 Compatible Keysight attenuators . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Table 1-4 Connecting accessories for Keysight 11713B/11713C . . . . . . . . . . . . 18
Table 2-1 Summary of switches and attenuators connections to 11713B/C . . . 30
Table 2-2 11713B/C front panel and back panel properties . . . . . . . . . . . . . . . . 32
Table 3-1 11713B/C supplemental characteristics . . . . . . . . . . . . . . . . . . . . . . . 38
Table 3-2 11713B/C environmental specifications . . . . . . . . . . . . . . . . . . . . . . . 39
Table 4-1 Recommended test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Table 5-1 Connection between the 11713B/C and programmable attenuators 51
Table 5-2 Connection between the 11713B/C and switches . . . . . . . . . . . . . . . 51
Keysight 11713B/C Operating and Service Manual 13
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight 11713B/C Operating and Service Manual
Keysight 11713B/C Attenuator/Switch Drivers Operating and Service Manual

1 Introduction

Key Features of Keysight 11713B/C Attenuator/Switch Drivers 16 Compatible Keysight Attenuators and Switches 17 Connecting Accessories 18 11713B Front and Rear Panels at a Glance 18 11713C Front and Rear Panels at a Glance 21
This chapter provides an overview of the Keysight 11713B/C attenuator/switch drivers which includes the instruments’ functions and capabilities, compatibility with Keysight switching components, and physical appearances.
15
1Introduction
NOTE

Key Features of Keysight 11713B/C Attenuator/Switch Drivers

The 11713B attenuator/switch driver is a GPIB compatible instrument that concurrently drives up to two four-section programmable step attenuators and two microwave coaxial switches, or up to 10 SPDT switches. The 11713B is fully backward compatible with 11713A in terms of functionality and fit. Connectivity using USB and LAN are optional.
The 11713C attenuator/switch driver is a GPIB/USB/LAN compatible instrument that concurrently drives up to four four-section programmable step attenuators and four microwave coaxial switches, or up to 20 SPDT switches. The 11713C comes with tri-voltage selection of +5 V, +15 V and +24 V and also permits user-defined voltage supply capability.
The 11713B/C attenuator/switch drivers output continuous current and do not support pulse drive. Please ensure your switching devices can withstand continuous current or have a built-in current interrupt feature.
Tab le 1 -1 Key features of 11713B/11713C
Key features 11713B 11713C
Manually-controlled using front panel push buttons Yes Yes
Automatically-control through: –GPIB –USB –LAN
Integrated LCD display Yes Yes
Self-contained power supply with current limiting Yes Yes
Common terminal supplies of: – +5 Vdc –+15 Vdc – +24 Vdc
– User-defined
TTL control
[a] For 11713C, maximum user-defined voltage supply is 30 Vdc.
[b] This TTL specification is 2.4 V at 1 mA.
[a]
[b]
Yes Optional Optional
No No Yes No
No Yes
Yes Yes Yes
Yes Yes Yes Yes
16 Keysight 11713B/C Operating and Service Manual

Compatible Keysight Attenuators and Switches

The 11713B/C attenuator/switch drivers are designed to drive the following Keysight attenuators and switches. If you are using attenuators and switches made by another supplier, check the switching characteristics against those specified in Chapter 3, "Specifications". Refer to Keysight 11713B/C Configuration Guide, literature number 5989-7277EN, for the most updated list.
Table 1-2 Compatible Keysight switches
Introduction 1
Keysight model number
8761A/B, 8765A/B/C/D/F (33314A/B/D), N1810UL SPDT, unterminated
8762A/B/C/F (33311A/B/C), N1810TL SPDT, terminated
8763A/B/C (33312A/B/C), N1811TL Bypass, 4-port, terminated
8764A/B/C (33313A/B/C), N1812UL Bypass, 5-port, unterminated
8766K (33366K) SP3T, unterminated
8767K (33367K), 8767M, L7204A/B/C SP4T, unterminated
87104A/B/C/D/E, 87204A/B/C/D/E, L7104A/B/C SP4T, terminated
8768K (33368K), 8768M SP5T, unterminated
8769K (33369K), 8769M, L7206A/B/C SP6T, unterminated
87106A/B/C/D/E, 87206A/B/C/D/E, L7106A/B/C SP6T, terminated
87222C/D/E, L7222C DPDT (transfer), unterminated
87406B Matrix, 4-port, terminated
87606B Matrix, 6-port, terminated
U9397A/C SPDT, terminated, solid state
[a] * Electromechanical switches unless specified
Description
[a]
Table 1-3 Compatible Keysight attenuators
Keysight model number Description
8494G/H (33320G/H), 84904K/L/M (33324K/L) 11 dB, 1 dB steps
8495G/H/K (33321 G/H/K), 84907K/L (33327K/L) 70 dB, 10 dB steps
8496G/H (33322G/H) 110 dB, 10 dB steps
8497K (33323K), 84906K/L (33326K/L) 90 dB, 10 dB steps
84905M 60 dB, 10 dB steps
84908M 65 dB, 5 dB steps
Keysight 11713B/C Operating and Service Manual 17
1Introduction

