Agilent 1159A Users Guide

User’s Guide
Publication number 01159-92000 February 2000
s1
For Safety information, Warranties, Regulatory information, and publishing information, see the pages at the back of this book.
© Copyright Agilent Technologies 2000
All Rights Rese rved.
Contents
Inspect the Probe 3 1159A Active Differential Probes Introduction 4 Probe Accessories Supplied 5 Attaching External Attenuators to the Probe 7
!
Specifications and Characteristics 8 To Connect the Probe to the Circuit under Test 12 Recommended Probe Configurations 14 Safety Considerations 17 Service Strategy 18 Performance Verification 19 Adjustment of 10:1 and 20:1 Attenuators 33 Performance Test Record 37
2
1159A 1GHz Active Differential Probe

Inspect the Probe

Inspect the Probe
Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has been checked mechanically and electrically.
Check the accessories.
Any accessories that were supplied with the probe are listed in “Probe Accessories Supplied” on page 5.
• If the contents are incomplete or damaged notify your Agilent Sales Office.
Inspect the instrument.
• If there is mechanical damage or defect, or if the probe does not operate properly or pass performance tests, notify your Agilent Sales Office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Agilent Sales Office. Keep the shipping materials for the carrier’s inspection. The Agilent office will arrange for repair or replacement at Agilent’s option without waiting for claim settlement.
3
1159A 1GHz Active Differential Probe

1159A Active Differential Probes Introduction

1159A Active Differential Probes Introduction
The 1159A is a wide-band differential active probe. The probe features low noise, low input capacitance, high common mode rejection, and Field Effect Transistor (FET) buffered inputs in the probe head. User-selectable offset gives the probe flexibility to measure a large range of signal types. Plug-on attenuators and AC coupling accessories further extend the application range. Included interconnect accessories allow connection to surface mount and through-hole components with minimal signal degradation. The input receptacles in the probe head are compatible with standard 0.025" (0.635 mm) square pins, which provide a convenient low-cost method of creating device characterization test fixtures.
The 1159A is ideal for acquiring high speed differential signals such as those found in disk drive read channels, differential LAN, video, and so on. The high impedance characteristics of both inputs allow you to use the probe as a FET probe to make single-ended measurements in digital systems without introducing a ground loop as a conventional FET probe would.

Differential Amplifiers and CMRR

The 1159A Differential Probe is a high input impedance amplifier. A characteristic of differential amplifiers is the ability to reject signals that are common to the two inputs. The common mode rejection ratio (CMRR) is the measurement of this ability. It is expressed as the ratio between the amplitudes of the common mode and differential signals that produce equal outputs. If the differential gain is known, these measurements can be referred to the probe input. CMRR is usually expressed in dB:
V common mode input Gain×

CMRR in dB 20
The ability to reject common mode signals depends on the balance designed into the probe amplifier. As the frequency of the common mode signal increases, it becomes harder to balance the amplifier parasitic parameters. This leads to degradation of the CMRR.
The CMRR of the 1159A Differential probe is specified from the probe tip. This method of specifying the probe CMMR eliminates the effects of source impedance, provided the connections from the probe tip to the signal source are symmetrical.
-------------------------------------------------------------------------
log=

V common mode output
4
1159A 1GHz Active Differential Probe

Probe Accessories Supplied

Probe Accessories Supplied
The following diagram and table show the accessories supplied with the 1159A Differential Probe.
1
4
5
Probe Accessories
Item Description Qty. Part Number
1 AC coupler 1 01154-82101 2 10:1 Attenuator 1 01159-82104 3 20:1 Attenuator 1 01159-82105 4 Header 1 N/A 5 Offset Pin 4 N/A 6 0.5 Grabber 2 N/A 7 0.8 Grabber 3 N/A 8 Ground Wire 1 N/A 9SMT Lead 4 N/A 10 Wire Lead 1 N/A
2
7
6
3
10
9
8
5
1159A 1GHz Active Differential Probe
Probe Accessories Supplied

To Order Replaceable Parts

To order a replaceable part, in the United States and Canada call our toll-free hotline at 1-877-447-7278, or call your local Agilent Technologies Sales Office.
Replaceable Parts
Item Description Qty. Part Number
1 AC coupler 1 01154-82101 2 10:1 Attenuator
3 20:1 Attenuator
4 Header 1 N/A 5 Offset Pin 4 N/A 6 0.5 Grabber 2 N/A 7 0.8 Grabber 3 N/A 8 Ground Wire 1 N/A 9SMT Lead 4N/A 10 Wire Lead 1 N/A
(includes an adjustment tool)
(includes an adjustment tool) Connection Kit 1 01154-60004
Trimmer Tool
(0.635 mm square head)
1 01159-82104
1 01159-82105
1 5063-2196

