Inspect the Probe 3
1159A Active Differential Probes Introduction 4
Probe Accessories Supplied 5
Attaching External Attenuators to the Probe 7
!
Specifications and Characteristics 8
To Connect the Probe to the Circuit under Test 12
Recommended Probe Configurations 14
Safety Considerations 17
Service Strategy 18
Performance Verification 19
Adjustment of 10:1 and 20:1 Attenuators 33
Performance Test Record 37
2
1159A 1GHz Active Differential Probe
Inspect the Probe
Inspect the Probe
❏ Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of
the shipment have been checked for completeness and the probe has been
checked mechanically and electrically.
❏ Check the accessories.
Any accessories that were supplied with the probe are listed in “Probe
Accessories Supplied” on page 5.
• If the contents are incomplete or damaged notify your Agilent Sales Office.
❏ Inspect the instrument.
• If there is mechanical damage or defect, or if the probe does not operate
properly or pass performance tests, notify your Agilent Sales Office.
• If the shipping container is damaged, or the cushioning materials show signs
of stress, notify the carrier as well as your Agilent Sales Office. Keep the
shipping materials for the carrier’s inspection. The Agilent office will arrange
for repair or replacement at Agilent’s option without waiting for claim
settlement.
3
1159A 1GHz Active Differential Probe
1159A Active Differential Probes Introduction
1159A Active Differential Probes Introduction
The 1159A is a wide-band differential active probe. The probe features low noise,
low input capacitance, high common mode rejection, and Field Effect Transistor
(FET) buffered inputs in the probe head. User-selectable offset gives the probe
flexibility to measure a large range of signal types. Plug-on attenuators and AC
coupling accessories further extend the application range. Included interconnect
accessories allow connection to surface mount and through-hole components
with minimal signal degradation. The input receptacles in the probe head are
compatible with standard 0.025" (0.635 mm) square pins, which provide a
convenient low-cost method of creating device characterization test fixtures.
The 1159A is ideal for acquiring high speed differential signals such as those
found in disk drive read channels, differential LAN, video, and so on. The high
impedance characteristics of both inputs allow you to use the probe as a FET
probe to make single-ended measurements in digital systems without
introducing a ground loop as a conventional FET probe would.
Differential Amplifiers and CMRR
The 1159A Differential Probe is a high input impedance amplifier. A
characteristic of differential amplifiers is the ability to reject signals that are
common to the two inputs. The common mode rejection ratio (CMRR) is the
measurement of this ability. It is expressed as the ratio between the amplitudes
of the common mode and differential signals that produce equal outputs. If the
differential gain is known, these measurements can be referred to the probe
input. CMRR is usually expressed in dB:
V common mode input Gain×
CMRR in dB20
The ability to reject common mode signals depends on the balance designed into
the probe amplifier. As the frequency of the common mode signal increases, it
becomes harder to balance the amplifier parasitic parameters. This leads to
degradation of the CMRR.
The CMRR of the 1159A Differential probe is specified from the probe tip. This
method of specifying the probe CMMR eliminates the effects of source
impedance, provided the connections from the probe tip to the signal source are
symmetrical.
To order a replaceable part, in the United States and Canada call our toll-free
hotline at 1-877-447-7278, or call your local Agilent Technologies Sales Office.
(includes an adjustment tool)
Connection Kit101154-60004
Trimmer Tool
(0.635 mm square head)
101159-82104
101159-82105
15063-2196
Using the Accessories
The 1159A Differential Probe and accessories provide a variety of ways to
connect to circuitry under test. Any method used to connect the probe signal
inputs to the circuit under test degrades the performance of the probing solution.
Take the following precautions to optimize common mode rejection.
• Maintain tip connection lead length as short as possible and the same
length.
• Follow the same path for wires used to connect the inputs of the probe
to the circuit under test.
• Probes do not have infinite input impedance and do load the circuit
under test. If the impedance of the test points is not identical, unequal
loading will occur. This degrades common mode rejection.
• The ground lead length is not usually critical with a differential probe.
• Carefully consider the ground potential relative to the oscilloscope
ground potential. The potential difference must be within the common
mode range of the probe.
• The DC potential between the AC coupling adapter and the oscilloscope ground must not exceed 42 Vpk.
• Do not cascade the external attenuators.
• Cascade the external AC coupling adapter in the following order: probe,
attenuator, and AC coupling adapter.
6
1159A 1GHz Active Differential Probe
Attaching External Attenuators to the Probe
Other Probe Accessories
The Agilent Wedge was designed to interface directly with the differential probe.