Connecting Accessories

Various types of connecting accessories are available to drive Keysight attenuators and switches using the 11713B/C attenuator/switch drivers. Tab le 1-4 lists available accessories.
Tab le 1 -4 Connecting accessories for Keysight 11713B/11713C
Connecting accessories Description
11716A
11716C
11713B-001 11713C-001
11713B-101 11713C-101
11713B-201 11713C-201
11713B-301 11713C-301
11713B-401 11713C-401
11713B-501 11713C-501
11713B-601 11713C-601
11713B-701 11713C-701
11713B-801 11713C-801
11713B-908 11713C-908
11713B-909 11713C-909
Interconnect kit (Type-N connectors)
Interconnect kit (SMA connectors)
11764-60004, Viking connector to 10-pin DIP connector, 60 inches long
8120-2703, Viking connector to Viking connector, 60 inches long
5061-0969, Viking connector to 12-pin conductor cable, bare wire (for five switches), 60 inches long
11761-60001, Viking connector to (4) ribbon cables, connect up to four switches, 60 inches long
11713-60042, dual Viking connector to 16-pin DIP connector, 60 inches long
11713-60043, Viking connector to (4) 9-pin Dsub connectors, connect up to four switches, 60 inches long
11713-60044, Viking connector to 16-pin DIP connector, 60 inches long
5064-7848, Viking connector to 14-pin DIP connector, 60 inches long
11713-60047, Viking connector to (4) 10-pin DIP connectors, connect up to 4 switches, 50 inches long
5063-9240, rack mount kit for one instrument
5061-9694 & 5063-9212, rack mount kit for two instruments
[a]
[a]
[a] Order this kit to connect two programmable step attenuators in series
Refer to Keysight 11713A/B/C Attenuator/Switch Driver Configuration Guide, literature number 5989-3703EN, for configuration details.

11713B Front and Rear Panels at a Glance

This section briefly describes the function of the front panel keys of 11713B.
18 Keysight 11713B/C Operating and Service Manual
Figure 1-1 11713B front panel features
1 2
3
4
5
6
7
89
10
11
12
1 LCD screen.
Introduction 1
2Softkeys. These unmarked keys are referred to by the text on display next to them.
3Menu/Enter. Press this key to select the highlighted parameter On/Off or select the highlighted
field or go back to the main menu.
4Preset. Press this key to preset the driver.
5Config. Press this key to access the configuration menu. You can set the attenuator type, supply
voltage, and TTL condition through this menu.
6Save/Recall. Press this key to save current settings or recall saved settings.
7Navigation buttons. The arrow keys are used to navigate parameters displayed on the LCD
screen or change parameters such as GPIB address.
8Switches. In the local mode, pushbutton switches 9 and 0 change the position of a coaxial
switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively.
9Attenuator Y. In the local mode, pushbuttons 5, 6, 7, and 8 change the attenuation setting of an
attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel.
10 Attenuator X. In the local mode, pushbuttons 1, 2, 3, and 4 change the attenuation setting of
an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel.
11 On/Standby. Press this key to switch between on and standby. When power is supplied, the
background LED is red. Pressing the key once switches the driver on and the background LED turns to green.
12 Local. Press this key to control the driver from the front panel when it is operating via the
remote interfaces.
Keysight 11713B/C Operating and Service Manual 19
1Introduction
1 2
3
4
7
8
9
10
11
5
6
This section briefly describes the function of the rear panel connectors of 11713B.
Figure 1-2 11713B rear panel features
1ATTEN X. Viking connector for connection to attenuator or switch(es).
2ATTEN Y. Viking connector for connection to attenuator or switch(es).
3S9 A/B. Banana jack connectors for connection to coaxial switch.
4 24 VDC COM. Banana jack connector to provide common +24 Vdc in driving the coaxial
switches connected to S9 and/or S0.
5S0 A/B. Banana jack connectors for connection to coaxial switch.
6 Receptacle. Matches transformer primary to line voltage via power cable.
7Alert symbol. This symbol is used to point out a necessary reference for the user.
8GPIB connector. The interface connector from a source device to a listening device for the
remote mode of operation.
9 LAN connector. The interface connector for LAN cable (option LXI only).
10 USB connector. The interface connector for Type mini B 5-pin USB cable (option LXI only).
11 Instrument markings.
20 Keysight 11713B/C Operating and Service Manual

