The 1152A 2.5-GHz Active Probe is a probe solution for high-frequency
applications. This probe is compatible with the AutoProbe Interface
which completely configures the Infiniium series of oscilloscopes for the
probe. See chapter 1 for full specifications and characteristics.
• A bandwidth of 2.5 GHz
• Input resistance of 100 kΩ
• Input capacitance of approximately 0.6 pF
• Dynamic range of ±5 V dc + peak ac
• Variable dc offset of ±20 V
• Excellent immunity to ESD and over-voltages
• Probe-tip ground connection not necessary for "browsing"
Accessories Supplied
The following accessories are supplied. See "Accessories supplied" in chapter 1
for a complete list.
• "Walking-stick" ground
• Box of small accessories
• Carrying case
Accessories Available
The following accessories can be ordered.
• 11880A Type N(m) to probe tip adapter and 50-Ω termination
• 10218A BNC(m) to probe tip adapter
• 10229A Hook tip adapter
ii
In This Book
This guide provides user and service inform ati on for the 1152A 2.5-GHz
Active Probe.
Chapter 1 gives you general information such as inspection, cleaning,
accessories supplied, and specifications and characteristics of the probe.
Chapter 2 shows you how to operate the probe.
Chapter 3 gives you information about some important aspects of probing and
how to get the best results with your probe.
Chapter 4 provides service information.
iii
iv
Contents
1 General Information
To inspect the probe 1-3
To clean the probe 1-3
Accessories supplied 1-5
Probe operating range 1-6
Performance Specifications 1-8
Characteristics 1-9
General Characteristics 1-10
2 Operating the Probe
Probe handling considerations 2-2
To bias the probe 2-3
To connect the probe 2-3
System bandwidth 2-4
Using probe accessories 2-5
Additional accessories 2-9
Service Strategy 4-2
To return the probe to Agilent Technologies for service 4-3
Troubleshooting 4-4
Failure Symptoms 4-4
Contents-1
Contents-2
1
General Information
Introduction
This chapter covers the following information:
• Inspection
• Cleaning
• Accessories supplied
• Probe operating range
• Performance specifications
• Performance characteristics
• General characteristics
1- 2
General Information
To inspect the probe
To inspect the probe
❏ Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents
of the shipment have been checked for completeness and the instrument has
been checked mechanically and electrically.
❏ Check the accessories.
Accessories supplied with the instrument are listed in "Accessories Supplied"
in table 1-1 later in this chapter.
• If the contents are incomplete or damaged notify your Agilent
Technologies Sales Office.
❏ Inspect the instrument.
• If there is mechanical damage or defect, or if the instrument does not
operate properly or pass performance tests, notify your Agilent
Technologies Sales Office.
• If the shipping container is damaged, or the cushioning materials show
signs of stress, notify the carrier as well as your Agilent Technologies
Sales Office. Keep the shipping materials for the carrier’s inspection.
The Agilent Technologies Office will arrange for repair or replacement
at Agilent Technologies’ option without waiting for claim settlement.
To clean the probe
If this probe requires cleaning, disconnect it from the oscilloscope and clean it
with a mild detergent and water. Make sure the instrument is completely dry
before reconnecting it to the oscilloscope.
1-3
Figure 1-1
General Information
To clean the probe
Walking-stick
ground
(supplied)
Included with the probe
is a box of small
accessories. See table
Accessories
1-1,
Supplied
chapter for a complete
list of accessories.
later in the
152A Active Probe
1- 4
Figure 1-2
General Information
Accessories supplied
Accessories supplied
The following figure and table illustrate the accessories supplied with the 1152A
Active Probe.
5
Table 1-1
234
6
1
7
8
91011
Accessories Supplied
Item DescriptionQty Agilent Part
1Walking-stick ground1 5960-2491
2Single-contact socket51 251-5185
3Standard probe re placement tip55 4701-26101
4Sharp probe tip25 081-7734
5200-Ω signal lead154701-81301
62-inch groun d extensi on lead, attac hable to walk ing-sti ck
ground
74-inch al ligator ground lead, attac hable to probe t ip ground101123-61302
8Nut Driver 3/32-in18710-1806
9Flexible Probe Adapter154701-63201
10Probe Socket15041-9466
11Coaxial Socket31250-2428
Number
10 1650-82103
1-5
Figure 1-3
General Information
Probe operating range
Probe operating range
Figure 1-3 shows the maximum input voltage for the active probe as a function
!
of frequency. This is the maximum input voltage that can be applied without
risking damage to the probe.
Maximum Input Voltage vs Frequency
Figure 1-4
Figure 1-4 shows the operating range of the probe. For the most accurate
measurements and safety for the probe, signals should be within the indicated
operating region.
Area of optimum
operation
Probe Operating Range
1- 6
Figure 1-5
General Information
Probe operating range
The curves in Figure 1-5 and Figure 1-6 represent the typical input signal limits
for several levels of second and third harmonic distortion in the output signal.
For input signals below a given curve, the level of harmonic distortion in the
output is equal to or below that represented by the curve. The dashed straight
line in each figure represents the operating range limit as shown in Figure 1-4
on the previous page.
Second harmonic
≤-20 dBc
Second harmonic
≤-30 dBc
Second harmonic
≤-40 dBc
Figure 1-6
Second Harmonic Distortion, Input Voltage vs Frequency
Third harmonic
≤-40 dBc
Third Harmonic Distortion, Input Voltage vs Frequency
1-7
General Information
Performance Specifications
Performance Specifications
Table 1-2 gives performance specifications for the active probe.