Using the Accessories

The 1159A Differential Probe and accessories provide a variety of ways to connect to circuitry under test. Any method used to connect the probe signal inputs to the circuit under test degrades the performance of the probing solution. Take the following precautions to optimize common mode rejection.
• Maintain tip connection lead length as short as possible and the same length.
• Follow the same path for wires used to connect the inputs of the probe to the circuit under test.
• Probes do not have infinite input impedance and do load the circuit under test. If the impedance of the test points is not identical, unequal loading will occur. This degrades common mode rejection.
• The ground lead length is not usually critical with a differential probe.
• Carefully consider the ground potential relative to the oscilloscope ground potential. The potential difference must be within the common mode range of the probe.
• The DC potential between the AC coupling adapter and the oscilloscope ground must not exceed 42 Vpk.
• Do not cascade the external attenuators.
• Cascade the external AC coupling adapter in the following order: probe, attenuator, and AC coupling adapter.
6
1159A 1GHz Active Differential Probe

Attaching External Attenuators to the Probe

Other Probe Accessories

The Agilent Wedge was designed to interface directly with the differential probe. These devices simplify connections to surface mount integrated circuits and have output pins compatible with the probe tip and attenuator sockets.
Attaching External Attenuators to the Probe
The external attenuators plug directly on to the probe tip. They are calibrated at the factory to provide the optimum common mode rejection and should not be swapped between probes.
1159A Attenuator
Div-By-10 1 GHz Div-By-20 1 GHz
1159A AC Coupler
Always Install Last
The 1159A probe’s best performance is achieved when the probe attenuation is set to /10 in the Infiniium’s Probes Setup dialog box.
7
1159A 1GHz Active Differential Probe

Specifications and Characteristics

Specifications and Characteristics
Specifications
Input Configuration Ground Connector
Input Coupling DC AC coupling obtained by installing an AC coupling
Gain Accuracy at 1 kHz 2%
!
Maximum Input Voltage
Either input from ground
CMRR at 70 Hz: 80dB
True Differential (+ and inputs), with shield
adapter
< ±42 V
at 1 MHz: 40dB at 100 MHz: 25dB at 500 MHz: 19dB at 1 GHz: 13dB
Range
Mode No attenuator ÷10 attenuator ÷20 attenuator
Differential < ±400 mV < ±4 V < ± 8V Common < ±16 V < ±42 V < ±42 V Offset (Common) < ±1.6 V < ±16 V < ±32 V
8
1159A 1GHz Active Differential Probe
Specifications and Characteristics
The following characteristics are valid for the 1159A probe after the probe has reached operating temperature, which is 20 minutes with power applied in a environment with stable ambient temperature. The probe must be operating
within the environmental conditions listed in the “Environmental Specifications” section on page 11, and must have been calibrated within the past 12 months in a ambient temperature of 23 5 C.
±°
Performance Characteristics
Probe Bandwidth (-3 dB) DC to 1 GHz Offset Range 1.6 V Rise Time (Probe only)
1:1 Attenuation Internal switched attenuation only
Input Resistance
(each side to ground)
Input Capacitance (between inputs)
1:1 Attenuation No external attenuators
Input Capacitance (each side to ground)
1:1 Attenuation No external attenuators
±
<350 ps
1 M
<0.85 pF
<1.5 pF
9
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Typical CMRR versus Frequency (Hz)
Typical Noise
10
1159A 1GHz Active Differential Probe
Specifications and Characteristics
Environmental Specifications
Operating Non-operating
Temperature 0 to 50 C -40 to 75 C Humidity Up to 80% RH at 40 C Up to 80% RH at 75 C Altitude Up to 4,600 meters
Vibration Random vibration 5 to
Weight Approximately 226 g Dimensions Refer to the drawing shown below
Dimensions
°°
°°
(15,000 feet)
500 Hz, 10 minutes per axis,
0.3 g
rms
Up to 15,000 meters (50,000 feet)
Random vibration 5 to 500 Hz, 10 minutes per axis, 2.41 g
. Resonant search 5 to 500 Hz
rms
swept sine, 1 octave/min. sweep rate, (0.75 g), 5 minutes resonant
dwell at 4 resonance’s per axis.
11
1159A 1GHz Active Differential Probe

To Connect the Probe to the Circuit under Test

To Connect the Probe to the Circuit under Test
The method you use to connect the probe to the circuit under test is critical for ensuring accurate measurements. The following examples examine the effect of using different lengths of wire at 100MHz to connect the signal source to the probe tip.

The Impedance of the Source

This is another instance where the symmetry of the differential circuit is important. The impedance of the source forms a network with the input impedance of the connection and the probe. This network determines the frequency response for the measurement. If each side of the differential source has a different impedance, the frequency response of each side will be different. This lack of balance is reflected in reduced CMRR. The higher the impedance of the source, the more critical these parasitic effects.

