AEROTRAK 7010 User Manual

AEROTRAK®+ REMOTE
RELIABLE MEASUREMENT
CONFIDENT RESULTS
ACTIVE AIR SAMPLER
MODEL 7010
TSI® AeroTrak®+ 7010 Remote Active Air Sampler (AAS) oers aseptic manufacturers condent and reliable microbial monitoring in pharmaceutical manufacturing Grade A and B environments with external vacuum systems. With active ow measurement and proactive ow alarms, the AAS allows cleanroom technicians to correct facility conditions before monitoring is aected to reduce production waste. The AAS integrates into TSI Facility Monitoring Software to put critical data at your ngertips – without worry of data interruption or loss.
Features & Benets
RELIABLE MEASUREMENT
+ Easy compliance with FDA cGMP and EU GMP regulations
by way of low d50 (0.8 µm) and active ow measurement
+ Quick corrective action driven by real-time ow measurement
warnings—e.g., kinked tubing, vacuum loss, etc.
+ Design specic for Grade A and B environments with traceable
materials and thoughtful accessories
+ Reduced interventions to change agar plates through
intermittent sampling
CONFIDENT RESULTS
+ Complete environmental monitoring system by TSI with
integrated total particle and microbial monitoring
+ Hassle-free data integrity with FMS Software integration
and automated, sample-based reporting
+ Secure shareability—OPC UA interface to LIMS
+ Distributed architecture without common points of failure
+ No lost data—sampling completes even if network fails
CONFIDENT, RELIABLE, NO-HASSLE COMPLIANCE & DATA INTEGRITY
TSI Complete Facility Monitoring Systems Learn more at www.tsi.com/reliablesystem
UNDERSTANDING, ACCELERATED
SPECIFICATIONS
AEROTRAK®+ REMOTE ACTIVE AIR SAMPLER MODEL 7010
Flow Rate 28.3 LPM (1.0 CFM)
with ±5% accuracy
Sampling Method Sieve impaction
d
50
Sampling Continuous or Intermittent
Vacuum Source External vacuum > 15 in.
Control Box Enclosure Stainless steel
Sample Head (Base and Cap) 316L SS
Recommended Tubing (Between Sample Head and Control Box) 0.5 in. ID x 0.625 in. OD
Agar Plate Recommended Dimensions 90 mm agar plate, deep ll (27 ml)
Standards CE
Operating Environment Indoor use only
Temperature 50° to 104°F (10° to 40°C) Relative Humidity 20% to 95% non-condensing Altitude <10,000 ft. (3,050 m) Pollution Degree 1
Communication TSI FMS5 Monitoring System
Status Indicator Power, ow, sample and ethernet
Data Storage 256,000 sample records
Dimensions
Sample Head 4.5 in. x 3.4 in.
Control Box (H x W x D) 5.6 in. x 4.5 in. x 2.6 in.
0.8µm per ISO 14698-1:2003
(38.1 cm) of Hg
20 m maximum length
(11.4 cm x 8.6 cm)
(14.2 cm x 11.4 cm x 6.7 cm)
Weight
Sample Head 1.3 lb. (0.59 kg) Control Box 2.3 lb. (1.05 kg)
Power Power-over-Ethernet (PoE
compliant with IEEE 802.3at) or 12-24 VDC @ 30W
Relay Load 0.5 A at 125 VAC; 2 A at 30 VDC Overvoltage Category II
Storage Range 14° to 122°F (-10° to 50°C) /
Up to 98% non-condensing
Included Accessories Power connector, 90 mm plate
standos, and operating manual and conguration utility on USB ash drive
Optional Accessories Sample cap, power supply, plate
holder, exhaust lter, sanitary tting inlet, tri-clamp ttings, alarm cable, sample tubing, vacuum tubing and mounting bracket
NO INTERRUPTIONS. NO DATA LOSS. NO-HASSLE COMPLIANCE.
Complete facility monitoring systems oered by TSI.
Learn more at www.tsi.com/reliablesystem
1
Monitor
OPC UA
Client/Server
UNDERSTANDING, ACCELERATED
TSI Incorporated
USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99 Germany Tel: +49 241 523030
P/N 5002612 Rev B ©2021 TSI Incorporated