Table 1-2
Table 1–2 Performance Specifications
Bandwidth (–3dB)>2.5 GHz
Attenuation Factor10:1
dc Input Resistance100 kΩ ±1%
dc Gain Accuracy±0.5% with 50Ω ±0.5Ω load
1- 8
Characteristics
The following characteristics are typical for the active probe.
General Information
Characteristics
Table 1-3
Characteristics
System bandwith
with 54845A and 54835A1.3 GHz
with 54810A/15A/20A/25A500 MHz
Rise time (10% to 90%)<140 ps calculat ed from
Input Capacitance0.6 pF (typical)
Flatness
<3 ns from rising edge±6% with input edge ≥170 ps
≥3 ns from rising edge±1% with input edge ≥170 ps
Dynamic Range
(<1.5% gain compression)
dc Offset Accuracy±1% of offset ±1 mV
Offset Adjustment Range±20 V at the probe tip
Offset Gain4.6 V/mA
RMS Output Noise
(dc to 2.5 GHz, input loaded in 50 Ω)
Propagation Delay7.5 ns (approxi ma tel y)
!
Maximum Input Voltage
ESD Tolerance
1
This is not the voltage measurement range. See Dynamic Range characteristic
1
±5 V dc + peak ac
<300 µV
±40 V [dc + peak ac(<20 MHz)], CAT I
(150 Ω/150 pF)
for measurement range.
tr = (0.35/Bandwidth)
1-9
General Information
General Characteristics
General Characteristics
The following general characteristics apply to the active probe.
Table 1-4
Figure 1-7
General Characteristics
Environmental Condi ti ons
OperatingNon-operating
Temperature0 °C to +55 °C (32 °F to +131 °F)−40 °C to +70 °C (−40 °F to +158 °F)
Humidityup to 95% relati ve humidity (non-
Altitudeup to 4,600 meters (15,000 ft) up to 15,3 00 me ter s (50,000 ft)
VibrationRandom vibration 5 to 500 Hz,
Power
Requirements
Weightapproximately 0.69 kg (1.52 lb)
DimensionsRefer to the outline drawi ng below.
condensing) at +40 °C (+104 °F)
10 minutes per axis, 0.3 g
+12 Vdc @ 5 mA max
–12 Vdc @ 95 mA max
+4 Vdc @ 90 mA max.
up to 90% relati ve humidity at +65 °C
(+149 °F)
Random vibration 5 to 500 Hz, 10
rms.
min. per axis, 2.41 grm s. Resonant
search 5 to 500 Hz swept sine, 1
octave/min. sweep rate, (0.75g), 5
min. resonant dwell at 4 resonances
per axis.
(voltages supplied from AutoProb e
Interface)
1152A Active Probe Dimensions
1- 10
2
Operating the Probe
Operating the Probe
Probe handling considerations
To use the probe
The 1152A Active Probe requires a dc voltage source for power, and a
dc bias voltage to offset any dc component in the target signal. The probe
must be properly biased to offset any dc component in the target system
or clipping occurs on the output.
The Infiniium family oscilloscopes provide both the power and the dc
bias through the front panel connector. The dc bias is directly
proportional to the vertical offset setting selected on the oscilloscope.
Probe handling considerations
This probe has been designed to withstand a moderate amount of physical and
electrical stress. However, with an active probe, the technologies necessary to
achieve high performance do not allow the probe to be unbreakable. Treat the
probe with care. It can be damaged if it is dropped onto a hard surface. This
damage is considered to be abuse and will void the warranty when verified by
Agilent Technologies service professionals.
• Exercise care to prevent the probe end from receiving mechanical
shock.
• Store the probe in a shock-resistant case such as the foam-lined
shipping case which came with the probe.
2- 2
Operating the Probe
To bias the probe
To bias the probe
The probe has limiting designed to avoid excessive power dissipation. The input
operating range of the probe is ±5 V, and up to ±20 Vdc can be biased out using
the oscilloscope or power module dc offset function. The dc offset function
sends a dc offset bias voltage to the probe. If the input plus the
non-compensated dc voltage exceeds +14 V relative to the probe tip, the output
of the probe will limit at +1.4 V. As the input plus the non-compensated dc
voltage reaches –14 V, the output will limit at –1.4 V; then, it will fold back to
approximately –0.8V as the input plus non-compensated DC voltage exceeds
–14 V. The output of the probe will remain at the limit voltage until the input
plus non-compensated DC voltage falls below approximately –8 Vdc.
To ensure that the output is in the active region of the probe, and not in one of
the two saturated limit regions, always set the oscilloscope vertical sensitivity
to ≥1.25 V/div, for a maximum fullscale reading of 10 V. If the waveform goes
offscale top or bottom, use the DC offset knob to position the waveform back
onto the screen. This places the probe back within its dynamic range.
To connect the probe
1 Connect the probe output to the instrument input.
2 Calibrate the oscilloscope and probe combination with the instrument
calibration routines.
When the probe has been calibrated, the dc gain, offset zero, and offset gain
will be calibrated. The degree of accuracy specified at the probe tip is
dependent on the oscilloscope system specifications.
2-3
Operating the Probe
System bandwidth
System bandwidth
Since the 1152A is an active probe, the bandwidth of the oscilloscope and probe
combination is a mathematical combination of their individual specifications.