The Ground Connection

A poorly located ground connection allows ground loops to add to the common mode signal. The differential probe measures the potential difference between two locations on a PC board. Usually, it is not necessary to ground the probe. Whether to ground the probe depends on the magnitude and frequency of the voltage difference between the oscilloscope ground and the board ground. It is good practice to maintain a board ground. Without this ground reference, you could easily exceed the common mode range of the probe.

Probe Offset

The amplifiers in the 1159A probe limit the Differential Mode Range to 400mV. If the input to the probe is approaching 400mV, there is little offset range available for positioning the trace on screen. There are two solutions to this problem:
• Attenuate the signal into the probe with the 1/10 or 1/20 attenuator. This mode of operation will induce some small loss in CMRR.
•Use Position to position the trace on screen.
The added feature of position control independent of Offset allows trace positioning without calculating how much probe offset range is available. The trace can be positioned by dragging the trace or positioning the trace under the Probe menu.
12
1159A 1GHz Active Differential Probe
To Connect the Probe to the Circuit under Test
Both inputs derived from a common signal at probe tip.
CMRR = 35.6dB
Both inputs derived from a common signal via 5cm coupling lead.
CMRR = 35dB
Probe coupli ng leads of different length.
Positive input 5.3cm. Negative input 7.5cm.
CMRR = 33dB
13
1159A 1GHz Active Differential Probe

Recommended Probe Configurations

Recommended Probe Configurations
For best performance, use the following configurations. They are presented in the recommended order from the most desirable to the least.
Note The use of the ground connection is optional for all configurations.

Direct Connection

1159A Probe Tip
Test Point Layout
See the “Test Point Layout” section for more information

AC Adapter/Attenuator

Use the attenuator shipped with the probe and marked with the same serial number for accurate measurements. Do not use the attenuators with other probes.
1159A Attenuator
Div-By-10 1 GHz Div-By-20 1 GHz
1159A AC Coupler
Always Install Last
Test Point Layout
See the “Test Point Layout” section for more information
14
1159A 1GHz Active Differential Probe
Recommended Probe Configurations

Offset Pins

You may use offset pins with any of the tip adapters.
Offset Pins Rotate to Adjust Spacing

SMT Lead

You may use SMT leads with any of the tip adapters.
SMT Leads
Maintain Equal Length
Solder Leads to Test Points
Ground
Ground Lead Opt
15
1159A 1GHz Active Differential Probe
Recommended Probe Configurations

Wire Leads

You may use wire leads with any of the tip adapters.
Connect Leads to Test Points
Ground
Opt

Grabbers

Using grabbers and wire leads results in significant lead length. Expect measurement quality degradation with fast signals.
0.5 mm or 0.8 mm Grabbers
Opt
Ground
16
1159A 1GHz Active Differential Probe

Safety Considerations

Test Point Layout

Safety Considerations
Read the Safety summary in the warranty pages at the back of this guide before servicing the instrument. Before performing any procedure, review the safety information for cautions and warnings.
WARNING Only trained service personnel aware of the hazards involved (for example, fire
!
and electric shock) should perform maintenance on the instrument. When maintenance can be performed without power applied, the power cord must be removed from the instrument.
17
1159A 1GHz Active Differential Probe

Service Strategy

Service Strategy
To return the 1159A 1 GHz Differential Probe to optimum performance requires factory repair. Return all probes to the service group for repair and calibration. If the probe is under warranty, normal warranty services apply. If the probe is not under warranty, you can exchange a failed probe for a reconditioned one at a nominal cost.

To Return the Probe to Agilent for Service

Call (877) 447-7278 for further details and the location of your nearest Agilent Service Office.
1
Write the following information on a tag and attach it to the probe.
• Name and address of owner.
• Probe model number.
• Probe serial number.
• Description of the service required or failure indications.
2 Return the following accessories with the probe:
• Attenuators
• AC coupling adapter
Retain all other accessories.
3
Return the probe in its case or pack the probe in foam or other shock absorbing material and place it in a strong shipping container.
You can use the original shipping materials or order materials from an Agilent Sales Office. If neither are available, place 3 to 4 inches of shock-absorbing material around the instrument and place it in a box that does not allow movement during shipping.
4
Seal the shipping container securely.
5 Mark the shipping container as FRAGILE.
In all correspondence, refer to the instrument by model number and full serial number.
18
1159A 1GHz Active Differential Probe

Performance Verification

Performance Verification
Use this procedure to verify the warranted characteristics of the 1159A 1 GHz Differential Probe. The recommended performance verification interval for the 1159A is one year. Perform the complete performance verification procedure as the first step of annual certification. You can complete performance verification without removing the probe covers. There are no user adjustments available for
calibration. Use the equipment shown in the “Test Equipment Required” section to complete the performance verification procedures.
1
Perform the steps listed in the “Preliminary Procedure” section on page 20.
2 Perform the steps listed in the “Test Gain Accuracy at 1 kHz” section
on page 25.
3 Perform the steps listed in the “Test Offset” section on page 26. 4 Perform the steps listed in the “Test Differential Mode Range” section
on page 26.
5 Perform the steps listed in the “CMRR Test Introduction” section on
page 27.