11713C Front and Rear Panels at a Glance

1 2
3
4
5
6
7
8
9
10
11
12131415
16
17
This section briefly describes the function of the front panel keys of 11713C.
Introduction 1
Figure 1-3 11713C front panel features
1 LCD screen.
2Softkeys. These unmarked keys are referred to by the text on display next to them.
3Navigation buttons. The arrow keys are used to navigate parameters displayed on the LCD
screen or change parameters such as GPIB address.
4Menu/Enter. Press this key to select the highlighted parameter On/Off or select the highlighted
field or go back to the main menu.
5Preset. Press this key to preset the driver.
6Config. Press this key to access the configuration menu. You can set the attenuator type, supply
voltage, and TTL condition through this menu.
7Save/Recall. Press this key to save current settings or recall saved settings.
8 Supply Voltage for Bank 1. Indicates supply voltage setting (background LED in red) for bank
1.
9 Supply Voltage for Bank 2. Indicates supply voltage setting (background LED in red) for bank
2.
10 Switches for Bank 1. In the local mode, pushbutton switches 9 and 0 change the position of a
coaxial switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively, for bank
1.
11 Switches for Bank 2. In the local mode, pushbutton switches 9 and 0 change the position of a
coaxial switch connected to rear panel banana jacks S9 A/B and S0 A/B respectively for, bank
2.
12 Attenuator Y for Bank 1. In the local mode, pushbuttons 5, 6, 7, and 8 change the attenuation
setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel, for bank 1.
Keysight 11713B/C Operating and Service Manual 21
1Introduction
13 Attenuator Y for Bank 2. In the local mode, pushbuttons 5, 6, 7, and 8 change the attenuation
setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN Y connector on the rear panel, for bank 2.
14 Attenuator X for Bank 1. In the local mode, pushbuttons 1, 2, 3, and 4 change the attenuation
setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel, for bank 1.
15 Attenuator X for Bank 2. In the local mode, pushbuttons 1, 2, 3, and 4 change the attenuation
setting of an attenuator or change the position of coaxial switch(es) connected to the ATTEN X connector on the rear panel, for bank 2.
16 On/Standby. Press this key to switch between on and standby. When power is supplied, the
background LED is red. Pressing the key once switches the driver on and the background LED turns to green.
17 Local. Press this key to control the driver from the front panel when it is operating via the
remote interfaces.
22 Keysight 11713B/C Operating and Service Manual
Introduction 1
1
2
3
4
7
8
9
10
11
5
6 12
13
14
15161718
This section briefly describes the function of the rear panel connectors of 11713C.
Figure 1-4 11713C rear panel features
1 ATTEN X Bank 1. Viking connector for connection to attenuator or switch(es), for bank 1.
2 ATTEN X Bank 2. Viking connector for connection to attenuator or switch(es), for bank 2.
3 ATTEN Y Bank 1. Viking connector for connection to attenuator or switch(es), for bank 1.
4 ATTEN Y Bank 2. Viking connector for connection to attenuator or switch(es), for bank 2.
5 S9 A/B Bank 1. Banana jack connectors for connection to coaxial switch, for bank 1.
6 S9 A/B Bank 2. Banana jack connectors for connection to coaxial switch, for bank 2.
7VDC COM Bank 1. Banana jack connector to provide common Vdc in driving the coaxial
switches connected to S9 and/or S0, for bank 1.
8VDC COM Bank 2. Banana jack connector to provide common Vdc in driving the coaxial
switches connected to S9 and/or S0, for bank 2.
9 S0 A/B Bank 1. Banana jack connectors for connection to coaxial switch, for bank 1.
10 S0 A/B Bank 2. Banana jack connectors for connection to coaxial switch, for bank 2.
11 External VDC. Banana jack connector to provide user-defined Vdc, for both banks.
12 Ground. Banana jack connector to provide grounding, for both banks.
13 Receptacle. Matches transformer primary to line voltage via power cable.
14 Alert symbol. This symbol is used to point out a necessary reference for the user.
15 GPIB connector. The interface connector from a source device to a listening device for the
remote mode of operation.
16 LAN connector. The interface connector for LAN cable.
17 USB connector. The interface connector for Type mini B 5-pin USB cable.
18 Instrument markings.
Keysight 11713B/C Operating and Service Manual 23
1Introduction
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 Keysight 11713B/C Operating and Service Manual
Keysight 11713B/C Attenuator/Switch Drivers Operating and Service Manual

2 Installation

Initial Inspection 26 Preparing for Use 27 Connecting to Keysight Attenuators and Switches 30
This chapter provides you important information on how to unpack and check your instrument, how to prepare your instrument for bench operation, and tips on configuring the 11713B/C with Keysight attenuators and switches.
25
2Installation

Initial Inspection

1 Unpack and inspect the shipping container and its contents thoroughly to ensure that nothing
was damaged during shipment. If the shipping container or cushioning material is damaged, the contents should be checked both mechanically and electrically.
2 If the contents are damaged or defective, contact your nearest Keysight Technologies Service
and Support Office. Refer to “Sales and Technical Support” on page 7. Keysight Technologies will arrange for repair or replacement of the damaged or defective equipment. Keep the shipping materials for the carrier's inspection.
3 If you are returning your instrument for service, repackaging the attenuator/switch driver
requires original shipping containers and materials or their equivalents. Keysight Technologies can provide packaging materials identical to the original materials. Refer to Sales and Technical
Support for the Keysight Technologies nearest to you.
26 Keysight 11713B/C Operating and Service Manual

Preparing for Use

Bench operation

Pull the handle outwards, and adjust it into one of the two positions illustrated in Figure 2-1.
– (Top) Handle placed underneath the instrument to assure self- alignment of the instruments
when stacked.
– (Bottom) Handle tilted to raise the front of the instrument for easier viewing of the front panel.
Installation 2
Figure 2-1 Handle positioning for bench operation
Keysight 11713B/C Operating and Service Manual 27