Equation 2-1
System Bandwidth
where
t
is the rise time of the oscilloscope
r1
t
is the rise time of the probe
r2
The probe has limiting designed to avoid excessive power dissipation. The input
operating range of the probe is ±5 V. If the input and offset exceeds +14 V relative
to the probe tip, the output of the probe will limit at +1.4 V. As the input plus offset
reaches –14 V, the output will limit at –1.4 V; then, it will fold back to approximately
–0.8 V as the input plus offset exceeds –14 V. The output of the probe will remain at
the limit voltage until the input plus offset falls below approximately –8 Vdc.
The following figure and table illustrate the accessories supplied with the 1152A
Active Probe.
5
Table 2-1
234
6
1
7
8
91011
Accessories Supplied
Item DescriptionQty Agilent Part
1Walking-stick ground1 5960-2491
2Single-contact socket51251-5185
3Standard probe re placement tip554701-26101
4Sharp probe tip25081-7734
5200-Ω signal lead154701-81301
62-inch groun d extensi on lead, attac hable to walk ing-sti ck
ground
74-inch al ligator ground lead, attac hable to probe t ip ground101123-61302
8Nut Driver 3/32-in18710-1806
9Flexible Probe Adapter154701-63201
10Probe Socket15041-9466
11Coaxial Socket31250-2428
Number
101650-82103
2-5
Operating the Probe
Using probe accessories
Walking-stick Ground
The walking-stick ground is the best ground for general probing. It is short, and
the ground wire includes a bead for damping probe resonance. This provides a
well maintained probe response for frequencies to 2.5 GHz.
Single-contact Socket
The single-contact sockets can be soldered into a circuit to provide a probe
point to hold the probe tip or ground. The socket accepts 0.018-inch to
0.040-inch pins. The sockets accept the probe tips, the walking-stick ground,
the 200-Ω signal lead, and the ground extension lead.
Probe Tips
There are two types of replaceable probe tips furnished with the probe. The
0.030-inch round standard probe tip is for general applications. It is made of a
material that will generally bend before breaking. The 0.025-inch round sharp
probe tip has a narrower point and is a harder material. It can be used to probe
constricted areas or penetrate hard coatings.
CAUTIONDo not solder the probe tip into circuitry. Excessive heat may damage the tip
or circuitry inside the probe. If you need to solder something into your
circuitry, use the single contact sockets, ground extension lead, or 200-Ω signal
lead. They are less easily damaged and less expensive to replace.
• To remove and replace probe tips, use the nut driver to unscrew the tip from
the end of the probe.
•
Be sure to screw the replacement tip all the way in or the probe may be
intermittent or appear ac coupled.
Nut Driver
The 3/32-in nut driver is provided for easier replacement of the probe tips.
2- 6
Operating the Probe
Using probe accessories
200-Ω Signal Lead
This 2-inch orange extension lead includes a molded-in resistor to dampen
resonance caused by the lead inductance. Use this lead and the ground
extension lead to provide a flexible connection to the circuit under test.
There is a trade-off when using the extension leads. To maintain a clean pulse
response, the probing system bandwidth is limited to 1.5 GHz. Probe resonance
is damped by the walking-stick bead and the resistor in the signal lead.
Ground Extension Lead
This 2.25-inch black ground lead can be used to extend ground from the
walking-stick to the circuit under test. When used with the walking-stick ground
the probe resonance is damped by the bead on the walking-stick.
Alligator Ground Lead
The alligator ground lead can be used in general applications when the
bandwidth of the signal is 350 MHz or lower. With no signal lead extension the
probe resonant frequency is about 650 MHz.
Flexible Probe Adapter
The flexible probe adapter provides a
high-quality connection between a
coaxial socket and the 1152A probe. The
right-angle connection allows the probe
to remain parallel to a PC board and the
flexibility prevents the leverage of the
probe and cable from damaging PC board
circuitry.
As with any cable-type interconnection,
always apply insertion and removal
forces to the connectors directly, and not
through the cable itself (see the
illustration).
2-7
Operating the Probe
Using probe accessories
Probe Socket
The probe socket is a direct fit to the
shield surface of the 1152A probe. Use
this socket and the single contact
socket to design the highest quality
probing of a PC board. The illustration
shows the socket and the PC board
layout needed to mount the parts.
Coaxial Socket
The coaxial socket is designed to fit the
standard mini-probe. When used with
the flexible probe adapter, it can be
installed in a circuit so you can probe
with the 1152A. The illustration shows
the socket and the PC board layout
needed to mount the socket to the
board.
See AlsoChapter 3, "Probing Considerations," for a more complete discussion about the
effects of probe connection techniques on signal fidelity.
2- 8
Operating the Probe
Additional accessories
Additional accessories
The following accessories enhance use of the active probe. For ordering
information, see "Replaceable Parts" in chapter 3.
Type-N to Probe Tip Adapter
The 11880-60001 Type-N(m) to probe tip adapter is
available to connect the input of the active probe to
Type-N connectors. It has an internal 50-Ω load. It can
be used for general testing and is specifically
recommended for testing the probe bandwidth. This
adapter must be ordered separately.
BNC to Probe Tip Adapter
The 10218A BNC(m) to probe tip adapter is
available to connect the input of the active probe to
BNC type connectors. It does not have an internal
load so it is not recommended for testing where the
full bandwidth of the probe is needed. This adapter
must be ordered separately.