Test Equipment Required

Description Minimum Requirements Part Number
Infinium Oscilloscope 1.5 GHz 54845A Digital AC/DC Voltmeter DC: 0.1% accuracy
AC: 0.2% accuracy 200 mV @ 1 kHz 5 1/2 digit resolution
Function Generator Sine Wave and Square Wave output
Sine Wave Generator 50 MHz to 1 GHz BNC T E9625A Probe Tip BNC Adapter 5063-2143 Calibration Fixture 01154-63801 Intelligent Interface 01154-63401 High Frequency 10:1 Resistor
Divider Probe 10:1 Resistor Divider Probe
(provided with the 54845A) Trimmer Tools
(included when attenuators are purchased separately)
waveforms
into 1 M
20 V
pp
70 Hz to 10 MHz
500 1163A
10 M 1161A
Flat blade 0.04” (1 mm) wide Square Head 0.025” (0.635 mm) 5063-2196
34401A
33120A
19
1159A 1GHz Active Differential Probe
µ
Performance Verification

Preliminary Procedure

1
Turn on the oscilloscope, the 1159A, and the other test equipment.
Allow 30 minutes for all test equipment to warm up.
2
Press Default Setup.
3 Perform Probe Tip Cal on the 1161A connected to CH4 of the
oscilloscope (54845A).
4 Save the setup files listed in the “Performance Verification Setup Files”
section on page 20 to the C drive of the 54845A.
These setups can also be saved to drive A and filed for future use.

Performance Verification Setup Files

Save the setup information shown below for the various performance verification tests to the specified file names on the C drive of the 54845A or to a floppy disk in drive A. Connect the probes to the specified channels before creating and storing the setup files.
Setup for Differential Mode Range
File Name: 1159CAL0
Probe Connected to CH1 1159A Probe Connected to CH4 1161A CH1 and CH4 Sensitivity 200 mV/div CH2 and CH3 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measurements: CH1 and CH4 V amplitude under Measure Voltage Timebase 500 Sec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 8 averages Trigger CH4; level 0 V + ve edge
20
1159A 1GHz Active Differential Probe
Setup for CMRR at 70 Hz
File Name: 1159CAL1
Probe Connected to CH1 1159A Probe Connected to CH4 1161A CH1 Sensitivity 2 mV/div CH4 Sensitivity 5 V/div Measure Amplitude CH1 and CH4 CH2 and CH3 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Timebase 10 mSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 64 averages Trigger CH4; level 0 V + ve edge
Setup for CMRR at 1 MHz
File Name: 1159CAL2
Probe Connected to CH1 1159A Probe Connected to CH4 1161A CH1 Sensitivity 10 mV/div CH4 Sensitivity 5 V/div CH2 and CH3 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 and CH4 Timebase 500 nSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 64 averages Trigger CH4; level 0 V + ve edge
Performance Verification
21
1159A 1GHz Active Differential Probe
Performance Verification
Setup for CMRR at 100 MHz
File Name: 1159CAL3a
Probe Connected to CH1 None Probe Connected to CH4 1163A CH1 Sensitivity 200 mV/div; 50 CH4 Sensitivity 200 mV/div; 50 CH2, CH3, and CH4 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 Timebase 5 nSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 64 averages Trigger CH4; 0 V + ve edge
Ω Ω
Setup for CMRR at 100 MHz
File Name: 1159CAL3b
Probe Connected to CH1 1159A Probe Connected to CH4 1163A CH1 Sensitivity 5 mV/div CH4 Sensitivity 200 mV CH2, CH3, and CH4 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 Timebase 5 nSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 64 averages Trigger CH4; 0 V+ ve edge
22
1159A 1GHz Active Differential Probe
Setup for CMRR at 500 MHz
File Name: 1159CAL4a
Probe Connected to CH4 1163A CH1 Sensitivity 200 mV/div; 50 CH4 Sensitivity 200 mV/div; 50 Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 Timebase 1 nSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 64 averages Trigger CH4; 0 V
Ω Ω
Setup for CMRR at 500 MHz
File Name: 1159CAL4b
Probe Connected to CH1 1159A Probe Connected to CH4 1163A CH1 Sensitivity 20 mV/div CH4 Sensitivity 200 mV/div; 50 Offset and Position CH1, CH2, CH3 and CH4 0 V offset Measure Amplitude CH1 Timebase 1 nSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 16 averages Trigger CH4; 0 V
Performance Verification
23
1159A 1GHz Active Differential Probe
Performance Verification
Setup for CMRR at 1 GHz
File Name: 1159CAL5a
Probe Connected to CH1 None Probe Connected to CH4 1163A CH1 Sensitivity 200 mV/div; 50 CH4 Sensitivity 200 mV/div; 50 CH2, CH3, and CH4 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 Timebase 500 pSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 16 averages Trigger CH4; 0 V
Ω Ω
Setup for CMRR at 1 GHz
File Name: 1159CAL5b
Probe Connected to CH1 1159A Probe Connected to CH4 1163A CH1 Sensitivity 50 mV/div CH4 Sensitivity 200 mV/div; 50 CH2, CH3, and CH4 Off Offset and Position CH1, CH2, CH3 and CH4 0 V Measure Amplitude CH1 Timebase 500 pSec/div Timebase Delay 0 Sec Setup Acquisition Equivalent time: 16 averages Trigger CH4; 0 V
24
1159A 1GHz Active Differential Probe
Performance Verification