These devices simplify connections to surface mount integrated circuits and have
output pins compatible with the probe tip and attenuator sockets.
Attaching External Attenuators to the Probe
The external attenuators plug directly on to the probe tip. They are calibrated
at the factory to provide the optimum common mode rejection and should not
be swapped between probes.
1159A Attenuator
Div-By-10 1 GHz
Div-By-20 1 GHz
1159A AC Coupler
Always Install Last
The 1159A probe’s best performance is achieved when the probe attenuation is set
to /10 in the Infiniium’s Probes Setup dialog box.
7
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Specifications and Characteristics
Specifications
Input ConfigurationGround Connector
Input CouplingDC AC coupling obtained by installing an AC coupling
Gain Accuracy at 1 kHz2%
!
Maximum Input Voltage
Either input from ground
CMRRat 70 Hz: 80dB
True Differential (+ and inputs), with shield
adapter
< ±42 V
at 1 MHz: 40dB
at 100 MHz: 25dB
at 500 MHz: 19dB
at 1 GHz: 13dB
The following characteristics are valid for the 1159A probe after the probe has
reached operating temperature, which is 20 minutes with power applied in a
environment with stable ambient temperature. The probe must be operating
within the environmental conditions listed in the “Environmental Specifications”
section on page 11, and must have been calibrated within the past 12 months in
a ambient temperature of 23 5 C.
±°
Performance Characteristics
Probe Bandwidth (-3 dB)DC to 1 GHz
Offset Range1.6 V
Rise Time (Probe only)
1:1 Attenuation
Internal switched attenuation only
Input Resistance
(each side to ground)
Input Capacitance (between inputs)
1:1 Attenuation
No external attenuators
Input Capacitance (each side to ground)
1:1 Attenuation
No external attenuators
±
<350 ps
1 M
Ω
<0.85 pF
<1.5 pF
9
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Typical CMRR versus Frequency (Hz)
Typical Noise
10
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Environmental Specifications
OperatingNon-operating
Temperature0 to 50 C-40 to 75 C
HumidityUp to 80% RH at 40 CUp to 80% RH at 75 C
AltitudeUp to 4,600 meters
VibrationRandom vibration 5 to
WeightApproximately 226 g
DimensionsRefer to the drawing shown below
Dimensions
°°
°°
(15,000 feet)
500 Hz, 10 minutes per axis,
0.3 g
rms
Up to 15,000 meters
(50,000 feet)
Random vibration 5 to
500 Hz, 10 minutes per axis, 2.41
g
The method you use to connect the probe to the circuit under test is critical for
ensuring accurate measurements. The following examples examine the effect of
using different lengths of wire at 100MHz to connect the signal source to the
probe tip.
The Impedance of the Source
This is another instance where the symmetry of the differential circuit is
important. The impedance of the source forms a network with the input
impedance of the connection and the probe. This network determines the
frequency response for the measurement. If each side of the differential source
has a different impedance, the frequency response of each side will be different.
This lack of balance is reflected in reduced CMRR. The higher the impedance
of the source, the more critical these parasitic effects.
The Ground Connection
A poorly located ground connection allows ground loops to add to the common
mode signal. The differential probe measures the potential difference between
two locations on a PC board. Usually, it is not necessary to ground the probe.
Whether to ground the probe depends on the magnitude and frequency of the
voltage difference between the oscilloscope ground and the board ground. It is
good practice to maintain a board ground. Without this ground reference, you
could easily exceed the common mode range of the probe.
Probe Offset
The amplifiers in the 1159A probe limit the Differential Mode Range to 400mV.
If the input to the probe is approaching 400mV, there is little offset range available
for positioning the trace on screen. There are two solutions to this problem:
• Attenuate the signal into the probe with the 1/10 or 1/20 attenuator. This
mode of operation will induce some small loss in CMRR.
•Use Position to position the trace on screen.
The added feature of position control independent of Offset allows trace
positioning without calculating how much probe offset range is available. The
trace can be positioned by dragging the trace or positioning the trace under the
Probe menu.
12
1159A 1GHz Active Differential Probe
To Connect the Probe to the Circuit under Test
Both inputs derived
from a common signal
at probe tip.
CMRR = 35.6dB
Both inputs derived
from a common signal
via 5cm coupling lead.
CMRR = 35dB
Probe coupli ng leads
of different length.
Positive input 5.3cm.
Negative input 7.5cm.
CMRR = 33dB
13
1159A 1GHz Active Differential Probe
Recommended Probe Configurations
Recommended Probe Configurations
For best performance, use the following configurations. They are presented in
the recommended order from the most desirable to the least.