2-9
2- 10
3
Probing Considerations
Introduction
This chapter gives you some guidance about the effects of probing and
how to get the best measurement results. The effect of the following
parameters are covered in this chapter:
• Resistive Loading
• Capacitiv e Loading
• Ground Inductance
• Bandwidth
Two important issues while measuring signals with probes are how the
probe/oscilloscope combination represents the signal at the probe tip
and how the probe affects the circuit during the measurement. When a
probe is connected to a circuit to measure a signal it becomes part of the
circuit. Probing a signal can be easy and successful if some forethought
is given to the nature of the circuit under test and what type of probe
best solves the measurement problem. Because of the wide variety of
signals that may be encountered, ranging from high bandwidth (fast rise
times) to high impedance, in a given situation one probe may do a better
job than another. Therefore, it is helpful to understand the different
effects caused by the interaction between the probed circuit and the
probe.
3- 2
Figure 3-1
Waveform 1
Probing Considerations
Resistive Loading Effects
Resistive Loading Effects
The two major effects caused by resistive loading are amplitude distortion and
changes in dc bias conditions in the circuit under test.
Amplitude Distortion
Amplitude distortion is depicted in Figure 3-1, where waveform 1 is the signal
before probing and waveform 2 is the signal while probing. (The baselines of
these signals have been overlaid to show the amplitude change. If the baseline
of a signal is not at zero volts it will shift when the signal is probed.)
Waveform 1
Equation 3-1
Oscilloscope Display Showing Amplitude Distortion
The cause of the error is the voltage divider developed between the source
resistance of the device under test and the input resistance of the probe being
used. Equation 3-1 calculates the error caused by the voltage divider.
R
source
Error %()
---------------------------------------- R
+
sourceRprobe
100×=
A probe with an input resistance ten times that of the source resistance of the
device under test causes a 9.09% error in the measurement. It is best to use a
probe with an input resistance at least ten times that of the source resistance.
3-3
Probing Considerations
Resistive Loading Effects
Bias Changes
Probes with low input resistance can cause bias changes in the device under
test. A good example of this effect can be seen when probing ECL circuits.
Figure 3-2 represents a typical ECL node with a 60-Ω bias resistor to −2 V. Ip
represents current that flows from ground into the circuit when the probe is
connected. The table shows the current that flows in each device at both the
high (−0.8 V) and low (−1.75 V) states, with and without a 500-Ω probe
connected.
Figure 3-2
High (-0.8 V)Low(-1.75 V)
Without
Probe
O 20 mA18.4 mA 4.2 mA 0.7 mA
I
R 20 mA20 mA4.2 mA 4.2 mA
I
P1.6 mA3.5 mA
I
With
Probe
Without
Probe
With
Probe
Probing ECL Circuits
Note that in the high state there is little difference in current flow with or
without the probe connected. However, in the low state the output stage is
closer to cutoff. Connecting the probe sources current into the output node,
which reduces the current sourced from the gate output. The output current
drops from 4.2 mA to 0.7 mA. The low output current can cause problems with
switching noise margins. The output gate will have difficulty reaching the low
threshold, so ac performance will suffer because the falling edge degrades. If a
larger bias resistor had been used to keep the current levels lower, when a 500-Ω
probe is attached the output gate could go into cutoff before it reaches the low
threshold.
Recommendation
Be careful not to use a probe just because it has the highest input resistance
available. High-resistance probes usually come with trade-offs in other
important parameters, such as higher capacitance, which also affect
measurement accuracy.
3- 4
Figure 3-3
1 MΩ/6pF probe
Probing Considerations
Capacitive Loading Effects
Capacitive Loading Effects
The input capacitance of a probe causes the overall input impedance to decrease
as a function of frequency. For this reason, input capacitance becomes one of
the most important parameters that affect high frequency measurements.
Figure 3-3 plots the probe impedance vs frequency for two probes: a 1-MΩ, 6-pF
probe and the 1152A probe (100 kΩ, 0.6 pF). It shows that because of the lower
input capacitance, the 1152A probe actually has a higher input impedance for
frequencies above 240 kHz. At frequencies above 2.65 MHz, it has as much as
10 times the impedance of the 1-MΩ probe.
1152 probe
Probe Impedance vs Frequency
The input capacitance of a probe forms an RC time constant with the parallel
combination of source impedance and probe input resistance. This can cause
an increase in the circuit rise time and a time delay in a pulse edge.
3-5
Figure 3-4
Plot 1
Plot 1
Plot 1
Probing Considerations
Capacitive Loading Effects
Figure 3-4 represents plots from three spice simulations showing this loading
effect. Plot 1 shows the signal edge before probing. Plot 2 shows the edge after
probing with a 6-pF probe and plot 3 after probing with a 15-pF probe.
Table 3-1
Spice Simulation Of Probe Capacitance Loading Effects
Table 3-1 summarizes the data. It shows that the 6-pF probe didn’t significantly
increase the rise time of the signal, but delayed it (referenced at the 50% point)
approximately 150 ps. The 15-pF probe not only slowed the rise time
approximately 33% but also delayed the edge 340 ps.
Probe Capacitance Loading Effects
PlotRise Time Delay
11 ns0.0 ps
21.067 ns150 ps
31.33 ns340 ps
The circuit used in the simulation is modeled after a 50-Ω system with a 1-ns
source and terminated transmission line with 500-ps delay. The inset shows the
spice model, with the probe point where the probe was connected. As signals
achieve faster rise times, probes with lower input capacitance are required to
make accurate timing and rise time measurements.
3- 6
Probing Considerations
Ground Inductance Effects
Ground Inductance Effects
Probe grounding techniques are an important factor in making accurate high
frequency measurements. The main limitation, probe resonance, is a function
of the input capacitance of the probe and the inductance of the ground return.
These two parameters in series form an LC resonant circuit that, when
connected to the circuit under test, becomes part of the circuit’s response.