Test Gain Accuracy at 1 kHz

Setup the test equipment as shown in the following diagram.
1
Select Auto Zero on the probe menu.
The probe tip must not be connected to a signal source when Auto Zero is performed.
2
Set the function generator as shown in the following table.
RMS (Sine wave) as measured with DVM.
Setting Value
Frequency 1 kHz Output Amplitude 200 mV
Connect the 1159A via Intelligent Interface to the oscilloscope.
3 4 Connect the function generator to BNC A of the calibration fixture. 5 Connect the probe tip to the probe input A of the calibration fixture. 6 Set and measure with DVM (RMS mode) output of the function
Sine Wave
rms
generator at the calibration fixture.
Record V1: ___________________ .
7
Measure the probe output with the same DVM at Intelligent Interface.
Record V2: ___________________ .
8
Calculate the % gain error.
(% Gain Error = 100(V2 - V1)/V1). Record % Gain Error: ___________________ .
25
1159A 1GHz Active Differential Probe
Performance Verification
9 Record the % Gain Error in the Performance Test Record on page 37.
Note Probe voltages at the calibration fixture and the output of the Intelligent
Interface.

Test Offset

This is an important step in the performance verification process. It should be performed, even though no data is recorded in the Performance Test Record.
1
Disconnect the probe tip from the calibration fixture.
2 Select Auto Zero under the probe menu. 3 Measure the probe output (at BNC T of the Intelligent Interface) with
DVM DC mode.
Note The DVM reading should be approximately 2mV or less. This is not a warranted
specification.

Test Differential Mode Range

Use setup file 1159CAL0 for this test. Set up the test equipment as shown in the following diagram.
1
Disconnect the intelligent interface.
2 Connect the 1159A to CH1 of the oscilloscope. 3 Load 1159CAL0.
26
1159A 1GHz Active Differential Probe
Performance Verification
4
Connect the 1161A to the output of the function generator and CH4 of the oscilloscope.
5 Setup the test equipment as shown in the following table. Connect the
1161A probe to CH4 of the oscilloscope. Measure the amplitude on CH4.
Test Equipment Setting
1161A Offset: 0 Function Generator Frequency: 1 kHz
Output Amplitude: 800 mV
6 Select Auto Zero under the probe menu.
Do not connect the probe tip to the fixture when Auto Zero is performed.
7
Connect a coaxial cable from the function generator to BNC A on the
pp
calibration fixture.
8 Connect the 1159A probe to the output of the calibration fixture CH A. 9 Record “Pass” in the “Differential Mode Range” section of the
Performance Test Record on page 37 if no clipping occurs. Record “Fail” if visible clipping occurs.