NoteThe use of the ground connection is optional for all configurations.
Direct Connection
1159A Probe Tip
Test Point Layout
See the “Test Point
Layout” section for
more information
AC Adapter/Attenuator
Use the attenuator shipped with the probe and marked with the same serial
number for accurate measurements. Do not use the attenuators with other
probes.
1159A Attenuator
Div-By-10 1 GHz
Div-By-20 1 GHz
1159A AC Coupler
Always Install Last
Test Point Layout
See the “Test Point
Layout” section for
more information
14
1159A 1GHz Active Differential Probe
Recommended Probe Configurations
Offset Pins
You may use offset pins with any of the tip adapters.
Offset Pins Rotate to
Adjust Spacing
SMT Lead
You may use SMT leads with any of the tip adapters.
SMT Leads
Maintain
Equal Length
Solder Leads
to Test Points
Ground
Ground
Lead Opt
15
1159A 1GHz Active Differential Probe
Recommended Probe Configurations
Wire Leads
You may use wire leads with any of the tip adapters.
Connect Leads
to Test Points
Ground
Opt
Grabbers
Using grabbers and wire leads results in significant lead length. Expect
measurement quality degradation with fast signals.
0.5 mm or 0.8 mm Grabbers
Opt
Ground
16
1159A 1GHz Active Differential Probe
Safety Considerations
Test Point Layout
Safety Considerations
Read the Safety summary in the warranty pages at the back of this guide before
servicing the instrument. Before performing any procedure, review the safety
information for cautions and warnings.
WARNING Only trained service personnel aware of the hazards involved (for example, fire
!
and electric shock) should perform maintenance on the instrument. When
maintenance can be performed without power applied, the power cord must be
removed from the instrument.
17
1159A 1GHz Active Differential Probe
Service Strategy
Service Strategy
To return the 1159A 1 GHz Differential Probe to optimum performance requires
factory repair. Return all probes to the service group for repair and calibration.
If the probe is under warranty, normal warranty services apply. If the probe is
not under warranty, you can exchange a failed probe for a reconditioned one at
a nominal cost.
To Return the Probe to Agilent for Service
Call (877) 447-7278 for further details and the location of your nearest Agilent
Service Office.
1
Write the following information on a tag and attach it to the probe.
• Name and address of owner.
• Probe model number.
• Probe serial number.
• Description of the service required or failure indications.
2 Return the following accessories with the probe:
• Attenuators
• AC coupling adapter
Retain all other accessories.
3
Return the probe in its case or pack the probe in foam or other shock
absorbing material and place it in a strong shipping container.
You can use the original shipping materials or order materials from an Agilent
Sales Office. If neither are available, place 3 to 4 inches of shock-absorbing
material around the instrument and place it in a box that does not allow
movement during shipping.
4
Seal the shipping container securely.
5 Mark the shipping container as FRAGILE.
In all correspondence, refer to the instrument by model number and full serial
number.
18
1159A 1GHz Active Differential Probe
Performance Verification
Performance Verification
Use this procedure to verify the warranted characteristics of the 1159A 1 GHz
Differential Probe. The recommended performance verification interval for the
1159A is one year. Perform the complete performance verification procedure as
the first step of annual certification. You can complete performance verification
without removing the probe covers. There are no user adjustments available for
calibration. Use the equipment shown in the “Test Equipment Required” section
to complete the performance verification procedures.
1
Perform the steps listed in the “Preliminary Procedure” section on
page 20.
2 Perform the steps listed in the “Test Gain Accuracy at 1 kHz” section
on page 25.
3 Perform the steps listed in the “Test Offset” section on page 26.
4 Perform the steps listed in the “Test Differential Mode Range” section
on page 26.
5 Perform the steps listed in the “CMRR Test Introduction” section on
page 27.
Test Equipment Required
DescriptionMinimum RequirementsPart Number
Infinium Oscilloscope 1.5 GHz54845A
Digital AC/DC VoltmeterDC: 0.1% accuracy
Function GeneratorSine Wave and Square Wave output
Sine Wave Generator50 MHz to 1 GHz
BNC TE9625A
Probe Tip BNC Adapter5063-2143
Calibration Fixture01154-63801
Intelligent Interface01154-63401
High Frequency 10:1 Resistor
Divider Probe
10:1 Resistor Divider Probe
(provided with the 54845A)
Trimmer Tools
(included when attenuators are
purchased separately)
Turn on the oscilloscope, the 1159A, and the other test equipment.