The probe resonance can cause overshoot and ringing on pulse edges that
contain energy in the same frequency band as the resonance. The true response
is masked, the false response gets transferred to the oscilloscope, and the
oscilloscope display shows an incorrect result. If overshoot and ringing added
by a probe during troubleshooting changes how the circuit functions, it can
produce an incorrect judgment about circuit operation. To minimize the
problem of ground ringing, use the shortest possible ground with a probe that
has the lowest possible input capacitance. Equation 3-2 can be used to calculate
the frequency where a certain probe and grounding technique resonates.
Equation 3-2
f
r
where
C is the probe input capacitance. (It is usually found in the probe data sheet.)
L is the inductance of the ground return. (It can be approximated using the
constant of 25 nH per inch.)
Figure 3-5 plots the probe impedance vs frequency for two probes: a 1-MΩ, 6-pF
probe and the 1152A probe (100 kΩ, 0.6 pF). It also plots the inductive
reactance vs frequency for three different values of ground inductance. The
5-nH inductance represents a PC board socket, the 20-nH inductance a spanner
ground, and the 100-nH inductance a 4-inch ground wire. Where the probe plots
cross the inductance plots gives the resonant frequency of the probe and ground
combination. You can see from the graphs that in all three cases the 6-pF probe
resonates at approximately one-third the frequency of the 1152A (0.6 pF). The
lower resonance means that the effect of the resonance is more likely to
influence the representation of the signal.
1
------------------ -=
2π LC
3-7
Figure 3-5
1-MΩ, 6- pF probe
1152A probe
Probing Considerations
Ground Inductance Effects
Probe Impedance and Resonance
3- 8
Figure 3-6
Waveform 1
Waveform 2
Waveform 3
Waveform 4
Probing Considerations
Ground Inductance Effects
Figure 3-6 shows waveforms measured by the 1152A (100 kΩ, 0.6 pf) and the
1-MΩ, 6-pF probe; both probes are connected to a 1-GHz oscilloscope.
Probe Resonance Effects
Waveform 1 shows the pulse response of a 6-pF probe measuring a 400-ps step.
The ringing on the pulse is caused by the input capacitance of the probe and by
the inductance of the ground return. The period of the ringing measures
1.72 ns, representing a frequency of 581 MHz. The circuit had a ground return
of 1/2 inch. Using Equation 3-2 to calculate the resonant frequency (12.5 nH
and 6 pF) results in 580 MHz. The measurement and the calculation yield the
same result, showing how probe resonance causes problems when probing high
speed signals.
Waveform 2 shows the pulse response when the same 6-pF probe measures an
800-ps edge. Notice that the overshoot and ringing are still present, but are
significantly reduced. This is because the slower signal edge has less energy at
the resonant frequency of the probe.
Waveform 3 shows the pulse response when the 6-pF probe measures a 1.25-ns
edge. The ringing is nearly subdued and doesn’t play a significant role in the
measurement.
Waveform 4 shows the 1152A 0.6-pF probe, with a one-inch ground lead,
measuring the 400-ps edge. Because of its much lower capacitance, and even
with a longer ground lead, its resonant frequency is much higher and it shows
no ringing in the response.
3-9
Probing Considerations
Ground Inductance Effects
The measurements from the first three waveforms lead to a rule of thumb:
To minimize signal distortion due to probe resonance, provide a two-to-one, or
greater, difference between the resonant frequency of the probe and the
bandwidth of the signal being measured.
For pulsed data applications, the rise time of a signal can be related to the
bandwidth by using a constant of 0.35 as shown in Equation 3-3. This equation
is derived from a first order RC response.
Equation 3-3
Bandwidth
0.35
----------=
t
r
ExampleThe 1.25-ns edge (waveform 3 in Figure 3-6) equates to a 280-MHz bandwidth.
Bandwidth
0.35
---------t
r
0.35
------------------------ -280 MHz===
1.25
×10
9–
This is approximately half the resonant frequency calculated for the 6-pF probe
with 1/2-inch ground, 580 MHz. Therefore the subdued ringing on waveform 3
validates the rule of thumb.
As noted before, waveform 4 shows the effect when a low-capacitance probe
measures a high-frequency signal. Because of the low capacitance the resonant
frequency is high. Therefore, there is less chance of the probing system affecting
the measurement of the signal.
3- 10
Probing Considerations
Probe Bandwidth
Probe Bandwidth
The bandwidth of the probe is often given much consideration during purchase,
then forgotten while making measurements. Error in m easurem ents occur when the frequency content (at the −3 dB point) of the signal being measured
approaches or exceeds the bandwidth of the probe. The probe can be modeled
as a low-pass filter for the signal.
If a 700-MHz probe is used to measure a 1-ns signal, the rise time error can be
calculated using equations 3-3 and 3-4. For this exercise assume that the
oscilloscope bandwidth is great enough not to contribute any errors.
Equation 3-4
t
r
tr1()2tr2()
2
+=
where
t
is the rise time of the probe,
r1
t
is the rise time of the signal.
r
2
Calculate the rise time of the 700-MHz probe (Equation 3-3).
1
t
r
0.35
---------------------------- Bandwidth
0.35
--------------------- -0.5 ns===
700 MHz
2 Calculate the rise time of the 1-ns signal as measured by the 700-MHz
probe (Equation 3-4).
0.5()21.0()
t
r
+1.251.12 ns===
The measurement error between the actual signal and what was measured is
12%. To keep measurement errors less than 6%, use a probe with a bandwidth
three or more times that of the signal.
3
Calculate the bandwidth of the 1-ns signal (Equation 3-3).