CMRR Test Introduction

You will use setup files 1159CAL1 through 1159CAL5b to perform the CMRR tests.
1
Set the output of the function generator to 20 Vpp.
2 Connect the 1161A probe to the output of the function generator using
BNC T.
27
1159A 1GHz Active Differential Probe
Performance Verification
The instrument setup is shown in the following diagram.
3
Follow the steps in the “Test CMRR at 70 Hz” section on page 28.
4 Follow the steps in the “Test CMRR at 1 MHz” section on page 29. 5 Follow the steps in the “Test CMRR at 100 MHz” section on page 30. 6 Follow the steps in the “Test CMRR at 500 MHz” section on page 31. 7 Follow the steps in the “Test CMRR at 1 GHz” section on page 31.
Test CMRR at 70 Hz Use setup file 11 59CAL1 in this test. If only a short time
has elapsed from the last AUTO-ZERO calibration this step may be omitted.
1
Set the function generator as shown in the following table.
Setting Value
Output Level 20 V Frequency 70 Hz
2 Connect the 1161A to the function generator output and CH4 of the
pp
oscilloscope.
3 Connect the 1159A probe to the CH C probe input of the calibration
fixture.
4 Load setup file 1159CAL1. 5 Select Auto Zero under the probe menu.
Do not connect the probe tip to the fixture during Auto Zeroing.
28
6 Connect the function generator to Channel C BNC of the calibration
fixture.
7 Measure the amplitude of the function generator output (CH4).
Record V1:_____________
8
Measure the maximum value of CH1.
Record V2:_____________
9
Calculate CMRR.
CMRR in dB 20
10 Record CMRR in the “CMRR at 70 Hz” section of the Performance Test
Record on page 37.
Note CMRR must 80dB or greater.
Test CMRR at 1 MHz Use setup file 1159CAL2 in this test.
1
Set the function generator as shown in the following table.
Setting Value
Output Level 20 V Frequency 1 MHz
pp
1159A 1GHz Active Differential Probe
V2 amplitude

--------------------------------
log=

V1 amplitude
Performance Verification
2 Disconnect the function generator from the BNC calibration fixture. 3 Load setup file 1159CAL2. 4 Perform Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
5
Connect the function generator to CHANNEL C BNC on the calibration fixture.
6 Measure the amplitude of the function generator output (CH4).
Record V1:_____________
7
Measure the maximum value of CH1.
Record V2:_____________
8
Calculate CMRR.
V2 amplitude

--------------------------------
CMRR in dB 20
log=

V1 amplitude
29
1159A 1GHz Active Differential Probe
Performance Verification
9 Record CMRR in the “CMRR at 1 MHz” section of the Performance Test
Record on page 37.
Note CMRR must 40dB or greater.
Test CMRR at 100 MHz Use setup files 1159CAL3a and 1159CAL3b in this test.
1
Set the signal generator as shown in the following table.
Setting Value
Output Level 5 dBm Frequency 100 MHz
2 Disconnect the 1159A probe from CH1 and connect it to CH2 to maintain
a constant temperature.
3 Load setup file 1159CAL3a. 4 Connect the signal generator to CH1 on the oscilloscope. 5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1. 8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
9
Connect the signal generator to BNC D of the calibration fixture.
10 Load setup file 1159CAL3b. 11 Connect the 1159A probe to D on the calibration fixture. 12 Measure the maximum value of CH1.
Record V2:_____________
13
Calculate CMRR.
CMRR in dB 20
14 Record CMRR in the “CMRR at 100 MHz” section of the Performance
Test Record on page 37.
V2 amplitude

--------------------------------
log=

V1 amplitude
Note CMRR must 25dB or greater.
30
1159A 1GHz Active Differential Probe
Performance Verification
Test CMRR at 500 MHz Use setup files 1159CAL4a and 1159CAL4b in this test.
1
Set the signal generator as shown in the following table.
Setting Value
Output Level 5 dBm Frequency 500 MHz
2 Disconnect the 1159A probe from the CH1 and connect it to CH2 to
maintain a constant temperature.
3 Load setup file 1159CAL4a. 4 Connect the signal generator to CH1 on the oscilloscope. 5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1. 8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
9
Connect the signal generator to BNC C of the calibration fixture.
10 Load setup file 1159CAL4b. 11 Connect the 1159A probe to D of the calibration fixture. 12 Measure the maximum value of CH1.
Record V2:_____________
13
Calculate CMRR.
V2 amplitude

--------------------------------
CMRR in dB 20
14 Record CMRR in the “CMRR at 500 MHz” section of the Performance
log=

V1 amplitude
Test Record on page 37.
Note CMRR must 19dB or greater.
Test CMRR at 1 GHz Use setup files 1159CAL5a and 1159CAL5b in this test.
1
Set the signal generator as shown in the following table.
Setting Value
Output Level 5 dBm Frequency 1 GHz
2 Disconnect the 1159A probe from the CH1 and connect it to CH2 to
maintain a constant temperature.
31
1159A 1GHz Active Differential Probe
Performance Verification
3 Load setup file 1159CAL5a. 4 Connect the signal generator to CH1 on the oscilloscope. 5 Measure the amplitude of the signal generator output.
Record V1:_____________
6
Remove the signal generator from CH1.
7 Remove the 1159A probe from CH2 and connect it to CH1. 8 Select Auto Zero under the probe menu.
Do not connect the probe tip to the calibration fixture during Auto Zeroing.
9
Connect the signal generator to BNC C of the calibration fixture.
10 Load setup file 1159CAL5b. 11 Measure the maximum value of CH1.
Record V2:_____________
12
Calculate CMRR.
CMRR in dB 20
13 Record CMRR in the “CMRR at 1 GHz” section of the Performance Test
Record on page 37.
Note CMRR must 13dB or greater.
V2 amplitude