Allow 30 minutes for all test equipment to warm up.
2
Press Default Setup.
3 Perform Probe Tip Cal on the 1161A connected to CH4 of the
oscilloscope (54845A).
4 Save the setup files listed in the “Performance Verification Setup Files”
section on page 20 to the C drive of the 54845A.
These setups can also be saved to drive A and filed for future use.
Performance Verification Setup Files
Save the setup information shown below for the various performance verification
tests to the specified file names on the C drive of the 54845A or to a floppy disk
in drive A. Connect the probes to the specified channels before creating and
storing the setup files.
Setup for Differential Mode Range
File Name: 1159CAL0
Probe Connected to CH11159A
Probe Connected to CH41161A
CH1 and CH4 Sensitivity200 mV/div
CH2 and CH3Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measurements: CH1 and CH4V amplitude under Measure Voltage
Timebase500 Sec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 8 averages
TriggerCH4; level 0 V + ve edge
20
1159A 1GHz Active Differential Probe
Setup for CMRR at 70 Hz
File Name: 1159CAL1
Probe Connected to CH11159A
Probe Connected to CH41161A
CH1 Sensitivity2 mV/div
CH4 Sensitivity5 V/div
Measure AmplitudeCH1 and CH4
CH2 and CH3Off
Offset and Position CH1, CH2, CH3 and CH40 V
Timebase10 mSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 64 averages
TriggerCH4; level 0 V + ve edge
Setup for CMRR at 1 MHz
File Name: 1159CAL2
Probe Connected to CH11159A
Probe Connected to CH41161A
CH1 Sensitivity10 mV/div
CH4 Sensitivity5 V/div
CH2 and CH3Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1 and CH4
Timebase500 nSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 64 averages
TriggerCH4; level 0 V + ve edge
Performance Verification
21
1159A 1GHz Active Differential Probe
Performance Verification
Setup for CMRR at 100 MHz
File Name: 1159CAL3a
Probe Connected to CH1None
Probe Connected to CH41163A
CH1 Sensitivity200 mV/div; 50
CH4 Sensitivity200 mV/div; 50
CH2, CH3, and CH4Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1
Timebase5 nSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 64 averages
TriggerCH4; 0 V + ve edge
Ω
Ω
Setup for CMRR at 100 MHz
File Name: 1159CAL3b
Probe Connected to CH11159A
Probe Connected to CH41163A
CH1 Sensitivity5 mV/div
CH4 Sensitivity200 mV
CH2, CH3, and CH4Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1
Timebase5 nSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 64 averages
TriggerCH4; 0 V+ ve edge
22
1159A 1GHz Active Differential Probe
Setup for CMRR at 500 MHz
File Name: 1159CAL4a
Probe Connected to CH41163A
CH1 Sensitivity200 mV/div; 50
CH4 Sensitivity200 mV/div; 50
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1
Timebase1 nSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 64 averages
TriggerCH4; 0 V
Ω
Ω
Setup for CMRR at 500 MHz
File Name: 1159CAL4b
Probe Connected to CH11159A
Probe Connected to CH41163A
CH1 Sensitivity20 mV/div
CH4 Sensitivity200 mV/div; 50
Offset and Position CH1, CH2, CH3 and CH40 V offset
Measure AmplitudeCH1
Timebase1 nSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 16 averages
TriggerCH4; 0 V
Ω
Performance Verification
23
1159A 1GHz Active Differential Probe
Performance Verification
Setup for CMRR at 1 GHz
File Name: 1159CAL5a
Probe Connected to CH1None
Probe Connected to CH41163A
CH1 Sensitivity200 mV/div; 50
CH4 Sensitivity200 mV/div; 50
CH2, CH3, and CH4Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1
Timebase500 pSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 16 averages
TriggerCH4; 0 V
Ω
Ω
Setup for CMRR at 1 GHz
File Name: 1159CAL5b
Probe Connected to CH11159A
Probe Connected to CH41163A
CH1 Sensitivity50 mV/div
CH4 Sensitivity200 mV/div; 50
CH2, CH3, and CH4Off
Offset and Position CH1, CH2, CH3 and CH40 V
Measure AmplitudeCH1
Timebase500 pSec/div
Timebase Delay0 Sec
Setup AcquisitionEquivalent time: 16 averages
TriggerCH4; 0 V
Ω
24
1159A 1GHz Active Differential Probe
Performance Verification
Test Gain Accuracy at 1 kHz
Setup the test equipment as shown in the following diagram.