Bandwidth
Use a probe with a bandwidth of 1.05 GHz (the rise time is 0.333 ns, Equation
3-3).
4
Calculate the rise time of the 1-ns signal measured by the 1.05-GHz
2
0.35
----------350 MHz==
1 ns
probe (Equation 3-4).
t
0.333()21.0()
r
Now, the measurement error is less than 6%.
2
+1.111.054 ns===
3- 11
Probing Considerations
Conclusion
Conclusion
In conclusion we can review the issues by using the effect the 1152A Active
Probe (100 kΩ, 0.6 pF) has while measuring a fast CMOS gate.
Resistive Loading
Resistive loading is caused by the input resistance of the probe. When the CMOS
output is high (5 V) the 100 kΩ input resistance of the probe draws 50 A. A
CMOS gate can drive many times this current, so the load is insignificant. In
addition, the output impedance of a CMOS gate is the on resistance of the output
FET. Whether high or low, this is typically less than 100 Ω. The voltage divider
of 100 Ω and 100 kΩ is also insignificant and will not change the value of either
state of the gate.
Capacitive Loading
CMOS gates typically have an input capacitance between 5 and 10 pF. The traces
between gates will contribute another 5 to 10 pF, which gives a total of 10 to 20
pF. The 0.6-pF input capacitance of the 1152A probe is about 3% to 6% that of
the circuit capacitance. It will not significantly change the time constant in the
node being probed.
Ground Inductance
The CMOS gate has a rise time approaching 1 ns. This equates to a bandwidth
of 350 MHz (Equation 3-3). If we use the walking-stick ground (about 20 nH)
provided with the 1152A probe, the probe resonance will be about 1.45 GHz
(Equation 3-2). We can see that the CMOS equivalent bandwidth (350 MHz) is
at less than half the resonant frequency of the probe. This fits within the rule
of thumb given previously, that to avoid ringing in the response, the resonance
of the probe should be at least twice the frequency of the energy in the signal.
Bandwidth
Although it was specifically not covered in this chapter, the bandwidth of the
probe and oscilloscope combination is also very important. As previously noted,
with CMOS signals of 1 ns rise times the signal bandwidth is 350 MHz. This
means for an accurate representation the probe and oscilloscope combination
should have at least a 3-to-1 margin in bandwidth, at least 1.05 GHz.
3- 12
4
Service
Introduction
This chapter provides service information for the 1152A Active Probe.
The following sections are included in this chapter:
• Service strategy
• Returning to Agilent Technologies for service
• Troubleshooting and failure symptoms
Service Strategy
The 1152A Active Probe is a high-frequency instrument with many critical
relationships between parts. For example, the frequency response of the
amplifier on the hybrid is trimmed to match the output coaxial cable. As a result,
to return the probe to optimum performance requires factory repair. All probes
must be returned to the factory for repair and calibration until January 1998.
After that time, if the probe is under warranty, normal warranty services apply.
If the probe is not under warranty, a failed probe can be exchanged for a
reconditioned one at a nominal cost.
4- 2
Service
To return the probe to Agilent Technologies for service
To return the probe to Agilent Technologies for service
Before shipping the instrument to Agilent Technologies, contact your nearest
Agilent Technologies Sales Office for additional details.
1
Write the following information on a tag and attach it to the instrument.
• Name and address of owner
• Instrument model number
• Instrument serial number
• Description of the service required or failure indications
2 Remove all accessories from the instrument.
Accessories include all cables. Do not include accessories unless they are
associated with the failure symptoms.
3
Protect the instrument by wrapping it in plastic or heavy paper.
4 Pack the instrument in foam or other shock absorbing material and
place it in a strong shipping container.
You can use the original shipping materials or order materials from an Agilent
Technologies Sales Office. If neither are available, place 3 to 4 inches of
shock-absorbing material around the instrument and place it in a box that does
not allow movement during shipping.
5
Seal the shipping container securely.
6 Mark the shipping container as FRAGILE.
In any correspondence, refer to instrument by model number and full serial
number.
4-3
Service
Troubleshooting
Troubleshooting
• If your probe is under warranty and requires repair, return it to Agilent
Technologies. Contact your nearest Agilent Technologies Service
Center.
• If the failed probe is not under warranty, you may exchange it for a
reconditioned probe. See "To Prepare the Probe for Exchange" in this
chapter.
Failure Symptoms
The following symptoms may indicate a problem with the probe or the way it is
used. Possible remedies and repair strategies are included.
The most important troubleshooting technique is to try different combinations
of equipment so you can isolate the problem to a specific instrument.
Probe Calibration Fails
Probe calibration failure with an oscilloscope is usually caused by improper
setup. If the calibration will not pass, check the following:
• Be sure the instrument passes calibration without the probe.
• Check that the probe passes a signal with the correct amplitude.
• If the probe is powered by the oscilloscope, check that the offset is
approximately correct. The probe calibration cannot correct major
failures.
Incorrect Frequency Response
Incorrect frequency response may be caused by a defective probe, plug-in or
oscilloscope mainframe, or an improper application such as poor connections
or grounding. Read chapter 2, "Probing Considerations," in this guide. If the
application is correct, try the probe with another oscilloscope.
If the probe appears ac coupled at a high frequency, check for a loose probe tip.
The frequency response of the probe is determined by the amplifier hybrid in
the probe and the probe cable. If the probe fails the bandwidth test, factory
repair is necessary. Also read "Incorrect Pulse Response" below.