--------------------------------
log=

V1 amplitude
32
1159A 1GHz Active Differential Probe

Adjustment of 10:1 and 20:1 Attenuators

Adjustment of 10:1 and 20:1 Attenuators
10:1 and 20:1 attenuators are supplied with the 1159A probe. This attenuator is matched to the probe and should require no further adjustment. Each attenuator is labeled with a serial number that matches the last four digits of the probe serial number. Identify matching probes and attenuators using this number. If you purchase new attenuators for the probe, you must adjust them to match the probe.

Optimizing CMRR for an Attenuator

For optimum CMRR when the attenuator is connected to the probe, the DC and HF attenuation should be the same for both active inputs of the probe. Three trimmers are provided in the attenuator to match the two active inputs. The adjustments are located under the serial number label. Apply a new label after
calibration and mark this label with the last four digits of the probe’s serial number. Only adjust an attenuator if it is not the original attenuator shipped with the probe.
33
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
The attenuator’s plastic covers may be removed. You must take care not to damage the probe attenuator pins.

10:1 and 20:1 Attenuator Adjustment Procedure

Use the following steps to adjust the 10:1 or 20:1 attenuators.
1
Connect the function generator, probe and oscilloscope as shown.
Connect the + active probe input to the function generator output for these tests. Connect the active input to ground with the probe ground socket.
2
Connect the 1159A Probe to CH1 of the oscilloscope.
3 Turn on the test equipment.
Allow 30 minutes for the probe to warm up.
34
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
4
Place the 10:1 or 20:1 attenuator on the 1159A probe.
Do not connect a signal to the probe at this time.
5
Set the function generator as shown in the following table.
Use a function with a flat top for this test.
Setting Value
Square Wave Output Amplitude 1 V Frequency 5 kHz Offset 0
6
Select Auto Zero on the probe menu.
7 Connect the 1159A probe as shown.
Use the clips to make these connections. The frequency being considered has a fundamental of 5 kHz. Connect the + input of the probe to the signal source. Connect the input to ground.
8
Set the oscilloscope time base to 500 nS/div.
9 Press Auto Scale.
The 54845A triggers off CH 4.
10
Use the square head trimmer tool to adjust the + LF comp to achieve
the best square corner and flat top of the displayed waveform.
Adjust the compensation to achieve the best waveform characteristics when you remove the tool.
11
Connect the + and probe inputs to the function generator output while
the probe ground is connected.
You can use the calibration fixture BNC to make these connections.
12
Set the function generator to 50Hz and the output to 10V amplitude.
The 1161A CH 4 will show this amplitude.
13
Set the oscilloscope time base to 5 mS/div.
14 Set CH1 sensitivity to maximum. 15 Adjust DC Att. Balance for the minimum square wave amplitude.
The phase of the signal will change by 180 as you adjust the balance control through zero.
16
Set the function generator to 5 kHz and amplitude to 1 Vpp.
17 Connect the Probe input to function generator’s output and the +
°
input to the ground. You can use BNC A on the calibration fixture.
Maintain the probe ground.
18
Press Auto Scale.
35
1159A 1GHz Active Differential Probe
Adjustment of 10:1 and 20:1 Attenuators
19
Use the square head trimmer tool to set the + LF compensation for a minimum of overshoot and undershoot at the leading edge of the waveform.
Adjust the compensation to achieve the best waveform characteristics when you remove the tool.
20
Set the oscilloscope time base to 500 nS/div
21 Connect the probe + and inputs to the function generator output.
You can use BNC C on the calibration fixture. Connect the probe ground to the function generator ground.
22
Set the amplitude of the function generator to 10 V and the frequency
to 5 kHz.
23 Set the oscilloscope sensitivity to maximum. 24 Use the square head trimmer tool to adjust the LF compensation for
minimum signal amplitude.
It is not possible to make the trace completely flat.
36