1
Select Auto Zero on the probe menu.
The probe tip must not be connected to a signal source when Auto Zero is
performed.
2
Set the function generator as shown in the following table.
RMS (Sine wave) as measured with DVM.
SettingValue
Frequency1 kHz
Output Amplitude200 mV
Connect the 1159A via Intelligent Interface to the oscilloscope.
3
4 Connect the function generator to BNC A of the calibration fixture.
5 Connect the probe tip to the probe input A of the calibration fixture.
6 Set and measure with DVM (RMS mode) output of the function
Sine Wave
rms
generator at the calibration fixture.
Record V1: ___________________ .
7
Measure the probe output with the same DVM at Intelligent Interface.
Record V2: ___________________ .
8
Calculate the % gain error.
(% Gain Error = 100(V2 - V1)/V1). Record % Gain Error: ___________________ .
25
1159A 1GHz Active Differential Probe
Performance Verification
9 Record the % Gain Error in the Performance Test Record on page 37.
Note Probe voltages at the calibration fixture and the output of the Intelligent
Interface.
Test Offset
This is an important step in the performance verification process. It should be
performed, even though no data is recorded in the Performance Test Record.
1
Disconnect the probe tip from the calibration fixture.
2 Select Auto Zero under the probe menu.
3 Measure the probe output (at BNC T of the Intelligent Interface) with
DVM DC mode.
NoteThe DVM reading should be approximately 2mV or less. This is not a warranted
specification.
Test Differential Mode Range
Use setup file 1159CAL0 for this test. Set up the test equipment as shown in the
following diagram.
1
Disconnect the intelligent interface.
2 Connect the 1159A to CH1 of the oscilloscope.
3 Load 1159CAL0.
26
1159A 1GHz Active Differential Probe
Performance Verification
4
Connect the 1161A to the output of the function generator and CH4 of
the oscilloscope.
5 Setup the test equipment as shown in the following table. Connect the
1161A probe to CH4 of the oscilloscope. Measure the amplitude on CH4.
Test EquipmentSetting
1161AOffset: 0
Function GeneratorFrequency: 1 kHz
Output Amplitude: 800 mV
6 Select Auto Zero under the probe menu.
Do not connect the probe tip to the fixture when Auto Zero is performed.
7
Connect a coaxial cable from the function generator to BNC A on the
pp
calibration fixture.
8 Connect the 1159A probe to the output of the calibration fixture CH A.
9 Record “Pass” in the “Differential Mode Range” section of the
Performance Test Record on page 37 if no clipping occurs. Record
“Fail” if visible clipping occurs.
CMRR Test Introduction
You will use setup files 1159CAL1 through 1159CAL5b to perform the CMRR
tests.
1
Set the output of the function generator to 20 Vpp.
2 Connect the 1161A probe to the output of the function generator using
BNC T.
27
1159A 1GHz Active Differential Probe
Performance Verification
The instrument setup is shown in the following diagram.
3
Follow the steps in the “Test CMRR at 70 Hz” section on page 28.
4 Follow the steps in the “Test CMRR at 1 MHz” section on page 29.
5 Follow the steps in the “Test CMRR at 100 MHz” section on page 30.
6 Follow the steps in the “Test CMRR at 500 MHz” section on page 31.
7 Follow the steps in the “Test CMRR at 1 GHz” section on page 31.
Test CMRR at 70 Hz Use setup file 11 59CAL1 in this test. If only a short time
has elapsed from the last AUTO-ZERO calibration this step may be omitted.
1
Set the function generator as shown in the following table.
SettingValue
Output Level20 V
Frequency70 Hz
2 Connect the 1161A to the function generator output and CH4 of the
pp
oscilloscope.
3 Connect the 1159A probe to the CH C probe input of the calibration
fixture.
4 Load setup file 1159CAL1.
5 Select Auto Zero under the probe menu.
Do not connect the probe tip to the fixture during Auto Zeroing.
28
6 Connect the function generator to Channel C BNC of the calibration
fixture.
7 Measure the amplitude of the function generator output (CH4).
Record V1:_____________
8
Measure the maximum value of CH1.
Record V2:_____________
9
Calculate CMRR.
CMRR in dB20
10 Record CMRR in the “CMRR at 70 Hz” section of the Performance Test
Record on page 37.
NoteCMRR must 80dB or greater.
Test CMRR at 1 MHz Use setup file 1159CAL2 in this test.
1
Set the function generator as shown in the following table.