4- 4
Service
Failure Symptoms
Incorrect Pulse Response (flatness)
If the probe’s pulse response shows a top that is not flat (incorrect ac gain), it
is most likely caused by an inaccurate 50-Ω load on the probe. The probe is
designed to work into a 50-Ω load that is accurate within 1.0% (±0.5 Ω). Check
the value of the load you are using before you suspect the probe. If the load is
accurate, the gain problem with the probe will have to be repaired by the factory.
If the probe appears ac coupled at a high frequency, check for a loose probe tip.
Incorrect dc Gain
The dc gain is a function of the values of internal parts. It is independent of the
load on the probe. Any failure of the accuracy of the dc gain requires factory
repair.
Incorrect Input Resistance
First, check that the probe tip is not loose. The input resistance is determined
in the amplifier hybrid in the probe and cannot be repaired in the field. The
probe must be returned to the factory for repair.
Incorrect Offset
Incorrect offset can be caused by a misadjusted offset zero (see "Offset Will Not
Zero" below) or lack of probe calibration with the oscilloscope.
Offset Will Not Zero
With no signal input and no offset setting, the dc output of the probe should be
within ±1 mV.
If the probe is connected to an Infiniium oscilloscope, the oscilloscope will
calibrate out an offset zero error during a probe calibration. If the offset error
can not be calibrated out, return it to Agilent Technologies for repair.
4-5
4- 6
Index
A
accessories
200-W signal lead 2-7
alligator ground lead 2-7
BNC to probe tip adapter 2-9
coaxial socket 2-8
flexible probe adaptor 2-7
ground extention lead 2-7
nut driver 2-6
probe socket 2-8
single-contact socket 2-6
Type-N to probe tip adapter 2-9
walking-stick ground 2-6
accessories available 1-5
accessories supplied 1-5
accessories, using 1-5, 2-5 to 2-8
B
bandwidth
of oscilloscope with probe 3-12
of probes 3-11
of signals 3-9 to 3-11
with oscilloscope 2-4
C
calibration
failure 4-4
probe with oscilloscope 2-3
capacitive loading 3-12
characteristics 1-9
cleaning 1-3
connecting to oscilloscope 2-3
D
dimensions 1-10
E
errors
amplitude distortion 3-3
capacitive loading 3-5 to 3-6
probe resonance 3-7 to 3-10
resistive loading 3-3 to 3-4
F
failure symptoms 4-4
G
general characteristics 1-10
ground inductance 3-7, 3-12
H
harmonic distortion 1-7
I
input capacitance 3-7
inspecting 1-3
L
limiting, probe offset 2-3
M
maximum input voltage 1-6
O
offset errors 4-5
offset limiting 2-3
offset zero
errors 4-5
operating environment 1-10
operating range 1-6 to 1-7
P
packing for return 4-3
power requirements 1-10
probes
capacitive loading 3-5 to 3-6
capacitive loading effects 3-6
ground inductance 3-7 to 3-10
grounding 3-7
high resistance 3-4
input impedance 3-5
resistive loading 3-3 to 3-4
resonance 3-7
R
repair 4-3
resistive loading 3-12
resonance
of probe 3-7 to 3-10
rule of thumb 3-10
resonant frequency 3-7
returning probe to Agilent Technologies
4-3
S
service strategy 4-2
specifications 1-8
storage environment 1-10
system bandwidth 2-4
T
troubl eshooting 4-4
W
weight 1-10
Index-1
Index-2
DECLARATION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014
Manufacturer’s Name:Agilent Technologies
Manufacturer’s Address:Colorado Springs Division
1900 Garden of the Gods Road
Colorado Springs, CO 80907, U.S.A.
declares, that the product
Product Name:Oscilloscope probe
Model Number(s):1152A
Product Option(s):All
conforms to the following Product Specifications:
Safety:IEC 1010-1:1990+A1 / EN 61010-1:1993
UL 3111
CSA-C22.2 No. 1010.1:1993
EMC:CISPR 11:1990 / EN 55011:1991 Group 1, Class A
IEC 555-2:1982 + A1:1985 / EN60555-2:1987
IEC 555-3:1982 + A1:1990 / EN 60555-2:1987 + A1:1991
IEC 801-2:1991 / EN 50082-1:1992 4 kV CD, 8 kV AD
IEC 801-3:1984 / EN 50082-1:1992 3 V/m, {1kHz 80% AM, 27-1000 MHz}
IEC 801-4:1988 / EN 50082-1:1992 0.5 kV Sig. Lines, 1 kV Power Lines
Supplementary Information:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC
Directive 89/336/EEC, and carries the CE-marking accordingly.
This product was tested in a typical configuration with Hewlett-Packard test systems.
Colorado Springs, 04/22/1997
Ken Wyatt, Quality Manager
European Contact: Your local Agilent Technologies Sales and Service Office or Agilent Technologies GmbH, Department ZQ / Standards
Reproduction, adaptation, or
translation without prior
written permission is
prohibited, except as allowed
under the copyright laws.
Restricted Rights Legend.
Use, duplication or disclosure
by the U.S. Government is
subject to restrictions as set
forth in subparagraph (c) (1)
(ii) of the Rights in Technical
Data and Computer Software
clause at DFARS
252.227-7013 for DOD
agencies, and subparagraphs
(c) (1) and (c) (2) of the
Commercial Computer
Software Restricted Rights
clause at FAR 52.227-19 for
other agencies.
Agilent Technologies 3000
Hanover Street Palo Alto,
California 94304 U.S.A.
Document Warranty
The information contained in
this document is subject to
change without notice.
Agilent Technologies
makes no warranty of any
kind with regard to this
material, including, but
not limited to, the implied
warranties of
merchantability or fitness
for a particular purpose.