Performance Test Record

1159A 1GHz Active Differential Probe
Performance Test Record
Agilent Technologies
Recommended Test Interval: 1 Year Recommended Date of Next Certification:_________ Certification Temperature:_____________________
1159A 1 GHz Differential Probe Serial No.:_______________________ Certification Date:_________________ Tested By:_______________________
_______________________________ _______________________________
Test Limit Minimum Results Limit Maximum
Accuracy at 1 kHz into 50 input of 54845 A
Differential Mode Range Att 1; Gain 1
Att 10; Gain 10
CMRR at 70 Hz 80 dB N/A CMRR at 1 MHz 40 dB N/A CMRR at 100 MHz 25 dB N/A CMRR at 500 MHz 19 dB N/A CMRR at 1 GHz 13 dB N/A
Min Gain 98% Max Gain 102%
No Clipping
Pass
_____
_____
Fail
_____
_____
N/A
37
38
DECLARATION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014
Manufacturer’s Name: Agilent Technologies Manufacturer’s Address: Colorado Springs Division
declares, that the product
Product Name: Active Differential Voltage Probe Model Number(s): 1159A Product Option(s): All
conforms to the following Product Specifications:
Safety: IEC 1010-1:1990+A1 / EN 61010-1:1993
UL 3111 CSA-C22.2 No. 1010.1:1993
EMC: CISPR 11:1990 / EN 55011:1991 Group 1, Class A
IEC 555-2:1982 + A1:1985 / EN60555-2:1987 IEC 555-3:1982 + A1:1990 / EN 60555-2:1987 + A1:1991 IEC 801-2:1991 / EN 50082-1:1992 4 kV CD, 8 kV AD IEC 801-3:1984 / EN 50082-1:1992 3 V/m, {1kHz 80% AM, 27-1000 MHz} IEC 801-4:1988 / EN 50082-1:1992 0.5 kV Sig. Lines, 1 kV Power Lines
Supplementary Information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly.
This product was tested in a typical configuration with Agilent test systems.
1900 Garden of the Gods Road Colorado Springs, CO 80907, U.S.A.
Colorado Springs, 1/12/2000
Ken Wyatt, Quality Manager
European Contact: Your local Agilent Technologies Sales and Service Office
© Copyright Agilent Technologies 2000. All Rights Reserved.
Reproduction, ad aptation, or translation without prior written permission is prohib ited, except as allowed under the copyright laws.
Restric ted Rights Legend.
Use, duplication or disclo sure by the U.S . Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227­7013 for DOD agencies, and subparagraphs (c) (1) and (c) (2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Agilen t Technologies
3000 Hanover Street
Palo Alto, California 94304 U.S.A.
Document Warranty
The information contained in this document is subject to change without notice.
Agilent makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability or fitness for a particular purpose.
Agilen t sha ll not be liabl e for errors contained herein or for damages in connection with the furnishing, perfo rmance, or use of this mate rial.
Product Warranty
This Agilent product has a warranty against defects in material and workmanship for a period of one year from date o f shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
For warranty service or repair, this product must be returned to a service facility designated by Agilen t.
For products retur ned to Agilent for warranty service, the Buyer shall prepay shipping charges to Agilent and Agilent shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent from another country.
Agilent warrants that its software and firmware designated by Agilen t for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent does not warrant that the operation of the instrument software, or firmware will be uninterru pted or erro r free.
Limitation of Warranty
The for egoin g warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buye r, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environ mental specifications for the product, or improper site preparation or maintenance.
No other warranty i s expressed or implied. Agilent specifically disclaims the implied warranties of merchantability or fitness for a particular purpose.
Exclusive Remedies
The remedies provided herein are the buyer's sole and exclusive remedies.
Agilent shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available for Agilent products .
For any assistance, contact your nearest Agilent Sales Office.
Certification
Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent fur ther certifies that its cal ibrat ion measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calib ration facility, and to the calibration facilities of other International S tandards Organization m embers.
Agilen t Technologies P.O. Box 2197 1900 Garden of the Gods Road Colorado Springs, CO 80901
Safety
This apparatus has been designed and tested in accordance with IEC Publication 1010, Safe ty Requirements for Measuring Apparat us, and has been supplied in a safe condition. To ensure safe operation and to keep the product safe, the information, cautions, and warnings in this operating manual must be heeded. In addition, note the external markin gs on the instrument that are described under "Safet y Symbols."
Safety Symbols
!
Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product.
Hazardous voltage symbol.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
WARNING
The Warning sign denotes a hazard . It calls at tention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyon d a Warning sign until the indicated conditions are fully understood and met.
CAUTION
The Caution sign denotes a hazard . It calls at tention to an operating procedure, practice, or the like, which, if not correctly performe d or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a Caution symbol until the indicated conditions are fully understood or met.
About this edition
This is the 1159 Active Differential Voltage Probe
User’s Guide.
Publication number 01159-92000, Feb. 2000
Printed in USA.
Print history is as follows:
01159-92000, Feb. 2000
New editions are complete revisions of the manual. Many product updates do not require manual changes; and, conversely, manual corrections may be done without accompanying product changes. Therefore, do n ot expect a one-to-one correspondence between product updates and manual updates.
s1
Agilent Technologies Printed in the USA
Manual Part Number 01159-92000

Loading...