SettingValue
Output Level20 V
Frequency1 MHz
pp
1159A 1GHz Active Differential Probe
V2 amplitude
--------------------------------
log=
V1 amplitude
Performance Verification
2 Disconnect the function generator from the BNC calibration fixture.
3 Load setup file 1159CAL2.
4 Perform Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
5
Connect the function generator to CHANNEL C BNC on the calibration
fixture.
6 Measure the amplitude of the function generator output (CH4).
Record V1:_____________
7
Measure the maximum value of CH1.
Record V2:_____________
8
Calculate CMRR.
V2 amplitude
--------------------------------
CMRR in dB20
log=
V1 amplitude
29
1159A 1GHz Active Differential Probe
Performance Verification
9 Record CMRR in the “CMRR at 1 MHz” section of the Performance Test
Record on page 37.
NoteCMRR must 40dB or greater.
Test CMRR at 100 MHz Use setup files 1159CAL3a and 1159CAL3b in this
test.
1
Set the signal generator as shown in the following table.
SettingValue
Output Level5 dBm
Frequency100 MHz
2 Disconnect the 1159A probe from CH1 and connect it to CH2 to maintain
a constant temperature.
3 Load setup file 1159CAL3a.
4 Connect the signal generator to CH1 on the oscilloscope.
5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1.
8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
9
Connect the signal generator to BNC D of the calibration fixture.
10 Load setup file 1159CAL3b.
11 Connect the 1159A probe to D on the calibration fixture.
12 Measure the maximum value of CH1.
Record V2:_____________
13
Calculate CMRR.
CMRR in dB20
14 Record CMRR in the “CMRR at 100 MHz” section of the Performance
Test Record on page 37.
V2 amplitude
--------------------------------
log=
V1 amplitude
NoteCMRR must 25dB or greater.
30
1159A 1GHz Active Differential Probe
Performance Verification
Test CMRR at 500 MHz Use setup files 1159CAL4a and 1159CAL4b in this
test.
1
Set the signal generator as shown in the following table.
SettingValue
Output Level5 dBm
Frequency500 MHz
2 Disconnect the 1159A probe from the CH1 and connect it to CH2 to
maintain a constant temperature.
3 Load setup file 1159CAL4a.
4 Connect the signal generator to CH1 on the oscilloscope.
5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1.
8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
9
Connect the signal generator to BNC C of the calibration fixture.
10 Load setup file 1159CAL4b.
11 Connect the 1159A probe to D of the calibration fixture.
12 Measure the maximum value of CH1.
Record V2:_____________
13
Calculate CMRR.
V2 amplitude
--------------------------------
CMRR in dB20
14 Record CMRR in the “CMRR at 500 MHz” section of the Performance
log=
V1 amplitude
Test Record on page 37.
NoteCMRR must 19dB or greater.
Test CMRR at 1 GHz Use setup files 1159CAL5a and 1159CAL5b in this test.
1
Set the signal generator as shown in the following table.
SettingValue
Output Level5 dBm
Frequency1 GHz
2 Disconnect the 1159A probe from the CH1 and connect it to CH2 to
maintain a constant temperature.
31
1159A 1GHz Active Differential Probe
Performance Verification
3 Load setup file 1159CAL5a.
4 Connect the signal generator to CH1 on the oscilloscope.
5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1.
8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto
Zeroing.
9
Connect the signal generator to BNC C of the calibration fixture.
10 Load setup file 1159CAL5b.
11 Measure the maximum value of CH1.
Record V2:_____________
12
Calculate CMRR.
CMRR in dB20
13 Record CMRR in the “CMRR at 1 GHz” section of the Performance Test
Record on page 37.
NoteCMRR must 13dB or greater.
V2 amplitude
--------------------------------
log=
V1 amplitude
32
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
Adjustment of 10:1 and 20:1 Attenuators
10:1 and 20:1 attenuators are supplied with the 1159A probe. This attenuator is
matched to the probe and should require no further adjustment. Each attenuator
is labeled with a serial number that matches the last four digits of the probe serial
number. Identify matching probes and attenuators using this number. If you
purchase new attenuators for the probe, you must adjust them to match the
probe.
Optimizing CMRR for an Attenuator
For optimum CMRR when the attenuator is connected to the probe, the DC and
HF attenuation should be the same for both active inputs of the probe. Three
trimmers are provided in the attenuator to match the two active inputs. The
adjustments are located under the serial number label. Apply a new label after
calibration and mark this label with the last four digits of the probe’s serial
number. Only adjust an attenuator if it is not the original attenuator shipped with
the probe.
33
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
The attenuator’s plastic covers may be removed. You must take care not to
damage the probe attenuator pins.
