Agilent 11683A Operations Guide

&-c~~.E~r PAiLK &e-d
P ERATING AND SERVICE MANUAL
11683A
Range ,.
Calibrator
General Information ~,
Installation Operation
Adjustments
Replaceable Parts
Service
Manual Changes
. ._........-.- -
- .-..._ .-.
COPYRIGHT AND DISCLAIMER NOTICE
Copyright - Agilent Technologies, Inc. Reproduced with the permission of Agilent
Technologies Inc. Agilent Technologies, Inc. makes no warranty of any kind with regard
to this material including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies, Inc. is not liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material or data.
OPERATING AND SERVICE MANUAL
11683A
Range Calibrator
SERIAL NUMBERS
This manual applies directly to instruments with serial
numbers prefixed 1719A. Withthechangesin theAppendix
added, this manual applies to instruments with serial
numer prefixes 1314A and 1551A.
For additional important informationabout serial numbers
see INSTRUMENTS COVERED BY MANUAL.
HP Model 11683A
Power Cable
(For HP Part Number refer to Figwe 4)
Figure 1. HP Model 11683A Calibrator and Accessories Supplied
I
Calibration Functions:
1, 3,10, 30, and 100 mW.
Range-to-Range
Dimensions:
Weight:
Net, 1.1 kg (2 lb, 8 oz.).
Calibration Uncertainty: i 0.25% in all ranges.
88.9 mm high, 133.3 mm wide, and 215.9 mm deep (3% x 5% x 8% in.).
Table 1. Specifications
Output voltage corresponding to meter readings at 3.10, 30,100, and 300 ,uW;
.Ol@
Model 11683A
Page 1
1. GENERAL INFORMATION
2. This operating and service manual contains information pertaining to incoming inspection,
operation, performance tests, adjustments, and service for the HP Model 11683A Range Calibrator.
3. Equipment recommended for use in perform­ance testsl adjustments, and service to the 11683A is listed in Table 2. Test equipment which meets or exceeds the critical specifications of Table 2 must be used for calibration if the I1683A is expected
to conform to the published specifications.
4. The 116t13A and all supplied accessories are shown in Figure 1. The published specifications are listed in Table 1.
5, Instrument Covered by Manual
6. This i&rum& has a two-part serial
number.
The first four digits and the letter comprise the serial number prefix. The last five digits form the
sequential suffix that is unique to each instrument.
The contents of this manual apply dtiectly to instruments having the same serial number pre­fix(es) as listed under SERIAL NUMBERS on the
title page.
7. An instrument’manufactured after the printing of this manual may have a serial prefix that is not
listed on the title page. This unlisted serial prefix
indicates that the instrument is different from those documented; in this manual. The manual for this instrument is isupplied with a yellow Manual Changes supplement that contains “change inform­ation” that documents the differences.
8. In addition to change information, the
supple-
ment may contain information for correcting
errors in the manual. To keep this manual as current and accurate as possible, Hewlett-Packard
recommends that you periodically request the
latest Manual Changes supplement. The supple­ment for this manual is keyed to this manual’s
print date and part number, both of which appear on the title page. Complimentary copies of the supplement are available from Hewlett-Packard.
9. For information concerning a serial number prefix not listed on the title page or in the Manual Changes supplement, contact your nearest Hewlett­Packard office.
10. Description
11. The 11683A Range Calibrator is used to verify
proper operation of compatible Power Meters such
Instrument
Digital Voltmeter
Oscilloscope
Four-Wire Cable
Table 2. Recommended
Test
-r
Critical Specifications
Readout: 5 digits
DC Measurements
Ranges: 100 mV to 100 V full-scale
Accuracy: i 0.02% Resistance Measurements
(four-wire measurement capability) Ranges: 100 R to 10 k.Q full-scale Sensitivity: 1 ma Accuracy: f. 0.02%
Vertical Amplifier
Bandwidth: DC to 5 MHz
Deflection Factor:
50 mV/divlsion minimum
Attenuator Accuracy: f 2% Time
Base
Time Span/division: 1 ms to 1 s Time base accuracy: i 3%
Recommended Length: 5 feet maximum
Equipment
Model
HP 3455A
HP lSOC/
I801A/ 1821A
(see Figure 2)
use*
P, A T
A, T
P