Agile nt Technologies shall not
be liable for errors contained
herein or for damages in
connection with the
furnishing, performance, or
use of this material.
Safety
This apparatus has been
designed and tested in
accordance with IEC
Publication 1010, Safety
Requirements for Measuring
Apparatus, an d has been
supplied in a safe condition.
This is a Safety Class I
instrument (provided with
terminal for protective
earthing). Before applying
power, verify that the correct
safety precautions are taken
(see the following warnings).
In addition, note the external
markings on the instrument
that are described under
"Safety Symbols."
Warning
Before turning on the
instrument, you must connect
the protective earth terminal
of the instrument to the
protective conductor of the
(mains) power cord. The
mains plug shall only be
inserted in a socket outlet
provided with a protective
earth contact. You must not
negate the protective action
by using an extension cord
(power cable) without a
protective conductor
(grounding). Grounding one
conductor of a two-conductor
outlet is not sufficient
protection.
• Only fuses with the required
rated current, voltage, and
specified type (normal blow,
time delay, etc.) should be
used. Do not use repaired
fuses or short-circuited
fuseholders. To do so could
cause a shock of fire hazard.
• Service instructions are for
trained service personnel. To
avoid dangerous electric
shock, do not perform any
service unless qualified to do
so. Do not attempt internal
service or adjustment unless
another person, capable of
rendering first aid and
resuscitation, is present.
• If you energize this
instrument by an auto
transformer (for voltage
reduction), make sure the
common terminal is
connected to the earth
terminal of the power source.
• Whenever it is likely that
the ground protection is
impaired, you must make the
instrument inoperative and
secure it against any
unintended operation.
• Do not operate the
instrument in the presence of
flammable gasses or fumes.
Operation of any electrical
instrument in such an
environment constitutes a
definite safety hazard.
• Do not install substitute
parts or perform any
unauthorized modification to
the instrument.
• Capacitors inside the
instrument may retain a
charge even if the instrument
is dis connected from its
source of supply.
• Use caution when exposing
or handling the CRT.
Handling or replacing the CRT
shall be done only by qu alified
maintenance personnel.
Safety Symbols
!
Instruction manual symbol:
the product is marked with
this symbol when it is
necessary for you to refer to
the instruction manual in
order to protect against
damage to the product.
Hazardous voltage symbol.
Earth terminal symbol: Used
to indicate a circuit common
connected to grounded
chassis.
WARNING
The Warning sign denotes a
hazard. It calls attention to a
procedure, practice, or the
like, which, if not correctly
performed or adhered to,
could result in personal
injury. Do not proceed
beyond a Warning sign until
the indicated conditions are
fully understood and met.
CAUTION
The Caution sign denotes a
hazard. It calls attention to an
operating procedure, practice,
or the like, which, if not
correctly performed or
adhered to, could result in
damage to or destruction of
part or all of the produ ct. Do
not pro ceed beyond a Caution
symbol until the indicated
conditions are fully
understood or met.
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901
Product Warranty
This Agilent Technologies
product has a warranty
against defects in material
and workmanship for a period
of three years from date of
shipment. During the
warranty period, Agilent
Technologies will, at its
option, either repair or
replace products that prove to
be defective. For warranty
service or repair, this produ ct
must be returned to a service
facility designated by Agilent
Technologies. For products
returned to Agilent
Technologies for warranty
service, the Buyer shall
prepay shipping charges to
Agilent Technologies and
Agile nt Techno logies s hall pay
shipping charges to return the
product to the Buyer.
However, the Buyer shall pay
all shipping charges, duties,
and taxes for products
returned to Agilent
Technologies from another
country. Agilent Technologies
warrants that its software and
firmware designated by
Agilent Technologies for use
with an instrument will
execute its programming
instructions wh en properly
installed on that instrument.
Agilent Technologies does not
warrant that the operation of
the instrument software, or
firmware will be
uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall
not apply to defects resulting
from improper or inadequate
maintenance by the Buyer,
Buyer- supplied software or
interfacing, unauthorized
modification or misuse,
operation outside of the
environmental specifications
for the product, or improper
site preparation or
maintenance.
No other warranty is
expressed or implied.
Agilent Technologies
specifically disclaims the
implied warranties of
merchantability or fitness
for a particular purpose.
Exclusive Remedies
The remedies provided herein
are the buyer’s sole and
exclusive remedies. Agilent
Technologies shall not be
liable for any direct, indirect,
special, incidental, or
consequential damages,
whether based on contract,
tort, or any other legal theory.
Assistance
Product maintenance
agreements and other
customer assistance
agreements are available for
Agilent Technologies
products. For any assistance,
contact your nearest Agilent
Technologies Sales Office.
Certification
Agilent Technologies certifies
that this product met its
published specificat ions at the
time of shipment from the
factory. Agilent Technologies
further certifies that its
calibration measurements are
traceable to the United States
National Institute of
Standards and Technology, to
the extent allowed by the
Institute’s calibration facility,
and to the calibration facilities
of other International
Standards Organization
members.
About this edition
This is the 1152 A 2.5-G Hz Active Probe U ser’s Guide.
Publication number
01152-97002, Jan. 2000
Print history is as follows:
01152-97001, May 1998
01152-97002, Jan. 2000
Printed in USA.
New editions are complete
revisions of the manual. Many
product updates do not
require manual changes; and,
conversely, manual
corrections may be done
without accompanying
product changes. Therefore,
do not expect a one-to-one
correspondence between
product updates and manual
updates.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.