10:1 and 20:1 Attenuator Adjustment Procedure
Use the following steps to adjust the 10:1 or 20:1 attenuators.
1
Connect the function generator, probe and oscilloscope as shown.
Connect the + active probe input to the function generator output for these tests.
Connect the active input to ground with the probe ground socket.
–
2
Connect the 1159A Probe to CH1 of the oscilloscope.
3 Turn on the test equipment.
Allow 30 minutes for the probe to warm up.
34
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
4
Place the 10:1 or 20:1 attenuator on the 1159A probe.
Do not connect a signal to the probe at this time.
5
Set the function generator as shown in the following table.
Use a function with a flat top for this test.
SettingValue
Square Wave Output Amplitude1 V
Frequency5 kHz
Offset0
6
Select Auto Zero on the probe menu.
7 Connect the 1159A probe as shown.
Use the clips to make these connections. The frequency being considered has a
fundamental of 5 kHz. Connect the + input of the probe to the signal source.
Connect the input to ground.
8
Set the oscilloscope time base to 500 nS/div.
9 Press Auto Scale.
The 54845A triggers off CH 4.
10
Use the square head trimmer tool to adjust the + LF comp to achieve
–
the best square corner and flat top of the displayed waveform.
Adjust the compensation to achieve the best waveform characteristics when you
remove the tool.
11
Connect the + and probe inputs to the function generator output while
–
the probe ground is connected.
You can use the calibration fixture BNC to make these connections.
12
Set the function generator to 50Hz and the output to 10V amplitude.
The 1161A CH 4 will show this amplitude.
13
Set the oscilloscope time base to 5 mS/div.
14 Set CH1 sensitivity to maximum.
15 Adjust DC Att. Balance for the minimum square wave amplitude.
The phase of the signal will change by 180 as you adjust the balance control
through zero.
16
Set the function generator to 5 kHz and amplitude to 1 Vpp.
17 Connect the Probe input to function generator’s output and the +
–
°
input to the ground. You can use BNC A on the calibration fixture.
Maintain the probe ground.
18
Press Auto Scale.
35
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
19
Use the square head trimmer tool to set the + LF compensation for a
minimum of overshoot and undershoot at the leading edge of the
waveform.
Adjust the compensation to achieve the best waveform characteristics when you
remove the tool.
20
Set the oscilloscope time base to 500 nS/div
21 Connect the probe + and inputs to the function generator output.
You can use BNC C on the calibration fixture. Connect the probe ground to the
function generator ground.
22
Set the amplitude of the function generator to 10 V and the frequency
–
to 5 kHz.
23 Set the oscilloscope sensitivity to maximum.
24 Use the square head trimmer tool to adjust the LF compensation for
–
minimum signal amplitude.
It is not possible to make the trace completely flat.
36
Performance Test Record
1159A 1GHz Active Differential Probe
Performance Test Record
Agilent Technologies
Recommended Test Interval: 1 Year
Recommended Date of Next Certification:_________
Certification Temperature:_____________________
Product Name:Active Differential Voltage Probe
Model Number(s):1159A
Product Option(s):All
conforms to the following Product Specifications:
Safety:IEC 1010-1:1990+A1 / EN 61010-1:1993
UL 3111
CSA-C22.2 No. 1010.1:1993
EMC:CISPR 11:1990 / EN 55011:1991 Group 1, Class A
IEC 555-2:1982 + A1:1985 / EN60555-2:1987
IEC 555-3:1982 + A1:1990 / EN 60555-2:1987 + A1:1991
IEC 801-2:1991 / EN 50082-1:1992 4 kV CD, 8 kV AD
IEC 801-3:1984 / EN 50082-1:1992 3 V/m, {1kHz 80% AM, 27-1000 MHz}
IEC 801-4:1988 / EN 50082-1:1992 0.5 kV Sig. Lines, 1 kV Power Lines
Supplementary Information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC
and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
This product was tested in a typical configuration with Agilent test systems.
1900 Garden of the Gods Road
Colorado Springs, CO 80907, U.S.A.
Colorado Springs, 1/12/2000
Ken Wyatt, Quality Manager
European Contact: Your local Agilent Technologies Sales and Service Office
Reproduction, ad aptation,
or translation without prior
written permission is
prohib ited, except as
allowed under the copyright
laws.
Restric ted Rights
Legend.
Use, duplication or
disclo sure by the U.S .
Government is subject to
restrictions as set forth in
subparagraph (c) (1) (ii) of
the Rights in Technical Data
and Computer Software
clause at DFARS 252.2277013 for DOD agencies, and
subparagraphs (c) (1) and
(c) (2) of the Commercial
Computer Software
Restricted Rights clause at
FAR 52.227-19 for other
agencies.
Agilen t Technologies
3000 Hanover Street
Palo Alto, California 94304
U.S.A.
Document Warranty
The information contained
in this document is subject
to change without notice.
Agilent makes no
warranty of any kind
with regard to this
material, including, but
not limited to, the
implied warranties of
merchantability or
fitness for a particular
purpose.
Agilen t sha ll not be liabl e
for errors contained herein
or for damages in
connection with the
furnishing, perfo rmance, or
use of this mate rial.
Product Warranty
This Agilent product has a
warranty against defects in
material and workmanship
for a period of one year
from date o f shipment.
During the warranty period,
Agilent Technologies will, at
its option, either repair or
replace products that prove
to be defective.
For warranty service or
repair, this product must be
returned to a service facility
designated by Agilen t.
For products retur ned to
Agilent for warranty
service, the Buyer shall
prepay shipping charges to
Agilent and Agilent shall
pay shipping charges to
return the product to the
Buyer. However, the Buyer
shall pay all shipping
charges, duties, and taxes
for products returned to
Agilent from another
country.
Agilent warrants that its
software and firmware
designated by Agilen t for
use with an instrument will
execute its programming
instructions when properly
installed on that
instrument. Agilent does
not warrant that the
operation of the instrument
software, or firmware will
be uninterru pted or erro r
free.
Limitation of Warranty
The for egoin g warranty
shall not apply to defects
resulting from improper or
inadequate maintenance by
the Buye r, Buyer-supplied
software or interfacing,
unauthorized modification
or misuse, operation outside
of the environ mental
specifications for the
product, or improper site
preparation or
maintenance.
No other warranty i s
expressed or implied.
Agilent specifically
disclaims the implied
warranties of
merchantability or
fitness for a particular
purpose.
Exclusive Remedies
The remedies provided
herein are the buyer's sole
and exclusive remedies.
Agilent shall not be liable
for any direct, indirect,
special, incidental, or
consequential damages,
whether based on contract,
tort, or any other legal
theory.
Assistance
Product maintenance
agreements and other
customer assistance
agreements are available for
Agilent products .
For any assistance, contact
your nearest Agilent Sales
Office.
Certification
Agilent Technologies
certifies that this product
met its published
specifications at the time of
shipment from the factory.
Agilent fur ther certifies that
its cal ibrat ion
measurements are traceable
to the United States
National Institute of
Standards and Technology,
to the extent allowed by the
Institute's calib ration
facility, and to the
calibration facilities of other
International S tandards
Organization m embers.
Agilen t Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901
Safety
This apparatus has been
designed and tested in
accordance with IEC
Publication 1010, Safe ty
Requirements for
Measuring Apparat us, and
has been supplied in a safe
condition. To ensure safe
operation and to keep the
product safe, the
information, cautions, and
warnings in this operating
manual must be heeded. In
addition, note the external
markin gs on the instrument
that are described under
"Safet y Symbols."
Safety Symbols
!
Instruction manual symbol:
the product is marked with
this symbol when it is
necessary for you to refer to
the instruction manual in
order to protect against
damage to the product.
Hazardous voltage symbol.
Earth terminal symbol:
Used to indicate a circuit
common connected to
grounded chassis.
WARNING
The Warning sign denotes a
hazard . It calls at tention to
a procedure, practice, or
the like, which, if not
correctly performed or
adhered to, could result in
personal injury. Do not
proceed beyon d a Warning
sign until the indicated
conditions are fully
understood and met.
CAUTION
The Caution sign denotes a
hazard . It calls at tention to
an operating procedure,
practice, or the like, which,
if not correctly performe d
or adhered to, could result
in damage to or destruction
of part or all of the product.
Do not proceed beyond a
Caution symbol until the
indicated conditions are
fully understood or met.
About this edition
This is the 1159 Active
Differential Voltage Probe
User’s Guide.
Publication number
01159-92000, Feb. 2000
Printed in USA.
Print history is as follows:
01159-92000, Feb. 2000
New editions are complete
revisions of the manual.
Many product updates do
not require manual
changes; and, conversely,
manual corrections may be
done without accompanying
product changes.
Therefore, do n ot expect a
one-to-one correspondence
between product updates
and manual updates.
s1
Agilent Technologies
Printed in the USA
Manual Part Number
01159-92000
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.